Corneal surgically induced astigmatism in resident surgeons

Timothy Sipos a,b,, Kendall Bicknell b
PMCID: PMC8682862  PMID: 34970027

Abstract

To evaluate surgically induced astigmatism (SIA) in resident surgeons in their first year of performing cataract surgery, a retrospective study was conducted involving three resident physicians from July 1, 2019, to June 30, 2020. Preoperative and postoperative corneal measurements were taken with the IOLMaster 500 and Pentacam. Mean SIA and centroid were determined with the Koch-Wang Excel spreadsheet. A total of 135 eyes were operated on, with Resident A performing 67 surgeries; Resident B, 60 surgeries; and Resident C, 8 surgeries. Resident A’s IOLMaster results revealed centroid values of 0.08 D @ 91° ± 0.37 diopters (D) and 0.16D @ 82° ± 0.32D and Pentacam centroid values of 0.13D @ 100° ± 0.34D and 0.24D @ 93° ± 0.38D for right and left eyes, respectively. Resident B’s IOLMaster results revealed centroid values of 0.23D @ 102° ± 0.23D and 0.29D @ 110° ± 0.26D and Pentacam centroid values of 0.21D @ 124° ± 0.33D and 0.17D @ 103° ± 0.51D for right and left eyes, respectively. In conclusion, centroid values of SIA with junior surgeons were more than the proposed established values, but likely not clinically significant, with at most a difference of <0.17D in the spectacle plane. To achieve more accurate refractive outcomes in toric intraocular lenses, all residents should calculate their personal SIA early in their surgical career.

Keywords: Cataract surgery, centroid, surgically induced astigmatism


Cataract surgery is the most common procedure performed, with over 3.6 million surgeries performed in 2015 alone.1 With improvements in intraocular lens (IOL) calculations and newer-generation formulas, 94% of patients achieve best corrected visual acuity of 20/30 or better, and 72% to 80% of patients end up within 0.5 diopters (D) of emmetropia.2,3 Surgically induced astigmatism (SIA) represents the amount of astigmatism created by the main wound construction. Several factors contribute to the amount of SIA created, such as length, width, location, and structure, with the width of the incision being the most influential.4–6 Abulafia et al found that the centroid value, the mean SIA magnitude and direction, can produce an SIA of around 0.1D as opposed to the traditional 0.35D.7 Junior surgeons typically lack the consistency and reproducibility of surgical approaches and wound outcomes compared to experienced surgeons. We were not aware of studies showing SIA outcomes by surgeons in training. This study assessed the amount of SIA of junior surgeons during early stages of training.

METHODS

This retrospective case series comprised a total of 135 eyes. All patients were operated on between July 1, 2019, and June 30, 2020, by second-year (postgraduate year 3) residents at the Central Texas Veterans Affairs Medical Center (VA). The study protocol followed the guidelines of the Declaration of Helsinki as well as the VA institutional review board. No consents were obtained, as the retrospective chart review posed minimal risk to patients and obtaining informed consent was not feasible, as many of these patients were no longer followed at the Temple VA. Preoperative keratometry and axis measurements were taken with the Zeiss IOLMaster 500 and the Oculus Pentacam. All surgeries were performed using 2.4 mm clear cornea incisions at a temporal position. Sutures were placed through the main wounds only in cases of wound leaks after adequate corneal hydration. In the instance of suture placement during cataract surgery, postoperative measurements were taken at least 3 weeks after suture removal; 4 to 6 weeks following cataract surgery, postoperative keratometry and axis measurements were repeated with IOLMaster 500 and Pentacam. The three residents were designated by the order (A, B, and C) in which they rotated at the VA Medical Center. Patient demographics were obtained, including the age at the time of surgery and gender.

Centroid and mean SIA were calculated using the Koch-Wang Excel spreadsheet available on the American Society Cataract and Refractive Surgery website.8 Data included the preoperative flat K, flat axis, steep K, and steep axis and postoperative flat K, flat axis, steep K, and steep axis. The data were split between right and left eyes. The results from the IOLMaster 500 and Pentacam were compared for any significant differences using a one-tailed unpaired t test. An unpaired t test was used since not every patient was able to obtain both forms of keratometry measurements.

Patients were excluded from the study if they were lost to follow-up or if a postoperative keratometry measurement was not performed. Other exclusion criteria included unreliable preoperative or postoperative keratometry measurements due, for example, to corneal scars or eyelid obstruction. Patients with any significant complications, such as endophthalmitis or ruptured posterior capsule, were also excluded.

RESULTS

From July 2019 through June 2020, Resident A performed 67 cataract surgeries, Resident B performed 60 cataract surgeries, and Resident C performed 8 cataract surgeries. Resident A operated on 36 right eyes (53.7%) and 31 left eyes (46.3%); Resident B operated on 28 right eyes (46.7%) and 32 left eyes (53.3%); and Resident C operated on 5 right eyes (62.5%) and 3 left eyes (37.5%) (Table 1).

Table 1.

Patient demographics for Residents A, B, and C

Resident Average age Male Female Right eyes Left eyes
Resident A 75 62 5 34 33
Resident B 74 57 3 28 32
Resident C 69 8 0 5 3

Of the 67 cataract surgeries performed by Resident A, 58 (33 right eye, 25 left eye) postoperative IOLMaster corneal measurements were taken. Figure 1 shows the double angle plot of the centroid values. Besides the 8 patients (8 eyes) lost to follow-up and the one patient unable to obtain IOLMaster measurements, no other patients were excluded from the Resident A data set. No significant complications were reported by Resident A.

Figure 1.

Figure 1.

Resident A double angle plot of centroid values and mean surgically induced astigmatism for both eyes from the IOLMaster 500.

Of the 60 cataract surgeries performed by Resident B, 27 (10 right eye, 17 left eye) postoperative IOLMaster corneal measurements were taken. Figure 2 shows the double angle plot of the centroid values. Eighteen patients (24 eyes) were lost to follow-up, with another 5 patients (9 eyes) unable to obtain IOLMaster measurements due to the IOLMaster undergoing repairs. One patient was excluded for wearing contact lenses during preoperative measurements. No significant complications were reported by Resident B.

Figure 2.

Figure 2.

Resident B double-angle plot of centroid values and mean surgically induced astigmatism for both eyes from the IOLMaster 500.

Due to COVID-19 elective surgery shutdowns at the VA, Resident C operated on 8 patients, with none returning for postoperative measurements. All 8 were thus excluded, and the results from Resident C were not included in the study. There were no significant complications reported by Resident C.

Table 2 shows the overall results of the study: IOLMaster centroid means ranged from 0.08D to 0.29D; IOLMaster mean SIA, from 0.23D to 0.37D; Pentacam centroid, from 0.13D to 0.24D; and Pentacam mean SIA, from 0.27D to 0.42D. The unpaired t test showed no significant difference between the IOLMaster centroid results and the Pentacam centroid results for both right eyes and left eyes, and no significant difference between mean SIA results for both right and left eyes.

Table 2.

SIA values calculated from the IOLMaster and Pentacam for residents

Resident Eyes IOLMaster
Pentacam
P value
Centroid Mean SIA Centroid Mean SIA Centroid Mean
A Right 0.08 @ 91° ± 0.37D 0.30D ± 0.17D 0.13D @ 100° ± 0.34D 0.27D ± 0.17D 0.57 0.48
  Left 0.16D @ 82° ± 0.32D 0.32D ± 0.16D 0.24D @ 93° ± 0.38D 0.41D ± 0.19D 0.42 0.07
B Right 0.23D @ 102° ± 0.23D 0.23D ± 0.07D 0.21D @ 124° ± 0.33D 0.30D ± 0.11D 0.87 0.09
  Left 0.29D @ 110° ± 0.26D 0.37D ± 0.10D 0.17D @ 103° ± 0.51D 0.42D ± 0.24D 0.38 0.43

SIA indicates surgically induced astigmatism.

DISCUSSION

As cataract surgery has become technologically more advanced, the correction of astigmatism has become ever more important. Factors influencing the correction of astigmatism include but are not limited to SIA, preoperative keratometry measurements, lens rotation, and the accepted range of manufacturer error of intraocular lenses. As beginning surgeons learn the intricacies of operating and wound creation, it is expected that there is less reproducibility in clear corneal incisions than that of experienced surgeons. As reported by Abulafia et al, the centroid value of 0.1D should be used instead of the traditional mean SIA of 0.35D.7 Significant ocular surface disease may result in >0.5D refractive surprise in up to 10% of patients.9 Toric IOLs may rotate after implantation, with most rotating <5° postoperatively.10 With every degree of rotation displaced, the toric lens loses 3.3% of its astigmatism-correcting power.11 Manufacturer lens errors may also provide a source of unexpected refractive outcomes. According to the international standard, spherical power may range 0.4D for lenses with power between 15 and 25D.12

This study provided interesting results, with both Resident A’s and Resident B’s mean SIA values falling in the accepted range of 0.30 to 0.50D.7,13 The results of Resident A and B were not markedly larger than the proposed centroid value of 0.12D suggested by Warren Hill.14 The results showed no statistical difference between the IOLMaster and Pentacam for both residents in terms of mean SIA and centroid values.

There are several limitations of this study, including the high rate of failure to follow-up. COVID-19 occurred during the study and caused 13 patients (18 eyes) of the 60 surgeries performed by Resident B and all of Resident C’s patients to not follow-up. It was assumed that all temporal incisions were performed at the 180° axis and the 0° axis for right and left eyes, respectively. Another limitation of this study is the lack of standard to compare the residents’ mean SIA and centroid values to that of an experienced surgeon as a control.

No study to date has looked at SIA values performed by residents. Although the resident centroid values were either near or slightly above what other studies have shown for experienced surgeons, it is still felt that residents should continue using the proposed centroid value of 0.12D. With other areas of potential error, such as poor preoperative keratometry measurements, lens rotation, and the accepted range of manufacturer error of intraocular lenses, it is felt that SIA does not pose as much of a risk for refractive surprise following cataract surgery. Residents have been shown to have successful outcomes following toric IOL implantation, with one study out of the University of Texas Southwestern showing that resident toric IOL cases result in a mean refractive cylinder of 0.38D.15 All residents are encouraged to calculate their own personal SIA for the most accurate refractive outcomes.

References

  • 1.Lindstrom R. Thoughts on cataract surgery. Review of Ophthalmology. 2015. https://www.reviewofophthalmology.com/article/thoughts-on--cataract-surgery-2015. Accessed September 10, 2021.
  • 2.Ewe SY, Abell RG, Oakley CL, et al. A comparative cohort study of visual outcomes in femtosecond laser-assisted versus phacoemulsification cataract surgery. Ophthalmology. 2016;123(1):178–182. doi: 10.1016/j.ophtha.2015.09.026. [DOI] [PubMed] [Google Scholar]
  • 3.Melles RB, Holladay JT, Chang WJ.. Accuracy of intraocular lens calculation formulas. Ophthalmology. 2018;125(2):169–178. doi: 10.1016/j.ophtha.2017.08.027. [DOI] [PubMed] [Google Scholar]
  • 4.Ermis SS, Inan UU, Ozturk F.. Surgically induced astigmatism after superotemporal and superonasal clear corneal incisions in phacoemulsification. J Cataract Refract Surg. 2004;30(6):1316–1319. [DOI] [PubMed] [Google Scholar]
  • 5.Dewey S, Beiko G, Braga-Mele R, et al. Microincisions in cataract surgery. J Cataract Refract Surg. 2014;40(9):1549–1557. doi: 10.1016/j.jcrs.2014.07.006. [DOI] [PubMed] [Google Scholar]
  • 6.Fukuda S, Kawana K, Yasuno Y, Oshika T.. Wound architecture of clear corneal incision with or without stromal hydration observed with 3-dimensional optical coherence tomography. Am J Ophthalmol. 2011;151(3):413–419.e1. doi: 10.1016/j.ajo.2010.09.010. [DOI] [PubMed] [Google Scholar]
  • 7.Abulafia A, Barrett GD, Kleinmann G, et al. Prediction of refractive outcomes with toric intraocular lens implantation. J Cataract Refract Surg. 2015;41(5):936–944. doi: 10.1016/j.jcrs.2014.08.036. [DOI] [PubMed] [Google Scholar]
  • 8.Koch DD, Wang L.. Surgically induced astigmatism. J Refract Surg. 2015;31(8):565 doi: 10.3928/1081597X-20150728-03. [DOI] [PubMed] [Google Scholar]
  • 9.Epitropoulos AT, Matossian C, Berdy GJ, Malhotra RP, Potvin R.. Effect of tear osmolarity on repeatability of keratometry for cataract surgery planning. J Cataract Refract Surg. 2015;41(8):1672–1677. doi: 10.1016/j.jcrs.2015.01.016. [DOI] [PubMed] [Google Scholar]
  • 10.Kessel L, Andresen J, Tendal B, Erngaard D, Flesner P, Hjortdal J.. Toric intraocular lenses in the correction of astigmatism during cataract surgery: a systematic review and meta-analysis. Ophthalmology. 2016;123(2):275–286. doi: 10.1016/j.ophtha.2015.10.002. [DOI] [PubMed] [Google Scholar]
  • 11.Novis C. Astigmatism and toric intraocular lenses. Curr Opin Ophthalmol. 2000;11(1):47–50. doi: 10.1097/00055735-200002000-00007. [DOI] [PubMed] [Google Scholar]
  • 12.International Organization for Standardization. Ophthalmic implants—intraocular lenses—part 2: Optical properties and test methods. International Organization for Standardization; 2014: p. ISO 11979–2. https://www.iso.org/standard/55682.html.
  • 13.Zhang W, Pasricha ND, Kuo AN, Vann RR.. Influence of corneal diameter on surgically induced astigmatism in small-incision cataract surgery. Can J Ophthalmol. 2019;54(5):556–559. doi: 10.1016/j.jcjo.2018.12.013. [DOI] [PubMed] [Google Scholar]
  • 14.Hill W. Toric IOL Pre-operative Planning. Presented at Cataract Surgery: Telling It Like It Is, Orlando, Florida, 2019. https://doctor-hill.com/CSTILIS_2019/Hill_2019_Toric.pdf.
  • 15.Roensch MA, Charton JW, Blomquist PH, Aggarwal NK, McCulley JP.. Resident experience with toric and multifocal intraocular lenses in a public county hospital system. J Cataract Refract Surg. 2012;38(5):793–798. doi: 10.1016/j.jcrs.2011.11.043. [DOI] [PubMed] [Google Scholar]