The utilization of atrial sensing dipole in single lead implantable cardioverter defibrillator for detection of new‐onset atrial high‐rate episodes or subclinical atrial fibrillation: A systematic review and meta‐analysis

Xuanming Pung 1, Daniel Zhihao Hong 2, Tzyy Yeou Ho 2, Xiayan Shen 1, Pei Ting Tan 3, Colin Yeo 1, Vern Hsen Tan 1,
PMCID: PMC8977580  PMID: 35387136

Abstract

This meta‐analysis aims to evaluate the performance of atrial sensing dipole in single lead implantable cardioverter defibrillator (VDD‐ICD) recipients in particular diagnosing new‐onset atrial high‐rate episodes (AHREs) defined as rate threshold of 200 beats per minute, or subclinical atrial fibrillation (SCAF) defined as device‐detected AF without symptoms. We comprehensively searched PubMed, Embase, and ClinicalTrials.gov. Studies comparing contemporary single‐ and dual‐chamber ICD (VVI‐/DDD‐ICD) versus VDD‐ICD were included. Restricted maximum likelihood method for random effect model and Mantel‐Haenszel method for fixed effect model were used to estimate the effect size of new‐onset AHREs, or SCAF detection in each group. Three prospective studies were identified and total of 991 participants were included. There were 330 (33.3%) in VDD‐ICD and 661 (66.7%) in VVI‐/DDD‐ICD. Most (78%) participants were men. Median follow‐up was from 365 days to 847 days. VDD‐ICD has a higher likelihood of detecting AHREs or SCAF as compared to VVI‐/DDD‐ICD [(OR random effect : 2.6; 95% CI: 1.2, 5.8; p = .018); I‐squared = 67.8%, = .019]. This difference was more apparently seen in the comparison between VDD‐ICD and VVI‐ICD [(OR random effect: 3.8; 95% CI: 2.1, 6.6, p < .001), I‐squared = 0.0%, = .518]. The result is same as fixed effect. Rate of AHREs detection observed in VDD‐ICD was not statistically different when compared to the only group with DDD‐ICD from SENSE trial. In conclusion, this meta‐analysis reveals that the use of floating atrial sensing dipole in VDD‐ICD increases the detection of new‐onset AHREs or SCAF when compared to VVI‐ICD, with similar atrial sensing performance to DDD‐ICD.

Keywords: atrial high rate episodes, atrial sensing dipole, implantable cardiac defibrillator, subclinical atrial fibrillation


This meta‐analysis reveals that the use of a floating atrial sensing dipole in VDD‐ICD increases the detection of new‐onset AHREs or SCAF when compared to VVI‐ICD alone, with similar atrial sensing performance to DDD‐ICD. However, longer follow‐up is needed to determine the reliability of the atrial sensing component in the VDD‐ICD.

graphic file with name JOA3-38-177-g001.jpg

1. INTRODUCTION

Implantable cardioverter defibrillator (ICD) has become established as a guideline‐directed recommendation for primary and secondary prevention of sudden cardiac death. 1 Data from international registries suggested that more patients received dual‐chamber ICD (DDD‐ICD) rather than single‐chamber ICD (VVI‐ICD) for primary prevention despite the indication for atrial pacing being absent. 2

There is no clear consensus to guide the selection between single‐ and dual‐chamber ICD. Advantages in implantation of dual‐chamber ICD were presumed better discrimination between supraventricular and ventricular arrhythmias to reduce inappropriate therapies, allow monitoring of atrial fibrillation (AF), and to avoid an additional procedure to place an atrial lead if the need for bradycardia pacing arises. Ueda et al showed that only two patients (6.0%) of VVI‐ICD cohort required to undergo additional atrial lead insertion for bradycardia indication. 3 Recent publication by Ahmed et al by looking at impact of insurance status on ICD implantation practice patterns based on National Cardiovascular Data Registry ICD registry found that among patients without a clear indication for pacing, the uninsured were more likely to receive single‐ versus dual‐chamber ICDs compared to those with insurance in which the apparent difference remains unclear, which requires further study. 4

Atrial tachyarrhythmias detected on atrial leads, such as device‐detected AF, occur frequently in patients with cardiac implantable electronic devices (CIEDs) and are associated with an increased risk of stroke. 5 This in turn influences decision relating to initiation of anticoagulation for stroke prophylaxis.

A single lead ICD (VDD‐ICD) system with a floating atrial sensing dipole [LinoxSmartDX(Biotronik SE & Co, Berlin, Germany)] has been developed in early 2000s in order to improve the diagnostic capacity of atrial arrhythmias. The atrial dipole spacing covers a relatively large area of atrial surface of 49 mm2 which provides flexibility in positioning within the atrium in order to improve stability of atrial signal. There are two different configurations of the lead design: a 15 and 17 cm version based on the distance between the distal lead tip and mid‐point of atrial dipole. The VDD‐ICD was designed to amplify and filter the signals in order to maximize atrial sensing and minimize far‐field oversensing of ventricular signals. The concept of atrial sensing dipole has been proven safe and functional, meeting predefined clinical safety and efficacy of significantly higher than 90%. 6

Hence, we performed this meta‐analysis to review the atrial sensing performance of VDD‐ICD using an atrial floating dipole in the detection of new‐onset atrial high‐rate episodes (AHREs) or subclinical atrial fibrillation (SCAF) and compare it with VVI‐ and/or DDD‐ICD.

2. METHODS

This meta‐analysis was performed in adherence to the Population, Intervention, Comparison, and Outcome (PICO) framework, PRISMA (Preferred Reporting Items for Systematic Reviews and Meta‐Analyses) guidelines 7 and MOOSE (Meta‐analysis of Observational Studies in Epidemiology) checklist on the quality of reporting of meta‐analyses as shown in the Supplementary table S1 and S3.

Two independent reviewers (DZH and TYH) searched PubMed, EMBASE, clinicaltrials.gov, and Google Scholar for relevant articles on VDD‐ICD compare to VVI‐ and/or DDD‐ICD for the detection of new‐onset AHREs or SCAF. The following search terms were used in the literature review: (Atrial fibrillation) AND [(Implantable Cardioverter‐Defibrillator) OR (VDD‐ICD) OR (A+‐ICD) OR (floating atrial sensing dipole ICD lead)]. They also searched the reference lists of all published studies and biographies of review articles to identify additional articles. No language or publication status restriction were applied. The search was conducted from December 13, 2020 to March 28, 2021 and included studies up to March 28, 2021. Only studies that compared VDD‐ICD against VVI‐ and/or DDD‐ICD for detection of new‐onset AHREs or SCAF were included. The corresponding authors of the studies were also contacted to provide their unpublished data if any.

Study selection involved the screening of titles and abstracts, followed by full‐text evaluation of the eligible studies. The inclusion criteria were: (a) study population comprising two groups of patients receiving VDD‐ICD compared to VVI‐ and/or DDD‐ICD; and (b) reported new‐onset AHREs or SCAF.

The primary outcome was new‐onset AHREs defined as rate threshold of 200 beats per minute in these studies, or SCAF defined as device‐detected AF occurring without symptoms.

Data were independently extracted by two of the study investigators (DZH and TYH) using a standardized protocol and reporting form (Supplementary table S2). Any disagreements were resolved by arbitration, and consensus was achieved after panel discussion consisting of three reviewers (DZH, VHT, and TYH).

2.1. Statistical analysis

Data were pooled and analyzed using Stata version 16 statistical software. Effect size of new‐onset AHREs or SCAF detection was setup using two‐group comparison of binary outcomes. Restricted maximum likelihood method was used for random effect model and the Mantel‐Haenszel method for fixed effect model because of small number of events.

The overall effect size is presented using random effects as odds ratio (OR) with 95% confidence interval (CI). All statistical tests were two sided and used a significance level of p < .05. Heterogeneity between studies was evaluated using I2 statistic. The I2 statistic allows easy interpretation and overcomes the shortcomings of the Cochran's Q test. 8 It is the proportion of total variation observed among the studies that is attributable to differences between studies rather than sampling error (chance), with I2  values corresponding to the following levels of heterogeneity: low (<25%); moderate (25%–75%); and high (> 75%). 9 A subgroup analysis comparing recipients of VVD‐ICD against VVI‐ICD was performed to assess the robustness of the result.

3. RESULTS

Our database search resulted in 9026 unique articles. Of those, 76 were deemed relevant based on the title and abstract screenings. The full text of these 76 studies were assessed for eligibility, of which two full texts and one abstract met the final inclusion criteria and were included in the main analysis of this study as illustrated in Figure 1.

FIGURE 1.

FIGURE 1

Flow diagram of selection. Progress through the systematic review.

The three selected studies comprised cohort studies. There were no randomized controlled trials. Both the SENSE trial 10 and the THINGS registry 11 were prospective, cohort‐controlled, and multicenter studies comprising patients who met standard indications for a primary or secondary prevention ICD. The SENSE trial included two control groups (VVI‐/DDD‐ICD groups), while the VDD‐ICD system was only compared with conventional VVI‐ICD cohort in the THINGS registry and Statuto et al (abstract only). 12

The characteristics of the patients in these three studies are shown in Table 1. A total of 991 participants were included in this meta‐analysis, of whom 330 (33.3%) were in VDD‐ICD system and 661 (66.7%) in VVI‐/DDD‐ICD group. Most (78%) participants were men. Median follow‐up was from 365 to 847 days. Patients had neither prior history of AHREs or AF, nor requirement for antibradycardia atrial pacing.

TABLE 1.

Characteristics of patients included.

Type of device: VDD‐ICD, n (%)

VVI‐ICD, n (%)

DDD‐ICD, n (%)

Male, n (%) Age b , years Follow up, days (median) Left ventricle ejection fraction b , (%) HTN, n (%) DM, n (%) CCF, n (%) CKD, n (%) CVA/TIA, n (%) AC, n (%) AP, n (%) BB, n (%) ACEI/ARB

THINGS registry Biffi M et al

(n = 376)

140 (37.2%)

236 (62.8%)

0 (0%)

307 (81.6%)

Median:

Total: 64.6 (53.8–72.7)

VDD: 63.0 (54.1–69.2)

VVI: 65.9 (52.5–74.1)

821

Median:

All: 30 (25–35)

VVI: 30 (25–35)

VDD: 30 (25–35)

213 (57.4%) 101 (27.5%) 173 (46.8%) 47 (12.8%) 101 (27.5%) 53 (14.5%) 272 (74.5%) 333 (91.2%) 237 (64.9%)

SENSE trial

Thomas G et al a

(n = 450)

150 (33.3%)

150 (33.3%)

150 (33.3%)

324 (72%)

Mean ± SD

VDD: 59 ± 13

VVI: 54 ± 17

DDD: 59 ± 13

365

Mean ± SD:

VDD: 33 ± 17

VVI:31 ± 16

DDD: 33 ± 16

318 (70.7%) 144 (32%) 353 (78.4%) NA 33 (7.3%) 36 (8.0%) 262 (58.2%) 412 (91.6%) 258 (57.3%)

Statuto G et al

(n = 165)

40 (24.2%)

125 (75.8%)

0 (0%)

(79%)

Median

Total: 63 (48–72)

847 NA

AC, anti‐coagulation; ACEI, angiotensin‐converting enzyme inhibitor; AP, anti‐platelet; ARB, angiotensin receptor blockers; BB, beta blocker; CCF, congestive cardiac failure; CKD, chronic kidney disease; CVA, cerebrovascular accidents; DDD‐ICD, Dual‐chamber implantable cardiac defibrillator; DM, diabetes mellitus; HTN, hypertension; NA, not available; TIA, transient ischaemic attacks; VDD‐ICD, single lead floating atrial sensing dipole implantable cardiac defibrillator; VVI‐ICD, single‐chamber implantable cardiac defibrillator without atrial lead or floating atrial sensing dipole.

a

SENSE trial included data on DDD‐ICD patients.

b

Data presented as either mean, mean ± standard deviation or median (interquartile range).

Pooled result showed that VDD‐ICD system has higher likelihood in detecting AHREs or SCAF when compared to conventional VVI‐/DDD‐ICD as illustrated in Figure 2A and B. (OR random effect: 2.6 95% CI: 1.2, 5.8, p = .018; I 2 = 67.8%; P‐heterogeneity = 0.019).

FIGURE 2.

FIGURE 2

(A) Overall analysis: Forest plot for random effect of VDD‐ICD versus VVI‐/DDD‐ICD in detecting new‐onset AHRE or SCAF. (B) Overall analysis: Forest plot for fixed effect of VDD‐ICD versus VVI‐/DDD‐ICD in detecting new‐onset AHRE or SCAF. (C) Subgroup analysis: Forest plot for random effect of VDD‐ICD versus VVI‐ICD in detecting new‐onset AHRE or SCAF. (D) Subgroup analysis: Forest plot for fixed effect of VDD‐ICD versus VVI‐ICD in detecting new‐onset AHRE or SCAF.

Subgroup analysis demonstrated that VDD‐ICD has higher likelihood in detecting AHREs or SCAF when compared to VVI‐ICD alone. This is shown in Figure 2C. and D. (OR random effect: 3.8 95% CI: 2.1, 6.6, p < .001; I 2 = 0%; P‐heterogeneity = 0.519). As heterogeneity is not high, results of random effect and fixed effects are similar. Both results are presented in this meta‐analysis.

The P wave amplitude in VDD‐ICD system was assessed with regards to its atrial sensing performance. From the SENSE trial, a reduction in P wave amplitude to <2 mV was found in 7% of patients and P wave amplitude <1 mV in 3% of patients during 12 months of follow‐up. 10 The THINGS registry showed a statistical but clinical negligible decrease in P wave amplitude from device implantation to 2‐year follow‐up; mean valve was 5.5 mV. 11 Two VDD‐ICD recipients (1.4% of study population) from the registry had inadequate atrial sensing intermittently when P wave amplitude decreased to mean values of 0.4 mV, corresponding to the detection threshold programmed in the device (standard setting). Table 2 demonstrates the stability of atrial sensing performance of the VDD‐ICD system from our selected studies.

TABLE 2.

VDD‐ICD atrial sensing performance (P wave amplitude, mV).

At implantation 1 year 2 years

SENSE trial

Thomas G et al

(n = 150)

8.0 ± 5.0 (mean) 7.3 ± 4.8 (mean) NA

THINGS registry

Biffi M et al

(n = 140)

5.5 (median), Interquartile range: 3.5‐10.1 NA 5.5 (median), Interquartile range: 2.9‐8.0

NA, not available.

Incidence of inappropriate detection of AHREs was reported at 13% in the SENSE trial and most commonly because of electromagnetic interference (88.9%). No patient required device upgrade with the addition of an atrial lead for atrial pacing or poor atrial sensing. No inappropriate ICD therapies were observed in the VDD‐ICD group from the SENSE trial. 10 Device‐related adverse events were also reported in this trial. The incidence of adverse events was similar between the VDD‐ICD and VVI‐ICD groups.

4. DISCUSSION

In this meta‐analysis, there was a statistically significant increase in the detection rate of AHREs, or SCAF in VDD‐ICD system as compared to VVI‐ICD alone. The AHREs detection rate was not significantly different between the VDD‐ICD and DDD‐ICD groups. There are several advantages of VDD‐ICD system including shortened procedural times and ability to detect AHREs or SCAF events leading to early diagnosis and management of AF.

The implantation of the VDD‐ICD system shortened the total procedural times because of less lead implantation time required. 13 It also reduces the need for an additional atrial lead and associated lead complications, such as dislodgement. Early displacement occurs more frequently than late displacement, and these usually affect atrial leads. 14 Atrial lead dislodgement was reported in 4% of DDD‐ICD recipients in the ADRIA study. 13 Higher incidences of periprocedural complications, in‐hospital mortality, and need for earlier generator change in patients with DDD‐ICD compared to VVI‐ICD recipients were shown in studies. 2 , 15 Procedural complications in DDD‐ICD were associated with increased rates of procedural complications within 90 days of device implant (mechanical complication requiring reoperation for system, generator, and/or lead revision, device‐related infection, need for ICD replacement). There was no difference in 1‐year likelihood of re‐hospitalization for heart failure or all‐cause mortality.

Atrial tachyarrhythmias commonly occur in patients with CIEDs and include episodes of AHREs, or SCAF, which are associated with stroke. 5 However, AHREs are defined differently by various studies in particular with regards to temporal cut‐off (duration and frequency of AHREs). 16 , 17 , 18 , 19 Based on a scientific statement from the American Heart Association, AHREs are defined as device‐detected atrial events, usually tachyarrhythmias, meeting programmed or other specified atrial high‐rate criteria (usually ranging between 175 and 220 bpm); SCAF refers to asymptomatic episodes of atrial fibrillation detected and confirmed by intracardiac electrograms and not previously detected by electrocardiographic or ambulatory monitoring. 20

Device‐detected AHREs occurred in up to 20% of the patients by various studies and stroke risk differs between device‐detected (subclinical) and clinical AF for patients with identical stroke risk scores. 21 Continuous atrial monitoring may identify patients with episodes of SCAF. However, these patients could have a different and potentially lower stroke risk compared to patients with clinical AF as reported in various stroke prevention trials. 16 , 22 , 23

Benefits of oral anticoagulation in new‐onset AHREs, or SCAF is uncertain because of lack of clear evidence on treatment threshold. 24 , 25 , 26 Data from the Veterans Health Administration showed wide variation in physician interpretation of device‐detected AF and 90‐day oral anticoagulation initiation across clinical practice. Highest risk of stroke was observed in patients with at least one episode of device‐detected AF >24 h, and the risk was reduced in those prescribed with anticoagulation. 27 Ongoing trials such as ARTESIA (Apixaban for the Reduction of Thrombo‐Embolism in Patients With Device‐Detected Sub‐Clinical Atrial Fibrillation) 28 and NOAH (Non‐Vitamin K Antagonist Oral Anticoagulants in Patients With Atrial High‐Rate Episodes) 19 will provide evidence to guide decision on anticoagulation for these AHREs and SCAF episodes detected.

On the other hand, there may be several potential issues with VDD‐ICD system in terms of atrial sensing. Marai et al. stated concern with long‐term reliability on P wave amplitude during the follow‐up period of 12 months. 29 It was a single center prospective experience that recruited all patients with P wave amplitude at implant of ≥0.8 mV. By 1‐year postimplantation, 11% of patients had a P wave amplitude <0.8, and one inappropriate shock was observed in a patient whose P wave amplitude dropped to 0.2 mV. Besides, the study also showed that the mean P wave amplitude was higher among the 15‐cm leads compared to the 17‐cm leads up to 12 months postimplantation. Similarly, the Linox DX study demonstrated that P wave in the 17 cm tip‐to‐ring configuration was significantly smaller than that observed in the 15 cm tip‐to‐ring distance and the rate of appropriate atrial sensing was also significantly lower. 6

However, our meta‐analysis showed that there was a nonclinically significant trend in the decrease in P wave amplitude observed in the VDD‐ICD over the period of follow‐up although none of the studies comparing the two different leads configuration (17 vs. 15 cm). Other studies have shown acceptable P wave amplitudes in the VDD‐ICD system during follow‐up, although none specifically assessed the frequency of P wave amplitudes <0.8 mV. 6 , 13 , 29 Worden et al. showed that satisfactory atrial sensing with floating atrial dipole is achievable and stable in time in the majority of patients. 30 Safak et al. also reported long‐term stability in both atrial and ventricular signals with the newer generation VDD‐ICD over the entire follow‐up period of nearly 2 years. 31 In term of P wave amplitude in comparing VDD‐ICD versus DDD‐ICD, based on SENSE trial, the amplified P wave amplitude at 1 year for VDD‐ICD remained stable (7.3 ± 4.8 mV) compared to during implantation (8.0 ± 5.0 mV). 10 For DDD‐ICD group (separate atrial lead), the mean sensed atrial amplitude was 8.0 ± 5.0 mV at implant and remained stable at 12‐month follow‐up (7.3 ± 4.8 mV).

Currently, all CIEDs carry detection algorithms with high sensitivities and specificities for discriminating atrial and ventricular signals derived from CIEDs. 32 , 33 , 34 Newer algorithmic approaches to AF detection have been developed. Multiple independent, real‐world analyses revealed an incidence rate of between 27% and 34% atrial fibrillation in the 6 months postimplant of single chamber Visia AF ICD. 35 , 36 In our meta‐analysis, there was a 13% newly detected AHREs in VDD group at 12 months from the SENSE trial, 10 and a 2‐year incidence of AT/AF diagnosis of 11.4% in VDD group from THINGS registry. 11  To date, no head‐to‐head comparison is available.

Increased frequency of inappropriate shocks has been associated with worse outcomes, 37 impairs quality of life, and increases the mortality rate. 38 Atrial sensing enables essential information from the atrium to be acquired and integrated into the detection algorithm to increase the accuracy in differentiating supraventricular arrhythmias from ventricular arrhythmias. Often, AF or SVT accounts for the leading cause of inappropriate shock therapy in ICD recipients, 39 despite algorithms present to discriminate ventricular tachycardia (VT) from AF or SVT. The implantation of DDD‐ICD provides advantage over VVI‐ICD in its capability to analyze both atrial and ventricular rates and atrio‐ventricular relationship. However, DDD‐ICD may introduce new SVT‐VT discrimination problems. 40 Evidence is conflicting as to whether dual‐chamber discrimination algorithms perform better than single chamber discriminators in reducing inappropriate shock therapy. 41

The landmark ADRIA study evaluated the role of the SMART detection enhancement algorithm in VDD‐ICD with floating atrial sensing dipole, comparing its efficacy to DDD‐ICD. 13 In that study, there was a higher prevalence of small atrial signals seen before amplification in the VDD‐ICDs, which can give rise to more episodes of under‐ or oversensing events. These episodes were predominantly P wave undersensing or far‐field R‐wave oversensing; they were mostly clinically silent. Studies reported the incidence of undersensing to be 5% ‐ 7%, the vast majority of which did not require any specific therapy. 42 , 43 , 44 Electromagnetic interference (particularly in contact with the mains) remained the most common cause of atrial oversensing in the VDD‐ICD. 45 , 46 This was likely attributed to floating atrial dipole sensor that was amplified up to four times coupled with sampling rate as well as proprietary Biotronik digital noise gating algorithm.

Another important aspect is inclusion of conference abstract in systematic review and meta‐analysis. Conference abstracts without full publications remain a controversy subject whether to be included in systematic review and meta‐analysis. However, Cochrane and the United States National Academy Sciences both recommend always searching for and including conference abstracts in systematic review. 47 , 48 In addition, this issue was discussed in depth and published recently by Scherer et al. 49 Essentially, Scherer et al concluded that it is worthwhile to search for and include results from conference abstracts in systematic reviews. Table 3 showed the resultant random effect for all studies when abstract was excluded which showed poor estimate (not significant) when the sample size was small as compared to fixed effect of all studies (Figure 3A—D). It became clearer that VDD‐ICD significantly increased detection of new‐onset atrial high‐rate episodes or subclinical atrial fibrillation (both random and fixed effect for all studies) when abstract was included.

TABLE 3.

Random and fixed effect from study analysis with and without abstract.

With abstract Without abstract
All studies included
Random effect Significantly favour VDD‐ICD Favour VDD‐ICD but not significant
Fixed effect Significantly favour VDD‐ICD Significantly favour VDD‐ICD
Single vs. DX a
Random effect Significantly favour VDD‐ICD Significantly favour VDD‐ICD
Fixed effect Significantly favour VDD‐ICD Significantly favour VDD‐ICD
a

Single referred to single lead implantable cardiac defibrillator without atrial lead or floating atrial sensing dipole; DX (VDD‐ICD) referred to single lead floating atrial sensing dipole implantable cardiac defibrillator.

FIGURE 3.

FIGURE 3

(A) Overall analysis without abstract: Forest plot for random effect of VDD‐ICD versus VVI‐/DDD‐ICD in detecting new‐onset AHRE or SCAF. (B) Overall analysis without abstract: Forest plot for fixed effect of VDD‐ICD versus VVI‐/DDD‐ICD in detecting new‐onset AHRE or SCAF. (c) Subgroup analysis without abstract: Forest plot for random effect of VDD‐ICD versus VVI‐ICD in detecting new‐onset AHRE or SCAF. (d) Subgroup analysis without abstract: Forest plot for fixed effect of VDD‐ICD versus VVI‐ICD in detecting new‐onset AHRE or SCAF.

5. LIMITATIONS

The studies included in this meta‐analysis are nonrandomized studies, which have weaknesses that are inherent to observational data. There may be selection bias given the use of VDD‐, VVI‐, or DDD‐ICD in these nonrandomized, noncontrolled data sources. There is an on‐going randomized clinical trial comparing recipients of VDD‐ICD and VVI‐ICD on the rate of SCAF detection. 50

Further studies could attempt to continue even longer term follow‐up of patients with the VDD‐ICD system to determine whether falling trend of atrial sensing over time from measurement of P wave amplitude translates into adverse clinical outcomes (e.g., undersensing and oversensing of atrial arrhythmias, inappropriate ICD shocks, or need for system revision).

Studies are also needed to establish the actual incidence of inappropriate AHREs or SCAF detection, and how we could mitigate these issues (e.g., electromagnetic interference, or lead dislodgement), which may lead to the incorrect prescription of anticoagulation therapy. With regards to cardiac resynchronization therapy (CRT), evidence is still limited in the utilization of a floating atrial sensing dipole as seen in the two‐lead CRT system.

6. CONCLUSION

In conclusion, this meta‐analysis reveals that the use of a floating atrial sensing dipole in VDD‐ICD increases the detection of new‐onset AHREs or SCAF when compared to VVI‐ICD alone, with similar atrial sensing performance to DDD‐ICD. However, longer follow‐up is needed to determine the reliability of the atrial sensing component in the VDD‐ICD.

DISCLOSURE

None.

Supporting information

Table S1‐S3

Pung X, Hong DZ, Ho TY, Shen X, Tan PT, Yeo C, et al. The utilization of atrial sensing dipole in single lead implantable cardioverter defibrillator for detection of new‐onset atrial high‐rate episodes or subclinical atrial fibrillation: A systematic review and meta‐analysis. J Arrhythmia. 2022;38:177–186. 10.1002/joa3.12675

REFERENCES

  • 1. Priori SG, Blomström‐Lundqvist C, Mazzanti A, Blom N, Borggrefe M, Camm J, et al. 2015 ESC Guidelines for the management of patients with ventricular arrhythmias and the prevention of sudden cardiac death: The Task Force for the Management of Patients with Ventricular Arrhythmias and the Prevention of Sudden Cardiac Death of the European Society of Cardiology (ESC). Endorsed by: Association for European Paediatric and Congenital Cardiology (AEPC). Eur Heart J. 2015;36(41):2793–867. [DOI] [PubMed] [Google Scholar]
  • 2. Peterson PN, Varosy PD, Heidenreich PA, Wang Y, Dewland TA, Curtis JP, et al. Association of single‐ vs dual‐chamber ICDs with mortality, readmissions, and complications among patients receiving an ICD for primary prevention. JAMA. 2013;309(19):2025–34. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 3. Ueda A, Oginosawa Y, Soejima K, Abe H, Kohno R, Ohe H, et al. Outcomes of single‐ or dual‐chamber implantable cardioverter defibrillator systems in Japanese patients. Journal of Arrhythmia. 2016;32(2):89–94. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 4. Ahmed I, Merchant FM, Curtis JP, Parzynski CS, Lampert R. Impact of insurance status on ICD implantation practice patterns: insights from the NCDR ICD registry. Am Heart J. 2021;05(235):44–53. 10.1016/j.ahj.2021.01.016 [DOI] [PubMed] [Google Scholar]
  • 5. Go AS, Mozaffarian D, Roger VL, Benjamin EJ, Berry JD, Blaha MJ, et al. Executive summary: heart disease and stroke statistics–2014 update: a report from the American Heart Association. Circulation. 2014;129(3):399–410. [DOI] [PubMed] [Google Scholar]
  • 6. Safak E, Schmitz D, Konorza T, Wende C, Ros JOD, Schirdewan A. Clinical efficacy and safety of an implantable cardioverter‐defibrillator lead with a floating atrial sensing dipole. Pacing Clin Electrophysiol. 2013;36(8):952–62. [DOI] [PubMed] [Google Scholar]
  • 7. Moher D, Liberati A, Tetzlaff J, Altman DG, Group P. Preferred reporting items for systematic reviews and meta‐analyses: the PRISMA statement. PLoS Med. 2009;6(7):e1000097. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 8. Huedo‐Medina TB, Sánchez‐Meca J, Marin‐Martinez F, Botella J. Assessing heterogeneity in meta‐analysis: Q statistic or I² index? Psychol Methods. 2006;11(2):193. [DOI] [PubMed] [Google Scholar]
  • 9. Higgins JPT, Thompson SG. Quantifying heterogeneity in a metaanalysis. Stat Med. 2002;21:1539–58. [DOI] [PubMed] [Google Scholar]
  • 10. Thomas G, Choi DY, Doppalapudi H, Richards M, Iwai S, Daoud EG, et al. Subclinical atrial fibrillation detection with a floating atrial sensing dipole in single lead implantable cardioverter‐defibrillator systems: Results of the SENSE trial. J Cardiovasc Electrophysiol. 2019;30(10):1994–2001. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 11. Biffi M, Iori M, De Maria E, Bolognesi MG, Placci A, Calvi V, et al. The role of atrial sensing for new‐onset atrial arrhythmias diagnosis and management in single‐chamber implantable cardioverter‐defibrillator recipients: Results from the THINGS registry. J Cardiovasc Electrophysiol. 2020;31(4):846–53. [DOI] [PubMed] [Google Scholar]
  • 12. Statuto G, Paolisso P, Massaro G, Angeletti A, Lorenzetti S, Ziacchi M, et al. Atrial fibrillation and single chamber ICDS: implications of atrial signal availability. European Heart Journal Supplements. 2019;21:J27–8. [Google Scholar]
  • 13. Sticherling C, Zabel M, Spencker S, Meyerfeldt U, Eckardt L, Behrens S, et al. Comparison of a novel, single‐lead atrial sensing system with a dual‐chamber implantable cardioverter‐defibrillator system in patients without antibradycardia pacing indications: results of a randomized study. Circ Arrhythm Electrophysiol. 2011;4(1):56–63. [DOI] [PubMed] [Google Scholar]
  • 14. Gul EE, Kayrak M. Common pacemaker problems: lead and pocket complications. Modern Pacemakers‐Present and Future/ed. MR Das. Rijeka: Croatia; 2011 Feb 14. p. 299–318.
  • 15. Dewland TA, Pellegrini CN, Wang Y, Marcus GM, Keung E, Varosy PD. Dual‐chamber implantable cardioverter‐defibrillator selection is associated with increased complication rates and mortality among patients enrolled in the NCDR implantable cardioverter‐defibrillator registry. J Am Coll Cardiol. 2011;58(10):1007–13. [DOI] [PubMed] [Google Scholar]
  • 16. Glotzer TV, Daoud EG, Wyse DG, Singer DE, Ezekowitz MD, Hilker C, et al. The relationship between daily atrial tachyarrhythmia burden from implantable device diagnostics and stroke risk: the TRENDS study. Circ Arrhythm Electrophysiol. 2009;2(5):474–80. [DOI] [PubMed] [Google Scholar]
  • 17. Hohnloser SH, Capucci A, Fain E, Gold MR, van Gelder IC, Healey J, et al. ASymptomatic atrial fibrillation and Stroke Evaluation in pacemaker patients and the atrial fibrillation Reduction atrial pacing Trial (ASSERT). Am Heart J. 2006;152(3):442–7. [DOI] [PubMed] [Google Scholar]
  • 18. Ip J, Waldo AL, Lip GY, Rothwell PM, Martin DT, Bersohn MM, et al. Multicenter randomized study of anticoagulation guided by remote rhythm monitoring in patients with implantable cardioverter‐defibrillator and CRT‐D devices: rationale, design, and clinical characteristics of the initially enrolled cohort: the IMPACT study. Am Heart J. 2009;158(3):364–70. [DOI] [PubMed] [Google Scholar]
  • 19. Kirchhof P, Blank BF, Calvert M, Camm AJ, Chlouverakis G, Diener H‐C, et al. Probing oral anticoagulation in patients with atrial high rate episodes: Rationale and design of the Non‐vitamin K antagonist Oral anticoagulants in patients with Atrial High rate episodes (NOAH‐AFNET 6) trial. Am Heart J. 2017;190:12–8. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 20. Noseworthy PA, Kaufman ES, Chen LY, Chung MK, Elkind MSV, Joglar JA, et al. American Heart Association Council on Clinical Cardiology Electrocardiography and Arrhythmias Committee; Council on Arteriosclerosis, Thrombosis and Vascular Biology; Council on Cardiovascular and Stroke Nursing; and Stroke Council. Subclinical and device‐detected atrial fibrillation: pondering the knowledge gap: a scientific statement from the American Heart Association. Circulation. 2019;140(25):e944–63. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 21. Ballatore A, Matta M, Saglietto A, Desalvo P, Bocchino PP, Gaita F, et al. Subclinical and asymptomatic atrial fibrillation: current evidence and unsolved questions in clinical practice. Medicina. 2019;55(8):497. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 22. Chen‐Scarabelli C, Scarabelli TM, Ellenbogen KA, Halperin JL. Device‐detected atrial fibrillation: what to do with asymptomatic patients? J Am Coll Cardiol. 2015;65(3):281–94. [DOI] [PubMed] [Google Scholar]
  • 23. Gorenek BC, Bax J, Boriani G, Chen SA, Dagres N, Glotzer TV, et al. Device‐detected subclinical atrial tachyarrhythmias: definition, implications and management‐an European Heart Rhythm Association (EHRA) consensus document, endorsed by Heart Rhythm Society (HRS), Asia Pacific Heart Rhythm Society (APHRS) and Sociedad Latinoamericana de Estimulacion Cardiaca y Electrofisiologia (SOLEACE). Europace. 2017;19(9):1556–78. [DOI] [PubMed] [Google Scholar]
  • 24. January CT, Wann LS, Alpert JS, Calkins H, Cigarroa JE, Cleveland JC, et al. 2014 AHA/ACC/HRS guideline for the management of patients with atrial fibrillation: a report of the American College of Cardiology/American Heart Association Task Force on practice guidelines and the Heart Rhythm Society. Circulation. 2014;130(23):e199–267. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 25. Kirchhof P, Benussi S, Kotecha D, Ahlsson A, Atar D, Casadei B, et al. 2016 ESC Guidelines for the management of atrial fibrillation developed in collaboration with EACTS. Eur Heart J. 2016;37(38):2893–962. [DOI] [PubMed] [Google Scholar]
  • 26. Martin DT, Bersohn MM, Waldo AL, Wathen MS, Choucair WK, Lip GYH, et al. Randomized trial of atrial arrhythmia monitoring to guide anticoagulation in patients with implanted defibrillator and cardiac resynchronization devices. Eur Heart J. 2015;36(26):1660–8. [DOI] [PubMed] [Google Scholar]
  • 27. Perino AC, Fan J, Askari M, Heidenreich PA, Keung E, Raitt MH, et al. Practice variation in anticoagulation prescription and outcomes after device‐detected atrial fibrillation. Circulation. 2019;139(22):2502–12. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 28. Lopes RD, Alings M, Connolly SJ, Beresh H, Granger CB, Mazuecos JB, et al. Rationale and design of the apixaban for the reduction of thrombo‐embolism in patients with device‐detected sub‐clinical atrial fibrillation (ARTESIA) trial. Am Heart J. 2017;189:137–45. [DOI] [PubMed] [Google Scholar]
  • 29. Marai I, Milman A, Diamante R, Gurevitz O, Barlev D, Lipchenka I, et al. The efficacy of the Linox(Smart) DX ICD lead from a single center experience. Indian Pacing Electrophysiol J. 2020;20(4):137–40. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 30. Worden NE, Alqasrawi M, Mazur A. Long‐term stability and clinical utility of amplified atrial electrograms in a single lead ICD system with floating atrial electrodes. Pacing Clin Electrophysiol. 2016;39:1327–34. [DOI] [PubMed] [Google Scholar]
  • 31. Safak E, D´Ancona G, Kaplan H, Caglayan E, Kische S, Öner A, et al. New generation cardioverter‐defibrillator lead with a floating atrial sensing dipole: Long‐term performance. Pacing Clin Electrophysiol. 2018;41(2):128–35. [DOI] [PubMed] [Google Scholar]
  • 32. Francia P, Balla C, Uccellini A, Cappato R. Arrhythmia detection in single‐ and dual‐chamber implantable cardioverter defibrillators: the more leads, the better? J Cardiovasc Electrophysiol. 2009;20(9):1077–82. [DOI] [PubMed] [Google Scholar]
  • 33. Friedman PA, McClelland RL, Bamlet WR, Acosta H, Kessler D, Munger TM, et al. Dual‐chamber versus single‐chamber detection enhancements for implantable defibrillator rhythm diagnosis: the detect supraventricular tachycardia study. Circulation. 2006;113(25):2871–9. [DOI] [PubMed] [Google Scholar]
  • 34. Sinha A‐M, Stellbrink C, Schuchert A, Mox B, Jordaens L, Lamaison D, et al. Clinical experience with a new detection algorithm for differentiation of supraventricular from ventricular tachycardia in a dual‐chamber defibrillator. J Cardiovasc Electrophysiol. 2004;15(6):646–52. [DOI] [PubMed] [Google Scholar]
  • 35. Boriani G, et al. Understanding the incidence of AF in single‐chamber ICD patients: a real‐world analysis. Presented at. Europace. 2017.
  • 36. Brown ML, et al. New AF occurrence in single‐chamber ICD patients: insights from a real‐world investigation. Presented at HRS. 2018.
  • 37. Ruwald AC, Pietrasik G, Goldenberg I, Kutyifa V, Daubert JP, Ruwald MH, et al. The effect of intermittent atrial tachyarrhythmia on heart failure or death in cardiac resynchronization therapy with defibrillator versus implantable cardioverter‐defibrillator patients: a MADIT‐CRT substudy (Multicenter Automatic Defibrillator Implantation Trial With Cardiac Resynchronization Therapy). J Am Coll Cardiol. 2014;63(12):1190–7. [DOI] [PubMed] [Google Scholar]
  • 38. Poole JE, Johnson GW, Hellkamp AS, Anderson J, Callans DJ, Raitt MH, et al. Prognostic importance of defibrillator shocks in patients with heart failure. N Engl J Med. 2008;359(10):1009–17. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 39. Peterson PN, Greenlee RT, Go AS, Magid DJ, Cassidy‐Bushrow A, Garcia‐Montilla R, et al. Comparison of inappropriate shocks and other health outcomes between single‐ and dual‐chamber implantable cardioverter‐defibrillators for primary prevention of sudden cardiac death: results from the cardiovascular research network longitudinal study of implantable cardioverter‐defibrillators. J Am Heart Assoc. 2017;6(11):e006937. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 40. Swerdlow CD. Supraventricular tachycardia‐ventricular tachycardia discrimination algorithms in implantable cardioverter defibrillators: state‐of‐the‐art review. J Cardiovasc Electrophysiol. 2001;12(5):606–12. [DOI] [PubMed] [Google Scholar]
  • 41. Gonçalves J, Pereira T. Inappropriate shocks in patients with ICDs: single chamber versus dual chamber. Arq Bras Cardiol. 2013;101:141–8. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 42. Kuzniec J, Golovchiner G, Mazur A, Battler A, Strasberg B. Atrial signal amplitude measurements in VDD systems according to body position. Pacing Clin Electrophysiol. 2004;27(10):1344–6. [DOI] [PubMed] [Google Scholar]
  • 43. Van Campen CM, De Cock CC, Huijgens J, Visser CA. Clinical relevance of loss of atrial sensing in patients with single lead VDD pacemakers. Pacing Clin Electrophysiol. 2001;24(5):806–9. [DOI] [PubMed] [Google Scholar]
  • 44. Wiegand UK, Nowak B, Reisp U, Peiffer T, Bode F, Potratz J, et al. Implantation strategy of the atrial dipole impacts atrial sensing performance of single lead VDD pacemakers. Pacing Clin Electrophysiol. 2002;25(3):316–23. [DOI] [PubMed] [Google Scholar]
  • 45. Stroobandt SY, Brieda M, Allocca G, Stroobandt RX. ICD sees what you do not see: how does it beat you? Pacing Clin Electrophysiol. 2015;38(4):529–33. [DOI] [PubMed] [Google Scholar]
  • 46. Konstantino Y, Kleiman A, Amit G. Home monitoring report from a single lead Lumax DX implantable cardioverter defibrillator: new observations in a new system. Journal of Arrhythmia. 2016;32(3):218–9. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 47. The Cochrane Collaboration . Cochrane handbook for systematic reviews of interventions version 5.1.0 [updated March 2011]. 2011.
  • 48. Balshem HSA, Ansari M, Norris S, Kansagara D, Shamliyan T, Chou R, et al. Methods Guide for Effectiveness and Comparative Effectiveness Reviews. Finding Grey Literature Evidence and Assessing for Outcome and Analysis Reporting Biases When Comparing Medical Interventions: AHRQ and the Effective Health Care Program. 2008. [PubMed] [Google Scholar]
  • 49. Scherer RW, Saldanha IJ. How should systematic reviewers handle conference abstracts? A view from the trenches. Syst Rev. 2019;8(1):264. doi: 10.1186/s13643-019-1188-0 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 50. Shurrab M, Janmohamed A, Sarrazin J‐F, Ayala‐Paredes F, Sturmer M, Williams R, et al. The Dx‐AF study: a prospective, multicenter, randomized controlled trial comparing VDD‐ICD to VVI‐ICD in detecting sub‐clinical atrial fibrillation in defibrillator patients. J Interv Card Electrophysiol. 2017;50(1):57–63. [DOI] [PubMed] [Google Scholar]

Associated Data

This section collects any data citations, data availability statements, or supplementary materials included in this article.

Supplementary Materials

Table S1‐S3


Articles from Journal of Arrhythmia are provided here courtesy of Japanese Heart Rhythm Society

RESOURCES