Skip to main content
Scientific Reports logoLink to Scientific Reports
. 2022 May 24;12:8748. doi: 10.1038/s41598-022-12673-9

Influence of Si wall thickness of CsI(Tl) micro-square-frustums on the performance of the structured CsI(Tl) scintillation screen in X-ray imaging

Zhixiang Sun 1, Mu Gu 1,, Xiaolin Liu 1, Bo Liu 1, Juannan Zhang 1, Shiming Huang 1, Chen Ni 1
PMCID: PMC9130278  PMID: 35610282

Abstract

To improve the detection efficiency of the structured scintillation screen with CsI(Tl) micro-square-frustums based on oxidized Si micropore array template in the case of a period as small as microns, the influence of Si wall thickness of the CsI(Tl) micro-square-frustums on the performance of the structured screen in X-ray imaging was investigated. The results show that when CsI(Tl) at the bottom of the screen is structured, the detective quantum efficiency (DQE) improves at almost all spatial frequency as the top thickness of the Si wall tSi decreases. However, when CsI (Tl) at the bottom of the screen is not structured, the DQE becomes better at low-frequency and worse at high-frequency as tSi decreases. The results can provide guidance for optimizing tSi according to the comprehensive requirements of detection efficiency and spatial resolution in X-ray imaging.

Subject terms: Imaging techniques, Optical materials and structures

Introduction

CsI(Tl) scintillation screen has been widely used in X-ray imaging because of its high light yield and detection efficiency1,2. To enhance detection efficiency, the thickness of the scintillation screen must be increased. As a result, the spatial resolution of X-ray imaging is reduced due to the lateral spread of scintillation light. To overcome this problem, the scintillation screen with columnar structure was developed. The scintillation light can be guided to propagate along the column channels and the lateral spread is restrained. The CsI(Tl) scintillation screens with needle-like columnar have been developed by vapor deposition method and the spatial resolution of X-ray imaging has been improved35. However, the improvement is imperfect and the lateral spread of the scintillation light between adjacent columns is still considerable because the columns are close to each other. To further suppress the lateral spread of scintillation light, the structured CsI(Tl) scintillation screen based on oxidized silicon micropore array template has been developed by filling the scintillator into the template611. The Si walls between the adjacent microcolumns are used to absorb the incoming scintillation light to prevent its lateral spread, and the SiO2 reflective layers grown on each side of the Si walls are used to guide the scintillation light with an incident angle θ larger than the critical angle θc to propagate along the column channels based on the principle of total reflection of light. The thickness of the SiO2 layers should not be less than 0.3 μm in consideration of the optical tunneling effect12. It has been reported that the spatial resolution of X-ray imaging has reached over 100 lp/mm by using the structured CsI(Tl) and CsI scintillation screens with microcolumn array period of 4 μm, respectively10,11. To achieve an effective optical isolation, the pore wall of the template needs to have a certain thickness, and is generally considered to be not less than 1 μm13. Therefore, when the array period of the structured scintillation screen is as small as microns, the detection efficiency of the screen will be restricted because the ratio of the filled CsI(Tl) area to that of the entire structured scintillation screen will be limited. To improve the detection efficiency, the structured CsI(Tl) scintillation screen with micro-conical-frustums was proposed14 and the frustums pore can be prepared by metal-assisted chemical etching15,16. The new structure with a cone angle of 2.4° and array period of 4 μm can increase the bottom light output by 14.5%. Compared with the columns, the frustums can guide more scintillation light into the total reflection and make it propagate along the channel.

This work intends to study the influence of the Si wall thickness on the X-ray imaging performance of the structured scintillation screen based on CsI(Tl) micro-square-frustum array to explore the possibility of further improving the detection efficiency of the scintillation screen. The performance of the scintillation screen was mainly characterized by the bottom light output (BLO), modulation transfer function (MTF) and detective quantum efficiency (DQE). The simulation was carried out by the Geant4 Monte Carlo simulation toolkit17,18, finite-difference time-domain method19 and Monte Carlo ray-tracing method20. The results can provide guidance for optimizing Si wall thickness and improving the performance of the structured CsI(Tl) screen in X-ray imaging.

Structure and methods

Screens structure

The structured scintillation screens are composed of oxidized silicon mircopore array templates filled with CsI(Tl) scintillator. The mircopores are 40 μm high, in a shape of a square-frustum, and arranged in a square array of 4.0 μm period, as shown in Fig. 1. The cone angle α of the micro-square-frustums was set to 2.4°, which is the maximum cone angle in the literature14. The walls between the micro-square-frustums consist of Si walls covered with a 0.3 μm thick SiO2 layer on each side. From the side view of the structured scintillation screen, there can be three structural layers from top to bottom. They are: (a) The micro-square-frustums are divided by Si walls covered by SiO2 layer on each side, and the height of the structural layer is represented by h1; (b) The micro-square-frustums are divided only by the SiO2 layers, and the height of the structural layer is represented by h2; (c) There is no separation wall between the micro-square-frustums, and the height of the structural layer is represented by h3. Figure 2 shows the variation of h1, h2, and h3 with tSi (top thickness of Si wall). When tSi > 1.68 μm, there is only the first structural layer and the thickness difference between the top and bottom of the Si wall is 1.68 μm. When 1.08 μm < tSi < 1.68 μm, there are the first and second structural layers. When tSi < 1.08 μm, all three structural layers exist. The variation of the filling ratio of CsI(Tl) with tSi is also shown in Fig. 2. It can be seen that the ratio rises quickly as tSi decreases.

Figure 1.

Figure 1

Side view of the CsI(Tl) micro-square-frustums structured scintillation screen with different tSi. As tSi decreases, the screen from top to bottom presents (a) one structural layer; (b) two structural layers; (c) three structural layers.

Figure 2.

Figure 2

Variation of h1, h2, and h3 with tSi, and variation of the filling ratio of CsI(Tl) with tSi.

BLO, MTF, and DQE

The bottom light output (BLO) is defined as the average number of scintillation photons generated by each X-ray photon that can reach the bottom of the structured scintillation screen. The BLO depends on the scintillation photon yield generated by CsI(Tl) and the light guide efficiency of the scintillation screen.

The modulation transfer function (MTF) can be used to characterize the resolution capability of an X-ray imaging system. The standard edge measurement method21 was used to calculate the MTF. An edge spread function (ESF) was derived from an edge phantom, which was the response of the X-ray imaging system to an opaque object with a straight edge. The MTF was calculated by Fourier transform of a line spread function (LSF), which was deduced from the differential of the ESF along the direction perpendicular to the straight edge.

LSFx=ddxESFx 1
MTFf=12π-LSFxe-2πifxdx 2

where x is the position coordinate perpendicular to the straight edge direction, f is the spatial frequency in x direction. The MTF was normalized at zero frequency. The spatial resolution of the detector is defined by the spatial frequency where the MTF = 0.1.

The detective quantum efficiency (DQE) is the frequency-domain spectral representation of the signal-to-noise characteristics of a given detector configuration. According to the International standard IEC 62220-122 it can be expressed as

DQEf=Q·MTF2fNPSf 3

where Q represents the X-ray photon flux incident on the scintillation screen and NPS is the output noise power spectrum of the X-ray imaging detector calculated using a uniformly exposed image. The exposed image was divided into several 256 × 256 pixel regions of interest (ROIs) that overlapped half the size of one another. The 2D NPS is given by

NPS(u,v)=Δx·ΔyM·256·256m=1Mi=1256j=1256(I(xi,yj)-S(xi,yj))·e-2πi(uxi+vyj)2 4

where x and y are the position coordinate of the pixel, i and j are the sequence number of the pixel, u and v are spatial frequencies in x and y directions, respectively, M is the number of ROIs, Δx and Δy are the pixel sizes, I(x, y) is the pixel value, and S(x, y) is the optionally fitted two-dimensional polynomial. The 1D NPS was obtained as the average of the NPS(u, v) data in the 7 rows or columns on either side of the corresponding axis over the frequency interval (f − 0.5 × fint ≤ f ≤ f + 0.5 × fint), where f = (u2 + v2)1/2 and fint = 0.01/pixel pitch (mm).

Simulation methods

The performance of the structured CsI(Tl) scintillation screen based on oxidized Si template in X-ray imaging was simulated using the version 10.0 of Geant4 Monte Carlo simulation toolkit17,18, finite-difference time-domain method19 and Monte Carlo ray-tracing method20. The interaction of X-ray with scintillation screen, the light generation process, transmission process and detection process need to be considered in the simulation. The Geant4 Monte Carlo simulation toolkit including the photoelectric effect, Compton scattering, coherent scattering, ionization, multiple scattering, bremsstrahlung and de-excitation was used to simulate the interaction of X-ray with scintillation screen, and the light generation process. The incidence of each X-ray photon in the simulation was treated as an event. The number of events was set to stabilize the results such as BLO and MTF. The event ended when all primary and secondary particles had disappeared. The transmission process of the scintillation light such as optical absorption and boundary processes was performed using the Monte Carlo ray-tracing method and finite-difference time-domain method because the ray optics cannot simulate the surface process of the scintillation light when the thicknesses of SiO2 and Si walls are less than or close to the wavelength of the light. For the Monte Carlo ray-tracing simulation, the probabilistic split rays trace mode was used. It means that the reflected and transmitted rays are traced. The transmittance and reflectance were computed based on Fresnel equations. The attenuation of ray power was computed based on the absorption lengths of the materials. The relative ray power threshold was set to 1%, which means that the ray will be terminated if the ray power is less than 1% of the original power. The ray would also be terminated if it was outside the scintillation screen. For the finite-difference time-domain simulation, the scintillation light incident on the surface of SiO2 layer from CsI(Tl) was represented by plane waves. The transverse electric wave and transverse magnetic wave were considered. The unit cell sizes in the directions parallel and perpendicular to the SiO2 interface were set to 1 and 3 μm, respectively. The mesh size was set to 1/34 of the wavelength. The boundary conditions were set as periodic boundary conditions in the direction perpendicular to the SiO2 interface, and perfectly matched boundary conditions in the direction parallel to the SiO2 interface.

The following parameters were needed in the simulation. The light yield and X-ray excited luminescence spectrum of CsI(Tl) were cited from Refs.1,23. The optical attenuation coefficient and refractive index of CsI(Tl) were cited from Ref.24. The refractive indices, optical attenuation coefficients of Si and SiO2 were cited from Refs.2527. The energy of the X-ray beam incident uniformly and vertically from above the scintillation screen is 20 keV. The intensity of X-rays incident on the screen was set to 4.0 × 106 mm−2. An ideal photodetector was placed at the bottom of the structured scintillation screen and a thin layer of silicone oil with a refractive index 1.465 is coupled between the scintillation screen and detector.

The same method has been used to study the performance of a CsI(Tl) scintillation screen with a dual-periodic structure based on oxidized silicon micro-pore array template in X-ray imaging28. The validity of the simulation method has been confirmed in our related preliminary experiments, which will be presented in our literature in the near future.

Results and discussion

Bottom light output

The bottom light output of the structured CsI(Tl) scintillation screen varies with the top thickness of the Si wall, as shown in Fig. 3. When tSi decreases from 2.00 to 0.70 μm, the BLO increases by 3.47 times. Specifically, when tSi decreases from 2.00 to 1.68 μm, the BLO increases by 38%; when tSi decreases from 1.68 to 1.08 μm, the BLO increases by 69%; when tSi decreases from 1.08 to 0.70 μm, the BLO increases by 49%. That is, when the second structural layer appears, the increase rate of BLO with the decrease of tSi is almost the same as that when only the first structural layer appears; when the third structural layer appears, the increase rate of BLO with the decrease of tSi is higher than that when the first structural layer appears or when the first and second structural layers appear simultaneously.

Figure 3.

Figure 3

Bottom Light output of the structured CsI(Tl) screen varies with the top thickness of Si wall.

The main reason for the increase in BLO is the corresponding increase in the ratio of the filled CsI(Tl) area to that of the entire structured scintillation screen. The reduced optical isolation performance caused by the thinning of the Si wall is the secondary reason for the enhancement of BLO, which accounts for about 15.4%-20.7% of the increment.

The normalized intensity distributions of BLOs of structured CsI(Tl) screens with different tSi were simulated when an narrow X-ray beam is incident along and on the central axis of a scintillation micro-square-frustum. Figure 4 gives the distributions on the horizontal axis with the centre of the micro-square-frustum as the origin. It can be seen that when tSi decreases from 2.00 to 1.08 μm, the distribution is relatively concentrated without obvious change. This shows that the optical isolation performance when both the first and second structural layers, as mentioned in Fig. 1, exist in the scintillation screen is basically the same as the performance when only the first structural layers exists. However, when tSi is less than 1.08 μm, which is when the third structural layer also appears, the distribution becomes more and more diffuse as tSi decrease. That is, the optical isolation performance becomes worse as tSi decreases. The normalized two-dimensional intensity distributions of BLOs with tSi of 0.700, 1.08, 1.68 and 2.00 μm are given in Fig. 5, so as to more intuitively show the influence of the top thickness of Si wall on the optical isolation.

Figure 4.

Figure 4

Normalized intensity distributions of BLOs of the structured CsI(Tl) screens with different tSi.

Figure 5.

Figure 5

Normalized two-dimensional intensity distribution of BLO with the tSi of 0.70, 1.08, 1.68 and 2.00 μm, respectively.

Modulation transfer function

The simulated MTFs for the structured CsI(Tl) with different tSi and the derived spatial resolutions of X-ray imaging are shown in Figs. 6 and 7, respectively. It can be seen that when tSi decreases from 2.00 to 1.08 μm, the MTF changes very little, and the spatial resolution remains good, only decreasing from 114 lp/mm to 110 lp/mm. However, when tSi is less than 1.08 μm, the MTF and spatial resolution deteriorate rapidly with the decrease of tSi. This is because the third structural layer without optical isolation appears and its height gradually increases accordingly. The variations of MTF and spatial resolution with tSi are completely in accordance with the variation of the normalized intensity distribution of BLO. When tSi decreases from 1.08 to 0.70 µm, h3 increases from 0.00 to 8.97 µm, and the corresponding spatial resolutions decreases from 110 to 76 lp/mm.

Figure 6.

Figure 6

MTFs for the structured CsI(Tl) screens with different tSi.

Figure 7.

Figure 7

Spatial resolutions for the structured CsI(Tl) screens with different tSi.

Detective quantum efficiency

The simulated DQEs for the structured CsI(Tl) screens with different tSi are shown in Fig. 8. When tSi decreases from 2.00 to 1.08 μm, the DQE increases at almost all frequency. This is because the BLO increases and MTF does not change significantly. However, when tSi is less than 1.08 μm, the DQE increases at low-frequency and decreases at high-frequency with the decrease of tSi. The reason is that the DQE in the low-frequency region is mainly affected by the X-ray detection efficiency and light guide efficiency, but as the frequency increases, the influence of the MTF on the DQE becomes important. Compared with the DQE when tSi is 1.08 μm, the variation of the turning frequency of DQE from rising to falling with tSi is shown in Fig. 9. It decreases as tSi decreases. The simulated results can give the optimal tSi according to the comprehensive requirements of detection efficiency and spatial resolution in X-ray imaging.

Figure 8.

Figure 8

DQEs for the structured CsI(Tl) screens with different tSi.

Figure 9.

Figure 9

Variation of the turning frequency of DQE from rising to falling with tSi.

Conclusions

In order to improve the bottom light output and spatial resolution of the structured scintillation screen composed of CsI(Tl) micro-square-frustum array, the influence of the Si wall thickness of the micro-square-frustums on the performance of the structured screen in X-ray imaging was studied. The array is a square array with a period of 4 μm. The CsI(Tl) micro-square-frustums are 40 μm high, the cone angle of the frustums is 2.4°, and the top thickness of the Si wall between the two nearest-neighbor frustums can vary from 0.70 to 2.00 µm. If viewed from the side of the structured scintillation screen, there can be three structural layers from top to bottom. They are: (1) The micro-square-frustums are divided by Si walls covered by SiO2 layer on each side; (2) The micro-square-frustums are divided only by the SiO2 layers; (3) There is no separation wall between the micro-square-frustums. When tSi > 1.68 μm, there is only the first structural layer. When 1.08 μm < tSi < 1.68 μm, there are the first and second structural layers. When tSi < 1.08 μm, all three structural layers exist. The performance of the structured CsI(Tl) screen in X-ray imaging was characterized by the bottom light output (BLO), modulation transfer function (MTF) and detective quantum efficiency (DQE). The simulation was carried out by the Geant4 Monte Carlo simulation toolkit, finite-difference time-domain method and Monte Carlo ray-tracing method.

When tSi decreases from 2.00 to 1.08 μm, the BLO of the structured CsI(Tl) screens increases by 233%. The normalized intensity distribution of the BLO is relatively concentrated without obvious change. The results show that the BLO increases with the decrease of tSi, and the optical isolation performance when both the first and second structural layers exist in the scintillation screen is basically the same as the performance when only the first structural layers exists. Meanwhile, the MTF changes very little, and the spatial resolution remains good which is only decreasing from 114 lp/mm to 110 lp/mm. Based on the performance of the BLO and MTF, the DQE increases as tSi decrease at almost all frequency. However, when tSi decreases from 1.08 to 0.70 μm, the third structural layer also appears and thickens. In this case, the BLO of the structured CsI(Tl) screens increases by 49%, but the distribution becomes more and more diffuse. Meanwhile, the MTF deteriorates rapidly, and the spatial resolution decreases from 110 lp/mm to 76 lp/mm. As a result, the DQE increases at low-frequency and decreases at high-frequency with the decrease of tSi. The reason is that the DQE in the low-frequency region is mainly affected by the X-ray detection efficiency and light guide efficiency, but as the frequency increases, the influence of the MTF on the DQE becomes important. Compared with the DQE when tSi is 1.08 μm, the turning frequency of DQE from rising to falling decreases as tSi decreases.

In conclusion, if there is only the first structural layer or the first and second structural layers in the structured scintillation screen composed of CsI(Tl) micro-square-frustum array, the smaller tSi, the better X-ray imaging performance of the screen. However, If there is also the third structural layer in the structured CsI(Tl) scintillation screen, the BLO increases and the MTF deteriorates as tSi decreases. As a result, the DQE increases at low-frequency and decreases at high-frequency with the decrease of tSi. In this case, if a higher image signal-to-noise ratio on a large scale is required, a smaller tSi is preferred, and if a higher image spatial resolution on a small scale is required, a larger tSi is preferred. The results can provide guidance for optimizing tSi according to the comprehensive requirements of detection efficiency and spatial resolution in X-ray imaging.

Acknowledgements

This work is supported by National Natural Science Foundation of China (NSFC) (11675121, 11475128, and 11775160) and the beamline BL13HB at the Shanghai Synchrotron Radiation Facility (2021-SSRF-PT-015196).

Author contributions

M.G. and Z.S. conceived the research, Z.S. conducted the simulation and wrote the main manuscript, M.G. revised the manuscript, X.L., J.Z., B.L., S.H., and C.N. gave simulation instructions, all authors discussed the results and reviewed the manuscript.

Data availability

The data generated or analysed during this study are included in this published article or its supplementary file.

Competing interests

The authors declare no competing interests.

Footnotes

Publisher's note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

References

  • 1.Lecoq P. Development of new scintillators for medical applications. Nucl. Instrum. Methods Phys. Res. A. 2016;809:130–139. doi: 10.1016/j.nima.2015.08.041. [DOI] [Google Scholar]
  • 2.Nikl M. Scintillation detectors for X-rays. Meas. Sci. Technol. 2006;17(4):37–54. doi: 10.1088/0957-0233/17/4/R01. [DOI] [Google Scholar]
  • 3.Nagarkar VV, et al. Structured CsI(Tl) scintillators for X-ray imaging applications. IEEE Trans. Nucl. Sci. 1998;45(3):492–496. doi: 10.1109/23.682433. [DOI] [Google Scholar]
  • 4.Fujieda I, et al. X-ray and charged particle detection with CsI(Tl) layer coupled to a-Si: H photodiode layers. IEEE Trans. Nucl. Sci. 1991;38(2):255–262. doi: 10.1109/NSSMIC.1990.693316. [DOI] [Google Scholar]
  • 5.Jing T, et al. Amorphous silicon pixel layers with cesium iodide converters for medical radiography. IEEE Trans. Nucl. Sci. 1994;41(4):903–909. doi: 10.1109/23.322829. [DOI] [Google Scholar]
  • 6.Simon M, Engel KJ, Menser B, Badel X, Linnros J. X-ray imaging performance of scintillator-filled silicon pore arrays. Med. Phys. 2008;35:968–981. doi: 10.1118/1.2839441. [DOI] [PubMed] [Google Scholar]
  • 7.Badel X, et al. Improvement of an X-ray imaging detector based on a scintillating guides screen. Nucl. Instrum. Methods A. 2002;487(1–2):129–135. doi: 10.1016/S0168-9002(02)00956-7. [DOI] [Google Scholar]
  • 8.Badel X, et al. Metallized and oxidized silicon macropore arrays filled with a scintillator for CCD-based X-ray imaging detectors. IEEE Trans. Nucl. Sci. 2004;51(3):1001–1005. doi: 10.1109/TNS.2004.829579. [DOI] [Google Scholar]
  • 9.Svenonius O, et al. Performance of an X-ray imaging detector based on a structured scintillator. Nucl. Instrum. Methods Phys. Res A. 2009;607(1):138–140. doi: 10.1016/j.nima.2009.03.146. [DOI] [Google Scholar]
  • 10.Hormozan Y, Sychugov I, Linnros J. High-resolution x-ray imaging using a structured scintillator. Med. Phys. 2016;43:696–701. doi: 10.1118/1.4939258. [DOI] [PubMed] [Google Scholar]
  • 11.Liu S, et al. Performance of pixelated CsI scintillation screen with hexagonal array arrangement prepared by vacuum melting injection method. Nucl. Instrum. Methods Phys. Res. A. 2018;903:18–24. doi: 10.1016/j.nima.2018.06.051. [DOI] [Google Scholar]
  • 12.Zhu S, Yu AW, Hawley D, Roy R. Frustrated total internal reflection: A demonstration and review. Am. J. Phys. 1986;54(7):601–607. doi: 10.1119/1.14514. [DOI] [Google Scholar]
  • 13.Chen H, et al. Influence of silicon wall thickness on the performance of structured CsI(Tl) scintillation screen based on oxidized silicon micropore array template in X-ray imaging. Opt. Express. 2019;27(10):14871–14880. doi: 10.1364/OE.27.014871. [DOI] [PubMed] [Google Scholar]
  • 14.Chen H, et al. Effect of CsI(Tl) micro-conical-frustums on the performance of the pixelated CsI(Tl) scintillation screen in X-ray imaging. Nucl. Instrum. Methods Phys. Res. A. 2019;921:18–21. doi: 10.1016/j.nima.2018.12.006. [DOI] [Google Scholar]
  • 15.Lee C, et al. Pore formation in silicon by wet etching using micrometre-sized metal particles as catalysts. J. Mater. Chem. 2008;18(9):1015–1020. doi: 10.1039/b715639a. [DOI] [Google Scholar]
  • 16.Chartier C, Bastide S, Lévy-Clément C. Metal-assisted chemical etching of silicon in HF-H2O2. Electrochim. Acta. 2008;53:5509–5516. doi: 10.1016/j.electacta.2008.03.009. [DOI] [Google Scholar]
  • 17.Agostinelli S, et al. Geant4: A simulation toolkit. Nucl. Instrum. Methods Phys. Res. A. 2003;506(3):250–303. doi: 10.1016/S0168-9002(03)01368-8. [DOI] [Google Scholar]
  • 18.Allison J, et al. Geant4 developments and applications. IEEE Trans. Nucl. Sci. 2006;53(1):270–278. doi: 10.1109/tns.2006.869826. [DOI] [Google Scholar]
  • 19.Taflove A. Computational Electromagnetics: The Finite-Difference Time-Domain Method. 3. Artech House; 2005. [Google Scholar]
  • 20.Stankevich D, Shkuratov Yu. Monte Carlo ray-tracing simulation of light scattering in particulate media with optically contrast structure. JQSRT. 2004;87:289–296. doi: 10.1016/j.jqsrt.2003.12.014. [DOI] [Google Scholar]
  • 21.Samei E, Flynn MJ, Reimann DA. A method for measuring the presampled MTF of digital radiographic systems using an edge test device. Med. Phys. 1998;25:102–113. doi: 10.1118/1.598165. [DOI] [PubMed] [Google Scholar]
  • 22.International Electrotechnical Commission. Medical Electrical Equipment-Characteristics of Digital X-ray Imaging Devices—Part 1–1: Determination of the Detective Quantum Efficiency-Detectors Used in Radiographic Imaging (International Standard, IEC 62220-1-1, 2015).
  • 23.Grassmann H, Lorenz E, Moser HG. Properties of CsI(TI): Renaissance of an old scintillation material. Nucl. Instrum. Methods Phys. Res. A. 1985;228(2–3):323–326. doi: 10.1016/0168-9002(85)90276-1. [DOI] [Google Scholar]
  • 24.Eldridge JE. Cesium iodide (Csl) In: Palik ED, editor. Handbook of Optical Constants of Solids. Academic; 1997. pp. 853–874. [Google Scholar]
  • 25.Edwards DF. Silicon (Si)*. In: Palik ED, editor. Handbook of Optical Constants of Solids. Academic; 1997. pp. 547–569. [Google Scholar]
  • 26.Aspnes DE, Studna AA. Dielectric functions and optical parameters of Si, Ge, GaP, GaAs, GaSb, InP, InAs, and InSb from 1.5 to 6.0 eV. Phys. Rev. B. 1983;27:985–1009. doi: 10.1103/PhysRevB.27.985. [DOI] [Google Scholar]
  • 27.Gao L, Lemarchand F, Lequime M. Refractive index determination of SiO2 layer in the UV/Vis/NIR range: spectrophotometric reverse engineering on single and bi-layer designs. J. Eur. Opt. Soc. Rap. Public. 2013;8:13010. doi: 10.2971/jeos.2013.13010. [DOI] [Google Scholar]
  • 28.Sun Z, et al. Performance of a CsI(Tl) scintillation screen with a dual-periodic structure based on an oxidized silicon micropore array template in X-ray imaging. Opt. Express. 2021;29:23752–23763. doi: 10.1364/OE.428145. [DOI] [PubMed] [Google Scholar]

Associated Data

This section collects any data citations, data availability statements, or supplementary materials included in this article.

Data Availability Statement

The data generated or analysed during this study are included in this published article or its supplementary file.


Articles from Scientific Reports are provided here courtesy of Nature Publishing Group

RESOURCES