Abstract
BACKGROUND:
Liver iron concentration (LIC) measured by MRI has become the clinical reference standard for managing iron overload in chronically transfused patients. Transverse relaxivity (R2 or R2*) measurements are converted to LIC units using empirically derived calibration curves.
HYPOTHESIS:
That flip angle (FA) error due to B1+ spatial heterogeneity causes significant LIC quantitation error. B1+ scale (b1, [FAactual/FAspecified]) variation is a major problem at 3T which could reduce the accuracy of transverse relaxivity measurements.
STUDY TYPE:
Prospective
POPULATION:
47 subjects with chronic transfusional iron overload undergoing clinically-indicated LIC assessment.
FIELD STRENGTH/SEQUENCE:
1.5T/3T dual-TR B1+ mapping sequence
ASSESSMENT:
We quantified the average/standard deviation b1 in the right and left lobes of the liver from B1+ maps acquired at 1.5T and 3T. The impact of b1 variation on spin-echo LIC estimates was determined using a Monte Carlo model.
STATISTICAL TESTS:
Mean, median, and standard deviation in whole liver and right and left lobes; two-sided t-test between whole-liver b1 means.
RESULTS:
Average b1 within the liver was 99.3%±12.3% at 1.5T versus 69.6%±14.6% at 3T and was independent of iron burden (p<0.05). Monte-Carlo simulations demonstrated that b1 systematically increased R2 estimates at lower LIC (<~25mg/g@1.5T, <~15mg/g@3T) but flattened or even inverted the R2-LIC relationship at higher LIC (≥~25mg/g@1.5T, ≥~15mg/g@3T); changes in the R2-LIC relationship were symmetric with respect to over and under excitation and were similar at 1.5T and 3T (for the same R2 value). The R2*-LIC relationship was independent of b1.
CONCLUSION:
Spin echo R2 measurement of LIC at 3T is error-prone without correction for b1 errors. The impact of b1 error on current 1.5T spin echo-based techniques for LIC quantification is large enough to introduce measurable inter-subject variability but the in-vivo effect size needs a dedicated validation study.
Introduction
Chronic anemias such as thalassemia and sickle cell disease represent the most common genetic disorders in the world1. Transfusion therapy in these patients produces severe iron deposition in the liver and other organs, leading to cardiac and endocrine dysfunction and liver cirrhosis2. Iron overload in transfusional siderosis cannot be treated with phlebotomy; instead, patients receive iron chelators that bind and remove iron with dosing specific to their tissue iron content, necessitating reliable iron quantitation techniques3,4. Before 2005, needle biopsies were required to quantify liver iron, with their attendant risks5 and sampling error6–8. Since then, MRI systems have become key tools for diagnosing and following iron overload disorders such as transfusional siderosis in sickle cell disease and thalassemia, hemochromatosis, liver disease, and a number of neurodegenerative disorders9,10. Both transverse relaxivity and magnetic susceptibility measurements have been used to estimate tissue iron concentrations in the liver, heart, endocrine glands, and brain11.
MRI-based iron quantification at 1.5 Tesla (T) is now standard of care12 but quantification of liver iron using high-field scanners (3T and above) is limited by excessively rapid signal decay and susceptibility artifacts. When increasing field strengths from 1.5T to 3T, field-dependent enhancement causes the slope of the relationship between LIC and R2* (1/T2*) decay to double, leading to transverse decay below 0.5 ms in organs such as the liver, well below the range standard gradient-echo techniques can reliably quantify13. Spin echo-based R2 (1/T2) estimates, which do not enhance as dramatically with increased field strength, have the potential to overcome these limitations at 3T and already see widespread use at 1.5T11. However, spin echo estimates are more susceptible to spatial variations in the radiofrequency transmit efficiency, B1+ heterogeneity, leading to incorrect flip angles throughout the image, a well-characterized imaging problem in high field scanners14,15. The ratio of the achieved flip angle to the target flip angle is known as the B1+ scale and is represented in this manuscript as b1. Quantifying and correcting for this heterogeneity could lead to more accurate liver iron concentration (LIC) estimates. Accordingly, this study aimed to measure b1 heterogeneity at 1.5T and 3T in patients over a broad range of iron burdens and to simulate the estimated error in R2-based LIC estimates at both field strengths using a Monte Carlo model.
Materials and Methods
Patient Population
Study participants were selected from a population of patients undergoing clinical treatment for iron overload primarily resulting from transfusional treatment of sickle cell disease, thalassemia, and other rare anemia syndromes. Participants were recruited and provided informed consent to participate in an IRB-approved study (CCI-14–00034). Each participant received a clinically indicated MRI assessment for iron overload via R2* on the 1.5T scanner; a similar research protocol was completed at 3T. All images were acquired on single-RF-transmit 1.5T and 3T Philips Achieva magnets (v3.2.2, Best, Netherlands) using 16-element SENSE-XL torso coils.
Patient Assessment
Imaging was performed on in 47 representative subjects (19M, 28F, 20.1±11.6 years of age, 19 β-Thalassemia, 21 Sickle Cell, 7 other rare anemia, LIC=0.5–41.5 mg FE/g liver tissue). Liver iron estimates were obtained at 1.5T using a 3-slice gradient echo acquisition with the following parameters: 16 linearly spaced echoes with TE/TR=0.96–11.47/50ms, FA=30°, matrix/voxel=72×67/1.25×1.25×10mm, BW=4409Hz/px. R2* estimates were made in hand-segmented regions of interest (ROI) in the liver, excluding major hilar vessels, and fit using an exponential+constant model to the data using a constrained Levenburg-Marquadt algorithm in custom MATLAB (Mathworks, Natick, MA) code; R2* values were converted to LIC estimates using a R2*-LIC calibration curve derived from biopsy16.
B1+ scale maps were obtained at 1.5T and 3T in a single slice using a dual repetition time (TR) b1 mapping sequence17 with TE/TR1/TR2=3ms/20ms/120ms, bandwidth of 2894Hz/px, voxel size of 4.2×4.2×8 mm, and one signal average. A second set of b1 maps were obtained at both fields using an echo time of one millisecond to ensure that we could sample severely iron loaded patients who demonstrate particularly rapid T2* decay. b1 maps were validated in a phantom (0–24 mM MnCl2 vials in a 0.5 mM MnCl2 bath) with transverse decay rates as low as 0.4 ms to ensure that the b1 maps had sufficient dynamic range to reliably map the patient population.
The maps were manually segmented by a biomedical engineering doctoral student (ED) with 5 years of experience to quantify the b1 heterogeneity found in this iron overloaded patient population. B1+ scale maps were assessed for dependence on patient size and clinical iron-load. Patient b1 maps were averaged after non-rigidly warping individual maps to a canonical liver shape to generate an aggregate b1 map. Because of imperfect registration, some pixels in the canonical liver shape received contributions from fewer than 5 patients and were excluded; each pixel was weighted according to the number of contributing patients. The warped b1 maps were used to estimate spatial statistics in the right and left lobe.
Simulation
An internally developed simulation framework written in MATLAB was used to create physiologically realistic liver iron geometries in-silico over a range of 1–50 mg/g (mg iron/g dry tissue weight, iron loads of=1,[5:5:50]mg/g). The framework was validated in human subjects at 1.5T and 3T13. Detailed descriptions of these models may be found in previous reports18–20. Briefly, liver tissue was modeled as 80 μm cubes of liver tissue containing 64 cuboidal hepatocytes and 18 cylindrical sinusoid regions18. Iron overload was modeled by distributing spheres of iron in the hepatocytes using Gamma distributions describing size and shape. From the iron distributions, magnetic field disturbances were calculated based on the field strength and magnetic susceptibility of iron19. Simulated protons were then diffused through the geometries using a random walk simulation obeying cell boundaries. The previous version of the simulation framework assumed perfect excitation and refocusing, ignored pulse bandwidth, and did not model longitudinal magnetization or T1 decay.
To model b1 heterogeneities, we implemented a full Bloch simulator to calculate the 3D magnetization of each proton at every time step. The simulations calculated rotations from instantaneous excitation and inversion pulses, realistic iron-free T2/T1 relaxation rates of 42ms/812ms at 3T and 50ms/576ms at 1.5T21, and the iron-based phase accrual from the proton’s random walk at each 0.05 μs time step. Iron-based T1 changes were ignored due to minimal longitudinal recovery over the duration of the echo trains simulated and lack of multiple TRs. Excitations and inversions were modeled as instantaneous pulses and achieved angles were scaled by b1 amplitude for each simulation. For example, a 90x-180y single spin echo sequence with 90% excitation efficiency would be simulated with an excitation pulse of 81° about x and inversion of 162° about y. Signals from 2500 separate spin cohorts were superposed to form a transverse signal for each combination of iron load and b1.
Simulations varied b1 from 53% to 127% in single echo and Carr-Purcell-Meiboom-Gill (CPMG) multiecho spin echo experiments with 90° excitations and 180° inversions; spin echo and CPMG echo times were [1.70, 2.56, 3.86, 5.82, 8.76, 13.2, 19.9, 30.0] ms and [2.0, 4.0, 6.0, 8.0, 10, 12, 14, 16, 18, 20] ms, respectively; the simulated b1 range was chosen empirically based on the range demonstrated by subject b1 maps. Complex signals were fit using a complex signal model with a non-negativity constraint to estimate the R2 relaxation rate of each sample. R2 as a function of LIC was determined for each simulated b1 by fitting individual results to a 3rd order polynomial. R2* was calculated from the simulation framework by fitting the envelope of the FID to a simple monoexponential.
Post-study investigation
Following the conclusion of study in patients and initial data analysis, we assessed the feasibility of overcoming issues identified at 3T using tools available on many clinical MRI systems. Following completion of the patient studies and the initial data analyses, b1 maps were obtained in 5 healthy volunteers using the same pulse sequences on a Philips 3T Ingenia platform with dual transmit capability. We attempted to correct the underexcitation by manually increasing the RF transmit gain proportional to the demonstrated underexcitation (gainnew = gainoriginal * 1/b1). We then tested adaptive RF shimming to determine if spatial variability could be improved.
Data availability
Code and data relating to this work is available in the following git repo:
Statistical Analysis
Mean B1 from acquired B1 maps was assessed for dependence on LIC and subject weight via linear regression. There was no significant dependence on LIC at either field strength. Mean B1 at 3T demonstrated a modest but significant dependence on weight; the relationship was insignificant at 1.5T.
LIC vs B1 regression
1.5T: B1mean = 0.087*LIC + 99.4 (p =0.143)
3T: B1mean = −0.178*LIC + 75.9 (p = 0.127)
Weight vs B1 regression
1.5T B1mean = 0.086*weight + 98.7 (p = 0.66)
3T B1mean = −0.24*weight + 86.5 (p < 0.05)
A two-sided t-test comparing 1.5T and 3T mean B1 values rejected the null hypothesis that the means were the same (p<0.05).
Results
Figure 1 demonstrates b1 heterogeneity in the liver at 1.5T and 3T. At 3T, mean b1 was low and the standard deviation was high (69.6%±14.6%). Excitation efficiency was higher in the right lobe than the left lobe (left mean=62.1%±12.8%, right mean=73.6%±13.9%) but there was considerable inter-subject variability. B1+ scale heterogeneity was independent of liver iron burden. Both the spatial homogeneity and the mean achieved b1 (99.3%±12.3%) were better at 1.5T compared to 3T; right and left lobe means were 103.3% and 92.0% respectively.
Figure 1-.

B1+ scale heterogeneity maps in patients demonstrate significantly different behavior at 1.5T and 3T. 1.5T maps demonstrate largely uniform excitation patterns and consistently complete excitation. 3T maps demonstrate significant spatial variation in b1 as well as a mean achieved b1 of only 69.6%.
Simulations results demonstrating the effects of varying b1 on multiple single echo R2 estimates and CPMG multiecho R2 estimates are shown in Figure 2 and Figure 3, respectively. The R2* iron calibration was robust to b1 errors at both 1.5T and 3T (not shown); no change in R2*-LIC calibration was detectable. Spin echo simulations demonstrated that calibration error increased noticeably with b1 error. At 1.5 T, b1 error caused single echo R2 to rise more quickly with iron concentration than the true calibration curve but then to flatten and even decrease at higher iron levels. The impact of b1 error was relatively symmetric about 100%, i.e. overexcitation and underexcitation produced similar changes to the R2-LIC relationship. When underexcitation was severe (b1 < 67%), increases in iron load greater 25 mg/g caused R2 estimates to decline.
Figure 2-.

Single echo simulations over a wide range of iron loads and b1 values demonstrates that R2, and therefore iron concentration, is overestimated in the presence of b1 at 1.5T. Further, inaccuracies increase with iron load for a given b1 value. Although at 1.5T R2 estimates tend to increase with b1 values, 3T maps show the opposite effect for LIC>~15mg/g due the combination of fast initial decay and shortened T1 resulting in artificially flat signal decay curves which are interpreted as having slow T2 decay when fitting is performed in the absence of additional constraints31.
Figure 3-.

CPMG simulations over a wide range of iron loads and b1 heterogeneities demonstrate that R2, and therefore iron concentration, is overestimated for all b1 values except with low iron loads at high b1 values. Further, inaccuracies increase with iron load for a given b1 value.
Single echo R2-LIC calibrations at 3T demonstrate a similar appearance with b1 error, but the effects were amplified. Inverse relationships between R2 and iron burden were observed at lower b1 (<80%) and higher iron burdens (>15 mg/g) (ie R2 estimates decreased as LIC increased). Using the b1 estimates derived from patient data (69.6%±14.6%), more than half of liver voxels corresponded to calibration curves that do not have a unique relationship between R2 and liver iron at 3T.
Multi-echo calibrations at 1.5T were qualitatively similar to single echo results, except for the expected blunting of R2 relaxation across all iron loads22. Multi-echo calibrations were more stable over b1 ranges of ±20% but demonstrated a larger proportional b1 effect than the single echo calibration for larger b1 values. At 3T, multi-echo R2 exhibited a more linear relationship with liver iron concentration. Spatial heterogeneities in b1 primarily changed the intercept of the R2-LIC relationships. R2 retained a 1:1 relationship with LIC until b1 was < 60%.
Supplemental Figure 1 demonstrates b1 mapping performed on a health volunteer at 3T with (bottom panel) and without (middle panel) adaptive RF shimming from dual RF transmission. The mean liver b1 was 76.2% ± 9.2% without dual transmit, similar to the average b1 values in our study. B1+ shimming corrected the systematic under-excitation but did not lower the spatial variability (102.2%±10.2%).
Supplementary Figure 2 restates the results from Figure 1 and Figure 2 as percent error of R2 versus LIC. The figure helps to demonstrate that as B1+ error increases, SE R2 is overestimated for lower iron loads and underestimated for higher iron loads. CPMG R2 estimates on the other hand always demonstrate an overestimate of R2 with B1+ error, though the error decreases with increasing LIC.
Supplementary Figure 3 demonstrates the long and short TE B1+ mapping sequences in a manganese chloride phantom. The images demonstrate that both sequences have sufficient dynamic range to perform B1+ mapping across the physiologically relevant range of liver iron.
Discussion
This study examined the magnitude of b1 error in the liver and its potential impact on clinical liver iron estimates. Iron chelation therapy is tuned based on LIC estimates and patient motivation to continue using chelating drugs is heavily motivated by their LIC estimates demonstrating efficacy over time. Thus, the accuracy of iron estimates is important for both clinicians and patients. This work has demonstrated that B1+ heterogeneity presents a quantitative challenge at 3T and, more surprisingly, that spatial flip angle variation at 1.5T may be sufficient to cause meaningful quantitative changes.
At 1.5T, the average excitation achieved in the liver was complete (ie matched the flip angle requested) but there was variation between the right and left lobes. Substantially larger spatial variations in b1 were observed in the liver at 3T. We had postulated a potential iron effect because B0 and B1+ heterogeneities can interact23. Because off-resonance from liver iron can exceed 1000 Hz for extreme clinical iron loads24, we expected the excitation bandwidth for standard b1 mapping sequences would to be too narrow, although the phantom study demonstrated a sufficiently short TE. Further, off-resonance can cause slice misplacement or distrotion25. In spite of this, excitation efficiency was uncorrelated with liver iron and manual visual evaluation of the images did not indicate that slices were meaningfully shifted.
At 3T, we observed a systematic underexcitation in the liver, averaging just ~70% of our specified flip angle in whole-liver ROIs. The effect was not observed at 1.5T. The underexcitation was observed both in patients and in healthy controls. Manual increases of the transmit gain improved whole-ROI average to near 100% at 3T, although the observed spatial heterogeneity remained, suggesting that the under-excitation may have been an unintended consequence of SAR modeling. We do not have access to abdominal 3T b1 maps from other imaging platforms to determine whether the observed underexcitation was vendor-specific, although this has been demonstrated across multiple vendors at 3T26. Further, a recent study of liver B1 inhomogeneity by Roberts, et.al noted similar spatial variation and average b1 scale at both 1.5T and 3T in a larger patient cohort27. Of note, adaptive RF shimming improved the systematic underestimation but did not lower the spatial variability in our small post-study investigation; however, Roberts et.al demonstrated underexcitation at 3T even with dual-transmit enabled on a different vendor.
In single echo simulations, the estimated error appeared to be primarily dependent on decay rate rather than field strength. The overestimation of liver R2 occurs up to 200–250 s−1 in both 1.5T and 3T simulation, representing approximate LIC burdens of 30 and 15 mg/g respectively. Note that the echo-sampling scheme (logarithmically spaced from 1.7 to 30 ms) should have been sufficient to capture rapid early signal decay. The increase in apparent R2 decay with b1 error arises from non-exponential signal loss in later echoes that prematurely collapses the decay curve. At decay rates over 250 s−1, the amplitude of the early echoes is significantly reduced as well, flattening the decay curve and lowering the R2 estimate. As iron load and b1 error increase, it becomes impossible to uniquely determine LIC from measured R2 values. Of note, the quantitative effects are symmetric for absolute b1 error due to the fact that spin echo signals are maximized when a 90° excitation and 180° inversion are achieved; both over-excitation and under-excitation lead to the same apparent signal for both single echo and CPMG sequences.
For CPMG simulation estimates, the phase cycling of the echo pulses leads to “sawtoothing” that envelopes the true decay curve: the odd echoes have lower amplitudes than subsequent even echoes. For smaller errors in b1, the error is mostly self-correcting. As b1 becomes large, a significant portion of the longitudinal magnetization remains after the excitation pulse and is recovered into the transverse plane when subsequent echo pulses are applied; similarly, transverse magnetization is stored in the longitudinal plane when incomplete echo pulses are applied. The sawtoothing eventually increases to the point that the first echo amplitude is smaller than all of the even echoes, even for very high LIC burdens; the resulting fits show reduced R2 estimates. The differing behavior of the 1.5T and 3T calibration curves in the presence of b1 error over ~25% is primarily a T1 effect. The longer T1 at 3T compared to 1.5T results in less longitudinal magnetization recovery, leading to a less pronounced addition of transverse signal over the course of the echo train. Later echoes at 3T therefore demonstrate more realistic decay than 1.5T echoes in the presence of large b1 error. Nonetheless, the inclusion of longitudinal magnetization and diminished first echo intensity lead to the apparent reduction of R2 for high iron loads in the presence of large b1 error.
At 1.5T, our study showed no systematic underexcitation, however the 95% confidence intervals were large enough to potentially impact signal quantification for clinical R2 imaging16,28. For a patient with true LIC of 10 mg/g, a b1 error of +/−10% will increase the predicted LIC by 30%. This level of error may not change the dosing of chelation therapy, but such variation could lead to incorrect assumptions about the efficacy of a particular course of chelation therapy. Inter-subject variation in b1 heterogeneity produces “noise” in R2-based LIC estimates while having no effect on R2*-based LIC estimates, contributing to disagreement between these methods29,30. B1+ variation could also contribute to the observed curvilinearity in the R2-LIC calibration curve28. With perfect excitation, the R2-LIC relationship is relatively linear over a wide range of LICs but even mild b1 errors introduce marked concavity to the calibration curve. B1+ scale heterogeneity could also potentially confound techniques that rely on proton-density estimators to stabilize LIC measurements at high iron concentrations28. Carefully controlled studies are warranted to determine whether B1+ correction could improve R2-based LIC measurements at 1.5T.
In contrast, correcting for b1 error, especially whole-liver underexcitation, is an absolute necessity for R2-based iron estimation at 3T. Our post-study investigations suggest that RF shimming using dual transmit RF systems can eliminate systematic underexcitation at 3T, without significantly impacting spatial variability, and should be considered a prerequisite for spin echo quantitation at 3T. Alternatively, one could limit R2 analysis only to those voxels having b1 of 90–110%, accepting the sampling limitations that this would impose.
Limitations
First, the B1+ mapping was performed only on Philips MR systems. Although the simulation has been well-validated, R2 estimates in patients were not performed. The simulation framework ignored iron-mediated T1 shortening and slice profile effects were not considered. Finally, the echo times chosen in the simulation are relatively fast and most clinical magnets may not be able to achieve such short TEs.
Conclusions
We observed clinically relevant b1 variability at both 1.5T and 3T as well as large global underexcitation at 3T. Simulation results demonstrated that even a modest amount of error in b1 leads to systematic LIC estimation error with spin echo protocols. Spin echo liver iron quantification cannot be performed reliably at 3T without RF shimming. Further work is necessary to validate the impact of b1 heterogeneity on current FDA-approved approaches for LIC quantification at 1.5T and to evaluate the performance of other models and vendors of MRI scanners.
Supplementary Material
Acknowledgements
Computation for the work described in this paper was supported by the University of Southern California’s Center for High-Performance Computing (hpc.usc.edu).
Grant Support
This research is supported by the National Institute of Diabetes, Digestion and Kidney Diseases of the National Institutes of Health, grant 1R01DK097115–01A1.
References
- 1.Baird PA, Anderson TW, Newcombe HB, Lowry RB. Genetic disorders in children and young adults: a population study. American Journal of Human Genetics 1988;42(5):677–693. [PMC free article] [PubMed] [Google Scholar]
- 2.Papakonstantinou O, Alexopoulou E, Economopoulos N, Benekos O, Kattamis A, Kostaridou S, Ladis V, Efstathopoulos E, Gouliamos A, Kelekis NL. Assessment of iron distribution between liver, spleen, pancreas, bone marrow, and myocardium by means of R2 relaxometry with MRI in patients with β-thalassemia major. Journal of Magnetic Resonance Imaging 2009;29(4):853–859. [DOI] [PubMed] [Google Scholar]
- 3.Jamuar SS, Lai AHM. Safety and efficacy of iron chelation therapy with deferiprone in patients with transfusion-dependent thalassemia. Therapeutic Advances in Hematology 2012;3(5):299–307. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 4.Wood JC, Ghugre N. Magnetic resonance imaging assessment of excess iron in thalassemia, sickle cell disease and other iron overload diseases. Hemoglobin 2008;32(1–2):85–96. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 5.Angelucci E, Baronciani D, Lucarelli G, Baldassarri M, Galimberti M, Giardini C, Martinelli F, Polchi P, Polizzi V, Ripalti M, et al. Needle liver biopsy in thalassaemia: analyses of diagnostic accuracy and safety in 1184 consecutive biopsies. British Journal of Haematology 1995;89(4):757–761. [DOI] [PubMed] [Google Scholar]
- 6.Ambu R, Crisponi G, Sciot R, Eyken PV, Parodo G, Ianneli S, Marongiu F, Silvagni R, Nurchi V, Costa V, et al. Uneven hepatic iron and phosphorus distribution in beta-thalassemia. Journal of Hepatology 1995;23(5):544–549. [DOI] [PubMed] [Google Scholar]
- 7.Emond MJ, Bronner MP, Carlson TH, Lin M, Labbe RF, Kowdley KV. Quantitative Study of the Variability of Hepatic Iron Concentrations. Clinical Chemistry 1999;45(3):340–346. [PubMed] [Google Scholar]
- 8.Villeneuve J-P, Bilodeau M, Lepage R, Côté J, Lefebvre M. Variability in hepatic iron concentration measurement from needle-biopsy specimens. Journal of Hepatology 1996;25(2):172–177. [DOI] [PubMed] [Google Scholar]
- 9.Miao X, Choi S, Tamrazi B, Chai Y, Vu C, Coates TD, Wood JC. Increased brain iron deposition in patients with sickle cell disease: an MRI quantitative susceptibility mapping study. Blood 2018;132(15):1618–1621. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 10.Wood JC. Magnetic resonance imaging measurement of iron overload. Current opinion in hematology 2007;14(3):183–190. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 11.Wood JC. Estimating tissue iron burden: current status and future prospects. British Journal of Haematology 2015;170(1):15–28. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 12.Wood JC. Diagnosis and management of transfusion iron overload: The role of imaging. American journal of hematology 2007;82(12 Suppl):1132–1135. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 13.Ghugre NR, Doyle EK, Storey P, Wood JC. Relaxivity-iron calibration in hepatic iron overload: Predictions of a Monte Carlo model. Magnetic Resonance in Medicine 2015;74(3):879–883. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 14.Greenman RL, Shirosky JE, Mulkern RV, Rofsky NM. Double inversion black-blood fast spin-echo imaging of the human heart: A comparison between 1.5T and 3.0T. Journal of Magnetic Resonance Imaging 2003;17(6):648–655. [DOI] [PubMed] [Google Scholar]
- 15.Hornak JP, Szumowski J, Bryant RG. Magnetic field mapping. Magnetic resonance in medicine 1988;6(2):158–163. [DOI] [PubMed] [Google Scholar]
- 16.Wood JC, Enriquez C, Ghugre N, Tyzka JM, Carson S, Nelson MD, Coates TD. MRI R2 and R2* mapping accurately estimates hepatic iron concentration in transfusion-dependent thalassemia and sickle cell disease patients. Blood 2005;106(4):1460–1465. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 17.Yarnykh VL. Actual flip-angle imaging in the pulsed steady state: a method for rapid three-dimensional mapping of the transmitted radiofrequency field. Magnetic Resonance in Medicine: Official Journal of the Society of Magnetic Resonance in Medicine / Society of Magnetic Resonance in Medicine 2007;57(1):192–200. [DOI] [PubMed] [Google Scholar]
- 18.Ghugre NR, Gonzalez-Gomez I, Shimada H, Coates TD, Wood JC. Quantitative analysis and modelling of hepatic iron stores using stereology and spatial statistics. Journal of microscopy 2010;238(3):265–274. [DOI] [PubMed] [Google Scholar]
- 19.Ghugre NR, Wood JC. Relaxivity-iron calibration in hepatic iron overload: probing underlying biophysical mechanisms using a Monte Carlo model. Magnetic Resonance in Medicine: Official Journal of the Society of Magnetic Resonance in Medicine / Society of Magnetic Resonance in Medicine 2011;65(3):837–847. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 20.Ghugre NR, Coates TD, Nelson MD, Wood JC. Mechanisms of tissue-iron relaxivity: nuclear magnetic resonance studies of human liver biopsy specimens. Magnetic resonance in medicine: official journal of the Society of Magnetic Resonance in Medicine / Society of Magnetic Resonance in Medicine 2005;54(5):1185–1193. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 21.Stanisz GJ, Odrobina EE, Pun J, Escaravage M, Graham SJ, Bronskill MJ, Henkelman RM. T1, T2 relaxation and magnetization transfer in tissue at 3T. Magnetic resonance in medicine: official journal of the Society of Magnetic Resonance in Medicine / Society of Magnetic Resonance in Medicine 2005;54(3):507–512. [DOI] [PubMed] [Google Scholar]
- 22.Ghugre NR, Wood JC. Modeling CPMG behavior in a realistic magnetically-heterogeneous tissue matrix :1. [Google Scholar]
- 23.Sung K, Nayak KS. Design and use of tailored hard-pulse trains for uniformed saturation of myocardium at 3 Tesla. Magnetic Resonance in Medicine 2008;60(4):997–1002. [DOI] [PubMed] [Google Scholar]
- 24.Hernando D, Cook RJ, Diamond C, Reeder SB. Magnetic Susceptibility as a B0 Field Strength Independent MRI Biomarker of Liver Iron Overload. Magnetic resonance in medicine : official journal of the Society of Magnetic Resonance in Medicine / Society of Magnetic Resonance in Medicine 2013;70(3):648–656. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 25.Harder JC den, Yperen GH van, Blume UA, Bos C. Off-resonance suppression for multispectral MR imaging near metallic implants. Magnetic Resonance in Medicine 2015;73(1):233–243. [DOI] [PubMed] [Google Scholar]
- 26.Bliesener Y, Zhong X, Guo Y, Boss M, Bosca R, Laue H, Chung C, Sung K, Nayak KS. Radiofrequency transmit calibration: A multi‐center evaluation of vendor‐provided radiofrequency transmit mapping methods. Medical Physics 2019;46(6):2629–2637. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 27.Roberts NT, Hinshaw LA, Colgan TJ, Ii T, Hernando D, Reeder SB. B0 and B1 inhomogeneities in the liver at 1.5 T and 3.0 T. Magnetic Resonance in Medicine 2021;85(4):2212–2220. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 28.Pierre TGS, Clark PR, Chua-anusorn W, Fleming AJ, Jeffrey GP, Olynyk JK, Pootrakul P, Robins E, Lindeman R. Noninvasive measurement and imaging of liver iron concentrations using proton magnetic resonance. Blood 2005;105(2):855–861. [DOI] [PubMed] [Google Scholar]
- 29.Wood JC, Cohen AR, Pressel SL, Aygun B, Imran H, Luchtman-Jones L, Thompson AA, Fuh B, Schultz WH, Davis BR, et al. Organ iron accumulation in chronically transfused children with sickle cell anaemia: baseline results from the TWiTCH trial. British Journal of Haematology 2016;172(1):122–130. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 30.Wood JC, Zhang P, Rienhoff H, Abi-Saab W, Neufeld E. R2 and R2* are equally effective in evaluating chronic response to iron chelation. American Journal of Hematology 2014;89(5):505–508. [DOI] [PubMed] [Google Scholar]
- 31.Doyle EK, Thornton S, Toy KA, Powell AJ, Wood JC. Improving CPMG liver iron estimates with a -corrected proton density estimator. Magnetic Resonance in Medicine [accessed 2021 Aug 26];n/a(n/a). [DOI] [PMC free article] [PubMed] [Google Scholar]
Associated Data
This section collects any data citations, data availability statements, or supplementary materials included in this article.
Supplementary Materials
Data Availability Statement
Code and data relating to this work is available in the following git repo: