Smart Blood Vessel Detection System for Laparoscopic Surgery

Ching-Chia Li 1, Bor-Shing Lin 2, Sheng-Chen Wen 1, Yuan-Teng Liang 3, Hung-Yu Sung 3, Jhen-Hao Jhan 1, Bor-Shyh Lin 3,
PMCID: PMC8939714  PMID: 35345534

Abstract

Objective: Compared with traditional surgery, laparoscopic surgery offers the advantages of smaller scars and rapid recovery and has gradually become popular. However, laparoscopic surgery has the limitation of low visibility and a lack of touch sense. As such, a physician may unexpectedly damage blood vessels, causing massive bleeding. In clinical settings, Doppler ultrasound is commonly used to detect vascular locations, but this approach is affected by the measuring angle and bone shadow and has poor ability to distinguish arteries from veins. To tackle these problems, a smart blood vessel detection system for laparoscopic surgery is proposed. Methods: Based on the principle of near-infrared spectroscopy, the proposed instrument can access hemoglobin (HbT) parameters at several depths simultaneously and recognize human tissue type by using a neural network. Results: Using the differences in HbT and StO2 between different tissues, vascular and avascular locations can be recognized. Moreover, a mechanically rotatable stick enables the physician to easily operate in body cavities. Phantom and animal experiments were performed to validate the system’s performance. Conclusion: The proposed system has high ability to distinguish vascular from avascular locations at various depths.

Keywords: Laparoscopic surgery, near infrared spectroscopy, hemoglobin parameters, neural network, vascular location


Clinical and Translational Impact Statement—A smart blood vessel detection system for laparoscopic surgery was proposed. Using the differences in HbT and StO2 values between different tissues, vascular and avascular locations can be successfully recognized. (Category: Preclinical Research).

I. Introduction

With the advent of laparoscopy, patients now experience less pain at a surgical site after the surgery. Compared with open surgery, laparoscopy offers a better cosmetic effect, quicker recovery, and comparable efficacy. However, laparoscopic procedures require a steeper learning curve than open procedures. In open surgery, the surgeon must confirm a vessel’s location through visual inspection and palpation. However, blood flow pulsation cannot be observed through palpation—the surgeon depends on their experience and visual inspection; thus, injuries such as massive bleeding due to accidental vessel rupture may occur during laparoscopic surgery.

In laparoscopic surgery, Doppler ultrasound is generally used to determine the location of blood vessels. The Doppler instrument [1] uses the frequency difference between the transmitted and reflected ultrasound signals to estimate blood flow change and determine the possible location of blood vessels. However, the tip direction of the ultrasound probe is difficult to control when the probe is passed through a trocar in the human body [2], and the quality of ultrasound imaging is lower at greater depth [3]. Moreover, Doppler ultrasound can only distinguish a change in blood flow but not distinguish between tissue types [4]. In 2007, Klein et al. designed a piezoelectric needle producing 1–8-kHz vibrations at the needle tip to assist the Doppler instrument in monitoring blood flow [5]. To improve the quality of color Doppler images, the needle must be inserted into the human body to vibrate the local tissue. Recently, the near-infrared vein visualization device was developed for real-time visualization of superficial vessels under the skin [6][8]; near-infrared light (wavelength: ~785 nm) is used to penetrate the skin and blood vessels and is partially absorbed at avascular locations but not at vascular locations. Using the difference in absorption between the vascular and avascular locations, the device determines the location of blood vessels. However, this technique cannot be employed for measurement depths exceeding 5 mm.

To address the aforementioned problems, a smart blood vessel detection system for laparoscopic surgery was proposed in this study for differentiating vascular and avascular locations in vivo. Based on the difference in absorption spectra for different tissues [9], [10], the proposed instrument can differentiate vascular and avascular locations at various depths of human tissue simultaneously. Moreover, the rotatable stick enables the physician to easily control the angle and direction of the optical probe in the human body cavity. Phantom and animal experiments were performed to validate the system’s performance; the experimental results confirmed that the proposed instrument has high performance in detecting blood vessels at different depths.

II. Methods

A. Design and Implementation of the Smart Blood Vessel Detection System

Using the difference in absorption spectra of different human tissues, tissue type can be recognized through the changes in optical intensity at a specific wavelength [11][14], and this approach can be employed for detecting blood vessels under adipose or lymphatic tissue. The basic structure and a photograph of the proposed smart blood vessel detection system for laparoscopic surgery are shown in Fig. 1 (a) and Fig. 1 (b), respectively; the proposed system mainly contains an optical vessel detection instrument and a backend monitoring platform.

FIGURE 1.

FIGURE 1.

(a) Basic structure and (b) photograph of the smart blood vessel detection system for laparoscopic surgery.

The optical vessel detection instrument consists of a rotatable stick, an optical probe, and a wireless optical signal acquisition module. The optical probe contains two photodiodes (PDs; PD15–22C; EVERLIGHT, Taipei, Taiwan) and two dual-wavelength light-emitting diodes (LEDs; SMT700/910; Marubeni Corp., Tokyo, Japan). The PDs are used to detect the light signal reflected by human tissue (adipose or lymphatic tissue), whereas the LEDs act as dual-wavelength light sources. When the LED light passes through human tissue, optical attenuation is caused by scattering and absorption. For most components of human tissue, absorption in the near-infrared wavelength region (650–1000 nm) is very small, and oxyhemoglobin (HbO2) and deoxyhemoglobin (Hb) are two of the primary absorbers. Therefore, for near-infrared light, the optical attenuation change Inline graphic can be simply expressed as

A.

where Inline graphic and Inline graphic are the light intensity of the incident light and light reflected by the human tissue, respectively; Inline graphic is the distance between the light source and detector; and Inline graphic is the modified optical path factor [15] corresponding to the wavelength Inline graphic. The parameters Inline graphic and Inline graphic are the molar extinction coefficients of Inline graphic and Hb, respectively. The hemoglobin changes Inline graphic and Inline graphic can be estimated from the multiwavelength optical attenuation changes by using the least square approximation method:

A.

where

A.

Herein, 640- and 910-nm-wavelength light is used. Thus, the isosbestic point of the absorption spectra of HbO2 and Hb is approximately 800 nm. Finally, the change in the total hemoglobin (HbT) concentration ( Inline graphic) and tissue oxygen saturation (StO2) can be obtained from Inline graphic and Inline graphic:

A.

Because the penetration depth is approximately half the distance between the light source and detector [16], the optical probe, as shown in Fig. 2 (a), is designed to contain three optical pairs for monitoring blood vessels at different depths simultaneously. The distances between the LEDs and PDs are set as 10 mm (optical pair 1), 20 mm (optical pair 2), and 30 mm (optical pair 3). The optical probe is fixed on the rotatable stick to assist insertion into the body cavity through a trocar. The front end of the rotatable stick can be spun at various angles, thereby enabling the optical probe to easily navigate the curved surfaces of human tissue.

FIGURE 2.

FIGURE 2.

(a) Placement of optical probe, and (b) block diagram of the designed wireless optical signal acquisition module.

The wireless optical signal acquisition module, illustrated in Fig. 2 (b), contains a microprocessor unit, a wireless transmission module, a photodiode amplification module, and an LED driving module. The LED driving module is designed to drive the LEDs by using the control command of a microprocessor unit (RX210, Renesas Electronics Corp., Tokyo, Japan) to turn on and off the dual-wavelength light source. The light reflected by human tissue is transformed into a voltage signal by the PDs and amplified by the PD amplification module. Next, the received light signal is digitized using the analog-to-digital converter in the microprocessor unit with a sampling rate of 60 Hz and then sent to the wireless transmission module for transmission to the backend monitoring platform.

The backend monitoring platform is designed for the Windows 10 operating system. A real-time monitoring program was developed using Microsoft Visual C# for estimating, displaying, and storing hemoglobin parameters as well as detecting blood vessels.

B. Classification of Vascular and Avascular Locations

The radial-basis-function neural network (RBFNN) [17], [18] is used to classify vascular and avascular locations. The RBFNN is a nonlinear data modeling approach that the hidden neurons can provide the non-linear transform function for the difference between the input vector and the hidden center vector, and the output neuron is contributed by a linear weighted combination of these hidden neuron outputs. The center vectors in the hidden neurons can be viewed as the prototype characteristic vectors taken from the training sets Therefore, it offers the advantages of simple network structure, fast training procedure, and excellent approximation capability. The network structure contains three layers, as illustrated in Fig. 3: an input layer ( Inline graphic neurons), a hidden layer ( Inline graphic neurons), and an output layer (1 neuron). HbT and StO2 are used as the input of the RBFNN. The output Inline graphic of the RBFNN at the Inline graphic-th trial input can be expressed as

B.

where Inline graphic is the weight vector and Inline graphic denotes the weight between the Inline graphic-th hidden neuron and output neuron.

FIGURE 3.

FIGURE 3.

Basic structure of the RBFNN.

Inline graphic is the output vector of the hidden neurons. The Inline graphic-th neuron output Inline graphic can be calculated using the Gaussian basis function:

B.

where Inline graphic and Inline graphic denote the Inline graphic-th trial input vector and its standard deviation respectively, and Inline graphic is the center vector of the Inline graphic-th hidden neuron. The operator Inline graphic denotes the Euclid norm between two vectors. A self-organized learning procedure (k-means clustering [19], [20]) and the normalized least-mean-square algorithm [21] are used to train the center vector Inline graphic and weight vector Inline graphic, respectively. In the learning stage, the output for the vascular and avascular locations is 1 and 0, respectively. Therefore, if the output of the RBFNN is larger than a given threshold, the corresponding location is classified as vascular; otherwise, it is classified as avascular.

C. Design of the Animal Experiment

In this study, six duroc Chinese native piglets with an average weight of approximately 21 kg were used. The animal use protocol was reviewed and approved by the Institutional Animal Care and Use Committee (IACUC), Kaohsiung Medical University, Kaohsiung, Taiwan. The IACUC protocol number is 108038. Zoletil IM (10–25 mg/kg) was administered as an intramuscular injection for general anesthesia. The animals were ventilated, intubated, and then maintained with 1%–3% sevoflurane (inhalation anesthetic). The animal experiment was performed as part of the in vivo study. Fig. 4 shows the animal experiment wherein the abdomen of the piglet was penetrated using trocars (Lagis, Shielded Bladed Trocar, Taiwan, and ENDOPATH XCEL® Bladeless Trocars, NJ, USA). The proposed smart blood vessel detection system was inserted through the trocar and detected various parts of the body, such as inferior tissue veins, inferior tissue arteries, the liver, and muscles. The proposed system was employed to measure the optical and hemoglobin parameters in different tissue parts, and locations possibly containing blood vessels were determined. Analysis of variance was used to analyze the experimental data, with significance set as p < 0.05.

FIGURE 4.

FIGURE 4.

Setup of the animal experiment.

III. Experimental Results

A. Performance in Detecting Blood Vessels in a Phantom

The performance of the proposed system was first validated using a phantom experiment. Two tissue phantoms made of polydimethylsiloxane (PDMS; SC-812, Hisn Han International Trading, New Taipei City, Taiwan) and dye (GY-IR-500, Go Yen Chemical Group, Kaohsiung, Taiwan) were employed. In tissue phantom I (size: 105 mm Inline graphic 40 mm Inline graphic 30 mm), a plastic tube (inside diameter: 2 mm) was placed at a depth of 5 mm to simulate a blood vessel. In tissue phantom II (size: 115 mm Inline graphic 70 mm Inline graphic 30 mm), three plastic tubes were placed at depths of 5, 10, and 15 mm to simulate three blood vessels, as illustrated in Fig. 5(a). All of these plastic tubes were filled with red ink to simulate blood. The use of this device has to be operated by hand scanning. Therefore, we divided the area of these phantoms into several parts for scanning test. The minimal scanning partition is about 21 mm Inline graphic 8 mm. The experimental results, shown in Fig. 5 (b), indicated that the values of HbT and StO2 at the location above the plastic tube in tissue phantom I were higher than at other locations. The values of HbT and StO2 estimated by optical pair 1 at depths >5 mm in tissue phantom II were relatively high in Fig. 5(c). However, all HbT and StO2 values estimated by optical pair 3 at depths of 5, 10, and 15 mm were also relatively high.

FIGURE 5.

FIGURE 5.

(a) Two types of tissue phantom placement, and measurement results for (b) tissue phantom I and (c) tissue phantom II.

B. Optical and Hemoglobin Parameters of Different Tissues in the Vivo Experiment

The optical densities and hemoglobin parameters for different tissues were investigated in the animal experiment. Fig. 6 (a) and Fig. 6 (b) show the optical density changes in various tissues corresponding to the wavelengths 700 and 910 nm, respectively. At 700 nm, the change in optical density for veins was slightly smaller than that for arteries, but at 910 nm, the change in the optical density for veins was much smaller. For the 700-nm wavelength, the optical density change was greatest for the liver. The change in the optical density was smallest for muscle at both 700 and 910 nm. The differences in the changes in optical density at 700 nm (p = 0.000) and at 910 nm (p = 0.000) for different tissues were significant.

FIGURE 6.

FIGURE 6.

(a) Plot of 700- and (b) 910-nm-wavelength optical density changes, and (c) HbT and (d) StO2 values for various tissues.

Fig. 6 (c) and Fig. 6 (d) show the changes in HbT and StO2 for various tissues. The HbT values for arteries, veins, and the liver were higher than those for skin and muscle. However, for arteries, veins, and the liver, the StO2 value was higher for arteries than veins, and the StO2 value for the liver was lowest. The StO2 value for muscle was higher than that for skin or the liver, but the muscle HbT value was smaller than that for skin. The differences in HbT (p = 0.000) and StO2 (p = 0.000) between the various tissues were also significant.

C. Performance in Classification of Vascular Versus Avascular Locations

The performance of the RBFNN in classifying vascular and avascular locations was investigated using several binary test parameters. True-positive (TP) means that a vascular group was correctly identified; false-positive (FP), an avascular group was incorrectly identified as a vascular group; true-negative (TN), an avascular group was correctly identified; and false-negative (FN), a vascular group was incorrectly identified as an avascular group. The F-measure—the harmonic mean of recall (sensitivity) and precision (positive predictive value, PPV)—was used to determine the optimal threshold.

C.

In total, 360 data sets (number of vascular groups = 180; number of avascular groups = 180) collect from different tissue in the animal experiment were used for training. Before classification, the number of hidden neurons and optimal threshold had to be determined in the training stage. The number of hidden neurons was set as 32, 64, and 128, and the threshold was set as 0–1. The optimal performance (F-measure = 94.74%, sensitivity = 95%, PPV = 94.48%, and accuracy = 94.72%) was obtained when the number of hidden neurons and threshold were set as 64 and 0.6, respectively. In the blind test stage, 20 data sets (vascular location groups = 10; avascular location groups = 10) were used. Here, the number of hidden neurons and threshold were set as 64 and 0.6, respectively. The experimental results revealed that the proposed system achieved high performance in classifying the two groups (F-measure = 90%, sensitivity = 90%, PPV = 90%, and accuracy = 90%). Fig. 7 shows the RBFNN outputs for different groups; the RBFNN output for vascular groups (0.901 ± 0.149) was significantly higher than that for avascular groups (0.095 ± 0.216; p = 0.000).

FIGURE 7.

FIGURE 7.

RBFNN outputs for different groups.

Fig. 8 shows the detection results for different locations in the abdominal cavity in the experiment on piglets. An ultrasonic instrument (Flex focus 8001202, BK medical, Denmark) and a robotic drop-in transducer (Type8826, BK medical, Denmark) were used to obtain ultrasound images at the same locations. The HbT and StO2 values in the middle locations of the abdominal cavity for optical pairs 1–3 were higher than those at the side locations. Moreover, the RBFNN outputs for the middle locations (0.639 ± 0.059) were higher than those for the right side (0.468 ± 0.000) and left side (0.423 ± 0.021). In the ultrasound image, the color part represents the effluence or influx state of bodily fluid, indicating blood flow. Therefore, the middle location may contain blood vessels.

FIGURE 8.

FIGURE 8.

Ultrasound image and hemoglobin parameters of the proposed system in the vivo experiment.

IV. Discussions

As shown in Fig. 5 (b), the HbT and StO2 values at the locations above the plastic tube were higher than those at other locations, and the experimental results agreed favorably with the absorption spectra for a plastic tube containing red ink. The higher HbT and StO2 values at the 5-mm-deep simulated blood vessel, shown in Fig. 5 (c), were estimated by optical pair 1 of the optical probe. For optical pair 3 of the optical probe, all HbT and StO2 values at the simulated blood vessels at depths of 5, 10, and 15 mm were higher because the penetration depth of the light was approximately half the distance between the light source and detector. When the distance between the light source and detector is greater, the light must pass through the shallow layer to the deeper layer so that absorbance information can be obtained for different layers. The HbT and StO2 values for arteries were higher than those for veins (Fig. 6), indicating that the volume of oxygenated blood in arteries was higher than that in veins, which accurately reflects human physiology. Moreover, the changes in optical density for 700- and 910-nm-wavelength light were respectively slightly lower and much lower for veins than for arteries (Fig. 6). However, for 700- and 910-nm-wavelength light, the absorbance of deoxyhemoglobin was higher and lower, respectively, than that of oxyhemoglobin. Therefore, the change in optical density at a specific wavelength was caused by not only absorbance but also blood volume. The changes in the 910-nm-wavelength optical density for arteries and veins were greater than those for skin and muscle; this may have been due to the higher absorption coefficient of arteries and veins compared with those of skin and muscle [22], [23]. The StO2 value of the liver was much lower than that of other tissues. The liver contains oxyhemoglobin, deoxyhemoglobin, and bile [24]. Bile has a specific absorption spectrum [25] and may influence the estimation of StO2. The HbT and StO2 values at vascular locations can be distinguished from those at avascular locations because blood vessels contain more blood than other human tissues, such as skin and muscle.

Several approaches have been proposed for detecting blood vessels; their comparison is summarized in Table 1. The vein visualization device [7] is generally used for screening the skin’s surface. The device emits near-infrared light (wavelength: 785 nm) and receives the light reflected from the skin’s surface. However, this device is unsuitable for laparoscopic surgery due to its shallow measuring depth. In clinical settings, the Doppler ultrasound system can be used for detecting blood flow and may be employed in laparoscopic surgery [26]. Such systems determine the frequency difference between transmitted and reflected ultrasound signals and estimate the blood flow rate to recognize vascular locations. However, the tip direction of the ultrasound probe is difficult to control in body cavities, and the blood flow estimation is strongly affected by the angle and certain body tissue barriers such as bones. Optical coherence tomography (OCT) [27] can also be used to obtain high-resolution images of tissue structure. However, the OCT technique can only provide information on tissue structure; the identification of arteries and veins requires the expertise of physicians. Moreover, the measuring depth of the OCT system is relatively small (~1.5 mm). In contrast to these methods, the rotatable stick in the proposed system is easy for a physician to operate in a body cavity, and the angle of the optical probe is easily adjusted to navigate rough tissue surfaces. Moreover, the specific LED-PD placement of the optical probe helps in detecting blood vessels at different depths simultaneously. Using the difference between HbT and StO2 values of different tissues, vascular and avascular locations can be successfully distinguished using a neural network. Although the resolution of this system is lower than that of the aforementioned systems, the system offers a greater measuring depth (~1.5–2 cm).

TABLE 1. System Comparison Between Proposed System and Other Systems.

Vein visualization device [7] Doppler ultrasound [26] SD-OCT 5000 [27] Proposed system
Sensing technique Near-infrared spectroscopy Ultrasound Optical coherence tomography Near infrared spectroscopy
Sensor type CCD camera Ultrasound Probe OCT probe Optical probe
Transmission mode USB USB Bluetooth
System size (cm3) Inline graphic Inline graphic Inline graphic Inline graphic
Wavelength of light source (nm) 785 840 700, 910
Analysis parameters 2-D image 2-D image 3-D image Hemoglobin parameters and optical density
System complexity Low High High Low
Applications Vein visualization Laparoscopy surgery Eye examination Laparoscopy surgery
Advantages Detail information of blood vessel image, easy operation, and lower cost Deeper measuring depth Higher-resolution image Multi-depths measurement, Good distinguishability of tissue types, easy operation, and lower cost
Limitations Lack of multi-depths measurement, and shallower measuring depth Low distinguishability of tissue type, influence of measuring angle and bone, requirement of experienced operator, and higher cost Low distinguishability of tissue type, shallower measuring depth, requirement of experienced operator, and higher cost Lower spatial resolution

V. Conclusion

In this study, a smart blood vessel detection system for laparoscopic surgery was proposed. Using the differences in HbT and Inline graphic values between different tissues, vascular and avascular locations can be successfully recognized. Moreover, the specific LED-PD placement of the optical probe helps in detecting blood vessels at different depths simultaneously. The proposed system was validated using phantom and animal experiments. The changes in the optical density at specific wavelengths and the HbT and Inline graphic values of various tissues—the skin, veins, arteries, the liver, and muscle—were significantly different. The proposed system can effectively distinguish vascular and avascular locations by using a neural network. The proposed system offers the advantages of low cost, easy operation, excellent ability to recognize various types of tissue, and multiple measuring depths. Therefore, the proposed system can be regarded as a favorable system prototype for vessel detection in laparoscopic surgery and may assist physicians in preventing damage to blood vessels during surgery.

Funding Statement

This work was supported in part by the Ministry of Science and Technology, Taiwan, under Grant MOST 110-2221-E-A49-096-MY3, Grant MOST 109-2221-E-305-001-MY2, and Grant MOST 110-2314-B-305-001; in part by the Higher Education Sprout Project of the National Yang Ming Chiao Tung University and Ministry of Education (MOE), Taiwan, Research Funding Sponsorship of Kaohsiung Medical University, University System of Taipei Joint Research Program under Grant USTP-NTPU-NTOU-110-01; and in part by the Faculty Group Research Funding Sponsorship by National Taipei University under Grant 2021-NTPU-ORDA-02.

References

  • [1].Ponzini R.et al. , “Womersley number-based estimates of blood flow rate in Doppler analysis: In vivo validation by means of phase-contrast MRI,” IEEE Trans. Biomed. Eng., vol. 57, no. 7, pp. 1807–1815, Jul. 2010. [DOI] [PubMed] [Google Scholar]
  • [2].Maslove D. and Mihm F., “Ultrasound-guided internal jugular vein cannulation,” New England J. Med., vol. 363, no. 8, pp. 796–797, Apr. 2010. [DOI] [PubMed] [Google Scholar]
  • [3].Martins W. P., Raine-Fenning N. J., Ferriani R. A., and Nastri C. O., “Quantitative three-dimensional power Doppler angiography: A flow-free phantom experiment to evaluate the relationship between color gain, depth and signal artifact,” Ultrasound Obstetrics Gynecol., vol. 35, no. 3, pp. 361–368, Mar. 2010. [DOI] [PubMed] [Google Scholar]
  • [4].Wijnberger L. D. E., Bilardo C. M., Hecher K., Stigter R. H., and Visser G. H. A., “Effect of antenatal glucocorticoid therapy on arterial and venous blood flow velocity waveforms in severely growth-restricted fetuses,” Ultrasound Obstetrics Gynecol., vol. 23, no. 6, pp. 584–589, Jun. 2004. [DOI] [PubMed] [Google Scholar]
  • [5].Klein S. M., Fronheiser M. P., Reach J., Nielsen K. C., and Smith S. W., “Piezoelectric vibrating needle and catheter for enhancing ultrasound-guided peripheral nerve blocks,” Anesthesia Analgesia, vol. 105, no. 6, pp. 1858–1860, 2007. [DOI] [PubMed] [Google Scholar]
  • [6].Iida T., Mihara M., Yoshimatsu H., Hara H., Narushima M., and Koshima I., “Versatility of a near-infrared vein visualization device in plastic and reconstructive surgery,” Plastic Reconstructive Surg., vol. 130, no. 4, pp. 636–638, 2012. [DOI] [PubMed] [Google Scholar]
  • [7].Yaprak E. and Sibel K.-Y., “Preliminary evaluation of near-infrared vein visualization technology in the screening of palatal blood vessels,” Medicina Oral, Patologia Oral Cirugia Bucal, vol. 23, no. 1, pp. 98–104, 2018. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [8].Nizamoglu M., Tan A., Gerrish H., Barnes D., and Dziewulski P., “Infrared technology to improve efficacy of venous access in burns population,” Eur. J. Plastic Surg., vol. 39, no. 1, pp. 37–40, Feb. 2016. [Google Scholar]
  • [9].Preuss L. E. and Profio A. E., “Optical properties of mammalian tissue: Introduction by the feature editors,” Appl. Opt., vol. 28, no. 12, pp. 2207–2209, 1989. [DOI] [PubMed] [Google Scholar]
  • [10].Wilson R. H., Nadeau K. P., Jaworski F. B., Tromberg B. J., and Durkin A. J., “Review of short-wave infrared spectroscopy and imaging methods for biological tissue characterization,” Proc. SPIE, vol. 20, no. 3, 2015, Art. no. 030901. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [11].Lin K.-D., Lin B.-S., Lin G.-A., and Lin B.-S., “Design of smart peripheral blood perfusion monitoring system for diabetics,” IEEE Sensors J., vol. 21, no. 8, pp. 10167–10173, Apr. 2021. [Google Scholar]
  • [12].Chou W., Wu P.-J., Fang C.-C., Yen Y.-S., and Lin B.-S., “Design of smart brain oxygenation monitoring system for estimating cardiovascular disease severity,” IEEE Access, vol. 8, pp. 98422–98429, 2020. [Google Scholar]
  • [13].Huang Y.-K., Chang C.-C., Lin P.-X., and Lin B.-S., “Quantitative evaluation of rehabilitation effect on peripheral circulation of diabetic foot,” IEEE J. Biomed. Health Informat., vol. 22, no. 4, pp. 1019–1025, Jul. 2018. [DOI] [PubMed] [Google Scholar]
  • [14].Kuo J.-R., Lin B.-S., Cheng C.-L., and Chio C.-C., “Hypoxic-state estimation of brain cells by using wireless near-infrared spectroscopy,” IEEE J. Biomed. Health Informat., vol. 18, no. 1, pp. 167–173, Jan. 2014. [DOI] [PubMed] [Google Scholar]
  • [15].Boas D. A., Gaudette T., Strangman G., Cheng X., Marota J. J. A., and Mandeville J. B., “The accuracy of near infrared spectroscopy and imaging during focal changes in cerebral hemodynamics,” Neuroimage, vol. 13, no. 1, pp. 76–90, Jan. 2001. [DOI] [PubMed] [Google Scholar]
  • [16].Feng S., Zeng F. A., and Chance B., “Photon migration in the presence of a single defect: A perturbation analysis,” Appl. Opt., vol. 34, no. 19, pp. 3826–3837, 1995. [DOI] [PubMed] [Google Scholar]
  • [17].Elanayar V. T. S. and Shin Y. C., “Radial basis function neural network for approximation and estimation of nonlinear stochastic dynamic systems,” IEEE Trans. Neural Netw., vol. 5, no. 4, pp. 594–603, Jul. 1994. [DOI] [PubMed] [Google Scholar]
  • [18].Lin B. S., Lin B. S., Chong F. C., and Lai F., “Higher-order-statistics-based radial basis function networks for signal enhancement,” IEEE Trans. Neural Netw., vol. 18, no. 3, pp. 823–832, May 2007. [DOI] [PubMed] [Google Scholar]
  • [19].Bezdek J. C., “Objective function clustering,” in Pattern Recognition With Fuzzy Objective Function Algorithms. New York, NY, USA: Plenum Press, 1981, pp. 43–93. [Google Scholar]
  • [20].Chen S., Mulgrew B., and Grant P. M., “A clustering technique for digital communications channel equalization using radial basis function networks,” IEEE Trans. Neural Netw., vol. 4, no. 4, pp. 570–590, Jul. 1993. [DOI] [PubMed] [Google Scholar]
  • [21].Liu W., Pokharel P. P., and Principe J. C., “The kernel least-mean-square algorithm,” IEEE Trans. Signal Process., vol. 56, no. 2, pp. 543–554, Feb. 2008. [Google Scholar]
  • [22].Garcia J. V., Zhang F., and Ford P. C., “Multi-photon excitation in uncaging the small molecule bioregulator nitric oxide,” Philos. Trans. Roy. Soc. A, Math., Phys. Eng. Sci., vol. 371, no. 1995, 2013, Art. no. 20120129. [DOI] [PubMed] [Google Scholar]
  • [23].Shiga T., Lin L., Takahashi M., Yamamoto K., Niwayama M., and Kudo N., “Influence of a fat layer on muscle oxygenation measurement using near-IR spectroscopy: Quantitative analysis based on two-layered phantom experiments and Monte Carlo simulation,” Frontiers Med. Biol. Eng., vol. 10, no. 1, pp. 43–58, 2000. [DOI] [PubMed] [Google Scholar]
  • [24].Carneiro I., Carvalho S., Henrique R., Oliveira L., and Tuchin V. V., “Measuring optical properties of human liver between 400 and 1000nm,” Quantum Electron., vol. 49, no. 1, pp. 13–19, Jan. 2019. [Google Scholar]
  • [25].Nachabé R.et al. , “Effect of bile absorption coefficients on the estimation of liver tissue optical properties and related implications in discriminating healthy and tumorous samples,” Biomed. Opt. Exp., vol. 2, no. 3, pp. 600–614, 2011. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [26].Perlmutter M. A., Hyams E. S., and Stifelman M. D., “Laparoscopic Doppler technology in laparoscopic renal surgery,” J. Soc. Laparoendoscopic Surgeons, vol. 13, no. 3, pp. 406–410, 2009. [PMC free article] [PubMed] [Google Scholar]
  • [27].Kashani A. H., Lee S. Y., Moshfeghi A., Durbin M. K., and Puliafito C. A., “Optical coherence tomography angiography of retinal venous occlusion,” Retina, vol. 35, no. 11, pp. 2323–2331, 2015. [DOI] [PubMed] [Google Scholar]