. 2022 Feb 23;148:105121. doi: 10.1016/j.jcv.2022.105121

A systematic review and meta-analysis of the accuracy of SARS-COV-2 IGM and IGG tests in individuals with COVID-19

Ana Cristina Lacerda Macedo a,, Gabriele da Silveira Prestes a, Tamy Colonetti a, Ana Claudia Rodrigues Candido a, Maria Laura Rodrigues Uggioni a, Ariane Cruz Gomes b, Fábio Rosa Silva a, Luciane Bisognin Ceretta c, Antonio José Grande d, Maria Inês da Rosa a,⁎⁎
PMCID: PMC8863416  PMID: 35245882

Abstract

Introduction

: Active SARS-CoV-2 infection is confirmed mainly through the detection of viral nucleic acid via the reverse transcriptase polymerase chain reaction (RT-PCR) technique. Methods to assess humoral responses contribute to the monitoring of the disease and confirmation of exposure to the virus.

Objective

: To evaluate the accuracy of tests for IgM and IgG antibodies for SARS-CoV-2 infection confirmed by RT-PCR and utility as complementary data for immunosurveillance.

Methods

: Literature research was performed by searching the terms “COVID-19”, “COVID-19 diagnostic testing” and “test” in the databases MEDLINE, EMBASE, Cochrane Library, Web of Science and Cumulative Index to Nursing and Allied Health Literature to search for potentially eligible observational studies without language restrictions published up to September 2020.

Results

: The pooled sensitivity and specificity, regardless of collection moment, was 80.0% (CI 95% 72.0–86.0) and 97.0% (CI 95% 94.0–98.0) for “IgM and/or IgG”, respectively. Serology considering immunoglobulins M and G together had a high accuracy performance on “fifteenth day and after”: sensitivity and specificity was 91.0% (CI 95% 85.0–94.0) and 98.0% (CI 95% 95.0–99.0) respectively, DOR 461 and AUC 0.98.

Conclusion

: This study shows that serology is a group of tests with high accuracy, mainly following the second week after infection.

Keywords: SARS-CoV-2, Diagnostic meta-analysis, Systematic review, Sensitivity, Specificity

1. Introduction

In December 2019, a new variety of coronaviruses, officially called Severe Acute Respiratory Syndrome by Coronavirus-2 (SARS-CoV-2) was identified through three patients who had atypical pneumonia in Wuhan, China [1,2]. The rapid spread and sustained transmission of the virus, gave rise to a pandemic within three months of identification of the virus [3]. Worldwide, SARS-Cov-2 has infected more than 146 billion people, with more than 3 billion deaths (data obtained on April, 25, 2021) [4]. Due to the overlap of manifestations, clinical diagnosis is problematic, especially during seasonal flu [3].

Confirmation of infection occurs mainly through the detection of the SARS-CoV-2 nucleic acid via the reverse transcriptase polymerase chain reaction (RT-PCR) technique [3]. However, this test can only detect active infection and has variable accuracy through the course of infection and therefore does not provide a complete picture of the dynamics of infection. Additionally, RT-PCRs requires laboratories with specific equipment and highly trained technicians to perform the test [5]. Moreover, this technique requires the presence of sufficient quantities of the viral genome at the sample site, and therefore an incorrect sample collection can limit the usefulness of the quantitative method [6]. To date, a number of studies have shown that the RNA of the virus in swab samples from the upper respiratory tract is detectable by PCR until 7 days after first symptoms, on average [7]. Testing programs in large-scale, rapid diagnosis and isolation associated with rigorous tracking and preventive self-isolation of close contacts are essential measures to reduce the burden of the COVID-19 pandemic.

The quick and accurate diagnosis of COVID-19 contributes to the management of diseases and outbreaks, allowing health surveillance to act in preventing and generating new public health measures [7]. Under these conditions, additional screening methods that can detect convalescent infection contribute to monitoring of the virus by informing the attack rate in the population, and the infection fatality rate. Such screening is commonly performed via serological tests.

Serological tests are common within laboratory routine and identify infection through the detection of viral antigens or circulating antibodies. The test for the detection of specific SARS-CoV-2 antibodies in the blood of a patient offers a simpler, cheaper and faster diagnosis [5,7]. Antigen-specific immunoglobulin M (IgM) is the first subtype to be detected in the blood, followed by the production of immunoglobulin G (IgG). Thus, serological test for SARS-CoV-2 can be used in the rapid screening of Covid-19, identifying symptomatic or asymptomatic carriers, on a large scale [5]. Therefore, the objective of this systematic review is to evaluate the accuracy of serology for SARS-CoV-2 infections in cases confirmed or not by RT-PCR.

2. Methods

This is a systematic review and meta-analysis synthesizing scientific evidence of the accuracy of IgM and IgG antibody tests (serology) for SARS-CoV-2 infection diagnosis compared to RT-PCR.

A protocol with explicitly defined objectives and methods was registered in the International Platform of Registered Systematic Review and Meta-analysis Protocols (registration number INPLASY202040099, doi: 10.37766/inplasy2020.4.0099) available online. We have reported the systematic review and meta-analysis in accordance with Preferred Reporting Items for Systematic Reviews and Meta-analyses – PRISMA statement [8].

2.1. Eligibility criteria

  • We included studies with patients who tested for SARS-Cov-19 using RT-PCR and who were evaluated using the IgM and IgG Antibody Test for SARS-CoV-2 Infection.

Population: patients tested by RT-PCR for SARS-CoV-2 and by a serologic test.

Intervention: IgM and IgG antibody test for SARS-CoV- 2 Infection.

  • Comparator: RT-PCR (reference test).

    Outcome: SARS-CoV-2 Infection.

  • Study designs: Cross-sectional, Cohort or Case-control.

2.2. Search strategy

A search strategy was developed using the following terms, both as text words and, as appropriate, Medical Subjects Heading (MeSH) or equivalent subject heading/thesaurus terms: “SARS-CoV” OR “COVID-19” OR “COVID” OR “COVID-19 diagnostic testing” AND “test”. The following databases were searched from their inception forwards for potentially eligible studies without language restrictions published up to September 2020: MEDLINE, EMBASE, Cochrane Library, Web of Science and Cumulative Index to Nursing and Allied Health Literature (CINAHL Database). Cross-referencing from retrieved studies was also conducted. We also performed searches for unpublished or ongoing trials. The search was performed on Octuber 2020. Before completing this review, we performed an additional search in each database and registration platform to guarantee that the most recent studies were included.

2.3. Studies selection

Two reviewers (ACM, GSP) screened all titles and abstracts for relevance in Rayyan (rayyan.qcri.org). Full texts of each potentially eligible study were retrieved and reviewed independently by the two reviewers (ACM, GSP). Disagreements between reviewers were resolved by a third reviewer (TC).

2.4. Data extraction

Data from all studies was independently extracted by two reviewers (ACM, GSP), and combined to construct a definitive dataset. Data was extracted from pre-piloted data extraction tables for study characteristics and placed on a 2 × 2 table for study results. However, when this was not possible, corresponding authors were contacted in order to obtain this data.

2.5. Quality assessment

All studies were assessed for their methodological quality using the QUADAS 2 (Quality Assessment of Diagnostic Accuracy Studies) criteria [9]. This was performed independently by two reviewers (TC and ACM). A figure representing each appraisal was provided in the results section.

2.6. Data synthesis

The primary analysis assessed the pooled sensitivity, specificity, diagnostic odds ratio (DOR), ROC curves, and negative and positive predictive values. Meta-analysis was conducted using the random effects model implemented with Metadisc 1.4 and STATA 16 softwares[10, 11]. We calculated the point estimate and 95% confidence interval of pooled sensitivity and pooled specificity data.

3. Results

3.1. Study identification and eligibility

Among the 1885 studies identified from electronic database searches and reference lists, 112 studies were excluded by duplication, leaving 1773 studies for reading titles and abstracts, and then we excluded 1703 published studies through title and abstract screening (Fig. 1 ). A total of 70 full-text studies were retrieved. Of those, 57 studies were excluded after further scrutiny: 9 studies included only confirmed cases, 31 were different design or editorial, 17 did not have serology and 1 did not have data for extraction. We included a total of 45 studies in the review (Li et al., 2020 [4], Guo et al., 2020[5], Jin et al., 2020[7], Adams et al., 2020 [12], Andrey et al., 2020 [13], Algaissi et al., 2020 [14], Ayouba et al., 2020 [15], Benavid et al., 2020 [16], Cassaniti et al., 2020 [17], Chan et al., 2020[18], Charlton et al., 2020 [19], Dohla et al. 2020 [20], Freeman et al., 2020 [21], Guedez-Lopez et al., 2020 [22], Halsemann et al., 2020 [23], Harb et al., 2020 [24], Herroelen et al., 2020 [25], Jia et al. 2020 [26], Lassauniere et al., 2020 [27], Lin et al. 2020 [28], Liu et al., 2020 (A) [29], Liu et al., 2020 (B)[30], Liu et al., 2020 (C) [31], Long et al., 2020 [32], Marinis et al., 2020[33], Naaber et al., 2020 [34], Pan et al., 2020 [35], Pancrazzi et al., 2020 [36], Paiva et al., 2020 [37], Paradisoet al., 2020 [38], Pegoraro et al., 2020 [39], Perera et al., 2020 [40], Perez-Garcia et al., 2020 [41], Pfluger et al., 2020 [42], Pieri et al., 2020 [43], Phipps et al., 2020 [44], Suhandynata et al., 2020 [45], Tan et al., 2020 [46], Tehrani et al., 2020 [47], Turbett et al., 2020 [48], Vásárhelyi et al., 2020 [49], Wang et al., 2020 [50], Yassine et al., 2020 [51], Zhao et al., 2020 [52], Zhang et al., 2020 [53], Zhong et al., 2020[54]). A complete list of excluded studies is available upon request.

Fig. 1.

Fig 1

PRISMA flow diagram of studies selections.

3.2. Study descriptions

Fourty-five primary studies were included in the analyses. A total of 13,650 patients met the inclusion criteria and were analyzed, including suspected cases, confirmed cases and healthy volunteers. The main characteristics of the included studies and samples are shown in Table 1 Table 2. shows the sum contingency table and analysis.

Table 1.

Characteristics of included studies.

Study Country Samples Ntotal N PCR+ N PCR- Antibody test Agemedia (range)years Collect periodmedia (range)Days
Adams et al., 2020 United Kingdom confirmed cases + pre-COVID samples 90 40 50 ELISA
LFIA
NI NI
Andrey et al., 2020 Switzerland confirmed cases + healthy volunteers (not PCR tested) 91 46 45 Augurix RDT, rIFA
Euroimmun (ELISA)

66 (50–76)
10 (5 −15)
discriminated by
(0–6) (7–14) (>15)
Algaissi et al., 2020 Saudi Arabia confirmed cases + pre-COVID samples 135 10 125 ELISA S1 / ELISA N 24 a 75
discriminated by
(0–6) (7–14) (>15)
Ayouba et al., 2020 France confirmed cases + pre-COVID samples 138 61 77 xMAP assay IgG 72 (33–99) >14
Benavid et al., 2020 USA confirmed cases +
pre-COVID samples
408 37 371 Premier Biotech (LFIA) NI (Zero - >65) NI
Cassaniti et al., 2020 Italy suspected cases +
healthy volunteers
110 68 42 Viva-DiagTM (LFIA) IU 73.5 (38–86)
ER 61.5 (33–97)
healthy: 38.5 (25–69)
7 (4–11)
Chan et al., 2020 USA confirmed cases + pre-COVID samples 155 99 56 Easy Check (LFIA) NI discriminated by
(0–6) (7–13) (>14)
Charlton et al., 2020 Canada confirmed cases + pre-COVID samples 92 42 50 Abbott IgG, Affinity, Bio-Rad (ELISA)
BTNX, Biolidics, Deep blue, Getein, Genrui
(LF (POC))

70.1 (34–102)
discriminated by
(0–14) (14–21) (>21)
Dohla et al. 2020 Germany suspected cases 49 22 27 rapid test 46 18,5
Freeman et al., 2020 USA confirmed cases + pre-COVID samples with virus infections 99 519 618 ELISA NI >10
Guedez-Lopez et al., 2020 Spain suspected cases 145 101 44 Sienna rapid Test (ICT)
Wondfo (ICT)
Prometheus (ICT)
50 Professionals 5 Patients 11
Guo et al., 2020 China suspected cases 140 82 58 ELISA rN NI (1–39)
Halsemann et al., 2020 Germany confirmed cases + healthy volunteers PCR negative 51 26 25 Epitope Ratio (ELISA)
Euroimmun (ELISA)
Elecsys
48 (20–73) NI
Harb et al., 2020 EUA confirmed cases + pre-COVID samples 447 65 382 Abbott Architect i2000
LIAISON®
Elecsys on cobas® e601
NI 2–45
Herroelen et al., 2020 Belgium confirmed cases + pre-COVID samples 225 169 56 Euroimmun (ELISA)
Diasorin
immunochromatographic Gene S/N and Innovita
71 (53–86) 0–39
Jia et al. 2020 China suspected cases 57 24 33 ICT NI NI
Jin et al., 2020 China suspected cases 66 43 33 chemiluminescence immunoassay,
Shenzhen YHLO Biotech
47 (7–74) 18 (11–23)
Lassauniere et al. 2020 Denmark confirmed cases + pre-COVID samples with virus infections 112 82 30 Euroimmun (ELISA)
POC: Dynamiker,
CTK Biotech
Autobio diagnostics, Artron, Acro Biotech, Alltest biotech
Controls: 18–64
Cases: NI
discriminated by
(0–6) (7–13)
(14–20) (>21)
Li et al., 2020 China suspected cases 525 397 128 Jiangsu Medomics (LFIA) NI (8–33)
Lin et al. 2020 China confirmed cases + healthy volunteers + samples with pulmonar infection 159 79 80 CLIA and ELISA 42 13
Liu et al., 2020 (A) China confirmed cases +
healthy volunteers
314 214 100 Lizhu (ELISA rN)
Hotgen (ELISA rS)
NI 15 (0–55)
discriminated by
(0–5) (6–15) (>16)
Liu et al., 2020 (B) China suspected cases 179 90 89 ICT controls: 56
cases: 76
NI
Liu et al., 2020 (C) China suspected cases 238 153 85 ELISA 55 (38–65) 0 - >16
Long et al., 2020 China confirmed cases +
close contacts
449 301 148 Bioscience (MCLIA) 47 (34–56) (0–19)
Marinis et al., 2020 Sweden confirmed cases + pre-COVID samples 76 37 39 LFIA Controls: 37,4 ± 8,3 Cases: 71 ± 8 >14
Naaber et al., 2020 Estonia confirmed cases + pre-COVID samples 197 97 100 SNIBE (CLIA)
Euroimmun (ELISA)
Abbott (CMIA)
Epitope Ratio (ELISA)
Diasorin (MCLIA)
Biosensor (ICT)
Lips-S-RBD Luciferase (ELISA)
Lips-N Luciferase (ELISA)
59 (21–100) >7
Pan et al., 2020 China
(Hong Kong)
suspected cases 38 31 7 GICA 58 (20–96) Discriminated by
(1–7) (8–14) (>15)
Pancrazzi et al., 2020 Italy suspected cases 516 73 413 Acro Biotech (POC) 53,7 NI
Paiva et al., 2020 USA confirmed cases + pre-COVID samples + healthy volunteers 1212 28 1184 STANDARD Q LFIA Abbott (CMIA)
Wondfo LFIA
59.5 ± 1.9 >14 dias
Paradisoet al., 2020 Italy suspected cases 190 120 70 Viva-DiagTM (LFIA) 58.5 0 - >15
Pegoraro et al., 2020 Italy suspected cases 226 159 67 Euroimmun (ELISA)
Maglumi (CLIA)
VivaDiag (ICT)
PRIMA (ICT)
control 49 ± 9
cases 58 ± 20
Discriminated by
Assyntomatic
(0–5) (6–8) (9–10)
(11–15) (>15)
Perera et al., 2020 China
(Hong Kong)
confirmed cases +
healthy volunteers
224 24 200 ELISA rS (28–80) Discriminated by
(5–9) (11–18)
Perez-Garcia et al., 2020 Spain confirmed cases +
pre-COVID samples + samples with pulmonar infection
163 55 108 AllTest Biotech, ICT 62 (51–74) Discriminated by
(1–7) (8–14) (>15)
Pfluger et al., 2020 Germany confirmed cases + pre-COVID samples 357
37
320
Euroimmun ELISA)
Diasorin (MCLIA)
Wantai (ELISA)
Siemens (MCLIA)
18–70 Discriminated by
(1–10) (>10)
Pieri et al., 2020 Italy suspected cases 80 40 40 SNIBE (CLIA)
Euroimmun (ELISA)
NI Discriminated by
(1–10) (11–45)
Phipps et al., 2020 USA suspected cases 172 76 97 Abbott 06R86 (CMIA) NI Discriminated by
(1–7) (8–14) (>15)
Suhandynata et al., 2020 USA suspected cases 289 54 235 Diazyme (MCLIA) (25–91) Discriminated by
(1–7) (8–14) (>15)
Tan et al., 2020 Singapure confirmed cases + pre-COVID samples 336 173 163 Abbott (CMIA)
Siemens (ELISA)
Beckman (CMIA)
Abbott (CMIA)
NI Discriminated by
(1–6) (7–13)
(14–20) (21–64)
Tehrani et al., 2020 EUA confirmed cases + pre-COVID samples 400 100 300 Chembio (LFIA) NI NI
Turbett et al., 2020 USA confirmed cases + pre-COVID samples 1338 70 1268 Abbott (CMIA) NI Discriminated by
(1–7) (8–14) (>15)
Vásárhelyi et al., 2020 Hungary workers screening 1029 31 998 Clungene (LFIA) NI NI
Wang et al., 2020 China confirmed cases +
excluded cases with comorbidities
86 14 72 ELISA NI (3–7)
Yassine et al., 2020 Qatar confirmed cases + pre-COVID sample 103 33 70
Epitope Ratio, AnshLabs
DiaPro, Nova tec
Lionex (ELISA)
controls: 36 (30–45)
cases: 48 (40–57)
Discriminated by
(1–6) (7–13) (>14)
Zhao et al., 2020 China confirmed cases +
healthy volunteers
481 69 412 ELISA NI NI
Zhang et al., 2020 China not specified 782 122 660 GICA NI NI
Zhong et al., 2020 China confirmed cases + healthy volunteers 347 47 300 ELISA rS- RBD
CLIA
48 (18–82) NI

Footnote: CLIA: Chemiluminescence ImmunoAssay; CMIA: Chemiluminescent Magnetic Immunological Assay; ELISA rN: enzyme-linked immunosorbent assay recombinant nucleocapsid protein-based kit; ELISA rS: enzyme-linked immunosorbent assay RBD of the recombinant S polypeptide-based kit; ELISA: Enzyme-linked immunossorbent assay; GICA: SARS-COV-2 ANTIGEN; ICT: immunochromatographic strip rapid serology tests; LFIA: Lateral Flow Immunoassay; MCLIA: magnetic chemiluminescence enzyme immunoassay; NI: Not Informed; PCR: Polymerase Chain Reaction; rIFA: rapid Immunofiltration Assay.

Table 2.

Accuracy of serology, reference standard: PCR. All studies pooled and discriminated by collect period and technique. Different techniques were not applied to the same sample.

All studies
Analysis IgM and/or IgG% (CI 95%) IgM only% (CI 95%) IgG only% (CI 95%)
Sensitivity 80.0 (72.0–86.0) 48.5 (52.0–65.5) 64.9 (53.0–75.2)
Specificity 97.0 (94.0–98.0) 95.6 (90.8–98.0) 97.3 (95.2–98.5)
DOR 131 (56–303) 20.59 (0.10–41.09) 66.76 (8.94–124.58)
AUC 0.96 (0.93–0.97) * *
PPV 84.5 (83.3 – 85.7) 74.5 (71.9–76.9) 85.8 (83.9–87.4)
NPV 90.1 (89.5 – 90.7) 83.0 (82.0–84.0) 87.5 (86.7- 88.2)
TP 2806 861 1305
FP 516 299 220
FN 982 802 893
TN 9346 4102 6448
N total 13,650 6064 8866
Studies included 45 24 30
Discriminated by collect period
Analysis Around 0–6th dayIgM and/or IgG% (CI 95%) Around 7th-14th daysIgM and/or IgG% (CI 95%) Around 15th day and afterIgM and/or IgG% (CI 95%)
Sensitivity 42.0 (24.0–63.0) 65.0 (64.9–77.3) 91.0 (85.0–94.0)
Specificity 97.0 (93.0–99.0) 97.0 (93.0–99.0) 98.0 (95.0–99.0)
DOR 25 (8–81) 72 (26–200) 461 (139–1534)
AUC 0.90 (0.87–0.92) 0.91 (0.88–0.93) 0.98 (0.96–0.99)
PPV 68.0 (61.5–73.9) 85.5 (81.7–88.6) 92.0 (90.2–93.5)
NPV 92.0 (90.9–92.9) 92.8 (91.7–93.7) 97.2 (95.7–96.7)
TP 144 340 946
FP 68 59 83
FN 199 181 102
TN 2324 2497 4156
N total 2735 3077 5287
Studies included 12 13 19
Discriminated by technique
Analysis ELISAIgM and/or IgG% (CI 95%) LFQIIgM and/or IgG% (CI 95%) MCLIAIgM and/or IgG% (CI 95%) Immunochroma-tographicIgM and/or IgG% (CI 95%)
Sensitivity 82.0 (72.0–89.0) 75.0 (59.0–87.0) 85.0 (71.0–93.0) 72.0 (48.0–88.0)
Specificity 96.0 (91.0–98.0) 97.0 (94.0–99.0) 98.0 (97.0–99.0) 88.0 (62.0–97.0)
DOR 112 (36–344) 115 (29–452) 334 (123–904) 19 (5–77)
AUC 0.95 (0.93–0.97) 0.96 (0.94–0.98) 0.99 (0.98–1.00) 0.86 (0.83–0.89)
PPV 86.0 (84.2–87.7) 94.1 (93.4–94.8) 96.6 (95.2–97.5) 78.4 (74.8–81.6)
NPV 85.1 (83.9–86.3) 91.3 (90.5–91.9) 87.6 (86.4–88.7) 77.2 (75.6–79.4)
TP 1330 834 906 410
FP 215 228 32 120
FN 531 277 371 256
TN 3048 4463 2634 972
N total 2072 5802 3943 758
Studies included 23 13 15 8
Discriminated by characteritics of patients included
Analysis Confirmed cases + pre-COVID samples OR healthy volunteers(PCR tested or not) Suspected cases
Sensitivity 83.0 (75.0–90.0) 74.0 (53.0–88.0)
Specificity 99.0 (98.0–99.0) 89.0 (73.0–96.0)
DOR 456 (214–972) 24 (14–57)
AUC 0.99 (0.98–1.00) 0.89 (0.86–0.92)
PPV 95.8 (94.6–96.7) 85.2 (83.3–86.9)
NPV 93.0 (92.3–93.6) 72.8 (70.8–74.8)
TP 1286 1217
FP 56 211
FN 414 531
TN 5682 1425
N total 7438 3384
Studies included 23 17
INCLUDED STUDIES: ALL STUDIES – IgM and/or IgG (Adams et al., 2020; Andrey et al., 2020; Algaissi et al., 2020; Ayouba et al., 2020; Benavid et al., 2020; Cassaniti et al., 2020; Chan et al., 2020; Charlton et al., 2020; Dohla et al. 2020; Freeman et al., 2020; Guedez-Lopez et al., 2020; Guo et al., 2020; Halsemann et al., 2020; Harb et al., 2020; Herroelen et al., 2020; Jia et al. 2020; Jin et al., 2020; Lassauniere et al. 2020; Li et al., 2020; Lin et al. 2020; Liu et al., 2020 (A); Liu et al., 2020 (B); Liu et al., 2020 (C); Long et al., 2020; Marinis et al., 2020; Naaber et al., 2020; Pan et al., 2020; Pancrazzi et al., 2020; Paiva et al., 2020; Paradisoet al., 2020; Pegoraro et al., 2020; Perera et al., 2020; Perez-Garcia et al., 2020; Pfluger et al., 2020; Pieri et al., 2020; Phipps et al., 2020; Suhandynata et al., 2020; Tan et al., 2020; Tehrani et al., 2020; Turbett et al., 2020; Vásárhelyi et al., 2020; Wang et al., 2020; Yassine et al., 2020; Zhao et al., 2020; Zhang et al., 2020; Zhong et al., 2020); IgM only (Andrey et al., 2020; Algaissi et al., 2020; Charlton et al., 2020; Guedez-Lopez et al., 2020; Guo et al., 2020; Herroelen et al., 2020; Jia et al., 2020; Jin et al., 2020; Li et al., 2020; Lin et al., 2020; Liu et al., 2020 (A); Liu et al., 2020 (B); Marinis et al., 2020; Pan et al., 2020; Pancrazzi et al., 2020; Paiva et al., 2020; Pegoraro et al., 2020; Perez-Garcia et al., 2020; Pieri et al., 2020; Phipps et al., 2020; Suhandynata et al., 2020; Tehrani et al., 2020; Vásárhelyi et al., 2020; Wang et al., 2020); IgG only (Andrey et al., 2020; Algaissi et al., 2020; Charlton et al., 2020; Guedez-Lopez et al., 2020; Halsemann et al., 2020; Harb et al., 2020; Herroelen et al., 2020; Jia et al., 2020; Jin et al., 2020; Lassauniere et al., 2020; Li et al., 2020; Lin et al., 2020; Liu et al., 2020 (A); Liu et al., 2020 (B); Marinis et al., 2020; Naaber et al., 2020; Pan et al., 2020; Pancrazzi et al., 2020; Paiva et al., 2020; Pegoraro et al., 2020; Perez-Garcia et al. 2020; Pfluger et al., 2020; Pieri et al., 2020; Phipps et al., 2020; Suhandynata et al., 2020; Tan et al., 2020; Tehrani et al., 2020; Turbett et al., 2020; Vásárhelyi et al., 2020; Yassine et al., 2020); DISCRIMINATED BY COLLECT PERIOD – Around 0–6th day (Andrey et al., 2020; Algaissi et al., 2020; Charlton et al., 2020; Liu et al., 2020 (A); Pan et al., 2020; Perez-Garcia et al., 2020; Phipps et al., 2020; Suhandynata et al., 2020; Tan et al., 2020; Turbett et al., 2020; Wang et al., 2020; Yassine et al., 2020); Around 7th-14th days (Andrey et al., 2020; Algaissi et al., 2020; Cassaniti et al., 2020; Guedez-Lopez et al., 2020; Liu et al., 2020 (A); Pan et al., 2020; Perera et al., 2020; Perez-Garcia et al., 2020; Phipps et al., 2020; Suhandynata et al., 2020; Tan et al., 2020; Turbett et al., 2020; Yassine et al., 2020); Around 15th day and after (Andrey et al., 2020; Algaissi et al., 2020; Ayouba et al., 2020; Chan et al., 2020; Charlton et al., 2020; Jin et al., 2020; Li et al., 2020; Liu et al., 2020 (A); Marinis et al., 2020; Pan et al., 2020; Paiva et al., 2020; Perera et al., 2020; Perez-Garcia et al., 2020; Pfluger et al., 2020; Phipps et al., 2020; Suhandynata et al., 2020; Tan et al., 2020; Turbett et al., 2020; Yassine et al., 2020); DISCRIMINATED BY TECHNIQUE – ELISA (Adams et al., 2020; Andrey et al., 2020; Algaissi et al., 2020; Charlton et al., 2020; Freeman et al., 2020; Guo et al., 2020; Halsemann et al., 2020; Herroelen et al., 2020; Lassauniere et al., 2020; Lin et al., 2020; Liu et al., 2020 (A); Liu et al., 2020 (C); Naaber et al., 2020; Pegoraro et al., 2020; Perera et al., 2020; Pfluger et al., 2020; Pieri et al., 2020; Tan et al., 2020; Tehrani et al., 2020; Yassine et al., 2020; Wang et al., 2020; Zhao et al., 2020; Zhong et al., 2020); LFQI (Adams et al., 2020; Benavid et al., 2020; Cassaniti et al., 2020; Chan et al., 2020; Charlton et al., 2020; Guedez-Lopez et al., 2020; Li et al., 2020; Marinis et al., 2020; Paiva et al., 2020; Paradisoet al., 2020; Tehrani et al., 2020; Turbett et al., 2020; Vásárhelyi et al., 2020); MCLIA (Halsemann et al., 2020; Harb et al., 2020; Herroelen et al., 2020; Lin et al., 2020; Jin et al., 2020; Long et al., 2020; Naaber et al., 2020; Paiva et al., 2020; Pegoraro et al., 2020; Pfluger et al., 2020; Pieri et al., 2020; Phipps et al., 2020; Suhandynata et al., 2020; Tan et al., 2020; Zhong et al., 2020); Immunochromatographic (Herroelen et al., 2020; Jia et al., 2020; Liu et al., 2020 (B); Pan et al., 2020; Pegoraro et al., 2020; Perez-Garcia et al., 2020; Wang et al., 2020; Zhang et al., 2020); DISCRIMINATED BY CHARACTERITICS OF PATIENTS INCLUDED – Confirmed cases + pre-COVID samples OR healthy volunteers (Adams et al., 2020; Algaissi et al., 2020; Andrey et al., 2020; Ayouba et al., 2020; Benavid et al., 2020; Chan et al., 2020; Charlton et al., 2020; Halsemann et al., 2020; Harb et al., 2020; Herroelen et al., 2020; Liu et al., 2020 (A); Marinis et al., 2020; Naaber et al., 2020; Paiva et al., 2020; Perera et al., 2020; Pfluger et al., 2020; Tan et al., 2020; Tehrani et al., 2020; Turbett et al., 2020; Yassine et al., 2020; Zhao et al., 2020; Zhong et al., 2020); Suspected case (Cassaniti et al., 2020; Dohla et al., 2020; Guedez-Lopez et al., 2020; Guo et al., 2020; Jia et al., 2020; Jin et al., 2020; Li et al., 2020; Liu et al., 2020 (B); Liu et al., 2020 (C); Long et al., 2020; Pan et al., 2020; Pancrazzi et al., 2020; Paradisoet al., 2020; Pegoraro et al., 2020; Pieri et al., 2020; Phippse et al., 2020; Suhandynata et al., 2020).

Legend: FLQI: lateral flow qualitative immunoassay; ELISA rN: enzyme-linked immunosorbent assay recombinant nucleocapsid protein-based; MCLIA: magnetic chemiluminescence enzyme immunoassay.

3.3. Quality assessment

QUADAS-2 was performed considering the following categories: selection of patients, index and reference test, flow and timing (Fig. 2 ). Regarding selection, three studies was considered "unclear"[15, 16, 54]. Regarding index fourteen studies were considered "unclear" [12, 14, 15, [19], [20], [21], [22], 24, 43, [46], [47], [48], 51, 54], since these studies did not described blinding of workers involved in test collection and analyses, but it was considered as a source of "low concern".  Regarding the reference tests, all studies received the same evaluation; limitations of PCR as a reference test are mentioned in the discussion. Regarding flow and timing, twenty-two studies performed "unclear" [7, 12, 13, 15, [21], [22], [23], [24], [25], 27, [32], [33], [34], [40], [41], [42], 44, [46], [47], [48], 51, 54] because did not show enough information about intervals between tests.

Fig. 2.

Fig 2

Assessment of risk of bias.

3.4. Characteristics of included studies

Eighteen studies were from Europe [12, 13, 15, 17, 20, 22, 23, 25, 27, 33, 34, 36, 38, 39, [41], [42], [43], 49]; China was the origin of fifteen studies [4, 5, 7, 26, [28], [29], [30], [31], [32], 35, 40, 50, [52], [53], [54]], two of these from Hong Kong [35,40]; there were nine studies from USA (United States of America) [16, 18, 21, 24, 37, 44, 45, 47, 48] and one from each of follow: Canada [19], Singapore [16], Qatar [51] and Saudi Arabia [14]. Regarding technique, thirteen studies used lateral flow qualitative immunoassay (LFQI) [4,12,[16], [17], [18], [19], 22, 33, 37, 38, [47], [48], [49]] by different manufactures; Twenty-three studies applied Enzyme-Linked Immunosorbent Assay (ELISA), using recombinant nucleocapsid protein-based (rN) and recombinant S polypeptide-based kit (rS) [5,[12], [13], [14],19,21,23,25,[27], [28], [29],31,34,39,40,42,43,46,47,[50], [51], [52],54]. Fifteen used chemiluminescence immunoassay (MCLIA) [7, [23], [24], [25], 28, 32, 34, 37, 39, [42], [43], [44], [45], [46], 54], by different manufactures, and eight used Colloidal gold-based immunochromatographic (ICG) strip assay [41, 25, 26, 30, 35, 39, 50, 53].

3.5. Accuracy of serology – IGM and IGG

Results were discriminated, as shown in primary studies, by immunoglobulin (“IgM” and “IgG” separately and “IgM, IgG or both” reactivity) but some kits only tested both subclasses together (“IgM, IgG or both”). The pooled sensitivity and specificity, regardless of collection moment, was 80,0% (95% CI 72.0–86.0) and 97.0% (95% CI 94.0–98.0) for IgM and/or IgG, respectively. Serology considering immunoglobulins M and G together had a high accuracy performance, as a DOR (Diagnostic Odds Ratio) 131 and AUC (Area Under the Curve) 0.96. NPV on this sample was 90.1% (95% CI 89.5 – 90.7) and PPV, 84.5% (95% CI 83.3 – 85.7). This data and additional ones are described in Table2.

3.6. Sensitivity analysis

We considered the importance of describing the results divided by technique, manufacturer, sample selection, and mainly by period of blood collection, considering the number of days since the first symptoms. However, analysis by brand was not possible since primary studies were done using a variety of them.

These analyses are explained on Table 2. A comparison of the techniques employed was not possible since most primary studies did not report results from more than one test applied to the same sample results shown on Table 2 must be considered separated one from de other. Considering the blood collection period, in cases where it happened around the fifteenth day and after, the test presented a better performance, with pooled sensitivity of 91.0% (95% CI 85.0–94.0), specificity of 98.0% (95% CI 95.0–99.0), DOR 461 and AUC 0.98, for immunoglobulins M and G together – Fig. 3;. Areas Under the Curve comparing performance evolution by blood collection period are on Fig. 4 . Subanalyses of rapid diagnostics test for point-of-care were not performed since in primary studies that used these tolls the sample was not whole blood, as in practice, but serum or plasma.

Fig. 3.

Fig 3

Forest plot.

Fig. 4.

Fig 4

SROC curve: summary receiver operating characteristics curve.

Subanalyses of rapid diagnostics test for point-of-care were not performed since in primary studies that used these tolls the sample was not whole blood, as in practice, but serum or plasma.

An important find was the subanalyses by “characteristics of patients included”. Mostly studies included only “suspected cases”, a cohort that configures a diagnostic scenario, and “confirmed cases + pre-COVID samples, or healthy volunteers (PCR tested or not)”, characterizing a theorical scenario with “true positives” and “true negatives” considering clinical feature. The AUC rised from 0.89 in “suspected cases” to 0.99 in “confirmed cases + pre-COVID samples”. It suggests that well-established clinical criteria screening will improve tests performance. Complete analysis shown on Table2.

3.7. Heterogeneity

The I² index aims to quantify the dispersion of effect sizes in a meta-analysis.

For the main analyses, including all techniques, I2 was substantial heterogeneity in sensitivity and specificity was found among studies in most analyses. In the Main analysis, “IgM, IgG or both all techniques pooled”, I2 was 96.56 (95%CI 96.02–97.10) Q 1337.77 (df 46.0) for sensitivity and 99.07 (95%CI 98.98–99.17) Q 4949.38 (df 46.0) for specificity. Regarding the period of sample collection, from 7th to 14th days, less heterogeneity was found, I2 was 84.80 (95%CI 77.85–91.75) Q 85.52 (df 13.0) for sensitivity and 96.57 (95%CI 95.96–97.58) Q 378.92 (df 13.0) for specificity; and to 15th day onwards I2 was 89.01 (95%CI 85.21–82.81) Q 172.88 (df 19.0) for sensitivity and 98.81 (95%CI 98.61–99.02) Q 1599.86 (df 19.0) for specificity. Egger's regression identified p = 0.00 for Main analyses, as Begg's test performed, suggesting that there was a “small study effect” for publish bias Fig. 5. shows funnel plot the Main analysis, “IgM, IgG or both all techniques pooled”.

Fig. 5.

Fig 5

Funnel plot.

This heterogeneity could be explained by the difference in technique, antigen, manufacturer and by period of blood collection. Exclusion of any study from analysis did not significantly decrease heterogeneity in the analysis. Meta-regression per continent, technique and characteristics of sample did not decreased heterogeneity as well.

4. Discussion

In response to severe acute respiratory syndrome coronavirus 2 (SARS–CoV-2), approximately 3 billion people were placed under social distancing measures. Around the world, strategies have been deployed to break the spread of the virus. Identifying who is infected and contagious is a key point underpinning success of those interventions [55].

Development of an antibody response is influenced by a number of host-dependent factors, including age, severity of disease, among others [56]. In the case of SARS–CoV-2, and early studies suggest that the majority of patients seroconvert from 7 to 11 days after infection, although some patients may develop antibodies sooner [57,58]. Serum IgG amounts can rise at the same time or earlier than those of IgM against SARS-CoV-2. Cochrane COVID-19 Diagnostic Test Accuracy Group published a Systematic Review exploring serological tests accuracy for COVID-19 including people suspected of current or previous SARS-CoV-2 infection, or where tests were used to screen for infection, with TR-PCR as reference test or not (clinical diagnostic criteria as reference). Despite these differences regarding selection and reference criteria, they identified the same sensitivity of this study in “15th day onwards” sample, 91.4% (95% CI 87.0–94.4) previous and 91.0% (85.0–94.0) the present study; and different result for “7th – 14th days” sample: 72.2% (95% CI 63.5–79.5) previous and 65.0% (64.9–77.3) the present study [59].

One special challenge when evaluating serological tests is the huge diversity of techniques and manufactures. To exemplify, in the middle of April of 2020, 91 manufacturers had notified the Food and Drug Administration (FDA), of the United States of America, that they were offering serologic tests for commercial use, and four products have received FDA Emergency Use Approval [60]. The present study includes data resulting from four different techniques, from different kits, which allows for a current external validation of these tests but without enough information regarding reagents and overall diversity among the tests. An important consideration of the analysis is that rapid tests when performed as point-of-care tests, utilize whole blood as sample, while the investigated studies utilized methods for sample fractionation that increase performance of the test. In July 2020, Lisboa Bastos et al [61]. published a systematic review that showed similar results regarding ELISA and discrepant regarding other techniques performances. For ELISA, sensitivity has 84.3% (CI 95% 75.6 - 90.9) and specificity was 97.6% (CI 95% 93.2–99.4) and the present study identified 82.0% (CI 95% 72.0–89.0) and 96.0% (CI 95% 91.0–98.0), respectively, with a similar number of participants. For MCLIA (chemiluminescence assay), sensitivity has 97.8% (CI 95% 46.2–100) and specificity was not estimated, and the present study identified 85.0% (CI 95% (71.0–93.0) and 98.0% (CI 97% 91.0–99.0), respectively, probably because of the included number of participants on this study was four times fold, comparing to previous one. Immunochromatographic was not performed by that Systematic Review.

It is not known for certain whether individuals infected with COVID-19 who subsequently recover will be protected from future infection or how long protective immunity may last.

5. Limitations

Regarding weaknesses of this systematic review, the main point is the reference test. The PCR-RT test identifies the presence of the virus itself through virus RNA replication and because of that it is considered the golden standard. However, it can identify only the acute phase of the disease, which compromises its specificity and can influence serology evaluation. Primary studies did not include patients with COVID‐19 who may have had a false negative result on PCR, it may have affected test accuracy, but it is impossible to identify by how much.

This study was careful to analyze Negative and Positive Prediction Values to improve the information provided. However, it is important to remember that these measures are influenced by the disease prevalence in the population. Even though most studies only included suspected cases, characterized in a hospital scenario, some studies included healthy volunteers with the aim of calculating specificity, as described in Table 1. Therefore, the global sample of this systematic review is the reflection of primary study choices. Primary studies included in this systematic review did not report measures of immunoglobulin A (IgA). Both previous systematic Reviews cited in this study have not shown quantitative data about heterogeneity, so that was not possible to compare it [59,61].

The severe acute respiratory syndrome coronavirus 2 is an emerging infection with many unknowns. Currently, the World Health Organization (WHO) does not recommend the use of rapid diagnostics test for point-of-care, but such tests contribute for disease surveillance and epidemiologic research [62]. Our analysis supports the use of serological tests in suspected patients and provides scientific evidence for guiding infection control policies and therapies. Serological tests may have a role complementing other testing in individuals presenting later, when RT‐PCR tests are negative, or are not done, mainly after 7 days of symptoms. Results of present study cannot assess the utility of these tests for seroprevalence surveys for public health management purposes. These results can be applied mostly for suspected cases, in hospital scenario.

6. Future research directions

It is important to point out that results consist of the basis of preliminary analysis and that further investigation including consecutive sampling, standardization of methods, and well-defined use-case must be incorporated to generate robust evidence, simultaneous to a rapidly evolving epidemiology

7. Conclusion

This study supports that serology is a group of tests with high accuracy, mainly following the second week after infection, however the high heterogeneity cautions for careful application. Well-defined use-case for the tests (e.g. point-of-care versus immunosurveillance and technique) are necessary to be in place in order to benefit from the results of serological tests.

Role of the funding source

The present study has no funding sources.

Declaration of Competing Interest

The present study has no conflicts of interest.

Footnotes

Supplementary material associated with this article can be found, in the online version, at doi:10.1016/j.jcv.2022.105121.

Appendix. Supplementary materials

mmc1.doc (63.5KB, doc)

References

  • 1.Chen T., Wu D., Chen H., Yan W., Yang D., Chen G. Clinical characteristics of 113 deceased patients with coronavirus disease: retrospective study. BMJ. 2020;368:1–12. doi: 10.1136/bmj.m1091. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 2.Chu D.K.W., Pan Y., Cheng S.M.S., et al. Molecular diagnosis of a novel coronavirus (2019-nCoV) causing an outbreak of pneumonia. Clin. Chem. 2020;66:549–555. doi: 10.1093/clinchem/hvaa029. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 3.Hoffman T., Nissen K., Krambrich J., Rönnberg B., Akaberi D., Esmaeilzadeh M. Evaluation of a COVID-19 IgM and IgG rapid test; an efficient tool for assessment of past exposure to SARS-CoV- 2, Infection. Ecol. Epidemiol. 2020;10:1–5. doi: 10.1080/20008686.2020.1754538. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 4.Johns Hopkins University and medicine. Coronavirus 2019-nCoV, CSSE. Coronavirus 2019-nCoV global cases by Johns Hopkins CSSE. (Available from: https://coronavirus.jhu.edu/map.html). April 2021.
  • 5.Guo L., Ren L., Yang S., Xiao M., Chang D., Yang F. Profiling early humoral response to diagnose novel coronavirus disease (COVID-19) Clin. Infect. Dis. 2020;71:778–785. doi: 10.1093/cid/ciaa310. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 6.Infantino M., Damiani A., Gobbi F.L., et al. Serological assay for SARS-CoV-2 infectious disease: benefits, limitations and perspectives. Isr. Med. Assoc. J. 2020;22:203–210. [PubMed] [Google Scholar]
  • 7.Jin Y., Wang M., Zuo Z., Fan C., Ye F., Cai Z. Diagnostic value and dynamic variance of serum antibody in coronavirus disease. Int. J. Infect. Dis. 2020;94:49–52. doi: 10.1016/j.ijid.2020.03.065. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 8.Page M.J., McKenzie J.E., Bossuyt P.M., et al. The PRISMA 2020 statement: an updated guideline for reporting systematic reviews. BMJ. 2021;372:1–9. doi: 10.1136/bmj.n71. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 9.Whiting P.F., Rutjes A.W., Westwood M.E., Mallett S., Deeks J.J., Reitsma J.B. QUADAS-2: a revised tool for the quality assessment of diagnostic accuracy studies. Ann. Intern. Med. 2011;155:529–536. doi: 10.7326/0003-4819-155-8-201110180-00009. [DOI] [PubMed] [Google Scholar]
  • 10.Zamora J., Abraira V., Muriel A., Khan K., Coomarasamy A. Meta-DiSc: a software for meta-analysis of test accuracy data. BMC Med. Res. Methodol. 2006;6:1–12. doi: 10.1186/1471-2288-6-31. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 11.Wang, D., Mou, Z.Y., Zhai, J.X., Zong, H.X., & Zhao, X.D. Application of Stata software to test heterogeneity in meta-analysis method. Zhonghua liu xing bing xue za zhi= Zhonghua liuxingbingxue zazhi. 29 (2008) 726–729. [PubMed]
  • 12.Adams E., Ainsworth M., Anand R., Andersson M.I., Auckland K., Baillie J.K., Whitehouse J. Evaluation of antibody testing for SARS-CoV-2 using ELISA and lateral flow immunoassays. medRxiv. 2020:1–24. doi: 10.1101/2020.04.15.20066407. [DOI] [Google Scholar]
  • 13.Andrey D.O., Cohen P., Meyer B., Torriani G., Yerly S., Mazza L. Geneva centre for emerging viral diseases. Diagnostic accuracy of Augurix COVID-19 IgG serology rapid test. Eur. J. Clin. Invest. 2020;50:1–7. doi: 10.1111/eci.13357. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 14.Algaissi A., Alfaleh M.A., Hala S., Abujamel T.S., Alamri S.S., Almahboub S.A., Hashem A.M. SARS-CoV-2 S1 and N-based serological assays reveal rapid seroconversion and induction of specific antibody response in COVID-19 patients. Sci. Rep. 2020;10:1–10. doi: 10.1038/s41598-020-73491-5. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 15.Ayouba A., Thaurignac G., Morquin D., Tuaillon E., Raulino R., Nkuba A., Peeters M. Multiplex detection and dynamics of IgG antibodies to SARS-CoV2 and the highly pathogenic human coronaviruses SARS-CoV and MERS-CoV. J. Clin. Virol. 2020;129:1–6. doi: 10.1016/j.jcv.2020.104521. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 16.Bendavid E., Mulaney B., Sood N, Shah S., Ling E. Bromley-Dulfano R. COVID-19 antibody seroprevalence in Santa Clara County, California. medRxiv. 2020:1–30. doi: 10.1101/2020.04.14.20062463. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 17.Cassaniti I., Novazzi F., Giardina F., Salivaro F., Sachs M., Perlini S. Performance of VivaDiagTM COVID-19 IgM/IgG rapid test is inadequate for diagnosis of COVID-19 in acute patients referring to emergency room department. J. Med. Virol. 2020;92:1724–1727. doi: 10.1002/jmv.25800. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 18.Chan C.W., Parker K., Tesic V., Baldwin A., Tang N.Y., van Wijk X.M., Yeo K.T.J. Analytical and clinical evaluation of the automated Elecsys anti–SARS-CoV-2 antibody assay on the Roche cobas e602 analyzer. Am. J. Clin. Pathol. 2020;154:620–626. doi: 10.1093/ajcp/aqaa155. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 19.Charlton C.L., Kanji J.N., Johal K., Bailey A., Plitt S.S., MacDonald C., Tipples G. Evaluation of six commercial mid-to high-volume antibody and six point-of-care lateral flow assays for detection of SARS-CoV-2 antibodies. J. Clin. Microbiol. 2020;58:1–13. doi: 10.1128/JCM.01361-20. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 20.Döhla M., Boesecke C., Schulte B., Diegmann C., Sib E., Richter E., Streeck H. Rapid point-of-care testing for SARS-CoV-2 in a community screening setting shows low sensitivity. Public Health. 2020;182:170–172. doi: 10.1016/j.puhe.2020.04.009. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 21.Freeman B., Lester S., Mills L., et al. Validation of a SARS-CoV-2 spike protein ELISA for use in contact investigations and serosurveillance. bioRxiv. 2020:1–12. doi: 10.1101/2020.04.24.057323. [Preprint] [DOI] [Google Scholar]
  • 22.Guedez-López G.V., Alguacil-Guillén M., González-Donapetry P., et al. SARS-CoV-2 working group. Evaluation of three immunochromatographic tests for rapid detection of antibodies against SARS-CoV-2. Eur. J. Clin. Microbiol. Infect. Dis. 2020;39:2289–2297. doi: 10.1007/s10096-020-04010-7. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 23.Haselmann V., Kittel M., Gerhards C., Thiaucourt M., Eichner R., Costina V., Neumaier M. Comparison of test performance of commercial anti-SARS-CoV-2 immunoassays in serum and plasma samples. Clin. Chim. Acta. 2020;510:73–78. doi: 10.1016/j.cca.2020.07.007. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 24.Harb R., Remaley A.T., Sacks D.B. Evaluation of three commercial automated assays for the detection of anti-SARS-CoV-2 antibodies. Clin. Chem. 2020;66:1351–1353. doi: 10.1093/clinchem/hvaa193. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 25.Herroelen P.H., Martens G.A., De Smet D., Swaerts K., Decavele A.S. Humoral immune response to SARS-CoV-2: comparative clinical performance of seven commercial serology tests. Am. J. Clin. Pathol. 2020;154:610–619. doi: 10.1093/ajcp/aqaa140. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 26.Jia X., Zhang P., Tian Y., Wang J., Zeng H., Wang J., Liu Y. Clinical significance of IgM and IgG test for diagnosis of highly suspected COVID-19 infection. MedRxiv. 2020:1–25. doi: 10.1101/2020.02.28.20029025. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 27.Lassaunière R., Frische A., Harboe Z.B., Nielsen A.C., Fomsgaard A., Krogfelt K.A., Jørgensen C.S. Evaluation of nine commercial SARS-CoV-2 immunoassays. MedRxiv. 2020:1–15. doi: 10.1101/2020.04.09.20056325. [DOI] [Google Scholar]
  • 28.Lin D., Liu L., Zhang M., Hu Y., Yang Q., Guo J., Zhang Z. Evaluations of the serological test in the diagnosis of 2019 novel coronavirus (SARS-CoV-2) infections during the COVID-19 outbreak. Eur. J. Clin. Microbiol. Infect. Dis. 2020;39:2271–2277. doi: 10.1007/s10096-020-03978-6. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 29.Liu W., Liu L., Kou G., Zheng Y., Ding Y., Ni W., Wang Q., Li Tan, Wu W., Tang S., Xiong Z., Zheng S. Evaluation of nucleocapsid and spike protein-based ELISAs for detecting antibodies against SARS-CoV-2. J. Clin. Microbiol. 2020;58:1–7. doi: 10.1128/JCM0,00461-20. A. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 30.Liu Y., Liu Y.P., Diao B., Ding J.Y., Yuan M.X., Ren F.F., Wang Y., Huang Q.C. Diagnostic indexes of a rapid immunoglobulin G/immunoglobulin M combined antibody test for severe acute respiratory syndrome coronavirus 2. Chin. Med. J. (Engl) 2020;134:475–477. doi: 10.1097/CM9.0000000000001204. 10.1097 /CM9.0000000000001204. (B) [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 31.Liu, L., Liu, W., Zheng, Y., Jiang, X., Kou, G., Ding, J., Zheng, S. A preliminary study on serological assay for severe acute respiratory syndrome coronavirus 2 (SARS-CoV-2) in 238 admitted hospital patients. Microbes and infection. 22(2020) 206-211, https://doi.org/10.1016/j.micinf.2020.05.008. (C). [DOI] [PMC free article] [PubMed]
  • 32.Long Q.X., Liu B.Z., Deng H.J., et al. Antibody responses to SARS-CoV-2 in patients with COVID-19. Nat. Med. 2020;26:845–848. doi: 10.1038/s41591-020-0897-1. [DOI] [PubMed] [Google Scholar]
  • 33.Marinis Y., Sunnerhagen T., Bompada P., et al. Serology assessment of antibody response to SARS-CoV-2 in patients with COVID-19 by rapid IgM/IgG antibody test. Infect. Ecol. Epidemiol. 2020;10:1–11. doi: 10.1080/20008686.2020.1821513. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 34.Naaber P., Hunt K., Pesukova J., Haljasmägi L., Rumm P., Peterson P., Sepp E. Evaluation of SARS-CoV-2 IgG antibody response in PCR positive patients: comparison of nine tests in relation to clinical data. PLoS ONE. 2020;15:1–10. doi: 10.1371/journal.pone.0237548. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 35.Pan Y., Li X., Yang G., Fan J., Tang Y., Zhao J., Long X., Guo S., Zhao Z., Liu Y., Hu H., Xue H., Li Y. Serological immunochromatographic approach in diagnosis with SARS-CoV-2 infected COVID-19 patients. medRxiv. 2020;81:28–32. doi: 10.1016/j.jinf.2020.03.051. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 36.Pancrazzi A., Magliocca P., Lorubbio M., Vaggelli G., Galano A., Mafucci M., Ognibene A. Comparison of serologic and molecular SARS-CoV 2 results in a large cohort in Southern Tuscany demonstrates a role for serologic testing to increase diagnostic sensitivity. Clin. Biochem. 2020;84:87–92. doi: 10.1016/j.clinbiochem.2020.07.002. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 37.Paiva K.J., Grisson R.D., Chan P.A., Huard R.C., Caliendo A.M., Lonks J.R., Lu S. Validation and performance comparison of three SARS-CoV-2 antibody assays. J. Med. Virol. 2020;93:916–923. doi: 10.1002/jmv.26341. [DOI] [PubMed] [Google Scholar]
  • 38.Paradiso A.V., De Summa S., Loconsole D., Procacci V., Sallustio A., Centrone F., Chironna M. Clinical meanings of rapid serological assay in patients tested for SARS-Co2 RT-PCR. MedRxiv. 2020:1–16. doi: 10.1101/2020.04.03.20052183. [DOI] [Google Scholar]
  • 39.Pegoraro M., Militello V., Salvagno G.L., Gaino S., Bassi A., Caloi C., Cascio G.L. Evaluation of three immunochromatographic tests in COVID-19 serologic diagnosis and their clinical usefulness. Eur. J. Clin. Microbiol. Infect. Dis. 2021;40:897–900. doi: 10.1007/s10096-020-04040-1. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 40.Perera R.A., Mok C.K., Tsang O.T., Lv H., Ko R.L., Wu N.C. Serological assays for severe acute respiratory syndrome coronavirus 2 (SARS-CoV-2) Eurosurveillance. 2020;25:1–9. doi: 10.2807/1560-7917. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 41.Pérez-García F., Tanoira R.P., Cabrera J.P.R., Serrano T.A., Herruz P.G., Gonzalez J.C. Rapid diagnosis of SARS-CoV-2 infection by detecting IgG and IgM antibodies with an immunochromatographic device: a prospective single-center study. MedRxiv. 2020:1–16. doi: 10.1101/2020.04.11.20062158. [DOI] [Google Scholar]
  • 42.Pflüger L.S., Bannasch J.H., Brehm T.T., Pfefferle S., Hoffmann A., Nörz D., Lütgehetmann M. Clinical evaluation of five different automated SARS-CoV-2 serology assays in a cohort of hospitalized COVID-19 patients. J. Clin. Virol. 2020;130:1–6. doi: 10.1016/j.jcv.2020.104549. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 43.Pieri M., Ciotti M., Carlozzi N., Frassanito M.L., Meloni A., Cistera A., Bernardini S. SARS-CoV-2 infection serology validation of different methods: usefulness of IgA in the early phase of infection. Clin. Chim. Acta. 2020;511:28–32. doi: 10.1016/j.cca.2020.09.033. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 44.Phipps W.S., SoRelle J.A., Li Q.Z., Mahimainathan L., Araj E., Markantonis J., Muthukumar A. SARS-CoV-2 antibody responses do not predict COVID-19 disease severity. Am. J. Clin. Pathol. 2020;154:459–465. doi: 10.1093/ajcp/aqaa123. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 45.Suhandynata R.T., Hoffman M.A., Kelner M.J., McLawhon R.W., Reed S.L., Fitzgerald R.L. Longitudinal monitoring of SARS-CoV-2 IgM and IgG seropositivity to detect COVID-19. J. Appl. Lab. Med. 2020;5:908–920. doi: 10.1093/jalm/jfaa079. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 46.Tan C.W., Chia W.N., Qin X., Liu P., Chen M.I.C., Tiu C., Wang L.F. A SARS-CoV-2 surrogate virus neutralization test based on antibody-mediated blockage of ACE2–spike protein–protein interaction. Nat. Biotechnol. 2020;38:1073–1078. doi: 10.1038/s41587-020-0631-z. [DOI] [PubMed] [Google Scholar]
  • 47.Tehrani Z.R., Saadat S., Saleh E., Ouyang X., Constantine N., DeVico A.L., Sajadi M.M. Specificity and performance of nucleocapsid and spike-based SARS-CoV-2 serologic assays. MedRxiv. 2020:1–12. doi: 10.1101/2020.08.05.20168476. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 48.Turbett S.E., Anahtar M., Dighe A.S., et al. Evaluation of three commercial SARS-CoV-2 serologic assays and their performance in two-test algorithms. J. Clin. Microbiol. 2020;59:1–10. doi: 10.1128/JCM.01892-20. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 49.Vásárhelyi B., Kristóf K., Ostorházi E., Szabó D., Prohászka Z., Merkely B. A specifikus IgM-és IgG-antitesteket detektáló gyorstesztek értéke a SARS CoV-2 vírusfertőzés kimutatásában (A COVID–19-pandémia orvosszakmai kérdései) Orv. Hetil. 2020;161:807–812. doi: 10.1556/650.2020.31859. [DOI] [PubMed] [Google Scholar]
  • 50.Wang Q., Du Q., Guo B., Mu D., Lu X., Ma Q. A method to prevent SARS-CoV-2 IgM false positives in gold immunochromatography and enzyme-linked immunosorbent assays. J. Clin. Microbiol. 2020;58:1–7. doi: 10.1128/JCM.00375-20. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 51.Yassine H.M., Al-Jighefee H., Al-Sadeq D.W., Dargham S.R., Younes S.N., Shurrab F., Nasrallah G.K. Performance evaluation of five ELISA kits for detecting anti-SARS-COV-2 IgG antibodies. Int. J. Infect. Dis. 2021;102:181–187. doi: 10.1016/j.ijid.2020.10.042. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 52.Zhao R., Li M., Song H., Chen J., Ren W., Feng Y. Early detection of SARS-CoV-2 antibodies in COVID-19 patients as a serologic marker of infection. Clin. Infect. Dis. 2020;71:2066–2072. doi: 10.1093/cid/ciaa523. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 53.Zhang P., Gao Q., Wang T., Ke Y., Mo F., Jia R., Zhao Y. Evaluation of recombinant nucleocapsid and spike proteins for serological diagnosis of novel coronavirus disease 2019 (COVID-19) MedRxiv. 2020:1–28. doi: 10.1101/2020.03.17.20036954. [DOI] [Google Scholar]
  • 54.Zhong L., Chuan J., Gong B., et al. Detection of serum IgM and IgG for COVID-19 diagnosis. Sci. China Life Sci. 2020;63:777–780. doi: 10.1007/s11427-020-1688-9. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 55.Petherick A. Developing antibody tests for SARS-CoV-2. Lancet. 2020;395:1101–1102. doi: 10.1016/S0140-6736(20)30788-1. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 56.Okba N.M.A., Muller M.A., Li W., Wang C., et al. SARS-COV-2 specific antibody responses in COVID-19 patients. Medxriv. 2020:1–18. doi: 10.1101/2020.03.18.20038059. [DOI] [Google Scholar]
  • 57.Theel E.S., Slev P., Wheeler S., Couturier M.R., Wong S.J., Kadkhoda K. The role of antibody testing for SARS-CoV-2: is there one? J. Clin. Microbiol. 2020;58:1–8. doi: 10.1128/JCM.00797-20. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 58.To K.K.W., Tsang O.T.Y., Leung W.S., Tam A.R., Wu T.C., Lung D.C. Temporal profiles of viral load in posterior oropharyngeal saliva samples and serum antibody responses during infection by SARS-CoV-2: an observational cohort study. Lancet Infect. Dis. 2020;20:565–574. doi: 10.1016/S1473-3099(20)30196-1. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 59.Deeks J.J., Dinnes J., Takwoingi Y., et al. Antibody tests for identification of current and past infection with SARS-CoV-2. Cochrane Database Syst. Rev. 2020;25:1–306. doi: 10.1002/14651858.CD013652. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 60.Patel R., Babady E., Theel E.S., Storch G.A., Pinsky B.A., George K.S. Report from the American society for microbiology COVID-19. ASM J. 2020;11:1–5. doi: 10.1128/mBio.00722-20. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 61.Lisboa Bastos M., Tavaziva G., Abidi S.K., et al. Diagnostic accuracy of serological tests for covid-19: systematic review and meta-analysis. BMJ. 2020;370:1–13. doi: 10.1136/bmj.m2516. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 62.WHO. Advice on the use of point-of-care immunodiagnostic tests for COVID-19. 2020;(April):1–6.

Associated Data

This section collects any data citations, data availability statements, or supplementary materials included in this article.

Supplementary Materials

mmc1.doc (63.5KB, doc)

Articles from Journal of Clinical Virology are provided here courtesy of Elsevier

RESOURCES