Skip to main content
. 2022 Jun 4;81:104095. doi: 10.1016/j.ebiom.2022.104095

Table 2.

Method validation for the precision (% RSD) and accuracy (% bias) of the UHPLC-MS/MS method for determination of GS-441524 in the maternal blood, placenta, fetus and amniotic fluid dialysate of pregnant rats.

Nominal concentration (ng/mL) Intra-day
(n=5)
Inter-day
(n=3)
Observed concentration (ng/mL) Precision (% RSD) Accuracy (% bias) Observed concentration (ng/mL) Precision (% RSD) Accuracy (% bias)
Maternal blood
5 5.04 ± 0.42 8.32 0.77 4.82 ± 0.13 2.76 -3.53
10 10.10 ± 0.65 6.42 0.98 10.04 ± 0.28 2.78 0.40
100 101.90 ± 4.39 4.30 1.90 96.76 ± 6.04 6.24 -3.24
500 501.3 ± 26.26 5.24 0.26 499.01 ± 3.50 0.70 -0.20
Placenta
5 5.07 ± 0.69 13.60 1.33 5.39 ± 0.87 16.20 7.87
10 10.14 ± 0.67 6.61 1.39 10.65 ± 0.75 7.00 6.47
100 101.80 ± 4.09 4.01 1.80 99.95 ± 3.61 3.62 -0.05
500 504.17 ± 4.70 0.93 0.83 501.25 ± 1.37 0.27 0.25
Fetus
5 5.76 ± 0.61 10.46 15.80 5.25 ± 0.15 2.86 5.00
10 10.39 ± 1.13 10.92 3.89 10.67 ± 1.16 10.87 6.67
100 103.17 ± 6.57 6.37 3.17 100.35 ± 2.23 2.22 0.35
500 504.01 ± 4.51 0.90 0.80 504.90 ± 7.51 1.49 0.98
Amniotic fluid
5 5.13 ± 0.27 5.17 2.57 5.29 ± 0.19 3.66 5.87
10 10.43 ± 0.92 8.82 4.34 10.60 ± 0.57 5.38 5.97
100 105.0 ± 2.98 2.84 5.00 98.28 ± 9.78 9.95 -1.72
500 500.98 ± 1.49 0.30 0.20 500.45 ± 1.47 0.29 0.09

Data have been expressed as mean ± SD.