Abstract
Current sit‐to‐stand methods measuring dynamic cerebral autoregulation (dCA) do not capture the precise onset of the time delay (TD) response. Reduced sit‐to‐stand reactions in older adults and individuals post‐stroke could inadvertently introduce variability, error, and imprecise timing. We applied a force sensor during a sit‐to‐stand task to more accurately determine how TD before the onset of dCA may be altered. Middle cerebral artery blood velocity (MCAv) and mean arterial pressure (MAP) were measured during two sit‐to‐stands separated by 15 min. Recordings started with participants sitting on a force‐sensitive resistor for 60 s, then asked to stand for 2 min. Upon standing, the force sensor voltage immediately dropped and marked the exact moment of arise‐and‐off (AO). Time from AO until an increase in cerebrovascular conductance (CVC = MCAv/MAP) was calculated as TD. We tested the sensor in four healthy young adults, two older adults, and two individuals post‐stroke. Healthy young adults stood quickly and the force sensor detected a small change in TD compared to classically estimated AO, from verbal command to stand. When compared to the estimated AO, older adults had a delayed measured AO and TD decreased up to ~53% while individuals post‐stroke had an early AO and TD increased up to ~14%. The stance time during the sit‐to‐stand has the potential to influence the TD before the onset of dCA metric. As observed in the older adults and participants with stroke, this response may drastically vary and influence TD.
Keywords: cerebral blood flow, cerebral blood velocity, dCA, middle cerebral artery blood velocity
The novel application of a force sensor during a sit‐to‐stand in a single healthy young adult (n = 1), older adult (n = 1), and individual post‐stroke (n = 1) was studied. Individuals were asked to perform the sit‐to‐stand at 60 s and a force sensor detected arise‐and‐off (AO) when participants stood from the chair. Cerebrovascular conductance (CVC) was used to measure the time delay before the onset of dynamic cerebral autoregulation.

1. INTRODUCTION
The brain requires constant cerebral blood flow, even during perturbations to the systemic circulation (Willie et al., 2014). Dynamic cerebral autoregulation (dCA) is the ability of the brain to maintain blood flow by altering cerebrovascular conductance (CVC) during perturbations to circulation, such as changes in peripheral blood pressure (Beek et al., 2008; Brassard et al., 2021; Roy & Sherrington, 1890). Recording the dCA response has become prevalent with the use of transcranial Doppler ultrasound to measure middle cerebral artery blood velocity (MCAv) and finger plethysmography to measure mean arterial pressure (MAP) (Beek et al., 2008; Panerai, 1998). A sit‐to‐stand transition is a common experimental perturbation utilized to measure the temporal regulatory response of dCA (Sorond et al., 2009). When performing a sit‐to‐stand, peripheral blood pressure transiently drops due to gravitational venous pooling and vasodilation (Olufsen et al., 2005). To counteract the drop in blood pressure, dCA responds by increasing the CVC of MCAv to the cerebrovascular system before MAP increases in the peripheral vascular system (Labrecque et al., 2017) The time delay (TD) before the onset of the dCA response is measured as the elapsed time before an increase in CVC (Labrecque et al., 2017; Lind‐Holst et al., 1985).
While previous studies in healthy young and older adults have performed the sit‐to‐stand technique to measure the temporal dCA response (Deegan et al., 2009; Labrecque et al., 2019, 2017; Lipsitz et al., 2000; Moir et al., 2020; Serrador et al., 2005; Sorond et al., 2009; Tzeng et al., 2010), the exact methods of characterizing dCA have varied between studies. While some report having the individual stand quickly (within 0–3 s) and mark the initiation of standing as when the verbal command was given (Labrecque et al., 2019, 2017; Moir et al., 2020; Tzeng et al., 2010), others report having the individual place their feet on a stool in front of them (legs at 90 degrees) and mark the initiation of standing from when their feet touch the floor (Deegan et al., 2009; Lipsitz et al., 2000; Serrador et al., 2005; Sorond et al., 2009). However, the moment of standing from a seated position is not objectively marked (Stevermer & Gillette, 2016). The start of the postural change, the stimulus which drops peripheral blood pressure, is critical for measuring the TD before the onset of the dCA response and has been indeterminate in prior protocols. Considering the sit‐to‐stand transition is the stimulus for the change in hemodynamics, not measuring the exact moment stance occurs could introduce error and imprecise timing of dCA onset.
Prior studies have implemented a tilt sensor to objectively measure the angle of the squatting motion during dCA repeated squat‐stand maneuvers (Barnes et al., 1985, 2017, 2018, 2021; Batterham et al., 2020; Panerai et al., 2021) and one study examined the differing transition time between young and older adults during a stand‐to‐sit with a goniometer (Klein et al., 2020). However, it was unclear if the transition time detected by the goniometer was measured simultaneously with the physiologic response to improve the accuracy of the TD before the onset of the dCA response (Klein et al., 2020). A force sensor to objectively measure the time of stance while simultaneously measuring the physiologic response during a single sit‐to‐stand is needed to improve the temporal precision of the TD response. Measuring the exact moment of stance is especially important in older adults and clinical populations, such as stroke who may present with reduced lower extremity strength, resulting in a delayed sit‐to‐stand reaction (Bohannon, 2006; Mentiplay et al., 2020; Mong et al., 2010). Current methodologies are unable to account for the potential delay between the initiation and the stance phase of a sit‐to‐stand (Stevermer & Gillette, 2016).
To address this gap in the literature, we developed a force sensor to determine the kinetic moment of arise‐and‐off (AO), or the exact moment of standing up from the seat (Stevermer & Gillette, 2016). Compared to estimating AO from a verbal command similar to previous work (Labrecque et al., 2019, 2017; Moir et al., 2020; Tzeng et al., 2010), we hypothesized that measuring the exact moment of AO, measured with a force sensor, would alter the calculation of the TD before the onset of the dCA response. We conducted the procedure twice separated by 15 min (T1 = first sit to stand and T2 = second sit to stand).
2. METHODS
This is an ongoing study, “Blood Flow Response and Acute Interval Exercise (BRAIN), focused on the characterization of the cerebrovascular hemodynamic response to a single bout of high‐ intensity interval exercise and the sit‐to‐stand dCA response before and after exercise (NCT04673994). We developed the force sensor prior to study initiation and present limited data to establish “proof of principle,” not as interim analyses of the a priori stated primary outcome of the clinical trial. We recruited and enrolled healthy young adults, older adults, and individuals post‐stroke. Healthy young adults were included if they were between the ages of 18–30 years old and classified as low cardiovascular risk using the American College of Sports Medicine (ACSM) Cardiovascular Risk Screen (Thompson et al., 2013). Older adults and individuals post‐stroke (6 months–5 years post‐stroke) were included if they were: (1) Between the ages of 40–80 years old, (2) performed less than 150 min of moderate‐intensity exercise per week, and (3) able to answer consenting questions and follow a two‐step command. Individuals were excluded if they (1) were unable to stand from a sitting position without physical assistance, (2) had a previous history of another neurological disease (i.e., multiple sclerosis, Alzheimer's disease, or Parkinson's disease), and (3) did not have a MCAv signal on the right or left temporal window using transcranial Doppler ultrasound (TCD). The study was approved by the University of Kansas Medical Center Institutional Review Board Human Subjects Committee.
Prior to study procedures, all participants provided written informed consent to participate and we collected demographic information. Participants also completed a standardized five time sit‐to‐stand assessment, to determine leg strength and balance by measuring how quickly an individual could stand up from a chair and sit down five times without the use of their arms (Mong et al., 2010; Ng, 2010; Tiedemann et al., 2008).
2.1. Setup
Consistent with previously published protocols, our laboratory room was dimly lit and kept at a controlled temperature of 22°C–24°C (Billinger et al., 1985). Prior to the visit, participants were asked to abstain from caffeine for 8 h (Addicott et al., 2009; Institute of Medicine, 2001; Perod et al., 2000), vigorous exercise for 24 h (Burma et al., 2020), and alcohol for 24 h (Mathew & Wilson, 1986). Participants were seated on a recumbent stepper (T5XR NuStep, Inc. Ann Arbor, MI) with the chair rotated 90° perpendicular. The participant sat with feet flat on the ground, legs at 90°, and an upright trunk posture. The participants rested for 20 min while the equipment was donned: (1) A 5‐lead electrocardiogram (ECG; Cardiocard, Nasiff Associates, Central Square, New York) to measure heart rate, (2) a left middle finger plethysmograph (Finometer, Finapres Medical Systems, Amsterdam, the Netherlands) to measure beat‐to‐beat MAP, and (3) bilateral TCD probes (2‐MHz, Multigon Industries Inc, Yonkers, New York) to measure MCAv. The TCD probes were secured with headgear to maintain the optimal position and angle. Consistent with previously published methodology measuring beat‐to‐beat MAP during a sit‐to‐stand, participants were instructed to place their left hand with the finger plethysmograph flat on the upper chest at heart level (Labrecque et al., 2017; Lipsitz et al., 2000; Sorond et al., 2009). Due to neuromuscular fatigue, an arm sling was used to help hold the Finometer in place during the sit‐to‐stand recording (Lipsitz et al., 2000). For individuals with upper extremity spasticity, the Finometer was placed on the non‐affected upper extremity.
2.2. Force sensor
Our custom force sensor was comprised of a force‐sensitive resistor (SEN‐09376, Mouser Electronics Inc., Tx) connected to an operational amplifier (Canaduino LM393, Universal‐Solder, Ontario, Canada) and a custom circuit housed on a breadboard (Inland 400 tie‐point), shown in Figure 1. The force‐sensitive resistor was placed on the seat under the visualized right ischial tuberosity and connected to the custom circuit by six feet of stranded wire. For individuals with stroke, the force‐sensitive resistor was placed underneath the non‐affected lower extremity due to asymmetrical weight bearing during a sit‐to‐stand (Brunt et al., 2002). When sitting, the force‐sensitive resistor converted the weight of the participant into a voltage, amplified at the breadboard, and digitized via an analog‐to‐digital unit (NI‐USB‐6212, National Instruments). Once participants stood from the chair the force‐sensitive resistor no longer detected the participant's weight and the voltage signal quickly decreased to 0 volts for the remainder of the recording. Therefore, the force sensor was able to detect the exact moment the individual lifted off the seat to stand. The voltage of our force sensor was captured simultaneously with MCAv and MAP during the sit‐to‐stand using custom‐written software. All measures were recorded at 500 Hz using a custom ‐written software within MATLab implementing the Data Acquisition Toolbox (R2019a, TheMathworks Inc, Natick, Massachusetts).
FIGURE 1.

Drawing of the custom force sensor measuring AO during sit‐to‐stands. For the operational amplifier (Op Amp), the pins correspond to: 8‐Vcc, 3‐Input, 1‐Output, 4‐Ground. The force‐sensitive resistor has a resistance that decreases with force. As the subject stands, force decreases, the resistance increases, and the voltage drop increases at the force‐sensitive resistor. This results in a decreased Op Amp input voltage due to decreased voltage drop across the 47 Ohm resistor
2.3. Sit‐to‐stand
Participants were familiarized with the sit‐to‐stand protocol prior to the recording. After a minute of seated rest, participants were given a 3‐s countdown and asked to stand at 60 s into the recording. Participants remained standing for 2 min to allow hemodynamic stabilization (Labrecque et al., 2019, 2017). The sit‐to‐stand recording was performed twice separated by 15 min (T1, T2).
Raw data were processed offline using the QRS complex within the ECG to calculate beat‐to‐beat MCAv and MAP (Moir et al., 2020). AO was defined as the minimum of the second derivative of the sensor voltage, corresponding with the acceleration of the decreasing voltage velocity during the sit‐to‐stand.
2.4. Data analysis
Based on the published methodology, the TD before the onset of the dCA response was evaluated by two trained researchers as the seconds between performing the sit‐to‐stand and a continuous increase in CVC without transient reduction (CVC = MCAv/MAP) (Labrecque et al., 2019, 2017). Rather than estimating AO in the recording as 60 s, our force sensor marks the exact AO time. We calculated the difference in the TD response when using the force sensor compared to estimating AO at 60 s, or when the participant was asked to stand.
3. RESULTS
A total of eight individuals completed the sit‐to‐stand transition with the force sensor to measure the TD before the onset of dCA. We included healthy young adults (n = 4), older adults (n = 2), and individuals post‐stroke (n = 2). Participant characteristics are shown in Table 1. The sit‐to‐stand AO time for all participants is shown in Table 2.
TABLE 1.
Participant characteristics
| Subject | Age (years) | Sex | 5 time sit to stand (seconds) |
|---|---|---|---|
| Young adults | |||
| 1 | 23 | F | 5.05 |
| 2 | 24 | F | 7.42 |
| 3 | 25 | M | 6.42 |
| 4 | 27 | M | 5.39 |
| Older adults | |||
| 5 | 70 | M | 6.54 |
| 6 | 71 | M | 10.85 |
| Individuals post‐stroke | |||
| 7 | 45 | M | 10.07 |
| 8 | 71 | M | 26.35 |
TABLE 2.
Sit‐to‐stand outcomes
| Subject | T1 AO | T1 TD | TD Difference with Force Sensor | T2 AO | T2 TD | TD Difference with Force Sensor |
|---|---|---|---|---|---|---|
| Young adults | ||||||
| 1 | 60.03 | 1.75 | −0.03 | 59.81 | 1.26 | +0.19 |
| 2 | 60.58 | 1.11 | −0.58 | 60.26 | 2.13 | −0.26 |
| 3 | 60.36 | 1.46 | −0.36 | 60.21 | 6.70 | −0.21 |
| 4 | 60.14 | 2.23 | −0.14 | 59.73 | 1.54 | +0.27 |
| Older adults | ||||||
| 5 | 61.43 | 6.79 | −1.43 | 59.93 | 2.89 | +0.07 |
| 6 | 60.32 | 2.33 | −0.32 | 62.52 | 2.23 | −2.52 |
| Individuals post‐stroke | ||||||
| 7 | 59.91 | 0.79 | +0.09 | 60.00 | 1.94 | 0 |
| 8 | 59.83 | 1.93 | +0.17 | 59.60 | 3.16 | +0.40 |
Time in seconds.
Abbreviations: AO, arise‐and‐off; T1, First sit‐to‐stand; T2, Second sit‐to‐stand (separated by 15 min); TD, time delay before the onset of the dCA response.
3.1. Healthy young adults
At T1, the AO time for the healthy young adults was 60.28 ± 0.24 s. The force sensor was able to account for a shorter TD by 0.28 ± 0.24 s compared to estimating AO from the verbal command to stand. At T2, the AO time was on average at 60.00 ± 0.27 s in the recording. However, because two individuals had an AO time before being asked to stand, the force sensor accounted for a change in the TD response of 0.003 ± 0.27 s.
3.2. Older adults
At T1, participant #5’s AO time was 61.43 s. Compared to estimating AO from the verbal command to stand, the force sensor was able to account for a decrease in the TD response by 1.43 s (~17%). At T2, the same participant #5 had a 1.50 s faster AO time compared to T1 and had an AO time before being asked to stand. In contrast, while participant #6 had a small change in the TD at T1 of 0.32 s, they were 2.20 s slower at T2. The force sensor was, therefore able to account for a decrease in the TD response by 2.52 s (~53%) at T2, compared to estimating AO from the verbal command to stand (shown in Figure 2).
FIGURE 2.

Novel application of a force sensor during a sit‐to‐stand in a single healthy young adult (n = 1), older adult (n = 1), and individual post‐stroke (n = 1). CVC, Cerebrovascular Conductance; Horizontal Dotted Line, Individuals were asked to perform the sit‐to‐stand at 60 s; AO, Arise‐and‐off when participants stood from the chair; TD, Time delay before the onset of dynamic cerebral autoregulation. Force sensor values are in arbitrary units
3.3. Individuals post‐stroke
At T1, participant #7’s AO time was 0.09 s before the verbal command to stand, which accounted for an increase (~13%) in the TD response, compared to estimating AO from the verbal command to stand. At T2, the same participant #7’s AO time was at exactly 60 s. In contrast, participant #8’s AO was before the verbal command to stand, which accounted for an increase in the TD response at T1 (~10%) and T2 (~14%), respectively.
4. DISCUSSION
Previously published sit‐to‐stand methodology measuring the TD before the onset of the dCA response does not report the true moment of AO from the chair and therefore does not measure the exact start of the stimulus that drops peripheral blood pressure. While force sensors have been used in the kinetic characterization of a sit‐to‐stand (Galli et al., 2008) and tilt sensors have been used during dCA repeated squat‐stands (Barnes et al., 1985, 2017, 2018, 2021; Batterham et al., 2020; Klein et al., 2020; Panerai et al., 2021), we are the first to apply a force sensor during a hemodynamic recording to improve the accuracy of the TD before the onset of the dCA response. While the force sensor does not assess the duration or acceleration of the sit‐to‐stand, the purpose of the force sensor was to detect the exact moment of AO to precisely time the start of the physiological response and the TD before the onset of dCA. The force sensor allowed for the simultaneous measurement of the sit‐to‐stand reaction behavior with the physiologic response. We hypothesized that measuring AO precisely with a force sensor would identify previously unobserved variability in standing dynamics compared to estimating AO from a verbal command or other methods (Labrecque et al., 2019, 2017; Lipsitz et al., 2000; Sorond et al., 2009). We have shown that while healthy young adults may demonstrate a faster transition reaction speed during a sit‐to‐stand dCA recording (Beek et al., 2008; Sorond et al., 2009), older adults and individuals post‐stroke may differ in their AO response. We have shown that the TD before the onset of dCA, which typically occurs within 10 s (Labrecque et al., 2019, 2017; Serrador et al., 2005; Sorond et al., 2009), could potentially be skewed in some individuals by more than 2.5 seconds. Therefore, implementing a force sensor during a sit‐to‐stand allows investigators to objectively identify when the stimulus occurs and may improve the accuracy of the TD response.
Previous reviews have also reported delayed sit‐to‐stand reaction time in individuals with stroke (Boukadida et al., 2015; Lecours et al., 2008; Mentiplay et al., 2020), which was supported by our results (see Table 1) reporting the five time sit‐to‐stand scoring two standard deviations away from the age‐normative values (Bohannon, 2006). While this would suggest an individual post‐stroke would have a delay in AO, our results show that the individuals with stroke began to stand prematurely during the 3 second countdown, shown in Figure 2. A possible explanation for an early AO could be that the individuals post‐stroke anticipated longer sit‐to‐stand transition times (Mong et al., 2010).
We show that the healthy young adult participants performed AO at approximately the same time between T1 and T2. Although definitive conclusions cannot be drawn, two healthy young adults stood early at T2, which could have been due to anticipation and learning effect (Crook et al., 2017; Haile et al., 2021). One older adult participant could have also had an increase in performance learning and stood faster at T2 (Bohannon et al., 2010). However, the other older adult took ~2 s longer to perform AO at T2, which may have been due to lower extremity weakness (Sloot et al., 2020). The individuals post‐stroke may have recognized their lower extremity weakness and anticipated needing longer to stand.
Imprecise temporal measures of dCA onset could potentially impact our interpretation of data, especially when comparing healthy individuals to clinical populations. A previous study performed repeated sit‐to‐stands every 10 s in healthy older adults compared to individuals with dementia and mild cognitive impairment (Heus et al., 2018). Contrary to their hypothesis, they reported individuals with mild cognitive impairment and dementia had a better autoregulation response (Heus et al., 2018). However, the individuals with cognitive impairment could have performed the sit‐to‐stands early and, therefore, could have skewed the response. Without measuring the exact start of the sit‐to‐stand stimulus, the wrong conclusions could unintentionally be drawn on the temporal response. Another example is a study comparing the sit‐to‐stand response after a 14‐week exercise intervention in pregnant women that reported no significant change in the TD before the onset of dCA (Skow et al., 2021). However, there could have been a large difference in the sit‐to‐stand response between the baseline measures taken during the second trimester of pregnancy and measures taken after the exercise intervention during the third trimester of pregnancy. Pregnant women within the third trimester could have had a slower sit‐to‐stand time that was not accounted for and could have inadvertently skewed the calculation of TD. Imprecise timing of the response, supported by our results showing ~10%–53% change in TD, could systematically impact study findings.
5. CONCLUSIONS
The implementation of a force sensor to measure AO during a hemodynamic sit‐to‐stand recording could improve the exact calculation of the TD before the onset of the dCA response. Due to the nature of detailing innovative methodology, few participants were included within this report. However, future studies are needed to implement a force sensor measuring the temporal onset of the dCA response between younger adults, older adults, and individuals with stroke. Applying the force sensor into the sit‐to‐stand dCA methodology has the potential to improve scientific rigor and reproducibility, especially in older adults and clinical populations such as stroke.
CONFLICT OF INTEREST
The author(s) report no conflict of interest.
ETHICAL APPROVAL
Ethics approval for this study was granted by the Human Subjects Committee and Institutional Review Board at the University of Kansas Medical Center. We provided verbal and written explanation of the experimental protocol and the associated risks were provided to all participants prior to obtaining written informed consent.
AUTHOR CONTRIBUTIONS
AW, EV, SB conceived and designed research; AW and SA performed experiments; AW, EV, SA, SB analyzed data; AW, EV, SA, AR, SB interpreted results of experiments; AW, EV, SA, AR, SB drafted manuscript; AW, EV, SA, AR, SB edited and revised manuscript; AW, EV, SA, AR, SB approved final version of manuscript; AW prepared figures.
ACKNOWLEDGMENTS
We thank Preston Judd for his contributions to editing and proof reading.
Whitaker, A. A. , Vidoni, E. D. , Aaron, S. E. , Rouse, A. G. , & Billinger, S. A. (2022). Novel application of a force sensor during sit‐to‐stands to measure dynamic cerebral autoregulation onset. Physiological Reports, 10, e15244. 10.14814/phy2.15244
Funding information
AW was supported by the Eunice Kennedy Shriver National Institute of Child Health & Human Development of the National Institutes of Health (T32HD057850). AW was also supported by the American Heart Association Predoctoral Fellowship Grant (898190). SA was supported by a CTSA grant from NCATS awarded to the University of Kansas for Frontiers: University of Kansas Clinical and Translational Science Institute (TL1TR002368). SB and EV were supported in part by the National Institute on Aging for the KU Alzheimer's Disease Research Center (P30AG072973). The article processing charges related to the publication of this article were supported by The University of Kansas (KU) One University Open Access Author Fund sponsored jointly by the KU Provost, KU Vice Chancellor for Research & Graduate Studies, and KUMC Vice Chancellor for Research and managed jointly by the Libraries at the Medical Center and KU ‐ Lawrence. REDCap at The University of Kansas Medical Center is supported by Clinical and Translational Science Awards (CTSA) Award # ULTR000001 from National Center for Research Resources (NCRR). The Georgia Holland Research in Exercise and Cardiovascular Health (REACH) laboratory space was supported by the Georgia Holland Endowment Fund. The content is solely the responsibility of the authors and does not necessarily represent the official views of the National Institutes of Health.
REFERENCES
- Addicott, M. A. , Yang, L. L. , Peiffer, A. M. , Burnett, L. R. , Burdette, J. H. , Chen, M. Y. , Hayasaka, S. , Kraft, R. A. , Maldjian, J. A. , & Laurienti, P. J. (2009). The effect of daily caffeine use on cerebral blood flow: How much caffeine can we tolerate? Human Brain Mapping, 30, 3102–3114. 10.1002/hbm.20732 [DOI] [PMC free article] [PubMed] [Google Scholar]
- Barnes, S. C. , Ball, N. , Haunton, V. J. , Robinson, T. G. , & Panerai, R. B. (2017). The cerebrocardiovascular response to periodic squat‐stand maneuvers in healthy subjects: A time‐domain analysis. American Journal of Physiology. Heart and Circulatory Physiology, 313, H1240–H1248. [DOI] [PubMed] [Google Scholar]
- Barnes, S. C. , Ball, N. , Haunton, V. J. , Robinson, T. G. , & Panerai, R. B. (2018). How many squat‐stand manoeuvres to assess dynamic cerebral autoregulation? European Journal of Applied Physiology, 118, 2377–2384. 10.1007/s00421-018-3964-2 [DOI] [PMC free article] [PubMed] [Google Scholar]
- Barnes, S. C. , Ball, N. , Panerai, R. B. , Robinson, T. G. , & Haunton, V. J. (1985). Random squat/stand maneuvers: a novel approach for assessment of dynamic cerebral autoregulation? Journal of Applied Physiology, 123(3), 558–566. 10.1152/japplphysiol.00316.2017 [DOI] [PubMed] [Google Scholar]
- Barnes, S. C. , Haunton, V. J. , Beishon, L. , Llwyd, O. , Robinson, T. G. , & Panerai, R. B. (2021). Extremes of cerebral blood flow during hypercapnic squat‐stand maneuvers. Physiological Reports, 9, e15021. 10.14814/phy2.15021 [DOI] [PMC free article] [PubMed] [Google Scholar]
- Batterham, A. P. , Panerai, R. B. , Robinson, T. G. , & Haunton, V. J. (2020). Does depth of squat‐stand maneuver affect estimates of dynamic cerebral autoregulation? Physiological Reports, 8, e14549. 10.14814/phy2.14549 [DOI] [PMC free article] [PubMed] [Google Scholar]
- Billinger, S. A. , Craig, J. C. , Kwapiszeski, S. J. , Sisante, J. V. , Vidoni, E. D. , Maletsky, R. , & Poole, D. C. (1985). Dynamics of middle cerebral artery blood flow velocity during moderate‐intensity exercise. Journal of Applied Physiology, 122(5), 1125–1133. 10.1152/japplphysiol.00995.2016 [DOI] [PMC free article] [PubMed] [Google Scholar]
- Bohannon, R. W. (2006). Reference values for the five‐repetition sit‐to‐stand test: a descriptive meta‐analysis of data from elders. Perceptual and Motor Skills, 103, 215–222. [DOI] [PubMed] [Google Scholar]
- Bohannon, R. W. , Bubela, D. J. , Magasi, S. R. , Wang, Y. C. , & Gershon, R. C. (2010). Sit‐to‐stand test: Performance and determinants across the age‐span. Isokinetics and Exercise Science, 18, 235–240. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Boukadida, A. , Piotte, F. , Dehail, P. , & Nadeau, S. (2015). Determinants of sit‐to‐stand tasks in individuals with hemiparesis post stroke: A review. Annals of Physical and Rehabilitation Medicine, 58, 167–172. [DOI] [PubMed] [Google Scholar]
- Brassard, P. , Labrecque, L. , Smirl, J. D. , Tymko, M. M. , Caldwell, H. G. , Hoiland, R. L. , Lucas, S. J. E. , Denault, A. Y. , Couture, E. J. , & Ainslie, P. N. (2021). Losing the dogmatic view of cerebral autoregulation. Physiological Reports, 9, e14982. 10.14814/phy2.14982 [DOI] [PMC free article] [PubMed] [Google Scholar]
- Brunt, D. , Greenberg, B. , Wankadia, S. , Trimble, M. A. , & Shechtman, O. (2002). The effect of foot placement on sit to stand in healthy young subjects and patients with hemiplegia. Archives of Physical Medicine and Rehabilitation, 83, 924–929. 10.1053/apmr.2002.3324 [DOI] [PubMed] [Google Scholar]
- Burma, J. S. , Copeland, P. , Macaulay, A. , Khatra, O. , Wright, A. D. , & Smirl, J. D. (2020). Dynamic cerebral autoregulation across the cardiac cycle during 8 hr of recovery from acute exercise. Physiological Reports, 8, e14367. 10.14814/phy2.14367 [DOI] [PMC free article] [PubMed] [Google Scholar]
- Institute of Medicine (2001). Caffeine for the sustainment of mental task performance: Formulations for military operations. The National Academies Press. [PubMed] [Google Scholar]
- Crook, S. , Busching, G. , Schultz, K. , Lehbert, N. , Jelusic, D. , Keusch, S. , Wittmann, M. , Schuler, M. , Radtke, T. , Frey, M. , Turk, A. , Puhan, M. A. , & Frei, A. (2017). A multicentre validation of the 1‐min sit‐to‐stand test in patients with COPD. European Respiratory Journal, 49(3), 1601871. 10.1183/13993003.01871-2016 [DOI] [PubMed] [Google Scholar]
- de Heus, R. A. A. , de Jong, D. L. K. , Sanders, M. L. , van Spijker, G. J. , Oudegeest‐Sander, M. H. , Hopman, M. T. , Lawlor, B. A. , Olde Rikkert, M. G. M. , & Claassen, J. (2018). Dynamic regulation of cerebral blood flow in patients with alzheimer disease. Hypertension, 72, 139–150. 10.1161/HYPERTENSIONAHA.118.10900 [DOI] [PubMed] [Google Scholar]
- Deegan, B. M. , Sorond, F. A. , Lipsitz, L. A. , Olaighin, G. , & Serrador, J. M. (2009). Gender related differences in cerebral autoregulation in older healthy subjects. Annual International Conference of the IEEE Engineering in Medicine and Biology Society, 2009, 2859–2862. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Galli, M. , Cimolin, V. , Crivellini, M. , & Campanini, I. (2008). Quantitative analysis of sit to stand movement: Experimental set‐up definition and application to healthy and hemiplegic adults. Gait & Posture, 28, 80–85. [DOI] [PubMed] [Google Scholar]
- Haile, S. R. , Fuhner, T. , Granacher, U. , Stocker, J. , Radtke, T. , & Kriemler, S. (2021). Reference values and validation of the 1‐minute sit‐to‐stand test in healthy 5–16‐year‐old youth: A cross‐sectional study. British Medical Journal Open, 11, e049143. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Klein, T. , Bailey, T. G. , Wollseiffen, P. , Schneider, S. , & Askew, C. D. (2020). The effect of age on cerebral blood flow responses during repeated and sustained stand to sit transitions. Physiological Reports, 8, e14421. 10.14814/phy2.14421 [DOI] [PMC free article] [PubMed] [Google Scholar]
- Labrecque, L. , Rahimaly, K. , Imhoff, S. , Paquette, M. , Le Blanc, O. , Malenfant, S. , Drapeau, A. , Smirl, J. D. , Bailey, D. M. , & Brassard, P. (2019). Dynamic cerebral autoregulation is attenuated in young fit women. Physiological Reports, 7, e13984. 10.14814/phy2.13984 [DOI] [PMC free article] [PubMed] [Google Scholar]
- Labrecque, L. , Rahimaly, K. , Imhoff, S. , Paquette, M. , Le Blanc, O. , Malenfant, S. , Lucas, S. J. E. , Bailey, D. M. , Smirl, J. D. , & Brassard, P. (2017). Diminished dynamic cerebral autoregulatory capacity with forced oscillations in mean arterial pressure with elevated cardiorespiratory fitness. Physiological Reports, 5, e13486. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Lecours, J. , Nadeau, S. , Gravel, D. , & Teixera‐Salmela, L. (2008). Interactions between foot placement, trunk frontal position, weight‐bearing and knee moment asymmetry at seat‐off during rising from a chair in healthy controls and persons with hemiparesis. Journal of Rehabilitation Medicine, 40, 200–207. 10.2340/16501977-0155 [DOI] [PubMed] [Google Scholar]
- Lind‐Holst, M. , Cotter, J. D. , Helge, J. W. , Boushel, R. , Augustesen, H. , Van Lieshout, J. J. , & Pott, F. C. (1985). Cerebral autoregulation dynamics in endurance‐trained individuals. Journal of Applied Physiology, 110(5), 1327–1333. 10.1152/japplphysiol.01497.2010 [DOI] [PubMed] [Google Scholar]
- Lipsitz, L. A. , Mukai, S. , Hamner, J. , Gagnon, M. , & Babikian, V. (2000). Dynamic regulation of middle cerebral artery blood flow velocity in aging and hypertension. Stroke, 31, 1897–1903. 10.1161/01.STR.31.8.1897 [DOI] [PubMed] [Google Scholar]
- Mathew, R. J. , & Wilson, W. H. (1986). Regional cerebral blood flow changes associated with ethanol intoxication. Stroke, 17, 1156–1159. 10.1161/01.STR.17.6.1156 [DOI] [PubMed] [Google Scholar]
- Mentiplay, B. F. , Clark, R. A. , Bower, K. J. , Williams, G. , & Pua, Y. H. (2020). Five times sit‐to‐stand following stroke: Relationship with strength and balance. Gait & Posture, 78, 35–39. [DOI] [PubMed] [Google Scholar]
- Moir, M. E. , Klassen, S. A. , Zamir, M. , & Shoemaker, J. K. (2020). Rapid changes in vascular compliance contribute to cerebrovascular adjustments during transient reductions in blood pressure in young, healthy adults. Journal of Applied Physiology, 129(1), 27–35. 10.1152/japplphysiol.00272.2020 [DOI] [PMC free article] [PubMed] [Google Scholar]
- Mong, Y. , Teo, T. W. , & Ng, S. S. (2010). 5‐repetition sit‐to‐stand test in subjects with chronic stroke: Reliability and validity. Archives of Physical Medicine and Rehabilitation, 91, 407–413. [DOI] [PubMed] [Google Scholar]
- Ng, S. (2010). Balance ability, not muscle strength and exercise endurance, determines the performance of hemiparetic subjects on the timed‐sit‐to‐stand test. American Journal of Physical Medicine & Rehabilitation, 89, 497–504. 10.1097/PHM.0b013e3181d3e90a [DOI] [PubMed] [Google Scholar]
- Olufsen, M. S. , Ottesen, J. T. , Tran, H. T. , Ellwein, L. M. , Lipsitz, L. A. , & Novak, V. (2005). Blood pressure and blood flow variation during postural change from sitting to standing: Model development and validation. Journal of Applied Physiology, 99, 1523–1537. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Panerai, R. B. (1998). Assessment of cerebral pressure autoregulation in humans–a review of measurement methods. Physiological Measurement, 19, 305–338. 10.1088/0967-3334/19/3/001 [DOI] [PubMed] [Google Scholar]
- Panerai, R. B. , Batterham, A. , Robinson, T. G. , & Haunton, V. J. (2021). Determinants of cerebral blood flow velocity change during squat‐stand maneuvers. American Journal of Physiology: Regulatory, Integrative and Comparative Physiology, 320, R452–R466. 10.1152/ajpregu.00291.2020 [DOI] [PubMed] [Google Scholar]
- Perod, A. L. , Roberts, A. E. , & McKinney, W. M. (2000). Caffeine can affect velocity in the middle cerebral artery during hyperventilation, hypoventilation, and thinking: A transcranial Doppler study. Journal of Neuroimaging, 10, 33–38. [DOI] [PubMed] [Google Scholar]
- Roy, C. S. , & Sherrington, C. S. (1890). On the regulation of the blood‐supply of the brain. Journal of Physiology, 11, 85–158 117. 10.1113/jphysiol.1890.sp000321 [DOI] [PMC free article] [PubMed] [Google Scholar]
- Serrador, J. M. , Sorond, F. A. , Vyas, M. , Gagnon, M. , Iloputaife, I. D. , & Lipsitz, L. A. (2005). Cerebral pressure‐flow relations in hypertensive elderly humans: Transfer gain in different frequency domains. Journal of Applied Physiology, 98, 151–159. [DOI] [PubMed] [Google Scholar]
- Skow, R. J. , Labrecque, L. , Rosenberger, J. A. , Brassard, P. , Steinback, C. D. , & Davenport, M. H. (2021). Prenatal exercise and cardiovascular health (PEACH) study: Impact of acute and chronic exercise on cerebrovascular hemodynamics and dynamic cerebral autoregulation. Journal of Applied Physiology, 132(1), 247–260. 10.1152/japplphysiol.00446.2021 [DOI] [PubMed] [Google Scholar]
- Sloot, L. H. , Millard, M. , Werner, C. , & Mombaur, K. (2020). Slow but steady: Similar sit‐to‐stand balance at seat‐off in older vs. younger adults. Frontiers in Sports and Active Living, 2, 548174. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Sorond, F. A. , Serrador, J. M. , Jones, R. N. , Shaffer, M. L. , & Lipsitz, L. A. (2009). The sit‐to‐stand technique for the measurement of dynamic cerebral autoregulation. Ultrasound in Medicine and Biology, 35, 21–29. 10.1016/j.ultrasmedbio.2008.08.001 [DOI] [PMC free article] [PubMed] [Google Scholar]
- Stevermer, C. A. , & Gillette, J. C. (2016). Kinematic and kinetic indicators of sit‐to‐stand. Journal of Applied Biomechanics, 32, 7–15. 10.1123/jab.2014-0189 [DOI] [PubMed] [Google Scholar]
- Thompson, P. D. , Arena, R. , Riebe, D. , & Pescatello, L. S. (2013). ACSM’s new preparticipation health screening recommendations from ACSM’s guidelines for exercise testing and prescription, ninth edition. Current Sports Medicine Reports, 12(4), 215–217. 10.1249/JSR.0b013e31829a68cf. [DOI] [PubMed] [Google Scholar]
- Tiedemann, A. , Shimada, H. , Sherrington, C. , Murray, S. , & Lord, S. (2008). The comparative ability of eight functional mobility tests for predicting falls in community‐dwelling older people. Age and Ageing, 37, 430–435. 10.1093/ageing/afn100 [DOI] [PubMed] [Google Scholar]
- Tzeng, Y. C. , Willie, C. K. , Atkinson, G. , Lucas, S. J. , Wong, A. , & Ainslie, P. N. (2010). Cerebrovascular regulation during transient hypotension and hypertension in humans. Hypertension, 56, 268–273. 10.1161/HYPERTENSIONAHA.110.152066 [DOI] [PubMed] [Google Scholar]
- van Beek, A. H. , Claassen, J. A. , Rikkert, M. G. , & Jansen, R. W. (2008). Cerebral autoregulation: An overview of current concepts and methodology with special focus on the elderly. Journal of Cerebral Blood Flow and Metabolism, 28, 1071–1085. [DOI] [PubMed] [Google Scholar]
- Willie, C. K. , Tzeng, Y. C. , Fisher, J. A. , & Ainslie, P. N. (2014). Integrative regulation of human brain blood flow. Journal of Physiology, 592, 841–859. 10.1113/jphysiol.2013.268953 [DOI] [PMC free article] [PubMed] [Google Scholar]