Dipole-spread-function engineering for simultaneously measuring the 3D orientations and 3D positions of fluorescent molecules

Tingting Wu 1,2, Jin Lu 1,2, Matthew D Lew 1,2,3,*
PMCID: PMC9122034  NIHMSID: NIHMS1801780  PMID: 35601691

Abstract

Interactions between biomolecules are characterized by both where they occur and how they are organized, e.g., the alignment of lipid molecules to form a membrane. However, spatial and angular information are mixed within the image of a fluorescent molecule–the microscope’s dipole-spread function (DSF). We demonstrate the pixOL algorithm for simultaneously optimizing all pixels within a phase mask to produce an engineered Green’s tensor–the dipole extension of point-spread function engineering. The pixOL DSF achieves optimal precision for measuring simultaneously the 3D orientation and 3D location of a single molecule, i.e., 4.1° orientation, 0.44 sr wobble angle, 23.2 nm lateral localization, and 19.5 nm axial localization precisions in simulations over a 700-nm depth range using 2500 detected photons. The pixOL microscope accurately and precisely resolves the 3D positions and 3D orientations of Nile red within a spherical supported lipid bilayer, resolving both membrane defects and differences in cholesterol concentration in 6 dimensions.

1. INTRODUCTION

The translational and rotational movements of molecules underlie almost all biological and chemical processes. For example, cell membranes are characterized by the organization and alignment of their lipid constituents; the folding conformation or structural disorder of a protein largely determines its interactions with neighbors; DNA must be unwound and accessible by an RNA polymerase for a gene to be expressed. Thus, to study biological function and dysfunction, the positions and conformations of biomolecules are both quantified by molecular dynamics simulations and visualized by experimental imaging techniques. However, imaging these dynamics in native biological environments is difficult. Electron microscopy has exquisite resolution, but cannot be used to image dynamic molecular motions in solution [1, 2]. Interferometric optical scattering can detect, image, and even measure the mass of single molecules (SMs) [3], but it is difficult to distinguish the scattering of one molecular species from another. Super-resolution microscopy can now achieve molecular resolution [46], but these techniques often intentionally reduce orientation sensitivity (e.g., MINFLUX) so that their position measurements are more robust. Thus, visualizing the position and orientation of SMs simultaneously, precisely, and robustly is difficult. The imaging task is inherently multidimensional, with 6 dimensions of information (3D position, 3D orientation, and rotational wobble) being conveyed by only hundreds or a few thousand fluorescence photons. Quantifying fundamental limits and engineering optimal methods for maximum measurement precision are topics of active research [710].

Many methods have been proposed for measuring either the 3D positions [11] or the 3D orientations [1214] of SMs. However, there are comparatively few methods experimentally demonstrated for measuring 3D position and 3D orientation simultaneously in single-molecule orientation-localization microscopy (SMOLM). The double-helix PSF [15] and bisected pupil [16] are early examples of PSFs designed for 3D single-molecule localization microscopy (SMLM), but both require relatively bright emitters. More recently, CHIDO uses a stress-engineered (birefringent) optic and polarizing beamsplitter (PBS) in the fluorescence detection path to measure 3D orientations and positions [17]. However, its measurement precision is strongly affected by optical aberrations. In addition, the vortex PSF measures SM 3D position and orientation by modulating fluorescence emission with a vortex phase plate, common in STED nanoscopy, and does not require a PBS [18]. However, its simple implementation requires comparatively bright emitters [13] or a PBS [19] to achieve high orientation measurement precision. Recent reports using estimation theory show that the aforementioned methods do not yet achieve optimal performance [13]. We hypothesize that more powerful optimization tools are required to explore the expansive space of possible Green’s tensors; dipole-spread function (DSF) engineering is distinguished from PSF engineering by the need to optimize for both 3D position and 3D orientation measurement precision–an inherently more complex task. We are inspired by recent methods that leverage automatic differentiation to engineer phase masks pixel-by-pixel to create PSFs with excellent performance [20].

In this article, we demonstrate the pixOL microscope, which achieves superior performance over state-of-the-art DSFs for simultaneously measuring the 3D orientations and 3D locations of SMs across an extended depth range (1 μm). We design an algorithm (pixOL) to simultaneously optimize all pixels of a phase mask to shape the dipole response from the microscope, i.e., engineer its Green’s tensor. Unlike optimization using Zernike polynomials [21], pixOL can directly take advantage of super-critical fluorescence arising from imaging SMs near a refractive index interface [22]. The resulting pixOL DSF measures simultaneously the 3D orientations and 3D positions of Nile red (NR) molecules transiently attached to spherical supported lipid bilayers (SLBs). In experiments, SMOLM using the pixOL DSF accurately resolves the 3D spherical shape of the lipid membrane. Further, SMOLM reconstructions show the presence or absence of cholesterol within the membrane through accurate orientation imaging of NR relative to the membrane surface. To our knowledge, these experiments are the first demonstrations of nanoscale super-resolved imaging with accurate molecular 3D position and 3D orientation determination over an entire extended object.

2. PIXEL-WISE OPTIMIZATION FOR SMOLM DSF DESIGN

We model a fluorescent molecule as a dipole-like emitter [2325] with a mean orientation [μx, μy, μz] = [sin θ cos ϕ, sin θ sin ϕ, cos θ] and a “wobble” solid angle Ω that characterizes its rotational diffusion [26, 27] during a camera frame (Fig. 1(b)). The image produced by the microscope is linearly proportional to a molecule’s orientational second-moment vector m=[μx2,μy2,μz2,μxμy,μxμz,μyμz]T6, given by (Eqn. S3)

I=s[Bxx,Byy,Bzz,Bxy,Bxz,Byz]m+b, (1)

where IN is the captured intensity on a camera with N pixels, s is the number of signal photons detected from the emitter, and b is the background in each pixel. The angle brackets 〈·〉 represent a temporal average over one camera frame. The matrices BilN×1 correspond to the imaging system’s response to each orientational second moment and can be calculated using vectorial diffraction theory [28, 29]. The DSF of any SM is a linear combination of these basis images, which are directly related to the Green’s tensor [27, 30].

Fig. 1.

Fig. 1.

pixOL phase mask and dipole spread function (DSF) for measuring the 3D orientations and 3D positions simultaneously of dipole-like emitters. (a) Imaging system schematic. A microscope objective is focused at a nominal focal plane (NFP, dotted black line) within water at a distance −z above (+z below) the coverslip at z = 0. The objective collects fluorescence photons from emitters at various locations (x,y,h), where h > 0 for an emitter above the coverslip. A polarization-sensitive 4f system, comprising 3 lenses and a polarizing beam splitter (PBS), is added after the microscope’s intermediate image plane (IIP) to place the pixOL phase mask at the back focal plane (BFP). Two cameras (or two regions of a single camera) capture x-polarized (red) and y-polarized (cyan) fluorescence. Arrows denote polarization of the light in each channel. (b) Orientation of a dipole-like emitter, parameterized by a polar angle θ ∈ [0°, 90°], azimuthal angle ϕ ∈ (−180°, 180°], and wobble solid angle Ω ∈ [0, 2π] sr. (c) Optimized pixOL phase mask. Arrows denote polarization of light in (a) relative to the phase mask. Colormap: phase (rad). Scalebar: 500 μm. (d-f) Simulated images of emitters located at (d) h = 800 nm, (e) h = 400 nm, and (f) h = 0 nm with orientations (θ, ϕ, Ω) shown in (b) (emitter 1: Ω = 2π sr, emitter 2: (0°, 0°, 0), emitter 3: (45°, 0°, 0), emitter 4: (90°, 0°, 0)), captured in the two polarization channels shown in (a) with the NFP at z = −580 nm. The intensities of each red-blue image pair are normalized. Scalebar: 500 nm.

A microscope directly encodes a dipole emitter’s lateral position into the location of its shift-invariant DSF, while an SM’s axial location (h) and 3D orientation (θ, ϕ, Ω) are hidden in the shape of the DSF. To achieve high precision for estimating 3D orientation and 3D location, the shape of the DSF must vary quickly as an emitter’s orientation and axial location changes. This measurement sensitivity can be quantified using the Fisher information matrix [31]. Its matrix inverse, the Cramér-Rao bound (CRB), gives a lower bound on the variance of any unbiased estimator.

We use the CRB matrix K, which quantifies the performance of estimating the orientational second moments m, to optimize a phase mask for a polarization-sensitive microscope (Figs. 1(a) and S1). Fluorescence emission is split into x- and y-polarized channels. Images from the two channels are collected simultaneously and are modeled computationally by concatenating them into a single intensity image I. Each basis image Bil in Eqn. 1 is assembled from a concatenation of x- and y-polarized images from our vectorial imaging model (SI Section 1). We specifically consider emitters that are located near a water-glass interface, which enables super-critical fluorescence to be captured by the imaging system, thereby boosting measurement sensitivity [22]. To best leverage this information, we simultaneously optimize all pixels of a phase mask Pn×n by minimizing the loss function

=minPmMdet (K(P,m)), (2)

where M denotes orientation space (SI Section 1) and det(·) represents the determinant of a matrix. To avoid poor lateral estimation precision, we force the algorithm to create DSFs smaller than 1.8 μm × 1.8 μm by ignoring photons diffracted outside of this region on the camera.

The optimized pixOL phase mask (Fig. 1(c)) breaks the symmetries within the images produced by the six orientational moments m at the back focal plane (Fig. S2). Notably, it modulates super-critical fluorescence differently from the rest of the BFP; if, instead, the super-critical light is modulated in the same manner as the sub-critical light, then measurement precision degrades significantly (Fig. S3). While a mask of similar design and performance can be optimized using a basis set of 55 Zernike polynomials, pixOL’s pixel-wise optimization converges to a high-performance mask using fewer iterations (Fig. S4). The resulting DSFs of SMs of various orientations exhibit easily discernible shapes and intensities across the x- and y-polarized imaging channels (Fig. 1(d–f)), owing to their 6 distinct basis images (Fig. S5). Simultaneously, pixOL also breaks the symmetry of defocus, rotating the DSF by 90° when an SM is above vs. below the focal plane (Fig. 1(d–f)).

Emitters often exhibit a broad distribution of signal to background ratios (SBRs), even within a single imaging experiment. To explore the stability and optimality of the pixOL design, we optimize new phase masks for three other SBR scenarios, namely 380:10, 3800:10, and 3800:2 (signal photons:background photons/pixel, Fig. S6(a-d)). The mask optimized for SBRs of 380:10 and 3800:10 exhibit a similar phase profile to the original pixOL mask (optimized for a 380:2 SBR). Further, all three masks show nearly identical 3D orientation measurement precisions across a range of SBR conditions (Fig. S6(e-l)). On the other hand, the mask optimized for high SBR imaging differs significantly from the others (Fig. S6(d)), showing that the pixOL optimization algorithm utilizes the increased photon budgets to achieve more uniform orientation measurement precision (Fig. S6(e-l)). To estimate the optimality of our design, we compared the performance of pixOL to that of direct imaging of the back focal plane, which was shown to perform close to the best-possible quantum CRB for measuring orientation under specific conditions [9]. By comparing the value of the loss function (Eqn. 2) for pixOL to that of BFP imaging BFP, we observe that pixOL phase masks optimized at SBRs of 3800:2 and 380:2 achieve 11% and 19% worse precision, respectively, than that of BFP imaging (Fig. S7). These data indicate that pixOL’s measurement performance is close to the global optimum.

Using CRB as a performance metric, we compare our pixOL DSF to other engineered DSFs designed for 3D orientation and 3D position measurements, namely the double helix [15], CHIDO [17], and unpolarized vortex DSFs [18] (SI Section 3, Table S1, see Fig. S8 for additional comparisons to DSFs only designed for orientation). We calculate the mean angular standard deviation σδ (MASD, Eqn. S23) as a combined precision of measuring 3D orientation (θ, ϕ) and the standard deviation σ of measuring wobble. For in-focus (Fig. 2(a,b)) emitters, pixOL shows the best precision for measuring 3D orientation (mean σδ = 0.80°, 10% better MASD than the next-best DSF, CHIDO, and mean σ = 0.16 sr, 11% better wobble precision than CHIDO). Over an 800-nm depth range, pixOL’s orientation precision degrades slightly (mean σδ = 1.14° and mean σ = 0.24 sr) and is comparable to CHIDO (Fig. S9). We also quantified the lateral localization precision σL and the axial localization precision σh for isotropic emitters across an axial range of 800 nm (Fig. 2(c,d)). The pixOL DSF has superior lateral precision compared to all other DSFs (mean σL = 8.17 nm over an 800-nm depth range, 26% better than CHIDO). It also has excellent axial localization precision (mean σh = 12.21 nm over an 800-nm depth range), outperformed only by the double-helix.

Fig. 2.

Fig. 2.

Measurement precision of the pixOL DSF. (a-d) Best-possible measurement precision of various engineered DSFs calculated using the Cramér-Rao bound (CRB) for emitters within water (1.33 refractive index) with 2500 signal photons and 3 background photons per pixel detected. Red: pixOL, yellow: double helix (DH) [15], grey: CHIDO [17], purple: unpolarized vortex [18]. (a) Mean angular standard deviation σδ (MASD) averaged uniformly over all θ. MASD quantifies the combined precision of measuring θ and ϕ as the half-angle of a cone representing orientation uncertainty (Eqn. S23) [32]. (b) Mean wobble angle precision σ averaged uniformly over all θ. (c,d) Localization precisions σL and σh for measuring (c) lateral position l and (d) axial location h above the interface, respectively. MASD and σ are calculated for in-focus SMs with fixed orientation (Ω = 0 sr); localization precisions are for isotropic emitters (Ω = 2π sr). (e) Position and emission anisotropy measurements of a fluorescent bead, scanned axially from z = −790 nm to z = 610 nm with a step size of 50 nm (11 camera frames per step). Red dot: estimated axial distance z between the bead and focal plane in each frame; green cross: expected stage position. Inset (i): Experimental axial precision σz at each scanning plane (mean precision σz¯=2.89 nm). Inset (ii): Experimental emission anisotropy precision σ at each scanning plane (average precision σΩ¯=0.097 sr).

For a system without aberration, the DSF produced by pixOL phase mask P is nearly identical to the DSF produced by its conjugate (pixOL*, −P) at the opposite axial position (Figs. S26(d) vs. S27(e)). In any experiment, optical aberrations (modeled as an additive phase mask Wn×n) will perturb the designed DSF and decrease estimation performance. Using a liquid-crystal spatial light modulator placed in the microscope back focal plane (Figs. 1 and S1), we compare the DSF produced by the experimental pixOL phase mask (Fig. S26(e)) to that of its experimental conjugate (Fig. S27(f)) in the presence of aberrations. Interestingly, the pixOL* DSF better matches the depth-dependent features of the ideal pixOL DSF. Thus, we use the pixOL* mask, an experimentally calibrated DSF model (SI Section 6), and a bespoke regularized maximum-likelihood estimator (SI Section 2) to jointly estimate the 3D orientations and 3D positions of all SMs within each FOV.

Monte Carlo simulations of our estimation algorithm show that the pixOL microscope achieves 23.2 nm lateral and 19.5 nm axial localization precisions on average throughout a 700-nm depth range with 2500 detected photons and 3 background photons per pixel (Figs. S16S18). Simultaneously, pixOL estimates the orientation and wobble of each SM with 4.1° and 0.44 sr precisions, respectively. Under higher background conditions (10 photons/pixel, Figs. S20S22), the pixOL microscope achieves similar precision (6.9° orientation, 0.59 sr wobble angle, 31.7 nm lateral, and 25.8 nm axial precision). However, similarly to other DSFs, the pixOL DSF’s footprint becomes larger for emitters located far away from the focal plane, leading to poorer detection rates for our estimation algorithm. At this lower SBR, our algorithm detects an average of 90.0% of fixed emitters at the coverslip, 99.8% of emitters in focus (at h = 400 nm), and 14.4% of emitters far away from the coverslip at h = 700 nm.

Scanning fluorescent beads (100-nm diameter) across an axial range of 1400 nm enables us to verify localization and orientation measurement precisions experimentally. The trajectory of defocus estimates z resolves the 50-nm stage movements very well (Fig. 2(e), average axial precision σz¯=2.89 nm in Fig. 2(e)(i)). Since the bead contains many fluorophores, we may quantify the bead’s emission pattern by measuring its effective “wobble” angle Ω. We find that its emission is largely isotropic (wobble angle Ω¯=1.72π±0.097 sr averaged over all steps, mean±std, Figs. 2(e)(ii) and S30(e)). In our implementation, experimental precisions for measuring axial location (Fig. 2(e)(i)) and orientation (Fig. 2(e)(ii)) can degrade slightly for certain emitter orientations when they are in focus (Fig. S17); this degradation can be corrected by using multiple calibrated phase masks, each tuned for a specific axial range. Data from other beads also show precise localization and orientation estimates over a 1400-nm axial range (average axial precision σz¯=4.64 nm and wobble angle Ω¯=1.61π±0.17 sr for the three beads in Figs. 2(e) and S30).

3. 6D SMOLM REVEALS MEMBRANE MORPHOLOGY AND COMPOSITION

To demonstrate accurate 6D imaging of molecular orientations (θ, ϕ, Ω) and positions (x, y, h), we adhere supported lipid bilayers (SLBs) to silica beads (2 μm diameter, Fig. 3(a), SI Section 7) [33, 34] using two lipid compositions: DPPC (di(16:0) phosphatidylcholine) only vs. a mixture of DPPC and cholesterol (chol). Previous studies [19, 35, 36] have shown that NR orients itself perpendicular to the membrane when a high concentration (40% used here) of chol is present. Thus, a spherical SLB enables us to validate simultaneously orientation and position imaging performance, quantifying both precision and accuracy. We use the transient binding and blinking of Nile red (NR) molecules to the SLBs [35, 37, 38] to facilitate SM detection and orientation-position measurements (Fig. 3(a), Movie 1). The beads are illuminated by a tilted, circularly polarized laser beam so that emitters can be excited efficiently regardless of their 3D orientation. To avoid position-dependent aberrations from the refractive index mismatch between the silica beads and imaging buffer, we focus our analyses on emitters located at the bottom half of each SLB.

Fig. 3.

Fig. 3.

SMOLM images of the 3D orientations and 3D locations of Nile red (NR) within spherical supported lipid bilayers (SLBs) collected by the pixOL DSF. (a) SLBs are adhered to 2 μm-diameter silica spheres, where θ represents the relative angle between the orientation of each NR molecule and the SLB’s surface normal and r is the distance between the 2D location (x, y) of a NR within a certain h slice and the center of the sphere. (b-d) 2D maps of polar orientation θ for each NR molecule, shown for the bottom half of each bead. (b) x-h and (c) x-y views for an SLB consisting of DPPC and 40% cholesterol (chol). (d) x-y view of a DPPC-only SLB. Colorbar: θ (deg). (e) x-y cross-sections of the bead in (b,c) depicting the 3D orientation (θ,ϕ) of each NR as a line segment. The length and direction of each line indicate in-plane magnitude (μx2+μy2)1/2 and azimuthal orientation ϕ, respectively. Colors represent azimuthal orientation ϕ. White box: a membrane defect. (f) NR orientation θ vs. wobble Ω for the (blue) DPPC+chol SLB and (yellow) DPPC-only SLB. Crosses indicate measurement medians. (g) NR orientations θ (green) across the entire DPPC+chol SLB and (purple) within the membrane defect in (e). (h,i) Measured (h) cross-sectional radius r and (i) apparent thickness of the spherical SLB, calculated as the peak and full-width at half-maximum (FWHM), respectively, of the distribution of NR lateral positions r, in each h slice in (e) and Figs. S32(d) and S33(e). (blue) DPPC + chol bead, (yellow) DPPC-only bead, and (red) ideal sphere. The theoretical FWHM in (i) accounts for the projection of the SLB into the xy plane and pixOL’s localization precision (Eqn. S27). Scale bars: 400 nm.

For the SLB containing chol, the 3D locations of NR form a sphere as expected (Fig. S32(a)). The orientations of NR change smoothly from being mostly parallel to the optical axis (small θ) at the bottom of the sphere to being within the xy-plane (θ approaching 90°) at the sphere’s waist (Fig. 3(b,c), Movie 2). Likewise, the azimuthal orientations ϕ of NR show that each molecule is perpendicular to the sphere’s surface within each h slice (Figs. 3(e) and S32(a,b), Movie 3). Note that the sphere’s shape and the symmetry of SM emission guarantee exactly two locations, on opposite sides of the sphere, where NR orientations are identical to one another. For example, the measured ϕ below the sphere’s equator (h < 900 nm) match the ϕ measurements on the opposite surface above the equator (h > 1100 nm, Fig. 3(e)).

Calculating the relative orientation θ (Fig. 3(a)) between each NR and the surface normal of the sphere shows that the molecules lie mostly parallel (small θ) to the lipid tails within the SLB (Fig. 3(f)), regardless of their location on the sphere (Fig. S32(c,d)). Moreover, each NR exhibits relatively little rotational diffusion, i.e., small wobble angle Ω (Figs. 3(f) and S32(e)), which is consistent with previous characterizations of NR within planar SLBs [19, 35, 36]. Across the SLB surface, we measure a mean angular bias θ⊥,bias of 12.7° and a mean wobble bias Ωbias of 0.51 sr (Fig. S32(d)(i,ii)), assuming that each NR should lie in a fixed orientation exactly normal to the SLB, which is a worst-case bias estimate. The NR data also show a mean angular standard deviation σδ of 20.2° and a mean wobble angle precision σ of 0.88 sr (Fig. S32(d)(iii,iv)); notably, these distributions convolve the true orientation distribution of Nile red with pixOL’s measurement precision.

Interestingly, we detect a defect in the membrane (white boxes in Fig. 3(e)) where NR orientations θ are more varied (Fig. 3(g)), showing the defect’s local disorganization. However, NR wobbling Ω within the defect is similar to other regions of the sphere (Fig. S32(e)), implying that chol is distributed uniformly throughout the membrane. Without cholesterol, DPPC molecules within the SLB exhibit greater intermolecular spacing. NR in contact with the DPPC-coated bead reveals the absence of cholesterol via more random orientations θ, ϕ, and θ (Figs. 3(d,f) and S33(a-d), Movies 2 and 3) than those of the SLB containing cholesterol (median θ˜=53 for DPPC only vs. θ˜=21 for DPPC+chol). NR also shows larger wobble angles (median Ω˜=0.48π sr for DPPC only vs. Ω˜=0.11π sr for DPPC+chol, Fig. 3(f))–another indication of less crowding within the pure DPPC SLB. Finally, we note that for dim NR in pure DPPC, severe Poisson shot noise can skew measurements toward small wobble angles Ω and particular ϕ angles (Fig. S33(f,g,h)). These artifacts can likely be solved by imaging brighter molecules or adopting an estimator that is more robust to severe shot noise.

We quantify the shape and apparent thickness of the SLB by calculating the (2D) radial distance r between each NR location and the sphere’s center within each h slice (Figs. S32(g), S33(e)). The estimated shapes of two beads match the expected cross-sectional radius of an ideal sphere accurately (Fig. 3(h)). We also compute the best-possible full-width at half-maximum (FWHM) that the pixOL* microscope can achieve, accounting for the curvature of the spherical surface and assuming that pixOL* achieves CRB-limited localization precision (SI Section 8). On average, the apparent SLB thickness measured by pixOL* is 55% larger than the best-possible precision of pixOL* (Eqn. S27) across a 1200 nm axial range (average FWHM is 129 nm for DPPC with cholesterol, 123 nm for DPPC only, and 82 nm for the theoretical distribution, Fig. 3(i)). This distribution is significantly broader than the CRB and likely stems from optical aberrations (Fig. S27) and precision and bias from our estimation algorithm (Figs. S16S22 and S24). Estimation performance can be improved via more detailed aberration calibrations and corrections [18], as well as more powerful estimation algorithms that robustly explore 6-dimensional position-orientation space with greater accuracy and computational efficiency (SI Section 4).

4. CONCLUSION

Here, we propose an algorithm (pixOL) for dipole-spread function (DSF) engineering, i.e., using vectorial diffraction theory to simultaneously optimize all pixels of a phase mask for measuring the 3D orientation of dipole-like emitters. The resulting pixOL DSF achieves superior orientation and localization precision over other techniques across a large axial range (Fig. 2). Pixel-wise engineering of its phase mask enables the pixOL DSF to leverage super-critical fluorescence (Fig. 1(c)) to improve orientation-measurement sensitivity [22]. Super-critical fluorescence is also beneficial for improving the pixOL DSF’s axial localization precision, as utilized by DONALD [39] and DAISY [40], since the amount of super-critical light can be used to measure h, the height of an emitter above an refractive index interface. In addition, one can easily modify pixOL’s imaging model to generate phase masks that are optimal for other imaging geometries (Figs. S4 and S6).

Notably, the use of DNA PAINT and DNA origami as molecular “rulers” is the gold standard for validating the accuracy of optical nanoscopic tools [41, 42]. However, due to practical issues with the robustness and precision of controlling both the 3D positions and 3D orientations of the labels in these samples, we adapted 3D spherical SLBs [33, 34] to demonstrate experimentally the accuracy and precision of the pixOL microscope for 6D SMOLM imaging. Visualizing SMOLM data of NR binding to these SLBs shows highly spherical membrane morphologies and dye orientations that are perpendicular to the spherical surface (Fig. 3). Thus, despite the presence of aberrations typical in optical microscopes, the diverse features of the pixOL DSF are detectable in the images of flashing NR (Fig. S32(f)) and convey the 3D orientations and 3D positions of fluorescent molecules accurately and precisely. Moreover, our imaging of the spherical SLBs shows that the pixOL DSF sensitively discerns membrane morphology and the composition of its lipid components through detailed measurements of NR positions, orientations, and rotational diffusion.

Since the pixOL microscope measures the 3D orientations and 3D locations simultaneously of SMs, we anticipate the technology enabling fascinating studies of biomolecular interactions away from the coverslip, e.g., the 3D growth of amyloid aggregates and their interactions with cellular membranes. However, additional developments can further improve SMOLM’s versatility for biological studies. Orientation measurement precision could be improved by considering alternate polarization projections [36] or optimizing both polarization splitting and phase modulation simultaneously within a birefringent optical component. In addition, DSFs capable of coping with high background autofluorescence, as is typical in cellular imaging [14], as well as advanced machine learning algorithms [20] able to distinguish and localize molecules whose images overlap on the camera, are needed. These topics, as well as the development of DSFs whose performance is closer to fundamental limits [810], remain exciting directions for future research.

Supplementary Material

Supplementary Material
Movie S1
Download video file (6.4MB, mp4)
Movie S2
Download video file (6.5MB, mp4)
Movie S3
Download video file (3.1MB, mp4)

Acknowledgments.

The authors thank Oumeng Zhang and Tianben Ding for helpful suggestions and comments.

Funding Information.

National Science Foundation (NSF) (ECCS-1653777); National Institute of General Medical Sciences (R35GM124858).

Footnotes

Disclosures.

The authors declare no conflicts of interest. See Supplement 1 for supporting content.

Data availability.

The pixOL algorithm, estimation algorithm, and data underlying the results presented in this paper are available via Ref. [43] and by request.

REFERENCES

  • 1.Fernandez-Leiro R and Scheres SHW, “Unravelling biological macromolecules with cryo-electron microscopy,” Nature 537, 339–346 (2016). [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 2.Luque D and Castón JR, “Cryo-electron microscopy for the study of virus assembly,” Nat. Chem. Biol 16, 231–239 (2020). [DOI] [PubMed] [Google Scholar]
  • 3.Young G and Kukura P, “Interferometric Scattering Microscopy,” Annu. Rev. Phys. Chem 70, 301–322 (2019). [DOI] [PubMed] [Google Scholar]
  • 4.Balzarotti F, Eilers Y, Gwosch KC, Gynnå AH, Westphal V, Stefani FD, Elf J, and Hell SW, “Nanometer resolution imaging and tracking of fluorescent molecules with minimal photon fluxes,” Science 355, 606–612 (2017). [DOI] [PubMed] [Google Scholar]
  • 5.Gwosch KC, Pape JK, Balzarotti F, Hoess P, Ellenberg J, Ries J, and Hell SW, “MINFLUX nanoscopy delivers 3D multicolor nanometer resolution in cells,” Nat. Methods 17, 217–224 (2020). [DOI] [PubMed] [Google Scholar]
  • 6.Schmidt R, Weihs T, Wurm CA, Jansen I, Rehman J, Sahl SJ, and Hell SW, “MINFLUX nanometer-scale 3D imaging and microsecond-range tracking on a common fluorescence microscope,” Nat. Commun 12, 1478 (2021). [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 7.Backlund MP, Shechtman Y, and Walsworth RL, “Fundamental Precision Bounds for Three-Dimensional Optical Localization Microscopy with Poisson Statistics,” Phys. Rev. Lett 121, 023904 (2018). [DOI] [PubMed] [Google Scholar]
  • 8.Zhang O and Lew MD, “Fundamental Limits on Measuring the Rotational Constraint of Single Molecules Using Fluorescence Microscopy,” Phys. Rev. Lett 122, 198301 (2019). [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 9.Zhang O and Lew MD, “Quantum limits for precisely estimating the orientation and wobble of dipole emitters,” Phys. Rev. Res 2, 33114 (2020). [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 10.Zhang O and Lew MD, “Single-molecule orientation localization microscopy I: fundamental limits,” J. Opt. Soc. Am. A 38, 277 (2021). [DOI] [PubMed] [Google Scholar]
  • 11.von Diezmann L, Shechtman Y, and Moerner WE, “Three-Dimensional Localization of Single Molecules for Super-Resolution Imaging and Single-Particle Tracking,” Chem. Rev 117, 7244–7275 (2017). [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 12.Backlund MP, Lew MD, Backer AS, Sahl SJ, and Moerner WE, “The Role of Molecular Dipole Orientation in Single-Molecule Fluorescence Microscopy and Implications for Super-Resolution Imaging,” ChemPhysChem 15, 587–599 (2014). [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 13.Zhang O and Lew MD, “Single-molecule orientation localization microscopy II: a performance comparison,” J. Opt. Soc. Am. A 38, 288 (2021). [DOI] [PubMed] [Google Scholar]
  • 14.Rimoli CV, Valades-Cruz CA, Curcio V, Mavrakis M, and Brasselet S, “4polar-STORM polarized super-resolution imaging of actin filament organization in cells,” Nat. Commun 13, 301 (2022). [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 15.Backlund MP, Lew MD, Backer AS, Sahl SJ, Grover G, Agrawal A, Piestun R, and Moerner WE, “Simultaneous, accurate measurement of the 3D position and orientation of single molecules,” Proc. Natl. Acad. Sci 109, 19087–19092 (2012). [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 16.Backer AS, Backlund MP, von Diezmann AR, Sahl SJ, and Moerner WE, “A bisected pupil for studying single-molecule orientational dynamics and its application to three-dimensional super-resolution microscopy,” Appl. Phys. Lett 104, 193701 (2014). [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 17.Curcio V, Alemán-Castañeda LA, Brown TG, Brasselet S, and Alonso MA, “Birefringent Fourier filtering for single molecule coordinate and height super-resolution imaging with dithering and orientation,” Nat. Commun 11, 5307 (2020). [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 18.Hulleman CN, Thorsen RØ, Kim E, Dekker C, Stallinga S, and Rieger B, “Simultaneous orientation and 3D localization microscopy with a Vortex point spread function,” Nat. Commun 12, 5934 (2021). [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 19.Ding T and Lew MD, “Single-Molecule Localization Microscopy of 3D Orientation and Anisotropic Wobble Using a Polarized Vortex Point Spread Function,” The J. Phys. Chem. B 125, 12718–12729 (2021). [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 20.Nehme E, Freedman D, Gordon R, Ferdman B, Weiss LE, Alalouf O, Naor T, Orange R, Michaeli T, and Shechtman Y, “DeepSTORM3D: dense 3D localization microscopy and PSF design by deep learning,” Nat. Methods 17, 734–740 (2020). [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 21.Shechtman Y, Sahl SJ, Backer AS, and Moerner WE, “Optimal point spread function design for 3D imaging,” Phys. Rev. Lett 113, 1–5 (2014). [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 22.Ding T, Wu T, Mazidi H, Zhang O, and Lew M, “Single-molecule orientation localization microscopy for resolving structural heterogeneities within amyloid fibrils,” Optica 7, 602–607 (2020). [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 23.Novotny L and Hecht B, Principles of Nano-Optics (Cambridge University Press, 2006). [Google Scholar]
  • 24.Chandler T, Shroff H, Oldenbourg R, and La Rivière P, “Spatio-angular fluorescence microscopy I Basic theory,” J. Opt. Soc. Am. A 36, 1334 (2019). [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 25.Chandler T, Shroff H, Oldenbourg R, and La Rivière P, “Spatio-angular fluorescence microscopy II. Paraxial 4f imaging,” J. Opt. Soc. Am. A 36, 1346–1360 (2019). [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 26.Stallinga S, “Effect of rotational diffusion in an orientational potential well on the point spread function of electric dipole emitters,” J. Opt. Soc. Am. A 32, 213 (2015). [DOI] [PubMed] [Google Scholar]
  • 27.Backer AS and Moerner WE, “Determining the rotational mobility of a single molecule from a single image: a numerical study,” Opt. Express 23, 4255 (2015). [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 28.Böhmer M and Enderlein J, “Orientation imaging of single molecules by wide-field epifluorescence microscopy,” J. Opt. Soc. Am. B 20, 554 (2003). [Google Scholar]
  • 29.Lieb MA, Zavislan JM, and Novotny L, “Single-molecule orientations determined by direct emission pattern imaging,” J. Opt. Soc. Am. B 21, 1210 (2004). [Google Scholar]
  • 30.Backer AS and Moerner WE, “Extending Single-Molecule Microscopy Using Optical Fourier Processing,” The J. Phys. Chem. B 118, 8313–8329 (2014). [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 31.Chao J, Sally Ward E, and Ober RJ, “Fisher information theory for parameter estimation in single molecule microscopy: tutorial,” J. Opt. Soc. Am. A 33, B36 (2016). [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 32.Chandler T, Mehta S, Shroff H, Oldenbourg R, and La Rivière PJ, “Single-fluorophore orientation determination with multiview polarized illumination: modeling and microscope design,” Opt. Express 25, 31309 (2017). [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 33.Bayerl T and Bloom M, “Physical properties of single phospholipid bilayers adsorbed to micro glass beads. A new vesicular model system studied by 2H-nuclear magnetic resonance,” Biophys. J 58, 357–362 (1990). [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 34.Sobrinos-Sanguino M, Zorrilla S, Monterroso B, Minton AP, and Rivas G, “Nucleotide and receptor density modulate binding of bacterial division FtsZ protein to ZipA containing lipid-coated microbeads,” Sci. Reports 7, 13707 (2017). [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 35.Lu J, Mazidi H, Ding T, Zhang O, and Lew MD, “Single-Molecule 3D Orientation Imaging Reveals Nanoscale Compositional Heterogeneity in Lipid Membranes,” Angewandte Chemie - Int. Ed 59, 17572–17579 (2020). [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 36.Zhang O, Zhou W, Lu J, Wu T, and Lew MD, “Resolving the Three-Dimensional Rotational and Translational Dynamics of Single Molecules Using Radially and Azimuthally Polarized Fluorescence.” Nano letters 22, 1024–1031 (2022). [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 37.Greenspan P and Fowler SD, “Spectrofluorometric studies of the lipid probe, nile red.” J. Lipid Res 26, 781–789 (1985). [PubMed] [Google Scholar]
  • 38.Sharonov A and Hochstrasser RM, “Wide-field subdiffraction imaging by accumulated binding of diffusing probes,” Proc. Natl. Acad. Sci. USA 103, 18911–18916 (2006). [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 39.Bourg N, Mayet C, Dupuis G, Barroca T, Bon P, Lécart S, Fort E, and Lévêque-Fort S, “Direct optical nanoscopy with axially localized detection,” Nat. Photonics 9, 587–593 (2015). [Google Scholar]
  • 40.Cabriel C, Bourg N, Jouchet P, Dupuis G, Leterrier C, Baron A, Badet-Denisot M-A, Vauzeilles B, Fort E, and Lévêque-Fort S, “Combining 3D single molecule localization strategies for reproducible bioimaging,” Nat. Commun 10, 1980 (2019). [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 41.Schmied JJ, Raab M, Forthmann C, Pibiri E, Wünsch B, Dammeyer T, and Tinnefeld P, “DNA origami–based standards for quantitative fluorescence microscopy,” Nat. Protoc 9, 1367–1391 (2014). [DOI] [PubMed] [Google Scholar]
  • 42.Lin R, Clowsley AH, Lutz T, Baddeley D, and Soeller C, “3D super-resolution microscopy performance and quantitative analysis assessment using DNA-PAINT and DNA origami test samples,” Methods 174, 56–71 (2020). [DOI] [PubMed] [Google Scholar]
  • 43.Lew MD and Wu T, “Dipole spread function engineering for simultaneously measuring the 3D orientation and 3D localization of dipole-like emitters,” OSF (2021), 10.17605/OSF.IO/97GMV. [DOI] [PMC free article] [PubMed] [Google Scholar]

Associated Data

This section collects any data citations, data availability statements, or supplementary materials included in this article.

Supplementary Materials

Supplementary Material
Movie S1
Download video file (6.4MB, mp4)
Movie S2
Download video file (6.5MB, mp4)
Movie S3
Download video file (3.1MB, mp4)

Data Availability Statement

The pixOL algorithm, estimation algorithm, and data underlying the results presented in this paper are available via Ref. [43] and by request.

RESOURCES