Repeatability and reproducibility test. Results are from plasma and liver (single batch) and pooled serum quality control (QC) samples (analyzed over three large batches of around 115 samples per batch). RSD: relative standard deviation.

Parameter Plasma Liver Batch 1 Batch 2 Batch 3
Number of metabolites 22.8 ±  0.5 37 ±  0 20.3 ± 1.9 20.2 ± 2.4 20.2 ± 2.2
Number of missing values 11 (0.8%) 0 (0%) 24 (11%) 25 (12%) 25 (12%)
Median RSD 16% 10% 21% 20% 19%
RSD range 11–28% 2–56% 3–42% 12–69% 13–39%