Skip to main content
NIHPA Author Manuscripts logoLink to NIHPA Author Manuscripts
. Author manuscript; available in PMC: 2023 Jul 1.
Published in final edited form as: J Cardiothorac Vasc Anesth. 2021 Oct 1;36(7):2012–2021. doi: 10.1053/j.jvca.2021.09.042

Continuous Noninvasive Blood Pressure Monitoring of Beat-by-Beat Blood Pressure and Heart Rate Using Caretaker® Compared to Invasive Arterial Catheter In the Intensive Care Unit

Younghoon Kwon 1, Patrick L Stafford 2, Kyle Enfield 2, Sula Mazimba 2, Martin C Baruch 3
PMCID: PMC8971137  NIHMSID: NIHMS1749013  PMID: 34666928

Abstract

Objective

We aimed to examine the accuracy of noninvasively-derived peripheral arterial BP by Caretaker device (CT) against invasively measured arterial BP. We also examined the fidelity of heart rate variability by CT compared to ECG derived data.

Design

Prospective cohort study.

Participants

Adult surgical and trauma patients admitted to the ICU.

Setting

Academic tertiary care medical center.

Interventions

In a prospective manner, beat-by-beat BP by CT was recorded simultaneously with invasive arterial BP measured in patients in the intensive care unit. Invasive arterial BPs were compared with those obtained by the CT system. All comparisons between the CT data, arterial catheter data, and ECG data were post-processed.

Measurements and Main Results

From 37 enrolled patients, 34 were included with satisfactory data that overlapped between arterial catheter and CT. A total of 87,757 comparative data points were obtained for the 40 minute time window comparisons of the 34 patients, spanning approximately 22.5 hours in total. Systolic BP and diastolic BP correlations (Pearson’s coefficient), as well as the mean difference (standard deviation) were, 0.92 and −0.36 (7.57) mmHg and 0.83 and −2.11 (6.00) mmHg, respectively. The overall inter-beat correlation was 0.99 with the mean difference between inter-beats obtained with the arterial BP and the CT was −0.056ms (6.0).

Conclusions

This study validates the non-invasive tracking of BP using the CT device and the pulse decomposition analysis approach is possible within the guidelines of the standard.

Keywords: Blood pressure, Heart rate variability, Continuous monitoring

Introduction

Continuous blood pressure (BP) monitoring is critical in the acute and intensive care setting, and the accurate assessment and tracking of BP are crucial in medical decision making. However, continuous BP measurement in a beat-by-beat fashion, while desirable, requires invasive arterial catheterization. Existing continuous noninvasive BP (cNIBP) technologies are nearly exclusively based on volume-clamp (vascular unloading) technique, or the “Penaz” method. This technique is based on the principle that BP can be estimated by measuring the finger cuff pressure required to maintain constant volume of blood in the finger1. Despite the significant efforts to develop more convenient and ubiquitous cNIBP options for cNIBP monitoring in clinical practice remain limited2,3.

The Caretaker® (CT) continuous noninvasive physiological monitor (Caretaker Medical LLC, Charlottesville, VA, USA), which has been described in detail elsewhere4,5, is a recently developed technology that provides beat-by-beat cNIBP, as well as high-resolution inter-beat interval information. This device uses a low pressure [~35–45 mmHg], pump-inflated finger cuff that pneumatically couples arterial pulsations via a pressure line to a custom-designed piezo-electric pressure sensor for detection and analysis. The CT device is depicted in Figure 1.

Figure 1.

Figure 1.

Depiction of the Caretaker BP device.

The aim of the present study was to assess the accuracy of the CT for two critical physiological variables, BP and heart rate (HR), against their respective gold standards, arterial catheter-derived BP and electrocardiography (ECG)-based HR, in intensive care unit (ICU) patients.

Patients and Methods

This study was approved by the University of Virginia Medical Center Review Board. We recruited patients hospitalized in the University of Virginia surgical/trauma ICU who were monitored using radial intra-arterial catheters. Invasive arterial BPs were compared with those obtained by the CT system, which collected pulse line shapes at the lower phalanx of the thumb of the ipsilateral hand.

BedMaster (Excel Medical, Jupiter, FL, USA) hardware and software were used to digitize and record intra-arterial waveform data and electrocardiography (ECG), with simultaneous time base at a sample rate of 240Hz (Unity Network, GE Healthcare, Chicago, IL, USA). The catheter/transducer system used consisted of Judkins-type catheters (6 French) and Meritrans pressure transducers (Merit Medical Systems, South Jordan, UT, USA). The frequency response of the system ranges from 0–500Hz with an accuracy of better than ±1mmHg. Representative data epochs of 20s were examined using Fourier analysis to determine the relative harmonic amplitude distribution. To ensure the absence of under-damping, which can introduce errors in systolic BP (SBP) exceeding 10mmHg, the amplitude ratio of the fundamental and 6th harmonic was verified to exceed 1.5 orders of magnitude6.

Caretaker Device and PDA model

The CT is FDA-cleared and CE-cleared for the measurement of cNIBP (FDA K151499), measurement of HR and respiratory rate, and self-calibration (FDA K163255). BP monitoring is accomplished via a pulse contour analysis algorithm called Pulse Decomposition Analysis (PDA), which analyzes the component pulses, specifically the left ventricular ejection pulse (P1) and its reflections, the renal reflection pulse (P2) and the iliac reflection pulse (P3), that constitute the arterial pressure pulse7. The model’s core BP parameter is the ratio of the amplitude of the renal reflection pulse (P2) to that of the primary systolic pulse (P1), referred to as P2P1. A linear model is used to convert from the P2P1 factor to the SBP and diastolic BP (DBP) components, as below:

Psystolic=αs(AS)(p2p1)+βs (1)
Pdiastolic=αd(AS)(p2p1)+βd (2)

The gain factors of the linear conversion, αs and αd, are functions of another PDA parameter, termed AS, that relates to arterial stiffness. The functional form is proprietary. The βs and βd parameters are individual patient-specific offsets that are determined during the calibration phase of the CT device. The AS parameter quantifies the spectral content of the arterial pressure pulse. The AS parameter is, in turn, related to arterial stiffness since the mechanical filtering of the arterial wall determines to what extent the structure of the component pulses, P1, P2, and P3, is resolved. Prior work by Callaghan et al. demonstrated that this filtering limits the upper observable frequency components in the peripheral arterial pressure pulse to approximately 20 Hz8. Preliminary validation tests indicate that the AS parameter tracks expected trends after the introduction of vasoactive agents, as well as age-related population trends. A detailed description of the calculation of the AS factor, as well the motivation therein, is provided elsewhere9.

By tracking the pulse wave forms, CT also allows HR measurement via inter-beat interval. HR measurement is another key component of vital sign monitoring in acute care setting. Moreover, as CT provides inter-beat interval information, HR variability (HRV) can be also derived. HRV is useful for assessment of autonomic system and prediction of cardiovascular risks10. While CT is FDA-cleared for a self-calibration procedure that involves an oscillometric sweep of the finger cuff pressure to obtain systolic and diastolic starting pressures for the PDA tracking algorithm, the goal of this study was to isolate and analyze the performance of the tracking accuracy in a post-process analysis.

Data Inclusion and Exclusion

Radial arterial catheter data were visually inspected and sections of obvious catheter failure, characterized by either continuous or spurious nonsensical readings, were excluded. Sections contaminated by excessive motion artifact such that the peak detection algorithm was no longer able to identify heart beats were also excluded. No separate inspection for the ECG data was performed, i.e. the same sections of arterial catheter and ECG data were discarded.

In the case of the CT data, a custom signal/noise factor (SNF) was used to identify poor quality data sections to be excluded. The factor is based on the ratio of the variances of the physiological signal band to the noise band and obtained using Fourier spectral analysis over an 8s window with 1s overlap. The frequency range of the band associated with the physiological signal was set to 1– 10Hz, based on data by the authors and results by others8, while the noise band was set to the 100–250Hz frequency range, which is subject to ambient noise but contains no signal relevant to the base band phenomena of the arterial pressure pulse or its propagation characteristics. Data sections with an SNF below 140 were excluded from the analysis. Across the 34 subjects, the percentage mean of excluded data was 3.4% with a median of excluded data of 1.5%.

Data Alignment & Calibration

All comparisons between the CT data, arterial catheter data, and ECG data were post-processed. The overlap of the CT data streams and the radial arterial catheter data streams was established after an initial alignment based on data collection system clocks. This was performed by matching time-based inter-beat interval series from the CT data and from the ECG gold standard data, thereby also time-aligning the arterial catheter data that were collected in parallel with the ECG data by the BedMaster system.

For each patient, the first 40 minute overlap section was used for the comparison. Stable overlap sections were defined as having an SNF of at least 140 for the CT data and having stable arterial catheter data, as described above. In a onetime procedure, a 15s window at the start of the 40 minute overlap section was used to calibrate the PDA pulse parameters. Patient-specific PDA constants, once established, were not changed, irrespective of subsequent hemodynamic changes. A 40 minute overlap window was chosen based on BP comparison studies performed by others11,12, the motivation being that tracking windows on that time order are reasonable for recalibration, whether internally or externally, in the clinical workflow.

Statistical Analysis

All continuous variables were expressed as mean (standard deviation [SD]). The normality of data distribution was tested with the Kolmogorov-Smirnov test. For both BP and inter-beat data series, we present graphical correlations and Bland-Altman comparisons including all patients13. Mean difference (bias) and 95% limits of agreement representing 2SD of the mean difference (precision) were shown in the Bland-Altman plot. Since the large number of datapoints made it difficult to appreciate the portion of the datapoint within and beyond the 95% limits of agreement line of the Bland-Altman plot, a histogram of the distribution of the counts is incorporated with each Bland-Altman plot. Representative correlations and Bland-Altman comparisons are presented for two individual patients. Consistency of the overall BP data was assessed by calculating Cronbach’s alpha and the corresponding correlation coefficients for different data ranges. Because the estimation of the difference between the methods was the outcome of interest, no power analyses for sample size estimates were calculated prior to the study. The cohort size was therefore driven by patient availability and the ANSI/AAMI/ISO 81060–2:2013 standard’s required lower limit of 15 patients when an arterial catheter is used for comparison14.

Results

A total of 37 patients who were approached were enrolled. Two patients with significant motion artifact resulting in invalid recordings were excluded. One patient, whose CT device became accidently disconnected early in the session, was also excluded. This resulted in 34 patients with sufficient data that overlapped between arterial catheter and CT. Patient characteristics and admission indications are presented in Table 1. A total of 87,757 comparative data points were obtained for the 40 minute time window comparisons of the 34 patients, spanning approximately 22.5 hours in total. Figures 2 and 3 present overlap examples in the case of a patient with dynamic BP for their inter-beat intervals and BP, respectively. Figures 4 and 5 demonstrate the agreement between SBP and DBP, the distribution of counts, and variation in SBP and DBP for example patients #17 and #21, respectively. The overall results, for CT measurements of SBP and DBP, respectively, in comparison with invasive BPs for the entire population are shown in Figure 6. Mean differences (SD) of the two BP measurements for SBP and DBP were −0.36 (7.57) and −2.11 (6.00) mmHg, respectively. The mean difference BP data were normally distributed (Figures 6B- top and bottom). As indicated by Figure 6, there was a wide range of BP manifested from the included patients for SBP and DBP, respectively. Correlations for both SBP and DBP were strong with an R=0.92 (p<0.001; adjusted R2 0.84) and R=0.83 (p<0.001; adjusted R2 0.69), respectively. Similarly, correlation and agreement were strong for the inter-beat interval. The correlation was 0.99 (p<0.001) (Figure 7A). The inter-beat interval data was normally distributed (Figure 7B). The mean difference between inter-beats obtained with the catheter system and the CT was −0.056 (6.0) ms (Figure 7C).

Table 1.

Baseline Patient Characteristics.

Characteristic Mean (SD) or N (%)
N = 34
Range
Age (years) 44.1 (13.9) 18 – 64
Male Sex 23 (67.6) --
Height (cm) 173.3 (9.4) 154.9 – 189
Weight (kg) 95.3 (27.4) 59.3 – 169.5
Body mass index (kg/m2) 32.0 (8.8) 20.5 – 55.2
Main Indications for Admission
Surgery Procedure Type N
Liver/Kidney Transplant 7/1
Gastric Surgery 5
Thoracic/Abdominal/Cardiac Surgery 5
Oncology/Osteosarcoma 1/1
Melanoma/Facial Reconstruction 1
Trauma Trauma Type N
Motor Vehicle Accident 10
Burn/Head Injury 1/1
Necrotizing Fasciitis 1
Patient Conditions
Aortic Regurgitation 0 --
Nephrectomy 0 --
Thoracoabdominal Aortic Stent or Graft 0 --

Figure 2.

Figure 2.

An example of overlap of ECG-based inter-beat intervals obtained from the Caretaker data stream. This presents both an overall overlay of the inter-beat intervals, as well as a 70s expanded section. The correlation (Pearson’s coefficient) is 0.99 (p<0.001), while the mean difference (SD) is 0.17 (2.75) ms.

Figure 3.

Figure 3.

An example of overlap of beat-by-beat systolic and diastolic BPs obtained from arterial catheter (A-line) and Caretaker (PDA) system. Correlations (Pearson’s coefficient) and mean difference SD for systole and diastole were, respectively, 0.93 (p<0.001) (4.41) mmHg and 0.91 (3.82) mmHg.

Figure 4.

Figure 4.

Correlations (panels A), count distributions (panels B), and Bland-Altman results (panels C) for systole (top) and diastole (bottom) for example patient #17. Results for systole – correlation: 0.83, Bland-Altman results: mean difference 0.89 mmHg (6.48). Results for diastole – correlation: 0.75, mean difference −1.04 mmHg (5.95).

Figure 5.

Figure 5.

Correlations (panels A), count distributions (panels B), and Bland-Altman results (panels C) for systole (top) and diastole (bottom) for example patient #21. Results for systole – correlation: 0.89, Bland-Altman results: mean difference 0.69 mmHg (4.3). Results for diastole – correlation: 0.86, mean difference 0.54 mmHg (3.1).

Figure 6.

Figure 6.

Correlations (panels A), count distributions (panels B), and Bland-Altman results (panels C) for systole (top) and diastole (bottom) for all patients. Results for systole – correlation: 0.92, Bland-Altman results: mean difference 0.36 mmHg (7.57). Results for diastole – correlation: 0.83, mean difference 2.11 mmHg (6.00).

Figure 7.

Figure 7.

Caretaker overall inter-beat interval data Correlations (panels A), count distributions (panels B), and Bland-Altman results (panels C). Correlation was 0.999 and adjusted R2 was 0.999. Bland-Altman results: mean difference −0.056 ms (6.00).

In Table 2, the internal consistency results are presented, specifically Cronbach’s alpha and the corresponding correlation coefficients for three different data ranges (early, middle and latter), as well as the corresponding concordances. Overall Cronbach’s alpha for SBP, DBP, and mean arterial pressure were 0.96, 0.90, and 0.93, respectively. The correlation was consistent throughout the three-time distributions.

Table 2.

Cronbach’s Alpha (correlation coefficients [R]) for systolic BP, diastolic BP, and mean arterial pressures divided into three time periods.

Overall Cronbach’s Alpha 1st third Cronbach’s Alpha (R) 2nd third Cronbach’s Alpha (R) 3rd third Cronbach’s Alpha (R)
Systolic BP (Overall) 0.96 0.98 (0.96) 0.97 (0.94) 0.88 (0.79)
Diastolic BP (Overall) 0.90 0.84 (0.78) 0.96 (0.94) 0.92 (0.84)
Mean Arterial Pressure (Overall) 0.93 0.90 (0.82) 0.97 (0.95) 0.84 (0.74)

Discussion

The principal finding of this study is the level of agreement between the beat-by-beat BP by CT physiological monitor and the gold standard of invasively measured arterial BP in critically ill patients in the ICU was well within the ANSI/AAMI/ISO 81060–2:2013 standard guidelines14. Further, the inter-beat intervals were sufficient for higher resolution HRV tracking than is available with non-ECG systems. Comparison values were obtained over considerable BP ranges in an ICU patient cohort with a broad range of medical issues, physiologies, and ages, supporting the feasibility of this non-invasive and minimally-intrusive approach to hemodynamic monitoring.

The purpose of the study was to validate the PDA-based new technology in its continuous beat-by-beat BP measurement directly against the invasive arterial BP measurement. While correlations for both SBP and DBP were high, stronger correlation was found with SBP than DBP. While the Bland-Altman analysis for both SBP and DBP showed comparable agreement, there were a few clusters of measurement where CT DBP underestimates in high DBP range and overestimated in the low DBP range. This methodology in reporting both correlations and Bland-Altman analysis is in line with prior validation studies of other technologies1517.

A number of studies have concluded that current non-invasive BP monitoring technologies are not accurate enough for clinical use when considering the ANSI/AAMI/ISO 81060–2:2013 standard18,19. However, it is important to note that there are different challenges associated with the validation of non-invasive BP monitoring technologies. One is related to the comparison Gold Standards, which in many cases are an automated cuff. There is no standardization of algorithmic approaches, involving instead different proprietary analyses of the oscillometric amplitude envelope20. Equally important, the challenge in validation is also related to the broad range of variances in human physiology, rather than to just technological shortcomings. In the comparison of BP obtained from different physiological monitoring sites, the lack of generally applicable defined functional forms relating them has been well documented2123. Indeed, a recent study revealed the lack of a consistent general relationship even between SBP measured invasively at the brachial and radial arteries in a cohort of 180 subjects21. In this study, while 43% of subjects presented with differences of <5mmHg between the two measurement sites, the remainder presented with larger differences. There was a wide range in the magnitude of difference in radially obtained SBP. There may therefore be a limit to which non-invasive BP monitors can match the results obtained invasively, particularly across large patient groups with commensurately varied physiologies and pathologies.

The agreement of CT bias (95% limits) of −0.36 (7.57) and −2.11 (6.00) mmHg for SBP and DBP, respectively, exceeded those of other commercially available cNIBP technologies18. Studies using finger cuff devices, such as Nexfin/Clearsight (Edwards Lifesciences, Irvine, California, USA), CNAP (CNSystems, Graz, Austria), and Finapres Nova (FMS, Enschede, Netherlands) are all based on volume clamping method (Penaz principal), and yielded mixed results in both medical and surgical ICUs2428. A meta-analysis by Kim et. al reported pooled bias (95% limits of bias) of CNAP against the standard BP that was invasively measured to be −1.8 ± 12.8 mmHg (−26.8 to 23.2 mmHg) and 7.2 ± 8.5 mmHg (−9.5 to 24.0 mmHg) for SBP and DBP, respectively18. The pooled bias (95% limits of bias) of Nexfin against the invasively measured standard BP was −1.6 ± 8.4 mmHg (−18.1 to 15.0 mmHg) and 5.1 ± 6.6 mmHg (−7.8 to 18.0 mmHg) for SBP and DBP, respectively. The improved accuracy may be explained in the context of mechanical coupling. Importantly, while the Penaz principle employs active monitoring modality, CT employs passive one, performing the mechanical coupling to arterial wall. A study by others showed more accurate pulse transit time measured by the CT when compared to a Finapres, in recording sessions before and after exercise29.

Studies testing another method using arterial tonometry, such as the T-Line System (Tensys Medical, San Diego, California, USA), have also yielded mixed results3032. The same meta-analysis reported pooled bias (95% limits of bias) of T-Line System against the standard BP to be −0.1 ± 8.4 mmHg (−16.5 to 16.3 mmHg) and 2.9 ± 6.7 mmHg (−10.2 to 16.0 mmHg) for SBP and DBP, respectively18. While the precision appears comparable to that of CT shown in our study, 95% limits were much wider. Of note, this device is no longer commercially available.

Much work is currently underway to develop novel cNIBP devices that use indirect and surrogate measures to track BP, such as pulse transit time or pulse analysis using photoplethysmography (PPG). However, a major limitation is that these measures do not solely depend on changes in BP4,33,34. Instead, the surrogate measures are subject to other hemodynamically relevant parameters, such as HR or arterial stiffness, which can mask BP-related effects. The PDA model proposes a physiologically-based explanation of the structure of the arterial pulse that models and uncovers the interplay of these masking effects. This is different from approaches that seek to correlate features on the pulse alone. Pulse features alone cannot be counted on to persist permanently across the continuum of inter- and intra-subject physiological states. In other words, there is a great variability in the structure and form of the pulse envelope among subjects and even within the same subject. The PDA model seeks to correct for these physiological changes to give a more precise output.

With regard to HR measurement, CT yielded accurate inter-beat interval when compared to ECG. Specifically, the CT tracked inter-beat intervals within a mean difference of <1ms (6.0ms) over a range of inter-beats from 0.4–1.4s, corresponding to HR from 43–150 bpm, in a physiologically challenged population across a wide range of ages. While ECG-derived HR is considered the most reliable and accurate, electrocardiac pathologies, such as atrial fibrillation, can introduce significant HR uncertainties compared to a monitoring methodology that measures the actually ejected pressure/flow pulse, be it mechanically like the CT that couples to the arterial pulsation or devices that rely on PPG. Aside from certain arrhythmias, there are issues of resolution and comfort. While ECG-based HR excels at the former, longer-term wear of the electrodes can lead to skin breakdown and irritation. PPG devices, on the other hand, are generally comfortable, but provide lower resolution, as several studies have demonstrated3537. The inter-beat comparison results presented here suggest that mechanically coupling to the arterial pulse can provide a beneficial compromise, providing less intrusive monitoring, if implemented using a low-pressure finger cuff, while also providing improved resolution.

Limitations of the study include motion artifact issues that necessitated the exclusion of 2 patients. These artifacts were principally associated with the accommodations necessary to perform research within the clinical workflow of the intensive care unit. While the vasopressor status is unknown for each patient, no vasopressor or intravenous anti-hypertensive medication was administered during the recording period of the CT device. Further, no adjustments in vasoactive medications that may have been infusing were recorded for any patient. No patient had significant pathology of the central arterial system, such as an aneurysm, that would influence results, but prospective screening was not performed. Effects due to possible other abnormal central arterial anatomy are unknown.

The technology’s relevance in those settings will further increase with additional hemodynamic parameters, such as stroke volume, cardiac output, and left ventricular ejection time, which can be readily modeled within the PDA formalism. Further, the effect of aberrant anatomy, such as abdominal aortic aneurysms or renal artery stenosis, could be investigated in relation to CT BP readings. Validity of the CT in the setting of more extreme BP range and during the titration of vasoactive medications would warrant further investigation.

In conclusion, we have presented evidence that the non-invasive tracking of BP and HRV using the CT device and the PDA approach is possible within the guidelines of the ANSI/AAMI/ISO 81060–2:2013 standard. The accuracy exceeded that of existing cNIBP technologies. Based on the results presented coupled with the convenience of use, the CT has the potential to extend cNIBP monitoring to a wider patient population. Future studies would benefit from involving more heterogenous patient population in various clinical settings.

Acknowledgements:

This work has not been previously presented on the whole or in part.

Funding:

This work was partly supported by research grant provided by Caretaker Medical Inc. This was an investigator initiated study.

Footnotes

Publisher's Disclaimer: This is a PDF file of an unedited manuscript that has been accepted for publication. As a service to our customers we are providing this early version of the manuscript. The manuscript will undergo copyediting, typesetting, and review of the resulting proof before it is published in its final form. Please note that during the production process errors may be discovered which could affect the content, and all legal disclaimers that apply to the journal pertain.

Conflicts of Interest: Martin Baruch is an employee of the Caretaker Medical Inc., which funded the study.

Declaration of interests

The authors declare the following financial interests/personal relationships which may be considered as potential competing interests:

Martin Baruch is an employee of the Caretaker Medical Inc, which sponsored the study.

This work was partly supported by research grant provided by Caretaker Medical Inc. This was an investigator initiated study.

References

  • 1.Imholz BP, Wieling W, van Montfrans GA, et al. : Fifteen years experience with finger arterial pressure monitoring: assessment of the technology. Cardiovasc. Res 38:605–16, 1998 [DOI] [PubMed] [Google Scholar]
  • 2.Gershengorn HB, Garland A, Kramer A, et al. : Variation of arterial and central venous catheter use in United States intensive care units. Anesthesiology 120:650–64, 2014 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 3.Lakhal K, Ehrmann S, Boulain T: Noninvasive BP Monitoring in the Critically Ill: Time to Abandon the Arterial Catheter? Chest 153:1023–39, 2018 [DOI] [PubMed] [Google Scholar]
  • 4.Solà J, Delgado-Gonzalo R: The Handbook of Cuffless Blood Pressure Monitoring: A Practical Guide for Clinicians, Researchers, and Engineers. Springer Nature, 2019, [Google Scholar]
  • 5.Baruch MC, Warburton DE, Bredin SS, et al. : Pulse Decomposition Analysis of the digital arterial pulse during hemorrhage simulation. Nonlinear Biomed. Phys 5:1, 2011 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 6.Hersh LT, Friedman B, Luczyk W, et al. : Evaluation of filtering methods for acquiring radial intra-artery blood pressure waveforms. J. Clin. Monit. Comput 29:659–69, 2015 [DOI] [PubMed] [Google Scholar]
  • 7.Gratz I, Deal E, Spitz F, et al. : Continuous Non-invasive finger cuff CareTaker® comparable to invasive intra-arterial pressure in patients undergoing major intra-abdominal surgery. BMC Anesthesiol. 17:48, 2017 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 8.Callaghan FJ, Babbs CF, Bourland JD, et al. : The relationship between arterial pulse-wave velocity and pulse frequency at different pressures. J. Med. Eng. Technol 8:15–8, 1984 [DOI] [PubMed] [Google Scholar]
  • 9.Baruch M Pulse Decomposition Analysis Techniques. In: Sola J, Delgado-Gonzalo R, eds. The Handbook of Cuffless Blood Pressure Monitoring. Springer International Publishing; 2019:75–106. [Google Scholar]
  • 10.Kleiger RE, Stein PK, Bigger JT: Heart Rate Variability: Measurement and Clinical Utility. Ann. Noninvasive Electrocardiol 10:88–101, 2005 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 11.Martina JR, Westerhof BE, van Goudoever J, et al. : Noninvasive continuous arterial blood pressure monitoring with Nexfin®. Anesthesiology 116:1092–103, 2012 [DOI] [PubMed] [Google Scholar]
  • 12.Epstein RH, Bartkowski RR, Huffnagle S: Continuous noninvasive finger blood pressure during controlled hypotension. A comparison with intraarterial pressure. Anesthesiology 75:796–803, 1991 [DOI] [PubMed] [Google Scholar]
  • 13.Bland JM, Altman DG: Measuring agreement in method comparison studies. Stat. Methods Med. Res 8:135–60, 1999 [DOI] [PubMed] [Google Scholar]
  • 14.Association for the Advancement of Medical Instrumentation. Non-invasive Sphygmomanometers–Part 2: Clinical Investigation of Automated Measurement Type. ANSI/AAMI/ISO 81060–2/ANSI-AAMI 2nd ed. Arlington, VA: AAMI; 2019. [Google Scholar]
  • 15.Balzer F, Habicher M, Sander M, et al. : Comparison of the non-invasive Nexfin® monitor with conventional methods for the measurement of arterial blood pressure in moderate risk orthopaedic surgery patients. J. Int. Med. Res. SAGE Publications Ltd, 44:832–43, 2016 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 16.van der Does Y, van Loon LM, Alsma J, et al. : Non-invasive blood pressure and cardiac index measurements using the Finapres Portapres in an emergency department triage setting. Am. J. Emerg. Med 31:1012–6, 2013 [DOI] [PubMed] [Google Scholar]
  • 17.Duclos G, Hili A, Resseguier N, et al. : Clearsight use for haemodynamic monitoring during the third trimester of pregnancy - a validation study. Int. J. Obstet. Anesth 36:85–95, 2018 [DOI] [PubMed] [Google Scholar]
  • 18.Kim S-H, Lilot M, Sidhu KS, et al. : Accuracy and precision of continuous noninvasive arterial pressure monitoring compared with invasive arterial pressure: a systematic review and meta-analysis. Anesthesiology 120:1080–97, 2014 [DOI] [PubMed] [Google Scholar]
  • 19.Picone DS, Schultz MG, Otahal P, et al. : Accuracy of Cuff-Measured Blood Pressure: Systematic Reviews and Meta-Analyses. J. Am. Coll. Cardiol 70:572–86, 2017 [DOI] [PubMed] [Google Scholar]
  • 20.Babbs CF: Oscillometric measurement of systolic and diastolic blood pressures validated in a physiologic mathematical model. Biomed. Eng. Online 11:56, 2012 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 21.Armstrong MK, Schultz MG, Picone DS, et al. : Brachial and Radial Systolic Blood Pressure Are Not the Same. Hypertens. Dallas Tex 1979 73:1036–41, 2019 [DOI] [PubMed] [Google Scholar]
  • 22.Moore C, Dobson A, Kinagi M, et al. : Comparison of blood pressure measured at the arm, ankle and calf. Anaesthesia 63:1327–31, 2008 [DOI] [PubMed] [Google Scholar]
  • 23.Sareen P, Saxena K, Sareen B, et al. : Comparison of arm and calf blood pressure. Indian J. Anaesth 56:83–5, 2012 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 24.Ameloot K, Van De Vijver K, Van Regenmortel N, et al. : Validation study of Nexfin® continuous non-invasive blood pressure monitoring in critically ill adult patients. Minerva Anestesiol. 80:1294–301, 2014 [PubMed] [Google Scholar]
  • 25.Meidert AS, Huber W, Müller JN, et al. : Radial artery applanation tonometry for continuous non-invasive arterial pressure monitoring in intensive care unit patients: comparison with invasively assessed radial arterial pressure. Br. J. Anaesth 112:521–8, 2014 [DOI] [PubMed] [Google Scholar]
  • 26.Hohn A, Defosse JM, Becker S, et al. : Non-invasive continuous arterial pressure monitoring with Nexfin does not sufficiently replace invasive measurements in critically ill patients. Br. J. Anaesth 111:178–84, 2013 [DOI] [PubMed] [Google Scholar]
  • 27.Wagner JY, Negulescu I, Schöfthaler M, et al. : Continuous noninvasive arterial pressure measurement using the volume clamp method: an evaluation of the CNAP device in intensive care unit patients. J. Clin. Monit. Comput 29:807–13, 2015 [DOI] [PubMed] [Google Scholar]
  • 28.Ilies C, Grudev G, Hedderich J, et al. : Comparison of a continuous noninvasive arterial pressure device with invasive measurements in cardiovascular postsurgical intensive care patients: a prospective observational study. Eur. J. Anaesthesiol 32:20–8, 2015 [DOI] [PubMed] [Google Scholar]
  • 29.Phillips A, Burr J, Cote A, et al. : Comparing the Finapres and Caretaker Systems for Measuring Pulse Transit Time Before and After Exercise. Int. J. Sports Med 33:130–6, 2012 [DOI] [PubMed] [Google Scholar]
  • 30.Szmuk P, Pivalizza E, Warters RD, et al. : An evaluation of the T-Line Tensymeter continuous noninvasive blood pressure device during induced hypotension. Anaesthesia 63:307–12, 2008 [DOI] [PubMed] [Google Scholar]
  • 31.Saugel B, Fassio F, Hapfelmeier A, et al. : The T-Line TL-200 system for continuous non-invasive blood pressure measurement in medical intensive care unit patients. Intensive Care Med. 38:1471–7, 2012 [DOI] [PubMed] [Google Scholar]
  • 32.Langwieser N, Prechtl L, Meidert AS, et al. : Radial artery applanation tonometry for continuous noninvasive arterial blood pressure monitoring in the cardiac intensive care unit. Clin. Res. Cardiol. Off. J. Ger. Card. Soc 104:518–24, 2015 [DOI] [PubMed] [Google Scholar]
  • 33.Lantelme P, Mestre C, Lievre M, et al. : Heart rate: an important confounder of pulse wave velocity assessment. Hypertens. Dallas Tex 1979 39:1083–7, 2002 [DOI] [PubMed] [Google Scholar]
  • 34.springerprofessional.de. Retrieved from: https://www.springerprofessional.de/en/performance-of-systolic-blood-pressure-estimation-from-radial-pu/12357506.
  • 35.Pietilä J, Mehrang S, Tolonen J, et al. in: Eskola H, Väisänen O, Viik J, et al. , eds. (ed): EMBEC NBC 2017. Singapore, Springer, [Google Scholar]
  • 36.Pinheiro N, Couceiro R, Henriques J, et al. : Can PPG be used for HRV analysis? Annu. Int. Conf. IEEE Eng. Med. Biol. Soc. IEEE Eng. Med. Biol. Soc. Annu. Int. Conf 2016:2945–9, 2016 [DOI] [PubMed] [Google Scholar]
  • 37.Weiler DT, Villajuan SO, Edkins L, et al. : Wearable Heart Rate Monitor Technology Accuracy in Research: A Comparative Study Between PPG and ECG Technology. Proc. Hum. Factors Ergon. Soc. Annu. Meet. SAGE Publications Inc, 61:1292–6, 2017 [Google Scholar]

RESOURCES