Abstract

We propose a rational electrode design concept for affinity biosensors based on electrochemical impedance spectroscopy to substantially suppress unexpected device-to-device variations. On the basis that the uniformity of the current distribution affects the variation, a novel micro-gap parallel plate electrode (PPE) was developed, where two planar electrodes with edges covered with a SiO2 layer were placed face to face. The structure provides a uniform current distribution over the planar electrode surface and maximizes the contribution of the planar electrode surface to sensing. For a comparative study, we also fabricated a micro-structured interdigitated electrode (IDE) that has been widely adopted for high-sensitivity measurement, although its current is highly concentrated on the electrode edge corner. Protein G (PrG) molecules were immobilized on both electrodes to prepare an immunoglobulin G (IgG) biosensor on which the specific binding of PrG–IgG can occur. We demonstrated that the IgG sensor with the PPE has small device-to-device variations, in strong contrast to the sensor with the IDE having large device-to-device variations. The results indicate that the current distribution on the electrode surface is important to fabricating electrochemical impedance spectroscopy biosensors with small device-to-device variations. Furthermore, it was found that the PPE allows ultrasensitive detection, that is, the sensor exhibited a linear range from 1 × 10–13 to 1 × 10–7 mol/L with a detection limit of 1 × 10–14 mol/L, which is a record sensitivity at low concentrations for EIS-based IgG sensors.
Introduction
Biosensing technology currently attracts much research interest in a wide variety of fields, such as clinical analysis (e.g., the analysis of biomarkers and hormones), food safety control, and environmental monitoring.1−3 There is an increasing need for an analytical technique that allows signal amplification and noise reduction and thus realizes low detection limits for a wide range of target analytes depending on the field.4,5 The enzyme-linked immunosorbent assay has generally been shown to be suitable in such a scenario and has been well developed for diagnosis and laboratory examinations.6−8 However, the technique suffers from several drawbacks, such as the time and effort required, the requirement for high operational skill, and the need for sophisticated instrumentation. Electrochemical impedance spectroscopy (EIS) has the advantages of a rapid and simple pretreatment, minimal manipulation, and uncomplicated instrumentation. EIS, which is an electrochemical technique, works by measuring the impedance change induced by a specific binding event of receptors immobilized on a surface through the relationship between applying a small sinusoidal perturbation potential and the derived current response.9−14 In recent decades, many efforts have been made to optimize the electrode structure for EIS measurements.15,16 In particular, the electrode design of an interdigitated microelectrode (IDE) array has been widely adopted. This structure comprises a series of parallel microband electrodes in which alternating microbands are connected and form a pair of interdigitated electrode fingers. The small gap between the cathodic and anodic electrodes results in the IDE having a high collection efficiency owing to the enhanced redox cycling between the electrodes, providing an advantage in the detection of a small amount of analyte.17,18 In addition, the need for a reference electrode can be eliminated in an equimolar environment such that [Fe(CN)6]3– and [Fe(CN)6]4– exist at the same molar concentration, which makes EIS easy and simple to carry out with the two-electrode system rather than the conventionally used three-electrode system.18,19 We have been developing several biosensors based on the IDE structure and demonstrated their highly sensitive and selective responses to target analytes.20−25 In the cited studies, we noted that a major drawback of an IDE-based affinity sensor is its low reproducibility, although the structure greatly improves the sensitivity. It has been frequently observed that individual IDE biosensors have different profiles for the calibration curves (sensing signal vs concentration) of individual chips even when they are from the same batch of IDE fabrication. Such a low reproducibility of the individual chips presents a serious obstacle to sensor application because a prior calibration procedure is required for each chip. A similar poor reproduction of the IDE has been reported in other studies,26,27 but its origin and solution remain open questions. It is thus an important issue for the design of a practical EIS biosensor to establish a novel electrode design structure that avoids this problem. We suppose that a high current density at the edge corner of the electrode is the most likely origin of the low reproducibility of an IDE. Several studies conducting finite element analysis showed that the electric flux between the adjacent electrodes of the IDE is highly concentrated on the edge corner of the electrodes.28−32 However, the edge part can be easily damaged in the lithography processes involving lift-off and chemical etching applied to the pattern IDE shape on a substrate. EIS measures the rate of charge transfer between the solution and the electrode that the electric current passes through and, therefore, EIS with IDE simply measures the rate of charge transfer on the disordered surface. We suppose that such a current focusing on the edge corner explains the low reproducibility of IDE-based EIS sensors.
One way to overcome the above drawback is to configure a new electrode structure that has an insulated electrode edge and enables the current to be uniformly distributed on a planar electrode surface. The simplest electrode design that provides the above characteristics is the parallel plate electrode (PPE) structure, where two electrode plates are placed face-to-face with a narrow gap. The electric current flows along the shortest path between the electrodes, and the electric flux is thus distributed uniformly on the planar surface. This is close to the ideal surface condition for EIS measurement. For this design, the gap between the two electrodes can be precisely controlled by varying the thickness of the insertion layer.
In this report, we demonstrate that the PPE structure offers greater reproducibility in EIS measurement than the IDE structure. We first simulate the current density for both the PPE and IDE to visualize current distributions depending on the electrode design. We then compare the EIS biosensor performance for the PPE and IDE, using the specific binding of immunoglobulin G (IgG) with protein G (PrG) as a test reaction,33−37 on the basis of Nyquist plots and calibration curves (Figure 1). We finally confirm that a non-specific binding event has no effect on the sensor response. The obtained results indicate that the PPE has a great advantage in terms of the reproducibility of EIS measurements as well as the ultrasensitive detection of IgG.
Figure 1.
Schematic of the experiment conducted to characterize the IgG sensor performance. (a) Top view of the working electrode (upper) and specific binding of IgG with PrG (lower). (b) Setting configuration of the PPE(upper) and schematic of the [Fe(CN)6]3–/4– redox cycle for EIS measurement (lower). (c) Rct evaluation with Nyquist plot. (d) Magnified image of the PPE edge part.
Results and Discussion
Simulation of the Current Density
The current density was simulated using COMSOL Multiphysics ver. 5.3 (COMSOL AB, Sweden), which adopts finite element analysis. In the simulation, the gap between the electrodes was set at a constant distance of 2 μm for both the IDE (10 μm band width) and PPE (3 mm working electrode and 5 mm counter electrode) models with a 10 mV voltage applied to the deionized water (σ = 5.5 × 10–6 S/m). The electrodes in the models consist of a 200 nm thickness Au layer deposited on a SiO2 substrate (σ = 1 × 10–14 S/m, εr = 4.2). The simulations were performed assuming that the parameters of electrical conductivity and dielectric constant are isotropic. Figure 2a shows the current density around a finger of the IDE. It is seen that the electric current is highly concentrated on the edge corner of the IDE and thinly distributed on the flat surface. Such a highly inhomogeneous distribution of the current is due to the closely spaced side-by-side electrode arrangement. Generally, the surface portion gathering a higher flux of current more strongly affects the EIS measurement. Accordingly, the EIS measurement with the IDE substrate reflects mainly the surface condition of the edge corner. However, the surface condition of the edge corner is considerably disordered or inhomogeneous because the edge is initially damaged in the etching process. In this sense, the EIS measurements with IDE substrates readily suffer from instability or low reproducibility. Figure 2b shows the current density around the PPE. In contrast with the current distribution on the IDE, the current distribution on the PPE is extremely uniform without there being any area of high or low current density on the electrode surface. A major reason for such a uniform distribution is that the two electrode plates are closely placed in a face-to-face geometry and, accordingly, the current flows directly along the shortest normal direction between the electrodes over the surface. Such a uniform current distribution on the flat surface is advantageous for obtaining stable EIS results because the surface has the most smooth and well-ordered structure on the electrode outside. The EIS measurement using the uniformly distributed current on the ideal condition surface should be highly reproducible.
Figure 2.

Simulated distribution of the current density around the electrodes of an IDE (a) and PPE (b). Geometric parameters of both electrodes were a thickness of 200 nm and a gap of 2 μm. The PPE edge was covered with a 1 μm-thick SiO2 layer.
Characterization of the IgG Sensor
The results of EIS measurements made with the PPE are shown as Nyquist plots (Zim vs −Zre) in Figure 3a. These results were obtained for successive immersion and measurement cycles at IgG concentrations ranging from 0 to 10–5 g/mL. The figure shows that a semicircular profile was obtained for each measurement. The diameter of the semicircle increased with the IgG concentration. The diameter of the semicircle represents the charge-transfer resistance (Rct), which is generally used as the sensor signal for a Faradic EIS biosensor. During each incubation of the sample in the IgG solution, there were specific binding events between IgG and PrG on the electrode surface. In the EIS measurements of the [Fe(CN)6]3–/[Fe(CN)6]4– redox probe solution, this binding of IgG prevents reversible electron transfer between the electrode and redox probe and thus increases Rct. Accordingly, the observed gradual increase in the diameter of the semicircle as a function of the IgG concentration reflects the IgG binding event on the electrode surface. Reliable Rct values were obtained by fitting the impedance spectra to the Randles equivalent circuit model (Figure 3b).38 In this modeling, the electrode–electrolyte interface is modeled as a parallel circuit of Rct and a constant phase element (CPE). The CPE expresses the frequency dependence of the double-layer capacitance (Cdl) arising from the capacitance dispersion. Rct is the main signal of the EIS measurement and indicates the degree of IgG binding on the electrode surface as described above. In the low-frequency range, the diffusion from the redox probe to the electrode can be slower than the charge transfer, and accordingly, the Warburg impedance (Zw) must be introduced in series with Rct to represent this behavior. Rsol is the resistance of the solution between the two electrodes. All these parameters were fit to each impedance spectrum and their values determined. The solid lines in Figure 3a are obtained from the best-fitting parameters and well reproduce all the experimental results, indicating that the obtained parameter sets well explain the interfacial behavior. The details of the fitting results are given in the Supporting Information (Figure S1 and Table S1).
Figure 3.

(a) Nyquist plots of the EIS spectra obtained using a PrG-modified PPE at various concentrations of IgG. The solid lines represent the fitting curves. (b) Equivalent circuit for data fitting.
The IgG sensor was quantitively assessed by the dependence of Rct/Rct (0 g/mL) on the IgG concentration, where Rct (0 g/mL) is Rct for no IgG. Figure 4 shows the calibration curves of Rct/Rct (0 g/mL) versus the logarithmic IgG concentration obtained from several samples of each electrode type. All the profiles show a linear increase of Rct/Rct (0 g/mL) in the IgG concentration range from 1 × 10–13 to 1 × 10–7 g/mL, which corresponds to IgG–PrG-specific binding on the electrode surface depending on the concentration. The lowest detectable concentration was 1 × 10–14 g/mL (see Supporting Information, Figure S2). Above this range, Rct/Rct (0 g/mL) remains at a constant value because the PrG binding sites on the electrode surface are saturated with IgG. Such behaviors are commonly observed for EIS-based biosensors. Figure 4 shows that the sensor with the PPE has better reproducibility than that with the IDE; that is, the four PPE samples have similar calibration curves, whereas the three IDE samples have different profiles. As described in the previous section, we suppose that this difference in reproducibility of the calibration curve stems from the current distribution and surface condition of the electrode. The uniform current distribution on the well-ordered surface of the PPE provides constant calibration curves, whereas the highly concentrated current on the disordered edge corner of the IDE can cause unstable profiles.39 The results clearly demonstrate that the PPE is more favorable in sensor applications than the IDE.
Figure 4.

Calibration curves for four individual PPE sensors (red filled symbols) and three individual IDE sensors (blue open symbols).
Selectivity of the IgG Sensor
We performed two experiments to assess the IgG selectivity of our biosensor using the PPE. We first investigated the interfering effect of IgA on the EIS measurement. IgA is the second-largest component of human immunoglobulin and has a protein structure similar to that of IgG. A series of EIS measurements were performed over the IgA concentration range of 1 × 10–13 to 1 × 10–5 g/mL, from the lowest to the highest concentration, with the same experimental conditions as for IgG. Figure 5a shows the obtained result together with the previous result for IgG. We see that Rct was almost independent of the concentration for IgA in contrast with that for IgG, which indicates that our IgG sensor does not respond to IgA at all.
Figure 5.

(a) Selectivity of the PPE sensor. Calibration curves obtained for IgG exposure (red circles) and IgA exposure (blue circles) are plotted. (b) Confirmation of non-specific binding. Calibration curves obtained for PPEss with (red triangle) and without (blue triangle) PrG immobilization are plotted.
We then prepared a blank surface through the immobilization of only bovine serum albumin (BSA). On this surface, there was no PrG receptor, and IgG only attached through non-specific binding (i.e., physical adsorption). We made the same series of IgG sensing performance measurements for the sample and obtained a calibration curve, which is shown in Figure 5b together with the previous result obtained with the PrG immobilized. It is seen that Rct remains at an almost constant low value even for solutions with a higher IgG concentration. We conclude from this result that non-specific binding is avoidable by passivating the surface with BSA.
Finally, the analytical performances of the PPE-based IgG sensor with those previously reported have been compared (Table 1). The data comparison shows that the PPE sensor is more sensitive than the other sensors reported. In particular, the linear range of the PPE sensor extends to 2 orders of magnitude lower concentrations than the other sensors, which is a record sensitivity at low concentrations for an EIS-based IgG sensor. These research results indicate that PPE-based EIS sensors have highly reproducible and ultrasensitive characteristics.40−44
Table 1. Comparison of Different Immunosensors for the Determination of IgG.
| method | linear range (g/mL) | LoD (g/mL) | ref. |
|---|---|---|---|
| EIS | 1.13 × 10–8–1.13 ×10–4 | (40) | |
| EIS | 5 × 10–10–2 × 10–7 | 3 × 10–11 | (41) |
| EIS | 5 ×10–10–1.25 × 10–7 | 2 × 10–11 | (42) |
| DPVa | 1 × 10–10–5 × 10–8 | 2 × 10–11 | (43) |
| DPVa | 1.2 × 10–11–3.52 × 10–7 | 6 × 10–12 | (44) |
| EIS | 1 × 10–13 – 1 × 10–7 | 1 × 10–14 | this work |
Differential pulse voltammetry.
Conclusions
Progressing toward a practical impedimetric biosensor, we developed a novel micro-gap PPE that has not only higher reproducibility than the traditional interdigitated electrode but also ultrasensitive characteristics. An important aspect of obtaining high reproducibility in our study was to cover the edge of the electrode with an insulator so that it does not contribute to the sensing. As a future perspective on the present work, we suppose that our novel PPE is suitable for a sensor array system because unexpected inherent sensor variability during the manufacturing process should be reduced for a sensor array system.45,46 The results of the study are useful for the further development of highly reproducible biosensors.
Methods
The sequence of the experiment is shown in Figure 1. A set of PPEs comprises two pieces of Au-patterned quartz glass (see Figure 1). The substrate was prepared adopting standard photolithography techniques. We fabricated two types of electrode with different diameters: a working electrode (3 mm) on which an IgG receptor PrG was immobilized and a counter electrode (5 mm) of a pure Au surface. On each electrode, a SiO2 layer with a thickness of 1 μm was formed around the patterned electrode through plasma-enhanced chemical vapor deposition. An important aspect of this process is that the edge corner of the electrode was covered with the SiO2 layer because the insulator covering the edge corner gives rise to a uniformly distributed current flux on the electrode. The PPE structure was then created by dropping a solution onto the working electrode plate and covering the solution with the counter electrode plate. For comparison, we fabricated IDE chips using similar photolithography techniques. The fabrication details are described in the Supporting Information (Figure S3).
The working electrode surface was modified with a self-assembled monolayer of a mixed molecular system (11-mercaptoundecanoic acid:6-mercapto-1-hexanol ratio of 1:3). PrG was covalently attached to the self-assembled monolayer. The residual active esters were blocked with BSA. The conditions of the surface modification are detailed in the Supporting Information (Figure S4).
The experimental procedure adopted to characterize the IgG sensor performance is shown in Figure 1b. First, the PrG-functionalized working electrode of the PPE was immersed in IgG having a fixed concentration for 20 min at 20 °C and then rinsed with distilled water and dried under a N2 flow. A phosphate buffered saline (PBS) solution containing an equimolar mixture (5 mM) of [Fe(CN)6]3– and [Fe(CN)6]4– was dropped on the working electrode, and the counter electrode was then placed to cover the drop, resulting in the thin liquid PBS layer (having a thickness of 2 μm) being sandwiched between the two electrodes. EIS measurements were carried out using this two-electrode system at an equilibrium potential of the [Fe(CN)6]3–/[Fe(CN)6]4– couple of 10 mV in the frequency range of 0.1–1 × 105 kHz. The apparatus used for the measurements was an Autolab PGSTAT128N (Metrohm, The Netherlands).
It is worth noting that the PPE fabricated in this study can be applied to two-electrode cyclic voltammetry (CV) measurements in the [Fe(CN)6]3–/[Fe(CN)6]4– equilibrium condition. We observed that the current in the CV measurements with PPE was significantly increased with a smaller electrode gap distance (Figure S5). This result indicates that the [Fe(CN)6]3–/4– redox cycle is enhanced between a narrow PPE gap.
Acknowledgments
This work was supported by the JSPS KAKENHI (grant no.20K05300). Part of this work was supported by the NIMS Nanofabrication Platform of the Nanotechnology Platform Project sponsored by the Ministry of Education, Culture, Sports, Science and Technology (MEXT), Japan.
Supporting Information Available
The Supporting Information is available free of charge at https://pubs.acs.org/doi/10.1021/acsomega.1c06942.
Numerical fitting parameters, lowest detection limit, PPE fabrication process, electrode functionalization process, and CV measurement with PPE (PDF)
The authors declare no competing financial interest.
Supplementary Material
References
- Kim J.; Campbell A. S.; de Ávila B. E.-F.; Wang J. Wearable Biosensors for Healthcare Monitoring. Nat. Biotechnol. 2019, 37, 389–406. 10.1038/s41587-019-0045-y. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Neethirajan S.; Weng X.; Tah A.; Cordero J. O.; Ragavan K. V. Nano-Biosensor Platforms for Detecting Food Allergens—New Trends. Sens. Bio-Sens. Res 2018, 18, 13–30. 10.1016/j.sbsr.2018.02.005. [DOI] [Google Scholar]
- Jarque S.; Bittner M.; Blaha L.; Hilscherova K. Yeast Biosensors for Detection of Environmental Pollutants: Current State and Limitations. Trends Biotechnol. 2016, 34, 408–419. 10.1016/j.tibtech.2016.01.007. [DOI] [PubMed] [Google Scholar]
- Purohit B.; Vernekar P. R.; Shetti N. P.; Chandra P. Biosensor Nanoengineering: Design, Operation, and Implementation for Biomolecular Analysis. Sensors 2020, 1, 100040. 10.1016/j.sintl.2020.100040. [DOI] [Google Scholar]
- Soleymani L.; Li F. Mechanistic Challenges and Advantages of Biosensor Miniaturization into the Nanoscale. ACS Sens. 2017, 2, 458–467. 10.1021/acssensors.7b00069. [DOI] [PubMed] [Google Scholar]
- Aydin S. A short history, principles, and types of ELISA, and our laboratory experience with peptide/protein analyses using ELISA. Peptides 2015, 72, 4–15. 10.1016/j.peptides.2015.04.012. [DOI] [PubMed] [Google Scholar]
- Ito E.; Iha K.; Yoshimura T.; Nakaishi K.; Watabe S.. Early diagnosis with ultrasensitive ELISA. 1st ed.; Elsevier Inc., 2021, Vol. 101, pp 121–133. [DOI] [PubMed] [Google Scholar]
- Klumpp-Thomas C.; Kalish H.; Drew M.; Hunsberger S.; Snead K.; Fay M. P.; Mehalko J.; Shunmugavel A.; Wall V.; Frank P.; Denson J. P.; Hong M.; Gulten G.; Messing S.; Hicks J.; Michael S.; Gillette W.; Hall M. D.; Memoli M. J.; Esposito D.; Sadtler K. Standardization of ELISA Protocols for Serosurveys of the SARS-CoV-2 Pandemic Using Clinical and at-Home Blood Sampling. Nat. Commun. 2021, 12, 113. 10.1038/s41467-020-20383-x. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Qi H.; Wang C.; Cheng N. Label-Free Electrochemical Impedance Spectroscopy Biosensor for the Determination of Human Immunoglobulin G. Microchim. Acta 2010, 170, 33–38. 10.1007/s00604-010-0382-5. [DOI] [Google Scholar]
- Daniels J. S.; Pourmand N. Label-Free Impedance Biosensors: Opportunities and Challenges. Electroanalysis 2007, 19, 1239–1257. 10.1002/elan.200603855. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Bogomolova A.; Komarova E.; Reber K.; Gerasimov T.; Yavuz O.; Bhatt S.; Aldissi M. Challenges of Electrochemical Impedance Spectroscopy in Protein Biosensing. Anal. Chem. 2009, 81, 3944–3949. 10.1021/ac9002358. [DOI] [PubMed] [Google Scholar]
- Luo X.; Davis J. J. Electrical Biosensors and the Label Free Detection of Protein Disease Biomarkers. Chem. Soc. Rev. 2013, 42, 5944–5962. 10.1039/c3cs60077g. [DOI] [PubMed] [Google Scholar]
- Lisdat F.; Schäfer D. The Use of Electrochemical Impedance Spectroscopy for Biosensing. Anal. Bioanal. Chem. 2008, 391, 1555–1567. 10.1007/s00216-008-1970-7. [DOI] [PubMed] [Google Scholar]
- Muñoz J.; Montes R.; Baeza M. Trends in Electrochemical Impedance Spectroscopy Involving Nanocomposite Transducers: Characterization, Architecture Surface and Bio-Sensing. Trac. Trends Anal. Chem. 2017, 97, 201–215. 10.1016/j.trac.2017.08.012. [DOI] [Google Scholar]
- Medina-Sánchez M.; Ibarlucea B.; Pérez N.; Karnaushenko D. D.; Weiz S. M.; Baraban L.; Cuniberti G.; Schmidt O. G. High-Performance Three-Dimensional Tubular Nanomembrane Sensor for DNA Detection. Nano Lett. 2016, 16, 4288–4296. 10.1021/acs.nanolett.6b01337. [DOI] [PubMed] [Google Scholar]
- Ding S.; Das S. R.; Brownlee B. J.; Parate K.; Davis T. M.; Stromberg L. R.; Chan E. K. L.; Katz J.; Iverson B. D.; Claussen J. C. CIP2A Immunosensor Comprised of Vertically-Aligned Carbon Nanotube Interdigitated Electrodes towards Point-of-Care Oral Cancer Screening. Biosens. Bioelectron. 2018, 117, 68–74. 10.1016/j.bios.2018.04.016. [DOI] [PubMed] [Google Scholar]
- Li D.; Batchelor-Mcauley C.; Chen L.; Compton R. G. Band Electrodes in Sensing Applications: Response Characteristics and Band Fabrication Methods. ACS Sens. 2019, 4, 2250–2266. 10.1021/acssensors.9b01172. [DOI] [PubMed] [Google Scholar]
- Min J.; Baeumner A. J. Characterization and Optimization of Interdigitated Ultramicroelectrode Arrays as Electrochemical Biosensor Transducers. Electroanalysis 2004, 16, 724–729. 10.1002/elan.200302872. [DOI] [Google Scholar]
- Arya S. K.; Chornokur G.; Venugopal M.; Bhansali S. Antibody Functionalized Interdigitated μ-Electrode (IDμE) Based Impedimetric Cortisol Biosensor. Analyst 2010, 135, 1941–1946. 10.1039/c0an00242a. [DOI] [PubMed] [Google Scholar]
- Ohno R.; Ohnuki H.; Wang H.; Yokoyama T.; Endo H.; Tsuya D.; Izumi M. Biosensors and Bioelectronics Electrochemical Impedance Spectroscopy Biosensor with Interdigitated Electrode for Detection of Human Immunoglobulin A. Biosens. Bioelectron. 2013, 40, 422–426. 10.1016/j.bios.2012.07.052. [DOI] [PubMed] [Google Scholar]
- Tsugimura K.; Ohnuki H.; Wu H.; Endo H.; Tsuya D.; Izumi M. Oriented Antibody Immobilization on Self-Assembled Monolayers Applied as Impedance Biosensors. J. Phys.: Conf. Ser. 2017, 924, 012015. 10.1088/1742-6596/924/1/012015. [DOI] [Google Scholar]
- Fujishiro S.; Ohnuki H.; Wu H.; Tsuya D.; Endo H. Regenerable Myoglobin Biosensor Based on Protein G Immobilized on Interdigitated Electrodes. Jpn. J. Appl. Phys. 2020, 59, SCCA05. 10.7567/1347-4065/ab48c6. [DOI] [Google Scholar]
- Tsugimura K.; Ohnuki H.; Endo H.; Tsuya D.; Izumi M. Protein-G-Based Human Immunoglobulin G Biosensing by Electrochemical Impedance Spectroscopy. Jpn. J. Appl. Phys. 2016, 55, 02BE06. 10.7567/jjap.55.02be06. [DOI] [Google Scholar]
- Wagh M. D.; Sahoo S. K.; Goel S. Laser-Induced Graphene Ablated Polymeric Microfluidic Device with Interdigital Electrodes for Taste Sensing Application. Sens. Actuators, A 2022, 333, 113301. 10.1016/j.sna.2021.113301. [DOI] [Google Scholar]
- Afsarimanesh N.; Mukhopadhyay S. C.; Kruger M. Performance Assessment of Interdigital Sensor for Varied Coating Thicknesses to Detect CTX-I. IEEE Sens. J. 2018, 18, 3924–3931. 10.1109/jsen.2018.2818718. [DOI] [Google Scholar]
- Bonyár A.; Harsányi G.. Typical Problems and Solutions in Electrochemical Measurement Cell Development. 33rd International Spring Seminar on Electronics Technology, ISSE 2010, 2010, pp 433–438.
- Navrátilová I.; Skládal P. The Immunosensors for Measurement of 2,4-Dichlorophenoxyacetic Acid Based on Electrochemical Impedance Spectroscopy. Bioelectrochemistry 2004, 62, 11–18. 10.1016/j.bioelechem.2003.10.004. [DOI] [PubMed] [Google Scholar]
- Singh K. V.; Bhura D. K.; Nandamuri G.; Whited A. M.; Evans D.; King J.; Solanki R. Nanoparticle-Enhanced Sensitivity of a Nanogap-Interdigitated Electrode Array Impedimetric Biosensor. Langmuir 2011, 27, 13931–13939. 10.1021/la202546a. [DOI] [PubMed] [Google Scholar]
- Jung H.-W.; Chang Y. W.; Lee G.-y.; Cho S.; Kang M.-J.; Pyun J.-C. A Capacitive Biosensor Based on an Interdigitated Electrode with Nanoislands. Anal. Chim. Acta 2014, 844, 27–34. 10.1016/j.aca.2014.07.006. [DOI] [PubMed] [Google Scholar]
- Zafarani H. R.; Mathwig K.; Sudhölter E. J. R.; Rassaei L. Electrochemical Redox Cycling in a New Nanogap Sensor: Design and Simulation. J. Electroanal. Chem. 2016, 760, 42–47. 10.1016/j.jelechem.2015.11.031. [DOI] [Google Scholar]
- Alayo N.; Fernández-Sánchez C.; Baldi A.; Esquivel J. P.; Borrisé X.; Pérez-Murano F. Gold Interdigitated Nanoelectrodes as a Sensitive Analytical Tool for Selective Detection of Electroactive Species via Redox Cycling. Microchim. Acta 2016, 183, 1633–1639. 10.1007/s00604-016-1792-9. [DOI] [Google Scholar]
- Singh K. V.; Whited A. M.; Ragineni Y.; Barrett T. W.; King J.; Solanki R. 3D Nanogap Interdigitated Electrode Array Biosensors. Anal. Bioanal. Chem. 2010, 397, 1493–1502. 10.1007/s00216-010-3682-z. [DOI] [PubMed] [Google Scholar]
- Tran T.; Eskilson O.; Mayer F.; Gustavsson R.; Selegård R.; Lundström I.; Mandenius C. F.; Martinsson E.; Aili D. Real-Time Nanoplasmonic Sensor for IGG Monitoring in Bioproduction. Processes 2020, 8, 1302. 10.3390/pr8101302. [DOI] [Google Scholar]
- Akerström B.; Brodin T.; Reis K.; Björck L. Protein, G: A Powerful Tool for Binding and Detection of Monoclonal and Polyclonal Antibodies. J. Immunol. 1985, 135, 2589–2592. [PubMed] [Google Scholar]
- Akerström B.; Björck L. A Physicochemical Study of Protein G, a Molecule with Unique Immunoglobulin G-Binding Properties. J. Biol. Chem. 1986, 261, 10240–10247. 10.1016/s0021-9258(18)67515-5. [DOI] [PubMed] [Google Scholar]
- Siwy Z.; Trofin L.; Kohli P.; Baker L. A.; Trautmann C.; Martin C. R. Protein Biosensors based on Biofunctionalized Conical Gold Nanotubes. J. Am. Chem. Soc. 2005, 127, 5000–5001. 10.1021/ja043910f. [DOI] [PubMed] [Google Scholar]
- Saha K.; Bender F.; Gizeli E. Comparative Study of IgG Binding to Proteins G and A: Nonequilibrium Kinetic and Binding Constant Determination with the Acoustic Waveguide Device. Anal. Chem. 2003, 75, 835–842. 10.1021/ac0204911. [DOI] [PubMed] [Google Scholar]
- MacDonald D. D. Reflections on the History of Electrochemical Impedance Spectroscopy. Electrochim. Acta 2006, 51, 1376–1388. 10.1016/j.electacta.2005.02.107. [DOI] [Google Scholar]
- Delgado-Arenas H. F.; Rodríguez-López A.; Rivera F.; Ramos K. J.; Reséndiz-Ramírez R.; Antano-Lopez R. Effect of Electrode Geometry on the Electrolyte Resistance Measurement over the Surface of a Skin Phantom in a Noninvasive Manner. Bioelectrochemistry 2019, 130, 107337. 10.1016/j.bioelechem.2019.107337. [DOI] [PubMed] [Google Scholar]
- Schrattenecker J. D.; Heer R.; Melnik E.; Maier T.; Fafilek G.; Hainberger R. Hexaammineruthenium (II)/(III) as Alternative Redox-Probe to Hexacyanoferrat (II)/(III) for Stable Impedimetric Biosensing with Gold Electrodes. Biosens. Bioelectron. 2019, 127, 25–30. 10.1016/j.bios.2018.12.007. [DOI] [PubMed] [Google Scholar]
- Cui M.; Gong Y.; Du M.; Wang K.; Li T.; Zhu X.; Wang S.; Luo X. An Antifouling Electrochemical Biosensor Based on a Protein Imprinted Hydrogel for Human Immunoglobulin G Recognition in Complex Biological Media. Sens. Actuators, B 2021, 337, 129820. 10.1016/j.snb.2021.129820. [DOI] [Google Scholar]
- Amouzadeh Tabrizi M.; Shamsipur M.; Mostafaie A. A High Sensitive Label-Free Immunosensor for the Determination of Human Serum IgG Using Overoxidized Polypyrrole Decorated with Gold Nanoparticle Modified Electrode. Mater. Sci. Eng., C 2016, 59, 965–969. 10.1016/j.msec.2015.10.093. [DOI] [PubMed] [Google Scholar]
- Axin Liang A.; Huipeng Hou B.; Shanshan Tang C.; Liquan Sun D.; Aiqin Luo E. An Advanced Molecularly Imprinted Electrochemical Sensor for the Highly Sensitive and Selective Detection and Determination of Human IgG. Bioelectrochemistry 2021, 137, 107671. 10.1016/j.bioelechem.2020.107671. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Barman S. C.; Zahed M. A.; Sharifuzzaman M.; Ko S. G.; Yoon H.; Nah J. S.; Xuan X.; Park J. Y. A Polyallylamine Anchored Amine-Rich Laser-Ablated Graphene Platform for Facile and Highly Selective Electrochemical IgG Biomarker Detection. Adv. Funct. Mater. 2020, 30, 1907297. 10.1002/adfm.201907297. [DOI] [Google Scholar]
- Zhang L.; Tian F.; Kadri C.; Xiao B.; Li H.; Pan L.; Zhou H. On-Line Sensor Calibration Transfer among Electronic Nose Instruments for Monitoring Volatile Organic Chemicals in Indoor Air Quality. Sens. Actuators, B 2011, 160, 899–909. 10.1016/j.snb.2011.08.079. [DOI] [Google Scholar]
- Zhang L.; Tian F.-C.; Peng X.-W.; Yin X. A Rapid Discreteness Correction Scheme for Reproducibility Enhancement among a Batch of MOS Gas Sensors. Sens. Actuators, A 2014, 205, 170–176. 10.1016/j.sna.2013.11.015. [DOI] [Google Scholar]
Associated Data
This section collects any data citations, data availability statements, or supplementary materials included in this article.

