Skip to main content
Journal of Pharmaceutical Analysis logoLink to Journal of Pharmaceutical Analysis
. 2021 Jan 22;11(6):739–745. doi: 10.1016/j.jpha.2021.01.002

Automatic analytical approach for the determination of 12 illicit drugs and nicotine metabolites in wastewater using on-line SPE-UHPLC-MS/MS

Jingyuan Wang a,b,1, Likai Qi a,b,1, Chenzhi Hou a,b, Tingting Zhang b,c,d, Mengyi Chen a,b, Haitao Meng e, Mengxiang Su a,b, Hui Xu b, Zhendong Hua b,c,d,, Youmei Wang b,c,d,∗∗, Bin Di a,b,∗∗∗
PMCID: PMC8740382  PMID: 35028179

Abstract

In this study, we developed a novel on-line solid phase extraction (SPE)-ultra-high-performance liquid chromatography tandem mass spectrometry (UHPLC-MS/MS)-based analytical method for simultaneously quantifying 12 illicit drugs and metabolites (methamphetamine, amphetamine, morphine, codeine, 6-monoacetylmorphine, benzoylecgonine, 3,4-methylenedioxymethamphetamine, 3,4-methylenedioxyamphetamine, cocaine, ketamine, norketamine, and methcathinone) and cotinine (COT) in wastewater samples. The analysis was performed by loading 2 mL of the sample onto an Oasis hydrophilic-lipophilic balance cartridge and using a cleanup step (5% methanol) to eliminate interference with a total run time of 13 min. The isotope-labeled internal standard method was used to quantify the target substances and correct for unavoidable losses and matrix effects during the on-line SPE process. Typical analytical characteristics used for method validation were sensitivity, linearity, precision, repeatability, recovery, and matrix effects. The limit of detection (LOD) and limit of quantification (LOQ) of each target were set at 0.20 ng/L and 0.50 ng/L, respectively. The linearity was between 0.5 ng/L and 250 ng/L, except for that of COT. The intra- and inter-day precisions were <10.45% and 25.64%, respectively, and the relative recovery ranged from 83.74% to 162.26%. The method was used to analyze various wastewater samples from 33 cities in China, and the results were compared with the experimental results of identical samples analyzed using off-line SPE. The difference rate was between 19.91% and −20.44%, and the error range could be considered acceptable. These findings showed that on-line SPE is a suitable alternative to off-line SPE for the analysis of illicit drugs in samples.

Keywords: Illicit drugs and metabolites, Wastewater analysis, On-line solid phase extraction, Ultra-high-performance liquid chromatography, Mass spectrometry

Graphical abstract

Image 1

Highlights

  • A new wastewater analysis method based on on-line SPE UHPLC-MS/MS was established.

  • The on-line SPE method showed excellent performance compared with off-line SPE.

  • High sensitivity and short analysis time were achieved using the on-line SPE method.

1. Introduction

Drug abuse has become a common problem in the international community. The 2019 World Drug Report [1] released by the United Nations Office on Drugs and Crime (UNODC) shows that in 2017, approximately 271 million individuals or 5.5% of the global population aged 15–64 years old had used drugs. Currently, the problem of drug abuse is a serious threat to human health and social security. Wastewater-based epidemiology (WBE) can be used to obtain accurate information on drug abuse and, therefore, has an important role in combating drug crime and reducing the harm caused by illicit drugs.

WBE was first proposed by Daughton [2] in 2001; it focuses on analyzing the concentrations of drug metabolites in wastewater drained from a specific city to estimate the consumption of a drug by the population of that city. Four years later, this wastewater analysis method was put into practice by Zuccato et al. [3]. They performed the first determination of cocaine (COC) and its main metabolite, benzoylecgonine (BZE) using solid phase extraction (SPE) liquid chromatography-tandem mass spectrometry (LC-MS/MS) on water samples collected from medium-sized urban wastewater treatment plants in Italy. Bones et al. [4] also supported this method and used it to evaluate the presence of drugs such as COC, BZE, and morphine (MOR) in Ireland. Since then, considerable research has led to the development of various analytical methods based on SPE for the determination of illicit drugs in wastewater samples from European countries (such as Croatia [5], France [6], Finland [7], and the Netherlands [8]) and North America (Canada [9] and the USA [10]). In China, the SPE-LC-MS/MS method was used for the first time to monitor four kinds of illicit drugs in major sewage treatment plants from Hong Kong [11]. Khan et al. [12] and Du et al. [13,14] used this method to estimate drug abuse in more major Chinese cities.

Much of the previous research on the quantification of illicit drugs in wastewater has focused on off-line SPE [[15], [16], [17], [18], [19]]. However, these methods have various acknowledged drawbacks. Off-line SPE requires large sample volumes of wastewater and a complex sample handling process. Furthermore, the multiple steps lead to a time- and organic solvent-consuming process with loss of analytes [20]. Hence, in 2008, the first fully automated method was developed by Postigo et al. [21]. The method was based on on-line SPE- LC-MS/MS for the determination of 17 compounds and metabolites in various influent and effluent samples from Spain [21]. On-line SPE has several advantages such as the reduction of sample handling and analysis time [22]. A large and growing body of literature has reported the application of on-line SPE [23,24]. In a follow-up study, Yao et al. [25] subsequently optimized this method for the detection of 10 illicit drugs in environmental samples from Chinese cities. Similarly, López-García et al. [26] studied the consumption of methamphetamine (MA), ketamine (KET), and methadone as well as their metabolites in Barcelona, Spain for one week.

Although the on-line SPE method has been relatively well established, it limits the existing research in two ways. On the one hand, the long total run time is not conducive to the rapid detection of numerous wastewater samples. On the other hand, the large matrix effects may lead to incomplete regeneration of the SPE column, which adversely affects the samples tested. In this work, these problems were resolved by reducing the analysis time and sample volume and optimizing the filtration condition for sample pre-treatment and SPE column regenerated solvent. In addition, the on-line SPE column, sample pH, and mobile phase were optimized for good sensitivity and accuracy.

The aim of this study was to develop an optimized method based on on-line SPE-LC-MS/MS for the determination of the following 13 target substances: MA, amphetamine (AM), MOR, codeine (COD), 6-monoacetylmorphine (6-MAM), BZE, 3,4-methylenedioxymethamphetamine (MDMA) 3,4-methylenedioxyamphetamine (MDA), COC, KET, norketamine (NK), and methcathinone (MC) and cotinine (COT) in various wastewater samples from 33 cities in China. In addition, we used the off-line SPE-UHLPC-MS/MS method [27] to analyze the collected wastewater samples and compared the results of the two methods concurrently. The optimized method was shown to have minor variation, good reliability, and substitutability.

2. Materials and methods

2.1. Reagent and chemicals

All reference standards were obtained as certified solutions in methanol from Cerilliant (Round Rock, TX, USA) at a concentration of 1 mg/mL. The studied illicit drugs and some of their studied metabolites were COT, MOR, MC, COD, AM, MA, MDA, BZE, NK, 6-MAM, MDMA, KET, and COC. Deuterated compounds COT-D3, MOR-D3, MC-D3, COD-D6, AM-D5, MA-D5, MDA-D5, BZE-D3, NK-D4, 6-MAM-D3, MDMA-D5, KET-D4, and COC-D3 used as isotope-labeled internal standards (IS) were also purchased from Cerilliant as 100 μg/mL solutions in methanol.

Mixed stock solutions were prepared with methanol at concentrations of 500 and 2500 ng/mL. Working standard solutions were prepared by appropriately diluting mixed stock solutions in methanol to different concentrations. All stock and working solutions were stored in the dark at −20 °C.

HPLC-grade methanol (MeOH), acetonitrile (ACN), and isopropyl alcohol (IPA) were purchased from Tedia (Fairfield, OH, USA). Formic acid (HCOOH) was obtained from Nanjing Chemical Reagent Co., Ltd. (Nanjing, China). Sodium hydroxide (NaOH) and sodium dihydrogen phosphate dihydrate (NaH2PO4⋅2H2O) were acquired from Xilong Scientific Co., Ltd. (Shantou, China). Deionized water was prepared using a Milli-Q water purification system (Nanjing Miaozhiyi Electronic Technology, Nanjing, China).

2.2. Sample collection and pre-treatment

A series of 24-h composite wastewater samples was collected from the influent of wastewater treatment plants (WWTP) in 33 different cities in China. All samples were collected in polyethylene terephthalate bottles and adjusted to pH 2 with hydrochloric acid. The samples were immediately transported to the laboratory under cool conditions and stored at −20 °C until analysis. Before extraction, the samples were thawed at room temperature and spiked with the IS. Next, the samples were centrifuged at 10,000 r/min for 5 min and filtered through 0.45-μm hydrophilic polyvinylidene fluoride (PVDF) syringe filters (Jinteng, Tianjin, China).

2.3. On-line SPE and UHPLC-MS/MS

The samples were pre-concentrated using an on-line SPE device, an automatic on-line extraction system (Shimadzu Corporation, Kyoto, Japan). The system consisted of an automated sample processor SIL-16P with a 5 mL sample loop and an LC-20AD pump unit with low-pressure gradient formation. The UHPLC system (Shimadzu Corporation, Kyoto, Japan) was equipped with two LC-30AD pumps, an autosampler SIL-30ACMP, and a CTO-20AC column oven (including proportional valve FCV-36AH). The 13 target substances were separated using a Shimadzu Shim-pack GISS C18 column (2.1 mm × 100 mm, 1.9 μm). The oven temperature was set to 40 °C.

The sample processing method comprised four steps: sample loading, cleanup, elution, and on-line SPE column regeneration, which was accomplished using a column switching operation of two on-line SPE columns [28]. All injection solutions (filtered) were adjusted to pH 5–7 with appropriate amounts of NaOH and NaH2PO4⋅2H2O. Then, a small amount of the sample solution of about 5 mL was injected into an Oasis hydrophilic-lipophilic balance (HLB) cartridge (2.1 mm × 30 mm, 20 μm, Waters Corporation, Wexford, Ireland) and washed with 5% MeOH at a flow rate of 3 mL/min to transfer all samples and remove interference such as water-soluble impurities.

After the cleanup, the analytes were eluted from the Oasis HLB into the UHPLC system using the chromatographic mobile phase (deionized water containing 0.1% formic acid/ACN) at a flow rate of 0.4 mL/min. The proportion of the organic phase was increased from 8% to 30% in the first 5 min and then to 100% in the following 13 min; subsequently, the column was washed with two different solvents (50% MeOH containing 0.1% HCOOH; MeOH:ACN:IPA (1:1:1, V/V/V) containing 0.1% HCOOH) to regenerate for 5 min.

The initial mobile phase conditions were maintained for 3 min to rebalance the chromatographic column between runs. During the elution and analysis of the first sample, the next sample was extracted simultaneously through another on-line SPE column. The total run time of the chromatographic analysis was 13 min per cycle. The detailed on-line SPE process is shown in Table S1, and a typical chromatogram of the sample analysis is displayed in Fig. 1.

Fig. 1.

Fig. 1

Typical chromatogram in the positive electrospray ionization (ESI) mode for 13 target substances obtained using the developed on-line solid phase extraction (SPE) method. COT: cotinine; MOR: morphine; MC: methcathinone; COD: codeine; AM: amphetamine; MDA: 3,4-methylenedioxyamphetamine; MA: methamphetamine; 6-MAM: 6-monoacetylmorphine; MDMA: 3,4-methylenedioxymethamphetamine; NK: norketamine; BZE: benzoylecognine; KET: ketamine; COC: cocaine.

The UHPLC system was coupled to an LCMS-8050 triple quadrupole mass spectrometer (Shimadzu Corporation, Kyoto, Japan) with an electrospray ionization (ESI) source in the positive mode. The MS operating conditions were as follows: capillary voltage, 3.5 kV; nebulizer gas (N2) flow rate and heating gas flow rate, 3 and 10 L/min, respectively; interface temperature, desolvation temperature and heated block temperature, 300, 250, and 400 °C, respectively; and drying gas (N2) flow rate, 10 L/min.

2.4. Method validation

The method was mainly validated for the limit of detection (LOD), limit of quantification (LOQ), linearity, accuracy, precision, repeatability, recovery, and matrix effect. All samples were quantified by peak areas and corrected using the IS method. The LOD and LOQ were defined as the concentrations in spiked blank wastewater with signal-to-noise ratios of 3 and 10, respectively. Because there were differences between the wastewater samples, they were mixed and subjected to SPE column processing, which absorbed the interfering substances to produce blank wastewater.

Linearity was established in spiked blank wastewater at different concentration levels with a seven-point calibration curve, using least-squares linear regression analysis. The LOQ was used as the lowest concentration point of the linear curve. The accuracy was investigated using spiked wastewater at three different concentration levels of 4, 80, and 200 ng/L, except for COT, which was used at 40, 800, and 2000 ng/L. The intra-day precision (expressed as repeatability), calculated as relative standard deviation (RSD), was studied in six replicates at the same concentration level. The inter-day precision was evaluated at the same level in 3 days. Recovery was divided into absolute recovery (AR) and relative recovery (RR), which were analyzed at the same concentration levels.

The AR was assessed by comparing the peak area of each target obtained from the on-line SPE analysis of the spiked wastewater with those obtained from samples directly injected into the UHPLC-MS/MS at equivalent amounts. The RR was calculated as the ratio between the concentration of each compound after it was corrected using the IS obtained from the on-line SPE treatment with spiked wastewater and spiked deionized water. The matrix effects were evaluated by comparing the peak area of each target obtained from the on-line SPE-UHPLC-MS/MS analysis in spiked wastewater at three concentration levels with those of the spiked deionized water.

3. Results and discussion

3.1. Method optimization

The on-line SPE and chromatographic conditions were optimized for the effective extraction and analysis of the target substances. To date, sample filtration has not been extensively discussed in the available published studies [25,26,29,30]. Optimized filtration methods play a crucial role in reducing the adsorption of particles to target substances and in protecting the instrument and reducing material losses.

In this work, the suitability of eight different 0.45-μm syringe filters was investigated: hydrophilic polytetrafluoroethylene, PVDF (both from Jinteng, Tianjin, China), mixed cellulose ester (MCE, Jinteng, Tianjin, China), regenerated cellulose (RCE), cellulose (CELL) (both from Shengze Technology Co., Ltd., Tianjin, China), polyethersulfone (PES), and nylon (both from Ameritech Scientific (Tianjin) Co., Ltd., Tianjin, China) syringe filters.

The experimental results showed that the 0.45-μm PES and nylon syringe filters adsorbed most of the tested substances compared to the unfiltered simulated samples, and the recovery of 6-MAM decreased when the 0.45-μm RCE, MCE, CELL, and nylon syringe filters were used. The results of 13 analytes were compared, and the 0.45-μm hydrophilic PVDF syringe filter was chosen as the sample filtering system. This two-stage method (centrifugation and filtration) not only removed the solid particles quickly and effectively but also exhibited a low adsorption rate, which ensured the accuracy of the target concentration.

In addition, the separation of the target compounds was investigated using three different SPE cartridges: the Oasis HLB (2.1 mm × 30 mm, 20 μm, Waters Corporation), XBridge C18 (2.1 mm × 30 mm, 10 μm, Waters Corporation), and XBridge C8 (2.1 mm × 30 mm, 10 μm, Waters Corporation). The results showed that using the Oasis HLB column for the separation enabled all the analytes to be separated within a short time and their chromatographic peaks were in good shape. Different proportions of MeOH (0%, 5%, and 10%) were tested, and HPLC-grade water containing 5% MeOH was chosen as the washing solvent because it minimized the matrix effect and increased the sensitivity of the method.

Furthermore, three different compositions of the mobile phase were evaluated: 1) HPLC-grade water/ACN, 2) HPLC-grade water containing 0.1% formic acid/ACN, and 3) HPLC-grade water containing 5 mM ammonium formate/ACN. HPLC-grade water containing 0.1% formic acid (solvent A) and ACN (solvent B) was selected as the most appropriate mobile phase because each target was adequately eluted and a low concentration of formic acid enhanced the response of the analytes. We also optimized the method using six different sample pH values (2, 3, 5, 7, 9, and 10), and three samples were examined under each condition. The results showed that the average peak areas of COT and COC decreased at pH 7–10 and the average peak areas of MOR, MC, and AM decreased at pH 2–5. The retention of COD, MDA, 6-MAM, MA, and KET on the on-line SPE was not affected by the pH of the sample solution. Based on the detection results of each target analyte, the optimal pH condition of the sample solution was 5–7.

In particular, adjusting the sample volume to 2 mL and the total run time to 13 min not only reduced the injection time and matrix effects but also resulted in sufficient sensitivity. In addition, the on-line SPE column regeneration step was re-optimized. Some previous studies only used a single organic solvent to wash the SPE column after the analysis [26,[30], [31], [32]], which could be a contributing factor in incomplete cleaning or even contamination of the next sample. In this experiment, different organic reagents and their mixtures were investigated depending on the properties of the target compounds. The following solvent was chosen for the SPE column regeneration: MeOH:ACN:IPA (1:1:1, V/V/V), containing 0.1% HCOOH, to ensure that the SPE column could be thoroughly cleaned and to guarantee the authenticity and accuracy of the sample determination.

The MS/MS conditions of the optimum performance were in the positive ESI mode. Data for each illicit drug were acquired in the multiple reaction monitoring (MRM) mode, where the transitions between the precursor ion and the two most abundant product ions were quantified and confirmed. To ensure high sensitivity, the collision energy of each selected product ion was optimized. All information on the chromatographic retention time and the relevant MRM conditions for the target analytes is displayed in Table S2.

3.2. Method performance

The results of the methodological evaluation are shown in Table 1. The method was validated by examining the LOD, LOQ, linearity, intra- and inter-day precision, AR, RR, and matrix effects, which guided the determination of the wastewater samples. The sensitivity was improved using the on-line SPE because all the samples were completely transferred to the chromatographic column without analyte loss, as observed during the off-line SPE method. The LOD and LOQ of each target were obtained at 0.20 ng/L and 0.50 ng/L, respectively. Linearity was evaluated at concentrations for all illicit drugs in the range of 0.5–250 ng/L (5–2500 ng/L for COT) using 1/y2 as the weighting factor.

Table 1.

Validation parameters of the proposed method.

Compound LOD (ng/L) LOQ (ng/L) Added concentration (ng/L) Accuracy (RE%) Precision (RSD%)
Absolute recovery (%) Relative recovery (%) Matrix effects (%)
Intra-day Inter-day
COT 0.2 0.5 40 −3.8 4.69 0.77 18.75 117.88 −88.50
800 −2.3 10.45 22.14 19.14 105.83 −92.50
2000 3.0 6.69 22.92 21.52 132.61 −92.20
MOR 0.2 0.5 4 −7.6 3.04 5.28 10.76 90.10 −73.80
80 6.1 7.57 15.59 10.55 89.07 −77.40
200 5.3 4.97 25.64 11.77 89.41 −78.90
MC 0.2 0.5 4 −7.5 2.73 2.35 15.49 97.73 −63.90
80 −3.5 2.76 7.40 18.54 99.56 −71.20
200 −1.6 3.48 6.32 27.01 111.77 −70.90
COD 0.2 0.5 4 −3.3 4.19 0.94 18.32 126.97 −73.40
80 2.0 4.20 0.41 21.62 135.05 −77.10
200 3.6 4.90 5.99 22.35 140.44 −80.90
AM 0.2 0.5 4 −5.3 3.21 3.75 21.21 83.74 −66.30
80 −8.5 2.34 19.31 20.30 84.31 −67.90
200 −4.1 2.71 9.04 20.06 88.36 −67.50
MDA 0.2 0.5 4 −5.1 3.55 6.82 41.41 112.99 −65.20
80 −4.0 2.81 11.7 42.10 110.79 −70.20
200 −3.1 3.00 2.89 50.54 115.22 −72.60
6-MAM 0.2 0.5 4 −6.2 3.97 2.19 20.47 109.26 −69.90
80 −3.8 5.09 11.20 24.83 100.82 −75.90
200 −3.7 3.94 3.59 29.43 102.81 −80.50
MA 0.2 0.5 4 −7.9 4.25 7.10 29.13 104.90 −39.20
80 3.3 2.67 20.04 54.82 130.62 −52.80
200 −2.1 2.92 22.79 65.14 161.84 −64.20
MDMA 0.2 0.5 4 −7.3 2.35 3.40 63.72 100.01 −59.40
80 −1.9 3.08 17.75 66.36 102.60 −74.50
200 1.0 3.28 15.70 75.98 144.52 −75.80
NK 0.2 0.5 4 −6.8 3.31 1.99 29.36 97.98 −42.20
80 −4.4 2.23 8.91 39.98 110.87 −63.70
200 −2.9 3.11 11.65 65.28 162.26 −70.50
BZE 0.2 0.5 4 −7.9 5.57 2.04 20.39 90.49 −57.00
80 −13.7 4.79 13.23 17.92 86.87 −60.60
200 −2.9 3.11 11.65 18.62 84.83 −63.30
KET 0.2 0.5 4 −5.1 3.38 1.75 33.44 101.63 −76.50
80 −0.8 2.80 13.97 34.25 101.21 −77.10
200 1.4 2.11 7.76 52.57 104.94 −76.20
COC 0.2 0.5 4 −7.6 2.01 1.06 30.75 93.72 −26.40
80 −2.2 2.16 11.07 59.20 140.97 −39.50
200 −4.3 2.80 3.53 70.97 160.08 −61.10

LOD: limit of detection; LOQ: limit of quantification; RE: relative Error; RSD: relative standard deviation; COT: cotinine; MOR: morphine; MC: methcathinone; COD: codeine; AM: amphetamine; MDA: 3,4-methylenedioxyamphetamine; 6-MAM: 6-monoacetylmorphine; MA: methamphetamine; MDMA: 3,4-methylenedioxymethamphetamine; NK: norketamine; BZE: benzoylecognine; KET: ketamine; COC: cocaine.

Linearity was assumed with the correlation coefficient (r2), which was higher than 0.99 for each substance. The intra-day precision and inter-day precision were expressed as RSD%. It can be seen that the intra-day precision was equal to or less than 10.45%, whereas the inter-day precision was consistently below 25.64%. The absolute recoveries of the 13 target compounds ranged between 10.55% and 75.98%, whereas the relative recoveries were between 83.74% and 162.26%. The results were attributed to losses in the analytical process and matrix effects, which can be corrected using the IS method. The matrix effects were evaluated using the following equation:

Matrix effects (%) = [(Asp ww−Aww)Asp water−1] × 100

where Asp ww is the analyte peak area in the spiked wastewater sample, Aww is the analyte peak area in the non-spiked wastewater sample, and Asp water is the analyte concentration in the spiked deionized water [21]. Different substrates had different effects on the analytes. The positive or negative values of the dates indicated that the responses of the wastewater samples to the analytes were enhanced or inhibited. The results showed that the method was suitable for a wide range of applications and met the requirements of actual sample determination.

3.3. Analysis of wastewater samples

The validated method was used to analyze 33 wastewater samples from WWTPs in 33 cities in China. Table 2 shows the results of the analysis of selected illicit drugs in the influent wastewater samples, and most of the target analytes were detected. COT and MA were found in 100% of the wastewater samples analyzed. MOR, COD, AM, and KET were detected in 94%, 94%, 56%, and 47% of the samples analyzed, respectively, whereas 6-MAM and MDA were not detected in any sample. The average concentrations of the other illicit drugs monitored increased in the following order: COC < MDMA < NK < BZE. In addition, MC levels were measured in four wastewater samples using the on-line SPE method.

Table 2.

Concentrations (ng/L) of compounds selected by on-line SPE method in influent wastewater samples from WWTPs in 33 cities in China.

Sample No. COT MOR MC COD AM MDA 6-MAM MA MDMA NK BZE KET COC
1 851.363 ND ND 1.732 1.274 ND ND 29.655 ND ND ND ND ND
2 1093.667 11.687 ND 3.945 ND ND ND 99.308 ND ND ND 1.244 ND
3 958.280 22.006 ND 3.521 ND ND ND 15.518 ND ND ND ND ND
4 1056.664 8.336 ND 2.559 ND ND ND 2.808 ND ND ND ND ND
5 1527.399 24.201 ND 4.720 ND ND ND 8.195 ND ND ND ND ND
6 2615.712 34.465 2.590 14.443 2.904 ND ND 47.571 ND ND ND ND ND
7 1011.889 10.318 ND 7.164 ND ND ND 4.060 ND ND ND ND ND
8 228.161 <LOQ ND ND ND ND ND 2.901 ND ND ND ND ND
9 897.603 <LOQ ND 3.135 1.315 ND ND 4.050 ND ND ND ND ND
10 1757.184 22.572 ND 12.460 2.664 ND ND 41.636 ND ND ND 0.982 ND
11 1194.711 44.974 ND 10.517 3.346 ND ND 83.456 ND 2.221 ND 11.297 ND
12 1570.679 28.585 ND 7.895 3.723 ND ND 70.857 ND 4.320 ND 29.117 ND
13 826.180 12.420 ND 6.125 5.907 ND ND 26.297 ND ND ND ND ND
14 2111.022 47.923 ND 14.919 5.036 ND ND 9.013 ND ND ND ND ND
15 1722.271 14.969 ND 2.910 ND ND ND 9.286 ND ND ND ND ND
16 1368.393 16.251 ND 3.203 ND ND ND 3.766 ND ND ND ND ND
17 572.188 <LOQ ND ND ND ND ND 6.914 ND ND ND ND ND
18 847.909 <LOQ ND 3.566 ND ND ND 3.109 ND ND ND 2.182 ND
19 1059.207 20.191 ND 8.674 2.892 ND ND 20.310 2.004 1.271 ND 6.586 ND
20 954.340 14.426 3.806 4.678 1.510 ND ND 11.922 ND ND ND 1.870 ND
21 673.691 18.838 ND 5.352 ND ND ND 12.954 1.096 ND ND ND ND
22 1373.990 31.616 8.241 6.848 7.107 ND ND 62.330 2.401 ND ND 5.215 1.712
23 1690.936 27.105 ND 14.039 3.221 ND ND 26.744 3.131 2.050 1.19 13.029 ND
24 922.929 62.893 ND 20.560 ND ND ND 36.773 ND ND ND 8.912 ND
25 1389.975 10.721 ND 8.650 ND ND ND 23.786 1.021 ND ND 5.628 ND
26 810.938 <LOQ ND 9.718 1.229 ND ND 41.863 ND ND ND ND ND
27 1178.995 17.993 ND 7.309 2.909 ND ND 28.543 1.586 ND 5.131 3.202 1.103
28 921.931 28.265 ND 13.336 2.069 ND ND 20.213 ND ND 3.051 0.754 ND
29 745.171 11.636 ND 4.692 ND ND ND 8.613 ND ND ND 8.881 ND
30 1020.312 15.439 ND 5.259 7.508 ND ND 23.890 ND ND ND 1.308 ND
31 1451.985 21.887 ND 6.822 2.513 ND ND 49.449 ND ND ND 0.708 ND
32 1189.114 31.775 5.879 10.846 17.012 ND ND 22.318 ND ND ND 4.352 ND
33 634.702 8.528 ND 3.230 ND ND ND 6.202 ND ND ND 3.234 ND

WWTP: wastewater treatment plant; ND: not detected.

Approximately 70%–80% of the nicotine absorbed by smokers is converted to COT and excreted in the urine [33]. The high concentration of COT in the different urban wastewater samples is due to the widespread consumption of tobacco and the large population in China. 6-MAM and MC have never been studied in wastewater using on-line SPE. 6-MAM is a minor but exclusive metabolite of heroin used as a specific detection marker for heroin abuse because MOR and COD cannot be acetylated to form 6-MAM in vivo [34]. MC is a psychoactive substance of the cathinone class that acts similarly to MA and can cause acute health problems and drug dependence [35]. In all samples, the detection rate of 6-MAM was zero, and the average concentration of MC was extremely low. Generally, the levels of COC, BZE, COD, MOR, and MDMA found in this study were lower than those previously reported in the UK and Spain [[36], [37], [38]].

All collected wastewater samples were determined simultaneously using the off-line SPE methods described by Wang et al. [27], and the results are presented in Table 3. The difference in concentration levels detected for each substance by the two data sets was calculated to evaluate the difference between the two methods. The difference in rates was defined as the percentage of the difference between the concentration of the same target substance detected using both SPE methods and compared with the detected concentration using the off-line SPE. The results showed that the difference in rate was between −19.91% and −20.44%, and the error range was acceptable, which proved that the optimized on-line SPE method had good applicability and could be used as an effective alternative method for the detection of illicit drugs in wastewater samples.

Table 3.

Concentrations (ng/L) of compounds selected by off-line SPE method in influent wastewater samples from WWTPs in 33 cities in China.

Sample No. COT MOR MC COD AM MDA 6-MAM MA MDMA NK BZE KET COC
1 767.822 ND ND 2.032 1.149 ND ND 33.222 ND ND ND ND ND
2 951.895 12.737 ND 3.903 ND ND ND 115.818 ND ND ND 1.313 ND
3 812.051 19.534 ND 4.031 ND ND ND 14.949 ND ND ND ND ND
4 892.543 7.859 ND 2.665 ND ND ND 3.506 ND ND ND ND ND
5 1312.350 21.892 ND 4.401 ND ND ND 7.286 ND ND ND ND ND
6 2308.760 34.268 2.152 16.558 3.222 ND ND 55.973 ND ND ND ND ND
7 840.149 9.747 ND 7.485 ND ND ND 4.082 ND ND ND ND ND
8 196.483 <LOQ ND ND ND ND ND 2.604 ND ND ND ND ND
9 783.933 <LOQ ND 3.680 1.121 ND ND 3.990 ND ND ND ND ND
10 1528.050 22.218 ND 12.155 2.225 ND ND 45.325 ND ND ND 0.975 ND
11 1015.570 49.118 ND 11.166 4.149 ND ND 100.860 ND 1.916 ND 12.050 ND
12 1313.910 28.139 ND 7.183 3.830 ND ND 79.278 ND 4.538 ND 30.277 ND
13 756.552 10.996 ND 5.950 6.592 ND ND 30.953 ND ND ND ND ND
14 1805.990 45.848 ND 14.769 5.256 ND ND 10.317 ND ND ND ND ND
15 1519.490 144.90 ND 3.249 ND ND ND 10.116 ND ND ND ND ND
16 1268.620 15.093 ND 3.121 ND ND ND 4.099 ND ND ND ND ND
17 480.057 <LOQ ND ND ND ND ND 6.577 ND ND ND ND ND
18 764.734 <LOQ ND 3.474 ND ND ND 3.373 ND ND ND 2.199 ND
19 920.622 19.019 ND 8.098 3.025 ND ND 22.775 1.896 1.057 ND 6.523 ND
20 862.541 16.584 3.195 5.595 1.374 ND ND 12.888 ND ND ND 1.835 ND
21 570.841 19.069 ND 4.641 ND ND ND 13.438 1.364 ND ND ND ND
22 1208.750 30.976 7.547 6.818 6.145 ND ND 67.218 2.966 ND ND 4.771 1.649
23 1413.080 24.596 ND 16.033 3.103 ND ND 30.642 3.792 1.994 1.433 13.897 ND
24 811.697 64.284 ND 20.303 ND ND ND 43.863 ND ND ND 9.741 ND
25 1161.170 11.874 ND 7.845 ND ND ND 26.763 0.990 ND ND 6.297 ND
26 802.387 <LOQ ND 11.403 1.075 ND ND 49.020 ND ND ND ND ND
27 1027.800 18.403 ND 7.341 3.153 ND ND 34.173 1.701 ND 6.058 3.419 1.181
28 849.668 31.941 ND 16.411 2.418 ND ND 22.584 ND ND 3.044 ND ND
29 642.471 13.593 ND 4.509 ND ND ND 10.683 ND ND ND 8.908 ND
30 900.915 15.930 ND 5.650 8.546 ND ND 26.075 ND ND ND 1.150 ND
31 1286.340 21.567 ND 7.149 2.413 ND ND 57.527 ND ND ND ND ND
32 1068.510 31.899 4.970 10.207 17.618 ND ND 25.564 ND ND ND 4.853 ND
33 618.671 8.342 ND 3.337 ND ND ND 5.441 ND ND ND 3.002 ND

4. Conclusions

In this work, a novel automated analytical method based on on-line SPE-UHPLC-MS/MS was developed and validated for the simultaneous analysis of 12 illicit drugs in wastewater samples collected from 33 cities in China. The method proved to be effective for the analysis of the selected target compounds. The major advantages of this method are small sample manipulation, high sensitivity, time and cost saving, and real-time information. Drug monitoring of public security departments based on this method could accurately provide information on drug abuse that would contribute to preventing and combating drug-related crimes. Furthermore, the application of this strategy could provide important technical support for the comprehensive implementation of all narcotics control work.

Declaration of competing interest

The authors declare that there are no conflicts of interest.

Acknowledgments

This work was supported by the National Key Research and Development Program of China (Grant No.: 2018YFC0807402) and the National Natural Science Foundation of China (Grant No.: 82073810).

Footnotes

Peer review under responsibility of Xi’an Jiaotong University.

Appendix A

Supplementary data to this article can be found online at https://doi.org/10.1016/j.jpha.2021.01.002.

Contributor Information

Zhendong Hua, Email: 28008085@qq.com.

Youmei Wang, Email: youmei_626@163.com.

Bin Di, Email: dibin@cpu.edu.cn.

Appendix A. Supplementary data

The following is the Supplementary data to this article:

Multimedia component 1
mmc1.docx (19.2KB, docx)

References

  • 1.World . United Nations Office on Drugs and Crime; https://wdr.unodc.org/wdr2019/: 2020. Drug Report. (Accessed 12 March 2020) [Google Scholar]
  • 2.Daughton C.G. Emerging pollutants and communicating the science of environmental chemistry and mass spectrometry: pharmaceuticals in the environment. J. Am. Soc. Mass Spectrom. 2001;12:1067–1076. doi: 10.1016/S1044-0305(01)00287-2. [DOI] [PubMed] [Google Scholar]
  • 3.Zuccato E., Chiabrando C., Castiglioni S., et al. Cocaine in surface waters: a new evidence-based tool to monitor community drug abuse. Environ. Health. 2005;4:14. doi: 10.1186/1476-069X-4-14. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 4.Bones J., Thomas K.V., Paull B. Using environmental analytical data to estimate levels of community consumption of illicit drugs and abused pharmaceuticals. J. Environ. Monit. 2007;9:701–707. doi: 10.1039/b702799k. [DOI] [PubMed] [Google Scholar]
  • 5.Terzic S., Senta I., Ahel M. Illicit drugs in wastewater of the city of Zagreb (Croatia)-Estimation of drug abuse in a transition country. Environ. Pollut. 2010;158:2686–2693. doi: 10.1016/j.envpol.2010.04.020. [DOI] [PubMed] [Google Scholar]
  • 6.Karolak S., Nefau T., Bailly E., et al. Estimation of illicit drugs consumption by wastewater analysis in Paris area (France) Forensic Sci. Int. 2010;200:153–160. doi: 10.1016/j.forsciint.2010.04.007. [DOI] [PubMed] [Google Scholar]
  • 7.Vuori E., Happonen M., Gergov M., et al. Wastewater analysis reveals regional variability in exposure to abused drugs and opioids in Finland. Sci. Total Environ. 2014;487:688–695. doi: 10.1016/j.scitotenv.2013.11.010. [DOI] [PubMed] [Google Scholar]
  • 8.Bijlsma L., Emke E., Hernández F., et al. Investigation of drugs of abuse and relevant metabolites in Dutch sewage water by liquid chromatography coupled to high resolution mass spectrometry. Chemosphere. 2012;89:1399–1406. doi: 10.1016/j.chemosphere.2012.05.110. [DOI] [PubMed] [Google Scholar]
  • 9.Metcalfe C., Tindale K., Li H., et al. Illicit drugs in Canadian municipal wastewater and estimates of community drug use. Environ. Pollut. 2010;158:3179–3185. doi: 10.1016/j.envpol.2010.07.002. [DOI] [PubMed] [Google Scholar]
  • 10.Gerrity D., Trenholm R.A., Snyder S.A. Temporal variability of pharmaceuticals and illicit drugs in wastewater and the effects of a major sporting event. Water Res. 2011;45:5399–5411. doi: 10.1016/j.watres.2011.07.020. [DOI] [PubMed] [Google Scholar]
  • 11.Lai F.Y., Bruno R., Leung H.W., et al. Estimating daily and diurnal variations of illicit drug use in Hong Kong: a pilot study of using wastewater analysis in an Asian metropolitan city. Forensic Sci. Int. 2013;233:126–132. doi: 10.1016/j.forsciint.2013.09.003. [DOI] [PubMed] [Google Scholar]
  • 12.Khan U., van Nuijs A.L.N., Li J., et al. Application of a sewage-based approach to assess the use of ten illicit drugs in four Chinese megacities. Sci. Total Environ. 2014;487:710–721. doi: 10.1016/j.scitotenv.2014.01.043. [DOI] [PubMed] [Google Scholar]
  • 13.Du P., Li K., Li J., et al. Methamphetamine and ketamine use in major Chinese cities, a nationwide reconnaissance through sewage-based epidemiology. Water Res. 2015;84:76–84. doi: 10.1016/j.watres.2015.07.025. [DOI] [PubMed] [Google Scholar]
  • 14.Du P., Zhou Z., Bai Y., et al. Estimating heroin abuse in major Chinese cities through wastewater-based epidemiology. Sci. Total Environ. 2017;605-606:158–165. doi: 10.1016/j.scitotenv.2017.05.262. [DOI] [PubMed] [Google Scholar]
  • 15.Kinyua J., Covaci A., Maho W., et al. Sewage-based epidemiology in monitoring the use of new psychoactive substances: validation and application of an analytical method using LC-MS/MS. Drug Test. Anal. 2015;7:812–818. doi: 10.1002/dta.1777. [DOI] [PubMed] [Google Scholar]
  • 16.Bijlsma L., Botero-Coy A.M., Rincón, R.J., et al. Estimation of illicit drug use in the main cities of Colombia by means of urban wastewater analysis. Sci. Total Environ. 2016;565:984–993. doi: 10.1016/j.scitotenv.2016.05.078. [DOI] [PubMed] [Google Scholar]
  • 17.Löve A.S.C., Baz-Lomba J.A., Reid M.J., et al. Analysis of stimulant drugs in the wastewater of five Nordic capitals. Sci. Total Environ. 2018;627:1039–1047. doi: 10.1016/j.scitotenv.2018.01.274. [DOI] [PubMed] [Google Scholar]
  • 18.Skee A.J., Foppe K.S., Loganathan B., et al. Contamination profiles, mass loadings, and sewage epidemiology of neuropsychiatric and illicit drugs in wastewater and river waters from a community in the Midwestern United States. Sci. Total Environ. 2018;631-632:1457–1464. doi: 10.1016/j.scitotenv.2018.03.060. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 19.Daglioglu N., Guzel E.Y., Kilercioglu S. Assessment of illicit drugs in wastewater and estimation of drugs of abuse in Adana Province, Turkey. Forensic Sci. Int. 2019;294:132–139. doi: 10.1016/j.forsciint.2018.11.012. [DOI] [PubMed] [Google Scholar]
  • 20.Bones J., Thomas K., Nesterenko P.N., et al. On-line preconcentration of pharmaceutical residues from large volume water samples using short reversed-phase monolithic cartridges coupled to LC-UV-ESI-MS. Talanta. 2006;70:1117–1128. doi: 10.1016/j.talanta.2006.02.026. [DOI] [PubMed] [Google Scholar]
  • 21.Postigo C., Lopez de Alda M.J.L., Barceló D. Fully automated determination in the low nanogram per liter level of different classes of drugs of abuse in sewage water by on-line solid-phase extraction-liquid chromatography-electrospray-tandem mass spectrometry. Anal. Chem. 2008;80:3123–3124. doi: 10.1021/ac702060j. [DOI] [PubMed] [Google Scholar]
  • 22.Rodriguez-Mozaz S., Lopez de Alda M.J., Barceló D. Advantages and limitations of on-line solid phase extraction coupled to liquid chromatography–mass spectrometry technologies versus biosensors for monitoring of emerging contaminants in water. J. Chromatogr. A. 2007;1152:97–115. doi: 10.1016/j.chroma.2007.01.046. [DOI] [PubMed] [Google Scholar]
  • 23.Heuett N.V., Ramirez C.E., Fernandez A., et al. Analysis of drugs of abuse by online SPE-LC high resolution mass spectrometry: communal assessment of consumption. Sci. Total Environ. 2015;511:319–330. doi: 10.1016/j.scitotenv.2014.12.043. [DOI] [PubMed] [Google Scholar]
  • 24.Mastroianni N., López-García E., Postigo C., et al. Five-year monitoring of 19 illicit and legal substances of abuse at the inlet of a wastewater treatment plant in Barcelona (NE Spain) and estimation of drug consumption patterns and trends. Sci. Total Environ. 2017;609:916–926. doi: 10.1016/j.scitotenv.2017.07.126. [DOI] [PubMed] [Google Scholar]
  • 25.Yao B., Lian L.S., Pang W., et al. Determination of illicit drugs in aqueous environmental samples by online solid-phase extraction coupled to liquid chromatography-tandem mass spectrometry. Chemosphere. 2016;160:208–215. doi: 10.1016/j.chemosphere.2016.06.092. [DOI] [PubMed] [Google Scholar]
  • 26.López-García E., Mastroianni N., Postigo C., et al. A fully automated approach for the analysis of 37 psychoactive substances in raw wastewater based on on-line solid phase extraction-liquid chromatography-tandem mass spectrometry. J. Chromatogr. A. 2018;1576:80–89. doi: 10.1016/j.chroma.2018.09.038. [DOI] [PubMed] [Google Scholar]
  • 27.Wang J., Hou C., Hua Z., et al. Simultaneous determination of illicit drugs and their metabolites in wastewater by SPE-UPLC-MS/MS. J. China Pharm. Univ. 2020;51:305–312. [Google Scholar]
  • 28.Quintana J.B., Miró M., Estela J.M., et al. Automated on-line renewable solid-phase extraction-liquid chromatography exploiting multisyringe flow injection-bead injection lab-on-valve analysis. Anal. Chem. 2006;78:2832–2840. doi: 10.1021/ac052256z. [DOI] [PubMed] [Google Scholar]
  • 29.López-Serna R., Pérez S., Ginebreda A., et al. Fully automated determination of 74 pharmaceuticals in environmental and waste waters by online solid phase extraction–liquid chromatography-electrospray–tandem mass spectrometry. Talanta. 2010;83:410–424. doi: 10.1016/j.talanta.2010.09.046. [DOI] [PubMed] [Google Scholar]
  • 30.Berset J., Brenneisen R., Mathieu C. Analysis of llicit and illicit drugs in waste, surface and lake water samples using large volume direct injection high performance liquid chromatography-Electrospray tandem mass spectrometry (HPLC-MS/MS) Chemosphere. 2010;81:859–866. doi: 10.1016/j.chemosphere.2010.08.011. [DOI] [PubMed] [Google Scholar]
  • 31.Trenholm R.A., Vanderford B.J., Snyder S.A. On-line solid phase extraction LC-MS/MS analysis of pharmaceutical indicators in water: a green alternative to conventional methods. Talanta. 2009;79:1425–1432. doi: 10.1016/j.talanta.2009.06.006. [DOI] [PubMed] [Google Scholar]
  • 32.Östman M., Fick J., Näsström E., et al. A snapshot of illicit drug use in Sweden acquired through sewage water analysis. Sci. Total Environ. 2014;472:862–871. doi: 10.1016/j.scitotenv.2013.11.081. [DOI] [PubMed] [Google Scholar]
  • 33.Bramer S.L., Kallungal B.A. Clinical considerations in study designs that use cotinine as a biomarker. Biomarkers. 2003;8:187–203. doi: 10.1080/13547500310012545. [DOI] [PubMed] [Google Scholar]
  • 34.Kintz P., Jamey C., Cirimele V., et al. Evaluation of acetylcodeine as a specific marker of illicit heroin in human hair. J. Anal. Toxicol. 1998;22:425–429. doi: 10.1093/jat/22.6.425. [DOI] [PubMed] [Google Scholar]
  • 35.Stepens A., Logina I., Liguts V., et al. A Parkinsonian syndrome in methcathinone users and the role of manganese. N. Engl. J. Med. 2008;358:1009–1017. doi: 10.1056/NEJMoa072488. [DOI] [PubMed] [Google Scholar]
  • 36.Baker D.R., Kasprzyk-Hordern B. Multi-residue analysis of drugs of abuse in wastewater and surface water by solid-phase extraction and liquid chromatography-positive electrospray ionisation tandem mass spectrometry. J. Chromatogr. A. 2011;1218:1620–1631. doi: 10.1016/j.chroma.2011.01.060. [DOI] [PubMed] [Google Scholar]
  • 37.Bijlsma L., Beltrán E., Boix C., et al. Improvements in analytical methodology for the determination of frequently consumed illicit drugs in urban wastewater. Anal. Bioanal. Chem. 2014;406:4261–4272. doi: 10.1007/s00216-014-7818-4. [DOI] [PubMed] [Google Scholar]
  • 38.González-Mariño I., Quintana J.B., Rodríguez I., et al. Determination of drugs of abuse in water by solid-phase extraction, derivatisation and gas chromatography-ion trap-tandem mass spectrometry. J. Chromatogr. A. 2010;1217:1748–1760. doi: 10.1016/j.chroma.2010.01.046. [DOI] [PubMed] [Google Scholar]

Associated Data

This section collects any data citations, data availability statements, or supplementary materials included in this article.

Supplementary Materials

Multimedia component 1
mmc1.docx (19.2KB, docx)

Articles from Journal of Pharmaceutical Analysis are provided here courtesy of Xi'an Jiaotong University

RESOURCES