Twelve lateral flow immunoassays (LFAs) to detect SARS-CoV-2 antibodies

Sophie I Owen a,1, Christopher T Williams a,1, Gala Garrod a, Alice J Fraser a, Stefanie Menzies a, Lisa Baldwin a, Lottie Brown a, Rachel L Byrne a, Andrea M Collins b,d, Ana I Cubas-Atienzar a, Margaretha de Vos c, Thomas Edwards a, Camille Escadafal c, Daniela M Ferreira d, Tom Fletcher b, Angela Hyder-Wright b,d, Grant A Kay a, Konstantina Kontogianni a, Jenifer Mason e, Elena Mitsi d, Tim Planche f,g, Jilian A Sacks c, Joseph Taylor e, Stacy Todd b, Caroline Tully c, Luis E Cuevas a, Emily R Adams a,
PMCID: PMC8664720  PMID: 34906597

Abstract

Background

There are an abundance of commercially available lateral flow assays (LFAs) that detect antibodies to SARS-CoV-2. Whilst these are usually evaluated by the manufacturer, externally performed diagnostic accuracy studies to assess performance are essential. Herein we present an evaluation of 12 LFAs.

Methods

Sera from 100 SARS-CoV-2 reverse-transcriptase polymerase chain reaction (RT-PCR) positive participants were recruited through the FASTER study. A total of 105 pre-pandemic sera from participants with other infections were included as negative samples.

Results

At presentation sensitivity against RT-PCR ranged from 37.4 to 79% for IgM/IgG, 30.3–74% for IgG, and 21.2–67% for IgM. Sensitivity for IgM/IgG improved ≥ 21 days post symptom onset for 10/12 tests. Specificity ranged from 74.3 to 99.1% for IgM/IgG, 82.9–100% for IgG, and 75.2–98% for IgM. Compared to the EuroImmun IgG enzyme-linked immunosorbent assay (ELISA), sensitivity and specificity ranged from 44.6 to 95.4% and 85.4–100%, respectively.

Conclusion

There are many LFAs available with varied sensitivity and specificity. Understanding the diagnostic accuracy of these tests will be vital as we come to rely more on the antibody status of a person moving forward, and as such manufacturer-independent evaluations are crucial.

Keywords: SARS-CoV-2, Lateral flow immunoassays, COVID-19, IgM, IgG

Introduction

As of May 2021, there have been over 100,000,000 confirmed cases of COVID-19 worldwide, however the total number of cases is much higher.1 This is due to, amongst other reasons, the lack of diagnostic testing worldwide in the first wave of the pandemic, the continued difficulties in testing in some lower-middle income countries and the number of asymptomatic infections that continue to go undetected.2,3 Detecting antibody responses to SARS-CoV-2 therefore could prove vital, both for understanding previous exposure on an individual level, but also at community and regional levels.

During SARS-CoV-2 infection, IgM and IgG titres begin to increase from around 10 days post-symptom onset with IgM titres generally declining earlier than IgG.4,5 The reference standard for detecting an antibody response, either IgM, IgG, or both, to SARS-CoV-2 is the immunoassay, usually either an enzyme-linked immunosorbent assay (ELISA) or chemiluminescence assay. Many immunoassays have been developed and commercialised for SARS-CoV-2 which are highly sensitive and specific.6 The process however is time-consuming, expensive, requires specialist laboratory equipment and trained professionals to perform.

To this end, large numbers of lateral-flow immunoassays (LFAs) have been developed that detect IgM and IgG responses. These are simple tests which require only a drop of blood and the addition of buffer to give a result in 10–15 min. Results are easy to interpret; if a test line appears the participant is considered positive, if no test line appears the participant is considered negative. LFAs are easy to mass-produce and are affordable making them ideal for mass-testing of populations, rapid identification of antibody response in travel situations, or for home-testing. Those that detect both IgM and IgG are also able to give an indication of when that person was likely infected, due to the dynamics of the antibody response.

Many LFAs are commercially available, and externally performed diagnostic accuracy studies, independent of the manufacturer, are required to generate robust performance data. Here, we present the evaluation of twelve LFAs and describe their diagnostic accuracy in a cohort of 100 confirmed SARS-CoV-2 positive participants with varying disease severity and 105 samples from participants confirmed as negative or collected pre-pandemic. This study forms part of a larger initiative to generate and share independent performance data on COVID-19 tests coordinated by FIND, the global diagnostics alliance.

Methods

Ethics statement

Research samples were provided with informed written consent. Participants were recruited through the Facilitating. A SARS CoV-2 TEst for Rapid triage (FASTER) study, approved by the National Health Service Research Ethics Committee (20/SC/0169) under the Integrated Research Application System no. 282,147.

Participants

A total of 142 serum/plasma samples from 125 SARS-CoV-2 RT-qPCR -positive participants were used in this study to assess the sensitivity of the LFAs. One hundred serum/plasma samples from RT-qPCR -positive participants were used for each LFA due to the limited quantity of some serum samples. Full details are given in Table S1. Briefly, 24 patients presenting at the Liverpool University Hospitals NHS Foundation Trust (Liverpool, UK) were recruited as part of the FASTER study and provided 41 serum samples collectively at different timepoints (D0, D2, D7, D28 post-admission). Sera from patients with RT-qPCR confirmed SARS-CoV-2 infection (n = 84) were provided by Liverpool Clinical Laboratories (LCL) as leftover diagnostic samples. Participants with RT-qPCR confirmed SARS-CoV-2 infection who did not result in hospital attendance (n = 12) were also recruited7. In addition, the NIBSC COVID-19 convalescent plasma panel, human (20/118), as well as NIBSC 20/130 plasma positive control were used for the evaluation. The COVID-19 convalescent plasma panel (NIBSC 20/118) and NIBSC 20/130 were obtained from the National Institute for Biological Standards and Control, UK.

A total of 105 SARS-CoV-2 negative serum samples were used to assess specificity of the LFAs. See Table S2 for full details. These consisted of 84 pre-pandemic sera collected from individuals diagnosed with influenza A (n = 20), tuberculosis (TB) (n = 10), human immunodeficiency virus (HIV) (n = 10), TB/HIV (n = 10), dengue virus (n = 10), parasitic diseases (n = 12), human coronavirus 229E (n = 10) and human coronavirus OC43 (n = 2). A panel of pre-pandemic plasma from participants with non-COVID-19-related fever (n = 20) were provided by FIND and an additional quality assurance sample (n = 1).

Enzyme-linked immunosorbent assay (ELISA) to detect IgG

The Anti-SARS-CoV-2 ELISA (IgG) kit (EI 2606–9601 G) (EuroImmun, Germany) was used to screen all serum samples for the presence of anti-SARS-CoV-2 IgG, as per the manufacturer's instructions. Samples with an OD value greater than the calibrator were considered positive, samples with OD value lower than the calibrator were considered negative.

Lateral flow immunoassays

Twelve LFAs (Table 1) were evaluated according to manufacturer's instructions. Briefly, 10–20 µl serum was required depending on the test, followed by 2–3 drops of buffer. Results were read independently by two people; if there was any disagreement a third person acted as a tiebreaker. Full details are given in Table 1. Of the 12 tests, 11 detected IgM and IgG separately, with only Beijing Wantai giving a ‘total antibody’ result. All tests were CE-IVD marked.

Table 1.

Details of LFAs evaluated.

Manufacturer Test name (manufacturer) Referred to herein as Product Code Lot Numbers Volume of sera (µl) Drops of buffer Time to result (minutes)
Beijing Wantai Biological Pharmacy Enterprise Co., Ltd WANTAI SARS-CoV-2 Ab Rapid Test Beijing Wantai WJ-2750 JNB20200408 10 2 15
Bionote Co., LTD. NowCheck COVID-19 IgM/IgG Test Bionote RB2901DG 2901D002 10 3 10
Core Technology Co., Ltd COVID-19 IgM/IgG Ab Test Core Technology B290–21 20,200,406 10 2 10
CTK Biotech Onsite COVID-19 IgM/IgG Rapid Test CTK Biotech R0180C F0507R1C00 10 2 10
Edinburgh Genetics Limited Edinburgh Genetics COVID-19 Colloidal Gold Immunoassay Testing Kit, IgM/IgG Combined Edinburgh Genetics TIL225AEGCV0055 2000555A 20 into 2 ml buffer 2–3 10
GenBody Inc. COVID-19 IgM/IgG GenBody COVI040, PQGB021 (reader) FJF029201 10 3 10
Jiangsu Bioperfectus Technologies Co., Ltd Novel Corona Virus(SARS-CoV-2)IgM/IgG Rapid Test Kit Jiangsu Bioperfectus SC30201W 20,200,401 10 3 10
PRIME4DIA Co., Ltd P4DETECT COVID-19 IgM/IgG PRIME4DIA CMG200701 10 3 10
Qingdao HIGHTOP Biotech Co., Ltd. SARS-CoV-2 IgM/IgG Ab Rapid Test Qingdao HIGHTOP H100 COV1252004C 10 2 15
Shanghai Kehua Bio-Engineering Co., Ltd Diagnostic Kit for SARS-CoV-2 IgM/IgG Antibody (Colloidal Gold) Shanghai Kehua R-423–20-C-CE 423,200,334 10 3 15
Shenzhen Bioeasy Biotechnology Co., Ltd 2019-Novel Coronavirus (2019-nCoV) IgM/IgG GICA Rapid Test Kit Shenzhen Bioeasy YRLG22301025 2003N104 10 2 10
Zhuhai Livzon Diagnostics Inc. Diagnostic Kit for IgM/IgG Antibody to Coronavirus (SARS-CoV-2) (Lateral Flow) Zhuhai Livzon CK2004240410 10 2 10

Data analysis

Sensitivity was calculated against RT-qPCR confirmed SARS-CoV-2 infections including sensitivity when stratified by days post-symptom onset. Specificity was calculated against RT-qPCR confirmed SARS-CoV-2 negative samples or samples collected pre-pandemic. Sensitivity was then calculated against RT-qPCR confirmed SARS-CoV-2 infections also positive by IgG ELISA. Specificity was calculated against IgG ELISA negative samples. Percentage agreement and Cohen's Kappa statistic against IgG ELISA were calculated.8 Data analyses were carried out in MedCalc for Windows, version 19.8 (MedCalc Software, Ostend, Belgium).

Results

Sensitivity and specificity against RT-qPCR

Sensitivity of the LFAs against RT-qPCR ranged from 37.4 to 79.0% for IgM/IgG, 30.3–74.0% for IgG only, and 21.2–67.0% for IgM only (Table 2, Fig. 1). The sensitivity for an IgM/IgG response increased in 10 out of 12 tests at > 21 days post-symptom onset, with a mean increase of 15.0% (Table 2, Fig. 1). The sensitivity for IgG increased > 21 days post-symptom onset, with a mean increase of 16.5% (Table 2, Fig. 1). For IgM, sensitivity was higher ≤ 21 days post-symptom onset in six of the LFAs and higher at > 21 days post-symptom onset for five tests (Table 2, Fig. 1).

Table 2.

Sensitivity and specificity of the 12 LFAs. Sensitivity was calculated using SARS-CoV-2 RT-qPCR positive sera/plasma and specificity determined on pre-pandemic sera/plasma.

Test Ig All samples ≤ 21 DAYS POST SYMPTOM ONSET Sensitivity vs. RT-qPCR (%) > 21 DAYS POST SYMPTOM ONSET Sensitivity vs RT-qPCR (%) More sensitive > 21 days?
Sensitivity vs RT-qPCR (%) [95% CI] Specificity vs pre-Pandemic panel (%) [95% CI]
Beijing Wantai IgG + IgM 69.7 (69/99 TP) [59.7–78.5] 99.1 (104/105 TN) [94.8–100] 66.7 (38/57) 71.1 (27/38) Yes
Bionote IgG + IgM 79.0 (79/100 TP) [69.7–86.5] 97.0 (97/100 TN) [91.5–99.4] 75.4 (46/61) 88.2 (30/34) Yes
IgG 65.0 (65/100 TP) [54.8–74.3] 100 (100/100 TN) [96.4–100] 52.5 (32/61) 88.2 (30/34) Yes
IgM 63.0 (63/100 TP) [52.8–72.4] 97.0 (97/100 TN) [91.5–99.4] 67.2 (41/61) 61.8 (21/34) No
Core Technology IgG + IgM 70.0 (70/100 TP) [60.0–78.8] 96.2 (101/105 TN) [90.5–99.0] 63.2 (36/57) 79.0 (30/38) Yes
IgG 67.0 (67/100 TP) [56.9–76.1] 100 (105/105 TN) [96.6–100] 59.7 (34/57) 76.3 (29/38) Yes
IgM 60.0 (60/100 TP) [49.7–69.7] 96.2 (101/105 TN) [90.5–99.0] 61.4 (35/57) 60.5 (23/38) No
CTK Biotech IgG + IgM 70.0 (70/100 TP) [60.0–78.8] 86.7 (85/98 TN) [78.4–92.7] 75.4 (46/61) 64.7 (22/34) No
IgG 51.0 (51/100 TP) [40.8–61.1] 99.0 (97/98 TN) [94.5–100] 49.2 (30/61) 55.9 (19/34) Yes
IgM 67.0 (67/100 TP) [56.9–76.1] 87.8 (86/98 TN) [79.6–93.5] 72.1 (44/61) 61.8 (21/34) No
Edinburgh Genetics IgG + IgM 58.0 (58/100 TP) [47.7–67.8] 87.6 (85/97 TN) [79.4–93.4] 50.8 (31/61) 73.5 (25/34) Yes
IgG 56.0 (56/100 TP) [45.7–65.9] 99.0 96/97 TN) [94.4–100] 49.2 (30/61) 70.6 (24/34) Yes
IgM 27.0 (27/100 TP) [18.6–36.8] 88.7 (86/97 TN) [80.6–94.2] 32.8 (20/61) 20.6 (7/34) No
GenBody Inc. IgG + IgM 37.4 (37/99 TP) [27.9–47.7] 92.4 (97/105 TN) [85.5–96.7] 42.1 (24/57) 29.0 (11/38) No
IgG 30.3 (30/99 TP) [21.5–40.4] 96.2 (101/105 TN) [90.5–99.0] 33.3 (19/57) 23.7 (9/38) No
IgM 21.2 (21/99 TP) [13.6–30.6] 94.3 (99/105 TN) [88.0–97.9] 29.8 (17/57) 10.5 (4/38) No
Jiangsu Bioperfectus IgG + IgM 72.0 (72/100 TP) [62.1–80.5] 88.6 (93/105 TN) [80.9–94.0] 66.7 (38/57) 79.0 (30/38) Yes
IgG 69.0 (69/100 TP) [59.0–77.9] 95.2 (100/105 TN) [89.2–98.4] 61.4 (35/57) 79.0 (30/38) Yes
IgM 61.0 (61/100 TP) [50.7–70.6] 91.4 (96/105 TN) [84.4–96.0] 59.7 (34/57) 63.2 (24/38) Yes
PRIME4DIA IgG + IgM 67.0 (67/100 TP) [56.9–76.1] 97.0 (98/101 TN) [91.6–99.4] 63.9 (39/61) 76.5 (26/34) Yes
IgG 56.0 (56/100 TP) [45.7–65.9] 100 (101/101 TN) [96.4–100] 52.5 (32/61) 64.7 (22/34) Yes
IgM 60.0 (60/100 TP) [49.7–69.7] 97.0 (98/101 TN) [91.6–99.4] 60.7 (37/61) 64.7 (22/34) Yes
Qingdao HIGHTOP IgG + IgM 71.0 (71/100 TP) [61.1–79.6] 96.0 (97/101 TN) [90.2–98.9] 63.9 (39/61) 85.3 (29/34) Yes
IgG 60.0 (60/100 TP) [49.7–69.7] 98.0 (99/101 TN) [93.0–99.8] 60.7 (37/61) 64.7 (22/34) Yes
IgM 67.0 (67/100 TP) [56.9–76.1] 98.0 (99/101 TN) [93.0–99.8] 57.4 (35/61) 85.3 (29/34) Yes
Shanghai Kehua IgG + IgM 78.0 (78/100 TP) [68.6–85.7] 74.3 (78/105 TN) [64.8–82.3] 66.7 (38/57) 94.7 (36/38) Yes
IgG 74.0 (74/100 TP) [64.3–82.3] 93.3 (98/105 TN) [86.8–97.3] 64.9 (37/57) 86.8 (33/38) Yes
IgM 55.0 (55/100 TP) [44.7–65.0] 75.2 (79/105 TN) [65.9–83.1] 57.9 (33/57) 52.6 (20/38) No
Shenzhen Bioeasy IgG + IgM 70.0 (70/100 TP) [60.0–78.8] 82.9 (87/105 TN) [74.3–89.5] 64.9 (37/57) 79.0 (30/38) Yes
IgG 65.0 (65/100 TP) [54.8–74.3] 82.9 (87/105 TN) [74.3–89.5] 59.7 (34/57) 76.3 (29/38) Yes
IgM 56.0 (56/100 TP) [45.7–65.9] 83.8 (88/105 TN) [75.4–90.3] 56.1 (32/57) 57.9 (22/38) Yes
Zhuhai Livzon IgG + IgM 70.0 (70/100 TP) [60.0–78.8] 86.0 (86/100 TN) [77.6–92.1] 70.5 (43/61) 76.5 (26/34) Yes
IgG 52.0 (52/100 TP) [41.8–62.1] 100 (100/100 TN) [96.4–100] 52.5 (32/61) 64.7 (22/34) Yes
IgM 66.0 (66/100 TP) [55.9–75.2] 86.0 (86/100 TN) [77.6–92.1] 60.7 (37/61) 64.7 (23/34) Yes

Fig. 1.

Fig. 1

Sensitivity (total, ≤ 21 and > 21 days post symptom onset) and specificity of the antibody response (IgM/IgG, IgG, IgM) to each LFA. Note, sensitivity was calculated against RT-PCR results and specificity was calculated using a pre-pandemic panel.

Specificity ranged from 74.3 to 99.1% for IgM/IgG, 82.9–100% for IgG only, and 75.2–98.0% for IgM only (Table 2, Fig. 1). For all tests, except for Beijing Wantai which is a combined IgM/IgG only, and Shenzhen Bioeasy where specificity for IgM/IgG and IgG were identical, IgG alone had a higher specificity than IgM/IgG (Table 2, Fig. 1). Four tests reported 100% specificity for IgG (Table 2). Nine LFAs had a higher IgG specificity than IgM (Table 2, Fig. 1).

Diagnostic accuracy of LFAs against IgG ELISA

Of the 142 participants positive by RT-qPCR, 90 (63.4%) were positive by IgG ELISA. Of the 105 pre-pandemic samples, 85 were tested by IgG ELISA due to sample availability, of which 82 (96.5%) were negative by IgG ELISA. Compared to IgG ELISA, LFA sensitivity was found to range between 44.6% and 95.4% (Table 3). Shanghai Kehua had the highest sensitivity at 95.4% (Table 3). Compared to IgG ELISA, LFA specificity was found to range between 85.4% and 100% (Table 3). The highest agreement between IgG measured by LFA and IgG measured by ELISA was seen with Shanghai Kehua (Table 4).

Table 3.

Sensitivity and specificity of LFAs against IgG ELISA.

Manufacturer Sensitivity (%) [95% CI] Specificity (%) [95% CI]
Beijing Wantai 92.3 (60/65 TP) [83.0–97.5] 98.8 (81/82 TN) [93.4–100]
Bionote 90.0 (54/60 TP) [79.5–96.2] 100 (78/78 TN) [95.4–100]
Core Technology 92.3 (60/65 TP) [83.0–97.5] 100 (82/82 TN) [95.6–100]
CTK Biotech 81.7 (49/60 TP) [69.6–90.5] 100 (76/76 TN) [95.3–100]
Edinburgh Genetics 80.0 (48/60 TP) [67.7–89.2] 100 (75/75 TN) [95.2–100]
GenBody 44.6 (29/65 TP) [32.3–57.5] 100 (76/76 TN) [95.6–100]
Jiangsu Bioperfectus 93.9 (61/65 TP) [85.0–98.3] 96.3 (79/82 TN) [89.7–99.2]
PRIME4DIA 86.7 (52/60 TP) [75.4–94.1] 100 (79/79 TN) [95.4–100]
Qingdao HIGHTOP 85.0 (51/60 TP) [73.4–92.9] 98.7 (78/79 TN) [93.2–100]
Shanghai Kehua 95.4 (62/65 TP) [87.1–99.0] 97.6 (80/82 TN) [91.5–99.7]
Shenzhen Bioeasy 83.1 (54/65 TP) [71.7–91.2] 85.4 (70/82 TN) [75.8–92.2]
Zhuhai Livzon 78.3 (47/60 TP) [65.8–87.9] 100 (78/78 TN) [95.4–100]

Table 4.

Agreement and Cohen's Kappa between IgG measured by LFA and IgG ELISA.

LFA ELISA Kappa [95% CI]
Positive Negative
Beijing Wantai Positive 60 1 0.9167 [0.8514–0.982]
Negative 5 81
Bionote Positive 54 0 0.9105 [0.8405–0.9805]
Negative 6 78
Core Technology Positive 60 0 0.9305 [0.8706–0.9904]
Negative 5 82
CTK Biotech Positive 49 0 0.8327 [0.7379–0.9275]
Negative 11 76
Edinburgh Genetics Positive 48 0 0.8163 [0.7171–0.9155]
Negative 12 75
GenBody Positive 29 0 0.4733 [0.3238–0.6228]
Negative 36 82
Jiangsu Bioperfectus Positive 61 3 0.9033 [0.8334–0.9732]
Negative 4 79
PRIME4DIA Positive 52 0 0.8808 [0.8006–0.961]
Negative 8 79
Qingdao HIGHTOP Positive 51 1 0.8510 [0.7620–0.9400]
Negative 9 78
Shanghai Kehua Positive 62 2 0.9309 [0.8714–0.9904]
Negative 3 80
Shenzhen Bioeasy Positive 54 12 0.6833 [0.5644–0.8022]
Negative 11 70
Zhuhai Livzon Positive 47 0 0.8034 [0.7017–0.9051]
Negative 13 78

Discussion

There is a plethora of LFAs available on the market today, all purporting to offer high sensitivity and specificity, but often without rigorous, manufacturer-independent evaluations. In this study, we evaluated 12 LFAs on serum samples collected from RT-qPCR -positive individuals and individuals with a wide range of diagnosed diseases pre-pandemic. We demonstrate the differences in sensitivity and specificity of the responses of combined IgM/IgG, IgG and IgM against a RT-qPCR and an IgG ELISA, in patients presenting with both acute and convalescent SARS-CoV-2 infections.

Bionote had the highest overall sensitivity (79.0% [95% CI: 69.7–86.5]), with a sensitivity of 88.2% at > 21 days post-symptom onset for an IgM/IgG response. Genbody Inc. had the lowest sensitivity with an overall sensitivity of 37.4% for an IgM/IgG response. Sensitivity of IgM/IgG and IgG improved for the majority of LFAs over 21 days post-symptom onset in agreement with other LFA evaluation studies.9 In this study, no test met the clinical sensitivity requirements of > 98% (95% CI: 96–100%) on samples collected ≥ 20 days post-symptom onset laid out in the target product profile (TPP) published by the UK government.10 However, our data are calculated on fewer than 200 confirmed positive cases as specified by the TPP.10 As expected with an earlier decline in IgM titres, fewer LFAs had improved sensitivity for IgM > 21 days post-symptom onset. It is important to note that the samples used in this study were collected before the roll out of any COVID-19 vaccine.

The large variation in performance in LFAs reported here is in accordance with other evaluations.9 The variations in diagnostic accuracy may in part be due to the antigen used to detect SARS-CoV-2 antibodies. The two main immunogenic antigens of SARS-CoV-2 are the nucleocapsid and the surface spike protein, split into domains S1 and S2, with the receptor-binding domain in S1. S1 is thought to be the most specific, with low-level cross-reactivity demonstrated for S2 and nucleocapsid.11 It is one or a combination of these antigens that are used for serological testing.

Not all manufacturers included in this evaluation disclose the antigen(s) used in their test. This information is key to testing during vaccine-rollout, with two vaccines, Pfizer-BioNTech and Moderna, containing mRNA encoding spike proteins to elicit an anti-spike immune response. LFAs that detect a response to spike antigens should prove useful for detecting both prior exposure to SARS-CoV-2, but also vaccinated individuals. Those that do not detect the relevant spike antigen may not prove as useful in detecting an immune response within vaccinated individuals but may have a role in identifying immune responses to breakout infections in vaccinated populations. Further studies should look at the use of LFAs in vaccinated individuals.

Our pre-pandemic negative panel consisted of serum from individuals diagnosed with a wide range of diseases, and in general, the false-positive results were found not to be linked to one disease. However, one sample from a returning traveler with malaria in 2005 gave a false positive result for 7 out of the 12 LFAs, which warrants further study. Of the 20 Influenza A samples, only 3 false positives were reported across all 12 LFAs. One sample with previous human coronavirus 229E exposure gave a false positive result for 4/12 LFAs as well as the EuroImmun IgG ELISA, and another gave a false positive result in 3/12 LFAs. Seven out of the twelve human coronavirus 229E and OC43 samples reported no false positives for any LFAs, whist 1 sample reported 2 false positives and the remaining 2 samples reported 1 false positive. This suggests cross-reactivity with immune responses to other human coronaviruses is possible but likely to be minimal, additionally pre-pandemic coronavirus patients often presented with other syndromes and coronavirus diagnosis has only been made due to multiplex molecular panels; this indicates samples used here may have unusual properties which may initiate cross-reaction as opposed to the coronaviruses present.

There is still no established ‘gold-standard’ serological test for SARS-CoV-2. The EuroImmun IgG ELISA was chosen at the time as it was one of the only CE-marked ELISA assays. We therefore use RT-qPCR as the main reference test in this study. Previous studies have reported false positives with the EuroImmun IgG ELISA, and we report here 3/105; one from a HIV patient in Nigeria in 2018, one from a patient with dengue virus in Brazil in 2015 and one with an individual with human coronavirus 229E in the UK in 2019. These were excluded from the ELISA/LFA specificity analysis. Cross-reactivity was seen in an evaluation of the EuroImmun Anti-SARS-CoV-2 NCP ELISA carried out in Nigeria in a negative control panel in which 50.2% of participants had the P. falciparum HRP2 antigen.12 The specificity of the EuroImmun ELISA in this study was 96.5%.

LFAs can be helpful in measuring exposure of a community to SARS-CoV-2, particularly in areas where testing of symptomatic individuals was not and is not readily available. Manufacturer-independent evaluations provide helpful data as to the accuracy of LFAs, as there is a large variation in the performance characteristics of these assays. Further evaluations are needed following the commencement of vaccination campaigns to evaluate the use of LFAs in vaccinated individuals.

Declaration of Competing Interest

Emily R. Adams is Director of Epidemics and NTDs at Mologic.

Acknowledgments

Sources of funding

This work is supported by grants from DFID/Wellcome Trust Epidemic Preparedness coronavirus grant (220764/Z/20/Z) to ERA and LEC, the National Institute for Health Research (NIHR; award CO-CIN-01), the NIHR Health Protection Research Unit (HPRU) in Emerging and Zoonotic Infections at University of Liverpool in partnership with Public Health England (PHE), in collaboration with Liverpool School of Tropical Medicine and the University of Oxford (award 200907), Pfizer (grant WI255862) to DMF, AHW, EM, and AMC, and MRC-DTP (MR/N013514/1) to SIO. This work was also funded as part of FIND's work as co-convener of the diagnostics pillar of the Access to COVID-19 Tools (ACT) Accelerator, including support from Unitaid [Grant No. 2019–32-FIND MDR], from UK Department for International Development [Grant number 300341–102], and the World Health Organization.

Acknowledgements

We would like to thank all those who participated in the FASTER study for their involvement in this research. We would also like to thank the teams from Liverpool University Hospitals National Health Services Foundation Trust, National Institute for Health Research and the Liverpool School of Tropical Medicine team for their assistance with the sample collection and processing: Kostas Liatsikos, Farah Shiham, Jesus Reine, Carla Solorzano Gonzalez, Elissavet Nikolaou, Ashleigh Howard, Lisa Hitchins, Katerina Cheliotis, Esther German, Sherin Pojar, Christopher Myerscough, Madi Farrar, Sharon Glynn.

References

  • 1.World Health Organization. World Health Organization Coronavirus (COVID-19) Dashboard [Internet]. 2021 [cited 2021 Mar 22]. Available from: https://covid19.who.int/.
  • 2.Giri A.K., Rana D.R.S.J.B. Charting the challenges behind the testing of COVID-19 in developing countries: Nepal as a case study. Biosaf Health. 2020;2:53–56. doi: 10.1016/j.bsheal.2020.05.002. [Internet][cited 2021 Mar 23] [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 3.Gao Z., Xu Y., Sun C., Wang X., Guo Y., Qiu S., et al. A systematic review of asymptomatic infections with COVID-19. J Microbiol Immunol Infect. 2020;54:12–16. doi: 10.1016/j.jmii.2020.05.001. Elsevier Ltd. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 4.To K.K.W., Tsang O.T.Y., Leung W.S., Tam A.R., Wu T.C., Lung D.C., et al. Temporal profiles of viral load in posterior oropharyngeal saliva samples and serum antibody responses during infection by SARS-CoV-2: an observational cohort study. Lancet Infect Dis. 2020;20(5):565–574. doi: 10.1016/S1473-3099(20)30196-1. [Internet]May 1 [cited 2021 Mar 23]Available from: www.thelancet.com/infection. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 5.Ma H., Zeng W., He H., Zhao D., Jiang D., Zhou P., et al. Serum IgA, IgM, and IgG responses in COVID-19. Cell Mol Immunol. 2020;17:773–775. doi: 10.1038/s41423-020-0474-z. [Internet][cited 2020 Oct 20] [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 6.Pickering S., Betancor G., Galão R.P., Merrick B., Signell A.W., Wilson H.D., et al. Comparative assessment of multiple COVID-19 serological technologies supports continued evaluation of point-of-care lateral flow assays in hospital and community healthcare settings. PLOS Pathog. 2020;16(9) doi: 10.1371/journal.ppat.1008817. Fouchier RAM, Editor. [Internet]Sep 24 [cited 2021 Mar 23] [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 7.Brown L., Byrne R.L., Fraser A., Owen S.I., Cubas Atienzar A.I., Williams C., et al. Self-sampling of capillary blood for serological testing of SARS-CoV-2 by COVID-19 IgG ELISA. medRxiv [Internet]. 2020 Jan 1;2020.09.25.20183459. Available from: http://medrxiv.org/content/early/2020/09/27/2020.09.25.20183459.abstract
  • 8.Landis J.R., Koch G.G. The measurement of observer agreement for categorical data. Biometrics. 1977;33(1):159. Mar. [PubMed] [Google Scholar]
  • 9.Whitman J.D., Hiatt J., Mowery C.T., Shy B.R., Yu R., Yamamoto T.N., et al. Evaluation of SARS-CoV-2 serology assays reveals a range of test performance. Nat Biotechnol. 2020;38(10):1174–1183. doi: 10.1038/s41587-020-0659-0. [Internet]Oct 1 [cited 2021 Apr 26] [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 10.Target Product Profile . Government UK; 2021. Antibody Tests to Help Determine if People have Recent Infection to SARS-CoV-2: Version 2. [Internet][citedMay 26]Available from: https://www.gov.uk/government/publications/how-tests-and-testing-kits-for-coronavirus-covid-19-work/target-product-profile-antibody-tests-to-help-determine-if-people-have-recent-infection-to-sars-cov-2-version-2. [Google Scholar]
  • 11.Li D., Li J. Immunologic testing for SARS-CoV-2 infection from the antigen perspective. J Clin Microbiol. 2020;59(5) doi: 10.1128/JCM.02160-20. [Internet]Dec 14 [cited 2021 Apr 26] [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 12.Steinhardt L.C., Ige F., Iriemenam N.C., Greby S.M., Hamada Y., Uwandu M., et al. Cross-reactivity of two SARS-CoV-2 serological assays in a malaria-endemic setting. J Clin Microbiol. 2021 doi: 10.1128/JCM.00514-21. [Internet]Apr 14 [cited 2021 Apr 27] [DOI] [PMC free article] [PubMed] [Google Scholar]

RESOURCES