Abstract
Purpose
We investigated occupational dose to the lens of the eye for physicians engaged in radiology procedures. We evaluated the potential for compliance with the new‐equivalent dose limits to the lens of the eye. Further, a “multiple radiation protection” protocol was proposed according to the basic principles of occupational health, and its effectiveness was estimated.
Methods
Physicians engaged in radiology procedure at medical facilities in Japan were included in this study. The eye lens dose (3‐mm dose equivalent: Hp(3)) for each participant was measured using a small radio‐photoluminescence glass dosimeter mounted on lead glasses. Physicians were directed to procedure multiple radiation protection measures to evaluate their usefulness.
Results
The Hp(3) was reduced by multiple radiation protection in all physicians. In particular, the Hp(3) reduced from 207.7 to 43.2 μSv/procedure and from 21.6 to 10.2 μSv/procedure in cardiovascular internal physician and cerebrovascular physician, respectively, after the implementation of the proposed multiple radiation protection measures. The dose reduction rate of these measures was 53% (range: 37%–79%).
Conclusions
The radiation doses received by the eye lenses of physicians engaged in radiology procedure may exceed the dose limits to the lens of the eye if radio‐protective equipment and imaging conditions are not properly controlled. However, based on the lens equivalent dose data, the implementation of “multiple radiation protection” according to the basic principles of occupational health can ensure compliance with the new‐equivalent dose limits to the lens of the eye without placing an undue burden on individual physicians or medical facilities.
Keywords: eye lens dose, multiple radiation protection, occupational exposure, radiology procedures, radio‐photoluminescence glass dosimeters
1. INTRODUCTION
In April 2011, the International Commission on Radiological Protection (ICRP) Statement on Tissue Reactions (Seoul Statement), reduced the threshold dose for cataracts to 0.5 Gy and issued the following recommendation for the eye lens equivalent dose limit for occupational exposure in planned exposure situations: “for occupational exposure in planned exposure situations the Commission now recommends an equivalent dose limit for the lens of the eye of 20 mSv/year, averaged over defined periods of 5 years, with no single year exceeding 50 mSv. 1 ” In response to this ICRP recommendation, the relevant Japanese national policy, Ordinance on Prevention of Ionizing Radiation Hazards, and the lens equivalent dose limit was revised from 150 to 100 mSv over 5 years and 50 mSv/year (revised in April 2021).
The occupational dose to the lens of physicians involved in radiology procedure has been reported to be significant in interventional radiology (IVR) procedures for cerebrovascular 2 , 3 , 4 , 5 and cardiovascular 3 , 6 medicine, tumors, 3 , 7 endoscopic retrograde cholangiopancreatography (ERCP), 3 , 8 , 9 and orthopedic surgery. 10 The lens equivalent dose limit is considered to have exceeded when cardiologists and gastroenterologists perform radiology procedure without radiation protection for the lens of the eye. 3 , 6 , 9 , 11 For this reason, the International Atomic Energy Agency disseminates information regarding the possibility of reducing exposure using lead glasses and ceiling‐mounted radiation shielding screens and educates the workers on the importance of eye lens protection. 12
ICRP reported that many physicians who perform radiology procedures have inadequate radiation protection. 13 , 14 Although the wear rate of lead aprons and neck guards by physicians performing radiology procedures is higher than 90%, 15 , 16 the wear rate of lead glasses is 30%–52%. 3 , 15 , 16 , 17 , 18 Moreover, despite the ability of lead‐containing ceiling‐mounted radiation shielding screens to reduce eye lens exposure by over 70%, 19 , 20 , 21 , 22 these screens are not always used appropriately in actual medical procedure, 17 putting physicians at risk of receiving high radiation doses to the lens of the eye.
In this multicenter study, by applying basic principles of occupational health, we proposed “multiple radiation protection” measures that did not place an undue physical burden on physicians and were less expensive for medical facilities and estimated the effectiveness of these measures. In addition, the occupational dose to the lens of the eye for physicians was measured on a case‐by‐case basis, and the potential for compliance with the new‐equivalent dose limits to the lens of the eye (ICRP: average annual limit, 20 mSv/year over 5 years) was assessed.
2. METHODS
2.1. Participants for measurement
Between April 2019 and July 2019, 15 physicians engaged in radiology procedure (angiography, non‐angiography, or IVR procedure) at 15 medical facilities in Japan were nominated by their respective societies (Japanese Society of Radiology and the Japanese Society of Interventional Radiology, Japanese Orthopedic Association, Japanese Society of Gastroenterology, and Japanese Society of Neuroendovascular Therapy). Eye lens doses (3‐mm dose equivalent, i.e., Hp(3)) were measured for each participant when they performed radiology procedure using conventional methods before implementing radio‐protective measures (before radiation protection measures, Table 1) and after the implementation of radio‐protective measures (after radiation protection measures, Table 1), taking into account the facility environment and the procedures in place at each medical facility. We confirmed the doctor's radiation protection method from the pre‐questionnaire and the photographs during the procedure. The personal dose values for the past 3 years and the number of procedures performed over the past year were also investigated. Since there was no evaluation of 3‐mm dose equivalent at that time, the personal dose values were defined as the 70‐µm dose equivalent of the skin or the 1‐cm dose equivalent of the effective dose, whichever is larger, as the eye lens dose.
TABLE 1.
Physician | Measures before/after | Simultaneous irradiation | Pulse rate reduction | Irradiation field | IR | Lead glasses | Shielding screens | Cu | Cs | R | Evacuation from the room |
---|---|---|---|---|---|---|---|---|---|---|---|
Cardiologist A | Before | ○ | ○ | ○ a | ● f | ○ | |||||
After | ○ | ○ | ○ | ○ a | ○ f | ○ | |||||
Cardiologist B | Before | ○ | ○ | ○ | ● g | ○ | |||||
After | ○ | ○ | ○ | ○ a | ○ g | ○ | |||||
Cardiologist C | Before | ○ | ○ | ○ | ○ b | ○ | |||||
After | ○ | ○ | ○ | ○ c | ○ g | ○ | |||||
Neurosurgery D | Before | ○ d | ○ | ||||||||
After | ○ | ○ | ○ | ○ d | ○ g | ○ | |||||
Neurosurgery E | Before | ○ | ○ a | ● g | ○ | ○ | |||||
After | ○ | ○ | ○ | ○ a | ○ g | ○ | ○ | ||||
Gastroenterologist F | Before | ○ | ○ a | ● h | |||||||
After | ○ | ○ | ○ a | ○ h | |||||||
Gastroenterologist G | Before | ||||||||||
After | ○ a | ||||||||||
Gastroenterologist H | Before | ○ a | ○ i | ||||||||
After | ○ | ○ e | ○ i | ||||||||
Gastroenterologist I | Before | ● h | |||||||||
After | ○ a | ○ h | |||||||||
Gastroenterologist J | Before | ○ h | |||||||||
After | ○ a | ○ h | |||||||||
Orthopedic Surgeon K | Before | ○ | ○ d | ● | |||||||
After | ○ | ○ | ○ | ○ d | ○ | ○ | ○ | ||||
Orthopedic Surgeon L | Before | ||||||||||
After | ○ d | ||||||||||
Orthopedic Surgeon M | Before | ||||||||||
After | ○ a | ||||||||||
Radiologist N | Before | ○ | ○ | ○ | ○ a | ○ g | ○ | ○ | |||
After | ○ | ○ | ○ | ○ a | ○ g | ○ | ○ | ||||
Radiologist O | Before | ○ | ○ | ○ | ○ a | ○ g | ○ | ○ | |||
After | ○ | ○ | ○ | ○ a | ○ g | ○ | ○ |
○ Radioprotective equipment was being used appropriately. ● Radioprotective equipment was used, but the method of use was inappropriate. Simultaneous irradiation = reduction of simultaneous frontal and lateral irradiation in fluoroscopy mode; Pulse rate reduction = appropriate selection/switching of fluoroscopy mode (changed from 15 to 7.5 pps); Irradiation field = restriction of irradiation field to target region; IR = dose reduction using iterative reconstruction (IR) during CT fluoroscopy; Shielding screens = ceiling‐mounted radiation‐shielding screens; Cu = under‐bed protective curtains; Cs = scatter‐radiation protection curtains; R = RADPAD®; Evacuation from the room = room evacuation during imaging mode.
Lead glasses (Panorama shield®; HF‐400 ultra‐light 0.07‐mm Pb; Toray).
Lead glasses (HAGOROMO Face Guard FG06‐110; 0.06‐mm Pb; Maeda).
Lead glasses (CROSSLINK 0.75‐mm Pb; Barrier technologies®).
Lead glasses (Panorama shield®; HF‐350 ultra‐light 0.07‐mm Pb; Toray).
Lead glasses (ProTech eyewear PT‐COMET 0.75‐mm Pb; Maeda).
Ceiling‐mounted radiation shielding screen 350 (0.5‐mm Pb; Kenex).
Ceiling‐mounted radiation shielding screen (MAVIG 0.5‐mm Pb; MAVIG GmbH).
Scatter‐radiation protection curtains (Scatter protection cloth NP, 0.125‐mm Pb; Maeda).
Scatter‐radiation‐protection curtains (self‐made scatter‐protection clothing, 0.25‐mm Pb).
2.2. X‐ray equipment and radio‐protective methods
Of the 15 participating medical institutions, six used biplane angiography, three used hybrid single‐plane angiography combined with X‐ray computed tomography (CT), and one used surgical X‐ray fluoroscopy. The remaining five centers used X‐ray fluoroscopy systems, of which four used over‐table X‐ray tube systems and one used an under‐table X‐ray tube system.
Radiation protection measures for physicians involved in radiology procedures included the use of reduction of simultaneous front‐to‐side irradiation during fluoroscopy, appropriate selection/switching of the fluoroscopy mode (switching from 15 pps to 7.5 pps), restriction of the irradiation field to the target range, dose reduction performed using iterative reconstruction (IR) in CT fluoroscopy in combination to avoid negatively influencing radiology procedure, lead glasses, ceiling‐mounted radiation‐shielding screens, scatter‐radiation–shielding curtain for over‐table X‐ray tube systems, and under‐bed protective curtains, RADPAD® (0.25 mmPb, Nippon Medical Readers Co., Ltd.), evacuation of the examination room during imaging. The lead glasses used were as follows: HF‐400 (0.07‐mm Pb‐equivalent; Toray Medical Inc., n = 11), HF‐350 (0.07‐mm Pb‐equivalent; Toray Medical Inc., n = 3), FG06‐110 (0.06‐mm Pb‐equivalent; Maeda, n = 1), CROSSLINK (0.75‐mm Pb; Barrier technologies®, n = 1) and PT‐COMET (0.75‐mm Pb‐equivalent; Maeda, n = 1). The ceiling‐mounted radiation‐shielding screens used were as follows: ceiling‐mounted radiation shield 350 (0.5‐mm Pb‐equivalent; Kenex, n = 1) and MAVIG (0.5‐mm Pb‐equivalent; MAVIG GmbH, n = 6). In addition, for the over‐table X‐ray tube systems, scatter protection was provided by NP cloth (0.125‐mm Pb‐equivalent; Maeda, n = 3) and facility‐made scatter‐protection curtain produced by each medical institution (0.25‐mm Pb‐equivalent, n = 1) (Table 1).
2.3. Details concerning radiation protection
We confirmed the physician's radiation protection method from the pre‐questionnaire and the photographs during the procedure. We have proposed the main protection methods based on the current situation. (Based on the three principles of external exposure protection, installing a ceiling‐mounted radiation shielding screen, reducing the pulse rate within the range that does not deteriorate the image quality, and if necessary, they were instructed to wear protective equipment.)
2.4. Method for measuring occupational dose to the lens of the eye
Physicians wore lead aprons and lead glasses. The Hp(3) to the lens of the eye was obtained from air kerma measurements obtained by radio‐photoluminescence glass dosimeters 23 (GD‐352M; Chiyoda Technol) attached to the inner and exterior sides of the lead glasses. The GD‐352M used in the measurements complied with the IEC62387 requirements for dosimetry systems with passive detectors and provided stable dose linearity in the low dose range (less than ±5.0% in the range of 0.01 to 50 mGy). 23 , 24 Before the start of this study, the coefficient of variation was confirmed to not exceed 3.0%. We modified a previously reported eye lens dosimeter clip 25 component and attached it to the left and right sides of the lead glasses and placed one GD‐352M unit each in fixed positions on the inner and outer sides of the lens (Figure 1A,B). For lead glasses where the eye lens dosimeter clips could not be used, GD‐352M units were attached to the left and right sides of the lens using adhesive tape (Figure 1C,D).
After the measurements were completed, the radio‐photoluminescence glass dosimeters were stored in a low‐background area outside the radiation‐controlled area and returned to the providing university by postal mail after the survey period. The data were then read and analyzed using a reading device (FGD‐1000; Chiyoda Technol) installed at our institution.
The eye lens dose Hp(3) in this study was calculated from air kerma measurements obtained using the radio‐photoluminescence glass dosimeters. Specifically, the air kerma to Hp(3) conversion coefficient K (Hp(3)/air kerma) on a cylindrical phantom (φ20 cm × 20 cm) was calculated using a Monte Carlo simulation from a previous report, 26 and the conversion was performed according to Equation (1). In this study, the effective energy used in radiology procedure was assumed to be 50 keV, and 1.590 Sv/Gy was adopted for K (Hp(3)/air kerma):
(1) |
where Hp(3) is the lens dose of the physicians eye per procedure (μSv), air kerma is the radio‐photoluminescent glass dosimeter measured value (μGy), K is the air kerma to Hp(3) conversion coefficient (Hp(3)/air kerma) (Sv/Gy).
For each physician, the Hp(3) values measured before the radiation protection measures were compared between the left and right eyes, and the value indicating a greater dose was recorded as the Hp(3) in this study. For evaluations after the radiation protection measures, the Hp(3) values for the dose on the same side as that measured before the measures were recorded.
2.5. Method for calculating Hp(3)rate to the lens of the eye
To determine the lens dose of the eye per unit time, the Hp(3) values obtained in Equation (1) were divided by the fluoroscopy time of the procedure to obtain the eye lens dose rate per unit time (Hp(3)rate) according to Equation (2) below:
(2) |
where Hp(3)rate is the lens dose of the eye per unit time (μSv/min), Hp(3) is the lens dose of the physicians’ eye per procedure (μSv), T is the fluoroscopy time during the procedure (min).
2.6. Calculation of number of possible annual radiology procedures by physicians
The median value of lens dose of the physicians’ eye (Hp(3)procedure) was calculated from the median value of the Hp(3)rate calculated in Equation (2) and the median fluoroscopy time of the procedures performed before and after the radiation protection measures. The number of possible annual radiology procedures was obtained based on Equations (3) and (4), respectively:
(3) |
(4) |
where Hp(3)procedure is the median value of lens dose of the physicians eye (μSv/procedure), Hp(3)rate median is the median value of eye lens dose rate per unit time (μSv/min), T median is the median value of fluoroscopy time before or after radiation protection measures (min), Hlens is the new‐equivalent dose limits to the lens of the eye (ICRP: average annual limit, 20 mSv/year over 5 years).
2.7. Analysis of the physician lens dose reduction rate with before and after radiation protection measures
To determine the dose reduction effect of radiation protection measures, the dose reduction rate (DRR) attributable to measures was calculated from before the measures Hp(3)procedure (Hp(3)procedure before) and the Hp(3) after the measures (Hp(3)procedure after) according to Equation (5).
(5) |
where DRR is the dose‐reduction rate, Hp(3)procedure before is before the measures Hp(3)procedure (μSv/procedure), Hp(3)procedure after is after the measures Hp(3)procedure (μSv/procedure).
2.8. Statistical analysis
Differences in Hp(3)rate before and after the radiation protection measures were confirmed using the Kruskal–Wallis one‐way analysis of variance. When the one‐way analysis of variance result was significant, the difference between the individual ‐before‐ and ‐after‐ radiation protection measures was evaluated using the Dunn test (with Bonferroni correction). Differences in fluoroscopy time before and after the radiation protection measures were confirmed using the Mann–Whitney U test. Differences were considered statistically significant at P < .05. All analyses were performed using the Statistical Package for the Social Science (version 25.0, IBM Corporation).
2.9. Ethical considerations
The study was approved by the Ethics Committee of the University of Occupational and Environmental Health, Kitakyushu, Japan (Protocol Number R1‐054).
3. RESULTS
One orthopedic surgeon who had incorrectly installed an eye lens dosimeter, lost a radio‐photoluminescence glass dosimeter, and failed to keep a dosimeter was excluded from the analysis, and the data from the remaining 14 physicians were analyzed.
3.1. Eye lens dose over the previous 3 years (past personal dose information)
Past personal dose information concerning the physicians eye lens dose is shown in Table 2. Since there was no evaluation of 3‐mm dose equivalent at that time, the personal dose values were defined as the 70‐µm dose equivalent of the skin or the 1‐cm dose equivalent of the effective dose, whichever is larger, as the eye lens dose. The proportion of doses exceeding the new‐equivalent dose limits to the lens of the eye was 27% (4/14). Particularly high Hp values were reported for cardiologist A and gastroenterologist F (42.3 and 75.3 mSv/year, respectively).
TABLE 2.
Physicians | FY2016 | FY2017 | FY2018 |
---|---|---|---|
Cardiologist A | 49.8 | 51.7 | 42.3 |
Cardiologist B | 18 | 15.2 | 31.0 |
Cardiologist C | N/A a | N/A a | N/A a |
Neurosurgeon D | 14.5 | 12.8 | 11.9 |
Neurosurgeon E | 12.1 | 14.5 | 12.7 |
Gastroenterologist F | 22.1 | 21.4 | 75.3 |
Gastroenterologist G | N/A a | N/A a | 3.6 |
Gastroenterologist H | 0.9 | 0.0 | 0.3 |
Gastroenterologist I | 8.9 | 15.3 | 9.9 |
Gastroenterologist J | N/A c | N/A c | 27.1 |
Orthopedic Surgeon K | N/A c | 0.5 | 0.8 |
Orthopedic Surgeon L | N/A a | N/A a | N/A a |
Orthopedic Surgeon M | N/A b | N/A b | N/A b |
Radiologist N | N/A c | 6.0 | 3.9 |
Radiologist O | N/A c | N/A c | 9.0 |
Equivalent dose limits for the lens of the eye >20 mSv/year are in boldface. Hp(0.07) = 70‐μm dose equivalent; FY = fiscal year.
Personal dosimeters had not been distributed by the hospital.
Personal dosimeters had been distributed by the hospital but were not being used.
No information on radiation dose at other hospitals before joining the hospital staff.
3.2. Number of radiology procedures and fluoroscopy time during which eye lens dosimetry was performed
The numbers of radiology procedures in which eye lens dosimetry measurements were taken were 5 ± 2 before the measures and 5 ± 2 after the measures (Table S1). Moreover, the fluoroscopy times recorded during the monitoring period were 10.3 (range: 0.4–114.8) min before the measures and 12.4 (range: 0.3–80.8) min after the measures. We performed the parametric test with fluoroscopy time before and after the measures and the result did not show any significant difference (P = .466, Mann–Whitney U test).
3.3. Details concerning radiation protection
Table 1 shows the status of radiation protection in each physicians before and after the radiation protection measures. We confirmed the physicians radiation protection method from the pre‐questionnaire and the photographs during the procedure. In assessments of the usage rate of lead glasses, 64% (9/14) of the physicians used lead glasses before the measures and 100% (14/14) did so after the measures. In this study, seven of the 14 facilities had a ceiling‐mounted radiation shielding screen. Although 29% (2/7) of the participants properly used a ceiling‐mounted radiation shielding screen before the measures, 100% (7/7) did so after the measures. With respect to the fluoroscopy devices used by gastroenterologists, 80% (4/5) used over‐table X‐ray tubes, whereas a scatter‐protection curtain was used by 100% (4/4) of the physicians to shield scatter radiation from patients.
3.4. Eye lens dose of physicians involved in radiology procedures
Median the Hp(3)rate of all physicians who participated in the study was reduced by the radiation protection measures. In particular, median the Hp(3)rate significantly reduced after the measures in the field of cardiovascular internal medicine (P = .011, Kruskal–Wallis test, Table 3) and neurology (P < .01, Kruskal–Wallis, Table 3). In contrast, the fluoroscopy time showed no differences among radiology procedure.
TABLE 3.
Physician | Before (After) a | Lg Y/N | Hp(3)rate [μSv/min] | ANOVA b (P value) | Fluoroscopy time [min] | P value c | |||
---|---|---|---|---|---|---|---|---|---|
Before‐measures | After‐measures | Before‐measures | After‐measures | ||||||
Outside | Inside | Inside | |||||||
Median [range] | Median [range] | Median [range] | Median [range] | Median [range] | |||||
Cardiologist A | 9 (9) | Y | 12.0 [3.1–72.8] | 5.2 [1.4–32.5] | 2.6 [0.6–11.5] | 15.3 [4.5–25.4] | 17.7 [8.6–34.4] | ||
Cardiologist B | 5 (2) | N | 7.6 [2.0 –23.1] | 3.9 [0.7–23.7] | 2.8 [0.7–4.9] | 23.2 [13.6–36.6] | 29.4 [22.1–36.6] | ||
Cardiologist C | 2 (5) | Y | 23.5 [13.7–33.3] | 13.4 [6.4–20.4] | 1.5 [0.8–37.9] | 15.5 [12.5–18.5] | 28.8 [7.3–44.1] | ||
Circulatory Internal Medicine | 16 (16) | 12.0 * [2.0–72.8]† | 5.0 * [0.7–32.5] | 2.6 * [0.6–38.0]† | .011 | 17.2 [4.5–36.6] | 23.2 [7.3–44.1] | .363 | |
Neurosurgeon D | 5 (5) | Y | 7.8 [1.3–16.7] | 3.6 [1.4–7.1] | 0.8 [0.5–2.2] | 35.0 [20.3–84.4] | 39.8 [26.9–79.3] | ||
Neurosurgeon E | 5 (5) | Y | 5.0 [4.0–11.6] | 2.1 [1.5–5.3] | 0.5 [0.4–1.0] | 10.3 [9.2–32.9] | 20.4 [10.9–21.9] | ||
Cerebrovascular Medicine | 10 (10) | 5.0 * [1.3–16.7]‡ | 2.5 * [1.4–7.1]§ | 0.7 * [0.4–2.2]‡§ | <.01 | 25.9 [9.2–84.4] | 24.4 [10.9–79.3] | .705 | |
Gastroenterologist F | 5 (5) | Y | 6.6 [4.2–14.6] | 3.7 [2.2–7.7] | 1.4 [0.6–4.9] | 13.8 [10.1–63.9] | 15.1 [6.9–28.8] | ||
Gastroenterologist G | 4 (4) | N | 1.1 [0.6–3.2] | — | 0.6 [0.33–1.5] | 22.7 [9.5–56.0] | 22.7 [9.5–56.0] | ||
Gastroenterologist H | 3 (3) | Y | 0.4 [0.0–0.9] | 0.2 [0.2–0.4] | 0.0 [0.0–1.3] | 2.5 [1.4–2.6] | 2.0 [1.8–2.5] | ||
Gastroenterologist I | 5 (5) | N | 8.3 [3.65–12.3] | — | 5.1 [2.0–9.1] | 3.0 [2.0–33.0] | 3.0 [2.0–33.0] | ||
Gastroenterologist J | 5 (5) | N | 2.2[0.6–3.2] | — | 1.2 [0.1–2.3] | 5.3 [1.9–10.8] | 5.3 [1.9–10.8] | ||
Gastroenterological Medicine | 22 (22) | 2.7 [0.0–15.0] | 2.5 [0.2–8.0] | 1.3 [0.0–4.9] | .073 | 9.3 [1.4–63.9] | 8.8 [1.8–56.0] | .743 | |
Orthopedic Surgeon K | 10 (7) | Y | 9.0 [0.0–41.9] | 8.0 [0.0–19.8] | 3.1 [0.0–13.7] | 1.1 [0.4–3.2] | 0.6 [0.3–3.1] | ||
Orthopedic Surgeon L | 5 (5) | N | 2.9 [0.6–4.9] | — | 1.1 [0.6–1.7] | 6.6 [2.4–7.3] | 6.6 [2.4–7.3] | ||
Orthopedic Surgery | 15 (12) | 6.4 [0.0–41.9] | 8.0 [0.0–19.8] | 2.5 [0.0–13.7] | .194 | 2.0 [0.4–7.3] | 1.6 [0.3–7.3] | .807 | |
Radiologist N | 6 (6) | Y | 1.9 [0.4–4.0] | 1.8 [0.5–1.4] | 0.7 [0.0–4.0] | 48.8 [8.2–114.8] | 40.6 [10.5–80.8] | ||
Radiologist O | 6 (4) | Y | 1.4 [0.2–5.6] | 0.7 [0.0–2.8] | 0.3 [0.0–1.0] | 12.6 [1.8–30.8] | 17.3 [0.3–43.0] | ||
Radiological Medicine | 12 (10) | 1.9 [0.2–5.6] | 0.8 [0.0–2.8] | 0.4 [0.0–4.0] | .219 | 21.5 [1.8–114.8] | 36.4 [0.3–80.8] | .997 |
The significant difference (P < .05.) is set in boldface.
Lg = lead glasses; Y/N = wear glasses/do not wear glasses; Hp(3)rate = lens dose of the eye per unit time (3‐mm dose equivalent).
Number of Before‐measures (After‐measures) procedures.
Differences in Hp(3)rate before and after the radiation protection measures were confirmed using the Kruskal–Wallis one‐way analysis of variance. When the one‐way analysis of variance result was significant, the difference between the individual before and after radiation protection measures was evaluated using the Dunn test (with Bonferroni correction).
Differences in fluoroscopy time before and after the radiation protection measures were confirmed using the Mann–Whitney U test. The same symbols indicate that Bonferroni‐corrected Dunn test multiple comparisons showed significant differences. All analyses were performed using the Statistical Package for the Social Science (version 25.0, IBM Corporation).
P < .05.
3.5. Annual number of cases indicated for radiology procedure per physician
Based on the new‐equivalent dose limits to the lens of the eye and the Hp(3)procedure before (μSv/procedure), the number of cases indicated for radiology procedure per year was calculated, and this number was lower than the number of radiology procedures performed before the radiation protection measures in fiscal year 2018 (FY2018) for three physicians (Table 4). However, after the implementation of the proposed radiation protection method (Table 1), the number of cases indicated for radiology procedure exceeded the number of radiology procedures performed in FY2018 for all physicians (Table 4). The DRR of this study was 53% (range: 37%–79%).
TABLE 4.
Physician | Number of radiology practice | Hp(3)procedure (μSv/procedure) | DRR c [%] | Number of possible annual radiology practices | ||
---|---|---|---|---|---|---|
FY2018 | Hp(3)procedure before a | Hp(3)procedure after b | Before‐measures d | After‐measures e | ||
Cardiologist A | 421 | 79.6 | 46.0 | 42 | 251 | 435 |
Cardiologist B | 26 | 176.3 | 82.3 | 53 | 113 | 243 |
Cardiologist C | 154 | 207.7 | 43.2 | 79 | 96 | 463 |
Neurosurgeon D | 120 | 126.0 | 79.0 | 37 | 159 | 253 |
Neurosurgeon E | 120 | 21.6 | 10.2 | 53 | 925 | 1961 |
Gastroenterologist F | 397 | 51.1 | 21.1 | 59 | 392 | 946 |
Gastroenterologist G | 40 | 24.8 | 13.5 | 45 | 807 | 1479 |
Gastroenterologist H | 70 | 1.0 | N/A f | — | 20 000 | — |
Gastroenterologist I | 111 | 24.9 | 15.3 | 39 | 803 | 1307 |
Gastroenterologist J | 304 | 11.7 | 6.4 | 46 | 1705 | 3145 |
Orthopedic Surgeon K | 97 | 8.8 | 2.0 | 78 | 2273 | 10 187 |
Orthopedic Surgeon L | Unknown | 19.1 | 7.3 | 62 | 1045 | 2755 |
Radiologist N | 202 | 87.8 | 28.4 | 68 | 228 | 704 |
Radiologist O | 222 | 8.8 | 5.2 | 41 | 2268 | 3854 |
Number of radiology practice FY2018 > number of possible radiology practices per year are set in boldface; FY2018 = fiscal year 2018; DRR = dose reduction rate.
The median value of eye lens dose before radiation protection measures.
The median value of eye lens dose after radiation protection measures.
The dose‐reduction rate (DRR) [%] = (1 − b/a) 100.
Number of possible annual radiology practices (before radiation protection measures) = The equivalent dose limits to the lens of the eye (ICRP: average annual limit, 20 mSv/year over 5 years) /a.
Number of possible annual radiology practices (after radiation protection measures) = The equivalent dose limits to the lens of the eye (ICRP: average annual limit, 20 mSv/year over 5 years)/b.
Less than the lower limit of dose measurement.
3.6. Case 1: Radiation protection measures for neurosurgeons
The number of IVRs in the field of head and neck medicine performed by neurosurgeon D, was 120 in FY2018 (Table 4). The equivalent eye lens dose calculated from the personal dosimeter attached to the neck was 11.9 mSv/year (Table 2), which was within the new‐equivalent dose limits to the lens of the eye. Even if neurosurgeon D maintains the dose (11.9 mSv/year), the total working period of 50 years is 595 (11.9 mSv/year × 50 years) mSv, which exceeds the radiation cataract threshold of 500 mGy. Therefore, it is necessary to adhere to the dose limit and optimize it in addition. As such, the radiation protection status of neurosurgeon D was investigated (Table 1). We found that although lead glasses were used, a ceiling‐mounted radiation shielding screen was not (Figure 2A). In addition, this physician performed procedures with simultaneous frontal and lateral irradiation and a high pulse rate (15 pps) during fluoroscopy. As such, the Hp(3)rate median before the measures was 3.6 μSv/min. The Hp(3)procedure before was 126.0 μSv/procedure based on this Hp(3)rate median, and the median fluoroscopy time was 35.0 min (median) before radiation protection measures. The resultant 159 IVR procedures targeting the head and neck region would exceed the new‐equivalent dose limits to the lens of the eye.
As such, we urged neurosurgeon D to use a ceiling‐mounted radiation shielding screen and to reduce the radiation dose (i.e., to reduce the simultaneous irradiation of the front side during fluoroscopy and to reduce the pulse rate during fluoroscopy to 7.5 pps), and the eye lens dose Hp(3) on the inner side of the lead glasses was evaluated under these conditions (Figure 2B). As a result, the Hp(3)rate median after the radiation protection measures was 0.8 μSv/min, and the Hp(3)procedure before was 79.0 μSv/procedure, calculated from the fluoroscopy time of 15.1 min (median) after the radiation protection measures. This indicates that 253 IVR procedures at the head and neck region could be performed, and the possibility of exceeding the new‐equivalent dose limits to the lens of the eye was low, even considering the 120 procedures performed in FY2018.
3.7. Case 2: Radiation protection measures for gastroenterologists
Gastroenterologist F, performed 397 radiology examinations as a part of ERCP procedures in FY2018 (Table 4). The eye lens equivalent dose calculated from the personal dosimeter attached to the neck was 75.3 mSv/year (Table 2), indicating that the exposure with the current radiation protection method exceeded the new‐equivalent dose limits to the lens of the eye. We investigated the radiation protection status of Gastroenterologist F (Table 1) and found that although the physician used scatter‐protection curtain and lead glasses, there may have been inadequacies in the use of scatter‐protection curtain. Moreover, gastroenterologist F was found to be performing the procedures at a high pulse rate (15 pps). As such, we measured the Hp(3)rate median before the radiation protection measures and found that it was 3.7 μSv/min., which was the highest among participating gastroenterologists using over‐table X‐ray tubes. Since the Hp(3)procedure before obtained from this Hp(3)rate median value and the median fluoroscopy time of 13.8 min before the radiation protection measures was 51.1 μSv/procedure, the new equivalent dose limits to the lens of the eye could be exceeded after 392 ERCP procedures. Therefore, we evaluated the Hp(3) on the inner side of the lead glasses after explaining the proper use of scatter‐protection curtain (Figures 2C,D) to gastroenterologist F and urging this physician to reduce the radiation dose by reducing the pulse rate during fluoroscopy to 7.5 pps. Subsequently, the Hp(3)rate median inside the lead glasses was 1.4 μSv/min, and the Hp(3)procedure before was 21.1 μSv/procedure, calculated from the median fluoroscopy time of 15.1 min after the radiation protection measures. At this exposure, 946 ERCP procedures could be performed under the new‐equivalent dose limits to the lens of the eye, and the possibility of exceeding the new equivalent dose limits to the lens of the eye was low even when considering the 397 procedures performed in FY2018.
4. DISCUSSION
In this multicenter study, we investigated the Hp(3) of physicians on a case‐by‐case basis and evaluated the potential for compliance with the new‐equivalent dose limits to the lens of the eye. In addition, we proposed “multiple radiation protection” measures for physicians engaged in radiology procedure according to the basic principles of occupational health and showed that adherence to these measures would ease compliance with the new‐equivalent dose limits to the lens of the eye without imposing an excessive burden on physicians or medical facilities.
Many physicians in the field of radiology perform procedures without adequate training and knowledge of radiation protection. 13 , 14 In this study, radiation protection was properly implemented only during procedures performed by radiologists who specialized in diagnostic imaging using radiation, and procedures performed in non‐radiology departments showed inadequate implementation of radiation protection measures (Table 1). The usage rate of lead glasses increased from 64% (9/14) to 100% (14/14) after‐measures. This result is higher than that reported in previous studies (30%–52%). 3 , 15 , 17 , 18 In addition, 50% (7/14) of the facilities had ceiling‐mounted radiation shielding screens, but only 29% (2/7) of the physicians (all physicians were radiologists) were able to use these screens appropriately. Although the shielding effect of the lead‐containing ceiling‐mounted radiation shielding screens is known to be high, 19 , 20 , 21 , 22 cardiologists and neurosurgeons in our study were unable to use the radiation shielding screens properly—the same as previous reports. 17 Therefore, we created a video explaining the proper use of ceiling‐mounted radiation shielding screens and encouraged the use of this protective equipment in close contact with patients (Video S1). After the measures, 100% (7/7) of the physicians used the ceiling‐mounted radiation shielding screen in an appropriate configuration. In addition, despite the fact that the scatter‐protection curtain used in gastroenterology procedure has been reported to have a high shielding effect, 27 , 28 some gastroenterologists who used scatter‐protection curtain had a high Hp(3)rate median. As a factor, the shielding effect may be reduced if the scatter‐protection curtain is pulled up during fluoroscopy to observe a patient's chest movement. 27 , 28 Since monitoring of the patient's respiratory status is essential in procedures such as ERCP, during which the patients are sedated, we recommend the use of devices such as pulse oximeters for respiratory management in such cases.
The Hp(3)rate median decreased in all medical fields. In particular, in the cardiovascular internal medicine (P = .011, Kruskal–Wallis, Table 3) and head and neck medicine (P < .01, Kruskal–Wallis, Table 3) fields, the Hp(3)rate median decreased significantly after the implementation of our proposed radiation protection measures. In addition to the appropriate use of a ceiling‐mounted radiation shielding screen (Figure 2B), the following measures were employed to reduce the overall radiation dose and significantly reduce the Hp(3): avoiding simultaneous front‐to‐side irradiation during fluoroscopy, using a low pulse rate (7.5 pps) to the extent that it did not affect radiology procedures, and restricting the irradiation field to the target imaging range. The Hp(3)rate median before and after the radiation protection measures in the field of orthopedic surgery were 6.4 and 2.5 μSv/min, respectively (Table 3), indicating a certain dose rate reduction effect (P = .194, Kruskal–Wallis, Table 3). The Hp(3)rate median in the field of orthopedic surgery was higher than that for gastroenterologists, who are known to have higher occupational eye lens doses, and the dose rate was comparable to that of cardiologists.
Radiation protection for physicians engaged in radiology procedures should be considered according to the three basic management principles of occupational health. 29 In this study, from the viewpoint of working environment management, we proposed the following measures: reducing the use of simultaneous front‐to‐side irradiation during fluoroscopy, appropriate selection/switching of the fluoroscopy mode (from 15 to 7.5 pps), restricting the irradiation field to the target area, and dose reduction using IR in CT fluoroscopy. Next, from the viewpoint of working management, we proposed the use of radiation protection equipment such as lead glasses, ceiling‐mounted radiation shielding screens, under‐bed protective curtains, and scatter radiation protection curtain and evacuation of the room during imaging. The implementation of the “multiple radiation protection” measures, which represented a combination of working environment management and working management protocols, did the DRR by 53% (range: 37%–79%) without placing any extraordinary burden on either medical staff or medical facilities. The number of cases indicated for radiology procedure calculated from the Hp(3)procedure after [μSv/procedure] after the radiation protection measures also suggested that all physicians could comply with the new eye lens dose equivalent (Table 4). Incorporating the three areas of the fundamentals of industrial health management into the concept of radiation protection measures can effectively reduce eye lens dose without overburdening medical facilities or individuals. These results clearly suggest that physicians eye lens doses can be significantly reduced by providing appropriate advice on radiation protection. However, to implement “multiple radiation protection,” radiologists, who are actually responsible for radiation exposure control in the medical field, must actively intervene to ensure compliance with radiation protection protocols and to improve the occupational health environment.
5. CONCLUSION
A case‐by‐case evaluation of the eye lens dose Hp(3) of physicians involved in radiology procedure revealed that 21% (3/14) of the physicians would be exposed to eye lens doses higher than the new‐equivalent dose limits to the lens of the eye if they performed the same number of radiology procedures as they did in FY2018 during their before‐measures radiation protection status. However, the findings also indicated that by adhering to the basic principles of occupational health, implementing the principle of “multiple radiation protection” through the review of fluoroscopy procedures and pulse rates, and using ceiling‐mounted radiation shielding screens, scatter‐protection curtains, and lead glasses, compliance with the new‐equivalent dose limits to the lens of the eye (ICRP: average annual limit, 20 mSv/year over 5 years) could be achieved without imposing an undue burden on either the physician or the medical facility.
DISCLOSURE
Approval of the research protocol: The study was approved by the Ethics Committee of the University of Occupational and Environmental Health, Kitakyushu, Japan (Protocol Number R1‐054). Informed consent: N/A. Registry and the registration no. of the study/trial: N/A. Animal studies: N/A. Conflict of interest: The authors declare that there is no conflict of interest.
AUTHOR CONTRIBUTIONS
Keisuke Nagamoto: Conceptualization, methodology, data curation, writing ‐ original draft, writing ‐ review & editing. Takashi Moritake: Conceptualization, formal analysis, writing ‐ review & editing, project administration, funding acquisition. Koichi Nakagami: Validation, investigation, visualization. Koichi Morota: Validation, data curation. Satoru Matsuzaki: Validation, investigation. Naoki Kunugita: Supervision, funding acquisition.
Supporting information
ACKNOWLEDGMENTS
The authors would like to thank Miho Itonaga, Tomomi Konari, Fumiko Kawagoe, Kimi Kihara, and Chisato Hori, for their technical support and assistance. The authors would also like to thank all medical staff who participated in this project.
Nagamoto K, Moritake T, Nakagami K, Morota K, Matsuzaki S, Kunugita N. A multicenter study of radiation doses to the eye lenses of clinical physicians performing radiology procedures in Japan. J Occup Health. 2021;63:e12305. doi: 10.1002/1348-9585.12305
Funding information
This work was supported in part by grants from the Japanese Ministry of Health, Labour and Welfare (Grant Numbers 180501‐01/200601‐01/210501‐01) and JSPS KAKENHI (Grant Numbers JP18K09959/JP19K10498).
DATA AVAILABILITY STATEMENT
On reasonable request, derived data supporting the findings of this study are available from the corresponding author.
REFERENCES
- 1. ICRP . ICRP statement on tissue reactions/early and late effects of radiation in normal tissues and organs – threshold doses for tissue reactions in a radiation protection context. Ann ICRP. 2012;41:1‐322. [DOI] [PubMed] [Google Scholar]
- 2. Moritake T, Matsumaru Y, Takigawa T, et al. Dose measurement on both patients and operators during neurointerventional procedures using photoluminescence glass dosimeters. AJNR Am J Neuroradiol. 2008;29:1910‐1917. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 3. Vanhavere F, Carinou E, Domienik J, et al. Measurements of eye lens doses in interventional radiology and cardiology: final results of the ORAMED project. Radiat Meas. 2011;46:1243‐1247. [DOI] [PubMed] [Google Scholar]
- 4. Wong JHD, Anem LEA, Tan S, et al. Eye lens dose of medical personnel involved in fluoroscopy and interventional procedures at a Malaysian Hospital. Phys Med. 2019;68:47‐51. [DOI] [PubMed] [Google Scholar]
- 5. Kato M, Chida K, Ishida T, et al. Occupational radiation exposure of the eye in neurovascular interventional physician. Radiat Prot Dosimetry. 2019;185:151‐156. [DOI] [PubMed] [Google Scholar]
- 6. Haga Y, Chida K, Kaga Y, et al. Occupational eye dose in interventional cardiology procedures. Sci Rep. 2017;7:569. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 7. O'Connor U, Walsh C, Gallagher A, et al. Occupational radiation dose to eyes from interventional radiology procedures in light of the new eye lens dose limit from the International Commission on Radiological Protection. Br J Radiol. 2015;88:20140627. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 8. O’Connor U, Gallagher A, Malone L, et al. Occupational radiation dose to eyes from endoscopic retrograde cholangiopancreatography procedures in light of the revised eye lens dose limit from the International Commission on Radiological Protection. Br J Radiol. 2013;86:20120289. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 9. Matsubara K, Takei Y, Mori H, et al. A multicenter study of radiation doses to the eye lenses of medical staff performing non‐vascular imaging and interventional radiology procedures in Japan. Phys Med. 2020;74:83‐91. [DOI] [PubMed] [Google Scholar]
- 10. Pomeroy CL, Mason JB, Fehring TK, et al. Radiation exposure during fluoro‐assisted direct anterior total hip arthroplasty. J Arthroplasty. 2016;31:1742‐1745. [DOI] [PubMed] [Google Scholar]
- 11. Mori H, Koshida K, Ishigamori O, et al. A novel removable shield attached to C‐arm units against scattered X‐rays from a patient’s side. Eur Radiol. 2014;24:1794‐1799. [DOI] [PubMed] [Google Scholar]
- 12. IAEA 10 Pearls: radiation protection of staff in fluoroscopy. 2021 [cited 2021 Sep.1]. Available from: https://www.iaea.org/sites/default/files/documents/rpop/poster‐staff‐radiation‐protection.pdf
- 13. ICRP . Education and training in radiological protection for diagnostic and interventional procedures. ICRP Publication 113. Ann ICRP. 2009;39. [DOI] [PubMed] [Google Scholar]
- 14. Rehani MM, Ciraj‐Bjelac O, Vañó E, et al. Radiological protection in fluoroscopically guided procedures performed outside the imaging department. Ann ICRP. 2010;40:1‐102. [DOI] [PubMed] [Google Scholar]
- 15. Shin JM, Lee TH, Park S‐H, et al. A survey of the radiation exposure protection of health care providers during endoscopic retrograde cholangiopancreatography in Korea. Gut and Liver. 2013;7:100‐105. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 16. EURADOS EURADOS Report 2012‐02; 2012.
- 17. Vano E, Kleiman NJ, Duran A, et al. Radiation‐associated lens opacities in catheterization personnel: results of a survey and direct assessments. J Vasc Interv Radiol. 2013;24:197‐204. [DOI] [PubMed] [Google Scholar]
- 18. Sethi S, Barakat MT, Friedland S, et al. Radiation training, radiation protection, and fluoroscopy utilization practices among US therapeutic endoscopists. Dig Dis Sci. 2019;64:2455‐2466. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 19. Thornton RH, Dauer LT, Altamirano JP, et al. Comparing strategies for operator eye protection in the interventional radiology suite. J Vascular Intervent Radiol. 2010;21:1703‐1707. [DOI] [PubMed] [Google Scholar]
- 20. Koukorava C, Carinou E, Simantirakis G, et al. Doses to operators during interventional radiology procedures: focus on eye lens and extremity dosimetry. Radiat Prot Dosimetry. 2010;144:482‐486. [DOI] [PubMed] [Google Scholar]
- 21. Fetterly KA, Magnuson DJ, Tannahill GM, et al. Effective use of radiation shields to minimize operator dose during invasive cardiology procedures. JACC Cardiovasc Interv. 2011;4:1133‐1139. [DOI] [PubMed] [Google Scholar]
- 22. Jia Q, Chen Z, Jiang X, et al. Operator radiation and the efficacy of ceiling‐suspended lead screen shielding during coronary angiography: an anthropomorphic phantom study using real‐time dosimeters. Sci Rep. 2017;7:42077. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 23. Silva EH, Knežević Ž, Struelens L, et al. Energy and angular dependence of radiophotoluminescent glass dosemeters for eye lens dosimetry. Radiat Prot Dosimetry. 2016;170:208‐212. [DOI] [PubMed] [Google Scholar]
- 24. Silva EH, Struelens L, Covens P, et al. Optimization of a radiophotoluminescent glass dosemeter for occupational eye lens dosimetry in interventional radiology/cardiology. Radiat Prot Dosimetry. 2018;182:177‐183. [DOI] [PubMed] [Google Scholar]
- 25. Nagamoto K, Moritake T, Nakagami K, et al. Occupational radiation dose to the lens of the eye of medical staff who assist in diagnostic CT scans. Heliyon. 2021;7:e06063. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 26. Daures J, Gouriou J & Bordy JM. Conversion coefficients from air kerma to personal dose equivalent Hp(3) for eye‐lens dosimetry. CEA Saclay, 2009.
- 27. Morishima Y, Chida K, Meguro T. Effectiveness of additional lead shielding to protect staff from scattering radiation during endoscopic retrograde cholangiopancreatography procedures. J Radiat Res. 2018;59:225‐232. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 28. Minami T, Sasaki T, Serikawa M, et al. Occupational radiation exposure during endoscopic retrograde cholangiopancreatography and usefulness of radiation protective curtains. Gastroenterology Research and Practice. 2014;2014:1‐5. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 29. Koshi S. A basic framework of working environment control for occupational health in Japan. Ind Health. 1996;34:149‐165. [DOI] [PubMed] [Google Scholar]
Associated Data
This section collects any data citations, data availability statements, or supplementary materials included in this article.
Supplementary Materials
Data Availability Statement
On reasonable request, derived data supporting the findings of this study are available from the corresponding author.