. 2022 Mar 28;26:85. doi: 10.1186/s13054-022-03961-x

Recruitment-to-inflation ratio measured with modern intensive care unit ventilators: How accurate is it?

Martin Cour 1,2,✉,#, Charlotte Biscarrat 1,#, Neven Stevic 1,2, Florian Degivry 1, Laurent Argaud 1,2, Claude Guérin 1,2
PMCID: PMC8962219  PMID: 35351182

The recruitment-to-inflation (R/I) ratio is a recent tool that has been developed to evaluate the potential for lung recruitment in patients with acute respiratory distress syndrome (ARDS) [1]. It is calculated as the ratio between the compliance of the recruited lung following the application of a high positive end expiratory pressure (PEEP) to that of the respiratory system measured at lower PEEP. This parameter can be easily measured at bedside with any intensive care unit (ICU) ventilator [1]. Identifying ARDS patients with high potential for lung recruitment is important to help choosing ventilatory settings, particularly the PEEP level [1]. In the landmark study by Chen et al. [1], a cut-off value of R/I at 0.5, i.e. the median value of the cohort, was proposed to define patients with low (R/I ≤ 0.5) or high (R/I > 0.5) potential for lung recruitment. Since then, the R/I ratio has been used for phenotyping ARDS [2] or to assess the effects of interventions (e.g. prone positioning [3], lung recruitment maneuvers [4]) according to the potential for lung recruitment. Before using this very promising tool at a large scale in trials, it is important to verify the accuracy and the consistency of its values across the ICU ventilators. Indeed, errors in measures of both volumes and pressures are common with most modern ICU ventilators, even after careful calibration, and often excess 10% [5]. In the present bench study, we aimed to assess accuracy of modern ICU ventilators for measuring R/I ratio set at 0.0, 0.5 and 1.0 on a lung model simulator.

We used an ASL-5000 lung model (Ingmar Medical, Pittsburgh, PA) to simulate PEEP-induced recruited lung volume (Vrec) by modifying the compliance of the lung model at high PEEP, in order to obtain R/I ratios equal to 0.0, 0.5 or 1.0. At low PEEP, the compliance of the test lung was set at 40 ml/cmH2O (i.e., a common value in ARDS). At high PEEP, the compliance was either unchanged (40 ml/cmH2O) or increased to 60 or 80 ml/cmH2O to obtain the abovementioned R/I ratios. Thus, for a 10 cmH2O difference between low and high PEEP, the expected Vrec was 0, 200 and 400 ml, respectively.

Five modern ICU ventilators were assessed: Carestation (General Electric, Fairfield, CO), Servo I (Maquet, Solna, Sweden), Hamilton C5 (Hamilton, Rhäzuns, Switzerland), Infinity C500 (Dräger, Lübeck, Germany) and Evita XL (Dräger, Lübeck, Germany). Ventilators were fully checked and calibrated according to the manufacturers’ specifications. The Y-piece of the double-limb ventilator tubing was directly connected to the ASL-5000. Ventilators were set in volume control mode with tidal volume (VT) 400 ml, inspiratory flow 60 l/min, respiratory rate 20 breaths/min, and FIO2 21%. The low and high PEEP were set at 5 and 15 cmH2O, respectively. R/I ratios were calculated from the data measured by the pressure transducers and flowmeters of the respirators, as previously described [1]. They included plateau pressure, PEEP total, expired VT, and end-expiratory lung volume change when PEEP was abruptly decreased from 15 to 5 cmH2O on a single breath (Table 1). Measures of parameters needed to calculate R/I ratio were performed twice with each ventilator. The differences between the measured and the theoretical R/I ratios were calculated for each ventilator.

Table 1.

Measures of R/I ratio with 5 ventilators according to theoretical R/I ratio equal to 0.0, 0.5 and 1.0

Theoretical R/I ratio Ventilator PEEP 5 cmH2O PEEP 15 cmH2O PEEP 15-to-5 Calculated values
VTe (ml) PPlat (cmH2O) PEEPt (cmH2O) VTe (ml) PPlat (cmH2O) PEEPt (cmH2O) EELV (ml) Vrec (ml) R/I Error in R/I
0.0 A 428 14 5 412 23 15 921 33 0.07 + 0.07
429 14 5 413 24 15 921 28 0.06 + 0.06
B 397 14 5 386 23 15 831 4 0.01 + 0.01
385 14 5 381 23 15 825 16 0.04 + 0.04
C 404 14.3 5.2 407 23.8 15.2 898 47 0.11 + 0.11
402 14.4 5.3 397 24.2 15.4 877 34 0.08 + 0.08
D 350 14.9 5.4 353 24.5 15.3 747 29 0.08 + 0.08
358 14.7 5.4 354 24.2 15.4 776 37 0.10 + 0.10
E 370 14 6 380 24 15 797 1 0.00 + 0.00
385 14 6 381 24 15 790 − 24 − 0.05 − 0.05
0.5 A 420 14 5 419 21 15 1172 286 0.61 + 0.11
429 14 5 414 21 15 1162 271 0.57 + 0.07
B 397 14 5 392 21 15 1060 227 0.51 + 0.01
385 14 5 384 21 15 1045 223 0.55 + 0.05
C 404 14.3 5.2 405 21.8 15.6 1144 277 0.60 + 0.10
402 14.4 5.3 398 21.0 15.1 1121 290 0.67 + 0.17
D 350 14.9 5.4 350 22.5 15.9 977 230 0.59 + 0.09
358 14.7 5.4 355 22.0 15.9 951 192 0.47 − 0.03
E 370 14 6 365 22 16 950 123 0.26 − 0.24
385 14 6 385 22 16 1010 144 0.30 − 0.20
1.0 A 428 14 5 416 20 16 1408 471 0.90 − 0.10
429 14 5 415 20 16 1406 467 0.89 − 0.11
B 397 14 5 393 20 15 1269 435 0.99 − 0.01
385 14 5 385 20 15 1258 445 1.04 + 0.04
C 404 14.3 5.2 397 20.4 15.9 1365 493 1.03 + 0.03
402 14.4 5.3 394 20.6 15.9 1322 460 0.98 − 0.02
D 350 14.9 5.4 353 22.1 16.4 1140 382 0.94 − 0.06
358 14.7 5.4 351 21.0 16.5 1137 359 0.84 − 0.16
E 370 14 6 387 21 16 1220 370 0.80 − 0.20
385 14 6 385 20 16 1220 353 0.73 − 0.27

Values are expressed as number and reported as displayed by the ventilator (i.e. with or without a decimal expansion). The ventilators were set in volume control mode with a tidal volume of 400 ml. Two independent set of measures are reported for each of the 5 ventilators (A: Carestation, B: Servo I, C: Hamilton C5, D: Infinity C500; E: EVITA XL)

R/I, recruitment-to-inflation ratio; PEEP, positive end-expiratory pressure; PEEP 15-to-5, decrease in PEEP from 15 to 5 cmH2O on a single breath; VTe, expired tidal volume; PPlat, plateau pressure; PEEPt, total PEEP; EELV, end-expiratory lung volume; Vrec, recruited lung volume

As shown in Table 1, R/I ratios were overestimated by 4/5 of the ventilators for theoretical R/I of 0.0 or 0.5 and underestimated by 3/5 of the ventilators for theoretical R/I of 1.0 (Table 1). For the theoretical R/I = 0.5 (i.e. the value commonly used in the literature to discriminate recruiters and non-recruiters), the error in the measured R/I was > 0.05 (> 10%) with 4/5 ventilators and > 0.1 (> 20%) with 3/5 ventilators (Table 1). For this condition, the highest overestimation of R/I was + 0.17 (+ 34%), the highest underestimation of R/I was − 0.24 (− 48%) and the highest error in Vrec was + 90 ml (+ 45%). The highest difference in R/I between 2 ventilators was 0.4.

The present study highlighted that clinically relevant underestimations or overestimations of the R/I ratio and/or of the Vrec are common when measured with modern ICU ventilators. Several clinical studies used a single cut-off value of R/I to discriminate groups of patients with low or high potential for lung recruitment [14]. However, our results showing large errors in the measurements of the true values of R/I, despite highly standardized bench conditions, suggest that using a single cut-off R/I value to individualize treatments of a given patient may be inappropriate and could even lead to opposite therapeutic strategies.

Clinicians should be aware of a range of values around a given cut-off R/I value (i.e. a grey zone) for which no conclusion may be drawn concerning potential for lung recruitment, especially if they use different models of ventilators in the same ICU. These insights need to be taken into account for interpreting and/or designing future studies on R/I ratios.

Acknowledgements

Not applicable.

Abbreviations

ARDS

Acute respiratory distress syndrome

ICU

Intensive care unit

PEEP

Positive end expiratory pressure

R/I

Recruitment-to-inflation ratio

Vrec

Recruited lung volume

Author contributions

Pr MC and Pr CG had full access to all of the data in the study and take responsibility for the integrity of the data and the accuracy of the data analysis. Pr MC, Dr CB, Dr NS, Mr FD, Pr LA, and Pr CG participated in acquisition, analysis, or interpretation of data. Pr MC and Pr CG designed the study and drafted the manuscript. Dr CB, Dr NS, Mr FD and Pr LA provided critical revision of the manuscript for important intellectual content. All authors read and approved the final manuscript.

Funding

Not applicable.

Availability of data and materials

The data that support the findings of this study are available from the corresponding author. The research team will provide an email address for communication of the data.

Declarations

Ethics approval and consent to participate

Not applicable (bench study).

Consent for publication

All authors consent to the publication of the manuscript in Critical Care, should the article be accepted by Editor-in-chief upon completion of the refereeing process.

Competing interests

The authors declare no competing interests.

Footnotes

Publisher's Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Martin Cour and Charlotte Biscarrat have contributed equally to this work

References

  • 1.Chen L, Del Sorbo L, Grieco DL, Junhasavasdikul D, Rittayamai N, Soliman I, et al. Potential for lung recruitment estimated by the recruitment-to-inflation ratio in acute respiratory distress syndrome. A clinical trial. Am J Respir Crit Care Med. 2020;201:178–187. doi: 10.1164/rccm.201902-0334OC. [DOI] [PubMed] [Google Scholar]
  • 2.Haudebourg A-F, Perier F, Tuffet S, de Prost N, Razazi K, Mekontso Dessap A, et al. Respiratory mechanics of COVID-19—versus non–COVID-19—associated acute respiratory distress syndrome. Am J Respir Crit Care Med. 2020;202:287–290. doi: 10.1164/rccm.202004-1226LE. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 3.Cour M, Bussy D, Stevic N, Argaud L, Guérin C. Differential effects of prone position in COVID-19-related ARDS in low and high recruiters. Intensive Care Med. 2021;47:1044–1046. doi: 10.1007/s00134-021-06466-3. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 4.Zerbib Y, Lambour A, Maizel J, Kontar L, De Cagny B, Soupison T, et al. Respiratory effects of lung recruitment maneuvers depend on the recruitment-to-inflation ratio in patients with COVID-19-related acute respiratory distress syndrome. Crit Care. 2022;26:12. doi: 10.1186/s13054-021-03876-z. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 5.Garnier M, Quesnel C, Fulgencio J-P, Degrain M, Carteaux G, Bonnet F, et al. Multifaceted bench comparative evaluation of latest intensive care unit ventilators. Br J Anaesth. 2015;115:89–98. doi: 10.1093/bja/aev028. [DOI] [PubMed] [Google Scholar]

Associated Data

This section collects any data citations, data availability statements, or supplementary materials included in this article.

Data Availability Statement

The data that support the findings of this study are available from the corresponding author. The research team will provide an email address for communication of the data.


Articles from Critical Care are provided here courtesy of BMC

RESOURCES