Abstract
Ultrasound localization microscopy (ULM) demonstrates great potential for visualization of tissue microvasculature at depth with high spatial resolution. The success of ULM heavily depends on robust localization of isolated microbubbles (MBs), which can be challenging in vivo especially within larger vessels where MBs can overlap and cluster close together. While MB dilution alleviates the issue of MB overlap to a certain extent, it drastically increases the data acquisition time needed for MBs to populate the microvasculature, which is already on the order of several minutes using recommended MB concentrations. Inspired by optical super-resolution imaging based on stimulated emission depletion (STED), here we propose a novel ULM imaging sequence based on MB uncoupling via transmit excitation (MUTE). MUTE “silences” MB signals by creating acoustic nulls to facilitate MB separation, which leads to robust localization of MBs especially under high concentrations. The efficiency of localization accomplished via the proposed technique was first evaluated in simulation studies with conventional ULM as a benchmark. Then, an in-vivo study based on the chorioallantoic membrane (CAM) of chicken embryos showed that MUTE could reduce the data acquisition time by half, thanks to the enhanced MB separation and localization. Finally, the performance of MUTE was validated in an in vivo mouse brain study. These results demonstrate the high MB localization efficacy of MUTE-ULM, which contributes to a reduced data acquisition time and improved temporal resolution for ULM.
Keywords: Microbubble (MB) separation, MB uncoupling, null excitation, super-resolution imaging, ultrasound localization microscopy (ULM)
I. INTRODUCTION
SUPER-RESOLUTION ultrasound localization microscopy (ULM) is an acoustic analog to optical super-resolution microscopy such as photoactivated localization microscopy (PALM) [1] or stochastic optical reconstruction microscopy (STORM) [2]. It was recently introduced for microvasculature imaging beyond the inherent acoustic spatial resolution limit while conserving the imaging depth of conventional ultrasound. The primary idea of ULM is to localize microbubbles (MBs) flowing in the vascular networks to achieve super-resolution, and then track the localized MBs over time to measure blood flow velocity [3]. ULM improves the conventional ultrasound spatial resolution by approximately tenfold and it showed promising results in various tissues including brain [4], kidney [5], liver [6], and tumor [7]. However, practical implementation of ULM is currently limited by its Achilles’ heel—slow temporal resolution, which is largely attributed to the long data acquisition time that is required to capture adequate, spatially separated MB signals for localization and tracking. Although increasing MB concentration alleviates slow MB perfusion and shortens data acquisition time, it causes more MB signals to overlap and become unlocalizable. As such, developing a method that is highly efficient at localizing MBs under high concentration is essential to accelerate ULM.
To this end, several techniques have been demonstrated to improve the super-resolution imaging speed, including super-resolution optical fluctuation imaging (SOFI)-based contrast-enhanced ultrasound (CEUS) imaging [8], and sparsity-based ultrasound super-resolution hemodynamic imaging (SUSHI) [9]. SOFI-based CEUS imaging use the high-order statistics of the fluctuating MBs’ signal. In contrast, SUSHI exploits the sparsity model by assuming the underlying vasculature consisting of point targets on higher pixel resolution. These techniques showed improvement of temporal resolution while providing a spatial resolution that is similar to ULM. However, because of the absence of MB tracking, these methods fall short of providing blood flow velocity measurements, which can be an important biomarker for various applications [4], [10], [11].
Recently, our group introduced an MB separation method that separates high-concentration MBs in the Fourier domain to facilitate localization of MBs under high concentration [10]. Thanks to the enhanced localization efficacy, MB separation leads to a reduced data acquisition time for ULM. However, by creating subsets of data with different MB flow characteristics, MB separation becomes computationally expensive. In addition, the performance of MB separation may be undermined by complex flow dynamics with high MB concentrations. Compressed sensing-based localization techniques also showed the capability of reducing data acquisition time for ULM under MB high concentration, but it has a similar issue of high computational cost as the MB separation method and it also does not provide blood flow velocity measurement [12].
Inspired by the optical super-resolution technique stimulated emission depletion (STED), in this article, we propose to implement a similar principle of contrast signal suppression in ultrasound imaging to promote MB separation and enhance ULM imaging performance. STED uses two transmissions including a Gaussian-shaped beam followed by a doughnut-shaped beam that depletes the fluorescence excited by the Gaussian beam. The partial depletion preserves a sharp focal spot in the donut center, where a small number of fluorophores are allowed to emit signals that are distinct from the surrounding fluorophores that have been depleted [13]. The distinction between the fluorophores inside and outside the donut center creates separation in fluorescence status in time, which manifests as super-resolution when imaging.
While the principle of STED is straightforward, translating it to ultrasound imaging is not because in ultrasound one needs to overcome the challenge of depleting MB signals, which is not as simple as in optics. For example, to deplete MB signal, one needs to apply high mechanical index (MI) acoustic pulses to disrupt MBs, which is not an efficient process and may cause tissue and probe heating. Also, unlike in optical STED where a fluorophore can be instantly and completely switched to the “OFF” or ground status by the stimulation photon, MB signal emission is less binary (e.g., either “ON” or “OFF”) and even disrupted MBs can emit strong acoustic signals [14]. As such, an alternative approach is necessary for ultrasound imaging to implement the principles of STED.
In this article, we introduce a method named MUTE (MB uncoupling via transmit excitation) that uses the principles of acoustic null generation and subtraction imaging to realize STED in ultrasound. MUTE accomplishes MB depletion in two steps: 1) obtain two images by first transmitting a regular plane wave (PW) without acoustic nulls and then another PW with acoustic nulls and 2) calculate the difference between the two images, and then MBs in the nonnull region will be effectively depleted and MBs in the null region will be enhanced and separated from the ones in the non-null region. Although similar ideas of using acoustic nulls to enhance ultrasound imaging resolution have been reported before, they were implemented on receive apodization which does not actively modulate the acoustic response of MBs by transmit excitation [15], [16]. In addition, these methods suffer from a low signal-to-noise ratio (SNR) due to the significantly reduced signal intensity induced by the processes of Heaviside apodization and subtraction [17]. With MUTE, one can create a similar ON/OFF status of MBs as in STED, which provides the opportunity to spatially separate MBs that are otherwise inseparable under traditional insonification. By promoting MB separation, MUTE facilitates robust MB localization for ULM, especially under high MB concentrations. The improved MB localization efficacy leads to an overall reduced data acquisition time for ULM.
The rest of this article is structured as follows: MUTE imaging sequence, postprocessing steps, and preparation of models for evaluation including simulation, chorioallantoic membrane (CAM) of a chicken embryo, and a mouse brain are presented in Section II. In Section III, we will demonstrate the improvement of MUTE-based ULM imaging with conventional ULM as reference. Finally, our results are discussed, and conclusions are drawn in Section IV.
II. MATERIALS AND METHODS
A. Principles of Microbubble Uncoupling via Transmit Excitation
Fig. 1(a) shows the conceptual schematic of MB uncoupling in tissue vasculature based on MUTE. Flowing MBs in vessels are activated or silenced depending on their spatial location with respect to the acoustic null. By calculating the difference between two consecutive images (i.e., an image with the acoustic null and the other without), MBs in the non-null region will be suppressed, and MBs in the null region will be enhanced and subsequently uncoupled from the surrounding MBs. In general, images with acoustic nulls can be generated by apodizing the transmit aperture with functions such as Heaviside [15], for example,
(1) |
where ATx,i is the transmit apodization weight for the ith element, and N is the total number of active elements in the transmit aperture. Fig. 1(b) shows the beam pattern of PW (left) and MUTE (right) acquired by adjusting the transmit apodization weight. The left and middle panels of Fig. 1(c) show B-mode images of three MBs spaced with 1.28λ (98.56 μm) distance, based on PW transmission and MUTE transmission, from which uncoupled MB signals can be obtained by calculating the difference between the two images. λ is the wavelength that was calculated using the transmit frequency of 20 MHz. To minimize image distortion by subtraction, an intensity weighted subtraction algorithm was adopted for the MUTE sequence [18]
(2) |
(3) |
where IPW and IMUTE are the intensity images (i.e., B-mode) obtained by PW and MUTE, respectively. εmax and εmin are the upper and lower bounds of the intensity values of the uncoupled MB image, respectively, in which we set −1 and +1, as the minimum and maximum intensity values. The bottom row in Fig. 1(c) shows the normalized intensity profiles in the lateral direction corresponding to each B-mode image. One can observe that the center MB in the acoustic null that was not distinguishable from the surrounding MBs now becomes distinct and localizable for ULM imaging.
B. MUTE Imaging Sequence and Data Acquisition
In practice, because a single acoustic null only uncouples MBs in a small-limited region, it becomes challenging to implement MUTE with an adequate frame rate for ULM imaging (e.g., one needs to translate the null to different lateral locations in different pulse-echo cycles). Also importantly, since a single PW does not provide transmit focusing and adequate imaging SNR, it is necessary to integrate spatial angular compounding into the MUTE sequence for practical use [19]. To address these practical issues, as shown in Fig. 2, we designed a pulse train of Heaviside functions for transmit apodization, which was then steered with various angles to realize spatial compounding by adjusting the transmit delay for each element. In detail, the MUTE beam includes a total number of 31 nulls with equal distances of 3.58λ (275.7 μm) in the lateral direction. The width of null is 0.45λ (34.6 μm). Furthermore, to cover the entire imaging field by nulls, the MUTE beam was translated laterally with a step size of 0.89λ (68.5 μm) corresponding to the pitch size of the transducer we used, resulting in the use of four MUTE beams at each steering angle. To minimize overlap of non-null regions between different steering angles while still providing ample angular compounding, a step steering angle of 2° was chosen throughout the study. As shown in Fig. 2, upon spatial compounding of each MUTE sequence, the acoustic nulls were preserved while the non-null regions were compounded with enhanced transmit focusing and SNR.
For the acquisition of ultrasound in-phase and quadrature (IQ) data, we used an ultrasound imaging system (Vantage 256, Verasonics Inc., Kirkland, WA, USA) with a high-frequency linear array transducer (L35–16vx, Verasonics Inc., Kirkland, WA, USA). The receiving demodulation frequency was set as 31.25 MHz. The transmit frequency of 20 MHz with a single sinusoidal wave and a transmit voltage of 10 V were used. Furthermore, ultrasound imaging was performed using three steering angles with a step steering degree of 2 (−2°, 0°, +2°). PW and four MUTE beams were sequentially generated at each steering angle, and coherent compounding was performed separately to each IQ sub-datasets. The postcompounded effective frame rate was 500 Hz. The total number of 5000 IQ datasets including each 1000 IQ subdatasets for PW and four MUTE sequences were stored per one acquisition, which corresponds to the data acquisition time of 2 s.
To measure the beam pattern and MI generated by the MUTE sequence, a needle hydrophone (1603, Precision Acoustic Ltd., Dorset, U.K.) with a customized acoustic intensity measurement system was used.
C. Postprocessing Steps for the Reconstruction of MUTE-ULM Image
For the reconstruction of MUTE-ULM image, the postprocessing steps are as shown in Fig. 3. A spatiotemporal singular value decomposition (SVD) filter was used to extract the MB signal from the surrounding tissue [20], [21]. A noise equalization process was applied throughout the imaging field of view to facilitate robust thresholding based on MB signal intensity [22]. MB separation was used in this study with three filter bands of 15– 50, 50–100, and 100–200 Hz [23]. After MB separation, a spatiotemporal nonlocal means (NLM) filter was used to further reduce noise that may result in false MB localizations [24]. We then performed MUTE image subtraction based on (2) followed by MB localization and tracking. For MB localization, MB images were spatially interpolated to 0.06λ (4.9 μm) pixel size. Two point-spread functions (PSFs) were then constructed for PW and MUTE sequences, respectively, using a 2-D multivariate Gaussian model
(4) |
where (cz, cx) represents the center location of MB. σz and σx were calculated in the axial and lateral directions, respectively, as follows:
(5) |
where the full-width at half-maximum (FWHM) of MBs was measured by averaging 10 random MB signals obtained at different imaging depths and frames. In this study, the average FWHM of MBs via PW in the lateral and axial directions was measured to be 138 and 86 μm, respectively; and 120 and 86 μm for MUTE. Note, that the null width of MUTE beam was measured to be ~50 μm.
After constructing appropriate PSFs, a similar process of MB localization and tracking as in our recent papers [24], [25] was performed.
D. Simulation Study
To evaluate the efficiency and error of localization depending on the concentration of MBs, we produced digital phantoms with randomly distributed MBs at concentrations of 5, 25, 50, 100, 125, and 150 MBs/mm2 using the Verasonics simulator. Furthermore, five phantoms were generated at each concentration to measure the average performance with standard deviations. The same MUTE imaging sequence as described in Fig. 2 was used in the simulation study. Besides, the same postprocessing steps were applied except for spatiotemporal dimension filters. The localized MBs were paired within 1λ (77 μm) distance with the true MBs generated in simulation using a bipartite graph-based pairing algorithm [24]. Then, the localization efficiency was measured by counting the number of pairings at each concentration. Simultaneously, the localization error was calculated by measuring the distance between paired MBs.
E. Chicken Embryo Chorioallantoic Membrane (CAM) Study
In this section, we used CAM microvessel with optical imaging ground truth to test the imaging performance of MUTE. Fertilized chicken eggs (white leghorn) were obtained from the University of Illinois Poultry Research Farm and stored in a 37 °C egg incubator (Digital Sportsman Cabinet Incubator 1502, GQF Manufacturing Inc., GA, USA). On the fourth day of embryonic development (EDD-04), the eggshells were peeled using a rotary Dremel tool with a cutting wheel, and embryos were transferred into weigh boats. Shell-less embryos were stored in a humidified incubator (HH09, Darwin Chambers Company, MO, USA) until the 16th day of embryonic development (EDD-16) to perform ultrasound imaging.
Food and Drug Administration (FDA)-approved lipid shell MBs (Definity®, Lantheus Medical Imaging Inc., N. Billerica, MA, USA) were used throughout in vivo imaging experiment. The final solution of MBs had a concentration of 1.2 × 1010 MBs/mL [26]. For MB injection, a high-order vein on the CAM was cannulated using a glass capillary needle to allow MBs’ bolus injection of 70 μL (8.4 × 108 MBs) intravenously. Then, the ultrasound transducer was placed at the side of the weigh boat. An optical image of the CAM vessels was acquired using a stereomicroscope (SMZ800N, Nikon Co., Tokyo, Japan) to provide the reference ground truth for imaging performance evaluations. For reconstruction of the ULM image of CAM, a total of 20 IQ datasets corresponding to 20 000 frames were acquired based on the ultrasound imaging setup we described in previous sections.
To measure the ULM imaging performance between PW and MUTE sequences, a normalized cross correlation (NCC) was computed between ULM image and optical image and used as a quantitative measure of similarity. To accomplish this, the optical image was first resized according to the dimension of the ULM image and then converted to grayscale. Then, a target patch with a size of 98 μm × 98 μm (20 pixels × 20 pixels) was selected from the optical image as the reference. A smaller patch (24.5 μm × 24.5 μm or 5 pixels × 5 pixels) was then selected from ULM image for NCC calculation This process was repeated for the entire imaging FOV (patch translation step size = 4.9 μm or 1 pixel), and the average NCC coefficients was used as the quantitative measurement of ULM imaging performance. In addition to similarity, we measured the vessel filling (VF) percentage for each imaging sequence. VF is defined as:
(6) |
where NGT denotes the total number of true vessel pixels, and NULM is the total number of pixels that have been correctly filled by ULM using MB locations. To obtain NGT, Otsu’s thresholding was applied to grayscale optical image followed by image dilation and erosion to extract vascular morphometric information.
F. In Vivo Mouse Brain Study
All experimental procedures on the mouse presented in this article were approved by the IACUC at the University of Illinois Urbana-Champaign (Protocol number: 19063). Anesthesia was induced using 4% isoflurane mixed with medical oxygen in a gas induction chamber. The mouse was then transferred to a nose cone supplying 2% isoflurane with oxygen for maintenance, and its head was secured to a stereotaxic imaging frame with ear bars. The scalp was removed with surgical scissors, and a cranial window on the left side of the skull was opened with a rotary Dremel tool to expose a hemisphere of the brain. The ultrasound transducer was fixed to a customized stereotaxic imaging stage and positioned above the cranial window to image the brain with a coronal imaging plane. A 30-gauge catheter was then inserted into the tail vein of the mouse for 50-μL MB injection (6 × 108 MBs). Additional 50-μL booster injections were provided throughout the imaging session to maintain MB concentration in the cerebral vasculature. For reconstruction of the ULM image of the mouse brain, a total of 40 IQ datasets corresponding to 40 000 frames were obtained based on the ultrasound imaging setup we described in previous sections during two bolus injections.
III. RESULTS
A. Measurement of the Beam Pattern and Mechanical Index of MUTE Sequence
We measured the acoustic null width, shift, and distance between the nulls of a MUTE beam using a needle hydrophone. The MUTE beam was generated with a transmit frequency of 20 MHz and a transmit voltage of 10 V and at a steering angle of 0°. Fig. 4(a) shows the 2-D MI image of the 1st MUTE beam. The needle hydrophone measurement confirmed the specifications of the designed MUTE beam: 31 null positions with an inter distance of 3.64λ (280 μm) in the lateral direction. The maximum MI was measured to be 0.14. Fig. 4(b) represents the lateral MI profiles acquired from the first (Black) and second (blue) MUTE beams at 5 mm in depth indicated by the white dashed line in Fig. 4(a). The width of null was measured at the MI of 0.04 indicated by a gray dashed line in Fig. 4(b). The average width of nulls was measured to be around 0.65λ (50 μm). Furthermore, it was confirmed that null locations of the MUTE beam could be laterally moved 0.91λ (70 μm) [Fig. 4(b)].
B. Evaluation of MUTE-Based Localization Efficacy in Simulation
We evaluated the performance of PW and MUTE-based localization using different concentrations of simulated MB signals. Fig. 5(a) and (b) shows the representative localization results based on each dataset obtained by PW and MUTE imaging sequences at a concentration of 50 MB/mm2. The MUTE-based localization had a higher MB detection rate than PW-based localization, especially for the overlapped MBs indicated by yellow arrows in Fig. 5(a) and (b). The average number of MB localizations was 26.4 ± 1.96 MB/mm2 for PW and 41.4 ± 1.74 MB/mm2 for MUTE at a concentration of 50 MB/mm2 [Fig. 5(c)]. The MUTE-based localization had higher efficacy with a similar localization error across all the MB concentrations tested [Fig. 5(c) and (d)]. In particular, as shown in Fig. 5(c), MUTE could localize over 80% of MBs up to 50 MB/mm2 concentration, whereas PW fell short of reaching 80% beyond 25 MB/mm2. Moreover, MUTE was largely capable of keeping up with the elevated MB concentration, while PW plateaued beyond 50 MB/mm2 concentration [Fig. 5(c)]. At the maximum concentration of 150 MB/mm2, MUTE was still able to localize over 50% of the MBs (81.8 MB/mm2). The maximum localization error for both PW- and MUTE-based localization was below 24.64 μm, which corresponds approximately to 0.32λ of the imaging frequency [Fig. 5(d)].
C. MUTE-ULM Imaging of CAM and Quantitative Analysis With Optical Microscopy
The optical image of the CAM in grayscale is shown in Fig. 6(a), and the corresponding binary image generated by Otsu’s thresholding is shown in Fig. 6(b). PW- and MUTE-ULM images are shown in Fig. 6(c) and (d), respectively. These images were reconstructed using 20 000 frames of ultrasound data corresponding to a total acquisition time of 40 s. Fig. 6(e) shows the NCC coefficient between ULM and the optical ground truth versus data acquisition time. It can be observed from Fig. 6(e) that MUTE-ULM generated images with greater similarity to the optical ground truth than did PW-ULM. The maximum NCC coefficient of PW-ULM was 0.59 at an acquisition time of 40 s, while MUTE-ULM achieved the same level of similarity with approximately half the time (22 s). Fig. 6(f) shows the VF percentage of the two techniques, and MUTE-ULM again outperformed PW-ULM across all data acquisition times. MUTE-ULM filled the vessels about twice as fast as did PW-ULM (i.e., 20 s versus 40 s) and was able to reach a higher VF percentage than PW-ULM (85.5% versus 79.9%). Furthermore, it was observed that the maximum NCC coefficient (i.e., a measure of similarities between the optical and ULM images of CAM) was improved from 0.51 to 0.55 from PW to MUTE without bubble separation, and from 0.51 to 0.59 from PW to PW with bubble separation. The VF percentage was enhanced from 67.6% to 72.4% from PW to MUTE without bubble separation, and from 67.6% to 79.9% from PW to PW with bubble separation [Fig. 6(e) and (f)].
Fig. 7 shows the local analysis results of ULM images in the yellow box in Fig. 6. As indicated by the different colored arrows in Fig. 7(a) and (b), MUTE-ULM demonstrated faster VF speed and higher overall MB count [Fig. 7(c)] than PW-ULM. MUTE-ULM could detect small vessel signals starting from 2 s, while PW-ULM could not detect the small vessels properly at the same acquisition time. As indicated by yellow arrows in Fig. 7(c), the second vessel was only detectable with MUTE-ULM and remained missing with PW-ULM until 40 s. Moreover, the fourth vessel indicated by the blue arrow was detected by MUTE-ULM at t = 10 s [Fig. 7(c)], detected the same vessel at t = 20 s. Furthermore, we compared the diameter of the four vessels that were measured at 40 s as labeled in Fig. 7(c). Based on MUTE-ULM, the diameter of vessels was measured to be 40.6, 29.7, 52, and 32 μm, respectively, for vessels 1–4, while the diameter of vessels was 39.23, 0 (missing), 50, and 30.77 from PW-ULM. The diameters measured from the optical ground-truth image were 36.8, 25.6, 42.7, and 28.8 μm. The average error in diameter measurement was 5.1 μm for MUTE-ULM and 9.33 μm for PW-ULM.
D. MUTE-ULM Imaging of the Mouse Brain in Vivo
Fig. 8 shows the ULM imaging results using MUTE and conventional PW on a mouse brain. A total of 40 000 frames (80 s) of ultrasound data were acquired during the experiment. We also carried out the local analysis at the regions indicated by the white box in Fig. 8(a). From Fig. 8, both PW- and MUTE-ULM were able to produce robust, super-resolved brain microvasculature maps throughout the full depth of the brain. MUTE-ULM showed higher intensity microvessel density maps that indicate higher MB count, which is most conspicuous in regions with smaller vessels between large vessels. Similar to the observations from CAM study, MUTE-ULM showed a faster VF speed than PW-ULM, as indicated by the yellow arrows in Fig. 8(c) and (d). The intensity profiles of vessels through the white dashed line indicated in Fig. 8(d) demonstrate that MUTE-ULM detected a higher MB count than PW-ULM, which is again consistent with the results in the CAM study. These results, thus, demonstrate that the proposed technique has the potential to shorten the data acquisition time for in vivo applications.
The velocity maps of mouse brain reconstructed via PW- and MUTE-ULM using entire datasets (80 s) are shown in Fig. 9(a) and (b). MUTE-ULM does not affect the measurement of the velocity map while assuring better detectability for the localization of high-concentration MBs as denoted by white arrows in Fig. 9(a) and (b).
IV. DISCUSSION
In this article, we introduced an acoustic analog to STED for super-resolution ultrasound microvascular imaging. MUTE enhances the performance of ULM by uncoupling overlapping MB signals in space based on acoustic nulls and image subtraction. We demonstrated that MUTE significantly improved the performance of ULM, especially under high MB concentrations. This has strong implications for the practical implementation of ULM because a higher MB concentration leads to a faster vascular filling time, which then translates to shorter data acquisition and higher temporal resolution for ULM.
Unlike in STED where optical fluorophores can be conveniently turned on or off by different wavelength emissions, MBs have a less binary response to acoustic waves and are difficult to transition between the ON and OFF status. Although high MI acoustic pulses can disrupt MBs and switch them OFF, the disruption process takes a long time and even the disrupted MBs (i.e., free gas) can be echogenic. Therefore, in this article, we resorted to using acoustic nulls and the principles of image subtraction to realize STED for ultrasound super-resolution imaging. Although the concept of using acoustic nulls to enhance ultrasound imaging resolution was reported before (e.g., NSI), these methods exclusively operated in the receive beamforming process and did not actively modulate MB response with the acoustic nulls.
In the simulation study, the nonuniform distribution at both sides of the null was observed when the full transducer aperture was used for null creation (i.e., transmit apodization: +1 for the left 64 elements and −1 for the right 64 elements), which induced high-level sidelobes [e.g., blue dashed line of the intensity profile in Fig. 1(c)]. However, as shown in Fig. 1(c), the side lobes could be effectively suppressed (e.g., red line of the intensity profile) using the intensity weighted subtraction algorithm [(2) and (4)]. Furthermore, the side lobes became less pronounced when using the sub-aperture technique (e.g., as shown in Fig. 4 where subapertures of eight elements were used for null generation), which was confirmed by hydrophone measurements.
We first evaluated the performance of MUTE in silico with randomly distributed MBs. We found that MUTE-based localization was highly effective with uncoupling the spatially overlapped MBs that are otherwise inseparable and unlocalizable by conventional ULM. MUTE also contributes to a more accurate MB localization when compared with conventional ULM [Fig. 5]. We then tested the performance of MUTE in vivo on a chicken embryo CAM model [Figs. 6 and 7] and a mouse brain [Figs. 8 and 9] with high concentration MB injections. In both the models, MUTE-ULM demonstrated significantly faster VF time and higher MB count than conventional ULM. These results suggest that MUTE can effectively mitigate the issue of slow imaging speed of ULM by promoting a more robust MB localization with high efficacy under high MB concentrations, which translates to a faster vascular MB filling rate and an overall reduced data acquisition time.
The lower precision of localization via the wrong system PSF could lead to an over-estimation of the diameter of the vessel (or blurring) in ULM image. In this study, we randomly selected 10 uncoupled MB images to derive a PSF to be used for MB localization as described in Section II–C. For example, five representative MBs after uncoupling by MUTE are shown in Fig. 10(a). Fig. 10(b) shows the averaged MB signal, which does resemble a 2-D Gaussian function [further confirmed by the 1-D axial and lateral profile plots in Fig. 10(c)]. One possible reason for the slightly increased localization error for MUTE is the distortion of the MB signal (i.e., from null subtraction) that results in a slight shift of the detected MB location from the true location. Nevertheless, the error was only slightly higher [e.g., 0.017λ (1.3 μm) in Fig. 5(d)] than conventional ULM and the maximum error was under 0.32λ (24.64 μm), which still assures the capability of super-resolution imaging over the diffraction limit.
The computational cost of MUTE-ULM including beamforming and post-processing is indeed fivefold more than PW-ULM. Beyond the postprocessing cost limitation, the low frame rate caused by the null transmission sequences is a limitation of MUTE-ULM in comparison to PW-ULM. One can acquire more PW data with a higher frame rate to accumulate more MB signals to improve PW-ULM performance. Besides, a higher frame rate is also helpful for better tracking results of MBs. However, the capability of localizing MBs in vessels with high MB concentrations does not necessarily improve with increased imaging frame rate or increased MB data. This is particularly true in larger vessels with high MB concentrations. As shown in Fig. 11(a), MUTE had much better performance in MB localization in the two parallel major vessels than PW-based ULM. Moreover, we plotted representative localization results using 200, 400, 600, 800, and 1000 frames for both MUTE-ULM and PW-ULM [Fig. 11(b)]. We examined the VF percentage at large vessels, which clearly shows that MUTE-ULM had a higher VF rate than conventional PW-ULM [Fig. 11(c)]. Combining these results with the observation of higher MB localization efficacy, the gain in ULM imaging performance from MUTE is primarily a result of enhanced MB localization instead of an acquired larger dataset.
Fig. 12 shows the histogram with a normal distribution fit of α values acquired under simulation, CAM, and mouse experiments. The mean and standard deviation of α were 0.49 ± 0.07, 0.51 ± 0.04, and 0.5 ± 0.04 for simulation, CAM, and mouse, respectively. Therefore, the value of α was fairly consistent in our study.
The filtering-based bubble separation is an additional postprocessing step to enhance the localization performance for situations with high-concentration MBs. MUTE is not postprocessing-based and directly uncouples MBs in a high MB density environment using null transmission. We used three filter banks to balance the tradeoff between MB separation performance and added computational costs associated with MB separation. As shown in Fig. 6(e) and (f), we elected to combine MB separation with MUTE because we found that the gain in MB localization performance was substantial with MB separation. For example, the maximum NCC coefficient was improved from 0.51 to 0.55 from PW to MUTE without bubble separation, and from 0.51 to 0.59 from PW to PW with bubble separation. Nevertheless, it is also clear that when combining MB separation with MUTE, ULM performance was the best.
From the in vivo CAM and mouse brain imaging results, we did not find MUTE-ULM to be particularly more susceptible to tissue motion than conventional PW-ULM. In this study, MUTE-ULM used five transmissions at each steering angle with a pulse repetition frequency of 22.7 kHz and achieved an effective frame rate of 500 Hz, which is similar to most ULM applications with compounding PW imaging as the detection sequence. However, it is unclear whether tissue motion would compromise the MB uncoupling performance (e.g., the null subtraction steps) for MUTE. We speculate that MB movement within the bloodstream (especially for large vessels with high flow speed) may be a more significant source of motion than tissue, but again it is unclear to what extent would it affect the MUTE performance within the context of MB uncoupling. For compounding PW, it is understood that such motion may decrease the coherence of MB signals from each transmission angle, which leads to suboptimal transmit focusing synthetization that deteriorates the MB signal (e.g., blurring). We expect a similar issue with MUTE but again based on the in vivo study results, we do not think MUTE is more susceptible to tissue or MB motion than conventional PW imaging.
For practical implementations of MUTE, there are several tradeoffs that need to be balanced. First, the number of nulls that can be simultaneously generated has a positive relationship with the imaging frame rate: the more the number of nulls that can be simultaneously generated in the FOV to cover a large region, the less the number of transmissions needed for constructing the full FOV, and hence the higher the imaging frame rate. In this article, we used eight elements to generate each null (31 nulls total for the 128-element transducer). The resulting image frame rate was 500 Hz (after three angle compounding and four lateral translations to cover the full FOV), which is adequate for most ULM applications. The second tradeoff is between the width and the number of the acoustic nulls and their relationship to the imaging frame rate. In theory, a larger number of elements could generate narrower nulls with lower acoustic pressure. However, the more the elements used for each null, the less the number of nulls that can be generated simultaneously, and consequently the slower the frame rate. In this study, the width of the acoustic nulls was measured at approximately 0.65λ (50 μm). Therefore, MBs with spacing less than 0.65λ (50 μm) may not get uncoupled effectively. Finally, one needs to note that the effect of uncoupling by MUTE is more significant in the lateral direction than in the axial direction. Although steering may introduce uncoupling in the axial direction, due to the limited steering angle that is achievable on conventional ultrasound transducers, MUTE still largely promotes MB separation along the lateral direction. Nevertheless, because the lateral resolution is intrinsically inferior to axial resolution in ultrasound imaging, improvement in the lateral direction is necessary and significant for robust ULM.
In contrast to a recent study using activatable phase-changing nanodroplets for super-resolution imaging (see [27]), MUTE “silences” MBs to enhance the imaging performance, while the nanodroplets-based method “evokes” MBs. Unlike optical contrast such as fluorophores, MBs cannot be switched between ON and OFF status repeatedly. For example, once a nanodroplet gets activated to become an MB (i.e., switch ON), one can only switch them OFF by disruption (e.g., using high MI acoustic pulses), after which it is impossible to switch them back ON again. However, one advantage of using phase-changing nanodroplets over conventional MBs is that one can selectively activate nanodroplets in a controlled-release fashion where local MB concentration can be maintained at a certain level that facilitates robust ULM performance (e.g., low MB density that facilitates better MB localization in larger vessels). This is similar to MUTE where local MB concentration is reduced by the acoustic nulls that suppress the MB signal. One downside of nanodroplets is that they can extravasate to extravascular space due to their small size, which confounds the microvascular imaging results. In addition, nanodroplets are not approved by the FDA and are not immediately available to be used for clinical applications. Nevertheless, nanodroplets do offer an enticing option for imaging beyond the microvasculature and providing new functional biomarkers such as vessel leakiness and targeted molecular imaging.
V. CONCLUSION
In this article, we demonstrate a new imaging sequence (MUTE—MB uncoupling via transmit excitation) that uses the principles of acoustic null and image subtraction to uncouple and separate spatially overlapping MB signals that are otherwise unlocalizable in conventional ULM imaging. We showed that MUTE could significantly improve the MB localization efficacy and imaging speed of ULM under high MB concentrations. These results indicate that MUTE has the potential to alleviate the issues of long data acquisition time and low temporal resolution of ULM, which is significant for practical implementations of ULM, especially under clinical settings.
ACKNOWLEDGMENT
The content is solely the responsibility of the authors and does not necessarily represent the official views of the National Cancer Institute and the National Institute of Biomedical Imaging and Bioengineering of the National Institute of Health.
This work was supported in part by the National Cancer Institute under Award R00CA214523 and in part by the National Institute of Biomedical Imaging and Bioengineering of the National Institutes of Health under Award R21EB030072.
This work involved human subjects or animals in its research. Approval of all ethical and experimental procedures and protocols was granted by the Institutional Animal Care and Use Committee (IACUC) at the University of Illinois Urbana–Champaign under Protocol No. 19063, and performed in line with the IACUC Guidelines.
Biography
Jihun Kim (Member, IEEE) was born in Changwon, South Korea, in 1990. He received the B.S. degree from the Department of Electronic Engineering, Hannam University, Daejeon, South Korea, in 2015, and the Ph.D. degree from the Department of Information and Communication Engineering, Daegu Gyeongbuk Institute of Science and Technology (DGIST), Daegu, South Korea, in 2019.
He was a Postdoctoral Research Associate with the Department of Aerospace and Mechanical Engineering, University of Notre Dame, Notre Dame, IN, USA. He is currently a Postdoctoral Research Associate with the Department of Electrical and Computer Engineering (ECE) and the Beckman Institute for Advanced Science and Technology, University of Illinois Urbana–Champaign (UIUC), Urbana, IL, USA. His research interests focus on the development of novel super-resolution ultrasound localization microscopy and shear wave elastography for tumor characterization in 3-D.
Mathew R. Lowerison received the Ph.D. degree in medical biophysics from the University of Western Ontario, London, ON, Canada, in 2017.
He is currently with the Department of Electrical and Computer Engineering and the Beckman Institute for Advanced Science and Technology, University of Illinois at Urbana–Champaign, Urbana, IL, USA. His current research interests include ultrasound microvessel imaging, super-resolution ultrasound localization microscopy, and ultrasonic characterization of tumor microenvironments.
Nathiya V. Chandra Sekaran received the Ph.D. degree in biochemistry from the University of Madras, Chennai, India, in 2015.
She is currently a Postdoctoral Research Fellow with the Daniel Llano Laboratory, Beckman Institute for Advanced Science and Technology, University of Illinois Urbana–Champaign, IL, USA. Her current research focuses on exploring cortical neurons plays in auditory system and imaging the aging rodent brain using super-resolution ultrasound microscopy.
Zhengchang Kou (Graduate Student Member, IEEE)received the B.S. and M.S. degrees in electrical engineering from the University of Illinois at Urbana-Champaign, Urbana, IL, USA, in 2017 and 2020, respectively. He is currently pursuing the Ph.D. degree with the Department of Electrical and Computer Engineering, University of Illinois at Urbana–Champaign, Urbana, IL, USA.
His current research interests are ultrasound communication and ultrasound imaging with programmable logics.
Zhijie Dong (Graduate Student Member, IEEE) received the B.Eng. degree in information engineering from Southeast University, Nanjing, China, in 2017, and the M.S. degree in electrical and computer engineering from the University of Michigan, Ann Arbor, MI, USA, in 2018. He is currently pursuing the Ph.D. degree with the Department of Electrical and Computer Engineering, University of Illinois at Urbana–Champaign, Urbana, IL, USA.
His current research interests are ultrafast 3-D imaging and signal processing.
Michael L. Oelze (Senior Member, IEEE) was born in Hamilton, New Zealand, in 1971. He received the B.S. degree in physics and mathematics from Harding University, Searcy, AR, USA, in 1994, and the Ph.D. degree in physics from The University of Mississippi, Oxford, MS, USA, in 2000.
From 2000 to 2002, he was a Postdoctoral Researcher with the Bioacoustics Research Laboratory, Department of Electrical and Computer Engineering (ECE), University of Illinois at Urbana–Champaign (UIUC), Urbana, IL, USA. From 2002 to 2004, he was an NIH Fellow conducting research in quantitative ultrasound techniques for biomedical ultrasound applications in cancer detection. In 2005, he joined the Faculty of ECE, UIUC, where he is currently a Professor and the Associate Head for Graduate Affairs in ECE, a Health Innovator Professor with the Carle Illinois College of Medicine, and the Frederick G. and Elizabeth H. Nearing Scholar, The Grainger College of Engineering. His research interests include biomedical ultrasound, quantitative ultrasound imaging for improving cancer diagnostics and monitoring therapy response, ultrasound tomography, ultrasound-based therapy, beamforming, and coded excitation and communications using ultrasound.
Dr. Oelze is currently a fellow of the AIUM, and a member of ASA. He is a member of the Technical Program Committee of the IEEE Ultrasonics Symposium. He currently serves as an Associate Editor for the IEEE TRANSACTIONS ON ULTRASONICS, FERROELECTRICS, AND FREQUENCY CONTROL, Ultrasonic Imaging, and IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING.
Daniel A. Llano received the M.D. and Ph.D. degrees from the University of Illinois at Urbana–Champaign, Urbana, IL, USA, in 2000 and 2002, respectively.
He completed neurology residency training at the Harvard Medical School, Boston, MA, USA, and postdoctoral training at the University of Chicago, Chicago, IL, USA. He is currently an Associate Professor of Molecular and Integrative Physiology and the Beckman Institute for Advanced Science and Technology, University of Illinois at Urbana–Champaign. His laboratory studies mechanisms of acoustic processing in the mammalian brain.
Pengfei Song (Senior Member, IEEE) received the B.S. degree in biomedical engineering from the Huazhong University of Science and Technology, Wuhan, China, in 2008, the M.S. degree in biological systems engineering from the University of Nebraska–Lincoln, Lincoln, NE, USA, in 2010, and the Ph.D. degree in biomedical engineering from the Mayo Clinic College of Medicine, Rochester, MN, USA, in 2014.
He is currently an Assistant Professor with the Department of Electrical and Computer Engineering, Beckman Institute for Advanced Science and Technology, Department of Bioengineering, and Carle Illinois College of Medicine, University of Illinois Urbana–Champaign, Urbana, IL, USA. His research interests include super-resolution ultrasound imaging, ultrafast ultrasound, deep learning, 3-D ultrafast imaging, functional ultrasound, and ultrasound shear wave elastography.
Dr. Song is a fellow of the American Institute of Ultrasound in Medicine, a Senior Member of the National Academy of Inventors, and a Full Member of the Acoustical Society of America.
Contributor Information
Jihun Kim, Department of Electrical and Computer Engineering, and the Beckman Institute for Advanced Science and Technology, University of Illinois Urbana–Champaign, Urbana, IL 61801 USA.
Mathew R. Lowerison, Department of Electrical and Computer Engineering, and the Beckman Institute for Advanced Science and Technology, University of Illinois Urbana–Champaign, Urbana, IL 61801 USA.
Nathiya V. Chandra Sekaran, Department of Molecular and Integrative Physiology, and the Beckman Institute for Advanced Science and Technology, University of Illinois Urbana–Champaign, Urbana, IL 61801 USA.
Zhengchang Kou, Department of Electrical and Computer Engineering, and the Beckman Institute for Advanced Science and Technology, University of Illinois Urbana–Champaign, Urbana, IL 61801 USA.
Zhijie Dong, Department of Electrical and Computer Engineering, and the Beckman Institute for Advanced Science and Technology, University of Illinois Urbana–Champaign, Urbana, IL 61801 USA.
Michael L. Oelze, Department of Electrical and Computer Engineering, and the Beckman Institute for Advanced Science and Technology, University of Illinois Urbana–Champaign, Urbana, IL 61801 USA.
Daniel A. Llano, Department of Molecular and Integrative Physiology, and the Beckman Institute for Advanced Science and Technology, University of Illinois Urbana–Champaign, Urbana, IL 61801 USA
Pengfei Song, Department of Electrical and Computer Engineering, and the Beckman Institute for Advanced Science and Technology, University of Illinois Urbana–Champaign, Urbana, IL 61801 USA.
REFERENCES
- [1].Betzig E et al. , “Imaging intracellular fluorescent proteins at nanometer resolution,” Science, vol. 313, no. 5793, pp. 1642–1645, 2006, doi: 10.1126/science.1127344. [DOI] [PubMed] [Google Scholar]
- [2].Rust MJ, Bates M, and Zhuang X, “Sub-diffraction-limit imaging by stochastic optical reconstruction microscopy (STORM),” Nature Methods, vol. 3, no. 10, pp. 793–796, 2006, doi: 10.1038/nmeth929. [DOI] [PMC free article] [PubMed] [Google Scholar]
- [3].Errico C et al. , “Ultrafast ultrasound localization microscopy for deep super-resolution vascular imaging,” Nature, vol. 527, no. 7579, pp. 499–502, Nov. 2015, doi: 10.1038/nature16066. [DOI] [PubMed] [Google Scholar]
- [4].Lowerison MR et al. , “Aging-related cerebral microvascular changes visualized using ultrasound localization microscopy in the living mouse,” Sci. Rep, vol. 12, no. 1, p. 619, Jan. 2022, doi: 10.1038/s41598-021-04712-8. [DOI] [PMC free article] [PubMed] [Google Scholar]
- [5].Foiret J, Zhang H, Ilovitsh T, Mahakian L, Tam S, and Ferrara KW, “Ultrasound localization microscopy to image and assess microvasculature in a rat kidney,” Sci. Rep, vol. 7, no. 1, p. 13662, Oct. 2017, doi: 10.1038/s41598-017-13676-7. [DOI] [PMC free article] [PubMed] [Google Scholar]
- [6].Zhang W, Lowerison MR, Dong Z, Miller RJ, Keller KA, and Song P, “Super-resolution ultrasound localization microscopy on a rabbit liver VX2 tumor model: An initial feasibility study,” Ultrasound Med. Biol, vol. 47, no. 8, pp. 2416–2429, Aug. 2021, doi: 10.1016/j.ultrasmedbio.2021.04.012. [DOI] [PMC free article] [PubMed] [Google Scholar]
- [7].Opacic T et al. , “Motion model ultrasound localization microscopy for preclinical and clinical multiparametric tumor characterization,” Nature Commun, vol. 9, no. 1, p. 1527, 2018, doi: 10.1038/s41467-018-03973-8. [DOI] [PMC free article] [PubMed] [Google Scholar]
- [8].Bar-Zion A, Tremblay-Darveau C, Solomon O, Adam D, and Eldar YC, “Fast vascular ultrasound imaging with enhanced spatial resolution and background rejection,” IEEE Trans. Med. Imag, vol. 36, no. 1, pp. 169–180, Jan. 2017, doi: 10.1109/TMI.2016.2600372. [DOI] [PubMed] [Google Scholar]
- [9].Bar-Zion A, Solomon O, Tremblay-Darveau C, Adam D, and Eldar YC, “SUSHI: Sparsity-based ultrasound super-resolution hemodynamic imaging,” IEEE Trans. Ultrason., Ferroelectr., Freq. Control, vol. 65, no. 12, pp. 2365–2380, Dec. 2018, doi: 10.1109/TUFFC.2018.2873380. [DOI] [PubMed] [Google Scholar]
- [10].Lowerison MR, Huang C, Lucien F, Chen S, and Song P, “Ultrasound localization microscopy of renal tumor xenografts in chicken embryo is correlated to hypoxia,” Sci. Rep, vol. 10, no. 1, p. 2478, Dec. 2020, doi: 10.1038/s41598-020-59338-z. [DOI] [PMC free article] [PubMed] [Google Scholar]
- [11].Perry DC et al. , “Abnormalities of vascular structure and function in children with perthes disease,” Pediatrics, vol. 130, no. 1, pp. e126–e131, Jul. 2012, doi: 10.1542/peds.2011-3269. [DOI] [PubMed] [Google Scholar]
- [12].Kim J, Wang Q, Zhang S, and Yoon S, “Compressed sensing-based super-resolution ultrasound imaging for faster acquisition and high quality images,” IEEE Trans. Biomed. Eng, vol. 68, no. 11, pp. 3317–3326, Nov. 2021, doi: 10.1109/TBME.2021.3070487. [DOI] [PMC free article] [PubMed] [Google Scholar]
- [13].Hell SW, “Far-field optical nanoscopy,” Science, vol. 316, no. 5828, pp. 1153–1158, May 2007. [DOI] [PubMed] [Google Scholar]
- [14].Couture O, Bannouf S, Montaldo G, Aubry JF, Fink M, and Tanter M, “Ultrafast imaging of ultrasound contrast agents,” Ultrasound Med. Biol, vol. 35, no. 11, pp. 1908–1916, 2009, doi: 10.1016/j.ultrasmedbio.2009.05.020. [DOI] [PubMed] [Google Scholar]
- [15].Agarwal A, Reeg J, Podkowa AS, and Oelze ML, “Improving spatial resolution using incoherent subtraction of receive beams having different apodizations,” IEEE Trans. Ultrason., Ferroelectr., Freq. Control, vol. 66, no. 1, pp. 5–17, Jan. 2019, doi: 10.1109/TUFFC.2018.2876285. [DOI] [PMC free article] [PubMed] [Google Scholar]
- [16].Savoia AS et al. , “Improved lateral resolution and contrast in ultrasound imaging using a sidelobe masking technique,” in Proc. IEEE Int. Ultrason. Symp, Sep. 2014, pp. 1682–1685, doi: 10.1109/ULTSYM.2014.0417. [DOI] [Google Scholar]
- [17].Kou Z, Lowerison M, Song P, and Oelze ML, “Fast high-resolution ultrasound microvessel imaging with null subtraction imaging-based beamforming,” in Proc. IEEE Int. Ultrason. Symp. (IUS), Sep. 2020, pp. 1–3, doi: 10.1109/IUS46767.2020.9251773. [DOI] [Google Scholar]
- [18].Korobchevskaya K et al. , “Intensity weighted subtraction microscopy approach for image contrast and resolution enhancement,” Sci. Rep, vol. 6, no. 1, p. 25816, May 2016, doi: 10.1038/srep25816. [DOI] [PMC free article] [PubMed] [Google Scholar]
- [19].Montaldo G, Tanter M, Bercoff J, Benech N, and Fink M, “Coherent plane-wave compounding for very high frame rate ultrasonography and transient elastography,” IEEE Trans. Ultrason., Ferroelectr., Freq. Control, vol. 56, no. 3, pp. 489–506, Mar. 2009, doi: 10.1109/TUFFC.2009.1067. [DOI] [PubMed] [Google Scholar]
- [20].Demené C et al. , “Spatiotemporal clutter filtering of ultrafast ultrasound data highly increases Doppler and fUltrasound sensitivity,” IEEE Trans. Med. Imag, vol. 34, no. 11, pp. 2271–2285, Nov. 2015, doi: 10.1109/TMI.2015.2428634. [DOI] [PubMed] [Google Scholar]
- [21].Song P, Manduca A, Trzasko JD, and Chen S, “Ultrasound small vessel imaging with block-wise adaptive local clutter filtering,” IEEE Trans. Med. Imag, vol. 36, no. 1, pp. 251–262, Jan. 2017, doi: 10.1109/TMI.2016.2605819. [DOI] [PubMed] [Google Scholar]
- [22].Song P, Manduca A, Trzasko JD, and Chen S, “Noise equalization for ultrafast plane wave microvessel imaging,” IEEE Trans. Ultrason., Ferroelectr., Freq. Control, vol. 64, no. 11, pp. 1776–1781, Nov. 2017, doi: 10.1109/TUFFC.2017.2748387. [DOI] [PMC free article] [PubMed] [Google Scholar]
- [23].Huang C et al. , “Short acquisition time super-resolution ultrasound microvessel imaging via microbubble separation,” Sci. Rep, vol. 10, no. 1, p. 6007, Apr. 2020, doi: 10.1038/s41598-020-62898-9. [DOI] [PMC free article] [PubMed] [Google Scholar]
- [24].Song P et al. , “Improved super-resolution ultrasound microvessel imaging with spatiotemporal nonlocal means filtering and bipartite graph-based microbubble tracking,” IEEE Trans. Ultrason., Ferroelectr., Freq. Control, vol. 65, no. 2, pp. 149–167, Feb. 2018, doi: 10.1109/TUFFC.2017.2778941. [DOI] [PMC free article] [PubMed] [Google Scholar]
- [25].Tang S et al. , “Kalman filter-based microbubble tracking for robust super-resolution ultrasound microvessel imaging,” IEEE Trans. Ultrason., Ferroelectr., Freq. Control, vol. 67, no. 9, pp. 1738–1751, Sep. 2020, doi: 10.1109/TUFFC.2020.2984384. [DOI] [PMC free article] [PubMed] [Google Scholar]
- [26].Qin S, Caskey CF, and Ferrara KW, “Ultrasound contrast microbubbles in imaging and therapy: Physical principles and engineering,” Phys. Med. Biol, vol. 54, no. 6, pp. R27–R57, Feb. 2009, doi: 10.1088/0031-9155/54/6/R01. [DOI] [PMC free article] [PubMed] [Google Scholar]
- [27].Zhang G et al. , “Fast acoustic wave sparsely activated localization microscopy: Ultrasound super-resolution using plane-wave activation of nanodroplets,” IEEE Trans. Ultrason., Ferroelectr., Freq. Control, vol. 66, no. 6, pp. 1039–1046, Jun. 2019, doi: 10.1109/TUFFC.2019.2906496. [DOI] [PubMed] [Google Scholar]