Abstract
There is limited evidence regarding the utility of fetal monitoring during pregnancy, particularly during labor and delivery. Developed countries rely on consensus ‘best practices’ of obstetrics and gynecology professional societies to guide their protocols and policies. Protocols are often driven by the desire to be as safe as possible and avoid litigation, regardless of the cost of downstream treatment. In high-resource settings, there may be a justification for this approach. In low-resource settings, in particular, interventions can be costly and lead to adverse outcomes in subsequent pregnancies. Therefore, it is essential to consider the evidence and cost of different fetal monitoring approaches, particularly in the context of treatment and care in low-to-middle income countries.
This article reviews the standard methods used for fetal monitoring, with particular emphasis on fetal cardiac assessment, which is a reliable indicator of fetal well-being. An overview of fetal monitoring practices in low-to-middle income counties, including perinatal care access challenges, is also presented. Finally, an overview of how mobile technology may help reduce barriers to perinatal care access in low-resource settings is provided.
Keywords: fetal monitoring, low-and middle-income country, fetal cardiac assessment, mobile applications, perinatal care
1. Introduction
Perinatal complications account for 40% of the perinatal and maternal deaths worldwide (World Health Organization 2016a). Low-and middle-income countries (LMICs) contribute approximately 90% of total births, and 98% of the total perinatal deaths (World Health Organization 1996, Save the Children 2001, Blencowe et al 2016, Wang et al 2016).
The perinatal mortality rate is defined as the sum of the number of stillbirths and deaths occurring during the first seven days of life, per 1000 live births. In 2018, this rate stood at 19 per 1000 in LMICs, whereas in upper-middle and high-income countries, there was an average of seven and three deaths per 1000 live births, respectively (UNICEF et al 2018). The highest perinatal mortality rates have been reported for countries in Sub-Saharan Africa and South-Asia (28% and 26%, respectively) (UNICEF et al 2018) and may be underreported (Lopez et al 2007, Pattinson et al 2009). At the beginning of the twentieth century, the perinatal mortality rate in high-income countries (HIC) was as alarmingly high as it currently is in LMICs, but was effectively reduced by the expansion of antenatal care coverage, extended indications for Cesarean sections, and the introduction of perinatal screening technologies (cardiotocography (CTG), ultrasound, amnioscopy, amniocentesis, and pH-meter) (Dražančić 2001, Lawn et al 2009, Flenady et al 2011, Goldenberg et al 2016).
The most common causes of perinatal deaths are preterm birth-related complications (35%), intrapartum-related events (24%), and sepsis (15%) (UNICEF et al 2019). Studies conducted in LMICs have reported significant issues with prematurity, birth asphyxia, maternal hypertensive disorders, and septicemia being the most common causes of perinatal death (Allanson et al 2015, Mahdizadeh et al 2019). Fetuses and newborns are also disproportionately affected by infections, including syphilis, malaria, and animal and vector-borne diseases, leading to elevated mortality and morbidity (Han et al 2010, Goldenberg et al 2010).
Asphyxia, one of the most common causes of death during childbirth (Goldenberg et al 2007, Lawn et al 2009, Wall et al 2010, Vogel et al 2013), involves oxygen deprivation arising from obstruction of the placental blood flow, which may be rooted in maternal pre-eclampsia, placental abruption, or umbilical cord accident. The high death rate associated with asphyxia is mainly due to poor delivery management. Signs of asphyxia can be identified via fetal heart rate monitoring (Figueras and Gardosi 2011), and timely detection and intervention can reduce the risk of irreversible organ damage and identify cases requiring rapid deliveries (Goldenberg et al 2010). However, this basic monitoring procedure is not often practiced in LMICs.
Low birth weight (LBW), common among preterm (<37 weeks) or small-for-gestational-age (SGA) babies, is documented in 70–80% of the perinatal deaths (Lawn et al 2014, Allanson et al 2015, Mahdizadeh et al 2019). In LMICs, approximately 60% of LBW newborns are SGA (Lee et al 2013), which, in these countries, is often ascribed to intrauterine growth restriction (IUGR) (de Onis et al 1998, Lee et al 2013). IUGR can develop as a consequence of maternal vascular problems, malnutrition, or placental malfunction (Yakoob et al 2009).
While fetal cardiac assessment has been in use over the past four decades to diagnose, monitor, or predict adverse fetal conditions throughout pregnancy (Liston et al 2007, World Health Organization 2016b), there is still insufficient evidence with regards to its contribution to improved perinatal outcomes (Signore et al 2009). As a result, the World Health Organization (WHO) does not currently recommend continuous cardiotocography during labor for assessment of fetal well-being in healthy pregnant women undergoing spontaneous labor, but rather a periodic point of care auscultation (World Health Organization 2018a).
To explore the potential future directions for fetal monitoring in low-resource settings, this review presents an insight into fetal cardiac assessment, briefly explaining the affordability and applicability to each stage of pregnancy. Finally, we provide an overview of how mobile technology may reduce barriers to access perinatal care in poor-resource settings.
2. Fetal cardiac circulation
The human heart develops within the first six weeks of gestation (Gittenberger-de Groot et al 2019), with the majority of its functionality achieved by the eighth week of gestation (Archer and Manning 2009). The development begins with a primary heart tube, which evolves into the four-chambered adult heart structure, composed of two atria and two ventricles. Fetal circulation is unique in that blood is oxygenated in the placenta rather than in the lungs (Stock and Vacanti 2001).
The oxygenated blood from the placenta (umbilical vein) flows into the fetal liver and later, the inferior vena cava via the ductus venosus. On the one hand, the majority of oxygenated blood flows directly from the right atrium to the left atrium through the foramen ovale, and subsequently to the left ventricular to be pumped to the aorta (Stock and Vacanti 2001, Freeman et al 2012). On the other hand, the remaining oxygenated blood passes from the right atrium to the right ventricle and subsequently to the pulmonary vein. As fetal lungs are non-functional, a significant percentage of the blood in the pulmonary vein passes into the aorta via the ductus arteriosus (Stock and Vacanti 2001). The blood sent to the aorta circulates to the fetal brain and tissues. Finally, deoxygenated blood is transported to the placenta via two umbilical arteries (Freeman et al 2012). After birth, the foramen ovale closes, resulting in occlusion of the ductus venosus and arteriosus, and to the separation of the pulmonary and circulatory functionalities (Stock and Vacanti 2001).
2.1. Control of fetal heart rate
The fetal heart rate (FHR) represents the reciprocal of the interval between two successive fetal heartbeats. The heartbeats are controlled by cardiac muscle cells located in the myocardium (Stock and Vacanti 2001). The cardiac cells are categorized as myocardial contractile cells and myocardial conductive cells. The contractile cells stimulate the contractions required to pump blood throughout the body, whereas the conductive cells are the autorhythmic cells responsible for the heart’s electrical activity.
The majority of the conducting cells are located in the sinoatrial node (SAN), also called the pacemaker. The SAN initiates action potentials, resulting in the contraction of the atria at the onset of systole (Stock and Vacanti 2001). The action potential is propagated via the atrioventricular (AV) node to the bundle branches and Purkinje fibers located within the ventricular walls. This impulse initiates ventricular contractility, which, in turn, pumps blood to the pulmonary veins and the aorta, to mark the end of systole. The impulse then leaves the ventricles, marking the onset of diastole, during which the ventricular walls are repolarized. The electrical and mechanical events of a heart contraction generate the cardiac cycle, which is measured as the number of beats per minute (bpm).
Throughout pregnancy, the pace of fetal cardiac activity is controlled by the autonomic nervous system (ANS), baroreceptors, and chemoreceptors (Freeman et al 2012). The ANS is comprised of the sympathetic and parasympathetic nervous systems. The sympathetic system accelerates the heart’s electrical activity, yielding a faster FHR. The parasympathetic system, on the other hand, has the opposite effect on the FHR.
The balance between the sympathetic and parasympathetic nervous systems sets the baseline of the heart rate. However, as the sympathetic system matures earlier than the parasympathetic system, the FHR is higher in the first months of gestation. At 15 weeks gestation, the average FHR is 60 bpm. With the advancement of pregnancy, and the evolvement of the parasympathetic system, the FHR increases to approximately 110–160 bpm (Pildner von Steinburg et al 2013).
Figure 1 illustrates how the sympathetic and parasympathetic systems affect the FHR across gestation, as reported by Wakai (2004). On analysis of traces of 61 healthy pregnant women, a short-term increase in FHR variability during the last trimester was noted. In contrast, long-term variability in FHR was most pronounced during the early gestational period.
Figure 1.

Illustration of the variation in FHR during gestation weeks 22–38. Note that the vertical axis has an arbitrary offset. FHR drops by about 15 bpm from week 25 to week 40 of gestation (Kapaya et al 2018). Adapted from Wakai (2004). Copyright © 2004 Elsevier Inc. All rights reserved.
The other two mechanisms that regulate the fetal heart rate are the baroreceptors and chemoreceptors. Barocepters are located in the aortic arch, carotid arteries, and brain stem. When blood pressure increases, baroreceptors signal the vagal nerve to slow down the heart rate, which then reduces blood pressure. In response to the blood pressure decrease, baroreceptors reduce the parasympathetic tone and stimulate an increase in the fetal heart rate and blood pressure.
The chemoreceptors, found in the aorta, carotid artery, and brain stem, impact the fetal heart rate via its oxygen level-sensing capacities. When the oxygen level decreases, the FHR is accelerated to increase the oxygen input rate from the placenta. However, when the oxygen level reduction is abrupt (hypoxemia), the chemoreceptors trigger a vagal response, resulting in a reduction in heart rate and an increment in blood pressure.
3. Fetal heart monitoring techniques
Fetal heart monitoring technologies can be categorized as either intermittent auscultation (IA) or electronic fetal monitoring (EFM) methods. IA techniques focus on verifying fetal cardiac performance by counting the number of beats over short periods, most commonly measured with a Pinard fetoscope, DeLee fetoscope, or hand-held Doppler device Blix et al (2019). EFM methods identify fetal stress or distress based on FHR variability, commonly performed via cardiotocography (CTG) (Miller et al 1984). It provides continuous information on the FHR, for a period of 10–60 min, using autocorrelation to obtain the average FHR over a specific window, which is generally every 3.75 s (Martis et al 2017).
EFM techniques can be categorized into invasive and non-invasive methods. In the invasive mode, the fetal electrocardiogram (fECG) is taken directly from the fetal scalp (Bakker et al 2004). Although the invasive technique is more accurate than non-invasive modalities, its use is limited to the intrapartum period, when the membranes are ruptured. In contrast, non-invasive methods are only employed during the antenatal period. Non-invasive methods extensively described in the literature include CTG, abdominal fECG, phonocardiography (fPCG), and fetal magnetocardiography (fMCG) (Gan et al 2009, Sameni and Clifford 2010), Wolfberg 2012, Ayres-de Campos et al 2015, Adithya et al 2017).
3.1. Fetal phonocardiogram
The fetal phonocardiogram (fPCG) is an electronic extension of the Pinard and DeLee stethoscopes. Similar to the stethoscope, fPCG is an IA technique in which a microphone is placed on the maternal abdomen to listen to fetal heart sounds (Sameni and Clifford 2010). The audible heart sounds correspond to the closure of the fetal valves during the cardiac cycle (Kovács et al 2011). The closure of the mitral and tricuspid valves generates a sound called S1, and the closure of the semilunar valves (pulmonary and aorta) generates a sound called S2. Both S1 and S2 have low acoustic energy and are affected by noises such as environmental noise, as well as other maternal and fetal physiological sounds, such as breathing, fetal movements, and maternal circulation (Várady et al 2003). The technique can extract cardiac timing and intensity of fetal heart sounds, which can carry useful diagnostic information (Adithya et al 2017).
The fPCG can be used during the antepartum phase (gestational week ≥24) (Várady et al 2003). Although fPCG is an alternative to the traditional ultrasound used in perinatal management (Hamelmann et al 2019), it is underutilized (Adithya et al 2017) and suffers from significant challenges related to signal acquisition and processing. Further research is needed to improve fPCG to compete with standard fetal monitoring methods, i.e. CTG and ultrasound imaging.
3.2. One-dimensional doppler ultrasound
One-dimensional Doppler ultrasound (1D-DUS) estimates FHR by measuring the Doppler shift between ultrasound beams transmitted and received from the mechanical heart movements and blood flow. The Doppler magnitude frequency shift fD, is described as (Kohler and Sumner 2014):
| (1) |
where f D is the measured change in frequency (Hz), f o the frequency of emitted ultrasound transducer in Hz, c the speed of sound in soft tissue in m s−1, V the velocity of the reflecting interface in m s−1 and θ is the angle between the ultrasound beam and the surface in radians.
The transmitted beam travels across various anatomical structures, from the skin surface, through the maternal skin and subcutaneous tissue, and then finally reaches the uterine muscles, the amniotic sac, and the fetal heart (Marieb and Hoehn 2007). The fetal heart movement reflects the ultrasound beam, and propagates the ultrasound waves in the reverse order. The distance between the DUS transducer and fetal heart depends on the maternal phenotype, which varies among nationalities (World Health Organization 2014a), socioeconomic status, as well as body mass index (Morgenstern et al 2009).
The shifted Doppler frequency is usually demodulated via the phase-quadrature demodulation, in which the received signals are mixed with the carrier signals sin 2πf ot and cos 2πf ot (Evans 1989). The demodulated signal is then autocorrelated to estimate the cycle period of the heartbeat rhythm (Shakespeare et al 2001).
Doppler ultrasound includes two different modes—the continuous wave (CW) and the pulsed-wave (PW). In CW, two piezoelectric crystals continuously monitor the reflection of the emitted wave. In PW, one piezoelectric crystal alternates between sending and receiving the sound waves. The dual functionality of CW Doppler enables the measurement of higher velocities. However, as velocities are measured in the same line of interrogation, it is impossible to know the origin of the velocity. In contrast, PW measures slower velocities, but the emitted sound waves are associated with the received waves, thus enabling the detection of the structure’s distance reflecting the wave.
CW is primarily integrated in hand-held Doppler transducers, while PW is used in standard CTG machines. Hand-held Doppler transducers are used during the intrapartum and antepartum periods after the 20th gestational week, to measure heart rate variability metrics as an indication of fetal wellbeing.
Mahomed et al (1994) carried out a randomized, controlled trial that showed that fetal monitoring using hand-held DUS transducers could detect a similar number of prolonged or late decelerations as ultrasonography. Additionally, the hand-held Doppler devices detected a substantially larger number of late and prolonged decelerations than the Pinard stethoscope. The potential of hand-held Doppler devices has also been reported by Devane et al (2017), following a randomized controlled study that demonstrated that they provide the same level of safety for screening and monitoring as cardiotocography in low-risk pregnancies.
3.3. Cardiotocography
Cardiotocography (CTG) is the simultaneous and continuous measurement of FHR and uterine pressure, often detected as uterine contractions (UC), and is a standard method for assessment of fetal wellbeing (Grivell et al 2015). To record the FHR, the medical assistant applies a gel on the maternal abdomen and the ultrasound transducer. The transducer is moved across the maternal abdomen while the technician listens for an audible version of the Doppler signal, in an attempt to identify the spot with maximum fetal heart rate impulse (as opposed to maternal arterial flow) (Freeman et al 2012).
CTG is used for fetal monitoring starting at the 20th week of gestation (Devane et al 2017), but most commonly indicated after the 28th week of gestation (Schneider and Maternal Fetal Medicine Study Group 2014). Although CTG is widely used, it suffers from high intra- and inter-interpreter variability (Todros et al 1996, Bernardes et al 1997, Blix et al 2003, Blackwell et al 2011, Hruban et al 2015), resulting in low specificity. To reduce this subjectivity, Dawes et al (1992) introduced a computerized version of the CTG. A Cochrane review of two studies (469 subjects) concluded that the mortality rate in a population monitored by computerized CTG was four times lower than in the population monitored by visual CTG (0.9% vs. 4.2%) (Grivell et al 2015).
Computerized CTG has also been used in recent years to develop artificial intelligence methods to detect abnormal FHR patterns, achieving comparable results to clinical assessment of the CTG (Chudáček et al 2011, Warrick et al 2012, Georgieva et al 2019). Notably, these artificial intelligence-based CTG systems have shown the potential to discriminate between normal and IUGR fetuses (Stroux et al 2017, Signorini et al 2020).
Although CTG is a standard method used for fetal monitoring in high-income countries, controlled clinical trials have not provided evidence of its benefits; CTG was associated with a 20% increase in Cesarean interventions with no improvement in fetal outcomes (Devane et al 2017). In addition, the use of CTG was not associated with statistically significant improvements in perinatal outcomes as compared to traditional intermittent auscultation methods (Kamala et al 2018b, Mdoe et al 2018b).
3.3.1. The nonstress test
The nonstress test (NST) monitors FHR patterns for at least 20 minutes, and is designed to identify accelerations associated with fetal movements (Malhotra et al 2014). This test calculates the baseline FHR, which is later used to measure long-term and short-term variability, episodes of high and low variation, acceleration, and deceleration. Test results are considered normal (reactive) when more than two accelerations occur within 20 minutes of observation. In contrast, a non-reactive result is when, no more than one acceleration occurs within 40 minutes (American College of Obstetricians 2000). The NST has a low false-negative (0.3%), but a high false-positive (50%) rate (Eden et al 1990). Of note, the test carries no risk of inducing any uterine contractions.
3.3.2. Contraction stress test
The contraction stress test (CST) is based on the premise that contractions, induced using oxytocin or nipple stimulation (Malhotra et al 2014), trigger a hypoxic state (Signore et al 2009). A healthy fetus can tolerate this hypoxic state, whereas a non-healthy fetus will respond with late FHR decelerations (Malhotra et al 2014). Although this method has a low false-negative rate (0.04%) (Freeman et al 1982) and a lower false-positive rate than NST (30% vs. 50%) (American College of Obstetricians 2000), it requires an intravenous intervention, which increases the risk of fetal hypoxia and of induction of preterm birth (Malhotra et al 2014).
3.3.3. Acoustic stimulation
Acoustic stimulation is a variation of NST, in which if there are no fetal movements, vibroacoustic stimulation (usually with a laryngeal stimulator) may be activated for 3 seconds on the maternal abdomen over the fetal head. This is performed to ‘awaken a sleeping fetus’, before initiation of the NST (Malhotra et al 2014). The artificial larynx produces a vibratory stimulus of 80 Hz that causes a healthy fetus to increase physical activity, as measured by an increase in FHR. Its advantages include shortening of the NST by 10 minutes (Clark et al 1987), and reduction of the number of non-reactive states, without affecting readability (Smith et al 1985, Smith et al 1986). In cases of a non-reactive result, the acoustic stimulation is repeated for 5 min. If the test is still non-reactive, a fetal biophysical profile or CST is indicated (Malhotra et al 2014).
3.4. Fetal electrocardiogram
The fECG records the complex electrical activity of the fetal heart. The main components of the ECG signals are P, Q, R, S, and T waves. The P wave represents atrial depolarization, which is followed by the atrial contraction (atrial systole). The atrial contraction is extended to the QRS complex, which corresponds with ventricular depolarization, with ventricles contracting at the peak (R wave). The ventricular contraction lasts until the ST-T wave, which corresponds with ventricular repolarization and relaxation. fECG can be captured in an invasive manner during the intrapartum period, when the cervix is dilated, and the fetus scalp is visible, or in a non-invasive manner, starting from the second trimester. The fECG is also used to complement CTG at intrapartum to reduce unnecessary Cesarean sections (Vijgen et al 2011, East et al 2014).
3.4.1. Invasive fetal electrocardiogram
The invasive fetal electrocardiogram (invasive-fECG) requires the rupture of the membranes to introduce electrodes, via the cervix, and to place them on the fetal scalp (Hon 1960, Hon and Hess 1960, Hon and Lee 1964). This technique processes the recorded signals to visualize the P and T waves, as well as the QRS complexes.
Scalp fECG has been used as a complementary technique during intrapartum FHR monitoring (Amer-Wåhlin et al 2001, Norén et al 2006). The morphology of the ST segment is analyzed to find patterns associated with uterine complications (Larks and Longo 1962, Larks and Larks 1966). Invasive fetal ST can be captured and analyzed from the 36th gestational week and is indicated in high-risk pregnancies when a non-reactive CTG is obtained, or labor is induced by oxytocin. Although its use has shown to effectively reduce neonatal encephalopathy (Norén et al 2003, Norén et al 2006), randomized control trials of this technology have yet to demonstrate a clear benefit.
3.4.2. Non-invasive fetal electrocardiogram
Non-invasive fECG measures the electrical activity of the fetal heart via electrodes which are placed on the maternal abdomen (Lai and Shynk 2002). This technique is indicated from the 18th week of gestation (Sameni and Clifford 2010), and therefore has much wider applicability than invasive ST analysis and can replace Doppler auscultation for the fetal heartbeat.
Although abdominal fECG signals have a relatively low amplitude (microvolts), it can provide a more accurate estimate of beat location when compared to the CTG, and hence a more accurate quantification of fetal heart rate variability indexes (Jezewski et al 2012, Clifford et al 2014, Jezewski et al 2017). The morphology and beat-to-beat heart rate variability estimated from fECG are established indicators of pre-eclampsia and IUGR. For instance, a study conducted on 106 patients (30 healthy, 44 mild pre-eclampsia, and 32 severe pre-eclampsia subjects) at 34–40 weeks of gestation, reported that FHR variability indices were associated with the suppression of fetal biophysical activity and the development of fetal distress in women suffering from severe pre-eclampsia Lakhno (2017). Similarly, Velayo et al (2017) assessed the impact of IUGR on FHR variability indices extracted from abdominal fECG recordings of 20 control and 15 IUGR singleton pregnant women. While the authors identified clear P-QRS-T complexes in all cases, prolonged QT intervals were measured in IUGR fetuses.
Over the last 30 years, a variety of methods have been proposed for extracting and processing fECG signals (Sameni and Clifford 2010), Jaros et al 2018). Methods range from adaptive filtering (Park et al 1992, Shao et al 2004, Martens et al 2007) to non-adaptive approaches such as, independent component analysis (Sameni et al 2006), principal component analysis (Al-Zaben and Al-Smadi 2006), wavelet transforms (Castillo et al 2013, Wu et al 2013) and neural networks (Assaleh 2007, Amin et al 2011, Behar et al 2014). Many of these techniques suffer from significant limitations due to causality and signal stability. Other approaches based on generalized eigenvalue decomposition have shown more promise (Sameni et al 2007). In Clifford et al (2009) and Behar et al (2016), it was shown that this approach was able to accurately resolve both QT interval and ST elevation/depression from non-invasive fECG. However, this promising result has yet to be applied in a randomized clinical trial to demonstrate efficacy.
3.5. Fetal magnetocardiography
Fetal magnetocardiography (fMCG) uses a sensitive, superconducting sensor to measure the magnetic field of fetal heart activity (Kariniemi and Hukkinen 1977, Sameni and Clifford 2010), Kiefer-Schmidt et al 2012). The fMCG provides a waveform almost identical to that of fECG, but at a higher signal-to-noise ratio, and with a higher resultant quality of waveform (Peters et al 2001, Kiefer-Schmidt et al 2012). The higher quality enables the classification of arrhythmias and detection of congenital heart diseases (Kähler et al 2001), as well as the ability to assess fetal neurological development (Wakai 2004).
The fMCG is used from the 20th gestational week (Peters et al 2001). Yet, although it provides good-quality waveforms, it is not routinely used in perinatal care due to its higher costs, i.e. the need for a shielded room, and highly skilled personnel (Peters et al 2001, Kiefer-Schmidt et al 2012). Alternative methods, such as the abdominal fECG or the hand-held Doppler, can be used at any time during pregnancy, and can even be performed at home by the patients themselves (Sameni and Clifford 2010).
4. Ultrasound imaging
Ultrasound imaging is considered the gold standard for fetal monitoring in high-income countries (Liston et al 2007, World Health Organization 2016c). It evaluates fetal growth, fetal cardiac structure and function, and fetal, uterine, and placental blood circulation. Ultrasound imaging is usually indicated in the second trimester, particularly after the 20th week of gestation, with a scan recommended by the WHO before week 24 (Whitworth et al 2015, World Health Organization 2016c, World Health Organization 2016b).
Ultrasound imaging is known to effectively assess pregnancy viability, estimate gestational age, detect multiple pregnancies, and determine placental position (Whitworth et al 2015). While there is no compelling evidence that ultrasound scans reduce perinatal mortality (Neilson 1998, Dražančić 2001), they can be used to validate suspicious diagnoses without invasive and risky interventions, reduce labor induction for post-term pregnancy, and detect fetal malformation (Whitworth et al 2015). Moreover, a review of 58 obstetric articles, concluded that ultrasound imaging provides appropriate clinical management in at least 30% of cases when used by skilled operators (Groen et al 2011).
Ultrasound imaging has also shown potential in the assessment of IUGR. A comparison between 38 IUGR and 32 appropriate for gestational age (AGA) fetuses showed that growth-restricted fetuses had a statistically significant thicker aortic wall than the AGA fetuses (1.9 mm vs. 1.15 mm) (Cosmi et al 2009). The median diameter of the abdominal aorta was also significantly higher in IUGR than in AGA fetuses. The thicker aortic wall in the IUGR fetus was also noted by Gomez-Roig et al (2015), who compared 35 IUGR fetuses with 49 AGA fetuses. In contrast to Cosmi et al (2009), they reported on the substantially lower diameter of the abdominal aorta for fetuses with IUGR (Gomez-Roig et al 2015).
There is no scientific evidence for, or consensus on how often ultrasound scans should be performed during pregnancy. Some obstetricians recommend at least four ultrasound scans during normal pregnancies, whereas others recommend only one, to be performed before the 24th gestational week (Papp and Fekete 2003). When four scans are performed during pregnancy, the first is conducted between weeks 10 and 14, to validate the pregnancy and estimate gestational age. The second scan is carried out between weeks 18 and 22, to detect fetal anomalies and confirm gestation age. The third scan is scheduled between weeks 30 and 34 of gestation, to assess fetal growth. The final scan is scheduled between weeks 36 and 38, and focuses on the fetal weight, position, and orientation/presentation, which helps to determine the optimal mode of delivery.
4.1. Fetal biometry
Ultrasound imaging enables measurement of different fetal organs, and estimation of gestational age and fetal weight. The most common measures are biparietal diameter (BDP), femur length (FL), head circumference (HC), crown rump length (CRL), and abdomen circumference (AC) (Malhotra et al 2014). Using a combination of these measurements, fetal weight estimates are within 5% of the actual weight in 50% of cases, and within 10% in 80% of the evaluations (Galan et al 2002).
Fetal biometry measurements have been shown to be more accurate during the first trimester. During the second and third trimesters, fetal measurement accuracy is impacted by genetic and nutritional factors (Reece et al 1989). Recently proposed formulas combining the transcerebellar diameter (TCD) with FL and AC are emerging as a solution for dating late pregnancies (after the 24th gestation week), with gestational age estimates within ±3 weeks of the CRL measurement taken between the 8th and 14th gestational weeks (Deb et al 2020).
Fetal biometry measurements can differentiate between fetuses that are IUGR and those that are constitutionally small (SGA) (Soothill et al 1993). Specifically, when an estimated weight is below the 10th percentile for gestational age, the fetus is considered growth-restricted, as defined by the American College of Obstetricians and Gynecologists (ACOG) guidelines (Uquillas et al 2017). However, a previous study, conducted by the Prospective Observational Trial to Optimize Pediatric Health (PORTO), found that only 2% of fetuses whose estimated birth weight was within the 3rd and 10th percentile, had an adverse perinatal outcome; the authors concluded that the threshold should be below the 3rd percentile (Unterscheider et al 2013).
Furthermore, fetal biometry measurements ignore the fact that 10% of the normal population is genetically predisposed to be small, thus increasing the false-positive rate (Galan et al 2002). Hence, to increase the accuracy in detecting IUGR, fetal biometry should be combined with methods assessing the fetal ANS physiology (Galan et al 2002). When IUGR is detected, the pregnancy is categorized as high-risk, as the condition has long-term consequences.
4.2. Doppler velocimetry
Doppler velocimetry assesses the blood flow in the umbilical arteries and vein to evaluate pregnancies at risk of fetal compromise (Trudinger et al 1986), such as growth restriction (Alfirevic and Neilson 1995) or cardiovascular abnormalities (Berkley et al 2012). In healthy pregnancies, the placental and fetal circulation transfers oxygen and nutrients, and eliminates fetal waste products (Malhotra et al 2014).
Umbilical flow is assessed using different indexes, such as systolic and diastolic ratio, pulsatility index and resistance index (American College of Obstetricians 2000). Higher indexes indicate significant vascular resistance, thus implying that fetal health is at risk (Berkley et al 2012, Signore et al 2009).
The resistance indexes are mainly measured on the umbilical artery (UA), the middle cerebral artery (MCA), and the ductus venosus (DV) (Mone et al 2015). Of these three areas, the UA Doppler is the only device that has been the subject of randomized controlled trials, which have supported its feasibility for fetal surveillance in high-risk pregnancies (Alfirevic et al 2017). The UA Doppler measures the resistance in fetoplacental circulation flow, providing a pulsatility index (PI). In a healthy fetus, the UA has a forward flow. However, increases in placental resistance obliterate the muscular arteries in the placental villi, resulting in a reduced diastolic flow (Berkley et al 2012), which then eliminates and later reverses the fetoplacental circulation flow. Both the absence and the reversal of flows can be visualized in the Doppler images. In the case of the absent end-diastolic flow (AEDF), the pronounced systolic peak is followed by an interruption, while in the reversed end-diastolic flow (REDF), the systolic peak is followed by a negative peak. In fetal growth-restricted pregnancies with AEDF or REDF, delivery is recommended at week 32 (Royal College of Obstetricians and Gynaecologists 2002).
Randomized controlled trials have demonstrated that a UA PI greater than the 95th percentile in restricted-growth fetuses is an indicator of a perinatal adverse outcome (Unterscheider et al 2013, O’Dwyer et al 2014). The use of UA Doppler was also shown to be effective in reducing the incidence of perinatal deaths and induced deliveries (Alfirevic et al 2017). MCA flow can be used to detect problems caused by fetal hypoxemia in IUGR. In a hypoxic state, most of the oxygenated blood is supplied to the brain, heart, and adrenal glands, affecting the peripheral circulation (Uquillas et al 2017). This phenomenon is called brain-sparing reflex and is observable in the waveform of the MCA Doppler. MCA Doppler is also a reliable indicator of anemia. Moreover, the MCA PI/UA ratio can indicate adverse perinatal outcomes (Mari and Hanif, 2008), which are related to an increment of the diastolic flow due to hypoxia (Morris et al 2012).
DV flow can be used to detect the cardiac failure in IUGR, particularly in cases of early-onset fetal growth restriction (Baschat 2010). It is a reliable marker of acidemia and stillbirth (Baschat 2010), which are caused by absent or reversed end-diastolic pressure at the ductus venosus. Although DV flow measurement displays moderate accuracy in detecting fetal compromise, previous works have suggested that DV Doppler alone is insufficient for fetal surveillance (Royal College of Obstetricians and Gynaecologists 2002). Furthermore, DV Doppler does not offer any added benefit over traditional CTG for fetal monitoring (Lees et al 2015). Nevertheless, delaying delivery until finding an abnormality using DV flow could prevent neurological impairment in the long-term (Ganzevoort et al 2017). Randomized controlled trials are still needed to more accurately assess the benefits of DV flow measurement.
Other anatomical areas useful in the management of fetal growth-restricted pregnancies are the uterine artery, the aortic isthmus, umbilical vein, and the atrioventricular valves (Mone et al 2015). The uterine artery flow is useful in identifying pre-eclampsia and SGA neonates in high-risk pregnancies (Royal College of Obstetricians and Gynaecologists 2002). The aortic isthmus measures the balance between the brain’s impedance and systemic circulation, indicating cardiac dysfunction when there is an abnormal balance (Cruz-Martinez et al 2011). Umbilical vein flow provides an indication of fetal venous circulation, where high values suggest increased venous pressure that results in right-sided heart failure and myocardial hypoxia (Nicolaides et al 2002).
4.3. Fetal echocardiography
Fetal echocardiography is a non-invasive ultrasonography technique that examines fetal cardiac anatomy and function (Godfrey et al 2012). The accuracy and speed of performing fetal cardiac assessment have improved in the last decades, following the introduction of advanced techniques such as color Doppler (DeVore et al 1987). The primary use of fetal echocardiography is in the detection of congenital heart diseases (CHDs), which are the most common abnormality in fetuses, with a prevalence of around 8 to 9 per 1,000 live births (Hoffman and Christianson 1978). The procedure assesses the heart structure, as well as the direction, pattern, volume, and velocity of flow (Allan 1986). The basic visualization of the chambers can be extended to include blood flow through the chambers, using a technique called ‘five chambers views’, which increases the sensitivity of detecting CHDs by 5%, achieving a final sensitivity rate of 65% (Allan 2000).
Fetal echocardiography also includes a pulse wave Doppler component, which is recommended for a complete evaluation of the fetal heart. The pulse wave shows the blood flow through the atrioventricular, mitral, and tricuspid valves (Abuhamad and Chaoui 2012). These valves generate a dual-peak Doppler waveform that comprises the E-wave, which is the passive diastolic filling, and the A-wave, which is the active diastolic filling (‘atrial kick’) (Abuhamad and Chaoui 2012). In healthy fetuses, the amplitude of the A-wave is greater than that of the E-wave, which increases throughout gestation. A higher increase in the E-wave/A-wave ratio is a sign of IUGR or congenital cystic adenomatoid malformation, which can lead to mitral or tricuspid regurgitation (Mari and Hanif 2008, Mäkikallio et al 2008).
Modern echocardiography techniques include three-dimensional (3D) and four-dimensional (4D) fetal heart assessment (Deng et al 1996), which enable real-time examination of the heart rate function, and a more accurate assessment of the heart structures (Chaoui et al 2004, Bennasar et al 2010, Ionescu 2010).
Although fetal echocardiography is considered one of the most relevant fetal cardiac assessment techniques, it is costly and requires qualified specialists to perform the examination (Caserta et al 2008). Therefore, fetal echocardiography is only provided when indicated by specific maternal and fetal conditions.
5. Comparison of fetal cardiac monitoring methods
Table 1 presents an comparison of the fetal cardiac monitoring techniques presented in sections 3 and 4, particularly with respect to the following four criteria:
Table 1.
Comparison of fetal cardiac monitoring methods. The first column presents a four-point ordinal scale of medical equipment cost, from low ($) to extremely high ($ $ $ $). The horizontal line indicates when, during pregnancy, the technology can be used. The color of the line indicates the time required for training operators (green: low; blue: moderate; cyan: considerable; red: high; magenta: extreme). The thickness of the line indicates the relative evidence for the utility of each technology.
|
GA ≥ 24 weeks (Várady et al 2003).
GA ≥ 20 weeks (Peters et al 2001).
GA ≥ 20 weeks (Peters et al 2001).
GA ≥ 20 weeks (Grivell et al 2015).
GA ≥ 18 weeks (Sameni and Clifford 2010).
Intrapartum (GA ≥ 36 weeks) (Norén et al 2006).
GA ≥ 20 weeks (World Health Organization 2016b).
GA ≥ 20 weeks (Peters et al 2001).
medical equipment cost;
operator training requirement;
gestational week at which the device can be used; and
evidence supporting the device’s utility.
5.1. Cost analysis and availability in LMICs
The fMCG is an expensive method, requiring specialized operator training, dedicated shielded rooms, each costing approximately $350 000 to construct (Strasburger et al 2008) and high maintenance (Peters et al 2001, Cannie et al 2006, Kiefer-Schmidt et al 2012), which have limited its use in HICs (Cannie et al 2006, Wakai 2014) and its introduction into and widespread integration in LMICs.
Even compact portable ultrasound equipment, such as the GE LOGIQ Book XP (General Electrics, Milwaukee, WI, USA), costs at least $10 000, and carries additional expenses such as maintenance, supplies, battery replacement, and staff training (World Health Organization 2014b). However, a recent review on the use of ultrasound in LMICs reported an expanding utilization of low-cost, portable imaging technology in low-resource settings (Stewart et al 2020).
The ultrasound devices most commonly used for fetal monitoring were provided by Sonosite Inc (Bothell, WA, USA) (Kimberly et al 2010, Greenwold et al 2014, Boamah et al 2014, Kozuki et al 2016) and General Electric (Ome-Kaius et al 2017, Goldenberg et al 2018). Despite these early indications of their increasing integration in fetal monitoring in LMICs, there is still a need to assess the benefits, trade-offs, and potential drawbacks of large-scale obstetric ultrasound implementation in these regions (Kim et al 2018).
Several randomized controlled studies using ultrasound imaging in antenatal care in LMICs, did not show any significant reductions in adverse perinatal outcomes (Neilson 1998, Dražančić 2001, Kim et al 2018, Goldenberg et al 2018, Franklin et al 2018). Moreover, this technique requires specialized skilled operators, who are in limited supply in LMICs (World Health Organization 2014b).
Scalp fECG is limited to intrapartum use and requires specialized training. The most common device used for invasive fECG is the STAN monitor (Neoventa Medical, Goteborg, Sweden) (Clifford et al 2014). Vijgen et al (2011) reported that the average cost of ST analysis in 2 827 deliveries in a Dutch hospital was €1345, which is not feasible for LMICs.
Abdominal fECG can be captured using low-cost equipment, which does not require skilled users (Behar et al 2016, Marzbanrad et al 2018). The Monica AN24 monitor (Monica Healthcare, Nottingham, UK) and the Meridian M100/M1000 monitors (MindChild Medical, North Andover, MA), both which have been approved by the Food and Drug Administration (FDA) and the European Commission (CE), are two commercial devices commonly used for non-invasive fECG (Behar et al 2016). However, although the cost of fECG devices is relatively low ($4500 for the Monica Novii Wireless Patch System (Cardo Medical 2020)), non-invasive fECG is still not widely used since the systems still require further testing to definitively demonstrate that the morphological analysis is similar to that provided by the scalp electrocardiography method (Smith et al 2019, Kahankova et al 2019).
On average, CTG machines cost at least $450 (World Health Organization 2016c) and require maintenance, supplies, and training, thereby limiting its use in low-resource settings (World Health Organization 2016b). Although obstetric protocols in HICs recommend CTG, its use in LMICs has not improved fetal outcomes in comparison to auscultation methods (Housseine et al 2018). In contrast, auscultation methods, particularly the fetoscope, have been shown to be associated with reduced perinatal deaths in LMICs (Wall et al 2010).
Of the various auscultation methods available, the Pinard stethoscope is the most available tool in resource-constrained regions due to its affordable cost (Jauniaux and Prefumo 2016). Before its introduction, midwives used a stethoscope in the labor ward to listen to the fetal heart rate for ten minutes every half hour (Mahomed et al 1994). However, auscultation with the stethoscope can be uncomfortable for both patients and practitioners. Additionally, the stethoscope provides unsatisfactory results due to confounding factors such as environmental noise (Mahomed et al 1994).
Hand-held Doppler transducers are simple to use and can be purchased for as little as $17 (Stroux et al 2016). They require less time than the Pinard stethoscope to assess the fetal heart rate (Mahomed et al 1994). Plotkin et al (2019) compared the performance of hand-held Doppler and Pinard devices for fetal monitoring in the intrapartum period by reviewing 19 studies conducted in India and African LMICs. The comparison showed that Doppler devices accurately detected more fetal abnormalities than the Pinard stethoscope. However, there was no statistically significant improvement in perinatal outcomes when an anomaly was detected. The authors suggested that the lack of progress resulted from the poor clinical management and protocol referral of abnormality events. The review also found that both patients and medical providers preferred Doppler hand-held devices than Pinard stethoscopes, thereby justifying the integration of Doppler in fetal monitoring protocols for LMICs.
6. Usage of devices in LMICs
In LMICs, there is limited availability of the life-saving and complex medical devices routinely used for fetal cardiac monitoring in developed countries. The technologies involve the use of ultrasound technology, telemedicine, CTG and other fetal monitoring techniques that may not be easily implementable in low-resource settings. However, multiple research studies have demonstrated the acceptable effectiveness of suitable and appropriate technologies in these regions for fetal cardiac monitoring (Banta and Thacker 2002, Wyatt 2008, Rahman et al 2012, Devane et al 2017, Alves et al 2019).
Mdoe et al (2018a) demonstrated the superiority of fetal heart rate monitoring using continuous Doppler when compared to intermittent fetoscope auscultation. In their study, there was an 8.1% incidence of abnormal fetal heart rate detection when using continuous Doppler, versus 3.0% with the use of intermittent auscultation.
Additional studies have provided evidence for the preferential use of Doppler ultrasound for fetal cardiac activity monitoring. It has also been shown to accurately determine the number of reported fetal demise and to classify them as stillbirths versus neonatal deaths (Lopez et al 2007). Additionally, the emotional reassurance among mothers associated with hearing the fetal heartbeat amplified by the Doppler is a positive experience that could influence positive outcomes. Compared to the Pinard stethoscope, the Doppler was noted to be superior in detecting abnormal intrapartum fetal heart rate but was not associated with improved perinatal outcomes (World Health Organization 2018a).
Britton et al (2019) demonstrated the use of ‘tele-ultrasound’ for fetal cardiac monitoring in low resource settings. In the study, fetal monitoring devices were delivered to participants, to overcome challenges of geographical distance, lack of facilities and inadequate healthcare personnel that are common in these areas. The ‘tele-ultrasound’ technique was found to be low-cost, and reliable and implementable in resource-limited settings.
World Health Organization (2018b) reported on improved perinatal outcomes in following the use of CTG during labor in LMICs. Others have associated CTG (cardiotocography) with higher Cesarean section rates, with no added benefit on perinatal outcomes, and therefore do not recommend its use in low-resource settings (Housseine et al 2019). High-quality evidence considering implementation barriers and enablers is needed to determine the optimal fetal monitoring technique in Low-resource settings. World Health Organization (2018b) noted that there are significant gaps between international recommendations and what is practically possible in most resource-constrained countries.
A study in Uganda highlighted critical challenges as shortage of staff and devices, institutional challenges, and maternal perceptions to monitoring Munabi-Babigumira et al (2019). Another study in Tanzania listed lack of strict protocol for use and misidentification of maternal heart rate as the challenges associated with the introduction of Moyo, an electronic strap on a fetal heart rate monitor (Lafontan et al 2019) developed to improve intrapartum fetal heart rate monitoring. More specifically, the quick, short, and unstructured assessments and inferences limited the initiation of various interventions due to indecisiveness. The introduction of CTG was associated with simpler and more efficient monitoring of labor, but no improved outcomes (Kamala et al 2018a, Lafontan et al 2019).
7. Telemonitoring for perinatal care, an alternative for LMICs
In recent years, telemonitoring applications have been developed to enhance maternal and fetal monitoring. These applications have been made possible by the high penetration of mobile telecommunications technologies in LMICs, with approximately 90% of the population owning a mobile phone device (Bergström et al 2015). This high coverage of mobile phone access can be exploited to overcome perinatal care access barriers in LMICs, such as low literacy levels, poor road infrastructure, and medical professionals and equipment deficiencies (Eysenbach and Consort-EHEALTH Group 2011, Sondaal et al 2016, Hasan et al 2017).
The feasibility of mobile health applications (mHealth) in improving antenatal care was presented by Feroz et al (2017) after reviewing 14 cases conducted in sub-Saharan Africa, Southeast Asia, and Middle-East countries. The authors found that mHealth solutions can improve perinatal care services by increasing the percentage of women attending the minimum WHO-recommended perinatal visits. They noted that the most effective mobile apps were those that used client education and behavior change communication via short messages and patient tracking to allow for patient follow-up in subsequent visits.
In a review of telemonitoring in obstetrics, researchers reported that mobile applications connected to external devices, such as electrodes, body sensors, and thermometers provided effective maternal and fetal monitoring (Alves et al 2019). The used external devices enabled the digitalization of data, which was later analyzed by medical professionals or artificial intelligence techniques to detect abnormalities. Among the available fetal monitoring mobile applications, Jatmiko et al (2015) developed a method to identify the location of the fetus from an ultrasound image. The algorithms tested with pictures taken in a public hospital in Indonesia, demonstrated 93% accuracy in the detection of the fetal head and abdomen. Similarly, Khan et al (2016) developed methods to calculate the mean abdominal diameter (MAD) from an ultrasound image. A mobile app prototype was tested on ultrasound images captured by professional midwives in a Norwegian hospital, and demonstrated a mean error of −0.06 mm.
Awiti et al (2016) developed an Android-based digital fetoscope prototype. Fetal heart sounds are acquired using a Pinard horn and a microphone and are later sent to a smartphone via Bluetooth technology. In the smartphone, the audio signals are processed to display a heartbeat. When testing the system on adults, and comparing the measurements with those of a standard electronic sphygmomanometer, the Android-based digital fetoscope achieved a root mean square error of 7.23 bpm with a standard deviation of 5.44 bpm.
Tapia-Conyer et al (2015) introduced a mobile maternal-fetal monitoring application in a resource-poor and educationally-limited community in Mexico. The project aimed to evaluate the feasibility of providing remote antenatal care. The staff in the rural medical center was trained to use the mobile fetal monitor, which comprised a fetal ultrasound heart monitor, a uterine tocodynamometer, and additional tools for recording maternal blood pressure, blood glucose, and urinary protein values. The researchers split the 125 volunteers into control and study groups. The study group received perinatal care at the local medical center using the mHealth system, whereas the control group received standard perinatal care at the main public hospital. Tapia-Conyer et al (2015) observed that volunteers using the mHealth system were more than twice as likely to adhere to antenatal care monitoring than those receiving standard of care. There were no statistically significant differences in adverse perinatal outcomes between the two groups, suggesting that the tested mobile technology did not compromise maternal and fetal health.
A low-cost fetal monitoring system introduced in a rural Guatemalan community in 2013, was later developed into a system that was tested in a randomized controlled trial in 2015–2017 (Stroux et al 2016, Martinez et al 2017, Martinez et al 2018). The mHealth system consists of a low-cost one-dimensional Doppler transducer and a blood pressure device connected to an Android-based smartphone running an app designed for low-literacy traditional birth attendants (TBAs). The TBAs were trained to use the mHealth system during home visits. When a TBA visited a patient, the app guided the TBA to find the fetus and record a Doppler recording of the fetal cardiac activity for up to 20 minutes. The TBA also recorded blood pressure readings from both arms. The app then guided the TBA through basic questions presented through appropriate pictograms and audio prompts, to assist in the identification of concerning signs and symptoms during pregnancy. In the event a risk factor was identified, the app connected the TBA to appropriate (local or remote) medical care through a voice call, to provide decisional support and onward referral to appropriate healthcare, if needed. The Doppler signal and maternal blood pressure recorded with the mHealth system has allowed the development of different modules for providing estimates of fHRV, gestational age, and hypertension (Valderrama et al 2017, Valderrama et al 2018, Valderrama et al 2019, Valderrama et al 2020a, Valderrama et al 2020b, Katebi et al 2020).
8. Discussion and conclusion
Fetal monitoring is performed with a variety of devices and approaches, with CTG and ultrasound imaging considered the ‘standard of care’ in high-income countries. Despite the paucity of evidence supporting the utility of these techniques in reducing perinatal mortality and morbidity (Neilson 1998, Devane et al 2017), their use may still be beneficial throughout pregnancy. Specifically, CTG may facilitate the detection of signs of hypoxia requiring Cesarean delivery (Grivell et al 2015), and ultrasound imaging can help estimate gestation age before gestational week 24, and to detect multiple pregnancies and fetal abnormalities requiring vigilance or/and interventions (Groen et al 2011, Whitworth et al 2015). However, in resource-constrained settings, CTG and Doppler imaging are scarce, due to their high cost and the need for trained operators (Dražančić 2001).
Among fetal monitoring techniques, 1D Doppler transducers provide an affordable option, with an outstanding balance between cost, clinical utility, and operator learning curve, thereby making its use most practical (see table 1). Moreover, Doppler transducers have been shown to be comparable to CTG in fetal heart rate monitoring in LMICs. Taken together, Doppler transducers, which are widely available in these regions (Becker et al 2016), are a reliable ‘standard of care’ in low-resource regions (Devane et al 2017, Housseine et al 2018).
Perinatal care access in LMICs can be facilitated by telemetry (Bergström et al 2015). Mobile technology can support fetal monitoring analysis and transmission of clinical information collected using low-cost devices, such as Doppler transducers, portable CTG, or auscultation methods. In this manner, economic and geographical barriers can be overcome to increase perinatal care coverage in LMICs.
The feasibility of the use and impact of mHealth mobile applications in fetal monitoring has been shown in several works conducted in LMICs (Tapia-Conyer et al 2015, Stroux et al 2017, Feroz et al 2017, Martinez et al 2018). With the advent of increasingly complex smartphones, particularly those with embedded ‘AI’ chipsets), mHealth applications may extend beyond data collection and decision support systems, by processing complex maternal and fetal information in an edge computing paradigm, which allows the use of mobile applications without relying on network communication. However, regulation (especially in the US) is likely to limit this development, and to drive the solutions to higher-cost, self-contained devices.
Increasingly higher bandwidth cellular networks in LMICs could mitigate this by driving the processing to the cloud. Still, the economics of providing high-bandwidth networks to the poorest and least populated parts of the globe will still leave significant disparities in fetal monitoring care. A store-and-forward approach may offer a reasonable solution to this, with text messaging and voice calls addressing immediate issues and use of a robust remote decision-support network of trained professionals and an integrated referral mechanism (see Stroux et al (2016) for example)
In summary, mHealth systems bear a significant potential to provide remote perinatal care and prevent many fetal complications, by removing many barriers prevailing in LMICs. Notably, such approaches can empower the frontline healthcare workers (and perhaps even mothers) to learn, and even improve the systems, and ultimately alleviate the ‘brain drain’ in the medical field in LMICs (Cometto et al 2013).
Acknowledgments
GC, PR and NK acknowledge the support of the National Institutes of Health, the Fogarty International Center and the Eunice Kennedy Shriver National Institute of Child Health and Human Development, Grant No. 1R21HD084114–01 (Mobile Health Intervention to Improve Perinatal Continuum of Care in Guatemala). GC has financial interest in Alivecor Inc, and receives unrestricted funding from the company. GC also is the CTO of Mindchild Medical and has ownership interests in Mindchild Medical. CV was funded by a Fulbright Scholarship.
Footnotes
Any further distribution of this work must maintain attribution to the author(s) and the title of the work, journal citation and DOI.
References
- Abuhamad AZ and Chaoui R 2012. A Practical Guide to Fetal Echocardiography: Normal and Abnormal Hearts (Philadelphia, PA: Lippincott Williams & Wilkins; ) [Google Scholar]
- Adithya PC, Sankar R, Moreno WA and Hart S 2017. Trends in fetal monitoring through phonocardiography: Challenges and future directions Biomed. Signal Process. Control 33 289–305 [Google Scholar]
- Al-Zaben A and Al-Smadi A 2005. Extraction of foetal ECG by combination of singular value decomposition and neuro-fuzzy inference system Phys. Med. Biol 51 137. [DOI] [PubMed] [Google Scholar]
- Alfirevic Z and Neilson JP 1995. Doppler ultrasonography in high-risk pregnancies: systematic review with meta-analysis Am. J. Obstet. Gynecol 172 1379–87 [DOI] [PubMed] [Google Scholar]
- Alfirevic Z, Stampalija T and Dowswell T 2017. Fetal and umbilical Doppler ultrasound in high-risk pregnancies Cochrane Database Systemat. Rev ( 10.1002/14651858.CD007529.pub4) [DOI] [PMC free article] [PubMed]
- Allan LD 1986. Manual of Fetal Echocardiography (Lancaster: MTP Press; ) [Google Scholar]
- Allan LD 2000. A practical approach to fetal heart scanning Seminars in Perinatol vol 24 pp 324–30 [DOI] [PubMed] [Google Scholar]
- Allanson ER, Muller M and Pattinson RC 2015. Causes of perinatal mortality and associated maternal complications in a South African province: challenges in predicting poor outcomes BMC Pregnancy Childbirth 15 37. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Alves DS, Times VC, da Silva EMA, Melo PSA and de Araújo Novaes M 2019. Advances in obstetric telemonitoring: a systematic review Int. J. Med. Inform 134 104004. [DOI] [PubMed] [Google Scholar]
- Amer-Wåhlin I et al. 2001. Cardiotocography only versus cardiotocography plus ST analysis of fetal electrocardiogram for intrapartum fetal monitoring: a Swedish randomised controlled trial Lancet 358 534–8 [DOI] [PubMed] [Google Scholar]
- American College of Obstetricians 2000. ACOG practice bulletin. Antepartum fetal surveillance. Number 9, October 1999. Clinical management guidelines for obstetrician-gynecologists Int. J. Gynaecol. Obstetrics 68 175–85 [PubMed] [Google Scholar]
- Amin M, Mamun M, Hashim F and Husain H 2011. Separation of fetal electrocardiography (ECG) from composite ECG using adaptive linear neural network for fetal monitoring Int. J. Phys. Sci 6 5871–6 [Google Scholar]
- Archer N and Manning N 2009. Fetal Cardiology (Oxford: Oxford University Press; ) [Google Scholar]
- Assaleh K 2007. Extraction of fetal electrocardiogram using adaptive neuro-fuzzy inference systems IEEE Trans. Biomed. Eng 54 59–68 [DOI] [PubMed] [Google Scholar]
- Awiti AM, Shifferaw BM, Byamukama MB, Kizito R and Mwikirize C 2016. Design and implementation of an Android based digital fetoscope 2016 IEEE-EMBS Int. Conf. on Biomedical and Health Informatics (BHI) (IEEE; ) pp 152–5 [Google Scholar]
- Ayres-de Campos D, Spong CY, Chandraharan E and FIGO Intrapartum Fetal Monitoring Expert Consensus Panel 2015. FIGO consensus guidelines on intrapartum fetal monitoring: Cardiotocography Int. J. Gynecol. Obstet 131 13–24 [DOI] [PubMed] [Google Scholar]
- Bakker P, Colenbrander G, Verstraeten A and Van Geijn H 2004. The quality of intrapartum fetal heart rate monitoring Eur. J. Obstet. Gynecol. Reprod. Biol 116 22–7 [DOI] [PubMed] [Google Scholar]
- Banta HD and Thacker SB 2002. Electronic fetal monitoring Int. J. Technol. Assessment Health Care 18 762. [DOI] [PubMed] [Google Scholar]
- Baschat AA 2010. Ductus venosus Doppler for fetal surveillance in high-risk pregnancies Clin. Obste. Gynecol 53 858–68 [DOI] [PubMed] [Google Scholar]
- Becker DM, Tafoya CA, Becker SL, Kruger GH, Tafoya MJ and Becker TK 2016. The use of portable ultrasound devices in low-and middle-income countries: a systematic review of the literature Tropical Medi. Int. Health 21 294–311 [DOI] [PubMed] [Google Scholar]
- Behar J, Andreotti F, Zaunseder S, Oster J and Clifford GD 2016. A practical guide to non-invasive foetal electrocardiogram extraction and analysis Physiol. Meas 37 R1. [DOI] [PubMed] [Google Scholar]
- Behar J, Johnson A, Clifford GD and Oster J 2014. A comparison of single channel fetal ECG extraction methods. Ann. Biomed. Eng 42 1340–53 [DOI] [PubMed] [Google Scholar]
- Bennasar M, Martinez J, Gomez O, Bartrons J, Olivella A, Puerto B and Gratacós E 2010. Accuracy of four-dimensional spatiotemporal image correlation echocardiography in the prenatal diagnosis of congenital heart defects Ultrasound Obstet. Gynecol 36 458–64 [DOI] [PubMed] [Google Scholar]
- Bergström A, Fottrell E, Hopkins H, Lloyd D, Stevenson O and Willats P 2015. mHealth: can mobile technology improve health in low-and middle-income countries Technical report UCL public policy
- Berkley E, Chauhan SP, Abuhamad A and Society for Maternal-Fetal Medicine Publications Committee 2012. Doppler assessment of the fetus with intrauterine growth restriction. Am. J. Obstet. Gynecol 206 300–8 [DOI] [PubMed] [Google Scholar]
- Bernardes J, Costa-Pereira A, Ayres-de Campos D, Van Geijn H and Pereira-Leite L 1997. Evaluation of interobserver agreement of cardiotocograms Int. J. Gynecol. Obstet 57 33–7 [DOI] [PubMed] [Google Scholar]
- Blackwell SC, Grobman WA, Antoniewicz L, Hutchinson M and Bannerman CG 2011. Interobserver and intraobserver reliability of the NICHD 3-tier fetal heart rate interpretation system Am. J. Obstet. Gynecol 205 378–e1 [DOI] [PubMed] [Google Scholar]
- Blencowe H et al. 2016. National, regional and worldwide estimates of stillbirth rates in 2015, with trends from 2000: a systematic analysis Lancet Global Health 4 e98–e108 [DOI] [PubMed] [Google Scholar]
- Blix E et al. 2019. Intermittent auscultation fetal monitoring during labour: A systematic scoping review to identify methods, effects and accuracy PloS One 14 1–21 [DOI] [PMC free article] [PubMed] [Google Scholar]
- Blix E, Sviggum O, Koss KS and Øian P 2003. Inter-observer variation in assessment of 845 labour admission tests: comparison between midwives and obstetricians in the clinical setting and two experts BJOG: Int. J. Obstet. Gynaecol 110 1–5 [PubMed] [Google Scholar]
- Boamah EA, Asante K, Ae-Ngibise K, Kinney PL, Jack DW, Manu G, Azindow IT, Owusu-Agyei S and Wylie BJ 2014. Gestational age assessment in the Ghana randomized air pollution and health study (graphs): ultrasound capacity building, fetal biometry protocol development and ongoing quality control JMIR Res. Protocols 3 e77. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Britton N, Miller MA, Safadi S, Siegel A, Levine A and McCurdy M 2019. Tele-ultrasound in resource-limited settings: A systematic review Front. Public Health 7 244. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Cannie M, Jani J, Dymarkowski S and Deprest J 2006. Fetal magnetic resonance imaging: luxury or necessity? Ultrasound Obstet. Gynecol 27 471–6 [DOI] [PubMed] [Google Scholar]
- Cardo Medical 2020. Fetal monitors: monica novii wireless patch system (https://cardomedical.com/product-category/fetal-monitors/) (Accessed: 27 September 2020)
- Caserta L, Ruggeri Z, D’Emidio L, Coco C, Cignini P, Girgenti A, Mangiafico L and Giorlandino C 2008. Two-dimensional fetal echocardiography: where we are J. Prenatal Med 2 31. [PMC free article] [PubMed] [Google Scholar]
- Castillo E, Morales DP, Botella G, García A, Parrilla L and Palma AJ 2013. Efficient wavelet-based ECG processing for single-lead FHR extraction Digit. Signal Process 23 1897–909 [Google Scholar]
- Chaoui R, Hoffmann J and Heling K 2004. Three-dimensional (3D) and 4D color Doppler fetal echocardiography using spatio-temporal image correlation (STIC) Ultrasound Obstet. Gynecol 23 535–45 [DOI] [PubMed] [Google Scholar]
- Chudáček V, Spilka J, Lhotská L, Janků P, Koucký M, Huptych M and Burša M 2011. Assessment of features for automatic CTG analysis based on expert annotation 2011 Annual Int. Conf. of the IEEE Engineering in Medicine and Biology Society (IEEE; ) pp 6051–4 [DOI] [PubMed] [Google Scholar]
- Clark S, Sabey P, Minton S and Stoddard R 1987. Non-stress testing with acoustic stimulation Seventh Annual Meeting of The Society of Perinatal Obstetricians (Lake Buena Vista FL, USA) ( 10.1016/S0002-9378(89)80062-6) [DOI] [Google Scholar]
- Clifford GD, Silva I, Behar J and Moody GB 2014. Non-invasive fetal ECG analysis Physiol. Meas 35 1521. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Clifford G, Sameni R, Ward J, Robinson J, Pettigrew C and Wolfberg A 2009. Clinically accurate fetal ECG parameters acquired from maternal abdominal sensors Am. J. Obstet. Gynecol 201 S242. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Cometto G, Tulenko K, Muula AS and Krech R 2013. Health workforce brain drain: from denouncing the challenge to solving the problem PLoS Medicine 10 1–3 [DOI] [PMC free article] [PubMed] [Google Scholar]
- Cosmi E, Visentin S, Fanelli T, Mautone AJ and Zanardo V 2009. Aortic intima media thickness in fetuses and children with intrauterine growth restriction Obstet. Gynecol 114 1109–14 [DOI] [PubMed] [Google Scholar]
- Cruz-Martinez R, Figueras F, Hernandez-Andrade E, Oros D and Gratacos E 2011. Changes in myocardial performance index and aortic isthmus and ductus venosus Doppler in term, small-for-gestational age fetuses with normal umbilical artery pulsatility index Ultrasound Obstet. Gynecol 38 400–5 [DOI] [PubMed] [Google Scholar]
- Dawes G, Lobb M, Moulden M, Redman C and Wheeler T 1992. Antenatal cardiotocogram quality and interpretation using computers BJOG: An Int. J. Obstet. Gynecol 99 791–7 [DOI] [PubMed] [Google Scholar]
- de Onis M, Blössner M and Villar J 1998. Levels and patterns of intrauterine growth retardation in developing countries Eur. J. Clin. Nutr 52 S5–15 Suppl 1 [PubMed] [Google Scholar]
- Deb S et al. 2020. Performance of late pregnancy biometry for gestational age dating in low-income and middle-income countries: a prospective, multicountry, population-based cohort study from the who alliance for maternal and newborn health improvement (AMANHI) study group Lancet Global Health 8 e545–e554 [DOI] [PMC free article] [PubMed] [Google Scholar]
- Deng J, Gardener JE, Rodeck CH and Lees WR 1996. Fetal echocardiography in three and four dimensions Ultrasound Med. Biol 22 979–86 [DOI] [PubMed] [Google Scholar]
- Devane D, Lalor JG, Daly S, McGuire W, Cuthbert A and Smith V 2017. Cardiotocography versus intermittent auscultation of fetal heart on admission to labour ward for assessment of fetal wellbeing Cochrane Database Syst. Rev 26 CD005122. [DOI] [PMC free article] [PubMed] [Google Scholar]
- DeVore GR, Horenstein J, Siassi B and Platt LD 1987. Fetal echocardiography: VII. Doppler color flow mapping: A new technique for the diagnosis of congenital heart disease Am. J. Obstet. Gynecol 156 1054–64 [DOI] [PubMed] [Google Scholar]
- Dražančić A 2001. Antenatal care in developing countries What should be done? J. Perinatal Med 29 188–98 [DOI] [PubMed] [Google Scholar]
- East CE, Begg L, Colditz PB and Lau R 2014. Fetal pulse oximetry for fetal assessment in labour Cochrane Database Syst. Rev CD004075. [DOI] [PMC free article] [PubMed]
- Eden RD, Boehm FH and Haire M 1990. Assessment and Care of the Fetus: Physiological, Clinical and Medicolegal Principles (Norwalk, CN: Appleton & Lange; ) [Google Scholar]
- Evans DH 1989. Doppler Ultrasound: Physics, Instrumentation and Clinical Applications (Hoboken, NJ: Wiley; ) [Google Scholar]
- Eysenbach G and 2011. CONSORT-EHEALTH: improving and standardizing evaluation reports of web-based and mobile health interventions J. Medical Internet Res 13 e126. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Feroz A, Perveen S and Aftab W 2017. Role of mhealth applications for improving antenatal and postnatal care in low and middle income countries: a systematic review BMC Health Ser. Res 17 704. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Figueras F and Gardosi J 2011. Intrauterine growth restriction: new concepts in antenatal surveillance, diagnosis and management Am. J. Obstet. Gynecol 204 288–300 [DOI] [PubMed] [Google Scholar]
- Flenady V et al. 2011. Stillbirths: the way forward in high-income countries Lancet 377 1703–17 [DOI] [PubMed] [Google Scholar]
- Franklin HL. et al. Factors influencing referrals for ultrasound-diagnosed complications during prenatal care in five low and middle income countries. Reprod. Health. 2018;15:204. doi: 10.1186/s12978-018-0647-8. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Freeman RK, Anderson G and Dorchester W 1982. A prospective multi-institutional study of antepartum fetal heart rate monitoring: I. Risk of perinatal mortality and morbidity according to antepartum fetal heart rate test results Am. J. Obstet. Gynecol 143 771–7 [DOI] [PubMed] [Google Scholar]
- Freeman RK, Garite TJ, Nageotte MP and Miller LA 2012. Fetal Heart Rate Monitoring (Philadelphia, PA: Lippincott Williams & Wilkins; ) [Google Scholar]
- Galan HL, Ferrazzi E and Hobbins JC 2002. Intrauterine growth restriction (IUGR): biometric and Doppler assessment Prenatal Diagnosis: Published Affiliation with the Int. Society Prenatal Diagnos 22 331–7 [DOI] [PubMed] [Google Scholar]
- Gan KB, Zahedi E and Ali MAM 2009. Transabdominal fetal heart rate detection using NIR photopleythysmography: instrumentation and clinical results IEEE Trans. Biomed. Eng 56 2075–82 [DOI] [PubMed] [Google Scholar]
- Ganzevoort W et al. 2017. How to monitor pregnancies complicated by fetal growth restriction and delivery before 32 weeks: post-hoc analysis of truffle study Ultrasound Obstet. Gynecol 49 769–77 [DOI] [PubMed] [Google Scholar]
- Georgieva A et al. 2019. Computer-based intrapartum fetal monitoring and beyond: A review of the 2nd Workshop on Signal Processing and Monitoring in Labor (October 2017, Oxford, UK) Acta Obstetricia et Gynecologica Scandinavica 98 1207–17 [DOI] [PMC free article] [PubMed] [Google Scholar]
- Gittenberger-de Groot AC, Jongbloed MR, de Ruiter M C, Bartelings M and E P R 2019. Cardiac morphogenesis Fetal Cardiology eds Yagel S, Silverman NH and Gembruch U (Boca Raton, FL: Taylor and Francis; ) ch 1 pp 1–18 [Google Scholar]
- Godfrey M, Messing B, Cohen S, Valsky D and Yagel S 2012. Functional assessment of the fetal heart: A review Ultrasound Obstet. Gynecol 39 131–44 [DOI] [PubMed] [Google Scholar]
- Goldenberg RL et al. 2018. Routine antenatal ultrasound in low-and middle-income countries: first look—a cluster randomised trial BJOG: An Int. J. Obstet. Gynecol 125 1591–9 [DOI] [PMC free article] [PubMed] [Google Scholar]
- Goldenberg RL, McClure EM and Bann CM 2007. The relationship of intrapartum and antepartum stillbirth rates to measures of obstetric care in developed and developing countries Acta Obstetricia et Gynecologica Scandinavica 86 1303–9 [DOI] [PubMed] [Google Scholar]
- Goldenberg RL, McClure EM, Saleem S and Reddy UM 2010. Infection-related stillbirths Lancet 375 1482–90 [DOI] [PMC free article] [PubMed] [Google Scholar]
- Goldenberg RL, Saleem S, Pasha O, Harrison MS and Mcclure EM 2016. Reducing stillbirths in low-income countries Acta Obstetricia et Gynecologica Scandinavica 95 135–43 [DOI] [PubMed] [Google Scholar]
- Gomez-Roig MD, Mazarico E, Valladares E, Guirado L, Fernandez-Arias M and Vela A 2015. Aortic intima-media thickness and aortic diameter in small for gestational age and growth restricted fetuses PLoS One 10 e0126842. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Greenwold N, Wallace S, Prost A and Jauniaux E 2014. Implementing an obstetric ultrasound training program in rural Africa Int. J. Gynecol. Obstet 124 274–7 [DOI] [PubMed] [Google Scholar]
- Grivell RM, Alfirevic Z, Gyte GM and Devane D 2015. Antenatal cardiotocography for fetal assessment Cochrane Database Syst. Rev 2015 CD007863. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Groen RS, Leow JJ, Sadasivam V and Kushner AL 2011. Indications for ultrasound use in low-and middle-income countries Tropical Med. Int. Health 16 1525–35 [DOI] [PubMed] [Google Scholar]
- Hamelmann P, Vullings R, Kolen AF, Bergmans JW, van Laar JO, Tortoli P and Mischi M 2019. Doppler ultrasound technology for fetal heart rate monitoring: a review IEEE Trans. Ultrasonics, Ferroelectr. Freq. Control 67 226–38 [DOI] [PubMed] [Google Scholar]
- Han YW, Fardini Y, Chen C, Iacampo KG, Peraino VA, Shamonki JM and Redline RW 2010. Term stillbirth caused by oral fusobacterium nucleatum Obstet. Gynecol 115 442. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Hasan MJ, Hossain MS, Arafat Y, Karim GMR Ahmed Z and Mafi MS 2017. Mobile application can be an effective tool for reduction of maternal mortality. Int. J. Percept. Public Health 1 92–4 [Google Scholar]
- Hoffman JI and Christianson R 1978. Congenital heart disease in a cohort of 19,502 births with long-term follow-up Am. J. Cardiol 42 641–7 [DOI] [PubMed] [Google Scholar]
- Hon E. The instrumentation of fetal heart rate and fetal electrocardiography. I. A fetal heart monitor. Connecticut Med. 1960;24:289. [PubMed] [Google Scholar]
- Hon E and Hess O 1960. The clinical value of fetal electrocardiography Am. J. Obstet. Gynecol 79 1012–23 [DOI] [PubMed] [Google Scholar]
- Hon E and Lee S 1964. Averaging techniques in fetal electrocardiography Med. Electr. Biol. Eng 2 71–6 [DOI] [PubMed] [Google Scholar]
- Housseine N et al. 2019. Delphi consensus statement on intrapartum fetal monitoring in low-resource settings Int. J. Gynecol. Obstet 146 8–16 [DOI] [PMC free article] [PubMed] [Google Scholar]
- Housseine N, Punt MC, Browne JL, Meguid T, Klipstein-Grobusch K, Kwast BE, Franx A, Grobbee DE and Rijken MJ 2018. Strategies for intrapartum foetal surveillance in low-and middle-income countries: a systematic review PLoS One 13 e0206295. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Hruban L et al. 2015. Agreement on intrapartum cardiotocogram recordings between expert obstetricians J. Eval. Clin. Pract 21 694–702 [DOI] [PubMed] [Google Scholar]
- Ionescu C. The benefits of 3D-4D fetal echocardiography. Maedica. 2010;5:45. [PMC free article] [PubMed] [Google Scholar]
- Jaros R, Martinek R and Kahankova R 2018. Non-adaptive methods for fetal ECG signal processing: a review and appraisal Sensors 18 3648. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Jatmiko W, Ma’Sum MA, Isa SM, Imah E, Rahmatullah R and Wiweko B 2015. Developing smart telehealth system in Indonesia: Progress and challenge 2015 Int. Conf. on Advanced Computer Science and Information Systems (ICACSIS) (IEEE; ) pp 29–36 [Google Scholar]
- Jauniaux E and Prefumo F 2016. Fetal heart monitoring in labour: from Pinard to artificial intelligence BJOG: An Int. J. Obstet. Gynecol 123 870. [DOI] [PubMed] [Google Scholar]
- Jezewski J, Matonia A, Kupka T, Roj D and Czabanski R 2012. Determination of fetal heart rate from abdominal signals: evaluation of beat-to-beat accuracy in relation to the direct fetal electrocardiogram Biomedizinische Technik/Biomed. Eng 57 383–94 [DOI] [PubMed] [Google Scholar]
- Jezewski J, Wrobel J, Matonia A, Horoba K, Martinek R, Kupka T and Jezewski M 2017. Is abdominal fetal electrocardiography an alternative to Doppler ultrasound for FHR variability evaluation? Front. Physiol 8 305. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Kähler C, Grimm B, Schleussner E, Schneider A, Schneider U, Nowak H, Vogt L and Seewald H-J 2001. The application of fetal magnetocardiography (FMCG) to investigate fetal arrhythmias and congenital heart defects (CHD) Prenatal Diagnosis: Published in Affiliation with the Int. Society for Prenatal Diagnosis 21 176–82 [DOI] [PubMed] [Google Scholar]
- Kahankova R, Martinek R, Jaros R, Behbehani K, Matonia A, Jezewski M and Behar JA 2019. A review of signal processing techniques for non-invasive fetal electrocardiography IEEE Rev. Biomed. Eng 13 51–73 [DOI] [PubMed] [Google Scholar]
- Kamala BA, Ersdal HL, Dalen I, Abeid MS, Ngarina MM, Perlman JM and Kidanto HL 2018a. Implementation of a novel continuous fetal Doppler (Moyo) improves quality of intrapartum fetal heart rate monitoring in a resource-limited tertiary hospital in tanzania: An observational study PLoS One 13 e0205698. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Kamala BA, Kidanto HL, Wangwe PJ, Dalen I, Mduma ER, Perlman JM and Ersdal HL 2018b. Intrapartum fetal heart rate monitoring using a handheld Doppler versus Pinard stethoscope: a randomized controlled study in Dar Es Salaam Int. J. Women’s Health 10 341. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Kapaya H, Jacques R and Anumba D 2018. Comparison of diurnal variations, gestational age and gender related differences in fetal heart rate (FHR) parameters between appropriate-for-gestational-age (AGA) and small-for-gestational-age (SGA) fetuses in the home environment PLoS One 13 1–12 [DOI] [PMC free article] [PubMed] [Google Scholar]
- Kariniemi V and Hukkinen K 1977. Quantification of fetal heart rate variability by magnetocardiography and direct electrocardiography Am. J. Obstet. Gynecol 128 526–30 [DOI] [PubMed] [Google Scholar]
- Katebi N, Marzbanrad F, Stroux L, Valderrama CE and Clifford GD 2020. Unsupervised hidden semi-Markov model for automatic beat onset detection in 1D Doppler ultrasound Physiol. Meas 41 085007. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Khan NH, Tegnander E, Dreier JM, Eik-Nes S, Torp H and Kiss G 2016. Automatic measurement of the fetal abdominal section on a portable ultrasound machine for use in low and middle income countries 2016 IEEE Int. Symp. (IUS) (IEEE; ) pp 1–4 [Google Scholar]
- Kiefer-Schmidt I, Lim M, Wacker-Gußmann A, Ortiz E, Abele H, Kagan KO, Kaulitz R, Wallwiener D and Preissl H 2012. Fetal magnetocardiography (FMCG): Moving forward in the establishment of clinical reference data by advanced biomagnetic instrumentation and analysis J. Perinatal Med 40 277–86 [DOI] [PubMed] [Google Scholar]
- Kim ET, Singh K, Moran A, Armbruster D and Kozuki N 2018. Obstetric ultrasound use in low and middle income countries: a narrative review Reprod. Health 15 129 [DOI] [PMC free article] [PubMed] [Google Scholar]
- Kimberly HH et al. 2010. Focused maternal ultrasound by midwives in rural Zambia Ultrasound Med. Biol 36 1267–72 [DOI] [PubMed] [Google Scholar]
- Kohler TR and Sumner DS 2014. Vascular laboratory: Arterial physiologic assessment Rutherford’s Vascular Surgery eds Cronenwett JL and Johnston KW (Philadelphia, PA: Elsevier Health Sciences; ) ch 15 pp 214–29 [Google Scholar]
- Kovács F, Horváth C, Balogh AT and Hosszú G 2011. Fetal phonocardiography-past and future possibilities Comput. Methods Programs Biomed 104 19–25 [DOI] [PubMed] [Google Scholar]
- Kozuki N. et al. Accuracy of home-based ultrasonographic diagnosis of obstetric risk factors by primary-level health workers in rural Nepal. Obstet. Gynecol. 2016;128:604. doi: 10.1097/AOG.0000000000001558. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Lafontan SR, Kidanto HL, Ersdal HL, Mbekenga CK and Sundby J 2019. Perceptions and experiences of skilled birth attendants on using a newly developed strap-on electronic fetal heart rate monitor in Tanzania BMC Preg. Childbirth 19 165. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Lai K-C and Shynk JJ 2002. A successive cancellation algorithm for fetal heart-rate estimation using an intrauterine ECG signal IEEE Trans. Biomed. Eng 49 943–54 [DOI] [PubMed] [Google Scholar]
- Lakhno I. Autonomic imbalance captures maternal and fetal circulatory response to pre-eclampsia. Clin. Hypertension. 2017;23:5. doi: 10.1186/s40885-016-0061-x. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Larks SD and Longo LD 1962. Electrocardiographic studies of the fetal heart during delivery Obstet. Gynecol 19 740–7 [PubMed] [Google Scholar]
- Larks S and Larks G 1966. Components of the fetal electrocardiogram and intrauterine electrical axis: Quantitative data Neonatology 10 140–52 [DOI] [PubMed] [Google Scholar]
- Lawn JE et al. 2014. Every newborn: Progress, priorities and potential beyond survival Lancet 384 189–205 [DOI] [PubMed] [Google Scholar]
- Lawn JE, Lee AC, Kinney M, Sibley L, Carlo WA, Paul VK, Pattinson R and Darmstadt GL 2009. Two million intrapartum-related stillbirths and neonatal deaths: where, why and what can be done? Int. J. Gynecol. Obstet 107 S5–S19 [DOI] [PubMed] [Google Scholar]
- Lee AC et al. 2013. National and regional estimates of term and preterm babies born small for gestational age in 138 low-income and middle-income countries in 2010 Lancet Global Health 1 e26–e36 [DOI] [PMC free article] [PubMed] [Google Scholar]
- Lees CC et al. 2015. 2 year neurodevelopmental and intermediate perinatal outcomes in infants with very preterm fetal growth restriction (TRUFFLE): a randomised trial Lancet 385 2162–72 [DOI] [PubMed] [Google Scholar]
- Liston R et al. 2007. Fetal health surveillance: antepartum and intrapartum consensus guideline J. Obstet. Gynecol. Canada 29 S3–S4 [PubMed] [Google Scholar]
- Lopez AD, AbouZahr C, Shibuya K and Gollogly L 2007. Keeping count: Births, deaths and causes of death Lancet 370 1744–6 [DOI] [PubMed] [Google Scholar]
- Mäkikallio K, Räsänen J, Mäkikallio T, Vuolteenaho O and Huhta J 2008. Human fetal cardiovascular profile score and neonatal outcome in intrauterine growth restriction Ultrasound Obstet. Gynecol 31 48–54 [DOI] [PubMed] [Google Scholar]
- Mahdizadeh J, Bouraghi H, Panahi SSGS, Mohammadpour A, Shargh AK, Mojarad MR and Kahouei M 2019. A theory map of the causes of perinatal death in a developing country Crescent J. Medical Biol. Sci 6 237–41 [Google Scholar]
- Mahomed K, Nyoni R, Mulambo T, Kasule J and Jacobus E 1994. Randomised controlled trial of intrapartum fetal heart rate monitoring BMJ 308 497–500 [DOI] [PMC free article] [PubMed] [Google Scholar]
- Malhotra N, Malhotra J, Mathur V, Tomar S, Singh K, Rao J, Gupta S and Malhotra N 2014. Antenatal assessment of fetal well-being Ultrasound in Obstetrics and Gynecology ed Malhotra N, Kumar P, Panchal S, Shah K, Acharya P and Malhotra J (London, UK: JP Medical Ltd; ) ch 26 pp 227–41 [Google Scholar]
- Mari G and Hanif F 2008. Fetal Doppler: umbilical artery, middle cerebral artery and venous system Seminars Perinatol 32 253–7 [DOI] [PubMed] [Google Scholar]
- Marieb EN and Hoehn K 2007. Human Anatomy Physiology (Boston, MA: Pearson Education; ) [Google Scholar]
- Martens SM, Rabotti C, Mischi M and Sluijter RJ 2007. A robust fetal ECG detection method for abdominal recordings Physiol. Meas 28 373. [DOI] [PubMed] [Google Scholar]
- Martinez B et al. 2017. Agile development of a smartphone app for perinatal monitoring in a resource-constrained setting J. Health Inform. Dev. Count 11 (http://www.jhidc.org/index.php/jhidc/article/view/158/212) [PMC free article] [PubMed] [Google Scholar]
- Martinez B. et al. mHealth intervention to improve the continuum of maternal and perinatal care in rural Guatemala: a pragmatic, randomized controlled feasibility trial. Reprod. Health. 2018;15:120. doi: 10.1186/s12978-018-0554-z. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Martis R, Emilia O, Nurdiati DS and Brown J 2017. Intermittent auscultation (IA) of fetal heart rate in labour for fetal well-being Cochrane Database Syst. Rev CD008680. [DOI] [PMC free article] [PubMed]
- Marzbanrad F, Stroux L and Clifford GD 2018. Cardiotocography and beyond: a review of one-dimensional Doppler ultrasound application in fetal monitoring Physiol. Meas 39 08TR01. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Mdoe PF, Ersdal HL, Mduma E, Moshiro R, Dalen I, Perlman JM and Kidanto H 2018a. Randomized controlled trial of continuous Doppler versus intermittent fetoscope fetal heart rate monitoring in a low-resource setting Int. J. Gynecol. Obstet 143 344–50 [DOI] [PubMed] [Google Scholar]
- Mdoe PF, Ersdal HL, Mduma ER, Perlman JM, Moshiro R, Wangwe PT and Kidanto H 2018b. Intermittent fetal heart rate monitoring using a fetoscope or hand held Doppler in rural Tanzania: a randomized controlled trial BMC Preg. Childbirth 18 1–8 [DOI] [PMC free article] [PubMed] [Google Scholar]
- Miller FC, Pearse KE and Paul RH 1984. Fetal heart rate pattern recognition by the method of auscultation Obstet. Gynecol 64 332–6 [PubMed] [Google Scholar]
- Mone F, McAuliffe FM and Ong S 2015. The clinical application of Doppler ultrasound in obstetrics Obstet. Gynaecol 17 13–19 [Google Scholar]
- Morgenstern M, Sargent JD and Hanewinkel R 2009. Relation between socioeconomic status and body mass index: evidence of an indirect path via television use Arch. Pediatr. Adolescent Med 163 731–8 [DOI] [PMC free article] [PubMed] [Google Scholar]
- Morris R, Say R, Robson S, Kleijnen J and Khan K 2012. Systematic review and meta-analysis of middle cerebral artery Doppler to predict perinatal wellbeing Eur. J. Obstet. Gynecol. Reprod. Biol 165 141–55 [DOI] [PubMed] [Google Scholar]
- Munabi-Babigumira S, Glenton C, Willcox M and Nabudere H 2019. Ugandan health workers’ and mothers’ views and experiences of the quality of maternity care and the use of informal solutions: a qualitative study PLoS One 14 e0213511. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Neilson JP. Ultrasound for fetal assessment in early pregnancy. Cochrane Database Syst. Rev. 1998. p. CD000182. [DOI] [PubMed]
- Nicolaides K, Rizzo G, Hecher K and Ximenes R 2002. Doppler in obstetrics The Fetal Medicine Foundation; London, UK [Google Scholar]
- Norén H, Amer-Wåhlin I, Hagberg H, Herbst A, Kjellmer I, Marşál K, Olofsson P and Rosén KG 2003. Fetal electrocardiogram in labor and neonatal outcome: data from the Swedish randomized controlled trial on intrapartum fetal monitoring Am. J. Obstet. Gynecol 188 183–92 [DOI] [PubMed] [Google Scholar]
- Norén H, Blad S, Carlsson A, Flisberg A, Gustavsson A, Lilja H, Wennergren M and Hagberg H 2006. STAN in clinical practice? The outcome of 2 years of regular use in the city of Gothenburg Am. J. Obstet. Gynecol 195 7–15 [DOI] [PubMed] [Google Scholar]
- O’Dwyer V et al. 2014. Defining the residual risk of adverse perinatal outcome in growth-restricted fetuses with normal umbilical artery blood flow Am. J. Obstet. Gynecol 211 420–e1 [DOI] [PubMed] [Google Scholar]
- Ome-Kaius M, Karl S, Wangnapi RA, Bolnga JW, Mola G, Walker J, Mueller I, Unger HW and Rogerson SJ 2017. Effects of plasmodium falciparum infection on umbilical artery resistance and intrafetal blood flow distribution: a Doppler ultrasound study from Papua New Guinea Malaria J 16 35. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Papp Z and Fekete T 2003. The evolving role of ultrasound in obstetrics/gynecology practice Int. J. Gynecol. Obstet 82 339–46 [DOI] [PubMed] [Google Scholar]
- Park Y, Lee K, Youn D, Kim N, Kim W and Park S 1992. On detecting the presence of fetal R-wave using the moving averaged magnitude difference algorithm IEEE Trans. Biomed. Eng 39 868–71 [DOI] [PubMed] [Google Scholar]
- Pattinson R, Kerber K, Waiswa P, Day LT, Mussell F, Asiruddin S, Blencowe H and Lawn JE 2009. Perinatal mortality audit: counting, accountability and overcoming challenges in scaling up in low-and middle-income countries Int. J. Gynecol. Obstet 107 S113–S122 (Supplement) [DOI] [PubMed] [Google Scholar]
- Peters M, Crowe J, Piéri JF, Quartero H, Hayes-Gill B, James D, Stinstra J and Shakespeare S 2001. Monitoring the fetal heart non-invasively: A review of methods J. Perinatal Med 29 408–16 [DOI] [PubMed] [Google Scholar]
- Pildner von SS, Boulesteix A, Lederer C, Grunow S, Schiermeier S, Hatzmann W, Schneider KM and Daumer M 2013. What is the ‘normal’ fetal heart rate? Peer J 1 e82. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Plotkin M, Kamala B, Ricca J, Fogarty L, Currie S, Kidanto H and Wheeler SB 2019. Systematic review of Doppler for detecting intrapartum fetal heart abnormalities and measuring perinatal mortality in low-and middle-income countries Int. J. Gynecol. Obstet 148 145–56 [DOI] [PMC free article] [PubMed] [Google Scholar]
- Rahman H, Renjhen P and Dutta S 2012. Reliability of admission cardiotocography for intrapartum monitoring in low resource setting Nigerian Med. J.: J. Nigeria Medical Association 53 145. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Reece EA, Gabrielli S, Degennaro N and Hobbins JC 1989. Dating through pregnancy: a measure of growing up Obstet. Gynecol. Survey 44 544–55 [DOI] [PubMed] [Google Scholar]
- Royal College of Obstetricians and Gynaecologists. Small-for-Gestational-Age Fetus, Investigation and Management. Green-top Guideline No. 31. 2002. [DOI] [PubMed]
- Sameni R and Clifford GD 2010. A review of fetal ECG signal processing; issues and promising directions Open Pacing, Electrophys. Ther. J 3 4–20 [DOI] [PMC free article] [PubMed] [Google Scholar]
- Sameni R, Jutten C and Shamsollahi MB 2006. What ICA provides for ECG processing: Application to noninvasive fetal ECG extraction 2006 IEEE Int. Symp. on Signal Processing and Information Technology (Vancouver, BC) pp 656–61 [Google Scholar]
- Sameni R, Shamsollahi MB, Jutten C and Clifford GD 2007. A nonlinear Bayesian filtering framework for ECG denoising IEEE Trans. Biomed. Eng 54 2172–85 [DOI] [PubMed] [Google Scholar]
- Save the Children. Technical Report. World Health Organization; Washington DC: 2001. State of the world’s newborns. [Google Scholar]
- Schneider K and Maternal Fetal Medicine Study Group 2014. S1-guideline on the use of CTG during pregnancy and labor Geburtshilfe und Frauenheilkunde 74 721–32 [DOI] [PMC free article] [PubMed] [Google Scholar]
- Shakespeare S, Crowe J, Hayes-Gill B, Bhogal K and James D 2001. The information content of Doppler ultrasound signals from the fetal heart Med. Biol. Eng. Comput 39 619–26 [DOI] [PubMed] [Google Scholar]
- Shao M, Barner KE and Goodman MH 2004. An interference cancellation algorithm for noninvasive extraction of transabdominal fetal electroencephalogram (TaFEEG) IEEE Trans. Biomed. Eng 51 471–83 [DOI] [PubMed] [Google Scholar]
- Signore C, Freeman RK and Spong CY 2009. Antenatal testing–a reevaluation: Executive summary of a Eunice Kennedy Shriver National Institute of Child Health and Human Development workshop Obstet. Gynecol 113 687. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Signorini MG, Pini N, Malovini A, Bellazzi R and Magenes G 2020. Integrating machine learning techniques and physiology based heart rate features for antepartum fetal monitoring Comput. Methods Programs Biomed 185 105015 [DOI] [PubMed] [Google Scholar]
- Smith CV, Phelan JP, Paul RH and Broussard P 1985. Fetal acoustic stimulation testing: A retrospective experience with the fetal acoustic stimulation test Am. J. Obstet. Gynecol 153 567–8 [DOI] [PubMed] [Google Scholar]
- Smith CV, Phelan JP, Platt LD, Broussard P and Paul RH 1986. Fetal acoustic stimulation testing: II. A randomized clinical comparison with the nonstress test Am. J. Obstet. Gynecol 155 131–4 [DOI] [PubMed] [Google Scholar]
- Smith V, Nair A, Warty R, Sursas JA, da Silva Costa F and Wallace EM 2019. A systematic review on the utility of non-invasive electrophysiological assessment in evaluating for intra uterine growth restriction BMC Preg. Childbirth 19 230. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Sondaal SFV, Browne JL, Amoakoh-Coleman M, Borgstein A, Miltenburg AS, Verwijs M and Klipstein-Grobusch K 2016. Assessing the effect of mHealth interventions in improving maternal and neonatal care in low-and middle-income countries: a systematic review PLoS One 11 e0154664. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Soothill P, Ajayi R, Campbell S and Nicolaides K 1993. Prediction of morbidity in small and normally grown fetuses by fetal heart rate variability, biophysical profile score and umbilical artery Doppler studies BJOG: An Int. J. Obstet. Gynecol 100 742–5 [DOI] [PubMed] [Google Scholar]
- Stewart KA, Navarro SM, Kambala S, Tan G, Poondla R, Lederman S, Barbour K and Lavy C 2020. Trends in ultrasound use in low and middle income countries: a systematic review Int. J. Maternal Child Health AIDS 9 103. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Stock UA and Vacanti JP 2001. Cardiovascular physiology during fetal development and implications for tissue engineering Tissue Eng 7 1–7 [DOI] [PubMed] [Google Scholar]
- Strasburger JF, Cheulkar B and Wakai RT 2008. Magnetocardiography for fetal arrhythmias Heart Rhythm 5 1073–6 [DOI] [PMC free article] [PubMed] [Google Scholar]
- Stroux L, Martinez B, Coyote E, King N, Hall-Clifford R, Rohloff P and Clifford GD 2016. An mHealth monitoring system for traditional birth attendant-led antenatal risk assessment in rural Guatemala J. Med. Eng. Technol 40 356–71 [DOI] [PMC free article] [PubMed] [Google Scholar]
- Stroux L, Redman CW, Georgieva A, Payne SJ and Clifford GD 2017. Doppler-based fetal heart rate analysis markers for the detection of early intrauterine growth restriction Acta Obstetricia et Gynecologica Scandinavica 96 1322–9 [DOI] [PMC free article] [PubMed] [Google Scholar]
- Tapia-Conyer R. et al. Improving perinatal care in the rural regions worldwide by wireless enabled antepartum fetal monitoring: a demonstration project. Int. J. Telemed. App. 2015;2015:3. doi: 10.1155/2015/794180. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Todros T, Preve C, Plazzotta C, Biolcati M and Lombardo P 1996. Fetal heart rate tracings observers versus computer assessment Euro. J. Obstet. Gynecol. Reprod. Biol 68 83–6 [DOI] [PubMed] [Google Scholar]
- Trudinger B, Cook C, Jones L and Giles W 1986. A comparison of fetal heart rate monitoring and umbilical artery waveforms in the recognition of fetal compromise BJOG: An Int. J. Obstet. Gynecol 93 171–5 [DOI] [PubMed] [Google Scholar]
- UNICEF, World Health Organization, the World Bank Group, and the United Nations 2019. Levels and trends in child mortality Report 2019 (https://data.unicef.org/resources/levels-and-trends-in-child-mortality/) (Accessed: 23 January 2020)
- UNICEF, World Health Organization, World Bank, and UN DESA Population Division 2018. Low birthweight. country, reginal and global estimates (https://data.worldbank.org/indicator/SH.DYN.NMRT?most_recent_value_desc=true) (Accessed: 21 January 2020)
- Unterscheider J et al. 2013. Optimizing the definition of intrauterine growth restrictionthe multicenter prospective PORTO study Am. J. Obstet. Gynecol 208 290–e1 [DOI] [PubMed] [Google Scholar]
- Uquillas KR, Grubbs BH, Prosper AE, Chmait RH, Grant EG and Walker DK 2017. Doppler us in the evaluation of fetal growth and perinatal health Radiographics 37 1831–8 [DOI] [PubMed] [Google Scholar]
- Várady P, Wildt L, Benyó Z and Hein A 2003. An advanced method in fetal phonocardiography Computat. Methods Programs Biomed 71 283–96 [DOI] [PubMed] [Google Scholar]
- Valderrama CE, Marzbanrad F, Hall-Clifford R, Rohloff P and Clifford GD 2020a. A proxy for detecting IUGR based on gestational age estimation in a Guatemalan rural population Front. Artific. Intell 3 56. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Valderrama CE, Marzbanrad F, Juarez M, Hall-Clifford R, Rohloff P and Clifford GD 2020b. Estimating birth weight from observed postnatal weights in a Guatemalan highland community Physiol. Meas 41 025008. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Valderrama CE, Marzbanrad F, Stroux L and Clifford GD 2017. Template-based quality assessment of the Doppler ultrasound signal for fetal monitoring Front. Physiol 8 511. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Valderrama CE, Marzbanrad F, Stroux L, Martinez B, Hall-Clifford R, Liu C, Katebi N, Rohloff P and Clifford GD 2018. Improving the quality of point of care diagnostics with real-time machine learning in low literacy LMIC settings Proc. 1st ACM Conf. on Computing and Sustainable Societies, COMPASS ‘18 (Menlo Park and San Jose, CA, USA) ( 10.1145/3209811.3209815) [DOI] [Google Scholar]
- Valderrama CE, Stroux L, Katebi N, Paljug E, Hall-Clifford R, Rohloff P, Marzbanrad F and Clifford GD 2019. An open source autocorrelation-based method for fetal heart rate estimation from one-dimensional Doppler ultrasound Physiol. Meas 40 025005. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Velayo CL, Funamoto K, Silao JNI, Kimura Y and Nicolaides K 2017. Evaluation of abdominal fetal electrocardiography in early intrauterine growth restriction Front. Physiol 8 437. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Vijgen SM et al. 2011. Cost-effectiveness of cardiotocography plus ST analysis of the fetal electrocardiogram compared with cardiotocography only Acta Obstetricia et Gynecologica Scandinavica 90 772–8 [DOI] [PubMed] [Google Scholar]
- Vogel JP, Souza JP and Gülmezoglu AM 2013. Patterns and outcomes of induction of labour in Africa and Asia: a secondary analysis of the WHO global survey on maternal and neonatal health PLoS One 8 e65612. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Wakai RT 2004. Assessment of fetal neurodevelopment via fetal magnetocardiography Exp. Neurol 190 65–71 [DOI] [PubMed] [Google Scholar]
- Wakai RT 2014. The atomic magnetometer: a new era in biomagnetism AIP Conf. Proc 1626 6–54 [Google Scholar]
- Wall SN et al. 2010. Reducing intrapartum-related neonatal deaths in low-and middle-income countries—what works? Sem. Perinatol 34 395–407 [DOI] [PubMed] [Google Scholar]
- Wang H et al. 2016. Global, regional, national and selected subnational levels of stillbirths, neonatal, infant and under-5 mortality, 1980–2015: a systematic analysis for the global burden of disease study 2015 Lancet 388 1725–74 [DOI] [PMC free article] [PubMed] [Google Scholar]
- Warrick PA, Hamilton EF, Kearney RE and Precup D 2012. A machine learning approach to the detection of fetal hypoxia during labor and delivery AI Mag 33 79 [Google Scholar]
- Whitworth M, Bricker L and Mullan C 2015. Ultrasound for fetal assessment in early pregnancy Cochrane Database Syst. Rev CD007058. [DOI] [PMC free article] [PubMed]
- Wolfberg A 2012. The future of fetal monitoring Rev. Obstet. Gynecol 5 e132–6 [PMC free article] [PubMed] [Google Scholar]
- World Health Organization 1996. Perinatal mortality: a listing of available information (https://apps.who.int/iris/handle/10665/60977) (Accessed: 11 January 2020)
- World Health Organization 2014a. Global status report on noncommunicable diseases 2014 (https://www.who.int/nmh/publications/ncd-status-report-2014/en/) (Accessed: 11 January 2020)
- World Health Organization 2014b. WHO compendium of innovative health technologies for low-resource settings (https://www.who.int/medical_devices/innovation/compendium/en/) (Accessed: 11 January 2020)
- World Health Organization 2016a. Maternal and perinatal health (https://www.who.int/maternal_child_adolescent/topics/maternal/maternal_perinatal/en/) (Accessed: 20 February 2017)
- World Health Organization 2016b. The neglected tragedy of stillbirths (https://www.who.int/reproductivehealth/topics/maternal_perinatal/stillbirth/Lancet-series/en/) (Accessed: 11 January 2020)
- World Health Organization 2016c. WHO recommendations on antenatal care for a positive pregnancy experience (https://www.who.int/reproductivehealth/publications/maternal_perinatal_health/anc-positive-pregnancy-experience/en/) (Accessed: 11 October 2019) [PubMed]
- World Health Organization 2018a. WHO recommendation on continuous cardiotocography during labour (https://extranet.who.int/rhl/topics/preconception-pregnancy-childbirth-and-postpartum-care/care-during-childbirth/care-during-labour-1st-stage/who-recommendation-intermittent-fetal-heart-rate-auscultation-during-labour) (Accessed: 4 May 2020)
- World Health Organization 2018b. WHO recommendation on intermittent fetal heart rate auscultation during labour (https://extranet.who.int/rhl/topics/preconception-pregnancy-childbirth-and-postpartum-care/care-during-childbirth/care-during-labour-1st-stage/who-recommendation-continuous-cardiotocography-during-labour) (Accessed: 4 May 2020)
- Wu S, Shen Y, Zhou Z, Lin L, Zeng Y and Gao X 2013. Research of fetal ECG extraction using wavelet analysis and adaptive filtering Comput. Biol. Med 43 1622–7 [DOI] [PubMed] [Google Scholar]
- Wyatt J 2008. Appropriate medical technology for perinatal care in low-resource countries Ann. Trop. Paediatr 28 243–51 [DOI] [PubMed] [Google Scholar]
- Yakoob MY, Menezes EV, Soomro T, Haws RA, Darmstadt GL and Bhutta ZA 2009. Reducing stillbirths: behavioural and nutritional interventions before and during pregnancy BMC Preg. Childbirth 9 S3. [DOI] [PMC free article] [PubMed] [Google Scholar]