Abstract
Given the extensive and routine use of cardiovascular devices, a major limiting factor to their success is the thrombotic rate that occurs. This both poses direct risk to the patient and requires counterbalancing with anticoagulation and other treatment strategies, contributing additional risks. Developing a better understanding of the mechanisms of device-induced thrombosis to aid in device design and medical management of patients is critical to advance the ubiquitous use and durability. Thus, mathematical and computational modelling of device-induced thrombosis has received significant attention recently, but challenges remain. Additional areas that need to be explored include microscopic/macroscopic approaches, reconciling physical and numerical timescales, immune/inflammatory responses, experimental validation, and incorporating pathologies and blood conditions. Addressing these areas will provide engineers and clinicians the tools to provide safe and effective cardiovascular devices.
Keywords: Cardiovascular, devices, thrombosis, modelling
1. Introduction
Blood-contacting medical devices are used to treat a variety of cardiovascular (CV) and cardiopulmonary diseases. These devices perturb hemostasis, resulting in complications that manifest in the patient as thrombosis and hemorrhage. Thrombosis is not only a frequent source of device failure [1], but excessive clotting and/or bleeding may also pose direct risks to the patient [2].
Nevertheless, devices have emerged as the standard-of-care for many cardiovascular disease conditions [3]. Current device use, incidence of adverse events, and material surfaces, and upcoming and novel/controversial devices, are summarized in Table 1.
Table 1.
Overview of blood-contacting medical devices
| Device | Refs | Clinical Indication for Use |
Rate of Failure or Adverse Events |
Material Surface(s) | Notes |
|---|---|---|---|---|---|
| Stents | [69,70] | Arterial Obstructions | Thrombosis <1%; restenosis 10% | metal (e.g. nitinol, stainless steel) ± coatings (e.g. carbon) | |
| Grafts | [71] | Vascular Repair or Bypass | 20-40% not patent at 5 years | saphenous vein, polymer (e.g. nylon, PTFE) | |
| Catheters | [72] | Central venous catheters 14-18%; peripherally-inserted central catheters 5-15% in-hospital | silicone rubber, polyurethane | ||
| Heart Valves | Valvular Heart Disease | ||||
| Bioprostethic Valves | [73] | Thrombosis ≤2-3%/year | fixed biologic tissues, can be on stent/graft scaffold | ||
| Mechanical Valves | [74] | Thrombosis ≤2-3%/year, up to 6%/year if insufficient anticoagulation | Requires lifelong anticoagulation | ||
| Transcatheter Aortic Valve Replacement (TAVR) | [75] | Aortic Stenosis | Thrombosis 16.9% | fixed biologic tissues on stent/graft scaffold | Increasing in use |
| Mechanical Circulatory Support (MCS) | Heart Failure | ||||
| Ventricular Assist Devices (VADs) | [76-78] | Thromboembolic events as high as 30% in HeartMate II (adult) and Berlin Heart EXCOR (pediatrics) | metals, ceramics, plastics | Decreases over time | |
| Intra-Aortic Pumps | [79] | 1% thrombosis, 27% bleeding | polyethylene, polyurethane | ||
| Cardiopulmonary Bypass | [80] | Cardiac Surgery | 2-10% myocardial infarction, 3% stroke | plastics | Many other complex complications, both surgical and related to critical illness |
| Total Artifical Hearts (TAHs) | [81,82] | 20% bleeding, 1.6% embolus, 2% stroke | polyurethane | ||
| Extracorporeal Membrane Oxygenation (ECMO) | [19,83] | Heart and/or Lung Failure | ≥90% some clot formation in circuitry; balance of bleeding risk with anticoagulation level | mainly plastics, metals | Notable pediatric patient population; encompasses many devices assembled into circuitry |
| Inferior Vena Cava (IVC) Filters | [84] | Pulmonary Embolism Risk | ≤30% thrombosis, DVT as high as 43% | metals (e.g. nitinol, stainless steel) | Intended to cause obstruction; reports of incidence vary widely |
| Endovascular Coils | [85] | Cerebral Aneurysm | 9.1% total complication rate, including rebleeding, ischemia, and rupture | metals (e.g. platinum alloy) | Intended to cause thrombosis |
| Resuscitative Endovascular Balloon Occlusion of the Aorta (REBOA) | [86] | Hemorrhage | Unclear; may increase mortality | plastics, employed by catheter/guidewire | Use still controversial |
| Dialysis | [87] | Kidney Disease | <1%/year | plastics | |
| Micro- and Nanoparticles | [88,89] | Thrombosis | n/a | charged polymers | Novel |
While disease etiologies and patient populations vary among blood-contacting devices and their applications, they all share the same overarching challenges: (1) device-induced thrombosis is not fully understood, and therefore is clinically unpredictable, and (2) clinical trials can be extremely difficult and risky in these patient populations, sometimes to the point of being prohibitively so. Therefore, in silico studies are a critical tool to complement in vitro, ex vivo, and in vivo studies to contribute to overcoming these challenges. Device thrombosis modelling can be used to inform on device design in the preclinical stage, and to inform on treatment and clinical handling of current patients, by developing guidelines and on demand treatments.
In this review, the initial focus is a discussion on the pathophysiology of device thrombosis. We then assess modelling efforts posterior to Fogelson and Neeves’ 2015 review [4], with a focus on model assumptions to facilitate discourse on the physiological relevance of different mathematical and computational approaches.
2. Device Perturbation of Hemostasis: Virchow’s Triad
As is widely appreciated, Virchow theorized, rather correctly, that hemostasis is a delicate balance between the blood state, surface, and flow [5]. The original Virchow’s Triad comprised stasis, vessel injury, and hypercoagulability. While there have been recent adaptations to his theory, the basic concept remains applicable to devices via a more liberal interpretation of the original construct, which we have summarized in Figure 1.
Figure 1. Virchow’s Triad for Device Thrombosis.
The driving mechanisms of device thrombosis can be viewed through the lense of Virchow’s Triad. The blood state, device surface, and device flow patterns have concomitant effects that ultimately result in thrombosis.
2.1. Blood state
Blood is a suspension containing erythrocytes, leukocytes, platelets, factors, ions, and glycoproteins, which all have critical roles to maintain hemostasis and perform other physiological functions [6]. Coagulation, a critical physiological function, occurs by initiation via either vascular-injury or foreign-body contact and is comprised of dozens of cellular species and nearly 100 reactions to form a stable thrombus [7]. With the introduction of a device in the CV system, the surface-mediated pathway (foreign-body) is stimulated along with the activation of the tissue-mediated pathway (vascular injury) as the endothelium is disrupted by the device during implantation. Whether these events are separate or simultaneous remains unknown, and while the question may be moot since both lead to thrombosis, elucidation of the interactions and timeline could inform therapeutic inhibition of either or both pathways.
In addition to coagulation, platelets are significant contributors to clotting and interface with other cellular components such as von Willebrand Factor (VWF) and fibrin(ogen).Their impact on devices can be substantial, including total device failure. Given the mechanical and chemical sensitivities of platelets, their response is critical to gauge device thrombogenicity. Platelets can be stimulated by high shear, the surface-mediated pathway, and activated by biochemical means [8]. For example, when platelets attach to VWF adsorbed to the device surface, this creates a foundational support for the clot (along with contributions from other clotting factors). VWF can also be compromised due to degradation, as has been shown in ECMO and VAD patients that develop bleeding events [9].
These cellular components are currently thought to have the most significant impact on devices, but as this area continues to be studied, more may be revealed. Thus the blood state, which includes its constituents, is quite integral to determining device success. However, the device surface itself can dramatically influence clotting.
2.2. Surface
When blood contacts a biomaterial surface, a series of complex systems initiate that potentiate thrombosis via the pathways described in 2.1. Proteins are adsorbed immediately to the surface [1,10,11], a process which is controlled initially by diffusion but dominated by protein-surface affinity over time (Vroman Effect, [12]). Device materials are typically hydrophobic, and have affinity with many proteins [13]. The most commonly adsorbed plasma proteins are albumin, fibrinogen, immunoglobulin G (IgG), fibronectin, and VWF. As mentioned previously, VWF and fibrinogen appear to be the most critical for platelet activity [14]. In the adsorption process, proteins undergo conformational changes to expose hydrophobic domains, which results in expression of receptor sites, causing subsequent immune crosstalk, and further potentiating thrombosis [15-17]. The phenotype of the end surface is dependent on the biomaterial, thus a multitude of unique surfaces exists.
However, the device surface is absorbed in its entirety, yet thrombosis tends to be localized [18,19]. Flow is thus the final key mechanistic driver.
2.3. Flow
Flow and its characteristics have been demonstrated to be quite potent in determining the effectiveness and long-term viability of a device. Fundamentally, flow regimes (i.e. laminar, transitional, and turbulent) can affect cellular responses by creating pathologically low and high stress regions, whereby shear stress can induce platelet activation [3,20], the combination of shear stress and its exposure time can induce hemolysis and thus potentiate coagulation [21], and recirculating flows and wakes can facilitate clotting [22,23]. Hemostatic and thrombotic mechanisms in vivo proceed via mechanisms that are flow dependent, resulting in gel-like coagulation “red” clots at low shear, and brittle platelet-rich “white” clots at high shear [24,25]. Experimental studies have demonstrated the impact of shear stress on platelet activation/aggregation [26-28] and thrombosis [29]. Pathologic flows created by the presence of devices and actions associated with them can influence hemostasis with examples of recirculating flows/wakes in ECMO circuitry and IVC filters, and extremely high shear stress associated with prosthetic heart valves and VADs (examples found in Table 1). Residence time of platelets and factors is another crucial factor that may enhance clotting [26].
Flow, the blood state, the biomaterial surface, and their concomitant effects are thus the essential elements to understanding the unique qualities of device thrombosis (Figure 1), and therefore determine their ultimate success.
3. Modelling Approaches Available
Despite much effort and tremendous progress, device thrombosis modelling remains extremely challenging due to the variety and complexity of the phenomena. Recently, comprehensive modelling studies (Table 2) have extensively covered hemostasis and thrombosis in the setting of the vasculature, and in the context of natural physiology and pathology [30]. In fact, some studies have demonstrated highly fidelic models of thrombosis in the setting of high shear [31]. However, as detailed in Section 2, device thrombosis has additional mechanisms which must be considered in the modelling effort, making the challenge even greater. Device modelling and understanding remains less elucidated than in vivo hemostasis and thrombosis.
Table 2.
Summary of recent approaches and assumptions
| Refs | Flow domain |
Coagulat ion |
Platelets | Clot | Contact system |
Injury site |
Gap to device | |
|---|---|---|---|---|---|---|---|---|
| High shear | [50] | 3D NS 10−3 m |
none | Particles | no | no | yes | Domain size // biochemistry // contact //clot |
| [20] | 3D DPD 10−5 m |
none | CGMD | no | no | yes | Domain size // biochemistry // contact // clot | |
| [51] | 3D DPD 10−4 m |
none | Particles | yes | no | yes | Domain size // biochemistry // contact // Ventricular assist devices | |
| [60] | 3D NS 10−2 - 10−1 m |
none | Pseudo particles |
yes | no | yes | Biochemistry // contact | |
| Low shear | [47,48] | 2D DPD 10−4 m |
8 species | none | yes | no | yes | 3D // domain size // contact system // platelets |
| [49] | 2D NS 10−2 m |
20 species | none | no | yes | no | 3D // platelets // clot // Generally device related | |
| All flow regimes | [55] | 2D NS 10−2 m |
1 species | 2 scalar quantities | yes | no | no | 3D // biochemistry // clot retraction on flow // Generally device related |
| [90] | 2D DPD 10−4 m |
3 species | Particles | yes | no | yes | 3D // domain size // contact // platelet-protein coupling | |
| [18,56,57] | 3D NS 10−1 m |
5 species | 5 scalar quantities | yes | no | no | Contact // Ventricular assist devices | |
| [54] | 3D NS 10−2 - 10−1 m |
1 species | 3 scalar quantities | no | no | no | Biochemistry // clot retraction | |
| [52] | 3D NS 10−4 m |
20 species | Particles | yes | yes w/o XII | yes | Domain size // FXII | |
| [91] | 2D NS 10−4 m |
8 species | Particles | yes | no | yes | 3D // domain size // contact | |
| [58] | 3D NS 10−2 - 10−1 m |
5 species | 3 scalar quantities | yes | no | no | Contact // Flow diverters | |
| [59] | 3D NS 10−2 m |
5 species | 2 scalar quantities | yes | no | no | Specific to stenosis geometries // contact |
3.1. Biomarkers Deduced from Pure CFD Results
The device’s presence in the CV system leads to thrombus formation after an unknown delay, which may be hours, days, or even months. Thus, to maximize clinical and/or design utility, a computational tool for device thrombosis should forecast clotting, and predict if, when, and where thrombosis occurs. Because some processes operate at very short time scales (1 ms, say, flow convection or fast biochemical reactions), practical approaches often neglect all the “details” and try to guess the thrombotic response from the analysis of the stationary flow field obtained from classical computational fluid dynamics (CFD). Typical fluid mechanics markers used to predict thrombosis in devices include: low velocity region [32]; residence time [33-38]; flow recirculation [39]; low wall shear stress [34,40,41]; washout [36,37]; kinetic energy density [37]; (platelets) stress accumulation [35,38,42]; (platelets) shear/convection/aggregation [43], and activation indices [44]. These approaches can provide relevant information at low cost about the flow structure and how geometric changes can reduce thrombogenicity [45]. Of course, they do not inform about the size, evolution, thrombus quality, and the clinical result.
An extension of the pure CFD approach was proposed [31,46] in order to predict the time of occlusion in high velocity stenotic channels. In this situation, VWF adsorption on the artificial surface, platelet adhesion via GPIb-VWF, and subsequent activation and aggregation of flowing platelets, lead to thrombus formation. Assuming that these processes can be abstracted by a correlation between the local hemodynamic shear rate and the thrombus growth rate, this approach relies on a set of pure CFD simulations to predict the occlusion time [31,46]. Of course, the approach can only be justified if the shear rate-growth rate correlation indeed exists, and if it can be calibrated in advance.
3.2. Physical-based models
Simulations aiming at describing thrombosis should include models for the coagulation cascade [47-49], and platelet dynamics [20,50,51]. Given the numerous interactions between fibrin formation and platelets during thrombosis, proper in silico models should consider these two ingredients. One of the most advanced formulation to date is likely that of Yazdani et al. [52], where a set of 24 partial differential equations (PDEs) are resolved for fluid flow, 20 biochemical species interacting in the clotting cascade are implemented (following the pattern set forth by Anand et al. [53]) with a set of Lagrangian particles (platelets) that can be chemically activated and produce agonists. The key element that makes this Eulerian-Lagrangian approach capable of representing thrombosis in low and high shear flows is a shear-dependent platelet adhesive model set to correctly reproduce data in vivo and in vitro. Surprisingly, FXII, whose activation on the artificial surface may trigger the intrinsic pathway, was not considered by Yazdani et al. [52]; instead, thrombus formation was triggered by imposing non-zero concentration of the TF-VIIa complex at a site of injury. The initiation of thrombin formation (and therefore fibrin) without explicitly defining an injury site, but rather by relying solely on the activation of factor XII, was made by Méndez Rojano et al. [49], where the backward facing step experiment was computed. A significant concentration of thrombin was present only in the region where thrombus formation had been observed experimentally by Taylor et al. [23], suggesting the flow-contact system interaction is a key element in thrombus location within devices, at least in the low shear regime.
The Eulerian-Lagrangian method developed in [52] is too computationally demanding to be applied to actual devices; the fundamental reason is the tremendous number of particles (1.5 μm each) that must be accumulated to form a thrombus of a typical size of 1 cm. A way to significantly reduce the overall computational load is to represent platelets by Eulerian fields [18,54-59], instead of particles. Wu et al. [18] applied 14 coupled PDEs: 4 for fluid flow, 5 for biochemical species to represent coagulation, and 5 for different platelet types. This approach has shown good potential for complex devices such as the Heartmate II VAD.
Recently, the concept of pseudo-platelets was introduced [60] to reduce the computational effort while keeping the Lagrangian description of platelets, by far, more realistic than the Eulerian framework. In this view, the spherical particles transported by the flow change their size (from 1.5 to 90 μm) once activated and adhered. This reduces the number of particles required to form the thrombus (by a factor up to (90/1.5)3 = 216000) and significantly reduces the associated computational load. This numerical treatment, although not physically based, produces reasonable thrombus formation according to Zheng et al. [60] and allows increasing size of the affordable computational domains from sub-millimetric, as implemented in [52], to centimetric [60]. In the latter study, a mapping from the particle aggregation step to a phase-field representation of the thrombus material properties and permeability was also first proposed. This methodology opens new perspectives in terms of thromboembolic event prediction, although necessary adaptations to accommodate device thrombosis remain.
4. Future Directions
As efforts continue to develop device-induced models, it should be noted that the U.S. Food and Drug Administration and other regulatory bodies, are keenly interested, and increasingly likely, to accept properly validated models as part of the device approval process. With this context, the future directions to facilitate impactful device modelling is paramount.
4.1. Microscopic vs. Macroscopic Approaches
Currently, there are two modelling approaches to device-induced thrombosis: either considering the individual cells and factors (microscopic) or considering the domain as a continuum and the cells as bulk concentrations (macroscopic). Both approaches have advantages and disadvantages with the associated limitations: the microscopic approach offers a more reliable description of the cell-based phenomena at the cost of (very) large computation load, while the macroscopic view is more efficient but assumes the interactions happen at a “gross” scale, losing the track of the individual interactions. However, developing a synergistic strategy and collaboration between the macroscopic approaches that could implement the microscopic, or even nanoscopic, results could yield significant progress to capture the truly salient features for thrombosis [61]. Within this context, careful consideration of the intent (and setting) of the proposed device is significant, considering endovascular coils aim to clot aneurysms while mechanical circulatory support devices must avoid thrombosis.
4.2. Reconciling Physical and Numerical Timescales
Thrombosis may be visible days, weeks, or months after device implantation, while multiphysics simulations representing flow-coagulation-platelets-surface interactions can only represent fractions of seconds or minutes depending on the model’s spatial scales. On top of purposely increasing reaction rates and/or diffusivity coefficients [18], current efforts to reduce the complexity of kinetics schemes [62,63] or model the near wall transport phenomena [64] may improve the situation, but will most probably be insufficient to fill the gap. A proper way to drastically increase the physical time that a simulation can demonstrate would be to have access to the thrombus growth rate. Then, the entire process could be represented by a set of simulations performed at different stages over the thrombus evolution, using the last computed growth rate to extrapolate from one instant to the following one. This strategy was successfully followed by Mehrabadi et al. [31] in the very simple case where the growth rate can be inferred from the local shear rate issued by pure CFD. This approach could be made more general by leveraging a multiphysics simulation at each step. Thus, future efforts should focus on modelling thrombus growth rate rather than simply their size.
4.3. Immune/Inflammatory
The persistent inflammatory stimuli resulting from blood contact with the device surface induces a perpetual immune response [15]. Adsorbed proteins, in addition to regulating activation of coagulation and platelets, also regulate the activation of complement and immune cells. Fibrinogen on the biomaterial surface can activate circulating monocytes, initiating the inflammatory response [15]. The presence of factors on the membranes of these and other cells can also contribute to activation and propagation of coagulation on the device surfaces. Complement is also activated via binding to the adsorbed surface [17], generating C3a and C5a at the device site [15], and both contribute to further inflammatory activation and stimulate coagulation. Devices can increase circulating platelet-leukocyte aggregates (PLAS, [65]), which are associated with thromboinflammatory diseases [65,66] and linked to cardiovascular events [67]. Including some description of the complement and inflammatory systems, and their crosstalk with thrombosis, should be considered in future modelling efforts.
4.4. Validation
Results from in vitro and in vivo experiments, as well as from clinical data, are necessary both to validate in silico efforts and to provide biological and mechanistic information as inputs into such models. The gold standard preclinical experiment is a large animal model with appropriately long timescales, however, this can be both practically challenging and expensive. Benchtop in vitro studies thus have less logistical and financial burden for implementation, thereby increasing throughput, but can encounter the same challenges discussed in 4.1 and 4.2: timescale and selection of parameters and endpoints. Both types of experiments can provide useful information to identify thrombogenic hotspots. Clinical data on the location and incidence of thrombosis in devices is ideal for validation but can be challenging to obtain. New methodologies for obtaining validation data, both in preclinical experiments and from device patients, are needed.
4.5. Incorporation of Pathologies and Blood State
Modelling is most frequently implemented with idealized parameters. Biological variability and patient pathophysiology may play key roles in the lack of translation of modelling results to the patient. Patients will likely be treated with pharmacological agents that modify hemostatic and thrombotic function, e.g. virtually all MCS patients will be anticoagulated. In addition, devices can alter the blood state, induce pathologies, and perturb the endothelial environment during implantation. This endothelial disruption can cause deleterious effects like restenosis of coronary stents and potentiate platelet activation and thrombus formation [68]. Incorporation of these clinically relevant blood states may increase the translatablitly of modelling efforts.
4.6. Conclusion
Due to the inherent complexity of the cardiovascular system, thrombosis modelling of high integrity is challenging. However, while the current tangible application is device design, computationally-guided patient care is the “science fiction” of today but will be the needed “reality” in the future.
Acknowledgements
K. Manning acknowledges support, in part, by U.S. National Institutes of Health grant HL136369. F. Nicoud acknowledges support by ANR (the French National Research Agency) under the "Investissements d’avenir" programme with the reference ANR-16-IDEX-0006.
Abbreviations
- CFD
computational fluid dynamics
- CV
cardiovascular
- ECMO
extracorporeal membrane oxygenation
- FXII
Factor XII
- GPIb
glycoprotein Ib
- IgG
immunoglobulin G
- IVC
inferior vena cava
- MCS
mechanical circulatory support
- PDE
partial differential equation
- REBOA
resuscitative endovascular balloon occlusion of the aorta
- TAH
total artificial heart
- TAVR
transcatheter aortic valve replacement
- VAD
ventricular assist device
- VWF
von Willebrand Factor
Footnotes
Declaration of Interest
None
Publisher's Disclaimer: This is a PDF file of an unedited manuscript that has been accepted for publication. As a service to our customers we are providing this early version of the manuscript. The manuscript will undergo copyediting, typesetting, and review of the resulting proof before it is published in its final form. Please note that during the production process errors may be discovered which could affect the content, and all legal disclaimers that apply to the journal pertain.
References
- 1.Jaffer IH, Fredenburgh JC, Hirsh J, Weitz JI: Medical device-induced thrombosis: what causes it and how can we prevent it? J Thromb Haemost 2015, 13:S72–S81. [DOI] [PubMed] [Google Scholar]
- 2.Susen S, Rauch A, Van Belle E, Vincentelli A, Lenting PJ: Circulatory support devices: Fundamental aspects and clinical management of bleeding and thrombosis. J Thromb Haemost 2015, 13:1757–1767. [DOI] [PubMed] [Google Scholar]
- 3.Slepian MJ, Sheriff J, Hutchinson M, Tran P, Bajaj N, Garcia JGN, Scott Saavedra S, Bluestein D: Shear-mediated platelet activation in the free flow: Perspectives on the emerging spectrum of cell mechanobiological mechanisms mediating cardiovascular implant thrombosis. J Biomech 2017, 50:20–25. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 4.Fogelson AL, Neeves KB: Fluid Mechanics of Blood Clot Formation. Annu Rev Fluid Mech 2015, 47:377–403. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 5.Virchow R: Gesammelte Abhandlungen Zur Wissenschaftlichen Medicin. 1856. [Google Scholar]
- 6.Fåhraeus R: The suspension stability of the blood. Physiol Rev 1929, 9:399–431. [Google Scholar]
- 7.Smith SA, Travers RJ, Morrissey JH: How it all starts: Initiation of the clotting cascade. Crit Rev Biochem Mol Biol 2015, 50:326–336. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 8.Yun SH, Sim EH, Goh RY, Park JI, Han JY: Platelet activation: The mechanisms and potential biomarkers. Biomed Res Int 2016, 2016. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 9.Rauch A, Susen S, Zieger B: Acquired von Willebrand syndrome in patients with ventricular assist device. Front Med 2019, 6. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 10.Basmadjian D, Sefton MV., Baldwin SA: Coagulation on biomaterials in flowing blood: Some theoretical considerations. Biomaterials 1997, 18:1511–1522. [DOI] [PubMed] [Google Scholar]
- 11.Sperling C, Fischer M, Maitz MF, Werner C: Blood coagulation on biomaterials requires the combination of distinct activation processes. Biomaterials 2009, 30:4447–4456. [DOI] [PubMed] [Google Scholar]
- 12.Vroman L, Adams AL: Identification of rapid changes at plasma–solid interfaces. J Biomed Mater Res 1969, 3:43–67. [DOI] [PubMed] [Google Scholar]
- 13.Thevenot P, Hu W, Tang L: Surface chemistry influences implant biocompatibility. Curr Top Med Chem 2008, 8:270–80. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 14.Grunkemeier JM, Tsai WB, Horbett TA: Co-adsorbed fibrinogen and von Willebrand factor augment platelet procoagulant activity and spreading. J Biomater Sci Polym Ed 2001, 12:1–20. [DOI] [PubMed] [Google Scholar]
- 15.Labarrere CA, Dabiri AE, Kassab GS: Thrombogenic and Inflammatory Reactions to Biomaterials in Medical Devices. Front Bioeng Biotechnol 2020, 8. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 16.Gorbet MB, Sefton MV.: Biomaterial-associated thrombosis: Roles of coagulation factors, complement, platelets and leukocytes. Biomater Silver Jubil Compend 2004, 25:5681–5703. [DOI] [PubMed] [Google Scholar]
- 17.Franz S, Rammelt S, Scharnweber D, Simon JC: Immune responses to implants - A review of the implications for the design of immunomodulatory biomaterials. Biomaterials 2011, 32:6692–6709. [DOI] [PubMed] [Google Scholar]
- 18.Wu WT, Yang F, Wu J, Aubry N, Massoudi M, Antaki JF: High fidelity computational simulation of thrombus formation in Thoratec HeartMate II continuous flow ventricular assist device. Sci Rep 2016, 6:1–11. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 19.Hastings SM, Ku DN, Wagoner S, Maher KO, Deshpande S: Sources of Circuit Thrombosis in Pediatric Extracorporeal Membrane Oxygenation. ASAIO J 2017, 63:86–92. [DOI] [PubMed] [Google Scholar]
- 20.Gupta P, Zhang P, Sheriff J, Bluestein D, Deng Y: A Multiscale Model for Recruitment Aggregation of Platelets by Correlating with In Vitro Results. Cell Mol Bioeng 2019, 12:327–343. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 21.Litvinov RI, Weisel JW: Role of red blood cells in haemostasis and thrombosis. ISBT Sci Ser 2017, 12:176–183. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 22.Yang L, Neuberger T, Manning KB: In vitro real-time magnetic resonance imaging for quantification of thrombosis. Magn Reson Mater Physics, Biol Med 2020, doi: 10.1007/s10334-020-00872-2. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 23.Taylor JO, Witmer KP, Neuberger T, Craven B a, Meyer RS, Deutsch S, Manning KB: In vitro quantification of time dependent thrombus size using magnetic resonance imaging and computational simulations of thrombus surface shear stresses. J Biomech Eng 2014, 136:1–11. [DOI] [PubMed] [Google Scholar]
- 24.Houdijk W, de Groot P, Nievelstein P, Sakariassen K, Sixma J: Subendothelial proteins and platelet adhesion. von Willebrand factor and fibronectin, not thrombospondin, are involved in platelet adhesion to extracellular matrix of human vascular endothelial cells. Arteriosclerosis 1986, 6:24–33. [DOI] [PubMed] [Google Scholar]
- 25.Cadroy Y, Horbett T, Hanson S: Discrimination between platelet-mediated and coagulation-mediated mechanisms in a model of complex thrombus formation in vivo. J Lab Clin Med 1989, 113:436–448. [PubMed] [Google Scholar]
- 26.Nobili M, Sheriff J, Morbiducci U, Redaelli A, Bluestein D: Platelet Activation Due to Hemodynamic Shear Stresses: Damage Accumulation Model and Comparison to In Vitro Measurements. ASAIO J 2008, 54:64–72. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 27.Shi X, Yang J, Huang J, Long Z, Ruan Z, Xiao B, Xi X: Effects of different shear rates on the attachment and detachment of platelet thrombi. Mol Med Rep 2016, 13:2447–2456. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 28.Navitsky MA, Taylor JO, Smith AB, Slattery MJ, Deutsch S, Siedlecki CA, Manning KB: Platelet adhesion to polyurethane urea under pulsatile flow conditions. Artif Organs 2014, 38:1046–1053. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 29.Casa LDC, Deaton DH, Ku DN: Role of high shear rate in thrombosis. J Vase Surg 2015, 61:1068–1080. [DOI] [PubMed] [Google Scholar]
- 30.Belyaev AV, Dunster JL, Gibbins JM, Panteleev MA, Volpert V: Modeling thrombosis in silico : Frontiers , challenges , unresolved problems and milestones. Phys Life Rev 2018, 1:1–39. [DOI] [PubMed] [Google Scholar]
- 31.Mehrabadi M, Casa LDC, Aidun CK, Ku DN: A Predictive Model of High Shear Thrombus Growth. Ann Biomed Eng 2016, 44:2339–2350. [DOI] [PubMed] [Google Scholar]
- 32.Bounouib M, Benakrach H, Es-Sadek Zeriab M, Taha-Janan M, Maazouzi W: Numerical study of a new ventricular assist device. Artif Organs 2020, 44:604–610. [DOI] [PubMed] [Google Scholar]
- 33.Zhang J, Chen Z, Griffith BP, Wu ZJ: Computational characterization of flow and blood damage potential of the new maglev CH-VAD pump versus the HVAD and HeartMate II pumps. Int J Artif Organs 2020, doi: 10.1177/0391398820903734. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 34.Chen HY, shin Tien W, Chambers SD, Dabiri D, Kassab GS: Search for an Optimal Design of a Bioprosthetic Venous Valve: In silico and in vitro Studies. Eur J Vasc Endovasc Surg 2019, 58:112–119. [DOI] [PubMed] [Google Scholar]
- 35.Chivukula VK, Beckman JA, Prisco AR, Dardas T, Lin S, Smith JW, Mokadam NA, Aliseda A, Mahr C: Left Ventricular Assist Device Inflow Cannula Angle and Thrombosis Risk. Circ Hear Fail 2018, 11. [DOI] [PubMed] [Google Scholar]
- 36.Liao S, Neidlin M, Li Z, Simpson B, Gregory SD: Ventricular flow dynamics with varying LVAD inflow cannula lengths: In-silico evaluation in a multiscale model. J Biomech 2018, 72:106–115. [DOI] [PubMed] [Google Scholar]
- 37.Liao S, Wu EL, Neidlin M, Li Z, Simpson B, Gregory SD: The Influence of Rotary Blood Pump Speed Modulation on the Risk of Intraventricular Thrombosis. Artif Organs 2018, 42:943–953. [DOI] [PubMed] [Google Scholar]
- 38.Aliseda A, Chivukula VK, McGah P, Prisco AR, Beckman JA, Garcia GJM, Mokadam NA, Mahr C: LVAD Outflow Graft Angle and Thrombosis Risk. ASAIO J 2017, 63:14–23. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 39.De Nisco G, Gallo D, Siciliano K, Tasso P, Lodi Rizzini M, Mazzi V, Calò K, Antonucci M, Morbiducci U: Hemodialysis arterio-venous graft design reducing the hemodynamic risk of vascular access dysfunction. J Biomech 2020, 100. [DOI] [PubMed] [Google Scholar]
- 40.Ghodrati M, Maurer A, Schlöglhofer T, Khienwad T, Zimpfer D, Beitzke D, Zonta F, Moscato F, Schima H, Aigner P: The influence of left ventricular assist device inflow cannula position on thrombosis risk. Artif Organs 2020, doi: 10.1111/aor.13705. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 41.Buck AKW, Groszek JJ, Colvin DC, Keller SB, Kensinger C, Forbes R, Karp S, Williams P, Roy S, Fissell WH: Combined in silico and in vitro approach predicts low wall shear stress regions in a hemofilter that correlate with thrombus formation in vivo. ASAIO J 2018, 64:211–217. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 42.Buck AKW, Goebel SG, Goodin MS, Wright NJ, Groszek JJ, Moyer J, Singh S, Bluestein D, Fissell WH, Roy S: Platelet stress accumulation analysis to predict thrombogenicity of an artificial kidney. J Biomech 2018, 69:26–33. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 43.Stiehm M, Wüstenhagen C, Siewert S, Ince H, Grabow N, Schmitz KP: Impact of strut dimensions and vessel caliber on thrombosis risk of bioresorbable scaffolds using hemodynamic metrics. Biomed Tech 2019, 64:251–262. [DOI] [PubMed] [Google Scholar]
- 44.Zlobina KE, Guria GT: Platelet activation risk index as a prognostic thrombosis indicator. Sci Rep 2016, 6. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 45.Chen HY, Diaz JA, Lurie F, Chambers SD, Kassab GS: Hemodynamics of venous valve pairing and implications on helical flow. J Vasc Surg Venous Lymphat Disord 2018, 6:517–522.e1. [DOI] [PubMed] [Google Scholar]
- 46.Ku DN, Casa LDC, Hastings SM: Choice of a hemodynamic model for occlusive thrombosis in arteries. J Biomech 2017, 50:110–113. [DOI] [PubMed] [Google Scholar]
- 47.Bouchnita A, Tosenberger A, Volpert V: On the regimes of blood coagulation. Appl Math Lett 2016, 51:74–79. [Google Scholar]
- 48.Bouchnita A, Galochkina T, Kurbatova P, Nony P, Volpert V: Conditions of microvessel occlusion for blood coagulation in flow. Int j numer method biomed eng 2017, 33:e2850. [DOI] [PubMed] [Google Scholar]
- 49.Méndez Rojano R, Mendez S, Nicoud F: Introducing the pro-coagulant contact system in the numerical assessment of device-related thrombosis. Biomech Model Mechanobiol 2018, 17:815–826. [DOI] [PubMed] [Google Scholar]
- 50.Tomita A, Tamura N, Nanazawa Y, Shiozaki S, Goto S: Development of virtual platelets implementing the functions of three platelet membrane proteins with different adhesive characteristics. J Atheroscler Thromb 2015, 22:201–210. [DOI] [PubMed] [Google Scholar]
- 51.Wang L, Chen Z, Zhang J, Zhang X, Wu ZJ: Modeling Clot Formation of Shear-Injured Platelets in Flow by a Dissipative Particle Dynamics Method. Bull Math Biol 2020, 82. [DOI] [PubMed] [Google Scholar]
- 52.Yazdani A, Li H, Humphrey JD, Karniadakis GE: A General Shear-Dependent Model for Thrombus Formation. 2017, doi: 10.1371/journal.pcbi.1005291. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 53.Anand M, Rajagopal K, Rajagopal KR: A model for the formation, growth, and lysis of clots in quiescent plasma. A comparison between the effects of antithrombin III deficiency and protein C deficiency. J Theor Biol 2008, 253:725–738. [DOI] [PubMed] [Google Scholar]
- 54.Taylor JO, Yang L, Deutsch S, Manning KB: Development of a platelet adhesion transport equation for a computational thrombosis model. J Biomech 2017, 50:114–120. [DOI] [PubMed] [Google Scholar]
- 55.Taylor JO, Meyer RS, Deutsch S, Manning KB: Development of a computational model for macroscopic predictions of device-induced thrombosis. Biomech Model Mechanobiol 2016, doi: 10.1007/s10237-016-0793-2. [DOI] [PubMed] [Google Scholar]
- 56.Wu WT, Zhussupbekov M, Aubry N, Antaki JF, Massoudi M: Simulation of thrombosis in a stenotic microchannel: The effects of vWF-enhanced shear activation of platelets. Int J Eng Sci 2020, 147. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 57.Wu WT, Jamiolkowski MA, Wagner WR, Aubry N, Massoudi M, Antaki JF: Multi-Constituent Simulation of Thrombus Deposition. Sci Rep 2017, 7. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 58.Sarrami-Foroushani A, Lassila T, Hejazi SM, Nagaraja S, Bacon A, Frangi AF: A computational model for prediction of clot platelet content in flow-diverted intracranial aneurysms. J Biomech 2019, 91:7–13. [DOI] [PubMed] [Google Scholar]
- 59.Hosseinzadegan H, Tafti DK: A Predictive Model of Thrombus Growth in Stenosed Vessels with Dynamic Geometries. J Med Biol Eng 2019, 39:605–621. [Google Scholar]
- 60.Zheng X, Yazdani A, Li H, Humphrey JD, Karniadakis GE: A three-dimensional phase-field model for multiscale modeling of thrombus biomechanics in blood vessels. PLoS Comput Biol 2020, 16. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 61.Zhang P, Zhang L, Slepian MJ, Deng Y, Bluestein D: A multiscale biomechanical model of platelets: Correlating with in-vitro results. J Biomech 2017, 50:26–33. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 62.Hansen KB, Shadden SC: Automated reduction of blood coagulation models. Int j numer method biomed eng 2019, 35. [DOI] [PubMed] [Google Scholar]
- 63.Mendez Rojano R, Mendez S, Lucor D, Nicoud F: Sensitivity Analysis, Model Reduction and Parameter Estimation for Thrombosis Modeling for CFD. In SIAM Conference on Computational Science and Engineering. .2019. [Google Scholar]
- 64.Hansen KB, Shadden SC: A reduced-dimensional model for near-wall transport in cardiovascular flows. Biomech Model Mechanobiol 2016, 15:713–722. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 65.Radovancevic R, Matijevic N, Bracey AW, Radovancevic B, Elayda M, Gregoric ID, Frazier OH: Increased leukocyte-platelet interactions during circulatory support with left ventricular assist devices. ASAIO J 2009, 55:459–464. [DOI] [PubMed] [Google Scholar]
- 66.Finsterbusch M, Schrottmaier WC, Kral-Pointner JB, Salzmann M, Assinger A: Measuring and interpreting platelet-leukocyte aggregates. Platelets 2018, 29:677–685. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 67.Liu C, Yang Y, Du L, Chen S, Zhang J, Zhang C, Zhou J: Platelet-leukocyte aggregate is associated with adverse events after surgical intervention for rheumatic heart disease. Sci Rep 2019, 9:1–10. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 68.Tesfamariam B: Endothelial Repair and Regeneration Following Intimal Injury. J Cardiovasc Transl Res 2016, 9:91–101. [DOI] [PubMed] [Google Scholar]
- 69.Ohkata I: Medical applications of superelastic nickel-titanium (Ni–Ti) alloys. In Shape Memory and Superelastic Alloys. . Elsevier; 2011:176–196. [Google Scholar]
- 70.Byrne RA, Joner M, Kastrati A: Stent thrombosis and restenosis: What have we learned and where are we going? the Andreas Grüntzig Lecture ESC 2014. Eur Heart J 2015, 36:3320–3331. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 71.Park KM, Kim YW, Yang SS, Kim DI: Comparisons between prosthetic vascular graft and saphenous vein graft in femoro-popliteal bypass. Ann Surg Treat Res 2014, 87:35–40. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 72.Wall C, Moore J, Thachil J: Catheter-related thrombosis: A practical approach. J Intensive Care Soc 2016, 17:160–167. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 73.Rashid HN, Gooley RP, Nerlekar N, Ihdayhid AR, McCormick LM, Nasis A, Cameron JD, Brown AJ: Bioprosthetic aortic valve leaflet thrombosis detected by multidetector computed tomography is associated with adverse cerebrovascular events: A meta-analysis of observational studies. EuroIntervention 2018, 13:e1748–e1755. [DOI] [PubMed] [Google Scholar]
- 74.Whitlock RP, Sun JC, Fremes SE, Rubens FD, Teoh KH: Antithrombotic and thrombolytic therapy for valvular disease: Antithrombotic therapy and prevention of thrombosis, 9th ed: American college of chest physicians evidence-based clinical practice guidelines. Chest 2012, 141:e576S–e600S. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 75.Yanagisawa R, Tanaka M, Yashima F, Arai T, Jinzaki M, Shimizu H, Fukuda K, Watanabe Y, Naganuma T, Higashimori A, et al. : Early and late leaflet thrombosis after transcatheter aortic valve replacement: A multicenter initiative from the OCEAN-TAVI registry. Circ Cardiovasc Interv 2019, 12. [DOI] [PubMed] [Google Scholar]
- 76.Jordan LC, Ichord RN, Reinhartz O, Humpl T, Pruthi S, Tjossem C, Rosenthal DN: Neurological complications and outcomes in the berlin heart EXCOR® pediatric investigational device exemption trial. J Am Heart Assoc 2015, 4:e001429–e001429. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 77.Ho A, Schultz J, John R, Goodwin K, Kluis A, Pritzker M, Missov E, Liao K, Cogswell R: Incidence of Early and Late Pump Thrombosis on HeartMate II LVAD Support. J Hear Lung Transplant 2018, 37:S369. [Google Scholar]
- 78.Upshaw JN, Kiernan MS, Morine KJ, Kapur NK, De Nofrio D: Incidence, management, and outcome of suspected continuous-flow left ventricular assist device thrombosis. ASAIO J 2016, 62:33–39. [DOI] [PubMed] [Google Scholar]
- 79.Zeymer U, Bauer T, Hamm C, Zahn R, Weidinger F, Seabra-Gomes R, Hochadel M, Marco J, Gitt A, Eberli F, et al. : Use and impact of intra-aortic balloon pump on mortality in patients with acute myocardial infarction complicated by cardiogenic shock: Results of the Euro Heart Survey on PCI. EuroIntervention 2011, 7:437–441. [DOI] [PubMed] [Google Scholar]
- 80.Ball L, Costantino F, Pelosi P: Postoperative complications of patients undergoing cardiac surgery. Curr Opin Crit Care 2016, 22:386–392. [DOI] [PubMed] [Google Scholar]
- 81.Copeland JG, Smith RG, Arabia FA, Nolan PE, McClellan D, Tsau PH, Sethi GK, Bose RK, Banchy ME, Covington DL, et al. : Total artificial heart bridge to transplantation: A 9-year experience with 62 patients. J Hear Lung Transplant 2004, 23:823–831. [DOI] [PubMed] [Google Scholar]
- 82.Copeland J, Copeland H, Nolan P, Gustafson M, Slepian M, Smith R: Results with an anticoagulation protocol in 99 syncardia total artificial heart recipients. ASAIO J 2013, 59:216–220. [DOI] [PubMed] [Google Scholar]
- 83.Hastings SM, Deshpande SR, Wagoner S, Maher K, Ku DN: Thrombosis in centrifugal pumps: Location and composition in clinical and in vitro circuits. Int J Artif Organs 2016, 39:200–204. [DOI] [PubMed] [Google Scholar]
- 84.Grewal S, Chamarthy MR, Kalva SP: Complications of inferior vena cava filters. Cardiovasc Diagn Ther 2016, 6:632–641. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 85.Imamura H, Sakai N, Satow T, lihara K: Factors related to adverse events during endovascular coil embolization for ruptured cerebral aneurysms. J Neurointerv Surg 2020, 12:605–609. [DOI] [PubMed] [Google Scholar]
- 86.Barnard EBG, Morrison JJ, Madureira RM, Lendrum R, Fragoso-Iñiguez M, Edwards A, Lecky F, Bouamra O, Lawrence T, Jansen JO: Resuscitative endovascular balloon occlusion of the aorta (REBOA): a population based gap analysis of trauma patients in England and Wales. Emerg Med J 2015, 32:926–932. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 87.Tennankore KK, D’Gama C, Faratro R, Fung S, Wong E, Chan CT: Adverse technical events in home hemodialysis. Am J Kidney Dis 2015, 65:116–121. [DOI] [PubMed] [Google Scholar]
- 88.Cicha I: Thrombosis: Novel nanomedical concepts of diagnosis and treatment. World J Cardiol 2015, 7:434. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 89.Griffin MT, Zhu Y, Liu Z, Aidun CK, Ku DN: Inhibition of high shear arterial thrombosis by charged nanoparticles. Biomicrofluidics 2018, 12. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 90.Tosenberger A, Ataullakhanov F, Bessonov N, Panteleev M, Tokarev A, Volpert V: Modelling of platelet–fibrin clot formation in flow with a DPD–PDE method. J Math Biol 2016, 72:649–681. [DOI] [PubMed] [Google Scholar]
- 91.Bouchnita A, Volpert V: A multiscale model of platelet-fibrin thrombus growth in the flow. Comput Fluids 2019, 184:10–20. [Google Scholar]

