Abstract
Objective:
Transcranial magnetic stimulation (TMS) can modulate brain function via an electric field (E-field) induced in a brain region of interest (ROI). The ROI E-field can be computationally maximized and set to match a specific reference using individualized head models to find the optimal coil placement and stimulus intensity. However, the available software lacks many practical features for prospective planning of TMS interventions and retrospective evaluation of the experimental targeting accuracy.
Approach:
The TMS targeting and analysis pipeline (TAP) software uses an MRI/fMRI-derived brain target to optimize coil placement considering experimental parameters such as the subject’s hair thickness and coil placement restrictions. The coil placement optimization is implemented in SimNIBS 3.2, for which an additional graphical user interface (TargetingNavigator) is provided to visualize/adjust procedural parameters. The coil optimization process also computes the E-field at the target, allowing the selection of the TMS device intensity setting to achieve specific E-field strengths. The optimized coil placement information is prepared for neuronavigation software, which supports targeting during the TMS procedure. The neuronavigation system can record the coil placement during the experiment, and these data can be processed in TAP to quantify the accuracy of the experimental TMS coil placement and induced E-field.
Main results:
TAP was demonstrated in a study consisting of three repetitive TMS sessions in five subjects. TMS was delivered by an experienced operator under neuronavigation with the computationally optimized coil placement. Analysis of the experimental accuracy from the recorded neuronavigation data indicated coil location and orientation deviations up to about 2 mm and 2°, respectively, resulting in an 8% median decrease in the target E-field magnitude compared to the optimal placement.
Significance:
TAP supports navigated TMS with a variety of features for rigorous and reproducible stimulation delivery, including planning and evaluation of coil placement and intensity selection for E-field-based dosing.
Keywords: TMS, MRI, coil, electric field, model, neuronavigation, targeting
1. Introduction
Precise placement and intensity selection for non-invasive brain stimulation devices, such as transcranial magnetic (TMS), offers promising avenues for understanding basic (e.g., [1]) and higher (e.g., [2]) cognitive brain functions. In neuroscience experiments, TMS coil placement is often landmark-referenced and supported by neuronavigation technology [3,4], targeting a particular brain region of interest (ROI). These ROIs can be determined individually by identifying anatomical brain structures and their connectivity with MRI [4–6] or from the brain activity or functional connectivity obtained with fMRI or EEG [2,7–16].
Prior to a TMS session, individualized computational modeling of head tissues and the electric field (E-field) induced in the ROI [17] allows for the identification of an optimal TMS coil placement on the scalp [12,18] as well as pulse intensity to deliver a specified E-field strength [2,36,44,45]. We previously developed the fast auxiliary dipole method (ADM) for software-assisted TMS targeting to maximize E-field delivery to an ROI [18], and ADM is now part of the SimNIBS E-field simulation software package. However, it remains challenging to combine ADM with individual functional or anatomical targeting data, subsequent experimental application using neuronavigation, and visualization and evaluation of the experimental accuracy. Previous publications have proposed proof-of-principle fMRI-based TMS targeting pipelines [12,19]. However, these efforts have several limitations with regard to coordinate transformations across software packages, coil placement constraints (e.g., TMS coil or cable in subject’s sight) and optimization, inclusion of hair thickness information, user interface for verification of ROI input information, and retrospective validation of TMS coil placement accuracy.
We present a TMS targeting and analysis pipeline (TAP) software that helps bridge the gaps between individual imaging data, SimNIBS, and neuronavigation. SimNIBS can robustly generate head tissue models [20] from MRI data. The ADM method in SimNIBS determines the optimal TMS coil position and orientation on each subject’s scalp to maximize the magnitude or a directional component of the ROI E-field [18]. In addition to SimNIBS’ basic functionality for adjustment and visualization of ROIs, the coil, and desired E-field parameters, TAP offers a graphical user interface (GUI), called ‘TargetingNavigator’.
TAP can be used for prospective targeting optimization before an experiment and/or retrospective targeting analysis after an experiment. The first step, in a prospective approach for TMS coil placement or E-field simulation, is to extract an ROI center from a volumetric target mask (i.e., a nifti file) for SimNIBS. In this approach, TAP generates specifically formatted ASCII text files readable by TMS neuronavigation software [21]. The presented implementation is for the Brainsight neuronavigation system, but TAP can be adapted to other systems as well. TAP can also control practical aspects of coil placement, such as the coil handle direction, which ideally should not block the TMS coil tracker or the subject’s sight, which may need adjustment to control the induced E-field direction in the cortex [22,23]. Retrospective analysis uses TMS coil placement data from an existing experiment recorded by neuronavigation software. In retrospective analysis, TAP can detect problematic coil distances from the scalp, such as a coil placement being too far from or entering the scalp surface due to inaccurate coil–scalp co-registration, or movement of the TMS coil during the stimulation; this analysis can incorporate measurements of the subjects’ hair thickness, if such are available.
In this paper, we describe TAP and illustrate its use in five TMS subjects who participated in a study of writer’s cramp dystonia.
2. Method
As shown in Figure 1A, TAP combines existing software packages, including SimNIBS v3.2 [20] for TMS-induced E-field simulations and Brainsight v2.5b2 [21] for neuronavigation, with custom MATLAB R2018a [24] code. The following sections describe how TAP was utilized in a complex study paradigm. Briefly, individual TMS target ROIs were derived from fMRI (Section 2.1), optimal coil placements were computed prospectively (Section 2.2) and applied in experimental TMS sessions (section 2.3). The experimental coil placements were recorded and analyzed to assess the targeting accuracy (Section 2.4).
2.1. Experimental participants, MRI, and target selection
2.1.1. Participants
The study [25] was open to all patients with writer’s cramp (WC) dystonia in their right hand who were not on any symptomatic treatments for more than three months and had no contraindications to MRI and TMS. Age-matched healthy volunteers (HV) who were right-hand dominant with no brain disorders were also recruited for the research study. Thirteen WC and 13 HV consented and completed the fMRI visit; of these, five subjects (four WC and one HV) consented and completed the three TMS visits as detailed below.
2.1.2. fMRI research design
All subjects completed a writing task in the MRI scanner during which they were presented with visual instructions to copy a sentence on an MRI-compatible digital tablet. The fMRI acquisition lasted for approximately 7.5 minutes and comprised twelve 20-second blocks of writing separated by periods of rest.
2.1.3. Image acquisition and preprocessing
A total of 13 WC and 13 HV patients completed the structural MRI and functional MRI (fMRI) sequences. A 3 Tesla GE scanner was used to collect structural T1-weighted (echo-planar sequence: voxel size = 1 × 1 × 1.3 mm3, repetition time (TR) = 6.836 s; echo time (TE) = 2.976 s; field of view (FOV) = 25.6 mm2, bandwidth 41.67 Hz/pixel), T2-weighted (echo-planar sequence voxel size = 1 × 1 × 1.3 mm3, TR = 1267.0 ms, TE = 20 ms, flip angle 10°, FOV = 25.6 mm2, bandwidth 31.26 Hz/pixel), and diffusion-weighted scans (voxel size = 1 × 1 × 1.3 mm3, TR = 9000 ms, TE = 92.1 ms, FOV = 25.6 mm, bandwidth 250 Hz/pixel, matrix size 128 × 128, B-value = 2000 s/mm2, diffusion directions = 36). Functional echo-planar images were acquired while subjects performed the writing task using the following parameters: voxel size = 3.5 × 3.5 × 4.0 mm3, TR = 2 s, TE = 30 ms, flip angle = 90°, FOV = 22 cm, bandwidth = 250 Hz/Pixel, interleaved slice order with whole brain coverage). The Cigal software was used to align each subject’s movement to a reference frame to monitor movement in real time [26]. All fMRI images were preprocessed using fMRIPrep software after discarding the first five acquired brain volumes [27].
2.1.4. TMS target selection
Preprocessed fMRI sequences were imported into the FSL software ([28], available at https://fsl.fmrib.ox.ac.uk/fsl/fslwiki) to perform general linear modeling (GLM) at the individual subject level (with spatial smoothing of 5 mm, cluster z-threshold of 2.3, and p-value of 0.05). The GLM model involved convolution of the writing blocks to the double-gamma hemodynamic response function to identify the regions of brain activation for each subject during the writing task compared to the rest period [26,28]. The z-stat brain activation maps of each subject’s writing task were then moved into a second-level group analysis using mixed effects FLAME1 modelling to generate the mean brain activation maps for 13 WC and 13 HV subjects during the writing task. The z-stat brain activation maps for each of the five subjects (four WC and one HV) were then visualized to develop customized ROIs for TMS stimulation. Specifically, each subject’s activation map was overlaid on the respective group activation map and a mask of the structural left premotor cortex or the primary sensory cortex, two regions involved in dystonia, as defined by the Harvard-Oxford MNI atlas [29]. ROI targets for TMS were defined by selecting two neighboring voxels, which were within the atlas-defined left premotor cortex or primary sensory cortex, had peak activation with z-statistic > 2.3, showed overlap between the individual and group activation maps, and were within 4 mm from the cortical surface. The 4 mm rule was used to ensure that the cortical ROI could be reached by the TMS E-field [3]. The individual ROIs were then reversed-transformed into individual native space using FLIRT linear registration with a sinc interpolation. The ROIs in the individual native MRI voxel space were then imported into the TAP workflow to identify the ROI center and transform to the computational model mesh (SimNIBS space) to determine the optimal TMS coil placements.
2.2. TMS targeting and analysis pipeline (TAP)
2.2.1. Coordinate system transformation
Generally, coordinate transformations are needed to integrate all different software pieces (MRI, SimNIBS, and Brainsight) in the pipeline, mediated by custom MATLAB code. The coordinate transformations are defined as 4 × 4 matrices (denoted as Ti,j with i, j = 1, …, 4), which contain submatrices for rotation (Ti,j with i, j = 1, 2, 3) defining axis rotations (with ∥T*,j∥ = 1 for j = 1, 2, 3 ) and translation (Ti,j with i = 1, 2, 3; j = 4 in units of mm; T4,4 = 1), adding a coordinate offset that moves the origin of the current coordinate system to the specified desired one. The matrix Ti,j is used to map locations such as the ROI and the TMS coil center, coordinate space rotations (i.e., colored arrows or letters in Figure 1A–C), and coil coordinate transformations from neuronavigation [21] or canonical coil space into SimNIBS modeling space [20], denoted as ‘Coil matrix’ in Figure 1A. The coordinate system transformations from MRI or Brainsight to SimNIBS space are performed based on mapping coordinates (expressed as world coordinates) relative to the center of the image volume of the respective MRI file created with SimNIBS (SubjectName_T1fs_conform.nii.gz) using either the mri2mesh or headreco pipeline. In more detail, SubjectName_T1fs_conform.nii.gz is present in the LAS coordinate/voxel-order system (except for the headreco option non-conform where it keeps the coordinate system of the input MRI), where the origin is at the LAS corner of the voxel space with positive voxels X, Y and Z axes, respectively. A 4 × 4 transformation matrix or quaternion transformation is stored in either MRI or ROI-defining image file (nifti file, extension: nii.gz) with additional information that is used by default to transform voxel coordinates to world coordinates. The nifti-header information can be modified by a variety of image-processing software, and some may transform any image to one specific voxel-order/coordinate system (e.g., to the RAS in Seg3D, [30]), creating the appearance of being at a proper location.
When reading original MRI and SimNIBS-modified MRI (i.e., SubjectName_T1fs_conform.nii.gz), TAP can detect differences in coordinate system orientation and voxel size (i.e., for models generated with mri2mesh SimNIBS 3.x and SimNIBS headreco with orientation parameter: keep/standard/custom) and determines a linear transformation matrix using FSL’s flirt command (cost function: mutual information) [28]. In the same way, targets defined on a brain atlas can be used or defined as MNI coordinate values for models generated with SimNIBS 3.x. The head models used in our experimental study were created using SimNIBS 2.0.1 mri2mesh pipeline and once detected, TAP converted them, including conductivity information, to be usable in SimNIBS 3.x. The mri2mesh pipeline in SimNIBS 2.x resamples the MRI, if needed, to a voxel size of 1 mm3 and clips the image volume to 2563 voxel image and aligns (compared to SimNIBS 3.x) with the original MRI in world coordinates.
To further ensure the correctness of the transformed ROI coordinate and alert the user to potential issues, TAP computes the shortest Euclidean distance to the next segmented gray matter volume in SimNIBS MRI-space (i.e., SubjectName_T1fs_conform.nii.gz) using gm_fromMesh.nii.gz SimNIBS v3 or gm_only.nii.gz for SimNIBS v2 mri2mesh runs. If this distance is larger than the voxel size, then the ROI and gray matter do not overlap. In this problematic case, which is most likely due to voxel-order/coordinate transformation misalignment. This could happen if FLIRT’s registration failed, i.e., in batch-mode without user interaction and TAP can be configured to systematically sign-flip the axes/origin to find a world coordinate transformation with minimal distance to a brain voxel and notify the user. And, if this alignment is not possible, TargetingNavigator is launched for manual inspection and adjustment.
Once the ROI voxels are in SimNIBS MRI-space (i.e., in SubjectName_T1fs_conform.nii.gz) as world coordinates (X, Y, Z), they need to be converted to SimNIBS model space (X′, Y′, Z′) by representing the coordinates relative to the image center and flip axis (X′ = −X, Y′ = −Y, Z′ = Z, see Figure 1B). The physical orientation of the TMS coil is defined in the SimNIBS space (denoted X′′, Y′′, Z′′) and Brainsight space (denoted as X′′′, Y′′′, Z′′′), while flipping by 180° the pitch (X′′′ = −X′′) and yaw coil orientation axes (i.e., Z′′′ = −Z′′). Coil orientations are visualized by black-colored coil representations in Figure 1B and 1C, respectively. For biphasic TMS pulses, TMS is most effective when the direction of the middle phase of the induced E-field points to the wall of the cortical sulci [2,23,31,32]. As described in section ‘Preparation and launch of SimNIBS’, TAP estimates this direction based on the gyral/sulcal transition zone in the ROI (i.e., through cortical curvature information in SimNIBS mri2mesh; for more details, see below) to find a representative location on the wall for which it computes a smoothed inward-pointing normal vector. This vector is then used as the ROI directional E-field constraint in the ADM algorithm to determine the optimal TMS coil setup. After ADM has finished, the optimal coil orientation is prepared for Brainsight considering the following experimental constraints: The Z′′′ axis is checked for pointing in the scalp-outward direction (indicated visually by a large green cylinder cone in Brainsight).
For TMS devices that allow electronic reversal of the coil current direction, there is the option to flip the coil orientation and current direction for most practical experimental coil placements. It is advantageous to maintain the orientation of the TMS coil handle, including any attached cable, pointing away from the subject’s face, and TMS tracking headgear. To achieve this, TAP projects the optimized TMS coil orientation (Y′′) onto the anterior–posterior (X′ − Z′) plane and then computes an angle α with the X′ axis (equal to the zero-degree axis and positive angles for clockwise rotation). If 90° < α < 270°, then the coil orientation is flipped automatically, and the necessity for coil current reversal is highlighted in the prepared Brainsight files.
2.2.2. Preparation and launch of SimNIBS
TMS coil placement optimization can be launched with TAP by executing the run_simnibs(s) command from the MATLAB prompt (or within a script as in TAP) providing a MATLAB struct object s with different data fields, which should be initialized with the default struct opt_struct(‘TMSoptimize’). If s has the data field entry s.target_direction, ADM or the direct solver in SimNIBS will find the coil placement that maximizes the average E-field induced in the ROI along the vector s.target_direction, or otherwise the overall magnitude, as specified. The ROI center location in the SimNIBS space is assigned to the data field. The ROI size (s.target_size) considers all tetrahedral gray matter elements that fall within this radius (from the point specified in s.target) as Euclidian distance in three dimensions. For our five TMS subjects, an ROI size value between 3 and 4 mm was chosen to ensure that the ROI covered a significant part of the sulcal wall, but not any part of a neighboring gyrus controlled for by visual inspection in TargetingNavigator. The ROI target direction was determined using the following main steps:
project the ROI center onto the smoothed brain surface.
project location from (1) to the closest scalp surface mesh point and compute its nodal normal (averaging the neighboring triangle normal vectors) representing the coil plane.
-
find the sulcal wall location/trajectory (convex–concave transition of sulcus and gyrus):
(3.1) mri2mesh: use location from (1) and Freesurfer’s surface curvature information (files: lh.sulc, rh.sulc; value range: between −2 and 2) to find the nearest surface mesh point that is numerically closest to zero.
(3.2) headreco: use location from (1) and its scalp projection to form a trajectory vector.
compute an outwards-pointing nodal normal by averaging the triangle normal vectors for the brain surface nodes.
-
find sulcal wall surface normal and adjust for scalp coil plane:
(5.1) mri2mesh: determine the nodal normal (4) for location (3.1).
(5.2) headreco: move up/down the trajectory vector to find a location for which its nodal normal is mostly parallel to the tangential plane at the closest scalp node (2).
The sulcal wall normal vector found in (5) for mri2mesh or headreco, respectively, is then adjusted by projection onto the tangential plane (using the closest scalp node normal vector) to create the desired ROI target direction vector, because the TMS-induced E-field is mostly tangential to the scalp surface. In many situations, this automatic approach to find the closest sulcus wall and a suitable s.target_direction vector performs well; however, if the local brain geometry is too complex, a human user may perform better by using TargetingNavigator.
For robust estimation of the scalp-tangential plane, we added a data field to the s structure (available in SimNIBS 3.2.5), namely s.scalp_normals_smoothing_steps (TAP default is 20), to adjust the coil tangentiality to the scalp and visualize it in TargetingNavigator. Even though this can avoid automatically obvious problems of a TMS coil being not scalp-tangential, our tests showed that even subtle differences in scalp-tangentiality of the coil can have significant effects in practice, and higher values for s.scalp_normals_smoothing_steps may be needed.
SimNIBS allows constraining the scalp search space for possible coil center positions and orientations, which TAP sets to the default values of s.search_radius = 25 mm, s.spatial_resolution = 1 mm, and s.search_angle = 1°. For a typical SimNIBS head model and the default values, the optimal coil placement search takes approximately 2 min on a regular laptop (Intel Core i5 1.6 GHz, 8 GB RAM) running ADM.
2.2.3. Prospective TMS dosing
For the five TMS subjects, we used mri2mesh/dwi2cond (SimNIBS 2.0.1) to generate a five-tissue head model from the subject’s T1-weighted MRI data set, and assigned to the resulting tetrahedral mesh elements isotropic conductivities for the scalp (0.465 S/m), skull (0.01 S/m), and cerebrospinal fluid (1.654 S/m), as well as anisotropic conductivities for the gray and white brain matter. The anisotropic conductivities were derived from the subject’s co-registered diffusion-weighted MRI, where the diffusion tensors are volume-normalized so that the geometric mean of the conductivity tensor eigenvalues equals literature-based values of 0.275 S/m and 0.126 S/m for gray and white matter, respectively.
The subject’s hair thickness is often unknown before the TMS session [2], but can be measured at the beginning of the session. Therefore, multiple runs of SimNIBS’s ADM are required for a range of hair thickness values. The default setting in TAP iterates through a reasonable range of hair thicknesses from 0 to 7.5 mm in steps of 0.5 mm. The optimal coil setup was chosen for TMS targeting (i.e., as loaded into Brainsight) while a TMS intensity related to the subject’s resting motor threshold was employed as described in Section 2.3.
The TMS pulse intensity, specified as the coil current rate of change (dI/dt), can be set as a parameter directly in SimNIBS (default: s.didt = 106 A/s = 1 A/μs) or through the TargetingNavigator interface. TAP also offers the option to determine the required dI/dt to reach a desired E-field strength in the ROI [2,34,35], which is an important feature for ROI-based dosing of TMS. The desired directional or magnitude E-field strength in the ROI can be specified in TAP by one of several metrics: the maximal value as well as statistically more robust metrics, such as the 50th, 75th, 90th, and 99th percentiles, to avoid numerical E-field outliers. Our previously proposed E20, E50, and E100 metrics, which define the minimum E-field strength within the 20, 50, and 100 voxels with the strongest E-field in the ROI [2,36]. The computed dI/dt can be entered into the TMS device, typically as a percentage of the maximum stimulator output (%MSO), based on the approximately linear relationship between dI/dt and %MSO [2,36]. This relationship can generally be described by the formula
(1) |
where coefficients a and b can be fitted to empirical data obtained from TMS devices that provide both %MSO and dI/dt [A/μs] readout. In this case, dI/dt should be measured across the full range of intensity for a specific coil, for example, with 5% MSO steps [2,36]. For devices that do not display dI/dt, the most practical approach is to set b = 0, and compute
(2) |
from the TMS device maximum capacitor voltage, VC,max, corresponding to 100 %MSO, and the specific coil inductance L, both of which can be obtained from the manufacturer. For example, VC,max is 1,800 V, 1,670 V, and 2,800 V for the standard MagVenture MagPro, Magstim Rapid2, and Magstim 2002 devices, respectively. The TargetingNavigator interface incorporates the ability to solve equation (1) for either dI/dt or %MSO with specified a and b, and provides values for these coefficients for some common TMS device and coil combinations.
2.2.4. Neuronavigation files
TAP generates different types of files that can be visualized in Brainsight:
scalp and brain surface of the SimNIBS head model converted to MRI/Brainsight space,
SimNIBS/Freesurfer preprocessed T1-weighted MRI data set,
ROI/scalp coil center converted to MRI/Brainsight space, and
optimal coil placement (fully defined by a 4 × 4 matrix) was converted to the Brainsight space and saved in a text file.
For retrospective analysis, the experimenter saved the coil placement data as text files (i.e., for Brainsight software) during the TMS session, which can then be read into MATLAB and visualized in TargetingNavigator.
2.3. Experimental TMS sessions
The optimal TMS coil placement information of the five subjects who received TMS was imported as a project file (comprising a cortical target for premotor cortex or primary sensory cortex, T1-weigthed MRI, scalp/brain surface, and hair thickness estimations) into the Brainsight software for the TMS sessions. All TMS sessions were performed in a double-blind, sham-controlled, crossover research design. Three TMS sessions per subject were performed in a randomized order: active TMS at the premotor cortex, sham TMS at the premotor cortex, and active TMS at the primary sensory cortex, with each session one week apart to allow washout of the rTMS neuromodulatory effects. Each session was divided into four blocks, each 5 min apart. During each block, subjects received 25 trains of 3.9 s, 10 Hz repetitive TMS with an intertrain interval of 10 s at 90% resting motor threshold with biphasic stimulation for a total of 1,000 and 4,000 pulses per block and session, respectively. TMS was delivered using a MagPro X100 stimulator with an A/P Cool-B65 coil (MagVenture, Alpharetta, GA, USA). A neuronavigation system (Brainsight, Rogue Research, Montreal, Canada; [21]) was used to co-register the T1-MRI and the fMRI-based TMS target developed with TAP to the subject’s scalp. The location of the subject’s hot spot over the left motor cortex that induced the strongest motor evoked potentials in the contralateral first dorsal interosseus (FDI) was determined and marked in Brainsight. The resting motor threshold (rMT, expressed as %MSO) was then determined using a maximum-likelihood-based estimation procedure (see Table 1) [37]. The subject’s hair thickness over the TMS target area was measured (see Table 1) [2] and the closest hair thickness value in the reference table provided by TAP was selected. The TMS coil position and orientation values corresponding to the selected hair thickness were then used for each TMS session. During sham stimulation, the A/P coil was flipped, and subjects received similar auditory stimulation and scalp sensation through two surface electrodes placed on the subject’s scalp around the TMS target, producing only approximately 7.7% of the active E-field stimulation in the brain [38]. For all TMS sessions, the experimenter positioned the coil tangentially to the scalp surface with its handle directed posteriorly with a goal angle deviation < 5°. The coil placement data were saved by Brainsight every 500 ms (every fifth pulse, to prevent any lags during stimulation) during the rTMS trains, totalling 200 recorded TMS coil setups per block, and exported as text files. Recording the coil position at this reduced rate compared to the pulse train rate was necessary for the neuronavigation system (Brainsight) to be able to save all data points correctly. No adverse events due to stimulation have been reported.
Table 1.
Subject | rTMS intensity setting (% MSO) | Hair thickness (mm) | ||||
---|---|---|---|---|---|---|
Session target & TMS condition | Session target & TMS condition | |||||
Premotor, Sham | Premotor, Active | Primary sensory, Active | Premotor, Sham | Premotor, Active | Primary sensory, Active | |
S1 | 52 | 53 | 47 | 0.5 | 0.5 | 0.5 |
S2 | 76 | 69 | 64 | 2.5 | 2.0 | 2.5 |
S3 | 39 | 41 | 38 | 1.5 | 1.5 | 2.5 |
S4 | 46 | 50 | 47 | 2.0 | 2.5 | 2.0 |
S5 | 55 | 58 | 45 | 2.0 | 2.0 | 3.5 |
2.4. Retrospective Analysis
The Brainsight neuronavigation system can digitize the coil position and orientation for each TMS pulse. TAP reads in (read_brainsight_file) the Brainsight-space exported text files and converts each line, corresponding to one digitized coil placement defined by a 4 × 4 transfer matrix, to SimNIBS modeling space (convert_coord_brainsight_2_simnibs). To avoid possible errors in the coil–scalp co-registration resulting in the coil center being inside the scalp, it is recommended to use Brainsight’s “Snap to Scalp” option when exporting [21]. TAP then extrudes these snapped coil centers outwards along the scalp normal to account for the measured hair thickness. If hair thickness measurement is unavailable, TAP results in a value of 2 mm [2]. The distance between the hair-thickness-extruded coil center and the original experimentally recorded coil center without snapping and hair-thickness extrusion was determined along the recorded coil normal direction (‘normal coil deviation’ in mm). The hair-thickness-extruded coil center is then compared against the computationally optimal coil center on the scalp’s tangential plane (‘tangential coil deviation’ in mm). TAP uses the computationally optimal coil placement to check if the rotational part of the experimental coil placement matrices aligns, and sign-flips the components for minimal angular discrepancy. The scalp normal direction was used to assess the deviation of the coil plane from the scalp-tangential plane (‘normal coil deviation’ in angular degree (°) equals to the arccosine of the dot product of the scalp and coil normal). Finally, the experimental coil placement is projected onto the scalp-tangential plane and the difference between the experimental and optimal coil orientation is computed (‘tangential coil deviation’ in angular degree (°)). When a targeted optimal coil placement is not available, the TAP can still analyze the experimental TMS coil placements recorded with Brainsight.
As a second step, TAP evaluates the difference in induced E-field between the optimal (if one is provided) and all experimentally recorded coil setups. Because one E-field simulation requires a computation time of several minutes, TAP avoids a very long total simulation time by choosing a median representative of the coil placements within an experimental block to simulate the E-field in the entire head model and then extracts the ROI E-field (using SimNIBS target.msh). The median coil representation is chosen by evaluating the medians for the X′′′, Y′′′ and Z′′′ coordinates separately and picking the recorded coil placement with the smallest Euclidean distance from these median coordinates. For the E-field simulations, the value of dI/dt is computed with Equation (1) from the individual rMT in units of %MSO, and entered into the s.didt field of the MATLAB structure inputted to SimNIBS. For the TMS coil and device in this study, coefficient values of a = 1.5662 and b = −2.3237 were used in Equation (1) based on prior calibration data [2]. Finally, TAP computes the deviation of magnitude and angle of the E-field vector in each tetrahedral ROI element between the simulations for the median experimental coil placement and the optimal coil placement.
2.5. Summary of steps for prospective and retrospective TAP use
Before TMS session:
Define volumetric ROI (*.nii.gz file)
Convert ROI center location to model space (SimNIBS)
Determine sulcal wall normal for E-field optimization (optional)
Optimize E-field (ADM) for particular hair thickness
Compute TMS intensity setting for specified E-field strength in ROI (optional)
Write TMS coil setup file for neuronavigation
Measure hair thickness over target and select setup file
Apply TMS and record TMS coil placement
After TMS session:
Convert recorded TMS coil placements to model space
Quantify coil placement deviations
Simulate and characterize E-field for coil placements
3. Results
Figure 2B–G quantifies the deviation between the computationally optimized coil placement that was set as a target during the TMS sessions and the actual placements recorded by the neuronavigation system. Across all subjects and sessions, the median coil position and orientation deviations were −1.55 mm and 1.28° relative to the scalp normal (Figure 2B, D) and 2.18 mm and −0.29° in the scalp tangential plane (Figure 2C, E), respectively. The corresponding median E-field magnitude and direction deviations were 7.69% and 0.65° (Figure 2F, G), respectively, even though some of the individual medians were significantly larger.
These values are comparable to the neuronavigation error and deviation estimates reported in the literature. TMS neuronavigation systems are known to have a coil placement accuracy of 5–6 mm on average [39,40]. Excluding errors from the registration and shifts of the head tracker relative to the head, the median tangential deviation of the coil position and orientation of 1.34 mm and 3.48°, respectively, was reported for a neuronavigated robotic coil holder [41]. Moreover, an inter-session position error of approximately 2.5 mm was reported for neuronavigated manual coil placement [42].
4. Discussion
The novel targeting and analysis pipeline (TAP) integrates the SimNIBS E-field simulation package with the Brainsight neuronavigation system using custom MATLAB code to improve the practicality of precise brain targeting in TMS experiments using E-field simulations. The proposed capabilities of TAP software go beyond what has been previously published (e.g., [12]). TAP enables prospective and retrospective TMS targeting analysis based on E-field simulations coupled with individual imaging and neuronavigation data. Prospective targeting computations include the optimization of the TMS coil placement as well as the selection of the TMS pulse intensity to deliver a specific E-field strength at the target ROI. A retrospective analysis of the experimental coil placement can quantify the overall accuracy of applying TMS. For example, retrospective coil placement analysis can be used to remove problematic TMS sessions, blocks, or pulses, perform statistical analyses to explain experimental outcomes, and draw more accurate conclusions from hypothesis-driven TMS research.
We demonstrated the use of TAP in an example study in which TMS coil positioning deviations did not exceed the error levels of neuronavigation systems. This, of course, depends on the precision of coil handling across experimental blocks, sessions, and subjects, and should be evaluated in each study. Moreover, the coil positioning deviations are detected only relative to the target placement set in the neuronavigation coordinates. The absolute precision relative to the actual brain target also depends on the accuracy of the neuronavigation registration. Therefore, the initial registration must be accurate, and it may be prudent to verify the registration of head landmarks and recorded scalp coil placements at the end and potentially within an experimental session [43].
TAP is an extension of SimNIBS, which is freely available. It is our hope that TAP will enable more researchers to utilize SimNIBS E-field simulation capabilities as part of rigorous TMS targeting and analysis methods. Deployment of TAP across various studies can help determine advantages and disadvantages of E-field-based targeting and dosing for TMS. TAP also facilitates readout from the Brainsight neuronavigation system to uncover potential issues regarding coil placement, including coil tracking and/or head registration. Presently, the TAP code supports only Brainsight, but we aim to extend compatibility with other neuronavigation systems in the future. Finally, we welcome input and contributions to the TAP code by users through the contact on the TAP GitHub page linked in the Data & Code Availability section.
5. Conclusion
TAP facilitates prospective planning and retrospective analysis of TMS coil placement and E-field delivery to individual brain targets. This functionality enables both the selection of the TMS coil placement and intensity to standardize E-field exposure as well as the quantification of variability in the coil placement and E-field dose across different subjects and experimental sessions. We demonstrated the use of TAP in an example study in which TMS coil positioning deviations did not exceed the error levels of neuronavigation systems. TAP can be a useful tool for TMS practitioners to precisely plan dosing and analyze experiments or interventions to ensure rigorous and reproducible stimulation delivery.
Acknowledgements
This work was supported by grants from the National Institutes of Health (RF1MH114268, RF1MH114253, U01AG050618, KL2TR002554), Dystonia Medical Research Foundation, Doris Duke Charitable Foundation, and Dystonia Coalition. The content is solely the responsibility of the authors and does not necessarily represent the official views of the funding agencies.
A. V. Peterchev has received research funding, travel support, patent royalties, consulting fees, equipment loans, hardware donations, and/or patent application support related to magnetic stimulation from Rogue Research, Magstim, MagVenture, Neuronetics, BTL Industries, and Advise Connect Inspire. The other authors report no relevant disclosures.
Footnotes
Ethics Approval, Patient Consent, and Clinical Trial Registration
The human subjects participating in the fMRI and TMS experiments were part of a multi-session study of dystonia, including structural and functional magnetic resonance imaging and behavioral testing. The study was approved by the Duke University Health System Institutional Review Board (#0094131), ensuring ethical guidelines were followed, and all subjects provided written informed consent to participate. The study was not registered as a clinical trial because it was a small pilot feasibility study; moreover, this paper focuses on methodological aspects of TMS targeting and does not report any clinical results.
Data and Code Availability
The software described herein as well as resulting data to recreate all figures are available on https://github.com/moritzdannhauer/TAP.git.
References
- [1].Jung J, Bungert A, Bowtell R and Jackson SR 2020. Modulating brain networks with transcranial magnetic stimulation over the primary motor cortex: a concurrent TMS/fMRI study Front. Hum. Neurosci 14 31. [DOI] [PMC free article] [PubMed] [Google Scholar]
- [2].Beynel L, Davis SW, Crowell CA, Dannhauer M, Lim W, Palmer H, Hilbig SA, Brito A, Hile C, Luber B, Lisanby SH, Peterchev AV, Cabeza R and Appelbaum LG 2020. Site-specific effects of online rTMS during a working memory task in healthy older adults Brain Sci. 10 255. [DOI] [PMC free article] [PubMed] [Google Scholar]
- [3].Aberra AS, Wang B, Grill WM and Peterchev AV 2020. Simulation of transcranial magnetic stimulation in head model with morphologically-realistic cortical neurons Brain Stimulat. 13 175–89 [DOI] [PMC free article] [PubMed] [Google Scholar]
- [4].Rusjan PM, Barr MS, Farzan F, Arenovich T, Maller JJ, Fitzgerald PB and Daskalakis ZJ 2010. Optimal transcranial magnetic stimulation coil placement for targeting the dorsolateral prefrontal cortex using novel magnetic resonance image-guided neuronavigation Hum. Brain Mapp. 31 1643–52 [DOI] [PMC free article] [PubMed] [Google Scholar]
- [5].Nummenmaa A, McNab JA, Savadjiev P, Okada Y, Hämäläinen MS, Wang R, Wald LL, Pascual-Leone A, Wedeen VJ and Raij T 2014. Targeting of white matter tracts with transcranial magnetic stimulation Brain Stimulat. 7 80–4 [DOI] [PMC free article] [PubMed] [Google Scholar]
- [6].Silva LM, Silva KMS, Lira-Bandeira WG, Costa-Ribeiro AC and Araújo-Neto SA 2021. Localizing the primary motor cortex of the hand by the 10–5 and 10–20 systems for neurostimulation: an MRI study Clin. EEG Neurosci. 52 427–35 [DOI] [PubMed] [Google Scholar]
- [7].Wang JX, Rogers LM, Gross EZ, Ryals AJ, Dokucu ME, Brandstatt KL, Hermiller MS and Voss JL 2014. Targeted enhancement of cortical-hippocampal brain networks and associative memory Science 345 1054–7 [DOI] [PMC free article] [PubMed] [Google Scholar]
- [8].Bergmann TO, Karabanov A, Hartwigsen G, Thielscher A and Siebner HR 2016. Combining non-invasive transcranial brain stimulation with neuroimaging and electrophysiology: current approaches and future perspectives Neuroimage 140 4–19 [DOI] [PubMed] [Google Scholar]
- [9].Zrenner C, Belardinelli P, Müller-Dahlhaus F and Ziemann U 2016. Closed-loop neuroscience and non-invasive brain stimulation: a tale of two loops Front. Cell. Neurosci 10 92. [DOI] [PMC free article] [PubMed] [Google Scholar]
- [10].Thut G, Bergmann TO, Fröhlich F, Soekadar SR, Brittain J-S, Valero-Cabré A, Sack AT, Miniussi C, Antal A and Siebner HR 2017. Guiding transcranial brain stimulation by EEG/MEG to interact with ongoing brain activity and associated functions: a position paper Clin. Neurophysiol 128 843–57 [DOI] [PMC free article] [PubMed] [Google Scholar]
- [11].Curtin A, Tong S, Sun J, Wang J, Onaral B and Ayaz H 2019. A systematic review of integrated functional near-infrared spectroscopy (fNIRS) and transcranial magnetic stimulation (TMS) studies Front. Neurosci. 13 84. [DOI] [PMC free article] [PubMed] [Google Scholar]
- [12].Balderston NL, Roberts C, Beydler EM, Deng Z-D, Radman T, Luber B, Lisanby SH, Ernst M and Grillon C 2020. A generalized workflow for conducting electric field–optimized, fMRI-guided, transcranial magnetic stimulation Nat. Protoc. 15 3595–614 [DOI] [PMC free article] [PubMed] [Google Scholar]
- [13].Cash RF, Weigand A, Zalesky A, Siddiqi SH, Downar J, Fitzgerald PB and Fox MD 2020. Using brain imaging to improve spatial targeting of transcranial magnetic stimulation for depression Biol. Psychiatry 90 689–700 [DOI] [PubMed] [Google Scholar]
- [14].Ding Z, Ouyang G, Chen H and Li X 2020. Closed-loop transcranial magnetic stimulation of real-time EEG based on the AR mode method Biomed. Phys. Eng. Express 6 035010. [DOI] [PubMed] [Google Scholar]
- [15].Esposito R, Bortoletto M and Miniussi C 2020. Integrating TMS, EEG, and MRI as an approach for studying brain connectivity The Neuroscientist 26 471–86 [DOI] [PubMed] [Google Scholar]
- [16].Zhang B, Liu J, Bao T, Wilson G, Park J, Zhao B and Kong J 2020. Locations for noninvasive brain stimulation in treating depressive disorders: A combination of meta-analysis and resting-state functional connectivity analysis Aust. N. Z. J. Psychiatry 54 582–90 [DOI] [PubMed] [Google Scholar]
- [17].Gomez LJ, Dannhauer M, Koponen LM and Peterchev AV 2020. Conditions for numerically accurate TMS electric field simulation Brain Stimulat. 13 157–66 [DOI] [PMC free article] [PubMed] [Google Scholar]
- [18].Gomez LJ, Dannhauer M and Peterchev AV 2021. Fast computational optimization of TMS coil placement for individualized electric field targeting Neuroimage 228 117696. [DOI] [PMC free article] [PubMed] [Google Scholar]
- [19].Opitz A, Fox MD, Craddock RC, Colcombe S and Milham MP 2016. An integrated framework for targeting functional networks via transcranial magnetic stimulation Neuroimage 127 86–96 [DOI] [PMC free article] [PubMed] [Google Scholar]
- [20].Thielscher A, Antunes A and Saturnino GB 2015. Field modeling for transcranial magnetic stimulation: a useful tool to understand the physiological effects of TMS? 2015 37th annual international conference of the IEEE engineering in medicine and biology society (EMBC) (IEEE) pp 222–5 [DOI] [PubMed] [Google Scholar]
- [21].Brainsight DT 2020. Brainsight (Montreal, Canada: Rogue Research Inc.) [Google Scholar]
- [22].Brasil-Neto JP, Cohen LG, Panizza M, Nilsson J, Roth BJ and Hallett M 1992. Optimal focal transcranial magnetic activation of the human motor cortex: effects of coil orientation, shape of the induced current pulse, and stimulus intensity. J. Clin. Neurophysiol. Off. Publ. Am. Electroencephalogr. Soc 9 132–6 [PubMed] [Google Scholar]
- [23].Balslev D, Braet W, McAllister C and Miall RC 2007. Inter-individual variability in optimal current direction for transcranial magnetic stimulation of the motor cortex J. Neurosci. Methods 162 309–13 [DOI] [PubMed] [Google Scholar]
- [24].MathWorks 2018 MATLAB (Natick, Massachusetts, USA: The MathWorks Inc.) [Google Scholar]
- [25].Bukhari-Parlakturk N, Fei M, Voyvodic J and Michael AM 2021. Data Driven Exploration of Network Connectivity in task-fMRI of Focal Hand Dystonia medRxiv 10.1101/2021.05.14.21257239 [DOI] [Google Scholar]
- [26].Voyvodic JT 1999. Real-time fMRI paradigm control, physiology, and behavior combined with near real-time statistical analysis Neuroimage 10 91–106 [DOI] [PubMed] [Google Scholar]
- [27].Esteban O, Markiewicz CJ, Blair RW, Moodie CA, Isik AI, Erramuzpe A, Kent JD, Goncalves M, DuPre E and Snyder M 2019. fMRIPrep: a robust preprocessing pipeline for functional MRI Nat. Methods 16 111–6 [DOI] [PMC free article] [PubMed] [Google Scholar]
- [28].Jenkinson M, Beckmann CF, Behrens TE, Woolrich MW and Smith SM 2012. Fsl Neuroimage 62 782–90 [DOI] [PubMed] [Google Scholar]
- [29].Desikan RS, Ségonne F, Fischl B, Quinn BT, Dickerson BC, Blacker D, Buckner RL, Dale AM, Maguire RP and Hyman BT 2006. An automated labeling system for subdividing the human cerebral cortex on MRI scans into gyral based regions of interest Neuroimage 31 968–80 [DOI] [PubMed] [Google Scholar]
- [30].CIBC 2015 Seg3D: Volumetric Image Segmentation and Visualization. Scientific Computing and Imaging Institute (SCI) [Google Scholar]
- [31].Corthout E, Barker A and Cowey A 2001. Transcranial magnetic stimulation Exp. Brain Res 141 128–32 [DOI] [PubMed] [Google Scholar]
- [32].Kammer T, Beck S, Thielscher A, Laubis-Herrmann U and Topka H 2001. Motor thresholds in humans: a transcranial magnetic stimulation study comparing different pulse waveforms, current directions and stimulator types Clin. Neurophysiol. 112 250–8 [DOI] [PubMed] [Google Scholar]
- [33].Bechthold B 2016. Violin Plots for Matlab, https://github.com/bastibe/Violinplot-Matlab
- [34].Caulfield KA, Li X and George MS 2021. Four electric field modeling methods of Dosing Prefrontal Transcranial Magnetic Stimulation (TMS): Introducing APEX MT dosimetry Brain Stimul. Basic Transl. Clin. Res. Neuromodulation 14 1032–4 [DOI] [PMC free article] [PubMed] [Google Scholar]
- [35].Caulfield KA, Li X and George MS 2021. A reexamination of motor and prefrontal TMS in tobacco use disorder: Time for personalized dosing based on electric field modeling? Clin. Neurophysiol 132 2199–207 [DOI] [PMC free article] [PubMed] [Google Scholar]
- [36].Beynel L, Dannhauer M, Plamer H, Hilbig SA, Crowell CA, Wang JE-H, Micheal AM, Wood EA, Luber B, Lisanby SH, Peterchev AV, Cabeza R, Davis SW and Appelbaum LG 2021. Network-based rTMS to modulate working memory: the difficult choice of effective parameters for online interventions. Brain and Behavior 11 e2361. [DOI] [PMC free article] [PubMed] [Google Scholar]
- [37].Awiszus F and Borckardt J 2011. TMS Motor Threshold Assessment Tool (MTAT 2.0)
- [38].Smith JE and Peterchev AV 2018. Electric field measurement of two commercial active/sham coils for transcranial magnetic stimulation J. Neural Eng 15 054001. [DOI] [PMC free article] [PubMed] [Google Scholar]
- [39].FDA 2020. Neuronal Navigator (New York City, USA: ) [Google Scholar]
- [40].Ruohonen J and Karhu J 2010. Navigated transcranial magnetic stimulation Neurophysiol. Clin. Neurophysiol 40 7–17 [DOI] [PubMed] [Google Scholar]
- [41].Goetz SM, Kozyrkov IC, Luber B, Lisanby SH, Murphy DL, Grill WM and Peterchev AV 2019. Accuracy of robotic coil positioning during transcranial magnetic stimulation J. Neural Eng. 16 054003. [DOI] [PMC free article] [PubMed] [Google Scholar]
- [42].Schönfeldt-Lecuona C, Thielscher A, Freudenmann RW, Kron M, Spitzer M and Herwig U 2005. Accuracy of stereotaxic positioning of transcranial magnetic stimulation Brain Topogr. 17 253–9 [DOI] [PubMed] [Google Scholar]
- [43].Goetz SM and Kammer T 2021. Neuronavigation The Oxford Handbook of Transcranial Stimulation. (Oxford University: Oxford University Press; ) [Google Scholar]
- [44].Zmeykina E, Mittner M, Paulus W, Zsolt T 2020. Weak rTMS-induced electric fields produce neural entrainment in humans. Sci. Rep. 10 1–16 [DOI] [PMC free article] [PubMed] [Google Scholar]
- [45].Caulfield KA, George MS 2021. Four electric field modelling methods of Dosing Prefrontal Transcranial Magnetic Stimulation (TMS). Brain Stimul. 14 1032–34 [DOI] [PMC free article] [PubMed] [Google Scholar]
Associated Data
This section collects any data citations, data availability statements, or supplementary materials included in this article.
Data Availability Statement
The software described herein as well as resulting data to recreate all figures are available on https://github.com/moritzdannhauer/TAP.git.