A 2‐year prospective clinical and bone density evaluation, with a subset undergoing radiostereometric analysis, using the Absolut cemented stem

Jay R Ebert 1,2,3,, Nils O Nivbrant 4, Victoria Petrov 2, Piers Yates 3,5, David J Wood 4,6
PMCID: PMC9303330  PMID: 35106881

Abstract

Background

Total hip arthroplasty (THA) is common though the investigation of new prostheses requires a practical, step‐wise introduction. This study reports the 2‐year clinical results and periprosthetic bone mineral density (BMD) changes, along with a subset undergoing Radiostereometric analysis (RSA), in patients undergoing primary cemented THA using a new highly polished, double tapered, collarless femoral stem (Absolut).

Methods

Between August 2013 and December 2016, 47 patients with a mean age of 74.2 years (range 36–89) underwent 51 THAs with the Absolut. All patients underwent clinical assessment pre‐surgery and at 6 weeks, 3, 12 and 24 months using the Oxford and Harris Hip Scores, as well as Dual Energy X‐ray Absorptiometry (DEXA) to assess BMD within 2–4 weeks post‐surgery, 12 and 24 months. RSA was undertaken in a patient subset (the first n = 30) early post‐surgery (1–2 days) and 3, 12 and 24 months.

Results

All clinical scores significantly improved (p < 0.05). RSA revealed a mean subsidence of 0.78 mm at 3 months, 1.23 mm at 12 months and 1.51 mm at 24 months. Anterior–posterior and medial‐lateral translation was negligible. A significant increase (p = 0.020) in BMD was observed in Gruen zone 1, though no significant changes were observed for any other zone up until 2 years. Two patients acquired an early post‐operative deep vein thrombosis that were treated accordingly and resolved, with no further complications or re‐operations.

Conclusion

The Absolut cemented femoral stem demonstrated good outcomes, BMD changes consistent with sound prosthesis integration and patterns of post‐operative micromotion observed in other successful cemented stems.

Keywords: absolut, cemented, clinical outcomes, dual energy X‐ray absorptiometry, radiostereometric analysis, total hip arthroplasty


Total hip arthroplasty (THA) is common though the investigation of new prostheses requires a practical, step‐wise introduction. In the current study, the Absolut cemented prosthesis, a highly polished, double tapered, collarless femoral stem (Absolut), demonstrated good outcomes, periprosthetic bone mineral density changes consistent with sound prosthesis integration and patterns of post‐operative micromotion observed in other successful cemented stems.

graphic file with name ANS-92-830-g002.jpg

Introduction

Total hip arthroplasty (THA) is one of the most common and effective operations for the treatment of arthritic hip disease with high levels of patient satisfaction. 1 Polished, tapered, cemented femoral implants stems have demonstrated success in forming a stable mechanical interaction between the cement interface and bone, while permitting limited stem subsidence within the cement mantle, with this subsidence creating compressive forces in the adjacent cement that are transferred to bone as hoop stresses according to the ‘taper slip’ principle. 2 The Exeter stem (Stryker, Mahwah, NJ), a collarless polished double tapered femoral prosthesis made of stainless steel, remains one of the most commonly used cemented femoral components with impressive and well‐reported survival rates, 3 , 4 , 5 , 6 as well as predictable subsidence patterns over time at the stem‐cement interface. 7 Stefánsdóttir et al. 8 reported a median distal migration of the Exeter of 1.34 mm and 1.77 mm at 2 and 5 years post‐surgery, respectively, evaluated via radiostereometric analysis (RSA). RSA is a highly accurate radiographic technique for assessing the stability and migration of the of the prosthetic implant, 9 , 10 , 11 and it has been shown that stem migration in satisfactory implants is rapid initially and then slows, 12 with those destined to fail early demonstrating rapid migration following this initial period. 13

The Absolut stem (Global Orthopaedic Technology Pty Ltd., Sydney, Australia) is a highly polished, double tapered collarless cemented prosthesis with design parameters similar to that of the Exeter Stem. It was designed after analysis of the clinical and radiological data from 10 and 15 year series of Exeter and CPT (collarless polished tapered, Zimmer, Warsaw, Indiana) stems. In comparison to the successful Exeter, the Absolut stem has a smaller shoulder and is made of a cobalt chromium (CoCr) alloy rather than stainless steel, allowing it to accommodate larger offsets and theorized to enable the surgeon to better reproduce joint geometry and biomechanics. It had also been designed to optimize cement mantles and minimize the risk of mantle defects, which may be associated with failure. 14 , 15 , 16 Unlike other stems, the Absolut stem has a true offset with no anthropomorphic variation, whereby the offset remains consistent and does not vary with increasing stem size. Hence, downsizing stems does not change offset, while allowing thicker and more complete cement mantles to be achieved easily. Compared with an offset range for the Exeter (range 31–50) and CPT (range 32–50) stems, the Absolut includes offsets of 37, 45 and 50, though without variation with stem size.

The aim of the current study was to evaluate the clinical outcome and periprosthetic bone change up until 2 years in a prospective series of patients undergoing primary THA for osteoarthritis with the Absolut cemented stem, together with an investigation of stem migration in a subset of the cohort. It was hypothesized that: (1) a significant improvement in clinical outcomes would be observed, (2) stem migration (and in particular stem subsidence) over the first 2 years would be similar to that reported for other successful cemented femoral stems and (3) no significant changes would be observed in periprosthetic bone over the evaluation period.

Methods

Between August 2013 and December 2016, 47 patients (51 hips) underwent primary total hip arthroplasty (THA) with the Absolut cemented femoral stem, undertaken by a single high‐volume arthroplasty surgeon (DW) at a single private hospital. For the prospective clinical study follow‐up, exclusion criteria included age ≥90 years, a body mass index (BMI) >40, inflammatory arthritis, an active local or systemic infection, mental compromise (i.e., currently being treated for a psychiatric disorder, senile dementia, Alzheimer's disease, presence of alcohol and/or substance abuse), or an inability to provide informed consent and/or living outside of the metropolitan area. No patient that was a candidate for (and underwent) primary THA with the Absolut stem was excluded from study participation. Patient demographics are shown in Table 1, while Figure 1 demonstrates the flow of patient recruitment and follow‐up over the 2‐year post‐operative assessment timeline. Ethics approval was provided by the relevant hospital human research ethics committee (HREC), with the study undertaken according to the Declaration of Helsinki and the consent of all participants obtained prior to pre‐operative clinical review and subsequent THA surgery.

Table 1.

Demographics of the patient sample that underwent surgery (n = 47, 51 hips)

Variable Mean (Range)
Age (years) 74.2 (36–89)
Males (n,%) 15, 29.4%
Height (m) 1.66 (1.48–1.86)
Weight (kg) 71.4 (46.5–109.6)
Body Mass Index 25.9 (18.9–36.5)

Fig. 1.

Fig. 1

Flowchart demonstrating patient recruitment and follow‐up throughout the 2‐year post‐operative timeline.

Operative technique and rehabilitation

All surgical procedures were performed using the minimally invasive, posterior approach with ‘tension dependent’ trans‐osseous repair of the short external rotator tendons. All patients underwent THA with the Absolut femoral stem (Global Orthopaedic Technology Pty Ltd., Sydney, Australia) and 32 mm diameter ceramic head and a cross‐linked polyethylene liner in either Trilogy or Continuum acetabular components (Zimmer, Warsaw, Indiana). Following surgery, all patients participated in a standardized post‐operative out‐patient rehabilitation program for a minimum of 6 weeks with a focus on early restoration of hip movement (within standard restrictions) and gait, followed by progressive strength and functional exercises with a focus on the restoration of functional capacity.

Clinical outcome evaluation

Clinical assessment was undertaken pre‐surgery and at 6 weeks post‐surgery, as well as 3, 12 and 24 months. Two patient‐reported outcome measures (PROMs) were employed including the Harris Hip Score (HHS) 17 and the Oxford Hip Score (OHS). 18 The HHS is scored out of 100 (a higher score being desirable), including subscales of pain (0–44), function (0–47), deformity (0–4) and motions (0–5). The OHS is scored from 0 to 48 (a higher score being desirable), with this PROM primarily evaluating perceived hip pain and function. Objectively, patients had their maximal hip range of motion (ROM) assessed pre‐ and post‐surgery (on the operated limb only) using either a hand‐held bubble inclinometer (hip flexion in supine, internal and external rotation in prone) or a long arm goniometer (hip adduction and abduction in supine, extension in standing). The same qualified therapist with more than 20 years of clinical experience undertook all ROM measures, though no intra‐observer reliability assessment was undertaken.

Dual energy X‐ray absorptiometry (DEXA) evaluation

DEXA was employed within the first 2–4 weeks post‐surgery (baseline) as well as 12 and 24 months post‐surgery, to evaluate bone changes during incorporation of the prosthesis using a Lunar Prodigy Vision DEXA machine (GE Medical Systems, Madison, WI). The periprosthetic bone mineral density (BMD, g/cm2) in the coronal plane was evaluated according to the seven Gruen zones. 19

Radiographic and radiostereometric analysis (RSA)

All patients underwent anteroposterior (AP) radiographs and the adequacy of cementing around the stem at 2‐years post‐surgery was determined as per the system reported by Barrack et al. 20 The minimum mantle thickness was measured within each of the seven Gruen zones surrounding the stem. Bone resorption from the proximal medial femur was measured according to Engh et al. 21

A subset of the cohort (the initial 30 patients) underwent RSA imaging within 1–3 days after surgery (30 hips) and at 3 (29 hips), 12 (28 hips) and 24 months (25 hips), using a standard protocol with an RSA calibration cage. At the time of surgery, 1 mm tantalum markers were inserted into the femur and cement, permitting evaluation of stem movement relative to the cement mantle. Translation of the femoral implant head centroid relative to the femur was calculated and reported in mm at 3, 12 and 24 months, in relation to the initial post‐operative (1–3 days) RSA films, using UmRSA 6.0 software (RSA Biomedical, Umeå, Sweden). Motion in all 3 planes was assessed: X medio‐lateral (medial positive), Y proximal‐distal (proximal positive) and Z antero‐posterior (anterior positive).

Data and statistical analysis

Means (SD) were calculated for all measures throughout the pre‐ and post‐operative timeline, including clinical scores, DEXA values (as per the seven Gruen zones) and RSA movement (as per the three evaluated planes of motion). Medians (and range) were also calculated and presented for RSA measures. Repeated measures Analysis of Variance (ANOVA) was employed to evaluate the change in clinical scores, BMD and RSA‐assessed translation over time. T‐tests were employed to evaluate the magnitude of stem subsidence based on post‐operative Barrack grade (Barrack grade A versus other), while Spearman's correlation coefficient was employed to evaluate the association between stem subsidence and mean cement mantle thickness across the seven Gruen zones. Statistical analysis was performed using SPSS software (SPSS, Version 23.0, SPSS Inc., USA), while statistical significance was determined at p < 0.05. The sample size was initially determined based on prior studies employing RSA to evaluate stem movement. RSA studies have generally reported outcomes in small cohorts of 22–26 patients. 7 , 8 , 22 Therefore, we sought to undertake an in‐depth RSA evaluation in the first 30 patients enrolled, with further recruitment (that excluded RSA evaluation) to n = 51.

Results

As per Figure 1, 47 patients (51 hips) were recruited and underwent primary cemented THA with the Absolut prosthesis. Of these, post‐operative clinical review was undertaken in 46 patients (50 hips) at 6 weeks, 43 patients (47 hips) at 3 months, 40 patients (44 hips) at 12 months and 39 patients (43 hips) at 24 months. Specifically for the initial patient subset that underwent early post‐operative RSA investigation (30 patients, 30 hips), 29, 28 and 25 patients underwent subsequent RSA imaging at 3, 12 and 24 months, respectively (Fig. 1).

Clinical results

A significant improvement (p < 0.05) was observed for all clinical scores (PROMs and hip range of motion measures) over the pre‐ and post‐operative timeline (Table 2).

Table 2.

Clinical outcomes over the pre‐ and post‐operative timeline including the Oxford Hip Score (OHS), Harris Hip Score (HHS) and hip range of motion measures (degrees). Shown are means (SD)

Time‐point OHS (0–48) HHS (0–100) Flexion Extension Abduction Adduction External rotation Internal rotation
Pre‐surgery 24.9 (7.0) 54.3 (16.1) 93.0 (23.1) 13.4 (3.3) 26.1 (8.7) 11.6 (6.3) 23.0 (8.6) 20.4 (8.3)
6 weeks 36.9 (5.7) 80.6 (9.1) 92.5 (14.4) 15.5 (3.5) 30.2 (8.9) 14.0 (5.1) 27.4 (8.7) 27.9 (11.9)
3 months 43.5 (3.7) 91.9 (6.8) 108.0 (11.0) 17.6 (2.9) 33.9 (7.2) 17.7 (4.3) 27.7 (9.1) 34.2 (11.3)
12 months 45.9 (2.3) 94.0 (6.6) 112.0 (12.5) 18.0 (3.5) 38.5 (7.8) 19.8 (3.9) 30.4 (7.8) 38.0 (11.7)
24 months 46.2 (2.2) 95.2 (5.2) 112.4 (12.8) 18.9 (2.6) 39.0 (5.5) 21.1 (3.8) 29.7 (8.5) 35.5 (10.5)
p value <0.0001 <0.0001 <0.0001 <0.0001 <0.0001 <0.0001 <0.0001 <0.0001

Radiographic and RSA results

A total of 39 hips (76.5%) demonstrated complete cement mantles (Barrack grade A), with 11 hips (21.5%) demonstrating any full thickness mantle defect (Barrack grade C). One hip demonstrated a mantle that did not cover the tip of the stem (Barrack grade D). At 2‐year radiographic review, 50 hips (98%) had either no or first degree resorption of the proximal medial femoral cortex, with one hip demonstrating second degree. 21 No significant difference (p = 0.510) was observed in 2‐year stem subsidence between those with Barrack grade A (1.58 mm) and C/D (1.45 mm), while no association was observed between stem subsidence and mean cement mantle thickness (r = 0.06, p = 0.341).

RSA analysis of the Absolut stem revealed a mean subsidence of 0.78 mm at 3 months, 1.23 mm at 12 months and 1.51 mm at 24 months (Table 3), which was significant over time (p = 0.003). When excluding a single outlier which demonstrated subsidence of 3.28, 4.20 and 4.58 mm at 3, 12 and 24 months (Fig. 2), a mean subsidence across the full cohort of 0.69 mm, 1.12 mm and 1.38 mm was observed across the three time‐points. A non‐significant change (p = 0.956) in mean posterior translation of the prosthesis head of was observed with 0.58, 0.51 and 0.66 mm at 3, 12 and 24 months, respectively, while medial‐lateral translation was negligible (p = 0.559) (Table 3).

Table 3.

Radiostereometric analysis (RSA) scores over the post‐operative timeline, including the change from 1 to 2 days to 3, 12 and 24 months post‐surgery. Shown are means (SD), medians and range

Post‐operative time‐point Variable Medial‐lateral (mm) Proximal‐distal (mm) Posterior–anterior (mm)
Medial (+ve), Lateral (−ve) Proximal (+ve), Distal (−ve) Anterior (+ve), Posterior (−ve)
3 months (n = 29) Mean (SD) −0.07 (0.29) −0.78 (0.55) −0.58 (1.06)
Median −0.07 −0.74 −0.47
Range −0.69 to 0.83 −3.28 to −0.19 −3.71 to 2.50
12 months (n = 28) Mean (SD) −0.03 (0.45) −1.23 (0.75) −0.51 (1.45)
Median −0.06 −1.09 −0.37
Range −1.11 to 0.76 −4.20 to ‐0.18 −4.01 to 2.89
24 months (n = 25) Mean (SD) −0.19 (0.85) −1.51 (0.96) −0.66 (2.58)
Median −0.16 −1.38 −0.79
Range −2.67 to 2.19 −4.58 to −0.63 −7.63 to 8.28
p‐value 0.559 0.003 0.956

Fig. 2.

Fig. 2

Stem subsidence (mm) evaluated via RSA for each of the Absolut stems at 3 (n = 29), 12 (n = 28) and 24 (n = 25) months post‐surgery.

DEXA results

With respect to the DEXA evaluation, a significant increase (p = 0.020) in BMD was observed in Gruen zone 1 (Table 4), though no significant changes (increase or decrease) were observed for any other zone up until 2 years, in comparison to the early DEXA assessment undertaken at 2–4 weeks post‐surgery.

Table 4.

Bone mineral density (BMD, g/cm2) scores for each of the seven Gruen zones over the post‐operative timeline. Shown are means (SD)

Post‐operative time‐point G1 G2 G3 G4 G5 G6 G7
2–4 weeks 1.23 (0.25) 2.19 (0.29) 2.35 (0.29) 2.47 (0.30) 2.33 (0.24) 1.97 (0.30) 1.36 (0.37)
12 months 1.29 (0.19) 2.17 (0.22) 2.33 (0.23) 2.43 (0.30) 2.31 (0.22) 1.98 (0.27) 1.35 (0.29)
24 months 1.44 (0.19) 2.33 (0.27) 2.52 (0.29) 2.68 (0.32) 2.48 (0.32) 2.11 (0.36) 1.33 (0.37)
p‐value 0.020 0.417 0.495 0.424 0.433 0.227 0.633

Complications, adverse events, re‐operations and loss to follow‐up

Two patients in the cohort acquired a deep vein thrombosis (DVT), diagnosed at 3 and 6 weeks post‐surgery, respectively, both of which were treated accordingly and resolved without further sequalae. No further complications were observed. At 24 months post‐surgery, 17% of patients (n = 8) were lost to follow‐up (one deceased due to an unrelated illness, two patients were undergoing cancer treatment for a recent diagnosis, and five patients requested study termination due to time and travel constraints).

Discussion

This is the first study that has reported clinical outcomes in patients undergoing primary THA with the Absolut stem, a highly polished, double tapered collarless cemented prosthesis. Good clinical outcomes over 2‐years were observed, along with a low complication and/or re‐operation rate, no evidence of significant periprosthetic bone change and subsidence and migration patterns similar to other reported successful cemented stems.

While the Absolut stem has design parameters similar to that of successful predecessors such as the Exeter 3 , 4 , 5 , 6 and the CPT stem, 23 , 24 it differs to the Exeter given its smaller shoulder and CoCr material. Compared with the Exeter stem which is made of stainless steel, the CoCr alloy accommodates a larger range of offsets, therefore better reproducing joint geometry and biomechanics. Yates et al. reported an 11% dislocation rate in their stainless steel CPT stem cohort (albeit in hips surviving beyond 10 years), 23 though did suggest that once of the reasons may have been related to the limited range of offsets initially available for use. The Absolut stem had also been designed to optimize cement mantles and minimize the risk of mantle defects, which may be associated with stem failure. 14 , 15 , 16 The current study demonstrated excellent mean clinical scores (HHS and OHS) at 1 and 2 years, with PROMs similar (or better) to those reported for other cemented stems such as the Exeter and CPT across various post‐operative time‐points. 24 , 25 , 26 , 27 A significant improvement over the pre‐ and post‐operative period was demonstrated, supporting the first hypothesis.

With cemented, polished, tapered stems it has been shown that stem migration in satisfactory implants is rapid initially and then slows. 12 In the current study, a mean stem subsidence of 1.12 mm (median 1.09 mm) and 1.38 mm (median 1.38 mm) was observed in the first 1 and 2 years, respectively, after excluding a single outlier. Yates et al. reported subsidence of the CPT stem into the cement mantle of 0.44 mm/y, with 1.08 mm in the first year and 2.18 mm over 5 years, with subsidence associated with cement mantle defects. 25 Stefánsdóttir et al. reported a median migration of the Exeter of 1.23 mm at 1 year, 1.34 mm at 2 years and 1.77 mm at 5 years, 8 while Ling et al. reported long‐term subsidence of the Exeter of 1.83 mm (excluding outliers). 6 Overall, the second hypothesis for the current study was largely supported with stem subsidence over the first 2 years similar to that reported for other successful cemented femoral stems, especially in considering previously reported subsidence for the Exeter and CPT stems.

There is generally good proximal bone preservation with tapered stems, 28 with subsidence acting to distribute load more evenly throughout the entire stem‐cement and cement‐bone interface and reduce distal cement stresses. 25 , 29 DEXA has been reported as the reference standard method to evaluate BMD after THA. 30 , 31 In the current study, Gruen zone 1 demonstrated a significant increase in BMD over the 2‐year period, which could be explained by the greater offsets permitting more efficient hip abductor activity. However, no other significant zonal changes and certainly no reduction in BMD in proximal zones were observed. Damborg et al. reported a small but significant increase in BMD in Gruen zone 1 following THA employing the Exeter, though a significant reduction in zones 2, 3, 6 and 7, although no changes in zones 4 and 5. 32 Broeke et al. reported a reduction in BMD over 2 years in all Gruen zones when comparing both the cemented SHP and Omnifit prostheses. 33 In comparing the SHP and Lubinus SP2 cemented prostheses, Nivbrant et al. reported a reduction in BMD across all Gruen zones. 34 In a study that sought to investigate whether femoral prosthetic geometry affected the pattern of strain‐adaptive bone remodelling in the proximal femur after THA, Jayasuriya et al. demonstrated that BMD loss was greatest in Gruen zone 7 (proximal medial femur) though similar in regional distribution and magnitude between the Charnley (composite‐beam), Exeter (double tapered) and C‐Stem (triple‐tapered) prostheses. 35 Overall, the third hypothesis of the current study was partially supported, whereby apart form a small though significant increase in BMD in Gruen zone 1 over the 2‐year period, no other significant changes were observed in periprosthetic bone BMD.

A number of study limitations are acknowledged. First, this study included a small sample size and a cohort of patients that underwent surgery by a single experienced surgeon via a mini‐incision posterior approach. Therefore, while this approach (single surgeon) lacks generalizability across varied operators and surgical approaches, it may also be considered an important first step in investigating a new device to reduce variability that may present from these other sources. It was felt that the cohort was sound for a prospective pilot cohort of patients undergoing with a new prosthesis, and given the accuracy of RSA reported RSA studies generally report outcomes in small cohorts of 22–26 patients. 7 , 8 , 22 Secondly, it should be acknowledged that patient and radiographic outcome should be continued well beyond the reported post‐operative 2 years, though the primary aims of the current study sought to report stem movement and periprosthetic bone changes over this initial period in a currently unreported prosthesis. Thirdly, we employed PROMs commonly reported in patients undergoing THA including the OHS and HHS, 17 , 18 though acknowledge that others exist and patient satisfaction was not specifically assessed.

Conclusion

The Absolut cemented femoral stem demonstrated good clinical outcomes and no re‐operations within 2‐years, BMD changes consistent with sound prosthesis integration and patterns of post‐operative micromotion observed in other successful cemented stems. The Absolut stem (in comparison to successful predecessors such as the Exeter and CPT stems) differs due to its smaller shoulder and CoCr material, accommodating a larger range of offsets for the surgeon and the ability to better reproduce joint geometry and biomechanics. Ongoing patient and radiographic review are required to assess longer‐term migration patterns and failures beyond 2‐years.

Conflict of interest

Global Orthopaedic Technology provided a research grant to assist this research project. There are no conflicts of interest associated with this publication and there has been no financial support for this work that could have influenced its outcome. Two authors (as outlined in the Cover Page) receive royalties for part‐development of the Absolut stem.

Author contributions

Jay R. Ebert: Conceptualization; data curation; formal analysis; investigation; methodology; project administration; supervision; writing – original draft; writing – review and editing. Nils O. Nivbrant: Conceptualization; data curation; formal analysis; methodology; resources; software; writing – original draft; writing – review and editing. Victoria Petrov: Data curation; formal analysis; investigation; methodology; resources; writing – original draft; writing – review and editing. piers yates: Conceptualization; data curation; formal analysis; methodology; supervision; writing – review and editing. David J. Wood: Conceptualization; data curation; formal analysis; funding acquisition; investigation; methodology; project administration; supervision; writing – original draft; writing – review and editing.

Acknowledgement

Open access publishing facilitated by The University of Western Australia, as part of the Wiley ‐ The University of Western Australia agreement via the Council of Australian University Librarians.

J. R. Ebert PhD; N. O. Nivbrant MBBS; V. Petrov BSc; P. Yates MBBS, BSc, MRCS, FRCS, FRACS, FAOrthA; D. J. Wood BSc MBBS, MS, FRCS, FRACS.

References

  • 1. Mancuso CA, Salvati EA, Johanson NA, Peterson MGE, Charlson ME. Patients' expectations and satisfaction with total hip arthroplasty. J. Arthroplasty 1997; 12: 387–96. [DOI] [PubMed] [Google Scholar]
  • 2. Shen G. Femoral stem fixation. An engineering interpretation of the long‐term outcome of Charnley and Exeter stems. J. Bone Joint Surg. Br. 1998; 80: 754–6. [DOI] [PubMed] [Google Scholar]
  • 3. Hook S, Moulder E, Yates PJ et al. The Exeter universal stem: a minimum ten‐year review from an independent Centre. J. Bone Joint Surg. Br. 2006; 88: 1584–90. [DOI] [PubMed] [Google Scholar]
  • 4. Lewthwaite SC, Squires B, Gie GA, Timperley AJ, Ling RSM. The Exeter Universal hip in patients 50 years or younger at 10‐17 years' followup. Clin. Orthop. Relat. Res. 2008; 466: 324–31. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 5. Carrington NC, Sierra RJ, Gie GA et al. The Exeter Universal cemented femoral component at 15 to 17 years: an update on the first 325 hips. J. Bone Joint Surg. Br. 2009; 91: 730–7. [DOI] [PubMed] [Google Scholar]
  • 6. Ling RS, Charity J, Lee AJ et al. The long‐term results of the original Exeter polished cemented femoral component: a follow‐up report. J. Arthroplasty 2009; 24: 511–7. [DOI] [PubMed] [Google Scholar]
  • 7. Alfaro‐Adrian J, Gill HS, Murray DW. Cement migration after THR. A comparison of charnley elite and Exeter femoral stems using RSA. J. Bone Joint Surg. Br. 1999; 81: 130–4. [DOI] [PubMed] [Google Scholar]
  • 8. Stefansdottir A, Franzen H, Johnsson R et al. Movement pattern of the Exeter femoral stem; a radiostereometric analysis of 22 primary hip arthroplasties followed for 5 years. Acta Orthop. Scand. 2004; 75: 408–14. [DOI] [PubMed] [Google Scholar]
  • 9. Karrholm J, Herberts P, Hultmark P et al. Radiostereometry of hip prostheses. Review of methodology and clinical results. Clin. Orthop. Relat. Res. 1997; 344: 94–110. [PubMed] [Google Scholar]
  • 10. Selvik G. Roentgen stereophotogrammetry. A method for the study of the kinematics of the skeletal system. Acta Orthop. Scand. Suppl. 1989; 232: 1–51. [PubMed] [Google Scholar]
  • 11. Ryd L. Roentgen stereophotogrammetric analysis of prosthetic fixation in the hip and knee joint. Clin. Orthop. Relat. Res. 1992; 276: 56–65. [PubMed] [Google Scholar]
  • 12. Kiss J, Murray DW, Turner‐Smith AR et al. Migration of cemented femoral components after THR. Roentgen stereophotogrammetric analysis. J. Bone Joint Surg. Br. 1996; 78: 796–801. [PubMed] [Google Scholar]
  • 13. Karrholm J, Borssen B, Lowenhielm G et al. Does early micromotion of femoral stem prostheses matter? 4‐7‐year stereoradiographic follow‐up of 84 cemented prostheses. J. Bone Joint Surg. Br. 1994; 76: 912–7. [PubMed] [Google Scholar]
  • 14. Star MJ, Colwell CW Jr, Kelman GJ, Ballock RT, Walker RH. Suboptimal (thin) distal cement mantle thickness as a contributory factor in total hip arthroplasty femoral component failure. A retrospective radiographic analysis favoring distal stem centralization. J. Arthroplasty 1994; 9: 143–9. [DOI] [PubMed] [Google Scholar]
  • 15. Massoud SN, Hunter JB, Holdsworth BJ, Wallace WA, Juliusson R. Early femoral loosening in one design of cemented hip replacement. J. Bone Joint Surg. Br. 1997; 79: 603–8. [DOI] [PubMed] [Google Scholar]
  • 16. Ebramzadeh E, Sarmiento A, McKellop HA et al. The cement mantle in total hip arthroplasty. Analysis of long‐term radiographic results. J. Bone Joint Surg. Am. 1994; 76: 77–87. [DOI] [PubMed] [Google Scholar]
  • 17. Harris WH. Traumatic arthritis of the hip after dislocation and acetabular fractures: treatment by mold arthroplasty. An end‐result study using a new method of result evaluation. J. Bone Joint Surg. Am. 1969; 51: 737–55. [PubMed] [Google Scholar]
  • 18. Murray DW, Fitzpatrick R, Rogers K et al. The use of the Oxford hip and knee scores. J. Bone Joint Surg. Br. 2007; 89: 1010–4. [DOI] [PubMed] [Google Scholar]
  • 19. Gruen TA, McNeice GM, Amstutz HC. “Modes of failure” of cemented stem‐type femoral components: a radiographic analysis of loosening. Clin. Orthop. Relat. Res. 1979; 141: 17–27. [PubMed] [Google Scholar]
  • 20. Barrack RL, Mulroy RD Jr, Harris WH. Improved cementing techniques and femoral component loosening in young patients with hip arthroplasty. A 12‐year radiographic review. J. Bone Joint Surg. Br. 1992; 74: 385–9. [DOI] [PubMed] [Google Scholar]
  • 21. Engh CA, Bobyn JD, Glassman AH. Porous‐coated hip replacement. The factors governing bone ingrowth, stress shielding, and clinical results. J. Bone Joint Surg. Br. 1987; 69: 45–55. [DOI] [PubMed] [Google Scholar]
  • 22. Onsten I, Akesson K, Besjakov J et al. Migration of the Charnley stem in rheumatoid arthritis and osteoarthritis. A roentgen stereophotogrammetric study. J. Bone Joint Surg. Br. 1995; 77: 18–22. [PubMed] [Google Scholar]
  • 23. Yates PJ, Burston BJ, Whitley E et al. Collarless polished tapered stem: clinical and radiological results at a minimum of ten years' follow‐up. J. Bone Joint Surg. Br. 2008; 90: 16–22. [DOI] [PubMed] [Google Scholar]
  • 24. Burston BJ, Barnett AJ, Amirfeyz R et al. Clinical and radiological results of the collarless polished tapered stem at 15 years follow‐up. J. Bone Joint Surg. Br. 2012; 94: 889–94. [DOI] [PubMed] [Google Scholar]
  • 25. Yates P, Gobel D, Bannister G. Collarless polished tapered stem: clinical and radiological follow‐up over 5 years. J. Arthroplasty 2002; 17: 189–95. [DOI] [PubMed] [Google Scholar]
  • 26. Keeling P, Howell JR, Kassam AM et al. Long‐term survival of the cemented Exeter universal stem in patients 50 years and younger: an update on 130 hips. J. Arthroplasty 2020; 35: 1042–7. [DOI] [PubMed] [Google Scholar]
  • 27. Burston BJ, Yates PJ, Hook S, Moulder E, Whitley E, Bannister GC. Cemented polished tapered stems in patients less than 50 years of age: a minimum 10‐year follow‐up. J. Arthroplasty 2010; 25: 692–9. [DOI] [PubMed] [Google Scholar]
  • 28. Mallory TH, Head WC, Lombardi AV Jr. Tapered design for the cementless total hip arthroplasty femoral component. Clin. Orthop. Relat. Res. 1997; 344: 172–8. [DOI] [PubMed] [Google Scholar]
  • 29. Huiskes R, Verdonschot N, Nivbrant B. Migration, stem shape, and surface finish in cemented total hip arthroplasty. Clin. Orthop. Relat. Res. 1998; 355: 103–12. [DOI] [PubMed] [Google Scholar]
  • 30. Trevisan C, Bigoni M, Cherubini R, Steiger P, Randelli G, Ortolani S. Dual X‐ray absorptiometry for the evaluation of bone density from the proximal femur after total hip arthroplasty: analysis protocols and reproducibility. Calcif. Tissue Int. 1993; 53: 158–61. [DOI] [PubMed] [Google Scholar]
  • 31. Kroger H, Miettinen H, Arnala I et al. Evaluation of periprosthetic bone using dual‐energy x‐ray absorptiometry: precision of the method and effect of operation on bone mineral density. J. Bone Miner. Res. 1996; 11: 1526–30. [DOI] [PubMed] [Google Scholar]
  • 32. Damborg F, Nissen N, Jorgensen HR et al. Changes in bone mineral density (BMD) around the cemented Exeter stem: a prospective study in 18 women with 5 years follow‐up. Acta Orthop. 2008; 79: 494–8. [DOI] [PubMed] [Google Scholar]
  • 33. Broeke RH, Harings SE, Emans PJ et al. Randomized comparison between the cemented scientific hip prosthesis and Omnifit: 2‐year DEXA and minimum 10‐year clinical follow‐up. J. Arthroplasty 2013; 28: 1354–61. [DOI] [PubMed] [Google Scholar]
  • 34. Nivbrant B, Karrholm J, Soderlund P. Increased migration of the SHP prosthesis: radiostereometric comparison with the Lubinus SP2 design in 40 cases. Acta Orthop. Scand. 1999; 70: 569–77. [DOI] [PubMed] [Google Scholar]
  • 35. Jayasuriya RL, Buckley SC, Hamer AJ et al. Effect of sliding‐taper compared with composite‐beam cemented femoral prosthesis loading regime on proximal femoral bone remodeling: a randomized clinical trial. J. Bone Joint Surg. Am. 2013; 95: 19–27. [DOI] [PubMed] [Google Scholar]

Articles from Anz Journal of Surgery are provided here courtesy of Wiley

RESOURCES