Comparative Analysis of Optoelectronic Accuracy in the Laboratory Setting Versus Clinical Operative Environment: A Systematic Review

Bryan W Cunningham 1,2,, Daina M Brooks 1
Editor: Paul C McAfee
PMCID: PMC8998481  PMID: 35393881

Abstract

Study Design:

Systematic review.

Objectives:

The optoelectronic camera source and data interpolation process serve as the foundation for navigational integrity in robotic-assisted surgical platforms. The current systematic review serves to provide a basis for the numerical disparity observed when comparing the intrinsic accuracy of optoelectronic cameras versus accuracy in the laboratory setting and clinical operative environments.

Methods:

Review of the PubMed and Cochrane Library research databases was performed. The exhaustive literature compilation obtained was then vetted to reduce redundancies and categorized into topics of intrinsic accuracy, registration accuracy, musculoskeletal kinematic platforms, and clinical operative platforms.

Results:

A total of 465 references were vetted and 137 comprise the basis for the current analysis. Regardless of application, the common denominators affecting overall optoelectronic accuracy are intrinsic accuracy, registration accuracy, and application accuracy. Intrinsic accuracy equaled or was less than 0.1 mm translation and 0.1 degrees rotation per fiducial. Controlled laboratory platforms reported 0.1 to 0.5 mm translation and 0.1 to 1.0 degrees rotation per array. Accuracy in robotic-assisted spinal surgery reported 1.5 to 6.0 mm translation and 1.5 to 5.0 degrees rotation when comparing planned to final implant position.

Conclusions:

Navigational integrity and maintenance of fidelity of optoelectronic data is the cornerstone of robotic-assisted spinal surgery. Transitioning from controlled laboratory to clinical operative environments requires an increased number of steps in the optoelectronic kinematic chain and error potential. Diligence in planning, fiducial positioning, system registration and intra-operative workflow have the potential to improve accuracy and decrease disparity between planned and final implant position.

Keywords: optoelectronic accuracy, spinal surgery, spinal biomechanics, imaging, navigation, robotics

Introduction

The fundamental technological challenge of navigation and robotic-assisted spinal surgery is the virtual world needs to clearly represent the physical, real time world. Among the multiple applications, variables and equipment utilized in navigation and robotic-assisted spinal surgery, the optoelectronic camera source and data interpolation process serves as the foundation for navigational integrity and accuracy, or lack thereof, in the surgical platform. The principles of optoelectronic measurement systems are founded on the basis of devices which have the capability to source, detect and control light and are typically considered a subdivision of photonics. The spectrum of optoelectronic technology platforms are quite diverse, with utilization in sports performance activities such as speed skating and soccer,1-4 human ergonomics,5,6 clinical gait and motion analysis,7-12 musculoskeletal kinematics,13-20 and clinical operative procedures.21-33 To this end, the degree of accuracy and errors acceptable across optoelectronic motion measurement platforms differ considerably based on application. 1 For example, fiducial arrays placed on anatomic pelvic landmarks of alpine skiers reported translation accuracy and errors of 8.37 ± 7.1 millimeters (mm). 34 Although considered adequate for the evaluation of positional or orientation-related differences in this athletic application, discrepancies of this magnitude would be unacceptable in the clinical operative setting. Technological advancements in the accuracy of optoelectronic marker-based systems over the past 20 years have facilitated the adoption and application of these platforms to the field of robotic-assisted spinal surgery.12,35,36 An ensuing plethora of journal publications have documented the safety, efficacy and technical accuracy of navigation and robotic systems,21-24,30-32,37-45 operative surgical applications,25,33,46,47 as well as challenges of process workflow, learning curve and training.28,30,31,39,40

Review of these publications, reveals what could be defined as a significant discrepancy when comparing optoelectronic accuracy in the laboratory setting versus clinical operative environment. An approximate 10-fold decrease in technical accuracy of final implant position (≤2 mm) in the clinical operative environment was observed compared to controlled musculoskeletal kinematic studies (≤0.2 mm), despite utilization of near identical optoelectronic camera systems. Hence, the objective of the current systematic review serves to provide a basis for the numerical disparity that exist when comparing intrinsic accuracy of optoelectronic cameras, accuracy observed in the laboratory setting and accuracy in the clinical operative environment. It is postulated that there exists a greater number of linkages in the optoelectronic kinematic chain when analyzing the clinical operative environment to laboratory setting. This increase in data interpolation, coupled with intraoperative challenges, reduces the degree of accuracy compared to that observed in controlled musculoskeletal kinematic laboratory investigations.

Methods

A comprehensive systematic review of the PubMed and Cochrane Library research databases was performed. The time interval was unrestricted, however, the majority of publications comprising the basis of this analysis ranged from 2000–present. Combinations of key search terms were stratified into the following: optoelectronic measurement systems, technical accuracy, experimental error, robotic assisted spinal surgery, spinal kinematics, and navigation. The search was limited to papers in the English language, indexed in peer-reviewed journals accessible through online searches, and all publications included required a bona fide PubMed identification (pmid) or digital object identifier (doi) citation. The exhaustive literature compilation obtained was then pooled in an EndNote file, vetted to reduce redundancies and categorized into topics pertinent to optoelectronic measurement system accuracy with specific reference to intrinsic accuracy, registration accuracy, musculoskeletal kinematic platforms, and clinical operative platforms. The primary tier for inclusion focused on publications which reported quantitative units of measure (microns, millimeters and degrees) for intrinsic camera accuracy and tolerances, accuracy obtained in a controlled laboratory setting and accuracy in the clinical operative setting. Musculoskeletal kinematic measurement studies were included if motions observed were greater than the standard error of measure (SEM). The second tier of inclusion criteria focused on publications highlighting the applications of optoelectronic measurement systems, percent accuracy of screw placement, inherent inaccuracies of surgical instruments, observational commentary leading to accuracy improvement, and methods to mitigate error potential.

Results

Three-Dimensional Cartesian Rigid Body Transformations

The reported optoelectronic measurements of accuracy, errors and methods to quantify these in the laboratory setting or clinical operative environment are based on a fixed 3-dimensional Cartesian coordinate system of rigid body transformation in millimeters (mm) translation and degrees (deg) rotation along 3 orthogonal axes—X, Y and Z.48-52 This is in accordance to the axial (Y), sagittal (Z) and coronal (X) anatomic planes as defined by Panjabi’s 3-dimensional conceptual framework for spinal kinematics (Figure 1A and B).53-55 From a nomenclature standpoint, accuracy is defined as a combination of trueness and precision according to the published ISO standard 5725-1. 56 Trueness refers to the difference between measured value and true position—typically represented by the mean value of repeated measurements. Precision is a measure of repeatability—typically represented by the standard deviation of repeated measurements and refers to random error and noise within the system. In addition to these standardizations, a useful key measure with regard to accuracy (trueness and precision) is the root mean square distance error (RMS) as given by ei being the 3-dimensional distance error of measurement i and N the number of measurements. 57

Figure 1.

Figure 1.

Cartesian coordinate system and conceptual fame for spinal kinematics—schematic representation of a fixed 3-dimensional cartesian coordinate system for calculation of rigid body transformation in millimeters (mm) translation and degrees (deg) rotation along 3 orthogonal axes—X, Y and Z (A). This is in accordance to the axial (Y), sagittal (Z), and coronal (X) anatomic planes as defined by Panjabi’s 3-dimensional conceptual framework for spinal kinematics (B).

eRMS=1N1Neiei

Optoelectronic Measurement Systems

Image guided surgery (IGS) is based on the principal of integration and registration of the operative field to pre- or intra-operative data set (e.g. CT or MRI), via amalgamation of an optoelectronic imaging system with robotic platform. 58 Although not necessarily involved in the execution of operative procedures, optoelectronic measurement systems are considered the gold standard in motion capture accuracy1,59 and provide 3-dimensional visualization and guidance; improving task execution, targeting accuracy while functioning in a semi-autonomous fashion. 60 Hence, objective accuracy and error assessments of optoelectronic-robotic interventional platforms is essential. Regardless of optoelectronic camera system, the fundamental triad of common denominators in assessing platform accuracy include: 1) intrinsic accuracy of the source device, 2) registration and tracking accuracy, and 3) application accuracy. Prior to addressing the basis for application accuracy across laboratory versus clinical platforms, the intrinsic and registration accuracy and potential for error propagation are of primary consideration.

Intrinsic (Technical) Accuracy

The initial link in the optoelectronic kinematic chain of data transference resides in the intrinsic accuracy of the camera source. Of the multiple factors affecting downstream optoelectronic accuracy in musculoskeletal kinematic and clinical operative platforms, the intrinsic camera components are most controllable. Mechanical compliance of the system, loose interconnection mechanisms, 58 variation in camera resolution, calibration, imperfect lenses, number of cameras, spatial orientation, noise, computer vision algorithms and jitter all represent sources of intrinsic error in optoelectronic systems.36,58,61-63 Topley and Richards 36 reported that optoelectronic cameras of higher resolution (Vicon 16MP) and number of cameras (n = 12) significantly improve the 3-dimensional spatial accuracy (0.080 ± 0.092 mm) compared to an equal number of cameras of lower resolution (OptiTrack 1.3MP) (0.259 ± 0.084 mm). Khadem et al 64 reported the intrinsic optical tracking error secondary to jitter. In the absence of intrinsic jitter, repeated static measurements from a camera source to fiducial would produce identical kinematic signatures along 3 orthogonal axes. Comparison of 5 different optoelectronic camera systems utilized in image based surgical navigation was performed [Image Guided Technologies (IGS), Boulder, CO and Northern Digital Imaging (NDI), Waterloo Canada]. The intrinsic static jitter (mean and standard deviation values) for the IGS Flashpoint systems ranged from 0.028 ± 0.012 mm (Flashpoint 300 mm), 0.051 ± 0.038 mm (FlashPoint 580 mm) to 0.059 ± 0.047 mm (Flashpoint 1 m). The NDI Polaris systems indicated mean values of 0.058 ± 0.037 mm (active LED) and 0.115 ± 0.075 (passive LED). When performing static single marker measurements according to ASTM guidelines 65 of the NDI Optotrak 3020, a motion analysis system commonly used in musculoskeletal kinematic platforms,13,17 Maletsky et al 66 reported the relative accuracy position between 2 rigid bodies at 0.03 mm translation and 0.04 degrees angulation, respectively. Elfring et al 57 evaluated the static volumetric single marker measurements of 3 commercially available optoelectronic tracking systems utilized in robotic assisted platforms—NDI Polaris P4, NDI Polaris Spectra (active and passive mode) and Stryker Navigation System II (Stryker Inc., Kalamazoo, MI). The Stryker Navigation System II camera and the Polaris Spectra (active mode), exhibited trueness values of 0.058 ± 0.033 mm and 0.089 ± 0.061 mm, respectively. The Polaris Spectra (passive mode) exhibited values of 0.170 ± 0.090 mm, and the Polaris P4 demonstrated the highest static measurement error of 0.272 ± 0.394 mm. As a baseline statement of comparison, the optoelectronic systems utilized in clinical operative or controlled laboratory platforms report only marginal differences in accuracy. Further, the contribution of intrinsic errors is of miniscule value in comparison to error(s) propagation secondary to registration, targeting tracking and application in controlled experimental and clinical operative platforms.

Registration Accuracy and Target Tracking

A second key step in the optoelectronic kinematic chain and highest probable link(s) of error propagation is the registration process. This intra-operative process integrates correlation and mapping algorithms to register the physical patient to the virtual patient via the navigation system, optoelectronic source, fiducial arrays in the operative field, and patient CT images. Accurate, close-to-ideal reference reproducibility of the dataset improves trueness and precision of subsequent intra-operative tracking. Multiple factors affect registration accuracy and target tracking, including optoelectronic camera source, passive versus active arrays, occlusions, distance between fiducial arrays and camera source, and anatomic locations of the coordinate reference fiducials.1,57,67-74

As reported by States and Pappas 75 and Simoes et al, 68 the NDI Certus 3 camera optoelectronic system demonstrated a significantly higher degree of registration target accuracy (0.1 mm translation and 0.13 degrees rotation 75 compared to the NDI Spectra 2 camera system when located 1.5 meters from camera source. 68 The surgical navigation principle of triangulation, necessary to quantify 3-dimensional fiducial array position, is void if 1 of 2 cameras is occluded.61,76,77 Marinetto et al 78 reported tracking errors secondary to camera and fiducial array occlusions when utilizing an 8-camera configuration. Occlusions of a single camera resulted in tracking errors from 0.2 mm to 0.5 mm, while occlusion of 5 cameras resulted in errors from 0.6 mm to 1.6 mm. Optical trackers utilizing multi-camera systems may lead to data redundancy but enables fidelity in data transference, despite fiducial occlusions within the application platform (Figure 2A-C).

Figure 2.

Figure 2.

Optoelectronic camera sources—comparison of 3 optoelectronic camera systems utilized in motion analysis. The NDI spectra 2 camera system (A), NDI certus optotrak 3 camera system (B), and viconvicon MX13 multiple camera system (C) (Vicon motion systems Ltd., Oxford, UK). The spectra NDI is commonly used in the clinical operative environment and latter 2 systems for musculoskeletal kinematics and biomechanics. The Vicon camera image and testing setup was generously provided by Prof. Dr. Hans Joachim Wilke, PhD, Institute of Orthopaedic Research and Biomechanics, University of Ulm.

The difference between active versus reflective passive markers systems also effect registration accuracy and target tracking. In the case of passive markers, the optoelectronic source floods the operative field and light is reflected back to the sensors via infrared-reflecting spheres. Active markers contain and emit infrared-emitting diodes (IREDs) and provide a more robust signal with increased accuracy compared to passive markers.57,72,79 Furthermore, the distance between optoelectronic camera source and operative field fiducial arrays effect accuracy. 64 Increasing the camera distance from 6 to 8 feet nearly triples the intrinsic registration error along the Z axis (maximum = 0.250 mm) for the Polaris passive fiducial array system. Hence, closer approximation of the optoelectronic camera source to the operative fiducial arrays (≤6 feet or approximately 1800 millimeters) minimizes jitter and improves precision. 64 For comparison, when performing static single marker measurements of the NDI Optotrak 3020 system, the precision in rotation degrades significantly when positioning the camera ≥ 2.5meters from fiducial arrays66,75 (Figure 3A and B).

Figure 3.

Figure 3.

Active versus passive marker arrays—comparison of active marker arrays used in the laboratory platform as shown attached to the vertebral elements. Note the active markers contain and emit infrared-emitting diodes (IREDs) via the attached wiring configurations (A). In the case of passive fiducial marker arrays (B), the optoelectronic source floods the operative field and light is reflected back to the camera source via infrared-reflecting spheres.

In addition to distance between camera source and fiducial arrays, Citak et al70,80 performed an investigation to determine the maximum acceptable distance between fixed reference arrays and dynamic “mobile” arrays within the operative field. The results demonstrated a mean registration error of 0.04 mm (0.04-0.05 mm) up to a distance of 200 mm from the patient reference array. When mobile fiducial arrays (e.g., end effector) exceeded 200 mm from the reference point, the registration error increased to 0.25 mm (0.24-0.26) (P < .0001). The initial registration and intra-operative working accuracy is significantly reduced with localization of arrays greater than 200 mm from the patient reference point.70,80 Moreover, the magnitude of errors secondary to array obstructions differ based on active versus passive markers and distance between arrays. 35 When comparing 2 active Vicon fiducial arrays versus distance, targeting errors increased from 0.47 mm to 1.2 mm, with corresponding stepwise decreases from 5 cm to 0 cm between arrays. In similarity but of greater magnitude, targeting errors for passive arrays (Qualisys AB, Goteborg Sweden) increased from 0.54 mm to 2.99 mm with decreases from 5 cm to 0 cm between arrays. Hence, as visualized by the camera source, increasing distances between fiducial arrays in the operative clinical setting reduces experimental targeting errors (Figure 4A-D).

Figure 4.

Figure 4.

Fiducial array occlusions—Schematic illustration comparing fiducial array configuration and occlusions secondary to optimal optoelectronic camera angle, steep camera angle, occluded sphere and interference between 2 fiducial arrays (A-D).

The effect of transitioning from static to dynamic array localization directly influences accuracy and targeting errors.44,45,69,81 A comparison of static and dynamic motion by Chassat and Lavallee 69 demonstrated a significant increase in array translation error when moving across a spectrum of conditions located 2 meters from the camera source. Using the best-in-class NDI Optotrak 3020 system, the static position error measured 0.28 mm, static position hand held 0.400 mm, and dynamic measure in translation at constant speed 1.29 mm. The magnitude in angular errors was less than translation but still increased significantly through dynamic localization of instrument arrays. 69 These findings were corroborated by Stancic et al 81 with static registration indicated errors of 0.11 mm while dynamic motion of the arrays resulted in errors ranging from 0.250 mm to 1.10 mm with increasing velocities. 81 Hence, higher displacement rates of the operative instrument arrays directly affect the camera’s tracking accuracy.

Two additional clinical factors influencing registration accuracy and targeting errors include the tool to tip distance and accuracy of pre- or intra-operative radiographic CT/MRI data.72,82-84 Static accuracy for a given marker array is on the order of 0.2 mm, however, errors increase significantly when extrapolated to the instrument tip, Wiles et al 72 reported that increasing the distance between fiducial array to instrument tip from 0 to 100 mm decreases static accuracy from 0.4 mm to 0.85 mm, respectively. 72 The array to tool tip distance on instruments utilized in robotic assisted platforms far exceed 100 mm (4 inches). In consideration of the linearity of Wiles et al accuracy degradation with distance calculations, 72 it is postulated that experimental targeting errors of longer instruments are of increased magnitude compared the reported 0.85 mm. Computed tomographic data input affects target accuracy and can be improved with resolutions of 1.0 mm and 2.0 mm versus 3.0 mm slice thickness.82,83

In summary, propagation of computational measurement errors in the optoelectronic kinematic chain has a compounding effect for the following transitions: 1) Measurement of the intrinsic image plane error (IPE) secondary to errors within the optoelectronic system, 2) transitioning from image error to fiducial location error (FLE), and 3) transition from fiducial location error to tracking target error (TRE). The mathematical expressions for these computational transformations are beyond the scope of the current publication but well documented by Fitzpatrick et al71,73 and Sielhorst et al. 61 The margins of error secondary to intrinsic and registration accuracy in optoelectronics are more manageable compared to unpredictable factors related to application in the laboratory versus dynamic clinical intraoperative environments.

Application Accuracy—Musculoskeletal Kinematics Laboratory Platform

Over the last 30 years, a plethora of spine publications have documented the biomechanical properties of the occipitocervical through lumbopelvic spine under controlled laboratory conditions utilizing Panjabi’s 3-dimensional conceptual framework for testing.17,53-55,85 In contrast to the challenges of the clinical operative environment, motion analysis of spinal implant and anatomic vertebral structures(s) in the controlled laboratory setting is performed utilizing a 6 degree of freedom musculoskeletal simulator interfaced with an optoelectronic measurement system. The fundamental principles pertinent to maximizing optoelectronic accuracy include mounting the specimen to a rigid testing platform, affixing active or passive fiducial arrays directly to implants or anatomic structures using screw-bolt fixation, and creating rigid body configurations parallel to the camera source. Multidirectional flexibility testing is typically performed along 3 predominant loading axes—flexion-extension, lateral bending, and axial rotation under controlled displacement rates of 1 to 3 degrees per second for multiple cycles. To this end, a series of laboratory investigations using the NDI Certus and Vicon MX13 camera systems (Vicon Motion Systems Ltd., Oxford, UK) reported the peak limits of optoelectronic accuracy when evaluating kinematics of the osteoligamentous spine13-19,86-92 (Figure 5A and B).

Figure 5.

Figure 5.

Application accuracy—musculoskeletal kinematics laboratory platform in contrast to the challenges of the clinical operative environment, motion analysis of spinal implant and anatomic vertebral structures(s) in the controlled laboratory setting is performed utilizing a 6 degree of freedom musculoskeletal simulator interfaced with an optoelectronic measurement system (A). The fundamental principles pertinent to maximizing optoelectronic accuracy include mounting the specimen to a rigid testing platform, affixing active or passive fiducial arrays directly to implants or anatomic structures using screw-bolt fixation, and creating rigid body configurations parallel to the camera source (B).

Cunningham et al 14 compared occipital plate versus intracranial anchors for reconstruction of the occipitocervical (O-C) junction. The reported differences (degrees) in axial rotation at the O-C junction based on optoelectronic measurements were 4.13 ± 2.05 (intact), 0.22 ± 0.13 (plate) and 0.30 ± 0.21 (anchor). Rotation of the plate and anchor, with respect to the occiput, in flexion-extension ranged from 0.06 ± 0.05 to 0.10 ± 0.08, respectively. Although not of clinical significance, the study quantified differences on the order of 0.1 degree between 2 methods of occipitocervical fixation. Bowden et al 86 described the variability in “quality of motion” of L4 relative to the L5 vertebral elements, following various methods of vertical preload. The optoelectronic data quantified differences in L4 anteroposterior translations of 0.6 mm and rotations of 0.6 degrees between different testing conditions. Ilharreborde et al 87 reported dynamic kinematic evaluation of the multi-segmental lumbar spine under dynamic loading conditions indicated peak accuracy to within 1.10 ± 0.18 degrees (L2-L3) and corresponding translations of 0.48 ± 0.06 mm rotation (L3-L4) with application of 7.5 Nm pure moment load. In a complex kinematic study utilizing a Vicon optoelectronic system, La Barbera et al18,19 investigated lumbar interbody cages with Ponte osteotomy versus pedicle subtraction osteotomy for severe sagittal imbalance, the peak accuracy of neutral zone measurements (degrees) across the intact L3-L5 segments demonstrated values of 0.7 (Range 0.3-1.9) in flexion-extension, 1.0 (Range 0.1-3.8) in lateral bending and 0.2 (Range 0.1-0.9) in axial rotation in flexion-extension.

From a kinematic standpoint, the sacroiliac junction (SIJ) presents formidable challenge to definitive and accurate measurement using optoelectronics—reaching the 0.1 degree accuracy error measurement limitations of most optoelectronic systems.18,19 Jeong et al 88 was able to differentiate the range of motion (degrees) of the SIJ in lateral bending when comparing the intact condition (1.5 ± 1.5), unilateral fusion (1.4 ± 1.6) and bilateral fusion (1.1 ± 1.0). Despite quantification of SIJ motion within a range of 0.5 degrees of accuracy, the comparisons were not significant. Osterhoff et al 89 quantified the effects of cement augmentation on sacroiliac screw position and fracture site motion using NDI optoelectronics. Screw tip positioning within the sacrum was quantified to an accuracy level of 0.7 mm (Range 0.5-1.3), with a corresponding vertical (Y axis) SIJ range of motion of 1.2 mm (Range 0.6-1.9) under cyclic compressive loads. In a comprehensive S2 alar iliac screw instrumentation study using 21 lumbopelvic specimens, Cunningham and co-workers 13 reported SIJ motion (degrees) to accuracies of 1.78 ± 0.96 (flexion-extension), 0.52 ± 0.34 (lateral bending) and 0.48 ± 0.32 degrees (axial rotation). In a similar study by Dall et al 15 evaluating SIJ fusion using lateral sacroiliac screws, optoelectronic tracking quantified the intact SIJ motion (degrees) under flexion-extension, lateral bending and axial rotation at 1.80 ± 1.62, 0.44 ± 0.35 and 1.13 ± 0.82, respectively. Additional studies on intact SIJ kinematics reported translations along the 3-dimensional coordinate axes on the order of 0.1 mm to 1.04 mm with corresponding rotations of 0.11 to 1.14 degrees.16,91,92

Laboratory workflow methods and conditions for experimental musculoskeletal kinematic studies are streamlined and optimized for maximizing optoelectronic accuracy. Factors of specimen stabilization, alignment, camera resolution, proximity to fiducials, planar visualization of the active arrays, and controlled motion application account for the high degree of accuracy reported in these studies. Moreover, in difference to the operative clinical environments, non-destructive testing procedures can be repeated multiple times on the same specimen to improve data accuracy. The collective effect of testing methodology and limited experimental coordinate transformations between data input/output reduces error propagation and maximizes optoelectronic accuracy (Figure 6).

Figure 6.

Figure 6.

Laboratory platform for optoelectronic data transference process—Schematic illustration demonstrating the laboratory workflow and process for data transference utilizing optoelectronic tracking. The camera source visualizes the active fiducial arrays affixed to the vertebral elements and transfers the data directly to the user interface for computational analysis. The collective effect of testing methodology and limited experimental coordinate transformations between data input/output reduces error propagation and maximizes optoelectronic accuracy.

Application Accuracy—Clinical Operative Platform

Transitioning from controlled laboratory conditions to the dynamic variability of a clinical operative environment presents a different set of application challenges for maintaining peak optoelectronic accuracy. Unique to robotic assisted spinal surgery and in difference to the laboratory setting, the intra-operative process requires considerably more steps in the transference of optoelectronic kinematic data. This complex process flow integrates correlation and mapping algorithms to register the physical patient to the virtual patient via the navigation system, optoelectronic source, surveillance markers, patient reference markers, end effector instruments in the operative field, and patient CT images. Accurate, close-to-ideal reference reproducibility and maintenance of this dataset is the primary intra-operative objective and challenge (Figure 7).

Figure 7.

Figure 7.

Application accuracy—clinical operative platform. Intra-operative images highlight the transition from controlled laboratory conditions to the dynamic variability of the clinical operative environment (A). A different set of application challenges are required for maintaining peak optoelectronic accuracy related to registering the physical patient to the virtual patient via the navigation system, optoelectronic source, surveillance markers, patient reference markers, end effector instruments in the operative field and workflow. The intra-operative images were generously provided by the operative surgeons, (A) Dr. Vladimir Sinkov, MD and (B) Dr. Bhavuk Garg, MD.

Despite utilization of near identical optoelectronic sources and fiducial arrays, a consistent disparity exists when comparing the reported technical accuracies in the laboratory setting versus clinical operative environment. An approximate 10-fold decrease in accuracy was observed when comparing the final implant position (≤2 mm) in the clinical operative environment to musculoskeletal kinematic studies (≤0.2 mm). An extensive number of peer reviewed journal publications have documented the use, efficacy safety and technical accuracy achieved with robotic-assisted spinal surgery.23,28,30,37,39-41,44-46,93-123 The focus in reviewing these publications was to highlight the technical accuracy observations and determine a basis for discrepancy between planned versus actual final implant position based on postoperative CT images. In case studies where quantitative measurements were not reported, the Gertzbein and Robbins 124 score (GRS) was adopted to calculate pedicle screw implant position. According to the GRS classification, screws centered within the pedicle are considered grade A; < 2 mm from center is a grade B; a breach from 2 mm to 4 mm is grade C; a breach from 4 mm to 6 mm is grade D; and > 6 mm is grade E. Grades of A and B (< 2 mm pedicle breach) are considered clinically acceptable, and all other grades indicate malposition.

Helm et al 94 performed a comprehensive literature review on the technical accuracy of 12 622 pedicle screws implanted using a variety of image guided surgery navigation systems. As reported, 11 830 were positioned perfectly according to pre-operative plan (A), 395 screws within less than 2 mm of plan (B), 92 breached between 2 mm to 4 mm off center (C), and 55 were within 4 mm to 6 mm of the pre-operative plan. The balance of 250 screws remained ungraded due to radiographic issues. Zhang et al 45 reported one of the largest compilations assessing screw accuracy in robotic assisted spinal surgery. A total of 23 studies including prospective, prospective randomized control trials and retrospective reviews comprised the basis of this publication with a total of 5,013 pedicle screw positions evaluated. The accuracy according to GRS grades of A and B (less than 2 mm cortical breach) was 4,781 screws (95.38%), while the balance of 232 screws (4.61%) ranged from 2 mm to 6 mm when comparing the planned to actual final positions. Solomiichuk et al 41 reported on 192 screws implanted in 35 patients. Trajectories were Grade A or B (less than 2 mm) in 162 (84.4%) of screws. The malposition rate of 2 mm to 6 mm was present in 30 of 192 screws (15.6%) with 23 of these occurring in the thoracic spine, where pedicle widths are significantly less than the lumbar region.103125 Devito et al 93 reported on the technical accuracy placement of 646 pedicle screws inserted in 139 patients using postoperative CT scans. 577 were centered in the pedicle, 58 were less than 2 mm off center, 9 breached 2 mm to 4 mm off center and 2 screws deviated greater than 4 mm from the pedicle wall. Schatlo et al 39 reported on the technical accuracy of 244 lumbar pedicle screws. 204 screws (83.6%) were graded as a perfect trajectory (A) compared to the pre-operative plan, 19 (7.8%) were less than 2 mm (B), 9 (3.7%) breached 2 mm to 4 mm off center (C), 4 (1.6%) breached from 4 mm to 6 mm (D), 2 (0.8%) were greater than 6 mm off center and 6 (2.5%) screws required revision.

A comprehensive study from Keric et al 107 reviewed the technical accuracy of 1857 screw positions based on postoperative CT scans. Of the 1857 screws, 1799 (96.9%) were graded as acceptable or good position, 38 screws (2%) exhibited deviations from 3 mm to 6 mm and 20 screws (1.1%) were greater than 6 mm from the planned trajectory. The 58 malpositioned screws (3 mm to 6 mm deviations) were located primarily in the upper and lower thoracic regions versus the lumbar spine. These deviations were considered secondary to smaller pedicle morphometry 126 and instrument skiving. 107

In consideration of the range for “acceptable” deviations (best case = <2 mm) between planned and actual positions in the clinical setting, attention must be directed toward the anatomical pedicle morphology from the cervical to lumbar regions. For typical cervical vertebrae (C3 to C6) the mean pedicle width was reported 4.9 ± 0.9 mm.127,128 In the thoracic region, pedicle width progressively decreases from T1 to T5 (mean 3.65 ± 0.40 mm) and increases from T6 though the T12 levels (mean 7.89 ± 0.70 mm).103,125 The lumbar vertebrae allow greater cross sectional areas for “ideal” pedicle screw placement, however, the proximity of adjacent exiting nerve roots ranges from 2.9 mm to 6.2 mm proximally and 0.8 mm to 2.8 mm distally. Moreover, the distance between the medial borders of pedicle to thecal sac ranges from 0.9 mm to 2.1 mm. 129

Basis for Disparity in Optoelectronic Accuracy

A key consideration pertaining to optoelectronic accuracy in the clinical environment compared to the laboratory setting is the dynamic nature of the operating room. The basis for disparity in accuracy when equating the laboratory versus clinical operative platforms is a result of the combined, cumulative errors secondary to the intraoperative workflow process, variability in anatomic morphology, and spinal flexibility. Of fundamental importance, and the crux of the matter, related to error propagation in navigation and robotic-assisted spinal surgery is the assumption that the workflow platform and patient’s spine is rigid, and as such, motion of any type is perceived as a rigid body transformation. Optoelectronic error reduction in the clinical flow requires stabilization of the camera source, rigid fixation of surveillance arrays in the iliac crest, stable attachment of patient reference and registration arrays to anatomic landmarks, and end effector instruments arrays which are inflexible. The end effector is the last link where the robotic enters the workspace and small rotations or translations in the array references can lead to large errors in instrument position. Although accurate, close-to-ideal reference reproducibility of these steps will reduce errors, the reality is that fixed arrays do move—leading to increased relative motions between arrays and subsequent error propagation and disparity between the physical, real time world and virtual world. Moreover, spatial errors can be further magnified due to geometrical distortion of preoperative images, and tracking error of the surgical instruments. 130 To register the physical patient to the virtual patient, Grunert et al 67 proposed a series of transformation matrices, including fiducial-based paired-point transformation, surface contour matching, and hybrid transformation. The hybrid transformation process is most applicable to robotic assisted spinal surgery as it includes the methods of surface-based and pair-point based methods with implanted fiducials. As such, tracing at least 3 anatomic landmarks with navigational confirmation serves to reduce error potential.

Several publications on optimizing clinical workflow process have been reported31,38,42,76130-133 Lieberman et al 31 provides an excellent description of the step-by-step workflow process in robotic assisted spinal surgery. The report provides a concise methodological approach to operative workflow, while at the same time providing a collective basis for potential error(s) propagation in the clinical setting. The sequential description of process flow/error potentials includes pre-operative and intra-operative registration, dislodgement of reference arrays, damaged or bent navigation tools134-136 and arrays occlusions (e.g., distance and blood), skiving or tool deflection secondary to sloped anatomic topology or muscle retraction, and untracked patient movement during the spinal destabilizing procedure. In addition to unintended motion or bending of fiducial arrays, the inherent differences in anatomic topology, bone mineral density and flexibility of the patient’s spine, both before and following destabilization and reconstruction procedures, cannot be overemphasized. The challenge is the spine is often flexible—the drill and robotic arm may be properly located, but highly mobile, multi-segmental spinal reconstructions with minimal deflection force leads to unintended rotation or translation of the operative vertebral elements, skiving or tool deflection, and effects precision rate during screw insertion.31,38132133135137 The basis for decreased technical accuracy in the clinical operative platforms is a result of combined, cumulative errors secondary to the intraoperative workflow process, number of kinematic linkages and variability in patient spinal morphology and flexibility (Figure 8).

Figure 8.

Figure 8.

Clinical platform for optoelectronic data transference process. Schematic illustration demonstrating the operative clinical workflow and process for data transference utilizing optoelectronic tracking. Unique to robotic assisted spinal surgery and in difference to the laboratory setting, the intra-operative process requires considerably more steps in the transference of optoelectronic kinematic data. This complex workflow process integrates correlation and mapping algorithms to register the physical patient to the virtual patient via the navigation system, optoelectronic source, surveillance markers, patient reference markers, end effector instruments in the operative field, and patient CT images. Accurate, close-to-ideal reference reproducibility and maintenance of this dataset is the primary intra-operative objective and challenge.

Discussion

In reviewing the intrinsic technical accuracy and registration accuracy, there exists a substantial burden of proof that the potential performance in optoelectronics is nearly identical between the 2 platforms—laboratory versus clinical operative—under static conditions. The downstream difference in optoelectronic technical accuracy and disparity between the 2 platforms is secondary to the dynamic factors unique to each. The laboratory workflow methods and array registration for experimental musculoskeletal kinematic studies are rigid, highly controlled, with limited experimental coordinate transformations between data input/output—reducing error propagation and maximizing optoelectronic accuracy. Unique to robotic-assisted spinal surgery and in difference to the laboratory setting, the dynamic intra-operative process necessitates considerably more steps in the transference of optoelectronic kinematic data. The complex data flow process integrates correlation and mapping algorithms to register the physical patient to the virtual patient via the navigation system, optoelectronic source, surveillance markers, patient reference markers, end effector instruments, and patient CT images. Essentially, this is a comparison of technical accuracy between a rigid, highly controlled setting and a variable environment with multiple data input factors. The collective effect results in an increased potential for error propagation from experimental coordinate transformations, data processing and optoelectronic kinematic linkages in the clinical setting.

Although workflow and patient related factors provide a basis for decreased accuracy in the clinical setting, with differences between planned versus actual final implant position ranging from 1.5 mm to 3 mm, it could be argued with some degree of confidence that these technical inaccuracies are inconsequential and of no clinical significance in anatomic zones permitting such deviation (e.g., L5 pedicle). However, in cases of cervical or thoracic operative procedures with pedicular dimensions less than 3.5 mm diameter and neural structures within 0.8 mm proximity to pedicular cortices, technical errors of 2 mm to 3 mm are significant, and navigational integrity and reliability of the data transformation process are of paramount importance. With this level of technical inaccuracy, there is basis for contraindication in the use of navigation and robotic-assisted spinal surgery depending on the indications presented and extent of confounding factors that may decrease registration accuracy and subsequent technical accuracy of implant placement.

The fundamental technological challenge of navigation and robotic-assisted spinal surgery is the virtual world needs to clearly represent the physical, real time world. Navigational integrity and maintenance of fidelity in the transference of optoelectronic data is the cornerstone of robotic-assisted spinal surgery. Transitioning from the controlled laboratory setting to clinical operative environment requires an increased number of steps in the optoelectronic kinematic chain and potential for error propagation in experimental coordinate transformations. Moreover, intra-operative challenges of array location, system registration, spinal flexibility, anatomic topography and workflow affect navigational integrity and provide a basis for the disparity of optoelectronic accuracy in the clinical environment compared to the controlled laboratory setting. A continuum of decreased accuracy is demonstrated when comparing the optoelectronic camera source itself to application in musculoskeletal platforms, and finally, clinical operative environment. Diligence in the areas of pre-operative planning, source camera and fiducial positioning, system registration and intra-operative process workflow have the potential to improve accuracy and decrease disparity between planned and final implant position.

Acknowledgments

The authors would like to credit Mir Hussain, Nobert Johnson (Globus Medical Corporation), and Dr. Vladimir Sinkov, MD, for their expertise and objective comments in the preparation of this manuscript.

Footnotes

Declaration of Conflicting Interests: The author(s) declared no potential conflicts of interest with respect to the research, authorship, and/or publication of this article.

Funding: The author(s) received no financial support for the research, authorship, and/or publication of this article.

ORCID iD: Bryan W. Cunningham, PhD Inline graphic https://orcid.org/0000-0002-4604-395X

References

  • 1.van der Kruk E, Reijne MM. Accuracy of human motion capture systems for sport applications; state-of-the-art review. Eur J Sport Sci. 2018;18(6):806–819. doi:10.1080/17461391.2018.1463397 [DOI] [PubMed] [Google Scholar]
  • 2.Linke D, Link D, Lames M. Validation of electronic performance and tracking systems EPTS under field conditions. PLoS One. 2018;13(7):e0199519. doi:10.1371/journal.pone.0199519 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 3.Linke D, Link D, Lames M. Football-specific validity of TRACAB’s optical video tracking systems. PLoS One. 2020;15(3):e0230179. doi:10.1371/journal.pone.0230179 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 4.Elliott B, Alderson J. Laboratory versus field testing in cricket bowling: a review of current and past practice in modelling techniques. Sports Biomech. 2007;6(1):99–108. doi:10.1080/14763140601058623 [DOI] [PubMed] [Google Scholar]
  • 5.Lindbeck L, Kjellberg K. Gender differences in lifting technique. Ergonomics. 2001;44(2):202–214. doi:10.1080/00140130120142 [DOI] [PubMed] [Google Scholar]
  • 6.Nimbarte AD, Sun Y, Jaridi M, Hsiao H. Biomechanical loading of the shoulder complex and lumbosacral joints during dynamic cart pushing task. Appl Ergon. 2013;44(5):841–849. doi:10.1016/j.apergo.2013.02.008 [DOI] [PubMed] [Google Scholar]
  • 7.Negrini S, Piovanelli B, Amici C, et al. Trunk motion analysis: a systematic review from a clinical and methodological perspective. Eur J Phys Rehabil Med. 2016;52(4):583–592. [PubMed] [Google Scholar]
  • 8.Bravi M, Gallotta E, Morrone M, et al. Concurrent validity and inter trial reliability of a single inertial measurement unit for spatial-temporal gait parameter analysis in patients with recent total hip or total knee arthroplasty. Gait Posture. 2020;76:175–181. doi:10.1016/j.gaitpost.2019.12.014 [DOI] [PubMed] [Google Scholar]
  • 9.Serrao M, Casali C, Ranavolo A, et al. Use of dynamic movement orthoses to improve gait stability and trunk control in ataxic patients. Eur J Phys Rehabil Med. 2017;53(5):735–743. doi:10.23736/S1973-9087.17.04480-X [DOI] [PubMed] [Google Scholar]
  • 10.Baker R. Gait analysis methods in rehabilitation. J Neuroeng Rehabil. 2006;3(1):4. doi:10.1186/1743-0003-3-4 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 11.Andújar D, Ribeiro Á, Fernández-Quintanilla C, Dorado J. Accuracy and feasibility of optoelectronic sensors for weed mapping in wide row crops. Sensors (Basel). 2011;11(3):2304–2318. doi:10.3390/s110302304 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 12.Colyer SL, Evans M, Cosker DP, Salo AIT. A review of the evolution of vision-based motion analysis and the integration of advanced computer vision methods towards developing a markerless system. Sports Med Open. 2018;4(1):24. doi:10.1186/s40798-018-0139-y [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 13.Cunningham BW, Sponseller PD, Murgatroyd AA, Kikkawa J, Tortolani PJ. A comprehensive biomechanical analysis of sacral alar iliac fixation: an in vitro human cadaveric model. J Neurosurg Spine. 2019;30(3):367–375. doi:10.3171/2018.8.SPINE18328 [DOI] [PubMed] [Google Scholar]
  • 14.Cunningham BW, Mueller KB, Mullinix KP, Sun X, Sandhu FA. Biomechanical analysis of occipitocervical stabilization techniques: emphasis on integrity of osseous structures at the occipital implantation sites. J Neurosurg. 2020;33(2):138–147. doi:10.3171/2020.1.spine191331 [DOI] [PubMed] [Google Scholar]
  • 15.Dall BE, Eden SV, Cho W, et al. Biomechanical analysis of motion following sacroiliac joint fusion using lateral sacroiliac screws with or without lumbosacral instrumented fusion. Clin Biomech (Bristol, Avon). 2019;68:182–189. doi:10.1016/j.clinbiomech.2019.05.025 [DOI] [PubMed] [Google Scholar]
  • 16.Mushlin H, Brooks DM, Olexa J, et al. A biomechanical investigation of the sacroiliac joint in the setting of lumbosacral fusion: impact of pelvic fixation versus sacroiliac joint fixation. J Neurosurg Spine. 2019;31(4):1–6. doi:10.3171/2019.3.SPINE181127 [DOI] [PubMed] [Google Scholar]
  • 17.Oxland TR. Fundamental biomechanics of the spine—what we have learned in the past 25 years and future directions. J Biomech. 2016;49(6):817–832. doi:10.1016/j.jbiomech.2015.10.035 [DOI] [PubMed] [Google Scholar]
  • 18.La Barbera L, Wilke HJ, Liebsch C, et al. Biomechanical in vitro comparison between anterior column realignment and pedicle subtraction osteotomy for severe sagittal imbalance correction. Eur Spine J. 2020;29(1):36–44. doi:10.1007/s00586-019-06087-x [DOI] [PubMed] [Google Scholar]
  • 19.La Barbera L, Wilke HJ, Ruspi ML, et al. Load-sharing biomechanics of lumbar fixation and fusion with pedicle subtraction osteotomy. Sci Rep. 2021;11(1):3595. doi:10.1038/s41598-021-83251-8 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 20.Camino Willhuber G, Zderic I, Gras F, et al. Analysis of sacro-iliac joint screw fixation: does quality of reduction and screw orientation influence joint stability? A biomechanical study. Int Orthop. 2016;40(7):1537–1543. doi:10.1007/s00264-015-3007-0 [DOI] [PubMed] [Google Scholar]
  • 21.Wang JQ, Wang Y, Feng Y, et al. Percutaneous sacroiliac screw placement: a prospective randomized comparison of robot-assisted navigation procedures with a conventional technique. Chin Med J (Engl). 2017;130(21):2527–2534. doi:10.4103/0366-6999.217080 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 22.Wang M, Li D, Shang X, Wang J. A review of computer-assisted orthopaedic surgery systems. Int J Med Robot. 2020;16(5):1–28. doi:10.1002/rcs.2118 [DOI] [PubMed] [Google Scholar]
  • 23.D’Souza M, Gendreau J, Feng A, Kim LH, Ho AL, Veeravagu A. Robotic-assisted spine surgery: history, efficacy, cost, and future trends. Robot Surg. 2019;6:9–23. doi:10.2147/RSRR.S190720 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 24.Staub BN, Sadrameli SS. The use of robotics in minimally invasive spine surgery. J Spine Surg. 2019;5(Suppl 1):S31–S40. doi:10.21037/jss.2019.04.16 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 25.Kim HJ, Jung WI, Chang BS, Lee CK, Kang KT, Yeom JS. A prospective, randomized, controlled trial of robot-assisted vs freehand pedicle screw fixation in spine surgery. Int J Med Robot. 2017;13(3):e1779. doi:10.1002/rcs.1779 [DOI] [PubMed] [Google Scholar]
  • 26.Han X, Tian W, Liu Y, et al. Safety and accuracy of robot-assisted versus fluoroscopy-assisted pedicle screw insertion in thoracolumbar spinal surgery: a prospective randomized controlled trial. J Neurosurg Spine. 2019:1–8. doi:10.3171/2018.10.SPINE18487 [DOI] [PubMed] [Google Scholar]
  • 27.Ahern DP, Gibbons D, Schroeder GD, Vaccaro AR, Butler JS. Image-guidance, robotics, and the future of spine surgery. Clin Spine Surg. 2020;33(5):179–184. doi:10.1097/BSD.0000000000000809 [DOI] [PubMed] [Google Scholar]
  • 28.Lieber AM, Kirchner GJ, Kerbel YE, Khalsa AS. Robotic-assisted pedicle screw placement fails to reduce overall postoperative complications in fusion surgery. Spine J. 2019;19(2):212–217. doi:10.1016/j.spinee.2018.07.004 [DOI] [PubMed] [Google Scholar]
  • 29.Sayari AJ, Pardo C, Basques BA, Colman MW. Review of robotic-assisted surgery: what the future looks like through a spine oncology lens. Ann Transl Med. 2019;7(10):224. doi:10.21037/atm.2019.04.69 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 30.Lieberman IH, Togawa D, Kayanja MM, et al. Bone-mounted miniature robotic guidance for pedicle screw and translaminar facet screw placement: part I—technical development and a test case result. Neurosurgery. 2006;59(3):641–650. Discussion 641-50. doi:10.1227/01.neu.0000229055.00829.5b [DOI] [PubMed] [Google Scholar]
  • 31.Lieberman IH, Kisinde S, Hesselbacher S. Robotic-assisted pedicle screw placement during spine surgery. JBJS Essent Surg Tech. 2020;10(2):e0020. doi:10.2106/JBJS.ST.19.00020 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 32.Shoham M, Lieberman IH, Benzel EC, et al. Robotic assisted spinal surgery—from concept to clinical practice. Comput Aided Surg. 2007;12(2):105–115. doi:10.3109/10929080701243981 [DOI] [PubMed] [Google Scholar]
  • 33.Ghasem A, Sharma A, Greif DN, Alam M, Maaieh MA. The arrival of robotics in spine surgery: a review of the literature. Spine (Phila Pa 1976). 2018;43(23):1670–1677. doi:10.1097/BRS.0000000000002695 [DOI] [PubMed] [Google Scholar]
  • 34.Spörri J, Schiefermüller C, Müller E. Collecting kinematic data on a ski track with optoelectronic stereophotogrammetry: a methodological study assessing the feasibility of bringing the biomechanics lab to the field. PLoS One. 2016;11(8):e0161757. doi:10.1371/journal.pone.0161757 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 35.Richards JG.The measurement of human motion: a comparison of commercially available systems. Hum Mov Sci. 1999;18(5):589–602. doi:10.1016/s0167-9457(99)00023-8 [Google Scholar]
  • 36.Topley M, Richards JG. A comparison of currently available optoelectronic motion capture systems. J Biomech. 2020;106:109820. doi:10.1016/j.jbiomech.2020.109820 [DOI] [PubMed] [Google Scholar]
  • 37.Kantelhardt SR, Martinez R, Baerwinkel S, Burger R, Giese A, Rohde V. Perioperative course and accuracy of screw positioning in conventional, open robotic-guided and percutaneous robotic-guided, pedicle screw placement. Eur Spine J. 2011;20(6):860–868. doi:10.1007/s00586-011-1729-2 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 38.Lonjon N, Chan-Seng E, Costalat V, Bonnafoux B, Vassal M, Boetto J. Robot-assisted spine surgery: feasibility study through a prospective case-matched analysis. Eur Spine J. 2016;25(3):947–955. doi:10.1007/s00586-015-3758-8 [DOI] [PubMed] [Google Scholar]
  • 39.Schatlo B, Molliqaj G, Cuvinciuc V, Kotowski M, Schaller K, Tessitore E. Safety and accuracy of robot-assisted versus fluoroscopy-guided pedicle screw insertion for degenerative diseases of the lumbar spine: a matched cohort comparison. J Neurosurg Spine. 2014;20(6):636–643. doi:10.3171/2014.3.SPINE13714 [DOI] [PubMed] [Google Scholar]
  • 40.Schatlo B, Martinez R, Alaid A, et al. Unskilled unawareness and the learning curve in robotic spine surgery. Acta Neurochir (Wien). 2015;157(10):1819–1823. doi:10.1007/s00701-015-2535-0 [DOI] [PubMed] [Google Scholar]
  • 41.Solomiichuk V, Fleischhammer J, Molliqaj G, et al. Robotic versus fluoroscopy-guided pedicle screw insertion for metastatic spinal disease: a matched-cohort comparison. Neurosurg Focus. 2017;42(5):E13. doi:10.3171/2017.3.FOCUS1710 [DOI] [PubMed] [Google Scholar]
  • 42.Roser F, Tatagiba M, Maier G. Spinal robotics: current applications and future perspectives. Neurosurgery. 2013;72(Suppl 1):12–18. doi:10.1227/NEU.0b013e318270d02c [DOI] [PubMed] [Google Scholar]
  • 43.Vadalà G, De Salvatore S, Ambrosio L, Russo F, Papalia R, Denaro V. Robotic spine surgery and augmented reality systems: a state of the art. Neurospine. 2020;17(1):88–100. doi:10.14245/ns.2040060.030 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 44.Zhang M, Wu B, Ye C, et al. Multiple instruments motion trajectory tracking in optical surgical navigation. Opt Express. 2019;27(11):15827–15845. doi:10.1364/OE.27.015827 [DOI] [PubMed] [Google Scholar]
  • 45.Zhang Q, Han XG, Xu YF, et al. Robotic navigation during spine surgery. Expert Rev Med Devices. 2020;17(1):27–32. doi:10.1080/17434440.2020.1699405 [DOI] [PubMed] [Google Scholar]
  • 46.Pechlivanis I, Kiriyanthan G, Engelhardt M, et al. Percutaneous placement of pedicle screws in the lumbar spine using a bone mounted miniature robotic system: first experiences and accuracy of screw placement. Spine (Phila Pa 1976). 2009;34(4):392–398. doi:10.1097/BRS.0b013e318191ed32 [DOI] [PubMed] [Google Scholar]
  • 47.Feng S, Tian W, Sun Y, Liu Y, Wei Y. Effect of robot-assisted surgery on lumbar pedicle screw internal fixation in patients with osteoporosis. World Neurosurg. 2019;125:e1057–e1062. doi:10.1016/j.wneu.2019.01.243 [DOI] [PubMed] [Google Scholar]
  • 48.Euler Angles. http://encyclopediaofmath.org/index.php?title=Euler_angles&oldid=34483 1748; Accessed April 19, 2021
  • 49.Wu G, Siegler S, Allard P, et al. ISB recommendation on definitions of joint coordinate system of various joints for the reporting of human joint motion—part I: ankle, hip, and spine. International Society of Biomechanics. J Biomech. 2002;35(4):543–548. doi:10.1016/s0021-9290(01)00222-6 [DOI] [PubMed] [Google Scholar]
  • 50.Euler L. Formulae generales pro translatione quacunque corporum rigidorum. Novi Commentari Academiae Scientiarum Imperalis Petropolitanae. 1775;20:189–207. [Google Scholar]
  • 51.Euler L. Nova methodus motum corporum rigidorum determinandi. Novi Commentari Academiae Scientiarum Imperalis Petropolitanae. 1775;20:208–238. [Google Scholar]
  • 52.Saha SK. Denavit and Hartenberg (DH) Parameters. Introduction to Robotics. 2nd ed. McGraw Hill Education; 2014. [Google Scholar]
  • 53.White AA, Panjabi MM. Clinical Biomechanics of the Spine. 2nd ed. J. B. Lippincott; 1990. [Google Scholar]
  • 54.Panjabi MM. Biomechanical evaluation of spinal fixation devices: I. A conceptual framework. Spine (Phila Pa 1976). 1988;13(10):1129–1134. doi:10.1097/00007632-198810000-00013 [DOI] [PubMed] [Google Scholar]
  • 55.Panjabi MM, Abumi K, Duranceau J, Crisco JJ. Biomechanical evaluation of spinal fixation devices: II. Stability provided by eight internal fixation devices. Spine (Phila Pa 1976). 1988;13(10):1135–1140. doi:10.1097/00007632-198810000-00014 [DOI] [PubMed] [Google Scholar]
  • 56.ISO 5725-1. Accuracy (Trueness and Precision) of Measurement Methods and Results. Part 1: General Principles and Definitions. 1994. [Google Scholar]
  • 57.Elfring R, de la Fuente M, Radermacher K. Assessment of optical localizer accuracy for computer aided surgery systems. Comput Aided Surg. 2010;15(1-3):1–12. doi:10.3109/10929081003647239 [DOI] [PubMed] [Google Scholar]
  • 58.Haidegger T, Kazanzides P, Rudas I, Benyó B, Benyó Z. The Importance of Accuracy Measurement Standards for Computer-Integrated Interventional Systems. IEEE; 2010:1–6. [Google Scholar]
  • 59.Corazza S, Mündermann L, Gambaretto E, Ferrigno G, Andriacchi TP. Markerless motion capture through visual hull, articulated ICP and subject specific model generation. Int J Comp Vision. 2010;87(1-2):156–169. doi:10.1007/s11263-009-0284-3 [Google Scholar]
  • 60.Simon D, O’Toole R, Blackwell M, Morgan F, DiGioia A, Kanade T. Accuracy validation in image-guided orthopaedic surgery. In: Proceedings of 2nd International Symposium on Medical Robotics and Computer Assisted Surgery. The Robotics Institute Carnegie Mellon University; 1995:185–192. [Google Scholar]
  • 61.Sielhorst T, Bauer M, Wenisch O, Klinker G, Navab N. Online estimation of the target registration error for n-ocular optical tracking systems. Med Image Comput Comput Assist Interv. 2007;10(Pt 2):652–659. doi:10.1007/978-3-540-75759-7_79 [DOI] [PubMed] [Google Scholar]
  • 62.Koivukangas T, Katisko JP, Koivukangas JP. Technical accuracy of optical and the electromagnetic tracking systems. Springerplus. 2013;2(1):90. doi:10.1186/2193-1801-2-90 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 63.Crouch DG, Kehl L, Krist JR. OPTOTRAK: at last a system with resolution of 10 μm. In: Proceedings Volume 1356, Image-Based Motion Measurement. Image-Based Motion Measurement; 1990. [Google Scholar]
  • 64.Khadem R, Yeh CC, Sadeghi-Tehrani M, et al. Comparative tracking error analysis of five different optical tracking systems. Comput Aided Surg. 2000;5(2):98–107. doi:10.1002/1097-0150(2000)5:2<98:: AID-IGS4>3.0.CO;2-H [DOI] [PubMed] [Google Scholar]
  • 65.ASTM E177-20. Standard Practice for Use of the Terms Precision and Bias in ASTM Test Methods. ASTM International; 2002. [Google Scholar]
  • 66.Maletsky LP, Sun J, Morton NA. Accuracy of an optical active-marker system to track the relative motion of rigid bodies. J Biomech. 2007;40(3):682–685. doi:10.1016/j.jbiomech.2006.01.017 [DOI] [PubMed] [Google Scholar]
  • 67.Grunert P, Darabi K, Espinosa J, Filippi R. Computer-aided navigation in neurosurgery. Neurosurg Rev. 2003;26(2):73–99. Discussion 100-1. doi:10.1007/s10143-003-0262-0 [DOI] [PubMed] [Google Scholar]
  • 68.Simoes R, Raposo C, Barreto J, Edwards P, Stoyanov D. Visual tracking vs optical tracking in computer-assisted intervention. IEEE Trans Biomed Eng. 2018. [Google Scholar]
  • 69.Chassat F, Lavallée S. Experimental protocol of accuracy evaluation of 6-D localizers for computer-integrated surgery: application to four optical localizers. In: Wells WM, Colchester A, Delp S, eds. Medical Image Computing and Computer-Assisted Intervention—MICCAI’98. 1998:277–284. Berlin, Heidelberg. doi:10.1007/bfb0056211 [Google Scholar]
  • 70.Citak M, Kendoff D, Wanich T, et al. The influence of distance on registration in ISO-C-3D navigation: a source of error in ISO-C-3D navigation. Technol Health Care. 2006;14(6):473–478. [PubMed] [Google Scholar]
  • 71.Fitzpatrick JM, West JB, Maurer CR. Predicting error in rigid-body point-based registration. IEEE Trans Med Imaging. 1998;17(5):694–702. doi:10.1109/42.736021 [DOI] [PubMed] [Google Scholar]
  • 72.Wiles AD, Thompson DG, Frantz DD. Accuracy Assessment and Interpretation for Optical Tracking Systems. SPIE; 2004:421. [Google Scholar]
  • 73.Fitzpatrick JM, West JB. The distribution of target registration error in rigid-body point-based registration. IEEE Trans Med Imaging. 2001;20(9):917–927. doi:10.1109/42.952729 [DOI] [PubMed] [Google Scholar]
  • 74.Windolf M, Götzen N, Morlock M. Systematic accuracy and precision analysis of video motion capturing systems—exemplified on the vicon-460 system. J Biomech. 2008-08-01. 2008;41(12):2776–2780. doi:10.1016/j.jbiomech.2008.06.024 [DOI] [PubMed] [Google Scholar]
  • 75.States RA, Pappas E. Precision and repeatability of the optotrak 3020 motion measurement system. J Med Eng Technol. 2006;30(1):11–16. doi:10.1080/03091900512331304556 [DOI] [PubMed] [Google Scholar]
  • 76.Min Z, Zhu D, Meng MQH. Accuracy Assessment of an N-ocular Motion Capture System for Surgical Tool Tip Tracking Using Pivot Calibration. IEEE; 2016. [Google Scholar]
  • 77.Li B, Zhang L, Sun H, Yuan J, Shen SG, Wang X. A novel method of computer aided orthognathic surgery using individual CAD/CAM templates: a combination of osteotomy and repositioning guides. Br J Oral Maxillofac Surg. 2013;51(8):e239–e244. doi:10.1016/j.bjoms.2013.03.007 [DOI] [PubMed] [Google Scholar]
  • 78.Marinetto E, Garcia-Mato D, Garcia A, Martinez S, Desco M, Pascau J. Multicamera optical tracker assessment for computer aided surgery applications. IEEE Access. 2018;6:64359–64370. doi:10.1109/access.2018.2878323 [Google Scholar]
  • 79.Gédet P, Thistlethwaite PA, Ferguson SJ. Minimizing errors during in vitro testing of multisegmental spine specimens: considerations for component selection and kinematic measurement. J Biomech. 2007;40(8):1881–1885. doi:10.1016/j.jbiomech.2006.07.024 [DOI] [PubMed] [Google Scholar]
  • 80.Citak M, Kendoff D, Wanich T, et al. The influence of metal artifacts on navigation and the reduction of artifacts by the use of polyether-ether-ketone. Comput Aided Surg. 2008;13(4):233–239. doi:10.3109/10929080802215292 [DOI] [PubMed] [Google Scholar]
  • 81.Stancic I, Grujic Supuk T, Panjkota A. Design, development and evaluation of optical motion-tracking system based on active white light markers. IET Sci Meas Technol. 2013;7(4):206–214. doi:10.1049/iet-smt.2012.0157 [Google Scholar]
  • 82.Brinker T, Arango G, Kaminsky J, et al. An experimental approach to image guided skull base surgery employing a microscope-based neuronavigation system. Acta Neurochir (Wien). 1998;140(9):883–889. doi:10.1007/s007010050189 [DOI] [PubMed] [Google Scholar]
  • 83.Dorward NL, Alberti O, Palmer JD, Kitchen ND, Thomas DGT. Accuracy of true frameless stereotaxy: in vivo measurement and laboratory phantom studies. J Neurosurg. 1999;90(1):160–168. doi:10.3171/jns.1999.90.1.0160 [DOI] [PubMed] [Google Scholar]
  • 84.Maurer CR, Maciunas RJ, Fitzpatrick JM. Registration of head CT images to physical space using a weighted combination of points and surfaces. IEEE Trans Med Imaging. 1998;17(5):753–761. doi:10.1109/42.736031 [DOI] [PubMed] [Google Scholar]
  • 85.Wilke H, Fischer K, Jeanneret B, Claes L, Magerl F. In-vivo-Messung der dreidimensionalen bewegung des iliosakralgelenks. Zeitschrift Fur Orthopadie Und Ihre Grenzgebiet. 2008;135(6):550–556. [DOI] [PubMed] [Google Scholar]
  • 86.Bowden AE, Guerin HL, Villarraga ML, Patwardhan AG, Ochoa JA. Quality of motion considerations in numerical analysis of motion restoring implants of the spine. Clin Biomech (Bristol, Avon). 2008;23(5):536–544. doi:10.1016/j.clinbiomech.2007.12.010 [DOI] [PubMed] [Google Scholar]
  • 87.Ilharreborde B, Zhao K, Boumediene E, Gay R, Berglund L, An KN. A dynamic method for in vitro multisegment spine testing. Orthop Traumatol Surg Res. 2010;96(4):456–461. doi:10.1016/j.otsr.2010.01.006 [DOI] [PubMed] [Google Scholar]
  • 88.Jeong JH, Leasure JM, Park J. Assessment of biomechanical changes after sacroiliac joint fusion by application of the 3-dimensional motion analysis technique. World Neurosurg. 2018;117:e538–e543. doi:10.1016/j.wneu.2018.06.072 [DOI] [PubMed] [Google Scholar]
  • 89.Osterhoff G, Dodd AE, Unno F, et al. Cement augmentation in sacroiliac screw fixation offers modest biomechanical advantages in a cadaver model. Clin Orthop Relat Res. 2016;474(11):2522–2530. doi:10.1007/s11999-016-4934-9 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 90.Osterhoff G, Tiziani S, Hafner C, Ferguson SJ, Simmen HP, Werner CM. Symphyseal internal rod fixation versus standard plate fixation for open book pelvic ring injuries: a biomechanical study. Eur J Trauma Emerg Surg. 2016;42(2):197–202. doi:10.1007/s00068-015-0529-5 [DOI] [PubMed] [Google Scholar]
  • 91.Hammer N, Scholze M, Kibsgård T, et al. Physiological in vitro sacroiliac joint motion: a study on three-dimensional posterior pelvic ring kinematics. J Anat. 2019;234(3):346–358. doi:10.1111/joa.12924 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 92.Baria D, Lindsey RW, Milne EL, Kaimrajh DN, Latta LL. Effects of lumbosacral arthrodesis on the biomechanics of the sacroiliac joint. JB JS Open Access. 2020;5(1):e0034. doi:10.2106/JBJS.OA.19.00034 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 93.Devito DP, Kaplan L, Dietl R, et al. Clinical acceptance and accuracy assessment of spinal implants guided with SpineAssist surgical robot: retrospective study. Spine (Phila Pa 1976). 2010;35(24):2109–2115. doi:10.1097/BRS.0b013e3181d323ab [DOI] [PubMed] [Google Scholar]
  • 94.Helm PA, Teichman R, Hartmann SL, Simon D. Spinal navigation and imaging: history, trends, and future. IEEE Trans Med Imaging. 2015;34(8):1738–1746. doi:10.1109/TMI.2015.2391200 [DOI] [PubMed] [Google Scholar]
  • 95.Ringel F, Stüer C, Reinke A, et al. Accuracy of robot-assisted placement of lumbar and sacral pedicle screws: a prospective randomized comparison to conventional freehand screw implantation. Spine (Phila Pa 1976). 2012;37(8):E496–E501. doi:10.1097/BRS.0b013e31824b7767 [DOI] [PubMed] [Google Scholar]
  • 96.Fan Y, Du JP, Liu JJ, et al. Accuracy of pedicle screw placement comparing robot-assisted technology and the free-hand with fluoroscopy-guided method in spine surgery: an updated meta-analysis. Medicine (Baltimore). 2018;97(22):e10970. doi:10.1097/md.0000000000010970 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 97.Fan M, Liu Y, He D, et al. Improved accuracy of cervical spinal surgery with robot-assisted screw insertion: a prospective, randomized, controlled study. Spine (Phila Pa 1976). 2020;45(5):285–291. doi:10.1097/BRS.0000000000003258 [DOI] [PubMed] [Google Scholar]
  • 98.Huang M, Tetreault TA, Vaishnav A, York PJ, Staub BN. The current state of navigation in robotic spine surgery. Ann Transl Med. 2021;9(1):86. doi:10.21037/atm-2020-ioi-07 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 99.Kalidindi KKV, Sharma JK, Jagadeesh NH, Sath S, Chhabra HS. Robotic spine surgery: a review of the present status. J Med Eng Technol. 2020;44(7):431–437. doi:10.1080/03091902.2020.1799098 [DOI] [PubMed] [Google Scholar]
  • 100.Elswick CM, Strong MJ, Joseph JR, Saadeh Y, Oppenlander M, Park P. Robotic-assisted spinal surgery: current generation instrumentation and new applications. Neurosurg Clin N Am. 2020;31(1):103–110. doi:10.1016/j.nec.2019.08.012 [DOI] [PubMed] [Google Scholar]
  • 101.Overley SC, Cho SK, Mehta AI, Arnold PM. Navigation and robotics in spinal surgery: where are we now? Neurosurgery. 2017;80(3S):S86–S99. doi:10.1093/neuros/nyw077 [DOI] [PubMed] [Google Scholar]
  • 102.Yu X, Xu L, Bi LY. Spinal navigation with intra-operative 3D-imaging modality in lumbar pedicle screw fixation. Zhonghua Yi Xue Za Zhi. 2008;88(27):1905–1908. [PubMed] [Google Scholar]
  • 103.Yu CC, Bajwa NS, Toy JO, Ahn UM, Ahn NU. Pedicle morphometry of upper thoracic vertebrae: an anatomic study of 503 cadaveric specimens. Spine (Phila Pa 1976). 2014;39(20):E1201–E1209. doi:10.1097/BRS.0000000000000505 [DOI] [PubMed] [Google Scholar]
  • 104.Yu L, Chen X, Margalit A, Peng H, Qiu G, Qian W. Robot-assisted vs freehand pedicle screw fixation in spine surgery—a systematic review and a meta-analysis of comparative studies. Int J Med Robot. 2018;14(3):e1892. doi:10.1002/rcs.1892 [DOI] [PubMed] [Google Scholar]
  • 105.Vardiman AB, Wallace DJ, Booher GA, et al. Does the accuracy of pedicle screw placement differ between the attending surgeon and resident in navigated robotic-assisted minimally invasive spine surgery? J Robot Surg. 2020;14(4):567–572. doi:10.1007/s11701-019-01019-9 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 106.Vardiman AB, Wallace DJ, Crawford NR, Riggleman JR, Ahrendtsen LA, Ledonio CG. Pedicle screw accuracy in clinical utilization of minimally invasive navigated robot-assisted spine surgery. J Robot Surg. 2020;14(3):409–413. doi:10.1007/s11701-019-00994-3 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 107.Keric N, Doenitz C, Haj A, et al. Evaluation of robot-guided minimally invasive implantation of 2067 pedicle screws. Neurosurg Focus. 2017;42(5):E11. doi:10.3171/2017.2.FOCUS16552 [DOI] [PubMed] [Google Scholar]
  • 108.Chen HY, Xiao XY, Chen CW, et al. Results of using robotic-assisted navigational system in pedicle screw placement. PLoS One. 2019;14(8):e0220851. doi:10.1371/journal.pone.0220851 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 109.Chen HY, Xiao XY, Chen CW, et al. A spine robotic-assisted navigation system for pedicle screw placement. J Vis Exp. 2020;(159). doi:10.3791/60924 [DOI] [PubMed] [Google Scholar]
  • 110.Chen L, Zhang F, Zhan W, Gan M, Sun L. Research on the accuracy of three-dimensional localization and navigation in robot-assisted spine surgery. Int J Med Robot. 2020;16(2):e2071. doi:10.1002/rcs.2071 [DOI] [PubMed] [Google Scholar]
  • 111.Togawa D, Kayanja MM, Reinhardt MK, et al. Bone-mounted miniature robotic guidance for pedicle screw and translaminar facet screw placement: part 2—evaluation of system accuracy. Neurosurgery. 2007;60(2 Suppl 1):ONS129–139. Discussion ONS139. doi:10.1227/01.NEU.0000249257.16912.AA [DOI] [PubMed] [Google Scholar]
  • 112.Macke JJ, Woo R, Varich L. Accuracy of robot-assisted pedicle screw placement for adolescent idiopathic scoliosis in the pediatric population. J Robot Surg. 2016;10(2):145–150. doi:10.1007/s11701-016-0587-7 [DOI] [PubMed] [Google Scholar]
  • 113.Fu W, Tong J, Liu G, et al. Robot-assisted technique vs conventional freehand technique in spine surgery: a meta-analysis. Int J Clin Pract. 2020;75(5):e13964. doi:10.1111/ijcp.13964 [DOI] [PubMed] [Google Scholar]
  • 114.Huntsman KT, Riggleman JR, Ahrendtsen LA, Ledonio CG. Navigated robot-guided pedicle screws placed successfully in single-position lateral lumbar interbody fusion. J Robot Surg. 2020;14(4):643–647. doi:10.1007/s11701-019-01034-w [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 115.Johnson N. Imaging, navigation, and robotics in spine surgery. Spine (Phila Pa 1976). 2016;41(Suppl 7):S32. doi:10.1097/BRS.0000000000001437 [DOI] [PubMed] [Google Scholar]
  • 116.Kochanski RB, Lombardi JM, Laratta JL, Lehman RA, O’Toole JE. Image-guided navigation and robotics in spine surgery. Neurosurgery. 2019;84(6):1179–1189. doi:10.1093/neuros/nyy630 [DOI] [PubMed] [Google Scholar]
  • 117.Kotani Y, Abumi K, Ito M, Minami A. Improved accuracy of computer-assisted cervical pedicle screw insertion. J Neurosurg. 2003;99(3 Suppl):257–263. doi:10.3171/spi.2003.99.3.0257 [DOI] [PubMed] [Google Scholar]
  • 118.Le X, Tian W, Shi Z, et al. Robot-assisted versus fluoroscopy-assisted cortical bone trajectory screw instrumentation in lumbar spinal surgery: a matched-cohort comparison. World Neurosurg. 2018;120:e745–e751. doi:10.1016/j.wneu.2018.08.157 [DOI] [PubMed] [Google Scholar]
  • 119.Sukovich W, Brink-Danan S, Hardenbrook M. Miniature robotic guidance for pedicle screw placement in posterior spinal fusion: early clinical experience with the SpineAssist. Int J Med Robot. 2006;2(2):114–122. doi:10.1002/rcs.86 [DOI] [PubMed] [Google Scholar]
  • 120.Vo CD, Jiang B, Azad TD, Crawford NR, Bydon A, Theodore N.Robotic spine surgery: current state in minimally invasive surgery. Global Spine J. 2020;10(2 Suppl):34S–40S. doi:10.1177/2192568219878131 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 121.Wallace DJ, Vardiman AB, Booher GA, et al. Navigated robotic assistance improves pedicle screw accuracy in minimally invasive surgery of the lumbosacral spine: 600 pedicle screws in a single institution. Int J Med Robot. 2020;16(1):e2054. doi:10.1002/rcs.2054 [DOI] [PubMed] [Google Scholar]
  • 122.West JB, Fitzpatrick JM, Toms SA, Maurer CR, Maciunas RJ. Fiducial point placement and the accuracy of point-based, rigid body registration. Neurosurgery. 2001;48(4):810–816. Discussion 816-7. doi:10.1097/00006123-200104000-00023 [DOI] [PubMed] [Google Scholar]
  • 123.West JB, Maurer CR. Designing optically tracked instruments for image-guided surgery. IEEE Trans Med Imaging. 2004;23(5):533–545. doi:10.1109/tmi.2004.825614 [DOI] [PubMed] [Google Scholar]
  • 124.Gertzbein SD, Robbins SE. Accuracy of pedicular screw placement in vivo. Spine (Phila Pa 1976). 1990;15(1):11–14. doi:10.1097/00007632-199001000-00004 [DOI] [PubMed] [Google Scholar]
  • 125.Verma V, Santoshi JA, Jain V, et al. Thoracic pedicle morphometry of dry vertebral columns in relation to trans-pedicular fixation: a cross-sectional study from central India. Cureus. 2020;12(5):e8148. doi:10.7759/cureus.8148 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 126.Holly LT. Image-guided spinal surgery. Int J Med Robot. 2006;2(1):7–15. doi:10.1002/rcs.69 [DOI] [PubMed] [Google Scholar]
  • 127.Gupta R, Kapoor K, Sharma A, Kochhar S, Garg R. Morphometry of typical cervical vertebrae on dry bones and CT scan and its implications in transpedicular screw placement surgery. Surg Radiol Anat. 2013;35(3):181–189. doi:10.1007/s00276-012-1013-0 [DOI] [PubMed] [Google Scholar]
  • 128.Herrero CF, Luis do Nascimento A, Maranho DAC, et al. Cervical pedicle morphometry in a Latin American population: a Brazilian study. Medicine (Baltimore). 2016;95(25):e3947. doi:10.1097/MD.0000000000003947 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 129.Attar A, Ugur HC, Uz A, Tekdemir I, Egemen N, Genc Y. Lumbar pedicle: surgical anatomic evaluation and relationships. Eur Spine J. 2001;10(1):10–15. doi:10.1007/s005860000198 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 130.Avrunin OG, Alkhorayef M, Farouk Ismail Saied H, Tymkovych MY. The surgical navigation system with optical position determination technology and sources of errors. J Med Imaging Health Inform. 2015;5(4):689–696. doi:10.1166/jmihi.2015.1444 [Google Scholar]
  • 131.Garg B, Mehta N, Malhotra R. Robotic spine surgery: ushering in a new era. J Clin Orthop Trauma. 2020;11(5):753–760. doi:10.1016/j.jcot.2020.04.034 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 132.Hu X, Ohnmeiss DD, Lieberman IH. Robotic-assisted pedicle screw placement: lessons learned from the first 102 patients. Eur Spine J. 2013;22(3):661–666. doi:10.1007/s00586-012-2499-1 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 133.Hu X, Lieberman IH. What is the learning curve for robotic-assisted pedicle screw placement in spine surgery? Clin Orthop Relat Res. 2014;472(6):1839–1844. doi:10.1007/s11999-013-3291-1 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 134.Abbasi HR, Grzeszczuk R, Chin S, et al. Clinical fluoroscopic fiducial-based registration of the vertebral body in spinal neuronavigation. Stud Health Technol Inform. 2001;81:1–7. [PubMed] [Google Scholar]
  • 135.Hüfner T, Geerling J, Oldag G, et al. Accuracy study of computer-assisted drilling: the effect of bone density, drill bit characteristics, and use of a mechanical guide. J Orthop Trauma. 2005;19(5):317–322. [PubMed] [Google Scholar]
  • 136.Messmer P, Gross T, Suhm N, Regazzoni P, Jacob AL, Huegli RW. Modality-based navigation. Injury. 2004;35(Suppl 1):S–A24–A29. doi:10.1016/j.injury.2004.05.007 [DOI] [PubMed] [Google Scholar]
  • 137.Rahmathulla G, Nottmeier EW, Pirris SM, Deen HG, Pichelmann MA. Intraoperative image-guided spinal navigation: technical pitfalls and their avoidance. Neurosurg Focus. 2014;36(3):E3. doi:10.3171/2014.1.focus13516 [DOI] [PubMed] [Google Scholar]

Articles from Global Spine Journal are provided here courtesy of SAGE Publications

RESOURCES