Abstract
This paper presents the first wireless and programmable neural stimulator leveraging magnetoelectric (ME) effects for power and data transfer. Thanks to low tissue absorption, low misalignment sensitivity and high power transfer efficiency, the ME effect enables safe delivery of high power levels (a few milliwatts) at low resonant frequencies (~250 kHz) to mm-sized implants deep inside the body (30-mm depth). The presented MagNI (Magnetoelectric Neural Implant) consists of a 1.5-mm2 180-nm CMOS chip, an in-house built 4 × 2 mm ME film, an energy storage capacitor, and on-board electrodes on a flexible polyimide substrate with a total volume of 8.2 mm3. The chip with a power consumption of 23.7 μW includes robust system control and data recovery mechanisms under source amplitude variations (1-V variation tolerance). The system delivers fully-programmable bi-phasic current-controlled stimulation with patterns covering 0.05-to-1.5-mA amplitude, 64-to-512-μs pulse width and 0-to-200-Hz repetition frequency for neurostimulation.
Index Terms—: Wireless neurostimulator, implantable device, bioelectronics, magnetoelectric effect, wireless power transfer
I. Introduction
Neurostimulation holds significant promise as a tool to modulate nerves for both neuroscience research and clinical therapies. Peripheral nerve stimulation (PNS) is a common approach to treat neuropathic pain. For example, devices can be implanted to deliver electrical pulses to the spinal cord, which can help prevent pain signals from reaching to the brain [1], [2].
A fundamental challenge in developing miniature neural implants is delivering power to devices inside the body. The use of a wired power supply causes failures for neural implants, as lead wires increase the risks of infections, restrict device deployment and affect subject mobility [3], [4]. Batteries add considerable weight and increases device footprint [5]. They are also required to be replaced or recharged frequently, which limits their long-term clinical applications. Compared to wired or battery powered implants, wirelessly powered battery-free neural stimulators have the potential to provide less invasive, longer lasting interfaces to nerves. Ideally, these implants would be programmable so that they can be reconfigured to suit user needs. Fig. 1 illustrates the concept of the proposed spinal cord stimulation system. The wireless implant receives power and data from the portable battery powered transmitter (TX) via magnetic field; the microcontroller, magnetic field driver, and the battery are assembled in a wearable belt.
While various wireless neural implants exploiting radio-frequency (RF) electromagnetic (EM) [6], [7], inductive coupling [8]–[16], ultrasonic [17]–[21], and optical [22], [23] power transfer have been reported, achieving safe and reliable wireless power transfer with the size and power constraints to neural implants is still challenging. Existing technologies cannot simultaneously satisfy all the desired properties as summarized in Fig. 2. Radio-frequency EM waves are capable of delivering power to implants deep in the tissue [6], [7]. However, they wrestle with size limitations of the receiver’s antenna since efficient power delivery with electromagnetic waves requires antenna sizes comparable to the wavelength. Higher frequency RF is necessary for mm-scale implants, but suffers from higher tissue absorption [24], limiting the amount of power that can be safely delivered. Near-field inductively coupling has been well developed in wireless power transfer [8]–[16]. It has less tissue absorption than RF because of the lower operating frequency. However, its power delivery is sensitive to perturbations in the distance and angle, especially when the coil is small. Ultrasound is another promising method to further reduce implant size and body absorption [18]–[21]. Compared to inductively coupled coils, its efficiency is more robust to the source-receiver misalignment [25]. However, it must overcome significant path loss caused by reflections at the boundaries between air, bone and tissue, which have different densities and acoustic properties. To alleviate the reflection between the air and the body, ultrasound gel is typically required for the transmitter. The need for frequent replacements of the ultrasound gel and the need for the transmitter to be in contact with the skin can be inconvenient and unreliable for long-term clinical treatments. Further, the gel cannot mitigate the in-body path loss, which is 22-dB cm−1 MHz−1 in the skull [26]. Optical power delivery is also advantageous in the miniaturization of neural implants [22], [23], but it suffers energy loss due to scattering and may have difficulties in supporting the neural stimulation with higher power requirements.
Among the various modalities of wireless power transfer, low-frequency magnetic field is believed to be one of the best mechanisms to safely deliver power deep inside body, because of its low absorption and strong penetration. Recently, magneothermal deep brain stimulation using magnetic nanoparticles has been demonstrated by [27]. However, the system has limited capabilities because of difficulties in the integration with other bio-electronics, especially with CMOS based devices. An alternative emerging approach to exploit low-frequency magnetic fields is by leveraging magnetoelectric (ME) effects [28], [29]. It promises several key merits, including: (1) low tissue absorption owing to the low carrier frequency; (2) less sensitivity to changes in the alignment (in comparison with inductive coupling); and (3) high output power and efficiency with miniaturization, which will be further discussed in Section II. The first use of ME laminates for wireless neural stimulation was recently demonstrated by Singer et al. [29]; however, like some other analog neural stimulators [11], [27], this proof-of-concept lacks robust control of the stimulation patterns and thus is highly sensitive to changes of coupling between the TX and the implants. For clinical applications, the stimulation timing and amplitude must be well controlled and programmable by the user to ensure the safety and reliability of the stimulator. Furthermore, due to the lack of energy storage and charge balancing techniques, this work has difficulties in providing large-power stimulating pulses and eliminating residual charge. Therefore, there is a critical need to create wireless neural stimulators that simultaneously achieve clinical safety, miniaturization, operation reliability and flexibility, and programmable stimulation parameters.
To meet all the desired properties and circumvent problems mentioned above, we present MagNI (Magnetoelectric Neural Implant), the first untethered and programmable neural stimulator that exploits ME effects for power and data transfer, which integartes a 1.5 mm2 180-nm CMOS system-on-chip (SoC), an in-house built 4 mm × 2 mm × 0.12 mm ME transducer, a single energy storage capacitor, and 1-mm2 on-board electrodes on a flexible polyimide substrate, as shown in Fig. 3. The proposed device features: (1) a miniature physical dimension of 8.2 mm3 and 28 mg, (2) adaptive system control and data transfer mechanisms robust under source amplitude variations, (3) a 90% chip efficiency due to its low static power down to 23.7 μW, and (4) the capability to perform fully programmable bi-phasic current stimulation covering 0.05 to 1.5 mA amplitude, 64 to 512 μs pulse width, and 0 to 200 Hz frequency ranges, making it appropriate for spinal cord stimulation to treat chronic pain.
This paper is an extended version of [30], with more comprehensive analysis, discussions, and measurements on the safety, robustness, and power efficiency of the proposed ME power transfer mechanism. The rest of the paper is organized as follows: Section II presents qualitative and quantitative analysis of safety, misalignment sensitivity and efficiency for the ME power transfer for miniaturized neural implants; Section III describes the detailed design and implementation of the proposed SoC and neural stimulator; Section IV gives experimental results, including stimulation variability, charge imbalance, impedance of on-board electrodes, and power transfer efficiency; Section V concludes this paper.
II. Magnetoelectric Power and Data Transfer for Neural Implants
A. Magnetoelectric Transducers
A magnetoelectric transducer converts an AC magnetic field to AC electrical voltage. It can be achieved by mm-scale ME laminates, which consist of a piezoelectric layer (pink), which is a nickel coated lead zirconate titanate (PZT), and a Metglas, a magnetostrictive layer (blue), as shown in Fig. 4(a). The Metglas is made up of magnetic grains, whose domain boundaries shift with applied AC magnetic field, which causes the overall material to change shape. The resulting vibrations are transferred to the PZT material, which develops a voltage in response to the induced strain [31], [32]. Thus, a thin-film laminate converts an applied low-frequency AC magnetic field into a voltage across the transducer via mechanical coupling between magnetostrictive and piezoelectric materials (Fig. 4(b)). The ME transducers are fabricated as laminates of Metglas (Metglas Inc.) and PZT (APC International, Ltd.). The Metglas is cut into the desired geometry from a thin sheet, while PZT is cut into the desired shape with a razor blade from a ceramic square plate. The two thin films are then coupled together with an adhesive non-conductive epoxy. Epoxy has been shown as an economical and simple solution to tightly couple the layers together for efficient energy transfer and yielding high ME voltage coefficients [33]. The resonant frequency of the ME transducers depends on the length and thickness ratio between the two laminates [34], [35]. Thus, the specific carrier frequency of ME power link can be selected by precisely controlling the film physical dimensions. Fig. 5 shows that at resonance, PZT/Metglas ME films generate high output voltage (greater than 7 Vpp) with low resistive source impedance (~ 800 Ω), under a magnetic field of 0.57 mT, making it suitable for efficient energy harvesting.
B. Safety Analysis
Body-area wireless power transfer via high frequency electric, magnetic or EM fields can cause safety issues such as heating, due to the power deposition in the human tissues. With such concerns, the field strength and frequency need to be chosen carefully to ensure safe operations in all cases. For example, when utilizing RF field to deliver power to implants, the carrier frequencies must be at GHz to match the resonant wavelength of small antennas [6], [7], which causes high EM absorption in the body and is subject to strict safety limits. Near-field inductively powered devices mostly work at 13.56 MHz or higher frequencies to improve the quality factor of the miniaturized receiver (RX) coil for better efficiency [11]–[16], which also results in energy absorbed by the body and requires operation under certain restrictions. The ME film works at a lower resonant frequency, which is around 250 kHz in this work. Low-frequency magnetic fields (100 kHz to 1 MHz) can penetrate the body without substantial body absorption, resulting in alleviated safety limitations for power delivery [36].
To assess the safety of ME power transfer for implants deep inside the body, we analyze the specific absorption rate (SAR) and induced electric fields with respect to a exposure to magnetic fields in COMSOL. The 4-layer tissue model representing spinal cord stimulation (skin, fat, muscle and bone) is built, as shown in Fig. 6 [37]. The flat TX coil with 5.6-cm diameter, 1-mm trace width and 15 turns is employed to create the 250-kHz magnetic field. A 0.55-mT magnetic field strength is achievable, which is adequate for the ME film to generate a 7-Vpp output, at a depth of 30 mm with a safe magnetic field shown in Fig. 7. In the 4-layer tissue model, the maximum SAR is in the muscle, which is 0.3 W/kg, and the strongest electric field of 51.9 V/m is at the surface of skin, both of which satisfy the IEEE standard for humans in unrestricted environments (2 W/kg, 52.2 V/m) [24]. For the 250-kHz magnetic field, safety limits are dominant by the maximum allowed electric field, while the maximum SAR is still far below the limit.
We also test the maximum allowed magnetic field strength under the SAR limitation [24] for the frequency of 13.56 MHz. The comparison given in Table I shows that the lower frequency magnetic field can be much stronger without exceeding the safety limits. Therefore, compared to inductive coupling and RF, the ME effect has great potential to safely deliver more power to deep implants.
Table I.
Frequency | Skin | Fat | Muscle | Bone |
---|---|---|---|---|
0.25 MHz | 5.79 mT | 4.27 mT | 2.73 mT | 0.56 mT |
13.56 MHz | 0.35 mT | 0.26 mT | 0.17 mT | 0.03 mT |
In addition, due to the fact that ME wireless power transfer couples magnetic fields, we wanted to assess the potential for ME power transfer to be used with magnetic resonance imaging (MRI). A study by Shellock et al., evaluated the MRI compatibility of the microstimulator RF BION, which is an inductively powered stimulator with a ~30-mg ferrite core [38]. The results demonstrate that patients can safely undergo MRI imaging at 1.5 T with minimal heating and artifacts being an issue only if the area of interest for imaging is proximal to the implant. In our case, Metglas will also generate artifacts; however, the amount of Metglas used in our device (~ 0.3 mg), which is 10% of the amount of ferrous material in BION, will not pose as a hazard to MRI patients implanted with the device. While not fully MRI compatible as imaging artifacts may need to be considered for certain applications, ME implants should be MRI safe and clinically viable.
C. Angular Misalignment Sensitivity Analysis
Misalignment can lead to failure for most wirelessly-powered implants by negatively affecting the stability of power and data links, especially for inductively coupled approaches. The direction of the RX to the TX affects the effective area of the RX coil to catch magnetic flux, and therefore changes the output voltage and power. Theoretically, in a uniform magnetic field, the output power of the RX coil is proportional to cos2(θ), where θ is the angle rotation. A misaligned device is often unable to maintain correct functionality due to the insufficient levels of received power or data errors.
ME laminates are deformed by the alternating magnetic field with the generation of stress by the magnetostrictive layer. This stress is then transferred to the PZT material, followed by crossing voltage changes, which means its output is primarily determined by the magnetic field strength instead of flux. On the contrary to inductive coils, the ideal alignment for ME is placing the laminate in parallel with magnetic induction lines. In addition, because of the high permeability (greater than 45000), the Metglas material has significant magnetic flux concentration effects, which enhance the magnetic flux density locally [39]. We simulate the magnetic flux concentration effects of Metglas in COMSOL. Fig. 8 shows that the flux density inside the laminate is much higher than that in the free space even with angle rotations, which means the decrease in flux density due to misalignment can be partially compensated by the magnetic flux concentration effect. As a result of these unique characteristics, compared to the inductive coil, the ME transducer is less sensitive to angular misalignment.
To test the angle dependence of the ME, we measure 12-mm2 ME transducers with a 25-mm cylindrical TX coil with angle rotations. For comparison, both simulations and measurements of inductive coils are conducted with the same TX but operating at 13.56 MHz. RX coils used for the experimental test are built in boards with a diameter of 4 mm, the number of turns of 10 and a quality factor of 30.6. The simulated and experimental results of the inductive coil match well with the theoretical value. Measurement results reported by [16] are also added as a reference, whose RX coil is 10.5 × 7 mm2, 7 turns with a Q of 23.05. Fig. 9 shows a 3X maximum improvement of ME in angular misalignment robustness compared to inductive coupling.
D. Power and Efficiency Analysis
Miniaturization of the wireless powered implant can influence the overall power efficiency of the link. With far-field power transfer taking advantage of EM wave radiations, the received power is quite weak when the antenna is smaller than one-tenth of the wavelength [41]. As another example, in the near field, the voltage and power of the RX coil is also highly dependent on the size. Faraday’s Law gives the induced voltage of the coil as
(1) |
where A is the effective area of the coil, B is the magnetic flux density and N is the number of turns. By assuming that the coil is flat and circular, an estimation of the output power can be given as
(2) |
where R is the diameter of the coil and μ is the permeability. Shrinking down the coil size not only reduces the amount of captured flux, but also degrades the quality factor, which causes a significant decrease in the received power.
Previous studies [34], [35], [42] developing the equivalent circuit model of ME laminated composites have shown that its voltage coefficient α (defined by dV/dH with units of V/Oe) is independent of width and length, which means scaling down the ME laminate area does not reduce the output voltage. However, the resonant frequency is inversely proportional to the length while the source impedance at resonance is inversely proportional to the width. As a result, the theoretical maximum output power of the ME laminate is linearly related to its width. Interestingly, in comparison to inductive coupling, the ME received power demonstrates less dependency on the transducer size, which demonstrates the potential for superior scalability in miniaturizing the device. These phenomena are verified by simulation results given in Fig. 10 and measurements of [29].
To evaluate the efficiency of ME against inductive coupling, both of which are based on low-frequency magnetic fields, we measure the ME and inductive coupling power transfer efficiency (PTE) across TX-RX distances in air (Fig. 11) using custom built 12-mm2 ME transducers and 12-mm2 10-turn coils on PCB. As expected, PTEs of both modalities decrease as the distance increases, but ME consistently presents a higher PTE across the range and a maximum efficiency of 1.1% is achieved when the ME transducer is at the center of the TX coil. Additionally, reported PTE of various miniaturized devices [7], [8], [16], [21], [43] are included in Fig. 11 for comparison. Among them, [7], [40] utilize RF, [8], [16] adopt inductive coupling power transfer and [21] establishes the power link by ultrasound. It is worth noting that the decrease of PTE at larger TX-RX distances is solely caused by the decline of magnetic field strength, the ME energy conversion efficiency (magnetic to mechanical, mechanical to electrical) that relies on the coupling between laminates depends solely on intrinsic material properties and thereby remains the same at varying distances [44].
III. Magni System-on-Chip Design
A. System Overview
The MagNI SoC, consisting of power management, data recovery, and stimulation modules, interfaces with a ME film to receive power and data and drives programmable stimulation, as shown in Fig. 12(a). The SoC cycles through charging, data receiving, and stimulation phases at the desired stimulation frequency set by the external transmitter (Fig. 12(b)).
1). Power Management:
AC voltage induced on the ME transducer is first converted to a DC voltage (Vrect) by the active rectifier (Fig. 13(a)) [45]. Energy is then stored in an off-chip 4.7-μF capacitor for the high-power stimulation. The low-dropout regulator (LDO) provides a constant 1-V supply VLDO for digital control and data transfer to reduce power consumption. Temperature and supply-invariant reference voltages for the entire system are generated on chip by a reference circuit with native NMOS and stacked diode-connected PMOS transistors [46]. Three different voltage references of 1 V, 0.6 V and 0.3 V are provided by the ultra low power circuit shown in Fig. 13(b).
2). Data Recovery:
Downlink data is transmitted by amplitude shift keying (ASK) modulation of the magnetic field and is used to program stimulation patterns. To save energy, we reuse the rectifier to detect data amplitudes. However, the large energy storage capacitor prohibits the rectified voltage change in a short time. To solve this problem, a dual path architecture is adopted, which has an auxiliary path enabled by a global control signal ENdata for data communication.
3). Stimulation:
The 5-bit current DAC, which is only enabled in the stimulation phase to save power, programs stimulation amplitudes with a 50-μA resolution. A finite-state machine (FSM), also configured by the received data, controls the stimulation timing. Its operating phases include initialization, anodic stimulus, inter-phasic pause, cathodic stimulus and electrodes shorting. Bi-phasic stimulation, unlike a mono-phasic pulse, is charge-balanced and prevents undesired electrochemical reactions on electrodes. It is realized by the H-bridge based stimulation driver controlled by Φ1 and Φ2. The inter-phase pause with a fixed duration of 32 μs allows the stimulation driver to stabilize after the phase switching and reduces the threshold for bi-phasic stimulation [47]. By the end of stimulation, electrodes are shorted by ENshort to completely remove the residual charge.
B. Design Considerations for Robust Operation
In order to maintain system operation robustness under varying conditions, such as amplitude changes due to varying distances and misalignment between device and transmitter, we make the device operation adaptive, calibration free, and fully controllable by the TX. Fig. 14 shows the proposed architecture to realize the robust operating scheme.
First, the enabling of the data transfer phase is controlled by changing the ME amplitude, so that scheduling of the implant is fully controlled by the external TX with accurate timing references and computation resources. Because the large energy storage capacitor prevents Vrect to quickly change, the existing comparator in the active rectifier is employed as a watchdog to monitor the ME induced voltage change. It generates a train of pulses WD tracking Vrect during charging phase and stops once the input’s amplitude is below the voltage of the energy capacitor. By comparing the watchdog signal with divided Vrect and extracting the envelop, a rail-to-rail start signal for data transfer (ENdata) is created, as shown in Fig. 15. Comparator offset may cause a failure in tracking Vrect and generate incorrect ENdata (Fig. 16(a)). To alleviate this, 4-input comparator [45] is employed to double the sensing margin and increase offset tolerance from 11mV to 23mV (Fig. 16(b)), effectively reducing the failure probability from 8% to 0.02% (calculated with simulated variance of offset).
Second, the system global clock is derived directly from the ME source by a low-power comparator, and therefore presents a process, supply and temperature-invariant frequency. The Monte Carlo simulation result given in Fig. 17(a) demonstrates a tight distribution of duty cycle with σ of 0.159%.
Third, to ensure correct data demodulation under ME voltage variations, the voltage threshold is generated online before every data transfer, using an alternating pilot tone sent by the TX. The threshold is extracted with a low-pass filter followed by track and hold circuits. Measurement results show successful data recovery without errors with a wide source amplitude range (minimum of 0.6 V for data “1” and minimum of 0.4 V for data “0”) and 0.2-V minimum amplitude difference (Fig. 17(b)). Furthermore, we validated the robust operation by showing that a fixed setting of the stimulation current stays around the expected 1.5 mA, when the ME-induced voltage varies from 2.6 to 3.6 V due to magnetic field variations (Fig. 17(c)). Maximum Istim falls slightly with the decrease of source amplitude reducing voltage headroom, until data recovery fails when the Vs is below 2.5 V.
IV. Experimental Results
A. Electrical Validation of the MagNI SoC
The SoC of the implant is fabricated in 180-nm CMOS technology, as shown in Fig. 18. To validate the functionalities of the proposed neural stimulating system, the device is tested with the ME film in air, and the AC magnetic field is generated by a portable magnetic field driver (Fig. 19).
Fig. 20 shows the measured waveforms of MagNI operating at different stimulation frequencies. Since transitions between operating phases are completely controlled by the TX, the frequency can be configured from 0 to 200 Hz with practically infinite resolution. As shown in Fig. 21, Vrect peaks at 2.95 V with an 83% voltage conversion ratio at 0.61-mT magnetic flux density in the charging phase and drops to 2.55 V after the 1.5-mA, 512-μs stimulation. ASK modulation, which is actually a frequency modulation at TX, is realized by slightly shifting the magnetic field frequency, which causes the ME film to vibrate off resonance with reduced amplitude. Considering the settling time of the ME film and the fact that high data rate is not demanded in this application, 32 cycles are required to reliably transmit one bit, resulting in a 7.8-kbps data rate.
Additionally, the waveform of bi-phasic stimulation current with various amplitudes is given in Fig. 22(a). Total stimulation charge at different settings with error bars indicating the variability, whose maximum is 2% of 1.5-mA amplitude, are shown in Fig. 22(b). While the variability of stimulation caused by supply voltage variations and current source settling is observed, the error of the total charge deposited to the neural tissue is negligible. Because stimulation effects are believed to mostly depend on the total charge deposited to the neural tissue [48], we believe the temporal stimulation amplitude variation will not affect the stimulation efficacy. The asymmetry of bi-phasic stimulation may introduce charge imbalance. Total charge of the measured bi-phasic stimulation with different amplitudes and pulse widths is calculated, which shows the worst charge imbalance of 6.5 nC when the stimulation amplitude is 1.5 mA, 512 μs (Fig. 22(c)). However, it should be noted that the electrodes are shorted after each stimulation, which removes the residual charge on the tissue and thus ensures the stimulation safety.
B. In-Vitro Experiments
To demonstrate the compatibility for implantation, a soak test is performed by immersing a MagNI in phosphate buffered saline (PBS), as illustrated in Fig. 23. The ME film is packaged in a 3D-printed 0.4-mm thick enclosure to eliminate environmental influences on its mechanical vibration before the entire implant is encapsulated with non-conductive epoxy.
The on-board electrodes are tested in PBS to evaluate the voltage compliance of the device. Fig 24 shows that the measured impedance of the bare electrodes with gold surface is 2100 Ω at 2 kHz (associated with the 500-μs stimulation). After a porous platinum coating, the impedance drops to 170 Ω [49]. Thus, the stimulation supply voltage, which is up to 3.3 V, is abundant to support the maximum amplitude of 1.5 mA. The order-of-magnitude impedance reduction by electrodes coating eliminates the needs of high-voltage circuits and improves the stimulation efficiency.
During start-up (Fig. 25(a)), the device is charged up to 2.6 V in 44 ms with 2.22-mW peak harvested power Pin,max and 0.435-% maximum power transfer efficiency, when at the center of the TX coil (Table II). Considering the 1-V source amplitude tolerance of the proposed implant, voltage drop with distance increase is allowed here to slow down the increment of TX power consumption and improve the overall efficiency. At a distance of 30 mm, Vrect of 2.05 V and Pin,max of 1.35 mW are achieved, which are sufficient to ensure SoC functionality. Pin,max is estimated based on the measured rectified voltage Vrect, as given by
(3) |
Fig. 25(c) shows the measured stimulation in saline with uncoated electrodes, which is 1.5 mA and 400 μs causing a voltage crossing electrodes of 2.53 V. Cathodic pulses compensate the built-in charge and residual charge is removed by shorting electrodes after stimulation. Effective charging-balance is realized by the bi-phasic stimulation shape and electrodes shorting phase.
Table II.
Distance (mm) | Pdriver (W) | Magnetic Flux Density (mT) | Vrect (V) | Pin,max (mW) | ηmax (%) |
---|---|---|---|---|---|
0 | 0.51 | 0.62 | 2.63 | 2.22 | 0.435 |
7.5 | 1.2 | 0.66 | 2.61 | 2.16 | 0.180 |
20 | 1.4 | 0.48 | 2.31 | 1.67 | 0.119 |
30 | 2.1 | 0.13 | 2.05 | 1.35 | 0.064 |
In addition to testing in PBS, the device is also tested in air and agar, a substrate used to emulate the brain’s mechanical properties, showing Vrect and Pin,max variations less than 0.14 V and 0.22 mW respectively when at the center of the TX coil (Fig. 26(a)), demonstrating the ME effect’s adaptability in different mediums. Throughout the one-week soak test, the device in PBS functions consistently with Pin,max fluctuating between 2.16-to-2.25 mW, indicating its long-term reliability for implantation (Fig. 26(b)).
C. Hydra Stimulation Tests
To further assess MagNI’s bio-stimulation capability, the device is validated using Hydra vulgaris as a model for excitable tissue [50] [51]. The transgenic Hydra strains used express GCaMP6s, a calcium-sensitive fluorescent protein, in the ectoderm, and naturally express voltage sensitive ion channels. Fig. 27 gives the diagram of the experimental setup using an inverted Nikon microscope with a 20x objective (NA = 0.75) and a constant excitation light of 460 nm. Fluorescence images are captured at 25 frames per second with 40 ms exposures and 2 × 2 binning using an Andor Zyla sCMOS camera. In order to facilitate simultaneous stimulation and fluorescent imaging of Hydra, stereotrodes are used as the stimulating electrodes as they are proportional to the size of our chosen animal model and can more easily provide targeted stimulation to the area of interest. A micromanipulator is used to position the stereotrode connected to the proposed device that is remotely powered by the ME effect (10 mm from the TX); a blue laser is used to image GCaMP6s activity in Hydra.
When we apply 5-sec biphasic pulse trains with 500-μs pulse widths at 100 Hz, we observe an increase larger than 200% in GCaMP6s fluorescence, which is consistent with electrical activation of voltage gated ion channels in Hydra tissue that result in muscle contractions, as shown in Fig. 28.
V. Conclusion
This work presents a proof-of-concept demonstration of MagNI, a magnetoelectrically powered and controlled neural implant. ME power link proves to be a viable technology to wirelessly deliver greater than 1 mW power to mm-sized devices deeply implanted in the body at 30-mm depth. The proposed spinal cord stimulator for neuropathic pain relief features miniaturized physical dimensions of 8.2 mm3 and 28 mg, adaptive operation and data transfer mechanisms (1-V source variation tolerance), and programmable bi-phasic current stimulation capability fully covering 0.05 to 1.5-mA amplitude, 64 to 512-μs pulse width, and 0 to 200-Hz repetition range. Table III summarizes the specifications of MagNI and compares with state-of-the-art mm-scale wireless neural stimulators. The use of low-frequency magnetic field (250 kHz) substantially alleviates tissue absorption and reflections in wireless power transfer, leading to higher deliverable power under safety limits (in Section II). Thanks to the high power transfer efficiency and chip efficiency, the proposed system achieves one of the highest 391-nC total stimulation charges among the comparisons, with 0.435% and 0.064% end-to-end efficiency at 0 and 30mm distance respectively. In addition, ME power link features robustness against alignment changes (20% less power with 50 degree angular mismatch between TX and implants).
Table III.
This Work | Neuron′17 [11] | ISSCC′18 [13] | TBioCAS′18 [14] | TBioCAS′19 [15] | Nat. Biomed. Eng.′20 [21] | |
---|---|---|---|---|---|---|
Application Target | Spinal Cord | Brain | Brain | PNS | Brain | PNS |
Process (nm) | 180 | Discrete | 350 | 180 | 130 | 65 |
Wireless Link | Magnetoelectric | Inductive | Inductive | Inductive | Inductive | Ultrasonic |
Carrier Freq. (MHz) | 0.25 | 13.56 | 60 | 198 | 1180 | 1.85 |
Implant Volume (mm3) | 8.2 | 98 | 12.2 | 97.5 | 0.009 | 1.7 |
Stimulation Mode |
Biphasic
Current |
Optical | Optical | Monophasic Voltage | Monophasic Voltage | Monophasic Current |
Max. Stim. Current (mA) / Resolution (bit) | 1.5 / 5 | N/A | N/A | 0.25 / Fixed | 0.04 / Fixed | 0.4 / 3 |
Max. Stimulation Charge (nC) | 391 | N/A | N/A | 17.5 | 0.66 | 157 |
Stimulating Channels | 1 | 1 | 16 | 1 | 1 | 1 |
SoC Power (μW) | 23.7 | N/A | 300 | 2.7 | N/A | 4 |
Max. Overall Efficiency (%) @ Distance (mm) |
0.435 @ 0
0.064 @ 30 |
N/A | N/A | N/A | 0.0019 @ 6.6 | 0.06 @ 18 |
Max. Operating Depth (mm) | 30 | 40 | 7 | 140 | 6.6 | 55 |
In-Vivo Test Model | Hydra | Rat | Rat | Rat | Rat | Rat |
Acknowledgment
The authors would like to thank Christophe Dupre and Dr. Rafael Yuste for sharing the transgenic Hydra strains. We would also like to thank Soonyoung Kim for the culturing and handling of Hydra and useful discussions of Hydra biology.
This work was supported in part by the National Institute of Biomedical Imaging and Bioengineering of the National Institutes of Health under award U18EB029353 and the National Science Foundation under award ECCS-2023849.
Biographies
Zhanghao Yu received the B.E. degree in Integrated Circuit Design and Integrated System from University of Electronic Science and Technology of China, Chengdu, China, in 2016, and the M.S. degree in Electrical Engineering from University of Southern California, Los Angeles, CA, in 2018. He is currently working toward his Ph.D. degree in Electrical and Computer Engineering at Rice University, Houston, TX.
His research interests include analog and mixed-signal integrated circuits design for power management, bio-electronics and security.
Joshua Chen received the B.S. degree in Bioengineering from the University of California, Berkeley in 2016. During his time at Berkeley, he was a member of the Berkeley Sensors and Actuators Center (BSAC) where he worked on developing technology for biomedical sensors and 3D printed microfluidics. After graduation, he worked at Verily Life Sciences prior to attending Rice University in Houston, TX to pursue the Ph.D. degree in Bioengineering.
His research interests are in neural engineering, wireless devices, biomedical implants, MEMS devices.
Fatima Alrashdan received a B.S and M.S degrees in Electrical Engineering from Jordan University of Science and Technology, Irbid, Jordan, in 2016 and 2018, respectively. She is currently working toward the Ph.D. degree in Electrical and Computer Engineering at Rice University, Houston, TX, USA.
Her research interests include bioelectronics, wireless-implantable biomedical systems, power electronics and control systems for biomedical applications.
Benjamin Avants received a B.S. in Electrical Engineering, a B.S. in Computer Engineering, and second majors in Math and Physics from the University of Memphis, Tennessee in 2012. He then joined the Robinson Lab at Rice University, where he has worked as a research engineer ever since.
Apart from electronics and programming, his interests also include 3D modeling, optics, imaging, and advanced fabrication on the nano, micro, and meso scale. He enjoys developing researchers’ dreams into realities whether it’s in the clean room or a maker space.
Yan He received the B.S degree in Electronic Science and Technology from the Zhejiang University, Hangzhou, China, in 2018. He is currently pursuing the Ph.D. degree in Electrical and Computer Engineering with the Rice University, Houston, TX, USA.
His current research interests include analog and mixed-signal integrated circuits design for power management and hardware security.
Amanda Singer received B.S. degree in Physics in 2014 from Saginaw Valley State University, University Center, MI, USA, and a M.A.Sc. degree in Applied Physics from Rice University, Houston, TX, USA, in 2018, and is currently pursuing a Ph.D. degree at Rice University, Houston, TX, USA.
Her current research interests include the development of wireless bioelectronics and the miniaturization of neural implants.
Jacob Robinson is an Associate Professor in Electrical Computer Engineering and Bioengineering at Rice University and an Adjunct Assistant Professor in Neuroscience at Baylor College of Medicine. Dr. Robinson earned a B.S. in Physics from UCLA and a Ph. D. in Applied Physics from Cornell. Following his Ph. D., he worked as a postdoctoral fellow in the Chemistry Department at Harvard University. Dr. Robinson joined Rice University in 2012 where he currently works on nanoelectronic, nanophotonic, and nanomagnetic technologies to manipulate and measure brain activity. Dr. Robinson is an IEEE Senior member and currently a co-chair of the IEEE Brain Initiative. He is also the recipient of an NSF NeuroNex Innovation Award, DARPA Young Faculty Award, and Materials Today Rising Star Award.
Kaiyuan Yang (S’13-M’17) received the B.S. degree in Electronic Engineering from Tsinghua University, Beijing, China, in 2012, and the Ph.D. degree in Electrical Engineering from the University of Michigan, Ann Arbor, MI, in 2017. His Ph.D. research was recognized with the 2016–2017 IEEE Solid-State Circuits Society (SSCS) Predoctoral Achievement Award.
He is an Assistant Professor of Electrical and Computer Engineering at Rice University, Houston, TX. His research interests include digital and mixed-signal circuits for secure and low-power systems, hardware security, and circuit/system design with emerging devices. Dr. Yang received the Distinguished Paper Award at the 2016 IEEE International Symposium on Security and Privacy (Oakland), the Best Student Paper Award (1st place) at the 2015 IEEE International Symposium on Circuits and Systems (ISCAS), the Best Student Paper Award Finalist at the 2019 IEEE Custom Integrated Circuits Conference (CICC), and the 2016 Pwnie Most Innovative Research Award Finalist.
References
- [1].Shealy CN, Mortimer JT, and Reswick JB, “Electrical Inhibition of Pain by Stimulation of the Dorsal Columns: Preliminary Clinical Report,” Anesthesia & Analgesia, vol. 46, pp. 489–491, July 1967. [PubMed] [Google Scholar]
- [2].Verrills P, Sinclair C, and Barnard A, “A review of spinal cord stimulation systems for chronic pain,” J Pain Res, vol. 9, pp. 481–492, July 2016. [DOI] [PMC free article] [PubMed] [Google Scholar]
- [3].Hargreaves DG, Drew SJ, and Eckersley R, “Kirschner Wire Pin Tract Infection Rates: A randomized controlled trial between percutaneous and buried wires,” The Journal of Hand Surgery: British & European Volume, vol. 29, pp. 374–376, August. 2004. [DOI] [PubMed] [Google Scholar]
- [4].Biran R, Martin DC, and Tresco PA, “The brain tissue response to implanted silicon microelectrode arrays is increased when the device is tethered to the skull,” Journal of Biomedical Materials Research Part A, vol. 82A, no. 1, pp. 169–178, 2007. [DOI] [PubMed] [Google Scholar]
- [5].Pinnell RC, Pereira de Vasconcelos A, Cassel JC, and Hofmann UG, “A Miniaturized, Programmable Deep-Brain Stimulator for Group-Housing and Water Maze Use,” Front. Neurosci, vol. 12, 2018. [DOI] [PMC free article] [PubMed] [Google Scholar]
- [6].Montgomery KL, Yeh AJ, et al. , “Wirelessly powered, fully internal optogenetics for brain, spinal and peripheral circuits in mice,” Nature Methods, vol. 12, pp. 969–974, October. 2015. [DOI] [PMC free article] [PubMed] [Google Scholar]
- [7].Leung VW, Lee J, Li S, et al. , “A CMOS Distributed Sensor System for High-Density Wireless Neural Implants for Brain-Machine Interfaces,” in ESSCIRC 2018 - IEEE 44th European Solid State Circuits Conference (ESSCIRC), pp. 230–233, September. 2018. [Google Scholar]
- [8].Biederman W, Yeager DJ, Narevsky N, et al. , “A Fully-Integrated, Miniaturized 0.125 mm2 10.5 μW Wireless Neural Sensor,” IEEE Journal of Solid-State Circuits, vol. 48, pp. 960–970, April. 2013. [Google Scholar]
- [9].Lee H-M, Kwon KY, Li W, and Ghovanloo M, “A Power-Efficient Switched-Capacitor Stimulating System for Electrical/Optical Deep Brain Stimulation,” IEEE Journal of Solid-State Circuits, vol. 50, pp. 360–374, January. 2015. [Google Scholar]
- [10].Lo Y-K, Chang C-W, Kuan Y-C, et al. , “A 176-channel 0.5 cm3 0.7 g wireless implant for motor function recovery after spinal cord injury,” in 2016 IEEE International Solid-State Circuits Conference (ISSCC), pp. 382–383, January. 2016. [DOI] [PMC free article] [PubMed] [Google Scholar]
- [11].Shin G, Gomez AM, Al-Hasani R, et al. , “Flexible Near-Field Wireless Optoelectronics as Subdermal Implants for Broad Applications in Optogenetics,” Neuron, vol. 93, pp. 509–521.e3, February. 2017. [DOI] [PMC free article] [PubMed] [Google Scholar]
- [12].Freeman DK, O’Brien JM, Kumar P, et al. , “A Sub-millimeter, Inductively Powered Neural Stimulator,” Front. Neurosci, vol. 11, 2017. [DOI] [PMC free article] [PubMed] [Google Scholar]
- [13].Jia Y, Mirbozorgi SA, Lee B, et al. , “A mm-sized free-floating wirelessly powered implantable optical stimulating system-on-a-chip,” in 2018 IEEE International Solid - State Circuits Conference - (ISSCC), pp. 468–470, February. 2018. [Google Scholar]
- [14].Lyu H, Wang J, La J-H, Chung JM, and Babakhani A, “An Energy-Efficient Wirelessly Powered Millimeter-Scale Neurostimulator Implant Based on Systematic Codesign of an Inductive Loop Antenna and a Custom Rectifier,” IEEE Transactions on Biomedical Circuits and Systems, vol. 12, pp. 1131–1143, October. 2018. [DOI] [PubMed] [Google Scholar]
- [15].Khalifa A, Liu Y, Karimi Y, et al. , “The Microbead: A 0.009 mm3 Implantable Wireless Neural Stimulator,” IEEE Transactions on Biomedical Circuits and Systems, vol. 13, pp. 971–985, October. 2019. [DOI] [PubMed] [Google Scholar]
- [16].Burton A, Obaid SN, Vázquez-Guardado A, et al. , “Wireless, battery-free subdermally implantable photometry systems for chronic recording of neural dynamics,” Proceedings of the National Academy of Sciences, vol. 117, pp. 2835–2845, February. 2020. [DOI] [PMC free article] [PubMed] [Google Scholar]
- [17].Charthad J, Weber MJ, Chang TC, and Arbabian A, “A mm-Sized Implantable Medical Device (IMD) With Ultrasonic Power Transfer and a Hybrid Bi-Directional Data Link,” IEEE Journal of Solid-State Circuits, vol. 50, pp. 1741–1753, August. 2015. [Google Scholar]
- [18].Seo D, Neely RM, Shen K, et al. , “Wireless Recording in the Peripheral Nervous System with Ultrasonic Neural Dust,” Neuron, vol. 91, pp. 529–539, August. 2016. [DOI] [PubMed] [Google Scholar]
- [19].Charthad J, Chang TC, Liu Z, et al. , “A mm-Sized Wireless Implantable Device for Electrical Stimulation of Peripheral Nerves,” IEEE Transactions on Biomedical Circuits and Systems, vol. 12, pp. 257–270, April. 2018. [DOI] [PubMed] [Google Scholar]
- [20].Ghanbari MM, Piech DK, Shen K, et al. , “A Sub-mm3 Ultrasonic Free-Floating Implant for Multi-Mote Neural Recording,” IEEE Journal of Solid-State Circuits, vol. 54, pp. 3017–3030, November. 2019. [Google Scholar]
- [21].Piech DK, Johnson BC, Shen K, et al. , “A wireless millimetre-scale implantable neural stimulator with ultrasonically powered bidirectional communication,” Nature Biomedical Engineering, vol. 4, pp. 207–222, February. 2020. [DOI] [PubMed] [Google Scholar]
- [22].Lee S, Cortese AJ, Gandhi AP, et al. , “A 250 μm × 57 μm Microscale Opto-electronically Transduced Electrodes (MOTEs) for Neural Recording,” IEEE Transactions on Biomedical Circuits and Systems, vol. 12, pp. 1256–1266, December. 2018. [DOI] [PMC free article] [PubMed] [Google Scholar]
- [23].Lim J, Moon E, Barrow M, et al. , “A 0.19 × 0.17 mm2 Wireless Neural Recording IC for Motor Prediction with Near-Infrared-Based Power and Data Telemetry,” in 2020 IEEE International Solid- State Circuits Conference - (ISSCC), pp. 416–418, February. 2020. [DOI] [PMC free article] [PubMed] [Google Scholar]
- [24].“IEEE Standard for Safety Levels with Respect to Human Exposure to Electric, Magnetic, and Electromagnetic Fields, 0 Hz to 300 GHz,” tech. rep, IEEE. ISBN: 9781504455480. [Google Scholar]
- [25].Denisov A and Yeatman E, “Ultrasonic vs. Inductive Power Delivery for Miniature Biomedical Implants,” in 2010 International Conference on Body Sensor Networks, pp. 84–89, June 2010. [Google Scholar]
- [26].Hoskins PR, Martin K, and Thrush A, Diagnostic Ultrasound: Physics and Equipment. Cambridge University Press, June 2010. [Google Scholar]
- [27].Chen R, Romero G, et al. , “Wireless magnetothermal deep brain stimulation,” Science, vol. 347, pp. 1477–1480, March. 2015. [DOI] [PubMed] [Google Scholar]
- [28].Zaeimbashi M, Lin H, Dong C, et al. , “NanoNeuroRFID: A Wireless Implantable Device Based on Magnetoelectric Antennas,” IEEE Journal of Electromagnetics, RF and Microwaves in Medicine and Biology, vol. 3, pp. 206–215, September. 2019. [Google Scholar]
- [29].Singer A, Dutta S, Lewis E, et al. , “Magnetoelectric Materials for Miniature, Wireless Neural Stimulation at Therapeutic Frequencies,” Neuron, June 2020. [DOI] [PMC free article] [PubMed] [Google Scholar]
- [30].Yu Z, Chen JC, Avants BW, et al. , “An 8.2 mm3 Implantable Neurostimulator with Magnetoelectric Power and Data Transfer,” in 2020 IEEE International Solid- State Circuits Conference - (ISSCC), pp. 510–512, February. 2020. [Google Scholar]
- [31].Dong S, Li J-F, and Viehland D, “Longitudinal and transverse magnetoelectric voltage coefficients of magnetostrictive/piezoelectric laminate composite: theory,” IEEE Transactions on Ultrasonics, Ferroelectrics, and Frequency Control, vol. 50, pp. 1253–1261, October. 2003. [DOI] [PubMed] [Google Scholar]
- [32].Zhai J, Dong S, Xing Z, Li J, and Viehland D, “Giant magnetoelectric effect in Metglas/polyvinylidene-fluoride laminates,” Applied Physics Letters, vol. 89, p. 083507, August. 2006. [Google Scholar]
- [33].Silva M, Reis S, et al. , “Optimization of the Magnetoelectric Response of Poly(vinylidene fluoride)/Epoxy/Vitrovac Laminates,” ACS Applied Materials & Interfaces, vol. 5, pp. 10912–10919, November. 2013. [DOI] [PubMed] [Google Scholar]
- [34].Dong S and Zhai J, “Equivalent circuit method for static and dynamic analysis of magnetoelectric laminated composites,” Chin. Sci. Bull, vol. 53, pp. 2113–2123, July 2008. [Google Scholar]
- [35].Zhou J-P, Ma Y-J, Zhang G-B, and Chen X-M, “A uniform model for direct and converse magnetoelectric effect in laminated composite,” Applied Physics Letters, vol. 104, p. 202904, May 2014. [Google Scholar]
- [36].Young J, Wang M-T, and Brezovich I, “Frequency/depth-penetration considerations in hyperthermia by magnetically induced currents,” Electronics Letters, vol. 16, pp. 358–359, May 1980. [Google Scholar]
- [37].Hasgall P, Neufeld E, Gosselin M-C, Klingenböck A, and Kuster N, “ITIS Database for thermal and electromagnetic parameters of biological tissues, Version 4.0,” May 2015. [Google Scholar]
- [38].Shellock FG, Cosendai G, Park S-M, and Nyenhuis JA, “Implantable Microstimulator: Magnetic Resonance Safety at 1.5 Tesla,” Investigative Radiology, vol. 39, pp. 591–599, October. 2004. [DOI] [PubMed] [Google Scholar]
- [39].Fang Z, Lu SG, Li F, et al. , “Enhancing the magnetoelectric response of Metglas/polyvinylidene fluoride laminates by exploiting the flux concentration effect,” Applied Physics Letters, vol. 95, p. 112903, September. 2009. [Google Scholar]
- [40].Chen X, Yeoh WG, Choi YB, Li H, and Singh R, “A 2.45-GHz Near-Field RFID System With Passive On-Chip Antenna Tags,” IEEE Transactions on Microwave Theory and Techniques, vol. 56, pp. 1397–1404, June 2008. [Google Scholar]
- [41].Landgren DW, Cook KR, Dykes DJP, et al. , “A wideband mmWave antenna element with an unbalanced feed,” in 2017 IEEE National Aerospace and Electronics Conference (NAECON), pp. 209–212, June 2017. [Google Scholar]
- [42].Truong BD and Roundy S, “Experimentally validated model and power optimization of a magnetoelectric wireless power transfer system in free-free configuration,” Smart Materials and Structures, May 2020. [Google Scholar]
- [43].Kuo N-C, Zhao B, and Niknejad AM, “Equation-Based Optimization for Inductive Power Transfer to a Miniature CMOS Rectenna,” IEEE Transactions on Microwave Theory and Techniques, vol. 66, pp. 2393–2408, May 2018. [Google Scholar]
- [44].Truong BD, “Fundamental Issues in Magnetoelectric Transducers: Magnetic Field Sensing Versus Wireless Power Transfer Systems,” IEEE Sensors Journal, vol. 20, pp. 5322–5328, May 2020. [Google Scholar]
- [45].Lam Y-H, Ki W-H, and Tsui C-Y, “Integrated Low-Loss CMOS Active Rectifier for Wirelessly Powered Devices,” IEEE Transactions on Circuits and Systems II: Express Briefs, vol. 53, pp. 1378–1382, December. 2006. [Google Scholar]
- [46].Seok M, Kim G, et al. , “A Portable 2-Transistor Picowatt Temperature-Compensated Voltage Reference Operating at 0.5 V,” IEEE Journal of Solid-State Circuits, vol. 47, pp. 2534–2545, October. 2012. [Google Scholar]
- [47].Gorman PH and Mortimer JT, “The Effect of Stimulus Parameters on the Recruitment Characteristics of Direct Nerve Stimulation,” IEEE Transactions on Biomedical Engineering, vol. BME-30, pp. 407–414, July 1983. [DOI] [PubMed] [Google Scholar]
- [48].Cogan SF, “Neural Stimulation and Recording Electrodes,” Annual Review of Biomedical Engineering, vol. 10, no. 1, pp. 275–309, 2008. [DOI] [PubMed] [Google Scholar]
- [49].Fan B, Rodriguez AV, Vercosa DG, et al. , “Sputtered porous Pt for wafer-scale manufacture of low-impedance flexible microelectrodes,” Journal of Neural Engineering, vol. 17, p. 036029, June 2020. [DOI] [PMC free article] [PubMed] [Google Scholar]
- [50].Wittlieb J, Khalturin K, et al. , “Transgenic Hydra allow in vivo tracking of individual stem cells during morphogenesis,” Proceedings of the National Academy of Sciences, vol. 103, pp. 6208–6211, April. 2006. [DOI] [PMC free article] [PubMed] [Google Scholar]
- [51].Szymanski JR and Yuste R, “Mapping the Whole-Body Muscle Activity of Hydra vulgaris,” Current Biology, vol. 29, pp. 1807–1817.e3, June 2019. [DOI] [PubMed] [Google Scholar]