Diagnostic accuracy of antibody-based rapid diagnostic tests in detecting coronavirus disease 2019: systematic review

Tiara Josephine Gracienta 1, Ryan Herardi 1, Frans Santosa 1, Taufiq Fredrik Pasiak 1, Yanto Sandy Tjang 1,
PMCID: PMC9266800  PMID: 35832707

Abstract

Introduction

The rapid transmission of coronavirus disease 2019 (COVID-19) requires a fast, accurate, and affordable detection method. Despite doubts of their diagnostic accuracy, rapid diagnostic tests (RDTs) are used worldwide due to their practicality. This systematic review aims to determine the diagnostic accuracy of antibody-based RDTs in detecting COVID-19.

Material and methods

A literature search was carried out on five journal databases using the PRISMA-P 2015 method. We included all studies published up to February 2021. The risk of bias was evaluated using the Joanna Briggs Institute (JBI) Critical Appraisal Checklist for Diagnostic Test Accuracy Studies. Data regarding peer-review status, study design, test kit information, immunoglobulin class, target antigen, and the number of samples were extracted and tabulated. We estimated the pooled sensitivity, specificity, positive predictive value (PPV), and negative predictive value (NPV) with a 95% confidence interval.

Results

Thirty-three studies met the eligibility criteria. The pooled data results showed that the combined detection method of IgM or IgG had the highest sensitivity and NPV, which were 73.41% (95% CI: 72.22–74.57) and 75.34% (95% CI: 74.51–76.16), respectively. The single IgG detection method had the highest specificity and PPV of 96.68% (95% CI: 96.25–97.07) and 95.97% (95% CI: 95.47–96.42%), respectively.

Conclusions

Antibody-based RDTs are not satisfactory as primary diagnostic tests but have utility as a screening tool.

Keywords: diagnostic accuracy, COVID-19, rapid diagnostic test,

Introduction

The World Health Organization officially declared coronavirus disease 2019 (COVID-19) a pandemic in March 2020. COVID-19, caused by severe acute respiratory syndrome-coronavirus-2 (SARS-CoV-2) infection, has an alarmingly high infection rate [1]. There are several tests to detect SARS-CoV-2, classified as molecular and serological tests. The RT-PCR method is the referral test for COVID-19, but it takes a long time to obtain the result [2]. A rapid diagnostic test (RDT) is the fastest alternative that detects either SARS-CoV-2 antigens or IgG/IgM antibodies [3]. Most antibody-based RDTs use blood (peripheral/serum) as a sample, which is faster and easier to obtain than the antigen-based RDTs. They are also preferred because other methods such as ELISA and CLIA require laboratory facilities.

RDTs for COVID-19 are still in development, and their reported diagnostic accuracy varies, possibly because of the differences of the manufacturers [4]. Despite that, however, antibody-based RDTs are popular and used worldwide, while antigen-based RDTs have been rarely used [510]. Limited resources, a high infection rate, and a high population compelled their use over other test methods. However, some publications have doubts about their diagnostic accuracy [8, 1113]. RDTs are not recommended for patient care but as a screening tool for suspects [14, 15]. SARS-CoV-2 antibody detection can provide fast results to complement the time-consuming RT-PCR test [16, 17]. Still, RDTs need further research to ascertain their performance and accuracy.

Therefore, this systematic review aims to determine the diagnostic accuracy of antibody-based RDTs in detecting COVID-19.

Material and methods

Search strategy

We conducted a literature search in five different databases, including one preprint source – PubMed, Prime PubMed, Science Direct, MedRxiv, and Google Scholar – according to the PRISMA-P 2015 method. We included all studies published up to February 2021. Preprints were updated as soon as they were published. The search strategy included two combinations: “COVID-19 OR SARS-CoV-2” AND “rapid diagnostic test OR rapid test” AND “diagnostic accuracy” and “COVID-19 OR SARS-CoV-2” AND “rapid diagnostic test OR rapid test” AND “sensitivity” AND “specificity”. Several articles were also obtained from the references of included studies.

Eligibility criteria

All articles were screened by titles and abstracts. Relevant articles were read in full and screened for duplications. We included all studies that report the diagnostic accuracy of antibody-based RDTs using the lateral flow immunoassay (LFIA) method, which involves the sensitivity, specificity, PPV, and NPV. We excluded reviews, case reports, editorials, articles not in English, and articles that lacked adequate test population data. We assessed the risk of bias with the Joanna Briggs Institute Critical Appraisal Checklist for Diagnostic Test Accuracy Studies. All potential articles (n = 33) met our eligibility criteria. The search process can be seen in Figure 1.

Figure 1.

Figure 1

PRISMA flow of literature search process

Data extraction and analysis

The risk of bias was evaluated using the Joanna Briggs Institute (JBI) Critical Appraisal Checklist for Diagnostic Test Accuracy Studies. Two reviewers (TJG, RH) performed an independent assessment of included articles. Conflicts were resolved through consensus, and unresolved ones were decided by the third reviewer (YST). Data regarding peer-review status, study design, test kit information, immunoglobulin class, target antigen, and the number of samples were extracted and tabulated. We estimated the pooled sensitivity, specificity, positive predictive value (PPV), and negative predictive value (NPV) with a 95% confidence interval.

Results

Data extraction

We found 1,751 articles from the literature search and only 33 met the eligibility criteria. Eight of the 33 articles were preprints. Out of thirty-three, 16 were cohort or cross-sectional studies, while the remaining 17 were case controls. Twenty-four used commercially registered RDT kits. Twenty studies used the single IgM detection method, 21 used single IgG detection, 19 used combined IgM or IgG detection, and 15 used a combined IgM and IgG method. Three studies used RDTs that targeted spike protein, two targeted nucleocapsid protein, 7 targeted both proteins, and 21 did not disclose the target antigen (Table I).

Table I.

Data extraction

Criteria No. of articles No. of samples
Peer-reviewed:
 Yes 25 5,334
 No 8 5,472
Study design:
 Cohort or cross-sectional 16 3,135
 Case control 17 7,671
Commercial kit:
 Yes 24 9,390
 No 9 1,416
Immunoglobulin class:
 IgM 20 8,288
 IgG 21 3,338
 IgM or IgG 19 3,873
 IgM and IgG 15 7,482
Target antigen:
 Spike protein 3 615
 Nucleocapsid 2 340
 Both 7 1,227
 N/A 21 8,624

Data synthesis

We evaluated single IgM and IgG detection, combined IgM or IgG detection, and combined IgM and IgG detection. In single detection, samples are positive if the kit detects the corresponding antibody. In combined IgM or IgG detection, samples are positive if at least one of either immunoglobulin is detected, while in combined IgM and IgG detection both immunoglobulins need to be detected.

Among 33 studies, the highest sensitivity for the single IgM detection was 88.75% (79.72–94.72%), the single IgG detection 100% (90.00–100%), the combined IgM or IgG 100% (90.00–100%) and the combined IgM and IgG 98.57% (92.30–99.96%) [18–20]. The combined IgM or IgG detection had the highest pooled sensitivity (73.41% (72.22–74.57%)), while the single IgM detection had the lowest (61.80% (60.79–62.80%)) (Table II) [2138].

Table II.

Pooled sensitivity

Studiesref. Sensitivity (95% CI)
IgM IgG IgM or IgG IgM and IgG
Bisoffi et al.21 35.88 (30.78–41.23) 33.53 (28.53–38.82)
Andrey et al.22 63.41 (65.19–74.55) 90.24 (85.83–93.65)
Nicol et al.12 81.56 (74.16–87.59) 78.01 (70.27–84.55) 81.56 (74.16–87.59)
Pérez-García et al.13 28.89 (19.82–39.40) 60.00 (49.13–70.19) 64.44 (53.65–74.26)
Elslande et al.8 54.06 (51.02–57.08) 64.99 (62.04–67.85) 70.49 (67.66–73.21) 48.55 (45.53–51.60)
Serre-Miranda et al.23 67.29 (63.32–71.09) 68.15 (64.20–71.91) 73.66 (69.51–77.53) 74.39 (67.00–80.88)
Pan et al.6 55.81 (44.70–66.52) 54.65 (43.55–65.42) 68.60 (57.70–78.19)
Whitman et al.9 61.81 (58.89–64.67) 58.86 (55.91–61.76) 67.23 (64.54–69.84)
Xiang et al.24 57.14 (46.34–67.47) 81.32 (71.78–88.72) 82.42 (73.02–89.60)
Haymond et al.10 50.00 (36.08–63.92) 79.63 (66.47–89.37)
Candel González et al.19 74.29 (56.74–87.51) 100 (90.00–100) 100 (90.00–100) 74.29 (56.74–87.51)
Li et al.25 82.62 (78.53–86.22) 70.53 (65.78–74.97) 88.66 (85.13–91.61) 64.48 (59.56–69.19)
Hoffman et al.26 68.97 (49.17–84.72) 93.10 (77.23–99.15)
Montesinos et al.27 61.98 (56.92–66.86) 61.72 (56.65–66.60) 70.57 (65.74–75.09)
Wu et al.28 44.44 (38.71–50.30) 67.34 (61.68–72.65) 73.48 (68.85–77.77)
Lou et al.18 88.75 (79.72–94.72) 86.25 (76.73–92.93) 97.50 (91.26–99.70)
Almeida et al.20 85.24 (79.71–89.74) 84.29 (78.65–88.93) 96.19 (92.63–98.34) 98.57 (92.30–99.96)
Dortet et al.29 52.34 (46.03–58.60) 48.83 (42.55–55.13) 52.34 (46.03–58.60)
Vauloup-Fellous et al.30 63.94 (62.19–65.66) 64.25 (62.52–65.95) 73.56 (72.10–74.98)
Garrod et al.17 73.78 (69.45–77.78) 86.22 (82.69–89.27) 95 (92.21–97.01)
Maya et al.15 92 (73.97–99.02)
Adams et al.31 61.17 (55.48–66.63)
Pellanda et al.5 77.11 (66.58–85.62)
Choe et al.32 92.86 (84.11–97.64) 65.71 (53.40–76.65)
Shen et al.7 71.13 (61.05–79.89)
Vidal-Anzardo et al.33 43.75 (19.75–70.12)
Sisay et al.34 72.94 (67.05–78.29)
Chan et al.11 90.91 (83.44–95.76)
Kızıloglu et al.35 57.75 (45.44–69.39)
Ying et al.3 85.56 (76.57–92.08)
Dellière et al.36 93.40 (86.87–97.30)
Carozzi et al.37 98.30 (95.10–99.65)
Pallett et al.38 90.81 (86.73–93.96)
Total 61.80 (60.79–62.80) 65.69 (64.70–66.66) 73.41 (72.22–74.57) 71.51 (70.43–72.58)

Twenty-one studies reported a specificity value of 100%. The single IgG detection had the highest pooled specificity (96.68% (96.25–97.07%)), while the combined IgM and IgG detection had the lowest (94.14% (93.47–94.76%)) (Table III).

Table III.

Pooled specificity

Studies Specificity (95% CI)
IgM IgG IgM or IgG IgM and IgG
Bisoffi et al.21 90.13 (88.16–91.88) 94.54 (92.98–95.84)
Andrey et al.22 98.67 (95.69–98.78) 98.00 (95.70–99.26)
Nicol et al.12 95.39 (90.74–98.13) 98.03 (94.34–99.59) 95.39 (90.74–98.13)
Pérez-García et al.13 100 (96.38–100) 100 (96.38–100) 100 (96.38–100)
Elslande et al.8 95.01 (93.15–96.48) 96.53 (94.92–97.74) 92.51 (90.34–94.32) 99.03 (97.90–99.64)
Serre-Miranda et al.23 96.46 (92.85–98.57) 96.97 (93.52–98.88) 94.51 (89.84–97.46) 96.67 (88.47–99.59)
Pan et al.6 63.64 (40.66–82.80) 40.91 (20.71–63.65) 36.36 (17.20–59.34)
Whitman et al.9 94.14 (92.46–95.55) 97.59 (96.41–98.47) 94.16 (92.58–95.50)
Xiang et al.24 100 (90.00–100) 100 (90.00–100) 100 (90.00–100)
Haymond et al.10 92.86 (66.13–99.82) 64.29 (35.14–87.24)
Candel González et al.19 100 (47.82–100) 100 (47.82–100) 100 (47.82–100) 100 (47.82–100)
Li et al.25 91.41 (47.82–100) 98.44 (94.47–99.81) 90.63 (84.20–95.06) 99.22 (95.72–99.98)
Hoffman et al.26 100 (97.07–100) 99.19 (95.59–99.98)
Montesinos et al.27 98.61 (95.99–99.71) 98.61 (95.99–99.71) 98.61 (95.99–99.71)
Wu et al.28 100 (97.90–100) 100 (97.90–100) 100 (98.42–100)
Lou et al.18 98.09 (95.17–99.48) 99.52 (97.36–99.99) 95.22 (91.38–97.68)
Almeida et al.20 88.44 (82.13–93.12) 95.92 (91.33–98.49) 85.03 (75.72–85.41) 100 (92.75–100)
Dortet et al.29 100 (92.89–100) 100 (92.89–100) 100 (92.89–100)
Vauloup-Fellous et al.30 95.01 (94.06–95.85) 96.19 (95.34–96.92) 94.04 (93.09–94.89)
Garrod et al.17 96.80 (94.86–98.16) 98.40 (96.87–99.31) 96 (93.59–97.70)
Maya et al.15 100 (86.28–100)
Adams et al.31 97.86 (96.43–98.82)
Pellanda et al.5 98.00 (92.96–99.76)
Choe et al.32 96.20 (89.30–99.21) 100 (95.44–100)
Shen et al.7 96.23 (87.02–99.54)
Vidal-Anzardo et al.33 89.91 (85.12–93.57)
Sisay et al.34 95.36 (92.58–97.33)
Chan et al.11 98.21 (90.45–99.95)
Kızıloglu et al.35 85.45 (77.46–91.45)
Ying et al.3 91.01 (83.05–96.04)
Dellière et al.36 100 (91.59–100)
Carozzi et al.37 85.25 (82.13–88.02)
Pallett et al.38 96.50 (92.92–98.58)
Total 94.82 (94.30–95.31) 96.68 (96.25–97.07) 94.63 (93.95–95.26) 94.14 (93.47–94.76)

Twenty-one studies reported a PPV value of 100%. Among the four detection methods, single IgG detection had the highest pooled PPV (95.97% (95.47–96.42%)), while the single IgM detection had the lowest (93.45% (92.83–94.03%)) (Table IV).

Table IV.

Pooled PPV

Studies Positive predictive value (PPV) (95% CI)
IgM IgG IgM or IgG IgM and IgG
Bisoffi et al.21 54.22 (48.43–59.90) 66.67 (59.86–72.84)
Andrey et al.22 97.50 (93.19–97.79) 97.37 (94.36–98.79)
Nicol et al.12 94.26 (88.81–97.14) 97.35 (92.26–99.12) 94.26 (88.81–97.14)
Pérez-García et al.13 100 (87.10–100) 100 (93.40–100) 100 (93.80–100)
Elslande et al.8 94.15 (92.09–95.69) 96.53 (94.97–97.62) 93.33 (91.52–94.77) 98.67 (97.26–99.36)
Serre-Miranda et al.23 98.25 (96.44–99.15) 98.51 (96.79–99.32) 97.55 (95.46–98.69) 98.39 (93.96–99.58)
Pan et al.6 85.71 (76.99–91.50) 78.33 (70.84–84.33) 80.82 (74.87–85.63)
Whitman et al.9 92.50 (90.50–94.11) 96.62 (95.01–97.73) 93.09 (91.34–94.51)
Xiang et al.24 100 (93.10–100) 100 (95.10–100) 100 (95.10–100)
Haymond et al.10 96.43 (80.03–99.45) 89.58 (80.79–94.62)
Candel González et al.19 100 (87.10–100) 100 (90.10–100) 100 (90.10–100) 100 (87.10–100)
Li et al.25 96.76 (94.42–98.13) 99.29 (97.25–99.82) 96.70 (94.47–98.05) 99.61 (97.32–99.94)
Hoffman et al.26 100 (83.90–100) 96.43 (79.27–99.48)
Montesinos et al.27 98.76 (96.26–99.59) 98.75 (96.24–99.59) 98.91 (96.70–99.64)
Wu et al.28 100 (97.20–100) 100 (98.10–100) 100 (98.70–100)
Lou et al.18 94.67 (87.02–97.92) 98.57 (90.69–99.80) 88.64 (80.97–93.46)
Almeida et al.20 91.33 (87.03–94.29) 96.72 (93.08–98.48) 90.18 (86.19–93.11) 100 (94.70–100)
Dortet et al.29 100 (97.20–100) 100 (97–100) 100 (97.20–100)
Vauloup-Fellous et al.30 94.16 (93.11–95.06) 95.58 (94.64–96.36) 94.24 (93.37–95)
Garrod et al.17 95.40 (92.74–97.12) 97.98 (96.06–98.97) 95.53 (92.97–97.19)
Maya et al.15 100 (85.70–100)
Adams et al.31 93.10 (88.86–95.81)
Pellanda et al.5 96.97 (88.98–99.22)
Choe et al.32 95.59 (87.70–98.50) 100 (92.30–100)
Shen et al.7 97.18 (89.80–99.27)
Vidal-Anzardo et al.33 24.14 (13.85–38.63)
Sisay et al.34 92.08 (87.75–94.97)
Chan et al.11 98.90 (92.80–99.84)
Kızıloglu et al.35 71.93 (60.97–80.78)
Ying et al.3 90.59 (83.17–94.94)
Dellière et al.36 100 (96.30–100)
Carozzi et al.37 66.54 (62.07–70.73)
Pallett et al.38 97.24 (94.45–98.65)
Total 93.45 (92.83–94.03) 95.97 (95.47–96.42) 94.09 (93.39–94.73) 94.15 (93.52–94.73)

Combined IgM or IgG detection had the highest pooled NPV (75.34% (74.51–76.16%)), while single IgM detection had the lowest (67.48% (66.90–68.07%)) (Table V).

Table V.

Pooled NPV

Studies Negative predictive value (NPV) (95% CI)
IgM IgG IgM or IgG IgM dan IgG
Bisoffi et al.21 81.19 (79.91–82.41) 81.37 (80.17–82.51)
Andrey et al.22 76.68 (76.54–81.58) 92.45 (89.33–94.71)
Nicol et al.12 84.80 (79.73–88.77) 82.78 (77.87–86.78) 84.80 (79.73–88.77)
Pérez-García et al.13 60.98 (57.80–64.06) 73.53 (68.32–78.15) 75.76 (70.29–80.50)
Elslande et al.8 58.20 (56.56–59.82) 64.99 (63.08–66.84) 67.85 (65.75–69.89) 56.44 (54.50–57.88)
Serre-Miranda et al.23 50.00 (47.02–52.98) 50.79 (47.77–53.82) 54.77 (50.95–58.53) 58.00 (51.44–64.29)
Pan et al.6 26.92 (19.88–35.36) 18.75 (11.72–28.64) 22.86 (13.57–35.86)
Whitman et al.9 67.82 (66.14–69.46) 66.98 (65.39–68.52) 71.07 (69.38–72.71)
Xiang et al.24 47.30 (41.45–53.22) 67.31 (57.28–75.97) 68.63 (58.37–77.34)
Haymond et al.10 32.50 (26.22–39.48) 45.00 (29.80–61.19)
Candel González et al.19 35.71 (24.03–49.38) 100 (56.60–100) 100 (56.60–100) 35.71 (24.03–49.38)
Li et al.25 62.90 (57.62–67.90) 51.85 (48.01–55.67) 72.05 (66.07–77.34) 47.39 (44.08–50.72)
Hoffman et al.26 93.23 (88.90–95.95) 98.40 (94.17–99.57)
Montesinos et al.27 59.33 (56.19–62.40) 59.17 (56.04–62.22) 65.34 (61.73–68.77)
Wu et al.28 51.33 (48.78–53.86) 64.21 (60.37–67.87) 68.84 (65.22–72.25)
Lou et al.18 95.79 (92.48–97.68) 94.98 (91.61–97.04) 99.00 (96.20–99.74)
Almeida et al.20 80.75 (75.09–85.37) 81.03 (75.75–85.41) 93.98 (88.75–96.87) 98 (87.50–99.71)
Dortet et al.29 29.07 (26.50–31.79) 27.62 (25.30–30.08) 29.07 (26.50–31.79)
Vauloup-Fellous et al.30 67.67 (66.60–68.72) 67.70 (66.64–68.74) 72.85 (71.76–73.92)
Garrod et al.17 80.40 (77.83–82.74) 88.81 (86.29–90.91) 95.52 (93.15–97.10)
Maya et al.15 92.59 (76.79–97.93)
Adams et al.31 84.19 (82.23–85.97)
Pellanda et al.5 83.76 (77.64–88.46)
Choe et al.32 93.83 (86.71–97.25) 76.70 (70.41–81.99)
Shen et al.7 64.56 (57.02–71.43)
Vidal-Anzardo et al.33 95.61 (93.38–97.11)
Sisay et al.34 82.66 (79.56–85.38)
Chan et al.11 85.94 (76.61–91.94)
Kızıloglu et al.35 75.81 (70.25–80.61)
Ying et al.3 86.17 (78.96–91.19)
Dellière et al.36 85.71 (74.57–92.47)
Carozzi et al.37 99.41 (98.20–99.81)
Pallett et al.38 88.53 (84.15–91.82)
Total 67.48 (66.90–68.07) 70.08 (69.47–70.67) 75.34 (74.51–76.16) 71.46 (70.68–72.23)

Discussion

Diagnostic tests are tools to determine the presence or absence of diseases in subjects, and their accuracy needs to be validated [39]. The sensitivity indicates the proportion of individuals who have the target disease and tested positive among the sick population. Specificity indicates the proportion of individuals who do not have the target disease and tested negative among those without the disease. PPV is the proportion of ill patients with a positive test result, while NPV is the proportion of healthy patients with a negative test result [40].

The single IgM detection method had the lowest pooled sensitivity (61.80% (60.79–62.80%)). Low SARS-CoV-2 antibody concentration below the detection threshold may cause false-negative results that influence the sensitivity, but there is no established threshold yet [25]. Different immune responses between individuals might play a role since the kinetics of antibody production during SARS-CoV-2 infection is still unexplained [23, 32].

The sensitivity of RDT was only 11.1% for IgM and 3.6% for IgG in the first 7 days, peaked at day 14, and decreased to 30% on day 22 [6, 13]. IgG could be detected in the blood of COVID-19 patients starting from the eighth day after infection. IgM is detected 3–6 days after infection. Consequently, a patient tested too soon or when the IgM level has declined will produce a false-negative result [25, 41]. A decrease in viral load to undetectable levels due to antiviral therapy may diminish the antibody response. The quality of kits varies due to differences in manufacturers that probably produce a false-negative result may increase the probability of spreading the infection [3, 6].

The single IgG detection method had the highest pooled specificity (96.68% (96.25–97.07%)). The prolonged circulation of IgG enables its detection over a more extensive period [42]. False-positive results could indicate a history of other coronavirus infections or completely different viruses such as flu, syphilis, herpes, metapneumovirus, and dengue [31, 43]. High specificity is crucial since healthy individuals who tested false-positive can be infected if quarantined together with a COVID-19 patient. Overdiagnosis leads to unnecessary quarantine and hospitalization [31].

The reliability of RDT depends on the disease prevalence, which can change over time and in different population groups [44]. RDTs are less reliable if we do not pay attention to the predictive values. The predictive values must be taken into account because they affect the overall test results (the groups with and without the disease) [39]. High PPV indicates that a positively tested individual has SARS-CoV-2 antibodies, while high NPV means that negatively tested individuals do not have the antibodies [45]. This systematic review had a PPV of 93.45–95.97%, so positively-tested individuals should have a high probability of having SARS-CoV-2 antibodies. In contrast, the NPV ranges from 67.48 to 75.34%, so individuals with a negative result do not necessarily have SARS-CoV-2 antibodies. Considering the various limitations, RDTs can still be used, albeit with a careful interpretation.

In consideration of its ease of use and rapid testing, an antibody-based RDT has utility as a point of care test (POCT), a screening tool for specific situations or populations (e.g., travelers), and triage of outpatients with a fever of unclear etiology [7, 14]. A serological test does not determine a person’s health status but it assists in determining the proportion of the population with a SARS-CoV-2 infection history (serological surveillance) [37]. The advantages of antibody-based RDTs are their simple uses and affordable prices compared to antigen-based RDTs and RT-PCR.

The diagnostic accuracy of an antibody-based RDT is not satisfactory as a primary diagnostic test, especially in the early stages of infection. The sensitivity of RDT is < 90%, in consideration of the time to test and the individual immune response. The test has higher sensitivity using the combined IgM or IgG detection method. In contrast, the RDT has a higher specificity using the single IgG detection method, provided that the test is carried out at least 7 days after the onset of symptoms. Also, the test may be interpreted more accurately in combination with other tests such as radiology and doctor assessment.

This systematic review evaluates PPV and NPV, which are not always available in various diagnostic accuracy studies, compares diagnostic test parameters from different study populations, and includes preprint articles updated as soon as they were published. There are several limitations, such as the fact that we excluded studies that lacked data and complete supplemental material. Research on the diagnostic accuracy of LFIA RDTs was rare and modest compared to other methods. Also, these studies do not use strict settings due to the pandemic situations.

In conclusion, this systematic review indicates the antibody-based RDTs had moderate accuracy. Combined IgM or IgG detection had the highest pooled sensitivity (73.41% (95% CI: 72.22–74.57%)) and NPV (75.34% (95% CI: 74.51–76.16%)). Single IgG detection had the highest pooled specificity (96.68% (95% CI: 96.25–97.07%)) and PPV (95.97% (95% CI: 95.47–96.42%)). The benefits of antibody-based RDTs are their utility as POCT, a screening tool in outpatient triage for a rapid result, and to assist with serological surveillance. The results of the RDTs are quite reliable, provided that the test is carried out at least 7 days after the onset of symptoms. Antibody-based RDTs are easier to use and more affordable than antigen-based RDTs and RT-PCR. If the accuracy can be improved, RDTs might become an accurate and accessible diagnostic tool during the COVID-19 pandemic.

Recommendation

Based on their diagnostic accuracy, antibody-based RDTs should be limited to specific situations that require massive or rapid testing instead of being a primary diagnostic tool. Antibody-based RDTs may be interpreted more accurately in combination with other tests such as radiology and doctor assessment. The time to conduct the test should take into consideration the antibody production time, as doing the test too soon or while the antibody level had declined will only yield a false-negative result.

Conflict of interest

The authors declare no conflict of interest.

References

  • 1.Susilo A, Rumende CM, Pitoyo CW, et al. Coronavirus Disease 2019 : Tinjauan Literatur Terkini Coronavirus disease 2019: review of current literatures. J Penyakit Dalam Indones 2020; 7: 45-67. [Google Scholar]
  • 2.Paradiso AV, Summa S De, Loconsole D, et al. Clinical meanings of rapid serological assay in patients tested for SARS-Co2 RT-PCR. medRxiv 2020; 2020.04.03.20052183. [Google Scholar]
  • 3.Ying L, Yue-ping L, Bo D, et al. Diagnostic indexes of a rapid IgG/IgM combined antibody test for SARS-CoV-2. medRxiv 2020; 2020.03.26.20044883. [Google Scholar]
  • 4.Swadźba J, Bednarczyk M, Anyszek T, Martin E, Sci AM. A comparison of 7 commercial anti-SARS-CoV-2 antibody immunoassays. Arch Med Sci 2020; DOI: 10.5114/aoms.2020.98361. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 5.Pellanda LC, Wendland EMda R, McBride AJA, et al. Sensitivity and specificity of a rapid test for assessment of exposure to SARS-CoV-2 in a community-based setting in Brazil. medRxiv 2020.05.06.20093476. [Google Scholar]
  • 6.Pan Y, Li X, Yang G, et al. Serological immunochromatographic approach in diagnosis with SARS-CoV-2 infected COVID-19 patients. J Infect 2020; 10: e28-32. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 7.Shen B, Zheng Y, Zhang X, et al. Clinical evaluation of a rapid colloidal gold immunochromatography assay for SARS-Cov-2 IgM/IgG. Am J Transl Res 2020; 12: 1348-54. [PMC free article] [PubMed] [Google Scholar]
  • 8.Van Elslande J, Houben E, Depypere M, et al. Diagnostic performance of seven rapid IgG/IgM antibody tests and the Euroimmun IgA/IgG ELISA in COVID-19 patients. Clin Microbiol Infect 2020; 26: 1082-7. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 9.Whitman JD, Hiatt J, Mowery CT, et al. Evaluation of SARS-CoV-2 serology assays reveals a range of test performance. Nat Biotechnol 2020; 38: 1174-83. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 10.Haymond A, Mueller C, Steinberg H, et al. Clinical utility of a highly sensitive lateral flow immunoassay as determined by titer analysis for the detection of anti-SARS-CoV-2 antibodies at the point-of-care. medRxiv 2020.07.30.20163824. [Google Scholar]
  • 11.Chan CW, Shahul S, Coleman C, Tesic V, Parker K, Yeo KTJ. Evaluation of the truvian easy check COVID-19 IgM/IgG lateral flow device for rapid anti-SARS-CoV-2 antibody detection. Am J Clin Pathol 2021; 155: 286-95. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 12.Nicol T, Lefeuvre C, Serri O, et al. Assessment of SARS-CoV-2 serological tests for the diagnosis of COVID-19 through the evaluation of three immunoassays: two automated immunoassays (Euroimmun and Abbott) and one rapid lateral flow immunoassay (NG Biotech). J Clin Virol 2020; 129: 104511. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 13.Pérez-García F, Pérez-Tanoira R, Romanyk J, Arroyo T, Gómez-Herruz P, Cuadros-González J. Alltest rapid lateral flow immunoassays is reliable in diagnosing SARS-CoV-2 infection from 14 days after symptom onset: a prospective single-center study. J Clin Virol 2020; 129: 104473. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 14.Kemenkes . Pedoman Pencegahan dan Pengendalian Corona Virus deases (Covid-19) Revisi 5. Kementrian Kesehatan. 2020; 5. [Google Scholar]
  • 15.Maya V, Kaviyil JE, Ponnambath DK, et al. Evaluation of diagnostic accuracy of developed rapid SARS-COV-2 IgG antibody test kit using novel diluent system. Virusdisease 2021; 32: 1-7. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 16.Liu C, Huang Q, Wang P, et al. COVID-19 disease: novel clinical manifestations and therapeutic exploration. Arch Med Sci 2020; DOI: 10.5114/aoms.2020.98401. [DOI] [Google Scholar]
  • 17.Garrod G, Owen SI, Baillie JK, et al. Comparative evaluation of ten lateral flow immunoassays to detect SARS-CoV-2 antibodies. Wellcome Open Res 2021; 6: 18. [Google Scholar]
  • 18.Lou B, Li TD, Zheng SF, et al. Serology characteristics of SARS-CoV-2 infection since exposure and post symptom onset. Eur Respir J 2020; 56: 2000763. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 19.Candel González FJ, Viñuela-Prieto JM, Del Castillo JG, et al. Utility of lateral flow tests in SARS-CoV-2 infection monitorization. Rev Esp Quimioter 2020; 33: 258-66. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 20.Almeida SM, Spalanzani RN, Nogueira MB, et al. Rapid serological tests for SARS-COV-2: diagnostic performance of four commercial assays. [preprint] Research Square. 2020. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 21.Bisoffi Z, Pomari E, Deiana M, et al. Sensitivity, specificity and predictive values of molecular and serological tests for COVID-19: a longitudinal study in emergency room. Diagnostics 2020; 10: 669. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 22.Andrey DO, Cohen P, Meyer B, et al. Head-to-head accuracy comparison of three commercial COVID-19 IgM/IgG serology rapid tests. J Clin Med 2020; 9: 2369. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 23.Serre-Miranda C, Nobrega C, Roque S, et al. Performance assessment of 11 commercial serological tests for SARS-CoV-2 on hospitalized COVID-19 patients. Int Infect J Dis 2021; 104: 661-9. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 24.Xiang J, Yan M, Li H, et al. Evaluation of enzyme-linked immunoassay and colloidal gold-immunochromatographic assay kit for detection of novel coronavirus (SARS-Cov-2) Causing an outbreak of pneumonia (COVID-19). medRxiv 2020; 2020.02.27.20028787. [Google Scholar]
  • 25.Li Z, Yi Y, Luo X, et al. Development and clinical application of a rapid IgM-IgG combined antibody test for SARS-CoV-2 infection diagnosis. J Med Virol 2020; 92: 1518-24. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 26.Hoffman T, Nissen K, Krambrich J, et al. Evaluation of a COVID-19 IgM and IgG rapid test: an efficient tool for assessment of past exposure to SARS-CoV-2. Infect Ecol Epidemiol 2020; 10: 1754538. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 27.Montesinos I, Gruson D, Kabamba B, et al. Evaluation of two automated and three rapid lateral flow immunoassays for the detection of anti-SARS-CoV-2 antibodies. J Clin Virol 2020; 128: 104413. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 28.Wu JL, Tseng WP, Lin CH, et al. Four point-of-care lateral flow immunoassays for diagnosis of COVID-19 and for assessing dynamics of antibody responses to SARS-CoV-2. J Infect 2020; 81: 435-42. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 29.Dortet L, Emeraud C, Vauloup-Fellous C, et al. Rapid determination of SARS-CoV-2 antibodies using a bedside, point-of-Care, serological test. Emerg Microbes Infect 2020; 9: 2212-21. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 30.Vauloup-Fellous C, Maylin S, Périllaud-Dubois C, et al. Evaluation of 31 commercial SARS-CoV-2 serology assays. SSRN Electronic J 2020; 10.2139/ssrn.3696824. [DOI] [Google Scholar]
  • 31.Adams ER, Ainsworth M, Anand R, et al. Antibody testing for COVID-19: a report from the National COVID Scientific Advisory Panel. Wellcome Open Res 2020; 5: 139. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 32.Choe JY, Kim JW, Kwon HH, et al. Diagnostic performance of immunochromatography assay for rapid detection of IgM and IgG in coronavirus disease 2019. J Med Virol 2020; 92: 2567-72. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 33.Vidal-Anzardo M, Solis G, Solari L, et al. Evaluation under field conditions of a rapid test for detection of IgM and IgG antibodies against SARS-CoV-2. Rev Peru Med Exp Salud Publica 2020; 37: 203-9. [DOI] [PubMed] [Google Scholar]
  • 34.Sisay A, Tesfaye A, Desale A, et al. Diagnostic performance of SARS-CoV-2 IGM/IGG rapid test kits for the detection of the novel coronavirus in ethiopia. J Multidiscip Healthc 2021; 14: 171-80. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 35.Kızıloglu İ, Şener A, Siliv N. Comparison of rapid antibody test and thorax computed tomography results in patients who underwent RT-PCR with the pre-diagnosis of COVID-19. Authorea; 2021. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 36.Dellière S, Salmona M, Minier M, et al. Evaluation of the covid-19 igg/igm rapid test from orient gene biotech. J Clin Microbiol 2020; 58: e01233-20. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 37.Carozzi FM, Cusi MG, Pistello M, et al. Detection of asymptomatic SARS-CoV-2 infections among healthcare workers: results from a large-scale screening program based on rapid serological testing. J Chem Inf Model 2020; 53: 1689-99. [Google Scholar]
  • 38.Pallett SJC, Rayment M, Patel A, et al. Point-of-care serological assays for delayed SARS-CoV-2 case identification among health-care workers in the UK: a prospective multicentre cohort study. Lancet Respir Med 2020; 8: 885-94. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 39.Wong HB, Lim GH. Measures of diagnostic accuracy: sensitivity, specificity, PPV and NPV. Proc Singapore Healthc 2011; 20: 316-8. [Google Scholar]
  • 40.Trevethan R. Sensitivity, specificity, and predictive values: foundations, pliabilities, and pitfalls in research and practice. Front Public Heal 2017; 5: 307. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 41.Lee HK, Lee BH, Seok SH, et al. Production of specific antibodies against SARS-coronavirus nucleocapsid protein without cross reactivity with human coronaviruses 229E and OC43. J Vet Sci 2010; 11: 165-7. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 42.Jacofsky D, Jacofsky EM, Jacofsky M. Understanding antibody testing for COVID-19. J Arthroplasty 2020; 35: S74-81. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 43.Kadkhoda K. COVID-19 serologic testing: FAQs and caveats. Cleve Clin J Med 2020; 87: 329-33. [DOI] [PubMed] [Google Scholar]
  • 44.Dortet L, Ronat JB, Vauloup-Fellous C, et al. Evaluating 10 commercially available SARS-CoV-2 rapid serological tests by use of the STARD (standards for reporting of diagnostic accuracy studies) method. J Clin Microbiol 2021; 59: 1-12. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 45.Manski CF. Bounding the Predictive Values of COVID-19 Antibody Tests. NBER Working Papers 27226. 2020. [Google Scholar]

Articles from Archives of Medical Science : AMS are provided here courtesy of Termedia Publishing