Abstract
Back-illuminated charged-coupled device (BI-CCD) arrays increase quantum efficiency but also amplify etaloning, a multiplicative, wavelength-dependent fixed-pattern effect. When spectral data from hundreds of BI-CCD rows are combined, the averaged spectrum will generally appear etalon-free. This can mask substantial etaloning at the row level, even if the BI-CCD has been treated to suppress the effect. This Note compares two methods of etalon correction, one with simple averaging and one with row-by-row calibration using a fluorescence standard. Two BI-CCD arrays, both roughened by the supplier to reduce etaloning, were used to acquire Raman spectra of murine bone specimens. For one array, etaloning was the dominant source of noise under the exposure conditions chosen, even for the averaged spectrum across all rows; near-infrared-excited Raman peaks were noticeably affected. In this case, row-by-row calibration improved the spectral quality of the average spectrum. The other CCD’s performance was shot-noise limited and therefore received no benefit from the extra calibration. The different results highlight the importance of checking for and correcting row-level fixed pattern when measuring weak Raman signals in the presence of a large fluorescence background.
Keywords: fixed pattern, etalon effect, calibration, Raman spectroscopy
Introduction
Raman spectroscopy is a well-established inelastic scattering technique that produces a molecular “fingerprint” of vibrational modes present in a sample, including relative abundances. The technique can be performed upon optically thick specimens in a noncontact, backscattering geometry, requiring minimal sample preparation compared to Fourier transform infrared spectroscopy.1 In the biomedical sciences, Raman spectroscopy has been used to study various diseases and cancers in many different tissues.2–5 Such work typically uses near-infrared (NIR) excitation in the 780–840 nm range in order to increase the Raman/autofluorescence ratio.
Current slit-imaging Raman spectrometers commonly disperse light onto two-dimensional charged-coupled device (CCD) arrays. Back-illuminated (BI) CCDs have higher quantum efficiency (QE) than front-illuminated CCDs because there is no obscuration from the front surface’s readout wires.6 Because the BI approach requires the photosensitive region to be thinned to the scale of the optical absorption length, some photons pass through the whole region and reflect back, creating interference that manifests as a sinusoidal oscillation in QE versus wavelength.7,8 This QE variation, called the “etalon effect”, affects all light, both the Raman scattering and the usually much stronger fluorescence background. The etalon effect becomes problematic in Raman spectroscopy when chipinduced modulation in the fluorescence amplitude becomes comparable to true Raman peaks’ amplitudes. This problem is larger in the near-IR because the detector’s absorption coefficient is lower, allowing more light to back-reflect.
To reduce the etalon effect, CCDs for imaging spectrometers are often surface-roughened to decorrelate the QE values between adjacent pixels in both the horizontal (wavelength) and vertical (slit-height) directions. This disrupts purely sinusoidal variations but does not change the standard deviation of QE fluctuations. The standard deviation can be reduced by suppressing back-reflection, either by increasing the depletion region thickness9 or by depositing an wavelength-dependent anti-reflective (AR) coating across the CCD.10 These options are rarely available commercially.
Rather than modifying the CCD physically, one can reduce etalon effects computationally by comparing multiple spectra acquired in different grating positions11–13 or at different laser excitation wavelengths.14,15 These approaches model the etalon contribution as additive and unchanged by shifting, such that subtraction will eliminate it but allow the Raman spectrum to be recovered. This is an approximate model; the etalon effect upon the spectrum is multiplicative, and shifting the wavelengths delivered to individual pixels (as in the grating shift approach) should alter the fixed pattern.
In many applications the slit’s height dimension encodes no relevant information. One such application, single-point Raman spectroscopy of ex vivo bone specimens, will be used here as an example. A laser illuminates a ~ 1 mm spot on the bone; the resulting Raman emission’s etendue is enough to fill a slit and CCD array both a few mm in height. The goal is to reduce the full 2D data matrix into a single spectrum representative of the optically interrogated bone region.
This Note compares two methods of mitigating fixed pattern in such situations. The first method simply sums the spectral rows (“vertical binning”). The other method employs row-by-row fixed-pattern correction, using a fluorescence standard as a pattern-free spectral reference. This approach attempts to eliminate the fixed pattern effect rather than relying solely upon noise-averaging.
Methods
Instrument
The Raman spectroscopy data were collected on a system described previously.16,17 Briefly, an 830-nm laser delivered 150 mW to the sample in a 1 mm diameter spot. The scattered light was imaged onto a circular, fixed bundle of multimode fibers (100 micron core diameter, 0.27 NA). Light delivery and collection were performed in a non-contact manner. The proximal end of the bundle rearranged the fibers linearly at the entrance of a f/1.8 imaging spectrograph with its slit removed (Kaiser HoloSpec). Two deep-depletion, back-thinned, fringe-suppressed CCDs were tested. “CCD1” is an Andor Model DU 420-BX-DD, antireflection-coated for 750–1000 nm, purchased in 2001. “CCD2” is an Andor Model DU-420-BEX2-DD, dual AR coating for UV-VIS-NIR broadband spectroscopy, purchased in 2018. The height of the detectors permitted spectral imaging of 40 fibers from the bundle.
Calibration
The spectra were pre-processed for readout and dark noise and calibrated for wavelength and Raman shift using a neon lamp and Tylenol, respectively, as described previously.18 Imaging distortions due to spectrograph optics and CCD alignment were corrected using a method described by Esmonde-White et al.19 830 nm laser excitation of a metal-ion-doped glass (Standard Reference Material 2241, National Institute of Standards and Technology [NIST]) designed for 785 nm provided spectrally smooth fluorescence using without Raman peaks. This enabled characterization of pixel-level fixed pattern effects, as described below. Each spectral row, n, of the resulting glass fluorescence data gn(λ) was divided by the NIST-supplied fluorescence emission spectrum f(λ) to produce a spectral throughput function tn(λ) in the form
| (1) |
This forced the rendered glass spectra from each row to have the same spectral lineshape. The deviation from absolute throughput correction due to the wavelength of the laser was immaterial to the subsequent analysis.
Using this terminology, Method I (fixed pattern suppressed solely by averaging) produces the output spectrum
| (2) |
where P denotes post-processing [subtraction of an estimated fluorescence background via continuous wavelet transform20 followed by Savitzky-Golay smoothing over a 3 pixel window21 to match the spectral resolution of 6 cm–1] and denotes the average of the tn(λ) spectral throughput functions. Similarly, Method II (fixed pattern corrected row by row) produces the output spectrum
| (3) |
Note that in Eq. 3 each row spectrum is corrected for its particular fixed pattern, unlike in Eq. 2 where all rows are corrected using the same function . As an optional intermediate step, signals from individual optical fibers were sometimes calculated. For qualitative analysis of fixed pattern variation across the CCD array, ten adjacent rows of an unprocessed spectral image were simply averaged at each of three locations (see Figure 2). For quantitative estimates of each fiber’s spectrum, a Gaussian-Lorentzian decomposition described by Dooley et al.22) corrected for spatial overlap between adjacent fibers’ images (see Figures 3 and 4).
Figure 2.

(A) Image collected from metal-ion-doped glass on CCD 2. Multiple adjacent rows are summed to create a spectrum from an optical fiber. The red boxes outline non-neighboring fibers that were used to characterize fixed pattern. (B) Residuals after subtracting a smoothed fit to three non-neighboring fiber spectra showing the fixed pattern is not uniform throughout the height of the CCD. (C) Zoomed in version of residuals shown in B with grey boxes highlighting a few locations showing the fixed pattern is not uniform.
Figure 3.

Data collected with CCD2 and calibrated using Method I (A,C,F,H) and Method II (B,D,E,G). (A-B) Full spectra (Raman plus fluorescence) from individual fibers. The etalon effect is seen for Method I but not for Method II. (C-D) Polynomial subtracted and normalized (to mean absolute deviation) spectra using (C) Method I and (D) Method II. (E-F) Final output spectra (E) sII(λ) and (F) sI(λ). The sI(λ)spectrum contains a residual amount of the etalon effect. (G-H) Zoomed in region demonstrating the distortion of minor Raman peaks from bone in sI(λ) due to uncorrected fixed pattern.
Figure 4.

Data collected with CCD1 and calibrated using Method I (A,C,F) and Method II (B,D,E). (A-B) Calibrated total-signal spectra (Raman plus fluorescence) from individual fibers using (A) Method I and (B) Method II. (C-D) Same spectra, polynomial-subtracted and normalized. (E-F) Final output spectra (E) sII(λ) and (F) sI(λ), with neither spectrum exhibiting the oscillatory component present in Figure 3F.
Specimens
The experimental data presented here were acquired in previous studies.16,23 Briefly, a tibia16 and femur23 were excised from C57BL/6J mice between the age of 4–24 weeks. Exposed bone specimens were stored in phosphate buffered saline and frozen before performing Raman spectroscopy.
Results
Figure 1 compares unprocessed fluorescence measurements of the same calibration glass using the two different CCD arrays. The summed spectra over all rows (panels A and B) exhibit smooth lineshapes without visible etalon effect artifacts for either array. Ripples characteristic of the etalon effect are evident in three adjacent row spectra from CCD2 (panel D); such ripples are absent for CCD1. High-pass-filtered spectra from two adjacent rows of CCD2 (panel F) share a common ripple effect from the etalon effect but have row-specific variations at the pixel level due to both fixed pattern and shot noise. Oscillations are not seen for pixels 1–400 because the spatial period is less than two pixels. Analogous spectra from CCD1 exhibit lower amplitude variations with no correlation between rows; successive exposures of the same row also indicated that fixed pattern was not the dominant noise source (data not shown).
Figure 1.

Comparisons of unprocessed metal-ion-doped glass data measured with CCD1 (A,C,E) and CCD2 (B,D,F) normalized to their maxima. Wavelength increases linearly with pixel number. (A,B) Sum of all rows of spectral data collected on CCD1 (A) and CCD2 (B). Spectra look similar with no visible fixed pattern. (C,D) Neighboring individual rows from CCD1 (C) and CCD2 (D) show a more distinct etalon effect present with data collected on CCD2. (E,F) High-pass filtered versions of two spectra from each of C,D. Traces are offset for clarity; all amplitudes of fluctuation are plotted on the same scale for direct comparison.
Figure 2A shows the uncalibrated CCD2 spectral image of glass fluorescence from which Figures 1B/D/F were calculated, with each horizontal stripe corresponding to one optical fiber’s output. Data from optical fibers at three different array heights (marked by red boxes) were vertically binned as noted previously and then high-pass filtered, producing the spectra shown in Fig.2B. Oscillatory patterns similar to Figure 1F are seen; the spectra are smoother because multiple rows are summed, averaging individual row’s pixel-level variations. Grey bars in Figure 2C emphasize that etalon oscillations are dephased between fibers at different array heights. This explains why etalon effects wash out in a simple sum over all fibers’ contributions (c.f. Figure 1B).
Methods I and II were applied to Raman spectroscopy of an ex vivo murine bone specimen collected using CCD2. Figure 3A–B shows total spectra (Raman plus fluorescence) from individual fibers; panels C-D are the corresponding background-subtracted (“Raman only”) versions. As expected, the Method I fiber spectra exhibit greater oscillatory patterns, due to the lack of local fixed pattern correction. Adding all fibers’ contributions together via the two methods produced the background-corrected spectra E and F in Figure 3; these are sI(λ) and sI(λ) from Eqs. 2 and 3. The main features of sII(λ) are those expected of a typical bone spectrum.24 sII(λ), by contrast, contains a noticeable oscillatory component throughout the spectrum. Plot H provides a magnified view of this additional noise in sI(λ) distorting the lower-amplitude Raman peaks from the bone.
Figure 4 shows analogous bone spectra measured from CCD1. In this instance the sII(λ) spectrum provided by Method II provided no obvious improvement over sI(λ).
Discussion
Back-illuminated CCDs for spectroscopy provide higher QE with the downside of greater QE variation due to etaloning. When hardware or software is set to bin over many spectral rows, the residual effect of etaloning can be impossible to perceive in a plot, as in the Method I spectrum of Figure 1A. A user might conclude that two CCDs have equivalent levels of fringe suppression.
In the case of Raman peaks riding on a fluorescence background, this metric for equivalency can be insufficient. Two CCDs, marketed similarly by a major supplier, exhibited qualitatively different levels of the etalon effect. Notably, the one manufactured more recently (CCD2) had the larger levels. In this instance, the fixed pattern from the etalon effect was readily visible in single rows (c.f. Figure 1) and contributed significant noise to the vertically binned spectrum sI(λ). Row-by-row calibration of the fixed pattern was required to render lower-amplitude Raman peaks more clearly (c.f. Figure 3). This indicates the importance of correcting fixed pattern when measuring Raman features in the presence of large fluorescence backgrounds, even when binning over the height of a slit.
On the other hand, CCD1 did not exhibit fixed pattern at the row level for the same experimental settings. There was therefore no expected benefit to using row-level calibration (c.f. Figure 4). In fact, if the noise amplitude had been a greater fraction of the total signal, Method II should have been noticeably detrimental, as it imprinted shot noise from the glass measurements onto the bone spectra. In this particular study, the signal to noise ratio was high enough that the two methods produced approximately equivalent results.
This work used a translucent glass calibration sample and an opaque, highly scattering bone specimen. The spatial distributions of light emission from these two targets’s surfaces were unequal, leading to different relative amounts of light collected by different optical fibers in the two cases. How to sum the different fiber’s contributions into a final spectrum is the user’s choice and depends upon the application. We note that illumination of a diffusive target with a spectrally calibrated broadband light source would work equally well for the calibration step.
Acknowledgements
The study was supported by grant number R01AR070613 from NIAMS/NIH. The content is solely the responsibility of the authors and does not necessarily represent the official views of the NIH.
Footnotes
Conflicts of Interest
The authors declare no potential conflicts of interest.
References
- 1.Moore DS, Sterling White J, and Allen Harbin B. ”Infrared sample preparation and interpretation using a knowledge based system”. Anal. Chimica Acta 1994. 294(1): 85–94. [Google Scholar]
- 2.Cordero E, Latka I, Matthäus C, Schie IW, and Popp J. ”In-vivo Raman spectroscopy: from basics to applications”. J. Biomed. Opt 2018. 23(7): 1–23. [DOI] [PubMed] [Google Scholar]
- 3.Cui S, Zhang S, and Yue S. ”Raman Spectroscopy and Imaging for Cancer Diagnosis”. J. Healthc. Eng. 2018 2018. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 4.Choo-Smith LP, Edwards HGM, Endtz HP, Kros JM, Heule F, Barr H, Robinson JS Jr., Bruining HA, Puppels GJ. ”Medical applications of Raman spectroscopy: From proof of principle to clinical implementation”. Biopolymers. 2002. 67(1): 1:9. [DOI] [PubMed] [Google Scholar]
- 5.Wen X, Ou Y, Bogatcheva G, Thomas G, Mahadevan-Jansen A, Singh B, Lin EC, and Bardhan R. ”Probing metabolic alterations in breast cancer in response to molecular inhibitors with Raman spectroscopy and validated with mass spectrometry”. Chem. Sci 2020. 11(36): 9863–9874. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 6.Naletto G, Pace E, Tondello G, and Boscolo A. ”Performance of a thinned back-illuminated ion-implanted CCD as detector for a normal incidence EUV spectrograph”. Meas. Sci. Technol 1994. 5(12): 1491–1500. [Google Scholar]
- 7.Ma L, Wei J, and Huang X. ”Research on and correction of interference fringes phenomenon in dispersive hyperspectral imaging spectrometer using back-illuminated CCDs in near-infrared band”. Spectroscopy and Spectral Analysis. 2014. 34(7). [PubMed] [Google Scholar]
- 8.Hu B, Zhang J, Cao K, Hao S, Sun D, and Liu Y. ”Research on the Etalon Effect in Dispersive Hyperspectral VNIR Imagers using Back-Illuminated CCDs”. IEEE Trans. Geosci. Electron 2018. 56(9): 5481–5494. [Google Scholar]
- 9.Groom DE, Holland SE, Levi ME, Palaio NP, Perlmutter S, Stover RJ, and Wei M. ”Quantum efficiency of a back-illuminated CCD imager: an optical approach”. Proc. SPIE 3649, Sensors, Cameras, and Systems for Scientific/Industrial Applications. 1999. 3649: 80–90. [Google Scholar]
- 10.Kelt A, Harris A, Jorden P, and Tulloch S. ”Optimised CCD Antireflection Coating”. In: Beletic JE, Beletic JW, Amico P editors. Scientific detectors for astronomy 2005. Netherlands: Springer, Dordrecht. 2005. 369–374. [Google Scholar]
- 11.Bell SEJ, Bourguignon ESO, and Dennis A. ”Analysis of luminescent samples using subtracted shifted Raman spectroscopy”. Anal. 1998. 123(8): 1729–1734. [Google Scholar]
- 12.O’Grady A, Dennis AC, Denvir D, McGarvey JJ, and Bell SEJ. ”Quantitative Raman Spectroscopy of Highly Fluorescent Samples Using Pseudosecond Derivatives and Multivariate Analysis”. Anal. Chem 2021. 73(9): 2058–2065. [DOI] [PubMed] [Google Scholar]
- 13.Auguié B, Reigue A, Ru ECL, and Etchegoin PG. ”Tiny Peaks vs Mega Backgrounds: A General Spectroscopic Method with Applications in Resonant Raman Scattering and Atmospheric Absorptions”. Anal. Chem 2012. 84(18): 7938–7945. [DOI] [PubMed] [Google Scholar]
- 14.Cooper JB, Abdelkader M, and Wise KL. ”Sequentially Shifted Excitation Raman Spectroscopy: Novel Algorithm and Instrumentation for Fluorescence-Free Raman Spectroscopy in Spectral Space”. Appl. Spectrosc 2013. 67(8): 973:984. [DOI] [PubMed] [Google Scholar]
- 15.Kloz M, Weißenborn J, Polívka T, Frank HA, and Kennis JTM. ”Spectral watermarking in femtosecond stimulated Raman spectroscopy: resolving the nature of the carotenoid S* state”. Phys. Chem. Chem. Phys 2016. 18(21): 14619–14628. [DOI] [PubMed] [Google Scholar]
- 16.Shu C, Chen K, Lynch M, Maher JR, Awad HA, and Berger AJ. ”Spatially offset Raman spectroscpy for in vivo bone strength prediction”. Biomed. Opt. Express 2018. 9(10): 4781–4791. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 17.Feng G, Ochoa M, Maher JR, Awad HA, and Berger AJ.”Sensitivity of spatially offset Raman spectroscopy (SORS) to subcortical bone tissue”. J. Biophotonics 10(8): 990–996. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 18.Maher JR and Berger AJ. ”Determination of Ideal Offset for Spatially Offset Raman Spectroscopy”. Appl. Spectrosc 2010. 64(1): 61–65. [DOI] [PubMed] [Google Scholar]
- 19.Esmonde-White FWL, Esmonde-White KA, and Morris MD. ”Minor Distortions with Major Consequences: Correcting Distortions in Imaging Spectrographs”. Appl. Spectrosc 2011. 65(1): 85–98. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 20.Zhang Z, Chen S, Liang Y, Liu Z, Zhang Q, Ding L, Ye F, and Zhou H. ”An intelligent background-correction algorithm for highly fluorescent samples in Raman spectroscopy”. J. Raman Spectrosc 2010. 41(6): 659–669. [Google Scholar]
- 21.Savitzky A and Golay MJE. ”Smoothing and Differentiation of Data by Simplified Least Squares Procedures”. Anal. Chem 1964. 36(8): 1627–1639. [Google Scholar]
- 22.Dooley KA, Esmonde-White FWL, and Morris MD. ”Optical fiber bundle coupling errors in Raman spectra: correction via data processing”. Proc. SPIE 7560, Biomedical Vibrational Spectroscopy IV: Advances in Research and Industry. 7560: 137–141. [Google Scholar]
- 23.Massie C, Knapp E, Chen K, Berger AJ, and Awad HA. ”Improved prediction of femoral fracture toughness in mice by combining standard medical imaging with Raman spectroscopy”. J. Biomech 2021. 116. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 24.Mandair GS and Morris M. ”Contributions of Raman spectroscopy to the understanding of bone strength”. BoneKEy Rep. 2015. 4: 620. [DOI] [PMC free article] [PubMed] [Google Scholar]