Abstract
Blood-based RRT, such as hemodialysis, requires access to the bloodstream and adequate blood flow to enable the requisite clearance. As such, nearly all RRT systems require two lumens, enabling a blood circuit that pulls blood from one lumen or needle and returns it via another lumen or needle. The proposed single lumen alternating micro-batch (SLAMB) technique uses a small single lumen to draw a “micro” batch of blood into a single reservoir. In the reservoir, the “batch” of blood is circulated at a high blood flow rate through a hemofilter or hemodialyzer, enabling efficient small- and middle-molecule clearance. Thereafter, the “purified” blood is returned to the patient and the cycle is repeated. Each batch comprises 20–300 ml of blood, which is adjusted to the vascular access, hemodynamic status, and size of the patient. Up to 15 cycles can be done per hour, allowing this system to achieve a blood clearance level comparable to modern continuous RRT systems. Because the system can function with a small-bore single lumen, this device can work with existing central lines, thus allowing for less invasive vascular access. Because of their size and relative simplicity, SLAMB-based systems are less expensive, smaller, and have improved portability. Lastly, a similar, manual SLAMB-hemofiltration kit, which requires no electricity or battery, can be developed at low cost (<$25) for use in austere medical conditions, thus expanding the availability of RRT for patients with AKI.
Keywords: dialysis, 0by25, batch dialysis, dialysis, hemodiafiltration, hemoperfusion, home dialysis, manual dialysis, renal replacement therapy, single needle, SLAMB
Introduction
Modern, blood-based RRTs, such as hemodialysis and hemofiltration, require access to the blood compartment and adequate blood flow to enable the required clearance for the RRT prescription. As such, nearly all RRT systems require two lumens, enabling an afferent and efferent limb to create a blood circuit that pulls blood from one lumen or needle and returns it via another lumen or needle. Additionally, achieving high levels of solute clearance requires high blood flow, thereby necessitating large-bore vascular access. As an example, to conduct a typical RRT treatment, a dual-lumen catheter with a diameter of 11–13 French (Fr), an arteriovenous graft, or a mature arteriovenous fistula is required. All of these vascular access types require maintenance to assure patency and are associated with potential complications.
The device concept proposed in this technical note is a modernization of the first hemodialysis ever performed. The first successful hemodialysis procedure was performed using a drum dialyzer, wherein a sausage casing made of cellophane was wrapped around a cylinder which was immersed in a large drum filled with dialysate (Figure 1). The patient was phlebotomized and the blood was placed into a jug with an anticoagulant (i.e., heparin). This “batch” of blood was then placed into the sausage casing membrane and submerged into the dialysate, thereby allowing a diffusive dialysis effect. After this procedure was completed, the dialyzed blood was drained into a second jug which was then transfused back into the patient. This procedure was repeated multiple times until an adequate clearance was achieved, and this series of “drain-refill” procedures comprised the first successful dialysis treatment(s) (1).
Figure 1.
Picture of a patient being treated with a Kolff drum dialyzer.
The treatment outlined above was cumbersome and took about 6 hours to complete. Therefore, an approach using an afferent/efferent circuit was pursued to improve the efficiency of the treatment, and this basic technique is still the basis of modern RRT. The device described below maintains the advantages but eliminates the cumbersome features of the initial batch concept by using modern pump, scale, and filter technology.
Description
To operate the device, an intravascular catheter that is patent and can reliably withdraw up to 300 ml of blood is all that is required. The single lumen alternating micro-batch (SLAMB) technique uses a small single lumen to draw a “micro” batch of blood into a single reservoir (Figure 2). If the patient is not already anticoagulated, a small amount of anticoagulant (e.g., heparin, citrate) is added to the reservoir first. A small batch of blood ranging from 20 to 300 cc of blood is added to the reservoir. Once the blood is in the reservoir, the blood is dialyzed and/or hemofiltered at a high blood flow rate and a high dialysate flow rate for 1–3 minutes. The blood is recirculated back into the reservoir after it is dialyzed, so the same blood can be reprocessed multiple times, thus allowing rapid and high-efficiency clearance of solutes and ultrafiltration as needed. Once the blood is purified, the blood is returned to the patient along with an anticoagulant-reversal agent if needed (i.e., protamine, calcium). This procedure is repeated until the desired level of clearance and ultrafiltration are achieved. The time for one full cycle (blood drawn in, hemodiafiltration, blood returned) is dependent on the size of the batch and the vascular access. Assuming a batch of 200 cc and a single lumen of a standard central line, the entire cycle would take 4–7 minutes, allowing eight to 15 batches per hour.
Figure 2.

Configuration of the SLAMB-hemodiafiltration (HDF). The confguration depicted can perform hemoperfusion, hemofiltration, hemodialysis, aquapheresis, or hemodiafiltrion. SLAMB, single lumen alternating micro-batch.
Results
Schematics and animation of the proposed SLAMB systems are shown in Figures 2 and 3 and Supplemental Figures 1–5.
Figure 3.

Configuration of SLAMB-hemofiltration (HF). Simplified SLAMB designed for hemofiltration only. This configuration only requires three pumps. SLAMB-HF, single lumen alternating micro-batch–hemofiltration.
Clearance rates and times have been modeled and shown in Supplemental Table 1 and Table 1. Supplemental Tables 2 and 3 outline the manual SLAMB dialysate and approximated costs, respectively.
Table 1.
SLAMB dosing models based on various prescriptions
| Batch Size (ml) | Total Cycle Time (min) | Reservoir Cycle Time (min) | Time to Achieve Dose of 20 cc/kg per Hour (h) | Intravenous Access Minimum Flow Rate Requirement (ml/min) |
| 100a | 3 | 2 | 17.9 | 200 |
| 100a | 3.5 | 2 | 20.8 | 133 |
| 100a | 4 | 2 | 23.8 | 100 |
| 150a | 4 | 2 | 15.9 | 150 |
| 150a | 4 | 2.5 | 15.9 | 200 |
| 150a | 5 | 2.5 | 19.8 | 120 |
| 150a | 6 | 2.5 | 23.8 | 85.7 |
| 200a | 4 | 3 | 11.9 | 400 |
| 200a | 5 | 3 | 14.8 | 200 |
| 200a | 6 | 3.5 | 17.8 | 160 |
| 200a | 7 | 3.5 | 20.8 | 114 |
| 200b | 3 | 2 | 8.9 | 400 |
| 250b | 3.5 | 2 | 8.3 | 400 |
| 250 | 5 | 3 | 11.8 | 250 |
| 250 | 6 | 4 | 14.2 | 250 |
| 250 | 7 | 4 | 16.6 | 167 |
| 300 | 6 | 4 | 11.8 | 300 |
| 300 | 7 | 4 | 13.8 | 200 |
| 300 | 8 | 4 | 15.7 | 150 |
The dosing models assume a 75-kg subject, no ultrafiltration, and that the distribution of urea is the same as the total body water. Dose goal is Kt/V of 0.8 (equivalent to 20 cc/kg per hour for 24 h) (8). SLAMB, single lumen alternating micro-batch.
SLAMB prescriptions that would likely be used for patients who are critically ill.
Prescriptions that could be used for home dialysis.
Discussion
A concept known as “single-needle dialysis” was proposed by Kopp et al. (2) in 1972, and this type of RRT has been used since that time with varying degrees of enthusiasm (3). Currently, there are two single-needle devices are available in Europe, the Dialog iQ (Braun) and FMC 5008 (Fresenius), both of which have vastly improved the original concept (4). The Dialog iQ and the FMC 5008 offer the benefit of a single needle, yet maintain the familiar arterial and venous circuit through which blood is dialyzed, and they use a large gauge needle (14 or 15 gauge) or large-bore central line. The Dialog iQ and FMC 5008 systems can both perform an intermittent therapy, albeit with reduced efficiency compared with standard intermittent machines (3,4). These devices are good options for patients who can only tolerate a single needle, but can have significant recirculation. In addition to the above systems, a device called the Newcastle Infant Dialysis and Ultrafiltration System (NIDUS) has been developed for neonates that uses a single-lumen system, comprising two reservoirs made up of two low-volume syringes to conduct precise hemofiltration. NIDUS is designed specifically for low-weight children (5).
The SLAMB-hemodiafiltration (HDF) system builds on these initial concepts and is designed for larger patients, a longer duration of therapy, and would be better suited for continuous RRT, aquapheresis, slow continuous ultrafiltration, periodic intermittent RRT, nocturnal dialysis, and/or daily home dialysis. There are multiple advantages to the proposed SLAMB-HDF technique over current dual-lumen RRT systems. First, this technique can occur via a small single-lumen catheter (i.e., one lumen of a standard triple-lumen central line or a small single needle), potentially easing the burden of vascular access in both acute and chronic RRT patients. Second, because the blood is held in a reservoir and recirculated, this system disentangles the usual need of large vascular access and high blood flow to achieve high middle-molecule and protein-bound uremic toxin clearance. Third, because the blood is held in a reservoir, this system can also be used for hemoperfusion, aquapheresis, and other extracorporeal therapeutics. Fourth, because blood batches are small and anticoagulated, a small hemofilter can be used, thus decreasing the cost. Fifth, the small batch and small filter can allow for a small device that takes up less space and increases portability. Sixth, the intravenous (IV) access attached to the SLAMB device can also be used to administer regular IV medications while performing RRT; this is possible because the device has three cycles: (1) “withdraw blood,” (2) “process blood,” and (3) “return blood” (Supplemental Figures 1–3). Thus, during the process-blood and return-blood phase, a separate IV pump synchronized to the SLAMB device can continue to administer other IV medications and pause infusion during the withdraw-blood phase. In essence, a patient with an existing central line can have RRT initiated without the loss of that lumen for giving other medications. Seventh, the absence of a dual-lumen catheter precludes the inefficiencies due to blood recirculation. Lastly, the batch size can be customized for the size and severity of illness of the patient. Theoretically, having a smaller amount of blood leave the circulation should decrease the hemodynamic instability often seen when a RRT session is initiated.
However, there are some disadvantages to the proposed system. First, the small batches preclude this device for being used for a standard intermittent RRT session, because the time for the push/pull and dialysis/hemofiltration to occur would take too long to generate the requisite clearance over the short period of time (e.g., 3–4 hours). Second, due to the batching, systemic anticoagulation or “regional” anticoagulation is required, which may be inappropriate in some patients.
SLAMB-Hemofiltration for Use in Austere Conditions
A simpler version of this device that uses only hemofiltration is shown in Figure 3. In this system, the blood from the patient is drawn into the reservoir, along with a small amount of anticoagulant and hemofiltration fluid (sterile crystalloid or similar hemofiltration fluid). The reservoir is then rapidly hemofiltered, and the blood is then returned to the patient. Because this system does not require high-flow pumps and relies exclusively on hemofiltration, the overall system is small and requires only three small pumps and one gravity scale. (Figure 3) This low-power system can run on a battery and its capacity to work with a small single-lumen catheter makes it appealing in places where full nephrology services are not available. The overall cost for this simple, minimally invasive system would be <$1000 per machine and can use known standard technology, making maintenance and repair inexpensive and tractable.
A fully manual syringe mode of the SLAMB-hemofiltration (SLAMB-HF) system is shown in Supplemental Figure 4. Although this system is labor intensive, it is “powered” manually with two 50-cc syringes, six three-way stopcocks, one hemofilter, two bags, and IV tubing. This manual version of the SLAMB-HF system does not require any electricity and would have a total cost of <$25 per kit. Although not ideal, a combination of 1 L 5% dextrose in normal saline and 4 L Hartmann solution (i.e., lactated Ringer solution) make an adequate hemofiltration solution to manage uremia, acidemia, and hyperkalemia (Supplemental Table 2). Because these sterile IV fluids are inexpensive and widely available, a simple, manual SLAMB-HF machine coupled with these fluids could allow for adequate RRT for patients with AKI who are likely to recover renal function (i.e., rhabdomyolysis). The manual SLAMB would require training to be implemented properly, but the technical operation is simple and could be taught to a combat medic, nurse, or other skilled healthcare professional, thus expanding the availability of RRT. An example where this might be useful is in the immediate aftermath of an earthquake. Many earthquake survivors suffer crush injury and associated rhabdomyolysis with severe AKI. The provision of RRT could prove lifesaving in many of these patients. An inexpensive RRT system that can be deployed with low-cost IV fluids and does not require power may be advantageous. In addition, the placement of a peritoneal catheter or a double-lumen catheter can be challenging in an austere environment because both require technical expertise. Because the SLAMB systems require more modest vascular access, RRT may be more feasible and safer with this system in an austere environment. The International Society of Nephrology has initiated the 0by25 program and, if the manual SLAMB-HF system is developed and deployed, it may improve the availability of RRT for patients with AKI and mitigate otherwise preventable deaths (6).
Dose, IV Access, and Anticoagulation Considerations
In general, the cycle time and, therefore, total dose of RRT requires an adequate IV access. A standard, double-lumen dialysis catheter of 11 Fr (each lumen of 5–5.5 Fr) can run a blood flow rate of 200–300 ml/min. Thus, a single-lumen catheter of at least 5 Fr can easily allow batches of 200–250 ml to be drawn into the SLAMB in 1 minute and returned in 1 minute. Single-pass urea clearance in hemodialysis ranges from between 85% and 90%, thus a batch of 200–250 ml can then be (re)cycled at 200–400 ml/min within the reservoir for 2–3 minutes to achieve 90%–95% clearance (7). Therefore, a conservative estimate of the total cycle time with a single-lumen catheter of 5 Fr is as follows: 5 minutes (1 minute ingress, 2–3 minutes of clearance, 1 minute blood return), which would allow 12–15 cycles per hour. If the urea distribution is assumed to be the same as total body water, the standard continuous RRT dose of 20 ml/kg per hour equates to a Kt/V of 0.8 (8). Modeling of different SLAMB prescriptions is shown in Table 1. In the above example, a 75-kg patient dosed with a SLAMB prescription of 200 cc batch, 5-minute cycle time, 0 ml of ultrafiltrate would achieve a Kt/V of 0.8 in 14.8 hours. If smaller/longer IV access were used, thus extending the cycle time, the time to achieve a Kt/V would be extended. In this example, if the ingress/egress of blood from the device were extended to 9 minutes (3 minutes cycling, 3 minutes for ingress, and 3 minutes for egress), a catheter would need to be able to enable a blood flow rate of at least 67 ml/min, and a Kt/V of 0.8 could be achieved in 24 hours.
Because the SLAMB system uses small batches that are resident in a reservoir, some element of anticoagulation is a prerequisite. However, the amount of anticoagulant (i.e., heparin/sodium citrate) per ml of blood to enable the batch operation is not yet known. Future research in the development of this proposed system will need to be done to understand the anticoagulation dosing parameters.
Conclusion
A SLAMB-HDF platform is a concept that may allow RRT to be conducted with a single and smaller vascular access. Systems based on this design are simpler than current RRT systems, making them less expensive, lighter, and more portable, thus increasing the options for patients who require RRT.
Disclosures
L. Chawla reports submitting patents on the SLAMB design and concept.
Funding
None.
Supplemental Material
This article contains the following supplemental material online at http://kidney360.asnjournals.org/lookup/suppl/doi:10.34067/KID.0001462020/-/DCSupplemental.
SLAMB dose modeling. Download Supplemental Table 1, PDF file, 1.3 MB (1.3MB, pdf)
Manual SLAMB-HF supplies and approximated costs. Download Supplemental Table 2, PDF file, 1.3 MB (1.3MB, pdf)
Hemofiltration fluid made with 5% dextrose/normal saline and lactated Ringers solution. Download Supplemental Table 3, PDF file, 1.3 MB (1.3MB, pdf)
SLAMB – HDF animated. Download Supplemental Figure 1, PDF file, 1.3 MB (1.3MB, pdf)
SLAMB – HF animated. Download Supplemental Figure 2, PDF file, 1.3 MB (1.3MB, pdf)
SLAMB – HD animated. Download Supplemental Figure 3, PDF file, 1.3 MB (1.3MB, pdf)
Manual syringe SLAMB system animated. Download Supplemental Figure 4, PDF file, 1.3 MB (1.3MB, pdf)
SLAMB – hemoperfusion. Download Supplemental Figure 5, PDF file, 1.3 MB (1.3MB, pdf)
Acknowledgments
The author would like to thank Dr. David Askenazi, Dr. Pat Brophy, Dr. Stuart Goldstein, Prof. Lui Forni, and Dr. Samir Patel for their constructive critiques and advice.
Author Contributions
L. Chawla conceptualized the study, was responsible for data curation, provided supervision, wrote the original draft, and reviewed and edited the manuscript.
References
- 1.Twardowski ZJ: History of hemodialyzers’ designs. Hemodial Int 12: 173–210, 2008 [DOI] [PubMed] [Google Scholar]
- 2.Kopp KF, Gutch CF, Kolff WJ: Single needle dialysis. Trans Am Soc Artif Intern Organs 18: 75–81, 1972 [DOI] [PubMed] [Google Scholar]
- 3.Vanholder R: Single needle hemodialysis: Is the past the future? J Nephrol 33: 49–58, 2020 [DOI] [PubMed] [Google Scholar]
- 4.Bieser W, Welsch M, Jakob M, Meibaum J, Rodriguez DA, Wolff H, Marcelli D, Barth C: Effectiveness of a new single-needle single-pump dialysis system with simultaneous monitoring of dialysis dose. Artif Organs 42: 814–823, 2018 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 5.Coulthard MG, Crosier J, Griffiths C, Smith J, Drinnan M, Whitaker M, Beckwith R, Matthews JN, Flecknell P, Lambert HJ: Haemodialysing babies weighing <8 kg with the Newcastle infant dialysis and ultrafiltration system (Nidus): Comparison with peritoneal and conventional haemodialysis. Pediatr Nephrol 29: 1873–1881, 2014 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 6.Mehta RL, Cerdá J, Burdmann EA, Tonelli M, García-García G, Jha V, Susantitaphong P, Rocco M, Vanholder R, Sever MS, Cruz D, Jaber B, Lameire NH, Lombardi R, Lewington A, Feehally J, Finkelstein F, Levin N, Pannu N, Thomas B, Aronoff-Spencer E, Remuzzi G: International Society of Nephrology’s 0by25 initiative for acute kidney injury (zero preventable deaths by 2025): A human rights case for nephrology. Lancet 385: 2616–2643, 2015 [DOI] [PubMed] [Google Scholar]
- 7.Macías N, Vega A, Abad S, Aragoncillo I, García-Prieto AM, Santos A, Torres E, Luño J: Middle molecule elimination in expanded haemodialysis: Only convective transport? Clin Kidney J 12: 447–455, 2018 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 8.Vijayan A, Palevsky PM: Dosing of renal replacement therapy in acute kidney injury. Am J Kidney Dis 59: 569–576, 2012 [DOI] [PMC free article] [PubMed] [Google Scholar]
Associated Data
This section collects any data citations, data availability statements, or supplementary materials included in this article.
Supplementary Materials
SLAMB dose modeling. Download Supplemental Table 1, PDF file, 1.3 MB (1.3MB, pdf)
Manual SLAMB-HF supplies and approximated costs. Download Supplemental Table 2, PDF file, 1.3 MB (1.3MB, pdf)
Hemofiltration fluid made with 5% dextrose/normal saline and lactated Ringers solution. Download Supplemental Table 3, PDF file, 1.3 MB (1.3MB, pdf)
SLAMB – HDF animated. Download Supplemental Figure 1, PDF file, 1.3 MB (1.3MB, pdf)
SLAMB – HF animated. Download Supplemental Figure 2, PDF file, 1.3 MB (1.3MB, pdf)
SLAMB – HD animated. Download Supplemental Figure 3, PDF file, 1.3 MB (1.3MB, pdf)
Manual syringe SLAMB system animated. Download Supplemental Figure 4, PDF file, 1.3 MB (1.3MB, pdf)
SLAMB – hemoperfusion. Download Supplemental Figure 5, PDF file, 1.3 MB (1.3MB, pdf)