Published in final edited form as: Phys Med Biol. 2022 Mar 29;67(7):10.1088/1361-6560/ac5c8f. doi: 10.1088/1361-6560/ac5c8f

Dosimetry of a Novel Focused Monoenergetic Beam for Radiotherapy

Jamshid Moradi-Kurdestany 1, Dirk A Bartkoski 1, Ramesh Tailor 1, Dragan Mirkovic 1, Ze’ev Harel 2, Aharon Bar-David 2, Michael Kleckner 2, Shirly Borukhin 2, Mohammad Salehpour 1
PMCID: PMC9020448  NIHMSID: NIHMS1793636  PMID: 35272282

Abstract

Objective.

A novel treatment modality is currently being developed that produces converging monoenergetic x-rays. Conventional application of dosimetric calibration as presented in protocol TG61 is not applicable. Furthermore, the dosimetry of the focal point of the converging beam is on the order of a few millimeters, requiring a high-resolution dosimeter. Here we present a procedure to calibrate radiochromic film for narrow-beam monoenergetic 60 keV photons as well as absolute dosimetry of monoenergetic focused x-rays. A study of the focal spot dose rate after passing through a bone-equivalent material was also done to quantify the effects of heterogeneous materials.

Approach.

This was accomplished by configuring a polyenergetic beam of equivalent energy using a clinical orthovoltage machine. Calibrated films were then used to perform absolute dosimetry of the converging beam by measuring the beam profile at various depths in water.

Main Results.

A method for calibrating radiochromic film has been developed and detailed that allows absolute dosimetry of a monoenergetic photon beam. Absolute dosimetry of a focused, mono-energetic beam resulted in a focal spot dose rate of ~30 cGy/min at a depth of 5 cm in water.

Significance.

This work serves to establish a dosimetry protocol for mono-energetic beam absolute dosimetry as well as the use of such a method for measurement of a novel teletherapy modality.

Keywords: Gafchromic Film Dosimetry, HVL, Energy Response, Orthovoltage Dosimetry

1. Introduction

A novel treatment modality currently in development at The University of Texas MD Anderson Cancer Center in collaboration with Convergent Radiotherapy and Radiosurgery (CRnR) has the potential of providing highly localized dose distributions with enhanced healthy tissue sparing. By using an annular configuration of aluminum single-crystal tiles, a diverging beam of x-rays may be converted to a converging beam with a focal spot size of a few millimeters, resulting in high-intensity localized dose distributions as illustrated in Figure 1.

Figure 1.

Figure 1.

Cross-sectional conceptual representation of a Bragg reflecting lens. A polyenergetic diverging beam, such as produced by an x-ray tube, is reflected from a lens structure using Bragg refraction with the central non-reflecting portion of the beam being blocked. The result is a converging monoenergetic beam with a highly localized dose distribution at the focal point. Figure taken from (Bartkoski et al 2021).

Bragg refraction, the process by which x-rays are reflected from metallic crystals, produces a nearly monoenergetic output spectrum from a polyenergetic input beam. The resulting conically shaped, converging, quasi-monoenergetic 60 keV photon beam, composed primarily of the Kα1 and Kα2 tungsten characteristic x-rays surrounded by an ~8 keV band x-ray energies, is unlike that of conventional teletherapy beams. A detailed description of the CRnR converging beam system including a discussion of physics principles and initial prototype measurements results may be found in (Bartkoski et al 2021).

Robotic manipulation of the delivery system will allow for the focal spot to be “painted” into 3D conformal doses. The unique combination of properties afforded by the CRnR beam make it ideal for tumors adjacent to highly radiosensitive structures such as the spinal column. Head and neck tumors also have the potential to benefit from the converging beam geometry as well as being a tool for stereotactic radiosurgery. Additionally, non-cancer related treatments are being considered such as age-related macular degeneration and denervation which require highly localized radiation delivery.

Physics aspects such as dose calibration, dosimetric measurements, and treatment planning of such a treatment modality as produced by the CRnR system present unique challenges. Dosimetry requirements for this beam include good spatial resolution, low energy dependence, and negligible dose-rate dependence. Radiographic EBT3 film appears to be a good candidate for absolute dosimetry. However, its dose-calibration protocol requires an Accredited Dosimetry Calibration Laboratory (ADCL)-calibrated ion chamber, preferably with low energy dependence. The use of radiochromic films for medical and industrial applications has been established and exhaustively discussed (Butson et al 2003, Williams and Metcalfe 2011, Schwob and Orion 2007, Niroomand-Rad et al 2020, Butson et al 2004). In this work, we describe a method for calibrating radiographic EBT3 film for a narrow-beam geometry, monoenergetic x-ray beam equivalent to the beam produced by the CRnR system. The calibrated films were then used to perform a variety of absolute dose measurements. These absolute dose measurements are used to further analyze the dose deposition characteristics of the CRnR beam. Relative dosimetery analysis of this unique beam has been reported in (Bartkoski et al 2021).

2. Methods

2.1. Equivalent Polyenergetic Beam

Current ADCL measurements do not include dose calibration for monoenergetic beams. Therefore, an equivalent polyenergetic beam was needed. A polyenergetic beam may be substituted for a monoenergetic beam by using the half-value layer (HVL). The HVL is the thickness at which half of the original photon intensity of a monoenergetic beam is attenuated by a material with attenuation coefficient μ and is given by

HVL=ln2μ. (1)

A polyenergetic beam is said to be approximately equivalent to a monoenergetic beam with the same first HVL. Furthermore, for a monoenergetic beam, the first and second half-value layers would be equal but beam hardening in polyenergetic beams result in longer second half-value layers. The homogeneity factor (HF) is the ratio of the first to second half-value layer and indicates how closely a polyenergetic beam approximates a monoenergtic beam with HF = 1. The mass attenuation coefficient of μ/ρ = 0.2778 cm2/g (Hubbell and Seltzer 1995) for aluminum with density ρ = 2.699 g/cm3 for a 60 keV mononenergetic beam results in an HVL of 9.2 mm.

The polyenergetic beam was configured on an Xstrahl 300 orthovoltage unit in narrow-beam geometry using a 180 kVp beam energy and corresponding filter No. 6 as specified by the manufacturer. Filter No. 6, which includes 1.5 mm Al and 0.15 mmCu, was modified with an additional 1 mm Al such that a voltage of 180 kVp at 15 mA produced the necessary attenuation to produce a first HVL = 9.0 mm and a measured HF of 0.76.

The process for calibrating film for a monoenergetic beam is as follows:

  1. An ion chamber is sent to ADCL for calibration at polyenergetic beam energies (E1 and E2) close to the desired beam energy E0.

  2. The ion chamber calibration is interpolated at E0 from E1 and E2.

  3. Using the aforementioned configured polyenergetic beam, a specified amount of radiation is delivered to the ion chamber, and the dose is recorded.

  4. Using the same configuration, the same amount of radiation is delivered to the film, and the film’s optical density (OD) is recorded.

  5. The process is repeated for varying amounts of radiation delivered, and a calibration curve of dose vs. OD is generated.

The film calibration was justified for a monoenergetic beam by repeating the film calibration process using the polyenergetic beam at various neighboring energies around the monoenergetic energy E0. A negligible energy dependence of the dose vs. OD calibration curves justified the usage for a monoenergetic source.

2.2. Dosimeter Selection

With an inherent spatial resolution <25 μm, dose rate dependence < 5%, and an energy dependence < 5%, the near-tissue-equivalent Gafchromic EBT3 film (lot No. 07281401, Ashland Inc., Wayne, NJ, USA) was chosen for absolute dosimetry of the converging beam focal spot. An Epson 10000XL flatbed scanner was employed in transmission mode to read the exposed film pieces. The scanner used 48-bit color with a spatial resolution of 72 dpi and was chosen because the scanner is the limiting factor in contributing to film resolution.

A Farmer-type ion chamber (NEL 2515/3, serial No. 2402) was used as the ADCL-calibrated dosimeter. Ion chamber calibration coefficient Nk were obtained in terms of air kerma from the ADCL for 100, 125, and 250 kVp from which a chamber calibration coefficient for 180 kVp was interpolated (see Table 1).

Table 1.

Beam energy, first half value layer, and ion chamber calibration coefficients NK for three ADCL beams obtained for ion chamber NEL2515/3 as well as the interpolated NK at 180 kVp.

Beam (kVp) HVL (mmAl) ADCL NK (cGy/nC) Interp NK (cGy/nC)
100 4.14 11.84 -
125 6.01 11.87 -
180 9.0 - 12.01
250 17.8 12.42 -

2.3. Film Calibration

EBT3 cross-calibrations were performed for three orthovoltage energies, 125 kVp (20 mA, filter No. 3, HVL 3.5 mm Al), 180 kVp (15 mA, modified filter No. 6, HVL 9.0 mm Al), and 250 kVp (10 mA, filter No. 7, HVL 1.3 mm Cu). Calibration protocol TG61 was employed (Ma et al 2001) to calibrate the beams output for dose to water in air. The following equation from TG-61 gives the absorbed dose to water in air at the phantom surface for low and medium energy × rays

Dw,surface=MNKBwPstem,air[(μ¯enρ)airw]air, (2)

where M is the free-in-air ion chamber reading corrected for temperature, pressure, ion recombination, and electrometer accuracy. For a given beam quality, NK is the ADCL-provided air-kerma calibration coefficient; the chamber stem correction factor Pstem,air was set to unity as the measurement field size was set to the calibration field size, and [(μ¯en/ρ)airw]air is the water-to-air ratio of the mean mass energy-absorption coefficients averaged over the incident photon spectrum. The backscatter factor Bw, which accounts for scatter from the phantom, was neglected due to the EBT3 film pieces being suspended in air on a 0.01 mm thick layer of polyethylene stretched over a 15×15 cm polystyrene foam frame. This allows the calculation of dose to a water medium in air, represented by the suspended water-equivalent EBT3 film.

A sheet of EBT3 was cut into 50 pieces, each 2 cm × 2 cm. Reading an exposed film is known to exhibit angular-dependence due to polarization effects (Butson et al., 2003); therefore, it is important to maintain the grain orientation while scanning film pieces. To achieve this, before cutting the parent film into pieces, ink lines were marked to show 2 cm × 2cm pieces, and ID numbers were marked consistently on bottom right corners. Film pieces were laid on the scanner bed maintaining the ID number orientation in a consistent manner.

Each film piece was irradiated in air to a known dose in reference geometry (field size 10 cm × 10 cm, source-to-film distance 50 cm). Seven dose levels were used, ranging from 0 to 1200 cGy. For each dose level, two film pieces (one at a time) were exposed to serve as repeat data. The same procedure was employed for each of the three beam energies.

2.4. Reading Exposed Films

Exposed EBT3 is known to exhibit an OD that continues to develop over time (Khachonkham et al 2018). Therefore, all exposed pieces were read after 1 day (± 2 hours).

To obtain optical density, each film piece was scanned before and after exposure. Two requirements are needed in order to achieve the highest possible OD precision. First, film grain orientation should be the same for all pieces, which is achieved by maintaining film ID number orientation. Second, a film piece should go to the same physical location on the scanner glass for pre- and post-exposure scans. This is achieved by employing a cardboard template with a cut-out window for placement of the film pieces. Each film ID number sits in the template window same way for pre- and post-exposure scans.

Three repeat scans were acquired for pre- and post-exposed films. All images were saved in TIFF format. TIFF images are read into the ImageJ software (developed at the U.S. National Institutes of Health and available at http://rsb.info.nih.gov/nih-image/) in RGB format, where the red channel is used for analysis. A region of interest (ROI) 0.5 cm × 0.5 cm at the center of each scan was employed to obtain the ROI mean pixel value representing the transmitted light intensity. This value along with the ROI standard deviation was recorded where the standard deviation is used to propagate error.

2.5. Transmitted Intensity to Optical Density Conversion

Using the notations Ipre and Ipost, respectively, for the films’ pre- and post-exposure transmitted light intensity, net optical density NetOD was calculated using the following equation (Devic et al 2004)

NetOD=log10(IpreIopqIpostIopq). (3)

Iopq, a scanner calibration factor that considers the light scattered into the scanner that can reach the irradiated film, is acquired by doing a blank scan with the scanner bed covered in black cloth. Film calibration curves report dose delivered in cGy versus net optical density.

2.6. Absolute Dosimetry Techniques

Films were irradiated in two different orientations. Transverse, or horizontal, films, where irradiated such that the plane of the film was perpendicular to the beam axis while longitudinal, or vertical, films were parallel to the beam axis. Vertical films were aligned such that the plane of the film passed through the beam’s focal spot. Horizontal films were placed at different positions along the beam path depending on the desired beam cross-section. Measurements were performed in air as well as in a 30 × 15 × 20-cm water-filled phantom.

The tissue-maximum ratio is the ratio of dose or dose rate to the maximum dose and gives a quantitative measure of how dose deposition changes with depth in the target volume. Since the focal spot was fixed in space, the depth in water was varied by moving the phantom longitudinally relative to the focal spot. In this manner, focal spot doses were recorded at depths from 0 to 12 cm.

To fully quantify the CRnR beam for treatment planning, it was important understand how the beam interacts with different materials. Polybutylene terephthalate (PBT), a.k.a. Teflon, is often used as bone equivalent material. A PBT sample was imaged with a Varian cone beam CT scanner at 100 kVp, 80 mAs, to compare the Hounsfield units to that of bone.

Measurements were performed to analyze the effect of beam passage through heterogeneous materials. A stack of materials, starting with the top, consisted of 0.5 cm-thick PBT, 5 cm solid water, EBT3 film, and 10 cm solid water were placed in the beam such that the film was at the focal spot. This simulated 5 cm depth in water after passing through 5 mm of bone. The 5 mm of PBT was then replaced with 5 mm of solid water and the measurement repeated.

3. Results and Discussion

An example of film pieces exposed to known doses is shown in Figure 2, and Figure 3 depicts an EBT3 calibration curve, measured in red channel, for the equivalent polyenergetic beam where the ion chamber was configured using a water-to-air conversion factor of 1.0535 interpolated for HVL 9.0 mmAl from Table IV in (Ma et al 2001), so dose is presented as dose to a water medium. A third order polynomial, forced to pass through origin, was fitted to the data with the following equation:

Dose(cGy)=a1(netOD)+a2(netOD)2+a3(netOD)3. (4)

Figure 2.

Figure 2.

Sample exposed films for dose-calibration. EPSON 10000XL scanner settings where: transmission mode, positive film, 75 dpi, and 48-bit color. The scanned images were analyzed with ImageJ software using red channel. The left two films were both exposed to 150.4 cGy with the right two films being exposed to 300 cGy.

Figure 3.

Figure 3.

EBT3 calibration curve for an equivalent polyenergetic beam configured using 180 kVp, 15 mA, modified filter No. 6, and HVL 9.0 mm Al. The line represents a third order polynomial fitted to the optical density data.

Coefficients a1, a2, and a3 are 897.6, −123.5, and 4639, respectively. The presented dose-response curve, measured for EBT3 lot No. 07281401, may not apply to different lots. Usage of films from a different lot requires recalibration.

3.1. Energy Dependence

To show energy dependence, dose calibrations at three energies (125 kVp, 180 kVp, and 250 kVp) are presented in Figure 4, and the fit coefficient values as well as beam quality parameters are reported in Table 2. The curves for the two lower energies are almost indistinguishable and differ by < 1%. The 250 kVp curve shows a slight over-response of ~3% over the 400–1200 cGy region. This discrepancy should be viewed within the perspective of measurement uncertainty, which will be discussed shortly. The minimal energy dependence of EBT3 resulting from measured data for the orthovoltage energy range and dose range of interest has previously been reported for polyenergetic beams (Guerda Massillon et al 2016) and monoenergetic beams (Brown et al 2012).

Figure 4.

Figure 4.

The results show that EBT3 film response is practically independent in the presented energy range, i.e., for a given spatial resolution, negligible difference between beam energies can be seen.

Table 2.

First half value layers and effective energies are shown for each of the beam energies used. Coefficients a1, a2, and a3 of the third order polynomial fit to the data for netOD to dose conversion for the beam energies.

Beam Energy HVL Eff. E (keV) a1 a2 a3
125 kVp 3.5 mmAl 35.5 933.1 −446.7 5147
180 kVp 9.0 mmAl 60.0 897.6 −123.5 4639
250 kVp 1.3 mmCu 89.1 834.3 260.2 4363

3.2. Dose Measurement Uncertainty

Various factors contribute to uncertainty in dose determination from film. A film calibration, as represented in Figure 3, is the result of two measured quantities; the calibrated dose delivered to the film and the optical density read from the film, each with their own uncertainties. Following is an analysis of the overall uncertainties and contributing factors to each.

3.2.1. Optical Density Uncertainty

Several steps were taken to minimize systematic uncertainties in the film results. There was no batch-to-batch variation as a single batch was employed. Light polarization effects were avoided by ensuring consistent film-grain orientation during scanning as previously described. Variation due to film’s self-development was minimized by maintaining the same time difference between radiation exposure and film scanning. Variation due to scanner-bed uniformity was avoided by using a template, which ensured that film pieces were positioned in the same location on the scanner bed.

Uncertainty in OD results from (a) film non-uniformity, (b) film-to-film variation, (c) non-uniformity of scanner-bed response over the region in which film pieces were placed, and (d) drift in scanner response for post- versus pre-exposure scans.

In order to evaluate the film non-uniformity, film-to-film variation, and non-uniformity of scanner-bed response uncertainties, three film sheets were chosen (identified as BA, AI, and AJ) from the same batch used for the calibration procedure. Each film was cut into 20 pieces (see Figure 5) and exposed to same dose of 600 cGy with a 180 kVp (15 mA, modified filter No. 6, HVL 9.0 mmAl) beam from an Xstrahl 300 orthovoltage unit. Exposed films were scanned on the Epson 10000XL after the same 24-hour waiting period used for film calibration and analyzed with ImageJ using the red channel. Pixel values were converted to OD and then to dose by employing the afore-mentioned film calibration. Results showing relative OD variations are presented in Figure 6 which shows and expanded uncertainty (coverage factor k=2) of 1.6%. Scanner-bed non-uniformity was determined to have an uncertainty of 0.8% (k=2). The fourth uncertainty component, the drift in scanner response, was tracked over ~1 year (data not presented) and was found to be negligible (< 0.1%). Added in quadrature, the overall uncertainty in OD with k=2 is 1.8%.

Figure 5.

Figure 5.

Site pattern used for measurement of EBT3 homogeneity.

Figure 6.

Figure 6.

Combed film homogeneity and film-to-film variations for three EBT3 (Lot 10231801) films. Expanded uncertainty (k=2) is shown as a highlighted box around data.

3.2.2. Dose Uncertainty

Uncertainty in delivered dose results from ion chamber calibration related uncertainties (a) from the National Institute of Standards and Technology (NIST), (b) ADCL contribution, as well as the (c) electrometer (d) setup uncertainty, and (e) beam output variation. The uncertainty budget for the air kerma calibration coefficient is reported to be 1.50% for the combined contributions of NIST and ADCL with a coverage factor of 2. Electrometer uncertainty in the range of 0.1–700 nC is 0.13%. Setup errors amount to 0.5%. Historical reports of monthly dose-output checks indicate a dose reproducibility of ± 0.4% (k=2) (Lim et al 2021). Therefore, an overall uncertainty in doses delivered to film pieces is estimated to be 1.64% (k=2). Uncertainty in film-measured dose is expected to be higher at lower doses, as signal OD gets closer to noise, or background OD.

3.3. Absolute Dosimetry

3.3.1. Tissue-Maximum Ratio

Presented in Figure 7 is the tissue-maximum ratio that shows a fairly linear decrease in dose rate in water at depths up to 8 cm. It is expected that the maximal clinical depth for the CRnR system is 7 to 8 cm in water.

Figure 7.

Figure 7.

Tissue-maximum ratio (TMR) at the focal spot for various depths for a beam produced with the CRnR system using 125 kVp, 28.6 mA source parameters. Polynomial fit and equation are also shown.

3.3.2. Depth Analysis

Vertical films were used produce longitudinal profiles of the beam at various depths in water. These measurements were also used to characterize other important aspects of the beam. Central axis depth dose curves are reported in figure 8 and display the evolution of the longitudinal profile of the focal spot at increasing depths in water. Measurement uncertainty results in peaks, i.e. the 11 cm peak, which do not decrease strictly monotonically.

Figure 8.

Figure 8.

Depth dose curves for focused monoenergetic 60 kV x-rays in water taken longitudinally down the central axis passing through the focal spot where the different curves represent the spot at different depths.

3.3.3. Heterogeneous Materials

Figure 9 shows the CT number comparison between H2O, solid water, and PBT. The water and solid water show CT number of ~0 while the Teflon is ~225. Bone CT numbers are anywhere from 200 to 1000 (Kalra 2018) meaning that PBT is a sufficient approximation for bone in the CRnR energy range.

Figure 9.

Figure 9.

(Left) CT image of a cup of water, polybutylene terephthalate (Teflon), and solid water. CT numbers along the indicated line are plotted. (Right) CT number across the three materials.

The ratio of the focal spot dose rate with the bone-equivalent to that without was 0.87, meaning that passing through 5 mm of PBT upstream of the focal spot attenuated the beam by 13%. A second measurement was performed similar to the first except that the material was stacked in the following order: 5 cm solid water, EBT3 film, 0.5 cm PBT, and 10 cm solid water. The measurement was repeated by replacing the bone-equivalent with 0.5 cm solid water. This measured the enhanced dose on the interface between PBT and tissue as a result of backscatter. The ratio of dose rate with and without the bone-equivalent was 1.2, resulting in an enhanced dose of 20%.

4. Conclusions

A method was devised for calibrating EBT3 film for monoenergetic 60 kV focused x-rays. Absolute dosimetry for the CRnR converging beam has been used to evaluate various beam characteristics in various materials. Future work will incorporate these results into analysis of optimal clinical sites for the CRnR system.

Acknowledgements

MD Anderson is in financial agreement with Convergent Radiotherapy and Radiosurgery to assist in clinical implementation.

Editorial support was provided by Bryan Tutt, Scientific Editor, Research Medical Library, MD Anderson Cancer Center

References

  1. Bartkoski DA, Bar-David A, Kleckner M, Mirkovic D, Tailor R, Moradi-Kurdestany J, Borukhin S, Harel Z, Burshtein Z, Zuck A and Salehpour M 2021. Analysis of a novel X-ray lens for converging beam radiotherapy Sci. Rep. 11 19180 Online: 10.1038/s41598-021-98888-8 [DOI] [PMC free article] [PubMed] [Google Scholar]
  2. Brown TAD, Hogstrom KR, Alvarez D, Matthews KL, Ham K and Dugas JP 2012. Dose-response curve of EBT, EBT2, and EBT3 radiochromic films to synchrotron-produced monochromatic x-ray beams Med. Phys 39 7412–7 Online: http://doi.wiley.com/10.1118/1.4767770 [DOI] [PubMed] [Google Scholar]
  3. Butson MJ, Cheung T and Yu PKN 2004. Radiochromic film: the new x-ray dosimetry and imaging tool Australas. Phys. Eng. Sci. Med 27 230–230 Online: http://link.springer.com/10.1007/BF03178654 [DOI] [PubMed] [Google Scholar]
  4. Butson MJ, Yu PKN, Cheung T and Metcalfe P 2003. Radiochromic film for medical radiation dosimetry Mater. Sci. Eng. R Reports 41 61–120 [Google Scholar]
  5. Devic S, Seuntjens J, Hegyi G, Podgorsak EB, Soares CG, Kirov AS, Ali I, Williamson JF and Elizondo A 2004. Dosimetric properties of improved GafChromic films for seven different digitizers Med. Phys 31 2392–401 [DOI] [PubMed] [Google Scholar]
  6. Guerda Massillon JL, Muñoz-Molina ID and Díaz-Aguirre P 2016. Optimum absorbed dose versus energy response of gafchromic EBT2 and EBT3 films exposed to 20–160 kV x-rays and 60Co gamma Biomed. Phys. Eng. Express 2 [Google Scholar]
  7. Hubbell JH and Seltzer SM 1995. Tables of X-Ray Mass Attenuation Coefficients and Mass Energy-Absorption Coefficients 1 keV to 20 MeV for Elements Z = 1 to 92 and 48 Additional Substances of Dosimetric Interest http://physics.nist.gov/PhysRefData/XrayMassCoef/cover.html Online: http://physics.nist.gov/PhysRefData/XrayMassCoef/cover.html
  8. Kalra A 2018. Developing FE Human Models From Medical Images Basic Finite Element Method as Applied to Injury Biomechanics ed Yang KH (Elsevier; ) pp 389–415 Online: 10.1016/B978-0-12-809831-8.00009-X [DOI] [Google Scholar]
  9. Khachonkham S, Dreindl R, Heilemann G, Lechner W, Fuchs H, Palmans H, Georg D and Kuess P 2018. Characteristic of EBT-XD and EBT3 radiochromic film dosimetry for photon and proton beams Phys. Med. Biol. 63 065007 Online: http://stacks.iop.org/0031-9155/63/i=6/a=065007?key=crossref.9c0c3405952e1a870964446240939921 [DOI] [PubMed] [Google Scholar]
  10. Lim TY, Mirkovic D, Wang X and Tailor R 2021. Devices for dosimetric measurements and quality assurance of the Xstrahl 300 orthovoltage unit J. Appl. Clin. Med. Phys 22 151–7 Online: https://onlinelibrary.wiley.com/doi/10.1002/acm2.13220 [DOI] [PMC free article] [PubMed] [Google Scholar]
  11. Ma CM, Coffey CW, DeWerd LA, Liu C, Nath R, Seltzer SM and Seuntjens JP 2001. AAPM protocol for 40–300 kV x-ray beam dosimetry in radiotherapy and radiobiology Med. Phys 28 868–93 [DOI] [PubMed] [Google Scholar]
  12. Niroomand-Rad A, Chiu-Tsao ST, Grams MP, Lewis DF, Soares CG, Van Battum LJ, Das IJ, Trichter S, Kissick MW, Massillon-JL G, Alvarez PE and Chan MF 2020. Report of AAPM Task Group 235 Radiochromic Film Dosimetry: An Update to TG-55 Med. Phys 47 5986–6025 [DOI] [PubMed] [Google Scholar]
  13. Schwob N and Orion I 2007. Film dosimetry calibration method for pulsed-dose-rate brachytherapy with an 192Ir source Med. Phys 34 1678–83 [DOI] [PubMed] [Google Scholar]
  14. Williams M and Metcalfe P 2011. Radiochromic film dosimetry and its applications in radiotherapy AIP Conf. Proc. 1345 75–99 [Google Scholar]

RESOURCES