CAN THE SIZE-SPECIFIC DOSE ESTIMATE BE DERIVED FROM THE BODY MASS INDEX? A FEASIBILITY STUDY

Beatrice Steiniger 1,, Chris Klippel 2, Ulf Teichgräber 3, Jürgen R Reichenbach 4, Martin Fiebich 5
PMCID: PMC9113340  PMID: 35443046

Abstract

Size-specific dose estimate (Inline graphic) index appears to be more suitable than the commonly used volume computed tomography dose index (Inline graphic) to estimate the dose delivered to the patient during a computed tomography (CT) scan. We evaluated whether an Inline graphic can be determined from the patient’s body mass index (Inline graphic) with sufficient reliability in the case that a Inline graphic is not given by the CT scanner. For each of the three most used examination types, CT examinations of 50 female and 50 male patients were analyzed. The Inline graphic values automatically provided by the scanner were compared with Inline graphic determined from Inline graphic and Inline graphic. A good accordance of Inline graphic and Inline graphic was found for the chest and abdominal regions. A low correlation was observed for the head region. The presented method is a simple and practically useful surrogate approach for the chest and abdominal regions but not for the head.

INTRODUCTION

The volumetric computed tomography dose index (Inline graphic) and the associated dose length product (Inline graphic) are routine dosimetry metrics that are currently used to monitor radiation exposure to patients undergoing computed tomography (CT)(1, 2). However, Inline graphic is of limited value as it does not represent the amount of exposure an individual patient receives but only indicates the amount of radiation that a cylindrical polymethyl-methacrylate (PMMA) phantom with a diameter of either 16 cm or 32 cm would receive(3–5). For this reason, the size-specific dose estimate (Inline graphic) index was introduced, which incorporates the water-equivalent skin-to-skin diameter of the patient(6, 7). This methodology is an IEC standard since 2019(8).

The advantage of Inline graphic is that individual patient size is used to estimate the dose. This is an important feature because Inline graphic, which is independent of patient size and scan length, would indicate too low a dose for a selected examination in slim patients, while it would be the opposite in obese patients. Dose management systems are currently being established to report radiation exposure for further evaluation when diagnostic reference values are significantly exceeded without reason(9). Improved estimates of the exposure are certainly necessary to automatically and reliably identify unjustified high doses of radiation. For the latter, Inline graphic seems well suited because it is patient size-specific.

However, the calculation of the Inline graphic may fail if the patient’s body contour cannot be clearly identified in the overview images or is outside of the overview images and the diameter in the center slice is therefore incorrectly determined or not determined at all. Although methods have been developed which recognize these problems and improve the calculations, they have not yet been implemented in the scanners currently in use(10). In addition, the calculation of the diameter and thus the Inline graphic from all CT slices is time-consuming and only possible afterward(11).

Therefore, it seems an interesting approach to estimate the Inline graphic without information about the water-equivalent diameter of the patient. Although other research groups have already addressed this issue, they have done so with different approaches, most of which did not differentiate by gender or include pediatric patients(12–16). Several research groups already addressed the correlation of body mass index (Inline graphic) and water-equivalent diameter but considered only a specific body region, limited themselves to individual types of CT examination or compared results from different scanners.

In this study, we evaluate a simple approach for female and male patients using Inline graphic to estimate the Inline graphic. Inline graphic is readily available and can often be retrieved from the hospital information system to estimate Inline graphic from Inline graphic for the most frequently examined body regions of head, chest and abdomen. When CT scanners and dose management systems do not provide this value, this could be beneficial.

MATERIALS AND METHODS

This retrospective study complies with the Declaration of Helsinki and was carried out with the approval of the Ethics Committee of the Friedrich Schiller University Jena. All patients provided written informed consent.

Patient population

For each of head, chest and abdomen, 100 CT examinations were retrospectively selected from the hospital database for 50 female and 50 male adult patients (see Table 1). The sample size was chosen to obtain a reasonable statistical result with a reasonable amount of information collection. The biometric data for calculating the Inline graphic were taken from the electronic health records and were not from the period of 2 weeks before to 2 weeks after the examination. None of the patients had any prosthesis or other implants within the scan field. The baseline patient data and the scan parameters are summarized in Table 1.

Table 1.

Baseline patient data and scan parameters.

Head
Female (n = 50) Male (n = 50) Scan parameter
Mean ± SD Range Mean ± SD Range Tube voltage (kVp) 120
Age 66.4 ± 14.1 21–92 67 ± 15.5 28–88 Effective mAs (mean) 313
BMI 26.7 ± 5.9 15.6–44.2 26.4 ± 4.7 16.2–43.3 Detector coverage (mm) 80
CTDI 53.8 ± 6.6 40.9–78.9 52.9 ± 4.4 42.0–82.9 Table increment (mm) 80
SSDE (CT) 49.8 ± 5.6 40.3–73.4 48.2 ± 4.0 39.7–57.2 Scanmode Step and shoot, axial
Chest
Female (n = 50) Male (n = 50) Scan parameter
Mean ± SD Range Mean ± SD Range Tube voltage (kVp) 120
Age 64.6 ± 12.4 36–89 65.1 ± 12.1 27–88 Effective mAs (mean) 96
BMI 26.8 ± 5.6 15.9–42.0 27.1 ± 4.0 20–38.4 Detector coverage (mm) 80
CTDI 5.0 ± 2.8 1.9–11.8 6.8 ± 2.4 2.8–12.9 Pitch factor 0.99
SSDE (CT) 5.1 ± 2.1 2.6–9.4 6.5 ± 1.7 3.3–9.6 Scanmode Spiral
Abdomen
Female (n = 50) Male (n = 50) Scan parameter
Mean ± SD Range Mean ± SD Range Tube voltage (kVp) 120
Age 66.2 ± 12.0 33–90 66.1 ± 10.9 37–82 Effective mAs (mean) 109
BMI 24.4 ± 4.9 17.2–37.6 27.0 ± 5.6 17.8–39.9 Detector coverage (mm) 80
CTDI 6.4 ± 1.8 4.4–11.1 8.6 ± 3.2 4.5–16.8 Pitch factor 0.99
SSDE (CT) 7.2 ± 1.1 5.7–9.8 8.4 ± 1.8 5.8–12.7 Scanmode Spiral

Technical specification

All scans had been performed with the same CT scanner (Revolution CT, General Electric Company, USA), which automatically provides Inline graphic and Inline graphic. The scanner is a 256 slice system and has a patient coverage in z-direction of up to 160 mm, resulting in 0.625 mm thickness per detector element.

Inline graphic and Inline graphic were determined by the scanner for patient-mimicking cylindrical PMMA phantoms with diameter of 16 and 32 cm for the head and body regions, respectively(3). By contrast, the American Association of Physicists in Medicine (AAPM) Report 220 proposes to use the patient’s effective water-equivalent diameter, Inline graphic, which can be extracted from localizer images in lateral and anterior–posterior directions or from cross-sectional CT images to calculate Inline graphic(7). The scanner used in our work calculates Inline graphic as

graphic file with name DmEquation1.gif (1)

using the patient’s diameter in lateral direction, Inline graphic, and in anterior–posterior direction, Inline graphic, as illustrated on a central slice of an exemplary scan region (center-slice method) in Figure 1.

Figure 1.

Figure 1

Localizer images for an abdominal scan; the black lines show the diameter of the patient in the central slice position; left: lateral direction (Inline graphic); right: anterior–posterior direction (Inline graphic); the Inline graphic values output by the CT scanner used in this study are calculated using this center-slice method.

In the AAPM Report 220, conversion factors, Inline graphic, are available for both PMMA phantoms for values of Inline graphic in the range from 8 to 45 cm to calculate Inline graphic from the Inline graphic values output from the scanner:

graphic file with name DmEquation2.gif (2)

The conversion factors, Inline graphic, are based on a curve fit between physical measurements and Monte Carlo simulations(6, 7). However, in cases where Inline graphic is not automatically calculated, we propose a simple method to derive Inline graphic from the patient’s Inline graphic and the Inline graphic output from the scanner, where the metric, Inline graphic (index Inline graphic indicates calculation based on the patient’s individual Inline graphic) is calculated as follows:

graphic file with name DmEquation3.gif (3)

The Inline graphic was calculated according to the equation

graphic file with name DmEquation4.gif (4)

To determine the conversion factors Inline graphic used by our CT scanner, the ratios between the available Inline graphic and Inline graphic values were calculated for all patients in each anatomic group (i.e. head, chest, and abdomen):

graphic file with name DmEquation5.gif (5)

In addition to Report 220, the AAPM released Report 293 for head CT(17). The conversion factor, Inline graphic, is calculated there with the formula

graphic file with name DmEquation6.gif (6)

Most of the currently used scanners calculating the Inline graphic still according to AAPM Report 220 for the head section, so we decided to consider both approaches for the head section for this work.

A similar functional relationship to the Inline graphic has now been examined. The approach should work if Inline graphic and Inline graphic are linearly related. This was initially checked visually as shown using a scatterplot. To distinguish, the conversion factor is now referred to as Inline graphic. To determine the functions for Inline graphic for each group, the calculated gender-separated values Inline graphic (n = 50 each) were plotted against the corresponding individual Inline graphic values and were then fitted to an exponential function of the form

graphic file with name DmEquation7.gif (7)

where Inline graphic and Inline graphic are fit parameters. MATLAB 2017a was used for these curve fits to extract the values of Inline graphic and Inline graphic for both genders and each target region (see Figure 4). With known parameters Inline graphic and Inline graphic, the fit function can be used to later calculate Inline graphic from Inline graphic according to Eq. (3).

Figure 4.

Figure 4

Scatterplots between Inline graphic and Inline graphic for all target regions; the lines indicate the fit to the data; the formulas developed of these fits can be used to calculate Inline graphic from the Inline graphic data of the patients; for the diagrams of the head section, light gray is the data set calculated according to AAPM 220, black stand for calculations with conversion factors as described in AAPM Report 293.

Statistical analysis

To assess the validity of our approach, the Inline graphic values were compared to the corresponding Inline graphic values by calculating the coefficient of determination R2 of the linear fit through the point of origin. A coefficient of determination of ≥0.25 was classified as weak correlation, ≥0.5 as moderate and ≥0.75 as substantial correlation(18).

In addition, Bland–Altman diagrams were created to show the variation in the results. For this purpose, the mean of two corresponding values (Inline graphic and Inline graphic) was plotted against the difference of the two values. For normally distributed differences, one would expect 95% of the differences to lie between two lines, calculated as

graphic file with name DmEquation8.gif (8)

where Inline graphic is the difference between Inline graphic and Inline graphic, and Inline graphic denotes the mean of all individual Inline graphic for each target region and gender. Almost all pairs of calculation of the two methods should ideally be closer to each other than these extreme values, which are called the ‘95% limits of agreement’(19).

RESULTS

Figure 2 shows the distributions of the water-equivalent diameter Inline graphic determined by CT and Inline graphic for all examinations. As expected, the values for Inline graphic for the head are distinctly less scattered compared to the corresponding values for Inline graphic, indicating the low correlation between head diameter and Inline graphic.

Figure 2.

Figure 2

Diagrams showing (a) the distribution of Inline graphic as calculated by the scanner and (b) the Inline graphic for all patient groups; each small horizontal line segment represents the data of one patient.

Figure 3 shows the scatterplots between Inline graphic and Inline graphic for the different body regions together with the linear regression lines. From these graphs, as well as from the R2 values of the linear fit functions, the poor correlation in the head region between Inline graphic and Inline graphic is clear. In order to show to what extent this affects the calculation of the Inline graphic, we have nevertheless continued the evaluation for the head region. The linear correlation is moderate for both the thoracic and abdominal regions, with a slightly better fit for the abdominal region in female patients and for the thoracic region in male patients.

Figure 3.

Figure 3

Scatterplots between Inline graphic and Inline graphic for all target regions and both genders; the black solid lines represent the linear fit; the goodness of that linear fit is given as R2 in the right edge of each diagram.

Figure 4 shows the scatterplots between the conversion factors Inline graphic derived from Eq. (5) and the corresponding Inline graphic for all patients and body regions together with the fit curves according to Eq. (7). The coefficients of determination R2 of the fit functions between Inline graphic and Inline graphic are also given for each equation. All results for chest and abdomen can be classified as moderate correlation. The fit result is best for the abdominal region in females and is best for the chest region in males. For the head region, the results corroborate the expected missing correlation between the conversion factors Inline graphic and Inline graphic. The results for estimations according to the older AAPM Report 220 are reproduced in gray color. Comparing the R2, we see that the current recommendation of the AAPM to calculate the Inline graphic with the method described in Report 293 does not improve the results with our method. Although the use of R2 is not optimal for non-linear models, as it then often simulates better results, it is still a quick and easy way to get an initial overview by comparing data series(20).

In Figure 5, the results for the calculated Inline graphic are compared with the Inline graphic. Correlations for the head region are again worst but can be classified as moderate. Although the previous findings of this study tend to suggest that the abdomen region might give the best results in the female group. Looking at R2 between Inline graphic and Inline graphic, it shows an substantial result in the chest region and only a weak correlation of both values for the abdominal region. The result of the assessment of the Inline graphic for the male group is substantial in the chest region and is moderate for the abdominal region.

Figure 5.

Figure 5

Scatterplots between Inline graphic and Inline graphic for all three anatomic target regions and both gender; the black dashed lines represent the lines of identity; R2 stands for the determination coefficients between Inline graphic and Inline graphic; for the diagrams of the head section, light gray is the data set calculated according to AAPM 220, black stand for calculations with conversion factors as described in AAPM Report 293.

As seen in Figure 6, 95% of the cases lie within the boundary lines (i.e. two to three values in this case) for all study groups. Thus, when assessing the results using this method, one can assume that an approximation of the Inline graphic with the help of the Inline graphic is quite practicable even for the head region. In view of the previous results, however, this should be questioned critically.

Figure 6.

Figure 6

Bland–Altman-diagrams comparing Inline graphic and Inline graphic; the dots represent the result of a corresponding pair of Inline graphic and Inline graphic; the solid lines indicate the mean of the differences between corresponding Inline graphic and Inline graphic, and the dashed lines represent the mean of the difference plus or minus 1.96 multiplied with the standard deviation of the differences; for the diagrams of the head section, light gray is the data set calculated according to AAPM 220, black stand for calculations with conversion factors as described in AAPM Report 293.

DISCUSSION

The proposed method is a quick and easy alternative to evaluate size-specific doses for patients with conspicuous Inline graphic values in situations where an automatically calculated Inline graphic is not provided by the CT scanner. It is not a substitute for an exact calculation, as recommended in the AAPM report, but can help physicians or radiographers to estimate the acceptability of an applied dose considering a reference level.

An advantage of our approach is that it uses the Inline graphic in combination with derived equations for different body regions without the need to know the patient’s effective diameter. However, this approach is also limited because the Inline graphic for a small person (child or small adult) could be the same as for a tall person, while the effective diameter and thus the exposed volume would most likely be significantly different. Parikh et al. investigated the extent to which body weight can also be used to determine Inline graphic in pediatric patients. The results differ depending on the patient’s body diameter(5). Therefore, to improve the approximation of our proposed Inline graphic to Inline graphic, more patients and groups including pediatric patients need to be investigated in future studies(21). Furthermore, the ratio of Inline graphic to Inline graphic can vary, depending on the proportions of muscle and fat tissue present and is age and gender effected(22). These differences can also affect the correct estimation of the Inline graphic. Another uncertainty for the calculated Inline graphic is possible weight loss, especially in oncology and intensive care patients, during the maximum allowable 2-week period between weight and height data collection and CT acquisition.

Considering only the Bland–Altman method to evaluate the results, one could come to the conclusion that the estimation of Inline graphic values for the head based on Inline graphic is unproblematic. However, in comparison with the other results, we cannot recommend this. The lack of correlation between Inline graphic and Inline graphic in the area of the head is particularly evident in Figure 3. The coefficient of determination here is slightly better for men than for women, but both values are extremely poor and show that a calculation using our method cannot produce any useful values for Inline graphic in the head region. There are recommendations to determine Inline graphic using specialized head scan band regardless of size and/or weight(23). For the head region, the equations used in our study to calculate Inline graphic show strikingly poor agreement with the values of Inline graphic, which is reflected in the low values for R2 (see Figure 4). The changes in AAPM Report 293 compared to Report 220 have only a small effect on our findings. It appears that estimating Inline graphic values of the head region from a patient’s Inline graphic is not accurate. As Figure 2 shows, the diameter of the head has only small variations in patients, although the Inline graphic varies relatively strongly, resulting in weak or no correlation. This observation is also supported by a study by Alikhani et al., who used an exponential relationship to calculate a size-specific Inline graphic from Inline graphic and assumed independence of both values for the head region(12). Mehdipour et al. have investigated how Inline graphic and lateral body diameter correlate. The correlation for the abdominal and for the chest region they found are also seen in our investigations. However, they did not include gender in their study, where our results show a difference between female and male patients(15). Xu et al. found that the Inline graphic is better suited than body weight to estimate Inline graphic. However, they calculated the Inline graphic from the patients water-equivalent diameter estimated from body weight and Inline graphic(16). In our study, we directly estimated the Inline graphic from the Inline graphic.

The results presented in Figure 5 show moderate correlations between Inline graphic and Inline graphic for the abdominal region. The best agreement with a substantial correlation was observed for the chest region, suggesting that Inline graphic for this region predicts Inline graphic well. The equations established with the scatterplots in Figure 4 provide feasible approaches for calculating Inline graphic in the chest and abdominal regions with almost similar accuracy as in the AAPM report. O’Neill et al. also investigated the applicability of Inline graphic to estimate the individual patient radiation dose by calculating an effective diameter from the Inline graphic and demonstrated a substantial correlation (r = 0.88) between these two values(13). However, they only evaluated abdominal scans in 50 patients (female and male varied) in their study.

Discrepancies between Inline graphic and Inline graphic may also be caused by the method implemented in the CT scanner to determine the effective diameter. The scanner used in our study determines the effective diameter from the central position of the scout images (see Figure 1), as described in the AAPM Task Group 204 report(6). Because of its higher accuracy, the AAPM recommends calculating Inline graphic using the slice-by-slice method from axial CT images(7, 24, 25). However, the disadvantage of the latter method is the longer calculation time and that the results are obtained only retrospectively. In most cases, Inline graphic calculated by the center-slice method is almost identical to the Inline graphic calculated by the slice-by-slice method, although it has been shown that they can differ by up to 3.7% in abdominal phantom studies(7). In patients, it could be even larger because the shape of the abdominal region varies more depending on weight, height and individual constitution. Boos et al. showed that the center-slice method results in a mean relative absolute error of 2–5% in pediatric and adult abdominal and chest CT compared with the slice-by-slice method(14). For pediatric thoracic, abdominal and pelvic scans, Öszoykal et al. determined a root-mean-squared error of 1.2–11% when comparing the Inline graphic from the center-slice method with the slice-by-slice determination(26). Furthermore, Inline graphic also depends on the positioning of the patient in the isocenter of the scanner gantry, which can lead to inaccuracies if performed incorrectly(7, 27, 28). The IEC standard for Inline graphic cites an included neck in the scanned anatomy, an actual scan length that exceeds the range of the overview radiogram, unilateral or bilateral limb scans or foreign objects located in the scan field as possible causes of miscalculation of the Inline graphic by the scanner software(8). These facts illustrate that the values for Inline graphic adopted as reference in our study also need to be reviewed and carefully evaluated.

In general, the use of Inline graphic is very helpful and diagnostic reference levels should be set to Inline graphic instead of Inline graphic. Although the Inline graphic is a useful indicator of radiation dose, it does not represent the dose received by the patient due to the neglect of size. Therefore, in very slim or obese patients, the Inline graphic should be compared with the Inline graphic to check the plausibility of the Inline graphic value output by the scanner. For the optimization process of various CT examinations, the Inline graphic is an important variable for adapting the radiation exposure to as many body constitutions as possible. As described by Lyra et al.(29), there is only a low correlation between the effective dose calculated with the Inline graphic compared to the Inline graphic. It can therefore be assumed that the value for the effective dose calculated from Inline graphic and Inline graphic should also be critically examined, especially for patients in whom Inline graphic and Inline graphic differ greatly from one another.

CONCLUSION

A patient’s Inline graphic can be used in CT scans of the chest and the abdomen to efficiently estimate Inline graphic This Inline graphic has an almost similar accuracy to that derived from images. For CT imaging of the skull or brain, using Inline graphic does not show good accuracy because the effective diameter of the head does not vary much in adults. When calculation of the Inline graphic from images is not possible, e.g. because body parts of the patient are not included in the image or localizer images, or because of missing software to calculate Inline graphic, the proposed methodology of using Inline graphic can be applied.

ACKNOWLEDGEMENTS

Not applicable.

Contributor Information

Beatrice Steiniger, Department of Diagnostic and Interventional Radiology, University Hospital Jena, Am Klinikum 1, Jena 07747, Germany.

Chris Klippel, Department of Diagnostic and Interventional Radiology, University Hospital Jena, Am Klinikum 1, Jena 07747, Germany.

Ulf Teichgräber, Department of Diagnostic and Interventional Radiology, University Hospital Jena, Am Klinikum 1, Jena 07747, Germany.

Jürgen R Reichenbach, Medical Physics Group, Department of Diagnostic and Interventional Radiology, University Hospital Jena, Philosophenweg 3, Jena 07743, Germany.

Martin Fiebich, Institute of Medical Physics and Radiation Protection, University of Applied Sciences Giessen, Wiesenstraße 14, Gießen 35390, Germany.

FUNDING

The authors state that this work has not received any funding.

References

  • 1. Shope, T. B., Gagne, R. M. and Johnson, G. C.  A method for describing the doses delivered by transmission x-ray computed tomography. Med. Phys.  8(4), 488–495 (1981). [DOI] [PubMed] [Google Scholar]
  • 2. McCollough, C.  et al.  Report no. 096—the measurement, reporting and management of radiation dose in CT. AAPM Rep. 6–17 (2008). [Google Scholar]
  • 3. IEC 60601–2–44:2017–03–08 . Medical electrical equipment Part 2–44: Particular requirements for the basic safety and essential performance of X-ray equipment for computed tomography. (2017).
  • 4. McCollough, C. H., Leng, S., Yu, L., Cody, D., Boone, J. M. and McNitt-Gray, M. F.  CT dose index and patient dose: they are NOT the same thing. Radiology  259(2), 311–316 (2011). [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 5. Parikh, R. A., Wien, M. A., Novak, R. D., Jordan, D. W., Klahr, P., Soriano, S., Ciancibello, L. and Berlin, S. C.  A comparison study of size-specific dose estimate calculation methods. Pediatr. Radiol.  48(1), 56–65 (2018). [DOI] [PubMed] [Google Scholar]
  • 6. Boone, J. M., Strauss, K. J., Cody, D. D., McCollough, C. H., McNitt-Gray, M. F., Toth, T. L., Goske, M. J. and Frush, D. P.  Report no. 204—size-specific dose estimates (SSDE) in pediatric and adult body CT examinations. AAPM Rep. 2–22 (2011). [Google Scholar]
  • 7. McCollough, C.  et al.  Use of water equivalent diameter for calculating patient size and size-specific dose estimates (SSDE) in CT: the report of AAPM task group 220. AAPM Rep.  2014, 6–23 (2014). [PMC free article] [PubMed] [Google Scholar]
  • 8. IEC62985 . Methods for calculating size specific dose estimates (SSDE) on computed tomography. (2019).
  • 9. Federal_office_of_radiation_protection Bekanntmachung der aktualisierten diagnostischen Refrenzwerte für diagnostische und interventionelle Röntgenanwendungen. ( BAnz AT 15.07.2016 B08; ) http://www.bfs.de/SharedDocs/Downloads/BfS/DE/fachinfo/ion/drw-roentgen.pdf?__blob=publicationFile&v=9  (Stand: 04.02.2022). 1–6 (2016). [Google Scholar]
  • 10. Anam, C., Mahdani, F. R., Dewi, W. K., Sutanto, H., Triadyaksa, P., Haryanto, F. and Dougherty, G.  An improved method for automated calculation of the water-equivalent diameter for estimating size-specific dose in CT. J. Appl. Clin. Med. Phys.  22(9), 313–323 (2021). [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 11. Anam, C., Haryanto, F., Widita, R., Arif, I. and Dougherty, G.  A fully automated calculation of size-specific dose estimates (SSDE) in thoracic and head CT examinations. J. Phys. Conf. Ser.  694, 012030 (2016). [Google Scholar]
  • 12. Alikhani, B., Getzin, T., Kaireit, T. F., Ringe, K. I., Jamali, L., Wacker, F., Werncke, T. and Raatschen, H. J.  Correlation of size-dependent conversion factor and body-mass-index using size-specific dose estimates formalism in CT examinations. Eur. J. Radiol.  100, 130–134 (2018). [DOI] [PubMed] [Google Scholar]
  • 13. O’Neill, S., Kavanagh, R. G., Carey, B. W., Moore, N., Maher, M. and O’Connor, O. J.  Using body mass index to estimate individualised patient radiation dose in abdominal computed tomography. Eur Radiol Exp.  2(1), 38 (2018). [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 14. Boos, J., Kropil, P., Bethge, O. T., Aissa, J., Schleich, C., Sawicki, L. M., Heinzler, N., Antoch, G. and Thomas, C.  Accuracy of size-specific dose estimate calculation from center slice in computed tomography. Radiat. Prot. Dosim.  178(1), 8–19 (2018). [DOI] [PubMed] [Google Scholar]
  • 15. Mehdipour, A., Parsi, M. and Khorram, F. S.  Patient dose survey based on size-specific dose estimate and acceptable quality dose in chest and abdomen/pelvis CT examinations. Radiat. Prot. Dosim.  185(2), 176–182 (2019). [DOI] [PubMed] [Google Scholar]
  • 16. Xu, J., Wang, X., Yang, P., Luo, K. and He, X.  Size-specific dose estimates of radiation based on body weight and body mass index for chest and abdomen-pelvic CTs. Biomed. Res. Int.  2020, 6046501 (2020). [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 17. Boone, J. M.  et al.  The report of AAPM task group 293: size-specific dose estimate (SSDE) for head CT. AAPM Rep. 6–19 (2019). [PMC free article] [PubMed] [Google Scholar]
  • 18. Hair, J. F., Ringle, C. M. and Sarstedt, M.  PLS-SEM: indeed a silver bullet. J. Mark. Theory Pract.  19(2), 139–152 (2011). [Google Scholar]
  • 19. Bland, J. M. and Altman, D. G.  Measuring agreement in method comparison studies. Stat. Methods Med. Res.  8(2), 135–160 (1999). [DOI] [PubMed] [Google Scholar]
  • 20. Spiess, A.-N. and Neumeyer, N.  An evaluation of R2 as an inadequate measure for nonlinear models in pharmacological and biochemical research: a Monte Carlo approach. BMC Pharmacol.  10(1), 6 (2010). [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 21. Menke, J.  Comparison of different body size parameters for individual dose adaptation in body CT of adults. Radiology  236(2), 565–571 (2005). [DOI] [PubMed] [Google Scholar]
  • 22. Zamboni, M., Zoico, E., Scartezzini, T., Mazzali, G., Tosoni, P., Zivelonghi, A., Gallagher, D., De Pergola, G., Di Francesco, V. and Bosello, O.  Body composition changes in stable-weight elderly subjects: the effect of sex. Aging Clin. Exp. Res.  15(4), 321–327 (2003). [DOI] [PubMed] [Google Scholar]
  • 23. Shohji, T., Kuriyama, K., Yanano, N. and Katoh, Y.  Development of a specialised tape measure to estimate the size-specific dose estimate. Radiat. Prot. Dosim.  191(3), 369–375 (2020). [DOI] [PubMed] [Google Scholar]
  • 24. Wang, J., Duan, X., Christner, J. A., Leng, S., Yu, L. and McCollough, C. H.  Attenuation-based estimation of patient size for the purpose of size specific dose estimation in CT. Part I. Development and validation of methods using the CT image. Med. Phys.  39(11), 6764–6771 (2012). [DOI] [PubMed] [Google Scholar]
  • 25. Wang, J., Christner, J. A., Duan, X., Leng, S., Yu, L. and McCollough, C. H.  Attenuation-based estimation of patient size for the purpose of size specific dose estimation in CT. Part II. Implementation on abdomen and thorax phantoms using cross sectional CT images and scanned projection radiograph images. Med. Phys.  39(11), 6772–6778 (2012). [DOI] [PubMed] [Google Scholar]
  • 26. Özsoykal, I., Yurt, A. and Akgungor, K.  Size-specific dose estimates in chest, abdomen, and pelvis CT examinations of pediatric patients. Diagn. Interv. Radiol.  24(4), 243–248 (2018). [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 27. Marsh, R. M. and Silosky, M. S.  The effects of patient positioning when interpreting CT dose metrics: a phantom study. Med. Phys.  44(4), 1514–1524 (2017). [DOI] [PubMed] [Google Scholar]
  • 28. Toth, T., Ge, Z. and Daly, M. P.  The influence of patient centering on CT dose and image noise. Med. Phys.  34(7), 3093–3101 (2007). [DOI] [PubMed] [Google Scholar]
  • 29. Lyra, M.  et al.  Effective dose and size-specific dose estimate (SSDE) of the torso in low dose MDCT protocol in multiple myeloma. Radiol Diagnostic Imaging  2(4) (2019). 10.15761/RDI.1000146. [DOI] [Google Scholar]

Articles from Radiation Protection Dosimetry are provided here courtesy of Oxford University Press