Abstract
Introduction:
Several computer-aided diagnosis systems for depression are suggested for use by clinicians to authorize the diagnosis. EEG may be used as an objective analysis tool for identifying depression in the initial stage to avoid it from reaching a severe and permanent state. However, artifact contamination reduces the accuracy in EEG signal processing systems.
Methods:
This work proposes a novel denoising method based on Empirical Mode Decomposition (EMD) (with Detrended Fluctuation Analysis (DFA) and wavelet packet transform. Initially, real EEG recordings corresponding to depression patients are decomposed into various mode functions by applying EMD. Then, DFA is used as the mode selection criteria. Further Wavelet Packets Decomposition (WPD)-based evaluation is applied to extract the cleaner signal.
Results:
Simulations were conducted on real EEG databases for depression to demonstrate the effects of the proposed techniques. To conclude the efficacy of the proposed technique, SNR and MAE were identified. The obtained results indicated improved signal-to-noise ratio and lower values of MAE for the combined EMD-DFA-WPD technique. Additionally, Random Forest and SVM (Support Vector Machine)-based classification revealed the improved accuracy of 98.51% and 98.10% for the proposed denoising technique. Whereas the accuracy of the EMD- DFA is 98.01% and 95.81% and EMD combined with DWT technique equaled
98.0% and 97.21% for the EMD- DFA technique for RF and SVM, respectively, compared to the proposed method. Furthermore, the classification performance for both classifiers was compared with and without denoising to highlight the effects of the proposed technique.
Conclusion:
Proposed denoising system results in better classification of depressed and healthy individuals resulting in a better diagnosing system. These results can be further analyzed using other approaches as a solution to the mode mixing problem of the EMD approach.
Keywords: EEG, Wavelets, Artifacts, Empirical Mode Decomposition (EMD), Depression
Highlights
Several computer-aided systems are suggested for diagnosing depression.
EEG may be used as an objective analysis tool for identifying depression.
Denoising system proposed in this study results in better classification of depressed and healthy individuals.
Plain Language Summary
Depression is a major global burden. For years, EEG-based computer-aided systems were a powerful tool for detecting neurological disorders. However, advanced healthcare facilities based on EEG for identifying depression in the early stage are required to avoid a severe and irreversible state. Thus, preprocessing is an exclusively required step in EEG signal analysis. EEG is sensitive to certain irrelevant sources as well as artifacts, like EOG and EMG. Denoising describes the procedure of removing noise present in the signal. At first, a linear digital filter-based reduction of superimposed noise on the tracings of the EEG was proposed. Then, the efficiency of the regression analysis was demonstrated for single trials of ERP signals and the average potentials. Denoising system proposed in this study results in better classification of depressed and healthy individuals and the results can be further analyzed using other approaches.
1. Introduction
Depression is a major global burden among societies worldwide. EEG-based computer-aided systems were a powerful tool for detecting numerous neurological disorders. Such systems have presented technical investigates concerning the neuroscience industry for exploring the neuronal mechanics underlying various symptoms of Depression (Acharya et al., 2015; Acharya et al., 2015; Cai et al., 2018; Sharma, Achuth, Deb, Puthankattil, & Acharya, 2018). Advanced healthcare facilities need to be designed using computational and mathematical methodologies based on EEG for identifying depression in the early stage to avoid a severe and irreversible state. Preprocessing is an exclusively required step in EEG signal analysis. The tentative measured signal can be represented as a linear combination of multiple brain events. EEG is sensitive to certain irrelevant sources as well as artifacts, like EOG (ocular artifacts), Electromyography (EMG) due to movement artifacts, and various other technical sources that interfere with the signal at particular scalp electrodeposition; thus, making control over the signal confused. Therefore, there is a necessity for preserving an adequate signal over the noise in biomedical applications (Kaur & Singh, 2015, 2016; Vaid, Singh, & Kaur, 2015).
Denoising describes the procedure of removing noise present in the signal. To challenge this problem, various algorithms were proposed, as shown in Figure 1. These include regression techniques (Awal, Mostafa, Ahmad, & Rashid, 2014; Wallstrom, Kass, Miller, Cohn, & Fox, 2004), autoregressive models (Hoffmann & Falkenstein, 2008), bandpass, and adaptive filtering, singular value decomposition, Kalman filtering, PCA, ICA and EMD (Barua & Begum, 2014; Cluitmans & Van De Velde, 2000; Molla, Tanaka, & Rutkowski, 2012).
Figure 1.
Existing denoising algorithms
In the initial years, a linear digital filter-based reduction of superimposed noise on the tracings of the EEG was proposed. Kalman filter was used as the optimal filter for removing the EMG noise. The performance of the filters is satisfactory; also from the clinical standpoint, obtaining a marked reduction of noise without distorting the useful information contained in the signal (Bartoli & Cerutti, 1983). Then, the efficiency of the regression analysis was demonstrated for single trials of ERP (Event-Related Potentials) signals and the average potentials (Semlitsch, Anderer, Schuster, & Presslich, 1986). In addition to providing the regression factors, it offers reduced coherence among the eye blink activity and ongoing EEG activity (Semlitsch et al., 1986).
The Blind Source Separation (BSS) algorithms, such as Principal Component Analysis (PCA) and Independent Component Analysis (ICA) were apparent as influential artifact removal methodologies. However, PCA and ICA have certain disadvantages, e.g., these fail to cope with high order statistical dependence. Moreover, these are inefficient for the removal of artifacts in case artifacts have the corresponding magnitude, like brain signals (Makeig, Bell, Jung, & Sejnowski, 1996). In ar study, a BSS-SVM-based artifact removal method for removing the ocular and ECG artifacts was presented (Shoker, Sanei, & Chambers, 2005). Here, SOBI-based components were identified from the raw EEG, and SVM was used to extract the features from these Independent Components (ICs). Then, a series of experiments on simulated EEG recordings for 5 different configurations of EEG electrodes found that SOBI is more effective than the other BSS-based algorithms for denoising (Kierkels, Van Boxtel, & Vogten, 2006; Kaur & Singh, 2016).
There is another technique of Wavelet Transform (WT)-based thresholding that provides more efficient multi-resolution exploration. It has been concluded to perform superior, compared to standard Low Pass Filters (LPF), median filters, and moving average filters (Lahmiri & Boukadoum, 2015). However, the limitation of Gibbs phenomena exists in WT. Additionally, other limitations of the wavelets include the manual setting of the level of decomposition and wavelet basis is needed that may add false harmonics as signals are nonlinear and non-stationary. The distortions might be introduced in the reconstructed signal that may be because of unsuitable breakdown, leading to less efficient denoising (Zeng, Song, Yan, & Qin, 2013). Discrete Wavelet Transform (DWT) was explored for ECG denoising for power line interference, the EMG, and the baseline drift (Alfoouri & Daqrouq, 2008).
The limitations in wavelet are overlapping spectrum and ICA are lacking redundancy in the number of signals, compared to sources. A large body of literature was conducted taking a combination of various techniques using wave-lets and ICA methods; accordingly, they reported the best performance for removing artifacts along with preserving the nominal data loss (Alfoouri & Daqrouq, 2008; Ghandeharion & Erfanian, 2010). Wavelet-Based thresholding is applied to demixed ICs rather than on the raw EEG data (Nazarpour, Wongsawat, Sanei, Chambers, & Oraintara, 2008). A more robust technique was offered to combine Wavelet and ICA without the need to identify the thresholds (Ghandeharion & Erfanian, 2010).
Another transform was growing for the applications of denoising, Empirical Mode Decomposition (EMD). The main advantage of EMD is no need to postulate the mother wavelet and the level of decomposition, compared to WT. EMD was successful for the removal of fractional and white Gaussian noise. However, it has the limitation of mode- mixing. Another restriction of EMD is in defining the stopping conditions of the sifting procedure (Mert & Akan, 2014; Zeng, Song, Yan, & Qin 2013). As a result, hybrid techniques, like EMD with wavelet thresholding and EMD- ICA, etc. were reported in the literature. For example, a study discovered a new technique where a noisy signal was decomposed using EMD then DWT thresholding was followed (Kabir & Shahnaz, 2012). Noise-Free Intrinsic Mode Functions (IMFs) and the residue were added to regenerate the signal. This leaves scope for additional upgrading. Like EMD, an unweighted summation of IMFs filtered after DWT thresholding may overlook the capability of carrying different structural information (Kabir & Shahnaz, 2012). The frequency and the effect of the decomposed signal decrease with an increase in the mode of IMF. Besides, residue contains a little bit of signal information; thus, adding it in the reconstructive step adds slight to the process of artifact removal.
Another finding proposed BSS-EMD based method to recover the loss of information. However, again, such performance is limited by dependence on the quality of ICA-separated ICs. Therefore, another study that used SSSA as a BSS algorithm along with EMD provided better results (Zeng et al., 2013). Mert et al. introduced Detrended Fluctuation Analysis (DFA) as stopping criteria for determining noisy IMFs obtained by EMD (Mert & Akan, 2014). Safieddine et al. proposed a comparison between deterministic (EMD & wavelet approaches) and stochastic (ICA & cross-correlation analysis, i.e., CCA) approaches which concluded that 2T- EMD should be preferred for denoising for lower SNR data (Safieddine et al., 2012).
Bono, Jamal, Das, and Maharatna (2014) introduced two-hybrid techniques of Wavelet Packet Transform (WPT)-ICA and WPT-EMD. Another study provided the comparison of EMD, WT, and Kalman filters (Salis et al., 2013).
A critical review of some of the existing systems for NFT is provided in Table 1. Several artifact removal techniques were presented. Regression-Based techniques were supported for denoising; however, they are limited by the disadvantage of bidirectional contamination. As a solution to the problem of bidirectional contamination, low pass filtering and adaptive filters were offered before applying the regression (Croft & Barry, 2000; Munia, Haider, Schneider, Romanick, & Fazel-Rezai, 2017; Salis et al., 2013; Suchetha & Kumaravel, 2013) our scope was a comparative analysis of the performance of three standard denoising methods like continuous Empirical Mode Decomposition (EMD. However, adaptive filters require defining reference techniques for modeling. Then, PCA found a growing attraction concerning denoising; however, in the case of approximately, the same magnitude with the brain signal of interest, more reliable algorithms of ICA were recognized as providing major contribution compared to PCA. Artifactual ICs identification in the case of ICA was considered in numerous investigations. To cover up these issues of artifactual ICs, wavelet transforms were offered. Another recent denoising methodology of EMD has been proposed afterward. In previous studies, using EMD inspired by wavelet transforms, ignoring various IMFs after wavelet-based thresholding could lead to ignoring information carrying the capacity of IMFs leading to inefficient denoising results. This work focuses on the performance comparison of EMD using DFA followed by WPD to denoise the EEG data with the conventional approaches; it was found more efficient than conventional approaches. A new classification method based on EMD and WPT was implemented. To assess the performance of the proposed algorithm, depression patients and normal individuals were classified using SVM and Random Forest.
Table 1.
A comparative analysis of existing EEG denoising methods
References | Method | Physiological Signals | Outcome |
---|---|---|---|
DiMatteo, Genovese and Kass (2001) | B- splines before applying regression and subspace projection | fMRI Image | Provides better results The problem of defining a reference prototype |
Wallstrom et al. (2004) | Regression-based and PCA, ICA component-based methods | EEG | Using the adaptive filter, the performance of regression-based artifact correction improves. PCA proved to be effective in denoising the EEG signal with minimum spectral distortion Limitation of spectral distortion from ICA based correction procedures and bidirectional contamination |
Shoker et al. (2005) | BSS- SVM | EEG | Efficient Denoising |
Kierkels et al. (2006) | BSS algorithms | EEG | SOBI effective than other BSS algorithms |
Castellanos and Makarov (2006) | wICA | EEG | Conserves both spectral as well as coherence characteristics unlike ICA leading to overestimation of power spectrum and underestimation of coherence property |
Phlypo et al. (2007) | JSSE | EEG | Reducing the distortion and interference of the artifacts than FastICA, SOBI, and JADE algorithms The problem of spectral distortion by ICA |
Alfoouri and Daqrouq (2008) | DWT | ECG | Better SNR and MSE The limitations of Gibbs phenomena, adding false harmonics and less efficient denoising |
Nazarpour et al. (2008) | Robust minimum variance beamformer (RMVB) | EEG | Low cost and more effective |
Wu and Huang (2009) | EEMD | Noise assisted Data | More accurate Lack of mathematical formulation |
Ghandeharion and Erfanian (2010) | Mutual information with ICA and wavelet denoising | EEG | No need to define the threshold values or offline training |
Suchetha and Kumaravel (2013) | EMD | ECG | Beter than adaptive filtering Mode- mixing problem |
Bono et al. (2014) | WPT- ICA & WPT- EMD | EEG | Suitable for artifact removal without any proper information about the artifacts |
Mert and Akan (2014) | EMD- DFA | - | Efficient at low SNR values |
Aneesh et al. (2015) | VMD | Power quality signal | Efficient denoising More accurate |
Kærgaard et al. (2016) | EEMD-BLMS and DWT-NN | ECG | Efficient denoising |
Liu et al. (2017) | VMD | Seismic Data | More robust and well- defined time- domain analysis The problem of defining the procedure of selection of modes |
Empirical mode decomposition
EMD is a recursive process of breaking down a signal into the sum of various finite intrinsic oscillatory functions called IMFs (Intrinsic Mode Functions), i.e., empirically identified based on their feature time scales in the signal. A signal S (t) can be represented as a finite sum of IMFs as in Equation (1).
(1) |
We define IMF as an AM-FM (amplitude modulation-frequency modulation) function written as Equation (2)
(2) |
It is assumed here that Øk’ (t) and Sk(t) are varying lower than. The sk IMF executes as a harmonic component. The algorithm is easily adjustable and the original function’s nonstationary part can be extracted. The stopping criteria are defined by a process called sifting that is carried out in the following steps (Kabir & Shahnaz, 2012; Kiamini, Alirezaee, Perseh, & Ahmadi, 2009; Krupa, Mohd Ali, & Zahedi, 2009).
Local maxima and local minima are defined from the input signal. Then, using the cubic spline line method, upper and lower envelopes were identified.
Take the average of an envelope to mean denoted as h(t). Subtract the input signal and the envelope mean and denote it as the first IMF if it satisfies the two conditions defined above to be met by IMFs. Else, take it as the next input for carrying the next iteration process to find the next IMF.
Repeat the above steps until a stopping criterion is met.
Wavelet packet decomposition
It has lately come into view in different field applications as a new helpful means for signal processing. WPT is the comprehensive structure of DWT. The standard wavelet transform is limited to wavelet bases that move towards the lower frequencies by a power of 2. Thus, it might not be able to give the finest results. However, some other combination of bases might give better desirable results. Discrete wavelet transform gives approximate transformation to the sampled or discrete signals. In the case of WPD, the sampling of low pass and high pass coefficients is conducted to attain d[n[ and an [n] as detail and approximate coefficients. This recursive process is performed with approximate coefficients till a preferred level of decomposition is attained. Wavelet packet decomposition was used in various applications related to emotion recognition in Brain-Computer Interface (BCI) applications. The technique of wavelet packet decomposition provided better results, compared to other existing methods in the terms of accuracy in the space time-frequency domain.
2. Methods
The main aim of this work was to perform decomposition of EEG signal into IMFs by using DFA-based stopping criteria. Then, these IMFs are further analyzed using wavelet packet decomposition. Finally reconstructed signal is analyzed for performance (Figure 2). For the present study, a real EEG dataset prepared by Hospital Universiti Sains Malaysia (HUSM), Kelantan, Malaysia was analyzed. It contains EEG signals of 34 MDD (Major Depressive Disorder) patients and 30 healthy subjects. The sampling rate of data is 256 Hz (Mumtaz, Xia, Yasin, Ali, & Malik, 2017).
Figure 2.
The flowchart of the proposed methodology
For performing the analysis, various values of signal-to-noise ratio are considered by adding the white Gaussian noise to the recorded signal. The additive white Gaussian noise is a basic prototype to present the behavior of naturally occurring random processes having the same intensity at various frequencies. First of all, the decomposed IMFs were selected according to scores defined by DFA. Then. the thresholded IMFs are further decomposed using WPD. Next, these wavelet denoised components are combined with selected IMFs to give the final output signal. To conclude the efficacy of the proposed technique, Signal to Noise Ratio (SNR) and Mean Absolute Error (MAE) were identified at different levels of added white Gaussian noise. If y(t) is the original input signal and is the denoised signal with sampling number represented as T.
SNR, a term encountered in signal processing, is an essential element in describing the quality of the neural information processed from the raw EEG signals. It is frequently used to assess the performance of various physiological systems, to compare and estimate denoising protocols, and to monitor the overall performance of the system. It is defined as the ratio of the related signal divided by the level of the noise. Here, the signal is the amplitude of the EEG signal and the noise is the residual unwanted background activity in the EEG signal that distorts the signal. Thus, SNR in decibels is defined by Equation (3):
(3) |
MAE is used similar to MSE to evaluate the denoising algorithm. MAE is the maximum value of the absolute error signal. It is also defined by using the aforementioned symbols as the Equation (4):
(4) |
Feature Extraction and Classification: For validation purposes, three features namely, Mean, Shannon entropy, and Hjorth parameter widely used in studies related to depression detection (Castellanos & Makarov, 2006; Phlypo, Boon, D’Asseler, & Lemahieu, 2007) followed by rejection of those deemed artificial. We show that a \”leak\” of cerebral activity of interest into components marked as artificial means that one is going to lost that activity. To overcome this problem we propose a novel wavelet enhanced ICA method (wICA were measured from the denoised signals. These features are briefly described as follows:
Mean
This time-domain feature is represented as the central point corresponding to a set of data points. If x(t) represents the data with T samples, the mean is defined as Equation (5):
(5) |
Another time-domain feature termed as Hjorth Parameters is defined using statistical calculations It consists of three parameters namely, activity, mobility, and complexity.
Activity (A): It is defined as Equation (6):
(6) |
Mobilit y (M): If the derivative of x(t) is X(t), then mobility is given by Equation (7)
(7) |
Complexity(C): It is defined as Equation (8)
(8) |
Shannon Entropy: It is the measurement of uncertainty or probability p of the signal value and is defined by Equation (9) as
(9) |
Classification: In this research work, Random Forest (RF) and Support Vector Machine (SVM) classifiers were used for classifying signals into depressed and healthy individuals.
The RF classifier is more accurate in generating the classification results even in the presence of noise. Other advantages of using this algorithm are higher operational efficiency which makes it more efficient for training on the EEG data. The RF classifier being the ensembled algorithm selects a random subset of a training set and generates a set of decision trees. Then, these decision trees are used to create the final test class.
SVM makes use of an assumed space in the form of linear functions based on optimization theory. It acts as a learning system that provides the best hyperplane acting as a separator between two classes of the input space. This system defines margin as the distance among hyperplane and adjoining array (known as support vector) of each class. The learning in SVM involves the power to trace the hyperplane.
To evaluate the performance of the denoising system for EEG signals of depression patients, the classification results are analyzed for these two classifiers. The results using various classifiers and the output are classified as depressed and normal. The parametric evaluation is conducted by calculating the classification accuracy and F1 score. Classification accuracy is defined as the number of accurate estimates made divided by the overall estimates made. More is the classification accuracy; more precise is the proposed system. Accuracy is measured by another metric known as the F1 score, i.e., calculated from precision and recall. A single value is assigned by calculating the harmonic mean from these two attributes. The F1 score was calculated along with the accuracy for this unbalanced class.
To better correlate the results, and to assess the performance of the proposed algorithm, statistical analysis using Repeated-Measures Analysis of Variance (RMANOVA) and a t-test analysis was performed on the denoising results. This statistical analysis was performed to check whether the proposed method outperforms other methods with the value of significance set at α=0.05. The significant differences among the techniques were evaluated using SNR and MAE as the dependent variable. The statistics were calculated for these two variables among the artifactual signals and the denoised signals.
3. Results
EEG signals with varying values of SNRs present that higher values of SNR and lower values of MAE are observed for the presented work. It indicates EMD-DFAWPD is a better denoising algorithm. Table 2 concludes the SNR and MAE values of the techniques for different SNR levels of white Gaussian noise denoted as σ.
Table 2.
Comparing the proposed denoising techniques with the conventional respecting SNR and MAE corresponding to different levels of white Gaussian noise
Variables | Denoising Method | σ= 0 dB | σ= 5 dB | σ= 10 dB | σ= 15 dB | σ=20 dB |
---|---|---|---|---|---|---|
SNR | DWT | 4.20 | 5.1492 | 10.18 | 15.82 | 20.12 |
EMD only | 4.21 | 5.03 | 12.32 | 15.34 | 21.29 | |
EMD- DWT | 1.23 | 5.944 | 12.04 | 16.582 | 22.583 | |
EMD-DFA- WPD | 20.24 | 18.54 | 18.04 | 16.29 | 17.06 | |
MAE | DWT | 13.7209 | 45.492 | 101.18 | 115.82 | 202.12 |
EMD only | 13.6731 | 45.03 | 101.32 | 115.34 | 202.29 | |
EMD- DWT | 13.23 | 44.944 | 101.04 | 116.582 | 202.583 | |
EMD-DFA- WPD | 12.24 | 44.04 | 100.04 | 115.89 | 203.06 |
The obtained results indicated improved signal-to-noise ratio and lower values of MAE for the combined EMD-DFA-WPD technique, compared to EMD, DWT, and EMD with DWT technique (Figures 3 & 4). EMD performs better than wavelet technique for lower SNR levels; however, EMD-DFA-WPD is providing higher SNR and lowest MAE than all the conventional techniques although its performance is better for lower levels of white Gaussian noise.
Figure 3.
The variation of SNR at the output corresponding to different levels of white Gaussian noise
Figure 4.
The variation of MAE at the output corresponding to different levels of white Gaussian noise
Furthermore, EMD is applied along with DFA with a value of Hurst exponent H for white Gaussian noise. The value of the Hurst exponent is defined accordingly and adjusted for analysis. The parameter α known as the scaling exponent represents the roughness of the series. Higher values of α represent smooth time series i.e., slow fluctuations (Mert & Akan, 2014). EMD based denoising requires a reliable threshold to determine which oscillations called intrinsic mode functions (IMFs. DFA slope α=0.5, α=1.0, and α=1.5 depending upon the type of noise to be white Gaussian noise, pink or Brownian noise respectively. To cope-up with the problem of mode-mixing, the value of the scaling exponent was set to be 0.75. The value of the Hurst exponent varies as 10.5, 1.0, and 1.5 for white Gaussian noise, pink, and Brownian noise. Figure 5 plots the performance of the proposed algorithm for different values of H demonstrating better performance at 0.37.
Figure 5.
The variation of SNR at different values of H
Table 3 lists the statistical analysis data to assess the performance of the proposed algorithm. The RM-ANOVA results revealed that the proposed algorithm outperforms at P<0.001. Additionally, t-test analysis provided the comparison of parameters where p1 represents EMDDFA-WPD vs. DWT, p2 represents EMD-DFA-WPD vs. EMD, p3 represents EMD-DFA-WPD vs. EMD-DWT. It concludes that the algorithm performs best compared to other algorithms. Better classification results are obtained for the proposed methodology. RF and SVM classifiers were used to assess the accuracy (Table 4). The best accuracy of 98.51% is achieved for RF and 98.07% for SVM for EMD-DFA-WPD than other approaches. EMD-DWT gives 98.01% and 95.81% accuracy values for RF and SVM. Additionally, the best F1 score values were observed for the proposed technique compared to the other conventional approaches. Moreover, the classification performance for both the classifiers was compared with and without denoising to highlight the effectiveness of the proposed technique (Table 5).
Table 3.
The statistical analysis of SNR, MAE between the reconstructed and original signal between different approaches
EEG Spectral Measure | ANOVA | t-test | ||
---|---|---|---|---|
| ||||
P | P1 | P2 | P3 | |
SNR | 0.03 | <0.001 | <0.001 | <0.001 |
MAE | 0.01 | <0.001 | <0.001 | <0.001 |
Table 4.
Accuracy evaluation of different denoising techniques using random forest and SVM based classification
Techniques | Random Forest | SVM | ||
---|---|---|---|---|
| ||||
F1 Score | Accuracy (%) | F1 Score | Accuracy (%) | |
DWT | 94.29 | 97.8 | 93.9 | 94.09 |
WPT | 92.7 | 96.7 | 90.1 | 94.7 |
EMD- DFA | 91.07 | 98.0 | 90.83 | 97.21 |
EMD-DWT | 96.83 | 98.01 | 93.89 | 95.81 |
EMD-DFA-WPD | 97.81 | 98.51 | 94.37 | 98.07 |
Table 5.
Classification performance, compared with and without Denoising using Random Forest and SVM-based classification
Classifier Performance | Without Denoising (%) | Denoising Using Proposed Technique (%) |
---|---|---|
RF | 96.98 | 98.51 |
SVM | 94.83 | 98.07 |
Although the proposed method yields better suppression of artifacts respecting the evaluation parameters of SNR for different values of white Gaussian noise added at different values of noise along with better classification results. Besides, more efficient algorithms can be developed by increasing the levels of decompositions. Furthermore, EMD lacks mathematical formulation and the mode-mixing problem. As a solution, various newer techniques were offered. Therefore, this analysis can be extended for analyzing other levels of decompositions. Moreover, more efficient algorithms than EMD, such as MEMD, EEMD, and VMD (Aneesh, Kumar, Hisham, & Soman, 2015; Kærgaard, Jensen, & Puthusserypady, 2016; Liu, Cao, & Wang, 2017; Molla et al., 2012). We propose to utilize recently developed a multivariate extension of Empirical Mode Decomposition (EMD can be used for further analysis to eliminate the mode mixing problem.
4. Discussion
There is a need to separate raw EEG signals from various noise sources using an appropriate artifact removal algorithm, leading to minimal neural information loss. Furthermore, there is insufficient evidence of denoising systems for EEG signals of depression patients. We addressed an approach for suppressing artifacts that imposes a challenge to the common preprocessing techniques in EEG processing systems corresponding to depression patients. The present study aimed to develop a reliable EEG preprocessing phase of removing the noise present in EEG signals of depression patients. The removal of these most common noise sources is critical to improving the performance of the EEG-based diagnosing systems for depression. EMD is gaining great success in the field of signal processing. In previous studies using EMD inspired by wavelet transforms, ignoring various IMFs after wavelet-based thresholding could lead to ignoring information carrying the capacity of IMFs leading to inefficient denoising results. In this paper, a denoising model was proposed for EEG signals using hybrid technique EMD and WPD, where the IMF selection criteria in EMD are identified by the DFA algorithm. Unlike the conventional EMD-based EEG denoising approaches that neglect multiple IMFs containing noise as well as neural information, we proposed to perform a windowing in the EMD domain to reduce the noise from a few IMFs, yielding a comparatively cleaner EEG signal. Compared to other conventional methodologies, the proposed method provides better SNR.
5. Conclusion
A new classification method based on EMD and wave-let packet transform was used. To assess the performance of the proposed algorithm, depression patients and healthy individuals were classified using SVM and Random Forest. Better accuracy is observed for the observed technique than the other approaches. In the future, more efficient algorithms can be developed by increasing the levels of decompositions and considering other partially variational algorithms to decrease the problem of mode mixing by EMD.
Ethical Considerations
Compliance with ethical guidelines
There were no ethical considerations to be considered in this research.
Footnotes
Funding
This research did not receive any grant from funding agencies in the public, commercial, or non-profit sectors.
Authors’ contributions
All authors equally contributed to preparing this article.
Conflict of interest
The authors declared no conflicts of interest.
References
- Acharya U. R., Sudarshan V. K., Adeli H., Santhosh J., Koh J. E. W., Puthankatti S. D., et al. (2015). A novel depression diagnosis index using nonlinear features in EEG signals. European Neurology, 74(1–2), 79–83. [DOI: 10.1159/000438457] [DOI] [PubMed] [Google Scholar]
- Acharya U. R., Sudarshan V. K., Adeli H., Santhosh J., Koh J. E. W., Adeli A. (2015). Computer-aided diagnosis of depression using EEG signals. European Neurology, 73(5–6), 329–36. [DOI: 10.1159/000381950] [DOI] [PubMed] [Google Scholar]
- Alfoouri M., Daqrouq Kh. (2008). ECG signal denoising by Wavelet transform thresholding. American Journal of Applied Science, 5(3), 276–81. https://thescipub.com/pdf/ajassp.2008.276.281.pdf [Google Scholar]
- Aneesh C., Kumar S., Hisham P. M., Soman K. P. (2015). Performance comparison of variational mode decomposition over empirical wavelet transform for the classification of power quality disturbances using support vector machine. Procedia Computer Science, 46, 372–80. [DOI: 10.1016/j.procs.2015.02.033] [DOI] [Google Scholar]
- Awal M. A., Mostafa S. S., Ahmad M., Rashid M. A. (2014). An adaptive level dependent wavelet thresholding for ECG denoising. Biocybernetics and Biomedical Engineering, 34(4), 238–49. [DOI: 10.1016/j.bbe.2014.03.002] [DOI] [Google Scholar]
- Bartoli F., Cerutti S. (1983). An optimal linear filter for the reduction of noise superimposed to the EEG signal. Journal of Biomedical Engineering, 5(4), 274–80. [DOI: 10.1016/0141-5425(83)90001-8] [DOI] [PubMed] [Google Scholar]
- Barua Sh., Begum Sh. (2014). A review on machine learning algorithms in handling EEG artifacts. Paper presented at The Swedish AI Society (SAIS) Workshop SAIS, 14, Stockholm, Sweden, 22–23 May 2014. https://www.diva-portal.org/smash/record.jsf?pid=diva2%3A759967&dswid=2641 [Google Scholar]
- Bono V., Jamal W., Das S., Maharatna K. (2014). Artifact reduction in multichannel pervasive EEG using hybrid WPTICA and WPT-EMD signal decomposition techniques. Paper presented at 2014 IEEE International Conference on Acoustics, Speech and Signal Processing (ICASSP), Florence, Italy, 4–9 May 2014. [DOI: 10.1109/ICASSP.2014.6854728] [DOI] [Google Scholar]
- Cai H., Han J., Chen Y., Sha X., Wang Z., Hu B., et al. (2018). A pervasive approach to EEG-based depression detection. Complexity, 2018, 5238028. [DOI: 10.1155/2018/5238028] [DOI] [Google Scholar]
- Castellanos N. P., Makarov V. A. (2006). Recovering EEG brain signals: Artifact suppression with wavelet enhanced independent component analysis. Journal of Neuroscience Methods, 158(2), 300–12. [DOI: 10.1016/j.jneumeth.2006.05.033] [DOI] [PubMed] [Google Scholar]
- Cluitmans P. J. M., Van De Velde M. (2000). Outlier detection to identify artefacts in EEG signals. Paper presented at Proceedings of the 22nd Annual International Conference of the IEEE Engineering in Medicine and Biology Society (Cat. No.00CH37143), Chicago, IL, USA, 23–28 July 2000. [DOI: 10.1109/iembs.2000.901453] [DOI] [Google Scholar]
- Croft R. J., Barry R. J. (2000). Removal of ocular artifact from the EEG: A review. Neurophysiologie Clinique/Clinical Neurophysiology, 30(1), 5–19. [DOI: 10.1016/S0987-7053(00)00055-1] [DOI] [PubMed] [Google Scholar]
- Dimatteo I., Genovese C. R., Kass R. E. (2001). Bayesian curve-fitting with free-knot splines. Biometrika, 88(4), 1055–71. [DOI: 10.1093/biomet/88.4.1055] [DOI] [Google Scholar]
- Ghandeharion H., Erfanian A. (2010). A fully automatic ocular artifact suppression from EEG data using higher order statistics: Improved performance by wavelet analysis. Medical Engineering & Physics, 32(7), 720–9. [DOI: 10.1016/j.medengphy.2010.04.010] [DOI] [PubMed] [Google Scholar]
- Hoffmann S., Falkenstein M. (2008). The correction of eye blink artefacts in the EEG: A comparison of two prominent methods. PLoS One, 3(8), e3004. [DOI: 10.1371/journal.pone.0003004] [DOI] [PMC free article] [PubMed] [Google Scholar]
- Kabir M. A., Shahnaz C. (2012). Denoising of ECG signals based on noise reduction algorithms in EMD and wavelet domains. Biomedical Signal Processing and Control, 7(5), 481–9. [DOI: 10.1016/j.bspc.2011.11.003] [DOI] [Google Scholar]
- Kærgaard K., Jensen S. H., Puthusserypady S. (2016). A comprehensive performance analysis of EEMD-BLMS and DWT-NN hybrid algorithms for ECG denoising. Biomedical Signal Processing and Control, 25, 178–87. [DOI: 10.1016/j.bspc.2015.11.012] [DOI] [Google Scholar]
- Kaur Ch., Singh P. (2015). EEG derived neuronal dynamics during meditation: Progress and challenges. Advances in Preventive Medicine, 2015, 614723. [DOI: 10.1155/2015/614723] [DOI] [PMC free article] [PubMed] [Google Scholar]
- Kaur Ch., Singh P. (2016). EEG artifact suppression based on SOBI based ICA using wavelet thresholding. Paper presented at 2015 2nd International Conference on Recent Advances in Engineering & Computational Sciences (RAECS), Chandigarh, India, 21–22 December 2015. [DOI: 10.1109/RAECS.2015.7453319] [DOI] [Google Scholar]
- Kiamini M., Alirezaee Sh., Perseh B., Ahmadi M. (2009). Elimination of ocular artifacts from EEG signals using the wavelet transform and empirical mode decomposition. Paper presented at 2009 6th International Conference on Electrical Engineering/Electronics, Computer, Telecommunications and Information Technology, Chonburi, Thailand, 6–9 May 2009. [DOI: 10.1109/ecticon.2009.5137235] [DOI] [Google Scholar]
- Kierkels J. J. M., van Boxtel G. J. M., Vogten L. L. M. (2006). A model-based objective evaluation of eye movement correction in EEG recordings. IEEE Transactions on Biomedical Engineering, 53(2), 246–53. [DOI: 10.1109/TBME.2005.862533] [DOI] [PubMed] [Google Scholar]
- Krupa B. N., Mohd Ali M. A., Zahedi E. (2009). The application of empirical mode decomposition for the enhancement of cardiotocograph signals. Physiological Measurement, 30(8), 729–43. [DOI: 10.1088/0967-3334/30/8/001] [DOI] [PubMed] [Google Scholar]
- Lahmiri S., Boukadoum M. (2015). Physiological signal denoising with variational mode decomposition and weighted reconstruction after DWT thresholding. Paper presented at 2015 IEEE International Symposium on Circuits and Systems (ISCAS), Lisbon, Portugal, 24–27 May 2015. [DOI: 10.1109/ISCAS.2015.7168756] [DOI] [Google Scholar]
- Liu W., Cao S., Wang Zh. (2017). Retracted: Application of variational mode decomposition to seismic random noise reduction. Journal of Geophysics and Engineering, 14(4), 888–99. [DOI: 10.1088/1742-2140/aa6b28] [DOI] [Google Scholar]
- Makeig S., Bell. A. J., Jung T. P., Sejnowski T. J. (1996). Independent component analysis of electroencephalographic data. In Touretzky D. S., Mozer M. C., Hasselmo M. E. (Eds.), Advances in neural information processing systems 8: Proceedings of the 1995 conference (pp. 145–51). Cambridge: MIT Press. https://books.google.com/books?id=ZkJrSots_SAC&lr=&source=gbs_navlinks_s [Google Scholar]
- Mert A., Akan A. (2014). Detrended fluctuation thresholding for empirical mode decomposition based denoising. Digital Signal Processing, 32, 48–56. [DOI: 10.1016/j.dsp.2014.06.006] [DOI] [Google Scholar]
- Molla M. K. I., Tanaka T., Rutkowski T. M. (2012). Multivariate EMD based approach to EOG artifacts separation from EEG. Paper presented at 2012 IEEE International Conference on Acoustics, Speech and Signal Processing (ICASSP), Kyoto, Japan, 25–30 March 2012. [DOI: 10.1109/ICASSP.2012.6287968] [DOI] [Google Scholar]
- Mumtaz W., Xia L., Mohd Yasin M. A., Azhar Ali S. S., Malik A. S. (2017). A wavelet-based technique to predict treatment outcome for Major Depressive Disorder. PLoS One, 12(2), e0171409. [DOI: 10.1371/journal.pone.0171409] [DOI] [PMC free article] [PubMed] [Google Scholar]
- Munia T. T. K., Haider A., Schneider Ch., Romanick M., Fazel-Rezai R. (2017). A novel EEG based spectral analysis of persistent brain function alteration in athletes with concussion history. Scientific Reports, 7, 17221. [DOI: 10.1038/s41598-017-17414-x] [DOI] [PMC free article] [PubMed] [Google Scholar]
- Nazarpour K., Wongsawat Y., Sanei S., Chambers J. A., Oraintara S. (2008). Removal of the eye-blink artifacts from EEGs via STF-TS modeling and robust minimum variance beamforming. IEEE Transactions on Biomedical Engineering, 55(9), 2221–31. [DOI: 10.1109/TBME.2008.919847] [DOI] [PubMed] [Google Scholar]
- Phlypo R., Boon P., D’Asseler Y., Lemahieu I. (2007). Removing ocular movement artefacts by a joint smoothened subspace estimator. Computational Intelligence and Neuroscience, 2007, 075079. [DOI: 10.1155/2007/75079] [DOI] [PMC free article] [PubMed] [Google Scholar]
- Safieddine D., Kachenoura A., Albera L., Birot G., Karfoul A., Pasnicu A., et al. (2012). Removal of muscle artifact from EEG data: Comparison between stochastic (ICA and CCA) and deterministic (EMD and wavelet-based) approaches. EURASIP Journal on Advances in Signal Processing, 2012, 127. [DOI: 10.1186/1687-6180-2012-127] [DOI] [Google Scholar]
- Salis C. I., Malissovas A. E., Bizopoulos P. A., Tzallas A. T., Angelidis P. A., Tsalikakis D. G. (2013). Denoising simulated EEG signals: A comparative study of EMD, wavelet transform and Kalman filter. Paper presented at 13th IEEE International Conference on BioInformatics and BioEngineering, Chania, Greece, 10–13 November 2013. [DOI: 10.1109/BIBE.2013.6701613] [DOI] [Google Scholar]
- Semlitsch H. V., Anderer P., Schuster P., Presslich O. (1986). A solution for reliable and valid reduction of ocular artifacts, applied to the P300 ERP. Psychophysiology, 23(6), 695–703. [DOI: 10.1111/j.1469-8986.1986.tb00696.x] [DOI] [PubMed] [Google Scholar]
- Sharma M., Achuth P. V., Deb D., Puthankattil S. D., Acharya U. R. (2018). An automated diagnosis of depression using three-channel bandwidth-duration localized wavelet filter bank with EEG signals. Cognitive Systems Research, 52, 508–20. [DOI: 10.1016/j.cogsys.2018.07.010] [DOI] [Google Scholar]
- Shoker L., Sanei S., Chambers J. (2005). Artifact removal from electroencephalograms using a hybrid BSS-SVM algorithm. IEEE Signal Processing Letters, 12(10), 721–4. [DOI: 10.1109/LSP.2005.855539] [DOI] [Google Scholar]
- Suchetha M., Kumaravel N. (2013). Empirical mode decomposition based filtering techniques for power line interference reduction in electrocardiogram using various adaptive structures and subtraction methods. Biomedical Signal Processing and Control, 8(6), 575–85. [DOI: 10.1016/J.BSPC.2013.05.001] [DOI] [Google Scholar]
- Vaid S., Singh P., Kaur Ch. (2015). Classification of human emotions using multiwavelet transform based features and random forest technique. Indian Journal of Science and Technology, 8(28), 1–7. [DOI: 10.17485/ijst/2015/v8i28/70797] [DOI] [Google Scholar]
- Vaid S., Singh P., Kaur Ch. (2015). EEG signal analysis for BCI interface: A review. Paper presented at 2015 Fifth International Conference on Advanced Computing & Communication Technologies, Haryana, India, 21–22 February 2015. [DOI: 10.1109/ACCT.2015.72] [DOI] [Google Scholar]
- Wallstrom G. L., Kass R. E., Miller A., Cohn J. F., Fox N. A. (2004). Automatic correction of ocular artifacts in the EEG: A comparison of regression-based and component-based methods. International Journal of Psychophysiology, 53(2), 105–19. [DOI: 10.1016/j.ijpsycho.2004.03.007] [DOI] [PubMed] [Google Scholar]
- Wu Zh., Huang N. E. (2009). Ensemble empirical mode decomposition: A noise-assisted data analysis method. Advances in Adaptive Data Analysis, 01(01), 1–41. [DOI: 10.1142/S1793536909000047] [DOI] [Google Scholar]
- Zeng H., Song A., Yan R., Qin H. (2013). EOG artifact correction from EEG recording using stationary subspace analysis and empirical mode decomposition. Sensors, 13(11), 14839–59. [DOI: 10.3390/s131114839] [DOI] [PMC free article] [PubMed] [Google Scholar]