. 2022 Apr 20;15:100133. doi: 10.1016/j.ahjo.2022.100133

Comparison of clinical symptoms and bioimpedance to pulmonary capillary wedge pressure in heart failure

Monica Polcz a, Jessica Huston b, Meghan Breed c,d, Marisa Case d, Philip Leisy d, Jeffrey Schmeckpeper e, Lexie Vaughn a, Jenna Helmer Sobey d, Colleen Brophy a, JoAnn Lindenfeld e, Kyle Hocking a,f, Bret Alvis d,f,
PMCID: PMC9119644  NIHMSID: NIHMS1804282  PMID: 35600671

Abstract

Introduction

Clinical symptoms of heart failure commonly include fatigue, edema, and shortness of breath. Unfortunately, clinical monitoring has proven unreliable in predicting congestion and the need for hospitalization. Biosensing wearables have been developed as a potential adjunct to clinical signs and symptoms to detect congestion before it becomes severe thus preventing a heart failure hospitalization.

Hypothesis

Clinical signs and symptoms of heart failure will correlate with thoracic bioimpedance measurements (ZOE®) and pulmonary capillary wedge pressure (PCWP).

Methods

One hundred and fifty-five subjects undergoing right heart catheterization (RHC) were prospectively enrolled. A Zo value (ohms) was obtained, jugular venous pressure (JVP) was estimated, edema graded, and shortness of breath (SOB) assessed in all subjects. RHC was performed by a scheduled cardiologist per routine. One-way ANOVA was performed to assess the relationship between variables. A Pearson correlation coefficient was used to compare the Zo value and PCWP.

Results

Neither estimated JVP (cmH2O) (p = 0.65, n = 110) nor edema scores (p = 0.12, n = 110) demonstrated a significant relationship to PCWP. The presence of subjective SOB also did not demonstrate a significant association with PCWP (p = 0.99, n = 110). There was no correlation between ZOE® and PCWP (r = −0.08, p = 0.56, n = 56).

Conclusions

These findings support the idea that traditional measures for monitoring heart failure patients are limited.

Keywords: Congestion, Heart failure, Bioimpedance, PCWP, Readmission


\

1. Introduction

Heart failure (HF) is a complex clinical syndrome that involves either impairment of ventricular filling or ejection of blood [1]. Clinical symptoms of heart failure commonly include fatigue, edema, and shortness of breath, but these often represent late manifestations of the disease [1], [2]. Heart failure symptoms significantly decrease health-related quality of life (HRQOL) and thus significant focus has been placed on preventing their onset [1], [3]. Accurately assessing heart failure severity before symptom onset is difficult and there is no single diagnostic test or data point that has proven reliable [1]. According to the ACCF/AHA Guidelines [4], volume status should be assessed at every patient encounter with serial assessment of weight, edema, jugular venous pressure (JVP), and orthopnea (SOB) [1]. Secondary analysis of data from the ESCAPE [5] trial demonstrated JVP as the only useful surrogate for predicting pulmonary capillary wedge pressure (PCWP) >22 mmHg [6], [7]. Orthopnea was only predictive of PCWP >28 mmHg [6], [7]. While clinical monitoring can detect current congestive symptoms, it is unreliable for predicting subclinical congestion and guiding therapeutic decisions to prevent acute decompensated heart failure and the need for hospital admission in patients with HF [8], [9], [10], [11], [12], [13].

Biosensing wearables have been developed as a potential adjunct to clinical symptoms for the assessment of heart failure, especially in the outpatient setting. Measurement of bioelectrical impedance (BI) is an example of such technology that multiple devices are based upon [14]. The ZOE® (Noninvasive Medical Technologies, Las Vegas, NV, USA) is a BI device cleared by the FDA in 2004 which measures the resistance of the thoracic cavity via a measurement of the time it takes a 100 kHz electric current to travel from the top to the bottom of the thorax [15], [16]. Resistance of tissue is altered by its fluid content, and the device outputs a Zo value that theoretically inversely correlates with the fluid content of the lungs [15]. Normal resistance has been determined to be between 19 and 30 Ω [15], [16]. To date, despite being used as a predicate device on multiple FDA 510 K applications/clearance, the ZOE® device has not been extensively studied in heart failure. It has demonstrated some success in correlating chest radiographic findings associated with pulmonary edema [16] but the relationship to important central hemodynamic parameters (i.e. PCWP) is limited [17], [18].

This report provides prospectively collected data examining the correlation between clinical signs and symptoms of heart failure using simultaneous bioimpedance measurements (ZOE®), and PCWP obtained in subjects with heart failure undergoing elective right heart catheterization (RHC) at a single large volume academic center.

2. Methods

This prospective, observational study was approved by the University of Alabama Birmingham Institutional Review Board through Vanderbilt University Medical Center Institutional Review Board. One hundred and fifty-five subjects undergoing elective RHC were approached for enrollment in the study. All subjects were positioned in the semi-recumbent position with the head of bed between 45 and 90 degrees and requested to remain still. ZOE® leads were placed per manufacturer's instructions, with hair shaved as needed for appropriate lead contact, and the Zo value (ohms) was obtained and recorded by one of the study personnel (MP, PL). Three values were obtained, one immediately after another, and averaged for each patient. Subjects were also questioned regarding subjective resting SOB and a focused physical exam was performed where edema grade was noted [19] and JVP was estimated [7], [20], [21], [22]. Demographic information and comorbidities were obtained from the medical record. RHC was performed by the scheduled interventional cardiologist per standard technique using central vein cannulation with a pulmonary artery catheter (PAC; Edwards Life Sciences Corporation, Irvine, CA, USA) inserted with the balloon tip inflated into the pulmonary artery and into “wedge” position. All waveforms were obtained at end expiration in spontaneously breathing patients and recorded to the electronic medical record. RHC tracings were then reviewed and PCWP values were assigned by a single board certified intensivist specialized (BA) in cardiovascular medicine who was blinded to Zo value and physical examination findings.

2.1. Statistical analysis

Clinical variables (JVP, edema grade, and subjective SOB) were evaluated for normality using a D'Agostino-Pearson normality test and then by one-way analysis of variance (ANOVA) with multiple comparisons across means of groups to observe the relationship between the dependent and independent variables. Pearson correlation coefficient between Zo value and PCWP was calculated using GraphPad Prism software (San Diego, CA). P-values less than 0.05 and a log worth of greater than 1 were considered significant. Multivariate analysis using linear regression on the variables the log worth and p-value were calculated in the ability to predict PCWP to compare all groups with principal component analysis to display as directional vectors by utilizing the eigenvectors of the covariance matrix. Sensitivity, specificity, positive and negative predictive values were also calculated using ROC curves for the ability of each clinical variable to predict PCWP >22 mmHg [6], [7].

3. Results

Of the 155 enrolled, only 56 subjects had ZOE® measurements and 126 had interpretable PCWP tracings. Clinical signs and symptoms of congestion were obtained in 110 patients prior to right heart catheterization. Demographic information is listed in Table 1. Neither estimated JVP (cmH2O) (n = 110, p = 0.65; Fig. 1A) nor edema scores (n = 110, p = 0.12; Fig. 1B) demonstrated a significant relationship to PCWP [one way analysis of variance (ANOVA)]. The presence of subjective SOB also did not demonstrate significant association with PCWP (p = 0.99; n = 110 parametric Student's t-test, Fig. 1C). There was no correlation between ZOE® and PCWP (r = −0.08, n = 56, p = 0.56; Fig. 2).

Table 1.

Demographic table of the outpatient RHC cohort. This table demonstrates the subject characteristics as median (iqr) or n(%). Abbreviations: BMI = body mass index, PAD = pulmonary artery diastolic pressure, PCWP = pulmonary capillary wedge pressure, CI = cardiac index, EF = ejection fraction, ASA = American Society of Anesthesiology, RHC = right heart catheterization, HF = heart failure.

All Male Female
Age 56 (44–68) 58 (45–68) 53 (43–64)
Sex 69 (62.7%)
BMI (kg/m2) 29.3 (25.7–33.5) 29.2 (25.8–33.2) 29.5 (25.8–34.7)
PAD (mmHg) 16 (11.5–22) 16 (10−22) 18 (13−22)
PCWP (mmHg) 13.5 (9.75–18.25) 14 (9–19) 13 (10–18)
PCWP > 22 mmHg 15 (14%) 12 (17%) 3 (7%)
CI (L/min/m2) 2.46 (2.06–2.87) 2.49 (2.12–2.85) 2.35 (1.80–2.87)
EF (%)
 <35 23 (21%) 18 (26.1%) 5 (12.2%)
 35–45 7 (6.4%) 2 (2.9%) 5 (12.2%)
 45–55 19 (17.3%) 10 (14.5%) 9 (22.0%)
 >55 61 (55.5%) 39 (56.5%) 22 (53.6%)
ASA status
 1 0 0 0
 2 4 (3.6%) 2 (2.9%) 2 (4.9%)
 3 101 (91.8%) 64 (92.8%) 37 (90.2%)
 4 5 (4.5%) 3 (4.3%) 2 (4.9%)
RHC (n; %)
HF diagnostic evaluation 18 (16.3%) 5 (7.3%) 13 (31.7%)
HF maintenance care evaluation 30 (27.3%) 25 (36.2%) 5 (12.2%)
HF post-transplant graft evaluation 62 (56.4%) 39 (56.5%) 23 (56.1%)
TOTAL (n) 110 69 41

Fig. 1.

Fig. 1

Box-and-Whisker Plot of Analysis of Variance (ANOVA) for Jugular Venous Pressure (JVP; A), Edema Score (B), and Shortness of Breath (SOB; C) to Pulmonary Capillary Wedge Pressure (PCWP) in heart failure patients. JVP measured at time of right heart catheterization demonstrated no statistically significance and no correlation (n = 110, p = 0.65, r = 0.17). Edema scores measured at time of right heart catheterization by study personnel demonstrated no statistically significance (P = 0.12) and no correlation (n = 110; R2 = 0.07). Whether a patient verbalized subjective shortness of breath at time of right heart catheterization by study personnel demonstrated no statistically significance (n = 110; P = 0.99) compared.

Fig. 2.

Fig. 2

Pearson Correlation of ZOE® measurements compared to measured Pulmonary Capillary Wedge Pressure (PCWP) in heart failure patients. Relationship of bioimpendence as measured by the ZOE® device (Ohms) at time of right heart catheterization (RHC) demonstrated no correlation (n = 56. r = −0.08; P = 0.56).

Clinical variables also did not demonstrate a reliable ability to predict PCWP >22 mmHg (Table 2 & Supplemental Table 1). JVP was measured less than 8 cmH2O in 91 out of 110 subjects (83%) and between 8 and 12 cmH2O in 19 (17%) subjects. 16% of patients in both groups (JVP < 8 cmH2O and 8–12 cmH2O) were found to have PCWP measurements >22 mmHg. Edema grade > 2 demonstrated a sensitivity of 28% and specificity of 86% in predicting PCWP >22 mmHg (Supplemental Table 1). JVP >8 demonstrated a sensitivity of 100% and specificity of 6% in predicting PCWP >22 mmHg. Subjective SOB demonstrated a sensitivity of 78% and specificity of 26% in predicting PCWP >22 mmHg. Zo value <20 demonstrated a sensitivity 33% and specificity of 80% in predicting a PCWP >22 mmHg (Supplemental Table 1). This table addresses the operating characteristics of clinical variables as a diagnostic tool for assessing PCWP >22 mmHg. None of the variables listed exhibited a balanced sensitivity and specificity for diagnosing this cutoff value, making them insensitive markers of a PCWP >22 mmHg. Finally, principal component analysis of the variables was conducted with PCWP to visualize the relation of the variables to one another. This sort of analysis is exploratory but can give insight to the underlying utility of variables in relation to one another. When this data was run in a stepwise regression to assess a multivariate approach to predicting PCWP >22 mmHg, no variables were included with either the stopping rule using minimum Bayesian Information Criterion. When a p-value threshold of 0.25 was used for the stepwise regression, edema was entered into the analysis and resulted in an insignificant nominal logistic fit to predicting PCWP >22 mmHg (p = 0.11, AUC = 0.60). Additionally, if all clinical signs and symptoms (JVP, edema, and dyspnea) were entered into a multivariate nominal logistic fit, edema displayed the greatest log worth in predicting PCWP >22 mmHg (Log worth = 0.9, p = 0.16) but none were statistically significant.

Table 2.

Number of subjects with designated history and physical examination findings and the percentage (%) of those with measured PCWP>22 mmHg. Abbreviations: JVP = jugular venous pressure; PCWP = pulmonary capillary wedge pressure.

N PCWP > 22 mmHg, %
JVP
 <8 91 16
 8–12 19 16
 13–16 0
 >16 0
Peripheral edema
 0 64 13
 1 28 18
 2 11 27
 3 5 20
 4 2 50
Dyspnea
 Yes 28 14
 No 82 17
ZOE
 <20 11 27
 20–30 31 10
 >30 14 14

4. Discussion

The primary findings of this study are: 1) common clinical signs and symptoms, JVP, edema, and SOB did not correlate with measured PCWP values and 2) thoracic bioimpedance, ZOE®, measurements did not correlate to measured PCWP values in those same patients. These data provide a look at how these recommended examination findings and a common non-invasive measurement provide insight into a patient's PCWP, the gold standard for volume status [17].

History and physical (H&P) examination remains central in the management of patients with HF [7]. Despite the declining emphasis on the H&P, the use has routinely been highlighted in the management of HF and is used by clinicians to assess underlying hemodynamic state [6], [7]. The determination of a patient having an elevated PCWP is routinely based on JVP, SOB and edema [7]. In the ESCAPE trial [5], a multivariable model with PCWP ≥22 mmHg as the dependent variable was performed and demonstrated that only an elevated JVP, as defined as ≥12 mmHg, demonstrated the associated with elevated PCWP, defined as ≥22 mmHg (JVP odds ratio, 3.3; 95% CI, 1.8, 6.1) in stable advanced heart failure patients [6], [23]. SOB became associated with elevated PCWP once PCWP was measured ≥28 mmHg [6]. In our data set, the patient's with JVP elevated between 8 and 12 cmH2O only 16% had elevated PCWP >22 mmHg. Half of the patients with 4+ pitting edema had PCWP>22 mmHg and only 14% of the patients with SOB had elevated PCWP>22 mmHg. Of note, the subjects included in this study demonstrated a wider range of HF severity including a lower proportion of patients with PCWP>22 mmHg compared to the population investigated in the ESCAPE trial. The logistic regression of the variables conducted in Supplemental Table 1 resulted in nonsignificant results of using each variable independently as a diagnostic predictor of an elevated PCWP >22 mmHg.

Our findings are more consistent with published studies that conclude that signs and symptoms of JVP, edema and/or SOB have limitations and may not be as reliable in determining the actual intra-vascular volume in patients [24], [25]. While these data are not presented to disparage the importance of a cornerstone of patient care, the H&P, they are simply investigated to better appreciate their abilities and limitations when used in the care of HF patients.

ZOE® has been used as the predicated device for multiple medical devices, specifically, for another thoracic impedance device ReDs (K150095). These data suggest that thoracic impedance is not an accurate measure of volume status as determined by PCWP. Interestingly, this point is visually clear in the principal component analysis demonstrating directional vectors of how increasing ZOE® measurements (less thoracic fluid) track along with the SOB findings (Supplemental Fig. 1). This does make clinical sense given what is measured. Thoracic bioimpedance measures the resistance (Ohms) of the thoracic region at 100 kHz. With a decrease in resistance representing “fluid” accumulation in the thoracic cavity, it stands to reason that decreasing ZOE® measurements would correlate to increasing SOB.

Supplemental Fig. 1.

Supplemental Fig. 1

Principle component analysis of investigational variables demonstrating directional vectors by utilizing the eigenvectors of the covariance matrix. This analysis visually demonstrates the relationship of increasing values for each variable to one another. Abbreviations: PCWP = pulmonary capillary wedge pressure; JVP = jugular venous pressure; CO = cardiac output.

There were limitations of this study. As with any observational data there were traditional limitations such as: prone to bias and cannot be used to demonstrate causality. The three averaged ZOE® measurements may not have represented the best way of data capture to analyze its efficacy. While less practical, continuous measurements for larger sampling rates may provide more accurate data. Bioimpedance also has inherent limitations with measurement effect from body positioning, body tissue (muscle and adiposity) composition, presence of cutaneous hair or sweat, and lead placement. Of note, ZOE® measurements were taken at only one time point and only on patients that agreed to have the necessary areas shaved to allow proper electrode conductance limiting the number of enrolled patients who had ZOE® readings. Additionally, while 110 subjects provided usable PCWP tracings only 15 had PCWP >22 mmHg, potentially under powering the study.

5. Conclusion

Given the findings of this observational study, JVP, edema, and SOB appear to be unreliable in their ability to estimate the PCWP. JVP may be of some utility at extreme values. Furthermore, low enrollment rates for ZOE measurements highlighted the perceived discomfort and inconvenience of the bioimpedance electrodes and the need for more comfortable and patient focused methods for acquiring bioimpedance data to better study its utility. In this study, there was no correlation between thoracic bioimpedance, a non-invasive alternative method for investigating a patient's congestion status, and PCWP measured with a PAC. These findings support the idea that traditional measures for monitoring heart failure patients are limited and there is the significant need for reliable and accurate non-invasive volume status monitoring that can translate to both the hospital and the home setting.

The following are the supplementary data related to this article.

Supplemental Table 1

Utility of the Clinical Findings in Detecting PCWP >22 mmHg. Abbreviation: LR = likelihood ratio; OR = odds ratio; CI = confidence interval.

mmc1.docx (72.6KB, docx)

Funding

This work was supported by two grants from the National Institutes of Health (BA: R01HL148244) and (KH: R44HL140669).

CRediT authorship contribution statement

Monica Polcz, MD: Conceptualization, Visualization, Methodology, Investigation, Data Curation, Writing – Original Draft, Review & Editing Jessica Huston, MD: Conceptualization, Visualization, Methodology, Investigation, Data Curation, Writing – Original Draft, Review & Editing Meghan Breed, MD: Methodology, Investigation, Data Curation, Writing – Original Draft, Review & Editing Marisa Case, RN: Investigation, Data Curation Phil Leisy, MD: Investigation, Writing – Review & Editing Jeffrey Schmeckpeper, MD: Investigation, Writing – Review & Editing Lexie Vaughn, MD: Investigation, Writing – Review & Editing Jenna Helmer Sobey, MD: Investigation, Writing – Review & Editing Colleen Brophy, MD: Project administration, Supervision, Writing – Review & Editing JoAnn Lindenfeld, MD: Project administration, Supervision, Writing – Review & Editing Kyle Hocking, PhD: Conceptualization, Visualization, Methodology, Data Curation, Writing –Review & Editing, Supervision, Funding acquisition Bret Alvis, MD: Conceptualization, Visualization, Methodology, Investigation, Data Curation, Writing – Original Draft, Review & Editing, Supervision, Project administration, Funding acquisition.

Declaration of competing interest

The authors declare the following financial interests/personal relationships which may be considered as potential competing interests:

The authors report no relationships that could be construed as a conflict of interest to the details of this study. Kyle Hocking, PhD, is Founder, CEO and President of VoluMetrix and an inventor on intellectual property in the field of venous waveform analysis assigned to Vanderbilt and licensed to VoluMetrix. Colleen Brophy, MD, is Founder and CMO of VoluMetrix and an inventor on intellectual property in the field of venous waveform analysis assigned to Vanderbilt and licensed to VoluMetrix. Bret Alvis, MD, CSO and is an inventor on intellectual property in the field of venous waveform analysis assigned to Vanderbilt and licensed to VoluMetrix and is married to the COO of VoluMetrix. The remaining authors have no disclosures to report.

Acknowledgements

The investigators appreciate the assistance of the care partners, nursing staff, and cardiologists at Vanderbilt University Medical Center.

Contributor Information

Monica Polcz, Email: Monica.polcz@vumc.org.

Jessica Huston, Email: hustonjh2@upmc.edu.

Meghan Breed, Email: meghan.breed@vumc.org.

Marisa Case, Email: marisa.case@vumc.org.

Philip Leisy, Email: Philip.j.leisy@vumc.org.

Jeffrey Schmeckpeper, Email: Jeffrey.schmeckpeper@vumc.org.

Lexie Vaughn, Email: lexie.h.vaughn.1@vumc.org.

Jenna Helmer Sobey, Email: jenna.m.helmer@vumc.org.

Colleen Brophy, Email: colleen.brophy@vumc.org.

JoAnn Lindenfeld, Email: joann.lindenfeld@vumc.org.

Kyle Hocking, Email: kyle.m.hocking@vumc.org.

Bret Alvis, Email: bret.d.alvis@vumc.org.

References

  • 1.Yancy C.W., Jessup M., Bozkurt B., Butler J., Casey D.E., Jr., Drazner M.H., et al. 2013 ACCF/AHA guideline for the management of heart failure: a report of the American College of Cardiology Foundation/American Heart Association Task Force on Practice Guidelines. J. Am. Coll. Cardiol. 2013;62:e147–e239. doi: 10.1016/j.jacc.2013.05.019. [DOI] [PubMed] [Google Scholar]
  • 2.Greenhalgh T., A'Court C., Shaw S. Understanding heart failure; explaining telehealth - a hermeneutic systematic review. BMC Cardiovasc. Disord. 2017;17:156. doi: 10.1186/s12872-017-0594-2. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 3.Heo S., Doering L.V., Widener J., Moser D.K. Predictors and effect of physical symptom status on health-related quality of life in patients with heart failure. Am. J. Crit. Care. 2008;17:124–132. [PubMed] [Google Scholar]
  • 4.Yancy C.W., Jessup M., Bozkurt B., Butler J., Casey D.E., Jr., Colvin M.M., et al. 2017 ACC/AHA/HFSA focused update of the 2013 ACCF/AHA guideline for the management of heart failure: a report of the American College of Cardiology/American Heart Association Task Force on Clinical Practice Guidelines and the Heart Failure Society of America. Circulation. 2017;136:e137–e161. doi: 10.1161/CIR.0000000000000509. [DOI] [PubMed] [Google Scholar]
  • 5.Binanay C., Califf R.M., Hasselblad V., O'Connor C.M., Shah M.R., Sopko G., et al. Evaluation study of congestive heart failure and pulmonary artery catheterization effectiveness: the ESCAPE trial. JAMA. 2005;294:1625–1633. doi: 10.1001/jama.294.13.1625. [DOI] [PubMed] [Google Scholar]
  • 6.Drazner M.H., Hellkamp A.S., Leier C.V., Shah M.R., Miller L.W., Russell S.D., et al. Value of clinician assessment of hemodynamics in advanced heart failure: the ESCAPE trial. Circ. Heart Fail. 2008;1:170–177. doi: 10.1161/CIRCHEARTFAILURE.108.769778. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 7.Thibodeau J.T., Drazner M.H. Reply: the role of the clinical examination in patients with heart failure. JACC Heart Fail. 2018;6:971. doi: 10.1016/j.jchf.2018.08.001. [DOI] [PubMed] [Google Scholar]
  • 8.Alvis B.D., Polcz M., Huston J.H., Hopper T.S., Leisy P., Mishra K., et al. Observational study of Non-Invasive Venous waveform Analysis (NIVA) to assess intracardiac filling pressures during right heart catheterization. J. Card. Fail. 2020;26(2):136–141. doi: 10.1016/j.cardfail.2019.09.009. [DOI] [PubMed] [Google Scholar]
  • 9.Adamson P.B., Ginn G., Anker S.D., Bourge R.C., Abraham W.T. Remote haemodynamic-guided care for patients with chronic heart failure: a meta-analysis of completed trials. Eur. J. Heart Fail. 2017;19:426–433. doi: 10.1002/ejhf.638. [DOI] [PubMed] [Google Scholar]
  • 10.Falcochio D.F., Da Costa A.C., Durigan C.P.I., Nascimento V.D.G., Santili C., Chakkour I. Epidemiological and clinical aspects of cleft hand: case series from a tertiary public hospital in Sao Paulo, Brazil. Hand (N Y) 2019;14:814–818. doi: 10.1177/1558944718778399. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 11.Gheorghiade M., Filippatos G., De Luca L., Burnett J. Congestion in acute heart failure syndromes: an essential target of evaluation and treatment. Am. J. Med. 2006;119:S3–S10. doi: 10.1016/j.amjmed.2006.09.011. [DOI] [PubMed] [Google Scholar]
  • 12.Ambrosy A.P., Pang P.S., Khan S., Konstam M.A., Fonarow G.C., Traver B., et al. Clinical course and predictive value of congestion during hospitalization in patients admitted for worsening signs and symptoms of heart failure with reduced ejection fraction: findings from the EVEREST trial. Eur. Heart J. 2013;34:835–843. doi: 10.1093/eurheartj/ehs444. [DOI] [PubMed] [Google Scholar]
  • 13.O'Connor C.M., Stough W.G., Gallup D.S., Hasselblad V., Gheorghiade M. Demographics, clinical characteristics, and outcomes of patients hospitalized for decompensated heart failure: observations from the IMPACT-HF registry. J. Card. Fail. 2005;11:200–205. doi: 10.1016/j.cardfail.2004.08.160. [DOI] [PubMed] [Google Scholar]
  • 14.Piccoli A. Bioelectric impedance measurement for fluid status assessment. Contrib. Nephrol. 2010;164:143–152. doi: 10.1159/000313727. [DOI] [PubMed] [Google Scholar]
  • 15.Technologies NM. ZOE Technical Specifications.
  • 16.Lasater M., Von Rueden K.T. Outpatient cardiovascular management utilizing impedance cardiography. AACN Clin. Issues. 2003;14:240–250. doi: 10.1097/00044067-200305000-00013. [DOI] [PubMed] [Google Scholar]
  • 17.Wise E.S., Hocking K.M., Polcz M.E., Beilman G.J., Brophy C.M., Sobey J.H., et al. Hemodynamic parameters in the assessment of fluid status in a porcine hemorrhage and resuscitation model. Anesthesiology. 2021;134:607–616. doi: 10.1097/ALN.0000000000003724. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 18.Hsu V.M., Moreyra A.E., Wilson A.C., Shinnar M., Shindler D.M., Wilson J.E., et al. Assessment of pulmonary arterial hypertension in patients with systemic sclerosis: comparison of noninvasive tests with results of right-heart catheterization. J. Rheumatol. 2008;35:458–465. [PubMed] [Google Scholar]
  • 19.Hassan Shaker M.E.E.D.T., Gawad Kamel A., Albert Gearge. Efficacy and safety assessment of a-chymotrypsin injection in postoperative and posttraumatic edema: a prospective, open-label, multicenter observational registry study in Egypt. <sb:contribution><sb:title>Egypt.</sb:title> </sb:contribution><sb:host><sb:issue><sb:series><sb:title>J. Surg.</sb:title></sb:series></sb:issue></sb:host>. 2017;36:88–91. [Google Scholar]
  • 20.Chiaco J.M.S.C., Parikh N.I., et al. The jugular venous pressure revisited. Cleve. Clin. J. Med. 2013;10:638–644. doi: 10.3949/ccjm.80a.13039. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 21.Shaker H.E.D.T.M., Gawad K.A., Albert G. Efficacy and safety assessment of α-chymotrypsin injection in postoperative and post-traumatic edema: a prospective, open-label, multicenter observational registry study in Egypt. Egypt.J.Surg. 2017;36:88–91. [Google Scholar]
  • 22.Sinisalo J., Rapola J., Rossinen J., Kupari M. Simplifying the estimation of jugular venous pressure. Am. J. Cardiol. 2007;100:1779–1781. doi: 10.1016/j.amjcard.2007.07.030. [DOI] [PubMed] [Google Scholar]
  • 23.Shah M.R., O'Connor C.M., Sopko G., Hasselblad V., Califf R.M., Stevenson L.W. Evaluation study of congestive heart failure and pulmonary artery catheterization effectiveness (ESCAPE): design and rationale. Am. Heart J. 2001;141:528–535. doi: 10.1067/mhj.2001.113995. [DOI] [PubMed] [Google Scholar]
  • 24.Howie-Esquivel J., Dracup K., Whooley M.A., McCulloch C., Jin C., Moser D.K., et al. Rapid 5 lb weight gain is not associated with readmission in patients with heart failure. ESC Heart Fail. 2019;6:131–137. doi: 10.1002/ehf2.12370. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 25.Lynga P., Persson H., Hagg-Martinell A., Hagglund E., Hagerman I., Langius-Eklof A., et al. Weight monitoring in patients with severe heart failure (WISH). A randomized controlled trial. Eur. J. Heart Fail. 2012;14:438–444. doi: 10.1093/eurjhf/hfs023. [DOI] [PubMed] [Google Scholar]

Associated Data

This section collects any data citations, data availability statements, or supplementary materials included in this article.

Supplementary Materials

Supplemental Table 1

Utility of the Clinical Findings in Detecting PCWP >22 mmHg. Abbreviation: LR = likelihood ratio; OR = odds ratio; CI = confidence interval.

mmc1.docx (72.6KB, docx)

Articles from American Heart Hournal Plus: Cardiology Research and Practice are provided here courtesy of Elsevier

RESOURCES