Abstract
Background
Wearable devices have been proposed as a novel method for monitoring patients after surgery to track recovery, identify complications early, and improve surgical safety. Previous studies have used a heterogeneous range of devices, methods, and analyses. This review aimed to examine current methods and wearable devices used for monitoring after abdominal surgery and identify knowledge gaps requiring further investigation.
Methods
A scoping review was conducted given the heterogeneous nature of the evidence. MEDLINE, EMBASE, and Scopus databases were systematically searched. Studies of wearable devices for monitoring of adult patients within 30 days after abdominal surgery were eligible for inclusion.
Results
A total of 78 articles from 65 study cohorts, with 5153 patients were included. Thirty-one different wearable devices were used to measure vital signs, physiological measurements, or physical activity. The duration of postoperative wearable device use ranged from 15 h to 3 months after surgery. Studies mostly focused on physical activity metrics (71.8 per cent). Continuous vital sign measurement and physical activity tracking both showed promise for detecting postoperative complications earlier than usual care, but conclusions were limited by poor device precision, adherence, occurrence of false alarms, data transmission problems, and retrospective data analysis. Devices were generally well accepted by patients, with high levels of acceptance, comfort, and safety.
Conclusion
Wearable technology has not yet realized its potential to improve postoperative monitoring. Further work is needed to overcome technical limitations, improve precision, and reduce false alarms. Prospective assessment of efficacy, using an intention-to-treat approach should be the focus of further studies.
Wearable devices have been proposed as a novel method for monitoring patients after surgery to track recovery, identify complications early, and improve surgical safety. A heterogeneous range of devices and methods have been used to monitor patients after abdominal surgery. Further work is needed to overcome technical limitations, improve precision, and reduce false alarms.
Introduction
Recovery after abdominal surgery is a high-risk interval, with up to one-third of patients suffering a major postoperative complication within 30 days of surgery1. Delayed recognition of complications and subsequent delays in the escalation of care may lead to further avoidable harm2,3. Many studies have identified the ‘failure to rescue’ patients from complications as a major contributor to perioperative mortality3–5, highlighting the importance of close postoperative monitoring. Even in patients who do not develop major complications, recovery can be challenging. Modern evidence-based enhanced recovery protocols have been shown to improve recovery, reduce complications, and postoperative duration of hospital of stay, but deviations from these protocols are common and are associated with poorer outcomes6,7.
Protocol-driven measurements of vital signs by nursing staff are the most common strategy used for postoperative monitoring on surgical wards, often in conjunction with an ‘early-warning score’ system for escalation8,9. However, these traditional recordings rely on intermittent and simplistic measurements of physiological function and may not identify early or intermittent signs of patient deterioration10,11.
Wearable devices such as ‘smart watches’ or ‘smart patches’ have been proposed as a novel method of monitoring patients after surgery to improve safety12–14. The activity metrics (such as step count and sleep) and physiological data (such as heart rate and respiratory rate) measured by these devices could be used to continuously monitor patients and track their recovery trajectory. Wearable devices have the potential to predict or detect the occurrence of postoperative complications and may also engage patients as active participants in the recovery process15; however, existing studies have used a diverse range of consumer- and research-grade wearable sensor devices and have employed heterogeneous methods of data collection and analysis.
This scoping review aimed to summarize the published literature investigating the use of wearable devices for patients during recovery after abdominal surgery, examine current methods and devices, and identify knowledge gaps requiring further investigation.
Methods
Trial design
This was a scoping systematic review conducted according to the Joanna Briggs Institute guidance for scoping reviews16. This review was reported according to the PRISMA 2020 guidelines and the extension for scoping systematic reviews17,18 (Appendix S1). Scoping reviews are not eligible for prospective registration on the PROSPERO database.
Search strategy
The MEDLINE (Ovid), Embase, and Scopus databases were searched from inception to 15 November 2021. Search terms related to surgery were combined with terms related to wearable devices and monitoring using Boolean ‘AND’ operators. The search strings used are included in Appendix S2.
Study selection
Inclusion criteria
Studies investigating wearable sensor devices in adult patients within 30 days of intra-abdominal surgery (including gastrointestinal, hepatobiliary, urological, gynaecological, and vascular surgery). Wearable sensors were defined as a device worn on the external body surface, unencumbered by wires, for the continuous and non-invasive detection of biosignals (such as movement, heart rate, respiratory rate, and oxygen saturation). There were no limits on sensor type, the specific metrics recorded, or the location of recordings (such as in a hospital or outpatient setting). Studies that used wearable sensors only in theatre or post-anaesthetic recovery units were excluded, as were studies that investigated the use of wearable devices with direct therapeutic intentions (such as electrostimulation).
There were no limits on indication for surgery or surgical approach. Studies conducted in obstetric cohorts (following Caesarean section) were excluded. Both observational and randomized studies were eligible for inclusion.
Studies reporting the use of wearable sensors in pre- or intraoperative settings were only included if they also reported postoperative use of wearable sensor devices.
Exclusion criteria
The exclusion criteria were:
Paediatric patients (aged under 18 years), or if most of the included patients did not undergo abdominal surgery.
Case reports, small case series (n < 10 patients), conference abstracts, and studies published in languages other than English.
Study protocols without publication of results.
Review articles, but the reference lists of all included studies and relevant review articles were manually screened to identify additional eligible papers for inclusion.
Data extraction and analysis
Records from the database search were exported and deduplicated in EndNote X9 (Clarivate, Philadelphia, Pennsylvania, USA) using the methods of Bramer et al.19. Two independent reviewers then used the Rayyan web application to screen the titles and abstracts for full-text review20. Discrepancies were settled by discussion between the reviewers as required. Authors were contacted by e-mail to clarify when it was unclear whether the study should be included.
Two investigators assessed and extracted relevant data from included full-text articles. Data and narrative summaries were extracted for each study with a pro forma developed specifically for the purposes of this review (Appendix S3).
It was expected that included papers would be too heterogeneous in their methods and inclusion criteria to perform a meaningful quantitative analysis. Therefore, this scoping review did not undertake any statistical analysis other than simple descriptive statistics used to report percentages or averages. A descriptive review of the included articles is presented.
Results
A total of 7138 records were screened, and 78 articles representing 65 study cohorts with a total of 5153 patients were included (Fig. 1 and Appendix S4). Studies were predominantly conducted in Europe (n = 33, 42.3 per cent) and North America (n = 28, 35.9 per cent), with a minority from Asia (n = 14, 17.9 per cent) and Australia/New Zealand (n = 3, 3.8 per cent). More than half the included studies were published in 2020 or 2021 (Fig. 2a).
Fig. 1.
PRISMA diagram
Fig. 2.
a Number of included articles published per year. b Percentage of studies investigating various metrics using wearable sensors
RR, respiratory rate; HR, heart rate; Temp, temperature; SpO2, oxygen saturation.
Most included articles were prospective observational studies (n = 54, 69.2 per cent); only 20 randomized studies were identified (25.6 per cent) (Table 1). Only six articles (7.7 per cent) were from multicentre studies; the majority were single-centre investigations. Studies mostly recruited mixed cohorts of patients undergoing abdominal surgery, and were predominantly conducted in elective patients, with only four studies including acute presentations.
Table 1.
Summary characteristics of included studies and patient cohorts
| Studies, n = 78 | Cohorts, n = 65 | |
|---|---|---|
| Specialty | ||
| Mixed cohort | 39 (50.0%) | 29 (44.6%) |
| Colorectal | 13 (16.7%) | 11 (16.9%) |
| Oesophagogastric | 8 (10.3%) | 7 (10.8%) |
| Hepatopancreaticobiliary | 7 (9.0%) | 7 (10.8%) |
| Gynaecology | 6 (7.7%) | 6 (9.2%) |
| Bariatric | 3 (3.8%) | 3 (4.6%) |
| Urology | 2 (2.6%) | 2 (3.1%) |
| Study design | ||
| Randomized trial | 20 (25.6%) | 18 (27.7%) |
| Non-randomized trial | 4 (5.1%) | 4 (6.2%) |
| Observational cohort | 52 (69.2%) | 42 (66.2%) |
| Number of centres | ||
| Single | 72 (92.3%) | 61 (93.8%) |
| Multiple | 6 (7.7%) | 4 (6.2%) |
| Surgical urgency | ||
| Elective only | 65 (83.3%) | 53 (81.5%) |
| Acute and elective | 4 (5.1%) | 3 (4.6%) |
| Not stated | 9 (11.5%) | 9 (13.8%) |
| Funding * | ||
| Academic | 54 (69.2%) | 43 (66.2%) |
| Philanthropic | 4 (5.1%) | 3 (4.6%) |
| Industry | 9 (11.5%) | 8 (12.3%) |
| Unfunded | 5 (6.4%) | 4 (6.2%) |
| Not stated | 14 (17.9%) | 13 (20.0%) |
| Conflicts of interest * | ||
| None | 42 (53.8%) | 34 (52.3%) |
| Wearable related | 11 (14.1%) | 10 (15.4%) |
| Not wearable related | 14 (17.9%) | 10 (15.4%) |
| Not stated | 13 (16.7%) | 12 (18.5%) |
Does not add up to 100% due to articles with multiple funding sources or conflicts of interest.
Wearable devices used
A total of 31 different wearable devices were used by the included studies to measure vital signs or other physiological measurements (Table 2), or physical activity metrics (Table 3). Most devices were commercial- or research-grade (n = 22, 71.0 per cent); only a minority were medical-grade wearable sensors with US Food and Drug Administration or CE mark approvals (n = 9, 29.0 per cent). Several studies used wearable sensors as one component of a larger mHealth or eHealth programme for postoperative monitoring21–28.
Table 2.
Wearable devices tracking vital signs and other physiology
| Name | Company | Number of studies | Location | Grade | US Food and Drug Administration | CE mark | Metrics |
|---|---|---|---|---|---|---|---|
| HealthPatch MD/Vital Patch | VitalConnect (California, USA) | 8 | Chest | Clinical | Yes | Yes | ECG, heart rate, heart rate variability, respiratory rate, skin temperature, accelerometery |
| SensiumVitals Patch | Sensium (UK) | 7 | Chest | Clinical | Yes | Yes | Heart rate, respiratory rate, axillary temperature |
| ViSi Mobile | Sotera Wireless (California, USA) | 6 | Wrist and chest | Clinical | Yes | Yes | Continuous non-invasive blood pressure, oxygen saturation, heart rate, pulse rate, respiratory rate, skin temperature, ECG, posture, fall detection |
| G-Tech Patch | G-Tech Medical (California, USA) | 2 | Abdomen | Research | No | No | Cutaneous electrical signals from the gastrointestinal tract |
| Orient Speck | Centre for Speckled Computing, University of Edinbugh (UK) | 2 | Chest/ Abdomen | Research | No | No | Respiratory rate |
| Aingeal | Renew Health (Ireland)* | 1 | Chest | Clinical | Yes | Yes | Respiratory rate, ECG, skin temperature, accelerometery |
| HealthDot | Phillips (The Netherlands) | 1 | Chest | Clinical | No | Yes | Heart rate, respiratory rate, body posture, activity |
| iThermonitor | Raiing Medical Company (China) | 1 | Axilla | Clinical | Yes | Yes | Axillary temperature |
| Radius- 7 | Masimo (California, USA) | 1 | Arm | Clinical | Yes | Yes | Oxygen saturation, pulse rate, perfusion index, pleth variability index, total haemoglobin, methaemoglobin, carboxyhaemoglobin, oxygen content, oxygen reserve index, acoustic respiration rate |
Formerly called Intelesens.
Table 3.
Wearable devices tracking activity metrics
| Name | Company | Number of studies | Location | Grade | US Food and Drug Administration | CE mark |
|---|---|---|---|---|---|---|
| Fitbit (various models *) | Fitbit (California, USA) | 18 | Wrist | Consumer | No | No |
| Vivofit (various models†) | Garmin (Switzerland) | 8 | Wrist | Consumer | No | No |
| UP MOVE | Jawbone (California, USA)† | 4 | Wrist | Consumer | No | No |
| ActiGraph (GT3X+ or GT9X) | Actigraph (Florida, USA) | 4 | Hip/Waist | Clinical | Yes | Yes |
| Active tracer AC-301 | GMS Co. (Tokyo, Japan) | 3 | Ankle | Research | No | No |
| E-care Fit | NEWEL (France) | 2 | Wrist | Research | No | No |
| Lifecorder | Suzuken Co. (Japan) | 2 | Waist | Research | No | No |
| acitvPAL3 micro | PAL Technologies (UK) | 1 | Thigh | Research | No | No |
| Active style Pro HJA-750C | Omron Healthcare (Japan) | 1 | Hip/Waist | Consumer | No | No |
| Actiwatch 64 | Mini Mitter/Respironics, (Oregon, USA) | 1 | Wrist | Clinical/Research | No | Yes |
| Apple Watch | Apple (California, USA) | 1 | Wrist | Consumer | No | No |
| Lifegram LA11M-BS | LG Electronics (South Korea) | 1 | Wrist | Consumer | No | No |
| Mini-Motion Logger Actigraph | Ambulatory Monitoring (New York, USA) | 1 | Wrist | Research | No | No |
| MTN/220 accelerometer | ACOS Co. (Japan) | 1 | Hip/Waist | Research | No | No |
| New Lifestyles NL-2000i | New Lifestyles (Michigan, USA) | 1 | Hip/Waist | Consumer | No | No |
| OMRON Walking Style Pro 2.0 | OMRON Medizintechnik (Germany) | 1 | Hip/Waist | Consumer | No | No |
| PAM AM101 accelerometer | PAM (The Netherlands) | 1 | Hip/Waist | Consumer | No | No |
| Polar Loop Activity Tracker | Polar Electro Oy (Finland) | 1 | Wrist | Consumer | No | No |
| Portable Sleep Monitor (PSM100A) | Chengdu Sealand Technology Co. (China) | 1 | Chest | Research | No | No |
| Positional Activity Logger | Gorman ProMed (Victoria, Australia)‡ | 1 | Thigh | Research | No | No |
| Samsung Gear | Samsung Group (South Korea) | 1 | Wrist | Consumer | No | No |
| Tractivity ankle pedometer | Kineteks Corporation (Canada)‡ | 1 | Ankle | Consumer/Research | No | No |
Alta HR, Inspire HR, Zip, Charge, Charge 2, Flex, Versa. †Vivofit, Vivofit 2, Vivofit 3.‡No longer active.
Most studies (n = 56, 71.8 per cent) reported on postoperative physical activity, predominantly measured as daily step counts (Fig. 2b). Respiratory rates and heart rates were the most measured vital signs and were reported by 19 (24.4 per cent) and 17 (21.8 per cent) studies respectively. Accelerometry data from these sensors were obtained from a range of body locations, including wrist, waist/hip, thigh, and ankle (Tables 2 and 3).
The duration of postoperative recordings was variable and ranged from 15 h to 3 months after surgery. Forty-five studies (57.7 per cent) used wearable sensors only during hospitalization, 6 studies (7.7 per cent) only at home, and 27 studies (34.6 per cent) had both hospital- and home-based recordings. In 36 studies (46.2 per cent), preoperative recordings were also used.
Patient recruitment and exclusion were often described poorly; 29 studies (37.2 per cent) reported the total number of patients screened, and 40 (51.3 per cent) reported the number of eligible patients, the number approached, and the number of eligible patients who declined to participate. Of the study cohorts who reported sufficient data, the mean rate of eligible patients who declined to participate was 30.2 ± 22.9 per cent (range 0–81.5 per cent). Patient characteristics were variably described; age and sex in 77 studies (98.7 per cent) each, ethnicity or race in 22 (28.2 per cent), BMI in 53 (67.9 per cent), and ASA score or other co-morbidity measures in 47 (60.3 per cent).
The reporting of adherence with the wearable device, rates of missing data, device failure, and strategies used to account for missing data were also poor. Adherence with the wearable device was only reported by 32 papers (41.0 per cent). Of those that did report adherence, this ranged from 49.2–100 per cent, though was defined variably between studies and could not be pooled. Two studies showed higher compliance with wrist-worn sensors during home-based recordings compared with in-hospital26,29, however, others showed no difference30. Several studies reported higher compliance with wearable use compared with other elements of an mHealth programme such as symptom reporting27,31. The rate of missing or unusable data or device failure or loss was only reported by 39 studies (50.0 per cent) and ranged from 3.0–51.4 per cent. Poor signal quality, data transmission, and connection problems were all reported to contribute to missing data by several studies24,32–34.
Physiological monitoring
Most studies on wearables to monitor postoperative physiology utilized in-hospital continuous vital sign monitoring. Pilot and feasibility trials implementing continuous vital sign monitoring systems suggested that this resulted in a shorter duration of hospital stay and fewer unplanned ICU admissions35,36. A continuous temperature monitoring device (iThermonitor) showed feasibility to identify fevers 4 h earlier and with a higher peak temperature than routine nurse measurements37. One small study implemented outpatient continuous vital sign monitoring following oesophagectomy; this showed no changes in clinical management but established the feasibility of home-based monitoring38.
Inspection of vital sign recordings from wearable sensors showed that they can detect selected postoperative complications, particularly postoperative atrial fibrillation39. Abnormal respiratory patterns and cyclical airway obstruction were common in patients receiving postoperative opioid analgesia40. Other studies showed episodes of hypotension and hypoxia are common in postoperative patients and often unrecognized by routine measurements41–43. Two studies recorded gastrointestinal electrical activity from the abdominal surface and reported that this had the potential to predict postoperative ileus and diet readiness44,45.
Several clinical validation studies compared the accuracy of wearable sensors with routine nursing measurements33,34,46,47, bedside monitors48–50, or other sensors51. Across the range of devices investigated, accuracy was generally acceptable with small errors in mean difference. However, the precision was poor with broad limits of agreement often outside clinically acceptable differences (Table 4). Several studies noted a ‘digit bias’ in nursing measurements of respiratory rate46,47,52. This implausible prevalence of respiratory rates of 16, 18, and 20 has been previously reported53, suggesting that routine nursing measurements may not be an appropriate gold-standard comparison for device validation.
Table 4.
Accuracy and precision of wearable devices reported in clinical validation studies
| Device | Study | Reference standard | Patients | Pairs of measurements | Mean difference (reference – device) | 95% Limits of agreement |
|---|---|---|---|---|---|---|
| Heart rate (b.p.m.) | ||||||
| HealthPatch | Breteler 201849 | Bedside monitor | 25 | 3986 | −1.2 | −5.7 to 3.2 |
| Breteler 202048 | Bedside monitor | 25 | 29 619 | 1.3 | −4.1 to 6.9 | |
| Weenk 201733 | Nursing measurements | 10 | 86 | −1.52 | −9.51 to 12.55 | |
| Weenk 201956 | Nursing measurements | 30 | NS | −1.00 | −11.11 to 13.11 | |
| SensiumVitals | Breteler 202048 | Bedside monitor | 25 | 16 917 | 1.0 | −14.6 to 16.7 |
| Downey 201946 | Nursing measurements | 51 | 1135 | −1.85 | −23.92 to 20.22 | |
| ViSi Mobile | Weenk 201733 | Nursing measurements | 10 | 86 | −0.20 | −11.06 to 10.66 |
| Weenk 201956 | Nursing measurements | 30 | NS | 0.69 | −17.48 to 18.86 | |
| Masimo | Breteler 202048 | Bedside monitor | 25 | 34 992 | −0.4 | −11.9 to 11.0 |
| Aingeal | Cheng 202134 | Nursing measurements | 35 | NS | 1.12 | −24.03 to 26.27 |
| Healthdot | Van der Stam 202150 | Bedside monitor | 25 | 237 928 | −0.23 | −7.43 to 6.97 |
| Respiratory rate (/min) | ||||||
| HealthPatch | Breteler 201849 | Bedside monitor | 25 | 4001 | −2.4 | −10.8 to 5.9 |
| Breteler 202048 | Bedside monitor | 25 | 29 135 | 4.4 | −5.8 to 14.7 | |
| Weenk 201733 | Nursing measurements | 10 | 86 | −0.64 | −10.32 to 9.04 | |
| Weenk 201956 | Nursing measurements | 30 | NS | −1.94 | −8.92 to 5.04 | |
| SensiumVitals | Breteler 202048 | Bedside monitor | 25 | 17 595 | −0.8 | −8.5 to 6.9 |
| Downey 201946 | Nursing Measurements | 51 | 1134 | 2.93 | −8.19 to 14.05 | |
| ViSi Mobile | Weenk 201733 | Nursing measurements | 10 | 86 | 1.19 | −5.53 to 7.91 |
| Weenk 201956 | Nursing measurements | 30 | NS | 0.84 | −5.88 to 7.56 | |
| Masimo | Breteler 202048 | Bedside monitor | 25 | 33 032 | 0.2 | −6.6 to 6.3 |
| Aingeal | Cheng 202134 | Nursing measurements | 35 | NS | 1.04 | −6.88 to 8.96 |
| Healthdot | Van der Stam 202150 | Bedside monitor | 21 | 263 742 | 0.28 | −5.19 to 5.74 |
| Temperature (°C) | ||||||
| HealthPatch | Weenk 201956 | Nursing measurements | 30 | NS | 2.76 | 1.02 to 4.50 |
| ViSi Mobile | Weenk 201956 | Nursing measurements | 30 | NS | 2.96 | 0.75 to 5.17 |
| Aingeal | Cheng 202134 | Nursing measurements | 35 | NS | −1.45 | −5.67 to 2.76 |
| SensiumVitals | Downey 201946 | Nursing measurements | 51 | 1132 | 0.82 | −1.13 to 2.78 |
| iThermonitor | Liu 202037 | Nursing measurements | 526 | 3621 | −0.03 | −0.73 to 0.63 |
| Oxygen saturation (%) | ||||||
| ViSi Mobile | Weenk 201956 | Nursing measurements | 30 | NS | 0.94 | −4.25 to 6.13 |
| Systolic blood pressure (mmHg) | ||||||
| ViSi Mobile | Weenk 201956 | Nursing measurements | 30 | NS | 5.42 | −22.5 to 33.4 |
NS, not stated.
Patient-reported evaluations of continuous vital sign monitoring generally reported high levels of acceptance, comfort, and safety34–38,54,55. However, in a randomized trial on the SensiumVitals patch, 24 per cent of patients chose to discontinue monitoring early35, usually due to adverse skin effects. Several studies reported a patient preference to go home with wearable monitoring55, potentially facilitating earlier discharge from the hospital38. Patients emphasized the importance of not losing opportunities for human contact with clinical staff, and concerns about devices not capturing other important aspects of the patient experience such as pain33,54,56,57.
Nurses and other clinicians often recognized the potential of wearable devices for continuous vital signs monitoring, but also expressed concerns regarding the number of false-positive alerts (when the wearable device triggered an alert for abnormal vital signs that were normal on manual review of the patient), increasing workload, and overload of data33,55,57.
Physical activity
Many papers investigated changes in physical activity perioperatively, showing reduced step counts after surgery, with a long return to baseline that may take weeks to months21,25–27,29–31,58–68. The recovery trajectory in physical activity differed depending on type of operation64, use of laparoscopy69–71, need for ICU admission64, as well as overall performance status72. Romain et al. showed that postoperative step counts were correlated with preoperative steps61. Kovar et al. showed step counts on postoperative day three could predict activity levels at 1 month after surgery68.
Multiple studies showed that greater postoperative physical activity was correlated with shorter length of stay28,58,73–75, and a reduced risk of readmission following discharge15,73,76,77. Higher postoperative step counts were associated with a lower risk of complications30,72–74,78–81, faster gastrointestinal recovery73, and lower long-term skeletal muscle loss82. Patients with postoperative delirium had similar mobilization in the early postoperative interval and had lower physical activity at 1 month after surgery83. Of these studies aiming to predict postoperative outcomes with postoperative physical activity measurements, only four accounted for patients’ baseline preoperative physical activity levels76,80,81,83. No studies analysed physical activity data in ‘real time’ to monitor recovery or identify complications.
Several studies investigated trends in activity to predict outcomes. Iida et al. investigated the impact of different recovery trajectories following hepatectomy, classifying patients into ‘steady increase’, ‘bell curve’, and ‘flat’ categories78,79. Patients with a ‘steady increase’ had a low risk of complications. Wound infections, pleural effusions, and ascites were more common in the bell curve group, and postoperative pneumonia was observed only in patients with a flat recovery profile79. Robinson et al. showed that a decrease in step count of more than 50 per cent over 2 days consecutively after surgery had a 79 per cent sensitivity and 90 per cent specificity for hospital readmission.
Several interventions to increase postoperative physical activity were studied, often with limited success84–91. One randomized trial investigating a targeted step count intervention showed no difference in duration of hospital stay and increased fatigue scores in patients with a wearable fitness tracker92. In comparison, a non-randomized trial showed a lower risk of pneumonia and shorter duration of hospital stay in a self-selected group of patients93. Feedback from wearable devices had mixed effects; some studies reported increased activity74,94, whereas others reported no effect63,95.
Five studies reported on postoperative sleep metrics derived from accelerometer data in combination with or separate from other physiological information (such as electrocardiogram signals)59,96–99. Sleep was generally poor during hospital stays96,97, predominantly driven by night-time awakenings, correlated with patient-reported symptoms59,98, and better sleep quality was associated with shorter duration of hospital stay99.
Discussion
Wearable technology has the potential to revolutionize postoperative monitoring and recovery after abdominal surgery (Fig. 3), but this possibility has not yet been realized. This scoping review identified a heterogeneous range of wearable sensors that have been studied in patients undergoing major abdominal surgery. Most studies were non-randomized and focused on the feasibility of using wearables to monitor physical activity or vital signs. Sensors were generally commercially available products and were not designed specifically for postoperative monitoring. Data were predominantly stored on the wearable device without wireless transmission, and rates of device failure and data loss were poorly reported. Adherence with the device was also infrequently described, and analysis was predominantly conducted retrospectively with a ‘per-protocol’ analysis rather than in real-time. Several studies suggested that measurements from wearable sensors were associated with clinical outcomes, including complications, duration of hospital stay, and readmission; however, while promising, the overall efficacy of these devices for early detection of complications compared with existing standards of care remains unclear.
Fig. 3.
Potential preoperative and postoperative uses of wearable technology for patients undergoing major abdominal surgery
RR, respiratory rate; HR, heart rate; Temp, temperature; SpO2, oxygen saturation.
Continuous vital sign and physiological monitoring
Wearable devices have the potential to improve surgical safety by identifying high-risk deteriorating patients for early intervention and ‘rescue’ from complications10,100. Vital sign changes are the fundamental components behind early-warning score systems that have been introduced worldwide to recognize deteriorating patients8; however, intermittent vital sign measurement often misses significant postoperative hypotension, hypoxemia, and apnoea41,101,102, with potential clinical consequences. Continuous monitoring of vital signs and early-warning score data in real-time may allow for earlier detection of vital sign changes, recognition of early signs of complications, and a faster time to intervention103–105. Using wearable devices is clearly preferable to traditional wired bedside monitors; however, technical challenges of accuracy, precision, and data transfer remain incompletely solved. Furthermore, it remains unclear whether earlier detection translates to meaningful clinical benefits. Pilot and feasibility randomized trials have shown promise in reducing unplanned ICU admission and duration of hospital stay35,36; however, these findings need to be replicated in adequately powered efficacy trials across a range of hospital settings.
Multiple sensors are available for continuous vital sign measurement, and these are rapidly advancing106. Patch-based sensors with electrodes for sensing heart rate and respiratory rate such as SenisumVitals, HealthDot, and Vital Patch offer a non-obstructive solution for monitoring with favourable acceptability to patients34,38,47,50,56. More complex systems such as the ViSi Mobile device include the use of finger plethomyosgraphy sensors with a wider range of metrics but may be more intrusive due to wired connections between components56. Although the accuracy of these devices was generally acceptable, precision was highly variable and often outside clinically acceptable limits. Addressing this will require further technical advancements in sensor design, signal processing, and validation against appropriate gold standards107, especially given the ‘digit bias’ evident in nursing measurements of respiratory rate, which are not consistent with true respiratory rates108,109.
Respiratory rate is crucial in identifying deteriorating patients110, and the accurate non-invasive measurement of this metric is paramount to the clinical applicability of wearable devices. In this review, respiration was measured by various techniques, including impedance pneumography, derivation from respiratory sinus arrhythmia changes in electrocardiogram signals, and accelerometers47. The relative accuracy and precision of these methods need further investigation. Furthermore, the assessment of respiratory function may be more complicated than measuring respiratory rate alone. Previous work has shown respiratory rate changes do not correlate with changes in either tidal volume or minute ventilation111. For other vital signs, it remains unclear whether skin temperature measurements from patch-based sensors can capture changes in core temperature, relevant for identifying fevers or other postoperative complications47.
Assessment and optimization of the patient experience will be essential for the implementation of continuous vital signs monitoring in clinical practice112. Although the devices were generally well accepted by the patients in these studies, it should be noted that on average up to one-third of eligible patients declined to participate or withdrew during the study, and the reasons for this remain largely unexplored. The use and implementation of wearable devices in vulnerable patients (those with cognitive impairment, communication difficulties, delirium, or low health literacy) should be explored as this setting poses unique challenges and opportunities for continuous monitoring. Optimal strategies for data processing and presentation to clinicians also remain unclear, as a high rate of false-positive alarms was identified as a barrier to clinical implementation of wearable sensors among nursing staff33,55,57. Optimization of device precision and methods for artefact filtering is essential to prevent alarm fatigue when data are presented in real-time to clinical staff. Averaging data over longer periods has been proposed as a solution to reduce the number of false alarms49, but this should be balanced against maintaining granularity of data to ensure that clinically significant episodes are detected. Prediction of patient deterioration using advanced data analytics may offer another solution to this problem and more accurately identify deteriorating patients113. Future studies should directly assess the impact of continuous monitoring on clinician workload, particularly for nursing staff.
Other potential avenues for wearable technology in postoperative patients include recording of gastrointestinal activity (with acoustic or electrical signals)44,45,114,115, continuous glucose monitoring for subclinical insulin resistance driven by the surgical stress response116,117, sensing of postoperative pain118,119, or other novel biomarkers of autonomic tone120. Continuous oximetry using wearable patches also remains an avenue of further development; however, care must be taken during sensor design to ensure compatibility with different skin tones and prevent the reinforcement of existing healthcare inequities121.
Postoperative physical activity monitoring
Early postoperative mobilization is a core tenant of enhanced recovery after surgery (ERAS) guidelines for all abdominal surgical specialties, and it therefore is a potential avenue for intervention to accelerate patient recovery through the utilization of digital technologies112. Wearables were used by several studies in this review to measure physical activity after surgery, either passively, or as part of other interventions aiming to increase mobilization.
Numerous studies showed that patients who mobilized less (both before and after surgery) were at a higher risk of complications, readmissions, and other adverse outcomes58,66; however, to what extent these findings represent the baseline frailty of patients rather than a potentially modifiable mediator of perioperative risk remains unclear. Notably, a recent randomized trial conducted within an established ERAS programme found that mobilization targets did not reduce patient complications, but increased levels of fatigue92.
Normalizing a patient’s postoperative physical activity relative to their baseline may be a more appropriate method for risk assessment76,80, and the characterization of mobilization ‘patterns’ as suggested by Iida et al. may offer more detailed insights into the prediction of specific postoperative complications78,79. The ability of physical activity to predict complications in ‘real time’, as opposed to retrospectively, also remains unclear, and requires further targeted investigation. Targeted feedback of physical activity data from wearable devices has the potential to change behaviour and decision-making for both patients and clinicians and optimal methods to help guide this should be explored122,123.
There are also several technical challenges with applying wearable activity trackers to monitor postoperative recovery that remain unsolved. Movement in postoperative patients may be characterized by shorter steps that are less purposeful, and concerns regarding the reliability of sensors in these populations have been raised, given few commercial activity monitors have been validated in hospitalized or postoperative patients124,125. A more sophisticated approach than measuring ‘step counts’ may be required for more accurate assessment of postoperative physical activity. Furthermore, concerns have been expressed regarding the reliability of wrist-based measurements of activity126, and this may vary with the location of device placement (such as wrist, ankle, or hip)127.
Limitations of this review
There are several limitations to this scoping review, including its focus on abdominal surgery, without considering other specialties; however, this was carried out as the principles of recovery for abdominal surgery are relatively homogenous across procedures. This review did not assess specialty-specific outcomes such as joint movement after orthopaedic surgery, although the applications of this technology have been described elsewhere128. Second, mobile applications, and environmental sensors also have potential roles in tracking postoperative recovery129–131, either in the hospital or after discharge, though evaluating these approaches was beyond the scope of this review. Finally, we were unable to perform a quantitative analysis due to the heterogeneous methods, devices, and populations included in the review.
Future research
Ongoing work in this field should be guided by the IDEAL framework132, clearly reported133, and initially developed/explored (stage 2a and 2b trials) to develop optimal devices and methods for postoperative monitoring, before moving to an adequately powered stage 3 randomized trial.
Technical advances in accuracy and reliability of wearable devices for physiological monitoring are needed, with consideration of appropriate gold-standard comparisons, and optimization of filtering and alarm thresholds107. Future clinical studies should clearly report adherence with wearable device use, reasons for refusal to participate, and aim to assess device acceptance by patients and nursing staff. Technical performance metrics, including the rates of missing data and device failure, should also be reported. Other authors have called for standardization in the quantification and analysis of data from wearable sensors134, and this remains an area requiring consensus.
‘Failure to rescue’ is an important concept in postoperative monitoring10 but is difficult to apply as a primary outcome given the relatively rare occurrence of postoperative mortality. Proxy measures, including the time of detection of complications compared with standard observations, the overall number, and burden of postoperative complications, rate of unplanned ICU admission, or duration of hospital stay could be considered by future studies investigating wearable devices.
Patient-reported outcome and experience measures should be assessed as part of overall postoperative recovery, in addition to adherence to enhanced recovery protocols and postoperative mobilization. Additionally, home-based continuous vital sign monitoring and ‘unsupervised’ use of devices also remains an area requiring further study38.
Supplementary Material
Acknowledgements
All authors contributed to the conception and design of the study. All authors contributed to drafting and revising of the manuscript. All authors have given final approval for the article to be published.
Disclosure. G.O.G. and A.G. are shareholders and members of Alimetry and hold intellectual property in the field of gastrointestinal electrophysiology. G.O.G. and I.B. are shareholders and members of the Insides Company. No commercial financial support was received for this study. The authors have no other conflicts of interest to declare.
Contributor Information
Cameron I. Wells, Department of Surgery, The University of Auckland, Auckland, New Zealand
William Xu, Department of Surgery, The University of Auckland, Auckland, New Zealand.
James A. Penfold, Department of Surgery, The University of Auckland, Auckland, New Zealand
Celia Keane, Department of Surgery, The University of Auckland, Auckland, New Zealand.
Armen A. Gharibans, Department of Surgery, The University of Auckland, Auckland, New Zealand Auckland Bioengineering Institute, The University of Auckland, Auckland, New Zealand.
Ian P. Bissett, Department of Surgery, The University of Auckland, Auckland, New Zealand Department of Surgery, Auckland District Health Board, Auckland, New Zealand.
Greg O’Grady, Department of Surgery, The University of Auckland, Auckland, New Zealand; Auckland Bioengineering Institute, The University of Auckland, Auckland, New Zealand; Department of Surgery, Auckland District Health Board, Auckland, New Zealand.
Funding
This work was supported by a Health Research Council of New Zealand Programme Grant. C.W. is supported by a Health Research Council of New Zealand Clinical Research Training Fellowship. The funder was not involved in any aspect of study design, data analysis, or manuscript preparation. There are no other sources of funding to report for this manuscript.
Supplementary material
Supplementary material is available at BJS Open online
Data availability
Data sharing is not applicable to this article as no new data were created or analysed in this study.
References
- 1. Simões CM, Carmona MJC, Hajjar LA, Vincent J-L, Landoni G, Belletti A et al. Predictors of major complications after elective abdominal surgery in cancer patients. BMC Anesthesiol 2018;18:49. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 2. Johnston MJ, Arora S, King D, Bouras G, Almoudaris AM, Davis R et al. A systematic review to identify the factors that affect failure to rescue and escalation of care in surgery. Surgery 2015;157:752–763 [DOI] [PubMed] [Google Scholar]
- 3. Burke JR, Downey C, Almoudaris AM. Failure to rescue deteriorating patients: a systematic review of root causes and improvement strategies. J Patient Saf 2022;18:e140–e155 [DOI] [PubMed] [Google Scholar]
- 4. Massarweh NN, Anaya DA, Kougias P, Bakaeen FG, Awad SS, Berger DH. Variation and impact of multiple complications on failure to rescue after inpatient surgery. Ann Surg 2017;266:59–65 [DOI] [PubMed] [Google Scholar]
- 5. Ghaferi AA, Dimick JB. Variation in mortality after high-risk cancer surgery: failure to rescue. Surg Oncol Clin N Am 2012;21:389–395, vii [DOI] [PubMed] [Google Scholar]
- 6. Wei IH, Pappou EP, Smith JJ, Widmar M, Nash GM, Weiser MR et al. Monitoring an ongoing enhanced recovery after surgery (ERAS) program: adherence improves clinical outcomes in a comparison of three thousand colorectal cases. Clin Surg 2020;5:2909. [PMC free article] [PubMed] [Google Scholar]
- 7. Messenger DE, Curtis NJ, Jones A, Jones EL, Smart NJ, Francis NK. Factors predicting outcome from enhanced recovery programmes in laparoscopic colorectal surgery: a systematic review. Surg Endosc 2017;31:2050–2071 [DOI] [PubMed] [Google Scholar]
- 8. Hollis RH, Graham LA, Lazenby JP, Brown DM, Taylor BB, Heslin MJ et al. A role for the early warning score in early identification of critical postoperative complications. Ann Surg 2016;263:918–923 [DOI] [PubMed] [Google Scholar]
- 9. Smith T, Den Hartog D, Moerman T, Patka P, Van Lieshout EMM, Schep NWL. Accuracy of an expanded early warning score for patients in general and trauma surgery wards. Br J Surg 2012;99:192–197 [DOI] [PubMed] [Google Scholar]
- 10. Sessler DI, Saugel B. Beyond ‘failure to rescue’: the time has come for continuous ward monitoring. Br J Anaesth 2019;122:304–306 [DOI] [PubMed] [Google Scholar]
- 11. Michard F, Kalkman CJ. Rethinking patient surveillance on hospital wards. Anesthesiology 2021;135:531–540 [DOI] [PubMed] [Google Scholar]
- 12. Abeles A, Kwasnicki RM, Darzi A. Enhanced recovery after surgery: current research insights and future direction. World J Gastrointest Surg 2017; 9:37–45 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 13. Appelboom G, Yang AH, Christophe BR, Bruce EM, Slomian J, Bruyère O et al. The promise of wearable activity sensors to define patient recovery. J Clin Neurosci 2014;21:1089–1093 [DOI] [PubMed] [Google Scholar]
- 14. Loftus TJ, Tighe PJ, Filiberto AC, Balch J, Upchurch GR Jr, Rashidi P et al. Opportunities for machine learning to improve surgical ward safety. Am J Surg 2020;220:905–913 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 15. Low CA, Bovbjerg DH, Ahrendt S, Choudry MH, Holtzman M, Jones HL et al. Fitbit step counts during inpatient recovery from cancer surgery as a predictor of readmission. Ann Behav Med 2018;52:88–92 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 16. Peters MDJ, Marnie C, Tricco AC, Pollock D, Munn Z, Alexander L et al. Updated methodological guidance for the conduct of scoping reviews. JBI Evid Synth 2020;18:2119–2126 [DOI] [PubMed] [Google Scholar]
- 17. Page MJ, McKenzie JE, Bossuyt PM, Boutron I, Hoffmann TC, Mulrow CD et al. The PRISMA 2020 statement: an updated guideline for reporting systematic reviews. Int J Surg;2021;88:105906. [DOI] [PubMed] [Google Scholar]
- 18. Tricco AC, Lillie E, Zarin W, O’Brien KK, Colquhoun H, Levac D et al. PRISMA extension for scoping reviews (PRISMA-ScR): checklist and explanation. Ann Intern Med 2018;169:467–473 [DOI] [PubMed] [Google Scholar]
- 19. Bramer WM, Giustini D, de Jonge GB, Holland L, Bekhuis T. De-duplication of database search results for systematic reviews in EndNote. J Med Libr Assoc 2016;104:240–243 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 20. Ouzzani M, Hammady H, Fedorowicz Z, Elmagarmid A. Rayyan—a web and mobile app for systematic reviews. Syst Rev 2016;5:210. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 21. Metcalf M, Glazyrine V, Glavin K, Dahlgren A, Michael C, Bechtel M et al. The feasibility of a health care application in the treatment of patients undergoing radical cystectomy. JUrol 2019;201:902–908 [DOI] [PubMed] [Google Scholar]
- 22. den Bakker CM, Huirne JA, Schaafsma FG, de Geus C, Bonjer HJ, Anema JR. Electronic health program to empower patients in returning to normal activities after colorectal surgical procedures: mixed-methods process evaluation alongside arandomized controlled trial. J Med Internet Res 2019;21:e10674. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 23. van der Meij E, Anema JR, Leclercq WKG, Bongers MY, Consten ECJ, Schraffordt Koops SE et al. Personalised perioperative care by e-health after intermediate-grade abdominal surgery: a multicentre, single-blind, randomised, placebo-controlled trial. Lancet 2018;392:51–59 [DOI] [PubMed] [Google Scholar]
- 24. van der Meij E, Huirne JA, Ten Cate AD, Stockmann HB, Scholten PC, Davids PH et al. A perioperative eHealth program to enhance postoperative recovery after abdominal surgery: process evaluation of a randomized controlled trial. JMed Internet Res 2018;20:e1. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 25. Lafaro KJ, Raz DJ, Kim JY, Hite S, Ruel N, Varatkar G et al. Pilot study of a telehealth perioperative physical activity intervention for older adults with cancer and their caregivers. Support Care Cancer 2020;28:3867–3876 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 26. Low CA, Danko M, Durica KC, Kunta AR, Mulukutla R, Ren Y et al. A real-time mobile intervention to reduce sedentary behavior before and after cancer surgery: Usability and feasibility study. JMIR Perioper Med 2020;3:e17292. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 27. Symer MM, Abelson JS, Milsom J, McClure B, Yeo HL. A mobile health application to track patients after gastrointestinal surgery: results from a pilot study. J Gastrointest Surg 2017;21:1500–1505 [DOI] [PubMed] [Google Scholar]
- 28. Wu J-M, Ho T-W, Chang Y-T, Hsu C, Tsai CJ, Lai F et al. Wearable-based mobile health app in gastric cancer patients for postoperative physical activity monitoring: focus group study. JMIR MHealth UHealth 2019;7:e11989. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 29. Barkley R, Khalil M, Shen P, Levine EA, Votanopoulos K, Clark CJ. Feasibility of low-cost accelerometers in measuring functional recovery after major oncologic surgery. J Surg Oncol 2019;121:279–285 [DOI] [PubMed] [Google Scholar]
- 30. Sun V, Dumitra S, Ruel N, Lee B, Melstrom L, Melstrom K et al. Wireless monitoring program of patient-centered outcomes and recovery before and after major abdominal cancer surgery. JAMA Surg 2017;152:852–859 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 31. Jonker LT, Hendriks S, Lahr MM, van Munster BC, de Bock GH,van Leeuwen BL. Postoperative recovery of accelerometer-based physical activity in older cancer patients. Eur J Surg Oncol 2020;46:2083–2090 [DOI] [PubMed] [Google Scholar]
- 32. Vix M, Rodriguez M, Ignat M, Marescaux J, Diana M, Mutter D. Postoperative remote monitoring with a transcutaneous biosensing patch: preliminary evaluation of data collection. Surg Innov 2020;27:320–327 [DOI] [PubMed] [Google Scholar]
- 33. Weenk M, van Goor H, Frietman B, Engelen LJ, van Laarhoven CJ, Smit J et al. Continuous monitoring of vital signs using wearable devices on the general ward: pilot study. JMIR Mhealth Uhealth 2017;5:e91. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 34. Cheng S-M, Chan JJI, Tan CW, Lu E, Sultana R, Sng BL. Use of wireless respiratory rate sensor monitoring during opioid patient-controlled analgesia after gynaecological surgery: a prospective cohort study. Indian J Anaesth 2021;65:146–152 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 35. Downey C, Randell R, Brown J, Jayne DG. Continuous versus intermittent vital signs monitoring using a wearable, wireless patch in patients admitted to surgical wards: pilot cluster randomized controlled trial. J Med Internet Res 2018;20:e10802. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 36. Downey CL, Croft J, Ainsworth G, Buckley H, Shinkins B, Randell R et al. Trial of remote continuous versus intermittent NEWS monitoring after major surgery (TRaCINg): a feasibility randomised controlled trial. Pilot Feasibility Stud 2020;6:183. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 37. Liu Y, Liu C, Gao M, Wang Y, Bai Y, Xu R et al. Evaluation of a wearable wireless device with artificial intelligence, iThermonitor WT705, for continuous temperature monitoring for patients in surgical wards: a prospective comparative study. BMJ Open. 2020; 10: e039474. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 38. Breteler MJM, Numan L, Ruurda JP, van Hillegersberg R, van der Horst S, Dohmen DAJ et al. Wireless remote home monitoring of vital signs in patients discharged early after esophagectomy: observational feasibility study. JMIR Perioper Med 2020;3:e21705. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 39. Breteler MJM, KleinJan E, Numan L, Ruurda JP, Van Hillegersberg R, Leenen LPH et al. Are current wireless monitoring systems capable of detecting adverse events in high-risk surgical patients? A descriptive study. Injury 2020;51(Suppl 2):S97–S105. [DOI] [PubMed] [Google Scholar]
- 40. Drummond GB, Bates A, Mann J, Arvind DK. Characterization of breathing patterns during patient-controlled opioid analgesia. Br J Anaesth 2013;111:971–978 [DOI] [PubMed] [Google Scholar]
- 41. Turan A, Chang C, Cohen B, Saasouh W, Essber H, Yang D et al. Incidence, severity, and detection of blood pressure perturbations after abdominal surgery: a prospective blinded observational study. Anesthesiology 2019;130:550–559. [DOI] [PubMed] [Google Scholar]
- 42. Shimada T, Cohen B, Shah K, Mosteller L, Bravo M, Ince I et al. Associations between intraoperative and post-anesthesia care unit hypotension and surgical ward hypotension. J Clin Anesth 2021;75:110495. [DOI] [PubMed] [Google Scholar]
- 43. Turan A, Essber H, Saasouh W, Hovsepyan K, Makarova N, Ayad S et al. Effect of intravenous acetaminophen on postoperative hypoxemia after abdominal surgery: the FACTOR randomized clinical trial. JAMA 2020;324:350–358 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 44. Dua MM, Navalgund A, Axelrod S, Axelrod L, Worth PJ, Norton JA et al. Monitoring gastric myoelectric activity after pancreaticoduodenectomy for diet ‘readiness.’. Am J Physiol Gastrointest Liver Physiol 2018;315:G743–G751 [DOI] [PubMed] [Google Scholar]
- 45. Navalgund A, Axelrod S, Axelrod L, Singhal S, Tran K, Legha P et al. Colon myoelectric activity measured after open abdominal surgery with a noninvasive wireless patch system predicts time to first flatus. J Gastrointest Surg 2019;23:982–989 [DOI] [PubMed] [Google Scholar]
- 46. Downey C, Ng S, Jayne D, Wong D. Reliability of a wearable wireless patch for continuous remote monitoring of vital signs in patients recovering from major surgery: a clinical validation study from the TRaCINg trial. BMJ Open 2019;9:e031150. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 47. Weenk M, Koeneman M, van de Belt TH, Engelen LJLPG, van Goor H, Bredie SJH. Wireless and continuous monitoring of vital signs in patients at the general ward. Resuscitation 2019;136:47–53 [DOI] [PubMed] [Google Scholar]
- 48. Breteler MJM, KleinJan EJ, Dohmen DAJ, Leenen LPH, van Hillegersberg R, Ruurda JP et al. Vital signs monitoring with wearable sensors in high-risk surgical patients: a clinical validation study. Anesthesiology 2020;132:424–439 [DOI] [PubMed] [Google Scholar]
- 49. Breteler MJM, Huizinga E, van Loon K, Leenen LPH, Dohmen DAJ, Kalkman CJ et al. Reliability of wireless monitoring using a wearable patch sensor in high-risk surgical patients at a step-down unit in the Netherlands: a clinical validation study. BMJ Open 2018;8:e020162. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 50. van der Stam JA, Mestrom EHJ, Scheerhoorn J, Jacobs F, de Hingh IHJT, van Riel NAW et al. Accuracy of vital parameters measured by a wearable patch following major abdominal cancer surgery. Eur J Surg Oncol 2021; DOI: 10.1016/j.ejso.2021.10.034 [Epub ahead of print] [DOI] [PubMed] [Google Scholar]
- 51. Drummond GB, Bates A, Mann J, Arvind DK. Validation of a new non-invasive automatic monitor of respiratory rate for postoperative subjects. Br J Anaesth 2011;107:462–469 [DOI] [PubMed] [Google Scholar]
- 52. Posthuma LM, Visscher MJ, Lirk PB, van Dijkum EJMN, Hollmann MW, Preckel B. Insights into postoperative respiration by using continuous wireless monitoring of respiratory rate on the postoperative ward: a cohort study. J Clin Monit Comput 2020;34:1285–1293 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 53. Mukkamala SG, Gennings C, Wenzel RP. R = 20: bias in the reporting of respiratory rates. Am J Emerg Med 2008;26:237–239 [DOI] [PubMed] [Google Scholar]
- 54. Downey CL, Brown JM, Jayne DG, Randell R. Patient attitudes towards remote continuous vital signs monitoring on general surgery wards: an interview study. Int J Med Inform 2018;114:52–56 [DOI] [PubMed] [Google Scholar]
- 55. Leenen JPL, Dijkman EM, van Dijk JD, van Westreenen HL, Kalkman C, Schoonhoven L et al. Feasibility of continuous monitoring of vital signs in surgical patients on a general ward: an observational cohort study. BMJ Open 2021;11:e042735. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 56. Weenk M, Bredie SJ, Koeneman M, Hesselink G, van Goor H, van de Belt TH. Continuous monitoring of vital signs in the general ward using wearable devices: randomized controlled trial. JMed Internet Res 2020;22:e15471. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 57. Areia C, King E, Ede J, Young L, Tarassenko L, Watkinson P et al. Experiences of current vital signs monitoring practices and views of wearable monitoring: a qualitative study in patients and nurses. J Adv Nurs 2022;78:810–822 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 58. Daskivich TJ, Houman J, Lopez M, Luu M, Fleshner P, Zaghiyan K et al. Association of wearable activity monitors with assessment of daily ambulation and length of stay among patients undergoing major surgery. JAMA Netw Open 2019;2:e187673. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 59. Bisgaard T, Kjaersgaard M, Bernhard A, Kehlet H, Rosenberg J. Computerized monitoring of physical activity and sleep in postoperative abdominal surgery patients. J Clin Monit Comput 1999;15:1–8 [DOI] [PubMed] [Google Scholar]
- 60. Gorzelitz J, Costanzo ES, Spencer RJ, Rumble M, Rose SL, Cadmus-Bertram L. Longitudinal assessment of post-surgical physical activity in endometrial and ovarian cancer patients. PLoS ONE 2019;14:e0223791. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 61. Romain B, Martin D, Fabacher T, Pache B, Hahnloser D, Demartines N et al. Comparison of footsteps using connected bracelets with the timed up-and-go test and the 6-minutes walking test in a prospective colorectal surgery cohort. Nutrients 2020;12:563. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 62. van der Meij E, van der Ploeg HP, van den Heuvel B, Dwars BJ, Meijerink WJHJ, Bonjer HJ et al. Assessing pre- and postoperative activity levels with an accelerometer: a proof-of-concept study. BMC Surg 2017;17:56. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 63. Waller GC, Kim TG, Perez S, Esper GJ, Srinivasan JK, Shaffer VO et al. Comparing activity trackers with vs. without alarms to increase postoperative ambulation: a randomized control trial. Am Surg 2021;87:1093–1098 [DOI] [PubMed] [Google Scholar]
- 64. Carmichael H, Overbey DM, Hosokawa P, Goode CM, Jones TS, Jr BC et al. Wearable technology: a pilot study to define ‘normal’ postoperative recovery trajectories. J Surg Res 2019;244:368–373 [DOI] [PubMed] [Google Scholar]
- 65. Guinan EM, Bennett AE, Doyle SL, O’Neill L, Gannon J, Foley G et al. Measuring the impact of oesophagectomy on physical functioning and physical activity participation: a prospective study. BMC Cancer 2019;19:682. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 66. Hedrick TL, Hassinger TE, Myers E, Krebs ED, Chu D, Charles AN et al. Wearable technology in the perioperative period: predicting risk of postoperative complications in patients undergoing elective colorectal surgery. Dis Colon Rectum 2020;63:538–544 [DOI] [PubMed] [Google Scholar]
- 67. Diniz-Sousa F, Veras L, Boppre G, Devezas V, Santos-Sousa H, Preto J et al. The effect of bariatric surgery on gravitational loading and its impact on bone mass. Bone 2021;153:116153. [DOI] [PubMed] [Google Scholar]
- 68. Kovar A, Carmichael H, Jones TS, Hosokawa P, Goode CM, Overbey DM et al. Early identification of patients at risk for delayed recovery of ambulation after elective abdominal surgery. Surg Endosc 2021;. DOI: 10.1007/s00464-021-08829-9 [Epub ahead of print] [DOI] [PubMed] [Google Scholar]
- 69. Inoue Y, Kimura T, Noro H, Yoshikawa M, Nomura M, Yumiba T et al. Is laparoscopic colorectal surgery less invasive than classical open surgery? Quantitation of physical activity using an accelerometer to assess postoperative convalescence. Surg Endosc 2003;17:1269–1273 [DOI] [PubMed] [Google Scholar]
- 70. Inouez Y, Kimura T, Fujita S, Noro H, Nishikawa K, Yumiba T et al. A new parameter for assessing postoperative recovery of physical activity using an accelerometer. Surg Today 2003;33:645–650 [DOI] [PubMed] [Google Scholar]
- 71. Takiguchi S, Fujiwara Y, Yamasaki M, Miyata H, Nakajima K, Sekimoto M et al. Laparoscopy-assisted distal gastrectomy versus open distal gastrectomy: a prospective randomized single-blind study. World J Surg 2013;37:2379–2386 [DOI] [PubMed] [Google Scholar]
- 72. Ward WH, Meeker CR, Handorf E, Hill MV, Einarson M, Alpaugh RK et al. Feasibility of fitness tracker usage to assess activity level and toxicities in patients with colorectal cancer. JCO Clin Cancer Inform 2021;5:125–133 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 73. Nevo Y, Shaltiel T, Constantini N, Rosin D, Gutman M, Zmora O et al. Activity tracking after surgery: does it correlate with postoperative complications? Am Surg 2021;88:226–232 [DOI] [PubMed] [Google Scholar]
- 74. Wolk S, Linke S, Bogner A, Sturm D, Meißner T, Müssle B et al. Use of activity tracking in major visceral surgery-the enhanced perioperative mobilization trial: a randomized controlled trial. J Gastrointest Surg 2019;23:1218–1226 [DOI] [PubMed] [Google Scholar]
- 75. Browning L, Denehy L, Scholes RL. The quantity of early upright mobilisation performed following upper abdominal surgery is low: an observational study. Aust J Physiother 2007;53:47–52 [DOI] [PubMed] [Google Scholar]
- 76. Kane WJ, Hassinger TE, Myers EL, Chu DL, Charles AN, Hoang SC et al. Wearable technology and the association of perioperative activity level with 30-day readmission among patients undergoing major colorectal surgery. Surg Endosc 2021;36:1584–1592 [DOI] [PubMed] [Google Scholar]
- 77. Robinson TN, Carmichael H, Hosokawa P, Overbey DM, Goode CM, Barnett CC Jr et al. Decreases in daily ambulation forecast post-surgical re-admission. Am J Surg 2021; DOI: 10.1016/j.amjsurg.2021.08.011 [Epub ahead of print] [DOI] [PubMed] [Google Scholar]
- 78. Iida H, Shimizu T, Maehira H, Kitamura N, Mori H, Miyake T et al. A pilot study: the association between physical activity level using by accelerometer and postoperative complications after hepatic resection. Exp Ther Med 2018;16:4893–4899 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 79. Iida H, Maehira H, Mori H, Takebayashi K, Kojima M, Ueki T et al. Usefulness of measuring temporal changes in physical activity levels using an accelerometer for prediction and early detection of postoperative complications after hepatectomy. HPB (Oxford) 2022;24:57–64 [DOI] [PubMed] [Google Scholar]
- 80. Martin D, Romain B, Pache B, Vuagniaux A, Guarnero V, Hahnloser D et al. Physical activity and outcomes in colorectal surgery: a pilot prospective cohort study. Eur Surg Res 2020;61:23–33 [DOI] [PubMed] [Google Scholar]
- 81. Rossi LA, Melstrom LG, Fong Y, Sun V. Predicting post-discharge cancer surgery complications via telemonitoring of patient-reported outcomes and patient-generated health data. J Surg Oncol 2021;123:1345–1352 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 82. Matsui K, Kawakubo H, Mayanagi S, Matsuda S, Irino T, Fukuda K et al. Exploratory prospective study of the influence of radical esophagectomy on perioperative physical activity in patients with thoracic esophageal cancer. Dis Esophagus 2022;35:doab043 [DOI] [PubMed] [Google Scholar]
- 83. Robinson TN, Kovar A, Carmichael H, Overbey DM, Goode CM, Jones TS. Postoperative delirium is associated with decreased recovery of ambulation one-month after surgery. Am J Surg 2021;221:856–861 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 84. Katogi M. Comparison of life-behavior-promoting mobilization care with walking-only mobilization care in post-gastrointestinal surgery patients: a quasi-experimental study. Jpn J Nurs Sci 2020;17:e12348. [DOI] [PubMed] [Google Scholar]
- 85. Jones ASK, Kleinstäuber M, Akroyd A, Mittendorf A, Bognuda P, Merrie AEH et al. Using animated visualization to improve postoperative mobilization: a randomized controlled trial. Health Psychol 2019;38:748–758 [DOI] [PubMed] [Google Scholar]
- 86. Klotz R, Seide SE, Knebel P, Probst P, Bruckner T, Motsch J et al. Continuous wound infiltration versus epidural analgesia for midline abdominal incisions - a randomized-controlled pilot trial (painless-pilot trial; DRKS Number: DRKS00008023). PLoS ONE 2020;15:e0229898 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 87. Porserud A, Aly M, Nygren-Bonnier M, Hagströmer M. Objectively measured mobilisation is enhanced by a new behaviour support tool in patients undergoing abdominal cancer surgery. Eur J Surg Oncol 2019;45:1847–1853 [DOI] [PubMed] [Google Scholar]
- 88. Strother M, Koepsell K, Song L, Faerber J, Bernard J, Malkowicz SB et al. Financial incentives and wearable activity monitors to increase ambulation after cystectomy: a randomized controlled trial. Urol Oncol 2020;39:434. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 89. Arunachalam D, Heit MH. Impact of postoperative instructions on physical activity following pelvic reconstructive surgery: a randomized controlled trial. Int Urogynecol J 2020;31:1337–1345 [DOI] [PubMed] [Google Scholar]
- 90. Guinan EM, Forde C, O’Neill L, Gannon J, Doyle SL, Valkenet K et al. Effect of preoperative inspiratory muscle training on physical functioning following esophagectomy. Dis Esophagus 2019;32:doy091. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 91. Wasowicz-Kemps DK, Slootmaker SM, Kemps HMC, Borel-Rinkes IHM, Biesma DH, van Ramshorst B. Resumption of daily physical activity after day-case laparoscopic cholecystectomy. Surg Endosc 2009;23:2034–2040 [DOI] [PubMed] [Google Scholar]
- 92. Steffens D, Solomon MJ, Beckenkamp PR, Koh CE, Yeo D, Sandroussi C et al. Individualised, targeted step count intervention following gastrointestinal cancer surgery: The Fit-4-Home randomised clinical trial. ANZ J Surg 2021; DOI: 10.1111/ans.17212 [Epub ahead of print] [DOI] [PubMed] [Google Scholar]
- 93. Honke J, Hiramatsu Y, Kawata S, Booka E, Matsumoto T, Morita Y et al. Usefulness of wearable fitness tracking devices in patients undergoing esophagectomy. Esophagus 2022;19:260–268 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 94. No JH, Kim K, Kim YB, Suh DH, Yang EJ, Hwang H et al. Effects of an activity tracker with feedback on physical activity in women after midline laparotomy: a randomized controlled trial. JObstet Gynaecol Res 2021;47:2544–2550 [DOI] [PubMed] [Google Scholar]
- 95. Reed B, Tabone LE, Tabone JK, Szoka N, Abunnaja S, Bailey K. The use of an activity tracker to objectively measure inpatient activity after bariatric surgery. Surg Obes Relat Dis 2021;17:90–95 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 96. Allen RW, Burney CP, Davis A, Henkin J, Kelly J, Judd BG et al. Deep sleep and beeps: Sleep quality improvement project in general surgery patients. J Am Coll Surg 2021;232:882–888 [DOI] [PubMed] [Google Scholar]
- 97. Li S, Song B, Li Y, Zhu J. Effects of intravenous anesthetics vs inhaled anesthetics on early postoperative sleep quality and complications of patients after laparoscopic surgery under general anesthesia. Nat Sci Sleep 2021;13:375–382 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 98. Low CA, Li M, Vega J, Durica KC, Ferreira D, Tam V et al. Digital biomarkers of symptom burden self-reported by perioperative patients undergoing pancreatic surgery: prospective longitudinal study. JMIR Cancer 2021;7:e27975. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 99. Yi Y, Sossenheimer PH, Erondu AI, Skowron KB, Rai V, Singer JM et al. Using wearable biosensors to predict length of stay for patients with IBD after bowel surgery. Dig Dis Sci 2021; DOI: 10.1007/s10620-021-06910-w [Epub ahead of print] [DOI] [PMC free article] [PubMed] [Google Scholar]
- 100. Downey CL, Chapman S, Randell R, Brown JM, Jayne DG. The impact of continuous versus intermittent vital signs monitoring in hospitals: a systematic review and narrative synthesis. Int J Nurs Stud 2018;84:19–27 [DOI] [PubMed] [Google Scholar]
- 101. Sun Z, Sessler DI, Dalton JE, Devereaux PJ, Shahinyan A, Naylor AJ et al. Postoperative hypoxemia is common and persistent: a prospective blinded observational study. Anesth Analg 2015;121:709–715 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 102. Voscopoulos C, Ladd D, Campana L, George E. Non-invasive respiratory volume monitoring to detect apnea in post-operative patients: case series. J Clin Med Res 2014;6:209–214 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 103. van Rossum MC, Vlaskamp LB, Posthuma LM, Visscher MJ, Breteler MJM, Hermens HJ et al. Adaptive threshold-based alarm strategies for continuous vital signs monitoring. J Clin Monit Comput 2021; DOI: 10.1007/s10877-021-00666-4 [Epub ahead of print] [DOI] [PMC free article] [PubMed] [Google Scholar]
- 104. Petit C, Bezemer R, Atallah L. A review of recent advances in data analytics for post-operative patient deterioration detection. J Clin Monit Comput 2018;32:391–402 [DOI] [PubMed] [Google Scholar]
- 105. McGillion MH, Duceppe E, Allan K, Marcucci M, Yang S, Johnson AP et al. Postoperative remote automated monitoring: need for and state of the science. Can J Cardiol 2018;34:850–862 [DOI] [PubMed] [Google Scholar]
- 106. Posthuma LM, Downey C, Visscher MJ, Ghazali DA, Joshi M, Ashrafian H et al. Remote wireless vital signs monitoring on the ward for early detection of deteriorating patients: a case series. Int J Nurs Stud 2020;104:103515. [DOI] [PubMed] [Google Scholar]
- 107. Xu W, Gharibans AA, Bissett IP, O’Grady G, Wells CI, Areia C et al. Continuous wireless postoperative monitoring using wearable devices: further device innovation is needed. Crit Care 2021;25:1–2 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 108. Badawy J, Nguyen OK, Clark C, Halm EA, Makam AN. Is everyone really breathing 20 times a minute? Assessing epidemiology and variation in recorded respiratory rate in hospitalised adults. BMJ Qual Saf 2017;26:832–836 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 109. Cooper S, Cant R, Sparkes L. Respiratory rate records: the repeated rate? J Clin Nurs 2014;23:1236–1238 [DOI] [PubMed] [Google Scholar]
- 110. Cretikos MA, Bellomo R, Hillman K, Chen J, Finfer S, Flabouris A. Respiratory rate: the neglected vital sign. Med J Aust 2008;188:657–659 [DOI] [PubMed] [Google Scholar]
- 111. van den Bosch OFC, Alvarez-Jimenez R, Stam MMH, den Boer FC, Loer SA. Variations in respiratory rate do not reflect changes in tidal volume or minute ventilation after major abdominal surgery. J Clin Monit Comput 2021;35:787–796 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 112. Michard F, Gan TJ, Kehlet H. Digital innovations and emerging technologies for enhanced recovery programmes. Br J Anaesth 2017;119:31–39 [DOI] [PubMed] [Google Scholar]
- 113. Bates DW, Levine D, Syrowatka A, Kuznetsova M, Craig KJT, Rui A et al. The potential of artificial intelligence to improve patient safety: a scoping review. NPJ Digit Med 2021;4:54. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 114. Erickson JC, Bruce LE, Taylor A, Richman J, Higgins C, Wells CI et al. Electrocolonography: non-invasive detection of colonic cyclic motor activity from multielectrode body surface recordings. IEEE Trans Biomed Eng 2020;67:1628–1637 [DOI] [PubMed] [Google Scholar]
- 115. Kaneshiro M, Kaiser W, Pourmorady J, Fleshner P, Russell M, Zaghiyan K et al. Postoperative gastrointestinal telemetry with an acoustic biosensor predicts ileus vs. uneventful GI recovery. J Gastrointest Surg 2016;20:132–139 [DOI] [PubMed] [Google Scholar]
- 116. Donatelli F, Corbella D, Di Nicola M, Carli F, Lorini L, Fumagalli R et al. Preoperative insulin resistance and the impact of feeding on postoperative protein balance: a stable isotope study. J Clin Endocrinol Metab 2011;96:E1789–E1797 [DOI] [PubMed] [Google Scholar]
- 117. Vriesendorp TM, DeVries JH, Holleman F, Dzoljic M, Hoekstra JBL. The use of two continuous glucose sensors during and after surgery. Diabetes Technol Ther 2005;7:315–322 [DOI] [PubMed] [Google Scholar]
- 118. Aqajari SAH, Cao R, Kasaeyan Naeini E, Calderon M-D, Zheng K, Dutt N et al. Pain assessment tool with electrodermal activity for postoperative patients: method validation study. JMIR Mhealth Uhealth 2021;9:e25258. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 119. Martini CH, Boon M, Broens SJL, Hekkelman EF, Oudhoff LA, Buddeke AW et al. Ability of the nociception level, a multiparameter composite of autonomic signals, to detect noxious stimuli during propofol-remifentanil anesthesia. Anesthesiology 2015;123:524–534 [DOI] [PubMed] [Google Scholar]
- 120. Natarajan A, Xu KS, Eriksson B. Detecting divisions of the autonomic nervous system using wearables. In: 2016 38th Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC), Orlando, FL, 2016. p.5761–5764. [DOI] [PubMed] [Google Scholar]
- 121. Hidalgo DC, Olusanya O, Harlan E. Critical care trainees call for pulse oximetry reform. Lancet Respir Med 2021;9:e37. [DOI] [PubMed] [Google Scholar]
- 122. Linton SC, De Boer C, Tian Y, Alayleh A, Bouchard ME, Figueroa A et al. Effect of consumer-grade wearable device data on clinician decision making during post-discharge telephone calls after pediatric surgery. J Pediatr Surg 2021; DOI: 10.1016/j.jpedsurg.2021.09.040 [Epub ahead of print] [DOI] [PubMed] [Google Scholar]
- 123. Patel MS, Asch DA, Volpp KG. Wearable devices as facilitators, not drivers, of health behavior change. JAMA 2015;313:459–460 [DOI] [PubMed] [Google Scholar]
- 124. Anderson JL, Green AJ, Yoward LS, Hall HK. Validity and reliability of accelerometry in identification of lying, sitting, standing or purposeful activity in adult hospital inpatients recovering from acute or critical illness: a systematic review. Clin Rehabil 2018;32:233–242 [DOI] [PubMed] [Google Scholar]
- 125. Beevi FHA, Miranda J, Pedersen CF, Wagner S. An evaluation of commercial pedometers for monitoring slow walking speed populations. Telemed J E Health 2016;22:441–449 [DOI] [PubMed] [Google Scholar]
- 126. Ryan W, Conigrave JH, Basarkod G, Ciarrochi J, Sahdra BK. When is it Good to use Wristband Devices to Measure HRV?: Introducing a New Method for Evaluating the Quality of Data from Photophlethysmography-based HRV Devices. https://psyarxiv.com/t3gdz/download (accessed 22 March 2021)
- 127. Clark CCT, Nobre GC, Fernandes JFT, Moran J, Drury B, Mannini A et al. Physical activity characterization: does one site fit all? Physiol Meas 2018;39:09TR02 [DOI] [PubMed] [Google Scholar]
- 128. Small SR, Bullock GS, Khalid S, Barker K, Trivella M, Price AJ. Current clinical utilisation of wearable motion sensors for the assessment of outcome following knee arthroplasty: a scoping review. BMJ Open 2019;9:e033832. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 129. Dykes PC, Lowenthal G, Lipsitz S, Salvucci SM, Yoon C, Bates DW et al. Reducing ICU utilization, length of stay, and cost byoptimizing the clinical use of continuous monitoring system technology in the hospital. Am J Med 2022;135:337–341 [DOI] [PubMed] [Google Scholar]
- 130. Dawes AJ, Lin AY, Varghese C, Russell MM, Lin AY. Mobile health technology for remote home monitoring after surgery: a meta-analysis. Br J Surg 2021;108:1304–1314 [DOI] [PubMed] [Google Scholar]
- 131. Knight SR, Ng N, Tsanas A, Mclean K, Pagliari C, Harrison EM. Mobile devices and wearable technology for measuring patient outcomes after surgery: a systematic review. NPJ Digit Med 2021;4:157. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 132. Hirst A, Philippou Y, Blazeby J, Campbell B, Campbell M, Feinberg J et al. No surgical innovation without evaluation: evolution and further development of the IDEAL framework and recommendations. Ann Surg 2019;269:211–220 [DOI] [PubMed] [Google Scholar]
- 133. Bilbro NA, Hirst A, Paez A, Vasey B, Pufulete M, Sedrakyan A et al. The IDEAL reporting guidelines: a Delphi consensus statement stage specific recommendations for reporting theevaluation of surgical innovation. Ann Surg 2021;273:82–85 [DOI] [PubMed] [Google Scholar]
- 134. Alharbi M, Straiton N, Smith S, Neubeck L, Gallagher R. Data management and wearables in older adults: a systematic review. Maturitas 2019;124:100–110 [DOI] [PubMed] [Google Scholar]
Associated Data
This section collects any data citations, data availability statements, or supplementary materials included in this article.
Supplementary Materials
Data Availability Statement
Data sharing is not applicable to this article as no new data were created or analysed in this study.