Dual-energy CT imaging with limited-angular-range data

Buxin Chen 1, Zheng Zhang 1, Dan Xia 1, Emil Y Sidky 1, Xiaochuan Pan 1,2
PMCID: PMC8704637  NIHMSID: NIHMS1764755  PMID: 34320478

Abstract

In dual-energy computed tomography (DECT), low- and high-kVp data are collected often over a full-angular range (FAR) of 360°. While there exists strong interest in DECT with low- and high-kVp data acquired over limited-angular ranges (LARs), there remains little investigation of image reconstruction in DECT with LAR data.

Objective:

we investigate image reconstruction with minimized LAR artifacts from low- and high-kVp data over LARs of ⩽180° by using a directional-total-variation (DTV) algorithm.

Methods:

image reconstruction from LAR data is formulated as a convex optimization problem in which data-2 is minimized with constraints on image’s DTVs along orthogonal axes. We then achieve image reconstruction by applying the DTV algorithm to solve the optimization problem. We conduct numerical studies from data generated over arcs of LARs, ranging from 14° to 180°, and perform visual inspection and quantitative analysis of images reconstructed.

Results:

monochromatic images of interest obtained with the DTV algorithm from LAR data show substantially reduced artifacts that are observed often in images obtained with existing algorithms. The improved image quality also leads to accurate estimation of physical quantities of interest, such as effective atomic number and iodine contrast concentration.

Conclusion:

our study reveals that from LAR data of low- and high-kVp, monochromatic images can be obtained that are visually, and physical quantities can be estimated that are quantitatively, comparable to those obtained in FAR DECT.

Significance:

as LAR DECT is of high practical application interest, the results acquired in the work may engender insights into the design of DECT with LAR scanning configurations of practical application significance.

Keywords: dual-energy CT, limited-angular-range reconstruction, directional-total-variation, atomic number, iodine cocentration

1. Introduction

In dual-energy CT (DECT), data are collected with low- and high-kVp spectra over a full-scan (or at least a short-scan) range (Alvarez and Macovski 1976); and images are reconstructed then directly or indirectly from low- and high-kVp data often by use of algorithms that are developed basing upon a linear-data model in conventional CT. The images reconstructed are used subsequently for estimation of basis images from which monochromatic images at given energies, and physical quantities of application interest, such as effective atomic numbers (simply referred to as atomic numbers hereinafter) and iodine concentrations, can be estimated (Johnson et al 2007, Maass et al 2009, Chandarana et al 2011, Goodsitt et al 2011, Faby et al 2015, Chen et al 2018).

DECT capable of limited-angular-range (LAR) imaging is of practical application interest because it may allow for the reduction of radiation dose and scanning time and for the design of scanning configurations avoiding possible collisions of the moving gantry, e.g. in a C-arm DECT, with patient or other components involved in the scanning. While image reconstruction is a key to enabling LAR DECT, it remains largely unexplored as only limited effort has been reported in the literature for highly special scanning configurations in which the sum of low- and high-kVp angular ranges is generally larger than 180° (Zhang and Xing 2016, Zhang et al 2020).

In this work, we investigate image reconstructions in LAR DECT by exploiting the directional-total-variation (DTV) algorithm (Zhang et al 2021) recently developed for image reconstruction in conventional CT with LAR data (Batenburg and Sijbers 2011, Liu et al 2016, Xu et al 2019, Zhang et al 2021). In the investigation, for each of the low- and high-kVp data sets collected over LARs in DECT, we first formulate image reconstruction as a convex optimization problem designed in which data-2 is minimized under image’s DTV constraints along orthogonal axes, and then use the DTV algorithm developed recently to solve the optimization problem for achieving image reconstructions. The DTV algorithm may allow for an efficient recovery of ‘invisible boundaries’ (Quinto 2017) along the scanning direction in reconstructed images from the dual-energy LAR data.

In LAR DECT, while images suffer from both LAR and beam-hardening (BH) artifacts, the LAR artifact is dominantly more significant than the BH artifact, as results in section 3 below show. The work focuses thus on LAR artifact correction in LAR DECT without explicit BH-artifact correction. In particular, results of LAR DECT obtained are compared against those of the full-angular-range (FAR) (with full- or short-scan range) DECT without suffering from LAR artifacts.

We carry out numerical studies with digital phantoms mimicking a suitcase and a breast, which are of relevance to industrial and clinical applications, respectively. Both noiseless and noisy data are generated from each phantom with low- and high-kVp spectra over arcs of LARs spanning from 14° to 180°, as well as over the FAR of 360°. Using the images reconstructed from low- and high-kVp LAR data, we estimate basis images, from which monochromatic images at energies of interest are formed. We then perform visual inspection and quantitative analysis of monochromatic images obtained, and estimate physical quantities including atomic numbers and iodine contrast concentrations from the basis images determined.

Following the introduction in section 1, we present materials and methods in section 2, including scanning configuration, data generation, image reconstruction, and image analysis. Image results and quantitative analysis are shown in section 3, while discussions and conclusion are followed in sections 4 and 5, respectively. We include in the appendices description of the methods for determination of basis images and for estimation of physical quantities to avoid distractions from the presentation flow in the main text.

2. Materials and methods

2.1. Scanning configuration

We consider a fan-beam-based scanning configuration, as shown in figure 1, while the study presented in sections 2.3 and 2.4 below can readily be extended to cone-beam geometries. The low- and high-kVp data are collected over two overlapping circular arcs of LAR α, from an object that is within the field of view of the configuration. The image grid is set up such that the circular scanning arc is symmetric relative to the y-axis. This setup with overlapping arcs might be of practical interest in situations where the total scanning angular range is physically limited due to workflow or safety concerns, and it can be implemented with current DECT techniques, such as sandwiched detectors, fast-kVp-switching x-ray tubes, and sequential scans (McCollough et al 2015).

Figure 1.

Figure 1.

Scanning configuration with overlapping arcs of LAR (thin and thick curves) for collecting low- and high-kVp data. The coordinate system of the image array is set up such that the circular scanning arc is symmetric relative to the y-axis.

In this work, we consider LARs α = 14°, 20°, 30°, 60°, 90°, 120°, 150°, and 180°, with an angular interval of 1° per view. Images are also obtained over a FAR of 360° by use of the DTV and FBP algorithms from noiseless data. They are used as the DTV- and FBP-reference images, respectively, against which we investigate how image quality from LAR data decreases. For suitcase phantom studies, the source-to-rotation distance (SRD) and source-to-detector distance (SDD) are 100 and 150 cm, and a linear detector of 32 cm consists of 512 bins, whereas for breast phantom studies, SRD and SDD are 36 and 72 cm, and a linear detector of 37.5 cm consists of 512 bins.

2.2. Dual-energy data

The suitcase and breast numerical phantoms shown in figure 2 are used in the numerical study, as the former is of potential interest in security-scan applications such as baggage screening and the latter mimicks the cross section of a breast in contrast-enhanced imaging. In the suitcase phantom, three bar-shaped regions of interest (ROIs) contain elements C, Al, and Ca, respectively, whereas rectangular and elliptical ROIs are filled with water, ammonium nitrate and fuel oil (ANFO) (Ying et al 2006), teflon, and PVC. In the breast phantom, the background, mixed with adipose and breast tissues, is embedded with three ROIs mimicking iodine contrast-enhanced vasculature and tumor of concentrations of 2, 2.5, and 5 mg ml−1, which are typical values of potential clinical relevance (Jong et al 2003, Volterrani et al 2020). The suitcase and breast phantoms are represented, respectively, by image arrays of 150 × 256 square pixels of size 0.7 mm and of 80 × 256 square pixels of size 0.7 mm.

Figure 2.

Figure 2.

Monochromatic images for the (a) suitcase phantom at 40 keV and (b) breast phantom at 34 keV. The numbers in each of the phantoms indicate ROIs of different materials. In the suitcase phantom, ROIs 0–6 contain C, Al, Ca, water, ANFO, teflon, and PVC, respectively. In the breast phantom, the darker and brighter background regions are adipose tissue and breast tissue (including skin), respectively; ROI 0 contains breast tissue, whereas ROIs 1–3 are filled with iodine contrast agent of concentration 5, 2, and 2.5 mg ml−1, respectively. Display windows for the suitcase and breast phantoms are [0.1, 0.65] cm−1 and [0.2, 0.35] cm−1, respectively.

For either phantom, each pixel is labeled with a material type, which is associated with a linear attenuation coefficient at a given energy. As such, we consider a nonlinear-data model incorporating the polychromatic x-ray spectrum (Chen et al 2017) as

gjs=lnmMqjms exp(iIajisfmi), (1)

where gjs is the model data for ray j, j = 1, 2, …, Js, within the low- (s = L) or high-kVp (s = H) scan; Js the total number of rays in the low- or high-kVp scan; qjms the normalized, low- or high-kVp x-ray spectrum (including detector response) for ray j at energy bin m, m = 1, 2, …, M; M the total number of energy bins; ajis the contribution to ray j, in the low- or high-kVp scan, from pixel i; fmi the linear attenuation coefficient at energy bin m for the labeled material at pixel i, i = 1, 2, …, I; and I the total number of image pixels. We can form matrix As of size Js × I with element ajis and refer to As as the discrete x-ray transform (DXT).

In the numerical studies in section 3 below, we use the nonlinear model in equation (1) to generate noiseless and Poisson-noisy (corresponding to 107 photons per ray in the air scan) data from the suitcase and breast phantoms with low- and high-kVp spectra, which are generated using the TASMIC model (Hernandez and Boone 2014). For the suitcase phantom, the low- and high-kVp spectra are set at 80 and 140 kVp’s, respectively, with an additional 5 mm Al filter for both. For the breast phantom, the low- and high-kVp spectra are set at 33 and 49 kVp’s, with 8 mm Al and 0.25 mm Cu filters, respectively.

2.3. Image reconstruction

We use vector gs[M] of size Js, where s = L or H, to denote measured data, with element gjs[M] indicating low- or high-kVp measurement with ray j in DECT, where j = 1, 2, …, Js. In the noiseless, numerical study below, gjs[M]=gjs in equation (1), as it is considered to be a reasonable data model in DECT, whereas in the noisy numerical study in the work, gjs[M] is obtained by addition of Poisson noise to gjs as described above.

In standard DECT, images are reconstructed often by use of an algorithm that is based upon a linear-data model instead of the nonlinear-data model in equation (1). We propose to use in the work a DTV algorithm that is also based upon a linear-data model for image reconstruction directly from low- and high-kVp data collected over arcs of LAR. The images reconstructed thus contain the BH effect inherent in data generated with the nonlinear-data model.

We formulate the reconstruction problem from either low- or high-kVp data over an arc of LAR as a convex optimization problem (Zhang et al 2021) given by:

fs=argminfs12gs[M]Asfs22 s.t. Dxfs1txs,Dyfs1tys, and fis0 (2)

where s = L or H; 22 operating on a vector denotes the squared 2-norm; vector fs of size I the image to be reconstructed; matrices Dx and Dy of size I × I denote two-point differences along the x- and y-axis, respectively; and Dxfs1 and Dyfs1 are 1 norms of the image partial derivatives along the x- and y-axis, respectively, also referred to as the image directional-total-variations (DTVs). In the formulation of equation (2), linear model As fs, i.e. the DXT of fs, is used to approximate low- or high-kVp data gs[M].

Basing upon the general primal-dual algorithm (Chambolle and Pock 2010, Sidky et al 2012) solving mathematically exactly convex optimization problems, we have developed a DTV algorithm tailored to solve equation (2) for reconstructing fs from low- or high-kVp data collected over an arc of LAR. As the detailed derivation of the DTV algorithm can be found in appendix A of Zhang et al (2021), we list below only the pseudo-code of the DTV algorithm.

2.

In the pseudo-code, stacked matrix Ks is defined as Ks=(As,ν1sDx,ν2sDy,μI), where superscript ‘⊤’ indicates a transpose operation; ∥ · ∥2, when applied to a matrix, computes the largest singular value of the matrix; I is an identity matrix of size I × I; vectors w(n) is of size Js, whereas vectors p(n), q(n), p(n), q(n), and t(n) are of size I; neg( · ) enforces the non-positivity constraint by thresholding; operator 1ballβ( · ) projects a vector onto the 1-ball of size β; and |q(n)| denotes a vector with entry j given by (|q(n)|)j=|qj(n)|, where qj(n) is the jth entry of vector q(n).

In DECT, monochromatic image fm, i.e. the linear attenuation coefficient distribution, at energy bin m is of interest, and it can be decomposed into a linear combination of two basis images b0 and b1 as

fm=μm0b0+μm1b1, (3)

where expansion coefficients μmk (k = 0 or 1) can be either calculated or looked up, and basis images bk can be estimated from low- and high-kVp images fL and fH reconstructed, as described in appendix B.

Image reconstruction with the DTV algorithm, like reconstructions with any algorithms, involves constraint parameters such as txs and tys whose selection can impact reconstruction quality. In the work, we select parameters txs and tys by visual evaluation of monochromatic images obtained with minimum artifacts, as shown in appendix A. In addition, the images are reconstructed with the FBP algorithm, along with a Hanning kernel and a cutoff frequency at 0.5, directly from the low- and high-kVp LAR data as they can provide a benchmark for the DTV reconstructions.

2.4. Analysis of monochromatic images obtained

We reconstruct images fL and fH directly from low- and high-kVp data generated over an arc of LAR by use of the DTV algorithm, and estimate two basis images from fL and fH by using either the interaction- or material-based method described in appendix B. With the basis images estimated, we then compose monochromatic images at energies of interest using equation (3), and perform visual inspection and quantitative analysis of the monochromatic images. Furthermore, we analyze DTV reconstructions in tasks of estimation of atomic number and contrast agent concentration within ROIs defined in figure 2.

2.4.1. Visual inspection and quantitative analysis of monochromatic images

We first perform visual inspection of the monochromatic images obtained at energy levels of interest to assess LAR artifacts. In addition to visual inspection, we compute two quantitative metrics, Pearson correlation coefficient (PCC) and normalized mutual information (nMI) (Pearson 1895, Pluim et al 2003, Bian et al 2010, Zhang et al 2021), for evaluating the visual correlation between a monochromatic image obtained from the LAR data and its corresponding reference image. Specifically, the maximum values of PCC and nMI are 1, and the higher the PCC and nMI values, the better the visual correlation between an image and its reference image. While the image and its reference are identical when PCC = 1 and nMI = 1, the image generally appears visually resembling the reference image even as PCC > 0.8 and nMI > 0.6–0.7.

2.4.2. Estimation of physical quantities

In addition to visual inspection and analysis, we also analyze the DTV reconstructions in two tasks of estimating physical quantities of interest as described below.

Estimation of atomic number:

The study involving the suitcase phantom is of potential interest to industrial/security CT applications, such as baggage screening, in which estimation of the atomic number of materials is used for explosive detection (Ying et al 2006). Using the interaction-based method on the DTV-reference image of the suitcase phantom, we obtain the 2 × 2 decomposition matrix in equation (B.2), which is used for estimating two basis images of photoelectric effect (PE) and Compton scattering (KN) components from fL and fH reconstructed throughout the studies with the suitcase phantom, as discussed in appendix B. Using the estimated basis images in the affine relationship in equation (C.1) in appendix C, we then estimate the atomic numbers within ROIs 3–6 of the suitcase phantom, as shown in figure 2(a). Constants c and n in equation (C.1) are fitted and calibrated using the image values within calibration ROIs 0–2 of the suitcase phantom from the DTV-reference image. The three ROIs correspond to three single-element common materials, C, Al, and Ca, which are picked as their atomic numbers, z = 6, 13, and 20, cover the range of atomic numbers of interest for the other materials contained in the suitcase phantom.

Estimation of iodine concentration:

In mammography, digital breast tomosynthesis (DBT), and breast CT, iodine-based contrast agents can be used for enhancing tumor contrast (Carton et al 2010, Samei and Saunders 2011, De Cecco et al 2012, Dromain et al 2012, Jochelson et al 2013, Zhang et al 2018). Quantitative estimation of iodine contrast concentration is of interest in breast tumor staging (Volterrani et al 2020) and capturing the contrast-uptake kinetics, as it may help differentiate between benign and malignant tumors (Jong et al 2003). In the study involving the breast phantom, the basis materials are selected as breast tissue and 5 mg ml−1 iodine contrast agent in calibration ROIs 0 and 1, respectively, as shown in figure 2(b). Similarly, using the material-based method on the DTV-reference image of the breast phantom, we first obtain the 2 × 2 decomposition matrix in equation (B.2), which is used for estimating two basis images of breast tissue and iodine contrast agent from fL and fH reconstructed throughout the studies with the breast phantom, as discussed in appendix B. Using the estimated basis image of iodine contrast agent in equation (C.2) in appendix C, we estimate the concentration of iodine contrast agent within ROIs 1–3. Constants γ and τ are fitted and calibrated using image values within ROIs 1–3 in the DTV-reference image of the breast phantom, together with their known concentrations.

3. Results

We reconstruct below fs from data gs[M] of the suitcase and breast phantoms in figure 2 from low- and high-kVp LAR data by using the DTV algorithm, where s = L and H. In each reconstruction, the DTV-constraint parameters are selected by use of the approach described in appendix A. Subsequently, using the interaction- or material-based method, as described in appendix B, we estimate the basis images from which monochromatic images fm at energy bin m are obtained by using equation (3). In addition to visual inspection and quantitative analysis of the monochromatic images, we estimate atomic numbers and iodine contrast concentrations within selected ROIs in the suitcase and breast phantoms, respectively, as described in appendix C.

3.1. Image reconstruction from noiseless data of the suitcase phantom

We reconstruct images from noiseless low- and high-kVp data of the suitcase phantom generated over arcs of LARs α = 14°, 20°, 30°, 60°, 90°, 120°, 150°, and 180° by use of the DTV algorithm, and then estimate basis images by using the interaction-based method with the DTV-reference image, as described in appendix B, from the images reconstructed. Subsequently, using the basis images estimated, we obtain monochromatic images at energy 40 keV for enhanced contrast in the images, and then compute atomic numbers within selected ROIs, as described in appendix C.

3.1.1. Visual inspection of monochromatic images

We first show in figure 3 monochromatic images and their respective zoomed-in ROI views at 40 keV obtained from the noiseless data generated over an arc of LAR 60° by use of the DTV and FBP algorithms, along with the DTV- and FBP-reference images from the noiseless data over 360°. The zoomed-in ROI is enclosed by the rectangular box depicted in the FBP-reference image (row 1, column 4) in figure 3. It can be observed that the DTV image from data over 60° displays significantly reduced LAR artifacts, which are otherwise observed and severely obscuring structures in the FBP image from the same data. The DTV image is also visually comparable to the DTV- and FBP-reference images. Furthermore, the contrast between water and ANFO and the edges in the three bar-shaped structures, as shown in the zoomed-in ROI views in the bottom row of figure 3, can be discerned in the DTV image of 60° as clearly as that observed in the DTV- and FBP-reference images.

Figure 3.

Figure 3.

Monochromatic images at 40 keV (top row) and their respective zoomed-in ROI views (bottom row) of the suitcase phantom obtained by use of the DTV (column 1) and FBP (column 2) algorithms from noiseless data generated over an arc of 60°, and the DTV-reference image (column 3) and FBP-reference image (column 4). The ROI is enclosed by the rectangular box depicted in the FBP-reference image. Display window: [0.1, 0.65] cm−1.

We show in figure 4 monochromatic images and their zoomed-in ROI views obtained by use of the DTV algorithm from noiseless data over, respectively, 7 arcs of LARs α = 14°, 20°, 30°, 60°, 90°, 120°, and 150°, along with the DTV-reference image. It can be observed that while the DTV images of α ⩽ 60° contain some visible artifacts as a result of the compound LAR and BH effects, the edges in the bar-shaped structures in ROIs 0–2, as well as other structures, in the suitcase phantom can be discerned as clearly as that observed in the DTV- and FBP-reference images. The remaining artifacts in the DTV images appear to be largely due to the BH effect. The monochromatic image from data over 180° is visually similar to the DTV-reference image and is thus not shown in this study and the following studies.

Figure 4.

Figure 4.

Monochromatic images (rows 1 and 3) of the suitcase phantom at 40 keV obtained from noiseless data generated over arcs of LARs 14°, 20°, 30°, 60°, 90°, 120°, 150°, and 360° by use of the DTV algorithm, along with their respective zoomed-in ROI views (rows 2 and 4). The zoomed-in ROI is enclosed by the rectangular box depicted in the FBP-reference image in figure 3. Display window: [0.1, 0.65] cm−1.

3.1.2. Quantitative analysis of monochromatic images

In addition to visual inspection, using the DTV-reference image (row 1, column 3) in figure 3, we compute metrics PCC and nMI of the DTV monochromatic images of the suitcase phantom, and display them in figure 5 as functions of LAR α. It can be observed that while the PCC and nMI drop understandably as α decreases, they remain generally above 0.9 and 0.5, respectively, suggesting that the DTV monochromatic images obtained with LAR data correlate reasonably well with the DTV-reference image. For providing a benchmark, we also obtain monochromatic images by use of the FBP algorithm for α = 14°, 20°, 30°, 60°, 90°, 120°, 150°, and 180°, but without showing them because the structures in the suitcase phantom are obscured by significant LAR artifacts in these FBP images, similar to those observed in the FBP image of 60° shown in column 2 of figure 3. Using the FBP-reference image (row 1, column 4) in figure 3, we compute metrics PCC and nMI of the FBP monochromatic images and plot them in figure 5. The results reveal that the FBP monochromatic images for α < 180° correlate poorly with their reference image.

Figure 5.

Figure 5.

Metrics PCC and nMI, as functions of LAR α, of monochromatic images of the suitcase at 40 keV obtained by use of the DTV (solid) and FBP (dotted) algorithms from noiseless data.

3.1.3. Estimation of atomic numbers

Using the interaction-based method described in appendices B and C, we compute atomic numbers for materials within ROIs 3–6 in the DTV images of the suitcase phantom as shown in figure 2(a). Specifically, using basis images estimated, and constants c and n in equation (C.1) fitted with calibration materials in ROIs 0–2 from the DTV-reference image, we obtain atomic numbers for materials water, ANFO, teflon, and PVC, respectively, in ROIs 3–6 and plot them as functions of angular range α in figure 6, along with the atomic numbers obtained from the DTV- and FBP-reference images. The results indicate that the atomic numbers obtained with the DTV algorithm for the LARs considered appear to agree well with those obtained from their reference images, only with slight deviations observed for angular ranges less than 90°. Due to the severe LAR artifacts in the corresponding FBP images, their basis images estimated can be negative, and equation (C.1) thus cannot be applied because it involves the computation of a logarithmic. Therefore, no atomic numbers can be estimated from images obtained by use of the FBP algorithm for a majority of the LARs considered in the work.

Figure 6.

Figure 6.

Atomic numbers (solid), as functions of LAR α, for (a) water, (b) ANFO, (c) teflon, and (d) PVC in the suitcase phantom estimated from images reconstructed by use of the DTV algorithm from noiseless data. The two horizontal lines, which are very close, indicate the atomic numbers estimated from the DTV-reference (dashed) and FBP-reference (dotted) images.

3.2. Image reconstruction from noisy data of the suitcase phantom

We repeat the study of section 3.1 except that noisy data are now used, which are obtained by addition of Poisson noise to the corresponding noiseless data, as described in section 2.2.

3.2.1. Visual inspection of monochromatic images

We show in figure 7 monochromatic images and their zoomed-in ROI views obtained by use of the DTV algorithm from noisy data over, respectively, 7 arcs of LARs α = 14°, 20°, 30°, 60°, 90°, 120°, and 150°, along with the FAR of 360°. It can be observed that the DTV images appear to contain only significantly reduced visual artifacts as a result of the compound LAR, BH, and noise, and that structures in the suitcase phantom can be discerned as clearly as those observed in the DTV- and FBP-reference images in figure 3.

Figure 7.

Figure 7.

Monochromatic images (rows 1 and 3) of the suitcase phantom at 40 keV obtained from noisy data generated over arcs of LARs 14°, 20°, 30°, 60°, 90°, 120°, 150°, and 360° by use of the DTV algorithm, along with their respective zoomed-in ROI views (rows 2 and 4). The zoomed-in ROI is enclosed by the rectangular box depicted in the FBP-reference image in figure 3. Display window: [0.1, 0.65] cm−1.

3.2.2. Quantitative analysis of monochromatic images

Using the DTV-reference image (row 1, column 3) in figure 3, we also compute metrics PCC and nMI of the noisy DTV monochromatic images and show them in figure 8 as functions of LAR α. It can be observed that while PCC and nMI drop understandably as α decreases, they remain generally above 0.9 and 0.5, respectively, suggesting that the DTV monochromatic images obtained with noisy LAR data correlate reasonably well with the DTV-reference image. For providing a benchmark, we also obtain monochromatic images by use of the FBP algorithm for LARs α = 14°, 20°, 30°, 60°, 90°, 120°, 150°, and 180°, but without showing them because the structures in the suitcase phantom are obscured by significant LAR artifacts in these FBP images, similar to those observed in the FBP image of 60° shown in column 2 of figure 3. Using the FBP-reference image (row 1, column 4) in figure 3, we compute metrics PCC and nMI of the FBP monochromatic images and plot them in figure 8. The results reveal that the FBP monochromatic images for α < 180° correlate poorly with their reference image.

Figure 8.

Figure 8.

Metrics PCC and nMI, as functions of LAR α, of monochromatic images of the suitcase at 40 keV obtained by use of the DTV (solid) and FBP (dotted) algorithms from noisy data.

3.2.3. Estimation of atomic numbers

Using the interaction-based method described in appendices B and C, we also compute atomic numbers for materials water, ANFO, teflon, and PVC, respectively, within ROIs 3–6 in the DTV images of the suitcase phantom, and plot them as functions of LAR α in figure 9, along with the atomic numbers obtained from the DTV- and FBP-reference images. The results indicate that the atomic numbers obtained with the DTV algorithm for the LARs considered appear to agree well with those obtained from their reference images, only with slight deviations observed for LARs less than 90°. Due to the severe artifacts in images reconstructed by use of the FBP algorithm from LAR data, their basis images estimated can be negative, and equation (C.1) thus cannot be applied because it involves the computation of a logarithmic. Therefore, no atomic numbers can be estimated from images obtained by use of the FBP algorithm for a majority of the LARs considered in the work.

Figure 9.

Figure 9.

Atomic numbers (solid), as functions of LAR α, for (a) water, (b) ANFO, (c) teflon, and (d) PVC in the suitcase phantom estimated from images reconstructed by use of the DTV algorithm from noisy data. The two horizontal lines, which are very close, indicate the atomic numbers estimated from the DTV-reference (dashed) and FBP-reference (dotted) images.

3.3. Image reconstruction from noiseless data of the breast phantom

We reconstruct images from noiseless low- and high-kVp data of the breast phantom generated over arcs of LARs α = 14°, 20°, 30°, 60°, 90°, 120°, 150°, and 180° by use of the DTV algorithm, and then estimate basis images by using the material-based method with the DTV-reference image, as described in appendix B, from the images reconstructed. Subsequently, using the basis images estimated, we obtain monochromatic image at energy 34 keV for enhanced iodine contrast, and compute iodine concentrations within the selected ROIs as described in appendix C.

3.3.1. Visual inspection of monochromatic images

We first show in figure 10 monochromatic images and their respective zoomed-in ROI views at 34 keV obtained from noiseless data over 60° by use of the DTV and FBP algorithms, along with the DTV- and FBP-reference images from the noiseless data over FAR of 360°. The zoomed-in ROI is enclosed by the rectangular box depicted in the FBP-reference image (row 1, column 4) in figure 10. It can be observed that the DTV image from 60° data displays significantly reduced LAR artifacts, which are otherwise observed and severely obscuring structures in the FBP image from data over 60°. The DTV image is also visually comparable to the DTV-reference image and appears sharper than the FBP-reference image.

Figure 10.

Figure 10.

Monochromatic images at 34 keV (top row) and their respective zoomed-in ROI views (bottom row) within the breast phantom obtained by use of the DTV (column 1) and FBP (column 2) algorithms from noiseless data generated over an arc of LAR 60°, and the DTV-reference image (column 3) and FBP-reference image (column 4). The zoomed-in ROI is enclosed by the rectangular box depicted in the FBP-reference image. Display window: [0.2, 0.35] cm−1.

We display in figure 11 monochromatic images and their zoomed-in ROI views obtained by use of the DTV algorithm from noiseless data over, respectively, 7 arcs of LARs α = 14°, 20°, 30°, 60°, 90°, 120°, and 150°, along with the DTV-reference image. It can be observed that while the DTV images of α ⩽ 30° contain some minor visible artifacts as a result of the compound LAR and BH effects, the DTV images visually resemble the DTV-reference image.

Figure 11.

Figure 11.

Monochromatic images (rows 1 and 3) of the breast phantom at 34 keV obtained from noiseless data generated over arcs of LARs 14°, 20°, 30°, 60°, 90°, 120°, 150°, and 360° by use of the DTV algorithm, along with their respective zoomed-in ROI views (rows 2 and 4). The zoomed-in ROI is enclosed by the rectangular box depicted in the FBP-reference image in figure 10. Display window: [0.2, 0.35] cm−1.

3.3.2. Quantitative analysis of monochromatic images

Using the DTV-reference image (row 1, column 3) in figure 10, we compute metrics PCC and nMI of the DTV monochromatic images of the breast phantom, and display them in figure 12 as functions of LAR α. It can be observed that while the PCC and nMI drop understandably as α decreases, they remain generally close to 1 and above 0.6, respectively, suggesting that the DTV monochromatic images obtained from LAR data correlate well with the DTV-reference image. For providing a benchmark, we also obtain monochromatic images by use of the FBP algorithm for α = 14°, 20°, 30°, 60°, 90°, 120°, 150°, and 180°, but without showing them because the structures in the breast phantom are obscured by significant LAR artifacts in these FBP images, similar to those observed in the FBP image from data over 60° shown in column 2 of figure 10. Using the FBP-reference image (row 1, column 4) in figure 10, we compute metrics PCC and nMI of the FBP monochromatic images and plot them in figure 12. The results reveal that the FBP monochromatic images for α < 180° correlate poorly with their reference image.

Figure 12.

Figure 12.

Metrics PCC and nMI, as functions of LAR α, of monochromatic images at 34 keV obtained by use of the DTV (solid) and FBP (dotted) algorithms from noiseless data of the breast phantom.

3.3.3. Estimation of iodine concentrations

Using the material-based method described in appendices B and C, we estimate iodine concentrations within ROIs 1–3 of the breast phantom, as shown in figure 2(b), using equation (C.2). Specifically, using basis images estimated from the DTV images, along with constants γ and τ in equation (C.2) fitted from the DTV-reference image, we obtain iodine concentrations for ROIs 1–3 of the breast phantom and plot them as functions of LAR α in figure 13, along with the iodine concentrations obtained from the DTV- and FBP-reference images. The results indicate that the iodine concentrations obtained with the DTV algorithm for the LARs considered appear to agree well with those obtained from their reference images, only with slight deviations observed for LARs less than 90°. Due to the severe LAR artifacts in images reconstructed by use of the FBP algorithm, the basis images estimated can be negative and cannot thus be interpreted meaningfully as iodine concentrations, which must physically be non-negative. Therefore, no iodine concentration is estimated from images reconstructed by use of the FBP algorithm for a majority of the LARs considered in the work.

Figure 13.

Figure 13.

Estimated iodine concentrations (solid), as functions of LAR α, for ROIs 1–3 in the breast phantom from images reconstructed by use of the DTV algorithm from noiseless data. The two horizontal lines, which are very close, indicate the iodine concentrations estimated from the DTV-reference (dashed) and FBP-reference (dotted) images.

3.4. Image reconstruction from noisy data of the breast phantom

We also repeat the study of section 3.3 except that noisy data are now used, which are obtained by addition of Poisson noise to the corresponding noiseless data, as described in section 2.2.

3.4.1. Visual inspection of monochromatic images

We show in figure 14 monochromatic images and their zoomed-in ROI views obtained by use of the DTV algorithm from noisy data generated over, respectively, 7 arcs of LARs α = 14°, 20°, 30°, 60°, 90°, 120°, and 150°, along with the FAR of 360°. It can be observed that the DTV images appear to visually resemble the DTV-reference image, while images from data over α ⩽ 30° contain some visible artifacts as a result of the compound effect of the LAR, BH, and noise.

Figure 14.

Figure 14.

Monochromatic images (rows 1 and 3) of the breast phantom at 34 keV obtained from noisy data generated over arcs of LARs 14°, 20°, 30°, 60°, 90°, 120°, 150°, and 360° by use of the DTV algorithm, along with their respective zoomed-in ROI views (rows 2 and 4). The zoomed-in ROI is enclosed by the rectangular box depicted in the FBP-reference image in figure 10. Display window: [0.2, 0.35] cm−1.

3.4.2. Quantitative analysis of monochromatic images

Using the DTV-reference image (row 1, column 3) in figure 10, we compute metrics PCC and nMI of the DTV monochromatic images obtained, and display them in figure 15 as functions of LAR α. It can be observed that while PCC and nMI drop understandably as α decreases, they remain generally close to 1 and above 0.5, respectively, suggesting that the DTV monochromatic images from LAR data correlate reasonably well with the DTV-reference image. For providing a benchmark, we also obtain monochromatic images by use of the FBP algorithm for α = 14°, 20°, 30°, 60°, 90°, 120°, 150°, and 180°, but without showing them because the structures in the breast phantom are obscured by significant LAR artifacts in these FBP images, similar to those observed in the FBP image of 60° shown in column 2 of figure 10. Using the FBP-reference image (row 1, column 4) in figure 10, we compute metrics PCC and nMI of the FBP monochromatic images and plot them in figure 15. The results reveal that the FBP monochromatic images for α < 180° correlate poorly with their reference image.

Figure 15.

Figure 15.

Metrics PCC and nMI, as functions of LAR α, of monochromatic images at 34 keV obtained by use of the DTV (solid) and FBP (dotted) algorithms from noisy data of the breast phantom.

3.4.3. Estimation of iodine concentrations

Using the material-based method described in appendices B and C, we also estimate iodine concentrations within ROIs 1–3 of the breast phantom using equation (C.2), and plot them as functions of LAR α in figure 16, along with the iodine concentrations obtained from the DTV- and FBP-reference images. The results suggest that the iodine concentrations obtained with the DTV algorithm appear to agree well with those obtained from their reference images, only with slight deviations observed for angular ranges less than 90°. Due to the severe LAR artifacts in images reconstructed by use of the FBP algorithm, the basis images estimated can be negative and cannot thus be interpreted meaningfully as iodine concentrations, which must physically be non-negative. Therefore, no iodine concentration is estimated from images reconstructed by use of the FBP algorithm for a majority of the LARs considered in the work.

Figure 16.

Figure 16.

Estimated iodine concentrations (solid), as functions of LAR α, for ROIs 1–3 in the breast phantom from images reconstructed by use of the DTV algorithm from noisy data. The two horizontal lines, which are very close, indicate the atomic numbers estimated from the DTV-reference (dashed) and FBP-reference (dotted) images.

4. Discussions

In the work, we have tailored the DTV algorithm developed previously to investigating image reconstructions with minimized LAR artifacts from low- and high-kVp data in LAR DECT. In particular, the results of LAR DECT are compared against that of FAR DECT, which by definition is free of LAR artifacts. The reconstruction problem is formulated as a convex optimization problem involving separate DTV constraints along orthogonal directions designed for a given LAR configuration. The DTV algorithm was applied to solving the optimization problem for achieving image reconstruction from low- and high-kVp data generated over arcs of LARs, ranging from 14° to 180°. From the images reconstructed, we estimate basis images and then obtain monochromatic images at energies of interest. Monochromatic images obtained have been visually inspected and quantitatively analyzed through comparison with their reference images in FAR DECT, revealing that monochromatic images obtained with the DTV algorithm are with substantially reduced artifacts that are observed often in monochromatic images obtained with existing algorithms in DECT. Additionally, using the basis images estimated, we have computed atomic numbers and iodine concentrations, and again compared them with their respective references obtained in FAR DECT. The study results reveal that the DTV algorithm can yield physical quantities such as atomic number and iodine concentration comparable to those estimated in FAR DECT. For comparison, we have applied also a standard isotropic total variation (ITV) algorithm (Zhang et al 2021) to reconstructing images from low- and high-kVp LAR data and show them in figure 17. It can be observed that the ITV images contain conspicuous LAR artifacts, which are, however, largely reduced in the corresponding DTV images in figures 4 and 11.

Figure 17.

Figure 17.

Monochromatic images of the suitcase phantom at 40 keV (row 1) and of the breast phantom at 34 keV (row 2) obtained from noiseless data generated over arcs of LARs 14° (columns 1), 20° (columns 2), 30° (columns 3), and 60° (columns 4) by use of the ITV algorithm. Display windows: [0.1, 0.65] cm−1 and [0.2, 0.35] cm−1 for the suitcase and breast phantoms, respectively.

The scanning configuration considered in the work includes completely overlapping arcs of LARs for the low- and high-kVp scans. However, our proposed approach can readily be applied to DECT scanning configuration that consists of two partially- or non-overlapping arcs of LARs of either same or different spans, because the DTV algorithm allows images to be reconstructed separately for low- and high-kVp data collected over LARs. We are investigating currently image reconstruction from low- and high-kVp data collected over two arcs of LARs that are not completely overlapping with each other (Chen et al 2021a).

While the optimization problem and DTV algorithm are designed in the work for two-dimensional DECT with a fan-beam scanning configuration, it is conceptually and mathematically straightforward to extend them to three-dimensional (3D) DECT with a cone-beam scanning configuration. The key to the extension is to design DTV constraints along orthogonal axes specific to a given 3D scanning geometry. We are investigating currently the design of a DTV algorithm for 3D image reconstruction from data collected over a circular segment in DBT with a cone-beam projection geometry.

Similar to many of the existing algorithms for image reconstruction in DECT, the DTV and FBP algorithms are based upon a linear-data model, i.e. the DXT that does not model the nonlinear-BH effect. As such, BH artifacts may be observed also in monochromatic images obtained with the DTV and FBP algorithms from FAR data. The BH effect may also result in estimation errors in physical quantities such as atomic number and iodine concentration. The work is not intended to correct for the BH artifacts; instead, it focuses on investigating the LAR effect on monochromatic images and physical quantities estimated relative to those obtained in FAR DECT without explicit BH-artifact correction. One can develop algorithms by basing upon the nonlinear-data model (see equation (1)) to correct for the BH artifacts and to improve the accuracy of physical quantity estimation (Zou and Silver 2008, Barber et al 2016, Chen et al 2021b). For DECT with completely overlapping arcs of LARs for collecting low- and high-kVp data, we are investigating currently the application of a data-domain method for BH-effect correction (Zou and Silver 2008) to low- and high-kVp LAR data. From the corrected basis projections, the DTV algorithm can then be tailored to reconstruct basis images and monochromatic images with BH artifacts corrected. Furthermore, one may develop a one-step algorithm (Chen et al 2021b) with DTV constraints basing upon the nonlinear-data model in equation (1) for reconstructing basis images without the BH artifacts directly from low- and high-kVp data in LAR DECT, leading to monochromatic images and physical quantities free from BH artifacts.

In the work, we have focused on investigating image reconstructions and physical quantity estimation in LAR DECT with computer-simulated data generated from phantoms of practical relevance. However, it would be necessary and important to evaluate the approach to image reconstruction in LAR DECT by use of real data collected in research or practical applications. Knowledge acquired in the work can be exploited to design and conduct extensive studies on image reconstructions from real LAR data collected in research and clinical DECT.

5. Conclusion

In this work, using the DTV algorithm developed previously for conventional LAR CT, we have investigated image reconstruction from low- and high-kVp data in LAR DECT. Results of our studies reveal that monochromatic images obtained from data collected over arcs of LARs as low as 60° appear visually comparable to their corresponding reference images obtained in FAR DECT and that the accuracy of atomic numbers and iodine concentrations estimated from data of LARs across 14°–180° is quantitatively comparable to that obtained in FAR DECT. The results acquired in the work may engender insights into the design of DECT with LAR scanning configurations of application significance.

Acknowledgments

This work was supported in part by NIH R01 Grant Nos. EB026282 and EB023968, and the Grayson-Jockey Club Research. The contents of this article are solely the responsibility of the authors and do not necessarily represent the official views of the National Institutes of Health.

Appendix A. Parameter selection for txs and tys

Like any reconstruction algorithms, the DTV algorithm investigated in the work for LAR DECT involves constraint parameters, such as txs and tys, that can impact the reconstruction. Without loss of generality, we consider an example of DTV reconstruction from low- and high-kVp data of the breast phantom collected over an arc of 60°, and illustrate how txs and tys are selected by basing upon visual inspection of monochromatic images obtained for artifacts reduction.

From data of the breast phantom over 60°, we perform image reconstructions for multiple values of txs and tys selected and show in figure A1 monochromatic images at 34 keV obtained with four different sets of selected values of txs and tys. It can be observed that the DTV-constraint parameters can impact image reconstruction. After sweeping through a range of the DTV parameters, we select empirically txs and tys that yield an image of minimized artifacts, as shown in column 1 of figure A1, for image reconstruction in the case. The parameters for the reconstructions with the DTV algorithm are selected consistently in this way throughout this work.

Figure A1.

Figure A1.

Monochromatic images at 34 keV of the breast phantom obtained by use of the DTV algorithm from 60° data with (txL,tyL,txH,tyH)=(123.5,224.0,80.8,156.0) (a), (169.0, 224.0, 102.0, 156.0) (b), (130.0, 224.0, 80.8, 165.8) (c), and (130.0, 238.0, 85.0, 175.5) (d), respectively. Display window: [0.2, 0.35] cm−1.

Appendix B. Basis and monochromatic images

Letting fmL and fmH denote monochromatic images at two different energy levels mL and mH, we obtain from equation (3)

(fmLifmHi)=(μmL0μmL1μmH0μmH1)(b0ib1i), (B.1)

where fmLi and fmHi denote values at pixel i of fmL and fmH, and bki the value at pixel i of basis image k. In the work, images fL and fH are reconstructed directly from low- and high-kVp spectral data by use of either the FBP or DTV algorithm, and basis images can be estimated approximately from fL and fH (or from their corrected versions (Maass et al 2009)). Using fiL and fiH at pixel i of fL and fH to approximate monochromatic images fmLi and fmHi in equation (B.1), we obtain estimates of basis images as

(b0ib1i)(μmL0μmL1μmH0μmH1)1(fiLfiH). (B.2)

As a result, if μmLk and μmHk (k = 0 and 1) can be estimated, they can be used in equation (B.2) for obtaining basis images bk and subsequently monochromatic image fm using equation (3). We consider in the work the interaction- and material-based methods, as described below, for computing μmLk and μmHk in estimation tasks and of atomic number and iodine concentration, respectively.

In the interaction-based method, the photoelectric effect and Compton scattering are the basis functions of choice, which cannot be mutually exclusively present within a ROI in the image. As a result, we estimate μmLk and μmHk by estimating the energy levels mL and mH of a common material within pre-determined calibration ROIs from knowledge of fL and fH. For a calibration ROI containing a common material, e.g. water, the mean values are first obtained by averaging fiL or fiH over pixels within the ROI. Using the mean values, we then determine effective energy levels mL and mH that yield the closest values of linear attenuation coefficients for the material within the ROI by using the NIST table (Hubbell and Seltzer 2004). Plugging μmLk and μmHk estimated into equation (B.2), we obtain basis images within the ROI. Moreover, while μmLk and μmHk are estimated only from knowledge of fL and fH within the calibration ROI, they are used immediately to compute the basis images at pixels that are not within the calibration ROI. Combination of the basis images estimated within and outside the calibration ROI yields the basis images for the entire image array. Finally, using the basis images estimated and μmk at energy level m, calculated using 1/E3 (where E is the energy corresponding to energy bin m) and the Klein–Nishina formula (Attix 2008), we then obtain a monochromatic image for all of the pixels in the image array by using equation (3).

In the material-based method, two basis images b0 and b1 represent two different materials, which are breast tissue and iodine in our study with the breast phantom. Two calibration ROIs 0 and 1, corresponding to these two basis materials, are first selected in the breast phantom. For calibration ROI 0 in the breast phantom containing breast tissue, the pixel values of b0 and b1 are 1 and 0, respectively. Conversely, for calibration ROI 1 in the breast phantom containing iodine, the pixel values of b0 and b1 are thus 0 and 1, respectively. Therefore, for calibration ROIs 0 and 1, individual estimates of μmLk and μmHk for k = 0 and 1 are obtained simply as the corresponding pixel values of fL and fH within calibration ROIs 0 and 1. We then compute the final estimates by averaging the individual estimates over pixels within the calibration ROIs, e.g. μmLk=iUkfiL/Ik, where Uk, with size Ik, denotes the set of those pixels within ROI k (k = 0 or 1). Plugging the final estimates of μmLk and μmHk into equation (B.2), we obtain estimates of basis images from knowledge of fL and fH for pixels outside the ROIs. Combination of the basis images estimated within and outside the calibration ROIs yields the basis images for the entire image array. Finally, using the basis images estimated and μmk at energy level m, looked up from the NIST table, we then obtain a monochromatic image for all of the pixels in the image array by using equation (3).

Appendix C. Estimation of physical quantities

Estimation of atomic number:

For the energy range of our interest, monochromatic image fm in equation (3) can be interpreted as a linear combination of two interaction-based basis images, i.e. the photoelectric effect (PE) (k = 0) and Compton scattering (KN) (k = 1) (Alvarez and Macovski 1976, Attix 2008). As a result, from the two basis images, we can estimate atomic number z^i at pixel i as (Alvarez and Macovski 1976, Ying et al 2006)

ln z^i=ln c+n ln(b0ib1i), (C.1)

where b0i and b1i are values at pixel i for basis images b0 and b1 corresponding to the PE and KN interaction types, respectively, which are estimated from knowledge of fL and fH by use of the interaction-based method described in appendix B, and constants c and n are related to the physical modeling of the cross sections for PE and KN (Alvarez and Macovski 1976, Attix 2008). As equation (C.1) depicts an affine relationship between the atomic number and the ratio of the basis image values in the log–log domain, constants c and n can be estimated using a linear regression in the log–log domain and calibration materials with known atomic numbers, such as single-element materials, in the imaged subject. For each material, a single atomic number is computed as the mean value of estimated atomic numbers within the ROI.

Estimation of iodine concentration:

In the study of the breast phantom, we use the material-based method described in appendix B to estimate b0 and b1 for breast tissue and iodine contrast agent from knowledge of fL and fH. An affine relationship is assumed between basis image b1 of the iodine contrast agent and the estimated iodine concentration β^i at pixel i as

β^i=γb1i+τ, (C.2)

where b1i is the image value at pixel i of the iodine basis, and γ and τ are the linear coefficients, which can be estimated using a linear regression in a calibration scan with known iodine concentrations. For each iodine contrast agent ROI, a single concentration level is computed as the mean value of estimated iodine concentrations in the ROI.

References

  1. Alvarez RE and Macovski A 1976. Energy-selective reconstructions in x-ray computerised tomography Phys. Med. Biol 21 733–44 [DOI] [PubMed] [Google Scholar]
  2. Attix FH 2008. Introduction to Radiological Physics and Radiation Dosimetry (New York: Wiley; ) [Google Scholar]
  3. Barber RF, Sidky EY, Schmidt TG and Pan X 2016. An algorithm for constrained one-step inversion of spectral CT data Phys. Med. Biol 61 3784–818 [DOI] [PMC free article] [PubMed] [Google Scholar]
  4. Batenburg KJ and Sijbers J 2011. DART: a practical reconstruction algorithm for discrete tomography IEEE Trans. Imag. Proc 20 2542–53 [DOI] [PubMed] [Google Scholar]
  5. Bian J, Siewerdsen JH, Han X, Sidky EY, Prince JL, Pelizzari CA and Pan X 2010. Evaluation of sparse-view reconstruction from flat-panel-detector cone-beam CT Phys. Med. Biol 55 6575–99 [DOI] [PMC free article] [PubMed] [Google Scholar]
  6. Carton A, Gavenonis S, Currivan J, Conant E, Schnall M and Maidment A 2010. Dual-energy contrast-enhanced digital breast tomosynthesis-a feasibility study Br. J. Radiol 83 344–50 [DOI] [PMC free article] [PubMed] [Google Scholar]
  7. Chambolle A and Pock T 2010. A first-order primal-dual algorithm for convex problems with applications to imaging J. Math. Imaging Vis 40 120–45 [Google Scholar]
  8. Chandarana H, Megibow AJ, Cohen BA, Srinivasan R, Kim D, Leidecker C and Macari M 2011. Iodine quantification with dual-energy CT: phantom study and preliminary experience with renal masses Am. J. Roentgenol 196 W693–700 [DOI] [PubMed] [Google Scholar]
  9. Chen B, Zhang Z, Sidky EY, Xia D and Pan X 2017. Image reconstruction and scan configurations enabled by optimization-based algorithms in multispectral CT Phys. Med. Biol 62 8763–93 [DOI] [PMC free article] [PubMed] [Google Scholar]
  10. Chen B, Zhang Z, Xia D, Sidky EY and Pan X 2018. Algorithm-enabled partial-angular-scan configurations for dual-energy CT Med. Phys 45 1857–70 [DOI] [PMC free article] [PubMed] [Google Scholar]
  11. Chen B, Zhang Z, Xia D, Sidky EY and Pan X 2021a. 90-plus-90 DECT imaging Proc. 16th Int. Meeting on Fully Three-Dimensional Image Reconstruction in Radiology and Nuclear Medicine submitted [Google Scholar]
  12. Chen B, Zhang Z, Xia D, Sidky EY and Pan X 2021b. Non-convex primal-dual algorithm for image reconstruction in spectral CT Comput. Med. Imaging Graph 87 101821. [DOI] [PMC free article] [PubMed] [Google Scholar]
  13. De Cecco CN, Darnell A, Rengo M, Muscogiuri G, Bellini D, Ayuso C and Laghi A 2012. Dual-energy CT: oncologic applications Am. J. Roentgenol 199 S98–105 [DOI] [PubMed] [Google Scholar]
  14. Dromain C, Thibault F, Diekmann F, Fallenberg EM, Jong RA, Koomen M, Hendrick RE, Tardivon A and Toledano A 2012. Dual-energy contrast-enhanced digital mammography: initial clinical results of a multireader, multicase study Breast Cancer Res. 14 1–18 [DOI] [PMC free article] [PubMed] [Google Scholar]
  15. Faby S, Kuchenbecker S, Sawall S, Simons D, Schlemmer H-P, Lell M and Kachelrieß M 2015. Performance of todayʼs dual energy CT and future multi energy CT in virtual non-contrast imaging and in iodine quantification: a simulation study Med. Phys 42 4349–66 [DOI] [PubMed] [Google Scholar]
  16. Goodsitt MM, Christodoulou EG and Larson SC 2011. Accuracies of the synthesized monochromatic CT numbers and effective atomic numbers obtained with a rapid kvp switching dual energy CT scanner Med. Phys 38 2222–32 [DOI] [PubMed] [Google Scholar]
  17. Hernandez AM and Boone JM 2014. Tungsten anode spectral model using interpolating cubic splines: unfiltered x-ray spectra from 20 kv to 640 kv Med. Phys 41 042101. [DOI] [PMC free article] [PubMed] [Google Scholar]
  18. Hubbell J and Seltzer S 2004. Tables of x-ray mass attenuation coefficients and mass energy-absorption coefficients (version 1.4) ( 10.18434/T4D01F) [DOI]
  19. Jochelson MS, Dershaw DD, Sung JS, Heerdt AS, Thornton C, Moskowitz CS, Ferrara J and Morris EA 2013. Bilateral contrast-enhanced dual-energy digital mammography: feasibility and comparison with conventional digital mammography and mr imaging in women with known breast carcinoma, Radiology 266 743–51 [DOI] [PMC free article] [PubMed] [Google Scholar]
  20. Johnson TRC et al. 2007. Material differentiation by dual energy CT: initial experience Eur. Radiol 17 1510–7 [DOI] [PubMed] [Google Scholar]
  21. Jong RA, Yaffe MJ, Skarpathiotakis M, Shumak RS, Danjoux NM, Gunesekara A and Plewes DB 2003. Contrast-enhanced digital mammography: initial clinical experience Radiology 228 842–50 [DOI] [PubMed] [Google Scholar]
  22. Liu Y, Beyer A, Schuetz P, Hofmann J, Flisch A and Sennhauser U 2016. Cooperative data fusion of transmission and surface scan for improving limited-angle computed tomography reconstruction, NDT&E Int. 83 24–31 [Google Scholar]
  23. Maass C, Baer M and Kachelriess M 2009. Image-based dual energy CT using optimized precorrection functions: a practical new approach of material decomposition in image domain Med. Phys 36 3818–29 [DOI] [PubMed] [Google Scholar]
  24. McCollough CH, Leng S, Yu L and Fletcher JG 2015. Dual- and multi-energy CT: principles, technical approaches, and clinical applications Radiol. 276 637–53 [DOI] [PMC free article] [PubMed] [Google Scholar]
  25. Pearson K 1895. Notes on regression and inheritance in the case of two parents Proc. R. Soc 58 240–2 (http://refhub.elsevier.com/S1361-8415(21)00076-1/sbref0019) [Google Scholar]
  26. Pluim JPW, Maintz JBA and Viergever MA 2003. Mutual-information-based registration of medical images: a survey IEEE Trans. Med. Imag 22 986–1004 [DOI] [PubMed] [Google Scholar]
  27. Quinto ET. Artifacts and visible singularities in limited data x-ray tomography. Sens. Imaging. 2017;18:9. [Google Scholar]
  28. Samei E and Saunders RS Jr 2011. Dual-energy contrast-enhanced breast tomosynthesis: optimization of beam quality for dose and image quality Phys. Med. Biol 56 6359–78 [DOI] [PMC free article] [PubMed] [Google Scholar]
  29. Sidky EY, Jorgensen JH and Pan X 2012. Convex optimization problem prototyping for image reconstruction in computed tomography with the Chambolle-Pock algorithm Phys. Med. Biol 57 3065–91 [DOI] [PMC free article] [PubMed] [Google Scholar]
  30. Volterrani L, Gentili F, Fausto A, Pelini V, Megha T, Sardanelli F and Mazzei MA 2020. Dual-energy CT for locoregional staging of breast cancer: preliminary results Am. J. Roentgenol 214 707–14 [DOI] [PubMed] [Google Scholar]
  31. Xu J, Zhao Y, Li H and Zhang P 2019. An image reconstruction model regularized by edge-preserving diffusion and smoothing for limited-angle computed tomography Inverse Probl. 35 085004 [Google Scholar]
  32. Ying Z, Naidu R and Crawford CR 2006. Dual energy computed tomography for explosive detection J. X-Ray Sci. Technol 14 235–56 (https://content.iospress.com/articles/journal-of-x-ray-science-and-technology/xst00163) [Google Scholar]
  33. Zhang H and Xing Y 2016. Reconstruction of limited-angle dual-energy CT using mutual learning and cross-estimation (mlce) Proc. SPIE 9783 978344 [Google Scholar]
  34. Zhang W, Wang L, Li L, Niu T, Li Z, Liang N, Xue Y, Yan B and Hu G 2020. Reconstruction method for DECT with one half-scan plus a second limited-angle scan using prior knowledge of complementary support set (Pri-CSS) Phys. Med. Biol 65 025005. [DOI] [PubMed] [Google Scholar]
  35. Zhang X, Zheng C, Yang Z, Cheng Z, Deng H, Chen M, Duan X, Mao J and Shen J 2018. Axillary sentinel lymph nodes in breast cancer: quantitative evaluation at dual-energy CT Radiology 289 337–46 [DOI] [PubMed] [Google Scholar]
  36. Zhang Z, Chen B, Xia D, Sidky EY and Pan X 2021. Directional-TV algorithm for image reconstruction from limited-angular-range data Med. Image Anal 70 102030. [DOI] [PMC free article] [PubMed] [Google Scholar]
  37. Zou Y and Silver MD 2008. Analysis of fast kv-switching in dual energy CT using a pre-reconstruction decomposition technique Proc. SPIE 6913 691313 [Google Scholar]

RESOURCES