Abstract
Background
In spinal oncology, titanium implants pose several challenges including artifact on advanced imaging and therapeutic radiation perturbation. To mitigate these effects, there has been increased interest in radiolucent carbon fiber (CF) and CF-reinforced polyetheretherketone (CFR-PEEK) implants as an alternative for spinal reconstruction. This study surveyed the members of the North American Spine Society (NASS) section of Spinal Oncology to query their perspectives regarding the clinical utility, current practice patterns, and recommended future directions of radiolucent spinal implants.
Methods
In February 2021, an anonymous survey was administered to the physicians of the NASS section of Spinal Oncology. Participation in the survey was optional. The survey contained 38 items including demographic questions as well as multiple-choice, yes/no questions, Likert rating scales, and short free-text responses pertaining to the “clinical concept”, “efficacy”, “problems/complications”, “practice pattern”, and “future directions” of radiolucent spinal implants.
Results
Fifteen responses were received (71.4% response rate). Six of the participants (40%) were neurosurgeons, eight (53.3%) were orthopedic surgeons, and one was a spinal radiation oncologist. Overall, there were mixed opinions among the specialists. While several believed that radiolucent spinal implants provide substantial benefits for the detection of disease recurrence and radiation therapy options, others remained less convinced. Ongoing concerns included high costs, low availability, limited cervical and percutaneous options, and suboptimal screw and rod designs. As such, participants estimated that they currently utilize these implants for 27.3% of anterior and 14.7% of all posterior reconstructions after tumor resection.
Conclusion
A survey of the NASS section of Spinal Oncology found a lack of consensus with regards to the imaging and radiation benefits, and several ongoing concerns about currently available options. Therefore, routine utilization of these implants for anterior and posterior spinal reconstructions remains low. Future investigations are warranted to practically validate these devices’ theoretical risks and benefits.
Keywords: Carbon fiber implants, Carbon fiber-reinforced PEEK, Radiolucent implants, Radiotherapy, Spinal metastases, Spine tumors, Spinal oncology
Introduction
With the recent advances in surgical options, radiation therapies, and chemotherapeutic options available for the treatment of primary and metastatic spinal lesions, treatment algorithms have evolved and now emphasize greater multidisciplinary collaboration [1,2]. Surgical intervention for resection and subsequent reconstruction is one such modality often used in patients with signs and symptoms of instability, mechanical pain, or neurological deficits [3,4]. Traditionally, reconstruction has been performed with the use of titanium instrumentation due to its reliable strength, stiffness, and ease of use given ubiquity throughout most facets of spine surgery. In the context of spinal oncology, however, titanium implants pose several challenges. Specifically, these implants impart significant artifact on advanced imaging modalities, thus decreasing the reliability of postoperative surveillance for disease recurrence or progression [5,6]. Furthermore, due to its higher density and significant compositional differences from cortical bone, titanium produces perturbation effects which may influence the dosimetry of adjuvant radiation therapies, in addition to complicating target delineation due artifact [7].
Recent years have seen increased interest in carbon fiber (CF) and CF reinforced (CFR) polyetheretherketone (PEEK) spinal implants (Fig. 1). Unlike titanium, CF and CFR-PEEK does not impede imaging or perturb radiation, improving the accuracy of dosing and delivery [8], [9], [10]. Furthermore, these radiolucent implants have a modulus of elasticity closer to that of compact cortical bone, decreasing the stress at the implant-host bone interface [6,11]. This is especially important among patients with osteoporotic or diseased bone as it decreases the risk for subsequent implant-related mechanical complications such as subsidence, loosening, and fracture [11,12]. However, the current literature on CF and CFR-PEEK radiolucent implants is sparse for both biomechanical and clinical studies, and these implants have their own set of disadvantages. For example, lower tensile strength may contribute to failure with compressive and shear forces, implant failure is difficult to assess, and the implant options available remain limited [13,14].
Fig. 1.
Preoperative (A) lateral radiograph, (B) sagittal CT, and (C) sagittal T2 MRI demonstrating a lytic lesion in the L2 vertebral body with pathologic fracture consistent with fibrous dysplasia. (D) Lateral radiograph 12 months postoperatively following L1 to L3 posterior instrumented fusion and L2 lateral extracavitary corpectomy with reconstruction using a carbon fiber radiolucent corpectomy cage. (E) Sagittal and (F) axial T2 MRI 5 months postoperatively demonstrating minimal artifact.
Given the limited data that currently exists pertaining to these radiolucent implants, this study sought to query the perspectives of the specialists in the North American Spine Society (NASS) section of Spinal Oncology with regards to their clinical utility, current practice patterns, and recommended future directions. Given the lack of data in the literature, our a priori hypothesis was that opinions wouldl vary widely even among these specialists.
Methods
This investigation was approved as exempt research by Rush University Medical Center's Institutional Review Board. In February of 2021, an anonymous survey was administered to the 21 physicians in the NASS section of Spinal Oncology via REDCap, an encrypted online database. Participants were spinal oncologists from various disciplines, including Neurological surgery, Orthopaedic surgery, and Radiation Oncology. Participation in the survey was optional.
The survey contained 38 items including demographic questions as well as multiple-choice, yes or no questions, rating scales, and short free-text responses pertaining to the “clinical concept”, “efficacy”, “problems/complications”, “practice pattern”, and “future directions” of radiolucent spinal implants. The rating system used a 10-point Likert scale wherein 10 corresponded to the greatest, strongest positive feeling regarding the question and 0 corresponded the least, most negative feeling about the question. Statistical analysis was performed using SPSS version 26.0 (IBM Corporation, Armonk, NY). Descriptive statistics were used to provide sample characteristics. Responses to categorical questions were reported as proportions and absolute counts, while responses to questions with scaled responses were reported as means with ranges.
Results
Demographics
Overall, 15 of the 21 NASS Spinal Oncology section members completed the questionnaire, for a response rate of 71.4%. Six of the participants (40%) were neurosurgeons, 8 (53.3%) were orthopedic surgeons, and 1 was a spinal radiation oncologist. There were 9 institutions represented by the 12 physicians who answered the optional field. The average number of years in practice among participants was 9.2 years (2-16) and oncology constituted 45.3% (20%-80%) of provider's spine practice. The participants’ responses to the survey are summarized by Table 1.
Table 1.
Summary of survey responses.
| Question | Response Mean (range) |
|---|---|
| Clinical Concept | |
|
|
|
6.5 (1-10) |
|
5.8 (2-9) |
|
6.3 (0-10) |
|
9.5 (9-10) |
|
7.0 (2-10) |
|
|
|
5.6 (2-10) |
|
6.3 (2-10) |
|
|
|
33.3% (5/15) |
|
13.3% (2/15%) |
|
53.3% (8/15) |
| Efficacy | |
|
|
|
5.1 (2-7) |
|
3.1 (0-7) |
|
|
|
46.7% (7/15) |
|
26.7% (4/15) |
|
20% (3/15) |
|
0% (0/15) |
|
|
|
60% (9/15) |
|
13.3% (2/15) |
|
40% (6/15) |
|
60% (9/15) |
|
0% (0/15) |
| Problems and Complications | |
|
|
|
26.7% (4/15) |
|
53.3% (8/15) |
|
60% (9/15) |
|
80% (12/15) |
|
|
|
7.7 (2-10) |
|
6.7 (1-10) |
|
6.2 (1-10) |
|
6.1 (1-10) |
|
7.1 (2-10) |
|
8.0 (6-10) |
| Practice Pattern | |
|
|
|
5.3 (1-10) |
|
7.3 (1-10) |
|
5.0 (1-9) |
|
Yes: 40% (6/15) |
|
|
|
6.7% (1/15) |
|
53.3% (8/15) |
|
13.3% (2/15) |
|
20.% (3/15) |
|
6.7% (1/15) |
|
27.3% (0%-100%) |
|
14.7% (0%-90%) |
|
Yes: 73.3% (11/15) |
| Future Directions | |
|
34% (0%-80%) |
Clinical concept
With regards to the importance of radiolucent implants, the ability to obtain postoperative MRI/CT imaging for early recurrence detection received an average score of 6.5 (1-10) out of 10, while the ability to use photons and protons without perturbation received an average score of 5.8 (2-9) and 6.3 (0-10), respectively. Two “other” responses were received, including a statement that the radiolucent “rods do not fracture” (9/10) and that radiolucent implants “allow precise visualization of the cord in the context of instrumentation without requiring a CT myelogram” (10/10). The importance of radiolucent spinal implants to anterior column reconstruction received an average score of 7.0 (2-10), while the importance of rods and screws for posterior column reconstruction received respective scores of 5.6 (2-10) and 6.3 (2-10). When asked which feature of pedicle screws was most importantly radiolucent, 5 (33.3%) said the shank, 2 (13.3%) said the tulip, and 8 (53.3%) said both.
Efficacy
Regarding the efficacy of radiolucent spinal implants, section members rated the quality of current “basic science/in vitro” research with an average score of 5.1 (2-7), while current “clinical outcomes” research received a score of 3.1 (0-7). Seven (46.7%) said they had seen imaging quality differences using radiolucent implants in their own practice, 4 (26.7%) had noticed a difference in radiation options, and 3 (20%) had noticed a difference in radiation efficacy. Furthermore, 9 (60%) noted clinical differences in strength between all-PEEK and CF/CFR-PEEK implants, 2 (13.3%) said there were differences in radiolucency, 6 (40%) in manufacturability, and 9 (60%) in implant cost.
Problems and complications
When asked if radiolucent implants pose problems with imaging, 4 section members (26.7%) said they make it difficult to detect proper positioning, 8 (53.3%) said difficulty in detecting migration/subsidence, 9 (60%) said difficulty detecting haloing/loosening, and 12 (80%) said implant fracture/failure is difficult to detect. Additionally, the average level of concern on a scale from 0 (least) to 10 (most) with radiolucent implants with regards to the following was: cost 7.7 (2-10), availability 6.7 (1-10), yield strength 6.2 (1-10), mechanical integrity 6.1 (1-10), and “too stiff”/inability to bend 7.1 (2-10). “Other” concerns included how to revise, extend, compress, distract, and bend (10/10) carbon fiber implants, as well as “technical concerns” with radiolucent implants (6/10).
Practice pattern
On average, section members felt that radiolucent implants are more important in the context of primary spine malignancy (7.3, range: 1-10) than for metastatic carcinoma (5.3, range: 1-10) and benign aggressive spinal tumors (5.0, range: 1-9). Six (40%) said their practice pattern had changed to increase usage over the past 5-10 years, but the majority said that they rarely use radiolucent implants, and their pattern has not changed. For anterior reconstructions, participants on average utilize a radiolucent strategy for 27.3% (0%-100%) of all cases. These implants are currently used even less frequently by participants for posterior reconstructions, reportedly for 14.7% (0%-90%) of cases. Eleven (73.3%) said they are hesitant to adopt a radiolucent strategy. The reasons were provided by 10 participants and are listed in Table 2.
Table 2.
Reasons for hesitancy to adopt a radiolucent strategy.
| Response |
|---|
|
|
|
|
|
|
|
|
|
|
Future directions
When asked by how much they anticipated to increase their usage of radiolucent over the next years based on current practice, section members projected an average increase of 34.0% (0%-80%). Table 3 lists free-text responses pertaining to future directions and advances that the section members would like to see made available for radiolucent implants going forward.
Table 3.
Future directions.
| Response |
|---|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
Discussion
Although radiolucent CF and CFR-PEEK spinal implants provide many theoretical benefits that make them an enticing option for spinal oncologists, there is a paucity of long-term clinical data in the literature, and little is known regarding current usage in routine clinical practice. Therefore, this study surveyed the NASS section of Spinal Oncology, finding mixed opinions even among advanced specialists. Specifically, while several participants believed that radiolucent spinal implants provide substantial benefits for the detection of disease recurrence and radiation therapy options, others remained less convinced. Furthermore, several ongoing concerns were noted, such as high costs, low availability, limited options for cervical reconstruction, and suboptimal design and rigidity, making them more difficult to work with than titanium. As such, participants estimated that they only utilize these implants for 27.3% of anterior and 14.7% of all posterior reconstructions.
A limitation of this study is the relatively small sample size of survey participants. Additionally, all the study's participants practice in large, tertiary academic medical institutions, limiting the generalizability of the findings. For example, only a small proportion of the participants currently utilize radiolucent implants routinely, citing high costs and differences in utilization criteria and institutional contracts between implant manufacturing companies as barriers. Therefore, the estimated usage of CF and CFR-PEEK instrumentation may in fact be inflated relative to the overall spinal oncology practice. Additionally, while societies exist for various spinal conditions, there is not a dedicated society focusing on spinal oncology. The NASS spine oncology section can thus help capture the current practice patterns and opinions of surgeons with clinical and academic interest in spinal oncology.
To date, several studies have evaluated the benefits of radiolucent spinal instrumentation on imaging artifact and radiation perturbation. Mastella et al. performed an in vitro study evaluating the dosimetric impact of CFR-PEEK devices on postoperative particle therapy when compared to titanium [8]. They found that CFR-PEEK screws caused little beam perturbation when compared to titanium screws and therefore a lower degree of dose degradation. The reduced artifacts on CT also improved dose calculation accuracy. Muller et al. retrospectively compared CT data on five patients with titanium and five patients with CFR-PEEK screws reinforced with a thin titanium coating and titanium tulips, also comparing the dosimetric impact of both implants for intensity modulated proton (IMPT) and volumetric arc photon therapy (VMAT) [9]. The authors found no difference in dosimetric quality for VMAT plans between implant types. However, CFR-PEEK implants demonstrated benefits for dosing accuracy on IMPT plans. In a series of 35 patients with spinal tumors, Ringel et al. demonstrated that posterior instrumentation with CFR-PEEK implants reduced implant-induced artifact when compared to standard titanium alloy and allowed for sufficient assessment of implant position and integrity [15]. The authors concluded that they are valuable and feasible options for patients with spinal tumors where imaging and radiation are crucial for survival outcomes.
A recent systematic review by Takayanagi et al. identified fourteen clinical studies in the literature discussing patient outcomes and complication following surgical reconstruction with CF and/or CFR-PEEK instrumentation, most of which were case reports or small series with limited follow-up under two years [16]. While the authors concluded on the imaging and radiation benefits previously discussed, they also noted several drawbacks of these devices identified by the studies, including limited intraoperative visualization, difficulty with rod bending, and substantial costs. In one of the more robust clinical studies, Boriani et al. reported the intraoperative, neurological, and survival outcomes of 34 patients with tumors (fourteen metastases, twenty primary tumors) to the thoracic and lumbar spine who had thoracolumbar fixation with CFR-PEEK implants [17]. There was one screw breakage intraoperatively among the 232 implanted screws, and two instances of screw loosening at nine and twelve months in multilevel constructs due to recurrence. Furthermore, six instances of recurrence were detected early, which the authors attributed to implant radiolucency, although they do not provide evidence to support this assertion. Comparing CFR-PEEK implants in 36 patients to standard titanium implants in 42 patients, Cofano et al. found no differences in axial pain and neurologic status functional recovery or clinical and hardware-related complication rates [18].
While there is evidence in support of radiolucent implants, there have been no large-scale clinical studies reporting long-term outcomes. As such, many spinal oncologists remain hesitant to adopt the higher costs and intraoperative challenges associated with these devices until equivocal results have been established. Among this study's participants, only 46.7% noted a benefit in imaging quality in clinical practice, while 26.7% and 20% saw advantages in radiation options and efficacy, respectively. However, these numbers may be skewed by the limited use of these implants in clinical practice.
Another common theme across the responses that were received was the lack of options for cervical reconstruction, percutaneous options for minimally invasive surgery, and rod options with the ability to contour. Furthermore, there remains a discrepancy in the availability of radiolucent anterior versus posterior implants. To date, few posterior systems have been approved by the Food and Drug Administration (FDA), likely influencing the rate of utilization. The findings that 73.3% of section members said they are hesitant to adopt a radiolucent strategy with an estimated use of only 14.7% for posterior reconstructions further emphasizes the need for better data on use criteria. For example, it remains unknown whether CF impacts convention radiation therapy or if it is specifically advantageous to postoperative stereotactic body radiation therapy (SBRT) and photon therapy. This would require preoperative planning for which radiation modality would be employed to avoid unnecessary utilization of radiolucent technology and the associated expenses should the patient be admitted to hospice without adjuvant radiation or undergo conventional fractionated radiation.
Although radiolucent implants provide the potential for a paradigm shift in the world of spinal oncology, there remain obstacles impeding their routine implementation. This study has emphasized the current limitations in the usability of these devices, as well as the need for additional large-scale prospective randomized trials comparing radiolucent implants to the current standard-of-care titanium.
Conclusion
This study surveyed the NASS section of Spinal Oncology for opinions on radiolucent CF and CFR-PEEK instrumentation, finding a lack of consensus with regards to the imaging and radiation benefits. Furthermore, several ongoing concerns were noted, such as higher costs, limited cervical and percutaneous options, and suboptimal screw and rod designs, making them more difficult to work with than titanium. Therefore, routine utilization of these implants for anterior and posterior spinal reconstructions remains low. Future large-scale longitudinal investigations are warranted to further delineate the benefits and complication rates associated with these devices.
Declarations of Competing Interests
The authors have no conflicts of interest to declare. All co-authors have seen and agree with the contents of the manuscript.
Funding disclosures
No funding sources were received for the completion of the work pertaining to this study.
Footnotes
Supplementary material associated with this article can be found, in the online version, at doi:10.1016/j.xnsj.2022.100105.
Appendix. Supplementary materials
References
- 1.Goodwin CR, Abu-Bonsrah N, Rhines LD, et al. Molecular markers and targeted therapeutics in metastatic tumors of the spine: changing the treatment paradigms. Spine. 2016;41:S218–SS23. doi: 10.1097/BRS.0000000000001833. [DOI] [PubMed] [Google Scholar]
- 2.Cofano F, Monticelli M, Ajello M, et al. The targeted therapies era beyond the surgical point of view: what spine surgeons should know before approaching spinal metastases. Cancer Control. 2019;26(1) doi: 10.1177/1073274819870549. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 3.Fisher CG, DiPaola CP, Ryken TC, et al. A novel classification system for spinal instability in neoplastic disease: an evidence-based approach and expert consensus from the Spine Oncology Study Group. Spine. 2010;35(22):E1221–E12E9. doi: 10.1097/BRS.0b013e3181e16ae2. [DOI] [PubMed] [Google Scholar]
- 4.Patchell RA, Tibbs PA, Regine WF, et al. Direct decompressive surgical resection in the treatment of spinal cord compression caused by metastatic cancer: a randomised trial. Lancet North Am Ed. 2005;366(9486):643–648. doi: 10.1016/S0140-6736(05)66954-1. [DOI] [PubMed] [Google Scholar]
- 5.Zimel MN, Hwang S, Riedel ER, Healey JH. Carbon fiber intramedullary nails reduce artifact in postoperative advanced imaging. Skeletal Radiol. 2015;44(9):1317–1325. doi: 10.1007/s00256-015-2158-9. [DOI] [PubMed] [Google Scholar]
- 6.Laux CJ, Hodel SM, Farshad M, Müller DA. Carbon fibre/polyether ether ketone (CF/PEEK) implants in orthopaedic oncology. World J Surg Oncol. 2018;16(1):1–6. doi: 10.1186/s12957-018-1545-9. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 7.Nevelsky A, Borzov E, Daniel S, Bar-Deroma R. Perturbation effects of the carbon fiber-PEEK screws on radiotherapy dose distribution. J Appl Clin Med Phys. 2017;18(2):62–68. doi: 10.1002/acm2.12046. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 8.Mastella E, Molinelli S, Magro G, et al. Dosimetric characterization of carbon fiber stabilization devices for post-operative particle therapy. Physica Med. 2017;44:18–25. doi: 10.1016/j.ejmp.2017.11.008. [DOI] [PubMed] [Google Scholar]
- 9.Müller BS, Ryang Y-M, Oechsner M, et al. The dosimetric impact of stabilizing spinal implants in radiotherapy treatment planning with protons and photons: standard titanium alloy vs. radiolucent carbon-fiber-reinforced PEEK systems. J Appl Clin Med Phys. 2020;21(8):6–14. doi: 10.1002/acm2.12905. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 10.Jackson JB, III, Crimaldi AJ, Peindl R, Norton HJ, Anderson WE, Patt JC. Effect of polyether ether ketone on therapeutic radiation to the spine: a pilot study. Spine. 2017;42(1):E1–E7. doi: 10.1097/BRS.0000000000001695. [DOI] [PubMed] [Google Scholar]
- 11.Li C, Liu L, Shi J-Y, Yan K-Z, Shen W-Z, Yang Z-R. Clinical and biomechanical researches of polyetheretherketone (PEEK) rods for semi-rigid lumbar fusion: a systematic review. Neurosurg Rev. 2018;41(2):375–389. doi: 10.1007/s10143-016-0763-2. [DOI] [PubMed] [Google Scholar]
- 12.Lill H, Hepp P, Korner J, et al. Proximal humeral fractures: how stiff should an implant be? Arch Orthop Trauma Surg. 2003;123(2):74–81. doi: 10.1007/s00402-002-0465-9. [DOI] [PubMed] [Google Scholar]
- 13.Fleck NA, Jelf PM. Deformation and failure of a carbon fibre composite under combined shear and transverse loading. Acta Metall Mater. 1995;43(8):3001–3007. [Google Scholar]
- 14.McDonnell JM, Nagassima Rodrigues dos Reis K, Ahern DP, Mahon J, Butler JS. Are carbon-fiber implants more efficacious than traditional metallic implants for spine tumor surgery? Clin Spine Surg. 2021;34(5):159–162. doi: 10.1097/BSD.0000000000001007. [DOI] [PubMed] [Google Scholar]
- 15.Ringel F, Ryang Y-M, Kirschke JS, et al. Radiolucent carbon fiber–reinforced pedicle screws for treatment of spinal tumors: advantages for radiation planning and follow-up imaging. World Neurosurg. 2017;105:294–301. doi: 10.1016/j.wneu.2017.04.091. [DOI] [PubMed] [Google Scholar]
- 16.Takayanagi A, Siddiqi I, Ghanchi H, et al. Radiolucent carbon fiber–reinforced implants for treatment of spinal tumors–clinical, radiographic, and dosimetric considerations. World Neurosurg. 2021 doi: 10.1016/j.wneu.2021.05.100. [DOI] [PubMed] [Google Scholar]
- 17.Boriani S, Tedesco G, Ming L, et al. Carbon-fiber-reinforced PEEK fixation system in the treatment of spine tumors: a preliminary report. Eur Spine J. 2018;27(4):874–881. doi: 10.1007/s00586-017-5258-5. [DOI] [PubMed] [Google Scholar]
- 18.Cofano F, Di Perna G, Monticelli M, et al. Carbon fiber reinforced vs titanium implants for fixation in spinal metastases: a comparative clinical study about safety and effectiveness of the new “carbon-strategy. J Clin Neurosci. 2020;75:106–111. doi: 10.1016/j.jocn.2020.03.013. [DOI] [PubMed] [Google Scholar]
Associated Data
This section collects any data citations, data availability statements, or supplementary materials included in this article.

