Skip to main content
PLOS One logoLink to PLOS One
. 2022 Mar 24;17(3):e0260855. doi: 10.1371/journal.pone.0260855

A novel autotransfusion device saving erythrocytes and platelets used in a 72 h survival swine model of surgically induced controlled blood loss

Kévin Schreiber 1,#, Benoit Decouture 2,#, Audrey Lafragette 1,#, Stéphane Chollet 2,, Marine Bruneau 2,, Maxence Nicollet 2,, Catherine Wittmann 2,, Francis Gadrat 2,, Alexandre Mansour 3,, Patricia Forest-Villegas 2,, Olivier Gauthier 1,4,#, Gwenola Touzot-Jourde 1,4,*,#
Editor: Ahmet Emre Eşkazan5
PMCID: PMC8947136  PMID: 35324911

Abstract

Background

The purpose of this study was to develop a swine model of surgically induced blood loss to evaluate the performances of a new autotransfusion system allowing red blood cells and platelets preservation while collecting, washing and concentrating hemorrhagic blood intraoperatively.

Methods

Two types of surgically induced blood loss were used in 12 minipigs to assess system performance and potential animal complications following autotransfusion: a cardiac model (cardiopulmonary bypass) and a visceral model (induced splenic bleeding). Animal clinical and hematological parameters were evaluated at different time-points from before bleeding to the end of a 72-hour post-transfusion period and followed by a post-mortem examination. System performances were evaluated by qualitative and quantitative parameters.

Results

All animals that received the autotransfusion survived. Minimal variations were seen on the red blood cell count, hemoglobin, hematocrit at the different sampling times. Coagulation tests failed to show any hypo or hypercoagulable state. Gross and histologic examination didn’t reveal any thrombotic lesions. Performance parameters exceeded set objectives in both models: heparin clearance (≥ 90%), final heparin concentration (≤ 0.5 IU/mL), free hemoglobin washout (≥ 90%) and hematocrit (between 45% and 65%). The device treatment rate of diluted blood was over 80 mL/min.

Conclusions

In the present study, both animal models succeeded in reproducing clinical conditions of perioperative cardiac and non-cardiac blood loss. Sufficient blood was collected to allow evaluation of autotransfusion effects on animals and to demonstrate the system performance by evaluating its capacity to collect, wash and concentrate red blood cells and platelets. Reinfusion of the treated blood, containing not only concentrated red blood cells but also platelets, did not lead to any postoperative adverse nor thrombogenic events. Clinical and comparative studies need to be conducted to confirm the clinical benefit of platelet reinfusion.

Introduction

Allogenic blood transfusions have a global acceptable safety profile in countries with well-developed blood banking systems, but are associated with the transmission of infectious disease, transfusion reactions, and immunosuppression [14]. Limited availability of donors and high processing costs place further constraints on the use of allogenic transfusions [58]. Consequently, safe and cost-efficient alternatives to allogenic transfusions are highly requested by health care providers and anesthesiologists for the patient’s benefit [9, 10].

Autotransfusion devices allow collection of shed blood during a surgical procedure and represent an attractive method for reducing allogenic blood transfusion. Benefits of autotransfusion have been demonstrated in multiple surgical contexts [11, 12]. Autotransfusion devices play an important role in patient blood management and are recommended by international guidelines [13, 14]. More precisely, the use of autotransfusion devices has been shown to reduce the need for allogenic blood transfusions in surgical procedures (cardiac, orthopedic, visceral, vascular) by 38% (RR 0.62; 95% CI 0.55–0.70) [11]. Once a minimal blood volume of 400–600 mL has been collected from the surgical site, it can be filtered, washed, concentrated, and re-transfused to the patient [15]. Currently available cell salvage devices process the collected blood mainly by centrifugation and produce packed red blood cells (RBCs) with a satisfactory safety profile for the average patient. However, centrifugation-based processing has been questioned due to damaged blood cells contributing to adverse outcomes in critical patients such as immunocompromised or neonatal patients [16]. Moreover, the process usually removes platelets and coagulation factors when used with standard programs intraoperatively and, when treating large volumes, can result in a dilutional coagulopathy [1719].

Platelets are critical regulators of hemostasis, serving as the matrix for initial vascular plug formation, creating a scaffold for the generation of fibrin clots, and releasing pro-wound-healing cytokines and procoagulant microparticles [20, 21]. Platelet concentrate reinfusion has been largely used in order to prevent or to treat perioperative hemorrhage in thrombocytopenic patients but also in combination with red blood cells and therapeutic plasmas for cases of massive bleeding due to trauma and surgical or obstetrical procedures [22, 23]. Many studies have been conducted to evaluate efficient treatments to quickly reach hemostasis and hence improve patient recovery and survival chances. In the past ten years, studies showed a regained interest for whole blood transfusion, specifically in cases of hemorrhagic syndromes [24]. As such, results obtained from the military field demonstrate the advantages of bringing as early as possible all factors, including platelets in combination with RBCs [25, 26]. Current recommendations state that platelet transfusion should be performed as soon as possible, even if it needs to be repeated in case of persistent bleeding, and in association with RBC concentrate transfusion as soon as necessary [27]. The need for efficient blood salvage processing and autotransfusion of concentrates comprised of both RBCs and platelets has been identified as well as the necessity to improve current technology and develop new technology [28, 29].

The novel autotransfusion device evaluated in this study is a medical device intended to collect and wash intra-operatively collected blood to obtain a concentrated blood product at the end of the process. The device integrates an innovative filtration technology through hollow fibers, allowing RBCs concentration but also platelets within the concentrated blood product and contaminant elimination. A preliminary in vitro study demonstrated the ability to efficiently wash and concentrate red blood cells, white blood cells and platelets without significant impact on cell integrity and function [30]. The dual objective of this study was first to develop a swine surgical model of controlled blood loss with a 72-hour post-operative follow-up, and then to evaluate the use of the autotransfusion device in this swine model. Device output, salvaged blood characteristics and a 72-hour post transfusion follow-up of the animals were used to assess suitability and performance.

Materials and methods

Autotransfusion device and study design

The autotransfusion system tested is a medical device (Same™, i-SEP, France) consisting of reusable equipment and disposable consumables. The reusable equipment is composed of a roller stand supporting a structure integrating the electrical, electronic, computer and mechanical elements allowing the device to treat the blood. Consumables consists in a suction and anticoagulation line, a blood collection reservoir and a treatment set. The suction line allows collection and anticoagulation of the blood obtained from the surgical site. A rough filtration before blood storage is performed in the blood collection reservoir. The treatment set consists mainly of tubings, a hollow fiber cartridge to separate blood cells from plasma, a blood treatment bag that ensures blood washing, a waste bag collecting plasma and contaminants, and a reinfusion bag to store filtered, washed and concentrated blood. A more detailed processing is described elsewhere [30].

The study was designed to develop two surgical models of controlled blood loss (abdominal and cardiac). The abdominal visceral model consisted of a splenic bleeding by surgically-induced capsule injuries representative of non-cardiac bleeding with aspiration of shed blood partially coagulated. Conversely, during the cardiac procedure, the blood was highly anticoagulated and collected from the right lateral thoracotomy surgical site and from the cardio-pulmonary bypass venous reservoir after ending an hour-long extracorporeal circulation. All collected blood were treated with the device to test the performance and evaluate a 72 h animal survival and post-mortem findings after autotransfusion. Taking into account the 3R recommendations, it was estimated that 4 to 6 animals per surgical model would be necessary to account for individual animal variation as well as surgical technique refinement and animal management optimization. Due to the early stage of the device testing on animal recovery, no randomization was used and the animal subjects were allocated to the cardiac group depending on their thoracic conformation that would facilitate the cardiac surgical approach by right lateral thoracotomy.

Animals

The animal study protocol received approval from the regional ethical committee for animal experimentation and was authorized by the French Ministry of Higher Education, Research and Innovation (Apafis number #11079-2017071317574824v1). Twelve adult female Yucatan mini-pigs were included in the study. The animal management, anesthesia and analgesia protocol as well as perioperative care, euthanasia and post-mortem examination are described in details in the S1 Appendix in S1 File. Briefly, after a week of acclimatization in an enriched environment, the animals were anesthetized by intramuscular injection for instrumentation, surgically induced controlled bleeding and autotransfusion. During anesthesia maintained with a balanced technique, controlled ventilation was instituted and jugular access was gained to place a central line for repeated blood sampling. Continuous monitoring during anesthesia included ECG, blood pressure measurement indirectly during instrumentation and invasively during the surgically induced bleeding and the following transfusion, pulse oximetry, inspired and expired gas analysis, arterial blood gases and body temperature. Anesthesia depth and intravenous fluid replacement were adjusted to the animal response to acute blood loss. Postoperative care comprising analgesia and supportive therapy was adapted to surgical models and each individual recovery during a 72 h follow-up post-transfusion. At the end of the period, general anesthesia was reinduced before euthanasia. Post-mortem examination was geared to assess thrombogenic risk according to the ISO standard 10933–4 [31] and the application of the FDA guide: Use of International Standard ISO 10993–1, "Biological evaluation of medical devices—Part 1: Evaluation and testing within a risk management process", section thrombogenicity, published in June 2016 and updated in 2021 [32].

Surgical models of controlled blood loss

Abdominal visceral model of controlled blood loss

A 20 cm long midline laparotomy incision was performed to exteriorize the spleen from the abdominal cavity. To counteract known porcine hypercoagulability [3335] and to facilitate blood loss and collection, a low dose of heparin (25 IU/kg IV) was administered just before initiating the splenic injury and repeated if needed during bleeding. This dosage has been used in previous studies of porcine intra-abdominal hemorrhage and has shown to bring the porcine coagulation profile within values found in humans [36]. ACT measurements (Activated Clotting Time, via Medtronic ACT II Coagulation Timer) were done before heparin administration, repeated during the blood collection time (target ACT 90–130 s) and just before transfusion to ensure a return to baseline ACT value.

Multiple lesions were created into the splenic parenchyma using digitoclastic technique and an additional lesion was done by severing the splenic vein. Spontaneous bleeding was let to drip into the abdominal cavity and then aspirated through the suction line (depression kept under 200 mbar to minimize hemolysis) [37] into the autotransfusion device blood collection reservoir. After reaching a 900 to 1100 ml volume of a blood and anticoagulation solution mixture in the collection reservoir, corresponding to a minimal of 20% of the estimated animal blood volume (61–68 ml/kg) [38], the bleeding was controlled by removal of the spleen using a tissue fusion device (Atlas Ligasure™ forceps 10 mm, Covidien, Medtronic). The surgical wound was sutured in different layers. A protective bandage consisting of an adhesive padded tape (Animal Polster, Snogg, Norway) was placed to cover the incision.

Cardiac model of controlled blood loss

A right lateral thoracotomy was performed in the third intercostal space. Cardio-pulmonary bypass (CPB) was set in place by cannulation of the right atrium and the ascending aorta. Extra-corporeal circulation (priming volume of 1250 mL of Ringer Lactate solution) was then initiated and maintained for an hour. Classic anticoagulation with heparin (5000 IU in the priming volume, 300 IU/kg IV to the animal) and ACT monitoring (> 400 s) were used. At the end of the bypass, adequate animal volemia was restored based on hemodynamic monitoring data. CBP cannulas were removed and cannulated vessels sutured. Heparin effect reversal was achieved with a slow IV infusion of protamine (1 mg/100 IU of heparin used during surgery). To allow elimination of residual air and fluids from the thoracic cavity, a chest tube was placed at the end of the procedure and connected to a low negative pressure drainage system. The surgical wound was closed in layers. A wound catheter was left in place to allow instillation of local anesthetics. A protective bandage consisting of an adhesive padded tape (Animal Poltser, Snogg, Norway) was placed to cover the incision. Hemorrhagic blood aspirated from the chest cavity (suction pressure kept under 200 mbar to minimize hemolysis) and blood remaining in the CPB circuit after stopping the extra-corporeal circulation were collected to be treated by the autotransfusion device (minimal volume of 700 mL).

Laboratory testing

Blood samples were collected from before surgery to the end of the 72-hour follow up to monitor possible variations in hematologic and coagulation parameters (Fig 1).

Fig 1. Schematic description of the study design.

Fig 1

The different times of sampling are during anesthesia before blood loss (T0), in the blood collection reservoir before the treatment (TR*), in the blood bag after the treatment (TB*), after autologous transfusion at the end of anesthesia (TP), then during the 72 h survival follow-up period (7 sampling times following the end of the transfusion: T2-6h, T6-12h, T12-24h, T24-36h, T36-48h, T48-60h and T60-72h). CPB: Cardio Pulmonary Bypass; CBC: Complete blood count. *TR and TB samplings were taken from the blood collection reservoir and the transfusion blood bag and not sampled directly in the animal bloodstream.

Times of sampling were before surgery once the central venous catheter was in place (T0), at the end of the transfusion (TP), then postoperatively between 2 h and 6 h (T2-6h), 6 h and 12 h (T6-12h), then every 12 hours until euthanasia at 72 h post-transfusion (T12-24h, T24-36h, T36-48h, T48-60h, T60-72h). Total blood volume collected was refined by establishing the minimal volume needed for laboratory testing ahead of the experiment. All blood samplings were performed by gentle aspiration through the central venous catheter which allowed stress-free repetitive interventions on the minipigs post-operatively. Additional blood samples were obtained from the blood collection reservoir (TR) and from the transfusion bag (TB) before reinfusion started. Blood analyses consisted of complete blood count (CBC) including RBC, WBC and platelet count, hematocrit and total hemoglobin (Procyte Dx Hematology Analyzer, IDEXX, Hoofddorp, The Netherlands), plasma free hemoglobin concentration measurement for the calculation of the hemolysis rate (Plasma/Low Hb photometer, HemoCue AB, Ängelholm, Sweden), plasma heparin concentration measurement through anti-Xa activity assay (Heparin standard-HNF, HemosIL Liquid anti-Xa, ACLTOP, IL, Werfen, Le Pré-Saint-Gervais, France) and ROTEM tests to evaluate coagulation status and platelet aggregation (ROTEM delta and ROTEM platelet, Werfen, Le Pré-Saint-Gervais, France).

Analyses of the coagulation status with ROTEM tests included EXTEM, INTEM and HEPTEM tests. HEPTEM is the same test as the INTEM but with additional heparinase. Both tests were performed to take into account the fact that heparin was administered to all animals during surgery and was used for the suction line anticoagulation. Analyzed parameters were clotting time (CT), amplitude at 30 minutes (A30) and maximum clotting firmness (MCF). Analyses of the platelet aggregations were performed with the ROTEM platelet module. Platelet activation was induced by the ADP agonist provided by Werfen.

Autotransfusion device evaluation

Qualitative and quantitative parameters were used to evaluate the performance of the autotransfusion system in washing and concentrating collected blood. Quality of the treated blood was assessed by measuring heparin and free hemoglobin concentrations and calculating their clearance from the washed blood. As defined by the American Association Blood Bank (AABB) [39], the threshold concentration of residual heparin in the concentrated blood is ≤ 0.5 IU/mL and the washout is expected to be ≥ 90%. Regarding hemolysis, the AABB states that the residual free hemoglobin concentrations may be quite high (5–10 g/L) despite greater than 90% removal (AABB recommendations). Washout, expressed as component clearance (%), was calculated with the following formula:

Heparinclearance%=InitialquantityofHeparinFinalquantityofHeparinInitialquantityofHeparinx100

Regarding free hemoglobin, a formula was adapted taking into account the free hemoglobin production during the treatment. The calculation is based on the eliminated free hemoglobin in the waste:

Freehemoglobinwashout%=FreehbinthewasteFreehbinthewaste+Freehbintheconcentratedbloodx100

Eliminated free hemoglobin (free hemoglobin in the waste bag) is calculated by the difference between the total amount of hemoglobin measured in the blood after the treatment compared to the total amount of hemoglobin measured in the blood before the treatment. Hemolysis rate is the percentage of the hemoglobin amount present in the extracellular medium (free hemoglobin) compared to the total amount of hemoglobin in the whole blood. The extracellular medium volume is calculated by subtracting the hematocrit percentage from the whole blood volume (100%). The formula to calculate hemolysis with the appropriate correction is therefore [40]:

Hemolysisrate=100HematocritxFreehemoglobinTotalhemoglobin

Quantitative performance parameters used to describe the treated blood product included cell yields (RBCs, Platelets and WBCs), the final hematocrit and the final hemoglobin concentration of the blood product ready to be transfused. Cell yields are calculated with the following formula:

Cellyield%=InitialvolumecontainingcellsxInitialconcentrationofcellsFinalvolumecontainingcellsxFinalconcentrationofcellsx100

Duration of a treatment cycle was used to characterize the technical efficacy of the device. Expected device performance was to obtain a heparin clearance ≥ 90% and a residual heparin concentration ≤ 0.5 IU/mL, a free hemoglobin washout ≥ 90%, a RBC yield ≥ 80%, a final hematocrit between 45% and 65% and a treatment rate ≥ 80 mL/min.

Statistical analysis

Due to the observational nature of the study on small animal samples, the results are reported by descriptive statistics, expressed as individual value or as median (min-max) for each surgical model. Data from the visceral model are represented with circles while data from the cardiac model are represented as squares. Each series of points represent a unique animal on figures.

Results

Data on animals

Twelve one year-old female Yucatan mini-pigs were included in the study with a median (min-max) weight of 46.9 (45.75–48.25) kg in the visceral model (n = 5) and 48 (41–53) kg in the cardiac model (n = 7). Each animal was individually identified by an ear tag (since a few days after birth) to guarantee traceability.

Three animals in the cardiac model were excluded due to premature death before administration of the collected and treated blood. Two minipigs died despite resuscitation attempts from ventricular fibrillation during the thoracotomy, one before the bypass started and one at the time of aortic cannula removal, and the third mini-pig had a fatal anaphylactic reaction during slow protamine infusion. All animals that underwent the complete surgical procedure and autotransfusion survived and were euthanized at the end of a 69 (68–71) h follow-up period after transfusion completion. Recovery from the visceral procedure was uneventful, mini-pigs stood up within the first two hours following the end of anesthesia, had all resumed eating and drinking at the time of the 6h post-transfusion clinical examination. No rescue analgesia was needed. In the cardiac model, nasal oxygen supplementation was needed for the first 8 to 18 hours of recovery and opioid analgesia could be stopped after the 24 to 36 hours of recovery. Mini-pigs stood up between 14 and 20 h post-transfusion and resumed eating shortly thereafter (0 to 4 h from standing).

Post-mortem examination in the visceral model showed gross and histological abnormalities that included a slight fibrino-congestive diffuse peritonitis in one animal and a slight focal perihepatitis in two animals, a hemorrhagic mediastinal lymph node in one animal, a renal infarct scar in two animals. Signs of inflammation and hemorrhage were considered to be results of the surgical trauma and the renal lesion was of chronic nature dating the injury well prior the surgery. In the cardiac model, lesions at the surgical site were identified in all animals: localized myocarditis, endocarditis and pericarditis at level of the right atrium, pericarditis in the left ventricle and right-sided pleuritis. One cardiac minipig had a multifocal granulomatous pneumonia associated with identification of inhaled vegetal particles. Gross and histologic examination failed in both models to reveal any thrombotic lesions consecutive to the autotransfusion.

Laboratory testing

The complete blood count and ROTEM results are detailed in the (S1 and S2 Tables in S1 File respectively).

RBC counts, hemoglobin and hematocrit

In the visceral model, red blood cell (RBC) count, hemoglobin and hematocrit tended to be lower at the T0 sample (during anesthesia and before surgery) compared to the TP sample taken at the end of transfusion (RBC count: 4.6 [4.3–6.1] 106/μL for T0 vs 5.4 [5.3–5.8] 106/μL for TP; Hemoglobin in g/L: 93 [85–119] for T0 vs 111 [105–114] for TP; Hematocrit in %: 29 [2739] for T0 vs 35 [3438] for TP). During follow-up, no major variation of the RBC counts, the hemoglobin concentration or the hematocrit was noticed (Fig 2, panel A, B and C and S1 Table in S1 File).

Fig 2. RBC count, hemoglobin and hematocrit animal values in the visceral model (n = 5).

Fig 2

RBC count (A), hemoglobin (B) and hematocrit (C) are represented for each individual animal. The different times of sampling are during anesthesia before blood loss (T0) and after autologous transfusion at the end of anesthesia (TP), then during the 72 h survival follow-up period (7 sampling times following the end of the transfusion).

In the cardiac model, minimal variations were seen on the RBC count, hemoglobin and hematocrit between T0 and TP samples (RBC count: 4.6 [4.3–5.1] 106/μL at T0 vs 4.3 [4.1–4.5] 106/μL at TP; Hemoglobin in g/L: 94.5 [90–105] at T0 vs 88.5 [85–93] at TP; Hematocrit in %: 30 [2933] at T0 vs 28 [2630] at TP) (Fig 3, panel A, B and C and S1 Table in S1 File). During follow-up, no major variation of red blood cell count, hemoglobin concentration nor hematocrit was noticed.

Fig 3. RBC count, hemoglobin and hematocrit animal values in the cardiac model (n = 4).

Fig 3

RBC count (A), hemoglobin (B) and hematocrit (C) are represented for each individual animal. The different times of sampling are during anesthesia before blood loss (T0), after autologous transfusion at the end of anesthesia (TP), then during the 72 h survival follow-up period (7 sampling times following the end of the transfusion).

WBC counts

All the minipigs in both surgical models developed a neutrophilic leukocytosis while under anesthesia or in the early phase of recovery with a peak between 2 and 12 hours after the end of transfusion lasting for 24 hours (22.7 [15.8–27.9] 106/μL at T2-6h and 28.2 [19.6–28.7] 106/μL at T6-12h for the visceral model and 21.5 [16.0–29.5] 106/μL at T2-6 and 26.3 [16.8–35.0] at T6-12h for the cardiac model) (Fig 4A and 4B; complete WBC values in S1 File).

Fig 4. WBC and neutrophil counts in the visceral model (A, n = 5) and the cardiac model (B, n = 4).

Fig 4

Each line represents one animal. The different times of sampling are during anesthesia before blood loss (T0), after autologous transfusion at the end of anesthesia (TP), then during the 72 h survival follow-up period (7 sampling times following the end of the transfusion).

Platelet counts

In both models, platelet counts decreased in post-operative blood samples. The nadir was reached between 6 and 12 hours post-operatively in both models. Thereafter, platelet count increased until a return close to normal count at 72 h post-operatively (see Fig 5 and S1 Table in S1 File)

Fig 5. Animal platelet counts in the visceral model (A, n = 4) and the cardiac model (B, n = 4).

Fig 5

Each line represents one animal. The different times of sampling are during anesthesia before blood loss (T0), after autologous transfusion at the end of anesthesia (TP), then during the 72 h survival follow-up period (7 sampling times following the end of the transfusion). One animal in the visceral model was excluded for the platelet count due to platelet aggregation verified on the blood smear and failure of the hematological analyzer.

Coagulation control in the visceral model

In the visceral model, baseline ACT were comprised between 69 and 96 s (median of 73 s) and increased between 102 and 133 s (median of 109 s) at 15 minutes post-heparin injection.

Coagulation evaluation by ROTEM

EXTEM and HEPTEM results are displayed in Tables 1 and 2 while INTEM results are to be found in S2 Table in S1 File. In the visceral model, EXTEM results showed a stable clotting time during the follow-up. In the cardiac model, EXTEM clotting time was more variable and tended to be slightly increased during the follow-up compared to before the surgery (T0). No clotting time exceeded 100 s. The HEPTEM results showed more variability in the clotting time that tended to increase slightly during the follow-up compared to before the surgery (T0; Table 2). The other parameters (A30, MCF and Alpha angle) showed minimal variations.

Table 1. ROTEM-EXTEM parameters during anesthesia and the 72 h survival follow-up period (9 sampling times).
Parameters/Time Surgical Models T0 TP T2-6h T6-12h T12-24h T24-36h T36-48h T48-60h T60-72h
EXTEM CT (s) VISC 56 (49–65) 59 (56–72) 62 (44–77) 62 (46–67) 65 (55–78) 62 (53–89) 75 (51–83) 79 (52–89) 67 (54–71)
CARD 56.5 (53–61) 51 (44–61) 81.5 (55–97) 70.5 (67–83) 61 (40–85) 71.5 (68–77) 62.5 (50–87) 70 (68–74) 85 (78–88)
A30 (mm) VISC 73 (70–76) 67 (61–69) 69 (63–69) 69 (63–70) 72 (71–74) 75 (70–75) 75 (63–76) 76 (73–79) 78 (72–80)
CARD 71.5 (67–81) 72 (68–75) 71 (68–73) 71.5 (67–73) 74 (72–77) 77 (74–79) 76.5 (74–81) 79.5 (74–83) 79.5 (61–83)
MCF
(mm)
VISC 74 (71–76) 68 (62–73) 70 (64–72) 71 (64–71) 75 (74–76) 77 (73–78) 78 (66–79) 79 (76–81) 80 (75–82)
CARD 73 (69–82) 72.5 (68–76) 71.5 (70–74) 73 (68–74) 76 (74–78) 78.5 (76–81) 78 (77–82) 81 (76–84) 80.5 (63–85)
Alpha Angle (°) VISC 79 (78–81) 77 (74–80) 77 (76–78) 76 (72–77) 77 (76–78) 78 (74–79) 78 (75–78) 78 (77–81) 80 (78–80)
CARD 78 (75–81) 76.5 (74–80) 76.5 (75–79) 77 (74–78) 77 (76–79) 77 (77–78) 77 (75–78) 77 (76–81) 78 (67–80)

Results expressed as median (min-max) for the visceral (VISC, n = 5) and cardiac models (CARD, n = 4)

The different times of sampling are during anesthesia before blood loss (T0), after autologous transfusion at the end of anesthesia (TP), then during the 72 h survival follow-up period (7 sampling times following the end of the transfusion). CT: clotting time, A30: amplitude at 30 minutes, MCF: maximum clotting firmness.

Table 2. ROTEM-HEPTEM parameters during anesthesia and the 72 h survival follow-up period (9 sampling times).
Parameters/Time Surgical Models T0 TP T2-6h T6-12h T12-24h T24-36h T36-48h T48-60h T60-72h
HEPTEM CT (s) VISC 147 (141–157) 154 (146–178) 132 (114–161) 141 (104–162) 158 (132–167) 168 (154–201) 181 (133–222) 177 (171–218) 164 (118–187)
CARD 189 (119–218) 170 (149–245) 171.5 (153–188) 146 (106–192) 148.5 (133–182) 148.5 (138–199) 171 (131–190) 187.5 (157–199) 195 (144–217)
A30 (mm) VISC 66 (65–67) 62 (58–63) 60 (58–63) 62 (58–64) 64 (63–65) 66.5 (65–68) 60 (48–65) 67 (64–71) 67 (66–72)
CARD 62.5 (46–68) 62.5 (61–68) 65.5 (63–67) 66 (64–67) 67 (45–68) 70 (70–70) 72.5 (69–79) 72.5 (69–73) 73 (67–74)
MCF (mm) VISC 69 (67–71) 64 (60–66) 65 (60–70) 65 (63–70) 70 (68–71) 72.5 (71–74) 70 (55–73) 72 (70–75) 75 (72–77)
CARD 66.5 (48–71) 66 (63–70) 68.5 (66–69) 69.5 (68–71) 72 (49–72) 75 (74–75) 77 (74–82) 76.5 (74–77) 78 (73–79)
Alpha Angle
(°)
VISC 81 (78–81) 78 (77–80) 78 (76–80) 77 (76–80) 78 (77–79) 77.5 (75–79) 75 (65–78) 76 (75–81) 79 (74–81)
CARD 78.5 (69–80) 78 (76–81) 79 (77–79) 77.5 (76–79) 78 (65–79) 79 (76–79) 79 (77–79) 78 (78–81) 79 (79–80)

Results expressed as median (min-max) for the visceral (VISC, n = 5) and cardiac models (CARD, n = 4)

The different times of sampling are during anesthesia before blood loss (T0), after autologous transfusion at the end of anesthesia (TP), then during the 72 h survival follow-up period (7 sampling times following the end of the transfusion).

CT: clotting time, A30: amplitude at 30 minutes, MCF: maximum clotting firmness

Additional ROTEM (INTEM) results are in the (S2 Table in S1 File).

ADP-induced platelet aggregation

Whole blood platelet aggregation was induced by ADP and measured with the ROTEM platelet module. Results showed that AUC related to ADP-induced platelet aggregation was not modified at the post-operative sample time (TP) or during the follow-up (from 2–6 h to 60–72 h) compared to the pre-operative value (T0), in both visceral and cardiac models (Table 3).

Table 3. Platelet aggregometry results.
Parameters/Time Surgical Models T0 TP T2-6h T6-12h T12-24h T24-36h T36-48h T48-60h T60-72h
ROTEM Delta ADP-induced aggregation Amplitude 6 minutes (ohm) VISC 24 (10–32) 16 (14–23) 20 (12–30) 25 (15–34) 22 (10–34) 14 (8–23) 19 (12–29) 17 (7–28) 23.5 (11–28)
CARD 23 (17–25) 15.5 (8–32) 20.5 (18–24) 19.5 (18–27) 14.5 (12–15) 12 (7–15) 20 (9–24) 12 (5–18) 19 (14–28)
AUC (A.U.) VISC 99 (40–130) 72 (60–94) 87 (54–119) 100 (64–136) 86 (45–135) 89 (45–97) 84 (70–122) 98 (55–121) 111.5 (50–119)
CARD 94 (77–106) 71.5 (42–130) 88.5 (80–105) 79.5 (79–115) 60.5 (55–64) 59.5 (45–90) 96 (73–102) 61 (52–76) 121 (82–122)

Aggregation amplitude at 6 min (ohm) and area under the curve in arbitrary unit (AUC, A.U.) are expressed as median (min-max) for the visceral (VISC, n = 5) and cardiac models (CARD, n = 4). The different times of sampling are during anesthesia before blood loss (T0), after autologous transfusion at the end of anesthesia (TP), then during the 72 h survival follow-up period (7 sampling times following the end of the transfusion).

Controlled blood loss and autotransfusion device evaluation

The splenic surgical injury resulted in an active hemorrhage and a blood collection in a median time of 31 (22–51) min. End volumes in the collection reservoir ranged between 925 and 1100 mL (median 1000 mL) in the visceral model corresponding to an estimated blood loss of 600 to 900 mL representing a minimum of 20% of the animal blood volume. Volume in the collection reservoir after the end of the bypass was comprised between 900 and 1425 mL (median 1125 mL) in the cardiac model. The hematocrit of the collected blood in the reservoir (TR) for the visceral and the cardiac model was respectively 24 (16–28) % and 19 (16–21) %, with a hemolysis rate of 2.1 (0.6–2.4) % and 0.2 (0.1–0.5) %.

The autotransfusion system performed a complete treatment cycle of 670 (670–755) mL in 5 min 54 s (5 min 27 s– 6 min 05 s) for the visceral group and 708 (670–755) mL in 6 min 10 s (4 min 0 s– 6 min 25 s) for the cardiac group. These treatment timings correspond to treatment rates at 121 (110–139) mL/min for the visceral group and 114 (106–189) mL/min for the cardiac group. Transfusion bags contained 229.4 (163.2–311.7) g of RBC solubilized in isotonic sodium chloride with a median hematocrit of 50 (45–60) %. Treated blood was transfused back to the animals in 23 (11–36) min in the visceral model and 29 (25–29) min in the cardiac model. The blood treatment was evaluated according to qualitative and quantitative parameters (Table 4). The quality of the reinfused blood was evaluated by measuring the concentration of unwanted substances, in particular heparin. Results show that the heparin concentration was always below 0.5 IU/mL except for one cardiac reinfused blood which was at 0.6 IU/mL (0.14 [0.00–0.36] IU/mL for the reinfused blood from visceral model and 0.46 [0.44–0.60] IU/mL for the reinfused blood from the cardiac model). Heparin concentration in the blood before treatment was at 4.5 [3.4–6.8] IU/mL in the visceral model and 9.6 [8.5–11.5] IU/mL in the cardiac model. Qualitative and quantitative performance parameters (heparin clearance, free hemoglobin washout, hematocrit and hemoglobin concentration) exceeded set objectives, except for RBC yield and hemolysis rate (Table 4). Red blood cell yield was 61.7 (50.1–71.9) % for the visceral model and 70.5 (65.4–72.1) % for the cardiac model.

Table 4. Qualitative and quantitative assessment of the autotransfusion device.

Parameters Visceral model
N = 5
Cardiac model
N = 4
Criteria
Quantification of blood elements and quality of collected blood before treatment (TR)
Red blood cell concentration (106/μL) 3.9 (2.6–4.6) 3.0 (2.5–3.3) NA
Hemoglobin concentration (g/L) 80 (52–92) 61 (51–67) NA
Hematocrit (%) 24 (16–28) 19 (16–21) NA
Hemolysis rate (%) 2.1 (0.6–2.4) 0.2 (0.1–0.5) NA
Platelet concentration (103/μL) 54 (16–83) 157 (115–192) NA
Quality of the blood to be reinfused (TB)–unwanted substances
Anticoagulant (heparin) concentration (IU/mL) 0.14 (0.00–0.36) 0.46 (0.44–0.60) ≤ 0.5
Anticoagulant clearance (%) 97.3 (93.6–100.0) 98.3 (95.6–99.0) ≥ 90%
Hemolysis rate (%) 1.2 (1.0–2.0) 1.4 (1.4–1.6) NA
Free hemoglobin washout (%) 94 (88–98) 90 (87–94) ≥ 90%
Quantification of blood elements to transfuse (TB) and device performance
Red blood cell concentration (106/μL) 7.5 (7.0–8.9) 7.2 (6.8–7.2) NA
Red Blood Cells Yield (%) 61.7 (50.1–71.9) 70.5 (65.4–72.1) ≥ 80%
Platelet concentration (103/μL) 109 (64–147) 305 (258–337) NA
Platelet yield (%) 74.4 (40.3–150.7) 58.2 (49.1–62.2) ≥ 40%
Final Hematocrit (%) 50 (45–60) 49 (45–50) 45% < value > 65%
Total hemoglobin concentration (g/L) 155 (141–189) 152 (139–159) NA
White Blood Cells yield (%) 69 (59–77) 83 (72–88) NA
Treatment rate (mL/min) 121 (110–139) 114 (106–189) ≥ 80 mL/min

Results expressed as median (min-max).

Platelet concentration was 54 (16–83) 103/μL and 157 (115–192) 103/μL in the collection reservoir (TR) and 109 (64–147) 103/μL and 305 (258–337) 103/μL in the transfusion bag, for the visceral and the cardiac model respectively. Platelet yield was 74.4 (40.3–150.7) % and 58.2 (49.1–62.2) % for the visceral and the cardiac model respectively (Table 4).

Coagulation properties of the blood from the collecting reservoir and in the transfusion bag were evaluated by ROTEM assays. EXTEM tests show that the clotting time was considerably increased in the transfusion bag compared to before filtration and washing for both models (for the visceral model, 56 [49–65] s before treatment vs 517 [291–3600] s after the treatment and for the cardiac model, 56.5 [5361] s before the treatment vs 1769 [119–2277] s after). Such as on the animal blood, HEPTEM results show variability in the clotting time and tended to increase after treatment compared to before treatment (147 [141–157] s before vs 354 [274–3600] s after treatment in the visceral model and 189 [119–218] before vs 986 [174–2085] s after treatment in the cardiac model).

Discussion

The aim of this study was to develop a minipig animal model of surgically induced hemorrhage reproducing the clinical conditions of perioperative blood loss and evaluate the performance of a new autotransfusion device allowing the concentration of not only red blood cells but also platelets using an innovative filtration technology. No clinical complication nor bleeding following the reinfusion and recovery from anesthesia were observed on the animals during the 72 h postoperative period. Device performance fulfilled the criteria that were defined either from regulatory limits, in particular for anticoagulant elimination and maximum acceptable hemolysis in the treated blood, or from scientific consensus in terms of performance on devices already available [39].

Several hemorrhagic models on different animal species have been used to evaluate the safety and performance of autotransfusion devices. Hofbauer et al. [41] described a canine model in which whole blood was collected on anticoagulated bags for further processing with an automated cell salvage device. Treated blood was not reinfused to the animals. Total recovery of RBCs was 80 ± 12%, but 70% of the platelets were washed out, while 57% of the leukocytes remained in the concentrated final product. Due to differences between canine and human blood [42] and for ethical reasons, this animal species was not considered as a potential model for the present study. Ovine and bovine models were not considered because ovine RBCs are smaller than human RBCs and data published using fresh bovine blood did not include any evaluation of in vivo reinfusion [4345].

In a porcine model, Vagianos et al. [46] evaluated hemorrhagic blood collection with citrated sponges instead of suction. Although the blood was not reinfused to the animals, they concluded that blood collection by means of surgical sponges may be a safe and an efficient method. Autotransfusion experiments in pigs were initially performed by reinfusion of a non-processed blood collected after a controlled hemorrhage that resulted in a 40% blood volume loss [47]. Animals were followed up for 48 hours without any mortality. The porcine model seems to be the most commonly used for blood transfusion studies [4850]. The minipig model (Sus Scrofa domesticus) has been extensively used in hemorrhagic models and cardiovascular research including testing medical devices [51, 52]. The Yucatan minipig has anatomical [53], physiological and blood characteristics comparable to those of humans, as well as a good homogeneity between the animals [54] which makes it a suitable model for in vivo biological evaluation of an autotransfusion device.

To reproduce as close as possible to the clinical setting, two surgical models were developed: (i) a cardiac model consisting of a cardiopulmonary bypass after a lateral thoracotomy and treatment of the remaining blood in the CPB circuit and (ii) an abdominal visceral model of surgically induced splenic hemorrhage and in situ blood aspiration from the surgical site. The present study confirmed that both models were relevant to evaluate the safety and the performance of the autotransfusion device not only perioperatively (thoracotomy, CPB implementation) but also postoperatively, with a 72 h follow-up after reinfusion.

Regarding the blood processing, the bleeding conditions in both models were adequate to collect enough blood to perform two cycle treatments by the device (first cycle volume of 700 mL and second of 500 mL). Both blood treatment and the subsequent reinfusion were performed in a very short time, compatible with the surgical time.

Animal clinical follow-up was limited to 72 hours +/- 4 hours postoperatively. This time covers the 24 hour period during which the acute complications related to transfusion are observed in the clinical practice [55]. A 72 h postoperative period was suitable to visualize the potential presence of large thrombi or organ infarction. These results are supported by the regression of the post-procedure inflammatory reaction seen through the decay of the neutrophilic leukocytosis after the first 24 hours.

The evaluated device demonstrated valuable performance related to its filtration technology. Heparin concentration in the treated blood was found below the regulatory threshold in both models (< 0.5 IU/mL) and heparin clearance was superior to 90% (97.4 ± 2.2%). Regarding the free hemoglobin removal, the AABB defined the threshold at 90% regardless of the free hemoglobin concentration [39]. In the present study, the removal of free hemoglobin was found to be superior to 90% (91.7 ± 3.5%), while the hemolysis rate in the treated blood was around 1.45%. Those data suggest that free hemoglobin was generated during the blood treatment but over 90% could be cleared. Hemolysis may also explain the lower removal performance of hemoglobin compared to the heparin washout (91.7% vs 97.4% for the free hemoglobin removal and the heparin washout, respectively). These results can be explained by the high hemolysis of porcine RBCs due to their fragility [56], while in human surgeries, hemolysis rate is about 1.3% [57]. The RBCs sensitivity to hemolysis has two consequences when compared to human blood treatment: hemolysis in the treated blood is higher and RBCs yield is lower. Indeed, in a preliminary in vitro study testing the system with human blood, RBC yield was over 88.1% and the final hemolysis was 0.12% [30]. A comparative study in swine with a system already available would allow confirmation of the fragility of porcine RBCs under these conditions and would rule out a specific impact of the i-SEP device. Hemolysis in the porcine model can be therefore considered as a worst-case scenario compared to human blood. The higher hemolysis rate in the visceral model compared to the cardiac model is explained by the blood shedding, the coagulation activation during abdominal bleeding and the suction conditions (2.07 [0.55–2.42] % for the visceral model and 0.15 [0.1–0.3] % for the cardiac model). In the cardiac model, the blood is collected directly from the extra-corporeal circulation and does not come into contact with the extravascular medium. Despite the hemolysis rate higher than the 0.8% threshold (as defined by the European guidelines [58]) and the RBCs yield below 80%, the treated blood reached the 45% to 65% hematocrit reference range and its reinfusion was well tolerated by the animals. No clinical sign which might have been related to the reinfusion of free hemoglobin was observed. However, the small number of animals included in the study does not allow conclusions to be drawn about the safety related to the reinfusion of free hemoglobin.

The i-SEP device is the first autotransfusion system which allows in the same time, automatically and in a time compatible with the surgery, to concentrate not only red blood cells but also platelets. Indeed, the obtained treatment rate was higher than 80 mL/min, which is faster than the rate encountered by machines treating blood per centrifugation, even when using the emergency mode [59], and allowed the treatment of 500 mL to 700 mL diluted shed blood in less than 6 minutes. Moreover, platelet yield was always greater than 40% and higher than that obtained in the preliminary in vitro study of i-SEP using human blood [30]. Platelet concentration yields with the i-SEP device is higher than the ones that are usually observed with devices using centrifugation as a concentration method in in vitro and in vivo studies [59]. Reinfused blood included 83.3 ± 44.9% and 56.9 ± 5.6% of platelets collected in the blood collection reservoir in the visceral and cardiac model respectively. As expected, the platelet yield obtained after the blood treatment in the cardiac model is lower than that observed in the visceral model. That is attributable to the known deleterious effect of cardiopulmonary bypass on platelets [6063] which therefore reduces the number of platelets to concentrate. In the visceral model, the presence of significant hemolysis in the collected blood produced red cells ghosts that may overestimate the platelet count. Some of the platelet yields were over 100% in the visceral model.

The reinfusion of platelets is considered as a potential prothrombotic risk. However, it was shown that the i-SEP autotransfusion system does not activate platelets during the treatment, and that platelets keep their ability to be activated after the treatment [30]. However, platelet activation was not evaluated in this animal study. Moreover, if platelets may be activated by the cardiopulmonary bypass in the cardiac model, the reinfusion of platelets during this study did not lead to any complication during the 72 h follow-up. No complication has been highlighted during the follow-up that could show a thrombogenic risk, in either model. After euthanasia, macroscopic examination of the different organs of interest and histological analysis did not reveal any evidence of thrombus formation. Moreover, even though the concentrated blood contains red blood cells and platelets, the coagulation power is annihilated because of the washing of coagulation factors in particular as described in a previous study [30] and as shown by the EXTEM and HEPTEM tests that demonstrated the inability of the treated blood to coagulate. However, the thrombotic risk was only evaluated during a postoperative period of 72 hours and therefore only the acute risks were assessed. Long-term follow-up on a larger number of individuals would allow conclusions to be drawn about safety, also regarding long-term risks.

The main limit of the study design was the descriptive nature of the results due to the limited number of animals per group that precluded any comparative statistical analysis. Moreover, the swine blood characteristics resulted in difficulties interpreting the hemolysis rate during blood treatment and the swine state of hypercoagulability required a partial anticoagulation with heparin in the visceral model that did not exactly replicate the clinical situation. As mentioned above, the evaluation of the thrombotic risk would be improved with a longer follow-up, a larger number of individuals in each group and a more complete hemostasis and inflammation laboratory testing (platelet function, inflammatory proteins, white blood cell activity). It was chosen to use a controlled hemorrhagic model with moderate blood loss (less than 30% of the total blood volume) that allowed blood reinfusion without a need of complementary plasma transfusion. Therefore, this study did not replicate a massive hemorrhagic situation.

Conclusions

This first in vivo study confirmed that this new autotransfusion system with an innovative filtration technology allowed to concentrate not only RBCs but also platelets, within a limited blood processing time. The concentrated blood did not have any pro-coagulant effect nor did cause any thrombotic effect after reinfusion to minipigs after cardiac or visceral bleeding. This autotransfusion device produced a concentrated blood ready for reinfusion in less than 6 minutes, with a high concentration of RBCs and platelets while removing efficiently heparin and free hemoglobin. Other studies should be implemented to compare the i-SEP device with other cell saver systems currently in use and evaluate clinical benefits of the transfusion of functional platelets during surgery.

Supporting information

S1 File

(DOCX)

Data Availability

All relevant data are within the manuscript and its Supporting Information files. Additional data may be ask directly to the corresponding author.

Funding Statement

The study was funded by BPi France (French Public Bank for investments) by the grant PSPC-402492 (BPI grants for structuring competitive research and development). The grant was awarded to the consortium Isep/Oniris/Université de Rennes. The funders had no role in study design, data collection and analysis, decision to publish, or preparation of the manuscript.

References

  • 1.Scott BH, Seifert FC, Grimson R. Blood transfusion is associated with increased resource utilisation, morbidity and mortality in cardiac surgery. Annals of Cardiac Anaesthesia. 2008;11: 15–19. doi: 10.4103/0971-9784.38444 [DOI] [PubMed] [Google Scholar]
  • 2.O’Brien SF, Yi QL, Fan W, Scalia V, Fearon MA, Allain JP. Current incidence and residual risk of HIV, HBV and HCV at Canadian Blood Services. Vox Sanguinis. 2012;103: 83–6. doi: 10.1111/j.1423-0410.2012.01584.x [DOI] [PubMed] [Google Scholar]
  • 3.Freedman J. The ONTraC Ontario program in blood conservation. Transfusion and apheresis science: official journal of the World Apheresis Association: official journal of the European Society for Haemapheresis. 2014;50: 32–6. doi: 10.1016/j.transci.2013.12.010 [DOI] [PubMed] [Google Scholar]
  • 4.Rohde JM, Dimcheff DE, Blumberg N, Saint S, Langa KM, Kuhn L, et al. Health care-associated infection after red blood cell transfusion: a systematic review and meta-analysis. JAMA. 2014;311: 1317–26. doi: 10.1001/jama.2014.2726 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 5.Tretiak R, Laupacis A, Rivière M, McKerracher K, Souêtre E. Cost of allogeneic and autologous blood transfusion in Canada. Canadian Cost of Transfusion Study Group. Canadian Medical Association Journal. 1996;154: 1501–1508. Available: /pmc/articles/PMC1487839/?report = abstract [PMC free article] [PubMed] [Google Scholar]
  • 6.Amin M, Fergusson D, Wilson K, Tinmouth A, Aziz A, Coyle D, et al. The societal unit cost of allogenic red blood cells and red blood cell transfusion in Canada. Transfusion. 2004;44: 1479–86. doi: 10.1111/j.1537-2995.2004.04065.x [DOI] [PubMed] [Google Scholar]
  • 7.Shander A, Hofmann A, Ozawa S, Theusinger OM, Gombotz H, Spahn DR. Activity-based costs of blood transfusions in surgical patients at four hospitals. Transfusion. 2010;50: 753–65. doi: 10.1111/j.1537-2995.2009.02518.x [DOI] [PubMed] [Google Scholar]
  • 8.Meybohm P, Straub N, Füllenbach C, Judd L, Kleinerüschkamp A, Taeuber I, et al. Health economics of patient blood management: a cost-benefit analysis based on a meta-analysis. Vox Sanguinis. 2020;115: 182–188. doi: 10.1111/vox.12873 [DOI] [PubMed] [Google Scholar]
  • 9.Callum JL, Waters JH, Shaz BH, Sloan SR, Murphy MF. The AABB recommendations for the Choosing Wisely campaign of the American Board of Internal Medicine. Transfusion. 2014;54: 2344–52. doi: 10.1111/trf.12802 [DOI] [PubMed] [Google Scholar]
  • 10.Lemke M, Eeson G, Lin Y, Tarshis J, Hallet J, Coburn N, et al. A decision model and cost analysis of intra-operative cell salvage during hepatic resection. HPB: The Official Journal of the International Hepato Pancreato Biliary Association. 2016;18: 428–435. doi: 10.1016/j.hpb.2016.02.002 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 11.Carless PA, Henry DA, Moxey AJ, O’Connell D, Brown T, Fergusson DA. Cell salvage for minimising perioperative allogeneic blood transfusion. The Cochrane database of systematic reviews. 2010. doi: 10.1002/14651858.CD001888.PUB3 [DOI] [PubMed] [Google Scholar]
  • 12.Meybohm P, Herrmann E, Steinbicker AU, Wittmann M, Gruenewald M, Fischer D, et al. Patient blood management is associated with a substantial reduction of red blood cell utilization and safe for patient’s outcome: a prospective, multicenter cohort study with a noninferiority design. Annals of Surgery. 2016;264: 203–11. doi: 10.1097/SLA.0000000000001747 [DOI] [PubMed] [Google Scholar]
  • 13.Klein AA, Bailey CR, Charlton AJ, Evans E, Guckian-Fisher M, McCrossan R, et al. Association of Anaesthetists guidelines: cell salvage for peri-operative blood conservation 2018. Anaesthesia. 2018;73: 1141–1150. doi: 10.1111/anae.14331 [DOI] [PubMed] [Google Scholar]
  • 14.Boer C, Meesters MI, Milojevic M, Benedetto U, Bolliger D, von Heymann C, et al. 2017 EACTS/EACTA Guidelines on patient blood management for adult cardiac surgery. Journal of Cardiothoracic and Vascular Anesthesia. 2018;32: 88–120. doi: 10.1053/j.jvca.2017.06.026 [DOI] [PubMed] [Google Scholar]
  • 15.Martin RC, Wellhausen SR, Moehle DA, Martin AW, McMasters KM. Evaluation of intraoperative autotransfusion filtration for hepatectomy and pancreatectomy. Annals of Surgical Oncology. 2005;12: 1017–24. doi: 10.1245/ASO.2005.12.018 [DOI] [PubMed] [Google Scholar]
  • 16.Lu M, Lezzar DL, Vörös E, Shevkoplyas SS. Traditional and emerging technologies for washing and volume reducing blood products. Journal of Blood Medicine. 2019;10: 37–46. doi: 10.2147/JBM.S166316 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 17.Samolyk KA, Beckmann SR, Bissinger RC. A new practical technique to reduce allogeneic blood exposure and hospital costs while preserving clotting factors after cardiopulmonary bypass: the Hemobag. Perfusion. 2005;20: 343–9. doi: 10.1191/0267659105pf831oa [DOI] [PubMed] [Google Scholar]
  • 18.Yarham G, Clements A, Oliver M, Morris C, Cumberland T, Bryan M, et al. Evaluating the “next generation” of cell salvage—Will it make a difference? Perfusion. 2011;26: 263–270. doi: 10.1177/0267659111399951 [DOI] [PubMed] [Google Scholar]
  • 19.Paparella D, Whitlock R. Safety of salvaged blood and risk of coagulopathy in cardiac surgery. Seminars in Thrombosis and Hemostasis. 2016;42: 166–171. doi: 10.1055/s-0035-1569067 [DOI] [PubMed] [Google Scholar]
  • 20.Ni H, Freedman J. Platelets in hemostasis and thrombosis: role of integrins and their ligands. Transfusion and Apheresis Science: Official Journal of the World Apheresis Association: Official Journal of the European Society for Haemapheresis. 2003;28: 257–64. doi: 10.1016/S1473-0502(03)00044-2 [DOI] [PubMed] [Google Scholar]
  • 21.Clemetson KJ. Platelets and primary haemostasis. Thrombosis Research. 2012;129: 220–4. doi: 10.1016/j.thromres.2011.11.036 [DOI] [PubMed] [Google Scholar]
  • 22.Gaydos LA, Freireich EJ, Mantel N. The quantitative relation between platelet count and hemorrhage in patients with acute leukemia. The New England Journal of Medicine. 1962;266: 905–9. doi: 10.1056/NEJM196205032661802 [DOI] [PubMed] [Google Scholar]
  • 23.McQuilten ZK, Crighton G, Brunskill S, Morison JK, Richter TH, Waters N, et al. Optimal dose, timing and ratio of blood products in massive transfusion: results from a systematic review. Transfusion Medicine Reviews. 2018;32: 6–15. doi: 10.1016/j.tmrv.2017.06.003 [DOI] [PubMed] [Google Scholar]
  • 24.Ledgerwood AM, Blaisdell W. Coagulation challenges after severe injury with hemorrhagic shock. The Journal of Trauma and Acute Care Surgery. 2012;72: 1714–1718. doi: 10.1097/TA.0b013e318245225c [DOI] [PubMed] [Google Scholar]
  • 25.Beckett A, Callum J, da Luz LT, Schmid J, Funk C, Glassberg E, et al. Fresh whole blood transfusion capability for Special Operations Forces. Canadian Journal of Surgery. 2015;58: 153–6. doi: 10.1503/cjs.012614 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 26.Spinella PC, Cap AP. Whole blood: back to the future. Current Opinion in Hematology. 2016;23: 536–542. doi: 10.1097/MOH.0000000000000284 [DOI] [PubMed] [Google Scholar]
  • 27.French Health Authority—Haute Autorité de Santé—Transfusion de plaquettes: produits, indications. 2015. Available: https://www.has-sante.fr/jcms/c_2571571/fr/transfusion-de-plaquettes-produits-indications
  • 28.Shen S, Zhang J, Wang W, Zheng J, Xie Y. Impact of intra-operative cell salvage on blood coagulation in high-bleeding-risk patients undergoing cardiac surgery with cardiopulmonary bypass: a prospective randomized and controlled trial. Journal of Translational Medicine. 2016;14: 228. doi: 10.1186/s12967-016-0986-6 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 29.Zhou J. A review of the application of autologous blood transfusion. Brazilian Journal of Medical and Biological Research. 2016;49:e5493. doi: 10.1590/1414-431X20165493 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 30.Mansour A, Decouture B, Roussel M, Lefevre C, Skreko L, Picard V, et al. Combined platelet and erythrocyte salvage: evaluation of a new filtration-based autotransfusion device. Anesthesiology. 2021;135: 246–257. doi: 10.1097/ALN.0000000000003820 [DOI] [PubMed] [Google Scholar]
  • 31.ISO—ISO 10993–4:2017—Biological evaluation of medical devices—Part 4: Selection of tests for interactions with blood. 2017. Available: https://www.iso.org/standard/63448.html
  • 32.Use of International Standard ISO 10993–1, “Biological evaluation of medical devices-Part 1: Evaluation and testing within a risk management process” Guidance for Industry and Food and Drug Administration Staff Preface Public Comment. 2020 Available: https://www.fda.gov/vaccines-blood-biologics/guidance-compliance-regulatory-information-biologics
  • 33.Olsen AK, Hansen AK, Jespersen J, Marckmann P, Bladbjerg EM. The pig as a model in blood coagulation and fibrinolysis research. Scandinavian Journal of Laboratory Animal Science. 1999;26: 214–224. [Google Scholar]
  • 34.Velik-Salchner C, Schnürer C, Fries D, Müssigang PR, Moser PL, Streif W, et al. Normal values for thrombelastography (ROTEM) and selected coagulation parameters in porcine blood. Thrombosis Research. 2006;117: 597–602. doi: 10.1016/j.thromres.2005.05.015 [DOI] [PubMed] [Google Scholar]
  • 35.Kessler U, Grau T, Gronchi F, Berger S, Brandt S, Bracht H, et al. Comparison of porcine and human coagulation by thrombelastometry. Thrombosis Research. 2011;128: 477–482. doi: 10.1016/j.thromres.2011.03.013 [DOI] [PubMed] [Google Scholar]
  • 36.Davidson AJ, Russo RM, Ferencz SE, Grayson JK, Williams TK, Galante JM, et al. A novel model of highly lethal uncontrolled torso hemorrhage in swine. The Journal of Surgical Research. 2017;218: 306–315. doi: 10.1016/j.jss.2017.06.045 [DOI] [PubMed] [Google Scholar]
  • 37.Gregoretti S. Suction-induced hemolysis at various vacuum pressures: implications for intraoperative blood salvage. Transfusion. 1996;36: 57–60. doi: 10.1046/j.1537-2995.1996.36196190516.x [DOI] [PubMed] [Google Scholar]
  • 38.McCrackin MA, Swindle MM. Biology, Handling, Husbandry and Anatomy. In Swindle MM, Smith AC editors Swine in the Laboratory: surgery, anesthesia, imaging and experimental techniques. 3rd edition. Boca Raton: CRC Press, 2015;1–36. [Google Scholar]
  • 39.American Association of Blood Banks (AABB). In: Standards for Perioperative Autologous Blood Collection and Administration 9th ed. 2021. Available: https://www.aabb.org/standards-accreditation/standards
  • 40.Naito K, Mizuguchi K, Nosé y. The need for standardizing the index of hemolysis. Artificial Organs. 1994;18: 7–10. doi: 10.1111/j.1525-1594.1994.tb03292.x [DOI] [PubMed] [Google Scholar]
  • 41.Hofbauer N, Windberger U, Schwendenwein I, Tichy A, Eberspächer E. Evaluation of canine red blood cell quality after processing with an automated cell salvage device. Journal of Veterinary Emergency and Critical Care. 2016;26: 373–383. doi: 10.1111/vec.12472 [DOI] [PubMed] [Google Scholar]
  • 42.Kellett-Gregory LM, Seth M, Adamantos S, Chan DL. Autologous canine red blood cell transfusion using cell salvage devices. Journal of Veterinary Emergency and Critical Care. 2013;23: 82–6. doi: 10.1111/vec.12017 [DOI] [PubMed] [Google Scholar]
  • 43.Yoon C, Noh S, Lee JC, Ko SH, Ahn W, Kim HC. Influence of the washing program on the blood processing performance of a continuous autotransfusion device. Journal of Artificial Organs: the Official Journal of the Japanese Society for Artificial Organs. 2014;17: 118–22. doi: 10.1007/s10047-013-0745-x [DOI] [PubMed] [Google Scholar]
  • 44.Gourlay T, Simpson C, Robertson CA. Development of a portable blood salvage and autotransfusion technology to enhance survivability of personnel requiring major medical interventions in austere or military environments. Journal of the Royal Army Medical Corps. 2018;164: 96–102. doi: 10.1136/jramc-2017-000789 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 45.Pelagalli A, Belisario MA, Tafuri S, Lombardi P, d’Angelo D, Avallone L, et al. Adhesive properties of platelets from different animal species. Journal of Comparative Pathology. 2003;128: 127–131. doi: 10.1053/jcpa.2002.0615 [DOI] [PubMed] [Google Scholar]
  • 46.Vagianos C, Steen S, Johansson S, Masson P, Tengborn L, Solem J. Intraoperative collection of shed blood with citrated compresses for autotransfusion. An experimental study in pigs. Acta Chirurgica Scandinavica. 1990;156: 121–6. [PubMed] [Google Scholar]
  • 47.Filos KS, Vagianos CE, Stavropoulos M, Tassoudis V, Patroni O, Fligou F, et al. Evaluation of the effects of autotransfusion of unprocessed blood on hemodynamics and oxygen transport in anesthetized pigs. Critical Care Medicine. 1996;24: 855–61. doi: 10.1097/00003246-199605000-00021 [DOI] [PubMed] [Google Scholar]
  • 48.Patel NN, Lin H, Toth T, Welsh GI, Jones C, Ray P, et al. Reversal of anemia with allogenic RBC transfusion prevents post-cardiopulmonary bypass acute kidney injury in swine. American Journal of Physiology Renal Physiology. 2011;301: F605–14. doi: 10.1152/ajprenal.00145.2011 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 49.Biagini S, Costa PA, Wendel S, Schettino G, Azevedo LCP. In vitro and in vivo validation of stored swine erythrocyte viability to establish an experimental model of homologous red blood cell transfusion: a pilot study. Revista Brasileira de Terapia Intensiva. 2014;26: 287–291. doi: 10.5935/0103-507x.20140040 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 50.Sondeen JL, Prince MD, Polykratis IA, Hernandez O, Torres-Mendoza J, Guzman R de, et al. Blood-banking techniques for plateletpheresis in swine. Journal of the American Association for Laboratory Animal Science: JAALAS. 2014;53: 307–316. Available: /pmc/articles/PMC4128570/ [PMC free article] [PubMed] [Google Scholar]
  • 51.Myers DD Jr, Diaz JA, Conte ML, Swindle MM. Cardiothoracic and vascular surgery/chronic intravascular catheterization. In Swindle MM, Smith AC: Swine in the Laboratory: surgery, anesthesia, imaging and experimental techniques. 3rd edition. Boca Raton: CRC Press, 2015;213–281. [Google Scholar]
  • 52.Hannon JP. Swine as models in biomedical research. In Swindle MM: Swine as models in biomedical research Iowa State University, Ames 1992; 197–245.
  • 53.Lelovas PP, Kostomitsopoulos NG, Xanthos TT. A comparative anatomic and physiologic overview of the porcine heart. Journal of the American Association for Laboratory Animal Science: JAALAS. 2014;53: 432–438. Available: /pmc/articles/PMC4181683/ [PMC free article] [PubMed] [Google Scholar]
  • 54.Bode G, Clausing P, Gervais F, Loegsted J, Luft J, Nogues V, et al. The utility of the minipig as an animal model in regulatory toxicology. Journal of Pharmacological and Toxicological Methods. 2010;62: 196–220. doi: 10.1016/j.vascn.2010.05.009 [DOI] [PubMed] [Google Scholar]
  • 55.Sharma S, Sharma P, Tyler LN. Transfusion of blood and blood products: indications and complications. American Family Physician. 2011;83: 719–724. Available: www.aafp.org/afp. [PubMed] [Google Scholar]
  • 56.Jain NC. Essentials of Veterinary Hematology. 1st edition, Philadelphia: Wiley-Blackwell. 1993; 417. [Google Scholar]
  • 57.Naumenko KS, Kim SF, Cherkanova MS, Naumenko SE. The Haemonetics® Cell Saver 5 washing properties: effect of different washing pump and centrifuge speeds. Interactive CardioVascular and Thoracic Surgery. 2008;7: 759–63. doi: 10.1510/icvts.2008.183582 [DOI] [PubMed] [Google Scholar]
  • 58.European Directorate for the Quality of Medicines and HealthCare of the Council of Europe (EDQM): Guide to the Preparation, Use and Quality Assurance of Blood Components. 20th ed. Strasbourg, France, Council of Europe Publishing; 2020.
  • 59.de Somer F, Bernolet H, Bouchez S, Bové T. Quality evaluation of the emergency program of three contemporary cell-washing machines. Perfusion. 2019;34: 318–322. doi: 10.1177/0267659118819927 [DOI] [PubMed] [Google Scholar]
  • 60.Kestin AS, Valeri CR, Khuri SF, Loscalzo J, Ellis PA, MacGregor H, et al. The platelet function defect of cardiopulmonary bypass. Blood. 1993;82: 107–117. doi: 10.1182/BLOOD.V82.1.107.BLOODJOURNAL821107 [DOI] [PubMed] [Google Scholar]
  • 61.Varghese SJ, Unni MK, Mukundan N, Rai R. Platelet functions in cardiopulmonary bypass surgery. Medical Journal Armed Forces India. 2005;61: 316–321. doi: 10.1016/S0377-1237(05)80053-X [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 62.Höfer J, Fries D, Solomon C, Velik-Salchner C, Ausserer J. A snapshot of coagulopathy after cardiopulmonary bypass. Official Journal of the International Academy of Clinical and Applied Thrombosis/Hemostasis. 2016;22: 505–11. doi: 10.1177/1076029616651146 [DOI] [PubMed] [Google Scholar]
  • 63.di Dedda U, Ranucci M, Porta A, Bari V, Ascari A, Fantinato A, et al. The combined effects of the microcirculatory status and cardiopulmonary bypass on platelet count and function during cardiac surgery. Clinical Hemorheology and Microcirculation. 2018;70: 327–337. doi: 10.3233/CH-180391 [DOI] [PubMed] [Google Scholar]

Decision Letter 0

Ahmet Emre Eşkazan

24 Aug 2021

PONE-D-21-23852

A novel autotransfusion device saving erythrocytes and platelets by filtration used in a 72h survival swine model of controlled blood loss: perioperative hematologic and coagulation assessments, salvaged blood characteristics and system performance

PLOS ONE

Dear Dr. Touzot-Jourde,

Thank you for submitting your manuscript to PLOS ONE. After careful consideration, we feel that it has merit but does not fully meet PLOS ONE’s publication criteria as it currently stands. Therefore, we invite you to submit a revised version of the manuscript that addresses the points raised during the review process.

Please submit your revised manuscript by Oct 08 2021 11:59PM. If you will need more time than this to complete your revisions, please reply to this message or contact the journal office at plosone@plos.org. When you're ready to submit your revision, log on to https://www.editorialmanager.com/pone/ and select the 'Submissions Needing Revision' folder to locate your manuscript file.

Please include the following items when submitting your revised manuscript:

  • A rebuttal letter that responds to each point raised by the academic editor and reviewer(s). You should upload this letter as a separate file labeled 'Response to Reviewers'.

  • A marked-up copy of your manuscript that highlights changes made to the original version. You should upload this as a separate file labeled 'Revised Manuscript with Track Changes'.

  • An unmarked version of your revised paper without tracked changes. You should upload this as a separate file labeled 'Manuscript'.

If you would like to make changes to your financial disclosure, please include your updated statement in your cover letter. Guidelines for resubmitting your figure files are available below the reviewer comments at the end of this letter.

If applicable, we recommend that you deposit your laboratory protocols in protocols.io to enhance the reproducibility of your results. Protocols.io assigns your protocol its own identifier (DOI) so that it can be cited independently in the future. For instructions see: http://journals.plos.org/plosone/s/submission-guidelines#loc-laboratory-protocols. Additionally, PLOS ONE offers an option for publishing peer-reviewed Lab Protocol articles, which describe protocols hosted on protocols.io. Read more information on sharing protocols at https://plos.org/protocols?utm_medium=editorial-email&utm_source=authorletters&utm_campaign=protocols.

We look forward to receiving your revised manuscript.

Kind regards,

Ahmet Emre Eşkazan

Academic Editor

PLOS ONE

Journal Requirements:

1. When submitting your revision, we need you to address these additional requirements.

Please ensure that your manuscript meets PLOS ONE's style requirements, including those for file naming. The PLOS ONE style templates can be found at 

https://journals.plos.org/plosone/s/file?id=wjVg/PLOSOne_formatting_sample_main_body.pdf and 

https://journals.plos.org/plosone/s/file?id=ba62/PLOSOne_formatting_sample_title_authors_affiliations.pdf

2. Thank you for stating the following in the Competing Interests section: 

"Dr. Benoit Decouture (BD) is currently employed as project manager by i-SEP (Nantes, France).

Stéphane Chollet (SC) is currently employed as technical director by i-SEP (Nantes, France).

Marine Bruneau (MB) was employed as research and development engineer by i-SEP (Nantes, France).

Maxence Nicollet (MN) was employed as a technician by i-SEP (Nantes, France).

Dr. Catherine Wittman (CW) was employed as a compliance and clinical affairs deputy head by i-SEP (Nantes, France).

Dr. Francis Gadrat (FG) is currently employed as medical director by i-SEP (Nantes, France).

Dr. Patricia Forest-Villegas (PVF) is currently employed as scientific director by i-SEP (Nantes, France)."

Please confirm that this does not alter your adherence to all PLOS ONE policies on sharing data and materials, by including the following statement: "This does not alter our adherence to  PLOS ONE policies on sharing data and materials.” (as detailed online in our guide for authors http://journals.plos.org/plosone/s/competing-interests).  If there are restrictions on sharing of data and/or materials, please state these. Please note that we cannot proceed with consideration of your article until this information has been declared. 

Please include your updated Competing Interests statement in your cover letter; we will change the online submission form on your behalf.

[Note: HTML markup is below. Please do not edit.]

Reviewers' comments:

Reviewer's Responses to Questions

Comments to the Author

1. Is the manuscript technically sound, and do the data support the conclusions?

The manuscript must describe a technically sound piece of scientific research with data that supports the conclusions. Experiments must have been conducted rigorously, with appropriate controls, replication, and sample sizes. The conclusions must be drawn appropriately based on the data presented.

Reviewer #1: Yes

Reviewer #2: Yes

Reviewer #3: Yes

**********

2. Has the statistical analysis been performed appropriately and rigorously?

Reviewer #1: Yes

Reviewer #2: Yes

Reviewer #3: Yes

**********

3. Have the authors made all data underlying the findings in their manuscript fully available?

The PLOS Data policy requires authors to make all data underlying the findings described in their manuscript fully available without restriction, with rare exception (please refer to the Data Availability Statement in the manuscript PDF file). The data should be provided as part of the manuscript or its supporting information, or deposited to a public repository. For example, in addition to summary statistics, the data points behind means, medians and variance measures should be available. If there are restrictions on publicly sharing data—e.g. participant privacy or use of data from a third party—those must be specified.

Reviewer #1: Yes

Reviewer #2: Yes

Reviewer #3: Yes

**********

4. Is the manuscript presented in an intelligible fashion and written in standard English?

PLOS ONE does not copyedit accepted manuscripts, so the language in submitted articles must be clear, correct, and unambiguous. Any typographical or grammatical errors should be corrected at revision, so please note any specific errors here.

Reviewer #1: Yes

Reviewer #2: Yes

Reviewer #3: Yes

**********

5. Review Comments to the Author

Please use the space provided to explain your answers to the questions above. You may also include additional comments for the author, including concerns about dual publication, research ethics, or publication ethics. (Please upload your review as an attachment if it exceeds 20,000 characters)

Reviewer #1: In the manuscript, the authors evaluated the performances of a new autotransfusion system which can collect bloods and platelets simultaneously using a swine model. The topic is both interesting and important. There are several minor problems need to be improved:

1. It is better if the autotransfusion system and the study design can be depict using a schematic diagram.

2. Obviously the greatest advantage of this device is it can collect bloods and platelets simultaneously, can you simply explain how it works?

Reviewer #2: This study described a new swine model of surgical blood loss that could be used to evaluate a cell saver or autotransfusion system. The study also described a new autotransfusion system. In some ways, this reads like two separate manuscripts. The manuscript is very lengthy and contains a lot of data and text. It also reads like two manuscripts have been combined.

Concerning the part on the new autotransfusion system, the tables and figures contain all of the important data, but the text of the results section does not mention some important information such as the recovery of platelets. This information is, however, mentioned in the discussion section.

Reviewer #3: The authors investigated on the performance of a new intra-operative autotransfusion system, i-Sep device, using two different models of controlled bleeding. It is a new system for the intraoperative blood salvage, which recovers not only red blood cells, but also platelets and leukocytes. The main point to consider is if transfusion of recovered platelets is effective in terms of hemostasis and if it can reduce transfusion of allogeneic platelets, and if they do not cause harm to the patient. Also, the consequences of recovering and transfusing leukocytes needs consideration.

Autologous blood transfusion is an important alternative in terms of patient blood management, and the intraoperative cell salvage is advantageous since it does not cause burden to the patients in collecting autologous blood pre-operatively.

The paper is quite confusing because there are two objectives in the study, first to develop pig models of controlled cardiac and visceral bleeding, and second to test the effectiveness of i-Sep as an autotransfusion system. It is important that both objectives are clearly stated, and conclusions are drawn for both the objectives.

It would be interesting to see the i-Sep results in comparison with an existing system, such as the Auto Saver, using established animal models.

There are many points that must be addressed before the paper can be considered for publication.

Major points:

1. What are the advantages of the established animal models compared to the existing ones, and how can they be applied in the new future?

2. In the cardiac model, what was the priming solution used for the cardio-thoracic bypass? A priming volume of 1250mL was used, which may affect the general results.

3. I understand on the importance of transfusing platelets together with RBC, especially in patients with massive bleeding, but I am not sure if platelets recovered from the cardio-thoracic bypass are effective enough to be transfused. It is very well-known that the cardio-thoracic bypass activates platelets, resulting in disturbed platelet function, as well as low platelet counts, which is compatible with their results of lower platelet yield in the cardiac model. There is no data, however, showing the number of platelets recovered from the cardio-thoracic bypass, nor on their in vitro activation or functional status. These data should be shown, at least for the cardiac model.

4. What is the advantage of recovering and transfusing leukocytes together with RBC and platelets? The universal leukoreduction has been implemented in various countries as a strategy to prevent alloimmunization and prevent non-hemolytic transfusion reactions. Since the transfused leukocytes are autologous, there is no risk of alloimmunization, and since autologous blood is not stored, I believe the risk of cytokine accumulation that occurs during storage would not be a problem, but it is necessary to compare leukoreduced and non-leukoreduced products, or at least discuss on the advantages/disadvantages of transfusing autologous leukocytes.

5. In both models, the authors performed blood sampling before surgery (T0), then at the end of transfusion (T1) and post-operatively. Did the authors perform blood sampling after the bleeding, that is between T0 and T1? What was the hemoglobin trigger to start the transfusion? How much did the hemoglobin levels drop after bleeding of more than 20% of the circulating blood volume? What were the platelet values before and after the bleeding, and how did they recover after the transfusion? This information is essential to conclude if their models are useful for the evaluation of transfusion practices in an animal model. Especially in the cardiac model, how much platelets were lost during the cardio-thoracic bypass, how was the function and activation status of the recovered platelets? Was the effect of blood transfusion poorer compared to the visceral model?

6. The massive transfusion protocol recommends transfusion of RBC: PLT: plasma at a 1:1:1 ratio, and recently, the pre-hospital transfusion of whole blood is being performed especially in trauma centers in the US, with very promising results. In the present study, RBC and platelets are transfused without plasma, that means without coagulation factors. What is the consequence of this?

7. Describe the limitations of the study at the end of Discussion.

Minor points:

1. Figure legends should be removed from the text and described in a separate section.

2. In line 48, 248, 271-2, 276, 364, 555-9, 563 (Table 4) and 631: There are descriptions of “UI” and “IU”. It should be expressed as “IU”.

3. In lines 133 and 681: The equipment is described as “I-Sep” and “i-Sep”. Which one is correct?

4. In line 173: “done just prior morning feeding” should be “done just prior to morning feeding”

5. In line 254: “insure” should be “ensure”

6. In line 322: “Iso” should be “ISO”.

7. In lines 322-324: There is need to close the quotation mark (“)

8. In lines 366, 441, 454, 563 (Table 4),Fig 1(B) and Fig 2(B): The units of hemoglobin are differently described, such as “(500-1000 mg/dL)” “Hemoglobin in mg/L” “Total hemoglobin concentration (g/dL)” “Hemoglobin concentration (g/L)” . Please check.

9. In line 360: “autotransfusion system washing” should be “autotransfusion system in washing”

10. In line 382: “Hematocrite” should be “Hematocrit”

11. In lines 440-442: There is need to close the quotation mark (“)

12. In lines 480-483: The descriptive statistics should be given for the platelet counts (median [min-max])

13. In lines 485-489 and Fig.4 (A): The data of 4 animals are shown and described in the visceral model. I believe there are 5 animals in this model. Please check and correct.

14. In line 503: “compared before” should be “compared to before”

15. In lines 506-507: In Table 1, some values are expressed in non-bold characters.

16. In line 530: In Table 3, should “Amplitude 6 minutes” be expressed in bold characters?

17. In lines 556-557and 563-564: The descriptions of heparin concentrations of “0.14 [0.00-0.36] UI/mL” from visceral model and “0.46 [0.44-0.60] UI/mL” from the cardiac model do not match the values described in Table 3 “0.16 ± 0.19” in visceral model and “0.50 ± 0.08” in cardiac model. Similarly, the hemolysis rate described in lines 652-653 does not match that described in Table 3.

18. In line 561: From Table 4, the criterium of RBC yield was not achieved, and there was high hemolysis, so it would be better to describe it correctly as follows: “performance parameters (heparin clearance, free hemoglobin washout, hemolysis rate, RBC yield, hematocrit and hemoglobin concentration) exceeded set objectives, except for RBC yield and hemolysis rate (Table 4).”

19. In line 564: I suppose the data in Table 4 is shown as mean ± SD, and not as “median” as described. Please check.

20. In line 569: “for both model” should be “for both models”

21. In Figure 4: I suppose there were 5 animals in the visceral model, but the figure shows data from 4 animals. Please check.

22. In line 600: “they concluded was that” should be “they concluded that”

23. In line 604: “The porcine model appears as” would be better “The porcine model seems to be”

24. In line 611: “as close as possible clinical indications” would be better “as close as possible to the clinical settings”

25. In line 613: “thoracotomy and blood treatment of the remaining blood in the CPB” would be better “thoracotomy and treatment of the remaining blood in the CPB”

26. In line 613-614: “(ii) a abdominal visceral model” should be “(ii) an abdominal visceral model”

27. In line 619: “blood treatment processing” should be “blood processing”

28. In line 620: “two cycle treatment” should be “two cycle treatments”

29. In line 622: “compatible with the surgery” would be better “compatible with the surgical time”

30. In line 625: “in human medicine” would be better “in the clinical practice”

31. In lines 636-637: “but was cleared over 90%” would be better “but over 90% could be cleared”

32. In lines 637-638: “This hemolysis phenomenon also explains the lesser performance of free hemoglobin removal compared to the heparin washout” would be better “Hemolysis may also explain the lower removal performance of hemoglobin compared to the heparin washout”

33. In lines 640-643: the sentence “Preliminary results with the present system…” is difficult to understand. Elaborate better

34. In line 644: “hemolysis in the treated blood was higher” should be “hemolysis in the treated blood is higher”

35. In line 645: “in vitro study on human blood testing the system” would be better “in vitro study testing the system with human blood”

36. In line 650: “compared to human blood treatment” should be “compared to human blood”

37. In line 651-652: “explained by blood shedding, coagulation activation during abdominal bleeding and suction conditions” should be “explained by the blood shedding, the coagulation activation during abdominal bleeding and the suction conditions”

38. In line 655-656: “Despite the hemolysis superior to the 0.8% threshold and RBCs yield below 80%” should be “Despite the hemolysis rate higher than the 0.8% threshold and the RBCs yield below 80%”

39. In lines 658-659: “free hemolysis” should be “free hemoglobin”

40. In line 663: “the surgery the concentration of red blood cells” should be “the surgery, to concentrate not only red blood cells”

41. In lines 667-670: “platelet yield always greater than 40% and greater than the one obtained in the preliminary i-Sep study on in vitro human blood” would be better “platelet yield was always greater than 40% and higher than that obtained in the preliminary in vitro study of i-Sep using human blood”

42. In line 678: “Some of the platelet yield over 100%” should be “Some of the platelet yields were over 100%”

43. In line 686: “whichever the animal model” would be better “in either model”

44. In lines 687-688: “any thrombus in any of them” would be better “any evidence of thrombus formation”

45. In line 691: “the incapacity of the treated blood to coagulate” would be better “the inability of the treated blood to coagulate”

46. In lines 699-700: “allowed not only RBCs concentration but also platelet one” would be better “allowed to concentrate not only RBCs but also platelets”

47. In line 701: “pro-coagulant effect nor cause” should be “pro-coagulant effect nor did cause”

48. In line 702: “cardiac or visceral” should be “cardiac or visceral bleeding”

**********

6. PLOS authors have the option to publish the peer review history of their article (what does this mean?). If published, this will include your full peer review and any attached files.

If you choose “no”, your identity will remain anonymous but your review may still be made public.

Do you want your identity to be public for this peer review? For information about this choice, including consent withdrawal, please see our Privacy Policy.

Reviewer #1: No

Reviewer #2: No

Reviewer #3: No

[NOTE: If reviewer comments were submitted as an attachment file, they will be attached to this email and accessible via the submission site. Please log into your account, locate the manuscript record, and check for the action link "View Attachments". If this link does not appear, there are no attachment files.]

While revising your submission, please upload your figure files to the Preflight Analysis and Conversion Engine (PACE) digital diagnostic tool, https://pacev2.apexcovantage.com/. PACE helps ensure that figures meet PLOS requirements. To use PACE, you must first register as a user. Registration is free. Then, login and navigate to the UPLOAD tab, where you will find detailed instructions on how to use the tool. If you encounter any issues or have any questions when using PACE, please email PLOS at figures@plos.org. Please note that Supporting Information files do not need this step.

PLoS One. 2022 Mar 24;17(3):e0260855. doi: 10.1371/journal.pone.0260855.r002

Author response to Decision Letter 0


2 Oct 2021

Dear Editor and Reviewers,

Please find below our responses to the reviewer comments and suggestion for the manuscript ONE-D-21-23852 entitled “A novel autotransfusion device saving erythrocytes and platelets by filtration used in a 72 h survival swine model of controlled blood loss: perioperative hematologic and coagulation assessments, salvaged blood characteristics and system performance”

To ease reading our answers, the text is displayed in blue right below the point raised or question asked buy the Reviewers

Journal Requirements:

1. When submitting your revision, we need you to address these additional requirements.

Please ensure that your manuscript meets PLOS ONE's style requirements, including those for file naming. The PLOS ONE style templates can be found at

https://journals.plos.org/plosone/s/file?id=wjVg/PLOSOne_formatting_sample_main_body.pdf and https://journals.plos.org/plosone/s/file?id=ba62/PLOSOne_formatting_sample_title_authors_affiliations.pdf

We have used the PLOS ONE template to submit our manuscript and checked its conformity to PLOS ONE style requirements.

2. Thank you for stating the following in the Competing Interests section:

"Dr. Benoit Decouture (BD) is currently employed as project manager by i-SEP (Nantes, France).

Stéphane Chollet (SC) is currently employed as technical director by i-SEP (Nantes, France).

Marine Bruneau (MB) was employed as research and development engineer by i-SEP (Nantes, France).

Maxence Nicollet (MN) was employed as a technician by i-SEP (Nantes, France).

Dr. Catherine Wittman (CW) was employed as a compliance and clinical affairs deputy head by i-SEP (Nantes, France).

Dr. Francis Gadrat (FG) is currently employed as medical director by i-SEP (Nantes, France).

Dr. Patricia Forest-Villegas (PVF) is currently employed as scientific director by i-SEP (Nantes, France)."

Please confirm that this does not alter your adherence to all PLOS ONE policies on sharing data and materials, by including the following statement: "This does not alter our adherence to PLOS ONE policies on sharing data and materials.” (as detailed online in our guide for authors http://journals.plos.org/plosone/s/competing-interests). If there are restrictions on sharing of data and/or materials, please state these. Please note that we cannot proceed with consideration of your article until this information has been declared.

Please include your updated Competing Interests statement in your cover letter; we will change the online submission form on your behalf.

We have implemented the changes and confirmed in the cover letter that the identified competing interests have not alter our ability to share data and materials from the study.

Reviewers' comments:

Reviewer's Responses to Questions

Comments to the Author

1. Is the manuscript technically sound, and do the data support the conclusions?

The manuscript must describe a technically sound piece of scientific research with data that supports the conclusions. Experiments must have been conducted rigorously, with appropriate controls, replication, and sample sizes. The conclusions must be drawn appropriately based on the data presented.

Reviewer #1: Yes

Reviewer #2: Yes

Reviewer #3: Yes

2. Has the statistical analysis been performed appropriately and rigorously?

Reviewer #1: Yes

Reviewer #2: Yes

Reviewer #3: Yes

3. Have the authors made all data underlying the findings in their manuscript fully available?

The PLOS Data policy requires authors to make all data underlying the findings described in their manuscript fully available without restriction, with rare exception (please refer to the Data Availability Statement in the manuscript PDF file). The data should be provided as part of the manuscript or its supporting information, or deposited to a public repository. For example, in addition to summary statistics, the data points behind means, medians and variance measures should be available. If there are restrictions on publicly sharing data—e.g. participant privacy or use of data from a third party—those must be specified.

Reviewer #1: Yes

Reviewer #2: Yes

Reviewer #3: Yes

4. Is the manuscript presented in an intelligible fashion and written in standard English?

PLOS ONE does not copyedit accepted manuscripts, so the language in submitted articles must be clear, correct, and unambiguous. Any typographical or grammatical errors should be corrected at revision, so please note any specific errors here.

Reviewer #1: Yes

Reviewer #2: Yes

Reviewer #3: Yes

5. Review Comments to the Author

Please use the space provided to explain your answers to the questions above. You may also include additional comments for the author, including concerns about dual publication, research ethics, or publication ethics. (Please upload your review as an attachment if it exceeds 20,000 characters)

Reviewer #1: In the manuscript, the authors evaluated the performances of a new autotransfusion system which can collect bloods and platelets simultaneously using a swine model. The topic is both interesting and important. There are several minor problems need to be improved:

1. It is better if the autotransfusion system and the study design can be depict using a schematic diagram.

A schematic diagram has been added to help visualizing the experimental steps (hemorrhage, transfusion, post-operative care) and times of blood samplings.

2. Obviously the greatest advantage of this device is it can collect bloods and platelets simultaneously, can you simply explain how it works?

The SAME device (Smart Autotransfusion for Me; i-SEP, France) was designed as an innovative filtration-based autotransfusion device able to salvage and wash both red blood cells and platelets. This new autotransfusion system integrates a hollow fiber filtration technology, comparable to the filters used for plasmapheresis or for ultrafiltration during cardiopulmonary bypass. Using a combination of washing and filtration of salvage blood, the device allows the concentration of red blood cells and platelets within the concentrated blood product, as well as the removal of heparin, free hemoglobin, coagulation factors, and inflammatory mediators such as complement proteins (already published in Mansour et al., Anesthesiology, 2021).

The SAME device is a medical device consisting of reusable equipment and disposable consumables (See digital content 1, http://links.lww.com/ALN/C621). The device innovative technology and process are described in the following patents: PCT/FR2018/053500 published as WO 2019/129973 on July 4, 2019 (corresponding to U.S. application no. 16/958,473); PCT/FR2018/053501 published as WO 2019/129974 on July 4, 2019 (corresponding to U.S. application no. 16/958,458); and PCT/ FR2020/051115 published as WO 2020/260836 on December 30, 2020. Consumables include a dual-lumen suction line (allowing both collection and anticoagulation of shed blood), a blood collection reservoir (including a 40-μm filter), and a treatment set. The treatment set includes tubing, a polyethersulfone hollow fiber cartridge that separates the blood cells from the plasma, a compliant blood treatment bag that ensures the blood washing, a waste bag that receives the plasma and contaminants, and a reinfusion bag that stores the filtered, washed, and concentrated blood cells (See digital content 2, http:// links.lww.com/ALN/C622). The reusable equipment is an electromedical medical device composed by several systems required for blood circulation and continuous measurements, including a continuous in-line hematocrit monitor. The i-SEP device is associated with a specific software to drive the different steps of the device installation and blood treatment.

During clinical use, the first stage of cell salvage with the i-SEP device is the collection of shed blood from the surgical field by the dual-lumen suction line, allowing the anticoagulation of shed blood by a heparinized saline drip. The shed blood is collected in the collection reservoir in which it undergoes a first filtration by the included 40-μm filter, allowing the removal of bone debris and microaggregates before blood treatment by the device.

Then the treatment set is filled with anticoagulated salvaged blood transferred from the blood collection reservoir when a sufficient volume is collected. During the treatment phase, the blood is processed by the i-SEP device, with simultaneous filtration and washing (Supplemental Digital Content 2, http://links.lww.com/ALN/C622). The volume of the treatment set (300 to 1,000 ml thanks to the compliant treatment bag) limits the amount of collected blood that can be processed in one time, hence defining a treatment cycle. Several simultaneous steps then constitute the innovative i-SEP process: wash solution (normal saline) is pumped into the treatment set; diluted salvaged blood circulates within the treatment set between the treatment bag and the polyethersulfone hollow fiber to allow microfiltration to occur; and fluid is continuously discarded from the treatment circuit into the waste bag through the effluent line. Once the continuously monitored hematocrit reaches the prespecified target, the device automatically transfers the processed blood from the treatment set into the reinfusion bag.

The reference of the device description (Mansour et al., Anesthesiology, 2021) has been added in the Materials and Methods.

Reviewer #2: This study described a new swine model of surgical blood loss that could be used to evaluate a cell saver or autotransfusion system. The study also described a new autotransfusion system. In some ways, this reads like two separate manuscripts. The manuscript is very lengthy and contains a lot of data and text. It also reads like two manuscripts have been combined.

The objectives of the study have been reformulated at the end of the introduction to better explain the rationale of the study. Indeed, the aim of this study was twofold and we specified it in the Introduction of the manuscript. The dual objective of this study was first to develop a swine surgical model of controlled blood loss with a 72 h post-operative follow-up that would reproduce the best possible clinically relevant surgical events, and then to evaluate the use of the autotransfusion device in this swine model. The afterward aim is to use this controlled blood loss surgical model to pursue the device development and its preclinical evaluation in a comparative study with a cell saver already on the market.

Concerning the part on the new autotransfusion system, the tables and figures contain all of the important data, but the text of the results section does not mention some important information such as the recovery of platelets. This information is, however, mentioned in the discussion section.

The result section corresponding to the tables and figures on the autotransfusion performance has been rewritten.

Reviewer #3: The authors investigated on the performance of a new intra-operative autotransfusion system, i-Sep device, using two different models of controlled bleeding. It is a new system for the intraoperative blood salvage, which recovers not only red blood cells, but also platelets and leukocytes. The main point to consider is if transfusion of recovered platelets is effective in terms of hemostasis and if it can reduce transfusion of allogeneic platelets, and if they do not cause harm to the patient. Also, the consequences of recovering and transfusing leukocytes need consideration.

Transfusion of recovered platelets

Reinfusion of platelets is indeed a crucial question as it could induce prothrombotic events in the patient blood circulation.

Thus, in our previous study on human blood (Mansour et al., Anesthesiology, 2021), we evaluated the platelet activation after blood treatment and showed that platelets had a low P-selectin expression after treatment (10.8%) and kept a strong response to thrombin-activating peptide. These results demonstrate that platelets were not activated during the treatment, even partially or reversibly, as they are not refractory to activation.

Moreover, in the HAS (Haute Autorité de Santé) and the ANSM (Agence Nationale de Sécurité du Médicament et des produits de santé) document about the good practice recommendation of platelet transfusion (“Transfusion de plaquettes: produits, indications” HAS and ANSM, argumentaire scientifique, October 2015), moderated disturbances of the platelet activity are detected in concentrated platelets from apheresis. Notably, the study of Metcalfe et al. shows between 25% and 30% of p-selectin positive platelets in apheresis platelets (Metcalfe et al., Br J Haematol, 1997). Recent studies show similar results (Bontekoe et al, Vox Sanguinis, 2018; Sperling et al, Hematology, 2018). Percentage of p-selectin positive platelets was between 15% and 25% after one day-storage, 39% after two days-storage and above 60% after five days-storage. In the concentrated blood obtained with the i-Sep Autotransfusion System, the percentage of p-selectin positive platelets (i.e. activated platelets by the treatment, 10.8%) was lower than those in the study of Metcalfe et al. Then, the results show that the platelets obtained with the i-Sep Autotransfusion System are compatible with transfusion.

Transfusion of recovered leukocytes

Leukocyte transfusion is also a crucial question as activated leukocytes could induce strong pro-inflammatory events which should be avoided in a recovering patient.

However, although the other autotransfusion devices claim that leukocytes are eliminated during blood treatment, literature showed high white blood cell recovery: the Sorin Xtra device white blood cell recovery is between 55% and 70% (Overdevest et al., Perfusion, 2012), the Fresenius Kabi devices white blood cell recovery are 65.6% for the CATSmart and 81.1% for the C.A.T.Splus (Alberts et al., J Extra Corpor Technol, 2017).

Regarding the i-SEP device, the previous study showed a 93.0% white blood cell recovery (Mansour et al., Anesthesiology, 2021). As the recovery was very high, white blood cell death or activation was evaluated. Leukocyte viability in pretreated blood (collection reservoir and treatment bag) was 97.6% for the treatment first cycle and 97.7% for the second cycle. After treatment, the leukocyte viability was changed by less than 1%. Regarding basal activation of leukocytes in the pretreated blood, respectively, for the first and second cycles: the percentages of HLA-DR positive/CD4- positive cells were 4.4% and 4.4%; the percentages of HLA-DR positive/CD8-positive cells were 13.2% and 12.7%; CD64 surface expression levels on neutrophils were 1,420 and 1,463; and CD64 surface expression levels on monocytes were 11,122 and 12,219. Cell recovery was not associated with significant leukocyte activation, either regarding CD4-positive cells (−0.1% for the first cycle; 0.0% for the second cycle) or CD8-positive cells (−3.2% for the first cycle; −2.0% for the second cycle). Last, blood treatment did not induce any significant increase in CD64 surface expression in post-treated compared to pretreated blood for neutrophils (8, for the first cycle; −30 for the second cycle) or monocytes (1,905 for the first cycle; 794 for the second cycle).

In conclusion, obtained results in the previous study on human blood showed that the leukocyte reinfusion from the i-SEP device comparable to the other autotransfusion devices already in use.

Autologous blood transfusion is an important alternative in terms of patient blood management, and the intraoperative cell salvage is advantageous since it does not cause burden to the patients in collecting autologous blood pre-operatively.

The paper is quite confusing because there are two objectives in the study, first to develop pig models of controlled cardiac and visceral bleeding, and second to test the effectiveness of i-Sep as an autotransfusion system. It is important that both objectives are clearly stated, and conclusions are drawn for both the objectives.

The objectives of the study have been reformulated at the end of the introduction to better explain the rationale of the study. Indeed, the aim of this study was twofold and we specified it in the Introduction of the manuscript. The dual objective of this study was first to develop a swine surgical model of controlled blood loss with a 72 h post-operative follow-up that would reproduce the best possible clinically relevant surgical events,, and then to evaluate the use of the autotransfusion device in this swine model.

It would be interesting to see the i-Sep results in comparison with an existing system, such as the Auto Saver, using established animal models.

This suggestion is indeed very relevant. A comparative study with a cell saver already on the market is indeed the next step.

There are many points that must be addressed before the paper can be considered for publication.

Major points:

1. What are the advantages of the established animal models compared to the existing ones, and how can they be applied in the new future?

The main indication of using an autotransfusion device in humans are surgical procedures with a potential for massive bleeding and cardiac surgery using extracorporeal circulation. The experiments intended to create with swine models a clinical situation as close as possible to the surgical situations in human operating room.

Cardiac model: Collecting diluted blood from the bypass circuit after patient blood volume adjustment at the end of the extracorporeal circulation followed retransfusion to the animal was therefore chosen to mimic the clinical situation.

Visceral model : Concerning the splenic model, the main objectives of the model were:

- to develop a model of visceral bleeding in which shed blood would be in contact with the visceral serosa, in order to reproduce a clinical situation to reproduce tissue bleeding in which blood contacts surrounding extravascular tissue;

- to create a visceral injury that allow massive bleeding while still being controllable by the bleeding organ resection;

- to have a survival model allowing to document hematological data during the postoperative period.

Therefore, classical controlled hemorrhagic models using venous or arterial blood aspiration even with an organ injury (liver lobe resection for example) were not deemed appropriate. Compared to liver trauma that are used in models of uncontrolled hemorrhage, the splenic hemorrhage was found to occur sufficiently rapidly and profusely to be equivalent to a liver trauma model while surgical hemostasis of the organ was found easy and predictable, which is not the case for the liver. The aim was also to be able to assess survival from the hemorrhage and retransfusion without having the confounding factors of a major traumatic vital organ resection.

2. In the cardiac model, what was the priming solution used for the cardio-thoracic bypass? A priming volume of 1250mL was used, which may affect the general results.

The priming volume chosen was a Ringer Lactate solution. Blood gas analysis during the on bypass pump time showed metabolic stability (normal pH, no to slight elevation of lactate but less than 4 mmol/l, stable electrolytes). It was chosen to minimally interfere (apart from the dilutional effect) with coagulation compared to a mixed priming solution that would contain colloids. The dilutional effect was more visible on the plasma protein concentration (by a decrease of about 20%) than on the hematocrit that inconstantly showed no to minor decrease (probable splenic contraction releasing RBC).

The information was added into the appropriate section.

3. I understand on the importance of transfusing platelets together with RBC, especially in patients with massive bleeding, but I am not sure if platelets recovered from the cardio-thoracic bypass are effective enough to be transfused. It is very well-known that the cardio-thoracic bypass activates platelets, resulting in disturbed platelet function, as well as low platelet counts, which is compatible with their results of lower platelet yield in the cardiac model. There is no data, however, showing the number of platelets recovered from the cardio-thoracic bypass, nor on their in vitro activation or functional status. These data should be shown, at least for the cardiac model.

It has been well described in the literature that the CBP is associated with platelet dysfunction, platelet activation and platelet loss (Muriithi EW et al., The effects of heparin and extracorporeal circulation on platelet counts and platelet microaggregation during cardiopulmonary bypass. J Thorac Cardiovasc Surg. 2000 Sep;120(3):538-43; Tamari Y et al., Functional changes in platelets during extracorporeal circulation. Ann Thorac Surg. 1975 Jun;19(6):639-47.). However, it has been shown that only a proportion of platelets are activated and lost, and that around 50% of the platelets remain in the circulating blood (Griffin BR, Bronsert M, Reece TB, et al. Thrombocytopenia After Cardiopulmonary Bypass Is Associated With Increased Morbidity and Mortality. Ann Thorac Surg. 2020). Platelet concentrations in the blood collection reservoir (TR) were 54 (16-83) 103/µL and 157 (115-192) 103/µL for the visceral and the cardiac model respectively. These data have been added in the text and in the Table 4.

Platelet function can also be affected by CPB and be reduced by up to 50% (Agarwal S et al., Pre- and Post-Bypass Platelet Function Testing With Multiple Electrode Aggregometry and TEG Platelet Mapping in Cardiac Surgery. J Cardiothorac Vasc Anesth. 2015 Oct;29(5):1272-6). Regarding platelets activity in the blood recovered from the extracorporeal circulation, the blood dilution in the collection reservoir did not allow to measure platelet (with platelet ROTEM) nor coagulation activity (ROTEM coagulation tests). However, after blood transfusion and all along the post-operative follow-up, platelet activity and coagulation tests on the animal blood samples were normal.

4. What is the advantage of recovering and transfusing leukocytes together with RBC and platelets? The universal leukoreduction has been implemented in various countries as a strategy to prevent alloimmunization and prevent non-hemolytic transfusion reactions. Since the transfused leukocytes are autologous, there is no risk of alloimmunization, and since autologous blood is not stored, I believe the risk of cytokine accumulation that occurs during storage would not be a problem, but it is necessary to compare leukoreduced and non-leukoreduced products, or at least discuss on the advantages/disadvantages of transfusing autologous leukocytes.

Leukocyte transfusion is also a crucial question as activated leukocytes could induce strong pro-inflammatory events which should be avoided in a recovering patient.

However, although the other autotransfusion devices claim that leukocytes are eliminated during the blood treatment, literature showed high white blood cell recovery: the Sorin Xtra device white blood cell recovery is between 55% and 70% (Overdevest et al., Perfusion, 2012), the Fresenius Kabi devices white blood cell recovery are 65.6% for the CATSmart and 81.1% for the C.A.T.Splus (Alberts et al., J Extra Corpor Technol, 2017).

Regarding the i-SEP device, the previous study showed a 93.0% white blood cell recovery (Mansour et al., Anesthesiology, 2021). As the recovery was very high, white blood cell death or activation was evaluated. Leukocyte viability in pretreated blood was 97.6% for the first cycle and 97.7% for the second cycle. After the treatment, the leukocyte viability was changed by less than 1%. Regarding basal activation of leukocytes in the pretreated blood, respectively, for the first and second cycles: the percentages of HLA-DR positive/CD4- positive cells were 4.4% and 4.4%; the percentages of HLA-DR positive/CD8-positive cells were 13.2% and 12.7%; CD64 surface expression levels on neutrophils were 1,420 and 1,463; and CD64 surface expression levels on monocytes were 11,122 and 12,219. Cell recovery was not associated with significant leukocyte activation, either regarding CD4-positive cells (−0.1% for the first cycle; 0.0% for the second cycle) or CD8-positive cells (−3.2% for the first cycle; −2.0% for the second cycle). Last, blood treatment did not induce any significant increase in CD64 surface expression in post-treated compared to pretreated blood for neutrophils (8, for the first cycle; −30 for the second cycle) or monocytes (1,905 for the first cycle; 794 for the second cycle). Thus, the absence of leukocyte activation strongly annihilates the risk of cytokine secretion and accumulation.

In conclusion, obtained results in the previous study on human blood showed that the leukocyte reinfusion from the i-SEP device comparable to the other autotransfusion devices already in use.

5. In both models, the authors performed blood sampling before surgery (T0), then at the end of transfusion (T1) and post-operatively. Did the authors perform blood sampling after the bleeding, that is between T0 and T1? What was the hemoglobin trigger to start the transfusion? How much did the hemoglobin levels drop after bleeding of more than 20% of the circulating blood volume?

Due to the repeated sampling times during the procedure and the 72 h follow-up for hematology and coagulation testing, we had to refine the numbers of samples to reduce the total volume of blood taken. As we had some prior data on hematocrit and plasma proteins while designing the visceral model, we felt that we didn’t need this extra information. It is however planned for the next study that will focus on the benefit of the transfusion compared to the other autotransfusion devices already on the market.

In preliminary experiments and in the present study, we saw that the surgically triggered splenic injury induced a splenic contraction (visible during the surgery). We also used a minimal fluid resuscitation strategy during the blood loss (enough IV Ringer lactate solution to maintain mean arterial blood pressure around 60 mmHg) to avoid hemodilution. The animals lost between 600 and 900 mL of blood and they received between 1L and 1,5 L of Ringer lactate solution. The combination of both events resulted in stable hematocrit and hemoglobin during bleeding and retransfusion in preliminary experiments.

What were the platelet values before and after the bleeding, and how did they recover after the transfusion? This information is essential to conclude if their models are useful for the evaluation of transfusion practices in an animal model. Especially in the cardiac model, how much platelets were lost during the cardio-thoracic bypass, how was the function and activation status of the recovered platelets? Was the effect of blood transfusion poorer compared to the visceral model?

As mentioned for the previous question, due to the repeated sampling times during the procedure and the 72 h follow-up for hematology and coagulation testing, we had to refine the numbers of samples to reduce the total volume of blood taken. During this study, we did not focus on the direct effect of transfusion (just before and just after the transfusion). This evaluation is however planned for the next study that will focus on the benefit of the transfusion compared to the other autotransfusion devices already on the market.

Data in the supplemental Table S1 allows to compare platelet count before and after surgery. Both models showed a very moderate decrease of the platelet count after the surgery (TP) compared to before (T0).

Regarding the platelet activity after the surgery, we also performed FIBTEM tests when using the ROTEM. FIBTEM is the same test as the EXTEM with a strong platelet inhibitor. Then, the difference between the EXTEM test and the FIBTEM test is the platelet contribution to the EXTEM clot formation. The Maximum Clot Firmness (MCF) is the evaluated parameter.

EXTEM MCF - FIBTEM MCF (mm) Model Parameter T0 TP T2-6 T6-12 T12-24 T24-36 T36-48 T48-60 T60-72

Cardiac (n=4) Median 53 57 53,5 49,5 50 46 44,5 43,5 36,5

Minimum 48 54 51 46 48 44 37 30 16

Maximum 53 60 54 52 55 48 57 46 45

Visceral (n=5) Median 49 53 51 50 50 46 45 47 43

Minimum 40 49 38 40 38 35 42 37 34

Maximum 55 54 57 53 51 48 51 52 51

Results show that platelets equally contributed to clot formation after the surgery (TP) when compared to before the surgery (T0). This way of evaluating platelet contribution to clot formation is not a classical way, thus these results were not added to the manuscript. However, platelet functionality will be the focus in the next studies.

6. The massive transfusion protocol recommends transfusion of RBC: PLT: plasma at a 1:1:1 ratio, and recently, the pre-hospital transfusion of whole blood is being performed especially in trauma centers in the US, with very promising results. In the present study, RBC and platelets are transfused without plasma, that means without coagulation factors. What is the consequence of this?

This comment is very interesting as the blood dilution in the circulation without plasma and coagulation factor transfusion could lead to a prothrombotic state.

However, the protocol performed in this study does not reproduce a massive bleeding nor a massive transfusion case. The volume of the blood and anticoagulation solution mixture in the collection reservoir corresponded to a minimal of 20% and a maximum of 30% of the estimated animal blood volume. We anticipated that this volume would not impact the hemostasis of the animals. This expectation was confirmed by the ROTEM assays that showed unchanged clotting time (CT) after the surgery (TP) when compared to before the surgery (T0).

We are currently planning a high bleeding volume model in which we plan to transfuse plasma volumes equivalent to washed plasma to avoid thrombotic events due to plasma dilution.

7. Describe the limitations of the study at the end of Discussion.

The limitations of the study have been rewritten and pooled at the end of the discussion

Minor points:

We have incorporated the appropriate changes into the manuscript following the minor comments that are listed below.

1. Figure legends should be removed from the text and described in a separate section.

2. In line 48, 248, 271-2, 276, 364, 555-9, 563 (Table 4) and 631: There are descriptions of “UI” and “IU”. It should be expressed as “IU”.

3. In lines 133 and 681: The equipment is described as “I-Sep” and “i-Sep”. Which one is correct?

4. In line 173: “done just prior morning feeding” should be “done just prior to morning feeding”

5. In line 254: “insure” should be “ensure”

6. In line 322: “Iso” should be “ISO”.

7. In lines 322-324: There is need to close the quotation mark (“)

8. In lines 366, 441, 454, 563 (Table 4),Fig 1(B) and Fig 2(B): The units of hemoglobin are differently described, such as “(500-1000 mg/dL)” “Hemoglobin in mg/L” “Total hemoglobin concentration (g/dL)” “Hemoglobin concentration (g/L)” . Please check.

9. In line 360: “autotransfusion system washing” should be “autotransfusion system in washing”

10. In line 382: “Hematocrite” should be “Hematocrit”

11. In lines 440-442: There is need to close the quotation mark (“)

12. In lines 480-483: The descriptive statistics should be given for the platelet counts (median [min-max]) � Added in the supplemental table 4.

13. In lines 485-489 and Fig.4 (A): The data of 4 animals are shown and described in the visceral model. I believe there are 5 animals in this model. Please check and correct.

14. In line 503: “compared before” should be “compared to before”

15. In lines 506-507: In Table 1, some values are expressed in non-bold characters.

16. In line 530: In Table 3, should “Amplitude 6 minutes” be expressed in bold characters?

17. In lines 556-557and 563-564: The descriptions of heparin concentrations of “0.14 [0.00-0.36] UI/mL” from visceral model and “0.46 [0.44-0.60] UI/mL” from the cardiac model do not match the values described in Table 3 “0.16 ± 0.19” in visceral model and “0.50 ± 0.08” in cardiac model. Similarly, the hemolysis rate described in lines 652-653 does not match that described in Table 3.

18. In line 561: From Table 4, the criterium of RBC yield was not achieved, and there was high hemolysis, so it would be better to describe it correctly as follows: “performance parameters (heparin clearance, free hemoglobin washout, hemolysis rate, RBC yield, hematocrit and hemoglobin concentration) exceeded set objectives, except for RBC yield and hemolysis rate (Table 4).”

19. In line 564: I suppose the data in Table 4 is shown as mean ± SD, and not as “median” as described. Please check.

20. In line 569: “for both model” should be “for both models”

21. In Figure 4: I suppose there were 5 animals in the visceral model, but the figure shows data from 4 animals. Please check.

22. In line 600: “they concluded was that” should be “they concluded that”

23. In line 604: “The porcine model appears as” would be better “The porcine model seems to be”

24. In line 611: “as close as possible clinical indications” would be better “as close as possible to the clinical settings”

25. In line 613: “thoracotomy and blood treatment of the remaining blood in the CPB” would be better “thoracotomy and treatment of the remaining blood in the CPB”

26. In line 613-614: “(ii) a abdominal visceral model” should be “(ii) an abdominal visceral model”

27. In line 619: “blood treatment processing” should be “blood processing”

28. In line 620: “two cycle treatment” should be “two cycle treatments”

29. In line 622: “compatible with the surgery” would be better “compatible with the surgical time”

30. In line 625: “in human medicine” would be better “in the clinical practice”

31. In lines 636-637: “but was cleared over 90%” would be better “but over 90% could be cleared”

32. In lines 637-638: “This hemolysis phenomenon also explains the lesser performance of free hemoglobin removal compared to the heparin washout” would be better “Hemolysis may also explain the lower removal performance of hemoglobin compared to the heparin washout”

33. In lines 640-643: the sentence “Preliminary results with the present system…” is difficult to understand. Elaborate better. The sentence was eliminated as these data were extracted from the feasibility experiments which cannot be exploited in this study.

34. In line 644: “hemolysis in the treated blood was higher” should be “hemolysis in the treated blood is higher”

35. In line 645: “in vitro study on human blood testing the system” would be better “in vitro study testing the system with human blood”

36. In line 650: “compared to human blood treatment” should be “compared to human blood”

37. In line 651-652: “explained by blood shedding, coagulation activation during abdominal bleeding and suction conditions” should be “explained by the blood shedding, the coagulation activation during abdominal bleeding and the suction conditions”

38. In line 655-656: “Despite the hemolysis superior to the 0.8% threshold and RBCs yield below 80%” should be “Despite the hemolysis rate higher than the 0.8% threshold and the RBCs yield below 80%”

39. In lines 658-659: “free hemolysis” should be “free hemoglobin”

40. In line 663: “the surgery the concentration of red blood cells” should be “the surgery, to concentrate not only red blood cells”

41. In lines 667-670: “platelet yield always greater than 40% and greater than the one obtained in the preliminary i-Sep study on in vitro human blood” would be better “platelet yield was always greater than 40% and higher than that obtained in the preliminary in vitro study of i-Sep using human blood”

42. In line 678: “Some of the platelet yield over 100%” should be “Some of the platelet yields were over 100%”

43. In line 686: “whichever the animal model” would be better “in either model”

44. In lines 687-688: “any thrombus in any of them” would be better “any evidence of thrombus formation”

45. In line 691: “the incapacity of the treated blood to coagulate” would be better “the inability of the treated blood to coagulate”

46. In lines 699-700: “allowed not only RBCs concentration but also platelet one” would be better “allowed to concentrate not only RBCs but also platelets”

47. In line 701: “pro-coagulant effect nor cause” should be “pro-coagulant effect nor did cause”

48. In line 702: “cardiac or visceral” should be “cardiac or visceral bleeding”

6. PLOS authors have the option to publish the peer review history of their article (what does this mean?). If published, this will include your full peer review and any attached files.

If you choose “no”, your identity will remain anonymous but your review may still be made public.

Do you want your identity to be public for this peer review? For information about this choice, including consent withdrawal, please see our Privacy Policy.

Reviewer #1: No

Reviewer #2: No

Reviewer #3: No

While revising your submission, please upload your figure files to the Preflight Analysis and Conversion Engine (PACE) digital diagnostic tool, https://pacev2.apexcovantage.com/. PACE helps ensure that figures meet PLOS requirements. To use PACE, you must first register as a user. Registration is free. Then, login and navigate to the UPLOAD tab, where you will find detailed instructions on how to use the tool. If you encounter any issues or have any questions when using PACE, please email PLOS at figures@plos.org. Please note that Supporting Information files do not need this step.

We have check that the figures meet the PLOS requirement using PACE diagnostic tool.

Attachment

Submitted filename: Response to Reviewers.docx

Decision Letter 1

Ahmet Emre Eşkazan

26 Oct 2021

PONE-D-21-23852R1A novel autotransfusion device saving erythrocytes and platelets by filtration used in a 72 h survival swine model of controlled blood loss: perioperative hematologic and coagulation assessments, salvaged blood characteristics and system performancePLOS ONE

Dear Dr. Touzot-Jourde,

Thank you for submitting your manuscript to PLOS ONE. After careful consideration, we feel that it has merit but does not fully meet PLOS ONE’s publication criteria as it currently stands. Therefore, we invite you to submit a revised version of the manuscript that addresses the points raised during the review process. Please submit your revised manuscript by Dec 10 2021 11:59PM. If you will need more time than this to complete your revisions, please reply to this message or contact the journal office at plosone@plos.org. When you're ready to submit your revision, log on to https://www.editorialmanager.com/pone/ and select the 'Submissions Needing Revision' folder to locate your manuscript file.

Please include the following items when submitting your revised manuscript:

  • A rebuttal letter that responds to each point raised by the academic editor and reviewer(s). You should upload this letter as a separate file labeled 'Response to Reviewers'.

  • A marked-up copy of your manuscript that highlights changes made to the original version. You should upload this as a separate file labeled 'Revised Manuscript with Track Changes'.

  • An unmarked version of your revised paper without tracked changes. You should upload this as a separate file labeled 'Manuscript'.

If you would like to make changes to your financial disclosure, please include your updated statement in your cover letter. Guidelines for resubmitting your figure files are available below the reviewer comments at the end of this letter.

If applicable, we recommend that you deposit your laboratory protocols in protocols.io to enhance the reproducibility of your results. Protocols.io assigns your protocol its own identifier (DOI) so that it can be cited independently in the future. For instructions see: https://journals.plos.org/plosone/s/submission-guidelines#loc-laboratory-protocols. Additionally, PLOS ONE offers an option for publishing peer-reviewed Lab Protocol articles, which describe protocols hosted on protocols.io. Read more information on sharing protocols at https://plos.org/protocols?utm_medium=editorial-email&utm_source=authorletters&utm_campaign=protocols.

We look forward to receiving your revised manuscript.

Kind regards,

Ahmet Emre Eşkazan, M.D.

Academic Editor

PLOS ONE

Journal Requirements:

Please review your reference list to ensure that it is complete and correct. If you have cited papers that have been retracted, please include the rationale for doing so in the manuscript text, or remove these references and replace them with relevant current references. Any changes to the reference list should be mentioned in the rebuttal letter that accompanies your revised manuscript. If you need to cite a retracted article, indicate the article’s retracted status in the References list and also include a citation and full reference for the retraction notice.

[Note: HTML markup is below. Please do not edit.]

Reviewers' comments:

Reviewer's Responses to Questions

Comments to the Author

1. If the authors have adequately addressed your comments raised in a previous round of review and you feel that this manuscript is now acceptable for publication, you may indicate that here to bypass the “Comments to the Author” section, enter your conflict of interest statement in the “Confidential to Editor” section, and submit your "Accept" recommendation.

Reviewer #2: All comments have been addressed

Reviewer #3: All comments have been addressed

**********

2. Is the manuscript technically sound, and do the data support the conclusions?

The manuscript must describe a technically sound piece of scientific research with data that supports the conclusions. Experiments must have been conducted rigorously, with appropriate controls, replication, and sample sizes. The conclusions must be drawn appropriately based on the data presented.

Reviewer #2: Yes

Reviewer #3: Yes

**********

3. Has the statistical analysis been performed appropriately and rigorously?

Reviewer #2: Yes

Reviewer #3: N/A

**********

4. Have the authors made all data underlying the findings in their manuscript fully available?

The PLOS Data policy requires authors to make all data underlying the findings described in their manuscript fully available without restriction, with rare exception (please refer to the Data Availability Statement in the manuscript PDF file). The data should be provided as part of the manuscript or its supporting information, or deposited to a public repository. For example, in addition to summary statistics, the data points behind means, medians and variance measures should be available. If there are restrictions on publicly sharing data—e.g. participant privacy or use of data from a third party—those must be specified.

Reviewer #2: Yes

Reviewer #3: Yes

**********

5. Is the manuscript presented in an intelligible fashion and written in standard English?

PLOS ONE does not copyedit accepted manuscripts, so the language in submitted articles must be clear, correct, and unambiguous. Any typographical or grammatical errors should be corrected at revision, so please note any specific errors here.

Reviewer #2: Yes

Reviewer #3: Yes

**********

6. Review Comments to the Author

Please use the space provided to explain your answers to the questions above. You may also include additional comments for the author, including concerns about dual publication, research ethics, or publication ethics. (Please upload your review as an attachment if it exceeds 20,000 characters)

Reviewer #2: (No Response)

Reviewer #3: I believe the paper was significantly improved with the revision.

There are however some points that need to be addressed, as follows.

Major points

The title is too long. Is it possible to shorten?

The Materials and Methods section is too long, and very exhaustive to read. It would benefit of shortening to about a half or 1/3, eventually by moving part to the supplemental material.

Compared to the visceral model, the platelet concentration was higher in the cardiac model, but as expected, the platelet yield was lower, which can be attributed to the deleterious effect of CBP on platelets, as discussed by the authors. The authors mention that there was no platelet activation during the treatment, and platelet kept the ability to be activated by treatment, confirmed in their previous study with human blood. Did the authors confirm this finding with the pig platelets? If yes, the results should be described, and if not, it should be mentioned in the Discussion.

In Page 30, Line 684, there is mention to the hemolysis rate higher than the 0.8% threshold. However, there is no mention to the expected hemolysis rate in the Materials and Methods section. It should be described in Page 17, together with the other expected device performance indexes.

Minor points

In Page 14, Lines 326-328: [“Use of International Standard ISO 10993-1”, “Biological evaluation of medical devices – Part 1: Evaluation and testing within a risk management process”] should be [Use of International Standard ISO 10993-1, “Biological evaluation of medical devices – Part 1: Evaluation and testing within a risk management process”]

In Page 19, Line 450: (Table S1) should be (Tables S1 and S2, respectively)

In Page 25, Lines 586 and 587: [61.7 % (50.1-71.9)] should be [61.7 (50.1-71.9) %] and [70.5 % (65.4-72.1)] should be [70.5 (65.4-72.1) %]

In Table 4: the hemoglobin concentration should be 80 and 61 (g/L) and not 8.0 and 6.1

In Page 27, Line 612: [complication or nor] should be [complication nor]

In Page 28, Line 625: [ovine RBCs cells] should be [ovine RBCs]

In Page 37, Line 857: is [Internet] necessary? Please check

In Page 37, Line 868: the link is incomplete. Please check

In Page 39, Line 902: the link is incomplete. Please check

In Page 39, Line 904: is [Internet] necessary? Please check

**********

7. PLOS authors have the option to publish the peer review history of their article (what does this mean?). If published, this will include your full peer review and any attached files.

If you choose “no”, your identity will remain anonymous but your review may still be made public.

Do you want your identity to be public for this peer review? For information about this choice, including consent withdrawal, please see our Privacy Policy.

Reviewer #2: No

Reviewer #3: No

[NOTE: If reviewer comments were submitted as an attachment file, they will be attached to this email and accessible via the submission site. Please log into your account, locate the manuscript record, and check for the action link "View Attachments". If this link does not appear, there are no attachment files.]

While revising your submission, please upload your figure files to the Preflight Analysis and Conversion Engine (PACE) digital diagnostic tool, https://pacev2.apexcovantage.com/. PACE helps ensure that figures meet PLOS requirements. To use PACE, you must first register as a user. Registration is free. Then, login and navigate to the UPLOAD tab, where you will find detailed instructions on how to use the tool. If you encounter any issues or have any questions when using PACE, please email PLOS at figures@plos.org. Please note that Supporting Information files do not need this step.

PLoS One. 2022 Mar 24;17(3):e0260855. doi: 10.1371/journal.pone.0260855.r004

Author response to Decision Letter 1


2 Nov 2021

Please find below the responses to Reviewer 3.

Thank you very much for your comments and suggestions.

We have done our best to follow them.

Best regards

Reviewer #3: I believe the paper was significantly improved with the revision.

There are however some points that need to be addressed, as follows.

Major points

- The title is too long. Is it possible to shorten?

The title has been shortened and is now as follow :

“A novel autotransfusion device saving erythrocytes and platelets used in a 72h survival swine model of surgically induced controlled blood loss”

- The Materials and Methods section is too long, and very exhaustive to read. It would benefit of shortening to about a half or 1/3, eventually by moving part to the supplemental material

The Material and Methods section has been shortened. The details on the anesthesia and postoperative management have been transferred in the supplemental data section.

- Compared to the visceral model, the platelet concentration was higher in the cardiac model, but as expected, the platelet yield was lower, which can be attributed to the deleterious effect of CBP on platelets, as discussed by the authors.

Thank you for the acknowledgement

- The authors mention that there was no platelet activation during the treatment, and platelet kept the ability to be activated by treatment, confirmed in their previous study with human blood. Did the authors confirm this finding with the pig platelets? If yes, the results should be described, and if not, it should be mentioned in the Discussion.

We could not test platelet activation in the transfusion bag after the treatment because we had no access to specific platelet activation evaluation test. Indeed, plasma protein are washed and eliminated in the treated blood. Then fibrinogen was eliminated and aggregation tests with the ROTEM platelets were not suitable. Flow cytometry could have been used as in the previous study on human blood. However, no flow cytometer was available on the experimentation site. The following sentence was added to the manuscript:

“The reinfusion of platelets is considered as a potential prothrombotic risk. However, it was shown that the i-SEP ATS does not activate platelets during the treatment, and that platelets keep their ability to be activated after the treatment [30]. However, platelet activation was not evaluated in this animal study.”

- In Page 30, Line 684, there is mention to the hemolysis rate higher than the 0.8% threshold. However, there is no mention to the expected hemolysis rate in the Materials and Methods section. It should be described in Page 17, together with the other expected device performance indexes.

The 0.8% refers to the maximal expected hemolysis in packed red cells defined by the European guidelines (European Directorate for the Quality of Medicines and HealthCare of the Council of Europe (EDQM): Guide to the Preparation, Use and Quality Assurance of Blood Components, 20th edition. Strasbourg, France, Council of Europe Publishing, 2020). The reference of the guidelines and the references are added to the manuscript as follow :

“Despite the hemolysis rate higher than the 0.8% threshold (as defined by the European guidelines, reference added) and the RBCs yield below 80 %, the treated blood reached the 45 % to 65 % hematocrit reference range and its reinfusion was well tolerated by the animals.”

Minor points

In Page 14, Lines 326-328: [“Use of International Standard ISO 10993-1”, “Biological evaluation of medical devices – Part 1: Evaluation and testing within a risk management process”] should be [Use of International Standard ISO 10993-1, “Biological evaluation of medical devices – Part 1: Evaluation and testing within a risk management process”]

In Page 19, Line 450: (Table S1) should be (Tables S1 and S2, respectively)

In Page 25, Lines 586 and 587: [61.7 % (50.1-71.9)] should be [61.7 (50.1-71.9) %] and [70.5 % (65.4-72.1)] should be [70.5 (65.4-72.1) %]

In Table 4: the hemoglobin concentration should be 80 and 61 (g/L) and not 8.0 and 6.1

In Page 27, Line 612: [complication or nor] should be [complication nor]

In Page 28, Line 625: [ovine RBCs cells] should be [ovine RBCs]

In Page 37, Line 857: is [Internet] necessary? Please check

In Page 37, Line 868: the link is incomplete. Please check

In Page 39, Line 902: the link is incomplete. Please check

In Page 39, Line 904: is [Internet] necessary? Please check

All the minor points have been addressed.

Concerning internet links, after a new careful reading of the guidelines, our interpretation is that the journal prefers having the full link. However it is a misreading on our part, we have identified the section to be remove in yellow in the document “revised manuscript”

Attachment

Submitted filename: Response to Reviewer.docx

Decision Letter 2

Ahmet Emre Eşkazan

18 Nov 2021

A novel autotransfusion device saving erythrocytes and platelets used in a 72 h survival swine model of surgically induced controlled blood loss

PONE-D-21-23852R2

Dear Dr. Touzot-Jourde,

We’re pleased to inform you that your manuscript has been judged scientifically suitable for publication and will be formally accepted for publication once it meets all outstanding technical requirements.

There some inconsistencies regarding the references, and please carefully check the reference numbers to be listed sequentially throughout the paper during the proof editing. Also there are some typos, which needs to be checked and corrected.

Within one week, you’ll receive an e-mail detailing the required amendments. When these have been addressed, you’ll receive a formal acceptance letter and your manuscript will be scheduled for publication.

An invoice for payment will follow shortly after the formal acceptance. To ensure an efficient process, please log into Editorial Manager at http://www.editorialmanager.com/pone/, click the 'Update My Information' link at the top of the page, and double check that your user information is up-to-date. If you have any billing related questions, please contact our Author Billing department directly at authorbilling@plos.org.

If your institution or institutions have a press office, please notify them about your upcoming paper to help maximize its impact. If they’ll be preparing press materials, please inform our press team as soon as possible -- no later than 48 hours after receiving the formal acceptance. Your manuscript will remain under strict press embargo until 2 pm Eastern Time on the date of publication. For more information, please contact onepress@plos.org.

Kind regards,

Ahmet Emre Eşkazan, M.D.

Academic Editor

PLOS ONE

Additional Editor Comments (optional):

Reviewers' comments:

Reviewer's Responses to Questions

Comments to the Author

1. If the authors have adequately addressed your comments raised in a previous round of review and you feel that this manuscript is now acceptable for publication, you may indicate that here to bypass the “Comments to the Author” section, enter your conflict of interest statement in the “Confidential to Editor” section, and submit your "Accept" recommendation.

Reviewer #3: All comments have been addressed

**********

2. Is the manuscript technically sound, and do the data support the conclusions?

The manuscript must describe a technically sound piece of scientific research with data that supports the conclusions. Experiments must have been conducted rigorously, with appropriate controls, replication, and sample sizes. The conclusions must be drawn appropriately based on the data presented.

Reviewer #3: Yes

**********

3. Has the statistical analysis been performed appropriately and rigorously?

Reviewer #3: N/A

**********

4. Have the authors made all data underlying the findings in their manuscript fully available?

The PLOS Data policy requires authors to make all data underlying the findings described in their manuscript fully available without restriction, with rare exception (please refer to the Data Availability Statement in the manuscript PDF file). The data should be provided as part of the manuscript or its supporting information, or deposited to a public repository. For example, in addition to summary statistics, the data points behind means, medians and variance measures should be available. If there are restrictions on publicly sharing data—e.g. participant privacy or use of data from a third party—those must be specified.

Reviewer #3: Yes

**********

5. Is the manuscript presented in an intelligible fashion and written in standard English?

PLOS ONE does not copyedit accepted manuscripts, so the language in submitted articles must be clear, correct, and unambiguous. Any typographical or grammatical errors should be corrected at revision, so please note any specific errors here.

Reviewer #3: Yes

**********

6. Review Comments to the Author

Please use the space provided to explain your answers to the questions above. You may also include additional comments for the author, including concerns about dual publication, research ethics, or publication ethics. (Please upload your review as an attachment if it exceeds 20,000 characters)

Reviewer #3: I believe the comments were appropriately addressed, and now it is easier to read.

There are only small corrections that need to be done before acceptance, as follows:

Due to restructuring/shortening of the Materials and Methods section, the references need to be renumbered. It jumps from [30] in Line 130 to [42] in Line 167.

In Line 480 (Table 4), the median Hemoglobin concentration value was appropriately corrected, but the min – max values also need correction: [8.0] [6.1] should be [80] [61]

In Line 607, reference [30] is cited 3 times [30][30][30] → please check

**********

7. PLOS authors have the option to publish the peer review history of their article (what does this mean?). If published, this will include your full peer review and any attached files.

If you choose “no”, your identity will remain anonymous but your review may still be made public.

Do you want your identity to be public for this peer review? For information about this choice, including consent withdrawal, please see our Privacy Policy.

Reviewer #3: No

Acceptance letter

Ahmet Emre Eşkazan

15 Mar 2022

PONE-D-21-23852R2

A novel autotransfusion device saving erythrocytes and platelets used in a 72 h survival swine model of surgically induced controlled blood loss.

Dear Dr. Touzot-Jourde:

I'm pleased to inform you that your manuscript has been deemed suitable for publication in PLOS ONE. Congratulations! Your manuscript is now with our production department.

If your institution or institutions have a press office, please let them know about your upcoming paper now to help maximize its impact. If they'll be preparing press materials, please inform our press team within the next 48 hours. Your manuscript will remain under strict press embargo until 2 pm Eastern Time on the date of publication. For more information please contact onepress@plos.org.

If we can help with anything else, please email us at plosone@plos.org.

Thank you for submitting your work to PLOS ONE and supporting open access.

Kind regards,

PLOS ONE Editorial Office Staff

on behalf of

Dr. Ahmet Emre Eşkazan

Academic Editor

PLOS ONE

Associated Data

    This section collects any data citations, data availability statements, or supplementary materials included in this article.

    Supplementary Materials

    S1 File

    (DOCX)

    Attachment

    Submitted filename: Response to Reviewers.docx

    Attachment

    Submitted filename: Response to Reviewer.docx

    Data Availability Statement

    All relevant data are within the manuscript and its Supporting Information files. Additional data may be ask directly to the corresponding author.


    Articles from PLoS ONE are provided here courtesy of PLOS

    RESOURCES