Skip to main content
Journal of Clinical Microbiology logoLink to Journal of Clinical Microbiology
. 2001 Oct;39(10):3781–3784. doi: 10.1128/JCM.39.10.3781-3784.2001

Optimal Inoculation Methods and Quality Control for the NCCLS Oxacillin Agar Screen Test for Detection of Oxacillin Resistance in Staphylococcus aureus

Jana M Swenson 1,*, Jean Spargo 2, Fred C Tenover 1, Mary Jane Ferraro 2
PMCID: PMC88434  PMID: 11574618

Abstract

To define more precisely the inoculation methods to be used in the oxacillin screen test for Staphylococcus aureus, we tested agar screen plates prepared in house with 6 μg of oxacillin/ml and 4% NaCl using the four different inoculation methods that would most likely be used by clinical laboratories. The organisms selected for testing were 19 heteroresistant mecA-producing strains and 41 non-mecA-producing strains for which oxacillin MICs were near the susceptible breakpoint. The inoculation method that was preferred by all four readers and that resulted in the best combination of sensitivity and specificity was a 1-μl loopful of a 0.5 McFarland suspension. A second objective of the study was to then use this method to inoculate plates from five different manufacturers of commercially prepared media. Although all commercial media performed with acceptable sensitivity compared to the reference lot, one of the commercial lots demonstrated a lack of specificity. Those lots of oxacillin screen medium that fail to grow heteroresistant strains can be detected by using S. aureus ATCC 43300 as a positive control in the test and by using transmitted light to carefully examine the plates for any growth. However, lack of specificity with commercial lots may be difficult to detect using any of the current quality control organisms.


The oxacillin agar screen test for detection of oxacillin resistance in Staphylococcus aureus was first included in the NCCLS dilution methods document, M7, in 1990 (11), although neither that edition nor the current edition (14) provides details on how the test should be inoculated. Furthermore, a variety of inoculation methods are recommended on the product labels of commercially prepared media. Despite this, many studies have shown that the oxacillin screen test performs well for detection of S. aureus strains that contain the mecA gene, which mediates oxacillin resistance (14, 68, 1517, 20). However, when studies have included strains whose resistance is heterogeneous, the test has been shown to perform less well (2, 15).

Recommendations for quality control of the test were not included in NCCLS documents until 1997 (13), when S. aureus ATCC 43300, a mecA-positive strain that is very heterogeneous in its expression of oxacillin resistance (10), was suggested as the quality control strain. The reason for recommending a strain that was difficult to detect was that, when adequate growth was obtained, the user would be assured that the test could detect heterogeneously resistant strains. Recent comments addressed to the NCCLS Subcommittee on Antimicrobial Susceptibility Testing (14) have questioned the appropriateness of this strain for this procedure. Other references (9) and package inserts from commercial plates have recommended the use of S. aureus ATCC 33591, a strain which expresses homogeneous resistance.

The present study was undertaken both to clarify the best inoculation methods for the NCCLS oxacillin agar screen test and to verify that S. aureus ATCC 43300 is adequate for use as a positive quality control strain for the test.

Study design.

The study was conducted in two laboratories, the Centers for Disease Control and Prevention (CDC) and Massachusetts General Hospital (MGH), in three phases. Phase 1 evaluated several inoculation methods in order to select a method that could be used in the subsequent phases. Phase 2 was undertaken to confirm that commercial lots of agar performed well with the chosen inoculation method. Phase 3 was performed to document the specificity of the test using a set of more commonly encountered susceptible organisms (see below).

Organisms.

In phases 1 and 2, the organisms chosen for testing represented a challenge set from the CDC culture collection. For the resistant organisms, those of a low expression class were selected (19), with oxacillin MICs ranging from 4 to 128 μg/ml; for the susceptible strains, organisms for which the oxacillin MICs were 1 to 8 μg/ml when tested by agar dilution using 4% NaCl were included. All strains used in phases 1 and 2 had been tested by PCR for the presence of the mecA gene and by population analysis (for oxacillin-resistant strains only) to determine the level of expression of oxacillin resistance (19). The organisms fell into four groups. Group 1 (n = 19) comprised mecA positive strains determined to be in expression class 1 or 2, i.e., they shared a high degree of heterogeneity. Group 2 (n = 16) comprised mecA-negative strains for which agar dilution MICs of oxacillin were ≥4 μg/ml when tested with 4% salt (5) but were ≤2 μg/ml with 2% salt. Group 3 (n = 5) comprised mecA-negative strains for which the oxacillin MICs were ≥4 μg/ml with both 4 and 2% salt. The five strains in group 3 were considered to have the “MOD” phenotype; two of the strains had been characterized previously and had penicillin-binding proteins (PBPs) with modified affinity to penicillin but did not contain PBP2a, the mecA gene product (18). Three additional strains in group 3 were classified as “MOD” only phenotypically, i.e., the oxacillin MICs for these strains were >2 μg/ml and were not lowered with the addition of clavulanic acid, and these strains did not contain mecA. The “MOD” strains were included to see how they would test; however, it was decided before beginning the study that because of the unknown prevalence of such strains, their unknown clinical significance, and the lack of molecular characterization of all the strains, the results of their testing would be considered separately in data analysis. The last group (group 4 [n = 20]) comprised mecA-negative strains for which the oxacillin MICs were 1 to 2 μg/ml with 4% salt.

In phase 3, additional, oxacillin-susceptible clinical isolates were tested along with a subset from phases 1 and 2. Combined, these included 64 susceptible strains: 47 strains for which the oxacillin broth microdilution MICs were ≤0.12 to 2 μg/ml, 15 susceptible crossover strains from groups 2 and 4 described above, and the 2 susceptible quality control strains (ATCC 29213 and ATCC 25923) that were blinded to the readers. Eight resistant strains were also included in phase 3: six less-challenging resistant strains (expression class 3 or 4) and the two oxacillin-resistant quality control strains (ATCC 43300 and ATCC 33591), also blinded. All strains were subcultured twice from the freezer on trypticase soy agar with 5% sheep blood (TSA-SB) before being tested.

Quality control strains.

Four S. aureus strains were used as controls on each day of testing: ATCC 43300, a heteroresistant strain; ATCC 33591, a homogeneously resistant strain; and two oxacillin-susceptible strains, ATCC 25923 and ATCC 29213.

Oxacillin screen agar medium.

For phase 1, plates containing 4% salt and 6 μg of oxacillin (Sigma, St. Louis, Mo.)/ml were prepared at CDC using the standard reference lot of Mueller-Hinton agar that is used by manufacturers to standardize their production lots (12). For phase 2, commercial plates were purchased from five different manufacturers, i.e., BD Biosciences (BBL, Cockeysville, Md.), PML Microbiologicals (Wilsonville, Oreg.), Hardy Diagnostics (Santa Maria, Calif.), Gibson Laboratories, Inc., (Lexington, Ky.), and Remel (Lenexa, Kans.); three manufacturers' plates were tested at MGH, and two manufacturers' plates were tested at CDC. In addition, for phases 2 and 3, plates were prepared and tested at CDC using Mueller-Hinton II agar (BBL). All plates were stored at 4 to 8°C for no longer than 4 weeks or until the manufacturer's expiration date. In the latter part of phase 2 and in phase 3, additional lots of commercial plates were obtained from the two most commonly used manufacturers (BBL and Remel).

Inoculation methods.

Inoculum suspensions were prepared from 18- to 24-h cultures grown on TSA-SB and adjusted to equal a 0.5 McFarland standard. Four inoculation methods were studied during phase 1: (i) spotting an area 10 to 15 mm in diameter using a cotton swab that was dipped in the suspension and from which excess fluid was expressed, (ii) streaking a quadrant of the plate using a swab prepared as above, (iii) spotting an area 10 to 15 mm in diameter using a 1-μl disposable loop, and (iv) spotting 10 μl using a micropipette. Four organisms were tested per plate, with only one inoculation method used per plate to avoid biasing the reading between plates. For phases 2 and 3, testing was done by spotting an area 10 to 15 mm in diameter using a 1-μl disposable loop. All plates were incubated at 35°C and read after 24 h.

Reading.

All readings were done by two independent readers using transmitted light at both sites, except for the testing of additional lots of media at the end of phase 2, where one reader only was used. All readers had had previous experience in reading oxacillin screen plates. Plates were read as positive (if confluent growth was observed), weak positive (if the growth was almost confluent but hazy or >20 colonies), the number of colonies (if <20), or negative. However, for data analysis purposes, any growth of >1 colony was considered positive.

Phase 1.

In phase 1, the abilities of the four inoculation methods to detect resistant strains were essentially equivalent (Table 1), with the method that would be expected to deliver the largest inoculum (the 10-μl pipette) achieving 100% sensitivity among all readers and laboratories. All the mecA-positive strains were detected by the other three methods by all readers in both laboratories, except for two strains (strains 107 and 351) which were not detected by all three methods in one laboratory. For these strains oxacillin MICs were 4 μg/ml by both conventional broth microdilution (2% salt) and agar dilution testing with increased salt (4%), and these strains were determined to be among the most heterogeneous of those tested (expression class 1; data not shown). Although the 10-μl-pipette inoculation method gave the best sensitivity across laboratories and readers, it resulted in the growth of a large number of susceptible strains (i.e., false positives), which resulted in a specificity of 88.9% in the best case and as low as 58% in the worst case. The swab and loop methods performed better in terms of specificity, ranging from 94 to 97%. The best combination of sensitivity (95 to 100%) and specificity (97%) among all readers and all laboratories was obtained with the 1-μl loop. However, the performances of all methods, except the 10-μl pipette, were not significantly different (P > 0.05) from each other (McNemar's chi-square test).

TABLE 1.

Phase 1 results of testing four inoculation methods in two laboratories using an in-house lot of oxacillin screen agar prepared with a reference standard lot of Mueller-Hinton agar containing 4% salt and 6 μg of oxacillin/ml

Inoculation method Sensitivitya
Specificityb
CDC
MGH
CDC
MGH
Reader 1 Reader 2 Reader 1 Reader 2 Reader 1 Reader 2 Reader 1 Reader 2
Swab, spotted 18 (94.7) 17 (89.5) 19 (100) 19 (100) 35 (97.2) 35 (97.2) 35 (97.2) 35 (97.2)
Swab, quadrant 18 (94.7) 19 (100) 19 (100) 19 (100) 34 (94.4) 34 (94.4) 35 (97.2) 35 (97.2)
1-μl loop 19 (100) 18 (94.7) 19 (100) 19 (100) 35 (97.2) 35 (97.2) 35 (97.2) 35 (97.2)
10-μl pipette 19 (100) 19 (100) 19 (100) 19 (100) 21 (58.3) 28 (77.8) 32 (88.9) 31 (86.1)
a

Expressed as number (percent) of 19 mecA-positive strains that grew. 

b

Expressed as number (percent) of 36 mecA-negative strains that did not grow, excluding 5 strains with the MOD resistance phenotype. 

Phase 2.

Of the five commercial lots tested (Table 2), the lowest sensitivity (84%) was obtained with the Gibson medium, although the performance of the in-house lot was similar (84 to 90%). There was no statistically significant difference among the sensitivity results when readers or medium lots were compared (P > 0.05 by McNemar's chi-square test). If the two highly heterogeneous strains (strains 107 and 351) are excluded, all the phase 2 media detected the remaining 17 resistant strains except for the lot from Gibson, which failed to grow one of the remaining 17 resistant strains (as noted by both readers), and the in-house lot, which was read as negative for one strain by one reader only. The Gibson lot also failed to grow the resistant control strain, ATCC 43300, and therefore results from that round of testing on that medium should not have been reported in a clinical laboratory.

TABLE 2.

Phase 2 results of testing five commercial lots and one in-house lot of oxacillin screen agar inoculated using a 1-μl loop

Manufacturer Testing laboratory Sensitivitya
Specificityb
Reader 1 Reader 2 Reader 1 Reader 2
In house (lot 2) CDC 17 (89.5) 16 (84.2) 36 (100) 36 (100)
BBL CDC 19 (100) 17 (89.5) 35 (97.2) 35 (97.2)
PML CDC 19 (100) 19 (100) 31 (86.1) 35 (97.2)
Hardy MGH 19 (100) 19 (100) 32 (88.9) 29 (80.6)
Gibson MGH 16 (84.2) 16 (84.2) 35 (97.2) 33 (91.7)
Remel MGH 18 (94.7) 19 (100) 21 (58.3) 17 (47.2)
a

Expressed as the number (percent) of 19 mecA-positive strains that grew. 

b

Expressed as the number (percent) of 36 mecA-negative strains that did not grow, excluding 5 strains with the MOD resistance phenotype. 

In phase 2, specificity was best with the BBL lot (97%) and the in-house lot (100%), but dropped to 47 to 58%, depending upon the reader, for the Remel medium (Table 2). In analyzing specificity results, statistically significant differences were found between the in-house lot and the lots from PML, Hardy, and Remel for reader 1, and between the in-house lot and the lots from Hardy and Remel for reader 2 (P < 0.05). When reader 1 was compared to reader 2, there were significant differences for both PML and Remel (P < 0.05).

To determine if the lower specificity of Remel media was related to a single lot, laboratory, or reader, we purchased and retested two additional lots from Remel and one from BBL along with a third in-house lot (data not shown). Sensitivity was very good (95 to 100%) for all additional lots tested. The specificity of the second lot from BBL was similar to that in the first testing in phase 2 (97%). Although differences could be attributed to the person performing the testing and how conservatively the test was read, one of the two additional lots from Remel performed slightly better in terms of specificity (83%); the other performed similarly (58%).

Phase 3.

In order to determine how well the media would perform in a routine fashion (presumably with less-challenging strains), additional lots of both BBL and Remel media (and a new in-house lot) were tested against a group of susceptible organisms for which the oxacillin MICs did not cluster around the resistance breakpoint and a subset of the susceptible challenge organisms from phases 1 and 2. For the susceptible strains, the specificity of the Remel lot improved but still remained below 90% (range, 84 to 88%). For eight mecA-positive strains, the sensitivity was 100% with all three media tested (data not shown).

Testing strains with modified PBPs.

Results for the MOD strains tested in phases 1 and 2 were dependent on the oxacillin MIC. Strains for which MICs were 8 μg/ml were more likely to give positive results than those for which MICs were 4 μg/ml and which displayed variable growth depending on the laboratory and reader (data not shown).

Quality control.

In phase 1, both oxacillin-resistant organisms grew with all inoculation methods. On two occasions, weak growth of one of the susceptible control strains, S. aureus ATCC 25923 (the strain recommended for disk diffusion quality control), was recorded as positive by one of the readers in one of the laboratories when the 10-μl inoculum was used. In phase 2, one of the commercial lots failed to grow S. aureus ATCC 43300, and growth of ATCC 25923 was detected on the PML lot by one reader. Growth of ATCC 25923 was also detected during phase 3 in one laboratory by both readers on one of two days of testing using Remel's medium. On this day growth was noted for both the quality control strain and the blinded quality control strain.

During all 3 phases, when the 1-μl loop inoculum was used, ATCC 43300 showed adequate growth on all media except one lot of commercial medium. The susceptible control strain ATCC 29213 also performed as expected, i.e., no growth on any of the medium lots. However, the susceptible strain ATCC 25923 did grow sporadically during phases 2 and 3 on some of the commercial medium lots.

In summary, three of four inoculation methods were found to perform adequately in the oxacillin screen test. With the commercial media tested, the specificity of the medium of one of the manufacturers was inferior. However, given the challenging nature of the organisms used in phases 1 and 2, lack of specificity is understandable. When strains that would be considered more representative of those likely to be encountered in a clinical laboratory were tested, the performance of that medium improved. However, laboratories need to be aware that some current commercial medium lots may overcall resistance. Given the expectation that the oxacillin screen test is comparable to MIC reference methods in reliability (14), if it is used without further confirmation of resistance, the potential exists for falsely labeling strains as oxacillin resistant. This could lead to inappropriate treatment for patients and unnecessary infection control measures. Although the potential also exists for missing truly resistant strains of low expression classes (as happened with two of the resistant strains in one laboratory in this study), proper attention to test conditions (e.g., inoculum and time of incubation), careful examination of plates for any growth using transmitted light, and the use of S. aureus ATCC 43300 as a resistant control to check medium and observer performance will decrease the chances of such errors. Although S. aureus ATCC 25923, a susceptible control, occasionally grew, its performance was not consistent enough to recommend its use as a negative control to detect media with the potential for decreased specificity. The use of an inoculum larger than 1 μl or larger than that achieved with a swab is not recommended because of greatly decreased specificity. Package inserts that accompany commercially available oxacillin screen agar should be revised to reflect proper inoculation procedures and use of quality control organisms.

Acknowledgments

We acknowledge Linda McDougal (CDC) and Lisa Gartland (MGH) for expert help as second readers in the study. We also thank Alaeddin Abu-Zant, a visiting fellow at CDC at the time of the study, for the population analysis studies of the mecA-producing strains.

REFERENCES

  • 1.Brakstad O G, Mæland J A, Tveten Y. Multiplex polymerase chain reaction for detection of genes for Staphylococcus aureus thermonuclease and methicillin resistance and correlation with oxacillin resistance. APMIS. 1993;101:681–688. doi: 10.1111/j.1699-0463.1993.tb00165.x. [DOI] [PubMed] [Google Scholar]
  • 2.Cavassini M, Wenger A, Jaton K, Blanc D S, Bille J. Evaluation of MRSA-Screen, a simple anti-PBP2a slide latex agglutination kit, for rapid detection of methicillin resistance in Staphylococcus aureus. J Clin Microbiol. 1999;37:1591–1594. doi: 10.1128/jcm.37.5.1591-1594.1999. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 3.Frebourg N B, Nouet D, Lemée L, Martin E, Lemeland J. Comparison of ATB Staph, Rapid ATB Staph, Vitek, and E-Test methods for detection of oxacillin heteroresistance in staphylococci possessing mecA. J Clin Microbiol. 1998;36:52–57. doi: 10.1128/jcm.36.1.52-57.1998. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 4.Gerberding J L, Miick C, Liu H H, Chambers H F. Comparison of conventional susceptibility tests with direct detection of penicillin-binding protein 2a in borderline oxacillin-resistant strains of Staphylococcus aureus. Antimicrob Agents Chemother. 1991;35:2574–2579. doi: 10.1128/aac.35.12.2574. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 5.Huang M B, Gay T E, Baker C N, Banerjee S N, Tenover F C. Two percent sodium chloride is required for susceptibility testing of staphylococci with oxacillin when using agar-based dilution methods. J Clin Microbiol. 1993;31:2683–2688. doi: 10.1128/jcm.31.10.2683-2688.1993. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 6.Knapp C C, Ludwig M D, Washington J A, Chambers H F. Evaluation of Vitek GPS-SA card for testing of oxacillin against borderline-susceptible staphylococci that lack mec. J Clin Microbiol. 1996;34:1603–1605. doi: 10.1128/jcm.34.7.1603-1605.1996. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 7.Kobayashi Y, Kizaki M, Kawakami Y, Uchida H, Ikeda Y. Assessment of oxacillin salt agar for detection of MRSA identified by presence of the mecA gene. J Hosp Infect. 1993;23:279–285. doi: 10.1016/0195-6701(93)90145-p. [DOI] [PubMed] [Google Scholar]
  • 8.Kohner P, Uhl J, Kolbert C, Persing D, Cockerill F. Comparison of susceptibility testing methods with mecA gene analysis for determining oxacillin (methicillin) resistance in clinical isolates of Staphylococcus aureus and coagulase-negative Staphylococcus spp. J Clin Microbiol. 1999;37:2952–2961. doi: 10.1128/jcm.37.9.2952-2961.1999. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 9.Leitch C, Boonlayangoor S. : H. D. Isenberg (ed.) 1992. Tests to detect oxacillin (methicillin)-resistant staphylococci with an oxacillin screen plate pp. 5.5.1–5.5.7. Clinical microbiology procedures handbook. American Society for Microbiology, Washington, D.C. [Google Scholar]
  • 10.Mackenzie A M R, Richardson H, Lannigan R, Wood D. Evidence that the National Committee for Clinical Laboratory Standards disk test is less sensitive than the screen plate for detection of low-expression-class methicillin-resistant Staphylococcus aureus. J Clin Microbiol. 1995;33:1909–1911. doi: 10.1128/jcm.33.7.1909-1911.1995. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 11.National Committee for Clinical Laboratory Standards. NCCLS document M7–A2. 2nd ed. Villanova, Pa: National Committee for Clinical Laboratory Standards; 1990. Methods for dilution antimicrobial susceptibility tests for bacteria that grow aerobically. Approved standard. [Google Scholar]
  • 12.National Committee for Clinical Laboratory Standards. NCCLS document M6-A. 1st ed. Wayne, Pa: National Committee for Clinical Laboratory Standards; 1996. Protocols for evaluating dehydrated Mueller-Hinton agar. Approved standard. [Google Scholar]
  • 13.National Committee for Clinical Laboratory Standards. NCCLS document M7–A4. 4th ed. Wayne, Pa: National Committee for Clinical Laboratory Standards; 1997. Methods for dilution antimicrobial susceptibility tests for bacteria that grow aerobically. Approved standard. [Google Scholar]
  • 14.NCCLS. NCCLS document M7–A5. 5th ed. Wayne, Pa: NCCLS; 2000. Methods for dilution antimicrobial susceptibility tests for bacteria that grow aerobically. Approved standard. [Google Scholar]
  • 15.Resende C A, Figueiredo A M. Discrimination of methicillin-resistant Staphylococcus aureus from borderline-resistant and susceptible isolates by different methods. J Med Microbiol. 1997;46:145–149. doi: 10.1099/00222615-46-2-145. [DOI] [PubMed] [Google Scholar]
  • 16.Skov R L, Palleson L V, Poulsen R L, Espersen F. Evaluation of a new 3-h hybridization method for detecting the mecA gene in Staphylococcus aureus and comparison with existing genotypic and phenotypic susceptibility testing methods. J Antimicrob Chemother. 1999;43:467–475. doi: 10.1093/jac/43.4.467. [DOI] [PubMed] [Google Scholar]
  • 17.Skulnick M, Simor A E, Gregson D, Patel M, Small G W, Kreiswirth B, Hathoway D, Low D E. Evaluation of commercial and standard methodology for determination of oxacillin susceptibility in Staphylococcus aureus. J Clin Microbiol. 1992;30:1985–1988. doi: 10.1128/jcm.30.8.1985-1988.1992. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 18.Tomasz A, Drugeon H B, de Lencastre H M, Jabes D, McDougal L, Bille J. New mechanism for methicillin resistance in Staphylococcus aureus: clinical isolates that lack the PBP 2a gene and contain modified penicillin-binding proteins with modified penicillin-binding capacity. Antimicrob Agents Chemother. 1989;33:1869–1874. doi: 10.1128/aac.33.11.1869. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 19.Tomasz A, Nachman S, Leaf H. Stable classes of phenotypic expression in methicillin-resistant clinical isolates of staphylococci. Antimicrob Agents Chemother. 1991;35:124–129. doi: 10.1128/aac.35.1.124. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 20.Unal S, Werner K, DeGirolami P, Barsanti F, Eliopoulos G. Comparison of tests for detection of methicillin-resistant Staphylococcus aureus in a clinical microbiology laboratory. Antimicrob Agents Chemother. 1994;38:345–347. doi: 10.1128/aac.38.2.345. [DOI] [PMC free article] [PubMed] [Google Scholar]

Articles from Journal of Clinical Microbiology are provided here courtesy of American Society for Microbiology (ASM)

RESOURCES