Abstract
Introduction
Implantation of cardiac implantable electronic devices (CIEDs) is an art of science. As the volume of implantation has increased worldwide, so has the rate of complications. Infection, fibrosis, lead and device erosion, lead displacement, right ventricle perforation, lead fracture, and insulation break are the common complications in the implantation process. This exposes the patient for reopening and threatens the implantation for further complication due to infection, fibrosis of veins, failure to retrieve the implanted wire, and failure to re-implant the device on the same site. We slightly changed our implantation technique to preserve the implantation site for future implantation and reduce the rate of complication in the index implantation.
Methods
This randomized control trial was conducted from January 2016 to September 2019 at Hayatabad Medical Complex Peshawar, Pakistan. A consecutive sampling technique was used to obtain a sample size of 602 patients keeping a 95% confidence interval and a 5% margin error. We adopted a strategy to take prick, for implantation of devices, inside the pocket, which reduces the number of sutures, hastens the procedure, prevents erosion, and minimizes the chance of subclavian crush syndrome and insulation break. We also selected the minimum possible length of leads. This will possibly decrease the chances of cumbersome fibrosis around the lead and device and will make future implantation convenient.
Results
There was a total of 602 procedures in the study period. About 253 (42%) procedures were done in the newly adopted strategy and 349 (58%) were performed in the conventional way. Our complication rate grossly reduces in the novel way of implantation in which we took our prick inside the pocket.
Conclusion
A slight modification in the implantation of CIEDs not only prevents the rate of complication in the index implantation but will also possibly preserve the site for future implantation.
Keywords: insulation break, lead fracture, right ventricle perforation, lead displacement, lead & device erosion
Introduction
Implantation of cardiac implantable electronic devices (CIEDs) has become an integral part of cardiology in general and of electrophysiology in particular [1]. Implantation of cardiac devices is not a one-time procedure. Most cardiac devices remain functional from 8-10 years after which the pulse generator needs to be replaced. People who become dependent on devices at a very early age need multiple implantations in their lifetime. Mostly the same leads from the previous implantation are used and only the device is replaced. Therefore, explantation needs to be carefully handled and the device and leads freed from adhesion and fibrosis to successfully implant the new device.
Cardiac device implantations are very prone to complications [2]. These complications range from very mild discomfort to serious consequences, which can put at risk the life of the patient. The most common complication with cardiac devices are pocket hematoma, erosion, lead displacement, insulation break, lead fracture, subclavian crush syndrome, right-sided cardiac chambers perforation, and endocarditis associated with fibrosis, infection, and lead [3]. Multiple implantations not only increase the rate of complication but also make future implantation difficult at the same site [4].
CIEDs implantation needs special care to make the index implantation successful, preserve the site for future implantation and replacement, and possibly reduce the rate of potential future complications [5]. These precautions start in the preoperative period when the device implantation is planned and stay throughout the postoperative period for the rest of the patient’s life. The dreaded complications are implantation site infection, fibrosis around the lead and devices, stenosis of the veins, and erosion of lead and devices. All these complications will make future implantation cumbersome [2].
Our big enemies in the implantation of CIEDs, in developing counties, are infections [6]. This not only spoils the index implantation but makes future implantation difficult and often endangers the life of the patient. It may be due to faulty implantation surgical skill, erosion of the devices, or trauma to the device’s site [7]. The other common complication that makes future implantation difficult or impossible is fibrosis around the device and inside the vein [8].
Preserving the implantation site is the key for future implantation. During our implantation of CIEDs, we experienced almost all the mentioned complications in the past two decades and, therefore, we continuously reviewed our implantation procedure, not only to minimize the rate of complication but to preserve the site for future implantation and make the re-implantation easy. In this article we share our own experience in the field of implantation and preservation of implantation sites.
Materials and methods
This randomized control trial was conducted from January 2016 to September 2019 at Hayatabad Medical Complex Peshawar, Pakistan. Using the reference sample size from a study published in 2018 [9] and the WHO calculator, a sample size of 602 patients was calculated with a 95% confidence interval and a 5% margin error. A consecutive sampling technique was used. Patients were randomly divided into two groups. Group “A” patients received CIEDs implantation in the conventional way and group “B” patients received the implantation of devices in the new way we adopted for device implantation.
The study was approved by the Scientific Review Committee and Ethical Committee (Human Research) of Post Graduate Medical Institute (PGMI), Peshawar, Pakistan.
Inclusion and exclusion criteria
All patients, irrespective of age, who presented for device implantation as elective cases or as emergency cases were included in our study. Patients who presented with device or lead erosion and patients who were having infection at the site of implantation were excluded from our study.
Data collection
All those patients who presented to our unit for an elective procedure of implantation of CIEDs were admitted a day before the procedure. Patients who presented to emergency room and were in need of temporary pacemakers, or some other emergency procedure, got admitted at the time of presentation.
Informed consent was obtained. Baseline investigations including renal function test (RFT), serum electrolytes, random blood sugar (RBS), virology, full blood count (FBC), liver function test (LFT), coagulation profile, and cardiac enzyme blood test were advised according to the hospital protocol. Night bath was offered to those who were able to do so. Male patients’ chests were shaved at night. Both male and female patient chests were painted with Pyodine. Night sedation with 3 mg bromazepam orally administered.
Patients were brought nil by mouth to the catheterization laboratory. Intravenous (IV) cannula were passed on both upper limbs. Premedication with IV dimenhydrinate, injection nalbuphine, and at time, injection midazolam was given. It is our usual practice to give IV antibiotics half an hour before surgery, and second dose during surgery or just at the end of procedure if implantation time is less 30 minutes.
After patient was scrubbed, draped venography was performed of the axillary vein on the side of implantation and local anesthesia with 2% lidocaine was infiltrated.
Group “A” patients: Devicewas implanted in the conventional way of implantation. After skin incision and pocket construction, guide wire was passed through Seldinger technique and sheath passed. After removal of wire and dilator, lead was implanted, battery attached, and pocket was closed. Lead berried in muscle and wound closed in layers.
Group “B” patients: Through the intact skin, axillary vein puncture was done by Seldinger technique and guide wire passed. About 2-3 cm superolateral to the puncture site, skin incision was made and self-retaining retractor applied. With blunt dissection, the subcutaneous tissues were separated using artery forceps till the deep fascia over the muscle was visible. Electric cattery was used to keep the field clean and dry. With tooth forceps, the flap of the skin is left and dissection extended infero-medially to construct the pocket for device. The preplaced guide wires were identified.
Once the guide wire came into the site, they were pulled in from inside the pocket, so that the outer end came in and out through the incision while the venous end was in place. In this way the puncture site goes inside the pocket. Now the sheath was passed over the wire and wire pulled out with dilator and PPM lead passed and secured. The lead was stabilized with muscle using 1/0 non-absorbable sutures. The pocket for pulse generator was irrigated with gentamicin so to remove any residual blood clots.
The device was attached with pacemakers leads, and battery pushed inside the pocket and stabilized with 1/0 silk inside the pocket. The pocket mouth was closed with 1/0 Vicryl and the skin was closed with interrupted non-absorbable sutures. Pressure dressing was applied for 24 hours to prevent any possible pocket hematoma. Patients were kept on IV antibiotics for five days according to our hospital microbiology protocol. Patients were discharged on oral antibiotics for five more days and skin stitches were removed on day 15 of procedure.
Data analysis
IBM SPSS Statistics for Windows, Version 22.0 (Released 2013, IBM Corp., Armonk, New York, United States) was used for statistical analysis involving quantitative variables.
Results
There were a total of 602 patients in the study. The patients whose procedure was performed in the conventional way numbered 349 (58%). The new strategy was tried in 253 (42%) patients. The demographic data of the patients are tabulated in Table 1. There were 325 (54%) male patients and 277 (46%) female patients. However, there was no significant difference in the rate of complication in both genders. The age distribution was from 10 years to 100 years. The mean age of implantation remains 60.26±16.88 years. Lead displacement and lead damage were seen more frequently in the young age group while pneumothorax complicated the procedure above 60 years of age. The prick site and incision line is demonstrated in Figure 1.
Table 1. Baseline characteristic of patients.
VVI/VVIR: ventricular demand pacing single chamber pacemaker; DDD/DDDR: dual chamber rate adaptive pacemaker
| DEMOGRAPHIC CHARACTERS | NUMBERS |
| Total Patients | 602 |
| Male | 325 (54%) |
| Female | 277 (46%) |
| Age | 10 to 100 (60.26±16.88)years |
| Tine Lead | 27 (4.5%) |
| Screwing Lead | 573 (95.2%) |
| Tine And Screwing Leads | 2 (0.3%) |
| VVI/VVIR | 432 (71.8%) |
| DDD/DDDR | 170 (28.2%) |
| New Implantation | 550 (91.4%) |
| Box Change | 19 (3.2%) |
| Upgradation Of Devices | 1 (0.2%) |
| Reposition Of Lead | 21 (3%) |
| Lead Abandoned | 1 (0.2%) |
| Explantation | 8 (1.3%) |
| Conventional Implantation Method | 349 (58%) |
| Novel /Change Strategy | 253 (42%) |
Figure 1. Demonstration of prick site and incision line.
The rate of pneumothorax was high in the conventional method of implantation, which is statistically significant. The comparison of complications in both our conventional method and novel changed strategy is tabulated in Table 2. The rate of almost all complications is more with dual-chamber pacemakers. These complications disturb the field of implantation and make future implantation cumbersome. However, single chambers devices have a very limited indication.
Table 2. Complications in procedure.
| COMPLICATION IN PROCEDURE | METHOD | |
| CONVENTIONAL | NOVEL | |
| Lead Displacement | 3 | 2 |
| Mild Pericardial Effusion | 0 | 1 |
| Haematoma | 2 | 0 |
| Infection | 1 | 0 |
| Pneumothorax | 7 | 2 |
| Lead Damage | 1 | 1 |
Tine leads are almost obsolete in today's practice, but we had 27 tine leads in our study population. There were 21 new implantations of tine leads and mostly they were implanted in atria, which were pre-shaped leads. So there is no chance of displacement in these leads. However, the retrieval rate is significantly low with these leads, and most of the time they are abandoned if there is damage to the leads.
Discussion
As CIEDs implantation is not a one-time procedure, re-implantation of CIEDs is a cumbersome procedure [10]. Re-implantation of the device by itself is not only difficult but it is much more prone to infection as compared to the primary procedure [11]. The recovery of the previous CIED and re-implantation success mostly depends on the first procedure. If the anatomy of the implantation site is destroyed at the time of primary implantation, the subsequent procedure will be very much difficult. Therefore preservation of the implantation site is of utmost importance. If the first procedure is performed in a way that there is hematoma formation, infection, and a lot of adhesion, then the repeat procedure will be not only difficult but may lead to failure of successful implantation [12].
Small hematoma and serous collection are common after implantation of CIEDs [13]. But large hematoma at the implantation site makes the site prone to dehisces, infection, adhesion, and delayed healing [14]. Apart from this, it will make the area unsuitable for subsequent procedures. Therefore, cautious hematoma prevention at first implantation will not only prevent failure of the first procedure but will make the following procedures a success. This will decrease the ratio of adhesion and the retrieval of the first lead and device will be easy.
Repeated unsuccessful attempts to prick the vein, and consequently hitting the axillary artery repeatedly, increase the chance of hematoma formation [15]. Therefore prior identification of the vein site by a venogram or ultrasound [16] and avoiding incidental pricking of the axillary artery will prevent hematoma. Plugging and securing the lead sleeve properly at the entry point in muscle also prevents oozing and collection of hematoma [17].
If the lead is lying directly in the line of incision, then it is necessary to bury it in muscles to avoid lead erosion, particularly in very young and elderly populations [2]. But this will increase the chance of adhesion in the future and retrieval of lead will be difficult. Similarly, the number of leads also affects the outcome. An increased number of leads will complicate the procedure in several ways. It increases the rate of procedure-related complications and also the increased number of hardware leads to increased adhesions and fibrosis [18]. But multiple leads are almost always needed in different morbidities and indication for a single chamber is very limited. Therefore, we adopted a technique to accommodate the whole hardware in the pocket in a way that it is not only in a well-protected position but also minimizes the procedure-related complication and saves the area for future implantation.
Increased numbers of suture material also increase the chance of infection and adhesion [19]. The suture martial was much reduced in our new technique. Only one suture is sufficient to bury the whole hardware in the pocket and there is less chance of infection and lead erosion. The chance of wound dehiscence due to pressure on the suture line is negligible.
The increased length of the lead and putting a lot of lead folded under the device also increase adhesion and subsequent retrieval difficulty [18]. Therefore, we selected the minimum possible length of the lead, which prevented the over folding of lead and decreased the cumbersome adhesions. This makes the retrieval easy subsequently and also prevents damage to the area for subsequent implantation.
Electrocautery reduces the chance of hematoma and keeps the implantation site dry [20]. Similarly, irrigating the pocket with saline and antibiotics to wash out any clotted blood also reduces the chance of infection and adhesions [21,22]. Applying pressure bandage to the wound site is very helpful in reducing collection and makes the recovery speedy [23]. Therefore, there is less chance of subsequent adhesion.
We do not practice the cut-down procedure of the cephalic vein [24] anymore, because there is more manipulation of structure, which ultimately increases fibrosis and makes future implantation more difficult. Skin incision at the time of explantation of the device is very important. Most operators cut the skin over the device, which will make the access easy to the previously implanted device but re-implantation becomes difficult at the same time and the pocket is needed to be extended downward to accommodate the hardware in the pocket. This process makes the procedure prone to further complications. Therefore, we give an incision at or near the old incision and carefully dissect the area to retrieve the device. This technique increases the time of explantation and re-implantation but preserves the same implantation site for future repeated implantation.
As infection is our biggest enemy, which not only spoils the index procedure but also leads to the destruction of the site for future implantation [25], we offer a night bath before the procedure if possible, paint the site with Pyodine solution at night, and give inject able broad-spectrum antibiotics an hour before the procedure and the second dose during the procedure. Patients remain on antibiotics for ten days to prevent infection. Patients with temporary pacemakers (TPM) are started on antibiotics at the time of implantation of TPM and remain on antibiotics till the patients get a permanent pacemaker (PPM).
No matter how careful and equipped we are, still, some unavoidable conditions pop up in the process of CIEDs implantation. Abnormal response of the human body to foreign materials is a cumbersome complication [26,27], which leads to exaggerated fibrosis and stenosis of veins. Leadless pacemakers [28] are best for the prevention of subclavian area and vein stenosis but it does not fulfill all the aims of CIEDs implantation. Future research of stem cell therapy will possibly be the right answer to this question.
Limitation of the study
Results are in favor of our change strategy as far as the procedure-related complications and efficacy of the procedure are concerned but we need time to compare the re-exploration results of the two approaches.
Conclusions
As CIEDs implantation is not a one-time procedure, therefore, the site of implantation needs to be preserved. If the rate of complication could be reduced, the area of implantation may be preserved for a long period of time. The complications rate is less with axillary vein prick and using an appropriate length of leads. Carefully positioning the lead at the time of implantation, so as to avoid displacement and repositioning will reduce complications. Last but not least, minimizing the chance of infection by appropriate use of antibiotics is the key to preventing erosion and preserving the site.
The content published in Cureus is the result of clinical experience and/or research by independent individuals or organizations. Cureus is not responsible for the scientific accuracy or reliability of data or conclusions published herein. All content published within Cureus is intended only for educational, research and reference purposes. Additionally, articles published within Cureus should not be deemed a suitable substitute for the advice of a qualified health care professional. Do not disregard or avoid professional medical advice due to content published within Cureus.
The authors have declared that no competing interests exist.
Human Ethics
Consent was obtained or waived by all participants in this study. Post Graduate Medical Institute (PGMI), Peshawar, Pakistan issued approval Ref ERC: PGMI-08-Jan16. This is to certify that ethical approval has been granted to Dr. Bakhtawar Shah, Medical Officer working in the Department of Cardiology, Hayatabad Medical Complex Peshawar, Peshawar, Pakistan, to the project titled “Novel Strategy for PPM Implantation”.
Animal Ethics
Animal subjects: All authors have confirmed that this study did not involve animal subjects or tissue.
References
- 1.Incidence and predictors of perioperative complications with transvenous lead extractions: real-world experience with national cardiovascular data registry. Sood N, Martin DT, Lampert R, Curtis JP, Parzynski C, Clancy J. Circ Arrhythm Electrophysiol. 2018;11:0. doi: 10.1161/CIRCEP.116.004768. [DOI] [PubMed] [Google Scholar]
- 2.Pacemaker lead erosion simulating "Loch Ness Monster": conservative management. Garg N, Moorthy N. Asian Cardiovasc Thorac Ann. 2012;20:718–720. doi: 10.1177/0218492312441357. [DOI] [PubMed] [Google Scholar]
- 3.Permanent leadless cardiac pacemaker therapy: a comprehensive review. Tjong FV, Reddy VY. Circulation. 2017;135:1458–1470. doi: 10.1161/CIRCULATIONAHA.116.025037. [DOI] [PubMed] [Google Scholar]
- 4.Complications of cardiac implantable electronic device placement in public and private hospitals. Ganesan AN, Moore K, Horton D, et al. Intern Med J. 2020;50:1207–1216. doi: 10.1111/imj.14704. [DOI] [PubMed] [Google Scholar]
- 5.Lead management and lead extraction. Love CJ. Card Electrophysiol Clin. 2018;10:127–136. doi: 10.1016/j.ccep.2017.11.013. [DOI] [PubMed] [Google Scholar]
- 6.Incidence and risk factors of surgical site infection in general surgery in a developing country. Alp E, Elmali F, Ersoy S, Kucuk C, Doganay M. Surg Today. 2014;44:685–689. doi: 10.1007/s00595-013-0705-3. [DOI] [PubMed] [Google Scholar]
- 7.Implantable device-related infection. VanEpps JS, Younger JG. Shock. 2016;46:597–608. doi: 10.1097/SHK.0000000000000692. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 8.Overcoming the challenge of venous occlusion for lead implantation. Burri H. Indian Pacing Electrophysiol J. 2015;15:110–112. doi: 10.1016/j.ipej.2015.07.006. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 9.Indication and frequency of implantation of permanent pacemaker in complete heart block patients in a tertiary care hospital in Rawalpindi. Raja TA, Kashif M, Rehman AU, Kiyani A, Shabbir F, Rajput TA. https://www.pafmj.org/index.php/PAFMJ/article/view/2319 PAFMJ. 2018;68:1339–1343. [Google Scholar]
- 10.Extraction of transvenous pacing and ICD leads. Smith MC, Love CJ. Pacing Clin Electrophysiol. 2008;31:736–752. doi: 10.1111/j.1540-8159.2008.01079.x. [DOI] [PubMed] [Google Scholar]
- 11.Timing of device reimplantation and reinfection rates following cardiac implantable electronic device infection: a systematic review and meta-analysis. Chew D, Somayaji R, Conly J, Exner D, Rennert-May E. BMJ Open. 2019;9:0. doi: 10.1136/bmjopen-2019-029537. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 12.Reimplantation and repeat infection after cardiac-implantable electronic device infections: experience from the MEDIC (multicenter electrophysiologic device infection cohort) database. Boyle TA, Uslan DZ, Prutkin JM, et al. Circ Arrhythm Electrophysiol. 2017;10:0. doi: 10.1161/CIRCEP.116.004822. [DOI] [PubMed] [Google Scholar]
- 13.The relationship between warfarin, aspirin, and clopidogrel continuation in the peri-procedural period and the incidence of hematoma formation after device implantation. Thal S, Moukabary T, Boyella R, Shanmugasundaram M, Pierce MK, Thai H, Goldman S. Pacing Clin Electrophysiol. 2010;33:385–388. doi: 10.1111/j.1540-8159.2009.02674.x. [DOI] [PubMed] [Google Scholar]
- 14.Challenges and management of surgical site occurrences. Gabriel A, Gupta S, Orgill DP. Plast Reconstr Surg. 2019;143:7–10. doi: 10.1097/PRS.0000000000005305. [DOI] [PubMed] [Google Scholar]
- 15.A case of unstable occult arterial bleeding post pacemaker implantation. Tan K, Tan LW, Chia PL, Foo D. J Arrhythm. 2019;35:773–775. doi: 10.1002/joa3.12223. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 16.Ultrasound-guided venous access for pacemakers and defibrillators. Seto AH, Jolly A, Salcedo J. J Cardiovasc Electrophysiol. 2013;24:370–374. doi: 10.1111/jce.12005. [DOI] [PubMed] [Google Scholar]
- 17.Evaluation of common suturing techniques to secure implantable cardiac electronic device leads: Which strategy best reduces the lead dislodgement risk? Rezazadeh S, Wang S, Rizkallah J. Can J Surg. 2019;62:0–3. doi: 10.1503/cjs.002418. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 18.Prolonged lead dwell time and lead burden predict bailout transfemoral lead extraction. Gould J, Sidhu BS, Porter B, et al. Pacing Clin Electrophysiol. 2019;42:1355–1364. doi: 10.1111/pace.13791. [DOI] [PubMed] [Google Scholar]
- 19.Bacterial adherence around sutures of different material at grafted site: a microbiological analysis. Mahesh L, Kumar VR, Jain A, et al. Materials (Basel) 2019;12:2848. doi: 10.3390/ma12182848. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 20.Pocket hematoma after pacemaker or implantable cardioverter defibrillator surgery: influence of patient morbidity, operation strategy, and perioperative antiplatelet/anticoagulation therapy. Wiegand UK, LeJeune D, Boguschewski F, Bonnemeier H, Eberhardt F, Schunkert H, Bode F. Chest. 2004;126:1177–1186. doi: 10.1378/chest.126.4.1177. [DOI] [PubMed] [Google Scholar]
- 21.Confounding factors on the prediction of opioid usage after thyroidectomy and parathyroidectomy surgery. Kang FG. Otolaryngol Head Neck Surg. 2019;161:371. doi: 10.1177/0194599819841606. [DOI] [PubMed] [Google Scholar]
- 22.Incidence of pocket infection postcardiac device implantation using antibiotic versus saline solution for pocket irrigation. Lakshmanadoss U, Nuanez B, Kutinsky I, Khalid R, Haines DE, Wong WS. Pacing Clin Electrophysiol. 2016;39:978–984. doi: 10.1111/pace.12920. [DOI] [PubMed] [Google Scholar]
- 23.Effectiveness of a non-taped compression dress in patients receiving cardiac implantable electronic devices. Chien CY, Chang YH, Wu YJ, et al. Acta Cardiol Sin. 2019;35:320–324. doi: 10.6515/ACS.201905_35(3).20190107A. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 24.Improved 'cut-down' technique for transvenous pacemaker lead implantation. Kolettis TM, Lysitsas DN, Apostolidis D, Baltogiannis GG, Sourla E, Michalis LK. Europace. 2010;12:1282–1285. doi: 10.1093/europace/euq173. [DOI] [PubMed] [Google Scholar]
- 25.Epicardial pacemaker as a bridge for pacemaker-dependent patients undergoing explantation of infected cardiac implantable electronic devices. Al-Maisary SS, Romano G, Karck M, De Simone R. J Card Surg. 2019;34:424–427. doi: 10.1111/jocs.14058. [DOI] [PubMed] [Google Scholar]
- 26.A rare case of recurrent pacemaker allergic reaction. Shittu M, Shah P, Elkhalili W, Suleiman A, Shaaban H, Shah PA, Shamoon F. Heart Views. 2015;16:59–61. doi: 10.4103/1995-705X.159222. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 27.Pacemaker contact sensitivity: clinical recognition and management. Abdallah HI, Balsara RK, O'Riordan AC. Ann Thorac Surg. 1994;57:1017–1018. doi: 10.1016/0003-4975(94)90228-3. [DOI] [PubMed] [Google Scholar]
- 28.Leadless pacemakers. El-Chami MF, Merchant FM, Leon AR. Am J Cardiol. 2017;119:145–148. doi: 10.1016/j.amjcard.2016.10.012. [DOI] [PubMed] [Google Scholar]