Incidence and Predictors of Infections and All-Cause Death in Patients with Cardiac Implantable Electronic Devices: The Italian Nationwide RI-AIAC Registry

Giuseppe Boriani 1,*,, Marco Proietti 2,3,4,, Matteo Bertini 5, Igor Diemberger 6, Pietro Palmisano 7, Stefano Baccarini 8, Francesco Biscione 9, Nicola Bottoni 10, Antonio Ciccaglioni 11, Alessandro Dal Monte 12, Franco Alberto Ferrari 13, Saverio Iacopino 14, Marcello Piacenti 15, Daniele Porcelli 16, Stefano Sangiorgio 17, Luca Santini 18, Michele Malagù 5, Giuseppe Stabile 19, Jacopo Francesco Imberti 1, Davide Caruso 20, Massimo Zoni-Berisso 20, Roberto De Ponti 21, Renato Pietro Ricci 22; on behalf of RI-AIAC Registry Investigators
Editor: Laura Adelaide Dalla Vecchia
PMCID: PMC8780465  PMID: 35055406

Abstract

Background: The incidence of infections associated with cardiac implantable electronic devices (CIEDs) and patient outcomes are not fully known. Aim: To provide a contemporary assessment of the risk of CIEDs infection and associated clinical outcomes. Methods: In Italy, 18 centres enrolled all consecutive patients undergoing a CIED procedure and entered a 12-months follow-up. CIED infections, as well as a composite clinical event of infection or all-cause death were recorded. Results: A total of 2675 patients (64.3% male, age 78 (70–84)) were enrolled. During follow up 28 (1.1%) CIED infections and 132 (5%) deaths, with 152 (5.7%) composite clinical events were observed. At a multivariate analysis, the type of procedure (revision/upgrading/reimplantation) (OR: 4.08, 95% CI: 1.38–12.08) and diabetes (OR: 2.22, 95% CI: 1.02–4.84) were found as main clinical factors associated to CIED infection. Both the PADIT score and the RI-AIAC Infection score were significantly associated with CIED infections, with the RI-AIAC infection score showing the strongest association (OR: 2.38, 95% CI: 1.60–3.55 for each point), with a c-index = 0.64 (0.52–0.75), p = 0.015. Regarding the occurrence of composite clinical events, the Kolek score, the Shariff score and the RI-AIAC Event score all predicted the outcome, with an AUC for the RI-AIAC Event score equal to 0.67 (0.63−0.71) p < 0.001. Conclusions: In this Italian nationwide cohort of patients, while the incidence of CIED infections was substantially low, the rate of the composite clinical outcome of infection or all-cause death was quite high and associated with several clinical factors depicting a more impaired clinical status.

Keywords: cardiac implantable electronic device, implantable defibrillator, pacemaker, infection, outcome

1. Introduction

In recent years, there has been a progressively increasing use of cardiac implantable electronic devices (CIEDs), both due to the constant technological advance and to the evidence showing a significant effectiveness in improving patients’ quality of life and long-term survival [1,2]. Over a large spectrum of clinical indications, permanent pacemakers (PMs), implantable cardioverter-defibrillators (ICDs) and cardiac resynchronization therapy (CRT-D (with defibrillator) and CRT-P (without defibrillator)) use have all progressively increased in the last 15 years [3,4].

In patients receiving CIEDs implantation the clinical course can be complicated by the occurrence of device infection. Rate of infections is usually ≤2%, but such event has an important impact on the clinical management of patients, with high healthcare-associated costs, prolonged hospitalisation and a high rate of mortality events (~20%) [5,6,7,8,9]. It is already known that, in determining the pathophysiological mechanism of the onset of CIEDs infection, various risk factors are involved, both clinical and procedural [10,11,12,13,14,15,16,17,18].

Given the relevant implications of the CIEDs infection, it became pivotal to identify those patients with a higher risk of developing such infection, to implement adequate preventive strategies. To stratify the infective risk, several clinical scores have been proposed, such as those proposed by Shariff and Kolek [19,20,21] and more recently the ‘Prevention of Arrhythmia Device Infection Trial’ (PADIT) score [5].

To provide a contemporary epidemiological assessment of the risk of CIEDs infection and associated clinical outcomes in Italy, we planned and performed a nationwide study about practices, clinical management and outcomes of patients receiving a CIED, irrespective of clinical indication and type of device, the ‘Ricerca sulle Infezioni Associate a ImpiAnto o sostituzione di CIED’ (RI-AIAC) Registry. Additionally, we proposed two novel contemporary scores for the risk stratification of CIEDs infection and clinical outcomes, to be compared to currently used risk scores, as proposed by Shariff, Kolek and the PADIT score [5,19,20,21].

2. Methods

The RI-AIAC Registry is an Italian nationwide multicentre prospective observational registry held by the ‘Italian Association of Arrhythmology and Cardiac stimulation’ (Associazione Italiana di Aritmologia e Cardiostimolazione(AIAC)). The registry is based on the AIAC nationwide electrophysiology practices network, promoted by the Cardiology Division at the University of Modena and Reggio Emilia, Policlinico di Modena, which was the coordinating centre. In the first phase of the study, the entire network was claimed for participation to the registry. Forty-three centres answered with an expression of interest. The study was then approved by the Institutional Review Board at AVEN Ethic Committee (Approval N. 9/16, protocol N. 1188/CE) on 01 April 2016. Overall, 18 practices finally entered the study and started the enrolment. Enrolment was undertaken from 1 September 2016 to 31 August 2019, with a progressive and constant accrual rate (Figure 1).

Figure 1.

Figure 1

RI-AIAC registry enrolment rates.

All consecutive patients ≥18 years old undergoing a CIED implant procedure at all sites, irrespective of clinical indication, type of procedure and type of CIED, were approached to obtain informed consent and included in the registry after signing for at least 30 days of enrolment. Baseline data about demographic information, clinical risk factors for CIED infection and procedure main characteristics were collected into an electronic case report form.

All patients enrolled entered a 12-months follow-up observation, with two consecutive visits at 6 and 12 months after the CIED implant procedure. Follow-up visits were performed at each electrophysiology practice. In case a CIED infection was detected, all relevant data were entered into a specific electronic case report form. In the case a death event was detected, date and data regarding the supposed cause of death were recorded separately.

As revision, we defined a procedure where no new prosthetic material was inserted, as upgrade, a procedure where the system characteristics were modified by adding new leads (e.g., from a single chamber ICD to a CRTD system), and as reimplantation the positioning of a new CIED system in a patient who had a previous CIED implant object of extraction.

2.1. Outcomes

The outcomes of interest were the occurrence of a CIED infection, as well as the occurrence of any death, irrespective of the cause. In this analysis, we considered the occurrence of CIED infection alone and a composite clinical event of infection or all-cause death.

2.2. Statistical Analysis

Continuous variables were reported as mean and standard deviation (SD) or median and interquartile range (IQR) and compared between groups with Independent Samples Student’s T test or Mann–Whitney U test, accordingly. Categorical variables were reported as counts and percentages and compared between groups with Chi-square test.

Incidence of outcomes was calculated by censoring the follow-up time at CIED infection and all-cause death occurrence and considering a total follow-up time of 365 days for each patient which completed the follow-up observation without reporting any event. All patients lost throughout follow-up were not considered for incidence calculation and follow-up analyses.

To establish clinical factors associated with outcomes occurrence, a logistic regression analysis was performed. Clinical factors significantly different at baseline between patients that reported and not reported the outcome were considered for univariate analysis. All factors with a p ≤ 0.10 at univariate analysis were included into the multivariate analysis. Results of the logistic regression analysis were reported as odds ratio (OR) and 95% confidence interval (CI).

Based on the coefficients reported at multivariate logistic analysis, we have obtained a clinical score by rounding and re-scaling beta-coefficients, corresponding to the logarithm of the OR estimated during regression procedure. We have elaborated and evaluated two differential scores, one about CIED infection occurrence and one about the composite clinical event occurrence.

The association between the clinical scores considered and the occurrence of the outcomes was evaluated using an adjusted regression analysis. The predictive ability of each score was examined and compared according to De Long and De Long and reported by c-index and 95% CI. A cut-off value for the new scores was established according to the Youden Index.

A two-sided p < 0.05 was considered statistically significant. All analyses were performed using SPSS statistical software version 27.0.1.0 (IBM, Armonk, NY, USA) and STATA/MP 16.1 (StataCorp, College Station, TX, USA) for MacOS.

2.3. External Validation

We performed an external validation of the two newly proposed scores in an independent cohort of 1017 consecutive patients receiving a CIED implant procedure in a third-level referral electrophysiology centre, collected consecutively from 1 January 2017 to 31 December 2019. We evaluated association between the clinical scores and the outcomes, as well as the predictive performance of each score using the same logistic model and the same methods used in the main analysis. All the patients were followed by the centre’s physicians, using similar procedures to those implemented in our study. Moreover, for this cohort 1 year follow-up data were assessed.

3. Results

A total of 2675 patients were enrolled throughout the study period in 18 active sites. Patients were enrolled prevalently in Northern Italy, with 1658 (62%) among 11 sites, while 571 (21.3%) were enrolled in Central Italy (5 centres) and 446 (16.7%) in Southern Italy (2 centres). Baseline characteristics are reported in Table 1. Overall, the study cohort was significantly old, with 1623 (60.7%) patients being ≥75 years old, and prevalently male (1720 patients (64.3%)). Most of the procedures were performed in a ward setting, with a majority of first implantation (70.1%). Regarding the type of device, 1785 (66.7%) patients received a PM, 450 (16.8%) an ICD, 106 (4.0%) a CRT-P and 329 (12.3) a CRT-D. Among the known risk factors for CIEDs infection, heart failure and the use of oral anticoagulants were the most common, being present in 27.7% and 30.7%, respectively.

Table 1.

Baseline characteristics of enrolled patients.

N = 2675
Age, years median (IQR) 78 (70–84)
Age Classes, n (%)
<65 years
65–74 years
≥75 years

422 (15.8)
630 (23.6)
1623 (60.7)
Male Sex, n (%) 1720 (64.3)
Admission, n (%)
Ward
Daily Service

2263 (84.6)
412 (15.4)
Procedure, n (%)
First Implantation
Replacement
Further Replacement
Contralateral Implantation
Upgrading
Revision
Other

1874 (70.1)
448 (16.7)
205 (7.7)
17 (0.6)
72 (2.7)
51 (1.9)
8 (0.3)
CIED Type, n (%)
Pacemaker
ICD
CRT-P
CRT-D
Other

1785 (66.7)
450 (16.8)
106 (4.0)
329 (12.3)
5 (0.2)
>2 Leads, n (%) 194 (7.3)
Early Revision, n (%) 38 (1.4)
Prolonged Temporary Pacing, n (%) 73 (2.7)
eGFR <60 mL/min, n (%) 622 (23.3)
Pre-Dialysis/Dialysis, n (%) 41 (1.5)
Diabetes Mellitus, n (%) 550 (20.6)
Heart Failure, n (%) 740 (27.7)
Use of Oral Corticosteroids, n (%) 72 (2.7)
Use of Oral Anticoagulants, n (%) 821 (30.7)
Use of Immunosuppressive Therapy, n (%) 18 (0.7)
24 h Pre-Implantation Fever, n (%) 12 (0.4)
Hospital-Acquired Infection, n (%) 44 (1.6)
Antibiotic Prophylaxis, n (%) 2636 (98.5)
Antibiotic Type, n (%) 2636
Cephalosporins
Clindamycin
Penicillin
Other

2186 (82.9)
70 (2.6)
250 (9.3)
130 (4.9)
Overall Infection Risk Factors, n median [IQR] 2 (1,2)
PADIT Score, median [IQR] 1 (0–4)
Kolek Score, median [IQR] 1 (0–2)
Shariff Score, median [IQR] 1 (0–2)

CIED = Cardiac Implantable Electronic Device; CRT-D = Cardiac Resynchronization Therapy Defibrillator; CRT-P = Cardiac Resynchronization Therapy Pacing; eGFR = Estimated Glomerular Filtration Rate; ICD = Implantable Cardioverter Defibrillator; IQR = Interquartile Range; PADIT = Prevention of Arrhythmia Device Infection Trial; SD = Standard Deviation.

Overall, our cohort had a median (IQR) of 2 (1,2) infection risk factors, with an overall low risk according to all the scores examined (Table 1).

Antibiotic prophylaxis was performed in almost all the patients enrolled, with cephalosporins being the agent most used. Among the overall cohort, only 323 (12.1%) patients were treated with intra-operatory pocket irrigation with antibiotics. Only 53 patients (2%) received at implant an absorbable antibacterial envelope.

3.1. Follow-Up and Incidence of Adverse Outcomes

Throughout the follow-up observation, a total of 2647 patients (99%) had available data, with 2523 (94.3%) with available data regarding the occurrence of infection, due to censoring of observation. Overall, there were 28 (1.1%) CIED infections and 132 (5%) deaths, with 152 (5.7%) composite clinical events. Cumulative risk of clinical outcomes is reported in Figure 2. Mean (SD) follow-up for CIED infection was 276.25 (143.61) days, while mean (SD) follow-up for composite clinical event was 273.57 (141.65) days.

Figure 2.

Figure 2

Kaplan–Meier curves for clinical outcomes occurrence. (Upper Figure) and (Cumulative risk for CIED infection; Lower Figure) Cumulative risk for composite clinical event; Legend: CIED = Cardiac Implantable Electronic Device.

Overall, the final incidence of CIED infection was equal to 1.1 per 100 patient-years, while the final incidence of the composite clinical event was 5.82 per 100 patient-years.

3.2. Clinical Factors Associated to Outcomes Occurrence

According to baseline characteristics, patients developing a CIED infection were younger than those not developing this complication (p = 0.041), even though there was no difference according to age classes (p = 0.538) (Table 2). Patients presenting a CIED infection had been more likely enrolled for a revision/upgrading/reimplantation with an early revision, receiving less likely a PM (Table 2). Furthermore, patients developing a CIED infection were more likely diabetic, with a severe kidney disease and an hospital-acquired infection (HAI) (Table 2).

Table 2.

Baseline characteristics according to Occurrence of Infection during Follow-Up.

No Infection
N= 2495
Infection
N= 28
p
Age, years median (IQR) 77 (69–83) 74 (66–77) 0.041
Age Classes, n (%)
<65 years
65–74 years
≥75 years

411 (16.5)
597 (23.9)
1487 (59.6)

5 (17.9)
9 (32.1)
14 (50)
0.538
Male Sex, n (%) 1601 (64.2) 19 (67.9) 0.686
Admission, n (%)
Ward
Daily Service

2112 (84.6)
383 (15.4)

23 (82.1)
5 (17.9)
0.715
Procedure, n (%)
First Implantation
Replacement
Further Replacement
Contralateral Implantation
Upgrading
Revision
Other

1745 (69.9)
428 (17.2)
186 (7.5)
14 (0.6)
67 (2.7)
48 (1.9)
7 (0.3)

15 (53.6)
2 (7.1)
4 (14.3)
3 (10.7)
2 (7.1)
1 (3.6)
1 (3.6)
<0.001
CIED Type, n (%)
Pacemaker
ICD
CRT-P
CRT-D
Other

1654 (66.3)
426 (17.1)
98 (3.9)
312 (12.5)
5 (0.2)

13 (46.5)
8 (28.6)
3 (10.7)
4 (14.3)
0
0.131
CIED Type Recode, n (%)
Pacemaker
Any Other CIED

1654 (66.3)
841 (33.7)

13 (46.4)
15 (53.6)
0.027
>2 Leads, n (%) 184 (7.4) 3 (10.7) 0.502
Early Revision, n (%) 34 (1.4) 2 (7.1) 0.010
Prolonged Temporary Pacing, n (%) 62 (2.5) 1 (3.6) 0.714
eGFR <60 mL/min, n (%) 553 (22.2) 6 (21.4) 0.926
Pre-Dialysis/Dialysis, n (%) 32 (1.3) 2 (7.1) 0.007
Diabetes Mellitus, n (%) 499 (20.0) 11 (39.3) 0.012
Heart Failure, n (%) 689 (27.6) 11 (39.3) 0.170
Use of Oral Corticosteroids, n (%) 62 (2.5) 1 (3.6) 0.714
Use of Oral Anticoagulants, n (%) 759 (30.4) 9 (32.1) 0.844
Use of Immunosuppressive Therapy, n (%) 17 (0.7) 0 1.000
24 h Pre-Implantation Fever, n (%) 10 (0.4) 0 1.000
Hospital-Acquired Infection, n (%) 37 (1.5) 2 (7.1) 0.016
Antibiotic Prophylaxis, n (%) 2455 (98.4) 28 (100) 0.499
Antibiotic Type, n (%) 2636
Cephalosporins
Clindamycin
Penicillin
Other

2055 (83.7)
61 (2.5)
224 (9.1)
116 (4.7)

19 (67.9)
2 (7.1)
5 (17.9)
2 (7.1)
0.120
Overall Infection Risk Factors, n median [IQR] 2 (1–2) 2 (1–3) 0.086
PADIT Score, median [IQR] 1 (0–4) 4 (1–5) 0.008
Kolek Score, median [IQR] 1 (0–2) 1 (0–2) 0.136
Shariff Score, median [IQR] 1 (0–2) 2 (1–3) 0.080

CIED = Cardiac Implantable Electronic Device; CRT-D = Cardiac Resynchronization Therapy Defibrillator; CRT-P = Cardiac Resynchronization Therapy Pacing; eGFR = Estimated Glomerular Filtration Rate; ICD = Implantable Cardioverter Defibrillator; IQR = Interquartile Range; PADIT = Prevention of Arrhythmia Device Infection Trial; SD = Standard Deviation.

According to logistic univariate analysis (Table 3), a multivariate analysis was performed, which identified the type of procedure (revision/upgrading/reimplantation vs. first implantation OR: 4.08, 95% CI: 1.38–12.08) and history of diabetes mellitus (OR: 2.22, 95% CI: 1.02–4.84) as main clinical factors associated to CIED infection occurrence, with the presence of an HAI showing a trend in association (OR: 3.96, 95% CI: 0.85–18.57, p = 0.080) (Table 3).

Table 3.

Logistic regression analysis for CIED infection occurrence.

Univariate Analysis Multivariate Analysis Score Point
OR (95% CI) p OR (95% CI) p
Age Classes
<65 years (ref.)
65–74 years
≥75 years

-
1.24 (0.41–3.72)
0.77 (0.28–2.16)

-
0.702
0.625

-
-
-

-
-
-

-
-
-
Procedure
First Implantation
Any Replacement
Revision/Upgrading/Reimplantation

-
1.14 (0.44–2.94)
5.99 (2.4–14.94)

-
0.792
<0.001

-
1.21 (0.46–3.18)
4.08 (1.38–12.08)

-
0.698
0.011

0
1
2
CIED
Pacemaker (ref.)
Any Other CIED

-
2.27 (1.08–4.79)

-
0.032

-
1.66 (0.75–3.68)

-
0.209

-
-
Early Revision 5.57 (1.27–24.40) 0.023 1.68 (0.30–8.87) 0.579 -
Pre-Dialysis/Dialysis 5.92 (1.35–26) 0.019 3.44 (0.74–15.94) 0.116 -
Diabetes Mellitus 2.59 (1.21–5.56) 0.015 2.22 (1.02–4.84) 0.045 1
Hospital-Acquired Infection 5.11 (1.17–22.32) 0.030 3.96 (0.85–18.57) 0.080 1

CI = Confidence Interval; CIED = Cardiac Implantable Electronic Device; OR = Odds Ratio.

Regarding the occurrence of the composite clinical event, at baseline those patients who developed an event were significantly older, more likely diabetic and with chronic kidney disease even at a severe stage and also were more likely to have experienced a prolonged period of temporary pacing and had presented more often an HAI. After univariate regression analysis, multivariate analysis showed that a progressively increasing age, a prolonged period of temporary pacing, use of oral corticosteroids and the presence of an HAI were significantly associated with the occurrence of a composite clinical event, with diabetes mellitus showing a strong trend in association (Table 4).

Table 4.

Logistic regression analysis for clinical events occurrence.

Univariate Analysis Multivariate Analysis Score Points
OR (95% CI) p OR (95% CI) p
Age Classes
<65 years (ref.)
65-74 years
≥75 years

-
2.32 (1.05–5.17)
4.04 (1.96–8.34)

-
0.039
<0.001

-
2.07 (0.92–4.65)
3.10 (1.45–6.64)

-
0.079
0.006

0
1
2
CIED
Pacemaker (ref)
Any Other CIED

-
0.7 (0.49–1.02)

-
0.062

-
0.95 (0.64–1.41)

-
0.798
Prolonged Temporary Pacing 3.06 (1.58–5.94) 0.001 2.9 (1.45–5.77) 0.002 1
eGFR <60 mL/min 2.55 (1.83–3.57) <0.001 2.03 (1.43–2.88) <0.001 1
Pre-Dialysis/Dialysis 1.62 (1.25–2.11) <0.001
Diabetes Mellitus 1.58 (1.09–2.28) 0.015 1.43 (0.98–2.08) 0.063 1
Use of Oral Corticosteroids 2.47 (1.2–5.08) 0.014 2.60 (1.25–5.40) 0.010 1
Hospital-Acquired Infection 3.21 (1.41–7.32) 0.006 3.29 (1.29–8.35) 0.012 1

CI = Confidence Interval; CIED = Cardiac Implantable Electronic Device; eGFR = Estimated Glomerular Filtration Rate; OR = Odds Ratio.

3.3. Association and Performance of Clinical Risk Scores

Based on the multivariate logistic regression analysis, we firstly developed a score regarding the occurrence of CIED infection, the RI-AIAC Infection score, by scoring 1 point for any CIED replacement, 2 points for revision / upgrading / reimplantation, 1 point for presence of diabetes mellitus and 1 point for reporting a HAI (Table 3).

Table 5 itemises simple clinical scores including age classes, presence of delayed temporary pacing, chronic kidney disease, diabetes mellitus and use of oral corticosteroids and HAI (Table 4). Scoring for each item was assigned according to the multivariate logistic regression analysis (Table 4). Distribution of patients according to the two scores is shown in Figure 3.

Table 5.

Predictive ability of clinical scores.

Infection Clinical Events
C-Index (95% CI) p C-Index (95% CI) p
PADIT Score 0.64 (0.53–0.76) 0.010 0.51 (0.47–0.56) 0.612
Kolek Score 0.56 (0.46–0.66) 0.261 0.63 (0.59–0.68) <0.001
Shariff Score 0.58 (0.47–0.68) 0.159 0.62 (0.58–0.67) <0.001
RI-AIAC Infection 0.64 (0.52–0.75) 0.015 - -
RI-AIAC Event * - - 0.67 (0.63–0.71) <0.001

* Adjusted for sex; † adjusted for age and sex; CI= Confidence Interval; OR= Odds Ratio; PADIT = Prevention of Arrhythmia Device Infection Trial; RI-AIAC = Ricerca sulle Infezioni Associate a ImpiAnto o sostituzione di CIED.

Figure 3.

Figure 3

RI-AIAC scores distribution; (Upper Figure RI-AIAC Infection Score; Lower Figure) RI-AIAC Event Score; RI-AIAC = Ricerca sulle Infezioni Associate a ImpiAnto o sostituzione di CIED.

We tested the association between the pre-existing scores, together with the new proposed ones (Table 6). After adjustments, while Kolek and Shariff scores were not associated with CIED infection occurrence, both the PADIT score and the RI-AIAC Infection score were significantly associated with CIED infections (Table 6), with the RI-AIAC infection score showing the strongest association (OR: 2.38, 95% CI: 1.60–3.55 for each point). Regarding the occurrence of the composite clinical event, while the PADIT score was not associated with the event, the Kolek and Shariff scores were both associated with the event, as well as the RI-AIAC Event score, which showed the strongest association (OR: 1.56, 95% CI: 1.38–1.75) (Table 6).

Table 6.

Logistic regression of clinical scores.

Infection Clinical Events
OR (95% CI) p OR (95% CI) p
PADIT Score * 1.28 (1.1–1.50) 0.002 1.01 (0.94–1.09) 0.748
Kolek Score † 1.55 (0.73–3.27) 0.251 1.36 (1.2–1.55) <0.001
Shariff Score † 1.69 (0.89–3.21) 0.112 1.32 (1.17–1.47) <0.001
RI-AIAC Infection † 2.38 (1.6–3.55) <0.001 - -
RI-AIAC Event * - - 1.56 (1.38-1.75) <0.001

* Adjusted for sex; † adjusted for age and sex; CI = Confidence Interval; OR = Odds Ratio; PADIT = Prevention of Arrhythmia Device Infection Trial; RI-AIAC = Ricerca sulle Infezioni Associate a ImpiAnto o sostituzione di CIED.

3.4. Predictive Ability of Clinical Scores

The four scores were tested for predictive capacity regarding the two clinical outcomes. For the occurrence of CIED infection, both the PADIT score and the RI-AIAC Infection score showed an overall modest prediction performance, while both Kolek and Shariff scores were not able to predict the outcome (Table 6), with no significant difference between the two scores (p = 0.87). Regarding the occurrence of the composite clinical event, Kolek score, Shariff score and RI-AIAC Event score all predicted the outcome occurrence (Table 6)., The RI-AIAC Event score showed a modest to good predictive performance (c-index: 0.67, 95% CI: 0.63–0.71), being superior both to Kolek score (p = 0.008) and to Shariff score (p = 0.006).

According to the Youden Index, a RI-AIAC Infection score ≥1 identified a patient at higher risk of CIED infection (Youden Index: 0.254, Sensitivity 36%, Specificity 90%). While a RI-AIAC Event score ≥2 (Youden Index: 0.272, Sensitivity 59%, Specificity 69%) identified a patient at higher risk of the composite clinical event. A RI-AIAC Infection score ≥1 was significantly associated with CIED infection occurrence (OR: 2.23, 95% CI: 1.02–4.85), after adjustment for age and sex. RI-AIAC Event score ≥2 was significantly associated with the composite clinical event occurrence (OR: 3.87, 95% CI: 2.29–6.56), after adjustment for sex.

3.5. External Validation Analysis

The external validation cohort consisted of 1017 patients, with a median (IQR) age of 83 (74–88) years and 630 (61.9%) male patients. Overall, 750 (73.7%) patients were ≥75 years and 622 (61.2%) patients were ≥80 years. Among the overall cohort, 819 (80.5%) patients received a PM implant, while the remaining 198 (19.5%) received an ICD implant. Median (IQR) PADIT score was 1 (0–3), with a maximum score of 12, median (IQR) Kolek score was 2 (1,2), median (IQR) Shariff score was 2 (1–3), median (IQR) RI-AIAC Infection score was 1 (0,1) and median (IQR) RI-AIAC Event score was 3 (2,3). Accordingly, a RI-AIAC Infection score ≥1 was found in 544 (53.5%) patients and a RI-AIAC Event score ≥2 was found in 866 (85.2%) patients. Throughout the 1-year observation, there were 14 (1.4%) CIED infections, with an overall incidence of 1.48 per 100 patient-years. Moreover, there were 121 (11.9%) composite clinical events, with an overall incidence of 12.72 per 100 patient-years.

Logistic regression analysis, adjusted for age and sex, showed that RI-AIAC Infection score was associated with occurrence of CIED infection (OR: 2.12, 95% CI: 1.09–4.12). Furthermore, the RI-AIAC Event score was associated with occurrence of the composite clinical event (OR: 1.84, 95% CI: 1.51–2.23), after adjustment for sex.

Results of the predictive ability analysis in this external validation cohort were reported in Table 6. Regarding the prediction of CIED infection occurrence, none of the scores resulted able to predict the outcome in this cohort (Table 6). Conversely, regarding the occurrence of the composite clinical event, the Shariff score showed a poor predictivity, while both the Kolek and the RI-AIAC Event scores showed modest to good predictivity (Table 7). Comparing the three scores, while the RI-AIAC Event score showed consistent better predictive ability than the Shariff score (p < 0.001), no difference was found between the RI-AIAC Event and the Kolek score (p = 0.200). Performing a complete evaluation of the two scores, while the Kolek score showed moderate sensitivity and specificity, the RI-AIAC Event score showed very good sensitivity (95.9%) but very low specificity (16.3%).

Table 7.

Predictive ability of clinical scores in external validation cohort.

Infection Clinical Events SE SP
C-Index (95% CI) p C-Index (95% CI) p
PADIT Score 0.53 (0.38–0.67) 0.746 0.49 (0.44–0.53) 0.600 - -
Kolek Score 0.64 (0.5–0.79) 0.065 0.65 (0.6–0.7) * <0.001 70.3% 51.2%
Shariff Score 0.62 (0.46–0.77) 0.131 0.56 (0.51–0.62) 0.025 - -
RI-AIAC Infection 0.58 (0.42–0.74) 0.292 - - - -
RI-AIAC Event - - 0.68 (0.63–0.72) <0.001 95.9% 16.3%

* No significant difference compared to RI-AIAC Event (De Long–De Long p = 0.2; significantly less predictive than RI-AIAC Event (De Long–De Long p < 0.001); CI = Confidence Interval; PADIT = Prevention of Arrhythmia Device Infection Trial; RI-AIAC = Ricerca sulle Infezioni Associate a ImpiAnto o sostituzione di CIED; SE = Sensitivity; SP = Specificity.

4. Discussion

In the Italian nationwide RI-AIAC registry, we showed that in a real-life contemporary cohort of patients receiving a CIED implantation the majority of procedures performed refers to PM implantation in an hospital setting. Patients included in the study were generally old, with several prevalent comorbidities and an overall low risk of infection. Throughout follow-up observation, we recorded a 1-year incidence of CIED infections of 1.1%, comparable with some recent prospectively collected data, such as the WRAP-IT trial [22] but lower than the rates of 2.2–2.4% reported in other studies [23,24] or the rate up to 5% reported by some centres in the EHRA survey [25]. Notably, in our study, the risk of CIED infection was higher at longer term from the device implant, with a different occurrence of early (occurring 1–30 days after CIED implant), mid-onset (occurring in the range 31–182 days after implant) and late-onset infections [26,27] with the latter being more common. In our cohort, the antibiotic envelope TYRX, already available at the time of our study initiation, was applied only to few selected cases, and we can hypothesize it was associated with a reduced occurrence of CIED infections in the first months after CIED implant, as shown in the WRAP-IT trial [22].

In this real-world cohort study, we found a high incidence of a composite clinical event of CIED infection or all-cause death. Previous studies that reported a high mortality, up to 20–25% at 1 year, were related to patients associated with CIED infection [28] or to patients with infections undergoing lead extraction in referral centres [29]. This appears to be in contrast with studies on long term outcomes after PM implant, showing a survival of around 55% at 10 years [30] or with the outcome of ICD or CRTD devices, which is characterized by a survival of around 60% at 5 years [31].

Our study derived a score for predicting both the risk of CIED infection and a composite clinical event of infection or all-cause death. We decided to keep separate the two assessments since only some of the independent predictors of the clinical composite end point were independent predictors of infections alone, and the overall incidence of the composite clinical score was more than 5-fold higher than the incidence of infection. The rationale for this choice may also be related to the consideration that age may act on different directions about some outcomes. Indeed, younger age was found to be a risk factor for CIED infections [11] and is actually included in the PADIT score [5]. Although we did not find an independent association between age classes and infections, we found conversely that a more advanced age was associated with the clinical composite outcome. Assessment of all-cause mortality appears necessary when evaluating a contemporary cohort of device patients in view of the important increase in ultra-octogenarian patients treated with PM implant in recent years [4]. Among the factors associated with the occurrence of the clinical outcomes, we found that most of the factors are associated with the presence of relevant comorbidities or with a more deprived clinical status. This evidence is reinforced by the analysis of the external validation cohort, which showed more than 60% of octogenarian patients, reporting a very high incidence of the composite clinical event, being more than 12% over 1 year of observation.

As reported above, previous literature reported a wide range for the incidence of CIED infections, with a trend towards a decline in most recent years [11,23,24,25]. Such trend is strengthened by our data, probably also underlining an improvement in the management of such patients resulting in a lower incidence of infections. Furthermore, even a recent epidemiological analysis underlined how several risk factors exist for CIED infection occurrence, whose risk can be strongly dependent on the cohort examined, being related to device, patient and procedural characteristics [11]. All these aspects likely influenced the predictive abilities of the scores tested to identify those patients at higher risk CIED infection occurrence. On one hand, the Kolek and the Shariff scores were derived from older cohorts with a different risk profile [19,20,21], on the other hand, the PADIT score derived from a clinical trial cohort stemmed from a highly selective cohort [5]. Moreover, the prevalence of PM implants was very low, differently from what can be found in real-life cohorts as those presented in our paper in which PM represents the vast majority. All these specific factors, together with a probably more complex pathophysiology such as that captured by a clinical risk score which is reductionist by definition, could have prevented a good discriminative ability for the previous scores and the proposed RI-AIAC Infection score. Such scenario is also reinforced by other data coming from different medical fields, which showed a large heterogeneity and a substantial failure of almost of all the proposed scores to identify patients at higher risk of developing an infective complication [32].

Contrarily, the occurrence of all-cause death seems to be increasing over time becoming even a more feared complication than the CIED infection itself [30,31]. In this regard, the predictive performance of the RI-AIAC Event score appears to be reassuring, reporting a moderate to good predictive ability both in the derivation and validation cohorts. Furthermore, the very high sensitivity showed underlines the ability to identify all those patients with the highest risk of reporting a significant clinical event, whether an infection or death, to adopt more intensive treatments and a stricter follow-up strategy to minimize the risk. At the same time, even though the Kolek score showed a similar predictive ability, the RI-AIAC Event has a lower number of variables included which are easy to collect bedside, and it is also specifically derived for the composite of major adverse outcomes from a contemporary cohort. Based on such similar predictive ability, the easiness of compilation for the RI-AIAC Event score can help physicians obtaining a simple and rapid risk stratification for patients undergoing a CIED implant in every clinical setting, to identify those patients for which a more intensive and specific management is deemed to reduce the risk of adverse outcomes. On these premises, we believe that the application of a clinical score to predict the occurrence of all the relevant clinical events, such as the RI-AIAC Event score which can provide a good balance between evidence, precision and practicality [33], would be beneficial in the management of these patients.

Strength and Limitations

In the field, many evaluations on the rate of infections were retrospective or derived from administrative datasets [9], but administrative datasets were found result in not really accurate assessments of infection rates with a high risk of CIED infections underreporting [34]. Our study is original in prospectively investigating a cohort of patients from daily clinical practice and in deriving a score both for infection at 1 year and, separately, for the clinical end point of infection or cardiac death. As recently stressed [11], the large majority of reports on CIED infections were retrospective, and also, the derived scores for predicting the risk of CIED infection were derived from retrospective analysis with the only exception of the PADIT score.

Among the limitations, we have to consider that the diagnosis of CIED infection was based on the diagnosis made by each investigator, as per standard care, without any centralized adjudication. Moreover, data regarding the cause of death were missing, and we could not deepen the analysis regarding the risk factors and the occurrence of death further. Furthermore, as this is an observational study, we cannot exclude the presence of other factors which could have influenced the occurrence of the clinical outcomes considered. Moreover, specific data regarding the occurrence of infective endocarditis were not collected in the study. Notwithstanding, data presented in this paper reflect the “real world” characterization of the present study, including a substantial amount of pacemaker implants and of CIED replacements [35]. Conversely, our analysis being limited to Italy, generalization of these results needs to be made with caution.

5. Conclusions

In this Italian nationwide cohort of patients receiving a CIED implant, we found that while incidence of CIED infections was substantially low, and the rate of the composite clinical outcome of infection or all-cause death was quite high and substantially associated with several clinical factors depicting a more impaired clinical status. A simple clinical score derived from those factors, the RI-AIAC Event score, has been found to be significantly associated with the occurrence of the composite clinical outcome, showing a modest to good predictive ability, also confirmed in the external validation cohort. Application of the RI-AIAC Event score would help to identify those patients at higher risk of reporting an adverse clinical event and adopt specific strategies to minimize this risk.

Acknowledgments

The list of RI-AIAC Investigators includes: University of Modena and Reggio Emilia, Policlinico di Modena, Modena, Italy: Giuseppe Boriani, Edoardo Casali, Vincenzo Turco, Erminio Mauro, Jacopo Francesco Imberti, Cristina Rolando, Riccardo Fontanesi, Davide Mei, Daria Sgreccia, Marco Vitolo; Ospedale C e G Mazzone, Ascoli Piceno, Italy: Procolo Marchese; Ospedale E. Muscatello, Augusta, Italy: Giovanni Licciardello; Ospedale Santa Maria Annunziata, Bagno a Ripoli, Italy: Antonio Fazi; Azienda Ospedaliero Universitaria Consorziale Policlinico, Bari, Italy: Stefano Favale; Casa di Cura Tricarico Rosano, Belvedere Marittimo, Italy: Giorgio Ventura; Ospedale di Bentivoglio, Bentivoglio, Italy: Giulio Boggian; Humanitas Gavazzeni, Bergamo, Italy: Giosuè Mascioli; Ospedale di Bergamo, ASST Papa Giovanni XXII, Bergamo, Italy: Marco Racheli; University of Bologna, Policlinico S. Orsola-Malpighi, Bologna: Igor Diemberger, Giulia Massaro, Jessica Frisoni, Matteo Ziacchi, Cristian Martignani, Mauro Biffi; Ospedale di Castrovillari, Castrovillari, Italy: Silvana De Bonis, Giovanni Bisignani; Ospedale di Chioggia, Chioggia, Italy: Michela Bevilacqua; Ospedale Città di Castello, Città di Castello, Italy: Federico Fedeli; Ospedale Civile di Civitanova Marche, Civitanova Marche, Italy: Diego Spagnolo; Maria Cecilia Hospital, Cotignola, Italy: Saverio Iacopino; Ospedale Maggiore di Crema, Crema, Italy: Erika Tavarelli; University of Ferrara, Ferrara, Italy: Matteo Bertini, Michele Malagù, Cristina Balla, Francesco Vitali; Ospedale di Fidenza, Fidenza, Italy: Stefano Baccarini, Paolo Pastori; Azienda Ospedaliero Universitaria Careggi, Firenze, Italy: Emilio Vincenzo Dovellini, Paolo Pieragnoli; Ospedale Padre Antero Miconi, Genova, Italy: Massimo Zoni-Berisso, Davide Caruso; Ospedale Villa Scassi, Genova, Italy; Massimo Zoni-Berisso, Matteo Laffi; Ospedale “Mater Salutis”, Legnago, Italy: Gabriele Zanotto; Ospedale Santa Croce Castelnuovo, Lucca, Italy: Pier Romano Mariani; Ospedale Fatebenefratelli e Oftalmico, Milano, Italy: Stefano Sangiorgio; Azienda Ospedaliera San Gerardo, Monza, Italy: Giovanni Rovaris; Clinica Mediterranea, Napoli, Italy: Antonio Stabile; Federico II University of Naples, Napoli, Italy; Antonio Rapacciuolo; Fondazione IRCCS Policlinico San Matteo, Pavia, Italy: Barbara Petracci; Fondazione Toscana Gabriele Monasterio, Pisa, Italy: Marcello Piacenti; Ospedale di Pistoia, Pistoia, Italy: Alberto Mandorli; Ospedale Santa Maria delle Croci, Ravenna, Italy: Alessandro Dal Monte; Arcispedale Santa Maria Nuova, Reggio Emilia, Italy: Nicola Bottoni; Ospedale di Rho, Rho, Italy: Franco Alberto Ferrari; Presidio Ospedaliero Santo Spirito in Sassia, Roma, Italy: Francesco Biscione; Policlinico Universitario Campus Bio-Medico, Roma, Italy: Vito Calabrese, Danilo Ricciardi; Sapienza University of Rome, Rome, Italy: Antonio Ciccaglioni, Agostino Piro; Ospedale San Pietro Fatebenefratelli, Roma, Italy: Daniele Porcelli; Ospedale G.B. Grassi, Roma, Italy: Luca Santini; Ospedale Nostra Signora di Bonaria, San Gavino Monreale, Italy: Gianfranco Delogu; Ospedale Galmarini, Tradate, Italy: Daniela Barbieri; ‘Card. Giovanni Panico’ Hospital, Tricase, Italy: Pietro Palmisano; Azienda Sanitaria Universitaria Giuliano Isontina, Trieste, Italy: Massimo Zecchin.

Author Contributions

Conceptualization: G.B., M.P. (Marco Proietti), M.B., I.D., P.P., S.B., F.B., N.B., A.C., A.D.M., F.A.F., S.I., M.P. (Marcello Piacenti), D.P., S.S., L.S., M.M., G.S., J.F.I., D.C., M.Z.-B., R.D.P., R.P.R. Formal analysis and data curation: M.P. (Marco Proietti) Writing—original draft preparation: G.B. and M.P. (Marco Proietti) Writing—review and editing: G.B., M.P. (Marco Proietti), M.B., I.D., P.P., S.B., F.B., N.B., A.C., A.D.M., F.A.F., S.I., M.P. (Marcello Piacenti), D.P., S.S., L.S., M.M., G.S., J.F.I., D.C., M.Z.-B., R.D.P. and R.P.R. All authors have read and agreed to the published version of the manuscript.

Funding

For this research, AIAC received a funding from Medtronic, without any interference of Medtronic on the conduct of the study or on analysis of the results (no profit research grant).

Institutional Review Board Statement

The study was approved by the Institutional Review Board at AVEN Ethic Committee (Approval N. 9/16, protocol N. 1188/CE).

Informed Consent Statement

Informed consent was obtained from all subjects involved in the study.

Data Availability Statement

Data will be provided upon reasonable request.

Conflicts of Interest

G.B.: small speaker’s fees from Medtronic, Boston, Biotronik, Boehringer and Bayer, outside of the submitted work. I.D.: maximal speaker fee from Medtronic, Boston and Biotronik. The other authors declare no conflicts of interest related to the present paper.

Footnotes

Publisher’s Note: MDPI stays neutral with regard to jurisdictional claims in published maps and institutional affiliations.

References

  • 1.Priori S.G., Blomström-Lundqvist C., Mazzanti A., Blom N., Borggrefe M., Camm J., Elliott P.M., Fitzsimons D., Hatala R., Hindricks G., et al. 2015 ESC Guidelines for the management of patients with ventricular arrhythmias and the prevention of sudden cardiac death. [(accessed on 26 August 2021)];Eur. Heart J. 2015 36:2793–2867. doi: 10.1016/j.rec.2016.01.001. Available online: https://academic.oup.com/eurheartj/article-lookup/doi/10.1093/eurheartj/ehv316. [DOI] [PubMed] [Google Scholar]
  • 2.Glikson M., Nielsen J.C., Kronborg M.B., Michowitz Y., Auricchio A., Barbash I.M., Barrabés J.A., Boriani G., Braunschweig F., Brignole M., et al. 2021 ESC Guidelines on cardiac pacing and cardiac resynchronization therapy. [(accessed on 7 November 2021)];Eur. Heart J. 2021 42:3427–3520. doi: 10.1093/eurheartj/ehab364. Available online: http://www.ncbi.nlm.nih.gov/pubmed/34455430. [DOI] [PubMed] [Google Scholar]
  • 3.Raatikainen M.J.P., Arnar D.O., Merkely B., Nielsen J.C., Hindricks G., Heidbuchel H., Camm J. A Decade of Information on the Use of Cardiac Implantable Electronic Devices and Interventional Electrophysiological Procedures in the European Society of Cardiology Countries: 2017 Report from the European Heart Rhythm Association. [(accessed on 26 August 2021)];Europace. 2017 19:ii1–ii90. doi: 10.1093/europace/eux258. Available online: http://academic.oup.com/europace/article/19/suppl_2/ii1/4100657/A-Decade-of-Information-on-the-Use-of-Cardiac. [DOI] [PubMed] [Google Scholar]
  • 4.Zecchin M., Torre M., Carrani E., Sampaolo L., Ciminello E., Ortis B., Ricci R., Proclemer A., Sinagra G. Seventeen-year trend (2001–2017) in pacemaker and implantable cardioverter-defibrillator utilization based on hospital discharge database data: An analysis by age groups. [(accessed on 7 November 2021)];Eur. J. Intern. Med. 2021 84:38–45. doi: 10.1016/j.ejim.2020.09.003. Available online: http://www.ncbi.nlm.nih.gov/pubmed/32933841. [DOI] [PubMed] [Google Scholar]
  • 5.Birnie D.H., Wang J., Alings M., Philippon F., Parkash R., Manlucu J., Angaran P., Rinne P., Coutu B., Low R.A., et al. Risk Factors for Infections Involving Cardiac Implanted Electronic Devices. [(accessed on 7 November 2021)];J. Am. Coll. Cardiol. 2019 74:2845–2854. doi: 10.1016/j.jacc.2019.09.060. Available online: https://linkinghub.elsevier.com/retrieve/pii/S0735109719379331. [DOI] [PubMed] [Google Scholar]
  • 6.Diemberger I., Lorenzetti S., Vitolo M., Boriani G. Infective endocarditis in patients with cardiac implantable electronic devices: Impact of comorbidities on outcome. [(accessed on 7 November 2021)];Eur. J. Intern. Med. 2019 66:e9–e10. doi: 10.1016/j.ejim.2019.06.015. Available online: http://www.ncbi.nlm.nih.gov/pubmed/31280907. [DOI] [PubMed] [Google Scholar]
  • 7.Diemberger I., Bonfiglioli R., Martignani C., Graziosi M., Biffi M., Lorenzetti S., Ziacchi M., Nanni C., Fanti S., Boriani G. Contribution of PET imaging to mortality risk stratification in candidates to lead extraction for pacemaker or defibrillator infection: A prospective single center study. [(accessed on 7 November 2021)];Eur. J. Nucl. Med. Mol. Imaging. 2019 46:194–205. doi: 10.1007/s00259-018-4142-9. Available online: http://www.ncbi.nlm.nih.gov/pubmed/30196365. [DOI] [PubMed] [Google Scholar]
  • 8.Diemberger I., Biffi M., Martignani C., Boriani G. From lead management to implanted patient management: Indications to lead extraction in pacemaker and cardioverter-defibrillator systems. [(accessed on 7 November 2021)];Expert Rev. Med. Devices. 2011 8:235–255. doi: 10.1586/erd.10.80. Available online: http://www.ncbi.nlm.nih.gov/pubmed/21381913. [DOI] [PubMed] [Google Scholar]
  • 9.Sgreccia D., Vitolo M., Valenti A.C., Manicardi M., Boriani G. Burden of disease and costs of infections associated with cardiac implantable electronic devices. [(accessed on 8 November 2021)];Expert Rev. Pharm. Outcomes Res. 2021 :1–10. doi: 10.1080/14737167.2021.1980386. Available online: http://www.ncbi.nlm.nih.gov/pubmed/34517745. [DOI] [PubMed] [Google Scholar]
  • 10.Polyzos K.A., Konstantelias A.A., Falagas M.E. Risk factors for cardiac implantable electronic device infection: A systematic review and meta-analysis. [(accessed on 8 November 2021)];Europace. 2015 17:767–777. doi: 10.1093/europace/euv053. Available online: https://academic.oup.com/europace/article-lookup/doi/10.1093/europace/euv053. [DOI] [PubMed] [Google Scholar]
  • 11.Han H.-C., Hawkins N.M., Pearman C.M., Birnie D.H., Krahn A.D. Epidemiology of cardiac implantable electronic device infections: Incidence and risk factors. [(accessed on 8 November 2021)];Europace. 2021 23:iv3–iv10. doi: 10.1093/europace/euab042. Available online: http://www.ncbi.nlm.nih.gov/pubmed/34051086. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 12.Traykov V., Bongiorni M.G., Boriani G., Burri H., Costa R., Dagres N., Deharo J.C., Epstein L.M., Erba P.A., Snygg-Martin U., et al. Clinical practice and implementation of guidelines for the prevention, diagnosis and management of cardiac implantable electronic device infections: Results of a worldwide survey under the auspices of the European Heart Rhythm Association. [(accessed on 8 November 2021)];Europace. 2019 21:1270–1279. doi: 10.1093/europace/euz137. Available online: http://www.ncbi.nlm.nih.gov/pubmed/31209483. [DOI] [PubMed] [Google Scholar]
  • 13.Diemberger I., Massaro G., Rossillo A., Chieffo E., Dugo D., Guarracini F., Pellegrino P.L., Perna F., Landolina M., De Ponti R., et al. Temporary transvenous cardiac pacing: A survey on current practice. [(accessed on 8 November 2021)];J. Cardiovasc. Med. 2020 21:420–427. doi: 10.2459/JCM.0000000000000959. Available online: http://journals.lww.com/10.2459/JCM.0000000000000959. [DOI] [PubMed] [Google Scholar]
  • 14.Rubini G., Ferrari C., Carretta D., Santacroce L., Ruta R., Iuele F., Lavelli V., Merenda N., D’Agostino C., Sardaro A., et al. Usefulness of 18F-FDG PET/CT in patients with cardiac implantable electronic device suspected of late infection. [(accessed on 8 November 2021)];J. Clin. Med. 2020 9:2246. doi: 10.3390/jcm9072246. Available online: https://www.mdpi.com/2077-0383/9/7/2246. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 15.Galea N., Bandera F., Lauri C., Autore C., Laghi A., Erba P.A. Multimodality imaging in the diagnostic work-up of endocarditis and cardiac implantable electronic device (CIED) infection. [(accessed on 8 November 2021)];J. Clin. Med. 2020 9:2237. doi: 10.3390/jcm9072237. Available online: https://www.mdpi.com/2077-0383/9/7/2237. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 16.Malagù M., Vitali F., Brieda A., Cimaglia P., De Raffaele M., Tazzari E., Musolino C., Balla C., Serenelli M., Cultrera R., et al. Antibiotic prophylaxis based on individual infective risk stratification in cardiac implantable electronic device: The PRACTICE study. [(accessed on 8 November 2021)];Europace. 2021 doi: 10.1093/europace/euab222. Available online: https://academic.oup.com/europace/advance-article/doi/10.1093/europace/euab222/6364843. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 17.Balla C., Brieda A., Righetto A., Vitali F., Malagù M., Cultrera R., Bertini M. Predictors of infection after “de novo” cardiac electronic device implantation. [(accessed on 8 November 2021)];Eur. J. Intern. Med. 2020 77:73–78. doi: 10.1016/j.ejim.2020.02.030. Available online: https://linkinghub.elsevier.com/retrieve/pii/S0953620520300844. [DOI] [PubMed] [Google Scholar]
  • 18.Bahamondez-Canas T.F., Heersema L.A., Smyth H.D.C. Current status of in vitro models and assays for susceptibility testing for wound biofilm infections. [(accessed on 8 November 2021)];Biomedicines. 2019 7:34. doi: 10.3390/biomedicines7020034. Available online: https://www.mdpi.com/2227-9059/7/2/34. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 19.Shariff N., Eby E., Adelstein E., Jan S., Shalaby A., Saba S., Wang N.C., Schwartzman D. Health and Economic Outcomes Associated with Use of an Antimicrobial Envelope as a Standard of Care for Cardiac Implantable Electronic Device Implantation. [(accessed on 8 November 2021)];J. Cardiovasc. Electrophysiol. 2015 26:783–789. doi: 10.1111/jce.12684. Available online: http://doi.wiley.com/10.1111/jce.12684. [DOI] [PubMed] [Google Scholar]
  • 20.Kolek M.J., Dresen W.F., Wells Q.S., Ellis C.R. Use of an antibacterial envelope is associated with reduced cardiac implantable electronic device infections in high-risk patients. [(accessed on 9 November 2021)];PACE-Pacing Clin. Electrophysiol. 2013 36:354–361. doi: 10.1111/pace.12063. Available online: http://doi.wiley.com/10.1111/pace.12063. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 21.Kolek M.J., Patel N.J., Clair W.K., Whalen S.P., Rottman J.N., Kanagasundram A., Shen S.T., Saavedra P.J., Estrada J.C., Abraham R.L., et al. Efficacy of a Bio-Absorbable Antibacterial Envelope to Prevent Cardiac Implantable Electronic Device Infections in High-Risk Subjects. [(accessed on 9 November 2021)];J. Cardiovasc. Electrophysiol. 2015 26:1111–1116. doi: 10.1111/jce.12768. Available online: http://doi.wiley.com/10.1111/jce.12768. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 22.Tarakji K.G., Mittal S., Kennergren C., Corey R., Poole J.E., Schloss E., Gallastegui J., Pickett R.A., Evonich R., Philippon F., et al. Antibacterial Envelope to Prevent Cardiac Implantable Device Infection. [(accessed on 9 November 2021)];N. Engl. J. Med. 2019 380:1895–1905. doi: 10.1056/NEJMoa1901111. Available online: http://www.ncbi.nlm.nih.gov/pubmed/30883056. [DOI] [PubMed] [Google Scholar]
  • 23.Henrikson C.A., Sohail M.R., Acosta H., Johnson E.E., Rosenthal L., Pachulski R., Dan D., Paladino W., Khairallah F.S., Gleed K., et al. Antibacterial Envelope Is Associated With Low Infection Rates After Implantable Cardioverter-Defibrillator and Cardiac Resynchronization Therapy Device Replacement: Results of the Citadel and Centurion Studies. [(accessed on 9 November 2021)];JACC Clin. Electrophysiol. 2017 3:1158–1167. doi: 10.1016/j.jacep.2017.02.016. Available online: http://www.ncbi.nlm.nih.gov/pubmed/29759500. [DOI] [PubMed] [Google Scholar]
  • 24.Essebag V., Verma A., Healey J.S., Krahn A.D., Kalfon E., Coutu B., Ayala-Paredes F., Tang A.S., Sapp J., Sturmer M., et al. Clinically significant pocket hematoma increases long-term risk of device infection: BRUISE CONTROL INFECTION study. [(accessed on 10 November 2021)];J. Am. Coll. Cardiol. 2016 67:1300–1308. doi: 10.1016/j.jacc.2016.01.009. Available online: http://www.ncbi.nlm.nih.gov/pubmed/26988951. [DOI] [PubMed] [Google Scholar]
  • 25.Boriani G., Vitolo M., Wright D.J., Biffi M., Brown B., Tarakji K.G., Wilkoff B.L. Infections associated with cardiac electronic implantable devices: Economic perspectives and impact of the TYRX™ antibacterial envelope. [(accessed on 9 November 2021)];Europace. 2021 23((Suppl. 4)):iv33–iv44. doi: 10.1093/europace/euab126. Available online: https://pubmed.ncbi.nlm.nih.gov/34160600/ [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 26.Daneman N., Homenauth E., Saskin R., Ng R., Ha A., Wijeysundera H.C. The predictors and economic burden of early-, mid- and late-onset cardiac implantable electronic device infections: A retrospective cohort study in Ontario, Canada. Clin. Microbiol. Infect. 2020;26:255.e1–255.e6. doi: 10.1016/j.cmi.2019.02.009. [DOI] [PubMed] [Google Scholar]
  • 27.Sohail M.R., Corey G.R., Wilkoff B.L., Poole J.E., Mittal S., Kennergren C., Greenspon A.J., Cheng A., Lande J.D., Lexcen D.R., et al. Clinical Presentation, Timing, and Microbiology of CIED Infections: An Analysis of the WRAP-IT Trial. [(accessed on 10 November 2021)];JACC Clin. Electrophysiol. 2021 7:50–61. doi: 10.1016/j.jacep.2020.07.021. Available online: http://www.ncbi.nlm.nih.gov/pubmed/33478712. [DOI] [PubMed] [Google Scholar]
  • 28.Greenspon A.J., Eby E.L., Petrilla A.A., Sohail M.R. Treatment patterns, costs, and mortality among Medicare beneficiaries with CIED infection. [(accessed on 10 November 2021)];PACE-Pacing Clin. Electrophysiol. 2018 41:495–503. doi: 10.1111/pace.13300. Available online: http://www.ncbi.nlm.nih.gov/pubmed/29411401. [DOI] [PubMed] [Google Scholar]
  • 29.Tarakji K.G., Wazni O.M., Harb S., Hsu A., Saliba W., Wilkoff B.L. Risk factors for 1-year mortality among patients with cardiac implantable electronic device infection undergoing transvenous lead extraction: The impact of the infection type and the presence of vegetation on survival. [(accessed on 11 November 2021)];Europace. 2014 16:1490–1495. doi: 10.1093/europace/euu147. Available online: http://www.ncbi.nlm.nih.gov/pubmed/25087154. [DOI] [PubMed] [Google Scholar]
  • 30.Brunner M., Olschewski M., Geibeli A., Bode C., Zehender M. Long-term survival after pacemaker implantation: Prognostic importance of gender and baseline patient characteristics. [(accessed on 11 November 2021)];Eur. Heart J. 2004 25:88–95. doi: 10.1016/j.ehj.2003.10.022. Available online: http://www.ncbi.nlm.nih.gov/pubmed/14683747. [DOI] [PubMed] [Google Scholar]
  • 31.Boriani G., Berti E., Belotti L.M.B., Biffi M., De Palma R., Malavasi V.L., Bottoni N., Rossi L., De Maria E., Mantovan R., et al. Cardiac device therapy in patients with left ventricular dysfunction and heart failure: ‘real-world’ data on long-term outcomes (mortality, hospitalizations, days alive and out of hospital) [(accessed on 11 November 2021)];Eur. J. Heart Fail. 2016 18:693–702. doi: 10.1002/ejhf.509. Available online: http://www.ncbi.nlm.nih.gov/pubmed/27060289. [DOI] [PubMed] [Google Scholar]
  • 32.Kunutsor S.K., Whitehouse M.R., Blom A.W., Beswick A.D. Systematic review of risk prediction scores for surgical site infection or periprosthetic joint infection following joint arthroplasty. [(accessed on 12 November 2021)];Epidemiol. Infect. 2017 145:1738–1749. doi: 10.1017/S0950268817000486. Available online: https://www.cambridge.org/core/product/identifier/S0950268817000486/type/journal_article. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 33.Proietti M., Mujovic N., Potpara T.S. Optimizing Stroke and Bleeding Risk Assessment in Patients with Atrial Fibrillation: A Balance of Evidence, Practicality and Precision. [(accessed on 7 December 2018)];Thromb. Haemost. 2018 118:2014–2017. doi: 10.1055/s-0038-1676074. Available online: http://www.thieme-connect.de/DOI/DOI?10.1055/s-0038-1676074. [DOI] [PubMed] [Google Scholar]
  • 34.Mull H.J., Stolzmann K.L., Shin M.H., Kalver E., Schweizer M.L., Branch-Elliman W. Novel Method to Flag Cardiac Implantable Device Infections by Integrating Text Mining With Structured Data in the Veterans Health Administration’s Electronic Medical Record. [(accessed on 12 November 2021)];JAMA Netw Open. 2020 3:e2012264. doi: 10.1001/jamanetworkopen.2020.12264. Available online: http://www.ncbi.nlm.nih.gov/pubmed/32955571. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 35.Palmisano P., Ziacchi M., Belotti G., Rapacciuolo A., Santini L., Stabile G., Zoni Berisso M., De Ponti R., Landolina M., Ricci R.P., et al. Clinical and organizational management of cardiac implantable electronic device replacements: An Italian Survey promoted by AIAC (Italian Association of Arrhythmology and Cardiac Pacing) [(accessed on 12 November 2021)];J. Cardiovasc. Med. 2019 20:531–541. doi: 10.2459/JCM.0000000000000817. Available online: http://www.ncbi.nlm.nih.gov/pubmed/31259858. [DOI] [PubMed] [Google Scholar]

Associated Data

This section collects any data citations, data availability statements, or supplementary materials included in this article.

Data Availability Statement

Data will be provided upon reasonable request.


Articles from Journal of Personalized Medicine are provided here courtesy of Multidisciplinary Digital Publishing Institute (MDPI)

RESOURCES