Abstract
Background
Measurement of oxygen saturation (SpO2) during the 6 minute walk test (6MWT) could be impacted by the measurement site.
Aims
To compare SpO2 and heart rate (HR) between forehead and finger sensors during the 6MWT. Sensor readings were also to be compared for signal quality and with capillary blood gas (CBG) pre and post 6MWT.
Method
80 subjects with pulmonary vascular disease (PVD) and/or interstitial lung disease (ILD) performed the 6MWT. Pulse oximetry was recorded at 30 s intervals. CBG was taken pre and post 6MWT to determine capillary oxygen saturation (SCO2).
Results
The forehead sensor recorded higher values for SpO2 (p < 0.001) and HR (p < 0.01) compared with the finger sensor during the 6MWT. For both sensors, the demonstrated bias compared to CBG post 6MWT was higher and more variable in subjects who desaturated. During the 6MWT there was a higher occurrence (p < 0.001) of poor signal quality in the finger sensor compared with the forehead sensor.
Conclusion
This study suggests that the sensor site can impact pulse oximetry readings. The variance in bias suggests pulse oximetry may not accurately reflect SCO2 measurements particularly in subjects who desaturate during 6MWT.
Keywords: Pulse oximetry, 6 minute walk test, oxygen desaturation, pulmonary vascular disease, interstitial lung disease
Introduction
There are two types of pulse oximetry: transmission and reflectance oximetry. Transmission pulse oximetry is most widely utilised, particularly the finger sensor. Transmission pulse oximetry can be less effective in poor peripheral circulation as it can be more prone to the effects of vasoconstriction and this can decrease accuracy of readings due to reduced photoplethysmography signal . 1 Reflectance pulse oximetry, as used in the forehead sensor, is less vulnerable to vasoconstriction . 2 Additionally, it can be more securely attached which could make it less prone to motion artefact and therefore more reliable during exercise testing.
The 6 minute walk test (6MWT) is clinically used to monitor a range of cardiorespiratory disorders including commonly pulmonary vascular disease (PVD) and interstitial lung disease (ILD). Guidelines for the 6MWT recommend pulse oximetry for detection of exercise-induced oxygen desaturation .3,4 The 6MWT has advantages such as increased sensitivity in detecting exercise-induced oxygen desaturation compared with maximal exercise testing. 5 Detection of significant exercise-induced oxygen desaturation from the 6MWT may aid decisions in prescribing ambulatory oxygen and additionally exercise-induced oxygen desaturation has shown to be prognostically significant in both PVD and ILD.6–8
A few investigations have demonstrated significantly higher SpO2 values recorded using the forehead sensor compared to the finger sensor with exercise.9,10 However, only one study investigated this during the 6MWT; this was in a scleroderma population where only pre and post SpO2 values were compared between sensors and accuracy of values were not compared to a blood gas standard. 10
The primary aim of this investigation was therefore to compare the differences in SpO2 values between the forehead and finger sensors recorded throughout the duration of the 6MWT in subjects with PVD and ILD
Secondary aims included comparing the agreement of sensor readings to a reference of arterialized capillary oxygen saturation (SCO2) pre and post 6MWT and assessing sensors signal quality during the 6MWT in subjects with PVD and ILD. Arterialized capillary blood sampling was selected as the technique to obtain reference of arterial oxygen saturation as it has advantages of being minimally invasive and low risk compared with arterial sampling.
Methods
Study design
This was a single centre study taking place in the Department of Respiratory Physiology at Royal Papworth Hospital, Cambridge. Ethical approval for the study (210,827) was given by the research committee and Health Research Authority. Written informed consent was obtained from all subjects.
Subjects
Royal Papworth Hospital is a specialist centre for referral of patients diagnosed with PVD and ILD. Many of these patients as part of their clinical evaluation at this hospital complete a 6MWT. Subjects were therefore recruited to the study that had a planned clinical 6MWT. In total, 80 patients were recruited with inclusion criteria of a diagnosis of PVD and/or ILD and aged between 18 and 80 years old. Individuals who used walking aids, ambulatory oxygen or currently smoked were excluded.
Six-minute walk test
The oximeter used was the Masimo Rad-5® with settings of an averaging time rate of 8 s and normal sensitivity setting. Two Rad-5 devices were attached to the patient simultaneously with the forehead sensor attached to one device and the finger sensor to the other device. A reusable forehead reflectance sensor (LNCS TF-1) was placed above the subject’s left eyebrow and secured with a headband. The reusable finger sensor (LNCS DC-I) was attached to the subject’s right index finger. No skin preparation was performed other than removal of nail varnish (if present) on the subject’s right index finger. The 6 minute walk test (6MWT) was performed according to the department protocol based on ATS guidelines 3 and with further guidance, particularly regarding criteria for termination of the test, determined by referring clinical teams. In ILD subjects, the 6MWT test was terminated early if SpO2 dropped to 75% or below using the forehead sensor or had significant symptoms as described in ATS guidance. In PVD subjects, the 6MWT test was only terminated early if the patient presented unwell with symptoms of anginal chest pain, extreme light headiness or unsteadiness not with a significant drop in SpO2 values in isolation. None of the subjects in the study had their 6MWT terminated early due to significant drop in SpO2 values or presence of significant symptoms. The subject performed their normal clinical 6MWT along a 10-m corridor with both devices and sensors attached. Both devices displayed measurements of SpO2 and heart rate (HR) continuously during the 6MWT and values were recorded from both devices at the same time by the investigator at 30 s intervals during the 6MWT. During the 6MWT the signal quality was continuously monitored on the Masimo Rad-5 device from the signal IQ bar displayed which provided a visual indicator of the signal quality. A high signal IQ bar indicated that the signal quality was good and when the IQ bar dropped to two bars or less and turned red this was poor signal quality and SpO2 values recorded at this point of the 6MWT was noted as an occurrence of poor signal.
Capillary blood gas
Capillary blood gas (CBG) samples were obtained at rest and post walk from the earlobes which were arterialised using Deep Heat (Mentholatum Corporation). Resting CBG was obtained from the subject seated and rested in normal ambient conditions. Post CBG was obtained from the subject immediately post walk test in a seated position in the walk corridor. Readings were recorded from the oximeter devices when half of the CBG capillary tube had been filled with blood.
Statistical analyses
Data analysis was performed using Statistical Package for Social Sciences (SPSS) version 25. The level of statistical significance was set at p < 0.05. A two-way repeated measure analysis of variance (ANOVA) was performed to investigate differences and interactions in SpO2 and HR readings between the sensors during the 6MWT. Within-subject factors entered in the ANOVA model were sensor and time. All readings recorded at 30 s intervals of the 6MWT were entered into the analysis. To compare the occurrences of poor signal between sensors during the 6MWT, the Wilcoxon Rank Sum test was performed. Agreement between the SpO2 and SCO2 was assessed using the Bland Altman (BA) method which calculated the bias (SpO2-SCO2) for both sensors 11 . As per BA method 95% limits of agreement was calculated and confidence intervals for the 95% limits of agreement. Linear regression was used to test for any change in the bias in the Bland Altman analysis with change in oxygen saturation range.
Results
Population characteristics
Mean demographics of the population were age was 59 ± 13 (range 28–80), male (n=40) female (n=40) and mean resting SCO2 was 94 ± 4.6% (range 74–99%).
In total, 80 subjects were enrolled into the study. Fifty three subjects had PVD and 24 subjects with ILD. Three subjects were diagnosed with both PVD and ILD. The disease and clinical characteristics of the population are presented in Supplementa1 Table 1
SpO2 and HR response during the 6MWT
To investigate for differences and interactions in SpO2 and HR values in the different cohorts between the sensors during the 6MWT, a two-way repeated measure ANOVA was performed. Within-subject factors entered in the ANOVA model were sensor and time. Results are presented in Table 1. Mean differences in SpO2 and HR values recorded between forehead and finger sensors are reported in Supplemental Table 2.
Table 1.
ANOVA results comparing differences and interaction of SpO2 and HR values between sensors during the 6MWT.
Whole population | |||
---|---|---|---|
— | Sensor | Time | Sensor/Time |
SpO2 | p < 0.001 | p <0.001 | NS |
HR | p <0.01 | p <0.001 | p <0.05 |
— | PVD population | ||
— | Sensor | Time | Sensor/Time |
SpO2 | p < 0.001 | p < 0.001 | NS |
HR | p <0.001 | p <0.001 | p <0.05 |
— | ILD population | ||
— | Sensor | Time | Sensor/Time |
SpO2 | p < 0.01 | p <0.001 | NS |
HR | p < 0.01 | p <0.001 | NS |
Within-Subject Factors.
Sensor – to determine if SpO2 and HR values were significantly different between sensors.
Time – to determine if SpO2 and HR values significantly changed over time of the 6MWT for both sensors.
Sensor/Time – to determine if any changes in SpO2 and HR over time of the 6MWT were significantly different between sensors.
ANOVA= Analysis of variance, p value of significance, NS= not significant, SpO2 = oxygen saturation, HR= heart rate, PVD= pulmonary vascular disease, ILD= interstitial lung disease.
For all cohorts, time was a significant factor, as expected, with SpO2 decreasing and HR increasing during the course of the 6MWT (p < 0.001) (see Table 1). Sensor was a significant factor for both SpO2 and HR values in the whole cohort and when cohorts were tested separately was significant in both PVD and ILD cohorts. The SpO2 recorded using the forehead sensor significantly higher compared with readings using the finger sensor (p < 0.001), although the average difference in SpO2 values at each time point of the 6MWT was clinically small ranging from 1 to 3% (See Figure 1(a)). The HR values were also significantly higher using the forehead sensor compared with readings using the finger sensor (p < 0.01). The average difference in HR values at each time point of the 6MWT ranged from 0 to 9 bpm as presented in Figure 1(b).
Figure 1.
Comparison of SpO2 response (A) and HR response (B) between forehead and finger sensors during the 6MWT for the whole population group (presented as mean and standard error).
The interaction of the factors sensor and time was only significant for HR (p < 0.05) and when cohorts were tested separately was only found in the PVD cohort (p < 0.05). This suggests that for the PVD cohort the change of HR response over the timespan for the 6MWT was significantly different between the 2 sensors.
Comparison of signal quality between sensors
The occurrence of poor signal during the 6MWT was compared between the sensors and these results are presented in Table 2. There was a higher occurrence (p < 0.001) of poor signal using the finger sensor (189/1040) compared with the forehead sensor (34/1040) during the 6MWT as shown by the Wilcoxon Rank Sum test in the whole population and additionally for the PVD cohort when tested separately
Table 2.
Occurrence of poor signal quality measurements during 6 MWT.
Occurrence of poor signal in total measurements | Z value | Significance (p value) | ||
---|---|---|---|---|
Whole population | Forehead | 34/1040 | −4.6 | p< 0.001 |
Finger | 189/1040 | |||
PVD population | Forehead | 33/728 | −4.6 | p <0.001 |
Finger | 176/728 | |||
ILD population | Forehead | 1/351 | −1.8 | p= 0.07 (NS) |
Finger | 30/351 |
The occurrence of poor signal is reported as the number of poor signal readings recorded out of the total number of recorded readings during the 6MWT; Abbreviations – PVD= pulmonary vascular disease, ILD= interstitial lung disease, Z value from Wilcoxon Rank Sum test, NS= not significant.
Agreement of sensor readings to capillary blood gas
Capillary blood collection
Average time to CBG collection was 86 ± 49 s. Resting CBG was obtained successfully in 65/80 patients and post walk in 60/80 patients. Samples were not obtained successfully in all subjects due to insufficient CBG collected for analysis due to subjects bleeding insufficiently. This was most frequently observed in subjects with poor circulation. Four patients presenting with exercise desaturation during 6MWT had resaturated to normal by time of CBG collection post walk due to fast recovery.
Comparisons of forehead and finger sensor readings of SpO2 with CBG SCO2 are presented for the different cohorts in Table 3 and in Figures 2 and 3 and also Supplemental Figures 1–4. The bias is presented, calculated using Bland Altman analysis which compares agreement between SpO2 and SCO2.
Table 3.
Sensor SpO2 compared with CBG ScO2 at rest and post 6 MWT.
Whole population group | PVD population group | ILD population group | |||||||||||
---|---|---|---|---|---|---|---|---|---|---|---|---|---|
Rest (n=65/80) | Post 6MWT (n=65/80) | Rest (n=41) | Post 6MWT (n=42) | Rest (n=19) | Post 6MWT (n=17) | ||||||||
Values (%) | Bias | Values (%) | Bias | Values | Bias | Values (%) | Bias | Values (%) | Bias | Values (%) | Bias | ||
SCO2 % | 95 (94–96) | n/a | 93 (88–96) | n/a | 95 (91–95) | n/a | 90 (85–96) | n/a | 96 (95–97) | n/a | 94 (90–97) | n/a | |
Forehead (SpO2)% | 97 (96–99) | 2.77 ± 2.13 | 95 (89–98) | 2.98 ± 3.16 | 97 (94–97) | 2.85 ± 2.29 | 93 (88–98) | 3.56 ± 3.70 | 99 (98–100) | 2.79 ± 1.51 | 97 (91–100) | 2.35 ± 2.34 | |
Finger (SpO2)% | 96 (93–97) | 0.68 ± 2.37 | 93 (88–96) | 0.85 ± 4.87 | 95 (92–97) | 1.00 ± 2.42 | 92 (85–96) | 1.85 ± 5.31 | 96 (99–97) | −0.10 ± 1.91 | 92 (88–97) | −0.88 ± 2.76 |
This table presents SpO2 readings from the different sensors and SCO2 obtained pre and post 6MWT.Abbreviations Values (%) for Rest and Post 6MWT presented as median, lower and upper interquartile range, n= number of patients in which CBG sample was obtained, n/a= not applicable, Bias (SpO2-ScO2).
Figure 2.
Bland Altman plots comparing forehead and finger sensor SpO2 readings to CBG ScO2 at rest for the whole population (presented as mean bias, upper and lower limits of agreement calculated as ± 1.96 X SD of the bias and 95% confidence intervals). A: Comparing baseline forehead SpO2 to SCO2 in subjects at rest. B: Comparing finger SpO2 to SCO2 in subjects at rest.
Figure 3.
Bland Altman plots comparing forehead and finger sensor SpO2 readings to CBG SCO2 post 6MWT for the whole population (presented as mean bias, upper and lower limits of agreement calculated as ± 1.96 X SD of the bias and 95% confidence intervals). A: Comparing baseline forehead SpO2 to SCO2 in subjects post 6MWT B: Comparing finger SpO2 to SCO2 in subjects post 6MWT.
At rest in the whole cohort the forehead sensor exhibited an average bias of 2.77 ± 2.13 whilst the finger sensor had an average bias of 0.68 ± 2.37 indicating closer agreement of finger sensor SpO2 with SCO2 than the forehead sensor as demonstrated by Figure 2. However, there was no systematic change in bias with change in oxygen saturation scale for both forehead sensor (r = 0.15, p > 0.05) and finger sensor (r = 0.06 p > 0.05). This indicates that at rest the level of agreement between both sensor readings to CBG was similar across the oxygen saturation scale range (85–100%).
Post 6MWT in the whole cohort the forehead sensor demonstrated an average bias of 2.98 ± 3.16 compared to an average bias of 0.85 ± 4.87 for the finger sensor again indicating closer agreement of finger sensor SpO2 with SCO2 than the forehead sensor as demonstrated by Figure 3 Bland Altman. There was however significant systematic change in bias with change in oxygen saturation scale for both forehead sensor (r = 0.33, p < 0.001) and finger sensor (r = 0.12 p < 0.01). This demonstrates that the level of agreement between sensor readings to CBG was different across the oxygen saturation scale range (65–100%). There was a higher bias in the lower oxygen saturation range indicating a lower level of agreement between sensor readings to CBG. Although the bias was lower for the finger sensor it was variable for both sensors, particularly for the finger sensor due to its wider limits of agreement.
Bland Altman analyses were additionally performed in PVD and ILD to compare agreement between SpO2 and SCO2 between the different sensors and these are presented in Supplemental Figures 1–4. The presented bias values were commonly higher and more variable in the PVD cohort compared to the ILD cohort where sensor values presented with more negative bias and mostly had closer agreement to SCO2.
Discussion
Pulse oximetry plays an important role in monitoring a range of patients with cardiorespiratory disorders during the 6MWT and may play a significant role in aiding clinical decisions in prescribing ambulatory oxygen. In this study, we investigated how different oximetry sensors influence the values recorded and compared these to a capillary blood gas standard. This study, we believe, is unique in comparing simultaneous measurements of SpO2 between forehead and finger sensors throughout the 6MWT.
Our findings demonstrated that the measurement site of pulse oximetry significantly impacted values recorded with significantly higher SpO2 using the forehead sensor compared with the finger sensor throughout the 6MWT in both our PVD and ILD cohorts. Average differences between sensor SpO2 readings, however, were clinically small at 1–3% and within the quoted accuracy error of the oximeter . 12 It is important to highlight though that in many subjects substantial clinical differences in sensor SpO2 readings during the 6MWT were observed and therefore not within the quoted accuracy of oximetry. Although we did not investigate the difference in sensor readings in a range of respiratory conditions we believe our oximetry findings are likely to be reflected in other respiratory conditions. Indeed, other investigations 9,10 have similarly reported significantly higher SpO2 values using the forehead sensor compared to the finger sensor with exercise in other cohorts such as COPD and scleroderma respectively. Our finding that the forehead sensor recorded higher SpO2 than the finger sensor will likely have several clinical implications. In particular, the type of sensor used during the 6MWT could impact the result obtained in terms of extent of exercise desaturation which may influence clinical decision making. Pulse oximetry is used during ambulatory oxygen assessment to help decide eligibility for supplemental oxygen therapy and the use of the forehead sensor may result in fewer patients, for example meeting these criteria. Possible reasons for differences in sensor readings could be due to several factors. The differences in SpO2 values could be due to differences in perfusion in the forehead compared with the finger. The finger sensor can be less effective in poor perfusion states. This is because the peripheral circulation is under control of the autonomic nervous system and in either low temperature or low cardiac output conditions these vessels can vasoconstrict in order to maintain central blood flow. Vasoconstriction can reduce the photoplethysmography signal thereby impacting the readings from this sensor . 1 In the forehead sensor, however, the light source and photodetector are positioned next to each other on the surface of the skin which is a likely advantage in conditions with poor peripheral circulation as this is less vulnerable to vasoconstriction. Peripheral vasoconstriction has demonstrated to impact measured SpO2 values in a number of clinical investigations .13-15 Differences in perfusion between the forehead and finger could have been a factor in differences in SpO2 values in our investigation due to a proportion of our cohort presenting with characteristics of poor peripheral circulation such as scleroderma, Raynaud’s and peripheral cyanosis. Additionally, differences in sensor readings may have been partly impacted by the different methodologies used to record pulse oximetry, that is reflectance and transmission. Transmission oximetry calculates SpO2 by the light emitting diode transmitting light through the vascular bed. In reflectance oximetry, the transmitted light will be reflected back to the photodetector on the same side of the vascular bed . 1 It is probable that the two measurement techniques may use different algorithms to calculate SpO2. Further investigation is needed to explore the causes of these differences in sensor readings.
Our findings also demonstrated higher recorded HR values in the forehead sensor compared with the finger sensor. The results from the ANOVA additionally demonstrated that the interaction of the factor sensor with time for HR readings was significant in the whole cohort and when cohorts were tested separately was significant for the PVD cohort. This indicates that in the PVD cohort the change in HR response over the timespan of the 6MWT was significantly different between the two sensors. A likely explanation of this finding is that a proportion of subjects with PVD were found to have a decrease in measured HR values from mid-point to the end of the 6MWT with the finger sensor which was not observed with the forehead sensor. This reduced HR response may have reflected occurrence of peripheral vasoconstriction in these subjects as peripheral vasoconstriction has been shown to reduce measured HR using the finger sensor . 13 This was a likely response to a reduction of cardiac output in these patients which can be commonly reduced in PVD . 16 The decrease in HR response in the finger sensor during progression of the walk was not observed in the ILD cohort. Some other investigators have examined the impact of different sensors on HR readings with exercise. Yamayha et al. found using the forehead sensor a better bias and precision for recording HR during exercise compared with the finger sensor and that the finger sensor significantly underestimated the HR when compared to electrocardiogram . 17 Conversely, Wilson et al. 9 found no difference in HR between sensors during exercise .
During the 6MWT, we observed a higher occurrence of poor signal using the finger sensor compared with the forehead sensor which was significant for the PVD but not for the ILD population. Lack of significance in ILD may be due to the smaller population size, as statistical significance was nearly achieved, although significantly the occurrence of poor signal was seen in certain ILD subjects such as those with scleroderma. The likely explanation in the PVD cohort, for poor signal to be more frequently observed using the finger sensor was likely due to peripheral vasoconstriction and would be expected to occur in subjects with poor circulation during walking. In poor circulatory conditions, there is a smaller amplitude signal which can be difficult to distinguish from background noise that may increase during walking due to motion thereby increasing chances of a poor signal with the finger sensor . 18 Clinically the finger sensor is the most widely used sensor for the 6MWT. Our findings significantly highlight that there is a higher probability of poor signal occurring in the finger sensor and therefore measurement error. Our findings suggest that the forehead sensor may be the preferred sensor choice especially in PVD and in ILD patients with poor circulation, for example those with scleroderma.
A few investigations have compared different sensors and how they agree to a blood gas standard with exercise ,9,17,19 but we are the first to compare forehead and finger sensor readings pre and post 6MWT to a blood gas standard. The forehead sensor both at rest and post 6MWT presented with a predominantly positive bias which indicated that the forehead sensor was recording above the SCO2. A limitation with capillary blood is that it is likely to be less oxygenated compared to arterial blood, and this could have been a partial factor in the forehead sensor predominately recording above the SCO2. In comparison, at rest and post 6MWT, the finger sensor also demonstrated on average a smaller but still positive bias, but also had a greater amount of negative bias and overall more variable bias. At rest for both sensors there was no systematic change in bias with change in oxygen saturation scale which indicated the level of agreement for both sensors to CBG was similar across the oxygen saturation scale range. Post 6MWT, however, there was a significant systematic change in bias with change in oxygen saturation scale for both sensors. This indicates that the level of agreement between sensor readings to CBG was different across the oxygen saturation scale ranges as higher and more variable bias values were seen in the lower oxygen saturation range levels indicating a lower level of agreement between both sensor readings to CBG. This result highlights that the agreement of both sensor types to SCO2 post 6MWT was lower and more variable in subjects who desaturated and this suggests that pulse oximetry measured by both forehead and finger sensor may not accurately reflect SCO2 in patients who desaturate during the 6MWT. This may be expected as pulse oximetry is less accurate in measuring SpO2 below 80% . 12 For both sensors, when looking at PVD and ILD cohorts separately the demonstrated bias values tended to be higher and more variable in the PVD cohort compared to the ILD cohort. This likely relates to characteristics of the PVD cohort with numerous subjects presenting with poor perfusion and additionally a number of subjects had congenital heart disease (CHD) where particularly high positive bias values were observed for both sensors and were least accurate with comparison to CBG readings. This may be expected as oximetry often over-estimates in CHD due to low oxygen saturation and perfusion . 20
Our findings importantly highlight the possible inaccuracy of pulse oximetry in reflecting the extent of exercise desaturation and we would suggest clinicians consider performing blood gas measurements alongside oximetry when making important clinical decisions such as when prescribing ambulatory oxygen. Despite the limitations of pulse oximetry, it is a clinically readily available tool to detect exercise-induced desaturation. Our findings showed that both sensors were able to detect exercise desaturation in all populations in accordance with the CBG. Blood gas collection techniques are technically difficult, and would unlikely be suitable to perform routinely with every 6MWT. As observed from our experience, CBG was not always obtained successfully especially in subjects with poor circulation or sufficiently quickly to reflect exercise desaturation in all our subjects due to fast recovery.
Limitations of study and future directions
The American Thoracic guidelines for 6MWT 3 state that the 6MWT should be performed on a 30 m corridor but due to insufficient corridor space the 6MWT was performed on a 10 m corridor. This a minor limitation as corridor length has demonstrated to affect walking distance and has shown not to impact SpO2 values during the 6MWT . 21
The main limitation of this study was that the gold standard, arterial blood gas sampling by radial artery cannulation, was not performed. This was due to the risks posed with cannulation and therefore arterialized capillary blood was collected as a lower risk alternative. We also had technical limitations with CBG collection in a small number of subjects as already discussed. Further investigation is needed to explore the causes of differences in sensor readings and compare with arterial blood.
Conclusion
To conclude, the measurement site of pulse oximetry significantly impacts values recorded during the 6MWT with significantly higher SpO2 and HR values occurring in the forehead compared with the finger sensor in our PVD and ILD cohorts. In our cohorts, the forehead sensor was more reliable in signal quality for measurement of SpO2 values during the 6MWT, but predominantly recorded higher than SCO2, whereas the finger sensor demonstrated a lower bias but, more variable level of agreement to SCO2. Significantly our results showed that both sensors may not accurately reflect SCO2 especially in those who desaturate.
Supplemental Material
Supplemental Material, sj-jpg-1-crd-10.1177_14799731211070844 for Comparison of forehead and finger oximetry sensors during the six minute walk test by Lucy Robertson, Rachel Lowry, Karl Sylvester, Helen Parfrey, Beverley Moseley, Karen Sheares and Katrina Oates in Chronic Respiratory Disease
Supplemental Material, sj-jpg-2-crd-10.1177_14799731211070844 for Comparison of forehead and finger oximetry sensors during the six minute walk test by Lucy Robertson, Rachel Lowry, Karl Sylvester, Helen Parfrey, Beverley Moseley, Karen Sheares and Katrina Oates in Chronic Respiratory Disease
Supplemental Material, sj-jpg-3-crd-10.1177_14799731211070844 for Comparison of forehead and finger oximetry sensors during the six minute walk test by Lucy Robertson, Rachel Lowry, Karl Sylvester, Helen Parfrey, Beverley Moseley, Karen Sheares and Katrina Oates in Chronic Respiratory Disease
Supplemental Material, sj-jpg-4-crd-10.1177_14799731211070844 for Comparison of forehead and finger oximetry sensors during the six minute walk test by Lucy Robertson, Rachel Lowry, Karl Sylvester, Helen Parfrey, Beverley Moseley, Karen Sheares and Katrina Oates in Chronic Respiratory Disease
Declaration of conflicting interests: The author(s) declared no potential conflicts of interest with respect to the research, authorship and/or publication of this article.
Funding: The author(s) received no financial support for the research, authorship and/or publication of this article.
Supplemental Material: Supplemental material for this article is available online.
ORCID iD
Lucy Robertson https://orcid.org/0000-0002-4635-9698
References
- 1.Nitzan M, Romen A, Koppel R. Pulse oximetry: fundamentals and technology update. Med Devices 2009; 8: 231–239. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 2.Macleod DB, Cortinez LI, Keifer JC, et al. The desaturation response time of finger pulse oximeters during mild hypothermia. Anesthesia 2005; 60(1): 65–71. [DOI] [PubMed] [Google Scholar]
- 3.American Thoracic Society . ATS statement: guidelines for the six-minute walk test. AM J Respir Crit Care Med 2002; 166(1): 111–117. [DOI] [PubMed] [Google Scholar]
- 4.Holland AE, Martijn AS, Troosters MA, et al. An official european respiratory society/american thoracic society technical standard: field walking tests in chronic respiratory disease. Eur Respir J 2014; 44. [DOI] [PubMed] [Google Scholar]
- 5.Poulain M, Durand F, Palomba B, et al. 6-minute walk testing is more sensitive than maximal incremental cycle testing for detecting oxygen desaturation in patients with COPD. Chest 2003; 123(5): 1401–1407. [DOI] [PubMed] [Google Scholar]
- 6.Paccico G, Martinez FJ, Bossone E, et al. Oxygen desaturation on the six-minute walk test and mortality in untreated primary pulmonary hypertension. Eur Respir J 2001; 17(4): 647–652. [DOI] [PubMed] [Google Scholar]
- 7.Santos M, Furtado I, Goncalves F, et al. Prognostic impact of oxygen saturation during the 6-minute walk test in pulmonary arterial hypertension. Eur Respir J 2016; 48: 60. [Google Scholar]
- 8.Lama VN, Flaherty KR, Toews GB, et al. Prognostic value of desaturation during a 6-Minute Walk Test in Idiopathic Interstitial Pneumonia. AM J Respir Crit Care Med 2003; 168(9): 1084–1090. [DOI] [PubMed] [Google Scholar]
- 9.Wilson S, Cecins N, Jenkins S, et al. Comparing finger and forehead sensors to measure oxygen saturation in people with chronic obstructive pulmonary disease. Respirology 2013; 18(7): 1143–1147. [DOI] [PubMed] [Google Scholar]
- 10.Wilsher M, Good N, Hopkins R, et al. The six-minute walk test using forehead oximetry is reliable in assessment of scleroderma lung disease. Respirology 2012; 17: 647–652. [DOI] [PubMed] [Google Scholar]
- 11.Bland JM, Altman DG. Statistical methods for assessing agreement between two methods of clinical measurement. Lancet 1986; 327: 307–310. [PubMed] [Google Scholar]
- 12.Masimo . Assessing the Clinical Accuracy of Pulse Oximeter in True Clinical Settings, 2007, https://www.masimo.co.uk/siteassets/uk/documents/pdf/clinicalevidence/whitepapers/lab4709a_whitepapers_pulse_oximetry_accuracy.pdf (accessed on 30 September 2021). [Google Scholar]
- 13.Talke P, Stapelfeldt C. Effects of peripheral vasoconstriction on pulse oximetry. J CLIN MONIT 2006; 20: 305–309. [DOI] [PubMed] [Google Scholar]
- 14.Hynson JM, Sessler DI, Belani K, et al. Thermoregulatory vasoconstriction during propofol/nitrous oxide anesthesia in humans: threshold and oxyhemoglobin saturation. Anaesth 1992; 75: 947–952. [DOI] [PubMed] [Google Scholar]
- 15.Schramm WM, Bartunek A, Gilly H. Effect of local limb temperature on pulse oximetry and the plethysmographic pulse wave. Int J Clin Monit Comput 1997; 14: 17–224. [DOI] [PubMed] [Google Scholar]
- 16.Lai YC, Potoka KC, Champion HC, et al. Pulmonary arterial hypertension: the clinical syndrome. Circ Res 2014; 115(1): 115–130. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 17.Yamaya Y, Bogaard HJ, Wagner PD, et al. Validity of pulse oximetry during maximal exercise in normoxia, hypoxia, and hyperoxia. J Appl Physiol 2002; 92: 162–168. [DOI] [PubMed] [Google Scholar]
- 18.Masimo . Monitoring SpO2 during Conditions Involving Low Peripheral Perfusion, 2006, http://www.masimo.de/siteassets/uk/documents/pdf/clinical-evidence/whitepapers/lab3692a_whitepapers_very_low_perfusion.pdf (accessed 7 March 2019). [Google Scholar]
- 19.Hanson LJ, Mottram CD, Scanlon PD. Comparison of the masimo pulse oximeter utilizing various probe sites at rest and with exercise against a laboratory oximeter using arterial blood. Respir Care 2003; 48: 1085. [Google Scholar]
- 20.Whitney GM, Tucker LR, Hall SR, et al. Clinical evaluation of the accuracy of masimo set and nellcor N-595 oximeters in children with cyanotic congenital heart disease. Anesthesiology 2005; 103: A1344. [Google Scholar]
- 21.Beekman E, Mesters I, Hendriks EJ, et al. Course length of 30 metres versus 10 metres has a significant influence on six-minute walk distance in patients with COPD: an experimental crossover study. J Physiother 2013; 59: 169–176. [DOI] [PubMed] [Google Scholar]
Associated Data
This section collects any data citations, data availability statements, or supplementary materials included in this article.
Supplementary Materials
Supplemental Material, sj-jpg-1-crd-10.1177_14799731211070844 for Comparison of forehead and finger oximetry sensors during the six minute walk test by Lucy Robertson, Rachel Lowry, Karl Sylvester, Helen Parfrey, Beverley Moseley, Karen Sheares and Katrina Oates in Chronic Respiratory Disease
Supplemental Material, sj-jpg-2-crd-10.1177_14799731211070844 for Comparison of forehead and finger oximetry sensors during the six minute walk test by Lucy Robertson, Rachel Lowry, Karl Sylvester, Helen Parfrey, Beverley Moseley, Karen Sheares and Katrina Oates in Chronic Respiratory Disease
Supplemental Material, sj-jpg-3-crd-10.1177_14799731211070844 for Comparison of forehead and finger oximetry sensors during the six minute walk test by Lucy Robertson, Rachel Lowry, Karl Sylvester, Helen Parfrey, Beverley Moseley, Karen Sheares and Katrina Oates in Chronic Respiratory Disease
Supplemental Material, sj-jpg-4-crd-10.1177_14799731211070844 for Comparison of forehead and finger oximetry sensors during the six minute walk test by Lucy Robertson, Rachel Lowry, Karl Sylvester, Helen Parfrey, Beverley Moseley, Karen Sheares and Katrina Oates in Chronic Respiratory Disease