Cytotoxicity of 3D‐printed, milled, and conventional oral splint resins to L929 cells and human gingival fibroblasts

Ralf Bürgers 1, Andrea Schubert 1, Jonas Müller 1, Sebastian Krohn 1, Matthias Rödiger 1, Andreas Leha 2, Torsten Wassmann 1,
PMCID: PMC9209804  PMID: 35570327

Abstract

Objectives

Evidence on the biocompatibility of three‐dimensional (3D)‐printed and milled resins for oral splints is limited. This in vitro study assessed the influence of the manufacturing method on the cytotoxicity of oral splint resins on L929 cells and human gingival fibroblasts (GF1).

Materials and Methods

Standardized specimens of four 3D‐printed, two‐milled, one‐thermoformed, and one‐pressed splint resin were incubated with L929 and GF1 cells for 24 h. Immunofluorescence and WST‐8 assay were performed to evaluate cytotoxic effects. One‐way analysis of variance and Tukey's multiple comparison test were applied with the variables “splint resin” and “manufacturing method” (p < .05).

Results

Immunofluorescence showed attachment of L929 and GF1 cells to the splint resins. Irrespective of the manufacturing method, the WST‐8 assay revealed significant differences between splint resins for the viability of L929 and GF1 cells. L929 cells generally showed lower viability rates than GF1 cells. The evaluation of cell viability by the manufacturing method showed no significant differences between 3D printing, milling, and conventional methods.

Conclusions

The cytotoxic effects of 3D‐printed, milled, and conventional oral splint resins were similar, indicating minor influence of the manufacturing method on biocompatibility. Cytotoxicity of the resins was below a critical threshold in GF1 cells. The chemical composition might be more crucial than the manufacturing method for the biocompatibility of splint resins.

Keywords: cytotoxicity, 3D printing, milling, splint resins

1. INTRODUCTION

Oral splints are widely used in dentistry for the treatment of temporomandibular disorders (Ebrahim et al., 2012; Klasser & Greene, 2009), to protect teeth from clenching or grinding (Melo et al., 2019), or for orthodontic alignment therapy (Borda et al., 2020; Weir, 2017). While acrylic resins remain the “gold standard” material for oral splints, conventional splint manufacturing methods have been increasingly replaced by computer‐aided designing and computer‐aided manufacturing (CAD/CAM) methods over the past decade (Alghazzawi, 2016; Dedem & Türp, 2016). Traditionally, oral splints have been manufactured by vacuum thermoforming or by pressing cold‐ or heat‐cured acrylic resins on plaster casts of dental arches. Digital workflows, instead, are performed by intraoral scanning of dental arches, software‐supported splint design, and computer‐aided splint manufacturing (Dedem & Türp, 2016; Salmi et al., 2013; Vandenberghe, 2020). The two common CAD/CAM techniques used to manufacture oral splints are milling and three‐dimensional (3D) printing. Milling is a subtractive process carried out with the aid of CNC (computerized numerical control) equipment, in which oral splints are carved from prefabricated resin blanks produced under high pressure (Marcel et al., 2020). 3D printing is an additive process in which liquid resin monomers are successively cured by visible light creating solid objects layer by layer (Dawood et al., 2015). CAD/CAM manufacturing is characterized by high accuracy, standardization, and reproducibility, while being time‐ and cost‐efficient (Beuer et al., 2008; Dedem & Türp, 2016; Marcel et al., 2020). Also, the mechanical properties of 3D‐printed and milled oral splints are suitable for clinical use (Berli et al., 2020; Huettig et al., 2017).

Oral splints are worn repeatedly and for several hours at a time, during which they interact with teeth, saliva, and the oral mucosa. Aside from mechanical and practical characteristics, biological aspects must be considered to fully evaluate their clinical performance. Biocompatibility is the ability of a material to perform its desired medical function without eliciting an undesirable systemic or local host response (Williams, 2008). The biocompatibility of acrylic resins has been discussed controversially in dental research (Ata & Yavuzyilmaz, 2009; Gautam et al., 2012; Goiato et al., 2015; Schweikl et al., 2006), and adverse host responses like local irritation and ulceration (Weaver & Goebel, 1980) and mucosal edema (Ruiz‐Genao et al., 2003), as well as burning mouth sensations (Ali et al., 1986), have been described in patients as a result of acrylic resin exposure.

Acrylic resins result from a polymerization process, in which monomers are converted into stable polymers via an addition reaction that is activated by heat, light, or chemical activators (Bayraktar et al., 2006; Braden, 1988). The conversion from monomers to polymers is never complete, inevitably resulting in residual monomers and other toxic chemical products such as methyl methacrylate, formaldehyde, methacrylic acid, benzoic acid, or dibutyl phthalate within the material (Gautam et al., 2012; Goiato et al., 2015; Singh et al., 2013; Wedekind et al., 2021). They leach into the surrounding saliva via diffusion and subsequently interact with the host mucosa where they may have cytotoxic effects (Kurt et al., 2018; Singh et al., 2013; Urban et al., 2009).

The extent of residual monomers is influenced by the polymerization method underlying the manufacturing process of the resin material (Bayraktar et al., 2006; Wedekind et al., 2021). Dental materials manufactured by milling from highly polymerized resin blanks show levels of residual monomers comparable to those of conventional materials (Steinmassl et al., 2017). In contrast, 3D‐printed dental devices cured by light in a stepwise process were found to contain higher levels of residual monomers (Alifui‐Segbaya et al., 2019; Wedekind et al., 2021). Therefore, it appears possible that 3D‐printed oral splint materials exert higher cytotoxicity than milled ones, but, surprisingly, scientific evidence is scarce.

The aim of the present in vitro study was to investigate the cytotoxicity of commercially available 3D‐printed, milled, and conventional resin‐based oral splint materials with particular consideration of the manufacturing method. Besides ISO‐10993‐5 conform L929 mice cells, human gingival fibroblasts (GF1) were investigated to increase the transferability of the results to human oral tissue.

2. MATERIALS AND METHODS

2.1. Specimen preparation

Cylindrical specimens with a diameter of 10 mm and a height of 2.5 mm were manufactured from four 3D‐printed, two‐milled and two conventional (one thermoformed, one pressed of cold‐cure resin) splint resins (Table 1). As previously described (Schubert et al., 2021), rods of each resin were produced according to the manufacturer's instructions and sliced into disks using a separating machine (Micracut 201; Metkon, Bursa, Turkey). Surface polishing of the specimens was performed with an automated grinding machine (Digiprep 251, Metkon) and silicon carbide grinding paper.

Table 1.

Specification of the oral splint resins assessed in this study

Manufacturing method Product Manufacturer
3D printing Med610 Stratasys, Eden Prairie, MN, USA
V‐Print splint Voco, Cuxhaven, Germany
FREEPRINT ortho 385 Detax, Ettlingen, Germany
Dental LT Clear Formlabs, Somerville, MA, USA
Milling M‐PM crystal Merz Dental, Luetjenburg, Germany
Therapon Transpa Zirkonzahn, Gais, Italy
Thermoforming Erkodur Erkodent, Pfalzgrafenweiler, Germany
Pressing PalaXpress ultra Kulzer, Hanau, Germany

2.2. Cell culture

L929 cells (lot. no. 85011425; Sigma‐Aldrich, St. Louis, MO, USA) complying with the ISO 10993‐5 standard for cytotoxicity testing were cultured in Dulbecco's modified Eagle's medium (Thermo Fisher Scientific, Waltham, MA, USA) with 10% fetal calf serum (Invitrogen, Darmstadt, Germany) and 1% penicillin–streptomycin (Fisher Scientific, Schwerte, Germany) under standard conditions. Immortalized human gingival fibroblasts (GF1) were established by the group for Oral Biology and Tissue Regeneration in our Department (ethic vote no. 16/6/2009) and have been published previously (Schubert et al., 2019).

2.3. Absorption assay

Cell viability of L929 and GF1 cells was assessed using WST‐8‐based assays (Cell Counting Kit 8 [CCK‐8]; Dojindo Molecular Technologies, Kumamoto, Japan). Under sterile conditions, specimens of the test resins (n = 25) were fixed to well bottoms of 24‐well plates using silicone. A total of 4 × 104 L929 or GF1 cells (passages 3–8), respectively, in 1 ml culture medium were added to each well. Adherence was verified via light microscopy. After 24 h of incubation, the medium was removed from the wells and replaced by 600 µl of CCK‐8 detection solution at a dilution of 1:10 in the culture medium. After 4 h of incubation, the supernatant was transferred to another 24‐well plate without test specimens. Absorption was measured using a plate reader (Fluostar Optima; BMG Labtech, Ortenberg, Germany) at 450 nm and a reference wavelength of 650 nm. Standard glass specimens served as controls.

2.4. Immunofluorescence

Exemplary fluorescence imaging of L929 and GF1 cells on the surface of the resins was performed via immunocytochemistry. Resin specimens were incubated with 2 × 104 of the respective cells for 24 h. After washing twice with phosphate‐buffered saline (PBS Tablets, Calbiochem, Merck, Darmstadt, Germany), cells were fixated using 2% paraformaldehyde solution for 15 min, treated with 0.25% Triton X‐100 (T8787; Sigma‐Aldrich) in PBS for 10 min, and blocked in 1% bovine serum albumin (BSA; Sigma‐Aldrich) in PBS for 15 min. A mouse monoclonal anti‐vinculin antibody (V9131; Sigma‐Aldrich) at a dilution of 1:1000 in 1% BSA in PBS was applied for 60 min at 37°C, followed by a polyclonal goat anti‐mouse secondary antibody to IgG (AlexaFluor 555, ab150114; Abcam, Cambridge, UK) and 4′,6‐diamidino‐2‐phenylindole (DAPI, 71‐03‐00; Kirkegaard & Perry Laboratories, Gaithersburg, MD, USA) at a dilution of 1:1000 in 1% BSA in PBS for 30 min at 37°C. Alexa Fluor 488 Phalloidin (A12379; Thermo Fisher Scientific) was incubated at a dilution of 1:400 in 1% BSA in PBS for 30 min at room temperature. Thorough washing with PBS was performed between all of the above steps. Specimens were dried and mounted on object slides for visualization via fluorescence microscopy (BZ‐X710; Keyence, Osaka, Japan). To reduce the autofluorescence of the test materials, haze reduction was applied to the images.

2.5. Statistical analysis

For the analysis of cell viability, values were tested for normal distribution using Q–Q plots. One‐way ANOVA was performed for the variables “splint product” and “manufacturing method.” Post hoc testing was performed via Tukey's multiple comparison test.

Statistical analyses were executed using GraphPad Prism 9 (GraphPad Software, San Diego, CA, USA). The overall level of significance was set at α = .05. Data from the WST‐8 assay are shown as medians with box–whisker plots.

3. RESULTS

Fluorescence imaging revealed the attachment of L929 and GF1 cells to all of the investigated splint resins after 24 h of incubation (Figure 1). Visualization of the actin filaments of the cytoskeleton via vinculin and phalloidin showed typical morphological characteristics of vital fibroblasts in monolayer culture. L929 cells were mostly round in shape, while GF1 cell bodies had a stretched out morphology with cell protrusions. Cell nuclei were visualized via DAPI.

Figure 1.

Figure 1

Fluorescence imaging of L929 and GF1 cells on the surface of dental splint resins after 24 h of incubation. Exemplary visualization shows adhesion of the cells to all of the tested resins. Morphology of GF1 cells includes more distinctive cell protrusions than L929 cells, which are generally rounder in shape. DAPI (blue) indicates cell nuclei, phalloidin (green), and vinculin (red) are associated with actin filaments of the cytoskeleton. DAPI, 4′,6‐diamidino‐2‐phenylindole.

Cell viability was quantified according to the relative absorption of a WST‐8‐based assay. For L929 cells (Figure 2a), the highest cell viability was found for Therapon Transpa (milling, arithmetic mean 0.960, standard deviation ± 0.236), and it was significantly higher (p < .05) than for FREEPRINT ortho 385 (3D printing, 0.724 ± 0.207) and Erkodur (thermoforming, 0.705 ± 0.216). Overall, Erkodur was associated with the lowest L929 viability. There were no significant differences within the 3D‐printed or milled resins. To investigate the influence of the manufacturing method on cell viability, splint resins were grouped according to the manufacturing method, and there were no significant differences between 3D printing, milling, thermoforming, and pressing (Figure 2b).

Figure 2.

Figure 2

Relative cell viability of L929 cells according to the Cell Counting Kit ‐8 assay after 24 h of incubation with the investigated oral splint resins. (a) Cell viability shows significant differences between resins indicated by asterisks (p < .05). (b) Arrangement of the data according to the manufacturing method shows no significant differences. Glass was used for normalization (=1.0).

For GF1 cells (Figure 3a), the highest (M‐PM crystal, 1.236 ± 0.312) and the lowest (Therapon Transpa, 0.813 ± 0.160) relative cell viability were found within the group of milled resins, and their difference was highly significant (p < .0001). No significant differences were revealed within the 3D‐printed resins. Erkodur (thermoforming, 0.922 ± 0.323) was associated with significantly lower GF1 cell viability than M‐PM crystal (milling) (p < .01). PalaXpress ultra (pressing, 1.201 ± 0.261) was related to significantly higher cell viability than Dental LT clear (3D printing, 0.921 ± 0.359, p < .01), Therapon Transpa (milling, p < .001), and Erkodur (thermoforming, p < .01). Arrangement of the results according to the manufacturing method indicated significantly higher (p < .05) cell viability of GF1 cells for pressing than for thermoforming (Figure 3b). 3D printing and milling showed no significant differences compared with conventional methods.

Figure 3.

Figure 3

Relative cell viability of GF1 cells according to the CCK‐8 assay after 24 h of incubation with the investigated oral splint resins. (a) Cell viability shows significant differences between resins. a p < .05 compared with Med610, b p < .01 compared with V‐print splint, c p < .05 compared with FREEPRINT ortho 385, d p < .01 compared with Dental LT clear, d p < .05 compared with Dental LT clear, e p < .0001 compared with M‐PM crystal, e p < .01 compared with M‐PM crystal, f p < .001 compared with Therapon Transpa, and g p < .05 compared with Erkodur. (b) Arrangement of the data according to the manufacturing method shows significant differences (*p < .05) between thermoforming and pressing. Glass was used for normalization (=1.0).

4. DISCUSSION

In the present study, the cytotoxic effect of 3D‐printed, milled, thermoformed, and pressed oral splint resins on L929 and GF1 cells was investigated. It has been suggested that 3D‐printed resins may be less biocompatible than milled or conventional resins due to higher amounts of residual monomers (Alifui‐Segbaya et al., 2019; Wedekind et al., 2021), but there is little evidence to support this. Therefore, the influence of the manufacturing method on biocompatibility of splint resins was evaluated for the first time.

In accordance with the current ISO 10993‐5 standard for cytotoxicity testing of medical devices, murine L929 cells were investigated in a monolayer cell culture model (Eljezi et al., 2017; Schmalz & Galler, 2017; Schubert et al., 2019; Tsuchiya et al., 1994; Wataha, 2012). In vivo, oral splints interact with the oral mucosa, which is composed of keratinocytes and underlying fibroblasts. Therefore, to simulate more tissue‐like conditions, we additionally assessed GF1 cells, a cell line of human gingival fibroblasts previously used to evaluate cytotoxic effects of composite resins (Schubert et al., 2019).

Fluorescence staining showed the attachment of L929 and GF1 cells to all of the investigated splint resins after 24 h. The actin cytoskeleton was visualized by labelling with phalloidin, a fluorescent toxin interacting with F‐actin filaments without affecting cell viability (Adtani et al., 2019; Barak et al., 1981; Chazotte, 2010; Wulf et al., 1979), and an antibody against vinculin, a small protein binding F‐actin in focal adhesions (Golji & Mofrad, 2013; Thievessen et al., 2013; Wilkins & Lin, 1982). Both cell lines showed the typical spindle morphology of fibroblasts in monolayer culture with L929 cells being smaller and slightly rounder in shape. Cell nuclei of L929 and GF1 cells were indicated by DAPI staining and showed full integrity. These findings suggest that none of the investigated resins exerted cytotoxic effects to an extent that inhibited cell attachment or affected morphology.

For the quantification of cytotoxic effects, CCK‐8 assay was performed. It contains WST‐8, a water‐soluble tetrazolium salt, which, in viable cells, is reduced to a colored formazan dye by dehydrogenase activity (Ishiyama et al., 1997). It represents a well‐established and reproducible method for the quantification of cell viability and cell proliferation (Camassa et al., 2021; Schubert et al., 2019; Zhang et al., 2019; Zhu et al., 2019), and it is more sensitive than other colorimetric assays (Failli et al., 2013; Miyamoto et al., 2002).

The cytotoxicity of resin products is influenced by the amount of leachable substances, especially residual monomers, that interact with the surrounding tissues. Residual monomers result from incomplete polymerization, and the degree of polymerization is affected by the method used to manufacture the resin (Bayraktar et al., 2006; Bural et al., 2011; Gautam et al., 2012; Schmalz & Galler, 2017). Layer‐by‐layer polymerization of resins by 3D printing has been associated with high rates of residual monomers, while milling from highly prepolymerized resin blanks has been related to low monomer levels (Alifui‐Segbaya et al., 2019; Steinmassl et al., 2017). Hence, it seemed possible that residual monomers leached from the 3D‐printed resins into the surrounding cell culture medium and affected biocompatibility in the present study. However, the biocompatibility of 3D‐printed and milled resins was similar and comparable to conventional resins. It is noteworthy that within the 3D printing group, the printing technology also had no decisive effect on biocompatibility: several technologies with importance to clinical dentistry were included in the present investigation, namely, poly jet modelling (Med610), digital light processing (V‐Print Splint, FREEPRINT ortho 385) and stereolithography (Dental LT clear) (Alghazzawi, 2016; Dawood et al., 2015), but showed no significant differences in cytotoxicity.

GF1 cell viability rates (arithmetic means) for all splint resins were greater than 75% compared to the control, and, therefore, their cytotoxic effect can be considered minor. Bural et al. (2011) L929 cells, however, showed a viability of greater than 50%, but less than 75% for FREEPRINT ortho 385 (3D printing), Dental LT clear (3D printing), M‐PM crystal (milling), and Erkodur (thermoforming), indicating slight cytotoxicity of these resins according to ISO 10993‐5.

Wedekind et al. (2021) showed that residual monomers and additives eluted from 3D‐printed, milled, and conventional splint resins had a substantial cytotoxic effect on human gingival fibroblasts. However, these data were obtained for a worst‐case scenario in terms of splint size, so comparability with our findings is limited.

Due to a lack of further evidence on oral splint resins, we related our results to studies on the biocompatibility of denture base resins as they are chemically similar. Srinivasan et al. (2018) showed that milled resins were equally biocompatible as conventional heat‐cured resins in an in vitro setting with human primary osteoblasts and mouse embryonic fibroblasts. In another study, cytotoxic effects of 3D‐printed denture base resins on L929 cells were negligible (Tzeng et al., 2021). In line with these findings, our data indicate that milling and 3D printing did not have a significant influence on the biocompatibility of splint resins. Instead, it can be assumed that the individual chemical composition of the studied splint resins was more relevant for cytotoxicity than the manufacturing method. This was evident in the milling group, where Therapon Transpa was significantly less cytotoxic to GF1 cells than M‐PM crystal. Dental resins differ by the type of monomers, the initiators used to start the polymerization reaction, and the additives incorporated to modify the mechanical properties, all of which may affect biocompatibility (De Matteis et al., 2019; Goiato et al., 2015; Jiao et al., 2015; Lee et al., 2020). Elution of unreacted chemicals was outside the scope of this study, but future investigations may help to understand the differences in resin toxicity that have been demonstrated.

L929 and GF1 cells responded differently to exposure to oral splint resins, with GF1 cells tending to exhibit higher relative cell viability. Several previous studies have shown that cells from different mammalian species and even between different donors within one species have varying sensitivities to eluted substances from dental resins (Engelmann et al., 2002; Schubert et al., 2019; Susila & Balasubramanian, 2016). Standardization of biocompatibility testing is therefore useful to improve the comparability of investigations, but the significance of results obtained from evaluations in murine L929 cells is limited because of their poor transferability to human cells. In addition, established in vitro cytotoxicity assays performed in monolayer cell cultures do not fully reflect the complex biological reactions in the oral mucosa (Schmalz & Galler, 2017; Wataha, 2012). Due to these limitations of the present study, our data must be considered preliminary. The need for improved biocompatibility testing could be addressed by using tissue‐engineered human oral mucosa with fibroblasts and keratinocytes in multilayer cell culture models that simulate organotypic conditions (Buskermolen et al., 2016; Izumi et al., 2000; Kinikoglu et al., 2009; Masuda, 1996; Moharamzadeh et al., 2008). In the long term, in vivo investigations will be required for final data validation.

Within the limitations of the present in vitro investigation, we conclude that the cytotoxic effects of 3D‐printed, milled, and conventional oral splint resins are comparable indicating minor effects of the manufacturing method on biocompatibility. In human gingival fibroblasts, cytotoxicity of all resins was below a critical threshold. Nonetheless, splint resins should be selected carefully for clinical application as individual biocompatibility may vary significantly, possibly due to different chemical compositions. Further in vitro and in vivo investigations are needed to validate our findings.

AUTHOR CONTRIBUTIONS

Ralf Bürgers: Conceptualization; resources; writing—original draft; project administration. Andrea Schubert: Conceptualization; methodology; writing—original draft; visualization. Jonas Müller: Formal analysis; investigation. Sebastian Krohn: Visualization; writing—review and editing. Matthias Rödiger: Validation; writing—review and editing. Andreas Leha: Validation; formal analysis. Torsten Wassmann: Conceptualization; methodology; resources; writing—review and editing; supervision; project administration.

CONFLICT OF INTEREST

The authors declare no conflict of interest.

ACKNOWLEDGMENTS

The authors would like to thank Dental Design Schnellbächer & Roland GmbH & Co. KG and Voco GmbH for providing technical support. In addition, the authors would like to thank Andreas Leha for his assistance with the statistical analysis. Open access funding enabled and organized by Projekt DEAL.

Bürgers, R. , Schubert, A. , Müller, J. , Krohn, S. , Rödiger, M. , Leha, A. , & Wassmann, T. (2022). Cytotoxicity of 3D‐printed, milled, and conventional oral splint resins to L929 cells and human gingival fibroblasts. Clinical and Experimental Dental Research, 8, 650–657. 10.1002/cre2.592

Ralf Bürgers and Andrea Schubert contributed equally to this study.

DATA AVAILABILITY STATEMENT

Data are available on reasonable request from the authors.

REFERENCES

  1. Adtani, P. , Narasimhan, M. , Ranganathan, K. , Punnoose, A. , Prasad, P. , & Natarajan, P. M. (2019, August). Characterization of oral fibroblasts: An in vitro model for oral fibrosis. Journal of Oral and Maxillofacial Pathology, 23(2), 198–202. [DOI] [PMC free article] [PubMed] [Google Scholar]
  2. Alghazzawi, T. F. (2016, April 1). Advancements in CAD/CAM technology: Options for practical implementation. Journal of Prosthodontic Research, 60(2), 72–84. [DOI] [PubMed] [Google Scholar]
  3. Ali, A. , Bates, J. F. , Reynolds, A. J. , & Walker, D. M. (1986, December 20). The burning mouth sensation related to the wearing of acrylic dentures: An investigation. British Dental Journal, 161(12), 444–447. [DOI] [PubMed] [Google Scholar]
  4. Alifui‐Segbaya, F. , Bowman, J. , White, A. R. , George, R. , & Fidan, I. (2019, December). Characterization of the double bond conversion of acrylic resins for 3D printing of dental prostheses. Compendium of Continuing Education in Dentistry, 40(10), e7–e11. [PubMed] [Google Scholar]
  5. Ata, S. O. , & Yavuzyilmaz, H. (2009, November). In vitro comparison of the cytotoxicity of acetal resin, heat‐polymerized resin, and auto‐polymerized resin as denture base materials. Journal of Biomedical Materials Research, Part B: Applied Biomaterials, 91(2), 905–909. [DOI] [PubMed] [Google Scholar]
  6. Barak, L. S. , Yocum, R. R. , & Webb, W. W. (1981, May). In vivo staining of cytoskeletal actin by autointernalization of nontoxic concentrations of nitrobenzoxadiazole‐phallacidin. Journal of Cell Biology, 89(2), 368–372. [DOI] [PMC free article] [PubMed] [Google Scholar]
  7. Bayraktar, G. , Guvener, B. , Bural, C. , & Uresin, Y. (2006). Influence of polymerization method, curing process, and length of time of storage in water on the residual methyl methacrylate content in dental acrylic resins. Journal of Biomedical Materials Research. Part B: Applied Biomaterials, 76B(2), 340–345. [DOI] [PubMed] [Google Scholar]
  8. Berli, C. , Thieringer, F. M. , Sharma, N. , Müller, J. A. , Dedem, P. , Fischer, J. , & Rohr, N. (2020, January 16). Comparing the mechanical properties of pressed, milled, and 3D‐printed resins for occlusal devices. Journal of Prosthetic Dentistry, 124, 780–786. [DOI] [PubMed] [Google Scholar]
  9. Beuer, F. , Schweiger, J. , & Edelhoff, D. (2008, May 10). Digital dentistry: An overview of recent developments for CAD/CAM generated restorations. British Dental Journal, 204(9), 505–511. [DOI] [PubMed] [Google Scholar]
  10. Borda, A. F. , Garfinkle, J. S. , Covell, D. A. , Wang, M. , Doyle, L. , & Sedgley, C. M. (2020, July 1). Outcome assessment of orthodontic clear aligner vs fixed appliance treatment in a teenage population with mild malocclusions. Angle Orthodontist, 90(4), 485–490. [DOI] [PMC free article] [PubMed] [Google Scholar]
  11. Braden, M. (1988, August). Some aspects of the chemistry and physics of dental resins. Advances in Dental Research, 2(1), 93–97. [DOI] [PubMed] [Google Scholar]
  12. Bural, C. , Aktaş, E. , Deniz, G. , Ünlüçerçi, Y. , & Bayraktar, G. (2011, August). Effect of leaching residual methyl methacrylate concentrations on in vitro cytotoxicity of heat polymerized denture base acrylic resin processed with different polymerization cycles. Journal of Applied Oral Science, 19((4), 306–312. [DOI] [PMC free article] [PubMed] [Google Scholar]
  13. Buskermolen, J. K. , Reijnders, C. M. A. , Spiekstra, S. W. , Steinberg, T. , Kleverlaan, C. J. , Feilzer, A. J. , Bakker, A. D. , & Gibbs, S. (2016). Development of a full‐thickness human gingiva equivalent constructed from immortalized keratinocytes and fibroblasts. Tissue Engineering. Part C: Methods, 22(8), 781–791. [DOI] [PMC free article] [PubMed] [Google Scholar]
  14. Camassa, L. M. A. , Ervik, T. K. , Zegeye, F. D. , Mdala, I. , Valen, H. , Ansteinsson, V. , & Zienolddiny, S. (2021, July). Characterization and toxicity evaluation of air‐borne particles released by grinding from two dental resin composites in vitro. Dental Materials, 37(7), 1121–1133. [DOI] [PubMed] [Google Scholar]
  15. Chazotte, B. (2010, May). Labeling cytoskeletal F‐actin with rhodamine phalloidin or fluorescein phalloidin for imaging. Cold Spring Harbor Protocols, 2010(5), 4947. [DOI] [PubMed] [Google Scholar]
  16. Dawood, A. , Marti Marti, B. , Sauret‐Jackson, V. , & Darwood, A. (2015, December). 3D printing in dentistry. British Dental Journal, 219(11), 521–529. [DOI] [PubMed] [Google Scholar]
  17. Dedem, P. , & Türp, J. C. (2016). Digital Michigan splint—From intraoral scanning to plasterless manufacturing. International Journal of Computerized Dentistry, 19(1), 63–76. [PubMed] [Google Scholar]
  18. De Matteis, V. , Cascione, M. , Toma, C. C. , Albanese, G. , De Giorgi, M. L. , Corsalini, M. , & Rinaldi, R. (2019). Silver nanoparticles addition in poly(methyl methacrylate) dental matrix: Topographic and antimycotic studies. International Journal of Molecular Sciences, 20(19):4691. [DOI] [PMC free article] [PubMed] [Google Scholar]
  19. Ebrahim, S. , Montoya, L. , Busse, J. W. , Carrasco‐Labra, A. , & Guyatt, G. H. , Medically Unexplained Syndromes Research Group . (2012, August). The effectiveness of splint therapy in patients with temporomandibular disorders: A systematic review and meta‐analysis. Journal of the American Dental Association, 143(8), 847–857. [DOI] [PubMed] [Google Scholar]
  20. Eljezi, T. , Pinta, P. , Richard, D. , Pinguet, J. , Chezal, J.‐M. , Chagnon, M.‐C. , Sautou, V. , Grimandi, G. , & Moreau, E. (2017, April). In vitro cytotoxic effects of DEHP‐alternative plasticizers and their primary metabolites on a L929 cell line. Chemosphere, 173, 452–459. [DOI] [PubMed] [Google Scholar]
  21. Engelmann, J. , Leyhausen, G. , Leibfritz, D. , & Geurtsen, W. (2002). Effect of TEGDMA on the intracellular glutathione concentration of human gingival fibroblasts. Journal of Biomedical Materials Research, 63(6), 746–751. [DOI] [PubMed] [Google Scholar]
  22. Failli, A. , Legitimo, A. , Orsini, G. , Castagna, M. , Spisni, R. , Miccoli, P. , & Consolini, R. (2013, March). Antiproliferative effects of 5‐fluorouracil and oxaliplatin in colon cancer cell lines: Comparison of three different cytotoxicity assays. Journal of Biological Regulators and Homeostatic Agents, 27(1), 275–284. [PubMed] [Google Scholar]
  23. Gautam, R. , Singh, R. D. , Sharma, V. P. , Siddhartha, R. , Chand, P. , & Kumar, R. (2012, July). Biocompatibility of polymethylmethacrylate resins used in dentistry. Journal of Biomedical Materials Research. Part B: Applied Biomaterials, 100(5), 1444–1450. [DOI] [PubMed] [Google Scholar]
  24. Goiato, M. C. , Freitas, E. , dos Santos, D. , de Medeiros, R. , & Sonego, M. (2015). Acrylic resin cytotoxicity for denture base—literature review. Advances in Clinical and Experimental Medicine, 24(4), 679–686. [DOI] [PubMed] [Google Scholar]
  25. Golji, J. , & Mofrad, M. R. K. (2013, April). The interaction of vinculin with actin. PLoS Computational Biology, 9(4), e1002995. [DOI] [PMC free article] [PubMed] [Google Scholar]
  26. Huettig, F. , Kustermann, A. , Kuscu, E. , Geis‐Gerstorfer, J. , & Spintzyk, S. (2017). Polishability and wear resistance of splint material for oral appliances produced with conventional, subtractive, and additive manufacturing. Journal of the Mechanical Behavior of Biomedical Materials, 75, 175–179. [DOI] [PubMed] [Google Scholar]
  27. Ishiyama, M. , Miyazono, Y. , Sasamoto, K. , Ohkura, Y. , & Ueno, K. (1997, July). A highly water‐soluble disulfonated tetrazolium salt as a chromogenic indicator for NADH as well as cell viability. Talanta, 44(7), 1299–1305. [DOI] [PubMed] [Google Scholar]
  28. Izumi, K. , Terashi, H. , Marcelo, C. L. , & Feinberg, S. E. (2000, March). Development and characterization of a tissue‐engineered human oral mucosa equivalent produced in a serum‐free culture system. Journal of Dental Research, 79(3), 798–805. [DOI] [PubMed] [Google Scholar]
  29. Jiao, Y. , Ma, S. , Li, J. , Shan, L. , Yang, Y. , Li, M. , & Chen, J. (2015). The influences of N‐acetyl cysteine (NAC) on the cytotoxicity and mechanical properties of poly‐methylmethacrylate (PMMA)‐based dental resin. PeerJ, 3, e868. [DOI] [PMC free article] [PubMed] [Google Scholar]
  30. Kinikoglu, B. , Auxenfans, C. , Pierrillas, P. , Justin, V. , Breton, P. , Burillon, C. , Hasirci, V. , & Damour, O. (2009, November). Reconstruction of a full‐thickness collagen‐based human oral mucosal equivalent. Biomaterials, 30(32), 6418–6425. [DOI] [PubMed] [Google Scholar]
  31. Klasser, G. D. , & Greene, C. S. (2009, February). Oral appliances in the management of temporomandibular disorders. Oral Surgery, Oral Medicine, Oral Pathology, Oral Radiology, and Endodontology, 107(2), 212–223. [DOI] [PubMed] [Google Scholar]
  32. Kurt, A. , Altintas, S. H. , Kiziltas, M. V. , Tekkeli, S. E. , Guler, E. M. , Kocyigit, A. , & Usumez, A. (2018, January 30). Evaluation of residual monomer release and toxicity of self‐adhesive resin cements. Dental Materials Journal, 37(1), 40–48. [DOI] [PubMed] [Google Scholar]
  33. Lee, M.‐J. , Kim, M.‐J. , Oh, S.‐H. , & Kwon, J.‐S. (2020). Novel dental poly (methyl methacrylate) containing phytoncide for antifungal effect and inhibition of oral multispecies biofilm. Materials (Basel), 13(2):371. [DOI] [PMC free article] [PubMed] [Google Scholar]
  34. Marcel, R. , Reinhard, H. , & Andreas, K. (2020, December). Accuracy of CAD/CAM‐fabricated bite splints: milling vs 3D printing. Clinical Oral Investigations, 24(12), 4607–4615. [DOI] [PMC free article] [PubMed] [Google Scholar]
  35. Masuda, I. (1996, June). An in vitro oral mucosal model reconstructed from human normal gingival cells. Kokubyo Gakkai Zasshi [The Journal of the Stomatological Society, Japan], 63(2), 334–353. [DOI] [PubMed] [Google Scholar]
  36. Melo, G. , Duarte, J. , Pauletto, P. , Porporatti, A. L. , Stuginski‐Barbosa, J. , Winocur, E. , Flores‐Mir, C. , & De Luca Canto, G. (2019, July). Bruxism: An umbrella review of systematic reviews. Journal of Oral Rehabilitation, 46(7), 666–690. [DOI] [PubMed] [Google Scholar]
  37. Miyamoto, T. , Min, W. , & Lillehoj, H. S. (2002, March). Lymphocyte proliferation response during Eimeria tenella infection assessed by a new, reliable, nonradioactive colorimetric assay. Avian Diseases, 46(1), 10–16. [DOI] [PubMed] [Google Scholar]
  38. Moharamzadeh, K. , Brook, I. M. , Van Noort, R. , Scutt, A. M. , Smith, K. G. , & Thornhill, M. H. (2008). Development, optimization and characterization of a full‐thickness tissue engineered human oral mucosal model for biological assessment of dental biomaterials. Journal of Materials Science: Materials in Medicine, 19(4), 1793–1801. [DOI] [PubMed] [Google Scholar]
  39. Ruiz‐Genao, D. P. , Moreno de Vega, M. J. , Sánchez Pérez, J. , & García‐Díez, A. (2003, May). Labial edema due to an acrylic dental prosthesis. Contact Dermatitis, 48(5), 273–274. [DOI] [PubMed] [Google Scholar]
  40. Salmi, M. , Paloheimo, K.‐S. , Tuomi, J. , Ingman, T. , & Mäkitie, A. (2013, January 13). A digital process for additive manufacturing of occlusal splints: A clinical pilot study. Journal of the Royal Society, Interface, 10(84), 20130203. [DOI] [PMC free article] [PubMed] [Google Scholar]
  41. Schmalz, G. , & Galler, K. M. (2017). Biocompatibility of biomaterials—Lessons learned and considerations for the design of novel materials. Dental Materials, 33(4), 382–393. [DOI] [PubMed] [Google Scholar]
  42. Schubert, A. , Bürgers, R. , Baum, F. , Kurbad, O. , & Wassmann, T. (2021). Influence of the manufacturing method on the adhesion of Candida albicans and Streptococcus mutans to oral splint resins. Polymers, 13(10):1534. [DOI] [PMC free article] [PubMed] [Google Scholar]
  43. Schubert, A. , Ziegler, C. , Bernhard, A. , Bürgers, R. , & Miosge, N. (2019, January 1). Cytotoxic effects to mouse and human gingival fibroblasts of a nanohybrid ormocer versus dimethacrylate‐based composites. Clinical Oral Investigations, 23(1), 133–139. [DOI] [PubMed] [Google Scholar]
  44. Schweikl, H. , Spagnuolo, G. , & Schmalz, G. (2006, October). Genetic and cellular toxicology of dental resin monomers. Journal of Dental Research, 85(10), 870–877. [DOI] [PubMed] [Google Scholar]
  45. Singh, R. D. , Gautam, R. , Siddhartha, R. , Singh, B. P. , Chand, P. , Sharma, V. P. , & Jurel, S. K. (2013). High performance liquid chromatographic determination of residual monomer released from heat‐cured acrylic resin. An in vivo study. Journal of Prosthodontics, 22(5), 358–361. [DOI] [PubMed] [Google Scholar]
  46. Srinivasan, M. , Gjengedal, H. , Cattani‐Lorente, M. , Moussa, M. , Durual, S. , Schimmel, M. , & Müller, F. (2018, July 29). CAD/CAM milled complete removable dental prostheses: An in vitro evaluation of biocompatibility, mechanical properties, and surface roughness. Dental Materials Journal, 37(4), 526–533. [DOI] [PubMed] [Google Scholar]
  47. Steinmassl, P.‐A. , Wiedemair, V. , Huck, C. , Klaunzer, F. , Steinmassl, O. , Grunert, I. , & Dumfahrt, H. (2017, June). Do CAD/CAM dentures really release less monomer than conventional dentures? Clinical Oral Investigations, 21(5), 1697–1705. [DOI] [PMC free article] [PubMed] [Google Scholar]
  48. Susila, A. V. , & Balasubramanian, V. (2016, March). Correlation of elution and sensitivity of cell lines to dental composites. Dental Materials, 32(3), e63–e72. [DOI] [PubMed] [Google Scholar]
  49. Thievessen, I. , Thompson, P. M. , Berlemont, S. , Plevock, K. M. , Plotnikov, S. V. , Zemljic‐Harpf, A. , Ross, R. S. , Davidson, M. W. , Danuser, G. , Campbell, S. L. , & Waterman, C. M. (2013). Vinculin–actin interaction couples actin retrograde flow to focal adhesions, but is dispensable for focal adhesion growth. Journal of Cell Biology, 202(1), 163–177. [DOI] [PMC free article] [PubMed] [Google Scholar]
  50. Tsuchiya, H. , Hoshino, Y. , Tajima, K. , & Takagi, N. (1994, June 1). Leaching and cytotoxicity of formaldehyde and methyl methacrylate from acrylic resin denture base materials. Journal of Prosthetic Dentistry, 71(6), 618–624. [DOI] [PubMed] [Google Scholar]
  51. Tzeng, J.‐J. , Yang, T.‐S. , Lee, W.‐F. , Chen, H. , & Chang, H.‐M. (2021, March 8). Mechanical properties and biocompatibility of urethane acrylate‐based 3D‐printed denture base resin. Polymers (Basel), 13(5), 822. [DOI] [PMC free article] [PubMed] [Google Scholar]
  52. Urban, V. M. , Machado, A. L. , Vergani, C. E. , Giampaolo, E. T. , Pavarina, A. C. , de Almeida, F. G. , & Cass, Q. B. (2009, May). Effect of water‐bath post‐polymerization on the mechanical properties, degree of conversion, and leaching of residual compounds of hard chairside reline resins. Dental Materials, 25(5), 662–671. [DOI] [PubMed] [Google Scholar]
  53. Vandenberghe, B. (2020, May). The crucial role of imaging in digital dentistry. Dental Materials, 36(5), 581–591. [DOI] [PubMed] [Google Scholar]
  54. Wataha, J. C. (2012, January 1). Predicting clinical biological responses to dental materials. Dental Materials, 28(1), 23–40. [DOI] [PubMed] [Google Scholar]
  55. Weaver, R. E. , & Goebel, W. M. (1980, February). Reactions to acrylic resin dental prostheses. Journal of Prosthetic Dentistry, 43(2), 138–142. [DOI] [PubMed] [Google Scholar]
  56. Wedekind, L. , Güth, J.‐F. , Schweiger, J. , Kollmuss, M. , Reichl, F.‐X. , Edelhoff, D. , & Högg, C. (2021, February 26). Elution behavior of a 3D‐printed, milled and conventional resin‐based occlusal splint material. Dental Materials, 37, 701–710. [DOI] [PubMed] [Google Scholar]
  57. Weir, T. (2017, March). Clear aligners in orthodontic treatment. Australian Dental Journal, 62(Suppl 1), 58–62. [DOI] [PubMed] [Google Scholar]
  58. Wilkins, J. A. , & Lin, S. (1982, January). High‐affinity interaction of vinculin with actin filaments in vitro. Cell, 28(1), 83–90. [DOI] [PubMed] [Google Scholar]
  59. Williams, D. F. (2008, July). On the mechanisms of biocompatibility. Biomaterials, 29(20), 2941–2953. [DOI] [PubMed] [Google Scholar]
  60. Wulf, E. , Deboben, A. , Bautz, F. A. , Faulstich, H. , & Wieland, T. (1979, September). Fluorescent phallotoxin, a tool for the visualization of cellular actin. Proceedings of the National Academy of Sciences of the United States of America, 76(9), 4498–4502. [DOI] [PMC free article] [PubMed] [Google Scholar]
  61. Zhang, C.‐Y. , Cheng, Y.‐L. , Tong, X.‐W. , Yu, H. , & Cheng, H. (2019). In vitro cytotoxicity of self‐adhesive dual‐cured resin cement polymerized beneath three different cusp inclinations of zirconia. BioMed Research International, 2019, 7404038. [DOI] [PMC free article] [PubMed] [Google Scholar]
  62. Zhu, N. , Chatzistavrou, X. , Ge, L. , Qin, M. , Papagerakis, P. , & Wang, Y. (2019, April). Biological properties of modified bioactive glass on dental pulp cells. Journal of Dentistry, 83, 18–26. [DOI] [PubMed] [Google Scholar]

Associated Data

This section collects any data citations, data availability statements, or supplementary materials included in this article.

Data Availability Statement

Data are available on reasonable request from the authors.


Articles from Clinical and Experimental Dental Research are provided here courtesy of Wiley

RESOURCES