Abstract
Background:
Dual-plate constructs have become an increasingly common fixation technique for midshaft clavicle fractures and typically involve the use of mini-fragment plates. The goal of this technique is to reduce plate prominence and implant irritation. However, limited biomechanical data exist for these lower-profile constructs. The study aim was to compare dual mini-fragment orthogonal plating with small-fragment clavicle plates for biomechanical noninferiority and to determine if an optimal plate configuration could be identified using a cadaveric model.
Methods:
Twenty-four cadaveric clavicles were randomized to 1 of 6 groups, stratified by computed tomography-based bone mineral content (BMC): precontoured superior or anterior fixation using a single 3.5-mm Locking Compression Plate (LCP), and 4 different dual-plating constructs utilizing 2.4-mm and 2.7-mm Adaptation plates or LCPs. An inferior butterfly fracture was created. Axial, torsional, and bending (anterior and superior surface loading) stiffnesses were determined through nondestructive cyclic testing, followed by a load-to-failure test in 3-point superior surface bending.
Results:
For axial stiffness, the 2 dual-plate constructs with a superior 2.4-mm and anterior 2.7-mm plate (either Adaptation or LCP) were significantly stiffer than the other 4 constructs (p = 0.021 and p = 0.034). For both superior and anterior bending, the superior 2.4-mm and anterior 2.7-mm plate constructs were significantly stiffer when compared with the 3.5-mm superior plate (p = 0.043). No significant differences were found in torsional stiffness or load to failure between the different constructs.
Conclusions:
Dual plating using mini-fragment plates is biomechanically superior for the fixation of midshaft clavicle fractures when compared with a single, superior, 3.5-mm plate and has biomechanical properties similar to those of a 3.5-mm plate placed anteriorly. With the exception of axial stiffness, no significant differences were found when different dual-plating constructs were compared with each other.
Clinical Relevance:
This study validates the use of dual plating for midshaft clavicle fractures.
Clavicle fractures are common injuries, with the middle one-third of the clavicle accounting for two-thirds of all fractures1-3. Nonoperative management remains the predominant treatment modality for these fractures. However, the treatment paradigm has changed over the past decade following numerous high-quality randomized controlled studies indicating that nonoperative outcomes are not as favorable as once believed4-7. Nonetheless, the recommended indications for surgical treatment are still conflicting. Most of the debate surrounds secondary operative procedures, as the number of patients who require a second operation (8.0% to 20.8%) is considerable6,8. The most common indication for a secondary surgical procedure is implant irritation8.
Although various fixation methods of midshaft clavicle fractures have been described, plate fixation remains the most established method9. Dual-plate fixation has become a more common technique and typically involves the use of mini-fragment plates10-17. It is a lower-profile construct in comparison with the traditionally used (contoured) small-fragment plates and could potentially reduce secondary surgical procedures because of less prominent implants13. In a recent systematic review and meta-analysis (278 patients), a 4.2% implant removal rate was reported for the dual-plating technique, and single-plate fixation was associated with a 3.9-fold increased implant removal rate (p = 0.003)8. This difference in implant removal rate is less pronounced with a single plate placed anteroinferiorly compared with superior plating18.
Limited biomechanical data exist regarding whether dual plating with smaller plate-screw constructs provides enough stability. Two biomechanical studies using synthetic clavicles showed noninferiority of dual plating and either superior or anteroinferior single plating using larger small-fragment plate-screw constructs in axial, torsional, and bending stiffnesses and in load to failure13,19. These findings have been further supported by a recent finite element analysis20 and a biomechanical study in a cadaveric model21. Although Ziegler et al.21 randomized their specimens, they were not able to correct for differences in bone mineral content (BMC) among the different fixation groups, possibly affecting their results. In addition, none of the previous studies reported on optimal plate configuration for the different constructs.
The hypothesis of the current study was that dual, mini-fragment, orthogonal plate fixation is biomechanically noninferior when compared with traditional, single, small-fragment clavicle plate fixation. The overall goal was to systematically compare different dual mini-fragment plate configurations to assess the noninferiority of the strength of each construct, using a clinically relevant cadaveric fracture model.
Materials and Methods
Specimen Preparation
The current study used 24 whole adult cadaveric clavicle specimens (12 pairs) without previous fractures or congenital anomalies. This study was approved by our institutional research ethics review board (REB19-0600) prior to obtaining the specimens. Using high-resolution clinical computed tomographic (CT) scans, BMC was determined for all specimens to perform a stratified random allocation of the specimens to 1 of the 6 different plate configuration groups. The characteristics of the clavicles in the different groups are listed in Table I.
TABLE I.
Specimen Characteristics for the Different Plate Configurations
| Groups | Donor Age* (yr) | Clavicle Length* (cm) | BMC* (g) |
|---|---|---|---|
| 3.5-mm LCP (7-hole) superiorly | 81.00 ± 3.56 | 14.75 ± 0.60 | 8.41 ± 1.71 |
| 3.5-mm LCP (7-hole) anteriorly | 77.50 ± 3.52 | 13.63 ± 0.13 | 8.14 ± 1.29 |
| 2.7-mm LCP (5-hole) superiorly and 2.4-mm Adaptation plate (12-hole) anteriorly | 78.25 ± 3.09 | 14.00 ± 0.54 | 8.55 ± 1.24 |
| 2.7-mm Adaptation plate (12-hole) superiorly and 2.4-mm LCP (5-hole) anteriorly | 86.00 ± 2.35 | 13.88 ± 0.24 | 8.21 ± 1.27 |
| 2.4-mm LCP (5-hole) superiorly and 2.7-mm Adaptation plate (12-hole) anteriorly | 80.00 ± 5.15 | 14.38 ± 0.72 | 8.44 ± 1.98 |
| 2.4-mm Adaptation plate (12-hole) superiorly and 2.7-mm LCP (5-hole) anteriorly | 75.25 ± 3.01 | 14.13 ± 0.52 | 8.28 ± 1.39 |
The values are given as the mean and the standard error.
Fracture Model and Fixation Technique
The 6 different plate configurations (4 per group), all from DePuy Synthes, included superior plating using a single 7-hole (110 mm in length), precontoured, 3.5-mm Superior Clavicle Locking Compression Plate (LCP); anterior plating using a single 7-hole (90 mm in length), 3.5-mm Medial Anterior Clavicle LCP; or 4 different dual-plating constructs (Table I) with superior and anteroinferior plating using a Modular Mini Fragment 5-hole 2.4-mm LCP (44 mm in length) or 2.7-mm LCP (49 mm) and a 12-hole Adaptation 2.4-mm plate (88 mm in length) or 2.7-mm plate (97 mm in length). These 4 different dual-plating constructs were chosen as they were the most commonly used in previous studies8 and in our current clinical practice. To allow for controlled osteotomy and reproducible fixation, the superior plate (or the anterior plate in the case of a single anterior plate configuration) was applied first before creating the inferior butterfly fragment (Fig. 1). No implants were used more than once.
Fig. 1.

A 2-cm inferior butterfly fracture fragment (OTA 15.2B) was created using a mini-sagittal saw, with the apex at the superior aspect of the clavicle, extending 1 cm lateral and medial to the midpoint along the inferior surface. In this specimen, a 2.4-mm LCP was placed anteriorly and a 2.7-mm Adaptation plate was placed superiorly.
Applying the standard compression plating technique, 3 bicortical 3.5-mm cortical screws were placed on either side of the fracture for both the superior-plating and anterior-plating constructs. For the dual-plating constructs, 2 bicortical 2.4-mm or 2.7-mm cortical screws were placed in the lateral and medial fragments through the corresponding plates. The butterfly fragment was reduced, and a drill-hole was made perpendicular to the fracture (superomedial to inferolateral). A single 3.5-mm interfragmentary screw was inserted for both superior and anterior plating; for the dual-plating constructs, a 2.7-mm interfragmentary screw was used. The interfragmentary screws were inserted through the plate, with the exception of the single anterior plate construct. The plates and screws were made of stainless steel.
Biomechanical Model
For biomechanical testing, a protocol similar to that described by Ziegler et al.21 was used. The specimens were stripped of all soft-tissue attachments. The central 11-cm portion of the clavicle was left exposed, and the remaining medial and lateral portions of each specimen were potted in polymethylmethacrylate (PMMA) after the placement of 2 screws in each of the lateral and medial ends to improve the adherence of the PMMA. A custom-made jig was used to ensure that the orientation of potting was parallel to the respective ends of the specimens (i.e., taking into account the anatomic curvature), so that no off-axis loads would be applied. Using a 858 Bionix system (MTS), axial, torsional, and bending (anterior and superior loading) stiffnesses were determined for each construct through nondestructive cyclic testing. This was followed by 3-point loading to failure by bending with superior loading. The order of testing was the same for all specimens. Loading rates were chosen to mimic non-traumatic physiological loading21. Nondestructive axial testing was performed by compressing each clavicle between 10 and 315 N at 0.25 Hz for 10 cycles (Fig. 2). Nondestructive torsional stiffness testing was performed along the long axis of the clavicle by rotating +1 Nm/degree and −1 Nm/degree at 0.25 Hz for 10 cycles. The 3-point anterior-load bending test was performed by placing the fulcrum at the dorsal aspect of the clavicle, 1 cm medial to the fracture site21 (Fig. 3). The clavicles were loaded cyclically from 10 to 60 N at 0.25 Hz for 10 bending cycles. This process was repeated with the fulcrum at the inferior aspect of the clavicle for the superior-load bending test21. The axial, torsional, and bending stiffnesses were measured during the tenth and final cycle of the test for all specimens. The setup for the loading to failure was identical for the superior-load bending test. The specimens were loaded at a rate of 15 mm/min until a fracture was observed either audibly or visually by 2 observers (J.K., K.P.). The load value immediately preceding this drop in peak load was used for the calculation of the load to failure for all specimens.
Fig. 2.
For nondestructive axial and torsional testing, the potted lateral end of the specimen was mounted parallel to the actuator, with the medial end mounted vertical to the testing table.
Fig. 3.
For the nondestructive bending tests, the medial end of the clavicle was perpendicular to the actuator. The force was applied at the lateral PMMA-potted end of the clavicle, with the actuator 10 cm lateral to the fulcrum.
Statistical Analysis
All specimens were randomly allocated to 1 of 6 different plate configuration groups, based on stratification by BMC. Construct stiffness and failure load were summarized using the mean, the standard error of the mean, and the 95% confidence interval (CI). Both the standard error and the 95% CI were corrected for BMC. To test for biomechanical noninferiority between the different groups, the nonparametric Kruskal-Wallis H test was applied. To determine if an optimal plate configuration could be identified, pairwise group comparisons were tested for significance using the nonparametric Mann-Whitney U test. A p value of <0.05 was considered significant. All analyses were performed using SPSS, version 25.0 (IBM).
Source of Funding
There was no external funding source.
Results
No significant differences were seen between the different groups with respect to age (mean [and standard error], 79.67 ± 1.46 years [range, 70 to 92 years]; p = 0.341), clavicle length (mean, 14.13 ± 0.20 cm [range, 13.0 to 16.5 cm]; p = 0.511), and BMC (mean, 8.34 ± 0.54 g [range, 4.23 to 13.76 g]; p = 1.000) (Fig. 4).
Fig. 4.
Box plot for BMC in grams determined for all specimens using high-resolution clinical CT scans (120 kVp, 150 mA, with an in-plane resolution of 0.263 × 0.263 mm and a slice thickness of 0.625 mm). The line within the box represents the median, and the whiskers indicate the range.
Axial Stiffness
The nondestructive cyclic testing for axial stiffness revealed a significant difference in stiffness (p = 0.006). The 2 dual-plate constructs with a superior 2.4-mm plate and an anterior 2.7-mm plate were significantly stiffer than the other 4 constructs when pairwise group comparisons were made (p = 0.021 for the 2.4-mm LCP superiorly and p = 0.034 for the 2.4-mm Adaptation plate superiorly) (Table II).
TABLE II.
Axial Stiffness by Plate Configuration Group
| Groups | Axial Stiffness* (N/mm) | Significance |
|---|---|---|
| 3.5-mm LCP superiorly | 520.71 ± 121.99 (119.72 to 896.19) | NS† |
| 3.5-mm LCP anteriorly | 914.07 ± 253.24 (201.39 to 1,813.26) | NS† |
| 2.7-mm LCP superiorly and 2.4-mm Adaptation plate anteriorly | 851.75 ± 182.86 (307.43 to 1,471.28) | NS† |
| 2.7-mm Adaptation plate superiorly and 2.4-mm LCP anteriorly | 812.02 ± 117.66 (409.34 to 1,421.85) | NS† |
| 2.4-mm LCP superiorly and 2.7-mm Adaptation plate anteriorly | 2,172.83 ± 342.70 (1,267.45 to 3,448.71) | S‡ |
| 2.4-mm Adaptation plate superiorly and 2.7-mm LCP anteriorly | 2,370.94 ± 392.56 (1,300.48 to 3,799.10) | S‡ |
The values are given as the mean and the standard error, with the 95% CI in parentheses.
NS = nonsignificant. Configurations are not significantly different from one another.
S = significant. There is a significant difference between this construct and ≥1 other constructs. The 2 dual-plate constructs with a superior 2.4-mm and anterior 2.7-mm plate were significantly stiffer than the other 4 constructs (p = 0.021 for the 2.4-mm LCP superiorly and p = 0.034 for the 2.4-mm Adaptation plate superiorly).
Torsional Stiffness
For the nondestructive cyclic testing for torsional stiffness, no significant differences (p = 0.324) were observed between the different plating constructs (Table III).
TABLE III.
Torsional Stiffness by Plate Configuration Group
| Groups | Torsional Stiffness* (N∙mm/deg) | Significance† |
|---|---|---|
| 3.5-mm LCP superiorly | 0.25 ± 0.04 (0.09 to 0.37) | NS |
| 3.5-mm LCP anteriorly | 0.37 ± 0.10 (0.04 to 0.70) | NS |
| 2.7-mm LCP superiorly and 2.4-mm Adaptation plate anteriorly | 0.39 ± 0.06 (0.17 to 0.58) | NS |
| 2.7-mm Adaptation plate superiorly and 2.4-mm LCP anteriorly | 0.36 ± 0.03 (0.27 to 0.45) | NS |
| 2.4-mm LCP superiorly and 2.7-mm Adaptation plate anteriorly | 0.45 ± 0.07 (0.26 to 0.69) | NS |
| 2.4-mm Adaptation plate superiorly and 2.7-mm LCP anteriorly | 0.59 ± 0.08 (0.32 to 0.85) | NS |
The values are given as the mean and the standard error, with the 95% CI in parentheses.
NS = nonsignificant. Configurations are not significantly different from one another.
Bending Stiffness
No significant differences were seen overall for the nondestructive cyclic testing of bending stiffness with both anterior loading (p = 0.095) and superior loading (p =0.079). However, when pairwise group comparisons were made between the different plate configurations, the 2 dual-plate constructs with a superior 2.4-mm plate and an anterior 2.7-mm plate were significantly stiffer than superior plating with both anterior loading (Table IV) and superior loading (Table V) (p = 0.043 for both). In addition, anterior plating was a significantly stiffer construct (p = 0.043), when compared with superior plating, in bending with superior loading (Table V).
TABLE IV.
Bending Stiffness with Anterior Loading by Plate Configuration Group
| Groups | Bending Stiffness* (N/mm) | Significance |
|---|---|---|
| 3.5-mm LCP superiorly | 25.98 ± 8.19 (2.63 to 54.76) | NS† |
| 3.5-mm LCP anteriorly | 40.73 ± 2.44 (32.97 to 48.48) | NS† |
| 2.7-mm LCP superiorly and 2.4-mm Adaptation plate anteriorly | 34.90 ± 6.81 (13.73 to 57.05) | NS† |
| 2.7-mm Adaptation plate superiorly and 2.4-mm LCP anteriorly | 60.04 ± 1.09 (55.00 to 64.37) | NS† |
| 2.4-mm LCP superiorly and 2.7-mm Adaptation plate anteriorly | 57.30 ± 6.33 (38.36 to 78.67) | S‡ |
| 2.4-mm Adaptation plate superiorly and 2.7-mm LCP anteriorly | 51.34 ± 3.14 (42.27 to 62.28) | S‡ |
The values are given as the mean and the standard error of the mean, with the 95% CI in parentheses.
NS = nonsignificant. Configurations are not significantly different from one another.
S = significant. There is a significant difference between this construct and ≥1 other constructs. The 2 dual-plate constructs with a superior 2.4-mm plate and an anterior 2.7-mm plate were significantly stiffer than superior plating (p = 0.043).
TABLE V.
Bending Stiffness with Superior Loading by Plate Configuration Group
| Groups | Bending Stiffness* (N/mm) | Significance |
|---|---|---|
| 3.5-mm LCP superiorly | 13.84 ± 2.58 (6.27 to 22.67) | NS† |
| 3.5-mm LCP anteriorly | 33.91 ± 10.86 (6.11 to 87.34) | S‡§ |
| 2.7-mm LCP superiorly and 2.4-mm Adaptation plate anteriorly | 24.35 ± 1.64 (18.04 to 28.50) | NS† |
| 2.7-mm Adaptation plate superiorly and 2.4-mm LCP anteriorly | 19.99 ± 4.36 (7.78 to 35.51) | NS† |
| 2.4-mm LCP superiorly and 2.7-mm Adaptation plate anteriorly | 27.80 ± 11.69 (2.17 to 72.27) | S‡# |
| 2.4-mm Adaptation plate superiorly and 2.7-mm LCP anteriorly | 31.22 ± 6.42 (11.10 to 51.97) | S‡# |
The values are given as the mean and the standard error, with the 95% CI in parentheses.
NS = nonsignificant. Configurations are not significantly different from one another.
S = significant. There is significant difference between this construct and ≥1 other constructs.
Anterior plating was a significantly stiffer construct than superior plating (p = 0.043).
The 2 dual-plate constructs with a superior 2.4-mm and anterior 2.7-mm plate were significantly stiffer than superior plating (p = 0.043).
Load to Failure
The results of the 3-point load to failure by bending with superior loading are displayed in Table VI. No significant differences (p = 0.360) were observed between the different groups.
TABLE VI.
Bending Load to Failure with Superior Loading by Plate Configuration Group
| Groups | Bending Load to Failure* (N/mm) | Significance† |
|---|---|---|
| 3.5-mm LCP superiorly | 254.75 ± 20.84 (192.56 to 325.20) | NS |
| 3.5-mm LCP anteriorly | 341.00 ± 55.74 (187.95 to 542.75) | NS |
| 2.7-mm LCP superiorly and 2.4-mm Adaptation plate anteriorly | 306.00 ± 15.99 (250.44 to 352.21) | NS |
| 2.7-mm Adaptation plate superiorly and 2.4-mm LCP anteriorly | 252.75 ± 26.84 (182.84 to 353.70) | NS |
| 2.4-mm LCP superiorly and 2.7-mm Adaptation plate anteriorly | 383.25 ± 40.59 (254.07 to 512.43) | NS |
| 2.4-mm Adaptation plate superiorly and 2.7-mm LCP anteriorly | 337.75 ± 23.53 (266.25 to 415.99) | NS |
The values are given as the mean and the standard error of the mean, with the 95% CI in parentheses.
NS = nonsignificant. Configurations are not significantly different from one another.
Location of Failure
For both superior plating (n = 2) and anterior plating (n = 4), the most frequent fracture location was at the interfragmentary screw (Table VII). In the superior-plating group, a fracture at the bone-screw interface was observed at the first medial screw in 2 specimens. For the 4 dual-plating constructs, no fractures were observed at the interfragmentary screw; most fractures were seen at the bone-screw interface at the first lateral screw (n = 3) or the first medial screw (n = 3) relative to the interfragmentary screw, followed by a fracture at the second medial screw (n = 1) or the second lateral screw (n = 1).
TABLE VII.
Location of Failure by Plate Configuration Group
| Groups | Fracture at Interfragmentary Screw | Fracture at Bone-Screw Interface | |||
|---|---|---|---|---|---|
| First Screw Medial | Second Screw Medial | First Screw Lateral | Second Screw Lateral | ||
| 3.5-mm LCP superiorly | 2 | 2 | |||
| 3.5-mm LCP anteriorly | 4 | ||||
| 2.7-mm LCP superiorly and 2.4-mm Adaptation plate anteriorly | 2 | 1 | 1 | ||
| 2.7-mm Adaptation plate superiorly and 2.4-mm LCP anteriorly | 3 | 1 | |||
| 2.4-mm LCP superiorly and 2.7-mm Adaptation plate anteriorly | 1 | 2 | 1 | ||
| 2.4-mm Adaptation plate superiorly and 2.7-mm LCP anteriorly | 1 | 3 | |||
Discussion
Dual plating using mini-fragment plates was biomechanically superior for fixation of midshaft clavicle fractures when compared with a single superior plate and had biomechanical properties similar to those of a single plate placed anteriorly. For axial stiffness, the 2 dual-plate constructs with a superior 2.4-mm plate and an anterior 2.7-mm plate were significantly stiffer than the other 4 constructs. This study used a cadaveric model with a priori random allocation, stratified by CT-based BMC to create homogenous groups and to minimize the effect of bone quality on our biomechanical outcomes.
Our findings are similar to those of the biomechanical study by Ziegler et al.21, who used 18 cadaveric clavicles. In this study, no significant differences were seen between dual plating (superior and anterior, 7-hole titanium, 2.7-mm LCP, 8 standard screws) and either superior or anterior single plating (7-hole titanium, 3.5-mm LCP, 6 standard screws) for axial, torsional, or superior bending stiffness or in bending load to failure. However, in accordance with our study, dual plating did seem to have a higher mean cyclical bending stiffness (14.65 N/mm) when compared with superior plating (6.75 N/mm) (p = 0.067). Those authors did not report on bending stiffness with an anterior load; consequently, no comparison could be made. Although an incomplete block design was used to randomize the specimens, they were not able to correct for differences in BMC among the different fixation groups; therefore, differences in cadaveric bone quality may have affected their results.
In a biomechanical and clinical study using synthetic clavicles (n = 19), Prasarn et al.13 reported similar biomechanical properties for dual plating (2.7-mm LCP superior and 2.4-mm reconstruction plate anterior, with 6 standard and 4 locking screws; material and plate length not reported) compared with superior plating and anterior plating (3.5-mm reconstruction plate, 6 locking screws; material and plate length not reported). For torsional and axial loading, no significant differences in construct stiffness were observed. With application of an anterior load, the dual-plate construct was significantly more rigid than an anterior plate, but was less rigid than a superior plate. The exact opposite was found when a superior load was applied. The authors concluded that single-plate constructs were least rigid when loaded parallel to the narrow dimension of the plate. Consequently, orthogonal plates may be better suited to resist multiplanar bending forces than a single plate13. This might be especially true in segmental or more comminuted clavicle fractures. Prasarn et al.13 used a transverse osteotomy (corresponding to a clinically less common fracture pattern), whereas, in our model and that of Ziegler et al.21, an inferior butterfly fragment was created. A systematic review of biomechanical studies on the surgical fixation of midshaft clavicle fractures stated that, for segmental or comminuted fractures, anterior plating was stiffer than superior plating in cantilever bending22. This finding was consistent with our results and the study by Ziegler et al.21. These findings support why we found that the 2 dual-plating constructs with a superior 2.4-mm plate combined with the more rigid 2.7-mm plate placed anteriorly are stiffer in cantilever bending when compared with the 2 opposite dual-plating constructs, and especially when compared with the single 3.5-mm superior plate.
In a biomechanical study, Boyce et al.19, using an inferior butterfly fragment in synthetic clavicles (n = 15), compared single superior plating (8-hole stainless steel 3.5-mm LCP with 6 locking screws), combination plating (an additional 10-hole titanium 2.8-mm LCP placed anteriorly and 9 locking screws), and dual mini-fragment plates (2 of the latter 2.8-mm plates placed superiorly and anteriorly, 18 locking screws). Not surprisingly, combination plating was the stiffest construct in torsion and cantilever bending. However, in terms of reducing implant prominence, this construct is likely suboptimal. Similar to the current study and that of Ziegler et al.21, Boyce et al. found that the location of failure in the majority of the single-plate constructs was at the fracture site. For numerous dual-plating constructs, they reported failure at the most lateral screw. Therefore, they recommended staggering the dual plates to minimize the stress riser created at the ends of the plates. However, for dual plating in both the current study and that of Ziegler et al.21, most failures occurred in either the first medial screw or the first lateral screw. This difference can be explained by the difference in surgical technique (18 locking screws compared with 8 standard screws). This difference is further illustrated by a recent finite element analysis (superior and anterior plating with a 6-hole, titanium, 3.5-mm LCP and 6 standard screws compared with dual plating with two 6-hole, 2.7-mm LCP and 8 standard screws). In this model, the concentration of stress found in the superior and anteroinferior single-plate constructs was located near the fracture gap in cantilever bending, axial compression, and axial torsion. In contrast, the force in the dual-plate construct was more equally distributed20. In terms of construct stability, dual plating exhibited the highest stiffness and the least micromotion in their models20. If a construct is too rigid, this could lead to stress-shielding and the clavicle might fracture at the periphery, as depicted by the study of Boyce et al.19. Ensuring that the dual plates are of different lengths could also reduce stress concentrations at the end of the plate-screw constructs. Additionally, a too-rigid construct might predispose to higher nonunion rates. However, a systematic review and meta-analysis described excellent union rates for dual plating (99.5%) and no significant differences were observed among all surgical fixation types8. It seems that dual plating is biomechanically superior for fixation of midshaft clavicle fractures when compared with a single superior plate. Dual plating had biomechanical properties similar to those of a single plate placed anteriorly, although it appears that orthogonal plates may be better suited to resist multiplanar bending and rotational forces, especially in more comminuted or segmental fractures, and therefore might better endure early weight-bearing.
The current study had several limitations. Although our biomechanical study had the largest number of specimens (n = 24), the limited number of clavicles per group (n = 4) made it possible to detect only large effects (leading to a potential type-II error). However, it is more than adequate to test for biomechanical noninferiority by applying nonparametric testing. Cadaveric clavicles were used, whereas synthetic clavicles provide a more consistent material and specimen size for biomechanical testing of the different configurations. However, by using human bone, the results are more realistic and clinically applicable. Our specimens were from patients older than those who would likely undergo surgical fixation. Although this does not warrant the use of locking screws22, it does increase the risk of poorer fixation in osteopenic bone and might amplify clinical differences. However, a priori random allocation, with stratification by BMC, ensured homogenous groups and minimized the effect of bone quality on our biomechanical outcomes.
Dual-plating configurations using mini-fragment plates were biomechanically superior for the fixation of midshaft clavicle fractures when compared with a single, superior, 3.5-mm plate and had biomechanical properties similar to those of a 3.5-mm plate placed anteriorly. It is therefore a viable treatment option, especially given that the use of these lower-profile implants could substantially reduce implant removal rates. With the exception of axial stiffness, no significant differences were found when the different dual-plating constructs were compared with each other. However, placing a 2.4-mm plate superiorly, in combination with a 2.7-mm plate anteriorly, seems to be the most rigid construct given its biomechanical superiority in cantilever bending over the 3.5-mm plate placed superiorly.
Acknowledgments
Note: The authors thank Ifaz Haider, Maria Beketskaia, Rohit Bansal, and W. Brent Edwards.
Footnotes
Investigation performed at the McCaig Institute for Bone and Joint Health, Cumming School of Medicine, University of Calgary, Calgary, Alberta, Canada
Disclosure: The Disclosure of Potential Conflicts of Interest forms are provided with the online version of the article (http://links.lww.com/JBJSOA/A359).
References
- 1.Canadian Orthopaedic Trauma Society. Nonoperative treatment compared with plate fixation of displaced midshaft clavicular fractures. A multicenter, randomized clinical trial. J Bone Joint Surg Am. 2007. Jan;89(1):1-10. [DOI] [PubMed] [Google Scholar]
- 2.DeFroda SF, Lemme N, Kleiner J, Gil J, Owens BD. Incidence and mechanism of injury of clavicle fractures in the NEISS database: athletic and non athletic injuries. J Clin Orthop Trauma. 2019. Sep-Oct;10(5):954-8. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 3.Kihlström C, Möller M, Lönn K, Wolf O. Clavicle fractures: epidemiology, classification and treatment of 2 422 fractures in the Swedish Fracture Register; an observational study. BMC Musculoskelet Disord. 2017. Feb 15;18(1):82. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 4.Ahrens PM, Garlick NI, Barber J, Tims EM; Clavicle Trial Collaborative Group. The Clavicle Trial: a multicenter randomized controlled trial comparing operative with nonoperative treatment of displaced midshaft clavicle fractures. J Bone Joint Surg Am. 2017. Aug 16;99(16):1345-54. [DOI] [PubMed] [Google Scholar]
- 5.Amer K, Smith B, Thomson JE, Congiusta D, Reilly MC, Sirkin MS, Adams MR. Operative versus nonoperative outcomes of middle-third clavicle fractures: a systematic review and meta-analysis. J Orthop Trauma. 2020. Jan;34(1):e6-13. [DOI] [PubMed] [Google Scholar]
- 6.Smith JR, Kitzen J, Buckley R. Midshaft clavicle fracture - nonoperative versus operative care. Injury. 2021. Aug;52(8):2049-51. [DOI] [PubMed] [Google Scholar]
- 7.Schneider P, Bransford R, Harvey E, Agel J. Operative treatment of displaced midshaft clavicle fractures: has randomised control trial evidence changed practice patterns? BMJ Open. 2019. Sep 4;9(9):e031118. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 8.You DZ, Krzyzaniak H, Kendal JK, Martin CR, Schneider PS. Outcomes and complications after dual plate vs. single plate fixation of displaced mid-shaft clavicle fractures: a systematic review and meta-analysis. J Clin Orthop Trauma. 2021. Apr 14;17:261-6. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 9.Wiesel B, Nagda S, Mehta S, Churchill R. Management of midshaft clavicle fractures in adults. J Am Acad Orthop Surg. 2018. Nov 15;26(22):e468-76. [DOI] [PubMed] [Google Scholar]
- 10.Allis JB, Cheung EC, Farrell ED, Johnson EE, Jeffcoat DM. Dual versus single-plate fixation of midshaft clavicular fractures: a retrospective comparative study. JB JS Open Access. 2020. Apr 1;5(2):e0043. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 11.DeBaun MR, Chen MJ, Campbell ST, Goodnough LH, Lai C, Salazar BP, Bishop JA, Gardner MJ. Dual mini-fragment plating is comparable with precontoured small fragment plating for operative diaphyseal clavicle fractures: a retrospective cohort study. J Orthop Trauma. 2020. Jul;34(7):e229-32. [DOI] [PubMed] [Google Scholar]
- 12.Zhuang Y, Zhang Y, Zhou L, Zhang J, Jiang G, Wu J. Management of comminuted mid-shaft clavicular fractures: comparison between dual-plate fixation treatment and single-plate fixation. J Orthop Surg (Hong Kong). 2020. Jan-Apr;28(2):2309499020915797. [DOI] [PubMed] [Google Scholar]
- 13.Prasarn ML, Meyers KN, Wilkin G, Wellman DS, Chan DB, Ahn J, Lorich DG, Helfet DL. Dual mini-fragment plating for midshaft clavicle fractures: a clinical and biomechanical investigation. Arch Orthop Trauma Surg. 2015. Dec;135(12):1655-62. [DOI] [PubMed] [Google Scholar]
- 14.Lee C, Feaker DA, Ostrofe AA, Smith CS. No difference in risk of implant removal between orthogonal mini-fragment and single small-fragment plating of midshaft clavicle fractures in a military population: a preliminary study. Clin Orthop Relat Res. 2020. Apr;478(4):741-9. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 15.Chen X, Shannon SF, Torchia M, Schoch B. Radiographic outcomes of single versus dual plate fixation of acute mid-shaft clavicle fractures. Arch Orthop Trauma Surg. 2017. Jun;137(6):749-54. [DOI] [PubMed] [Google Scholar]
- 16.Czajka CM, Kay A, Gary JL, Prasarn ML, Choo AM, Munz JW, Harvin WH, Achor TS. Symptomatic implant removal following dual mini-fragment plating for clavicular shaft fractures. J Orthop Trauma. 2017. Apr;31(4):236-40. [DOI] [PubMed] [Google Scholar]
- 17.Shannon SF, Chen X, Torchia M, Schoch B. Extraperiosteal dual plate fixation of acute mid-shaft clavicle fractures: a technical trick. J Orthop Trauma. 2016. Oct;30(10):e346-50. [DOI] [PubMed] [Google Scholar]
- 18.Nourian A, Dhaliwal S, Vangala S, Vezeridis PS. Midshaft fractures of the clavicle: a meta-analysis comparing surgical fixation using anteroinferior plating versus superior plating. J Orthop Trauma. 2017. Sep;31(9):461-7. [DOI] [PubMed] [Google Scholar]
- 19.Boyce GN, Philpott AJ, Ackland DC, Ek ET. Single versus dual orthogonal plating for comminuted midshaft clavicle fractures: a biomechanics study. J Orthop Surg Res. 2020. Jul 9;15(1):248. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 20.Zhang F, Chen F, Qi Y, Qian Z, Ni S, Zhong Z, Zhang X, Li D, Yu B. Finite element analysis of dual small plate fixation and single plate fixation for treatment of midshaft clavicle fractures. J Orthop Surg Res. 2020. Apr 15;15(1):148. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 21.Ziegler CG, Aman ZS, Storaci HW, Finch H, Dornan GJ, Kennedy MI, Provencher MT, Hackett TR. Low-profile dual small plate fixation is biomechanically similar to larger superior or anteroinferior single plate fixation of midshaft clavicle fractures. Am J Sports Med. 2019. Sep;47(11):2678-85. [DOI] [PubMed] [Google Scholar]
- 22.Hulsmans MH, van Heijl M, Houwert RM, Burger BJ, Verleisdonk EJM, Veeger DJ, van der Meijden OA. Surgical fixation of midshaft clavicle fractures: a systematic review of biomechanical studies. Injury. 2018. Apr;49(4):753-65. [DOI] [PubMed] [Google Scholar]