Abstract
In this paper, the classical least-squares (CLS) method with molecular absorption spectrophotometric measurement was used to determine simultaneously paracetamol (PAR), ibuprofen (IBU), and caffeine (CAF) in tablets. The absorbance spectra of the standard solutions and samples were measured over a wavelength from 220 to 300 nm with a 0.5 nm step. The concentration of PAR, IBU, and CAF in the sample solutions was calculated by using Visual Basic for Applications (VBA) and a program called CLS-Excel written in Microsoft Excel 2016. The method and the CLS-Excel program were tested on mixed standard laboratory samples with different PAR, IBU, and CAF concentration ratios, and they showed only small errors and a satisfying repeatability. An analytical procedure for tablets containing PAR, IBU, and CAF was developed. The reliability of the procedure was proved via the recovery and repeatability of the analysis results with an actual tablet sample and by comparing the mean contents of active substances in the tablets obtained from the analytical procedure with the HPLC method. The procedure is simple with a reduced cost compared with the HPLC standard method.
Keywords: paracetamol, ibuprofen, caffeine, classical least-square, simultaneous, spectroscopy
1. Introduction
Paracetamol (PAR), ibuprofen (IBU), and caffeine (CAF) are the main active ingredients widely used in multicomponent pharmaceuticals. PAR is a common pain reliever and fever reducer. IBU is a nonsteroidal anti-inflammatory drug with good analgesic and antipyretic effects. CAF is a methylated xanthine that stimulates the central nervous system, reduces feelings of fatigue and drowsiness, and increases brain excitement and sensory perception, thereby helping humans to work more effectively. The combination of these ingredients in tablets enhances the healing effect [1].
Quality control of multicomponent pharmaceutical products requires fast and reliable analytical techniques. The UV-Vis spectroscopy method is commonly used in laboratories due to its simplicity and low equipment cost. However, the quantitative analysis of pharmaceutical products containing many components having overlap spectra is often difficult. To analyze them by conventional UV-Vis method, we must often extract specific substance or mask substances which interfere with the analytical procedure. Thus, the procedure becomes complicated, consuming much time, chemicals, and solvent, with poor reliability. Currently, numerous UV-Vis spectroscopy methods combined with chemometrics have been developed to analyze simultaneously substances with overlapping absorption spectra. These methods often use entire spectrum data and computer programs to calculate, eliminate measurement errors, and statistically assess a large amount of data to give reliable and useful information. In particular, they allow us to calculate the concentrations of substances in multicomponent solutions with high accuracy without separation or masking. This advantage enables researchers to design simple, low-cost, short analytical procedures with high reliability. The UV-Vis molecular absorption spectrophotometric method is coupled with chemometrics used for simultaneous determination of substances in multicomponent pharmaceuticals. Such chemometrics include the classical least-squares (CLS) [1,2], partial least-squares (PLS) [2,3,4], principal component regression (PCR) [2,4], artificial neural network (ANN) [1,4], derivative [5,6,7], and Kalman filter [8] methods, as well as others. As we know, the absorption spectra of PAR, IBU, and CAF overlap to a great extent in the ultraviolet region. Many methods have been developed for the determination of PAR, IBU, and CAF in multicomponent drugs, including standard methods [9,10], spectroscopy [3,5,6,7], and chromatography [11]. To the best of our knowledge, there is no published paper concerning the use of the CLS method with full spectrum to simultaneously determine PAR, IBU, and CAF in drugs. Thus, in this paper we apply the CLS method for the full spectrum to simultaneously determine PAR, IBU, and CAF in drugs using Visual Basic for Applications (VBA) and a self-written program called CLS-Excel written on Microsoft Excel 2016.
2. Apparatus, Materials and Methods
2.1. Apparatus and Chemicals
2.1.1. Apparatus
A Cary 60 UV-Vis spectrophotometer (Agilent, Santa Clara, CA, USA) was in a wavelength range of 190–990 nm, connected to a computer with Cary WinUV software for storing spectral data as an excel spreadsheet. Other basic laboratory equipment was also used.
2.1.2. Chemicals
Paracetamol 99.9%, ibuprofen 100.1%, and caffeine 99.9% conforming with Vietnamese pharmaceutical standards were supplied from the Central Institute for Drug Testing, Vietnam.
A drug sample of Ibuparavic, containing paracetamol (300 mg/tablet), ibuprofen (200 mg/tablet), and caffeine (20 mg/tablet), was purchased from Thanh Nam Pharmaceutical Manufacturing and Trading Co., Ltd., Ho Chi Minh City, Vietnam. The production batch number was 601119, and the sample was produced on 11 January 2019 and expired on 10 January 2022. A box has 10 blisters with 10 hard capsules each, and the registration number is GC 318-19.
Distilled water and methanol (Merck) were also used.
Paracetamol, Ibuprofen, and Caffeine Standard Solutions
First, stock solutions with a 50 µg/mL concentration were prepared as follows: precisely 12.5 mg of each preparation was placed in a 250 mL volumetric flask with methanol, appropriately shaken, and made up to the mark. Then 50 mL of each solution was transferred to a 100 mL volumetric flask and made up to the mark with methanol to obtain a 25 µg/mL working solution. Finally, 10 mL of the working solution was placed in a 25 mL volumetric flask and made up to the mark with methanol to a 10 µg/mL PAR and IBU standard solution. For preparing a 5 µg/mL CAF standard solution, 5 mL of the working solution was used.
Mixed Experimental Solutions
The working solution of PAR, IBU, and CAF was mixed with different volume ratios. The standard and working solutions were used to verify the reliability of the method.
2.2. Analytical Procedure
The theoretical basis of the classical least-squares method is as follows:
For multicomponent systems, the absorbance is cumulative. We use Beer’s law for a system of n components and m wavelengths (m > n). Let ei = εi × b, Yi = Ai, and xi = Ci, where εiis the molecular absorptivity of the i-th component; Ci is the concentration of the ith component in the mixture; and Ai is the absorbance of the mixed solution measured at the ith wavelength. A system of linear equations is obtained with m equations and n unknowns as follows:
| (1) |
The molecular absorbance measured at the jth wavelength is yj. This parameter is often erroneous, and it is different from the actual value Yj by a value sj, where sj is the measurement residual:
| sj = yj − Yj | (2) |
The function representing the total squared error S is
| (3) |
For S to be minimized, the derivative of S with respect to xi must be 0. If we take the derivative of S with respect to x1 and let the derivative equal 0, we get the following equation:
Transforming this equation, we get
| (4) |
Similarly, we also take the derivative S with respect to the remaining xi and let these derivatives equal 0. Combining this equation with Equation (4), we get the following system of equations:
| (5) |
Let
| (6) |
where ; .
The system of equations can be summarized as follows:
| (7) |
The values of aki and bk in the system of Equation (7) are calculated from the initial experimental values of eji by using Equation (6). The system of Equation (7) is a system of linear equations consisting of n equations with n unknowns. Solving this system of equations with the Gaussian reduction method, we have the concentration of the components xi. The concentration of the components in the sample solution was calculated by using the CLS-Excel program.
The advantage of this method is that it uses all spectral data to create a system of linear equations with more equations than unknowns. Then, by transforming this system of equations with the least-squares technique, we obtain a system with an equal number of equations and unknowns. As a result, the error becomes minimal, thus enhancing accuracy. The concentration of the substances in the sample solution is determined rapidly thanks to the program. The method can be applied to the substances in the mixtures with the components’ complex absorption spectra overlapping.
The steps for measuring and calculating the concentration of substances are as follows:
Preparing standard solutions of each component to be determined and the sample solutions containing their mixtures.
Scanning the spectrum of the solutions at an appropriate wavelength range to obtain CSV files in the form of an excel spreadsheet.
Running the CLS-Excel program for the data from the excel files to calculate the concentration of components in the mixed solution and their relative error.
2.3. Statistical Parameters
2.3.1. Relative Error
The relative error between the determined concentration and the preparation concentration (RE%) was calculated according to Equation (8)
| (8) |
where C is the determined concentration (µg/mL) and C0 is the concentration of the known standard solution (µg/mL).
2.3.2. Repeatability
Repeatability was assessed by using the relative standard deviation value (RSD%):
| (9) |
where S is the standard deviation and Cmean is the mean concentration after n measurements (µg/mL). For in-laboratory quality control, method repeatability is satisfactory when the RSD% values obtained are less than 1/2RSDHorwitz [12,13]
| RSDHorwitz = 2(1−0.5×lgC) | (10) |
where C is the concentration expressed as a power (for example, C = 5 µg/mL = 5 × 10–6).
2.3.3. Accuracy
-
a.
Recovery
The recovery of the method was calculated based on the standard addition according to Equation (11)
| (11) |
where C2 (µg/mL) is the determined concentration of the sample solution after standard addition; C1 (µg/mL) is the determined concentration of the sample solution before standard addition; and Cadd (µg/mL) is the standard addition concentration [14].
-
b.
Comparison of the proposed method with the HPLC standard method
The basic information of the HPLC standard method [10] to analyze the tablet containing PAR, IBU, and CAF is as follows:
First: Determination of IBU only: Stationary phase: C18 (100 × 4.6 mm; 5 µm); Mobile phase: Acid phosphoric 0.01 M: Acetonitril 60:40 (V:V); Detector: Diode Array, UV at λ = 224 nm; Flowrate: 1.0 mL/min;
Second: simultaneous determination of PAR and CAF: Stationary phase: C18 (100 × 4.6 mm; 5 µm); Mobile phase: Water-methanol-glacial acid acetic (69:28:3) (V); Detector: Diode Array at λ = 275 nm; Flowrate: 2.0 mL/min.
According to [15], to determine the method’s accuracy, we analyze the same sample repeatedly with the proposed method and the standard method. Then we compare the two sample mean values by using the Student’s t-test.
| (12) |
where texp is the experimental student value; and are the mean value of methods A and B; nA and nB are the number of repeat measurements of methods A and B; and , are the variance of the two methods.
Finally, we compare the texp value with the theoretical student value t(α, ν), where α is the significance level (usually taken as 0.05) and ν is the degrees of freedom determined above. If texp < t(α,ν), the mean values of the two methods are not significantly different.
2.4. Actual Sample Treatment and Calculation of the Content of Substances
2.4.1. Sample Treatment
Twenty tablets from the same production batch were weighed, and the average weight was determined (M). Then the tablets were ground to fine powder in an agate mortar. An amount of powder equal to 0.7 to 1.0 of the average tablet weight was placed into a 250 mL beaker containing 150 mL of methanol. The content of the beaker was sonicated for 30 min and quantitatively transferred to a 250 mL volumetric flask, made up to the mark with methanol and thoroughly mixed. The solution was then filtered through blue-band filter paper; the first 10 mL of the filtrate was discarded. Next, 10 mL of the filtrate was transferred to a 100 mL volumetric flask, made up to the mark with methanol, and thoroughly mixed (solution 1). Again, 10 mL of solution 1 was diluted to 100 mL with methanol to obtain solution 2. Finally, solution 2 was subjected to UV-Vis absorption determination, and the CLS-Excel program (please see the Supplementary Materials) was used to calculate the concentration of the active substances. The concentration of the active ingredients from another 20 pills from the same batch was determined simultaneously with the HPLC method.
2.4.2. Calculation of the Content of Substances
The content of active ingredients in one tablet was determined from the formula (13)
| H (mg/tablet) = Cm × 100 × (100/10) × (250/10) × (1/1000) × (M/m) = 25 × Cm × (M/m) | (13) |
where Cm (µg/mL) is the concentration of each active ingredient determined in the sample solution; m is the weight of the sample (mg); and M is the average tablet weight (mg).
3. Results and Discussion
3.1. Accuracy and Repeatability of the Analytical Method on Laboratory Samples
From the working standards of PAR 25 μg/mL, IBU 25 μg/mL, and CAF 25 μg/mL, prepare individually 10 µg/mL PAR and IBU standard solutions and 5 µg/mL CAF as described in Section 2.1.2 and their mixture solutions at different concentration ratios (Table 1). The standard solutions and the mixture solutions were measured three times. The solutions were spectroscopically scanned in the range of 220–300 nm with 0.5 nm intervals. The relative error between the determined concentration and the preparation concentration of PAR, IBU, and CAF in the mixed solutions was calculated according to the CLS-Excel program, and the corresponding relative standard deviation (RSD%) of the analytical results was also calculated. The absorption spectra of the standard solutions and laboratory mixture solutions are illustrated in Figure 1. The concentrations of PAR, IBU, and CAF in the mixtures and statistical data are presented in Table 1.
Table 1.
Concentration of PAR, IBU, and CAF in the mixture, and their RE and statistics.
| Sample | Conc. Ratio (μg/mL) PAR/IBU/CAF | Run Order | PAR | IBU | CAF | ||||||
|---|---|---|---|---|---|---|---|---|---|---|---|
| CPAR (μg/mL) | RE (%) | Statistics | CIBU (μg/mL) | RE (%) | Statistics | CCAF (μg/mL) | RE (%) | Statistics | |||
| M1 | 16:12:7 | 1 2 3 |
16.091 16.179 16.161 |
0.57 1.12 0.82 |
Cmean = 16.144 RSD (%) = 0.288 1/2RSDH = 5.270 REmean (%) = 0.84 |
11.858 11.892 11.885 |
−1.18 −0.93 −0.96 |
Cmean = 11.878 RSD (%) = 0.151 1/2RSDH = 5.504 REmean (%) = −1.02 |
6.989 6.989 7.000 |
−0.16 −0.16 0.00 |
Cmean = 6.993 RSD (%) = 0.091 1/2RSDH = 5.969 REmean (%) = −0.11 |
| M2 | 14:10:6 | 1 2 3 |
13.992 14.059 14.013 |
−0.06 0.42 0.09 |
Cmean = 14.021 RSD (%) = 0.244 1/2RSDH = 5.378 REmean (%) = 0.15 |
10.004 9.981 10.002 |
0.04 −0.19 0.02 |
Cmean = 9.996 RSD (%) = 0.127 1/2RSDH = 5.657 REmean (%) = −0.07 |
5.991 5.980 5.990 |
−0.15 −0.33 −0.17 |
Cmean = 5.987 RSD (%) = 0.102 1/2RSDH = 6.109 REmean (%) = −0.22 |
| M3 | 12:8:5 | 1 2 3 |
11.943 12.008 11.964 |
−0.47 0.07 −0.30 |
Cmean = 11.992 RSD (%) = 0.277 1/2RSDH = 5.504 REmean (%) = −0.23 |
8.023 7.996 7.989 |
0.29 −0.05 −0.14 |
Cmean = 8.003 RSD (%) = 0.224 1/2RSDH = 5.850 REmean (%) = 0.03 |
4.957 4.958 4.974 |
−0.86 −0.84 −0.52 |
Cmean = 4.963 RSD (%)= 0.192 1/2RSDH = 6.279 REmean (%) = −0.71 |
| M4 | 10:6:4 | 1 2 3 |
9.979 10.035 10.004 |
−0.21 0.35 0.04 |
Cmean = 10.006 RSD (%) = 0.280 1/2RSDH = 5.657 REmean (%) = 0.06 |
6.040 6.052 6.044 |
0.67 0.87 0.73 |
Cmean = 6.045 RSD (%) = 0.101 1/2RSDH = 6.109 REmean (%) = 0.76 |
3.961 3.959 3.969 |
−0.98 −1.03 −0.78 |
Cmean = 3.963 RSD (%) = 0.134 1/2RSDH = 6.493 REmean (%) = −0.91 |
| M5 | 8:4:3 | 1 2 3 |
7.899 7.937 7.909 |
−1.26 −0.79 −1.14 |
Cmean = 7.918 RSD (%) = 0.247 1/2RSDH = 5.850 REmean (%) = −1.02 |
4.023 4.024 4.014 |
0.58 0.60 0.35 |
Cmean = 4.020 RSD (%) = 0.129 1/2RSDH = 6.493 REmean (%) = 0.50 |
2.971 2.967 2.971 |
−0.97 −1.10 −0.97 |
Cmean = 2.970 RSD (%) = 0.089 1/2RSDH = 6.781 REmean (%) = −0.93 |
| M6 | 6:2:2 | 1 2 3 |
5.973 6.027 6.007 |
−0.45 0.45 0.12 |
Cmean = 6.002 RSD (%)= 0.455 1/2RSDH = 6.109 REmean (%) = 0.04 |
1.998 2.002 1.992 |
−0.10 0.10 −0.40 |
Cmean = 1.997 RSD (%) = 0.252 1/2RSDH = 7.207 REmean (%) = −0.13 |
2.003 1.999 2.003 |
0.15 −0.05 0.15 |
Cmean = 2.002 RSD (%) = 0.115 1/2RSDH = 7.207 REmean (%) = 0.08 |
| M7 | 4:10:1 | 1 2 3 |
4.009 4.012 4.010 |
0.23 0.30 0.25 |
Cmean = 4.010 RSD (%)= 0.038 1/2RSDH = 6.493 REmean (%) = 0.26 |
10.053 9.989 10.035 |
0.53 −0.11 0.35 |
Cmean = 10.026 RSD (%) = 0.329 1/2RSDH = 5.657 REmean (%) = 0.26 |
1.006 0.997 1.005 |
0.60 −0.30 0.50 |
Cmean = 1.003 RSD (%) = 0.492 1/2RSDH = 8.000 REmean (%) = 0.27 |
| M8 | 15:10:1 | 1 2 3 |
14.929 15.060 14.876 |
−0.47 0.40 −0.83 |
Cmean = 14.955 RSD (%)= 0.633 1/2RSDH = 5.322 REmean (%) = −0.30 |
9.984 10.010 9.915 |
−0.16 0.10 −0.85 |
Cmean = 9.970 RSD (%) = 0.492 1/2RSDH = 5.657 REmean (%) = 0.10 |
1.008 0.997 0.986 |
0.80 −0.30 −1.40 |
Cmean = 0.997 RSD (%) = 1.103 1/2RSDH = 8.000 REmean (%) = −0.30 |
| M9 | 8:5:0.5 | 1 2 3 |
8.004 7.981 8.044 |
0.05 −0.24 0.55 |
Cmean = 8.010 RSD (%)= 0.398 1/2RSDH = 5.849 REmean (%) = 0.12 |
4.974 5.023 4.981 |
−0.52 0.46 −0.38 |
Cmean = 4.993 RSD (%) = 0.531 1/2RSDH = 6.280 REmean (%) = −0.15 |
0.507 0.497 0.498 |
1.40 −0.60 −0.40 |
Cmean = 0.501 RSD (%) = 1.099 1/2RSDH = 8.877 REmean (%) = 0.133 |
Note: The number of decimal places is taken to represent the calculation result.
Figure 1.
UV absorption spectra of standard solutions and laboratory mixture solutions with different concentration ratios (Standard solution (µg/mL): PAR 10, IBU 10, CAF 5; PAR/IBU/CAF mixed solution (µg/mL): M1 (16:12:7); M2 (14:10:6); M3 (12:8:5); M4 (10:6:4); M5 (8:4:3); M6 (6:2:2); M7 (4:10:1); M8 (15:10:1; M9 (8:5:0.5)).
Table 1 shows that at different concentration ratios the errors of the concentrations of PAR, IBU, and CAF determined with the CLS method are from −1.40 to 1.12% and that the RSD% values are also small (RSD%max = 1.103) and less than 1/2RSDHorwitz. Therefore, the method’s accuracy and repeatability are satisfactory for mixed laboratory solutions with different concentration ratios.
3.2. Simultaneous Quantification of PAR, IBU, and CAF in Drug Samples
The characteristics of ibuparavic tablets were described in Section 2.1.2 with the average tablet weight of 0.5265 g.
The samples were treated as described in Section 2.4.1 with precisely 526.5 mg of powder. The entire spectrum of the sample solution was scanned in the wavelength range of 220–300 nm, with a 0.5 nm step. The concentration of PAR, IBU, and CAF in the sample solutions was determined with the CLS-Excel program, and their content was calculated from the Formula (13).
The absorption spectra of the standard solutions and sample solutions of ibuparavic are presented in Figure 2, and the content of the active ingredient is displayed in Table 2.
Figure 2.
Absorption spectra of standard solutions and sample solutions of ibuparavic (standard solution 10 µg/mL: PAR (1), IBU (2); standard solution 5 µg/mL: CAF (3); S1: Ibuparavic drug sample solution (4)).
Table 2.
Concentration of PAR, IBU, and CAF in sample solutions and their drug content in Ibuparavic tablets.
| Sample | PAR | IBU | CAF | |||
|---|---|---|---|---|---|---|
|
CPAR (µg/mL) |
Content (mg/Tablet) |
CIBU (µg/mL) |
Content (mg/Tablet) |
CCAF (µg/mL) |
Content (mg/Tablet) |
|
| S1 | 11.467 | 286.68 | 7.765 | 194.13 | 0.812 | 20.30 |
| S2 | 11.503 | 287.58 | 7.749 | 193.73 | 0.795 | 19.88 |
| S3 | 11.463 | 286.58 | 7.825 | 195.63 | 0.802 | 20.05 |
| Mean | 11.478 | 286.95 | 7.780 | 194.50 | 0.803 | 20.08 |
| RSD% | 0.192 | 0.515 | 1.052 | |||
| 1/2RSDH | 5.541 | 5.875 | 8.269 | |||
| %H* | 95.65 | 97.25 | 100.40 | |||
Note: %H*: % active ingredient compared with labelled content.
The data show that the method is highly reproducible with all three components (RSD% < 1.2). The content of each substance in the ibuparavic tablets is as follows: PAR: 286.95 ± 1.37 mg, IBU: 194.50 ± 2.49 mg, and CAF 20.08 ± 0.52 mg. This content is consistent with that reported on the label of these tablets and also agrees with the quality standards required by Vietnam’s Ministry of Health: PAR 300 mg ± 5% (285–315 mg), IBU 200 mg ± 5% (190–210 mg), and CAF 20 mg ± 5% (19–21 mg).
3.3. Accuracy Verification
3.3.1. Recovery
Four batches of the sample powder equal to 0.7 times the average tablet weight were weighed. No standard addition was carried out for the first batch. The remaining three batches were added with PAR, IBU, and CAF with increasing amounts of standard. The samples were treated as described in Section 2.4.1. Measurements were carried out for the spectra of standard solutions PAR 10 µg/mL, IBU 10 µg/mL, and CAF 5 µg/mL, the sample solution without standard (S0), and sample solutions after adding standards (S1, S2, S3). The concentration of PAR, IBU, and CAF in the standard and sample solutions was calculated with the CLS-Excel program. The spectra of the standard solutions and the sample solutions are shown in Figure 3. The concentration of the standard additions and that of the sample without and with the added standard is presented in Table 3. The recovery of the CLS-Excel method was calculated from Equation (11).
Figure 3.
Absorbance spectra of standard solutions, sample solution, and standard addition solutions (standard solution (µg/mL): PAR 10, IBU 10, CAF 5; S0: drug sample without standard S0 (m = 0.7 Mtablets); PAR/IBU/CAF mixed solution (µg/mL): S1 (2:2:0.5); S2 (4:4:1); S3 (6:6:1.5)).
Table 3.
Recovery of CLS-Excel method applied for analyzing Ibuparavic tablets.
| Repeated Sample | PAR | IBU | CAF | ||||||
|---|---|---|---|---|---|---|---|---|---|
|
Cadded (µg/mL) |
Cmeasured (µg/mL) |
Rev (%) |
Cadded (µg/mL) |
Cmeasured (µg/mL) |
Rev (%) |
Cadded (µg/mL) |
Cmeasured (µg/mL) |
Rev (%) |
|
| S01 | 0 | 8.045 | 0 | 5.462 | – | 0 | 0.562 | ||
| S02 | 8.000 | – | 5.502 | 0.568 | – | ||||
| S03 | 7.987 | 5.483 | 0.564 | ||||||
| Statistics | Cmeasured (mean) = 8.011 | Cmeasured (mean) = 5.482 | Cmeasured (mean) = 0.565 | ||||||
| RSD% = 0.380 | RSD% = 0.360 | RSD% = 0.540 | |||||||
| S11 | 2.000 | 9.901 | 92.80 | 2.000 | 7.475 | 100.65 | 0.500 | 1.035 | 94.60 |
| S12 | 9.916 | 95.80 | 7.543 | 102.05 | 1.057 | 97.80 | |||
| S13 | 9.944 | 98.30 | 7.564 | 104.05 | 1.064 | 100.00 | |||
| Statistics | Cmeasured (mean) = 9.920 | Cmeasured (mean) = 7.527 | Cmeasured (mean) = 1.052 | ||||||
| RSD% = 0.22 | RSD% = 0.61 | RSD% = 1.44 | |||||||
| Revmean (%) = 95.63 | Revmean (%) = 102.25 | Revmean (%) = 97.47 | |||||||
| S21 | 4.000 | 11.824 | 94.48 | 4.000 | 9.551 | 102.23 | 1.000 | 1.578 | 101.60 |
| S22 | 11.713 | 92.83 | 9.538 | 100.90 | 1.585 | 101.70 | |||
| S23 | 11.711 | 93.10 | 9.536 | 101.33 | 1.582 | 101.80 | |||
| Statistics | Cmeasured (mean) = 11.749 | Cmeasured (mean) = 9.542 | Cmeasured (mean) = 1.582 | ||||||
| RSD% = 0.55 | RSD% = 0.09 | RSD% = 0.22 | |||||||
| Revmean (%) = 93.47 | Revmean (%) = 101.48 | Revmean (%) = 101.70 | |||||||
| S31 | 6.000 | 13.893 | 97.46 | 6.000 | 11.247 | 96.42 | 1.500 | 2.074 | 100.80 |
| S32 | 13.777 | 96.28 | 11.246 | 95.73 | 2.093 | 101.67 | |||
| S33 | 13.774 | 96.45 | 11.238 | 95.92 | 2.093 | 101.80 | |||
| Statistics | Cmeasured (mean) = 13.815 | Cmeasured (mean) = 11.238 | Cmeasured (mean) = 2.087 | ||||||
| RSD% = 0.49 | RSD% = 0.04 | RSD% = 0.52 | |||||||
| Revmean (%) = 96.73 | Revmean (%) = 96.02 | Revmean (%) = 101.42 | |||||||
Note: The number of decimal places is taken to represent the calculation result.
Table 3 shows that the method’s recovery is satisfactory: 92.80–98.30% for PAR, 95.73–104.05% for IBU, and 94.60–101.80% for CAF. All recovery values are within the allowable range required by AOAC [13].
3.3.2. Comparison of CSL-Excel and HPLC Methods
To objectively evaluate the accuracy of our method, we compared the content of the active ingredients in Ibuparavic tablets with those determined with the standard HPLC method performed by the Centre for Drug, Food, and Cosmetic Testing in Thua Thien Hue, Vietnam [10]. The comparison was carried out statistically [15] (Table 4).
Table 4.
Comparison of CLS and HPLC methods.
| N0 | Content (H, mg/Tablet) | |||||
|---|---|---|---|---|---|---|
| PAR | IBU | CAF | ||||
| CLS | HPLC | CLS | HPLC | CLS | HPLC | |
| 1 | 286.68 | 287.98 | 194.13 | 193.02 | 20.30 | 20.37 |
| 2 | 287.58 | 285.26 | 193.73 | 195.90 | 19.88 | 20.19 |
| 3 | 286.58 | 289.18 | 195.63 | 195.69 | 20.05 | 19.98 |
| Hmean | 286.95 | 287.47 | 194.50 | 194.87 | 20.08 | 20.18 |
| RSD (%) | 0.19 | 0.73 | 0.51 | 0.83 | 1.05 | 0.99 |
| tcal | 0.438 | 0.342 | 0.622 | |||
| ttheory (0.05;4) | 2.78 | 2.78 | 2.78 | |||
| p | 0.68 | 0.75 | 0.57 | |||
The results in Table 4 show that the calculated t-values are smaller than the t-theory values, indicating that the CLS and HPLC methods are statistically identical at α = 0.05. Thus, we can say that the method has a satisfying accuracy.
4. Conclusions
An analytical procedure for simultaneous determination of PAR, IBU, and CAF in tablets was developed by using the molecular absorption spectrophotometric method with the entire spectrum, coupled with the classical least-squares technique. The concentration of PAR, IBU, and CAF in the sample solutions was calculated by using Visual Basic for Applications (VBA) and a self-made program called CLS-Excel written in Microsoft Excel 2016. The analytical procedure has satisfactory repeatability with an RSD% less than or equal to 1.052. The recoveries obtained for PAR, IBU, and CAF ranged from 92.80 to 98.30, 95.73 to 104.05, and 94.60 to 101.80%, respectively. The content of PAR, IBU, and CAF in the drug sample Ibuparavic analyzed with the procedure is consistent with that of the HPLC method at the 0.05 significance level.
Supplementary Materials
The following supporting information can be downloaded at: https://www.mdpi.com/article/10.3390/molecules27092657/s1, Kalman and CLS-Excel Program.csv; EXP1(Lab Mix).csv.
Author Contributions
Data curation, A.M.T.T.; Formal analysis, B.T.T.; Investigation, Q.T.T.N.; Methodology, B.T.T. and Q.T.T.N.; Software, B.T.T.; Supervision, T.N.T.; Writing—original draft, G.C.D.N.; Writing—review & editing, B.T.T. All authors have read and agreed to the published version of the manuscript.
Funding
This paper was funded by the Vietnam Ministry of Education and Training with the code no. B2020-DHH-16.
Data Availability Statement
The data used to support the finding of this study are available from the corresponding author upon request.
Conflicts of Interest
The authors declare that they have no conflict of interest.
Sample Availability
Samples of the Ibuparavic tablets containing PAR, IBU and CAF are available from the authors.
Footnotes
Publisher’s Note: MDPI stays neutral with regard to jurisdictional claims in published maps and institutional affiliations.
References
- 1.Khoshayand M.R., Abdollahi H., Shariatpanahi M., Saadatfard A., Mohammadi A. Simultaneous spectrophotometric determination of paracetamol, ibuprofen and caffeine in pharmaceuticals by chemometric methods. Spectrochim. Acta. 2008;70:491–499. doi: 10.1016/j.saa.2007.07.033. [DOI] [PubMed] [Google Scholar]
- 2.Basu D., Mahalanabis K.K., Roy B. Application of least-squares method in matrix form-simultaneous determination of ibuprofen and paracetamol in tablets. J. Pharm. Biomed. Anal. 1998;16:809–812. doi: 10.1016/S0731-7085(97)00111-8. [DOI] [PubMed] [Google Scholar]
- 3.Darwish H.W., Bakheit A.H., Abdelhameed A.S., Mothana R.A. Application of Classical Least Squares, Principal Component Regression and Partial Least Squares Methods for Simultaneous Spectrophotometric Determination of Rutin and Ascorbic Acid in Their Combined Dosage Form. Life Sci. J. 2013;10:1680–1686. [Google Scholar]
- 4.Aktaş A.H., Kitiş F. Spectrophotometric Simultaneous Determination of Caffeine and Paracetamol in Commercial Pharmaceutical by Principal Component Regression, Partial Least Squares and Artificial Neural Networks Chemometric Methods. Croat. Chem. Acta. 2014;87:69–74. doi: 10.5562/cca2214. [DOI] [Google Scholar]
- 5.Dang H.V., Thu H.T.T., Ha L.D.T., Mai H.N. RP-HPLC and UV Spectrophotometric Analysis of Paracetamol, Ibuprofen, and Caffeine in Solid Pharmaceutical Dosage Forms by Derivative, Fourier, and Wavelet Transforms: A Comparison Study. J. Anal. Methods Chem. 2020;2020:8107571. doi: 10.1155/2020/8107571. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 6.Hajian R., Afshari N. The Spectrophotometric Multicomponent Analysis of a Ternary Mixture of Ibuprofen, Caffeine and Paracetamol by the Combination of Double Divisor-Ratio Spectra Derivative and H-Point Standard Addition Method. E-J. Chem. 2012;9:1153–1164. doi: 10.1155/2012/621263. [DOI] [Google Scholar]
- 7.Alkhafaji S.L., Mahood A.M. First-order derivative and UV-spectrophotometric methods for simultaneous determina-tion of paracetamol, ibuprofen, and caffeine in bulk and pharmaceutical formulations. Int. J. Pharm. Res. 2018;25:1–14. doi: 10.9734/JPRI/2018/46503. [DOI] [Google Scholar]
- 8.Trang N.T.Q., Van Hop N., Chau N.D.G., Tran T.B. Simultaneous Determination of Amlodipine, Hydrochlorothiazide, and Valsartan in Pharmaceutical Products by a Combination of Full Spectrum Measurement and Kalman Filter Algorithm. Adv. Mater. Sci. Eng. 2019;2019:8107571. doi: 10.1155/2019/5719651. [DOI] [Google Scholar]
- 9.British Pharmacopoeia 2014. British Pharmacopoeia Commission; London, UK: 2014. [Google Scholar]
- 10.Vietnamese Pharmacopoeia V. Vietnamese Pharmacopoeia Commission V, Ministry of Health, Medicine Publisher; Hanoi City, Vietnam: 2017. [Google Scholar]
- 11.Cunha R.R., Chaves S.C., Ribeiro M.M.A.C., Torres L.M.F.C., Munoz R.A.A., Dos Santos W.T.P., Richter E.M. Simultaneous determination of caffeine, paracetamol, and ibuprofen in pharmaceutical formulations by high-performance liquid chromatography with UV detection and by capillary electrophoresis with conductivity detection. J. Sep. Sci. 2015;38:1657–1662. doi: 10.1002/jssc.201401387. [DOI] [PubMed] [Google Scholar]
- 12.Thompson M., Lowthian P.J. A Horwitz-like function describes precision in a proficiency test. Analyst. 1995;120:271–272. doi: 10.1039/an9952000271. [DOI] [Google Scholar]
- 13.AOAC International . AOAC® Guidelines for Single Laboratory Validation of Chemical Methods for Dietary Supplements and Botanicals. AOAC International; Gaithersburg, MD, USA: 2012. [Google Scholar]
- 14.Brereton R.G. Chemometrics: Data Analysis for the Laboratory and Chemical Plant. John Wiley & Sons Ltd.; Chichester, UK: 2003. [Google Scholar]
- 15.Harvey D. Modern Analytical Chemistry. 1st ed. McGraw Hill; New York, NY, USA: 2000. pp. 80–93. [Google Scholar]
Associated Data
This section collects any data citations, data availability statements, or supplementary materials included in this article.
Supplementary Materials
Data Availability Statement
The data used to support the finding of this study are available from the corresponding author upon request.