Abstract
Introduction:
Hyperglycemia following meals is a recurring challenge for people with type 1 diabetes, and even the most advanced available automated systems currently require manual input of carbohydrate amounts. To progress toward fully automated systems, we present a novel control system that can automatically deliver priming boluses and/or anticipate eating behaviors to improve postprandial full closed-loop control.
Methods:
A model predictive control (MPC) system was enhanced by an automated bolus system reacting to early glucose rise and/or a multistage MPC (MS-MPC) framework to anticipate historical patterns. Priming was achieved by detecting large glycemic disturbances, such as meals, and delivering a fraction of the patient’s total daily insulin (TDI) modulated by the disturbance’s likelihood (bolus priming system [BPS]). In the anticipatory module, glycemic disturbance profiles were generated from historical data using clustering to group days with similar behaviors; the probability of each cluster is then evaluated at every controller step and informs the MS-MPC framework to anticipate each profile. We tested four configurations: MPC, MPC + BPS, MS-MPC, and MS-MPC + BPS in simulation to contrast the effect of each controller module.
Results:
Postprandial time in range was highest for MS-MPC + BPS: 60.73 ± 25.39%, but improved with each module: MPC + BPS: 56.95±25.83 and MS-MPC: 54.83 ± 26.00%, compared with MPC: 51.79 ± 26.12%. Exposure to hypoglycemia was maintained for all controllers (time below 70 mg/dL <0.5%), and improvement came primarily from a reduction in postprandial time above range (MS-MPC + BPS: 39.10 ± 25.32%, MPC + BPS: 42.99 ± 25.81%, MS-MPC: 45.09 ± 25.96%, MPC: 48.18 ± 26.09%).
Conclusions:
The BPS and anticipatory disturbance profiles improved blood glucose control and were most efficient when combined.
Keywords: automated insulin delivery, artificial pancreas, meal detection, behavioral patterns, disturbance mitigation
Introduction
Preventing hyperglycemia following eating is still a persistent issue for people with type 1 diabetes (T1D), even while using an automated insulin dosing (AID) system which delivers insulin variably based on an internal closed-loop algorithm. 1 The unwanted increase in plasma blood glucose (BG) following eating for people with T1D is primarily a result of the dynamics associated with subcutaneously injected insulin and ingested carbohydrates. Currently available subcutaneous insulin analogs have slower time constants than consumed carbohydrates, thus necessitating the use of feedforward control to prevent hyperglycemia. Anticipation and automatic bolusing in an advanced control system framework have proven to compensate for the inherent absorption delay from current insulin analogs. 2 In this contribution, we designed an in silico experiment using real-world eating behavior to contrast the impact of an automatic bolus priming system (BPS) and anticipatory glycemic disturbance profiles in our control system.
Fully Closed-Loop AID
Several AID systems that do not require users to input meal amounts have been tested. In 2014, Harvey et al evaluated a fully closed-loop (FCL) AID system on 12 adult participants with T1D in an inpatient clinical trial. 3 Overall, this system produced glycemic control with 80% of all BG readings in the 70- to 180-mg/dL range (time in range [TIR]) but caused high postprandial glucose values following unannounced meals. Forlenza et al presented the performance of an FCL that preempts predefined postprandial excursions in a 72-hour hotel-based study, reporting a TIR of 63.6 ± 9.2%. 4 However, in the four hours following unannounced meals, the time above range (>180 mg/dL, TAR) was 60.9 ± 23.3%. A recent study by Dovc et al used two formulations of insulin, FIASP and aspart, in a 27-hour inpatient admission study of their FCL AID system, 5 reporting TIR of 53.3% and 57.9% for aspart and FIASP insulins, respectively. Their approach relied on a meal detection algorithm but similarly faced difficulty with unannounced meals. Multihormone approaches have reported better performances, with recent early outpatient results. 6 In 2021, Haidar et al conducted a 30-participant open-label crossover trial and showed that an AID system using simple meal announcements and empagliflozin was noninferior to a hybrid system with meal announcements (mean BG: 153 ± 25.2 vs 153 ± 27.0 mg/dL). 7 Majdpour et al demonstrated the noninferiority of an insulin-pramlintide FCL AID to a hybrid AID (TIR: 81% vs 83%) in another 2021 pilot study. 8
Cameron’s 2012 manuscript describes a methodology for an AID system that uses multiple BG prediction models, each informed by different disturbances. 1 When the system determines that one disturbance is more likely, additional weight is given to that model’s predictions. In addition, information was included regarding the likely timing of meal disturbances based on normal mealtimes, the time of the last meal, and sleep schedule. Using this approach, Cameron et al were able to reduce the two-hour prediction error by 45% without meal detection and 18% with meal detection. The three-hour prediction error was reduced by 60% without meal detection and 30% with it.
Cameron et al extended these findings in a 2014 clinical trial. 2 The so-called multiple model probabilistic controller (MMPC) was used by ten patients in an inpatient study where they consumed five unannounced meals. For the six patients who used the final version of the controller, the mean continuous glucose monitor (CGM) TIR was 78%. During the admission, there was only one controller-induced hypoglycemia.
The MMPC was evaluated on ten patients in an inpatient clinical study where the mean TIR was 142 mg/dL overall and 125 mg/dL overnight. 3 A different version of the algorithm that was tested in a hotel-based study with 15 subjects achieved an overall mean BG of 152 mg/dL and a mean overnight BG of 139 mg/dL.
Use of Behavioral Patterns to Anticipate Glycemic Disturbances
Several simulation and clinical experiments conducted by our group have shown how anticipatory profiles integrated into AID systems can reduce the unwanted effects of glycemic disturbances. Simulation experiments have demonstrated that multistage model predictive controllers (MS-MPC) informed by disturbance profiles to anticipate moderate exercise’s effects can reduce hypoglycemia,9 -11 results that were later confirmed in a randomized crossover clinical trial with 15 adult participants: 12 There were fewer hypoglycemic events (9 vs 33), and the percent time where BG was <70 mg/dL (time below range [TBR]) was 1.3% lower while the participants used the MS-MPC system compared with a well-tuned standard MPC. The overall reduction in hypoglycemia resulted in no significant increase in TAR.
We also demonstrated in silico the capacity of the MS-MPC framework to anticipate glycemic disturbances caused by meals. 2 In that work, disturbance profiles were generated from a representative real subject and then used to perform closed-loop experiments using the 100 adult cohort of the University of Virginia (UVA)/Padova simulator. Results showed an average increase in TIR of 1.6% when using a hybrid closed-loop approach and 16.4% when using an FCL approach.
Our goal in this work is to build upon our past experiences and further study the impact of automated priming bolus and disturbance anticipation using the MS-MPC framework. To test the effects of each component of the control system, we implemented a 2 × 2 experimental design (each module on or off) resulting in four treatment strategies: model MPC, MS-MPC, MPC + BPS, and MS-MPC + BPS.
Methods
Control Strategy
Our modular control strategy is depicted in Figure 1 below. Within our legacy modular architecture (ie, data management, safety system module, autocorrection, state estimation), we consider four different combinations of modules to reject large positive disturbances. Details of each module are provided below.
Core controller
Rather than a single control strategy, MPC encompasses a general control paradigm; it integrates the predictions from an explicit mathematical model of the user’s glucose-related metabolic state within a real-time optimization problem that aims to find the best insulin injection subject to possible constraints over the model variables. In this setup, the MPC part commands optimized basal insulin injections through the pump every five minutes to reject internal metabolic disturbances. Multistage model predictive controller is a robust control strategy that considers N parallel MPC controllers, each perturbed by a particular disturbance realization. In this case, the disturbance realizations are composed of N different subject-specific disturbance profiles representing typical eating behavior. The final control action is reached after a consensus of the N controllers by the nonanticipatory constraint. 2
Bolus priming system
The BPS was designed to deliver doses of insulin that are comparatively higher than basal infusions when the probability of sizeable positive glycemic disturbance is elevated. This feedforward action is intended to work together with the feedback aspects of the controller to reduce postprandial BG excursions. At each five-minute interval, a second-order polynomial is fit on the last 30 minutes of CGM data using least squares regression. The coefficients of this equation roughly approximate the average value, first derivative, and second derivative of the most recent CGM values. These features are then used in a logistic regression-based disturbance detection algorithm.
The logistic regression coefficients were found by training a logistic regression classifier on a data set where 15 subjects using CGMs were observed but not supervised at home in a clinical trial data collection period based at UVA (NCT03394352). 13 In this data set, periods following meals were labeled in the positive class, and all other times were considered the negative class to train the classifier.
The output of the logistic regression, π , is then used to dictate if a priming bolus should be delivered. Because the disturbance’s size is unknown, the BPS priming bolus amounts are based on the user’s average total daily insulin (TDI). At the escalating thresholds for π , the BPS delivers different amounts of the patient’s TDI,
(1) |
The BPS dose, , is then determined based on the user’s TDI and the amount of insulin that was previously injected by the BPS system. This equation is,
(2) |
where was the amount of insulin on board determined using a 6-hour curve from previous BPS boluses. 14
Subject-specific disturbance profile generation
To construct a manageable ensemble of disturbance profiles for the MS-MPC to anticipate, we need to reduce the current ensemble of signatures to a few representative profiles. To do so, we cluster the occurrence of noticeable disturbances and then construct a representative disturbance input for each cluster. This necessitates (1) the construction of disturbance signatures, (2) detection of disturbance events, (3) clustering daily patterns of detected events, and (4) the construction of signatures that represent each cluster. We detail this process below.
Historical disturbance estimation
Disturbance signatures, , are estimated for each day of the historical record using a Kalman filter with the same internal model as the controller (ie, based on CGM and insulin only, not the meal record). In this case, the operating point for the linearization of the model is defined based on that subject’s mean CGM value for the day, the basal rate, and the steady-state solution of the model equations. Kalman filter tuning was performed by modifying the covariance matrices for measurement and modeling noises. Variations on and the diagonal elements of matrix in the range were tested to determine values that, in addition to a smoother estimation of the signature, adequately represented a system input to reflect the meal-related fluctuations in glucose observed in the CGM. Thus, values for and were fixed in such a way that had suitable behavior representing food intake during high glucose excursions. Figure 2 shows an example of the estimation of for a subject based on daily CGM data and the meal record. Kalman filter design is detailed in the Supplemental Appendix.
Disturbance detection
As records of meals would be absent in an FCL system, we reconstructed meal-like disturbances retrospectively from CGM and insulin data. To filter out any small deviation and focus on disturbances worth anticipating, we propose an automated disturbance detector using features characterizing the estimated disturbance values, , and continuous glucose measurements, , for each day of historical data collected. These features were selected to represent the changes in the glucose and disturbance values characteristic of glycemic disturbances, such as meals. A description of the features which were calculated for each five-minute sample is listed in Supplemental Table 1.
Once features were generated for each five-minute interval, , the probability of a large disturbance, π , is determined using logistic regression (Supplemental Table 2). The detection times are determined by finding the local minimums of CGM values when the detection vector was equal to one at least 60 minutes apart (or the first index if there were no local minimums). If detection times were within one hour of each other, the first detection time was used.
Disturbance profiles generation
Once major glycemic disturbances are detected, daily indicator signals are defined to group similar days into clusters (equal to one in the two hours following disturbance detections and zero otherwise). Using k-means with the hamming distance measure, these signals are clustered with .15,16 The number of clusters, , for each individual is based on which produced the highest Calinski-Harabasz score, maximizing cluster separation and cohesion. 17 Once days of data were grouped, the profile trace, , for each cluster, , at each five-minute interval of the day, , is determined from the average of each day in the cluster’s disturbance signal in that five-minute interval, .
(3) |
with as the number of days grouped into cluster . These profiles are then smoothed using a centered moving average over an hour. The values are multiplied by a weighting function shown in Figure 3 to deemphasize profiles overnight, allowing for the profile probabilities to return to their prior value at the beginning of each day.
We then estimate the prior probability value for each cluster, π , as the proportion of days of data that were assigned to that given cluster.
(4) |
where was the number of days of data in cluster and represented the total number of days considered. These prior probability values serve as a starting point so that the initial weight of each profile in the MS-MPC is related to historical data.
Figure 4 shows an example of one subject’s profiles. Here, it can be seen that this individual’s disturbance profiles are elevated following typical times for breakfast (6-8 am), lunch (12-2 pm), and dinner (6-8 pm). The associated prior probabilities indicate that this person eats earlier on 31% of days, and on 46% of the days of data used, the subject ate later. On 23% days, they had less of a discernable pattern of eating determined from their detected disturbances.
Profile probability estimation for online MS-MPC weighing
The probability value of each profile is updated in real-time based on the current disturbance measurement, which was found using the same technique applied to the retrospective data. This probability, π , is estimated using the method developed by Patek. 18 This process allows each profile’s probabilities to be shifted dynamically following the disturbance currently being experienced by the user.
We further add a “night mode,” allowing the profile probabilities real-time estimates to reset back to their prior (so each day is not influenced by the previous one). From 11 pm to 1 am, the probabilities devolved linearly from the value before the beginning of night mode, π , to the prior,
(5) |
For the, , intervals after the beginning of night mode,
(6) |
Finally, to ensure that the currently observed disturbance is prioritized over anticipation of expected disturbances, all the profile probability values are scaled based on an index, , derived from the level of the current disturbance estimate, . The calculation for is detailed in the Supplemental Appendix. The final adjusted probabilities, π , were found by multiplying the profile probabilities by .
(7) |
Experimental Design
Data collected during the unsupervised at-home portion of a large-scale pivotal trial conducted at UVA (NCT03563313) were used to evaluate this method. 19 These data are from 124 adult and adolescent participants with T1D over 6 months, during which they used an AID system with meal announcements. One hundred clinical subjects’ data were randomly selected and paired to an in silico subject in the UVA/Padova T1D simulator platform. 20 A total of 80% of the data were used to create meal profile clusters (see above). Seven days with at least one recorded meal were randomly selected from the remaining month of collected data. This week of meal record, scaled by the bodyweight, was then used as the meal protocol of the simulation experiment, which included intraday insulin sensitivity variability. 20 In this implementation of the simulation platform, circadian changes to insulin sensitivity and dawn phenomenon are simulated using subject-specific (ie, class-specific) time-varying profiles, which impacts endogenous glucose production and insulin-dependent glucose utilization. The use of real eating records allowed for more realistic behaviors in simulation and the existence of eventual meal patterns. This experimental configuration was then tested with the four configurations of the control system: MPC, MS-MPC, MPC + BPS, and MS-MPC + BPS. Treatments were compared overall and during the four hours after meals using the relevant metrics described by Maahs et al’s criteria for evaluating AID systems. 21 We provide individual P values (level: .05) for mean difference testing between control strategies for the primary outcome (TIR: 70-180 mg/dL); to note, due to the absence of random factors in our in silico experiment, statistical testing is potentially misleading; 22 it is therefore less emphasized than for a similar clinical experiment.
Results
The simulation experiment results suggest that the use of the anticipatory profiles in the MS-MPC and the BPS reduced BG values overall. The primary outcome (TIR over the entire simulation) improved from 72 ± 17.7% with MPC only to 73.4 ± 17.4% with anticipation and 75.5 ± 17.1% with priming bolus (all comparisons with P < .001). Maximum effect was seen with the combination of priming bolus and anticipation with TIR reaching 77.2 ± 16.7%, or +5.2% over MPC alone (P < .001). On average, when delivered, BPS boluses were 27.64 ± 29.70 minutes after the actual mealtime; 4% of the patients’ TDI was delivered after 64.11% of meals, 7% was delivered after 36.72%, and 10% was delivered after 22.18%. No BPS dose was delivered after 35.89% of meals. The overall false positive rate was 0.69%.
The mean BG for the MS-MPC + BPS was the lowest (155.14±31.88 mg/dL). The other controller configurations had higher mean BG values of 165.49 ± 33.49, 161.61 ± 33.60, and 159.65 ± 33.20 mg/dL for the MPC, MS-MPC, and MPC + BPS, respectively. This trend in lower BG values was represented similarly in the TBRs (ie, <50, <60, and <70 mg/dL), time in tight range (ie, 70-140 mg/dL), and time above ranges (ie, >180, >250, and >300 mg/dL). The BPS and anticipatory profiles also reduced the standard deviation of BG values and caused the system to deliver more insulin overall. The overall results of the experiment are listed in Table 1.
Table 1.
Controller | MPC | MS-MPC | MPC + BPS | MS-MPC + BPS | ANOVA P value |
---|---|---|---|---|---|
<50 mg/dL (%) | 0.03 ± 0.21 | 0.07 ± 0.40 | 0.02 ± 0.15 | 0.14 ± 0.54 | <.001 |
<60 mg/dL (%) | 0.06 ± 0.41 | 0.23 ± 0.85 | 0.10 ± 0.45 | 0.33 ± 1.01 | <.001 |
<70 mg/dL (%) | 0.14 ± 0.64 | 0.47 ± 1.37 | 0.24 ± 0.81 | 0.65 ± 1.58 | <.001 |
70-140 mg/dL (%) | 49.16 ± 17.94 | 50.35 ± 19.25 | 51.92 ± 17.61 | 53.81 ± 18.89 | .02 |
70-180 mg/dL (%) | 72.02 ± 17.67 | 73.37 ± 17.38 | 75.50 ± 17.07 | 77.17 ± 16.73 | <.001 |
>180 mg/dL (%) | 27.85 ± 17.47 | 26.16 ± 17.25 | 24.27 ± 16.84 | 22.18 ±16.45 | <.001 |
>250 mg/dL (%) | 9.92 ± 11.51 | 9.31 ± 11.25 | 7.66 ± 10.71 | 7.02 ± 10.28 | <.001 |
>300 mg/dL (%) | 4.81 ± 8.29 | 4.52 ± 8.03 | 3.74 ± 7.86 | 3.43 ± 7.36 | <.001 |
Mean (mg/dL) | 165.5 ± 33.5 | 161.6 ± 33.6 | 159.7 ± 33.2 | 155.1 ± 31.9 | <.001 |
SD (mg/dL) | 53.90 ± 33.49 | 53.62 ± 32.72 | 49.51 ± 34.32 | 48.91 ± 32.82 | <.001 |
CV (%) | 30.46 ± 12.09 | 31.12 ± 11.89 | 28.85 ± 12.01 | 29.44 ± 11.88 | <.001 |
Total daily insulin (U) | 36.72 ± 17.33 | 37.65 ± 17.57 | 38.09 ± 18.04 | 39.20 ±18.34 | <.001 |
The ANOVA P value tests the hypothesis that at least one condition is different from the others.
Abbreviations: MPC, model predictive control; MS, multistage; BPS, bolus priming system; ANOVA = analysis of variance; SD, standard deviation; CV, coefficient of variation.
In the four hours following meals, the effect of the anticipatory profiles and BPS was more evident. During this period, the TIR was 60.73 ± 25.39% for the MS-MPC + BPS. The mean TIR was 3.78%, 5.9%, and 8.94% less for the MPC + BPS, MS-MPC, and MPC configurations, respectively. Postprandial mean BG was also 4.51 to 10.35 mg/dL lower for the MS-MPC + BPS than it was for the other controller setups. Time below range for all controller setups was less than 1% during this timeframe. Table 2 lists the results during the postprandial period.
Table 2.
Controller | MPC | MS-MPC | MPC + BPS | MS-MPC + BPS | ANOVA P value |
---|---|---|---|---|---|
<50 mg/dL (%) | 0.00 ± 0.02 | 0.00 ± 0.02 | 0.00 ± 0.01 | 0.03 ± 0.15 | 1.00 |
<60 mg/dL (%) | 0.01 ± 0.08 | 0.03 ± 0.16 | 0.01 ± 0.06 | 0.08 ± 0.33 | <.001 |
<70 mg/dL (%) | 0.03 ± 0.17 | 0.08 ± 0.33 | 0.06 ± 0.32 | 0.17 ± 0.53 | <.001 |
70-140 mg/dL (%) | 23.02 ± 18.78 | 27.79 ± 22.09 | 25.67 ± 18.92 | 31.24 ± 22.28 | <.001 |
70-180 mg/dL (%) | 51.79 ± 26.12 | 54.83 ± 26.00 | 56.95 ± 25.83 | 60.73 ± 25.39 | <.001 |
>180 mg/dL (%) | 48.18 ± 26.09 | 45.09 ± 25.96 | 42.99 ± 25.81 | 39.10 ± 25.32 | <.001 |
>250 mg/dL (%) | 17.87 ± 19.85 | 16.61 ± 19.28 | 13.77 ± 18.46 | 12.46 ± 17.91 | <.001 |
>300 mg/dL (%) | 8.68 ± 14.73 | 8.13 ± 14.34 | 6.69 ± 13.77 | 6.19 ± 13.16 | <.001 |
Mean (mg/dL) | 197.2 ± 54.4 | 191.1 ± 54.2 | 188.6 ± 54.4 | 181.7 ± 52.3 | <.001 |
SD (mg/dL) | 53.87 ± 30.73 | 54.59 ± 30.37 | 49.67 ± 32.03 | 50.00 ± 30.96 | <.001 |
CV (%) | 25.69 ± 9.09 | 26.97 ± 8.95 | 24.61 ± 8.96 | 25.83 ± 8.76 | <.001 |
Insulin delivered (U) | 3.52 ± 1.92 | 3.61 ± 1.98 | 3.67 ± 2.07 | 3.79 ± 2.13 | <.001 |
The ANOVA P value tests the hypothesis that at least one condition is different from the others.
Abbreviations: MPC, model predictive control; MS, multistage; BPS, bolus priming system; ANOVA = analysis of variance; SD, standard deviation; CV, coefficient of variation.
Discussion
This simulation, which encompasses realistic eating behaviors in T1DM, indicates that mean BG values were lowest for the MS-MPC + BPS, followed in order by the MPC + BPS, the MS-MPC, and the MPC both overall and after eating. This relationship was maintained in terms of the percent time where BG was in the euglycemic ranges (ie, 70-140 mg/dL and 70-180 mg/dL) and the hypoglycemic ranges (ie, <50, <60, and <70 mg/dL). It was reversed in the amount of time where BG was in the hyperglycemic ranges (ie, >180, >250, and >300 mg/dL). This shows that both the anticipatory profiles and the BPS had the effect of lowering BG values overall. Overall, the MPC + BPS had a 5.84-mg/dL lower mean BG and TIR 3.38% higher than the MPC. The MS-MPC case resulted in a 1.35% greater TIR compared with the MPC, where no profiles were used. The MPC + BPS and MS-MPC + BPS cases had a difference of 1.67% in TIR. This indicates a synergetic interaction between BPS and the anticipatory disturbance profiles. While statistically significant, the differences between MPC versus either MS-MOC or MPC + BPS were incremental, while MS-MPC + BPS showed a potentially clinically relevant difference above 5% TIR.
Across the four configurations, there was no meaningful change in the amount of hypoglycemia overall. Comparing the MPC and MS-MPC + BPS shows that the modules may be responsible for increasing the amount of TBR by less than five minutes, while increasing TIR by more than 5% overall, which is clinically relevant.
The most insulin was used in the case where both the BPS and profiles were active (MS-MPC + BPS). This was less when only BPS was used (MPC + BPS), then even less when just the profiles were used (MS-MPC), and the least when the standard MPC. Interestingly, the MS-MPC case had the highest average coefficient of variation (CV), followed by MPC, MS-MPC + BPS, and MPC + BPS.
The postprandial TIR was increased by nearly 10% when the anticipatory profiles and BPS were used compared with the standard MPC. In addition, there was an increase in the 70- to 140-mg/dL range, but this amount was slightly smaller. This difference resulted from a reduction in TAR and a reduction in the mean BG of roughly 16 mg/dL. The combination of the MS-MPC structure and BPS had its greatest effect during the postprandial period by lowering BG values without increasing hypoglycemia.
A limitation of this work is that this is a simulation study and does not include some variability inherent in free-living conditions (eg, stress, hormonal changes). An additional limitation is that there was no physical activity which could cause both a risk for hypoglycemia and hyperglycemia in the simulation. Because the system does not know future actions like a person might, physical activity could be dangerous if an increase in automatic insulin delivery preceded it. In future versions of this work, we will address this issue by incorporating our past methodologies that consider inputs related to negative disturbances in the prediction model. Other improvements involve using activity sensors to detect eating and exercise events. Model personalization could also improve the generation of the disturbance estimation procedure.
Conclusion
This experiment shows that in a simulation environment, both the BPS and the disturbance profiles positively impact the amount of time that BG values are in the euglycemic range while also decreasing hyperglycemia. These modules are the most impactful during the four hours following eating. Independently, both modules improve TIR and do even better in combination. The reduction in mean BG values attributed to the BPS and disturbances profiles did increase hypoglycemia, but only to a degree that is not clinically meaningful. The BPS and disturbance profiles seem to impact the variability of glucose, although the cause of this is unclear. Future work will determine how this system is impacted by situations that may create risk for hypoglycemia and hyperglycemia.
Supplemental Material
Supplemental material, sj-docx-1-dst-10.1177_19322968211059159 for Using an Online Disturbance Rejection and Anticipation System to Reduce Hyperglycemia in a Fully Closed-Loop Artificial Pancreas System by John P. Corbett, Jose Garcia-Tirado, Patricio Colmegna, Jenny L. Diaz Castaneda and Marc D. Breton in Journal of Diabetes Science and Technology
Acknowledgments
We would like to acknowledge the clinical, regulatory, and data management staff at the University of Virginia Center for Diabetes Technology
Footnotes
Abbreviations: AID, automated insulin dosing; BG, blood glucose; BPS, bolus priming system; CGM, continuous glucose monitor; CV, coefficient of variation; FCL, fully closed-loop; MPC, model predictive control; MS-MPC, multistage model predictive control; SAP, sensor-augmented pump; SD, standard deviation, TAR, time above range; TIR, time in range; TDI, total daily insulin; T1D, type 1 diabetes.
Declaration of Conflicting Interests: The author(s) declared the following potential conflicts of interest with respect to the research, authorship, and/or publication of this article: M.D.B. has received honoraria and travel reimbursement from Dexcom and Tandem, and research support from Dexcom, Novo Nordisk, and Tandem. The other authors have no conflicts of interest to disclose.
Funding: The author(s) disclosed receipt of the following financial support for the research, authorship, and/or publication of this article: This work has been supported by the American Diabetes Association (1-19-PMF-014), the National Institutes of Health (1UC4 DK108483-01), and the University of Virginia Strategic Investment Fund.
ORCID iDs: John P. Corbett https://orcid.org/0000-0001-5674-3696
Jose Garcia-Tirado https://orcid.org/0000-0002-9970-2162
Patricio Colmegna https://orcid.org/0000-0001-9074-8634
Marc D. Breton https://orcid.org/0000-0001-7645-2693
Supplemental Material: Supplemental material for this article is available online.
References
- 1. Gingras V, Taleb N, Roy-Fleming A, Legault L, Rabasa-Lhoret R. The challenges of achieving postprandial glucose control using closed-loop systems in patients with type 1 diabetes. Diabetes Obes Metab. 2018;20(2):245-256. doi: 10.1111/dom.13052. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 2. Corbett JP, Colmegna P, Garcia-Tirado J, Breton MD. Anticipating meals with behavioral profiles in an artificial pancreas system—an informed multistage model predictive control approach. IFAC-PapersOnLine. 2020;53(2):16305-16310. doi: 10.1016/j.ifacol.2020.12.652. [DOI] [Google Scholar]
- 3. Harvey RA, Dassau E, Bevier WC, et al. Clinical evaluation of an automated artificial pancreas using zone-model predictive control and health monitoring system. Diabetes Technol Ther. 2014;16(6):348-357. doi: 10.1089/dia.2013.0231. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 4. Forlenza GP, Cameron FM, Ly TT, et al. Fully closed-loop multiple model probabilistic predictive controller artificial pancreas performance in adolescents and adults in a supervised hotel setting. Diabetes Technol Ther. 2018;20(5):335-343. doi: 10.1089/dia.2017.0424. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 5. Dovc K, Piona C, Mutlu GY, et al. Faster compared with standard insulin aspart during day-and-night fully closed-loop insulin therapy in type 1 diabetes: a double-blind randomized crossover trial. Diabetes Care. 2020;43(1):29-36. doi: 10.2337/dc19-0895. [DOI] [PubMed] [Google Scholar]
- 6. Blauw H, Onvlee AJ, Klaassen M, van Bon AC, DeVries JH. Fully closed loop glucose control with a bihormonal artificial pancreas in adults with type 1 diabetes: an outpatient, randomized, crossover trial. Diabetes Care. 2021;44(3):836-838. doi: 10.2337/dc20-2106. [DOI] [PubMed] [Google Scholar]
- 7. Haidar A, Yale J-F, Lovblom LE, et al. Reducing the need for carbohydrate counting in type 1 diabetes using closed-loop automated insulin delivery (artificial pancreas) and empagliflozin: a randomized, controlled, non-inferiority, crossover pilot trial. Diabetes Obes Metab. 2021;23:1272-1281. doi: 10.1111/dom.14335. [DOI] [PubMed] [Google Scholar]
- 8. Majdpour D, Tsoukas MA, Yale J-F, et al. Fully automated artificial pancreas for adults with type 1 diabetes using multiple hormones: exploratory experiments [published online ahead of print February 20, 2021]. Can J Diabetes. doi: 10.1016/j.jcjd.2021.02.002. [DOI] [PubMed] [Google Scholar]
- 9. Garcia-Tirado J, Corbett JP, Boiroux D, Jørgensen JB, Breton MD. Closed-loop control with unannounced exercise for adults with type 1 diabetes using the Ensemble Model Predictive Control. J Process Control. 2019;80:202-210. doi: 10.1016/j.jprocont.2019.05.017. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 10. Garcia-Tirado J, Colmegna P, Corbett JP, Ozaslan B, Breton MD. Ensemble model predictive control strategies can reduce exercise hypoglycemia in type 1 diabetes: in silico studies. In: 2019 American Control Conference (ACC), vol. 2019; 2019:4752-4758. doi: 10.23919/acc.2019.8814728. [DOI] [Google Scholar]
- 11. Garcia-Tirado J, Colmegna P, Corbett JP, Ozaslan B, Breton MD. In silico analysis of an exercise-safe artificial pancreas with multistage model predictive control and insulin safety system. J Diabetes Sci Technol. 2019;13(6):1054-1064. doi: 10.1177/1932296819879084. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 12. Garcia-Tirado J, Brown SA, Laichuthai N, et al. Anticipation of historical exercise patterns by a novel artificial pancreas system reduces hypoglycemia during and after moderate-intensity physical activity in people with type 1 diabetes. Diabetes Technol Ther. 2020;23(4):277-285. doi: 10.1089/dia.2020.0516. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 13. Ozaslan B, Brown SA, Pinnata J, et al. Safety and feasibility evaluation of step count informed meal boluses in type 1 diabetes: a pilot study [published online ahead of print April 1, 2021]. J Diabetes Sci Technol. doi: 10.1177/1932296821997917. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 14. Ellingsen C, Dassau E, Zisser H, et al. Safety constraints in an artificial pancreatic beta cell: an implementation of model predictive control with insulin on board. J Diabetes Sci Technol. 2009;3(3):536-544. doi: 10.1177/193229680900300319. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 15. Hamming RW. Error detecting and error correcting codes. Bell Syst Tech J. 1950;29(2):147-160. doi: 10.1002/j.1538-7305.1950.tb00463.x. [DOI] [Google Scholar]
- 16. Lloyd SP. Least squares quantization in PCM; 1982. https://sites.cs.ucsb.edu/~veronika/MAE/kmeans_LLoyd_Least_Squares_Quantization_in_PCM.pdf. Accessed August 15, 2019.
- 17. Caliñski T, Harabasz J. A dendrite method foe cluster analysis. Commun Stat. 1974;3(1):1-27. doi: 10.1080/03610927408827101. [DOI] [Google Scholar]
- 18. Patek SD. Open-loop feedback control under multiple disturbance function hypotheses; 2010. doi: 10.1109/CDC.2010.5717476. [DOI] [Google Scholar]
- 19. Brown SA, Kovatchev KP, Raghinaru D, et al. Six-month randomized, multicenter trial of closed-loop control in type 1 diabetes. N Engl J Med. 2019;381(18):1707-1717. doi: 10.1056/NEJMoa1907863. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 20. Visentin R, Campos-Náñez E, Schiavon M, et al. The UVA/Padova type 1 diabetes simulator goes from single meal to single day. J Diabetes Sci Technol. 2018;12:273-281. doi: 10.1177/1932296818757747. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 21. Maahs DM, Buckingham BA, Castle JR, et al. Outcome measures for artificial pancreas clinical trials: a consensus report. Diabetes Care. 2016;39(7):1175-1179. doi: 10.2337/dc15-2716. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 22. White JW, Rassweiler A, Samhouri JF, Stier AC, White C. Ecologists should not use statistical significance tests to interpret simulation model results. Oikos. 2014;123(4):385-388. doi: 10.1111/j.1600-0706.2013.01073.x. [DOI] [Google Scholar]
Associated Data
This section collects any data citations, data availability statements, or supplementary materials included in this article.
Supplementary Materials
Supplemental material, sj-docx-1-dst-10.1177_19322968211059159 for Using an Online Disturbance Rejection and Anticipation System to Reduce Hyperglycemia in a Fully Closed-Loop Artificial Pancreas System by John P. Corbett, Jose Garcia-Tirado, Patricio Colmegna, Jenny L. Diaz Castaneda and Marc D. Breton in Journal of Diabetes Science and Technology