Significance
Although MRI is a powerful method for visualizing features within organisms and materials, the relatively low signal-to-noise ratio (SNR) of the NMR signals that are used to construct an image makes MRI with isotropic spatial resolution below 3.0 μm impractical at room temperature. Here we show that SNR enhancements available from a combination of low temperatures and dynamic nuclear polarization allow MRI with 1.7-μm isotropic resolution. These results may enable informative MRI studies of eukaryotic cells, cell clusters, and small tissue samples.
Keywords: magnetic resonance imaging, magnetic resonance microscopy, hyperpolarization, dynamic nuclear polarization
Abstract
Spatial resolution in MRI is ultimately limited by the signal detection sensitivity of NMR, since resolution equal to ρiso in all three dimensions requires the detection of NMR signals from a volume ρiso3. With inductively detected NMR at room temperature, it has therefore proven difficult to achieve isotropic resolution better than ρiso = 3.0 μm, even with radio-frequency microcoils, optimized samples, high magnetic fields, optimized pulse sequence methods, and data acquisition times around 60 h. Here we show that spatial resolution can be improved and data acquisition times can be reduced substantially by performing MRI measurements at 5 K and using dynamic nuclear polarization (DNP) to enhance sensitivity. We describe the experimental apparatus and methods, and we report images of test samples with ρiso = 2.6 μm and ρiso = 1.7 μm, with signal-to-noise ratios greater than 15, acquired in 31.5 and 81.6 h, respectively. Image resolutions are verified by quantitative comparisons with simulations. These results establish a promising direction for high-resolution MRI of small samples. With further improvements in the experimental apparatus and in paramagnetic dopants for DNP, DNP-enhanced low-temperature MRI with ρiso < 1.0 μm is likely to become feasible, potentially enabling informative studies of structures within typical eukaryotic cells, cell clusters, and tissue samples.
MRI is a powerful and widely used technique for visualizing structures and characterizing other properties within organisms and heterogeneous materials. Spatial resolution in MRI can be limited by translational diffusion (1, 2) or by magnetic field gradient strengths, but, more generally, is limited by the signal-to-noise ratio (SNR) of the NMR signals from which MRI images are constructed. Due primarily to SNR considerations, inductively detected 1H MRI with isotropic spatial resolution ρiso better than ∼3.0 μm has been difficult to achieve, even with the use of radio-frequency (RF) microcoils to improve NMR detection sensitivity, paramagnetic doping to increase signal-averaging rates, and optimization of MRI pulse sequences (3–5). Consequently, although studies of individual biological cells have been a goal since the early days of MRI (6–17), progress in this direction has not been rapid.
In principle, the SNR of NMR can be improved by cooling both the sample and the RF circuitry, with an expected T−3/2 dependence on temperature T. If all other factors were unchanged, reducing T from 296 K to 5 K could increase the SNR by a factor of 455, allowing a reduction in ρiso by a factor of 4551/3 = 7.7. At low temperatures, dynamic nuclear polarization (DNP) based on the cross-effect mechanism (18, 19) could increase the SNR of 1H MRI by an additional factor of 100 or more, allowing a net reduction of ρiso by a factor greater than 30 relative to experiments at 296 K. In principle, MRI with submicron resolution seems possible, enabled by low temperatures and DNP (20).
However, various complications arise in practice when 1H MRI is performed at low temperatures with DNP. Most obviously, samples of interest become solids, implying that 1H–1H dipole–dipole couplings are no longer averaged out by molecular motions. Pulse sequences that include homonuclear dipolar decoupling techniques (21–26) must be employed to overcome the severe NMR line broadening and signal losses during phase-encoding periods that would otherwise outweigh the SNR advantages described above. We have shown that inclusion of Lee–Goldburg (LG) decoupling (27) during phase encoding and pulsed spin locking (PSL) during NMR signal detection (28) allows 1H MRI images with ρiso = 2.8 μm to be obtained at T = 28 K without DNP (5, 29).
A second complication is that paramagnetic compounds must be added to the sample at relatively high concentrations to enable cross-effect DNP, typically nitroxide-based biradicals (30, 31) or triradicals (32, 33). Fluctuating hyperfine fields from these compounds produce large reductions in transverse 1H spin relaxation times (T2H), again leading to signal losses during phase-encoding and NMR detection periods that cancel out the signal gains from DNP. These signal losses can be minimized by operating at T < 10 K, where the flip-flop transitions among unpaired electron spins that produce short T2H values are largely suppressed because the electron spins become strongly polarized (34). A third complication, discussed further below, is that build-up times for nuclear spin polarization (τDNP) are long at low temperatures, increasing the time between scans during MRI data acquisition.
Here we demonstrate that, despite the various complications, 1H MRI with DNP can be performed successfully at T = 5 K, thereby permitting substantial reductions in total image acquisition times and/or improvements in spatial resolution. Specifically, we report 1H MRI images of test samples with ρiso = 2.6 μm, acquired in 31.5 h, and with ρiso = 1.7 μm, acquired in 80.6 h. The latter value represents a reduction in voxel volume by a factor of 5.5 relative to the room temperature 1H MRI image with the highest isotropic resolution in the current literature (ρiso = 3.0 μm) (4).
Methods
Apparatus and Samples.
MRI experiments were performed at 9.39 T with a Tecmag Redstone console, operating at a 1H NMR frequency of 399.2 MHz. Gradient current pulses up to 100 A were available from AE Techron model 7548 gradient amplifiers. As previously described (5, 29), the MRI apparatus was contained in a Janis Supertran ST-200 continuous flow cryostat (Fig. 1A), with liquid helium flow from a pressurized tank (170-mbar gauge pressure) being controlled manually by a needle valve in the helium transfer line (Cryo Industries model HTL-3123-2). The temperature of the copper cold finger block within the cryostat was below 4.5 K, indicated by a temperature sensor (Lake Shore Cryotronics, CX-1070) and controller (Stanford Research Systems model CTC100). The actual sample temperature was 5 K, determined from 1H NMR signal amplitudes without DNP. Liquid helium consumption during acquisition of MRI data was 1.2 l/h.
Fig. 1.
(A) Cryostat used for DNP-enhanced MRI at low temperatures, photographed on a stand in the hallway outside the laboratory. The cryostat inserts into a vertical-bore (89 mm) superconducting magnet from the top. (B) MRI apparatus mounted on the temperature-controlled cold finger of the cryostat. Microwaves reach the sample from below the magnet through a hole in the sapphire plates (yellow arrow). (C) View of the RF microcoil through the hole for microwave irradiation. The microcoil is wound on a capillary tube that contains the sample and is soldered to RF leads on either side. (D) Equipment for microwave irradiation, consisting of two microwave sources with perpendicular polarizations (1 and 2) and a quasi-optical system for combining their outputs (3), comprising wire grid polarizers, elliptical mirrors, and roof mirrors. The combined output is transmitted to the cryostat through a corrugated waveguide (4). Additional details of the apparatus are shown in SI Appendix, Fig. S1.
The MRI apparatus consisted of a stack of sapphire plates containing gradient coils and the RF microcoil, with the design and dimensions reported previously (5, 29). The stack of sapphire plates was mounted on the cold finger block (Fig. 1B). Gradient coils were glued to the sapphire plates with Stycast epoxy. The RF microcoil was wound around a fused silica capillary tube with 105-μm outer diameter and 40-μm inner diameter (Molex/Polymicro, TSP040105), using bare copper wire with 20-μm diameter, and was held in place with cyanoacrylate glue (Fig. 1C). Additional details of the cryostat and MRI apparatus are shown in SI Appendix, Fig. S1.
Prior to winding the microcoil, the capillary tube was loaded with monodisperse amorphous silica beads with 9.2-μm diameter (Cospheric, SiO2MS-2.0), suspended in partially deuterated glycerol/water (1:3:6 volume ratio of H2O, D2O, and d8-glycerol, 10 mM Tris buffer, pH ≈ 9) containing 12 mM 4-{N,N-di-[2-succinyl,3-(TEMPO-4'-oxy)-propyl]}-amino-TEMPO (succinyl-DOTOPA) (33) as the paramagnetic dopant for DNP. Details of sample preparation and loading are given in SI Appendix, Supporting Methods and Fig. S2 A and B.
Gradient strengths were calibrated with one-dimensional (1D) images at room temperature and were found to be 0.385 T/m-A, 0.453 T/m-A, and 0.454 T/m-A in the x, y, and z directions, respectively, equivalent to 16.4 Hz/μm-A, 19.3 Hz/μm-A, and 19.3 Hz/μm-A for 1H nuclei. The x and z directions are defined to be the directions of the capillary and the external magnetic field, respectively. Measurements at 5 K indicated that gradient strengths were reduced by ∼9%, 14%, and 4%, respectively, presumably due to eddy currents. Gradient currents were adjusted accordingly.
A quasi-optical system, comprising wire grid polarizers, elliptical mirrors, and roof mirrors (Thomas Keating Ltd.), combined the outputs of two independent microwave sources (Virginia Diodes, Inc.) with perpendicular planes of linear polarization and with nominal output powers of 50 mW (source #1) and 30 mW (source #2). The combined outputs were focused on the entrance of a corrugated waveguide, which transmitted the microwaves to the cryostat (Fig. 1D and SI Appendix, Figs. S1 and S2C). After passing through a Teflon window at the lower end of the cryostat, the microwaves entered a tapered horn that was mounted on the bottom of the cryostat’s inner radiation shield. After the microwave beam diameter was reduced from 12 mm to 3.0 mm by the horn, microwaves were transmitted to the microcoil within the MRI apparatus by a tapered Teflon rod, with a final diameter of 1.8 mm (SI Appendix, Fig. S1J).
The output frequency of microwave source #1 could be modulated at 2.0 kHz with a symmetric triangular waveform. The output frequency of source #2 was variable but could not be modulated rapidly. The orientation of the cryostat relative to the microwave beam and the output frequencies of the two sources were adjusted to maximize DNP-enhanced 1H NMR signals, as shown in SI Appendix, Fig. S2 D and E and discussed further below. Images described below were obtained with source #1 sweeping between 262.75 GHz and 263.05 GHz and source #2 set to 262.9 GHz.
MRI Pulse Sequence.
We used the pulse sequence in Fig. 2A to acquire MRI data. The sequence began with a train of 64 RF pulses to destroy 1H spin polarization and suppress background signals (200-μs pulse lengths, 156-kHz RF amplitude, 1.0-ms delay between pulses). Microwaves were then applied for a period τDNP, optimized as discussed below, followed by a 100-ms delay for reequilibration of electron spin polarization to maximize transverse 1H spin relaxation times (34). To create a field of view in the x direction (FOVx) that was experimentally feasible, as discussed below, slice selection in the x direction was performed by applying a train of eight π pulses in the presence of a strong x gradient pulse (1.12 kHz/μm), with 50-μs delays for dephasing of transverse 1H polarization between pulses and with RF amplitudes alternating between 167 and 83 kHz. As shown in SI Appendix, Fig. S3, the longitudinal 1H polarization after the π pulse train is strongly peaked within a ±50-kHz range of offsets from the RF carrier frequency (∼90-μm range of x) and is nearly zero to ±150 kHz outside that range (∼270-μm range of x). For images described below, the RF carrier frequency was offset from the nominal 1H NMR frequency (i.e., the frequency at the center of the 1H NMR line in the absence of gradients) by 30.00 kHz, so that the selected slice was displaced from the center of the x gradient by 27 μm.
Fig. 2.
(A) Pulse sequence for DNP-enhanced MRI, as described in the text. (B) The 1H NMR signal at 5 K and k = 0, acquired with pulsed spin locking. (C) Signal after 11-Hz Lorentzian apodization and Fourier transformation. Spectrum in cyan is scaled up by a factor of 100 relative to the spectrum in blue to show the noise level. (D) Dependence of the NMR signal amplitude on τDNP from measurements with both microwave sources off (red points, scaled up by a factor of 32 relative to other data), with source #1 at 262.9 GHz and source #2 off (green points), with source #1 sweeping over a 0.3-GHz range around 262.9 GHz (cyan points) and source #2 off, and with source #1 sweeping and source #2 at 262.9 GHz (blue points). Color-coded curves are fits with the expression , yielding the indicated best-fit values of the buildup time τbu and asymptotic signal S∞. (E) Dependence of the NMR signal amplitude on τLG (blue points), fitted with the expression , yielding best-fit values T2LG = 688 ± 4 μs and P = 1.57 ± 0.02.
In earlier work, we demonstrated an alternative approach to slice selection for 1H MRI of frozen samples, based on RF phase modulation during LG irradiation (35). In this alternative approach, signals outside the desired slice are suppressed in a difference image, obtained by subtracting data with phase modulation from data without phase modulation. This approach could not be used in the current experiments because small signal fluctuations, apparently due to minor pulse sequence instabilities, prevented full cancellation of the large undesired signals in the difference data, leading to excess noise in the final images. With the slice selection method in Fig. 2A and SI Appendix, Fig. S3, undesired signals were directly suppressed in each scan, without requiring subtraction of signals in pairs of scans. Minor instabilities therefore did not contribute significant noise to the images.
After the slice selection period in Fig. 2A, the 3D phase-encoding period used LG irradiation to average out 1H–1H dipole–dipole couplings that would otherwise cause rapid signal decay. A π pulse in the middle of the phase-encoding period refocused gradient-independent frequency offsets. Effective LG decoupling field amplitudes were 250 kHz (204.1-kHz RF field with phase y or −y, 144.3- or −144.3-kHz carrier frequency offsets for “+LG” or “−LG” irradiation). Variations in LG precession frequencies due to RF inhomogeneity were refocused by alternating +LG and −LG periods, with equal length τLG/4. The 1H polarization was prepared initially perpendicular to the +LG effective field by a pulse with phase −x and flip angle , with . At the end of the phase-encoding period, polarization components that correspond to real or imaginary parts of the MRI data were stored as longitudinal polarization by applying either a single θ pulse with phase x or a θ pulse with phase x followed by a π/2 pulse with phase y. To maximize sensitivity, NMR signals were detected in the tPSL period with PSL, consisting of a π/2 pulse followed by a train of 32,768 pulses with phase y and flip angle α, with signal digitization after each pulse. To suppress ring-down and background signals, the phase of the π/2 pulse and the receiver alternated between x and −x on successive scans. With a 1.0-μs pulse length and a 3.8-μs delay between pulses, α was optimized at ∼72° as previously described (5). Under these conditions, 1H NMR signals at T = 5 K decayed exponentially during the PSL period with a time constant of 29 ms, corresponding to an 11-Hz linewidth (Fig. 2 B and C).
When strong gradients are applied, the angle between the effective LG field and the z axis of the usual rotating frame deviates from its ideal value in regions of the sample that are far from the center of the gradients, according to and for +LG and −LG periods, where ν1LG and ΔLG are the magnitudes of the RF field and frequency offset, respectively, g is the gradient vector, and r is the position within the sample relative to the center of the gradients. Therefore, the θ pulse at the beginning of the phase-encoding period in Fig. 2A cannot prepare 1H polarization perpendicular to the effective LG field at all sample positions when gradients are applied. Moreover, polarization that is perpendicular to the effective field at the end of a +LG period is not entirely perpendicular to the effective field in a subsequent −LG period. These effects lead to nonnegligible polarization components from r ≠ 0 that do not precess around the effective LG field (i.e., are spin locked along the effective field during the phase-encoding period) and consequently appear in images as strong artifacts near r = 0. As shown by simulations (SI Appendix, Fig. S4A), artifacts from spin-locked polarization can be eliminated by deliberately introducing nonzero rise and fall times for gradient pulses and by reducing gradient amplitudes close to zero around transitions between +LG and −LG periods. In this way, at least in principle, the θ pulse can prepare all 1H polarization perpendicular to the effective LG field initially. As gradients turn on and effective LG field directions change, the polarization follows adiabatically and remains perpendicular to the effective field. In experiments, rise and fall times for gradient pulses were roughly 50 μs, determined by the properties of the gradient amplifiers (SI Appendix, Fig. S4B).
Data Acquisition and Processing.
Points in k space were acquired in order of increasing |k|, which allowed preliminary images to be examined as data acquisition progressed. Data were stored as a set of 2D blocks in the Tecmag software, with 1H signals as a function of tPSL in the direct dimension and 128 combinations of gradient amplitudes (representing various k vectors) in the indirect dimension. For each k, real and imaginary parts of S(k) were acquired consecutively. Sixteen dummy scans were applied at the beginning of each 2D block, to suppress background 1H NMR signals from outside the sample.
It was not possible to place the center of the sample precisely in the center of the gradients. Consequently, strong gradient pulses produced large 1H NMR frequency shifts even at the center of the sample, which reduced the effectiveness of LG decoupling. To compensate for inevitable errors in sample positioning, corrections proportional to the gradient amplitudes were applied to the values of ΔLG and −ΔLG in +LG and −LG periods as previously described (5). In addition, to compensate for minor drifts in RF amplitudes during data acquisition, the value of ν1LG was optimized after each group of 10 2D blocks by maximizing the signal with all gradient amplitudes set to zero (SI Appendix, Fig. S5).
After exponential apodization and Fourier transformation with respect to tPSL, the amplitudes of the zero-frequency peaks in the resulting spectra were saved as the real or imaginary part of S(k). For images in Figs. 3 and 4, baseline corrections were applied in k space as described below. Data were zero filled to 256 points in kx and to 128 points in ky and kz before Fourier transformation. Cyclic permutations were then applied to the images in x, y, and z to bring the regions of greatest intensity to the center of the field of view.
Fig. 3.
DNP-enhanced 1H MRI image of 9.2-μm silica spheres in glycerol/water at 5 K within a capillary with 40-μm inner diameter. The image has nominal isotropic resolution of 2.6 μm and SNR ≈ 17. Data were obtained in 31.5 h. (A) The 2D slices perpendicular to x with 3.86-μm increments (points 60 through 228 out of 256 in the x direction). (B) The 2D slices perpendicular to z with 2.25-μm increments (points 30 through 100 out of 128 in the z direction). The 1D sections are shown above each panel at positions indicated by dashed red lines.
Fig. 4.
DNP-enhanced 1H MRI image with 1.7-μm nominal isotropic resolution and SNR ≈ 15, obtained in 80.6 h. (A) The 2D slices perpendicular to x with 2.35-μm increments (points 56 through 200 out of 256 in the x direction). (B) The 2D slices perpendicular to z with 2.13-μm increments (points 32 through 102 out of 128 in the z direction). The 1D sections are shown above each panel at positions indicated by dashed red lines.
Results
Selection of Measurement Conditions.
Following standard definitions, the nominal resolution of an MRI image acquired with the pulse sequence in Fig. 2A is , where , with gmax being the maximum average gradient amplitude (units of hertz per micrometer, assumed to be the same in x, y, and z) during the LG irradiation period τLG. The field of view in direction u is , where δku is the increment of , so that and for integers mu. In our experiments, we acquired data S(k) within a sphere in k space, including all points that satisfied and , with u = x, y, z. If we had acquired all data within a cube in k space, the number of complex data points would have been . With spherical k-space sampling, the number of complex data points is reduced by a factor of π/6. With four scans per k-space point (two scans each for the real and imaginary parts of S(k)), the total time to acquire one image is then
| [1] |
where tscan is the time per scan.
DNP-enhanced 1H NMR signals at 5 K are very large (compared to signals at room temperature) but build up slowly during τDNP, on a 6.2-s time scale with optimized microwave irradiation (Fig. 2D). Although the SNR of the final image with a given ttotal would be maximized by using τDNP ≈ 7.7 s, ttotal would exceed 5 d if FOVu/ρiso ≥ 30 for u = x, y, and z and tscan = 8 s. Hence, we used τDNP = 2.0 s to accelerate data acquisition at the expense of SNR.
In terms of the SNR of the k = 0 data point, called SNR0, assuming four scans per k-space point, and with the additional assumption that the total signal in the final image is divided equally among a fraction f of the voxels, the SNR of the final image can be expressed as
| [2] |
SNR0 decreases with increasing τLG, as shown in Fig. 2E. From Fig. 2C, with τDNP = 2.0 s and τLG = 600 μs, SNR0 ≈ 2,400. Assuming f ≈ 0.3, and with ≈ 5 × 104, Eqs. 1 and 2 predict SNR ≈ 18 for ttotal ≈ 60 h. In principle, with τLG = 600 μs, our gradient system allowed values of ρiso as small as ∼1.45 μm with 100-A current pulses, including the scaling of effective gradient amplitudes by under LG irradiation (5, 29) and reductions in gmax due to the nonnegligible rise and fall times of gradient pulses.
Image with 2.6-μm Isotropic Resolution.
Fig. 3 shows 2D planes perpendicular to x and to z from a 3D image of a sample of 9.2-μm silica beads in a capillary with 40.0-μm inner diameter, filled with partially protonated glycerol/water and doped with succinyl-DOTOPA triradicals as described above. Larger sets of 2D planes are shown in SI Appendix, Fig. S6. This image was acquired at 5 K with a total of 12,548 complex k-space points (−21 ≤ mx ≤ 20, −11 ≤ my, mz ≤ 10, with spherical sampling as described above), four scans per complex point, τDNP = 2.0 s, and a total data acquisition time of 31.5 h (not including overhead for data transfer and reoptimization of RF amplitudes). Gradient increments were chosen to give equal resolution in x, y, and z, based on calibrations described above. Slice selection conditions were chosen to shift the region of nonzero image intensity away from x = 0 (to avoid artifacts discussed below) and to keep this region within FOVx.
The actual fields of view in y and z were determined from the image itself by equating the number of points in the capillary diameter with 40 μm, resulting in FOVy = FOVz = 57.5 μm and implying a nominal spatial resolution of 2.6 μm. Assuming the same nominal spatial resolution in x, FOVx = 109.8 μm.
Silica spheres are clearly seen and clearly resolved from one another in Fig. 3. The actual SNR value is 16.6, defined as the ratio of the maximum image intensity within the capillary to the rms intensity in regions outside the capillary that contain only noise. It should be noted that ρiso = 2.6 μm is a small improvement over our previous results without DNP, in which ρiso = 2.8 μm was achieved at 28 K, using a microcoil with 150-μm inner diameter (5). The previous results required 208 h of measurements to produce SNR ≈ 11, with 0.50 s per scan and a Dy3+-doped sample.
Image with 1.7-μm Isotropic Resolution.
Fig. 4 shows 2D planes from a higher-resolution image of a similar sample, acquired at 5 K with a total of 32,102 complex k-space points (−30 ≤ mx ≤ 29, −16 ≤ my, mz ≤ 15, with spherical sampling), four scans per complex point, τDNP = 2.0 s, and a total data acquisition time of 80.6 h. Larger sets of 2D planes are shown in SI Appendix, Fig. S7. Again, gradient increments were chosen to give equal resolution in x, y, and z; slice selection conditions were chosen to shift the region of nonzero image intensity away from x = 0 and to keep this region within FOVx; k-space data were zero-filled to 256 × 128 × 128; and the real-space image was cyclically permuted to center the region of nonzero intensity. Maximum x, y, and z gradient currents were 88.0, 79.5, and 68.3 A, respectively. From the known capillary diameter of 40 μm, fields of view were calculated to be FOVy = 52.3 μm and FOVz = 54.6 μm, implying a nominal spatial resolution of 1.7 μm. Assuming the same nominal spatial resolution in x, FOVx = 100.2 μm.
The inner wall of the capillary and the surfaces of the spheres are more sharply defined in Fig. 4 than in Fig. 3, consistent with higher spatial resolution. The actual SNR value of the image in Fig. 4 is 15.5. Compared with our previous results at 28 K without DNP (5), the SNR value is greater by a factor of 1.4, the data acquisition time is smaller by a factor of 2.6, and the voxel volume is smaller by a factor of 4.5.
Baseline Corrections in k Space.
Images in Figs. 3 and 4 show minor artifacts near the origin (i.e., x = y = z = 0) due to low-frequency baseline distortions in the k-space data. In Fig. 3, the origin is shifted from image point (129, 65, 65) to image point (74, 69, 60) by the cyclic permutation described above. In Fig. 4, the origin is shifted to image point (199, 60, 97). Baseline distortions in k space arise from two sources. First, as gradient pulse amplitudes increase, the nonevolving, spin-locked component of nuclear spin polarization in the phase-encoding period varies due to changes in NMR frequency offsets that produce changes in the effective RF field direction during LG irradiation. Although gradient pulse shapes were modified to reduce spin locking along the effective field, as described above, RF inhomogeneity during θ pulses and other pulse imperfections can still produce significant spin-locked components. This contribution is different for different gradient directions, and hence depends on both the magnitude and the direction of k. Second, as data acquisition proceeds, slow drifts in RF amplitudes and temperature within the cryostat also cause the spin-locked component of nuclear spin polarization in the phase-encoding period to vary. Because our data were acquired in an “onion shell” manner in k space, proceeding monotonically from small |k| to large |k|, this contribution to baseline distortions in k space is largely independent of the direction of k.
The maximum amplitude of the spin-locked component is less than 5% of the total polarization, leading to baseline distortions that are less than 5% of the maximum NMR signal. Nonetheless, these distortions cause severe image artifacts if no correction is done (SI Appendix, Fig. S8A). To minimize these artifacts, k-space data were first symmetrized by replacing S(k) with [S(k)+S(-k)*]/2. Next, functions proportional to kx2, ky2, kz2, kxky, kxkz, kx4, and kx2ky2 were subtracted from the real part of S(k), and functions proportional to kx, ky, and kz were subtracted from the imaginary part of S(k), with coefficients that were optimized to minimize the total k-space signal power in the range 0.89kmax ≤ |k| ≤ kmax. These direction-dependent corrections reduced, but did not eliminate, the image artifacts (SI Appendix, Fig. S8B). Finally, a direction-independent baseline-fitting function F(|k|) was derived by calculating the average values of the real part of S(k) within each of 40 (Fig. 3) or 100 (Fig. 4) ranges of k-space radii, equally spaced in |k|2 from zero to kmax2, then performing cubic spline interpolation and smoothing of these average values as a function of |k| (SI Appendix, Fig. S8 C and D). After subtraction of F(|k|) from S(k), final images with only minor artifacts were obtained (Figs. 3 and 4 and SI Appendix, Fig. S8E).
Verification of Image Resolution by Comparisons with Simulations.
Although the images in Figs. 3 and 4 have nominal isotropic spatial resolutions equal to 2.6 and 1.7 μm, respectively, the true resolution could be lower if the samples or gradient coils moved in response to strong gradient pulses, if NMR signals decreased over the many hours of data acquisition due to instrumental instabilities, or for other reasons. It is therefore important to assess the true resolution by comparing the experimental images with simulations. Two types of comparisons were performed, with results shown in Figs. 5 and 6. Additional details of image simulations are given in SI Appendix.
Fig. 5.
Assessment of spatial resolution by comparison of experimental and simulated images. (A) Examples of 2D slices of simulated images with Gaussian blurring to produce the indicated values of the isotropic resolution. Slices are perpendicular to the x direction at the indicated x values. Cyan circles have 40-μm diameters to match the inner capillary diameter in experiments. Deviations between experimental and simulated images were calculated within the dashed red circles, which have 34.4-μm diameters. (B) Total squared deviation between the experimental image in Fig. 3 and simulations with various values of the isotropic resolution, evaluated over two different ranges of x. (C) Total squared deviation between the experimental image in Fig. 4 and simulations with various values of the isotropic resolution, evaluated over four different ranges of x. Error bars are SDs from four independent simulations for each resolution value, with separate optimizations of sphere positions in each simulation. When not shown, error bars are smaller than the symbols.
Fig. 6.
Assessment of spatial resolution by comparison of simulated and experimental k-space power spectra. (A and B) Examples of 2D slices perpendicular to the z and x directions from simulated images, generated from simulated k-space data that were truncated at a kmax value corresponding to ρiso = 1.7 μm. As described in the text, simulated sphere positions were optimized to match those in experimental images. (C and D) Comparisons of k-space power spectra in 2D planes perpendicular to x or z, for a simulated image with ρiso = 2.6 μm (red squares) and the experimental image in Fig. 3, over ranges of x or z that contain spheres (blue circles) or only noise (cyan triangles). (E and F) Same, for a simulated image with ρiso = 1.7 μm and the experimental image in Fig. 4. Dashed red lines in A indicate 39.8 μm < x < 55.2 μm, the interval used for E.
For Fig. 5, simulated images of silica spheres in glycerol/water were generated, using Gaussian point spread functions (i.e., Gaussian “blurring” of the spheres) with various values of ρiso. According to the standard Sparrow criterion (36), in this case, where Φ is the full width at half maximum of the Gaussian function. The number of spheres and their approximate locations in the experimental images were first determined by running simulated annealing calculations to minimize the total squared deviation between experimental and simulated image intensities, using simulated images with ρiso = 1.0 μm, a sphere diameter of 8.0 μm, a variable number of spheres, and variable coordinates for each sphere. Simulated annealing calculations used the standard Monte Carlo algorithm (37) to accept or reject changes in the number of spheres or their coordinates. The y and z coordinates were constrained to remain within the circular cross-section of the capillary. In a series of separate calculations, the x coordinates were constrained to remain within intervals of length 12.9 μm (30 points), starting at x points 95, 105, 115, 125, 135, or 145 (out of 256) for the image in Fig. 3, or within intervals of length 15.7 μm (40 points), starting at x points 90, 100, 110, or 120 for the image in Fig. 4. The overall scaling of intensities in the simulated images was optimized in each Monte Carlo step.
From multiple simulated annealing runs, coordinates for 36 spheres for the image in Fig. 3 and 32 spheres for the image in Fig. 4 were obtained. These coordinates were then used as initial conditions for the final simulated annealing calculations, in which the number of spheres was kept constant, sphere diameters were set to 9.2 μm, and spheres were constrained to remain within 3.0 μm of their initial positions. To prevent spheres from overlapping, repulsive potentials among all spheres, proportional to for spheres u and v when , were included in the function that was minimized by the simulated annealing algorithm.
In the final calculations, the ρiso value in the simulated images was varied from 1.7 μm to 3.5 μm for comparisons with the experimental image in Fig. 3, and from 0.8 μm to 2.6 μm for comparisons with the experimental image in Fig. 4. Four independent runs were performed with each value of ρiso. The squared deviation between experimental and simulated images was minimized within two (for Fig. 3) or four (for Fig. 4) separate ranges of x values in separate runs. Fig. 5A shows examples of 2D planes from the final simulated images with various values of ρiso. Fig. 5B shows the dependence of the minimized squared deviation on ρiso for the experimental image in Fig. 3. The optimal value of ρiso is about 2.3 μm to 2.7 μm, consistent with the nominal resolution of 2.6 μm explained above. Fig. 5C shows the dependence of the minimized squared deviation on ρiso for the experimental image in Fig. 4. The optimal value of ρiso is about 1.6 μm to 1.9 μm, consistent with the nominal resolution of 1.7 μm explained above. Thus, a quantitative comparison of simulated and experimental images, using a Gaussian point spread function in the simulations, indicates no discrepancy between true and nominal spatial resolution values.
In the second approach to assessing the true resolution, we calculated the k-space data Ssim(k) that would produce simulated images with 9.2-μm-diameter silica spheres at positions determined in the final simulated annealing calculations described above. Ssim(k) was truncated to include only the k-space points that were sampled in experimental measurements. Zero filling and Fourier transformation of Ssim(k) resulted in simulated images that appeared quite similar to the experimental images within the range of x coordinates where spheres were located, as shown in Fig. 6 A and B. Note that simulated images in Fig. 6 A and B differ from simulated images in Fig. 5A because the point spread function in Fig. 6 A and B matches the experimental point spread function, which is not a Gaussian function. We then calculated average 2D k-space power spectra Wx(|k|) and Wz(|k|) for the experimental and simulated images in planes perpendicular to x and z, as shown in Fig. 6 C–F. These power spectra are defined by
| [3] |
where x1 ≤ x ≤ x2 and z1 ≤ z ≤ z2 are the ranges of image points within which the power spectra are averaged, FTx and FTz represent Fourier transforms with respect to x and z, and the δ-functions select combinations of ky and kz or kx and ky with a given |k|. Values of |k| are binned for the plots in Fig. 6 C–F. For calculations of Wz(|k|) from Ssim(k), intensities in the simulated images were multiplied by a Gaussian function of x to approximate the effects of slice selection in the experimental images.
Simulated power spectra in Fig. 6 (red squares) were multiplied by overall scaling factors to produce agreement with the experimental power spectra (blue circles) at small values of |k|/kmax. Importantly, good agreement at larger values is also observed in all cases until the k-space power from background noise in the experimental images (cyan triangles) becomes comparable to the “signal power” from real intensity variations from the silica spheres. The experimental power then exceeds the simulated power, because the experimental power is the sum of signal power and noise power.
Plots in Fig. 6C show that the experimental signal power is substantially greater than the noise power at all values of |k|/kmax. Together with the good agreement between experimental and simulated power spectra, this implies that the true resolution of the image in Fig. 3 must be essentially the same as the nominal resolution (at least in yz planes). Plots in Fig. 6E show that the signal power exceeds the noise power when |k|/kmax ≤ 0.9. This implies that the true resolution of the image in Fig. 4 is at least (1.7/0.9) μm ≈ 1.9 μm. For |k|/kmax > 0.9, the power in the experimental image is dominated by the noise contribution and therefore exceeds the power in the simulated image, because the k-space signal power from 9.2-μm silica spheres above |k|/kmax = 0.9 is inherently small.
Effects of k-Space Truncation on Experimental and Simulated Images.
A surprising feature of the experimental image in Fig. 4 is the presence of bright spots in the centers of each silica sphere. This feature is also present in the simulated images in Fig. 6 A and B. As shown in SI Appendix, Fig. S9, strong peaks of positive intensity in the centers of the spheres result from truncation of S(k) when the value of kmax = π/ρiso is such that the ratio of ρiso to the sphere radius R is ∼0.39 ± 0.02. These peaks are due to “truncation wiggles” that extend inward from the surface of each sphere in all directions and combine constructively at the center. At other values of kmax, the image intensity at the centers of the spheres is negative, for example, when ρiso/R = 0.565 as in Fig. 3. As also shown in SI Appendix, Fig. S9, the values of ρiso/R at which truncation of S(k) produces either positive or negative intensities at the centers of the spheres are different if the data are sampled within a cube (i.e., −kmax ≤ kx,ky,kz < kmax) rather than within a sphere. Different sampling methods in k space result in point spread functions with different symmetries in the final images.
Discussion
Summary of Results and Comparison with Previous Work.
Images in Figs. 3 and 4 demonstrate that 3D MRI images can be obtained at low temperatures with signal enhancements from DNP. DNP-enhanced 1D images obtained at 30 K were reported in an earlier paper from our laboratory (35). As shown in Figs. 3 and 4, DNP-enhanced 3D images with SNR ≈ 16 and with ρiso = 2.6 μm or 1.7 μm can be obtained at 5 K in 31.5 h or 80.6 h with our current apparatus and measurement conditions. Comparisons with simulations in Figs. 5 and 6 indicate good agreement between nominal and actual values of ρiso. Bright spots in the centers of dark spheres in Fig. 4 are also an indication that the experimental point spread function matches expectations for ρiso = 1.7 μm.
The highest resolution in 1H MRI near 300 K was obtained by Weiger et al. (4), who reported ρiso = 3.0 μm and SNR ≈ 43 for an image of glass fibers in Cu2+-doped H2O, acquired at 18.8 T in 58 h with 0.10 s per scan. Somewhat earlier, Ciobanu et al. (3) reported an image of polymer beads in Cu+-doped H2O with 3.7 μm × 3.3 μm × 3.3 μm resolution and SNR ≈ 5, acquired at 9.0 T in 30 h with 0.35 s per scan. Our laboratory subsequently reported an image of glass beads in Dy3+-doped, fully protonated glycerol/water with ρiso = 2.8 μm and SNR ≈ 11, acquired at 28 K and 9.39 T in 208 h with 0.50 s per scan (5). Coil volumes and other experimental details were similar but not identical in these three studies. Our DNP-enhanced MRI results therefore represent a reduction in voxel volume by factors of 5.5 to 8.2 relative to the best previous results at 300 K (liquid state MRI) and a factor of 4.5 relative to the best previous results at low temperature (solid-state MRI). Our DNP-enhanced MRI results also represent increases in SNR per voxel volume by factors of 2 to 26 relative to the best previous results at 300 K and a factor of 6.5 relative to the best previous results at low temperature.
Prospects for Further Improvements.
A calculation by Abragam (38) shows that, under ideal conditions, the SNR of inductively detected NMR signals from spin S nuclei with gyromagnetic ratio γ (radians per second per gauss), NMR frequency νNMR (hertz), linewidth w (hertz), and density N (per cubic centimeter) in a solenoidal coil with volume V (cubic centimeters), filling factor η, and quality factor Q is after a single scan, with T being the temperature of the sample and the coil, and assuming spin polarization equal to its thermal equilibrium value. For 1H NMR at 9.39 T and with Q = 100, the ideal SNR after n scans becomes , where c is the spin concentration (moles per liter), assuming full equilibration of the spin polarization before each scan. In MRI, η is the ratio of one voxel volume to the total coil volume. Assuming T = 300 K, c = 111 M (as in pure water), V = 3.14 × 10−6 cm3 (100-μm diameter by 400-μm length), w = 10 Hz, and η = 8.6 × 10−6 (3-μm voxel width in x, y, and z directions), roughly 1,000 scans would be required to achieve SNR ≈ 10 under ideal conditions. In practice, the number of scans required to achieve similar SNR values in micrometer-scale MRI with microcoils is much greater, on the order of 1 × 105 at 18.8 T (4) or 2 × 105 at 9.0 T with a somewhat larger value of η (3). The discrepancy factor of ∼30 between ideal and real SNR values is attributable to nonideal behavior of solenoidal coils as their diameter decreases below 1 mm (39, 40), additional noise from resistance in other RF circuit elements, reduction in the real value of η due to RF fields outside the RF coil, signal loss from transverse spin relaxation during the MRI pulse sequence, and a factor of 21/2 from the need for separate acquisition of the real and imaginary parts of k-space data.
Compared with MRI images at 300 K with similar values of Q, η, V, and ttotal, DNP-enhanced MRI images at 5 K could, in principle, have greater SNR by a factor of , where subscripts 5 and 300 refer to the temperatures, and εDNP is the factor by which nuclear spin polarization is enhanced. With ≈ 100, ≈ 0.1 (due to the slower buildup of polarization at 5 K compared with spin-lattice relaxation in paramagnetically doped water at 300 K), and ≈ 0.1, this expression predicts improvements in SNR by a factor of ∼4,000. The predicted improvement should be reduced by factors of 2.4 from signal decay during τLG, 1.5 from the fact that signals were sampled during only 42% of tPSL, 3.6 from our use of a suboptimal value of τDNP, ∼4 from our use of a standard preamplifier in the NMR spectrometer's receiver with a noise temperature around 80 K, and ∼10 from our use of partially protonated glycerol/water. With these corrections, the predicted improvement in SNR becomes a factor of ∼8. In fact, we have achieved an improvement by a factor on the order of 5 to 20 relative to previous micrometer-scale MRI experiments at room temperature (3, 4).
This analysis indicates several directions for further improvements: 1) Lower sample temperatures may lead to slower signal decays during τLG and tPSL (34), resulting in larger signals; 2) a cryogenic preamplifier may result in substantially lower noise levels; 3) paramagnetic dopants that are more efficient at very low temperatures may allow smaller values of τDNP and provide larger values of εDNP. More efficient dopants may also allow higher protonation levels. A further increase in SNR by a net factor of 5 would permit ρiso = 1.0 μm.
Potential Applications.
With further improvements as outlined above, we believe it will become possible to apply DNP-enhanced, low-temperature MRI to real biological samples. One goal is to perform MRI on individual eukaryotic cells with sufficient resolution to visualize the internal structures of membrane-enclosed organelles, as well as phase-separated entities known as “membraneless organelles” (41). Obviously, a great deal is already known about the internal organization of eukaryotic cells from well-established methods of optical microscopy, fluorescence microscopy, and electron microscopy. Nonetheless, because contrast mechanisms in MRI are different from contrast mechanisms in other forms of microscopy, it seems possible that MRI will provide information that is not otherwise accessible. In DNP-enhanced 1H MRI, contrast may be generated from variations in local concentrations of paramagnetic dopants, variations in nuclear spin relaxation properties, or variations in 1H–1H dipole–dipole coupling strengths, in addition to variations in local 1H densities. Double-resonance techniques may also be used to produce contrast, for example, by taking advantage of dipole–dipole couplings between 31P and 1H couplings to select 1H NMR signals from regions with higher or lower 31P concentrations. DNP-enhanced MRI may also be useful for characterizing the arrangement of different cell types within cell clusters, such as the α, β, and δ cells within pancreatic islets (42). Applications to small tissue samples, such as samples from biopsies, may be possible and useful. DNP-enhanced MRI studies of small organisms that are commonly used as model systems in biology, or portions thereof, such as Caenorhabditis elegans worms and Drosophila melanogaster brains, may also prove to be fruitful. Extensive exploratory studies are needed before the true feasibility and scientific significance of these applications can be assessed.
Supplementary Material
Acknowledgments
This work was supported by the Intramural Research Program of the National Institute of Diabetes and Digestive and Kidney Diseases, NIH. We thank Dr. Kent R. Thurber for helpful discussions and Dr. Wai-Ming Yau for synthesizing succinyl-DOTOPA.
Footnotes
The authors declare no competing interest.
This article is a PNAS Direct Submission.
This article contains supporting information online at https://www.pnas.org/lookup/suppl/doi:10.1073/pnas.2201644119/-/DCSupplemental.
Data Availability
Raw and processed data for Figs. 3 and 4, as well as Fortran programs for data processing and for simulations in Figs. 5 and 6, are available at https://doi.org/10.17632/vk344gdnwp.1.
References
- 1.McFarland E. W., Time-independent point-spread function for NMR microscopy. Magn. Reson. Imaging 10, 269–278 (1992). [DOI] [PubMed] [Google Scholar]
- 2.Hyslop W. B., Lauterbur P. C., Effects of restricted diffusion on microscopic NMR imaging. J. Magn. Reson. 94, 501–510 (1991). [Google Scholar]
- 3.Ciobanu L., Seeber D. A., Pennington C. H., 3D MR microscopy with resolution 3.7 microm by 3.3 microm by 3.3 microm. J. Magn. Reson. 158, 178–182 (2002). [DOI] [PubMed] [Google Scholar]
- 4.Weiger M., et al. , NMR microscopy with isotropic resolution of 3.0 μm using dedicated hardware and optimized methods. Concepts Magn. Reson. Part B 33B, 84–93 (2008). [Google Scholar]
- 5.Chen H. Y., Tycko R., Low-temperature magnetic resonance imaging with 2.8 μm isotropic resolution. J. Magn. Reson. 287, 47–55 (2018). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 6.Aguayo J. B., Blackband S. J., Schoeniger J., Mattingly M. A., Hintermann M., Nuclear magnetic resonance imaging of a single cell. Nature 322, 190–191 (1986). [DOI] [PubMed] [Google Scholar]
- 7.Bowtell R. W., Brown G. D., Glover P. M., McJury M., Mansfield P., Resolution of cellular structures by NMR microscopy at 11.7 T. Philos. Trans. R. Soc. Lond. Ser. A Math. Phys. Eng. Sci. 333, 457–467 (1990). [Google Scholar]
- 8.Johnson G. A., et al. , Histology by magnetic resonance microscopy. Magn. Reson. Q. 9, 1–30 (1993). [PubMed] [Google Scholar]
- 9.Hsu E. W., Aiken N. R., Blackband S. J., Nuclear magnetic resonance microscopy of single neurons under hypotonic perturbation. Am. J. Physiol. 271, C1895–C1900 (1996). [DOI] [PubMed] [Google Scholar]
- 10.Grant S. C., et al. , NMR spectroscopy of single neurons. Magn. Reson. Med. 44, 19–22 (2000). [DOI] [PubMed] [Google Scholar]
- 11.Wind R. A., et al. , An integrated confocal and magnetic resonance microscope for cellular research. J. Magn. Reson. 147, 371–377 (2000). [DOI] [PubMed] [Google Scholar]
- 12.Ciobanu L., Pennington C. H., 3D micron-scale MRI of single biological cells. Solid State Nucl. Magn. Reson. 25, 138–141 (2004). [DOI] [PubMed] [Google Scholar]
- 13.Purea A., Neuberger T., Webb A. G., Simultaneous NMR microimaging of multiple single-cell samples. Concepts Magn. Reson. Part B 22B, 7–14 (2004). [Google Scholar]
- 14.Lee S. C., et al. , In vivo magnetic resonance microscopy of differentiation in Xenopus laevis embryos from the first cleavage onwards. Differentiation 75, 84–92 (2007). [DOI] [PubMed] [Google Scholar]
- 15.Flint J. J., et al. , Cellular-level diffusion tensor microscopy and fiber tracking in mammalian nervous tissue with direct histological correlation. Neuroimage 52, 556–561 (2010). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 16.Baxan N., et al. , Microcoil-based MR phase imaging and manganese enhanced microscopy of glial tumor neurospheres with direct optical correlation. Magn. Reson. Med. 68, 86–97 (2012). [DOI] [PubMed] [Google Scholar]
- 17.Flint J. J., Menon K., Hansen B., Forder J., Blackband S. J., Visualization of live, mammalian neurons during Kainate-infusion using magnetic resonance microscopy. Neuroimage 219, 116997 (2020). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 18.Carver T. R., Slichter C. P., Experimental verification of the Overhauser nuclear polarization effect. Phys. Rev. 102, 975–980 (1956). [Google Scholar]
- 19.Farrar C. T., Hall D. A., Gerfen G. J., Inati S. J., Griffin R. G., Mechanism of dynamic nuclear polarization in high magnetic fields. J. Chem. Phys. 114, 4922–4933 (2001). [Google Scholar]
- 20.Thurber K. R., Tycko R., Prospects for sub-micron solid state nuclear magnetic resonance imaging with low-temperature dynamic nuclear polarization. Phys. Chem. Chem. Phys. 12, 5779–5785 (2010). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 21.Chingas G. C., Miller J. B., Garroway A. N., NMR images of solids. J. Magn. Reson. 66, 530–535 (1986). [Google Scholar]
- 22.Cory D. G., Miller J. B., Garroway A. N., Time-suspension multiple-pulse sequences: Applications to solid-state imaging. J. Magn. Reson. 90, 205–213 (1990). [Google Scholar]
- 23.Cho H. M., Lee C. J., Shykind D. N., Weitekamp D. P., Nutation sequences for magnetic resonance imaging in solids. Phys. Rev. Lett. 55, 1923–1926 (1985). [DOI] [PubMed] [Google Scholar]
- 24.Deluca F., Lugeri N., Motta S., Cammisa G., Maraviglia B., Full-bandwidth parameter distribution in line-narrowing solid-state imaging. J. Magn. Reson. A 115, 1–6 (1995). [Google Scholar]
- 25.Weigand F., Blümich B., Spiess H. W., Application of nuclear magnetic resonance magic sandwich echo imaging to solid polymers. Solid State Nucl. Magn. Reson. 3, 59–66 (1994). [DOI] [PubMed] [Google Scholar]
- 26.Tarasek M. R., Goldfarb D. J., Kempf J. G., Enhancing time-suspension sequences for the measurement of weak perturbations. J. Magn. Reson. 209, 233–243 (2011). [DOI] [PubMed] [Google Scholar]
- 27.Lee M., Goldburg W. I., Nuclear magnetic resonance line narrowing by a rotating RF field. Phys. Rev. 140, 1261–1271 (1965). [Google Scholar]
- 28.Ostroff E. D., Waugh J. S., Multiple spin echoes and spin locking in solids. Phys. Rev. Lett. 16, 1097–1098 (1966). [Google Scholar]
- 29.Moore E., Tycko R., Micron-scale magnetic resonance imaging of both liquids and solids. J. Magn. Reson. 260, 1–9 (2015). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 30.Hu K. N., Yu H. H., Swager T. M., Griffin R. G., Dynamic nuclear polarization with biradicals. J. Am. Chem. Soc. 126, 10844–10845 (2004). [DOI] [PubMed] [Google Scholar]
- 31.Sauvée C., et al. , Highly efficient, water-soluble polarizing agents for dynamic nuclear polarization at high frequency. Angew. Chem. Int. Ed. Engl. 52, 10858–10861 (2013). [DOI] [PubMed] [Google Scholar]
- 32.Yau W. M., Thurber K. R., Tycko R., Synthesis and evaluation of nitroxide-based oligoradicals for low-temperature dynamic nuclear polarization in solid state NMR. J. Magn. Reson. 244, 98–106 (2014). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 33.Yau W. M., Jeon J., Tycko R., Succinyl-DOTOPA: An effective triradical dopant for low-temperature dynamic nuclear polarization with high solubility in aqueous solvent mixtures at neutral pH. J. Magn. Reson. 311, 106672 (2020). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 34.Chen H. Y., Tycko R., Temperature-dependent nuclear spin relaxation due to paramagnetic dopants below 30 K: Relevance to DNP-enhanced magnetic resonance imaging. J. Phys. Chem. B 122, 11731–11742 (2018). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 35.Chen H. Y., Tycko R., Slice selection in low-temperature, DNP-enhanced magnetic resonance imaging by Lee-Goldburg spin-locking and phase modulation. J. Magn. Reson. 313, 106715 (2020). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 36.Sparrow C. M., On spectroscopic resolving power. Astrophys. J. 44, 76–86 (1916). [Google Scholar]
- 37.Metropolis N., Ulam S., The Monte Carlo method. J. Am. Stat. Assoc. 44, 335–341 (1949). [DOI] [PubMed] [Google Scholar]
- 38.Abragam A., Principles of Nuclear Magnetism (Oxford University Press, New York, 1961), pp. 71–84. [Google Scholar]
- 39.Peck T. L., Magin R. L., Lauterbur P. C., Design and analysis of microcoils for NMR microscopy. J. Magn. Reson. B. 108, 114–124 (1995). [DOI] [PubMed] [Google Scholar]
- 40.Seeber D. A., Cooper R. L., Ciobanu L., Pennington C. H., Design and testing of high sensitivity microreceiver coil apparatus for nuclear magnetic resonance and imaging. Rev. Sci. Instrum. 72, 2171–2179 (2001). [Google Scholar]
- 41.Gomes E., Shorter J., The molecular language of membraneless organelles. J. Biol. Chem. 294, 7115–7127 (2019). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 42.Bosco D., et al. , Unique arrangement of α- and β-cells in human islets of Langerhans. Diabetes 59, 1202–1210 (2010). [DOI] [PMC free article] [PubMed] [Google Scholar]
Associated Data
This section collects any data citations, data availability statements, or supplementary materials included in this article.
Supplementary Materials
Data Availability Statement
Raw and processed data for Figs. 3 and 4, as well as Fortran programs for data processing and for simulations in Figs. 5 and 6, are available at https://doi.org/10.17632/vk344gdnwp.1.