Failure mode and effect analysis (FMEA) to identify and mitigate failures in a hospital rapid response system (RRS)

Ehsan Ullah a,b, Mirza Mansoor Baig c, Hamid GholamHosseini c, Jun Lu a,d,e,f,g,
PMCID: PMC8857483  PMID: 35243066

Abstract

We performed FMEA on the existing RRS with the help of routine users of the RRS who acted as subject matter experts and evaluated the failures for their criticality using the Risk Priority Number approach based on their experience of the RRS. The FMEA found 35 potential failure modes and 101 failure mode effects across 13 process steps of the RRS. The afferent limb of RRS was found to be more prone to these failures (62, 61.4%) than the efferent limb of the RRS (39, 38.6%). Modification of calling criteria (12, 11.9%) and calculation of New Zealand Early Warning Scores (NZEWS) calculation (11, 10.9%) steps were found to potentially give rise to the highest number of these failures. Causes of these failures include human error and related factors (35, 34.7%), staff workload/staffing levels (30, 29.7%) and limitations due to paper-based charts and organisational factors (n = 30, 29.7%). The demonstrated electronic system was found to potentially eliminate or reduce the likelihood of 71 (70.2%) failures. The failures not eliminated by the electronic RRS require targeted corrective measures including scenario-based training and education, and revised calling criteria to include triggers for hypothermia and high systolic blood pressure.

Keywords: Failure modes and effects analysis, Rapid response system, Clinical deterioration, Vital signs, Early warning score


Failure modes and effects analysis; Rapid response system; Clinical deterioration; Vital signs; Early warning score.

1. Introduction

The Rapid Response System (RRS) as shown in Figure 1 acts as the surveillance mechanism used by healthcare organisations to monitor patients admitted to general hospital wards outside the critical care settings with repeated vital signs observations and Early Warning Scores such as NZEWS. The values of vital signs and NZEWS determine the escalation trigger or calling criteria i.e., when patients require escalation of care such as a rapid response by a Medical Emergency Team (MET) or equivalent [1].

Figure 1.

Figure 1

Schematic representation of a RRS.

In New Zealand, most public hospitals use paper based vital signs charts in the general hospital ward settings to drive the afferent limb of the RRS, and a specialised team of nurses called Patient at-Risk (PaR) nurses and MET constitute the afferent limbs of the RRS [2].

The literature on the RRS from New Zealand and elsewhere have mainly reported the epidemiology of RRS activations [2, 3, 4, 5], outcomes of the in-hospital cardiac arrests [6, 7, 8] and comparison of various models of the RRS to recognise deteriorating patients in general hospital wards, outside critical care settings [9, 10]. There is a severe shortage of literature on how the RRS could be improved using systems approaches. We found only one publication using Root Cause Analysis to examine the causes of failures in patient monitoring and escalation of care for deteriorating patients [11].

Failure Mode and Effect Analysis (FMEA) has been widely used in high-risk industries to evaluate and mitigate process weaknesses [1213, 14, 15]. FMEA has been effectively applied to examine and mitigate risks and failure modes in many healthcare processes [16, 17]. FMEA has not been applied to systematically assess and address RRS failures despite such failures being widely reported [18, 19]. This study applies FMEA methodology to identify and address potential failures within an RRS.

2. Methods

2.1. Setting

The FMEA was performed at Taranaki Base Hospital, Taranaki District Health Board, New Plymouth, New Zealand between January and July 2021. The stages of the FMEA are shown in Figure 2.

Figure 2.

Figure 2

Stages of the failure mode and effect analysis.

2.2. Selection of high-risk process

RRS, as a unifying term to describe the process from vital signs monitoring to a rapid response type care delivered by MET or equivalent [20], was selected as the high-risk process for the FMEA study.

2.3. Selection of experts

FMEA methodology utilizes hands-on knowledge of the users of a process whereby a diverse group of users working on various parts of the process are recruited as subject matter experts (SMEs) to get insights into the potential ways a process may fail (failure modes). Then, these insights lead to the analysis of the effects of such failure modes, their frequency, and the ability of process controls on that frequency using standard FMEA framework. Therefore, SMEs doesn't represent a sample of users and no statistical tests are applied to validate selection of SMEs [12]. For this FMEA, we recruited six regular users of the RRS process as (SMEs who participated on a voluntary basis. Two SMEs were registered nurses (RNs), two were resident medical doctors, one was a specialised critical care outreach nurse, locally known as a ‘Patient-at-risk nurse or PAR nurse’ and one was a senior medical officer, a general physician or hospitalist. The SMEs attended a training session on FMEA methodology and participated in the demonstration of Vital Signs Monitoring and the Decision Support System [21] developed by Precision Driven Health – a system that automatically calculates NZEWS and has the ability to send alerts to cellular or landline phones and pager devices in a similar way the current paging system works for MET activations, and in addition enables them to view the vital signs values, trends and NZEWS on smartphones or tablet devices.

2.4. Mapping the process steps of RRS

A process diagram was drawn up to illustrate all the process steps of the existing rapid response system as shown in Figure 3.

Figure 3.

Figure 3

Flow diagram of steps involved in a rapid response system.

2.5. Risk analysis

Potential failure mode effects were identified at each process step shown in Figure 3. The criticality of these failures was determined using a Risk Priority Number (RPN). A RPN is a numerical score quantifying the severity level (SL), occurrence level (OL) and detection level (DL) of failures according to the rating scale shown in Table 1 (RPN = severity x occurrence x detectability) of failures. The theoretical minimum RPN is 1∗1∗1 = 1. The theoretical maximum RPN is 3∗10∗5 = 150.

Table 1.

Rating scale for severity, occurrence and detection of failure modes in a RRS.

Severity Levels (SL) of the Effect of the Failure Modes Rating
No harm to patient/no effect on detection of patient deterioration 1
Non-documented vital signs or NZEWS or incorrect calculated NZEWS 2
Actual or potential delay to or lack of detection of patient deterioration 3
Occurrence Level (OL) of the Failure Modes Rating
Once in more than a year 1
Once in a year 2
Once in six months 3
Once in three months 4
Once a month 5
Once a week 6
Once every 3 days 7
Once per day 8
One per 8-hour shift 9
More than once per 8-hour shift 10
Detection Level (DL) of the Failure Modes when they occur Rating
100% detection 1
>50% detection 2
11–50% detection 3
<10 % detection 4
0% detection 5

This rating scale (Table 1) was adapted from Rezaee et al., [22] and Buja et al., [23]. The RPN or criticality of a failure increases with higher severity and occurrence levels, and with lower detection levels.

The SMEs assigned the SL, OL and DL to each failure mode effect based on their day-to-day experience about existing RRS activities at the study site.

2.6. Developing recommendations

The SMEs were asked whether the electronic system would eliminate or reduce the risk of failure modes identified within existing rapid response system components. The responses of the SMEs were recorded within FMEA data collection sheet. Specific recommendations were formulated targeting the failure modes of which the risk was not deemed to be eliminated or reduced by the electronic system.

2.7. Institutional review board statement

The study was conducted according to the guidelines of the Declaration of Helsinki and approved by Auckland University of Technology Ethics Committee (AUTEC) approved the research dated 13 March 2019, application number 19/37.

3. Results

The FMEA identified a total of 35 failure modes and 101 failure effects distributed over 13 process steps as shown in Table 2. Whether the demonstrated electronic RRS will potentially reduce or eliminate the risk and likelihood of these failures is also tabulated in the last column of Table 2.

Table 2.

Failure modes, their effects and causes across the process steps of a RRS.

Serial No Process Step Failure Mode Failure Mode Effects Causes RPN Will electronic RRS reduce the risk?
1 1.Timeliness of vital signs observations Delay in undertaking vital signs Delay in detecting possible derangements in vital signs and/or NZEWS which may lead to delayed review and/or rapid response hence increased chances of adverse events Lack of reinforcing function/mechanism to remind staff when vital sign observations become due based on patient's previous NZEWS value and/or minimum 4-hourly vital signs monitoring 108 No
2 Non-compliance with protocols such as minimum 4-hourly vital signs monitoring on general wards 108 No
3 Staff allocated to the patient is traceable through clinical records data and can be held responsible for this omission 108 No
4 2. Pulse Rate (PR) Non-measurement of the PR Delay in detecting possible derangements in vital signs and/or NZEWS Too busy staff or inadequate staff allocation, memory lapse 120 Yes
5 Non-compliance with protocols such as minimum 4-hourly vital signs monitoring, and due to inability to calculate NZEWS score 120 Yes
6 Staff allocated to the patient is traceable through clinical records data and can be held responsible for this omission 120 Yes
7 Non-documentation of the PR Delay in detecting possible derangements in vital signs and/or NZEWS Interruptions due to other more urgent tasks 18 Yes
8 Non-compliance with protocols 18 Yes
9 Staff allocated to the patient is traceable through clinical records data and can be held responsible for this omission 18 Yes
10 No trigger for rhythm abnormalities Heart rhythm abnormalities that do not cause haemodynamic instability are not recognised and managed Not included in the monitoring protocols 25
11 3. Systolic blood pressure (SBP) Non-measurement of the SBP Delay in detecting possible derangements in vital signs and/or NZEWS Incorrect or less frequent measurements due to previous set of vital signs or NZEWS being miscalculated or staff not able to take measurement due to being busy, distracted or memory lapse or non-cooperation of patient 120 Yes
12 Non-compliance with protocols 120 Yes
13 Staff allocated to the patient is traceable through clinical records data and can be held responsible for this omission unless non-cooperation of patient is documented in clinical notes, if relevant 120 Yes
14 Non-documentation of the SBP Delay in detecting possible derangements in vital signs and/or NZEWS Interruptions due to other more urgent tasks 18 Yes
15 Non-compliance with protocols 18 Yes
16 Staff allocated to the patient is traceable through clinical records data and can be held responsible for this omission unless non-cooperation of patient is documented in clinical notes, if relevant 18 Yes
17 No trigger until too high SBP Significant high Systolic BP (<220 mmHg) alone may continue for hours to days and could be indication of significant illness without any change in NZEWS Protocol definition issue 90 No
18 4.Temperature Non-measurement of the temperature Delay in detecting possible derangements in vital signs and/or NZEWS Too busy staff or inadequate staff allocation, memory lapse 70 No
19 Non-compliance with protocols 70 No
20 Staff allocated to the patient is traceable through clinical records data and can be held responsible for this omission 70 No
21 Non-documentation of the temperature Delay in detecting possible derangements in vital signs and/or NZEWS Interruptions due to other more urgent tasks 12 No
22 Non-compliance with protocols 12 No
23 Staff allocated to the patient is traceable through clinical records data and can be held responsible for this omission 12 No
24 No trigger for hypothermia Significant illness such as sepsis may not be recognised Protocol definition issue 75 No
25 5. Respiratory rate (RR) Non-measurement of the RR Delay in detecting possible derangements in vital signs and/or NZEWS Too busy staff or inadequate staff allocation, memory lapse 120 No
26 Non-compliance with protocols 120 No
27 Staff allocated to the patient is traceable through clinical records data and can be held responsible for this omission 120 No
28 Errors in measurement the RR Delay in detecting possible derangements in vital signs and/or NZEWS Interruptions due to other more urgent tasks 36 No
29 Non-compliance with protocols 36 No
30 Staff allocated to the patient is traceable through clinical records data and can be held responsible for this omission 36 No
31 Non-documentation of the RR Delay in detecting possible derangements in vital signs and/or NZEWS Interruptions due to other more urgent tasks 18 No
32 Non-compliance with protocols 18 No
33 Staff allocated to the patient is traceable through clinical records data and can be held responsible for this omission 18 No
34 6. Oxygen saturation Non-measurement of oxygen saturation Delay in detecting possible derangements in vital signs and/or NZEWS Incorrect or less frequent measurements due to previous set of vital signs or NZEWS being miscalculated or staff not able to take measurement due to being busy, inadequate staffing levels or memory lapse 105 Yes
35 Non-compliance with protocols 105 Yes
36 Staff allocated to the patient is traceable through clinical records data and can be held responsible for this omission 105 Yes
37 Non-documentation of oxygen saturation Delay in detecting possible derangements in vital signs and/or NZEWS Interruptions due to other more urgent tasks 18 Yes
38 Non-compliance with protocols 18 Yes
39 Staff allocated to the patient is traceable through clinical records data and can be held responsible for this omission 18 Yes
40 7. Oxygen requirement Non-measurement of oxygen requirement Delay in detecting possible derangements in vital signs and/or NZEWS Incorrect or less frequent measurements due to previous set of vital signs or NZEWS being miscalculated or staff not able to take measurement due to being busy, distracted or memory lapse 90 Yes
41 Non-compliance with protocols 90 Yes
42 Staff allocated to the patient is traceable through clinical records data and can be held responsible for this omission 90 Yes
43 Non-documentation of oxygen requirement Delay in detecting possible derangements in vital signs and/or NZEWS Too busy, interrupted by other more urgent task, memory lapse 18 Yes
44 Non-compliance with protocols 18 Yes
45 Staff allocated to the patient is traceable through clinical records data and can be held responsible for this omission 18 Yes
46 8. Level of consciousness Errors in interpretation of the level of consciousness Delay in detecting possible derangements in vital signs and/or NZEWS Human error, complacency, lack of awareness of patient's sleeping pattern 75 No
47 Non-compliance with protocols 75 No
48 Staff allocated to the patient is traceable through clinical records data and can be held responsible for this omission 75 No
49 Non-documentation of the level of consciousness Delay in detecting possible derangements in vital signs and/or NZEWS Too busy staff or inadequate staff allocation, memory lapse 72 No
50 Non-compliance with protocols 72 No
51 Staff allocated to the patient is traceable through clinical records data and can be held responsible for this omission 72 No
52 9. NZEWS calculation Non-calculation of NZEWS Delay in detecting possible derangements in vital signs and/or NZEWS Time-consuming, difficult to calculate when staff working with huge cognitive load, interruptions 135 Yes
53 Non-compliance with protocols 135 Yes
54 Staff allocated to the patient is traceable through clinical records data and can be held responsible for this omission 135 Yes
55 Incorrect calculation of NZEWS Delay in detecting possible derangements in vital signs and/or NZEWS Time-consuming, difficult to calculate when staff working with huge cognitive load, interruptions 135 Yes
56 Non-compliance with protocols 135 Yes
57 Staff allocated to the patient is traceable through clinical records data and can be held responsible for this omission 135 Yes
58 Incorrect NZEWS calculated for type and age of patient such as adult NZEWS calculated when patient required a paediatric or maternal early warning score 72
59 Non-compliance with protocols 72 Yes
60 Non-documentation of NZEWS Delay in detecting possible derangements in vital signs and/or NZEWS Time-consuming, difficult to calculate when staff working with huge cognitive load, interruptions 150 Yes
61 Non-compliance with protocols 150 Yes
62 Staff allocated to the patient is traceable through clinical records data and can be held responsible for this omission 150 Yes
63 10. Escalation of care Revised frequency of observations does not match the NZEWS protocol Delay in detecting possible derangements in vital signs and/or NZEWS Incorrect or less frequent measurements due to previous set of vital signs or NZEWS being miscalculated or staff not able to take measurement due to being busy, distracted or memory lapse 120 Yes
64 Non-compliance with protocols 120 Yes
65 Staff allocated to the patient is traceable through clinical records data and can be held responsible for this omission 120 Yes
66 Secondary responders are not informed about deterioration in a timely manner Delay in detecting possible derangements in vital signs and/or NZEWS Need to use phone or pager or task manager application, multiple devices and technology which creates inconsistency, causes delays in decision making 90 Yes
67 Non-compliance with protocols 90 Yes
68 Staff allocated to the patient is traceable through clinical records data and can be held responsible for this omission 90 Yes
69 Secondary responders are not able to review patient in a timely manner Delay in detecting possible derangements in vital signs and/or NZEWS Limited secondary responder resource, busy responding to other patients elsewhere, not involved in a timely manner 105 Yes
70 Non-compliance with protocols 105 Yes
71 Staff who are paged, tasked, or sent a RRS activation call (777 call) are traceable and could be held accountable for delayed review or response 105 Yes
72 11. Modification of triggers Non documentation of modifications Delay in detecting or respond to possible derangements in vital signs or NZEWS because of the masking effect of modifications Complacency, lack of safety net/defensive barriers and culture 105 Yes
73 Non-compliance with protocols 105 Yes
74 Staff making inappropriate modifications, leaving incomplete documentation of the modifications, or not authorising the modifications at appropriate level of seniority may be held accountable 105 Yes
75 Lack of sufficient space on NZEWS chart to document modifications every 24 hours Delay in detecting or respond to possible derangements in vital signs or NZEWS because of the masking effect of modifications Paper-based charts giving way documentation of modifications elsewhere in clinical records 75 Yes
76 Non-compliance with protocols 75 Yes
77 Staff making inappropriate modifications, leaving incomplete documentation of the modifications, or not authorising the modifications at an appropriate level of seniority may be held accountable 75 Yes
78 Workarounds are common (modifications are validated for entire admission) Delay in detecting or respond to possible derangements in vital signs or NZEWS because of the masking effect of modifications Culture, lack of interdisciplinary dialogue, lack of space on paper-based charts to accommodate frequent modifications, lack of reinforcing functions 60 No
79 Non-compliance with protocols 60 No
80 Staff making inappropriate modifications, leaving incomplete documentation of the modifications, or not authorising the modifications at an appropriate level of seniority may be held accountable 60 No
81 Handwritten modification may be illegible Delay in detecting or respond to possible derangements in vital signs or NZEWS because of the masking effect of modifications Paper-based charts 60 Yes
82 Non-compliance with protocols 60 Yes
83 Staff making inappropriate modifications, leaving incomplete documentation of the modifications, or not authorising the modifications at an appropriate level of seniority may be held accountable 60 Yes
84 12. Call to response time Limited information shared by pager/phone Delay in detecting possible derangements in vital signs and/or NZEWS Technological limitation 30 Yes
85 Non-compliance with protocols 30 Yes
86 Staff who are paged, tasked, or sent a RRS activation call (777 call) are traceable and could be held accountable for delayed review or response 30 Yes
87 Responder cannot access vital signs and NZEWS chart remotely Delay in detecting possible derangements in vital signs and/or NZEWS Paper-based charts and technological limitation of the mode of communication used 30 Yes
88 Non-compliance with protocols 30 Yes
89 Staff who are paged, tasked, or sent a RRS activation call (777 call) are traceable and could be held accountable for delayed review or response 30 Yes
90 Non-documentation of call to response time Delay in detecting possible derangements in vital signs and/or NZEWS Paper-based charts, documentation also paper based and is separated physically from NZEWS charts and likely to be missed in busy environment 90 Yes
91 Non-compliance with protocols 90 Yes
92 Staff who are paged, tasked, or sent a RRS activation call (777 call) are traceable and could be held accountable for delayed review or response 90 Yes
93 13. Secondary response Limited information shared by pager/phone and rarely a phone advice only is appropriate Delay in detecting possible derangements in vital signs and/or NZEWS Paper-based charts and technological limitation of the mode of communication used 30 Yes
94 Non-compliance with protocols 30 Yes
95 Staff who are paged, tasked, or sent a RRS activation call (777 call) are traceable and could be held accountable for delayed review or response 30 Yes
96 Secondary responders do not communicate or initiate actions remotely Delay in detecting possible derangements in vital signs and/or NZEWS Paper-based charts, complex information, not suitable to be effectively communicated by phone call and pager system not capable of passing long messages 60 Yes
97 Non-compliance with protocols 60 Yes
98 Staff who are paged, tasked, or sent a RRS activation call (777 call) are traceable and could be held accountable for delayed review or response 60 Yes
99 Non-documentation of the actions taken by secondary responder Non-availability of the actions undertaken by one staff member to another, leading to delays in response and/or duplication of work Paper-based charts, documentation also paper based and is separated physically from NZEWS charts and likely to be missed in busy environment 45 Yes
100 Non-compliance with protocols 45 Yes
101 Staff who are paged, tasked, or sent a RRS activation call (777 call) are traceable and could be held accountable for delayed review or response 45 Yes

Authors plan to calculate RPN for each failure mode post implementation of electronic RRS for comparison when/if this is possible.

The most common causes of the failure modes (n = 35, 34.7%) were related to human error, memory lapses, lack of reinforcement or reminders. Another 30 (29.7%) failures were related to staffing levels, too busy staff/workload, and work-related interruptions. The third largest group of failures (n = 30, 29.7%) was caused by limitations of paper-based vital signs charts, technological limitations of modes of communication, organisational culture, workarounds, and other organisational factors. A small proportion of failures was caused by mere task complexity (3, 3%) and protocol weaknesses (3, 3%).

The demonstrated electronic system was assessed by the SMEs against each failure to determine whether it could potentially eliminate or reduce the likelihood of those failures. The response of the SMEs was recorded and presented against the process step as summarised in Table 3 which shows that the demonstrated electronic system could potentially eliminate or reduce the likelihood of 71 (70.2%) failures.

Table 3.

Elimination or reduction of failures by demonstrated electronic system across process steps.

Process steps All failures
Eliminated Not eliminate Sub-total
1. Timing of vital signs 0 3 3
2. Heart Rate 6 1 7
3. Blood pressure 6 1 7
4.Temperature 0 7 7
5. Respiratory rate 0 9 9
6. Oxygen saturation 6 0 6
7. Oxygen requirement 6 0 6
8. Level of consciousness 0 6 6
9. NZEWS calculation 11 0 11
10. Escalation of care 9 0 9
11. Modification of calling criteria 9 3 12
12. Accessing afferent inputs (by responders) 9 0 9
13. Timeliness and documentation of rapid response 9 0 9
Total 71 30 101

4. Discussion and recommendations

We reported the first FMEA applied to identify failures, determine their criticality, locate those to the process steps and limbs of RRS, and decide whether those failures could be eliminated or reduced by the electronic system or other specific measures. We found that some process steps of the RRS tend to encounter a higher number of failures whereas failures at other process steps have a higher tendency to be critical failures. We saw the afferent limb of the RRS as a more failure-prone component, and relatively less likely to be rendered free of critical failures by implementation of an electronic system alone.

The root cause analysis of 49 unplanned critical care admissions by van Galen et al. [11] found that 46% of the root causes were human-related, which predominantly included failures within monitoring and interventions. The FMEA presented in the current report shows that 48% of the critical failures and over 34% of the total failures were related to memory lapses/human errors. The majority of failures (>60%) in our study were found within the monitoring/afferent limb of the RRS. Other findings of van Galen et al. are not suitable for comparison with our findings and vice versa, yet van Galen et al. seem to be the only relevant literature for us to compare a small portion of our findings with. This means we present unique insights into RRS through FMEA which will help adoption of electronic RRS to replace the RRS utilizing paper-based vital signs and early warning score charts and offers to point out the failures which may not be addressed by implementing electronic RRS alone, and hence require specific remedial actions at policy or procedure level. Following paragraphs elaborate these points.

As shown in the Results section, the demonstrated electronic system offers eliminating or reducing the likelihood of a majority (71, 70.3%) of the failures within the existing RRS which is driven by paper-based vital signs charts. The electronic system would eliminate or reduce these failures by reducing human error in simple calculations, applying calling criteria and activating MET upon meeting the criteria, and possibly by adding reminders or reinforcements for staff to undertake vital signs observations. The electronic system would reduce delays in response by enabling remote access to vital signs charts and removing the need to locate paper charts in time-critical situations. We recommend using an electronic system similar to the demonstrated electronic system [21] to replace paper-based vital sign charts as it offers mitigation of the majority of failures encountered in the RRS driven by paper charts. Implementing any change should follow evidence-based methods [24] and take into account lessons learnt from similar implementations elsewhere [25] if applicable.

The failures not addressed by the electronic system, e.g., delay in undertaking vital signs; lack of triggers for hypothermia (potential failure to recognise serious illness such as sepsis [26]) and systolic blood pressure (no trigger until SBP is above 220 mmHg [27]); and omissions, lack of measurement and recording of respiratory rate and the level of consciousness were related to the afferent limb of the RRS – another finding that matches the Root Cause Analysis presented by van Galen et al. [11]. We propose that the delay in vital signs observations could possibly be reduced by adding a reminder within the electronic system about the timing of the next set of vital signs observations, or by choosing an electronic system that measures all vital signs automatically. In this regard, the authors have reviewed electronic RRS applications [28] and have provided a summary of the features and functionalities of each application. We propose that the education and training with scenario-based learning and simulation could mitigate the failures associated with assessment of respiratory rate and the level of consciousness. We suggest that the two protocol-related failures (lack of triggers for hypothermia and high SBP) should be considered at the time of future revision of the NZEWS protocols. We think increasing the frequency of internal and external audits, mandating reporting of error and omission rates and patient monitoring might add to the workload – a factor involved in about one third of failures and therefore we do not recommend these measures.

Declarations

Author contribution statement

Ehsan Ullah: Conceived and designed the experiments; Performed the experiments; Analyzed and interpreted the data; Contributed reagents, materials, analysis tools or data; Wrote the paper.

Mirza Mansoor Baig and Hamid GholamHosseini: Conceived and designed the experiments; Wrote the paper.

Jun Lu: Conceived and designed the experiments; Contributed reagents, materials, analysis tools or data; Wrote the paper.

Funding statement

This work was supported by Auckland University of Technology.

Data availability statement

Data will be made available on request.

Declaration of interests statement

The authors declare no conflict of interest.

Additional information

No additional information is available for this paper.

Acknowledgements

Authors acknowledge medical and nursing staff at Taranaki Base Hospital, Taranaki District Health Board, New Plymouth, New Zealand who helped authors complete failure mode and effect analysis.

References

  • 1.Rihari-Thomas J., Newton P.J., Sibbritt D., Davidson P.M. Rapid response systems: where we have come from and where we need to go? J. Nurs. Manag. 2018;26(1):1–2. doi: 10.1111/jonm.12533. [DOI] [PubMed] [Google Scholar]
  • 2.Psirides A., Hill J., Jones D. Rapid Response Team activation in New Zealand hospitals—a multicentre prospective observational study. Anaesth. Intensive Care. 2016;44(3):391–397. doi: 10.1177/0310057X1604400314. [DOI] [PubMed] [Google Scholar]
  • 3.Jones D. The epidemiology of adult rapid response team patients in Australia. Anaesth. Intensive Care. 2014;42(2):213–219. doi: 10.1177/0310057X1404200208. [DOI] [PubMed] [Google Scholar]
  • 4.Kim S.W., Lee H.Y., Han M.R., Lee Y.S., Kang E.H., Jang E.J., et al. Epidemiology and clinical characteristics of rapid response team activations. Kor. J. Critic. Care Med. 2017;32(2):124. doi: 10.4266/kjccm.2017.00199. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 5.Psirides A., Hill J., Hurford S. A review of rapid response team activation parameters in New Zealand hospitals. Resuscitation. 2013;84(8):1040–1044. doi: 10.1016/j.resuscitation.2013.01.022. [DOI] [PubMed] [Google Scholar]
  • 6.Fennessy G., Hilton A., Radford S., Bellomo R., Jones D. The epidemiology of in-hospital cardiac arrests in Australia and New Zealand. Intern. Med. J. 2016;46(10):1172–1181. doi: 10.1111/imj.13039. [DOI] [PubMed] [Google Scholar]
  • 7.Høybye M., Stankovic N., Lauridsen K.G., Holmberg M.J., Andersen L.W., Granfeldt A. Pulseless electrical activity vs. asystole in adult in-hospital cardiac arrest: predictors and outcomes. Resuscitation. 2021;165:50–57. doi: 10.1016/j.resuscitation.2021.05.036. [DOI] [PubMed] [Google Scholar]
  • 8.Nolan J.P., Berg R.A., Andersen L.W., Bhanji F., Chan P.S., Donnino M.W., et al. Cardiac arrest and cardiopulmonary resuscitation outcome reports: update of the Utstein resuscitation registry template for in-hospital cardiac arrest: a consensus report from a task force of the international Liaison committee on resuscitation (American heart association, European resuscitation Council, Australian and New Zealand Council on resuscitation, heart and stroke foundation of Canada, InterAmerican heart foundation, resuscitation Council of southern africa, resuscitation Council of asia) Circulation. 2019;140(18):e746–e757. doi: 10.1161/CIR.0000000000000710. [DOI] [PubMed] [Google Scholar]
  • 9.Bunch J.L., Groves P.S., Perkhounkova Y. Realistic evaluation of a rapid response system: context, mechanisms, and outcomes. West. J. Nurs. Res. 2019;41(4):519–536. doi: 10.1177/0193945918776310. [DOI] [PubMed] [Google Scholar]
  • 10.Heller A.R., Mees S.T., Lauterwald B., Reeps C., Koch T., Weitz J. Detection of deteriorating patients on surgical wards outside the ICU by an automated MEWS-based early warning system with paging functionality. Ann. Surg. 2020;271(1):100–105. doi: 10.1097/SLA.0000000000002830. [DOI] [PubMed] [Google Scholar]
  • 11.van Galen L.S., Struik P.W., Driesen B.E.J.M., Merten H., Ludikhuize J., van der Spoel J.I., et al. Delayed recognition of deterioration of patients in general wards is mostly caused by human related monitoring failures: a root cause analysis of unplanned ICU admissions. PLoS One. 2016;11(8) doi: 10.1371/journal.pone.0161393. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 12.Liu H.-C. Springer; 2019. FMEA for Proactive Healthcare Risk Analysis: A Systematic Literature Review. Improved FMEA Methods for Proactive Healthcare Risk Analysis; pp. 15–45. [Google Scholar]
  • 13.Vázquez-Valencia A., Santiago-Sáez A., Perea-Pérez B., Labajo-González E., Albarrán-Juan M.E. Usefulness of failure mode and effects analysis for improving mobilization safety in critically ill patients. Rev. Colomb. Anestesiol. 2018;46(1):3–10. [Google Scholar]
  • 14.Sharma K.D., Srivastava S. Failure mode and effect analysis (FMEA) implementation: a literature review. J. Adv. Res. Aeronaut. Space Sci. 2018;5:1–17. [Google Scholar]
  • 15.Niv Y., Itskoviz D., Cohen M., Hendel H., Bar-Giora Y., Berkov E., et al. The utility of failure modes and effects analysis of consultations in a tertiary, academic, medical center. Qual. Manag. Health Care. 2018;27(2):69–73. doi: 10.1097/QMH.0000000000000166. [DOI] [PubMed] [Google Scholar]
  • 16.Huang J., You J.-X., Liu H.-C., Song M.-S. Failure mode and effect analysis improvement: a systematic literature review and future research agenda. Reliab. Eng. Syst. Saf. 2020:106885. [Google Scholar]
  • 17.Moradi L., Emami Sigaroudi A., Pourshaikhian M., Heidari M. Risk assessment of clinical care in emergency departments by health failure modes and effects analysis. J. Holistic Nurs. Midwifery. 2020;30(1):35–44. [Google Scholar]
  • 18.Welcn S. 2021. Implementation of a Proactive Rapid Response Team Nurse Rounding Protocol to Address Afferent Limb Failure in a Mature Rapid Response System. [Google Scholar]
  • 19.Smith G.B., Redfern O., Maruotti A., Recio-Saucedo A., Griffiths P., Group T.M.C.S. The association between nurse staffing levels and a failure to respond to patients with deranged physiology: a retrospective observational study in the UK. Resuscitation. 2020;149:202–208. doi: 10.1016/j.resuscitation.2020.01.001. [DOI] [PubMed] [Google Scholar]
  • 20.Devita M.A., Bellomo R., Hillman K., Kellum J., Rotondi A., Teres D., et al. Findings of the first consensus conference on medical emergency teams. Crit. Care Med. 2006;34(9):2463–2478. doi: 10.1097/01.CCM.0000235743.38172.6E. [DOI] [PubMed] [Google Scholar]
  • 21.Vital Signs Monitoring and Decision Support System (VitalsAssist) Precision Driven Health; 2021. [Google Scholar]
  • 22.Rezaei F., Yarmohammadian M.H., Haghshenas A., Fallah A., Ferdosi M. Revised risk priority number in failure mode and effects analysis model from the perspective of healthcare system. Int. J. Prev. Med. 2018;9:7. doi: 10.4103/2008-7802.224046. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 23.Buja A., Manfredi M., De Luca G., Zampieri C., Zanovello S., Perkovic D., et al. Using failure mode, effect and criticality analysis to improve safety in the COVID mass vaccination campaign. Vaccines. 2021;9(8):866. doi: 10.3390/vaccines9080866. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 24.Connelly L.M. Using the PDSA model correctly. Medsurg Nurs. 2021;30(1):61–64. [Google Scholar]
  • 25.Razmak J., Bélanger C.H., Refae G.A.E., Farhan W. Electronic medical records: taking young generations of patients and physicians through innovative technology and change management. Int. J. Electron. Healthc. 2021;11(3):208–238. [Google Scholar]
  • 26.Yamaga S., Kawabata A., Hosokawa K., Shime N. Hypothermia, poorly recognized by clinicians, is associated with higher mortality among critically ill patients with infections: a retrospective observational study. J. Infect. Chemother. 2021;27(3):540–543. doi: 10.1016/j.jiac.2020.12.016. [DOI] [PubMed] [Google Scholar]
  • 27.New Zealand Early Warning Score Vital Sign Chart User Guide 2017 [Internet] Health Quality & Safety Commission; 2017. https://www.hqsc.govt.nz/assets/Deteriorating-Patient/PR/Vital_sign_chart_user_guide_July_2017_.pdf Available from: [Google Scholar]
  • 28.Baig M.M., GholamHosseini H., Afifi S., Lindén M. A systematic review of rapid response applications based on early warning score for early detection of inpatient deterioration. Inf. Health Soc. Care. 2021:1–10. doi: 10.1080/17538157.2021.1873349. [DOI] [PubMed] [Google Scholar]