A comparison of first-attempt cannulation success of peripheral venous catheter systems with and without wings and injection ports in surgical patients—a randomized trial

Rudolf Mörgeli 1,#, Katrin Schmidt 1,#, Tim Neumann 2, Jochen Kruppa 1, Ulrich Föhring 2, Pascal Hofmann 3, Peter Rosenberger 3, Elke Falk 1, Willehad Boemke 1, Claudia Spies 1,
PMCID: PMC8969381  PMID: 35361115

Abstract

Background

A peripheral venous catheter (PVC) is the most widely used device for obtaining vascular access, allowing the administration of fluids and medication. Up to 25% of adult patients, and 50% of pediatric patients experience a first-attempt cannulation failure. In addition to patient and clinician characteristics, device features might affect the handling and success rates. The objective of the study was to compare the first-attempt cannulation success rate between PVCs with wings and a port access (Vasofix® Safety, B. Braun, abbreviated hereon in as VS) with those without (Introcan® Safety, B. Braun, abbreviated hereon in as IS) in an anesthesiological cohort.

Methods

An open label, multi-center, randomized trial was performed. First-attempt cannulation success rates were examined, along with relevant patient, clinician, and device characteristics with univariate and multivariate analyses. Information on handling and adherence to use instructions was gathered, and available catheters were assessed for damage.

Results

Two thousand three hundred four patients were included in the intention to treat analysis. First-attempt success rate was significantly higher with winged and ported catheters (VS) than with the non-winged, non-ported design (IS) (87.5% with VS vs. 78.2% with IS; PChi < .001). Operators rated the handling of VS as superior (rating of “good” or “very good: 86.1% VS vs. 20.8% IS, PChi < .001). Reinsertion of the needle into the catheter after partial withdrawal—prior or during the catheterization attempt—was associated with an increased risk of cannulation failure (7.909, CI 5.989–10.443, P < .001 and 23.023, CI 10.372–51.105, P < .001, respectively) and a twofold risk of catheter damage (OR 1.999, CI 1.347–2.967, P = .001).

Conclusions

First-attempt cannulation success of peripheral, ported, winged catheters was higher compared to non-ported, non-winged devices. The handling of the winged and ported design was better rated by the clinicians. Needle reinsertions are related to an increase in rates of catheter damage and cannulation failure.

Trial registration

ClinicalTrials.gov, Identifier: NCT02213965, Date: 12/08/2014.

Supplementary Information

The online version contains supplementary material available at 10.1186/s12871-022-01631-7.

Keywords: Catheters, Catheterization, Peripheral, Cannulation, Operating room

Introduction

A peripheral venous catheter (PVC) is the most widely used device for obtaining continuous vascular access, allowing the administration of fluids and medication to a variety of patients. It is estimated that 30–80% of all hospitalized patients receive a PVC during their stay [14]. For procedural and safety reasons, vascular access in the surgical settings is virtually indispensable. Along with attachment of monitoring devices, the insertion of a PVC usually occurs immediately upon arrival in the induction or operating room, allowing for the administration of fluids, induction of anesthesia and management of potential complications.

Since establishing venous access occurs while the patient is awake, and usually precedes administration of any anesthetic agents, PVC placement can be an uncomfortable and painful procedure. Repeated and unsuccessful attempts to place a PVC can be stressful for both patient and clinician, causing localized pain and swelling, creating a portal of entry for micro-organisms, and forcing clinicians to seek alternative sites, where cannulation may be more difficult or dangerous [57]. Therefore, a successful PVC placement in the first attempt is always desirable. Reports indicate that up to a quarter of adult patients, and half of pediatric patients experience a first-attempt cannulation failure [810].

Cannulation success rates can be influenced by the venous conditions and body-mass index (BMI) of the patient, as well as the level of experience of the healthcare provider [11], but also the type of PVC used may impact success rates. There are several types of PVCs available in the European market, and features such as size, bevel type, and the presence of notches, wings and ports might affect the handling of these devices [12]. Ported and winged catheters are widespread in German hospitals, but there is no evidence that they are superior to non-ported catheters, especially in relation to first-attempt cannulation success. The objective of this study is to compare the rate of first-attempt placement success between two PVC designs, namely a ported, winged catheter (Vasofix® Safety, B. Braun Melsungen AG, Germany, abbreviated hereon in as VS) and a non-ported, non-winged catheter (Introcan® Safety, B. Braun Melsungen AG, Germany, abbreviated hereon in as IS) (see Fig. 1).

Fig. 1.

Fig. 1

Introcan® Safety (IS) and Vasofix® Safety (VS), B. Braun Melsungen AG, Germany (images retrieved from https://www.bbraun.de/de/products/b0/vasofix-safety.html and https://www.bbraun.de/de/products/b/introcan-safety.html, adapted and printed with permission from B. Braun Melsungen)

Methods

This open-label, multicenter study randomized patients to receive one of two distinct types of PVCs prior to elective surgery. This sponsor-initiated study compared two CE-marked, Class II medical devices in accordance with their intended purpose and use instructions [13, 14]. Data was collected between December 2014 and March 2016 by the anesthesiology departments of two German tertiary care university hospitals, namely the Charité – Universitätsmedizin Berlin with its three separate sites, Campus Charité Mitte (CCM), Campus Virchow Klinikum (CVK), and Campus Benjamin Franklin (CBF), and the Universitätsklinikum Tübingen. In accordance with inclusion and exclusion criteria (see Supplemental Fig. 1), adult and pediatric patients undergoing elective surgery were recruited by the anesthesiology staff during the routine pre-surgical assessment, and written informed consent was obtained from all participants or their proxies. Peripheral venous cannulation was required for surgery, so that patients were not subjected to any additional procedure or distress in the context of this investigation. All catheters were placed immediately prior to surgery, at the induction or operating room. The study was approved by the ethic committees of both participating universities (Berlin: EA2/104/14; Tübingen: 494/2014BO2), as well as by their respective data protection officers. The study was conducted in accordance with the Declaration of Helsinki and its amendments and was registered at ClinicalTrials.gov (NCT 02213965, 12/08/2014).

After providing informed consent, patients were randomly selected to receive a ported, winged fluorethylene propylene catheter (VS) or a non-ported, non-winged polyurethane catheter (IS). Study personnel prepared and conducted randomization via sealed envelopes, stratified by study site in a 1:1 ratio. General patient data was then collected, including age, gender, BMI, American Society of Anesthesiologists Physical Status (ASA-PS) score, as well as venous status (subjective assessment: good, moderate, or poor). Whenever possible, study personnel accompanied patients to the operating theater to ensure that venous cannulation took place with the designated catheter and that documentation was completed appropriately. The staff member placing the PVC (operator) completed a paper-based case report form (CRF), which included questions regarding the number and sites of cannulation attempts, subjective reasons for failure, adherence to use instructions (assessed by documenting needle reinsertion into the catheter, i.e. the practice of moving the needle stylet in the cannula prior to puncture or during the catheterization attempt; the manufacturer’s instructions for use (IFU) explicitly warn against this practice, and the variable was included as a marker for misuse), catheter handling information (e.g. 6-point Likert scale, blood spillage, use of ports), as well as their professional function and level of experience with venous catheterization. In case of a failed cannulation attempt, the catheter was collected and sent to the Bergische Universität Wuppertal, Faculty of Mechanical and Safety Engineering for a macro- and microscopic examination for signs of damage. Observation continued until patient discharge from the operating or recovery room, whereas any adverse events (e.g. catheter occlusion, displacement, or extravasation) were noted and the respective catheters collected. Used catheters were also collected following decannulation of ambulatory patients. No treatment changes occurred, all standard operating procedures remained in place throughout the investigation, and the attending physician had ultimate authority in all treatment decisions.

The primary outcome was successful cannulation at first attempt. Secondary outcomes included overall number of cannulation attempts, subjective reasons for cannulation failure, operators’ evaluation of catheter handling, frequency of needle reinsertion into the catheter, rate of PVC damage, and adverse events.

The sample size calculation, which was aimed for a power of 80%, an α of 0.05, and a 5% delta for a 75% first-attempt cannulation success for the control group (VS), yielded 1134 patients per trial arm (n = 2268). Additionally, in order to retain the prognostic balance provided by the randomization process [15], all data were analyzed according to the intention-to-treat (ITT) principle. Metric data was not normally distributed, so that data was reported as median and interquartile ranges and differences among groups were evaluated using the Wilcoxon-Mann–Whitney U test. Ordinal and nominal data were reported using frequencies and percentages, and groups were compared using Chi-Square Test, or Fisher’s Exact Test when small cell exceptions were present. Missing data was treated as such, with no replacement by estimates. Univariate and multiple logistic regression analyses was conducted to determine the most relevant factors for cannulation success at the first attempt, adjusting for possible confounders. The stepwise regression model was based on the available literature, which was supplemented by input from the stakeholders, who were familiar with the complaints and expectations from users in the field. The included variables depict patient (age, gender, BMI, cannulation site and venous condition) [16], operator (function and experience) [17], and product (catheter type and size) [17, 18] characteristics that might affect the cannulation success rates. Additionally, as part of an exploratory analysis, rate of needle reinsertion into catheters (any, before, or during the catheterization attempt) was added among operator variables as a misuse indicator, along with a detailed risk assessment for cannulation failure and catheter damage. Statistical analysis was performed with SPSS (Version 27.0, Armonk, NY: IBM Corp., USA).

Results

Among the four study sites, 2304 patients were included in the ITT analysis, as shown in Supplemental Fig. 1 (Consort Diagram). Of these patients, 1133 received an IS catheter and 1171 received a VS catheter. Overall, there were no statistically relevant differences among patients in terms of age, gender, BMI, ASA score, venous conditions, or cannulation site. However, significant differences were observed in terms of catheter size (see Table 1). Patient characteristics among sites are provided in Supplemental Tables 1 and 2.

Table 1.

All data shown as frequencies and percentages. BMI body-mass index, ASA American Society of Anesthesiologists. Missing data is treated as such

Patient Characteristics N All Introcan Vasofix P
(n = 1134) (n = 1175)
Age (years) 2304 .15
  1 to 6 34 (1.5%) 16 (1.4%) 18 (1.5%)
  7 to 12 25 (1.1%) 13 (1.1%) 12 (1.0%)
  13 to 17 43 (1.9%) 28 (2.5%) 15 (1.3%)
  18 to 44 659 (28.6%) 311 (27.4%) 348 (29.7%)
  45 to 64 840 (36.5%) 432 (38.1%) 408 (34.8%)
  ≥ 65 703 (30.5%) 333 (29.4%) 370 (31.6%)
Gender 2279 .43
  Female 1097 (48.1%) 532 (47.3%) 565 (49.0%)
BMI (adults only) 2175 .62
  < 18.5 kg/m2 51 (2.3%) 25 (2.4%) 26 (2.3%)
  18.5 to 24.9 kg/m2 951 (43.7%) 471 (44.4%) 480 (43.1%)
  25 to 29.9 kg/m2 719 (33.1%) 337 (31.7%) 382 (34.3%)
  ≥ 30 kg/m2 454 (20.9%) 229 (21.6%) 225 (20.2%)
ASA Score 2242 .65
  1 600 (26.8%) 303 (27.7%) 297 (25.8%)
  2 1287 (57.4%) 623 (57.0%) 664 (57.8%)
  3 349 (15.6%) 165 (15.1%) 184 (16.0%)
  4 6 (0.3%) 2 (0.2%) 4 (0.3%)
Venous Status 2277 .58
  Good 1583 (69.5%) 778 (69.4%) 805 (69.6%)
  Moderate 568 (24.9%) 286 (25.5%) 282 (24.4%)
  Poor 126 (5.5%) 57 (5.1%) 69 (6.0%)
Catheter Size 2295 .045
  ≤ 16 G 86 (3.7%) 32 (2.8%) 54 (4.6%)
  18 G 1185 (51.6%) 587 (52.1%) 598 (50.5%)
  20 G 951 (41.4%) 478 (42.5%) 473 (40.5%)
  ≥ 22 G 73 (3.2%) 29 (2.6%) 44 (3.8%)
Site of Cannulation 2290 .44
  Back of hand 1814 (79.2%) 905 (80.3%) 909 (78.2%)
  Forearm 339 (14.8%) 153 (13.6%) 186 (16.0%)
  Antecubital fossa 110 (4.8%) 55 (4.9%) 55 (4.7%)
  Other 27 (1.2%) 14 (1.2%) 13 (1.1%)
Operator Function 2275 .88
  Physician 834 (36.7%) 416 (37.1%) 418 (36.2%)
  Registered Nurse 1250 (54.9%) 610 (54.4%) 640 (55.5%)
  Other Specialist 191 (8.4%) 95 (8.5%) 96 (8.3%)
Operator Experience 2166 .31
  < 3 years 486 (22.4%) 248 (23.4%) 238 (21.5%)
  > 3 years 1680 (77.6%) 813 (76.6%) 867 (78.5%)

Overall, the first-attempt cannulation success was achieved in 82.9% of cases and was higher with VS (87.5%) than with IS (78.2%) (P < 0.001). This significance was confirmed in a per protocol group analysis (87.5% vs. 80.5% success rate; n = 2267, P < 0.001), as well as in a sub-analysis including only patients ≥ 18 years of age (88.2% vs. 78.0% success rate; n = 2202, P < 0.001). Relevant factors influencing first-attempt success rates are shown in Table 2. While the success rates with VS were significantly higher in patients with at least 45 years of age, differences ceased to be significant for those aged between 18 and 44, and observed success rates did not differ significantly for patients below the age of 18. There were no differences in success rates among patients with a BMI < 18.5 kg/m2, although VS was superior in all other BMI categories. VS was also superior in good and moderate vein conditions, while the success rate was not significantly different in patients with poor venous status. The majority of venous catheters were placed on the back of the hand (78.7%), where success rates with VS were significantly higher than at other cannulation sites. Here, a success rate of 89.8% was achieved with VS, compared to 78.9% with IS. No statistical differences were detected between success rates for cannulas placed in the forearm or antecubital fossa. Compared to IS, first-attempt success rates were significantly higher with VS for nurses and physicians alike (P < 0.001), although this difference could not be observed among other groups (i.e. medical students, interns). There were differences among centers regarding operator function. While physicians performed the majority of cannulations in CBF and UCT (60.7% and 54.6%, respectively), nurses performed the majority of cannulations in CCM and CVK (83.7% and 61.4%, respectively). 72.9% of operators had over 3 years of work experience, placing an average of 17 (± 9) catheters per week. Here, nurses placed significantly higher number of catheters per week in comparison to physicians (P < 0.001), and only 7% reported having less than 3 years of experience, compared to 50% of physicians. The multivariate analysis (n = 1157) identified venous conditions, BMI, operator function and experience, adherence to IFU, catheter size and catheter type as the most significant determinants of cannulation success at the first attempt (see Table 3). In a regression model without misuse markers (n = 1956, Supplemental Table 3), operator function ceased to be significant.

Table 2.

All data shown as frequencies and percentages. ITT intention to treat, BMI body-mass index. Missing data is treated as such

Successful Cannulation Introcan (IS) Vasofix (VS) P
n total n successful % n total n successful %
Total (ITT) 1133 886 78.2% 1171 1025 87.5%  < .001
Age (years)
  1 to 6 16 12 75.0% 18 10 55.6% .24
  7 to 12 13 12 92.3% 12 9 75.0% .32
  13 to 17 28 23 82.1% 15 13 86.7% .70
  18 to 44 311 257 82.6% 348 305 87.6% .07
  45 to 64 432 330 76.4% 408 365 89.5%  < .001
   ≥ 65 333 252 75.7% 370 323 87.3%  < .001
Gender
  Female 532 418 78.6% 565 474 83.9% .02
  Male 593 463 78.1% 589 538 91.3%  < .001
BMI (adults only)
  < 18.5 kg/m2 25 21 84.0% 26 23 88.5% .70
  18.5 to 24.9 kg/m2 471 365 77.5% 480 427 89.0%  < .001
  25 to 29.9 kg/m2 337 266 78.9% 382 334 87.4% .002
  ≥ 30 kg/m2 229 175 76.4% 225 197 87.6% .002
Venous Condition
  Good 778 657 84.4% 805 745 92.5%  < .001
  Moderate 286 198 69.2% 282 234 83.0%  < .001
  Poor 57 23 40.4% 69 35 50.7% .25
Site of Cannulation
  Back of hand 905 714 78.9% 909 816 89.8%  < .001
  Forearm 153 121 79.1% 186 155 83.3% .38
  Antecubital fossa 55 40 72.7% 55 41 74.5% .83
Operator Function
  Physicians 416 312 75.0% 418 364 87.1%  < .001
  Nurses 610 483 79.2% 640 560 87.5%  < .001
  Others 95 80 84.2% 96 88 91.7% .11
Operator Experience
  > 3 years 813 655 80.6% 867 762 87.9%  < .001
  < 3 years 248 175 70.6% 238 207 87.0%  < .001
Catheter Size
  ≤ 16 G 32 28 87.5% 54 52 96.3% .19
  18 G 587 472 80.4% 598 546 91.3%  < .001
  20 G 478 356 74.5% 473 399 84.4%  < .001
  ≥ 22 G 29 24 82.8% 44 26 59.1% .03

Table 3.

Multivariate analysis; VS Vasofix® Safety, IS Introcan® Safety; adult participants only; All data analyzed as categorical variables

Multivariate Analysis (Risk of Cannulation Failure) n P Odds Ratio 95% Confidence Interval
Patient Variables
  Age 1157 .74 0.998 0.988 … 1.008
  Gender, female (0) / male 619/538 .18 1.275 0.893 … 1.820
  Body Mass Index
   < 18.5 kg/m2 29 .54 0.691 0.210 … 2.279
  18.5 to 24.9 kg/m2 514 .07
  25 to 29.9 kg/m2 391 .92 0.980 0.669 … 1.434
   ≥ 30 kg/m2 223 .01 0.557 0.351 … 0.885
Venous Condition
  Good (0) 787  < .001
  Moderate 295  < .001 2.837 1.945 … 4.138
  Poor 75  < .001 10.780 5.665 … 20.512
Site of Cannulation
  Back of hand (0) 887 .43
  Forearm 184 .37 1.230 0.781 … 1.938
  Antecubital fossa 71 .23 1.494 0.776 … 2.876
  Other 15 .32 1.843 0.552 … 6.152
Operator Variables
  Operator Function
    Nurses (0) 701 .02
    Physicians 397 .005 1.756 1.188 … 2.597
    Others 59 .98 0.989 0.445 … 2.199
  Experience
    > 3 years (0) / < 3 years 886/271 .21 1.311 0.858 … 2.003
  Needle Reinsertion
    Pre-puncture, No (0) / Yes 891/266  < .001 7.663 5.353 … 10.970
    Post-puncture, No (0) / Yes 1106/51  < .001 15.401 6.271 … 37.824
Product Variables
  Catheter type, VS (0) / IS 587/570  < .001 2.996 2.134 … 4.207
  Catheter size
    ≤ 16 G 32 .03 0.208 0.050 … 0.867
    18 G (0) 652 .003
    20 G 468 .03 1.478 1.044 … 2.093
    ≥ 22 G 5 .046 10.895 1.049 … 113.181

The most commonly reported reasons for cannulation failure were poor vein status (48.8% of respondents, n = 313), poor handling (27.8% or respondents, n = 127; 90.8% IS vs. 9.2% VS, P < 0.001), and blunt cannulas (21.4% of respondents, n = 67; 76.1% IS vs. 23.9% VS, P = 0.73). Handling of VS was rated by operators as “good” or “very good” in 86.1% of cases, while only 20.8% rated IS as such (P < 0.001) (see Table 4). Accordingly, VS operators said that the wings and ports influenced catheter placement in 25.7% of cases. The port of the VS was used in 65.0% of cases for flushing, and/or drug administration.

Table 4.

Handling characteristics. Subjective assessment by corresponding operators

Handling Characteristics Total Introcan Vasofix Pchi
N (all) N Perc N (all) N Perc N (all) N Perc
Was backflow clearly/quickly visible? 2293  < .001
  Yes 2150 93.8% 1126 1025 91.0% 1167 1125 96.4%
  No 143 6.2% 101 9.0% 42 3.6%
Was there blood spillage during placement? 2275  < .001
  Yes 285 12.5% 1113 185 16.6% 1162 100 8.6%
  No 1990 87.5% 928 83.4% 1062 91.4%
Dressing status 2216  < .001
  Clean, dry 2131 96.2% 1068 1002 93.8% 1148 1129 98.3%
  Soiled by blood 33 1.5% 26 2.4% 7 0.6%
  Soiled by other fluids 44 2.0% 34 3.2% 10 0.9%
  Soiled by blood and other fluids 8 0.4% 6 0.6% 2 0.2%
Did the port or wings influence catheterization?
  No 1138 846 74.3% -
  Yes, the port 71 6.2%
  Yes, the wings 81 7.1%
  Yes, the port and wings 140 12.3%
Overall Assessment 2302  < .001
  Very good 572 24.8% 1133 41 3.6% 1169 531 45.4%
  Good 671 29.1% 195 17.2% 476 40.7%
  Satisfactory 404 17.5% 303 26.7% 101 8.6%
  Sufficient 312 13.6% 268 23.7% 44 3.8%
  Poor 240 10.4% 228 20.1% 12 1.0%
  Unsatisfactory 103 4.5% 98 8.6% 5 0.4%

Data on misuse parameters was limited with a 58.9% response rate. Partial needle withdrawal and reinsertion into the cannula prior to the catheterization attempt (to ensure mobility of the needle stylet) was reported in 318 of 1355 documented cases (23.5%), while the same maneuver during the catheterization attempt (i.e. partial needle withdrawal to test whether a flashback in the cannula occurs, followed by needle reinsertion and a further catheterization attempt) was reported in 61 of 1348 cases (4.5%). These maneuvers were associated with an 8- and 23-fold risk of cannulation failure, respectively (see Table 5; a detailed analysis by catheter is available in Supplemental Table 4). Of the 461 catheters returned for laboratory evaluation, 32.5% displayed signs of damage, and needle reinsertion prior or during the catheterization attempt were associated with a twofold increase in the risk of catheter damage (see Table 5; a detailed analysis by catheter is available in Supplemental Table 5).

Table 5.

Catheter misuse in relation to risk of cannulation failure and catheter damage. Catheter misuse refers to needle movement, the practice of moving the needle in the cannula prior to puncture or during the catheterization attempt (the manufacturer’s use instructions explicitly warn against such maneuvers). Catheter damage refers to abnormalities in catheter structure, macroscopic or microscopic, such as cuts, tears, compressions or loss of material

Needle Reinsertion (Misuse) N Odds Ratio 95% Confidence Interval P
Risk of Cannulation Failure Failure Rate
  Needle Reinsertion 1358
    None (proper use) 1019 (75.0%) 162 (15.9%)
    Yes (misuse) 339 (25.0%) 219 (64.3%) 9.531 7.214 … 12.592  < .001
  Prior to Catheterization Attempt 1355
    None (proper use) 1037 (76.5%) 181 (13.4%)
    Yes (misuse) 318 (23.5%) 199 (62.6%) 7.909 5.989 … 10.443  < .001
  During Catheterization Attempt 1348
    None (proper use) 1287 (95.5%) 323 (25.1%)
    Yes (misuse) 61 (4.5%) 54 (88.5%) 23.023 10.372 … 51.105  < .001
Risk of Catheter Damage Damage Rate
  Needle Reinsertion 461
    None (proper use) 253 (54.9%) 65 (25.7%)
    Yes (misuse) 208 (45.1%) 85 (40.9%) 1.999 1.347 … 2.967 .001
  Prior to Catheterization Attempt 461
    None (proper use) 270 (58.6%) 73 (27.0%)
    Yes (misuse) 191 (41.4%) 77 (40.3%) 1.823 1.228 … 2.705 .003
  During Catheterization Attempt 459
    None (proper use) 412 (89.8%) 127 (30.8%)
    Yes (misuse) 47 (10.2%) 23 (48.9%) 2.151 1.170 … 3.954 .01

Following an unsuccessful first attempt, two further attempts were usually required to secure venous access (75.9%; median 3, IQR [3, 3]), regardless of the type of cannula used. Overall, adverse events took place in 77 (3.3%) cases, whereas 65 of these events involved the IS cannula (P < 0.001). No serious adverse events were observed and only one needle stick injury took place during the trial (with a VS PVC), which took place prior to patient (or blood) contact.

Discussion

In general, success rates were significantly higher with the ported, winged catheter VS than with the non-ported, non-winged IS. With 83% in the ITT and 84% in the per protocol (PP) populations, the success rates observed were higher than the 73–74% success rates reported in emergency departments [9, 11], but failed to reach the desired 90% threshold [19]. The results are comparable to recent studies including surgical patients, which report a 79–85% first-attempt success rate [12].

Key factors affecting first-attempt success rates of peripheral cannulation were described by Carr et al. as patient, operator, and product variables [20]. Patient factors included age, BMI, and venous status, while operator characteristics included function and experience. Product variables included catheter size, but this study could demonstrate that handling and differences in product design can also affect success rates. For the surgical cohort, the most significant factors in the multifactorial analysis were venous status, catheter type, higher BMI, and operator experience. Additionally, adherence to IFU (i.e. no needle reinsertion) was shown to be a significant factor, although the analysis encompassed only a subpopulation of the cohort. The operator-reported reasons for cannulation failure were most often poor venous status, poor handling of the PVC (primarily IS), and blunt needles.

Overall, the VS design was shown to be more favorable. This PVC design, which has wings and an injection port, appears to facilitate catheter placement and fixation, as indicated by the consistently higher successful cannulation rates, improved backflash visibility, as well as lower rates of spillage and soiled dressings. The majority of operators (86.1%) rated the VS PVC as “good” or “very good”, while IS received such ratings from only 20% of operators. The injection port also appears to be well accepted, being utilized during the surgical procedure for flushing and/or drug administration in approximately 2/3 of cases, despite the availability of 3-way stopcocks. It must be noted, however, that such ports have been associated with a greater risk of infection, likely due to the inability to properly decontaminate the port area [21].

Interestingly, the superiority of the VS PVC was diminished with younger age and lower BMI. The slimmer and smaller IS design seems to be more suitable for children and patients with BMI < 18.5 kg/m2, although these differences remained statistically insignificant, likely due to the smaller size of these subgroups. Nevertheless, IS PVCs were associated with 65 of the 77 (84.4%) catheter-related adverse events observed in this investigation. These included primarily extravasate infusions, dislocations, and occlusions, factors that are rather related to care and maintenance of the PVCs, and not the placement procedure itself.

The most commonly chosen site for cannulation in this surgical collective was the backside of the hand, which usually allows ease of access for anesthesiologists throughout the surgical procedure. Furthermore, should a cannulation failure occur, proximal alternative sites remain available. This remains the primary cannulation site despite evidence that placement in the forearm reduces catheter-associated complication rates [22].

Physicians and nurses achieved similar success rates. Nurse involvement increased when moderate or poor vein conditions were reported, whereas nurses were also significantly more experienced and placed a higher number of catheters per week in comparison to physicians, suggesting an advantage in terms of routine and experience. More experienced physicians are underrepresented, likely because they frequently supervise longer and more complex operations (often one per day), whereas less experienced colleagues usually accompany several short and simple procedures during a workday. When taking experience into consideration, there were no differences in success (or misuse) between physicians and nurses. Interestingly, the best success rates were achieved by students and interns, despite no differences in the subjective assessment of the patients’ venous status. This is possibly due to less time pressure and a selected patient group, but it must be noted that in the larger, more representative logistic regression (Supplemental Table 3), no significant differences were observed.

The PVCs were not always used as intended. The pervasive custom of moving the needle inside the plastic cannula prior to puncture took place in nearly a quarter of cases, being associated with an increase in the risk of failure from 16 to 64%. Likewise, partially withdrawing the needle through the cannula after a puncture, then reinserting it in a rescue attempt to redirect the cannula was seen in nearly 5% of cases, with an 88.5% likelihood of failure. The IFU specifically warn users not to reinsert needles through the cannula, and there are reports of intravenous catheter fractures and embolisms [23]. In this investigation, such maneuvers were related to a twofold increase in the rate of catheter damage in the laboratory analysis. While VS showed higher rates of damage, with or without needle reinsertion, IS appears to be more susceptible to damage by misuse, possibly due to the polyurethane material (see Supplemental Table 5). This information must be emphasized in training sessions in an effort to reduce such practices. It is also worthy of note that 25.7% of PVCs that were used properly still showed signs of damage.

It is important to note that an increasing number of tools have been developed for predicting difficult venous access, both for pediatric [2426] and adult patients [19, 2729]. Additionally, there has been a marked increase in the availability and acceptance of ultrasound-guided assessment and cannulation [30, 31], sparking the formation of vascular access teams [32]. The European Society of Anesthesiologist published guidelines recommending the early identification of difficult venous access through validated tools, as well as the use of ultrasound [33]. The implementation of such tools has the potential to improve patient care by adapting clinical processes and allowing an appropriate allocation of personnel and materials, as well as provide an objective basis for the comparison in clinical studies.

Limitations

Several limitations must be taken into account. Although an open-label investigation was unavoidable, there was an attempt to minimize bias through implementation of a randomization process and the use of CRFs. Nevertheless, an observation bias cannot be ruled out, and nor can a performance bias (involvement of more experienced colleagues by particularly difficult venous status). Although patients were accompanied by study personnel whenever possible, self-reported data from unsupervised operators may be an additional source of bias. The venous status of patients was subjective and may vary considerably, depending on the operator’s experience. Although the adult version of the Difficult Intravenous Access (DIVA) scale [27] was not available at the beginning of this trial, the lack of an objective scoring system is a clear limitation of this study. No identifiable information was collected from the operators, and no data is available regarding the number of operators involved nor the frequency of their participation. No follow-up attempt was made after discharge from the recovery room, and the short duration of observation may have affected the reported rate of complications. There might be additional factors affecting cannulation success that were not considered in this study, particularly in the multivariate analyses. Although both products were newly introduced, it is important to note that winged and ported catheters are far more common in Germany, so that operators were likely to be more familiar with this design type. Additionally, due to the high number of missing values, data on needle reinsertion and damage must be regarded with caution. Data generalizability is limited, as the study was conducted exclusively in a university setting.

Conclusions

First-attempt cannulation success was more likely with ported, winged VS catheters compared to non-ported, non-winged IS devices. The larger VS design allowed for improved handling and fixation, although this superiority could not be observed in younger patients or those with low BMI. The overall rate of successful puncture at first attempt in the operating room was 83%. Among clinicians, nurses were more experienced and placed the majority of catheters. The backside of the hand was the preferred site, and the most commonly reported reasons for failure were poor venous status and poor handling of the catheter. This trial identified the reinsertion of the needle stylet into the cannula, prior or during the cannulation attempt, as novel clinician-related markers that were associated with higher incidences of catheter damage and cannulation failure. This information must be highlighted in training sessions to limit these dangerous practices and improve patient safety. Standardized tools for the identification of difficult venous access are increasingly common, and the implementation of ultrasound techniques and dedicated vascular access teams have the potential to improve care and reduce catheter-related complications.

Supplementary Information

12871_2022_1631_MOESM1_ESM.docx (2.6MB, docx)

Additional file 1: Supplemental Figure 1. Consort diagram, flowchart; *For one study site (Universitätsklinik Tübingen), only information about completed case report forms are available; # Other reasons include screening failure, urgent operations, and missing documentation; ## Other reasons include missing/incomplete data and drop-out participants.

12871_2022_1631_MOESM2_ESM.docx (698.9KB, docx)

Additional file 2: Supplemental Table 1. Patient characteristics by center. All data shown as frequencies and percentages. ASA  American Society of Anesthesiologists, CBF Campus Benjamin Franklin, CVK Campus Virchow Klinikum, CCM Campus Charité Mitte; UCT Universitätsklinikum Tübingen; P-values represent Chi-Square Test, or Fisher’s Exact Test when small cell exceptions were present; missing data is treated as such.

12871_2022_1631_MOESM3_ESM.docx (808.3KB, docx)

Additional file 3: Supplemental Table 2. Patient characteristics by center (continued). All data shown as frequencies and percentages. CBF Campus Benjamin Franklin, CVK Campus Virchow Klinikum, CCM Campus Charité Mitte, UCT Universitätsklinikum Tübingen. Missing data is treated as such. Vein status is a subjective variable assigned by the corresponding operator; P-values represent Chi-Square Test, or Fisher’s Exact Test when small cell exceptions were present; missing data is treated as such.

12871_2022_1631_MOESM4_ESM.docx (334.6KB, docx)

Additional file 4: Supplemental Table 3. Multivariate analysis without misuse markers; VS: Vasofix® Safety, IS: Introcan® Safety; adult participants only; All data analyzed as categorical variables.

12871_2022_1631_MOESM5_ESM.docx (252.8KB, docx)

Additional file 5: Supplemental Table 4. Catheter misuse in relation to risk of cannulation failure, detailed by type of catheter. Catheter misuse refers to needle movement, the practice of moving the needle in the cannula prior to puncture or during the catheterization attempt (the manufacturer’s use instructions explicitly warn against such maneuvers).

12871_2022_1631_MOESM6_ESM.docx (247.8KB, docx)

Additional file 6: Supplemental Table 5. Catheter misuse in relation to risk of catheter damage, detailed by type of catheter. Catheter misuse refers to needle movement, the practice of moving the needle in the cannula prior to puncture or during the catheterization attempt (the manufacturer’s use instructions explicitly warn against such maneuvers). Catheter damage refers to abnormalities in catheter structure, macroscopic or microscopic, such as cuts, tears, compressions or loss of material.

Acknowledgements

The authors thank Dr. Kathrin Scholz for the clinical monitoring and Prof. Dr.- Ing. Andreas Wittmann and his team from the Bergische Universität Wuppertal, Fakultät für Maschinenbau und Sicherheitstechnik, Fachbereich Technischer Infektionsschutz for the catheter damage assessment. The authors also thank the study nurses Beatrix Laetsch, Susanne Sonnemann and Tuba Gülmez, as well as Dr. Jörn Kiselev for the critical review of the manuscript.

Abbreviations

PVC

Peripheral venous catheter

BMI

Body-mass index

VS

Vasofix® Safety

IS

Introcan® Safety

CBF

Charité Campus Benjamin Franklin

CVK

Charité Campus Virchow Klinikum

CCM

Charité Campus Charité Mitte

UCT

Universitätsklinikum Tübingen

CRF

Case report form

ITT

Intention-to-treat

PP

Per protocol

ASA

American Society of Anesthesiologists

OR

Odds Ratio

CI

Confidence Interval

Authors’ contributions

RM: writing-original draft, writing-review & editing, validation, formal analysis; KS: conceptualization, investigation, validation, supervision, writing-review & editing, project administration; TN: investigation, validation, supervision, formal analysis, writing-review & editing; JK: validation, formal analysis; UF: resources, supervision, project administration; PH: investigation, data curation; PR: resources, supervision, project administration; EF: investigation, data curation; WB: conceptualization, methodology, supervision; CS: conceptualization, methodology, resources, project administration, funding acquisition. All authors have read and approved the final version of the manuscript.

Funding

Open Access funding enabled and organized by Projekt DEAL. This study was sponsored by B.Braun Melsungen AG under the direction of Dr. Heike Fries. The sponsorship did not influence the analysis or reporting of study outcomes.

Availability of data and materials

The datasets analyzed during the current study are not publicly available due to constrains imposed in the consent forms, but an anonymized version is available from the corresponding author on reasonable request.

Declarations

Ethics approval and consent to participate

Written informed consent was obtained from all participants or their proxies. The study was approved by the ethic committees of the Charité – Universitätsmedizin Berlin (EA2/104/14) and the Universitätsklinikum Tübingen (494/2014BO2), as well as by their respective data protection officers. The study was conducted in accordance with the Declaration of Helsinki and its amendments and was registered at ClinicalTrials.gov (NCT 02213965).

Consent for publication

B. Braun has granted permission to publish Fig. 1.

Competing interests

CS, UF, TN and PR have received material institutional support from B.Braun Melsungen AG for the conduction of this trial. The sponsorship did not influence the analysis or reporting of study outcomes. All other authors have no competing interests to report.

Footnotes

Publisher’s Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Rudolf Mörgeli and Katrin Schmidt shared first authorship.

References

  • 1.Waitt C, Waitt P, Pirmohamed M. Intravenous therapy. Postgrad Med J. 2004;80(939):1–6. doi: 10.1136/pmj.2003.010421. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 2.Malach T, Jerassy Z, Rudensky B, Schlesinger Y, Broide E, Olsha O, et al. Prospective surveillance of phlebitis associated with peripheral intravenous catheters. Am J Infect Control. 2006;34(5):308–312. doi: 10.1016/j.ajic.2005.10.002. [DOI] [PubMed] [Google Scholar]
  • 3.Pujol M, Hornero A, Saballs M, Argerich MJ, Verdaguer R, Cisnal M, et al. Clinical epidemiology and outcomes of peripheral venous catheter-related bloodstream infections at a university-affiliated hospital. J Hosp Infect. 2007;67(1):22–29. doi: 10.1016/j.jhin.2007.06.017. [DOI] [PubMed] [Google Scholar]
  • 4.Alexandrou E, Ray-Barruel G, Carr PJ, Frost S, Inwood S, Higgins N, et al. International prevalence of the use of peripheral intravenous catheters: Prevalence of the Use of PIVCs. J Hosp Med. 2015;10(8):530–533. doi: 10.1002/jhm.2389. [DOI] [PubMed] [Google Scholar]
  • 5.Rickard CM, Marsh N, Webster J, Playford EG, McGrail MR, Larsen E, et al. Securing All intraVenous devices Effectively in hospitalised patients–the SAVE trial: study protocol for a multicentre randomised controlled trial. BMJ Open. 2015;5(9):e008689. doi: 10.1136/bmjopen-2015-008689. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 6.Moayedi S, Witting M, Hirshon JM, George N, Burke A, Schenkel S. Prospective, randomized controlled comparison of a flash-tip catheter and a traditional intravenous catheter in an urban emergency department. J Vasc Access. 2018;19(4):387–391. doi: 10.1177/1129729817747530. [DOI] [PubMed] [Google Scholar]
  • 7.Castillo MI, Larsen E, Cooke M, Marsh NM, Wallis MC, Finucane J, et al. Integrated versus nOn-integrated Peripheral inTravenous catheter. Which is the most effective systeM for peripheral intravenoUs catheter Management? (The OPTIMUM study): a randomised controlled trial protocol. BMJ Open. 2018;8(5):e019916. doi: 10.1136/bmjopen-2017-019916. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 8.Sabri A, Szalas J, Holmes KS, Labib L, Mussivand T. Failed attempts and improvement strategies in peripheral intravenous catheterization. Biomed Mater Eng. 2013;23(1–2):93–108. doi: 10.3233/BME-120735. [DOI] [PubMed] [Google Scholar]
  • 9.Lapostolle F, Catineau J, Garrigue B, Monmarteau V, Houssaye T, Vecci I, et al. Prospective evaluation of peripheral venous access difficulty in emergency care. Intensive Care Med. 2007;33(8):1452–1457. doi: 10.1007/s00134-007-0634-y. [DOI] [PubMed] [Google Scholar]
  • 10.Hess HA. A biomedical device to improve pediatric vascular access success. Pediatr Nurs. 2010;36(5):259–263. [PubMed] [Google Scholar]
  • 11.Carr PJ, Rippey JCR, Cooke ML, Trevenen ML, Higgins NS, Foale AS, et al. Factors associated with peripheral intravenous cannulation first-time insertion success in the emergency department. A multicentre prospective cohort analysis of patient, clinician and product characteristics. BMJ Open [Internet]. 2019 Apr 2 [cited 2020 Mar 22];9(4). https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6500093/Available from [DOI] [PMC free article] [PubMed]
  • 12.van Loon FH, Timmerman R, den Brok GP, Korsten EH, Dierick-van Daele AT, Bouwman AR. The impact of a notched peripheral intravenous catheter on the first attempt success rate in hospitalized adults: Block-randomized trial. J Vasc Access. 2021;2:1129729821990217. doi: 10.1177/1129729821990217. [DOI] [PubMed] [Google Scholar]
  • 13.Introcan Safety® IV Catheter Straight [Internet]. [cited 2021 Sep 14]. https://www.bbraunusa.com/en/products/b/introcan-safety-ivcatheterstraight.htmlAvailable from
  • 14.Vasofix® Safety [Internet]. [cited 2021 Sep 14]. https://www.bbraun.de/de/products/b0/vasofix-safety.htmlAvailable from
  • 15.McCoy CE. Understanding the Intention-to-treat Principle in Randomized Controlled Trials. West J Emerg Med. 2017;18(6):1075–1078. doi: 10.5811/westjem.2017.8.35985. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 16.Sebbane M, Claret P-G, Lefebvre S, Mercier G, Rubenovitch J, Jreige R, et al. Predicting peripheral venous access difficulty in the emergency department using body mass index and a clinical evaluation of venous accessibility. J Emerg Med. 2013;44(2):299–305. doi: 10.1016/j.jemermed.2012.07.051. [DOI] [PubMed] [Google Scholar]
  • 17.Jacobson AF, Winslow EH. Variables influencing intravenous catheter insertion difficulty and failure: an analysis of 339 intravenous catheter insertions. Heart Lung. 2005;34(5):345–359. doi: 10.1016/j.hrtlng.2005.04.002. [DOI] [PubMed] [Google Scholar]
  • 18.Fields JM, Piela NE, Au AK, Ku BS. Risk factors associated with difficult venous access in adult ED patients. Am J Emerg Med. 2014;32(10):1179–82. doi: 10.1016/j.ajem.2014.07.008. [DOI] [PubMed] [Google Scholar]
  • 19.van Loon FHJ, van Hooff LWE, de Boer HD, Koopman SSHA, Buise MP, Korsten HHM, et al. The Modified A-DIVA Scale as a Predictive Tool for Prospective Identification of Adult Patients at Risk of a Difficult Intravenous Access: A Multicenter Validation Study. J Clin Med. 2019;8(2):E144. doi: 10.3390/jcm8020144. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 20.Carr PJ, Higgins NS, Cooke ML, Rippey J, Rickard CM. Tools, Clinical Prediction Rules, and Algorithms for the Insertion of Peripheral Intravenous Catheters in Adult Hospitalized Patients: A Systematic Scoping Review of Literature. J Hosp Med. 2017;12(10):851–858. doi: 10.12788/jhm.2836. [DOI] [PubMed] [Google Scholar]
  • 21.Hadaway L. Short peripheral intravenous catheters and infections. J Infus Nurs. 2012;35(4):230–240. doi: 10.1097/NAN.0b013e31825af099. [DOI] [PubMed] [Google Scholar]
  • 22.Wallis MC, McGrail M, Webster J, Marsh N, Gowardman J, Playford EG, et al. Risk factors for peripheral intravenous catheter failure: a multivariate analysis of data from a randomized controlled trial. Infect Control Hosp Epidemiol. 2014;35(1):63–68. doi: 10.1086/674398. [DOI] [PubMed] [Google Scholar]
  • 23.Glassberg E, Lending G, Abbou B, Lipsky AM. Something’s Missing: Peripheral Intravenous Catheter Fracture. The Journal of the American Board of Family Medicine. 2013;26(6):805–806. doi: 10.3122/jabfm.2013.06.130097. [DOI] [PubMed] [Google Scholar]
  • 24.Yen K, Riegert A, Gorelick MH. Derivation of the DIVA score: a clinical prediction rule for the identification of children with difficult intravenous access. Pediatr Emerg Care. 2008;24(3):143–147. doi: 10.1097/PEC.0b013e3181666f32. [DOI] [PubMed] [Google Scholar]
  • 25.Riker MW, Kennedy C, Winfrey BS, Yen K, Dowd MD. Validation and refinement of the difficult intravenous access score: a clinical prediction rule for identifying children with difficult intravenous access. Acad Emerg Med. 2011;18(11):1129–1134. doi: 10.1111/j.1553-2712.2011.01205.x. [DOI] [PubMed] [Google Scholar]
  • 26.Girotto C, Arpone M, Frigo AC, Micheletto M, Mazza A, Da Dalt L, et al. External validation of the DIVA and DIVA3 clinical predictive rules to identify difficult intravenous access in paediatric patients. Emerg Med J. 2020;37(12):762–767. doi: 10.1136/emermed-2020-209658. [DOI] [PubMed] [Google Scholar]
  • 27.van Loon FHJ, Puijn LAPM, Houterman S, Bouwman ARA. Development of the A-DIVA Scale: A Clinical Predictive Scale to Identify Difficult Intravenous Access in Adult Patients Based on Clinical Observations. Medicine (Baltimore) 2016;95(16):e3428. doi: 10.1097/MD.0000000000003428. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 28.Angles E, Robin F, Moal B, Roy M, Sesay M, Ouattara A, et al. Pre-operative peripheral intravenous cannula insertion failure at the first attempt in adults: Development of the VENSCORE predictive scale and identification of risk factors. J Clin Anesth. 2021;75:110435. doi: 10.1016/j.jclinane.2021.110435. [DOI] [PubMed] [Google Scholar]
  • 29.Civetta G, Cortesi S, Mancardi M, De Pirro A, Vischio M, Mazzocchi M, et al. EA-DIVA score (Enhanced Adult DIVA score): A new scale to predict difficult preoperative venous cannulation in adult surgical patients. J Vasc Access. 2019;20(3):281–289. doi: 10.1177/1129729818804994. [DOI] [PubMed] [Google Scholar]
  • 30.O’Reilly-Shah VN, Franz A, Groenewald CB, Collins M, Patak LS. Training the Trainers in Ultrasound-guided Access to Improve Peripheral Intravenous Catheter Placement among Children Presenting for Anesthesia. Pediatr Qual Saf. 2021;6(3):e406. doi: 10.1097/pq9.0000000000000406. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 31.Gras S, Roy-Gash F, Bruneau B, Salvi N, Colas A-E, Skhiri A, et al. Reducing the time to successful intravenous cannulation in anaesthetised children with poor vein visibility using a near-infrared device: A randomised multicentre trial. Eur J Anaesthesiol. 2021;38(8):888–894. doi: 10.1097/EJA.0000000000001467. [DOI] [PubMed] [Google Scholar]
  • 32.Bell JA, Spencer TR. Implementing an emergency department vascular access team: A quality review of training, competency, and outcomes. J Vasc Access. 2021;22(1):81–89. doi: 10.1177/1129729820924554. [DOI] [PubMed] [Google Scholar]
  • 33.Lamperti M, Biasucci DG, Disma N, Pittiruti M, Breschan C, Vailati D, et al. European Society of Anaesthesiology guidelines on peri-operative use of ultrasound-guided for vascular access (PERSEUS vascular access) Eur J Anaesthesiol. 2020;37(5):344–376. doi: 10.1097/EJA.0000000000001180. [DOI] [PubMed] [Google Scholar]

Associated Data

This section collects any data citations, data availability statements, or supplementary materials included in this article.

Supplementary Materials

12871_2022_1631_MOESM1_ESM.docx (2.6MB, docx)

Additional file 1: Supplemental Figure 1. Consort diagram, flowchart; *For one study site (Universitätsklinik Tübingen), only information about completed case report forms are available; # Other reasons include screening failure, urgent operations, and missing documentation; ## Other reasons include missing/incomplete data and drop-out participants.

12871_2022_1631_MOESM2_ESM.docx (698.9KB, docx)

Additional file 2: Supplemental Table 1. Patient characteristics by center. All data shown as frequencies and percentages. ASA  American Society of Anesthesiologists, CBF Campus Benjamin Franklin, CVK Campus Virchow Klinikum, CCM Campus Charité Mitte; UCT Universitätsklinikum Tübingen; P-values represent Chi-Square Test, or Fisher’s Exact Test when small cell exceptions were present; missing data is treated as such.

12871_2022_1631_MOESM3_ESM.docx (808.3KB, docx)

Additional file 3: Supplemental Table 2. Patient characteristics by center (continued). All data shown as frequencies and percentages. CBF Campus Benjamin Franklin, CVK Campus Virchow Klinikum, CCM Campus Charité Mitte, UCT Universitätsklinikum Tübingen. Missing data is treated as such. Vein status is a subjective variable assigned by the corresponding operator; P-values represent Chi-Square Test, or Fisher’s Exact Test when small cell exceptions were present; missing data is treated as such.

12871_2022_1631_MOESM4_ESM.docx (334.6KB, docx)

Additional file 4: Supplemental Table 3. Multivariate analysis without misuse markers; VS: Vasofix® Safety, IS: Introcan® Safety; adult participants only; All data analyzed as categorical variables.

12871_2022_1631_MOESM5_ESM.docx (252.8KB, docx)

Additional file 5: Supplemental Table 4. Catheter misuse in relation to risk of cannulation failure, detailed by type of catheter. Catheter misuse refers to needle movement, the practice of moving the needle in the cannula prior to puncture or during the catheterization attempt (the manufacturer’s use instructions explicitly warn against such maneuvers).

12871_2022_1631_MOESM6_ESM.docx (247.8KB, docx)

Additional file 6: Supplemental Table 5. Catheter misuse in relation to risk of catheter damage, detailed by type of catheter. Catheter misuse refers to needle movement, the practice of moving the needle in the cannula prior to puncture or during the catheterization attempt (the manufacturer’s use instructions explicitly warn against such maneuvers). Catheter damage refers to abnormalities in catheter structure, macroscopic or microscopic, such as cuts, tears, compressions or loss of material.

Data Availability Statement

The datasets analyzed during the current study are not publicly available due to constrains imposed in the consent forms, but an anonymized version is available from the corresponding author on reasonable request.