In vivo dosimetry for testicular and scalp shielding in total skin electron therapy using a radiophotoluminescence glass dosimeter

Hiroyuki Okamoto, Kae Okuma , Hiroki Nakayama, Satoshi Nakamura, Kotaro Iijima, Takahito Chiba, Mihiro Takemori, Kyohei Fujii, Shohei Mikasa, Tetsu Nakaichi, Ako Aikawa, Shouichi Katsuta, Hiroshi Igaki
PMCID: PMC8776692  PMID: 34718685

Abstract

Mycosis fungoides (MF) is a common, low-grade non-Hodgkin’s lymphoma of skin-homing T lymphocytes that can be treated via skin-directed radiotherapy. Our institution has implemented total skin electron therapy (TSET) with a 4.3 m source-to-surface distance (SSD) and 6 MeV electron beams with a beam spoiler. A 35-year-old male undergoing TSET desired to avoid radiotherapy-induced hair loss and temporary infertility; therefore, leakage dose to scalp and testicles was reduced with a special radiation shield composed of stacked lead sheets. The shields for the scalp and scrotal were of 3 mm and 6 mm, respectively. To assess leakage doses, a radiophotoluminescence glass dosimeter (RPLD) was placed at every fraction. The difference dose between the measured and prescribed dose at the calibration point was 2%. The top of the head and scrotal surface exhibited 18 cGy and 10 cGy, respectively. Thus, the dose to the scrotal surface was not beyond the testicular tolerance dose of 20 cGy. Results of semen analysis two months postradiotherapy were normal. There was no hair loss during or after radiation therapy. Therefore, the RPLD is a useful in vivo dosimeter that provides technical information on radiation shielding to allow for completion of TSET without hair loss or temporary infertility.

Keywords: in vivo dosimetry, radiophotoluminescence glass dosimeter (RPLD), total skin electron therapy (TSET), mycosis fungoides (MF)

INTRODUCTION

Mycosis fungoides (MF) is a common, low-grade non-Hodgkin’s lymphoma of skin-homing T lymphocytes. Radiotherapy can be directed to a specific skin area or the entire skin (total skin electron therapy [TSET]), which, when performed with the Stanford technique, provides standard irradiation to the entire skin [1, 2]. The standard dose of TSET is 30–36 Gy over 9–10 weeks. Common acute toxicities in TSET involve erythema/desquamation, blisters, hyperpigmentation, skin pain, and skin infection, which should be carefully managed. Low doses have been proposed with the aim to reduce acute toxicity [3].

The in vivo dosimetry in TSET should be performed to determine the distribution of dose to the patient’s skin and to verify the delivered prescribed dose to the patient’s skin is correct [2]. TSET with in vivo dosimetry [4] has demonstrated that flat regions such as the chest agreed with the prescribed dose, whereas special sites such as the ankle, medial thigh and axilla exhibited an underdose. Patient position and characteristics (height, obesity index) influence variations in the measured dose [5, 6].

The typical in vivo dosimetry consists of a diode detector, a thermoluminescent dosimeter (TLD), optically stimulated luminescent dosimeter, or radiophotoluminescence glass dosimeter (RPLD), and a radiochromic film. In vivo dosimetry with a diode detector has been used widely as it provides real-time measurement and is easy to handle (real-time measurements with TLDs and film are not available). However, TLDs and film have been used in in vivo dosimetry for a long time, and stable measurements may be obtained without complications.

In this case report, a patient undergoing TSET wanted to prevent radiotherapy-associated hair loss and infertility. The threshold dose for temporary infertility loss is 20 cGy [7], and the threshold dose for temporary alopecia in a single fraction is 2 Gy [8]. Radiation shielding can limit leakage doses but must be monitored accurately. With the aim to complete TSET without hair loss or temporary infertility, we used an RPLD as an in vivo dosimeter because it provides accurate reading and linearity, with fewer variation responses, fading, or overall measurement uncertainty compared with other detectors [9, 10].

MATERIAL AND METHODS

A 35-year-old male was first diagnosed with atopic dermatitis in 2009 and then with Stage IIB MF (T3N0M0, ISCL/EORTC [11]) in 2017. He underwent narrowband ultraviolet light B phototherapy. Treatment with oral psoralen plus ultraviolet A (PUVA) and bexarotene was continued >2 years. In 2019, facial and body skin tumor lesions emerged, which were locally irradiated by an electron beam. His doctors considered an allogeneic hematopoietic stem cell transplant, but the patient requested sperm preservation. PUVA and bexarotene were temporarily stopped, and the patient was referred to our hospital for TSET to control the lesions. The patient wanted to prevent radiotherapy-associated alopecia and fertility loss for semen cryopreservation. The radiation oncologist blocked radiation to the scalp and testicles to prevent hair loss and infertility.

Our institution implemented the Stanford University technique with 4.3 m source-to-surface distance (SSD) and 6 MeV electron beams using Varian Clinac iX (Supplementary Fig. 1). A prescribed dose of 10 Gy was administered in cycles of 2 Gy over 2 days. Fig. 1a shows a craniocaudal dose profile for dual fields of ±11° gantry angle with an acrylic beam spoiler. The patient’s soles of the feet corresponds to −120 cm (patient’s height 174 cm). The dose can be uniformly delivered to the whole body except the patient’s foot. Fig. 1b compares the percentage depth dose (PDD) with 4.3 m SSD in a water-equivalent phantom (Tough Water, Kyoto Kagaku, Kyoto, Japan) for dual fields with electron energy values of 6 MeV with a beam spoiler. All PDDs were obtained via the measured percentage depth ionization curve with a plane-parallel chamber (Roos®, PTW-Freiburg GmbH, Freiburg, Germany) [12].

Fig. 1.

Fig. 1

(a) The cranio–caudal dose profile for the dual-fields. (b) PDD in water equivalent phantom.

Radiation shields were composed of lead with a physical density of 11.7 g/cm3 (Orion RadSafe Medical Co., Ltd., Tokyo, Japan). Optimum thickness was verified, and transmissions of the 3-mm- and 6-mm-thick lead sheets were 0.8% and 0.6%, respectively, as measured with an electron chamber (Roos®).

Fig. 2 shows the scalp helmet and bowl-shaped testicular shield were 3 mm thick. Three lead sheets were wrapped around the patient’s waist, and the lead shield for the testicles was 6 mm thick (Supplementary Fig. 2). We ensured patient comfort and ensured no gap between the radiation shield and patient’s skin at the groin for every fraction.

Fig. 2.

Fig. 2

The lead shield for the head (A) and the testicles (B and C).

The RPLD dosimeter (GD-302 M, AGC Technology Solutions Co., Ltd., Kanagawa, Japan) had a 12 mm longitudinal length with 1.5 mm diameter. RPLDs were placed at a depth of 10 mm in the water-equivalent phantom and were calibrated to obtain the absorbed dose for each dosimeter. No other correction factors [13] were performed. Encapsulated RPLDs were attached by tape to the patient’s skin (Fig. 3). RPLD measurements were performed for every fraction at eight locations (Table 1 and Supplementary Fig. 2). This report was approved by the institutional ethics review board in National Cancer Center Hospital (approval number: 2017–091).

Fig. 3.

Fig. 3

The attachment of the capsuled RPLDs to the phantom.

Table 1.

The in vivo dosimetry using RPLDs

Cycle #1 #2 #3 #4 #5 Sum
Number of fraction 1 2 3 4 5 6 7 8 9 10
Top of the head (Shielding) 4 2 2 2 3 2 3 2 3 2 18
Chest (Calibration point) NA. 129 63 123 76 128 66 131 66 132 915
Chest per a cycle NA 186 204 197 198 785**
Left scrotal surface* 1 1 1 4 1 2 1 1 1 1 11
Right scrotal surface* 1 1 1 1 1 1 1 1 1 1 9
Left groin* 2 2 2 2 2 2 2 2 2 2 15
Right groin* 2 2 2 2 2 2 3 2 2 2 17
Left underside of the foot 13 11 8 10 12 9 11 12 12 9 82
Right underside of the foot 11 10 10 12 9 20 10 13 12 11 97

NA, Not Available; Unit of measurement is cGy;

*Shielding measurement points.

**Total dose was derived from sum of the measured dose from 2nd to last cycles, because the measurement at the 1st fraction could not be done by misplacement.

RESULTS AND DISCUSSION

Results of in vivo dosimetry with RPLDs are shown in Table 1. Dosimetry for the chest at the 1st fraction could not be performed due to misplacement. Therefore, the total measured dose at the chest location was 789 cGy from the 2nd cycle dose, corresponding to the delivered dose of 800 cGy. The difference between the two was within 2%. The head and scrotal surface for all fractions exhibited 18 and 10 cGy leakage doses, respectively, which were lower than the threshold [7, 8]. Semen test results according to the WHO criteria [14] two months postradiotherapy were normal (Supplementary Table 2). Hair loss was not observed.

TLDs are used most widely for in vivo dosimetry in TSET [5]. RPLD for in vivo dosimetry in TSET requires testing. To the best of our knowledge, this is the first report to examine the testicular dose through in vivo dosimetry using RPLD in TSET. RPLDs overcome the stem effect when measuring leakage doses [15]. In this study, the signal-to-noise ratio of the RPLDs was 500. RPLDs measurements for the chest were in good agreement with the prescribed dose and similar to in vivo dosimetry using TLD in TSET [5].

Supplementary Table 1 summarizes measurement uncertainty (5.2%) of RPLDs at the calibration point. For other uncertainties, angular dependency should be considered [10, 16]. In this work, placement of RPLDs at the chest was almost the same as the calibration points. However, it was impossible to place RPLDs in the same direction as the calibration points at the top of the head and scrotal surface. The effects of these remain unclear. Moreover, materials in RPLDs have a high atomic number, which might cause energy dependency and increased sensitivity of measurements as photon energy decreases [13, 17]. Energy spectra changes under radiation shielding remains unclear. However, primary radiation is likely shielded effectively under lead sheets, and bremsstrahlung radiation or scattered radiation mainly causes leakage doses. These effects lead to safe estimation of the exposure dose.

Leakage doses at the head and scrotal surface was 1.8% and 1.0%, respectively, and slightly greater than the phantom measurements (0.8% and 0.6%). That was because the shapes of the lead sheets were different to cover the scrotal and head skin surfaces and six different standing positions were required; radiation shielding might be inadequate especially for the back of the patient’s body. However, lead shielding successfully maintained leakage doses below the tolerance dose. Our work demonstrates that a RPLD is a reliable and useful in vivo dosimeter for TSET.

Supplementary Material

supplementary_material_rrab100

ACKNOWLEDGEMENTS

The authors thank all of the staff involved in radiation therapy in our institution.

CONFLICT OF INTEREST

The authors declare they have no conflicts of interest.

FUNDING

This study was partly supported by the National Cancer Center Research and Development Fund (30–A–14) and a Grant-in-Aid for Young Scientists (B) from the Ministry of Education, Culture, Sports, Science and Technology (Grant Number 18 K15613).

References

  • 1. Page  V, Gardner  A, Karzmark  CJ. Patient dosimetry in the electron treatment of large superficial lesions. Radiology  1970;94:635–41. [DOI] [PubMed] [Google Scholar]
  • 2. Karzmark  CJ  et al.  Total Skin Electron Therapy: Technique and Dosimetry. AAPM report no. 23. American Association of Physicists in Medicine by the American Institute of Physics  1988. [Google Scholar]
  • 3. Kroeger  K, Elsayad  K, Moustakis  C  et al.  Low-dose total skin electron beam therapy for cutaneous lymphoma. Strahlenther Onkol  2017;193:1024–30. [DOI] [PubMed] [Google Scholar]
  • 4. Antolak  JA and Cundiff  JH, and Ha  CS. Utilization of thermoluminescent dosimetry in total skin electron beam radiotherapy of mycosis fungoides, Int J Radiat Oncol Biol Phys.  1998;40:101–8. [DOI] [PubMed] [Google Scholar]
  • 5. Dhivya  S, Anuradha  C, Murali  V, Ramasubramanian  V.  In-vivo dosimetry in total skin electron therapy: literature review, Journal of Radiotherapy in Practice  2021;20:357–360. [Google Scholar]
  • 6. Weaver  RD, Gerbi  BJ, Dusenbery  KE. Evaluation of dose variation during total skin electron irradiation using thermoluminescent dosimeters. Int J Radiat Oncol Biol Phys.  1995;33:475–478. [DOI] [PubMed] [Google Scholar]
  • 7. Kinsella  TJ, Trivette  G, Rowland  J  et al.  Long-term follow-up of testicular function following radiation therapy for early-stage Hodgkin's disease. J Clin Oncol  1989;7:718–24. [DOI] [PubMed] [Google Scholar]
  • 8. Hamilton  CS, Potten  CS, Denham  JW  et al.  Response of human hair cortical cells to fractionated radiotherapy. Radiother Oncol  1997;43:289–92. [DOI] [PubMed] [Google Scholar]
  • 9. Wesolowska  PE, Cole  A, Santos  T  et al.  Characterization of three solid state dosimetry systems for use in high energy photon dosimetry audits in radiotherapy. Radiat Meas  2017;106:556–62. [Google Scholar]
  • 10. Rah  JE, Hwang  UJ, Jeong  H  et al.  Clinical application of glass dosimeter for in vivo dose measurements of total body irradiation treatment technique. Radiat Meas  2011;46:40–45. [Google Scholar]
  • 11. Kim  YH, Willemze  R, Pimpinelli  N  et al.  TNM classification system for primary cutaneous lymphomas other than mycosis fungoides and Sezary syndrome: a proposal of the International Society for Cutaneous Lymphomas (ISCL) and the Cutaneous Lymphoma Task Force of the European Organization of Research and Treatment of Cancer (EORTC). Blood  2007;110:479–84. [DOI] [PubMed] [Google Scholar]
  • 12. Standard Dosimetry of Absorbed Dose to Water in External Beam Radiotherapy (Standard Dosimetry 12). Tokyo: Tsusho-Sangyo-Kenkyusha, 2012. [Google Scholar]
  • 13. Mizuno  H, Kanai  T, Kusano  Y  et al.  Feasibility study of glass dosimeter postal dosimetry audit of high-energy radiotherapy photon beams. Radiother Oncol  2008;86:258–63. [DOI] [PubMed] [Google Scholar]
  • 14. World Health Organization . WHO Laboratory Manual for the Examination of Human Semen, 5th edn. Geneva: World Health Organization, 2010. [Google Scholar]
  • 15. Das  IJ, Copeland  JF, Bushe  HS. Spatial distribution of bremsstrahlung in a dual electron beam used in total skin electron treatments: errors due to ionization chamber cable irradiation. Med Phys  1994;21:1733–8. [DOI] [PubMed] [Google Scholar]
  • 16. Araki  F, Moribe  N, Shimonobou  T, Yamashita  Y. Dosimetric properties of radio-photoluminescent glass rod detector in high-energy photon beams from a linear accelerator and cyber-knife. Med Phys  2004;31:1980–6. [DOI] [PubMed] [Google Scholar]
  • 17. Kadoya  N, Shimomura  K, Kitou  S  et al.  Dosimetric properties of radio-photoluminescent glass detector in low-energy photon beams. Med Phys  2012;39:5910–6. [DOI] [PubMed] [Google Scholar]