Skip to main content
PLOS ONE logoLink to PLOS ONE
. 2022 Feb 3;17(2):e0263244. doi: 10.1371/journal.pone.0263244

The effectiveness of scoring balloon angioplasty in the treatment of chronic thromboembolic pulmonary hypertension

Masao Takigami 1, Hideo Tsubata 1, Naohiko Nakanishi 1,*, Yuki Matsubara 1, Noriyuki Wakana 1, Kenji Yanishi 1, Kan Zen 1, Takeshi Nakamura 1, Satoaki Matoba 1
Editor: Yoshiaki Taniyama2
PMCID: PMC8812943  PMID: 35113935

Abstract

Background

Balloon pulmonary angioplasty (BPA) is an effective treatment for inoperable chronic thromboembolic pulmonary hypertension (CTEPH). The purpose of this study is to evaluate the therapeutic effect and safety of the non-slip element percutaneous transluminal angioplasty (NSE PTA) scoring balloons in BPA.

Methods

108 pulmonary artery branches in 14 CTEPH patients who underwent BPA using NSE PTA scoring balloon (the NSE PTA group) or plain balloon (the POBA group) and pressure gradient evaluation were analyzed. We compared the improvement of the pressure ratios after BPA (Δ Pressure ratio) of both groups.

Results

There was no significant difference in the Δ Pressure ratios of the two groups (0.241 ± 0.196 POBA, 0.259 ± 0.177 NSE PTA, p = 0.63). No complications occurred in the NSE PTA group, while 3 episodes of hemoptysis were seen in the POBA group. This, however, was not found to be significant (p = 0.27). In the cases where balloon-to-vessel ratio exceeded 1.0 (n = 35), multivariate analysis showed that the use of NSE PTA scoring balloon and pressure ratio before BPA were significantly correlated with Δ Pressure ratio (β coefficient: 0.047, 95% CI: 0.0016 to 0.093, p = 0.043 and β coefficient: −0.60, 95% CI: −0.78 to −0.42, p < 0.01, respectively).

Conclusions

Although NSE PTA scoring balloon was safe, there was no significant pressure gradient improvement with NSE PTA scoring balloon compared to conventional BPA. Nevertheless, the NSE PTA scoring balloon showed effective blood-flow improvement in the case of large balloon-to-vessel ratio.

Introduction

Chronic thromboembolic pulmonary hypertension (CTEPH) is a progressive disease associated with increased pulmonary vascular resistance and pulmonary hypertension (PH) as a result of organized thromboses, clinically classified as Group 4 PH [1]. In spite of therapeutic developments, the prognosis of CTEPH remains poor. Pulmonary endarterectomy (PEA) reduces pulmonary arterial pressure and improves symptoms and prognosis in patients with surgically accessible CTEPH [25]. For inoperable patients or patients with residual pulmonary hypertension after PEA, pulmonary vasodilators such as soluble guanylate cyclase stimulators are indicated [6, 7]. In recent years, the efficacy of balloon pulmonary angioplasty (BPA) has been reported for the treatment of inoperable CTEPH. BPA has been shown to improve symptoms, exercise tolerance, right heart function, and long-term prognosis in patients with CTEPH [810].

In the pulmonary artery of CTEPH patients a mesh or slit-like organized thrombus, known as a web lesion, can form and inhibit pulmonary blood flow [11]. In BPA, a balloon expands and shifts the organized thrombus to the vascular wall in order to improve blood flow [12]. Unfortunately, over-dilatation injures of the pulmonary artery can occur and lead to parenchymal hemorrhage and hemoptysis. Currently, auxiliary devices such as intravascular ultrasound (IVUS) and pressure monitoring devices can be used to reduce such complications [1315]. At this time, the best way to maximize the therapeutic effect while limiting the complications of BPA has not yet been well established.

Currently, there are many studies showing the usefulness of the scoring balloon in the treatment of coronary artery disease [16, 17]. Non-slip element percutaneous transluminal angioplasty balloon (NSE PTA) is a scoring balloon with three nylon elements and can prevent the balloon from slipping during expansion (Fig 1). It has also been reported that hardened lesions, such as those that are calcified, can be satisfactorily expanded by applying concentrated force to the elements [18]. Additionally, in peripheral vascular treatment, the NSE PTA scoring balloon has been reported to have better dilation and less vessel dissociation [19, 20]. In BPA treatment, NSE PTA scoring balloons may be able to successfully dilate an organized thrombus and obtain improved blood flow, but there have not been any studies on its efficacy and safety thus far.

Fig 1. Non-slip element percutaneous transluminal angioplasty balloon (NSE PTA).

Fig 1

The NSE PTA scoring balloon is a non-slip peripheral angioplasty catheter with three nylon scoring elements for a controlled scoring of the vessel wall and aimed to reduce slipping during balloon inflation.

Therefore, the purpose of this study is to evaluate the therapeutic effect and safety of NSE PTA scoring balloon in CTEPH patients who underwent BPA. We found that although the NSE PTA scoring balloon was safe, there was no significant pressure gradient improvement with the NSE PTA scoring balloon versus the conventional balloon during BPA in most cases. However, the NSE PTA scoring balloon did show superior blood-flow improvement in the cases of large balloon-to-vessel ratios.

Methods

Study population and collection of clinical data

This study is a retrospective and single center study approved by The Institutional Review Board of the Kyoto Prefectural University of Medicine. From December 2017 to January 2020, 108 pulmonary artery branches in 14 consecutive patients with CTEPH who underwent BPA using NSE PTA scoring balloon or plain balloon and had pressure gradient evaluation with pressure microcatheter were analyzed. Clinical data, such as age, sex, World Health Organization functional class (WHO-FC), six-minute walking distance (6MWD), brain natriuretic peptide (BNP) and medications, were collected from the electronic medical record at the time of treatment. Hemodynamic characteristics were assessed with right heart catheterization before BPA and right atrium pressure (RAP), pulmonary artery wedge pressure (PAWP), pulmonary artery pressure (PAP) and cardiac output (CO) with thermodilution were measured. Since this study is retrospective observational study, IRB of our institute waived the requirement for informed consent. All data were anonymized for collection.

BPA procedures

BPA procedures were performed via femoral vein or jugular vein approach. 8-Fr sheath was inserted into the vein and 6-Fr ParentPlus guiding sheath (MEDIKIT, Tokyo, Japan) was advanced to the main pulmonary artery through the 8-Fr sheath using a 0.035-inch wire (Radifocus Guide Wire M; Terumo, Tokyo, Japan). We selected a branch of the pulmonary artery by a 6-Fr guiding catheter (Profit MP, JR4.0 or AL1.0; NIPRO, Osaka, Japan). Pulmonary angiography was performed manually using half contrast medium diluted with saline. A 0.014-inch guidewire (B-pahm; Japan Lifeline, Tokyo, Japan) was crossed under the pulmonary angiography, and IVUS (Eagle Eye Platinum; Volcano, San Diego, CA) was used for the detection of vessel diameter. Vessel diameter was defined as short diameter assessed by IVUS. The balloon size, the expansion pressure and the choice of NSE PTA scoring balloon or plain old balloon angioplasty (POBA) were at the discretion of the operator. Balloon-to-vessel (B/V) ratio was defined as the balloon diameter divided by vessel diameter assessed with IVUS.

Evaluation of pressure gradient of the target lesion

Distal and proximal pressure readings of the target lesion were measured before and after balloon dilatation using microcatheter with an optical pressure sensor (Navvus MicroCatheter; ACIST Medical Systems, Inc., Eden Prairie, MN). This microcatheter has been reported to show better correlation between microcatheter and pressure wire fractional flow reserve measurements [21]. This information was used to calculate the pressure ratio between the two readings. Since this pressure microcatheter was a rapid exchange type catheter, there was no need to change the guidewire and the pressure gradient could be measured while maintaining the guidewire’s position throughout the BPA procedure. The pressure microcatheter on the guidewire was advanced to the tip of guiding catheter where the 2 pressures, a tip of guiding catheter and pressure microcatheter, were equalized after saline flush. Then the pressure microcatheter was advanced and positioned distal to the target lesion. The pressure ratio was defined as the distal pressure of the pressure microcatheter (Pd) divided by the proximal pressure of guiding catheter (Pa) which was recorded before and after BPA (Fig 2). Δ Pressure ratio was defined as the difference between the pressure ratio before and after BPA. Finally, the ratio of pressure ratios was defined as the ratio between the before BPA pressure ratio and the after BPA pressure ratio.

Fig 2. Assessment of pressure ratio by pressure catheter.

Fig 2

(A) Measuring of blood pressure. Pa was measured by guiding catheter and Pd was measured by pressure microcatheter. (B) Angiography and pressure ratio before and after balloon dilatation. Pressure ratio was defined as Pd / Pa.

Statistical analysis

We evaluated whether each valuable was normally distributed. Continuous data were expressed as the mean ± standard deviation (SD). Data that were not normally distributed were presented by median (interquartile range). We compared the baseline characteristics, pressure ratio and Δ Pressure ratio at the target lesion between the NSE PTA group and the POBA group before and after BPA procedure using mixed models for repeated measures. Categorical variables were compared using the chi-square test. Valuables that were not normally distributed were analyzed using non-parametric test. Correlations between pressure ratio before and after BPA in the POBA group and the NSE group were evaluated respectively. We performed univariate analysis using regression analysis with mixed models for repeated measures to find the relationship of Δ Pressure ratio in whole patients and patients with B/V ratio over 1.0 according to Soga’s report [22]. Multivariable analysis, adjusting for balloon type, B/V ratio and the pressure ratio before BPA, was performed using multiple linear regression. P-value of < 0.05 was considered statistically significant. The statistical analyses were performed using JMP software version 11.0.0 (SAS Institute Inc., Cary, NC, USA).

Results

Patients and procedure characteristics

Table 1 shows the baseline population characteristics of the 14 patients. Mean age was 67.9 ± 15.9 years-old and most of patients were female (93%). Six-minute walking distance (6MWD) was 298.7 ± 100 m. BNP level was 225.3 pg/mL (73.1–495.6 pg/mL) and most of patients had WHO-FC II or III. Eleven patients (79%) were treated by soluble guanylate cyclase stimulator. Hemodynamics parameters showed that the mean pulmonary artery pressure (mPAP) was 39.1 ± 10.3 mmHg, the cardiac output (CO) was 3.6 ± 1.0 L/min and the resulting pulmonary vascular resistance (PVR) was 9.0 ± 4.5 Wood Unit. Fluid balance was well controlled; mean right atrium pressure (RAP) was 8.0 ± 4.0 mmHg, and pulmonary artery wedge pressure (PAWP) was 9.4 ± 3.5 mmHg.

Table 1. Baseline population characteristics.

All patients
(n = 14)
Age, years 67.9 ± 15.9
Female, n (%) 13 (93)
WHO-FC I/II/III/IV, n 0/5/7/2
6MWD, m 298.7 ± 100.0
BNP, pg/mL 225.3 (73.1–495.6)
Medication, n (%)
 sGC stimulator 11 (79)
 Diuretics 8 (57)
Hemodynamics
 RAP, mmHg 8.0 ± 4.0
 PAWP, mmHg 9.4 ± 3.5
 sPAP, mmHg 67.6 ± 18.2
 dPAP, mmHg 24.4 ± 7.5
 mPAP, mmHg 39.1 ± 10.3
 CO, L/min 3.6 ± 1.0
 CI, L/min/m2 2.5 ± 0.7
 PVR, WU 9.0 ± 4.5

Data are presented as n (%), mean ± SD or median (interquartile ranges). WHO-FC, World Health Organization functional class; 6MWD, 6-minute walk distance; BNP, brain natriuretic peptide; sGC, soluble guanylate cyclase; RAP, right atrial pressure; PAWP, pulmonary artery wedge pressure; sPAP, systolic pulmonary artery pressure; dPAP, diastolic pulmonary artery pressure; mPAP, mean pulmonary artery pressure; CO, cardiac output; CI, cardiac index; PVR, pulmonary vascular resistance.

Procedure characteristics involving the 108 lesions are presented in Table 2. The plain balloon was used in 65 cases (POBA group) and the NSE PTA scoring balloon was used in 43 cases (NSE PTA group). Treated branches and lesion types were well matched between both groups. In the majority of cases, we selected the right lower lobe due to easy access. Most lesions were web lesions (74% in the POBA group, 63% in the NSE PTA group, respectively). Although vessel diameter and balloon size tended to be smaller in the NSE PTA group, there did not appear to be statistically significant differences. Balloon pressure was higher in the NSE PTA group (8.0 (6.0–8.0) atm in the POBA group, and 8.0 (8.0–8.0) atm in the NSE PTA group, p = 0.01). Overall, balloon to vessel (B/V) ratio was similar between two groups (0.93 ± 0.13 in the POBA group and 0.93 ± 0.15 in the NSE PTA group, p = 0.92).

Table 2. Procedure characteristics.

POBA
(n = 65)
NSE PTA
(n = 43)
P value
Treated site 0.56
 Right PA
  Upper lobe, n (%) 8 (12) 10 (23)
  Middle lobe, n (%) 10 (15) 10 (23)
  Lower lobe, n (%) 28 (43) 13 (30)
 Left PA
  Upper lobe, n (%) 2 (4) 1 (2)
  Lingular, n (%) 5 (8) 3 (7)
  Lower lobe, n (%) 12 (18) 6 (14)
Lesion type 0.40
 Ring-like, n (%) 10 (15) 10 (23)
 Web, n (%) 48 (74) 27 (63)
 Subtotal, n (%) 5 (8) 6 (14)
 Total occlusion, n (%) 2 (3) 0
 Tortuous, n (%) 0 0
Vessel diameter, mm 4.49±1.23 4.09±1.03 0.10
Balloon
 Size, mm 4.0 (3.0–5.0) 4.0 (3.0–4.0) 0.07
 Pressure, atm 8.0 (6.0–8.0) 8.0 (8.0–8.0) 0.01
Balloon-to-Vessel (B/V) ratio 0.93±0.13 0.93±0.15 0.92

Data are presented as n (%), mean ± SD or median (interquartile ranges). PA, pulmonary artery; POBA, plain old balloon angioplasty; NSE PTA, non-slip element percutaneous transluminal angioplasty.

Pressure ratio of target lesion before and after BPA

Fig 3 shows the relationship between pressure ratios before and after BPA in the POBA group and the NSE PTA group. Both groups had a positive correlation between the pressure ratio before BPA and the pressure ratio after BPA (r = 0.68 in the POBA group; p < 0.01, r = 0.70 in the NSE group; p < 0.01). In addition, a negative correlation was shown between the pressure ratio before BPA and the degree of improvement in pressure gradient ratio (Δ Pressure ratio) (r = −0.80 in the POBA group; P < 0.01, r = −0.54 in the NSE group; p < 0.01); however, there was no significant differences between two groups (p = 0.26). These results suggest that severe lesions have a larger pressure gain after BPA procedure.

Fig 3. The relationship between pressure ratio before and after balloon pulmonary angioplasty (BPA).

Fig 3

Pressure ratio before BPA and pressure ratio after BPA plot (A), pressure ratio before BPA and Δ Pressure ratio plot (B).

Table 3 shows a comparison of distal to proximal pressure ratios before and after BPA. Although pressure ratio before BPA tended to be lower in the NSE PTA group, there was not statistically significant difference (0.526 ± 0.267 in the POBA group, 0.415 ± 0.243 in the NSE PTA group, p = 0.08). There was a significant difference in pressure ratio after BPA (0.766 ± 0.162 in the POBA group, 0.675 ± 0.208 in the NSE PTA group, p = 0.03). Despite this finding, there was no significant difference in the Δ Pressure ratio between the two groups (0.241 ± 0.196 in the POBA group, 0.259 ± 0.177 in the NSE PTA group, p = 0.63) (Fig 4). Because of the negative correlation between the pressure ratio before BPA and the Δ Pressure ratio, we investigated the degree of improvement in the pressure gradient ratio based on the pressure ratio before BPA. The ratio of pressure ratios also had no significant difference between the two groups (1.29 (1.12–2.06) in the POBA group, 1.51 (1.23–2.24) in the NSE PTA group, p = 0.95).

Table 3. Pressure ratio of target lesion before and after BPA.

POBA
(n = 65)
NSE PTA
(n = 43)
P value
Pressure ratio before BPA 0.526 ± 0.267 0.415 ± 0.243 0.08
Pressure ratio after BPA 0.766 ± 0.162 0.675 ± 0.208 0.03
Δ Pressure ratio 0.241 ± 0.196 0.259 ± 0.177 0.63
Ratio of pressure ratio 1.29 (1.12–2.06) 1.51 (1.23–2.24) 0.95

Data are presented as mean ± SD, or median (interquartile ranges). BPA, balloon pulmonary angioplasty; POBA, plain old balloon angioplasty; NSE PTA, non-slip element percutaneous transluminal angioplasty; Δ Pressure ratio, pressure ratio after BPA—pressure ratio before BPA; Ratio of pressure ratio, ratio of pressure ratio after BPA-to-pressure ratio before BPA.

Fig 4. Δ Pressure ratio in the plain old balloon angioplasty (POBA) group and the non-slip element (NSE) group.

Fig 4

The data are given as the mean ± SD.

Predictors of Δ Pressure ratio

We analyzed the predictive factors regarding the improvement of Δ Pressure ratio (Table 4). In univariate analysis, only pressure ratio before BPA demonstrated significant relationships with Δ Pressure ratio. Because the groups were not well matched and pressure gradient before BPA was significantly different at baseline, we explored the usefulness of NSE PTA scoring balloon by multivariate analysis. In multivariate analysis using the multiple linear regression model, only pressure ratio before BPA was associated with Δ Pressure ratio (β coefficient: −0.55, 95%CI: −0.66 to −0.44, P < 0.01). In total, the NSE PTA scoring balloon did not have a significant pressure gradient improvement effect when compared with conventional plain balloon in BPA procedure.

Table 4. Univariate and multivariate analysis of Δ Pressure ratio.

Univariate Multivariate
β 95% CI P value β 95% CI P value
Balloon type, NSE -0.016 −0.046 to 0.013 0.18 -0.018 −0.046 to 0.0101 0.20
Vessel diameter, mm −0.077 −0.17 to 0.016 0.063
Balloon Size, mm 0.082 −0.023 to 0.19 0.10
B/V ratio −0.19 −0.65 to 0.28 0.37 0.16 −0.0022 to 0.349 0.084
Web lesion 0.022 −0.0046 to 0.048 0.08
Pressure ratio before BPA −0.57 −0.68 to −0.45 <0.01 −0.55 −0.66 to −0.44 <0.01

Δ Pressure ratio, pressure ratio after BPA—pressure ratio before BPA; NSE PTA, non-slip element percutaneous transluminal angioplasty; POBA, plain old balloon angioplasty; BPA, balloon pulmonary angioplasty; B/V ration, Balloon-to-Vessel ratio.

Safety of the NSE PTA scoring balloon in BPA procedure

Complications of BPA are presented in Table 5. No complication occurred in the NSE PTA group while 3 hemoptysis episodes were seen in the POBA group due to wire injury, however, there was no significant difference between the two groups (p = 0.27). Additionally, other complications including dissection, perforation, use of noninvasive positive pressure ventilation, need for intubation and death were not seen in either group. At least, the NSE PTA scoring balloon did not increase complications compared to the conventional balloon in BPA.

Table 5. Incidence of complications.

POBA
(n = 65)
NSE PTA
(n = 43)
P value
Haemoptysis, n (%) 3 (5) 0 0.27
Dissection, n (%) 0 0 1.00
Perforation, n (%) 0 0 1.00
NPPV, n (%) 0 0 1.00
Intubation, n (%) 0 0 1.00
Death, n (%) 0 0 1.00

Data are presented as n (%). POBA, plain old balloon angioplasty; NSE PTA, non-slip element percutaneous transluminal angioplasty; NPPV, noninvasive positive pressure ventilation.

Subgroup analysis of Δ Pressure ratio when B/V ratio >1.0

We explored the effectiveness of NSE PTA scoring balloon in the cases where B/V ratio exceeds 1.0 (n = 35) because it has been reported that, in this situation, the effect of the scoring balloon may be greater [22]. Univariate analysis showed a significant correlation with the Δ Pressure ratio and the pressure ratio before BPA (beta coefficient: −0.60, 95% CI: −0.80 to −0.41, p < 0.01) (Table 6). However, multivariate analysis showed that the use of the NSE PTA scoring balloon and the pressure ratio before BPA were significantly correlated with Δ Pressure ratio (β coefficient: 0.047, 95% CI: 0.0016 to 0.093, p = 0.043 and β coefficient: -0.60, 95% CI: −0.78 to −0.42, p < 0.01, respectively). Therefore, the NSE PTA scoring balloon shows superior improvement in blood-flow after BPA over the conventional balloon when the B/V ratio is greater than 1.0.

Table 6. Univariate and multivariable analysis of Δ Pressure ratio in B/V ratio >1.0 (n = 35).

Univariate Multivariate
β 95% CI P value β 95% CI P value
Balloon type, NSE 0.057 −0.0063 to 0.12 0.07 0.047 0.0016 to 0.093 0.043
Vessel diameter, mm 0.23 −0.36 to 0.83 0.43
Balloon Size, mm −0.21 −0.77 to 0.34 0.44
B/V ratio 0.523 −1.15 to 2.20 0.53 −0.088 −0.5 to 0.28 0.57
Web lesion −0.0001 −0.047 to 0.047 1.0
Pressure ratio before BPA −0.60 −0.80 to -0.41 <0.01 −0.60 −0.78 to −0.42 <0.01

Δ Pressure ratio, pressure ratio after BPA—pressure ratio before BPA; NSE PTA, non-slip element percutaneous transluminal angioplasty; POBA, plain old balloon angioplasty; BPA, balloon pulmonary angioplasty; B/V ration, Balloon-to-Vessel ratio.

Discussion

Our findings in this study were as follows: There was no significant difference in Δ Pressure ratio between the POBA group and the NSE PTA group. NSE PTA scoring balloon was safe to use in BPA. In cases where the B/V ratio was higher than 1.0, the Δ Pressure ratio was significantly higher in the NSE PTA group.

CTEPH is known as a progressive disease with a poor prognosis, and with the recent development of BPA, it has been reported that BPA improves pulmonary hypertension, prevents the development of right heart failure, and improves the prognosis and quality of life of the patient [12, 23, 24]. However, there are still few reports regarding the best strategy for BPA in this population, including the selection of balloons. The NSE PTA scoring balloon is a non-slip peripheral angioplasty catheter with three nylon scoring elements which produce controlled scoring of the vessel wall and work to reduce slipping during balloon inflation. The triangular cross section of the scoring elements provides a higher and more concentrated pressure transfer to the vessel wall, which contributes to the reduction in balloon slipping during expansion. In percutaneous coronary intervention (PCI), the ELEGANT study reported that acute gain was significantly higher than that of normal balloons in in-stent restenosis of coronary arteries [17]. In addition, Soga et al. reported the usefulness of NSE PTA in experimental model was enormously observed in the group with B/V ratio more than 1.0 [22]. In this study, there was no significant difference in Δ Pressure ratio between the NSE PTA group and the POBA group, but the Δ Pressure ratio increased significantly in the NSE PTA group when the B/V ratio was greater than 1.0. This suggests that expansion with an NSE balloon adjusted to the blood vessel diameter is more effective in applying concentrated force due to the non-slip element, which results in an adequate pressure gradient improvement. However, high pulmonary pressure and high perfusion pressure were a risk of reperfusion pulmonary injury [14, 23]. Recently, BPA with undersized balloons and repeated sessions are recommended to get enough improvement with reducing complications [25]. Therefore, NSE PTA may be suitable for consolidation treatment with sufficient low pulmonary artery pressure after several BPA treatments rather than the initial dilation of complete occlusion or severe stenotic lesions with high pulmonary artery pressure. In addition, NSE PTA is more expensive compared with conventional plain balloon. Therefore, we do not recommend routine use of NSE PTA scoring balloon in BPA procedures, and should be used in the case that we want to dilate pulmonary artery intensively in the last session.

In particular, venous blood vessels are highly compliant, and it is important to select a balloon that is suitable for the diameter of the blood vessel. Initially it was expected that NSE PTA would be more effective for lotus root-like lesions, called web lesions, because the non-slipping element seemed to be forced into the filamentous organized thrombus, but this study did not show a significant difference from conventional balloons whether lesion was a web lesion or not. Because of a small sample size, we could not analyze the treatment effect among the lesion type in detail. Further research will be needed to determine the efficacy of the NSE PTA scoring balloon according to thrombotic lesion type.

Regarding safety concerns, a multicenter registry of BPA in Japan found that complications occurred in 36.3% of patients [9]. In the cases of our study, no procedural complications, such as hemoptysis, vessel dissection or perforation, were observed in the NSE PTA group. Scoring balloon has been reported to reduce the incidence of severe dissection in superficial femoral artery angioplasty [26]. Since the cross section of this element is wedge-shaped, high stress is efficiently concentrated to the vessel wall. This may prevent severe dissection that reaches the media. In addition, because the NSE PTA scoring balloon has three nylon elements, it seems to have less penetrating force into the vessel wall than compared to the cutting balloon with metallic blades. Venous tissue, such as the pulmonary artery, is more fragile than arterial tissue, therefore, NSE PTA scoring balloon might effectively apply pressure and could prevent vascular injury.

There were several limitations in this study. First, this study was conducted retrospectively at a single center with a limited sample size, not a randomized trial. There were several biases including patient and procedure selection. Therefore, both groups are not well matched at baseline. Second, the balloon size, the expansion pressure and the choice of NSE PTA scoring balloon or POBA were at the discretion of the operator. A multicenter and randomized trial with a larger number of patients is needed to validate our findings. Third, since BPA is often performed on many lesions during one session, it is not possible to compare major adverse cardiovascular events and the improvement of mean pulmonary artery pressure by balloon selection alone. In addition, we evaluated only immediately after BPA. It has been reported that treated pulmonary arteries may take time to dilate sufficiently. We need to investigate follow-up data to conclude the effectiveness of scoring balloon in BPA procedures. Finally, the pressure gradient depends on peripheral vascular resistance caused by vasculopathy. We could not put the presence of vasculopathy in the analysis as a confounder.

Conclusion

NSE PTA scoring balloon did not show a significant improvement in pressure gradient compared to conventional balloon. Nevertheless, NSE PTA scoring balloon showed effective blood-flow improvement in the case of large balloon-to-vessel ratio. NSE PTA scoring balloon was safe to use in BPA procedure. Our results suggest that the use of the NSE PTA scoring balloon should be considered as one of the treatment options for BPA, especially in the case that we want to dilate pulmonary artery intensively in the last session.

Acknowledgments

We would like to thank Dr. Sean Delue for proofreading in English. We gratefully acknowledge the work of past and present members of our hospital.

Data Availability

All relevant data are within the paper.

Funding Statement

The authors received no specific funding for this work.

References

  • 1.Kim NH, Delcroix M, Jais X, Madani MM, Matsubara H, Mayer E, et al. Chronic thromboembolic pulmonary hypertension. Eur Respir J. 2019;53(1). doi: 10.1183/13993003.01915-2018 . [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 2.Bonderman D, Skoro-Sajer N, Jakowitsch J, Adlbrecht C, Dunkler D, Taghavi S, et al. Predictors of outcome in chronic thromboembolic pulmonary hypertension. Circulation. 2007;115(16):2153–8. doi: 10.1161/CIRCULATIONAHA.106.661041 . [DOI] [PubMed] [Google Scholar]
  • 3.Condliffe R, Kiely DG, Gibbs JS, Corris PA, Peacock AJ, Jenkins DP, et al. Improved outcomes in medically and surgically treated chronic thromboembolic pulmonary hypertension. Am J Respir Crit Care Med. 2008;177(10):1122–7. doi: 10.1164/rccm.200712-1841OC . [DOI] [PubMed] [Google Scholar]
  • 4.Freed DH, Thomson BM, Berman M, Tsui SS, Dunning J, Sheares KK, et al. Survival after pulmonary thromboendarterectomy: effect of residual pulmonary hypertension. J Thorac Cardiovasc Surg. 2011;141(2):383–7. doi: 10.1016/j.jtcvs.2009.12.056 . [DOI] [PubMed] [Google Scholar]
  • 5.Jamieson SW, Kapelanski DP, Sakakibara N, Manecke GR, Thistlethwaite PA, Kerr KM, et al. Pulmonary endarterectomy: experience and lessons learned in 1,500 cases. Ann Thorac Surg. 2003;76(5):1457–62; discussion 62–4. doi: 10.1016/s0003-4975(03)00828-2 . [DOI] [PubMed] [Google Scholar]
  • 6.Simonneau G, D’Armini AM, Ghofrani HA, Grimminger F, Hoeper MM, Jansa P, et al. Riociguat for the treatment of chronic thromboembolic pulmonary hypertension: a long-term extension study (CHEST-2). Eur Respir J. 2015;45(5):1293–302. doi: 10.1183/09031936.00087114 . [DOI] [PubMed] [Google Scholar]
  • 7.Ghofrani HA, D’Armini AM, Grimminger F, Hoeper MM, Jansa P, Kim NH, et al. Riociguat for the treatment of chronic thromboembolic pulmonary hypertension. N Engl J Med. 2013;369(4):319–29. doi: 10.1056/NEJMoa1209657 . [DOI] [PubMed] [Google Scholar]
  • 8.Fukui S, Ogo T, Morita Y, Tsuji A, Tateishi E, Ozaki K, et al. Right ventricular reverse remodelling after balloon pulmonary angioplasty. Eur Respir J. 2014;43(5):1394–402. doi: 10.1183/09031936.00012914 . [DOI] [PubMed] [Google Scholar]
  • 9.Ogawa A, Satoh T, Fukuda T, Sugimura K, Fukumoto Y, Emoto N, et al. Balloon Pulmonary Angioplasty for Chronic Thromboembolic Pulmonary Hypertension: Results of a Multicenter Registry. Circ Cardiovasc Qual Outcomes. 2017;10(11). doi: 10.1161/CIRCOUTCOMES.117.004029 . [DOI] [PubMed] [Google Scholar]
  • 10.Taniguchi Y, Matsuoka Y, Onishi H, Yanaka K, Emoto N, Nakai H, et al. The role of balloon pulmonary angioplasty and pulmonary endarterectomy: Is chronic thromboembolic pulmonary hypertension still a life-threatening disease? Int J Cardiol. 2021;326:170–7. doi: 10.1016/j.ijcard.2020.10.026 . [DOI] [PubMed] [Google Scholar]
  • 11.Nakanishi N, Fukai K, Tsubata H, Ogata T, Zen K, Nakamura T, et al. Angioscopic Evaluation During Balloon Pulmonary Angioplasty in Chronic Thromboembolic Pulmonary Hypertension. Heart Lung Circ. 2019;28(4):655–9. doi: 10.1016/j.hlc.2018.08.008 . [DOI] [PubMed] [Google Scholar]
  • 12.Sugimura K, Fukumoto Y, Satoh K, Nochioka K, Miura Y, Aoki T, et al. Percutaneous transluminal pulmonary angioplasty markedly improves pulmonary hemodynamics and long-term prognosis in patients with chronic thromboembolic pulmonary hypertension. Circ J. 2012;76(2):485–8. doi: 10.1253/circj.cj-11-1217 . [DOI] [PubMed] [Google Scholar]
  • 13.Shimokawahara H, Ogawa A, Mizoguchi H, Yagi H, Ikemiyagi H, Matsubara H. Vessel Stretching Is a Cause of Lumen Enlargement Immediately After Balloon Pulmonary Angioplasty: Intravascular Ultrasound Analysis in Patients With Chronic Thromboembolic Pulmonary Hypertension. Circ Cardiovasc Interv. 2018;11(4):e006010. doi: 10.1161/CIRCINTERVENTIONS.117.006010 . [DOI] [PubMed] [Google Scholar]
  • 14.Kinutani H, Shinke T, Nakayama K, Taniguchi Y, Otake H, Takaya T, et al. High perfusion pressure as a predictor of reperfusion pulmonary injury after balloon pulmonary angioplasty for chronic thromboembolic pulmonary hypertension. Int J Cardiol Heart Vasc. 2016;11:1–6. doi: 10.1016/j.ijcha.2015.11.006 . [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 15.Inami T, Kataoka M, Shimura N, Ishiguro H, Yanagisawa R, Fukuda K, et al. Pressure-wire-guided percutaneous transluminal pulmonary angioplasty: a breakthrough in catheter-interventional therapy for chronic thromboembolic pulmonary hypertension. JACC Cardiovasc Interv. 2014;7(11):1297–306. doi: 10.1016/j.jcin.2014.06.010 . [DOI] [PubMed] [Google Scholar]
  • 16.Matsukawa R, Kozai T, Tokutome M, Nakashima R, Nishimura R, Matsumoto S, et al. Plaque modification using a cutting balloon is more effective for stenting of heavily calcified lesion than other scoring balloons. Cardiovasc Interv Ther. 2019;34(4):325–34. doi: 10.1007/s12928-019-00578-w . [DOI] [PubMed] [Google Scholar]
  • 17.Aoki J, Nakazawa G, Ando K, Nakamura S, Tobaru T, Sakurada M, et al. Effect of combination of non-slip element balloon and drug-coating balloon for in-stent restenosis lesions (ELEGANT study). J Cardiol. 2019;74(5):436–42. doi: 10.1016/j.jjcc.2019.04.011 . [DOI] [PubMed] [Google Scholar]
  • 18.Sugawara Y, Ueda T, Soeda T, Watanabe M, Okura H, Saito Y. Plaque modification of severely calcified coronary lesions by scoring balloon angioplasty using Lacrosse non-slip element: insights from an optical coherence tomography evaluation. Cardiovasc Interv Ther. 2019;34(3):242–8. doi: 10.1007/s12928-018-0553-6 . [DOI] [PubMed] [Google Scholar]
  • 19.Karashima E, Yoda S, Yasuda S, Kajiyama S, Ito H, Kaneko T. Usefulness of the "Non-Slip Element" Percutaneous Transluminal Angioplasty Balloon in the Treatment of Femoropopliteal Arterial Lesions. J Endovasc Ther. 2020;27(1):102–8. doi: 10.1177/1526602819887954 . [DOI] [PubMed] [Google Scholar]
  • 20.Horie K, Tanaka A, Taguri M, Inoue N. Impact of Scoring Balloons on Percutaneous Transluminal Angioplasty Outcomes in Femoropopliteal Lesions. J Endovasc Ther. 2020;27(3):481–91. doi: 10.1177/1526602820914618 . [DOI] [PubMed] [Google Scholar]
  • 21.Fearon WF, Chambers JW, Seto AH, Sarembock IJ, Raveendran G, Sakarovitch C, et al. ACIST-FFR Study (Assessment of Catheter-Based Interrogation and Standard Techniques for Fractional Flow Reserve Measurement). Circ Cardiovasc Interv. 2017;10(12). doi: 10.1161/CIRCINTERVENTIONS.117.005905 . [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 22.Soga Y, Ando K. Effect of an NSE PTA balloon in experimental lesion models. Cardiovasc Interv Ther. 2018;33(1):35–9. doi: 10.1007/s12928-016-0430-0 . [DOI] [PubMed] [Google Scholar]
  • 23.Feinstein JA, Goldhaber SZ, Lock JE, Ferndandes SM, Landzberg MJ. Balloon pulmonary angioplasty for treatment of chronic thromboembolic pulmonary hypertension. Circulation. 2001;103(1):10–3. doi: 10.1161/01.cir.103.1.10 . [DOI] [PubMed] [Google Scholar]
  • 24.Mizoguchi H, Ogawa A, Munemasa M, Mikouchi H, Ito H, Matsubara H. Refined balloon pulmonary angioplasty for inoperable patients with chronic thromboembolic pulmonary hypertension. Circ Cardiovasc Interv. 2012;5(6):748–55. doi: 10.1161/CIRCINTERVENTIONS.112.971077 . [DOI] [PubMed] [Google Scholar]
  • 25.Lang I, Meyer BC, Ogo T, Matsubara H, Kurzyna M, Ghofrani HA, et al. Balloon pulmonary angioplasty in chronic thromboembolic pulmonary hypertension. European respiratory review: an official journal of the European Respiratory Society. 2017;26(143). Epub 2017/03/31. doi: 10.1183/16000617.0119-2016 . [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 26.Hong H, Park UJ, Roh YN, Kim HT. Scoring Balloon Reduces the Severity of Dissection and Stent Implantation Rate in Superficial Femoral Artery Angioplasty Compared to Plain Balloon. Vasc Endovascular Surg. 2021;55(2):135–42. doi: 10.1177/1538574420968977 . [DOI] [PubMed] [Google Scholar]

Decision Letter 0

Yoshiaki Taniyama

10 Dec 2021

PONE-D-21-25699The effectiveness of scoring balloon angioplasty in the treatment of chronic thromboembolic pulmonary hypertensionPLOS ONE

Dear Dr. Nakanishi,

Thank you for submitting your manuscript to PLOS ONE. After careful consideration, we feel that it has merit but does not fully meet PLOS ONE’s publication criteria as it currently stands. Therefore, we invite you to submit a revised version of the manuscript that addresses the points raised during the review process.

Both of the reviewers regard this experiment important, so please revise the manuscript.

Please submit your revised manuscript by Jan 24 2022 11:59PM. If you will need more time than this to complete your revisions, please reply to this message or contact the journal office at plosone@plos.org. When you're ready to submit your revision, log on to https://www.editorialmanager.com/pone/ and select the 'Submissions Needing Revision' folder to locate your manuscript file.

Please include the following items when submitting your revised manuscript:

  • A rebuttal letter that responds to each point raised by the academic editor and reviewer(s). You should upload this letter as a separate file labeled 'Response to Reviewers'.

  • A marked-up copy of your manuscript that highlights changes made to the original version. You should upload this as a separate file labeled 'Revised Manuscript with Track Changes'.

  • An unmarked version of your revised paper without tracked changes. You should upload this as a separate file labeled 'Manuscript'.

If you would like to make changes to your financial disclosure, please include your updated statement in your cover letter. Guidelines for resubmitting your figure files are available below the reviewer comments at the end of this letter.

If applicable, we recommend that you deposit your laboratory protocols in protocols.io to enhance the reproducibility of your results. Protocols.io assigns your protocol its own identifier (DOI) so that it can be cited independently in the future. For instructions see: https://journals.plos.org/plosone/s/submission-guidelines#loc-laboratory-protocols. Additionally, PLOS ONE offers an option for publishing peer-reviewed Lab Protocol articles, which describe protocols hosted on protocols.io. Read more information on sharing protocols at https://plos.org/protocols?utm_medium=editorial-email&utm_source=authorletters&utm_campaign=protocols.

We look forward to receiving your revised manuscript.

Kind regards,

Yoshiaki Taniyama, MD, PhD

Academic Editor

PLOS ONE

Journal Requirements:

When submitting your revision, we need you to address these additional requirements.

1. Please ensure that your manuscript meets PLOS ONE's style requirements, including those for file naming. The PLOS ONE style templates can be found at

https://journals.plos.org/plosone/s/file?id=wjVg/PLOSOne_formatting_sample_main_body.pdf and https://journals.plos.org/plosone/s/file?id=ba62/PLOSOne_formatting_sample_title_authors_affiliations.pdf

2. In your ethics statement in the manuscript and in the online submission form, please ensure that you have discussed whether all data/samples were fully anonymized before you accessed them and/or whether the IRB or ethics committee waived the requirement for informed consent. If patients provided informed written consent to have data/samples from their medical records used in research, please include this information.

3.  In your Data Availability statement, you have not specified where the minimal data set underlying the results described in your manuscript can be found. PLOS defines a study's minimal data set as the underlying data used to reach the conclusions drawn in the manuscript and any additional data required to replicate the reported study findings in their entirety. All PLOS journals require that the minimal data set be made fully available. For more information about our data policy, please see http://journals.plos.org/plosone/s/data-availability.

"Upon re-submitting your revised manuscript, please upload your study’s minimal underlying data set as either Supporting Information files or to a stable, public repository and include the relevant URLs, DOIs, or accession numbers within your revised cover letter. For a list of acceptable repositories, please see http://journals.plos.org/plosone/s/data-availability#loc-recommended-repositories. Any potentially identifying patient information must be fully anonymized.

Important: If there are ethical or legal restrictions to sharing your data publicly, please explain these restrictions in detail. Please see our guidelines for more information on what we consider unacceptable restrictions to publicly sharing data: http://journals.plos.org/plosone/s/data-availability#loc-unacceptable-data-access-restrictions. Note that it is not acceptable for the authors to be the sole named individuals responsible for ensuring data access.

We will update your Data Availability statement to reflect the information you provide in your cover letter.

[Note: HTML markup is below. Please do not edit.]

Reviewers' comments:

Reviewer's Responses to Questions

Comments to the Author

1. Is the manuscript technically sound, and do the data support the conclusions?

The manuscript must describe a technically sound piece of scientific research with data that supports the conclusions. Experiments must have been conducted rigorously, with appropriate controls, replication, and sample sizes. The conclusions must be drawn appropriately based on the data presented.

Reviewer #1: Yes

Reviewer #2: No

**********

2. Has the statistical analysis been performed appropriately and rigorously?

Reviewer #1: Yes

Reviewer #2: No

**********

3. Have the authors made all data underlying the findings in their manuscript fully available?

The PLOS Data policy requires authors to make all data underlying the findings described in their manuscript fully available without restriction, with rare exception (please refer to the Data Availability Statement in the manuscript PDF file). The data should be provided as part of the manuscript or its supporting information, or deposited to a public repository. For example, in addition to summary statistics, the data points behind means, medians and variance measures should be available. If there are restrictions on publicly sharing data—e.g. participant privacy or use of data from a third party—those must be specified.

Reviewer #1: Yes

Reviewer #2: No

**********

4. Is the manuscript presented in an intelligible fashion and written in standard English?

PLOS ONE does not copyedit accepted manuscripts, so the language in submitted articles must be clear, correct, and unambiguous. Any typographical or grammatical errors should be corrected at revision, so please note any specific errors here.

Reviewer #1: Yes

Reviewer #2: Yes

**********

5. Review Comments to the Author

Please use the space provided to explain your answers to the questions above. You may also include additional comments for the author, including concerns about dual publication, research ethics, or publication ethics. (Please upload your review as an attachment if it exceeds 20,000 characters)

Reviewer #1: Plain old balloon angioplasty (POBA) group vs the non-slip element (NSE) manuscript review.

Thank you so very much for the opportunity to review this interesting manuscript.

I am in agreement that we need to explore options for best delivery of balloon angioplasty to our chronic thromboembolic pulmonary hypertension patients.

I found this very interesting and helpful.

I do agree that additional study on a larger scale would be very helpful.

Your description of methods and statistics for sound and I do not recommend any significant change.

Comments regarding cost and implications regarding that may be helpful to the reader. That may be something to consider to add regarding catheter choice.

I agree with your comments and findings regarding effective blood flow improvement in the case of large balloon two-vessel ratio is something that needs to be explored further in the future as catheter selection may be based on pre-BPA assessment and does require further study to determine if this would improve outcomes.

I do recommend to review your conclusions as I think expanding this could be helpful to the reader.

Conclusion

2 NSE PTA scoring balloon did not show a significant improvement in pressure

3 gradient compared to conventional balloon. Nevertheless, NSE PTA scoring

4 balloon showed effective blood-flow improvement in the case of large balloon-to-

5 vessel ratio. NSE PTA scoring balloon was safe to use in BPA procedure. Our

6 results suggest that the use of the NSE PTA scoring balloon should be considered

7 as one of the treatment options for BPA.

Additional comments....................

When reading I did find a few typos noted below:

Page 10

6 pressure microcatheter on the guidewire was advanced to the tip of guiding

7 catheter where the 2 pressures, a tip of guiding catheter and pressure

8 microcatheter, were equalized after saline flash.

**FLUSH

Page 10

Δ Pressure ratio was defined as the difference between the pressure

13 ratio before and after BPA. Finally, the ratio of pressure ratios was defined as the

14 ratio between the before BPA pressure ratio and the after BPA pressure ration.

**RATIO

Question:

Page 13

11 was better controlled; mean right atrium pressure (RAP) was 8.0 ± 4.0 mmHg,

12 and pulmonary artery wedge pressure (PAWP) was 9.4 ± 3.5 mmHg.

**THIS NOTES BETTER CONTROLLED but likely should just be controlled as not comparing

Page 21

11 95% CI: -0.76 to -0.43, p < 0.01, respectively). Therefore, the NSE PTA scoring

12 balloon shows superior improvement in blood-flow after BPA over the

13 conventional balloon when the B/V ration is greater than 1.0.

**RATIO

Page 24

1 In addition, Soga et al. reported the usefulness of NSE PTA in experimental model

2 was enormously observed in the group with B/V ratio more than 1.0 [22].

**Would replace ENORMOUSLY here.

Page 25

Since the cross section of this element is wedge-shaped, high stress is

13 efficiently concentrated to the vessel wall. This may prevent severe dissociation

14 that reaches the media.

**DISSECTION

Page 26

1 There were several limitations in this study. First, this study was

2 conducted retrospectively at a single center with a limited sample size, not

3 randomized trial.

**A RANDOMONIZED TRIAL

Page 26

. In addition, we evaluated

11 only immediately after BPA. It has been reported that treated pulmonary arteries

12 are took time to dilate sufficiently.

**MAY TAKE TIME

Page 26

We need to investigate follow-up data to

13 conclude the effectiveness of scoring balloon in BPA procedure.

**PROCEDURES

Reviewer #2: PONE-D-21-25699: statistical review

SUMMARY. This is a retrospective study that compares the effect of the non-slip element percutaneous transluminal angioplasty (NSE PTA) on blood pressure ratios to the effect of plain-balloon treatments (POBA). The statistical analysis relies on t-test methods for comparing the NSE PTA and the POBA groups and on regression analysis for evaluating the effect of the use of NSE PTA scoring balloon on the pressure ratio in a subset of data where the balloon-to-vessel ratio exceeded 1.0. I have several major concerns about the statistical methods that have been employed in this paper.

MAJOR ISSUES

1. The statistical analysis focuses on the improvements of pressure ratios observed after Balloon pulmonary angioplasty (BPA) in 108 lesions, which are treated as independent observations. However, if I understood correctly, these lesions are clustered within 14 subjects and, as such, they may not be assumed as independent cases. Instead, they should be treated as repeated measures within a sample of 14 subjects. Repeated measures methods are available in SAS, which is the software used by the authors. Notice that the repeated measures structure of the data must be accounted for in both the ANOVA analysis that compares the NSE PTA and the POBA groups and in the regression analysis that focuses on the effect of the NSE PTA treatment.

2. This is a retrospective study and, as a result, subjects have not been randomly associated to the NSE PTA and the POBA groups. The authors should therefore provide evidence that these two groups do not significantly differ with respect to the biometrical variables that are summarized in Table 1. Otherwise, results could be biased by group-specific differences

3. The statistical analysis relies on the assumption that the dependent variables are normally distributed. The authors should provide some evidence that such assumption is realistic, otherwise all the p-values of the paper could be biased.

**********

6. PLOS authors have the option to publish the peer review history of their article (what does this mean?). If published, this will include your full peer review and any attached files.

If you choose “no”, your identity will remain anonymous but your review may still be made public.

Do you want your identity to be public for this peer review? For information about this choice, including consent withdrawal, please see our Privacy Policy.

Reviewer #1: No

Reviewer #2: No

[NOTE: If reviewer comments were submitted as an attachment file, they will be attached to this email and accessible via the submission site. Please log into your account, locate the manuscript record, and check for the action link "View Attachments". If this link does not appear, there are no attachment files.]

While revising your submission, please upload your figure files to the Preflight Analysis and Conversion Engine (PACE) digital diagnostic tool, https://pacev2.apexcovantage.com/. PACE helps ensure that figures meet PLOS requirements. To use PACE, you must first register as a user. Registration is free. Then, login and navigate to the UPLOAD tab, where you will find detailed instructions on how to use the tool. If you encounter any issues or have any questions when using PACE, please email PLOS at figures@plos.org. Please note that Supporting Information files do not need this step.

PLoS One. 2022 Feb 3;17(2):e0263244. doi: 10.1371/journal.pone.0263244.r002

Author response to Decision Letter 0


26 Dec 2021

Response to Reviewer comments (PONE-D-21-25699):  

We would like to thank the Editors and Reviewers for their comments on our manuscript, including their very useful suggestions for improvement. We have endeavored to respond to their comment and suggestions, and we believe that we have adequately modified the revised manuscript to address their concerns.

Responses to all queries are highlighted with Track Changes in the revised manuscript.

Reviewer #1:

Plain old balloon angioplasty (POBA) group vs the non-slip element (NSE) manuscript review.

Thank you so very much for the opportunity to review this interesting manuscript.

I am in agreement that we need to explore options for best delivery of balloon angioplasty to our chronic thromboembolic pulmonary hypertension patients. I found this very interesting and helpful. I do agree that additional study on a larger scale would be very helpful. Your description of methods and statistics for sound and I do not recommend any significant change. Comments regarding cost and implications regarding that may be helpful to the reader. That may be something to consider to add regarding catheter choice. I agree with your comments and findings regarding effective blood flow improvement in the case of large balloon two-vessel ratio is something that needs to be explored further in the future as catheter selection may be based on pre-BPA assessment and does require further study to determine if this would improve outcomes.

I do recommend to review your conclusions as I think expanding this could be helpful to the reader.

Conclusion

NSE PTA scoring balloon did not show a significant improvement in pressure gradient compared to conventional balloon. Nevertheless, NSE PTA scoring balloon showed effective blood-flow improvement in the case of large balloon-to-vessel ratio. NSE PTA scoring balloon was safe to use in BPA procedure. Our results suggest that the use of the NSE PTA scoring balloon should be considered as one of the treatment options for BPA.

Response:

We very much appreciate your valuable comments and suggestions. As you mentioned, the cost-benefit is an important issue of balloon choice. NSE PTA is more expensive compared with conventional plain balloon. Therefore, we do not recommend routine use of NSE PTA scoring balloon in BPA procedures, and should be used in the case that we want to dilate pulmonary artery intensively in the last session. According to your comments, we have revised the manuscript as follows.

(Discussion, p24, line 15-18)

In addition, NSE PTA is more expensive compared with conventional plain balloon. Therefore, we do not recommend routine use of NSE PTA scoring balloon in BPA procedures, and should be used in the case that we want to dilate pulmonary artery intensively in the last session.

(Conclusion, p28, line 5-8)

Our results suggest that the use of the NSE PTA scoring balloon should be considered as one of the treatment options for BPA, especially in the case that we want to dilate pulmonary artery intensively in the last session.

Additional comments....................

When reading I did find a few typos noted below:

Page 10

pressure microcatheter on the guidewire was advanced to the tip of guiding catheter where the 2 pressures, a tip of guiding catheter and pressure microcatheter, were equalized after saline flash.

**FLUSH

Response:

Thanks to your pointing. We corrected it as below:

(Methods, p10, line 5-8)

The pressure microcatheter on the guidewire was advanced to the tip of guiding catheter where the 2 pressures, a tip of guiding catheter and pressure microcatheter, were equalized after saline flush.

Page 10

ΔPressure ratio was defined as the difference between the pressure ratio before and after BPA. Finally, the ratio of pressure ratios was defined as the ratio between the before BPA pressure ratio and the after BPA pressure ration.

**RATIO

Response:

We greatly appreciate your comment. We corrected it as below:

(Methods, p10, line 12-14)

Δ Pressure ratio was defined as the difference between the pressure ratio before and after BPA. Finally, the ratio of pressure ratios was defined as the ratio between the before BPA pressure ratio and the after BPA pressure ratio.

Question:

Page 13

was better controlled; mean right atrium pressure (RAP) was 8.0 ± 4.0 mmHg, and pulmonary artery wedge pressure (PAWP) was 9.4 ± 3.5 mmHg.

**THIS NOTES BETTER CONTROLLED but likely should just be controlled as not comparing

Response:

We thank to your pointing. As you mentioned, these are not compared data and we corrected it as below:

(Results, p13, line 10-12)

Fluid balance was well controlled; mean right atrium pressure (RAP) was 8.0 ± 4.0 mmHg, and pulmonary artery wedge pressure (PAWP) was 9.4 ± 3.5 mmHg.

Page 21

95% CI: -0.76 to -0.43, p < 0.01, respectively). Therefore, the NSE PTA scoring balloon shows superior improvement in blood-flow after BPA over the conventional balloon when the B/V ration is greater than 1.0.

**RATIO

Response:

Thanks to your pointing. We corrected it as below:

(Results, p21, line 8-10)

Therefore, the NSE PTA scoring balloon shows superior improvement in blood-flow after BPA over the conventional balloon when the B/V ratio is greater than 1.0.

Page 24

In addition, Soga et al. reported the usefulness of NSE PTA in experimental model was enormously observed in the group with B/V ratio more than 1.0 [22].

**Would replace ENORMOUSLY here.

Response:

We greatly appreciate your comment. We deleted ‘ENORMOUSLY’ as below:

(Discussion, p24, line 1-2)

In addition, Soga et al. reported the usefulness of NSE PTA in experimental model was observed in the group with B/V ratio more than 1.0 [22].

Page 25

Since the cross section of this element is wedge-shaped, high stress is efficiently concentrated to the vessel wall. This may prevent severe dissociation that reaches the media.

**DISSECTION

Response:

We thank to your pointing. We modified it as below:

(Discussion, p25, line 15-17)

Since the cross section of this element is wedge-shaped, high stress is efficiently concentrated to the vessel wall. This may prevent severe dissection that reaches the media.

Page 26

There were several limitations in this study. First, this study was conducted retrospectively at a single center with a limited sample size, not randomized trial.

**A RANDOMONIZED TRIAL

Response:

We greatly appreciate your comment. We corrected it as below:

(Discussion, p26, line 4-6)

There were several limitations in this study. First, this study was conducted retrospectively at a single center with a limited sample size, not a randomized trial.

Page 26

. In addition, we evaluated only immediately after BPA. It has been reported that treated pulmonary arteries are took time to dilate sufficiently.

**MAY TAKE TIME

Page 26

We need to investigate follow-up data to conclude the effectiveness of scoring balloon in BPA procedure.

**PROCEDURES

Response:

Thanks to your pointing. We modified it as below:

(Discussion, p26, line 14-16)

It has been reported that treated pulmonary arteries may take time to dilate sufficiently. We need to investigate follow-up data to conclude the effectiveness of scoring balloon in BPA procedures.

Reviewer #2: PONE-D-21-25699: statistical review

SUMMARY. This is a retrospective study that compares the effect of the non-slip element percutaneous transluminal angioplasty (NSE PTA) on blood pressure ratios to the effect of plain-balloon treatments (POBA). The statistical analysis relies on t-test methods for comparing the NSE PTA and the POBA groups and on regression analysis for evaluating the effect of the use of NSE PTA scoring balloon on the pressure ratio in a subset of data where the balloon-to-vessel ratio exceeded 1.0. I have several major concerns about the statistical methods that have been employed in this paper.

Response:

We thank Reviewer #2 for the kind and constructive comments. I have described below our responses to the reviewers' comments and the changes to the revised manuscript, tables and figures. I hope that these responses will satisfy Reviewer #2.

MAJOR ISSUES

1. The statistical analysis focuses on the improvements of pressure ratios observed after Balloon pulmonary angioplasty (BPA) in 108 lesions, which are treated as independent observations. However, if I understood correctly, these lesions are clustered within 14 subjects and, as such, they may not be assumed as independent cases. Instead, they should be treated as repeated measures within a sample of 14 subjects. Repeated measures methods are available in SAS, which is the software used by the authors. Notice that the repeated measures structure of the data must be accounted for in both the ANOVA analysis that compares the NSE PTA and the POBA groups and in the regression analysis that focuses on the effect of the NSE PTA treatment.

Response: 

I appreciate your constructive comments. In the BPA procedures, many pulmonary branches are treated in one patient in several sessions. In this study, 108 lesions are different branches in 14 patients. Since each lesion were dilatated by either plain balloon or NSE PTA balloon, we consider these lesions were independent respectively. However, as you mentioned, these lesions were treated repeatedly in same patients. Therefore, we used mixed models for repeated measures as you pointed out, and confirmed the almost same results. We corrected the revised tables and manuscript.

Manuscript:

(Method, p11, line 8-11):

We compared the baseline characteristics, pressure ratio and Δ Pressure ratio at the target lesion between the NSE PTA group and the POBA group before and after BPA procedure using mixed models for repeated measures.

(Method, p11, line 15-18):

We performed univariate analysis using regression analysis with mixed models for repeated measures to find the relationship of Δ Pressure ratio in whole patients and patients with B/V ratio over 1.0 according to Soga's report [22].

(Result, p16-17, line 16-4):

Although pressure ratio before BPA tended to be lower in the NSE PTA group, there was not statistically significant difference (0.526 ± 0.267 in the POBA group, 0.415 ± 0.243 in the NSE PTA group, p = 0.08). There was a significant difference in pressure ratio after BPA (0.766 ± 0.162 in the POBA group, 0.675 ± 0.208 in the NSE PTA group, p = 0.03). Despite this finding, there was no significant difference in the Δ Pressure ratio between the two groups (0.241 ± 0.196 in the POBA group, 0.259 ± 0.177 in the NSE PTA group, p = 0.63)

(Result, p18, line 11-12):

In univariate analysis, only pressure ratio before BPA demonstrated significant relationships with Δ Pressure ratio.

2. This is a retrospective study and, as a result, subjects have not been randomly associated to the NSE PTA and the POBA groups. The authors should therefore provide evidence that these two groups do not significantly differ with respect to the biometrical variables that are summarized in Table 1. Otherwise, results could be biased by group-specific differences

Response:

We apologize for confusing you. In the BPA procedures, many pulmonary branches are treated in one patient in several sessions. Each vessel was treated with either POBA or NSE PTA. Therefore, both plain balloon and NSE PTA balloon were used in same patient. We added additional information about BPA procedures and lesion selection that were analyzed as follows.

(Abstract, p3, line 7-9)

108 pulmonary artery branches in 14 CTEPH patients who underwent BPA using NSE PTA scoring balloon (the NSE PTA group) or plain balloon (the POBA group) and pressure gradient evaluation were analyzed.

(Method, p8, line 5-7)

108 pulmonary artery branches in 14 consecutive patients with CTEPH who underwent BPA using NSE PTA scoring balloon or plain balloon and had pressure gradient evaluation with pressure microcatheter were analyzed.

3. The statistical analysis relies on the assumption that the dependent variables are normally distributed. The authors should provide some evidence that such assumption is realistic, otherwise all the p-values of the paper could be biased.

Response:

We greatly appreciate your comment. As a result of the verification, we consider that all variables except ‘Balloon size’, ‘Balloon Pressure’, and ‘Ratio of pressure ratio’ are normally distributed from each histogram and Q-Q plot. As a representative validation, the histogram and the Q-Q plot of ΔPressure ratio are shown below. Valuables that were not normally distributed were analyzed by using nonparametric analysis. According to your recommendation, we revised our manuscript about the evaluation of normal distribution and Tables.

(Method, p11, line 6)

We evaluated whether each valuable was normally distributed.

(Method, p11, line 12-13)

Valuables that were not normally distributed were compared using non-parametric test.

(Result, p14, line 1-4)

Although vessel diameter and balloon size tended to be smaller in the NSE PTA group, there did not appear to be statistically significant difference. Balloon pressure was higher in the NSE PTA group (8.0 (6.0–8.0) atm in the POBA group and 8.0 (8.0–8.0) atm in the NSE PTA group, p = 0.01).

Attachment

Submitted filename: Response to Reviewers 211222.docx

Decision Letter 1

Yoshiaki Taniyama

17 Jan 2022

The effectiveness of scoring balloon angioplasty in the treatment of chronic thromboembolic pulmonary hypertension

PONE-D-21-25699R1

Dear Dr. Nakanishi,

We’re pleased to inform you that your manuscript has been judged scientifically suitable for publication and will be formally accepted for publication once it meets all outstanding technical requirements.

Within one week, you’ll receive an e-mail detailing the required amendments. When these have been addressed, you’ll receive a formal acceptance letter and your manuscript will be scheduled for publication.

An invoice for payment will follow shortly after the formal acceptance. To ensure an efficient process, please log into Editorial Manager at http://www.editorialmanager.com/pone/, click the 'Update My Information' link at the top of the page, and double check that your user information is up-to-date. If you have any billing related questions, please contact our Author Billing department directly at authorbilling@plos.org.

If your institution or institutions have a press office, please notify them about your upcoming paper to help maximize its impact. If they’ll be preparing press materials, please inform our press team as soon as possible -- no later than 48 hours after receiving the formal acceptance. Your manuscript will remain under strict press embargo until 2 pm Eastern Time on the date of publication. For more information, please contact onepress@plos.org.

Kind regards,

Yoshiaki Taniyama, MD, PhD

Academic Editor

PLOS ONE

Additional Editor Comments (optional):

The author responded the problems.

Reviewers' comments:

Reviewer's Responses to Questions

Comments to the Author

1. If the authors have adequately addressed your comments raised in a previous round of review and you feel that this manuscript is now acceptable for publication, you may indicate that here to bypass the “Comments to the Author” section, enter your conflict of interest statement in the “Confidential to Editor” section, and submit your "Accept" recommendation.

Reviewer #2: All comments have been addressed

**********

2. Is the manuscript technically sound, and do the data support the conclusions?

The manuscript must describe a technically sound piece of scientific research with data that supports the conclusions. Experiments must have been conducted rigorously, with appropriate controls, replication, and sample sizes. The conclusions must be drawn appropriately based on the data presented.

Reviewer #2: (No Response)

**********

3. Has the statistical analysis been performed appropriately and rigorously?

Reviewer #2: (No Response)

**********

4. Have the authors made all data underlying the findings in their manuscript fully available?

The PLOS Data policy requires authors to make all data underlying the findings described in their manuscript fully available without restriction, with rare exception (please refer to the Data Availability Statement in the manuscript PDF file). The data should be provided as part of the manuscript or its supporting information, or deposited to a public repository. For example, in addition to summary statistics, the data points behind means, medians and variance measures should be available. If there are restrictions on publicly sharing data—e.g. participant privacy or use of data from a third party—those must be specified.

Reviewer #2: (No Response)

**********

5. Is the manuscript presented in an intelligible fashion and written in standard English?

PLOS ONE does not copyedit accepted manuscripts, so the language in submitted articles must be clear, correct, and unambiguous. Any typographical or grammatical errors should be corrected at revision, so please note any specific errors here.

Reviewer #2: (No Response)

**********

6. Review Comments to the Author

Please use the space provided to explain your answers to the questions above. You may also include additional comments for the author, including concerns about dual publication, research ethics, or publication ethics. (Please upload your review as an attachment if it exceeds 20,000 characters)

Reviewer #2: (No Response)

**********

7. PLOS authors have the option to publish the peer review history of their article (what does this mean?). If published, this will include your full peer review and any attached files.

If you choose “no”, your identity will remain anonymous but your review may still be made public.

Do you want your identity to be public for this peer review? For information about this choice, including consent withdrawal, please see our Privacy Policy.

Reviewer #2: No

Acceptance letter

Yoshiaki Taniyama

20 Jan 2022

PONE-D-21-25699R1

The effectiveness of scoring balloon angioplasty in the treatment of chronic thromboembolic pulmonary hypertension

Dear Dr. Nakanishi:

I'm pleased to inform you that your manuscript has been deemed suitable for publication in PLOS ONE. Congratulations! Your manuscript is now with our production department.

If your institution or institutions have a press office, please let them know about your upcoming paper now to help maximize its impact. If they'll be preparing press materials, please inform our press team within the next 48 hours. Your manuscript will remain under strict press embargo until 2 pm Eastern Time on the date of publication. For more information please contact onepress@plos.org.

If we can help with anything else, please email us at plosone@plos.org.

Thank you for submitting your work to PLOS ONE and supporting open access.

Kind regards,

PLOS ONE Editorial Office Staff

on behalf of

Dr. Yoshiaki Taniyama

Academic Editor

PLOS ONE

Associated Data

    This section collects any data citations, data availability statements, or supplementary materials included in this article.

    Supplementary Materials

    Attachment

    Submitted filename: Response to Reviewers 211222.docx

    Data Availability Statement

    All relevant data are within the paper.


    Articles from PLoS ONE are provided here courtesy of PLOS

    RESOURCES