Abstract
A stable contact force (CF) is correlated with more effective radiofrequency (RF) ablation (RFA) lesions and long-term procedural outcomes. Efforts to improve catheter stability include jet ventilation, pacing, steerable sheaths, and CF-sensing ablation catheters. This study compares CF stability and effective RF lesions between two commercially available steerable sheaths. Thirty patients underwent first-time RFA at a single center using the Agilis™ NxT (Abbott, Chicago, IL, USA) or SureFlex™ (Baylis Medical, Montreal, Canada) steerable sheath. High-power short-duration RFA was utilized, targeting a 10-Ω drop. Sheath performance was assessed for the entire procedure and around each pulmonary vein (PV) in terms of mean CF, CF variability, RF time per lesion, and inefficient contact lesions (defined as lesions with a CF of less than 5 g for at least 10% of the RF delivery time). The operator-targeted mean CF was achieved using both sheaths; however, the overall CF variability was 12.8% lower when using the SureFlex™ sheath (p = 0.08). The CF variability was generally 16% greater in the right PVs than the left PVs (p = 0.001) but trended lower with the SureFlex™ sheath. There were 8% more inefficient contact lesions created when using the Agilis™ sheath as compared to the SureFlex™ sheath (p = 0.035), especially in the right inferior PV (p = 0.009). The RF time per lesion was, on average, 12% (1.4 seconds) shorter when using the SureFlex™ sheath than the Agilis™ sheath (p < 0.05). The choice of steerable sheath may affect both catheter stability and lesion quality, especially in the right PVs.
Keywords: Ablation, atrial fibrillation, catheter stability, contact force, pulmonary vein isolation, steerable sheath
Introduction
While radiofrequency (RF) catheter ablation for atrial fibrillation (AF) is associated with acute pulmonary vein (PV) isolation (PVI) in over 90% of cases, long-term success of a single ablation for paroxysmal AF has been reported to be 69% at one year1 and only 54% beyond three years.2 Long-term PVI durability is often a function of continuity and transmurality of formed lesions3 and has been associated with the amount of RF energy delivered. The relative RF energy received by the tissue and, consequently, lesion quality are dependent on consistent coupling between the ablation catheter tip and the target tissue.4,5 Efforts to improve catheter–tissue contact and stability include the use of high-frequency jet ventilation to minimize respiratory excursion, cardiac pacing, electroanatomic mapping, contact force (CF)-sensing ablation catheters, and sheath selection.6–11 Steerable sheaths significantly enhance CF stability, facilitate mapping and ablation, and reduce procedure times when compared to fixed curve sheaths.12 Steerable sheaths are the active component maintaining stability during mapping and ablation compared to the passive ablation catheter. The loads these sheaths face are a function of left atrial (LA) dwell time and position. If steerable sheaths respond to these loads differently, the response to RF ablation (RFA) could be affected. In this study, CF stability and effective RF lesions were compared between two different commercially available steerable sheaths.
Materials and methods
Study design
A retrospective analysis was performed on 30 consecutive RFA procedures conducted at a single center by a single operator (D. P. S.) between February and June 2019. Patients with a history of previous ablation or cardiac surgery, as well as cardiac implants, were excluded from the series. An initial 15 consecutive patients undergoing PVI procedures using the 8.5-French (Fr), 72.5-cm, bidirectional Agilis™ NxT Steerable Introducer (Abbott, Chicago, IL, USA) were compared to a subsequent 15 consecutive patients undergoing PVI procedures using the 8.5-Fr, 72.5-cm, bidirectional SureFlex™ Steerable Guiding Sheath (Baylis Medical, Montreal, Canada). Standard informed consent was obtained prior to each procedure. Institutional approval was obtained for the retrospective chart review.
Radiofrequency ablation procedure
Procedures were performed under general anesthesia and as per the usual protocol. Percutaneous access was obtained from the right and left femoral veins for all catheters. Intravenous heparin was administered to maintain an activated clotting time of approximately 350 seconds. Transseptal puncture was performed under intracardiac echocardiography and fluoroscopy guidance using the NRG Transseptal Needle (Baylis Medical) with the TorFlex Transseptal Guiding Sheath (Baylis Medical) or the SL-1 Transseptal Guiding Introducer (Abbott, Chicago, IL, USA). Three-dimensional electroanatomic mapping (EnSite Precision™ mapping system; Abbott) was used for catheter guidance and CF measurement. The Agilis™ NxT or SureFlex™ steerable sheath was used to position the ablation catheter (TactiCath Contact Force Ablation Catheter; Abbott) for point-by-point circumferential ablation using a high-power short-duration protocol of approximately 50 W RF energy. A target CF of 10 g to 15 g was applied until an impedance drop of approximately 10 Ω and a lesion size index of 4 to 6 were achieved in both sheath groups. The left superior and inferior PVs (LSPV and LIPV, respectively) were isolated before attempting to isolate the right superior and inferior PVs (RSPV and RIPV, respectively). The esophageal temperature was monitored during the ablation to remain below approximately 38°C. Acute PV reconnection was assessed by monitoring electrocardiograms, using a PV catheter (Advisor™ HD-Grid; Abbott) for electrical isolation and voltage maps for re-appearance of PV potentials within a waiting time of approximately 10 minutes.
Data analysis
CF was recorded using the EnSite mapping system at approximately 10-ms intervals during each episode of RF application and was exported from the mapping system for further analysis. To account for procedural complexity, number of lesions, and overall procedure time between patients in each sheath group, CF parameters were evaluated on a per-lesion basis. The RF time per lesion and mean CF per lesion were used to assess the procedural efficiency. Stability in CF was evaluated in terms of CF variability within an individual lesion; standard deviation was calculated for the mean CF within each lesion (Figure 1). Inefficient contact lesions were defined as those whereby CF dropped below 5 g (minimum level to ensure catheter–tissue contact12) for at least 10% of the total RF application time for the lesion.13,14 Analyses were performed for the overall procedure, as well as by correlating each lesion to individual PVs on the electroanatomic map.
Statistical analysis
Baseline patient characteristics were compared using a T-test between the two steerable sheath groups. A hierarchical regression analysis was performed for the procedural parameters (mean CF, RF time per lesion, CF variability) using a linear mixed-effects model by adjusting for individual lesions and each patient in the R software (version 1.1; R Foundation for Statistical Computing, Vienna, Austria). Time-sequence data to determine inefficient contact lesions were analyzed using MATLAB (version 9.4; MathWorks, Natick, MA, USA); significance was evaluated using an F-test on R.
Results
Contact force analysis by location
A total of 3,157 lesions were analyzed with a mean of 105.2 ± 33.1 lesions per patient. Overall, the mean CF achieved during the ablation of right-sided PVs was 17% higher (p = 0.001; Figure 2A) than for the left-sided PVs. The CF variability within each lesion was 19% higher in the right-sided veins than in the left-sided veins (p = 0.001; Figure 2B).
Contact force analysis by steerable sheath
Baseline patient characteristics were similar in both sheath groups except for a significantly higher body mass index (BMI) (p = 0.01) and female population (p = 0.03) in the SureFlex™ group than in the Agilis™ group (Table 1). A total of 1,354 lesions were analyzed in the Agilis™ group and 1,803 lesions in the SureFlex™ group. There was no significant difference in the mean CF between the Agilis™ sheath and the SureFlex™ sheath at the overall procedure level, as well as around each PV. However, the CF variability was 13% lower with the SureFlex™ sheath over the entire procedure compared to the Agilis™ sheath (p = 0.043; Figure 3A). Further analysis showed a trend of 13% to 14% lower CF variability in the LSPV and RSPV with the SureFlex™ sheath than the Agilis™ sheath (significant at the α = 0.1 level; Figure 3B).
Table 1:
Characteristics | SureFlex™ (n = 15) | Agilis™ (n = 15) | p-value |
---|---|---|---|
Age, average, years | 60.2 ± 11.9 | 64.7 ± 11.0 | 0.30 |
Male sex, % | 40 | 80 | 0.025 |
BMI, average | 35.9 ± 8.8 | 28.8 ± 3.6 | 0.01 |
Hypertension, % | 80 | 67 | 0.43 |
Diabetes, % | 0 | 13 | 0.16 |
ASCVD, % | 40 | 67 | 0.15 |
Heart failure, % | 20 | 40 | 0.25 |
Pacemaker, % | 13 | 0 | 0.16 |
Paroxysmal/persistent AF, % | 60/27 | 53/47 | 0.72/0.27 |
Ejection fraction ≤ 50%, % | 21 | 25 | 0.84 |
LVH, % | 33 | 47 | 0.47 |
LA size—volume/BSA, average, cm2 | 38.7 ± 18.7 | 39.8 ± 17.9 | 0.89 |
Mitral regurgitation, % | 73 | 67 | 0.70 |
AF: atrial fibrillation; ASCVD: atherosclerotic cardiovascular disease; BMI: body mass index; BSA: body surface area; LA: left atrial; LVH: left ventricular hypertrophy.
Inefficient contact lesions
The SureFlex™ sheath maintained better catheter–tissue contact than the Agilis™ sheath, as demonstrated by 20% overall fewer inefficient contact lesions (ie, lesions with a CF of less than 5 g for more than 10% of the ablation time; p < 0.001; Figure 4A). The odds ratio (OR) for inefficient contact lesions was 0.605 for SureFlex™ versus Agilis™ [95% confidence interval (CI): 0.371–0.976; p = 0.035], revealing a 39% lower chance of inefficient contact lesions with the SureFlex™ sheath than the Agilis™ sheath. While this reduction was consistent among all PVs, the difference between the SureFlex™ and Agilis™ sheaths was significant in the right PVs, with 17% and 45% fewer inefficient contact lesions in the RIPV and RSPV, respectively (p < 0.01; Figure 4B). The OR for the SureFlex™ sheath versus the Agilis™ sheath for RIPV was 0.607 (95% CI: 0.35–1.03; p = 0.009) and that for RSPV was 0.583 (95% CI: 0.27–1.24; p = 0.15).
Radiofrequency time per lesion
The overall RFA time per lesion was 12% (1.4 seconds per lesion) shorter with the SureFlex™ sheath when compared to the Agilis™ sheath (p = 0.002; Figure 5A). A 9% to 21% reduction in RFA time was observed in the LSPV, RIPV, and RSPV when using the SureFlex™ sheath compared to the Agilis™ sheath (p < 0.05; Figure 5B).
Procedural outcomes
PVI was successfully achieved in all 30 patients with paroxysmal (57%) and persistent (37%) AF. No major procedure-related adverse events, such as cardiac tamponade, stroke, or esophageal injury, occurred.
Discussion
Several steerable sheaths are commercially available but their relative influence on CF and catheter stability remains unknown. In this study, a retrospective evaluation of data from 30 patients who underwent PVI procedures using two different steerable sheaths indicated that procedural efficiency measures such as CF stability and RF time per lesion may be impacted by the choice of sheath. A consistent ablation strategy was used whereby PVs were ablated in sequence, with the left veins before the right veins, suggesting that any observed trends are attributed to both anatomy and/or sheath fatigue over time. The overall CF achieved during the ablation of the right-sided PVs was higher than the CF achieved for the left-sided PVs. Greater CF variability reached statistical significance between the right PVs and left PVs, but this difference trended lower when using the SureFlex™ sheath than the Agilis™ sheath. Both steerable sheaths in this study achieved a similar mean CF, ruling out potential procedural bias. Further, despite more extensive operator experience with the Agilis™ sheath than the SureFlex™ sheath prior to this case series, the SureFlex™ sheath maintained an overall 13% greater CF stability than the Agilis™ sheath, with similar trends in individual PVs.
Maintaining a minimum CF throughout the RF application is necessary for effective lesions and improving ablation outcomes,13,14 whereby insufficient tissue CF has been correlated with inadequate lesion formation15–17 and a higher rate of AF recurrence.13,18,19 The SureFlex™ sheath led to significantly fewer lesions with poor contact (< 5 g for more than 80% of the ablation time). This trend was further pronounced in the RIPV and RSPV, which are typically more difficult to navigate. Reduced CF stability may necessitate longer RFA to reach the desired lesion endpoint, potentially reducing the efficiency and increasing the risk of coagulum formation and steam pops.5,16,20,21 Although the total number of ablation lesions was greater in the SureFlex™ group, the RF time was assessed per lesion, targeting the same 10-Ω impedance drop. To account for the difference in ablation lesions, statistical analysis using a mixed-effects model was used. The RF time per lesion was 12% shorter in the SureFlex™ group than in the Agilis™ group, suggesting improved RF delivery to achieve the desired lesion endpoint.
To the best of our knowledge, this is the first study that compares the procedural performance of two commercially available sheaths with the goal of improving CF stability for more effective RFAs. Both the SureFlex™ and Agilis™ sheaths are bidirectional steerable sheaths, with comparable outer and inner diameters, 180° deflection clockwise, and 90° deflection counterclockwise. However, proprietary differences between these braided sheath shafts could theoretically lead to more controlled steering as well as less sheath fatigue with the SureFlex™ sheath over the duration of the procedure, thereby resulting in differences in CF variability and effective RF lesions observed in this study.
Limitations
This single-operator nonrandomized study retrospectively evaluated the procedural parameters in a small patient population, which limited the statistical power of the analyses. While a consecutive series of patients undergoing first-time ablation was selected, the BMI and the number of women were higher in the SureFlex™ group, which may have introduced procedural complexity and technical challenges. Although baseline left atrium echo parameters were comparable between the two groups, a larger percentage of women and potentially a thinner left atrial wall could have resulted in the shorter RF duration seen with SureFlex™. The operator was not blinded to the sheath being used; however, the same method of circumferential point-by-point ablation by the same operator attempts to mitigate variability between the groups. Also, a similar mean CF was achieved using both sheaths, suggesting a lack of operator bias. Future studies assessing CF stability and effective RF lesions with other steerable sheaths as well as fixed curve sheaths and their correlation to long-term clinical outcomes will be important to explore.
Conclusion
This study demonstrates that the choice of steerable sheath may affect CF variability and the effectiveness of RFA lesions.
References
- 1.Murray MI, Arnold A, Younis M, Varghese S, Zeiher AM. Cryoballoon versus radiofrequency ablation for paroxysmal atrial fibrillation: a meta-analysis of randomized controlled trials. Clin Res Cardiol. 2018;107(8):658–669. doi: 10.1007/s00392-018-1232-4. [CrossRef] [PubMed] [DOI] [PubMed] [Google Scholar]
- 2.Ganesan AN, Shipp NJ, Brooks AG, et al. Long-term outcomes of catheter ablation of atrial fibrillation: a systematic review and meta-analysis. J Am Heart Assoc. 2013;2(2):e004549. doi: 10.1161/JAHA.112.004549. [CrossRef] [PubMed] [DOI] [PMC free article] [PubMed] [Google Scholar]
- 3.Hocini M, Condie C, Stewart MT, Kirchhof N, Foell JD. Predictability of lesion durability for AF ablation using phased radiofrequency: power, temperature, and duration impact creation of transmural lesions. Heart Rhythm. 2016;13(7):1521–1526. doi: 10.1016/j.hrthm.2016.02.012. [CrossRef] [PubMed] [DOI] [PubMed] [Google Scholar]
- 4.Kumar S, Barbhaiya CR, Balindger S, et al. Better lesion creation and assessment during catheter ablation. J Atr Fibrillation. 2015;8(3):1189. doi: 10.4022/jafib.1189. [PubMed] [DOI] [PMC free article] [PubMed] [Google Scholar]
- 5.El Haddad M, Taghji P, Phlips T, et al. Determinants of acute and late pulmonary vein reconnection in contact force–guided pulmonary vein isolation: identifying the weakest link in the ablation chain. Circ Arrhythm Electrophysiol. 2017;10(4):e004867. doi: 10.1161/CIRCEP.116.004867. [CrossRef] [PubMed] [DOI] [PubMed] [Google Scholar]
- 6.Okumura Y, Watanabe I, Kofune M, et al. Effect of catheter tip-tissue surface contact on three-dimensional left atrial and pulmonary vein geometries: potential anatomic distortion of 3D ultrasound, fast anatomical mapping, and merged 3D CT-derived images. J Cardiovasc Electrophysiol. 2013;24(3):259–266. doi: 10.1111/jce.12062. [CrossRef] [PubMed] [DOI] [PubMed] [Google Scholar]
- 7.Goode JS, Jr, Taylor RL, Buffington CW, Klain MM, Schwartzman D. High-frequency jet ventilation: utility in posterior left atrial catheter ablation. Heart Rhythm. 2006;3(1):13–19. doi: 10.1016/j.hrthm.2005.09.013. [CrossRef] [PubMed] [DOI] [PubMed] [Google Scholar]
- 8.Hutchinson MD, Garcia FC, Mandel JE, et al. Efforts to enhance catheter stability improve atrial fibrillation ablation outcome. Heart Rhythm. 2013;10(3):347–353. doi: 10.1016/j.hrthm.2012.10.044. [CrossRef] [PubMed] [DOI] [PubMed] [Google Scholar]
- 9.Aizer A, Cheng AV, Wu PB, et al. Pacing mediated heart rate acceleration improves catheter stability and enhances markers for lesion delivery in human atria during atrial fibrillation ablation. JACC Clin Electrophysiol. 2018;4(4):483–490. doi: 10.1016/j.jacep.2017.12.017. [CrossRef] [PubMed] [DOI] [PubMed] [Google Scholar]
- 10.Ali A, Plettenburg DH, Breedveld P. Steerable catheters in cardiology: classifying steerability and assessing future challenges. IEEE Trans Biomed Eng. 2016;63(4):679–693. doi: 10.1109/TBME.2016.2525785. [CrossRef] [PubMed] [DOI] [PubMed] [Google Scholar]
- 11.Piorkowski C, Eitel C, Rolf S, et al. Steerable versus nonsteerable sheath technology in atrial fibrillation ablation: a prospective, randomized study. Circ Arrhythm Electrophysiol. 2011;4(2):157–165. doi: 10.1161/CIRCEP.110.957761. [CrossRef] [PubMed] [DOI] [PubMed] [Google Scholar]
- 12.Ullah W, Hunter RJ, McLean A, et al. Impact of steerable sheaths on contact forces and reconnection sites in ablation for persistent atrial fibrillation. J Cardiovasc Electrophysiol. 2015;26(3):266–273. doi: 10.1111/jce.12573. [CrossRef] [PubMed] [DOI] [PubMed] [Google Scholar]
- 13.Natale A, Reddy VY, Monir G, et al. Paroxysmal AF catheter ablation with a contact force sensing catheter: results of the prospective, multicenter SMART-AF trial. J Am Coll Cardiol. 2014;64(7):647–656. doi: 10.1016/j.jacc.2014.04.072. [CrossRef] [PubMed] [DOI] [PubMed] [Google Scholar]
- 14.Deyell MW, Wen G, Laksman Z, et al. The impact of steerable sheaths on unblinded contact force during catheter ablation for atrial fibrillation. J Interv Card Electrophysiol. 2020;57(3):417–424. doi: 10.1007/s10840-019-00514-1. [CrossRef] [PubMed] [DOI] [PubMed] [Google Scholar]
- 15.Ullah W, McLean A, Tayebjee MH, et al. Randomized trial comparing pulmonary vein isolation using the SmartTouch catheter with or without real-time contact force data. Heart Rhythm. 2016;13(9):1761–1767. doi: 10.1016/j.hrthm.2016.05.011. [CrossRef] [PubMed] [DOI] [PubMed] [Google Scholar]
- 16.Vassallo F, Cunha C, Serpa E, et al. Comparison of high-power short-duration (HPSD) ablation of atrial fibrillation using a contact force-sensing catheter and conventional technique: initial results. J Cardiovasc Electrophysiol. 2019;30(10):1877–1883. doi: 10.1111/jce.14110. [CrossRef] [PubMed] [DOI] [PubMed] [Google Scholar]
- 17.Calkins H, Kuck KH, Cappato R, et al. 2012 HRS/EHRA/ECAS expert consensus statement on catheter and surgical ablation of atrial fibrillation: recommendations for patient selection, procedural techniques, patient management and follow-up, definitions, endpoints, and research trial design: a report of the Heart Rhythm Society (HRS) task force on catheter and surgical ablation of atrial fibrillation. Developed in partnership with the European Heart Rhythm Association (EHRA), a registered branch of the European Society of Cardiology (ESC) and the European Cardiac Arrhythmia Society (ECAS); and in collaboration with the American College of Cardiology (ACC), American Heart Association (AHA), the Asia Pacific Heart Rhythm Society (APHRS), and the Society of Thoracic Surgeons (STS). Endorsed by the governing bodies of the American College of Cardiology Foundation, the American Heart Association, the European Cardiac Arrhythmia Society, the European Heart Rhythm Association, the Society of Thoracic Surgeons, the Asia Pacific Heart Rhythm Society, and the Heart Rhythm Society. Heart Rhythm. 2012;9(4):632–696.e21. doi: 10.1016/j.hrthm.2011.12.016. [CrossRef] [PubMed] [DOI] [PubMed] [Google Scholar]
- 18.Chinitz LA, Melby DP, Marchlinski FE, et al. Safety and efficiency of porous-tip contact-force catheter for drug-refractory symptomatic paroxysmal atrial fibrillation ablation: results from the SMART SF trial. Europace. 2018;20(FI_3):f392–f400. doi: 10.1093/europace/eux264. [CrossRef] [PubMed] [DOI] [PubMed] [Google Scholar]
- 19.Neuzil P, Reddy VY, Kautzner J, et al. Electrical reconnection after pulmonary vein isolation is contingent on contact force during initial treatment: results from the EFFICAS I study. Circ Arrhythm Electrophysiol. 2013;6(2):327–333. doi: 10.1161/CIRCEP.113.000374. [CrossRef] [PubMed] [DOI] [PubMed] [Google Scholar]
- 20.Nakagawa H, Kautzner J, Natale A, et al. Locations of high contact force during left atrial mapping in atrial fibrillation patients: electrogram amplitude and impedance are poor predictors of electrode-tissue contact force for ablation of atrial fibrillation. Circ Arrhythm Electrophysiol. 2013;6(4):746–753. doi: 10.1161/CIRCEP.113.978320. [CrossRef] [PubMed] [DOI] [PubMed] [Google Scholar]
- 21.Chinitz JS, Kapur S, Barbhaiya C, et al. Sites with small impedance decrease during catheter ablation for atrial fibrillation are associated with recovery of pulmonary vein conduction. J Cardiovasc Electrophysiol. 2016;27(12):1390–1398. doi: 10.1111/jce.13095. [CrossRef] [PubMed] [DOI] [PubMed] [Google Scholar]