Abstract

The paper extends applicability of the internal standard method published in 2009 (Fuguet E. et al., J. Chromatogr. A2009,1216(17), 3646). Although the original capillary zone electrophoresis method was suggested to determine thermodynamic acidity constants of compounds sparingly soluble in aqueous solutions by carrying out only runs at two different pH values (i.e., without the need to perform many experiments over the appropriate pH range including the form of a low-ionized analyte), we proved that the approach also virtually overcomes any interactions of the analyte in mixed solvents, so that the experiments can be carried out in a methanol–water buffer where the solubility is much better. Applicability of the extended method is illustrated on six selected β-blockers.
Introduction
A common method for determining pKa of monoprotic weak bases by capillary zone electrophoresis (CZE) is based on changes in the analyte mobility with the variation of buffer pH: a series of experiments with electrolytes are conducted over the appropriate pH range (≈pKa ± 2) at a constant ionic strength.1−3 The theory of electrophoretic mobility states that4
| 1 |
where μeff is the effective electrophoretic mobility, μBH+ is the electrophoretic mobility of the fully protonated base, pKa´ is the negative decadic logarithm of the mixed acidity constant. Equation 1 indicates that by plotting the observed electrophoretic mobility (calculated from migration times in the electropherogram) against pH, a sigmoidal curve is obtained where its inflection point represents pKa´. There are many spreadsheet calculators that can help to calculate pKa´ by fitting the curve.5
Calculation of the effective electrophoretic mobility μeff in capillary zone electrophoresis is based on measurement of two migration times: apparent migration time (of the analyte) tm and a migration time of the electroosmotic flow (an EOF marker) tEOF
| 2 |
where LD is the length from the capillary inlet to the center of the detection window, LT is the total capillary length, and U is the applied separation voltage.
Because the experiments are carried out in an electrolyte, typically an aqueous buffer, in order to get the thermodynamic pKa, the obtained value should be corrected to the activity coefficient for ions in dilute (up to 0.075 mol/L) electrolyte solutions at 25 °C according to the Debye–Hückel theory of nonideality of electrolyte solution. For bases, it holds
| 3 |
where γ is the activity coefficient
of the buffer species, calculated as
where z is a charge number
and I is the ionic strength of the solution.
Compounds that are slightly soluble in water may require experiments
in mixed solvents. A mixture of water and methanol is usually employed.6 However, in a mixed solvent (e.g., of the volume fraction ϕ of methanol in
water), an experimentally accessed acidity constant
(solvent–water acidity
constant)
is related to a pure organic solvent
(solvent–solvent acidity
constant)
by a formula7
| 4 |
where
.
Therefore,
a way to obtain the thermodynamic pKa =
(water–water acidity constant)
from
is a correction to solvation effects.
The
literature suggests several extrapolations to estimate
. A common approach is the Yasuda–Shedlovsky
equation that relates
with reciprocal relative electric
permittivity
of an aqueous binary solvent.8 In this
case, an extrapolation to zero content of organic solvent is performed
from series of experiments with different amounts of water. Alternatively,
an empirical linear equation was suggested for acids belonging to
the same family when specific solvation effects in solvent (S) and
in water (W) can be linearly related with the acidity of the acid
| 5 |
where
parameters
S and bS can be calculated from the organic
solvent amount (for bases
or acids from data of Rived et al.(9)). Another approach to the conversion of methanolic pKa values to
for structurally similar compounds
was
also presented.10
Internal Standard in Capillary Zone Electrophoresis for pKa Determination
In 2009, Fuguet et al. introduced a method for the determination of acidity constants by capillary zone electrophoresis (CZE) with an internal standard (IS): only two pairs of electrophoretic runs are required to determine the acidity constant: (i) at pH, where the analyte and internal standard are fully ionized and (ii) at a different pH where both of them are partially ionized. The authors emphasized that the main advantage of the method is that it is not pH-dependent, so there is no need to know the exact pH of the buffer solutions—it is only important that the pH is identical for runs with an analyte and the internal standard. They measured acidity constants of various amines and phenols (pKa range 7.1–9.6) and compared them to the literature.11
In a study, the authors measured pKa of weak acids12 and proposed a set of 24 monoprotic weak acids of various structures as internal standards. Later, the same authors established a set of 25 basic internal standards13 and the method was extended for polyprotic compounds.14 The authors claimed the IS-CE method suitable also for sparingly soluble compounds,15 as other reference methods require the use of aqueous–organic solvent buffers and extrapolation (corrections) to obtain a thermodynamic pKa. Temperature variations in CZE were studied by the same team in 201316 with a conclusion that the IS-CE method also compensates uncontrolled temperature fluctuations (e.g., due to Joule heat) inside the capillary. The authors obtained reliable acidity constant values at the desired temperatures. Cabot et al.(17) enhanced the IS-CE method as a high-throughput method (3 min runs) by calculating pH from electrophoretic mobilities of multiple internal standards and applying pressure. Despite depletion of BGE (“buffer instability”), the authors confirmed that the method eliminates this systematic error. Later, the authors introduced an automated analyzer for pKa determination.18
The goal of this paper is to demonstrate that the IS-CE method in principle can eliminate the influence of the contingent interactions of an analyte with nonrecommended buffers and even compensate the solvation effect of an analyte in mixed solvents, which means that, within experimental precision, the method yields correct values of thermodynamic acidity constants, though the data are measured in methanol–water mixed solvent and no corrections are taken.
Theory
Many popular electrolytes used in capillary electrophoresis are Good’s buffers (derivatives of ethane–sulfonic acids), mainly for separation purposes, where the goal is to obtain the resolution of compounds with close mobilities, for example, MES, Bis-Tris, ACES, MOPS, HEPES, CHES, and TAPS.19
Several authors studied electrophoretic mobilities measured in common buffers and found that some common inorganic buffers may exhibit unpredictable migration behavior (e.g., phosphate20). Buffers suitable for pKa determination by CZE were reported by Poole et al., who recommended mostly inorganic buffers for electrophoretic pKa determination: sodium phosphate, acetate, and boric, phosphoric acid, acetic, and formic acid (for pH > 10 butylamine).2 Later, other researchers concluded that “phosphate and borate buffers should be avoided to determine the mobility of amines with aqueous pKa higher than 8, at least in solutions with high methanol content”.6 Critical evaluation of buffers for capillary electrophoresis was presented in 2008 by Fuguet et al. who did not recommend ammonium salts, organoammonium salts, and hydrogen phosphate/phosphate because they may interact with a wide range of compounds. Also, dihydrogen phosphate/hydrogen phosphate, MES, HEPES, and borates showed specific interaction.21 In 2009, Fuguet et al. suggested for pKa determination the following set of buffers: formate, acetate, Bis-Tris, CHES, and CAPS.11 Also, for pKa determination, the use of univalent anionic/cationic buffers with only one counterion (sodium/chloride) was recommended.22,23 Later, Cabot et al. observed systematic electrophoretic mobility deviations of weak bases at pH > 9 in some buffers (TAPS and CHES).12 Nevertheless, their observations proved that the IS-method showed a better performance compared to the common approach because such a deviation was compensated.
Principle of the Internal Standard Capillary Electrophoresis Method (IS-CE Method)
This method requires in principle two electrophoretic runs: a first one at a pH, where both analyte and internal standard are totally ionized (as protonated bases, pH < pKa – 2) to calculate their actual ionic mobilities and a second one at another pH where both are partially ionized (pH ≈ pKa); the mobility of the partially ionized form should be approximately 50% lower compared to the totally ionized form in order to calculate pKa correctly. As noted above, the method is not pH-dependent, so an accurate measure of the pH of the buffer solutions is not needed because the solution, where both the compounds are measured, has identical pH and composition.11
As the authors stated “One of the main advantages of using an internal standard is that some systematic errors are compensated”. The following equations will show the calculation of the IS-CE method and how it eliminates the activity coefficient correction. In an analogous manner, it can eliminate the corrections for the solvation effect in mixed solvents.
Activity Coefficient Correction
For a base, eq 1 can be rearranged introducing a variable Q > 0 (μBH+ > μeff)
| 6 |
and in combination with eq 3 we get
| 7 |
Because eq 7 holds for both the analyte (AN) and internal standard (IS), log γ is subtracted13
| 8 |
which proves that pKa(AN) is pH- and γ-independent because the activity coefficients of the buffer are identical for the IS and analyte. Such an elimination of the activity coefficient may fail at basic analytes with acidic internal standards, which was also discussed in Fuguet 2011;13 however, using a weak acid as an internal standard for pKa, determination of a base is not a common approach.
Water-Solvent pH Scale Correction
From eq 4,
can be easily estimated (calculated)
from
any experimental value
knowing the methanol volume fraction
ϕ.
Clearly, the correction δ is identical for both the internal
standard (IS) and the analyte (AN), thus after a rearrangement with
a help of 8 we get
| 9 |
which proves that
the experimental data (a
calculated difference of log Qs) will
directly give the difference of
and
without the
presence of δ because eq 9 turns into eq 8.
Mixed Solvent (Solvation) Correction
Calculation of
the coefficients aS and bS for amines gives9
and
(v is the volume fraction
of methanol in the mixture with water). As shown in Figure 1, we plot the course of eq 5 on methanol content for
two bases
= 9.48 (e.g., propranolol)
and a hypothetical base with
= 9.00.
Figure 1.
Graph of eq 5 (correction
to solvation effects for methanol–water solvent) for two hypothetical
bases. Two lower traces are plots of parameters
S (dotted line) and bS (dash-dot
line); the upper two traces show the course
of
for bases with thermodynamic
acidity constants
of
9.00 (dashed line) and 9.48 (solid
line),
respectively.
In Figure 1, one
can see (i) the coefficient aS is practically
constant and close to 1 (dotted line) and (ii) the graphs of
course for both bases (solid and
dashed
lines, resp.) decrease in parallel lines. A calculated difference
of both
is 0.48–0.46 within the
range of
0–70% (v/v) of methanol. Therefore
| 10 |
Again, this leads to an elimination of bS and practically also
. It means that
the difference of log Qs can be used for a direct
calculation of
in methanol–water solutions
because eq 10 turns
into eq 8 (within the
experimental error).
Results and Discussion
Buffer Choice
Selection of a buffer and its concentration for experiments in CE is practically limited due to Joule heating; to keep Ohḿs law valid (constant resistance of the solution), high concentrations of multiple-charged species should be avoided. In this work, the course of Ohḿs law for the buffers (concentration 0.025 M) was recorded at a continuous increase of voltage and showed deviations from a linear course for U > 15 kV.
Our starting experiments about an effect of voltage on mobility +5, +10, and +20 kV (gradient 151–606 V/cm) proved that at +20 kV, calculated electrophoretic mobilities exhibited higher values (approx. by +10%) in comparison to +10 or +5 kV (also after correction to voltage ramp24) for all the analytes and common buffers tested. Because the effect was observed also for electro-osmotic flow mobility, and even after setting the thermostat to 15 °C, it is likely that excessive Joule heating and inefficient heat dissipation caused the viscosity decrease inside the capillary, which affected the species electrophoretic movement. Despite the fact that the IS-CE method should eliminate such a shift similar to the temperature effect,16 the voltage +10 kV (where the Ohḿs plot was strictly linear) was selected for all the following runs for pKa determination in order to avoid any unpredictable migration behavior. Because the compounds studied were monoprotic bases with pKa around 9.5, pairs of buffers with pH values between 6.0 and 9.5 were always chosen (c = 25 mM).
Ammonium Buffer and Triethylamine Buffer
The acidity constant of atenolol 9.5425 is close to that of propranolol 9.48,25 so one would expect their electrophoretic mobilities to be similar, which was confirmed by experiments with all the β-blockers in the carbonate buffer (pH = 9.5) (≈+13 × 10–9 m2/V·s, data not shown). However, our additional experiments with other buffers showed that ammonium buffer pH = 9.5 exhibited systematically higher electrophoretic mobility at all the voltages for all the β-blockers, which was mostly pronounced for atenolol (≈+17 × 10–9 m2/V·s) in comparison to propranolol (≈+15 × 10–9 m2/V·s). An explanation can be the presence of the amide functional group of atenolol in contrast to propranolol.
Further experiments at +10 kV with different buffers (pH = 9.5) revealed a systematic positive shift in electrophoretic mobility (by 60–100%) of all the β-blockers in BGE of triethylamine (TEA) buffer (Figure 2, the dashed line). Another interesting systematic increase in mobility (≈3.5 × 10–9 m2/V·s) was also observed for MES at pH = 6.0 in comparison to bicarbonate at pH = 6.0, suggesting an interaction of the protonated bases with MES. This is in a general agreement with findings of Fuguet et al.(21) where the authors concluded, among others, that ammonium and alkylammonium buffers are not recommended for pKa determination by CZE (see Theory above).
Figure 2.
Electrophoretic mobilities of all the analytes at +10 kV in various buffers (10–25 mM). The dashed trace (full diamonds) of analytes in TEA is compared to other buffers of pH = 9.5 (full trace, CHES—closed triangles and carbonate—closed squares) and at pH = 6.0 (MES—open squares, bicarbonate—open circles). The lines connect points for clarity only.
An important consequence of the observations for determining the pKa of weak bases is that TEA buffer (pH = 9.5) cannot be combined with, for example, bicarbonate buffer (pH = 6.0) by the IS-CE method, as the algorithm would fail, because the electrophoretic mobility around pKa would be higher than the mobility of the fully protonated base and the variable Q becomes negative (see eq 6).
pKa Determination
Several series of measurements of six β–blockers (N = 7–14) were carried out with aqueous buffers and buffers in mixed solvents 10–50% (v/v) with propranolol as the internal standard (pKa = 9.48 of propranolol was taken as an average from a review25). The experimental values are graphically shown in Figure 3. The calculations were performed according to eq 8 without any correction to activity coefficient or solvent interactions and were statistically evaluated (Tables 1, 2).
Figure 3.
Box-and-whiskers plots with all the individual experimental values of pKa determination for six β-blockers. Values obtained in aqueous buffers are shown in (a) and those obtained in methanol–water buffers 10–50% (v/v) are in (b). All the measured data are shown at each box (N = 7–14).
Table 1. Statistical Evaluation of a Linear Fit of a Dependence of pKa on Methanol Content in BGE [10–50% (v/v)].
| R2 | p (F-test) | |
|---|---|---|
| acebutolol | 0.010 | 0.655 |
| atenolol | 0.121 | 0.243 |
| alprenolol | 0.077 | 0.318 |
| betaxolol | 0.041 | 0.419 |
| celiprolol | 0.073 | 0.249 |
| nadolol | 0.174 | 0.156 |
Table 2. Statistical Evaluation of Results in Aqueous BGE vs Methanol–Water BGEa.
| p (t-test) | p (MW U test) | |
|---|---|---|
| acebutolol | 0.51 | 0.71 |
| atenolol | 0.85 | 0.94 |
| alprenolol | 0.43 | 0.45 |
| betaxolol | 0.53 | 0.53 |
| celiprolol | 0.40 | 0.26 |
| nadolol | 0.49 | 0.28 |
Table 1 compares coefficients of determination (R2) of pKa vs methanol % (v/v) in BGE for individual β-blockers. All R2 are close to zero and p-values were always ≫0.05, which means that, at level α = 0.05, the slope was NOT significantly different from zero and there was no statistically significant correlation (Table 1).
This
finding suggests that there is no systematic change in pKa values in the mixed solvents (no increase/decrease
in pKa) depending on the methanol content
in BGE as predicted from Figure 1 (eq 5) for
of an individual base.
In Table 2, the pKa values for each β-blocker of the two groups (group 1 = aqueous buffers and group 2 = methanol–water buffers) were statistically tested by independent sample tests of equality (t-test and Mann–Whitney U test). Both the parametric and nonparametric tests proved no statistically significant differences at level α = 0.05. Therefore, both the data sets belong to the same populations and they could be pooled. Then, average acidity constants calculated from the pooled data (N = 13–22) were compared to values from the literature (Table 3).
Table 3. Comparison of the Determined pKa to the Literaturea.
| pKa values (reference) | this work | |
|---|---|---|
| propranolol | 9.53, 9.40, 9.57, 9.51, 9.32, 9.72, 9.43, 9.45, 9.23, 9.40, 9.50, 9.7, 9.45, 9.5926 | 9.48 (IS) |
| acebutolol | 9.40, 9.67, 9.67, 9.4, 9.5226 | 9.47 ± 0.01 |
| atenolol | 9.60, 9.58, 9.25, 9.56, 9.54, 9.54, 9.55, 9.6, 9.6, 9.6026 | 9.55 ± 0.02 |
| alprenolol | 9.6, 9.63, 9.6226 | 9.56 ± 0.02 |
| betaxolol | 9.21 | 9.59 ± 0.02 |
| celiprolol | 9.7 | 9.70 ± 0.01 |
| nadolol | 9.39, 9.67, 9.4 | 9.75 ± 0.03 |
pKa values in the second column were mostly found in a review,25 if not stated otherwise. The half-widths of the confidence interval in the last column were calculated according to Student (α = 0.05).
Conclusions
The results showed that triethylamine buffer cannot be recommended as a background electrolyte for measuring the pKa of weak bases by capillary electrophoresis because extreme values of electrophoretic mobility in the basic region may exceed values for electrophoretic mobility of the fully protonated form and the IS-CE algorithm fails.
If a suitable internal standard is selected, the IS-CE method can be used even for (i) other buffers that are not recommended for the traditional approach because contingent interactions with BGE can be compensated and (ii) analytes with low solubility in water because the runs can be safely performed in methanol–water mixed solvents. The latter advantage may overcome problems with acidity constant determination of many newly synthetized compounds with limited water solubility.
Based on error propagation, the experimental error of the determined acidity constant (calculated according to eqs 2, 6, and 8) is only by 0.02 higher than the uncertainty of the internal standard pKa.
Experimental Section
The CE experiments were carried out using an Agilent CE G-1600 equipped with DAD (190–600 nm) (Agilent Technologies, Waldbronn, Germany) and data software supplied by the manufacturer (Chemstation). An untreated fused silica capillary of 50 μm internal diameter (Simplus Capillaries MicroSolv, USA) was used with a total length 33 cm, effective length 24.5 cm. For buffer preparation, a pH meter Orion 370 (Thermo Electron Corp., USA) was utilized.
Chemicals were purchased from various manufacturers: nadolol, atenolol, betaxolol hydrochloride, and alprenolol hydrochloride from EDQM (Strasbourg, France), propranolol hydrochloride, acebutolol hydrochloride, CHES, MES, and mesityl oxide (MSO, an EOF marker) from Fluka (Buchs, Switzerland), sodium bicarbonate, celiprolol hydrochloride, sodium dihydrogen phosphate, citric acid, and methanol HPLC grade from Sigma Aldrich (St. Louis, MO, USA), triethyl amine (TEA), and ammonium hydroxide from Lach:ner (Czech Republic). Standards of β-blockers (propranolol, atenolol, alprenolol, nadolol, acebutolol, celiprolol, and betaxolol) of concentration 0.2 mg/mL were prepared in a buffer of pH = 6 and dissolver in an ultrasonic bath. Injection was performed for 2 s at 20 mbar, the diode-array detector was set to 240 nm (MSO maximum) and 204 nm (for several β-blockers where the wavelength 240 nm was not sensitive enough).
Various concentrations of methanol in buffers MES (pH = 6.0) and CHES (pH = 9.5) (10–50% v/v) were prepared by dissolution of the buffer salt in water and methanol.
Acknowledgments
Dr. Iva Kapustíková (Comenius University in Bratislava, Faculty of Pharmacy, Slovakia) is acknowledged for fruitful discussion on solvation correction in mixed solvents.
Author Contributions
The manuscript was written through contributions of all authors.
The authors declare no competing financial interest.
References
- Gluck S. J.; Cleveland J. A. Investigation of experimental approaches to the determination of pKa values by capillary electrophoresis. J. Chromatogr. A 1994, 680, 49–56. 10.1016/0021-9673(94)80051-0. [DOI] [Google Scholar]
- Poole S. K.; Patel S.; Dehring K.; Workman H.; Poole C. F. Determination of acid dissociation constants by capillary electrophoresis. J. Chromatogr. A 2004, 1037, 445–454. 10.1016/j.chroma.2004.02.087. [DOI] [PubMed] [Google Scholar]
- Nowak P.; Woźniakiewicz M.; Kościelniak P. Application of capillary electrophoresis in determination of acid dissociation constant values. J. Chromatogr. A 2015, 1377, 1–12. 10.1016/j.chroma.2014.12.032. [DOI] [PubMed] [Google Scholar]
- Ishihama Y.; Oda Y.; Asakawa N. Microscale Determination of Dissociation Constants of Multivalent Pharmaceuticals by Capillary Electrophoresis. J. Pharm. Sci. 1994, 83, 1500–1507. 10.1002/jps.2600831025. [DOI] [PubMed] [Google Scholar]
- Malý M.; Boublík M.; Pocrnić M.; Ansorge M.; Lorinčíková K.; Svobodová J.; Hruška V.; Dubský P.; Gaš B. Determination of thermodynamic acidity constants and limiting ionic mobilities of weak electrolytes by capillary electrophoresis using a new free software AnglerFish. Electrophoresis 2020, 41, 493–501. Article 10.1002/elps.201900283. [DOI] [PubMed] [Google Scholar]
- de Nogales V.; Ruiz R.; Rosés M.; Ràfols C.; Bosch E. Background electrolytes in 50% methanol/water for the determination of acidity constants of basic drugs by capillary zone electrophoresis. J. Chromatogr. A 2006, 1123, 113–120. 10.1016/j.chroma.2006.05.008. [DOI] [PubMed] [Google Scholar]
- Rosés M.; Bosch E. Influence of mobile phase acid-base equilibria on the chromatographic behaviour of protolytic compounds. J. Chromatogr. A 2002, 982, 1–30. 10.1016/S0021-9673(02)01444-9. [DOI] [PubMed] [Google Scholar]
- Avdeef A.; Comer J. E. A.; Thomson S. J. pH-Metric log P. 3. Glass electrode calibration in methanol-water, applied to pKa determination of water-insoluble substances. Anal. Chem. 1993, 65, 42–49. 10.1021/ac00049a010. [DOI] [Google Scholar]
- Rived F.; Canals I.; Bosch E.; Rosés M. Acidity in methanol-water. Anal. Chim. Acta 2001, 439, 315–333. 10.1016/S0003-2670(01)01046-7. [DOI] [Google Scholar]
- Ehala S.; Míšek J.; Stará I. G.; Starý I.; Kašička V. Determination of acid-base dissociation constants of azahelicenes by capillary zone electrophoresis. J. Sep. Sci. 2008, 31, 2686–2693. 10.1002/jssc.200800227. [DOI] [PubMed] [Google Scholar]
- Fuguet E.; Ràfols C.; Bosch E.; Rosés M. Fast high-throughput method for the determination of acidity constants by capillary electrophoresis. J. Chromatogr. A 2009, 1216, 3646–3651. 10.1016/j.chroma.2008.12.090. [DOI] [PubMed] [Google Scholar]
- Cabot J. M.; Fuguet E.; Ràfols C.; Rosés M. Fast high-throughput method for the determination of acidity constants by capillary electrophoresis. II. Acidic internal standards. J. Chromatogr. A 2010, 1217, 8340–8345. 10.1016/j.chroma.2010.10.060. [DOI] [PubMed] [Google Scholar]
- Fuguet E.; Ràfols C.; Rosés M. A fast high throughput method for the determination of acidity constants by capillary electrophoresis. 3. Basic internal standards. J. Chromatogr. A 2011, 1218, 3928–3934. 10.1016/j.chroma.2011.04.054. [DOI] [PubMed] [Google Scholar]
- Cabot J. M.; Fuguet E.; Ràfols C.; Rosés M. Determination of acidity constants by the capillary electrophoresis internal standard method. IV. Polyprotic compounds. J. Chromatogr. A 2013, 1279, 108–116. 10.1016/j.chroma.2013.01.018. [DOI] [PubMed] [Google Scholar]
- Cabot J. M.; Fuguet E.; Rosés M. Determination of acidity constants of sparingly soluble drugs in aqueous solution by the internal standard capillary electrophoresis method. Electrophoresis 2014, 35, 3564–3569. 10.1002/elps.201400353. [DOI] [PubMed] [Google Scholar]
- Gravador R. S.; Fuguet E.; Ràfols C.; Rosés M. Temperature variation effects on the determination of acidity constants through the internal standard-capillary electrophoresis method. Electrophoresis 2013, 34, 1203–1211. 10.1002/elps.201200584. [DOI] [PubMed] [Google Scholar]
- Cabot J. M.; Fuguet E.; Rosés M. Internal Standard Capillary Electrophoresis as a High-Throughput Method for pKa Determination in Drug Discovery and Development. ACS Comb. Sci. 2014, 16, 518–525. 10.1021/co500059p. [DOI] [PubMed] [Google Scholar]
- Cabot J. M.; Fuguet E.; Rosés M.; Smejkal P.; Breadmore M. C. Novel Instrument for Automated pKa Determination by Internal Standard Capillary Electrophoresis. Anal. Chem. 2015, 87, 6165–6172. 10.1021/acs.analchem.5b00845. [DOI] [PubMed] [Google Scholar]
- Good N. E.; Izawa S.. [3] Hydrogen ion buffers. In Methods in Enzymology; Academic Press, 1972; Vol. 24, pp 53–68. [DOI] [PubMed] [Google Scholar]
- Gebauer P.; Pantůiková P.; Bocek P. Capillary zone electrophoresis in phosphate buffer--known or unknown?. J. Chromatogr. A 2000, 894, 89–93. 10.1016/S0021-9673(00)00707-X. [DOI] [PubMed] [Google Scholar]
- Fuguet E.; Reta M.; Gibert C.; Rosés M.; Bosch E.; Ràfols C. Critical evaluation of buffering solutions for pKadetermination by capillary electrophoresis. Electrophoresis 2008, 29, 2841–2851. 10.1002/elps.200700869. [DOI] [PubMed] [Google Scholar]
- Šlampová A.; Křivánková L.; Gebauer P.; Boček P. Standard systems for measurement of pKs and ionic mobilities. J. Chromatogr. A 2008, 1213, 25–30. 10.1016/j.chroma.2008.08.101. [DOI] [PubMed] [Google Scholar]
- Šlampová A.; Křivánková L.; Gebauer P.; Boček P. Standard systems for measurement of pK values and ionic mobilities. J. Chromatogr. A 2009, 1216, 3637–3641. 10.1016/j.chroma.2009.01.018. [DOI] [PubMed] [Google Scholar]
- Williams B. A.; Vigh G. Effect of the initial potential ramp on the accuracy of electrophoretic mobilities in capillary electrophoresis. Anal. Chem. 1995, 67, 3079–3081. 10.1021/ac00113a051. [DOI] [Google Scholar]
- Brittain H. G.Profiles of Drug Substances, Excipients and Related Methodology: Critical Compilation of pKa Values for Pharmaceutical Substances; Elsevier, 2007. [DOI] [PubMed] [Google Scholar]
- Shalaeva M.; Kenseth J.; Lombardo F.; Bastin A. Measurement of Dissociation Constants (pKa Values) of Organic Compounds by Multiplexed Capillary Electrophoresis Using Aqueous and Cosolvent Buffers**Advanced Analytical Technologies, Inc. (Formerly CombiSep), 2711 South Loop Drive, Suite 4200, Ames, IA 50010. J. Pharm. Sci. 2008, 97, 2581–2606. 10.1002/jps.21287. [DOI] [PubMed] [Google Scholar]