Skip to main content
Frontiers in Oncology logoLink to Frontiers in Oncology
. 2022 Apr 20;12:841761. doi: 10.3389/fonc.2022.841761

Synthetic MRI for Radiotherapy Planning for Brain and Prostate Cancers: Phantom Validation and Patient Evaluation

Pierrick Gouel 1,2,*, Sebastien Hapdey 1,2, Arthur Dumouchel 2, Isabelle Gardin 1,2,3, Eva Torfeh 3, Pauline Hinault 1, Pierre Vera 1,2, Sebastien Thureau 1,2,3, David Gensanne 1,2,3
PMCID: PMC9065558  PMID: 35515105

Abstract

Purpose

We aimed to evaluate the accuracy of T1 and T2 mappings derived from a multispectral pulse sequence (magnetic resonance image compilation, MAGiC®) on 1.5-T MRI and with conventional sequences [gradient echo with variable flip angle (GRE-VFA) and multi-echo spin echo (ME-SE)] compared to the reference values for the purpose of radiotherapy treatment planning.

Methods

The accuracy of T1 and T2 measurements was evaluated with 2 coils [head and neck unit (HNU) and BODY coils] on phantoms using descriptive statistics and Bland–Altman analysis. The reproducibility and repeatability of T1 and T2 measurements were performed on 15 sessions with the HNU coil. The T1 and T2 synthetic sequences obtained by both methods were evaluated according to quality assurance (QA) requirements for radiotherapy. T1 and T2 in vivo measurements of the brain or prostate tissues of two groups of five subjects were also compared.

Results

The phantom results showed good agreement (mean bias, 8.4%) between the two measurement methods for T1 values between 490 and 2,385 ms and T2 values between 25 and 400 ms. MAGiC® gave discordant results for T1 values below 220 ms (bias with the reference values, from 38% to 1,620%). T2 measurements were accurately estimated below 400 ms (mean bias, 8.5%) by both methods. The QA assessments are in agreement with the recommendations of imaging for contouring purposes for radiotherapy planning. On patient data of the brain and prostate, the measurements of T1 and T2 by the two quantitative MRI (qMRI) methods were comparable (max difference, <7%).

Conclusion

This study shows that the accuracy, reproducibility, and repeatability of the multispectral pulse sequence (MAGiC®) were compatible with its use for radiotherapy treatment planning in a range of values corresponding to soft tissues. Even validated for brain imaging, MAGiC® could potentially be used for prostate qMRI.

Keywords: synthetic MRI, radiotherapy planning, quantitative MRI, T1 mapping, T2 mapping

1 Introduction

MRI displays contrast that is principally dependent on 3 intrinsic parameters: T1, T2, and ρ. For diagnostic purposes, weighted sequences aim to obtain a contrast depending on the nature of the tissue studied by highlighting one of these 3 parameters. The drawback of such an approach is the complex nature of the signal. It differs from one manufacturer to another for the same types of weighted images, and it is not possible to extract the quantitative values of intrinsic parameters.

Meanwhile, quantitative magnetic resonance imaging (qMRI) (1) has been proposed for many years to obtain a mapping of the measurement of one of the intrinsic parameters. These sequences are based on a known mathematical expression of the signal, depending on only one of the 3 intrinsic parameters, to calculate the intrinsic parameter at the voxel level by solving the signal expression. qMRI has not been widely used because of the prohibitive time required for image acquisition, measurement uncertainties (2), and diagnostic success of weighted sequences.

qMRI allows the characterization of tissues and pathologies for a wide range of diagnostic applications (1, 3). It can provide quantitative T1 and T2 tissue mappings at high spatial resolution (4), which can be useful for radiotherapy to differentiate recurrent tumors from benign tissues (5), to improve contouring (6) and optimize treatment planning (7), or to detect early effects of irradiation (8).

Another application of qMRI after the acquisition of T1 or T2 mapping is in mathematically generating synthetic T1- or T2-weighted sequences (synMRI). This technique allows the operator to modify the repetition time (TR) and echo time (TE) values in order to obtain a multi-contrast MRI signal from a single acquisition. The main advantages of this technique are the reduction of the scan time and improvement of patient throughput. Evaluations of this new technique are essentially based on brain studies (9). Several authors have also proposed synMRI for the study of the knee (10, 11), lumbar intervertebral disc degeneration (12), and prostate cancer (13), but always with the acquisition parameters for diagnostic purposes favoring the signal-to-noise ratio (SNR).

The first quantitative methods proposed in the literature were inspired by the methods used in nuclear magnetic resonance (NMR) (1). For T1 measurement, the reference technique is an inversion recovery (IR) sequence that remains too time-consuming for routine use. To reduce the acquisition time, a variable flip angle (VFA) method has been proposed (14). This method allows a rapid return to equilibrium magnetization of the repeated sequences with low tilt angles of longitudinal magnetization. At the end of each sequence, the signal is measured after the application of a gradient echo (GRE). The homogeneity of the B1 field must be checked and corrected when necessary (14). For T2 measurement, the reference sequence is based on a single spin echo (SE) sequence with signal recording after each multi-echo (ME-SE) (1).

The previous methods allow mapping of only one intrinsic parameter (T1 or T2), but not all three (T1, T2, and ρ). In recent years, several authors have proposed multispectral pulse sequences that enable recording the signal several times in a single sequence in order to calculate the 3 intrinsic parameters while limiting the acquisition time. The characteristics are then mapped, and each voxel represents the value of T1, T2, or ρ. Warntjes et al. (15) proposed a quicker method based on a multi-parametric pulse sequence (QRAPMASTER) from a ME saturation recovery acquisition using a turbo spin echo (TSE) readout to generate T1, T2, and ρ mappings applied to brain imaging. At the same time, measurement of the B1 field is performed to correct the inhomogeneity of the amplitude radiofrequency (RF) emission. Inspired by the QRAPMASTER sequence, the magnetic resonance image compilation (MAGiC®) sequence, available on General Electric Healthcare systems (GE Healthcare, Milwaukee, WI, USA), has been made commercially available (16).

Several publications evaluated the performance of MAGiC® compared to conventional sequences (IR and ME-SE) with a phantom on a 3-T MRI, with discordant results (10, 17, 18). Although the SNR is lower than that at 3 T (19), MRI at 1.5 T has advantages, such as the reduction of artifacts and a better homogeneity of the radiofrequency field, which has a direct impact on the qMRI image quality and which are of major interest for the purpose of anatomical segmentation radiotherapy planning. Furthermore, Guarnaschelli et al. showed that 3-T MRI may reveal a significantly smaller tumor volume for high-grade gliomas and that target volume segmentation for radiation treatment may be better at 1.5 T (20). West et al. (21) studied the similarities and differences of the QRAPMASTER® sequence at 1.5- and 3-T field strengths in brain tissue segmentations. This study showed that most of the different brain tissues were classified identically at both field strengths, although some regional differences were observed, such as variations in the segmented tissue volumes. They also noticed a better repeatability of measurements at 1.5 T.

To our knowledge, no study has evaluated the quality of MAGiC® measurements at 1.5 T compared to conventional qMRI measurement methods to obtain multi-contrast synthetic sequences applied to anatomic segmentation for radiotherapy treatment planning, given its mandatory quality assurance (QA) requirements (7, 22). In this work, we evaluated the accuracy, the range of use, and the reproducibility of T1 (MAGiC® and GRE-VFA) and T2 (MAGiC® and ME-SE) based on phantom studies. Performance was evaluated with two coils, a head and neck unit (HNU) coil and a BODY coil with the acquisition parameters for radiotherapy treatment. We also compared the in vivo T1 and T2 measurements in human brain and prostate tissues for radiotherapy application.

2 Materials and Methods

2.1 Phantoms

The Magphan® SMR170 (The Phantom Laboratory, Salem, NY, USA) was used (Figure 1A). It has conventional cylindrical housing for test groups and a removable end plate for internal access. The acrylic cylinder has an outer diameter of 20 cm and an inner diameter of 19 cm. It contains features that allow comprehensive testing of achievable MRI scanner QA parameters (QAp). Geometric distortion, spatial linearity, pixel size, slice increment, slice thickness, high-contrast detectability, low-contrast detectability, and uniformity were measured. These parameters have already been described in previous publications (7, 2224).

Figure 1.

Figure 1

(A) Magphan® SMR170 (The Phantom Laboratory, Salem, NY, USA). (B) Synthetic T2-weighted image of the homemade phantom used in this study. The blue volumes of interest (VOIs) (B) represent the different inserts measured in this study.

To cover a wider range of T1 and T2 values, a homemade phantom was developed (Figure 1B). It is a box composed of a 0.3-cm-thick plastic wall filled with water. It is composed of eight inserts, 5 cm high and 1 cm wide. Each sample contained a solution of 1.5 ml of gadoteric acid (DOTAREM®, 0.5 mmol/ml). The concentrations and the relaxation times (T1 and T2) were, respectively, for insert 1 (I1) = 8 × 10−3 mmol/ml (29.1 and 24.7 ms), I2 = 4 × 10−3 mmol/ml (57.7 and 48.8 ms), I3 = 2 × 10−3 mmol/ml (113.1 and 95.5 ms), I4 = 1.10−3 mmol/ml (217.8 and 181.7 ms), I5 = 4 × 10−4 mmol/ml (489.7 and 399.6 ms), I6 = 2 × 10−4 mmol/ml (838.9 and 665.6 ms), I7 = 1 × 10−4 mmol/ml (1,303.8 and 997.5 ms), and I8 = 0 mmol/ml (2,385.7 and 1,311.3 ms). The effective concentrations of each insert were verified by an MRI spectrometer (Biospec 47/40 imager; Bruker, Billerica, MA, USA) under clinical conditions at 1.5 T to define the values of the T1 and T2 reference relaxation times for a temperature of 22°C after the end of the measurement phantom study.

2.2 T1 and T2 Measurements

2.2.1 Phantom Data Acquisitions

All acquisitions were performed on a 1.5-T Optima MR450w clinical MRI (GE, software version DV26.0-R01) at a temperature of 22°C (±0.5°C) maintained by a continuous air conditioning system. The sequences were acquired on the phantoms in 2D axial orientation according to the acquisition parameters given in Table 1. It consisted of the MAGiC® sequence for generating T1, T2, and B1 mappings (16) and conventional sequences. The conventional methods used were four GRE-VFA with a destructive gradient of the residual magnetization between each change of the tilt angle (14) for the measurement of T1 and an ME-SE sequence with eight echoes for the measurement of T2 (1). To compare the performance of the methods, the acquisition parameters of the MAGiC® sequence were similar to those of the conventional sequences (Table 1). Two coils were used for the qMRI evaluation: the HNU coil and the BODY coil in 24AA2 configuration (combination of coils anterior and posterior). Due to its size, the Magphan® SMR170 phantom could not be imaged with the HNU coil for the synMRI QA analysis.

Table 1.

Acquisition parameters of the phantom images.

Sequences MAGiC® GRE-VFA (T1) ME-SE (T2)
No. of TE 2 1 8
TE values (ms) 21.8–87.1 2.2 8.1–16.2–24.3–32.2–40.3–48.4–56.4–64.5
TR values (ms) 4,000 8 1,000
Bandwidth (kHz) 31.25 31.25 31.25
Slice thickness (mm) 2.5 2.5 2.5
Flip angle (deg) 120–180–90 3–10–20–30 90
Pixel size (mm) 1.02 × 1.02 1.02 × 1.02 1.02 × 1.02

TE, echo time; TR, repetition time.

2.2.2 qMRI Measurements From the MAGiC® Sequence

T1 and T2 mappings from the MAGiC® sequence were calculated using SyMRI® software (version 100.1.1; SyntheticMR AB, Linköping, Sweden) available on the MRI acquisition station. The expression of the signal, SMAGiC, is given in Equation 1.

SMAGiC=KGE·ρ·eTE/T2·1[1cos(B1θ)]·exp(TI/T1)cos(B1θ)·exp(TR/T1)1cos(B1α)cos(B1θ)·exp(TR/T1)   (1)

where KGE is a global intensity scaling factor taking into account coil sensitivity, RF chain amplification, and voxel volume that is specific to MRI. ρ is the proton density, B1 is the radiofrequency field, TI is the macroscopic magnetization reversal time, TR is the repetition time, and TE is the echo time.

The algorithm for calculating T1 and T2 mappings is based on a least-squares adjustment of the signal intensity for each voxel. The volumes of interest (VOIs) of 200 voxels were positioned on the homemade phantom on each sample at slice 13/26. The mean T1 and T2 values, as well as the standard deviation (SD), were calculated for each VOI.

2.2.3 qMRI Measurements From Conventional Sequences

For T1 measurements using the GRE-VFA sequence, the expression of the signal is given in Equation 2.

ST1=KGE·ρ·sin(θ)·1e(TRT1)·e(TE/T2*)1cos(θ)xexp(TR/T1)   (2)

The B1 field uniformity correction using the Bloch–Siegert offset method (25) was evaluated on phantom once a week for 6 weeks for both coils. The mean values were calculated; the obtained calibration curves are given as supplementary data (Supplementary Figures S1, S2). Since the systematic errors for both coils were between −1.6% and 1.8% whatever the coil used, no correction of the B1 field uniformity was applied to the data obtained with the T1 mappings generated from the conventional sequences on phantoms and on patients.

For the T2 measurement, the signal expression verified Equation 3.

ST2=KGE·ρ·1eTR/T1·eTE/T2   (3)

T1 and T2 mappings were calculated using the OleaNova+® module of Olea Sphere® software (version 3.0; OLEA MEDICAL, La Ciotat, France) by solving Equations 2 (T1) and 3 (T2). The VOIs of 180 voxels on T1 and T2 mappings were defined on the homemade phantom on each sample at slice 13/26. The mean T1 and T2 values, as well as the SD, were calculated for each VOI.

2.2.4 synMRI Measurements

From the quantitative maps obtained, T1- and T2-weighted images were synthesized from these maps with the syMRI® software for the MAGiC® sequence and OleaNova+® for the conventional sequences. For both methods, values of TE = 25 ms and TR = 600 ms were defined for the T1-weighted synthetic sequence and TE = 65 ms and TR = 1,900 ms for the T2-weighted synthetic sequence. All sequences were exported and saved in Artiscan® software (version 4.1.18; AQUILAB, Loos, France) for QA analysis. For each sequence, the 2D geometric distortion, spatial linearity, pixel size, slice increment, slice thickness, high-contrast detectability, low-contrast detectability, and uniformity were measured (24, 26).

2.2.5 In Vivo Measurements

After giving their free and informed consent (Institutional Review Board no. 2103B), 10 patients (2 women and 8 men) with oncologic diseases accepted additional acquisitions with MAGIC® and conventional sequences. Five patients imaged on the brain (mean age = 56 years, range = 39–76 years) had indications of glioblastoma, and five patients imaged on the pelvis (mean age = 67 years, range = 45–80 years) had indications of prostate cancer. The same acquisition parameters were used (Table 1), except for the pixel size, which was increased to 2 × 2 mm for both methods to make the acquisition time acceptable (5 min, 36 s for MAGiC® and 2 min, 36 s GRE-VFA + 3 min, 38 s for ME-SE). T1 and T2 mappings were calculated from the MAGiC® sequence and conventional sequences using the same procedures as in the phantom experiments. The VOIs measured for both patients on T1 and T2 mappings corresponded to the theoretical range of application of the MAGiC® sequence. For the brain and prostate, 3 VOIs (oval and identical of 44 voxels and size 2 × 2 × 2.5 mm) were manually positioned on the same anatomical and non-pathological areas at the identical slice number for each T1 and T2 mapping obtained with both methods. For the brain, the anatomical areas were white matter (WM), gray matter (GM), and cerebrospinal fluid (CSF). For 1 of the 5 patients, a VOI was positioned over the known calcium lesion. For the pelvis, anatomical areas of the prostate, gluteal muscle, and subcutaneous fat of the buttocks were studied.

2.3 Statistical Analysis

A descriptive statistic was carried out from the T1 and T2 maps obtained using the phantom at 1.5 T. This analysis focused on the calculation of the mean and relative percentage difference between the measured and the reference value of the VOIs positioned on the image of each insert. The standard deviation and coefficient of variation (CV) (27) representing the voxel-to-voxel measurement uncertainty were given as percentages in each insert for all measurement methods and for each coil. The agreement of the mean values of the T1 and T2 maps was compared to the reference values calculated at 1.5 T using a Bland–Altman graph (28) for each coil configuration. The mean relative difference of 14 repeated measurements (in percent) with respect to the first measurement was calculated and plotted in boxplot graphs (29). Repeatability was quantified with the CV within the VOIs over the 15 scan sessions, and the 95% confidence interval of the CV was calculated. A comparison was performed on the T1 and T2 maps obtained on the brain and prostate between the two measurement methods using a Bland–Altman graph (28) with the 95% limits of agreement as the mean difference (a low limit represents the mean − 1.96 × SD and a high limit represents the mean + 1.96 × SD). The mean VOIs measured with the minimum and maximum values of the different anatomical areas, as well as the average of the relative differences with the minimum and maximum values, were given for analysis.

All statistical analyses were performed using Python® software (version 3.7.0).

3 Results

3.1 Phantom Measurements

3.1.1 Accuracy and Range of T1 and T2 Measurements

Regarding the T1 measurement, Figure 2 and Supplementary Tables S1, S2 show that the GRE-VFA sequence with the HNU coil provided a relatively accurate estimation of T1 (relative differences ranged from −7.9% to +16.0%). Larger biases were measured with the BODY coil, which systematically gave lower measurements with maximum errors from −2.8% to +18.8%. The MAGIC® sequence gave T1 measurements consistent with the expected values, with biases less than 20% in the range 490–2,386 ms. On the other hand, for T1 ≤218 ms, the measurements were highly overestimated (see outliers from 113 ms and below in Figure 2). The HNU coil gave better results than the BODY coil in the T1 measurement from 490 to 2,386 ms.

Figure 2.

Figure 2

Bland–Altman graph representing the relative difference (in percent) between the T1 measurement with the MAGIC® and the gradient echo with variable flip angle (GRE-VFA) sequences for the head and neck unit (HNU) coil (A) and the BODY coil (B) compared to the reference measurements.

For the ME-SE sequence, between 25 and 400 ms, the HNU coil provided accurate T2 measurements (relative differences ranged from −13.6% to +6.8%) (Figure 3 and Supplementary Tables S1, S2). For higher T2 (>400 ms), the measurement was systematically underestimated, with maximum errors from −86.1% to −31.8%. With the BODY coil, the ME-SE sequence also gave good agreement for T2 measurements ≤400 ms (relative differences ranged from +3.9% to +8.2%). The MAGIC® sequence gave T2 measurements with a bias less than 20% for T2 ≤400 ms (relative differences ≤15.7%) with the HNU coil. With the BODY coil, the MAGiC® sequence was less efficient, with biases lower than 20% on T2 values lower than 95.2 ms.

Figure 3.

Figure 3

Bland–Altman graph representing the relative difference (in percent) between the T2 measurements with the MAGIC® and the multi-echo spin echo (ME-SE) sequences for the head and neck unit (HNU) coil (A) and the BODY coil (B) compared to the reference measurements.

The known ranges of relaxation times for human biological tissues at 1.5 T are listed in a database from various articles published in the literature (30) and presented in Figure 4. This figure shows that, in our study, the T1 and T2 values estimated with the conventional and MAGiC® methods with error less than 20% covered a wide range of known relaxation times of human biological tissues at 1.5 T. In particular, those of interest for radiotherapy treatment planning are between 30 and 2,883 ms and between 25 and 182 ms for T1 and T2 measurements, respectively. The MAGiC® method showed higher biases than those of the conventional method for T1 values <490 ms and for T2 values >95 ms.

Figure 4.

Figure 4

Comparisons of the results of the relative differences ≤10% (except inserts 5, 7, and 8, where the results were slightly higher and given on the diagram) of the T1 and T2 mappings for the MAGiC® method (thick orange line) and the conventional method (thick blue line) for the range of values measured on phantom compared to the reference values (dotted line mark), to the range of human biological tissues (30) at 1.5 T (thick red line mark), and to the range of human biological tissues of interest for radiotherapy planning at 1.5 T (green line mark).

3.1.2 Reproducibility and Repeatability of T1 and T2 Measurements

Figures 5A , B give the evolution of the T1 and T2 measurements over time for each insert for the conventional and MAGiC® methods using the HNU coil. The relative difference with the associated median, the first and third quartiles, and the minimum and maximum values are given between the first measurement and each of the 14 measurements repeated over 5 months. This figure shows that T1 measurement with the MAGIC® sequence was stable over 5 months, with a median close to zero for all inserts, except for insert 2 (median error greater than −25%), and a dispersion between ±10%, but with some extreme measurements. On the other hand, T1 measurements with the GRE-VFA method were less reproducible for the first four inserts. The T2 measurement of the inserts was globally reproducible over 5 months with the ME-SE method, with a median error less than 6% compared to the MAGiC® method, which showed worse results, especially for the last four samples (medians errors between −13% and −63%). The dispersion and extreme values of the relative differences of each sample increased as the T2 values increased for both methods.

Figure 5.

Figure 5

Box plot of the repeated measurements (n = 14) with the head and neck unit (HNU) coil of T1 mappings for the MAGiC® sequence and gradient echo with variable flip angle (GRE-VFA) (A) and T2 mappings for the MAGiC® sequence and multi-echo spin echo (ME-SE) (B). The values of the relative differences (in percent) are given with the associated medians, first and third quartiles, and the maximum and minimum values.

Figures 6A , B give the CVs of repeated measurements (n = 15) over 5 months with the HNU coil of T1 mapping for MAGiC ®and GRE-VFA and T2 mapping for MAGiC® and ME-SE, with the associated 95% confidence intervals. The figure shows that the CVs were elevated for T1 measurement with the MAGiC® method for the first two inserts (29% and 66%). For the other inserts, all CVs were similar and lower than 16% for both measurement methods. The CVs were similar and comparable for the T2 measurement for both measurement methods. The CV values were less than 10% for the first four inserts. Then, they degraded for the last four inserts with a maximum of 34% for MAGiC®, whose CV values were higher than those of ME-SE for the last three inserts. An increase in the CV confidence interval for both T1 and T2 measurement methods was observed with higher relaxation times studied.

Figure 6.

Figure 6

Coefficients of variation of the repeated measurements (n = 15) with the head and neck unit (HNU) coil of T1 mappings for MAGiC ®and gradient echo with variable flip angle (GRE-VFA) (A) and T2 mapping for MAGiC® and multi-echo spin echo (ME-SE) (B). Vertical lines show the 95% confidence intervals.

3.1.3 QA Analysis on Synthetic T1- and T2-Weighted Images

Synthetic T1- and T2-weighted images were obtained with the Magphan® SMR170 phantom. These images were synthesized from the T1 and T2 mappings with the syMRI® software for the MAGiC® sequence and OleaNova+® for the conventional sequences with the same TE and TR values. For each sequence, the geometric distortion, spatial linearity, pixel size, slice increment, slice thickness, high-contrast detectability, low-contrast detectability, and uniformity were measured and are shown in Table 2. The expected optimal values of each parameter were given for comparison to the measurements of each sequence obtained from T1 and T2.

Table 2.

Values of the quality assurance (QA) parameters for phantom.

Optimal values MAGiC® T1 GRE-VFA MAGiC® T2 ME-SE
Geometric distorsion Mean diameter (mm) 190.00 190.38 189.47 190.11 189.47
% 0.00 0.04 0.14 0.37 0.14
Spatial linearity (%) 0.00 −0.13 −1.46 −0.16 −1.51
Pixel size x (mm) 1.02 1.00 1.02 1.00 1.02
y (mm) 1.02 1.02 1.02 1.02 1.01
Slice increment (mm) 2.5 2.47 2.24 2.47 2.26
Slice thickness (mm) 2.5 2.22 2.86 2.37 2.9
High-contrast detectability Resolution (pl/cm) Minimal value 6.95 6.77 7.55 6.80
Low-contrast detectability SNR Maximal value 58.64 267.02 79.29 353.94
Noise (%) 0.00 3.73 5.98 5.52 5.41
Uniformity (%) 100.00 92.08 89.30 89.34 89.35

GRE-VFA, gradient echo with variable flip angle; ME-SE, multi-echo spin echo; SNR, signal-to-noise ratio.

The results showed a distortion lower than 0.5% for both T1 and T2 measurement methods. The measurement of spatial linearity showed good results (<1.5%), but was better for the MAGiC® measurement method for T1 and T2. Pixel size was comparable for both T1 and T2 measurement methods. The slice thickness was slightly below the expected optimal value for the MAGiC® method (between −0.22 and −0.13 mm), while it was higher for the conventional method in T1 and T2 (+0.4 mm). The high-contrast detectability was comparable for both measurement methods in T1 and worse for the MAGiC® method (7.55 pl/cm) compared to the conventional method (6.8 pl/cm). The conventional method showed better low-contrast detectability in T1 and T2 compared to the MAGiC® method with a similar noise level. The measured uniformity was good for both methods in T1 and T2 (>89%).

3.2 In Vivo Measurements

Figure 7 shows an example of a synthetic image obtained from the MAGiC® sequence for the study of the brain and prostate. The images offered a level of visual image quality that allows the study of the different tissues for radiotherapy treatment planning.

Figure 7.

Figure 7

(A) T1 maps, T2 maps, SynT1w, and SynT2w of the brain in a 39-year-old man with known calcified lesion. Stars represent the volume of interest (VOI) of the mean measured values, white matter (WM), gray matter (GM), and cerebrospinal fluid (CSF), for the MAGiC® method. (B) T1 maps, T2 maps, SynT1w, and SynT2w of the pelvis in a 72-year-old man with prostate cancer. Stars represent the VOI of the mean measured fat, muscle, and prostate for the MAGiC® method and conventional sequences.

3.2.1 Comparison of T1 and T2 Measurements in the Human Brain

In Figures 8A, B, the overall results showed that the average differences between the two measurement methods for the different anatomical areas of the brain studied were −7% for the T1 measurement and 7.2% for the T2 measurement. All measurements were within the limits of agreement, except for one extreme measurement for a high T2 value (>30%) and one measurement at the limit of agreement for a high T1 value (−19.8%). The results showed that the two measurement methods are comparable for obtaining T1 and T2 mappings. In Figure 7A, a brown star represents the measurement of a calcium lesion found in 1 of the 5 patients. We showed a significant difference in the T1 measurement with a relative difference of +69.6% between the MAGiC® and GRE-VFA methods, whereas the T2 measurement was comparable for both methods (22 ms for MAGiC® versus 21 ms for ME-SE).

Figure 8.

Figure 8

Bland–Altman graph representing the relative difference (in percent) between the T1 (A) and T2 (B) maps with the MAGIC® method and conventional sequences for the five patients scanned on the brain with the head and neck unit (HNU) coil. The red solid line marks the mean of the difference; the red dotted lines mark the mean ± 1.96 standard deviation of the difference.

3.2.2 Comparison of T1 and T2 Measurements in the Human Prostate

The overall results (Figures 9A, B) showed that the average differences between the two measurement methods for the different anatomical areas of the prostate studied were −3.2% for the T1 measurement and 5.7% for the T2 measurement. All measurements were within the limits of agreement. The results showed that the two measurement methods were similar for obtaining T1 and T2 mappings.

Figure 9.

Figure 9

Bland–Altman graph representing the relative difference (in percent) between the T1 (A) and T2 (B) maps with the MAGIC® method and conventional sequences for the five patients scanned on the prostate with the BODY coil. The red solid line marks the mean of the difference; the red dotted lines mark the mean ± 1.96 standard deviation of the difference.

4 Discussion

Our study compared, at 1.5 T, the performance of T1 and T2 measurements using a combined method enabling a single acquisition to obtain T1 and T2 mappings (MAGiC®) and conventional sequences (GRE-VFA and ME-SE) with acquisition parameters adapted to radiotherapy planning. The results of the homemade phantom showed good agreement between the two measurement methods with respect to the reference values for a wide range of T1 values, compatible with the biological tissues encountered in radiotherapy. The QAp evaluated on the T1 and T2 synthetic weighted images for both measurement methods are in agreement with the recommendations of imaging for contouring purposes for radiotherapy planning (7, 22), with a very low average distortion (<1%) and spatial linearity (<1.5%) and an excellent uniformity (>89%). For both brain and prostate tissues, the estimations by the two qMRI methods of T1 and T2 mappings were comparable.

To our knowledge, this is the first study evaluating the feasibility of the MAGiC® method for radiotherapy imaging of the male pelvis and, in particular, the prostate (Figure 6B). However, these results obtained on 5 patients need to be further evaluated on a larger number of patients.

For T1 measurement, Figure 2 demonstrates the good agreement between the reference values and the GRE-VFA sequence for the whole range of measurements (relative error, <18.8%), with a reproducibility close to 20% over 5 months for both coils. The MAGiC® sequence gave discordant values (>20%) for low T1 values, below 218 ms. A more detailed analysis between 200 and 500 ms would refine this analysis, but caution should be taken when using this sequence for the study of low T1 values. This issue was also found for the calcified lesion in one patient in the brain group (see brown star in Figure 7A), where the T1 values measured by the MAGiC® method were 500 and 152 ms for GRE-VFA. However, it was not found for the measurement of subcutaneous fat in the patients of the prostate group (Supplementary Table S3), with average measurement differences between the MAGiC® method and EG-VFA ranging from 6% to 15%. Our results are in agreement with those of Li et al. (17) at 3 T, who found comparable results between the 2 methods in the range of T 1 values they studied (For T1 values measured above 1131 ms). Our results are also in agreement with the results of Kim et al. (18), showing that, at 3 T, the 2 methods gave comparable T1 measurements below 433 ms. For T1 values measured between 1,131 and 2,117 ms and T2 values between 59 and 230 ms. This study showed relative differences between −8% and 25% for T1 and −27% and +10% for T2 compared to the phantom reference values. Kim et al. (18) also compared MAGiC® with conventional sequences (SE multi-TR for T1 measurement and ME-SE for T2 measurement) over a range of measured T1 values from 70 to 2,875 ms and T2 values from 51 to 1,437 ms. The results showed that the T1 measurements were comparable for both methods over a range of values between 433 and 2,875 ms and were significantly different between the two methods for T1 values measured between 70 and 382 ms and over the range of T2 values. Lee et al. (10), whose phantom evaluation of T2 measurements performed with the MAGiC® sequence in a value range between 30 and 110 ms, showed very good linear regression (y = 1.022x + 0.9903, R2 = 0.9985) compared to the ME-SE sequence (B0 = 3 T).

The accuracy and discrepancy between the 2 methods in the T2 measurements above 95 ms were not adequate for radiotherapy planning (Figure 4). This result is well known in NMR, where it has been shown that T2 measurement is very sensitive to magnetic field heterogeneities (31). The use of SE sequences can partly correct this effect, but the gradients required in imaging have a significant effect on transverse relaxation. Our results are in agreement with the results of Kim et al. (18) at 3 T, but discordant with those of Lee et al. (10), who found a very good linear relationship between the 2 methods for T2 values between 30 and 110 ms.

Due to the measurement uncertainties of qMRI, reported by Nunez-Gonzalez et al. (2), and our final application for radiotherapy treatment planning for prostate and brain tissues (Figure 4), considering a quantitative accuracy below 20% in the T1 and T2 measurements seems to be sufficient. Indeed, Zavalla et al. (32) proposed a decision tree for the automatic segmentation–classification of prostate tissues in 3 T from T1 and T2 mappings. The application of this decision tree to our 5 patients who had pelvic scans was performed with the manually obtained VOIs of the different tissues on non-pathological areas (Supplementary Table S3). By considering that the relaxation times of T1 and T2 between 1.5 and 3 T were different (2, 33), our results showed correct classification of the prostate, fat, and muscle tissues regardless of the qMRI method. For other quantitative imaging modalities (positron emission tomography), a reproducibility error of 25%–30% is accepted when interpreting the response to treatment of cancerous lesions (34). In addition, Stikov et al. (14) showed that the variability of T1 measurements is about 10% higher in vivo than in the phantom.

The values measured with the MRI scanner were highly erroneous in relation to the theoretical values of relaxation times. The measurement accuracy of T1 and T2 degraded for high values of T1 and T2, showing that there is significant scope for MRI to improve the accuracy of relaxation time measurement. However, these underestimates of T2 measurements from theoretical values at 1.5 T of 500 ms can be explained by the choice of the TR value used for the ME-SE method, which does not allow correct sampling of the signal decay in T2. Additionally, to obtain correct T2 measurements with this sequence, it would be necessary to increase the value of TE, and thus TR, at the detriment of the acquisition time. From a methodological point of view, we chose very similar acquisition times between MAGiC® and the conventional sequences for a relevant comparison. This duration corresponded to a clinically compatible acquisition time. For these same reasons, the GRE-VFA method was used for the conventional method instead of the IR method, which is acknowledged as the gold standard method for T1 measurement. The VFA method may overestimate the T1 values compared to IR (35) and affect the measurements if there is not a perfect destruction of the transverse magnetization before each flip angle; otherwise, the signal may deviate from the expected value as the flip angle increases. But this method has the advantage of speed and can be applied in 3D (14), which is interesting for a radiotherapy planning use. Recently, Nunez-Gonzalez et al. (2) have compared the MAGiC® sequence to modified conventional methods based on a very short acquisition time (36). The results on phantoms were similar to ours, with 3 times faster acquisition time. However, this very promising sequence has to be evaluated under acquisition conditions dedicated to radiotherapy planning with an adapted voxel size and QAp analysis.

The results of our study showed the stability over time of qMRI measurement with MRI at 1.5 T for repeated measurements, which is a prerequisite for oncology follow-up. Our results suggest that a regular evaluation through quality control of the measurement of values over time is essential for such an application. The harmonization of this quality control on several MRI scanners would allow the comparison of the signal measurements in multicenter studies (37). Our results showed extreme values for T1 measurements obtained with both methods between 489.7 and 1,303.8 ms at the time of one of our 14 measurements. Although the values varied in the same direction and modified the contrast between tissues, without altering the quality of the delineation, we believe that the insertion of a calibration object in the field of view (FOV) at the time of the acquisition of a patient could quantify the possible drift and allow an immediate correction of the obtained values. Thus, a fast quality control in real time could be performed before releasing the patient and performing a rescan, if necessary.

The QAp results revealed a slice thickness slightly away from the expected optimal value for both methods. This result can be very influential, especially in brain metastases for stereotactic radiosurgery (38), but can be compensated by a registration on a treatment planning CT (39). The SNR of the conventional method was significantly higher than that of the MAGIC® method for low-contrast detectability. It is known in MRI that the flip angle error due to imperfection of the slice profile and local changes in the B1 field is a source of bias for quantitative MR (40). It has been shown that, in the conventional method, the distribution of flip angles across the slice can induce slice profile distortions and a large excess of signal from the slice edges in subsequent RF pulses and that this could be ameliorated by discarding the subsequent RF views (41). Other promising acquisition methods are currently being investigated, which could improve these results (2, 42). Geometric distortion was only evaluated in 2D in this study, whereas 3D sequences are recommended for radiotherapy to more easily reconstruct in multiple planes and to limit the occurrence of 2D-related artifacts (22) because the MAGiC® method is currently only available in 2D on MRI systems. Further evaluation of 3D distortion should be performed upon implementation in MRI machines. Furthermore, the distortion measurement would have been more complete by measuring it in all x and y directions in addition to the phantom diameter measurement, but the Magphan® SMR170 version does not allow for this more accurate assessment.

In this work, quantitative MRI relaxometry techniques were evaluated for integration into radiotherapy treatment planning of the brain and prostate. We believe that this imaging technique could provide significant assistance to the radiation therapist in tumor contouring. For radiotherapy planning, quantitative MRI relaxometry allows the generation of several synthetic MRI contrasts that can distinguish several close contrast structures from a single acquisition (Videos S1 and S2). It thus offers the double advantage of contouring assistance and workflow optimization because it can replace the acquisition of several sequences requiring a longer time. qMRI could also evaluate changes induced by irradiation delivered in radiotherapy, as recently shown with MRI imaging of myelin water by multicomponent T2 relaxometry (43). The MAGiC® method was initially proposed for the study of the brain. The results given in Figures 7B, 9 and Supplementary Table S3 showed that its use seems possible on the prostate. This result is interesting for prostate cancer, given that several studies (44, 45) have shown that T2 mapping has good diagnostic performance and could provide an indication of the aggressiveness of the disease. The preliminary results on five patients presented here are encouraging and seem more favorable than those on phantoms compared to reference samples (Figure 3B). This result is counterintuitive as it is expected in MRI that the results on patients are worse than those on phantoms. The differences between the in vitro and in vivo measurements may be due to the design of the homemade phantom. Indeed, as the vials used were 1 cm in diameter, the proton environment may have been more sensitive to local variations in T1 and T2 relaxation times. This type of variation was not present in the in vivo measurements. Regardless, these results on patients must be confirmed by a clinical study on a larger number of patients. The sample concentrations of the T1 and T2 values of the homemade phantom were <2,386 and <24.7 ms, respectively, whereas minimum concentrations of 4,627 ms for T1 and 6 ms for T2 were required for the completeness of the range of human biological tissues. However, only the CSF for T1 and the lung for T2 had relaxation times at 1.5 T in this range (30). We verified on phantom that it was not necessary to apply a B1 field uniformity correction from the GRE-VFA method. However, Stikov et al. (14) showed that the variability of the T1 measurements was affected by the B1 heterogeneity. However, it was not possible with our MRI system to parameterize the B1 ( 25) measurement method used at a slice thickness less than 5 mm. Furthermore, this would have increased the global acquisition time of the conventional method. Errors in the absolute values of quantitative relaxometry MRI techniques may also arise from the software used to generate the T1 maps (14, 46, 47). But an analysis of the performance of the software generating the maps is beyond the scope of this work. The study of the two patient groups of brain and male pelvis only included a small population. These results need to be confirmed by a study on a larger sample of patients, and it would be interesting to study other applications for radiotherapy such as dose calculation and treatment simulation. In addition, we compared the relaxation time measurements between the two measurement methods by manually drawing VOIs in each anatomical area. Although we used average VOI values for our analyses, this method may have measurement variabilities between the different qMRI mappings obtained intra- and inter-patient.

5 Conclusions

Our results showed good agreement between the MAGiC® method and the conventional sequences with respect to the reference values for T1 values between 500 and 1,304 ms and T2 values between 25 and 95 ms, with good reproducibility and repeatability at 1.5 T. Along with the QAp results, our work showed that it is now possible to use these imaging methods for anatomical segmentation for radiotherapy treatment planning. The results of the patient study are encouraging for the use of the MAGiC® sequence with a BODY coil for prostate qMRI and should be confirmed in a study with a larger population.

Data Availability Statement

The original contributions presented in the study are included in the article/Supplementary Material. Further inquiries can be directed to the corresponding author.

Ethics Statement

The studies involving human participants were reviewed and approved by IRB Henri Becquerel Center. The patients/participants provided written informed consent to participate in this study. Written informed consent was obtained from the individual(s) for the publication of any potentially identifiable images or data included in this article.

Author Contributions

PG, IG, and DG: conceived and designed this study. PG, AD, and DG: material preparation and data collection. PG and ET: statistical analysis. PG, SH, ST, PV, PH, and DG: visualization, manuscript writing, and submission. All authors contributed to the article and approved the submitted version.

Conflict of Interest

A scientific collaboration between Henri Becquerel Center and Olea Medical® allowed the use of OleaNova+® software for this study.

The authors declare that the research was conducted in the absence of any commercial or financial relationships that could be construed as a potential conflict of interest.

Publisher’s Note

All claims expressed in this article are solely those of the authors and do not necessarily represent those of their affiliated organizations, or those of the publisher, the editors and the reviewers. Any product that may be evaluated in this article, or claim that may be made by its manufacturer, is not guaranteed or endorsed by the publisher.

Acknowledgments

The authors thank Dr Lionel Nicol of SC IRM Petit Animal-IRIB Inserm U1096, Faculty of Medicine and Pharmacy, Rouen, France, and Sonia Saï of the Department of Radiology (Henri Becquerel Center), for their help in this study.

Supplementary Material

The Supplementary Material for this article can be found online at: https://www.frontiersin.org/articles/10.3389/fonc.2022.841761/full#supplementary-material

Supplementary Video 1

MRI synthetic contrast obtained with the MAGiC® sequence on a brain.

Supplementary Video 2

MRI synthetic contrast obtained with the MAGiC® sequence on a prostate.

References

  • 1. Mills AF, Sakai O, Anderson SW, Jara H. Principles of Quantitative MR Imaging With Illustrated Review of Applicable Modular Pulse Diagrams. RadioGraphics (2017) 37(7):2083–105. doi:  10.1148/rg.2017160099 [DOI] [PubMed] [Google Scholar]
  • 2. Nunez-Gonzalez L, Kotek G, Gomez PA, Buonincontri G, Vogel M, Krestin GP, et al. Accuracy and Repeatability of QRAPMASTER and MRF-vFA. Magn Reson Imag (2021) 83(October 2021):196–207. doi:  10.1016/j.mri.2021.09.004 [DOI] [PubMed] [Google Scholar]
  • 3. Krauss W, Gunnarsson M, Andersson T, Thunberg P. Accuracy and Reproducibility of a Quantitative Magnetic Resonance Imaging Method for Concurrent Measurements of Tissue Relaxation Times and Proton Density. Magn Reson Imag (2015) 33(5):584–91. doi:  10.1016/j.mri.2015.02.013 [DOI] [PubMed] [Google Scholar]
  • 4. Palma G, Tedeschi E, Borrelli P, Cocozza S, Russo C, Liu S, et al. A Novel Multiparametric Approach to 3D Quantitative MRI of the Brain. PloS One (2015) 10(8):1–20. doi:  10.1371/journal.pone.0134963 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 5. Dinis Fernandes C, van Houdt PJ, Heijmink SWTPJ, Walraven I, Keesman R, Smolic M, et al. Quantitative 3T Multiparametric MRI of Benign and Malignant Prostatic Tissue in Patients With and Without Local Recurrent Prostate Cancer After External-Beam Radiation Therapy. J Magn Reson Imag (2019) 50(1):269–78. doi:  10.1002/jmri.26581 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 6. West J, Warntjes JBM, Lundberg P. Novel Whole Brain Segmentation and Volume Estimation Using Quantitative MRI. Eur Radiol (2012) 22(5):998–1007. doi:  10.1007/s00330-011-2336-7 [DOI] [PubMed] [Google Scholar]
  • 7. Kurz C, Buizza G, Landry G, et al. Medical Physics Challenges in Clinical MR-Guided Radiotherapy. Radiat Oncol (2020) 15(1):1–16. doi:  10.1186/s13014-020-01524-4 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 8. Foltz WD, Wu A, Chung P, Catton C, Bayley A, Milosevic M, et al. Changes in Apparent Diffusion Coefficient and T2 Relaxation During Radiotherapy for Prostate Cancer. J Magn Reson Imag (2013) 37(4):909–16. doi:  10.1002/jmri.23885 [DOI] [PubMed] [Google Scholar]
  • 9. Gonçalves FG, Serai SD, Zuccoli G. Synthetic Brain MRI: Review of Current Concepts and Future Directions. Top Magn Reson Imag (2018) 27(6):387–93. doi:  10.1097/RMR.0000000000000189 [DOI] [PubMed] [Google Scholar]
  • 10. Lee SH, Lee YH, Song H-T, Suh J-S. Quantitative T2 Mapping of Knee Cartilage: Comparison Between the Synthetic MR Imaging and the CPMG Sequence. Magn Reson Med Sci (2018) 17(4):1–6. doi:  10.2463/mrms.tn.2017-0121 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 11. Kumar NM, Fritz B, Stern SE, Warntjes M, Chuah Y, Fritz J. Synthetic MRI of the Knee: Phantom Validation and Comparison With Conventional MRI. Radiology (2018) 289(2):465–77. doi:  10.1148/radiol.2018173007 [DOI] [PubMed] [Google Scholar]
  • 12. Jiang Y, Yu L, Luo X, Lin Y, He B, Wu B, et al. Quantitative Synthetic MRI for Evaluation of the Lumbar Intervertebral Disk Degeneration in Patients With Chronic Low Back Pain. Eur J Radiol (2020) 124:108858. doi:  10.1016/j.ejrad.2020.108858 [DOI] [PubMed] [Google Scholar]
  • 13. Cui Y, Han S, Liu M, Wu P, Zhang W, Zhang J, et al. Diagnosis and Grading of Prostate Cancer by Relaxation Maps From Synthetic MRI. J Magn Reson Imag (2020) 52(2):552–64. doi:  10.1002/jmri.27075 [DOI] [PubMed] [Google Scholar]
  • 14. Stikov N, Boudreau M, Levesque IR, Tardif CL, Barral JK, Pike GB. On the Accuracy of T1mapping: Searching for Common Ground. Magn Reson Med (2015) 73(2):514–22. doi:  10.1002/mrm.25135 [DOI] [PubMed] [Google Scholar]
  • 15. Warntjes JBM, Dahlqvist Leinhard O, West J, Lundberg P. Rapid Magnetic Resonance Quantification on the Brain: Optimization for Clinical Usage. Magn Reson Med (2008) 60(2):320–9. doi:  10.1002/mrm.21635 [DOI] [PubMed] [Google Scholar]
  • 16. HealthCare GE. MAGiC Technical Overview. Milwaukee, WI, USA: GE Healthcare; (2016). [Google Scholar]
  • 17. Li C-W, Hsu A-L, Huang C-WC, Yang S-H, Shieh C-C, Chan W-P, et al. Reliability of Synthetic Brain MRI for Assessment of Ischemic Stroke With Phantom Validation of a Relaxation Time Determination Method. J Clin Med (2020) 9(6):1857. doi:  10.3390/jcm9061857 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 18. Kim S-H, Park J-S, Park C-R, Jang J-S, Kwon S-Y, Yang H-J. Accuracy Confirmation of Relaxation Time Mapping in Synthetic MRI. J Korean Phys Soc (2019) 74(9):892–900. doi:  10.3938/jkps.74.892 [DOI] [Google Scholar]
  • 19. Chang KJ, Kamel IR, Macura KJ, Bluemke DA. 3.0-T MR Imaging of the Abdomen: Comparison With 1.5 T. Radiographics (2008) 28(7):1983–98. doi:  10.1148/rg.287075154 [DOI] [PubMed] [Google Scholar]
  • 20. Guarnaschelli JN, Vagal AS, McKenzie JT, et al. Target Definition for Malignant Gliomas: No Difference in Radiation Treatment Volumes Between 1.5T and 3T Magnetic Resonance Imaging. Pract Radiat Oncol (2014) 4(5):e195–201. doi:  10.1016/j.prro.2013.11.003 [DOI] [PubMed] [Google Scholar]
  • 21. West J, Blystad I, Engström M, Warntjes JBM, Lundberg P. Application of Quantitative MRI for Brain Tissue Segmentation at 1.5 T and 3.0 T Field Strengths. PloS One (2013) 8(9):e74795. doi:  10.1371/journal.pone.0074795 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 22. Johnstone RI, Guerrero-Urbano T, Michaelidou A, et al. Guidance on the Use of MRI for Treatment Planning in Radiotherapy Clinical Trials. Br J Radiol (2020) 93(1105). doi:  10.1259/bjr.20190161 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 23. Firbank MJ, Harrison RM, Williams ED, Coulthard A. Quality Assurance for MRI: Practical Experience. Br J Radiol (2000) 73(868):376–83. doi:  10.1259/bjr.73.868.10844863 [DOI] [PubMed] [Google Scholar]
  • 24. Stupic KF, Ainslie M, Boss MA, et al. A Standard System Phantom for Magnetic Resonance Imaging. Magn Reson Med (2021) 86(3):1194–211. doi:  10.1002/mrm.28779 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 25. Sacolick LI, Wiesinger F, Hancu I, Vogel MW. B 1 Mapping by Bloch-Siegert Shift. Magn Reson Med (2010) 1322:1315–22. doi:  10.1002/mrm.22357 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 26. Keenan KE, Biller JR, Delfino JG, et al. Recommendations Towards Standards for Quantitative MRI (qMRI) and Outstanding Needs. J Magn Reson Imaging (2019) 49(7):e26–e39. doi:  10.1002/jmri.26598 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 27. Vangel MG. Confidence Intervals for a Normal Coefficient of Variation. Am Stat (1996) 50(1):21–6. doi:  10.1080/00031305.1996.10473537 [DOI] [Google Scholar]
  • 28. Bland JM, Altman DG. Measuring Agreement in Method Comparison Studies. Stat Methods Med Res (1999) 8(2):135–60. doi:  10.1177/096228029900800204 [DOI] [PubMed] [Google Scholar]
  • 29. Tukey JW. Exploratory Data Analysis, Vol. 2. Reading, Mass: Addison-Wesley Pub. Co. (1977). [Google Scholar]
  • 30. Hasgall PA, Di Gennaro F, Baumgartner C, Neufeld E, Lloyd B, Gosselin MC, et al. It’is Database for Thermal and Electromagnetic Parameters of Biological Tissues. , Version 4.0. doi:  10.13099/VIP21000-04-0. [DOI] [Google Scholar]
  • 31. Shah B, Anderson SW, Scalera J, Jara H, Soto JA. Quantitative MR Imaging: Physical Principles and Sequence Design in Abdominal Imaging. Radiographics (2011) 31(3):867–80. doi:  10.1148/rg.313105155 [DOI] [PubMed] [Google Scholar]
  • 32. Zavala J, Stéphanie B, François B, Paul B, Lalande A. A Novel Alternative to Classify Tissues From T 1 and T 2 Relaxation Times for Prostate MRI. Magn Reson Mater Physics Biol Med (2016) 29(5): 777–88. doi:  10.1007/s10334-016-0562-3 [DOI] [PubMed] [Google Scholar]
  • 33. Ullrich T, Quentin M, Oelers C, Dietzel F, Sawicki LM, Arsov C, et al. Magnetic Resonance Imaging of the Prostate at 1.5 Versus 3.0 T: A Prospective Comparison Study of Image Quality. Eur J Radiol (2017) 90:192–7. doi:  10.1016/j.ejrad.2017.02.044 [DOI] [PubMed] [Google Scholar]
  • 34. Koopman D, Jager PL, Slump CH, Knollema S, van Dalen JA. SUV Variability in EARL-Accredited Conventional and Digital PET. EJNMMI Res (2019) 9(1). doi:  10.1186/s13550-019-0569-7 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 35. Keenan K, Boos M, Jackson EF, Kown S-J, Jennings D, Russek S. NIST/ISMRM MRI System Phantom T1 Measurements on Multiple MRI System. Proc 21th Sci Meet Int Soc Magn Reson Med (2013) 21:4338. [Google Scholar]
  • 36. Ma D, Gulani V, Seiberlich N, Liu K, Sunshine J-L, Duerk J-L, et al. Magnetic Resonance Fingerprinting. Nature (2013) 495(7440):187–92. doi:  10.1038/nature11971 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 37. Captur G, Bhandari A, Brühl R, Ittermann B, Keenan K-E, Yang Y, et al. T 1 Mapping Performance and Measurement Repeatability: Results From the Multi-National T 1 Mapping Standardization Phantom Program (T1MES). J Cardiovasc Magn Reson (2020) 22(1):1–17. doi:  10.1186/s12968-020-00613-3 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 38. Thrower SL, Al Feghali KA, Luo D, et al. The Effect of Slice Thickness on Contours of Brain Metastases for Stereotactic Radiosurgery. Adv Radiat Oncol (2021) 6(4):100708. doi:  10.1016/j.adro.2021.100708 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 39. Paštyková V, Novotný J, Veselský T, Urgošík D, Liščák R, Vymazal J. Assessment of MR Stereotactic Imaging and Image Co-Registration Accuracy for 3 Different MR Scanners by 3 Different Methods/Phantoms: Phantom and Patient Study. J Neurosurg (2018) 129(Suppl1):125–32. doi:  10.3171/2018.7.GKS181527 [DOI] [PubMed] [Google Scholar]
  • 40. Wang J, Mao W, Qiu M, Smith MB, Constable RT. Factors Influencing Flip Angle Mapping in MRI: RF Pulse Shape, Slice-Select Gradients, Off-Resonance Excitation, and B0 Inhomogeneities. Magn Reson Med (2006) 56(2):463–8. doi:  10.1002/mrm.20947 [DOI] [PubMed] [Google Scholar]
  • 41. Deppe MH, Teh K, Parra-Robles J, Lee KJ, Wild JM. Slice Profile Effects in 2D Slice-Selective MRI of Hyperpolarized Nuclei. J Magn Reson (2010) 202(2):180–9. doi:  10.1016/j.jmr.2009.11.003 [DOI] [PubMed] [Google Scholar]
  • 42. Ma D, Coppo S, Chen Y, McGivney D, Jiang Y, Pahwa S, et al. Slice Profile and B1 Corrections in 2D Magnetic Resonance Fingerprinting (MRF). Magn Reson Imag (2017) 78):1781–9. doi:  10.1002/mrm.26580.Slice [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 43. Bontempi P, Scartoni D, Amelio D, Cianchetti M, Turkaj A, Amichetti M, et al. Multicomponent T(2) Relaxometry Reveals Early Myelin White Matter Changes Induced by Proton Radiation Treatment. Magn Reson Med (2021) 86(6):3236–45. doi:  10.1002/mrm.28913 [DOI] [PubMed] [Google Scholar]
  • 44. Mai J, Abubrig M, Lehmann T, Hilbert T, Weiland E, Grimm M-O, et al. T2 Mapping in Prostate Cancer. Invest Radiol (2019) 54(3):146–52. doi:  10.1097/RLI.0000000000000520 [DOI] [PubMed] [Google Scholar]
  • 45. Chatterjee A, Devaraj A, Mathew M, et al. Performance of T2 Maps in the Detection of Prostate Cancer. Acad Radiol (2018) 7:1–7. doi:  10.1016/j.acra.2018.04.005 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 46. Head J. Cooperative NRD. A Multi-Institutional Comparison of Dynamic Contrast-Enhanced Magnetic Resonance Imaging Parameter Calculations. Sci Rep (2017) 7(1):11185. doi:  10.1038/s41598-017-11554-w [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 47. Wang Y, Tadimalla S, Rai R, Goodwin J, Foster S, Liney G, et al. Quantitative MRI: Defining Repeatability, Reproducibility and Accuracy for Prostate Cancer Imaging Biomarker Development. Magn Reson Imaging (2020) 77:169–79. doi:  10.1016/j.mri.2020.12.018 [DOI] [PubMed] [Google Scholar]

Associated Data

This section collects any data citations, data availability statements, or supplementary materials included in this article.

Supplementary Materials

Supplementary Video 1

MRI synthetic contrast obtained with the MAGiC® sequence on a brain.

Supplementary Video 2

MRI synthetic contrast obtained with the MAGiC® sequence on a prostate.

Data Availability Statement

The original contributions presented in the study are included in the article/Supplementary Material. Further inquiries can be directed to the corresponding author.


Articles from Frontiers in Oncology are provided here courtesy of Frontiers Media SA

RESOURCES